1. 2014 11 01 3 23 IPO2L08 MEDICATLANTIC The bed is a medical device and must not be modified under any circumstances You must ensure Its traceability including that of the boards and its accessories If you assemble different types of medical devices you must conduct a risk analysis and make the CE declaration The electric parts Jack supply box wired control etc shall only be repaired by the manufacturer Linak The bed is not suitable for use with an inflammable anaesthetic mixture with air or oxygen or nitrous oxide The loads permitted see bed characteristics must be distributed evenly over the bed base Do not activate all the motors at the same time when the patient is in the bed only one motor is authorised at one time except elevation by 2 motors or simultaneous function After each use and while care is being administered to the patient the brakes must be activated We recommend putting the bed in its low position after every use and while the patient is resting to reduce the height of falls by a confused or agitated person Remember to lock the function s if the option is available On change of height or angle of the parts of the bed make sure that there are no objects and no parts of the patient s or carer s body caught between the bed the boards the accessories and the ground or between the boards and base or between the cross braces Do not sit down on the side of the back rest or leg rest if this is not fl
2. and take hold of the back rest handle with one hand Push or lift to compensate the pressure exerted by the patient and unhook the clip by the connecting rod with the other hand The back rest jack will then pivot downwards d Put the headboard back Version with handle on the back rest Version without handle on the back rest gt If the bed is fitted with an emergency release for the back rest Cardio Pulmonary Resuscitation 1 Grasp the back rest with one hand 2 With the other hand activate one of the two handles on the back rest while lowering If the handle is released the back rest will stop moving Release handles 2014 11 01 13 23 IPO2L08 MEDICATLANTIC 4 MAINTENANCE 4 1 Instructions for dismantling the motors AN Disconnect the 230 volts connection before dismantling Dismantle when the bed is empty or in the side position If dismantling in any other position keep a firm hold of the moving parts to avoid any shearing Unblock the safety clips O unplug the motor leads and remove them from the securing seals Put the motors back in place and put in the same direction as at the beginning O Safety clip 4 2 Maintenance Quality control of medical beds will be made by technical staff or trained biomedical and taking into account the normal conditions of use specified in the user guide on a bed with its specific security barriers The bed must be available to perform all quality control at le
3. contrary to the safety instructions made by the user since the bed s installation Remind the users of the safety instructions All installation and preventive maintenance operations must be recorded See table model below This record must be kept in a designated area throughout the bed s lifetime 2014 11 01 14 23 IPO2L08 MEDICATLANTIC QUALITY INSPECTION OF MEDICAL BEDS IDENTIFICATION OF MEDICAL BED TYPE MODEL TRADEMARK SERIAL NO DATE OF MANUFACTURE Type model Identification serial no Mass continuity tester 1 VISUAL CHECKS General condition User s manual available Headboard and footboard present Good overall condition head and footboards bed corners protective Stops General cleanliness Acceptable state of corrosion given the requirements of the user department Identification label serigraphy in good condition Mechanical condition Lifting pole in good condition positioning and strap 1 Mechanical leads in good condition Sleeping surface in good condition bed base 1 Boards lock and tighten well head and footboards 1 Chest rest functions properly Leg rest functions proper Manual leg rest functions properly TR RTR positions function properl Bed base extension functions properly me Castors function properly pivoting rolling etc including the steering castor where applicable Bed immobilises properly castor brakes etc me Verification of ti
4. for the supply box voltage Also ensure that the power lead and the remote control lead are positioned correctly to prevent any risks of getting caught between the moving parts of the bed V YYY Check that the bed operates properly after installing it in accordance with the check list appended in this document Test all of its functions Users must be trained in how to use the equipment Inform the patient and his visitors of the safety instructions to be observed 2014 11 01 5 23 IPO2L08 MEDICATLANTIC 3 3 Use 3 3 1 Easy Move boards gt Installing an Easy Move bed board gt Removing an Easy Move bed board fm 4 f I to 2 cm 2014_11_01 6 23 IPO2L08 MEDICATLANTIC 3 3 2 Metal side rails gt To positionning the side rail Head side Feet side 1 insert amount of the side rail in the grip jaw on the head side of the bed 2 Tighten the Rondo screws of the grip jaws on the bed base To remove the side rail in reverse operatior 2 and operatio 1 by pulling the release pin on the grip jaw of the head side If the side rail is poorly positioned safety of the patient may be endangered or a malfunctioning may occur The side rails must not be used when the patient is a child under 12 or if s he is too small lt 146 cm gt To lower the side rail Feet side Head side gt Toraise the side rail Check that the side rail is locked by trying to fold it w
5. transmitters and the bed as recommended below according to the maximum output power of the communications equipment Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz d 1174P d 117VP d 2 33VP Rated maximum power of transmitter W 0 01 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer 0 1 Note 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies Note 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people 3 4 2 Equipotentiality Under the head half of the bed base Sn rm you will find an equipotentiality socket Label identified by the label enabling you to connect any electromedical devices The leads of these devices must pass through the head end and not the sides Equipotentiality T gt a 2014 11 01 12 23 IPOZLO8 MEDICATLANTIC 3 5 Putting the back rest flat In the event of a power cut or failure flatten the back rest as follows Disconnect the power supply b Dismantle the headboard Stand at the head of the bed
6. 4 Electrical connection diagram si 21 6 5 USE OT SPECIFIC GlEMENS Leke iiesddiiendee 22 8 5 1 Centralised Drake ccccccccccccccccccccssseeeccccccccaussseeeccccccccauuseseececcccauausseeececccecaaauateecececeeeaaaeeees 22 8 5 2 Wooden side rails iiiseeecceeeeeeeeeseeeceeeeeeeeeceeeeecceeeeeeenes 22 8 5 3 Protective covers iii eeceeeeneeeecceceeeeeeeceeeeeeeeecceeeeeeeeeeeeeeeneeeee 22 6 A0 Me OCCIIIC technical dal ecc 23 SOL NO 23 002 NN PR E E A E E E 23 8 6 3 Dimensional si iiiseeeceeeeneseecceeeeeceeeceeeeeeeeecceeeeeeeeeeeeeeeneeeee 23 2014 11 01 2 23 IPO2L08 MEDICATLANTIC Dear Sir Madam You have acquired a MEDICATLANTIC medical bed equipped with its accessories and we thank you for your custom Our beds and their accessories are designed and manufactured in compliance with the essential requirements of the European Directive 93 42 EEC and 2007 47 EEC They are tested in conformity with standard EN 60601 2 52 2010 in their commercial configurations including the boards and accessories that we manufacture so as to ensure you maximum safety and performance As a result maintenance of the contracted good s warranty depends on compliance with the conditions for use recommended by MEDICATLANTIC and the use of original accessories which also guarantees you safe use of the medical bed and its accessories 1 TRAN
7. FUNCTION TURN THE KEY SLIGHTLY TO THE IS UNLOCKED RIGHT TO LOCK THE FUNCTION Key to lock a function 2014 11 01 9 23 IPO2L08 MEDICATLANTIC 3 3 7 Leg rest gt Manual crank version 11 To lift lift the leg rest using the wire handle at the end To lower relieve the leg rest slightly or to its maximum with the hand to release the catch then lower the leg rest gt Version with electric folding 08 Memory folding This function keeps a position of the tibia section horizontal when the jack is activated upwards To use this function the 1 crank catch must be engaged when the leg rest is flat Folding without memory the end of the tibia section stays in contact with the bed base 3 4 Technical characteristics 3 4 1 Electrical data Bed code Class II double insulation NS Ol XXXXXXX TT 08 liquid penetration D IP XX qe Current type Protection level against 230V FACTEUR DE SERVICE EE EN XX Yo max gt xmin gt xmin q Operating time PUISSANCE ABSORBEIE xxx VA Absorbed power PROTECTION TYPE INDEX VOLTAGE FREQUENCY LINAK jack LA27 LA24 LA34 IP66 24VDC Supply box c6 cBis imee ovac som console Wired control He72 HBe74 pes vo Lockable wired control HL72 HL74 IP 54 24V DC me Infrared control HB21 IP 21 3V DC mm A Maximum operating time Read the recommendations on the electrical label on the bed Protection level against electric shocks typ
8. SPORT AND STORAGE For transport the bed should be in its low position on a pallet and strapped and protected The wired control and supply lead should be attached to the bed base The head and footboards are protected and strapped to the sleeping surface The bed should be transported upright when in its original packaging in compliance with the instructions printed on the packaging It is strictly forbidden to stack packages weighing over 60kg m whatever position they are in Before transporting or dismantling the bed make sure the back and leg rests are fixed to the frame of the bed base 2 BED ENVIRONMENT CONDITIONS The bed along with the boards and accessories must be transported stored and used at a room temperature of between 100 and 50 and relative humidity of between 30 and 75 I Atmospheric pressure between 700hPa and 1060hPa 3 GENERAL USE Observe the specified environmental conditions 3 1 Precautions for use Before use it is essential to read these instructions carefully They contain advice on using and looking after the bed to guarantee optimum safety A The user and staff must be trained and aware of the risks associated with using the bed and children confused or disorientated persons must not be allowed on it Although the bed is conforming with Electromagnetic Compatibility some devices may alter how it functions in which case they must be used at a distance or not used at all
9. ast once a year but also on special request and corrective maintenance on the performance that could be affected by the intervention However to save time this may be associated with preventive maintenance In this case it is not useful to make a further examination of already controlled performance RECOMMENDATIONS FOR PREVENTIVE MAINTENANCE Preventive maintenance should be carried out in accordance with our specifications and at least once a year by the organisation or person who installed the bed Between two maintenance sessions and at least once a year the following should be carried out Verification that the electrical leads are connected all along the metal jambs to prevent shearing of these leads when the variable height is being activated Verification that all of the electrical leads and plugs are in good condition Replacement if there is the slightest alteration wear shearing damage etc Verification of the external appearance traces of damp and good overall condition of protective covers in particular and that the motors and jacks function properly Verification that the bed is in good working order test all functions Verification that the frame bed base and mechanical joints are all in good condition When maintenance is carried out at the patient s home as part of a long term contract the installer must also Check that the bed is properly installed check to see that there hasn t been any modification
10. at In the case of a prolonged more than 50 tilt bust semi sitting position it is recommended to vary the position of the person in bed every 2 hours When the bed is being moved keep the power lead well away from the ground and wheels When use of an adaptor extension lead or connection plug proves necessary you must check that its characteristics are suitable for the bed Connection to the supply box must be done using a mains complying with the standards in force and corresponding to a voltage of use of 230 V The mains plug must be disconnected before the bed is moved Do not pull on the mains leads to disconnect the mains plug During any handling try not to catch the leads of the motors and remote control and do not get them knotted The wired control must be hooked to the headboard when not in use MEDICATLANTIC prohibits the establishment of two beds in a room or in a too close environment as infrared remote command the two beds together The condition of the leads must be checked frequently If the slightest modification is observed the person in charge for maintaining the bed must be contacted to carry out the necessary repairs If repairs are required the person in charge of maintenance must be contacted Moreover the telephone number of the company to be contacted for any repairs is given in this document When using side rails the distance between the top of the rail and the uncompressed top surface of the mattress
11. ck load well Visual and sound alarms in good working order oo o Quantitative aspects Bed functions properly using the battery functions properly using the batter Check the scale of movements Maximum angle when propped Maximum angle of specifications claimed by the manufacturer 29 Maximum height Maximum height of specifications claimed by the manufacturer 20 mm 2014 11 01 16 23 IPO2L08 MEDICATLANTIC Comments Operational is the safety of the patient carers and technical staff at risk me Plan of action see comments 3 me NAME Establishment gt DATE Signature S O 1 Not Applicable 2 If the height measured does not comply with the standard the health manager responsible for correct application must be informed Failure to comply is not a criterion for a non operational status 3 The manager decides on the actions to take and which people to contact depending on the results of the quality inspection and the comments made 4 3 Cleaning and disinfection High pressure cleaning is forbidden Unplug the mains lead Check that all the electrical parts are connected together All the sockets of the supply box must be used otherwise its watertightness is not guaranteed Clean the electric covers of the jacks and wired control straightaway if any bodily fluids particularly urine have sprayed on to them The medical bed is a non critical appliance requiring Low Level disinfection W
12. e B ar Essential performances The bed will not move automatically when subject to electromagnetic disturbances within the limit of the values indicated below 2014 11 01 10 23 IPO2L08 MEDICATLANTIC Manufacturer s declaration and guide electromagnetic emissions The medical bed see references in contents has been designed for use in the electromagnetic environment specified below The user should ensure that it is used in Such an environment RF emissions Group 1 The medical bed see references in contents uses RF energy only for its internal functions Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions Class B The medical bed see references in contents can be used in all domestic environments CISPR 11 including those directly connected to the public low voltage power supply network that supplies buildings for domestic purpose Harmonic emissions Class A EN 61000 3 2 Voltage fluctuations Flicker Applicable EN 61000 3 3 RF emissions Compliant The medical bed see references in contents has not been designed for connection to other CISPR 14 1 equipment Manufacturer s declaration and guide electromagnetic immunity CISPR 11 The medical bed see references in contents has been designed for use in the electromagnetic environment specified below The user should ensure that it is used in such an en
13. e draw your attention to the fact that the recommendations below are drawn up according to the rules of good practice but are not a protocol Contact the hospital hygiene department gt AIM To recondition the bed and prevent the transmission of germs from one patient to another To eliminate all organic soiling by physical action cleaning chemical action disinfection gt INDICATION Physical and bacteriological cleanliness of the bed and its accessories 2014 11 01 17 23 JPO2L08 MEDICATLANTIC gt EQUIPMENT Microfiber wipes Detergent or Detergent Disinfectant Surface DD with CE mark and surface Disinfectant Chlorine substances alcohol base lt 30 Attention DD products and Javel water must not be used undiluted A remanence time must be applied according to the disinfectant manufacturer s instructions the drying time without human presence is often the same as the remanence time SANIVAP steam appliance with accessories gt TECHNICAL Daily maintenance with a surface DD product in one operation Maintenance when the patient leaves or periodically by the process known as Bio cleaning observing the 3 operations o Cleaning is done by means of a cloth soaked in a surface detergent or Detergent Disinfectant DD solution o Rinsing is done with cloth rinsed in clean water o Disinfection is done by means of a cloth soaked in a surface disinfectant solution Specific maintenance by specialist contractors after rem
14. eceeeceeeeeceecceeeeeeeeceeeeneeeeeeee 9 A LOO TOST a 10 3 4 Technical characteristics si iiiseeeeeeeeeesceeeeeeeeeeceeeeeeeeeeeeee 10 3 4 1 Electrical data 0 cccccccccccccecccccccccssscccccccccscecccccccssececccsccuseececesccstececcesccsceseccsscaseeseecsecaseceeeesecseeees 10 EE LOT EET EE 12 3 5 Putting the back rest flat 13 gt Ifthe bed is fitted with an emergency release for the back rest Cardio Pulmonary FUNN hr 13 4 MAINTENANCE errennnnnnnnrnnnnnnvrrnrnnnrrrnnnnnerrnrnerrrnnnnvveeereenerrrnnnnnesseeneesssnnnnneesseenesssnnnnneeseeneeessnnnnnneseee 14 4 1 Instructions for dismantling the motors 14 LETE TG 14 4 3 Cleaning and disinfection ss 17 4 A 5 LA 19 4o CHAU UTE ANIM see ce acres tect sc E sea net sm epi ence csvset sa std se mea E 19 5 CONDITIONS FOR SCRAPPING ccsssececccccccccsaassecececececeaasssseececcceeeaaaaseeeesecs 19 6 BED BOARDS 80cm WIDTH COMPATIBLE eee 20 7 COMPATIBLE ACCESSORIES rrrrnvvvrrrrnnnnnrrnnnnnnnnrnnnnnrrnnnnnnesnennnnerrnnnnneesenneessvnnnnneeseeenenssnnnnnneseee 20 8 SPECIFIC USE ooo eeeecccccccccessssscecccccccceassssesececececesaaarseececececeeaaaasseeeeececeeaaaaeteecesecs 21 Gils UPD OSE On CO ais a et eases 21 8 2 Gen ral EIN a dada E ni ee ec ee nn ne 21 8 3 Specific precautions for USe sisi 21 8
15. ghtenings diverse nuts and bolts pins pivot IV stand mere 2014 11 01 15 23 IPO2L08 MEDICATLANTIC Qualitative aspects Verification that welds are in good condition 2 that welds are in good condition Absence of sound disturbances squeaking lubrications Electrics hydraulics and pneumatics Electrical leads plugs and connectors are in good condition not sheared not caught etc Electrical parts in good condition leads motors boxes etc me Hydraulic and pneumatic parts in good condition pumps compressors jacks dampers etc Remote controls displays and lights in good condition Pf To o Bed specific side rails YES NO 1 The rails in place and specific to the bed and or comply with the CRE manufacturer s specifications Properly positioned and secured Side rail lockina functions properly in raised position nn FT Check that the height measured from the top of the barrier to the uncompressed mattress surface excluding therapeutic mattresses more than or equal to 220 mm complies with the standard in force 1 Locking of operational functions sss of operational functions Inactivation of variable height control pedals Cardio Pulmonary Check that the headboard extracts or ER Resuscitation CPR retracts properly in an emergency emergency flattening of the Check that the chest rest emergency EE back rest flattening function works properly Withstands ja
16. in a typical commercial or hospital environment 50 60 Hz NB Ur is the nominal value of power voltage applied during the test Manufacturer s declaration and guide electromagnetic immunity The medical bed see references in contents has been designed for use in the electromagnetic environment specified below The user should ensure that it is used in Such an environment IEC 60601 Compliance level Immunity test P Electromagnetic environment Guide Severity level Portable and mobile RF communications equipment should be used no closer to the medical bed see references in contents including leads than the recommended separation distance calculated using equations applicable to the frequency of the transmitter Recommended separation distance Conducted RF 3 Vrms 3 V d 117VP EN 61000 4 6 150 kHz to 80 MHz Radiated RF 3 V m 3 V m d 1 17VP 80MHzto 800 MHz EN 61000 4 3 80 MHz to 2 5 GHz 80 to 800 MHz d 2 334 P 800 MHz to 2 5 GHz eo HE where P is the maximum output power rating of the transmitter in watts W according to 2014_11_01 11 23 IPO2L08 MEDICATLANTIC 800 MHz to the transmitter manufacturer and d the recommended separation distance in meters m 2 GHz The field strengths transmitted by fixed RF transmitters determined by an electromagnetic measurement of the site must be less than the conformity level in each range of frequencies a Disturbances can occur near devices mar
17. ithout using the release pin There must be at least 220 mm between the top of the side rail and uncompressed mattress surface 2014 11 01 1123 IPOZLO8 3 3 3 Wooden barriers MEDICATLANTIC See the attached instructions for fitting the wooden side rail gt Toraise the side rail Raise the top side rail with both hands until it locks Check that it is properly slot in gt To lower the side rail Raise the top side rail with one hand Press on the unlocking button with the other hand Support the rail as it lowers Repeat steps to for the other side 3 3 4 Angled lifting pole and IV stand Check that the high guide engage in the right direction in the low guide The lifting pole is intended to help the patient lift him herself up and change position in the bed It is not meant to help with transferring gt a i w 2014 11 01 8 23 IPO2L08 MEDICATLANTIC 3 3 5 Separate braking Check that the wheels are locked by trying to move the bed If this is not done the patient or another person who leans on the bed may fall 3 3 6 Remote control Carry out a test cycle when the bed is empty to familiarise yourself with the bed functions Lifting and lowering of variable height Lifting and lowering of electric folding Simultaneous lifting and lowering of back rest and folding KEY IN VERTICAL POSITION THE
18. ked with this symbol Note 1 At 80 MHz and 800 MHz the upper frequency range applies Note 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the medical bed see references in contents is used exceeds the applicable RF compliance level above the normal operation of the bed must be checked If abnormal performance is observed additional measures may be necessary such as re orienting or relocating the medical bed Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3 V m Recommended separation distances between portable and mobile RF communications equipment and the medical bed See references in contents The medical bed See references in contents is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled The user of the bed can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment
19. low keep away from heat sources and avoid contact with eyes 2014 11 01 18 23 IPOZLO8 MEDICATLANTIC 4 4 Warranties gt All of our products carry a warranty against any manufacturing defect provided the normal conditions for use and maintenance are complied with Labour costs due to changes in structures or parts under warranty are not taken into account Please refer to the standard terms of sale for the specific terms of warranty for each product Every time you contact us for possible maintenance you must quote us the information on the bed identification label and on the electric parts if these are concerned Original parts shall be supplied for replacement within the term of warranty by our customer sales network determining the beginning of the term of warranty Defective parts must be returned to ensure proper application of this warranty and also to avoid any invoicing 4 5 Identification Product identification Sa ANG Manufacturer CE marking Serial number ape oe Refer to user s manual and date of TT Manufacturer s address manufacture oral Refer to user s manual gt A kg lt Maximum patient en m kg lt Safe operational load 5 CONDITIONS FOR SCRAPPING The product must be scrapped if the main requirements are no longer met particularly when the product no longer has its original characteristics and has not been subject to corrective action during the manufacturing pr
20. nd cross Leg lift handle pieces 8 3 Specific precautions for use The bed should only be used for stretcher work if it is fitted with centralised brakes Back lift jack Leg lift jack Height adjustment jack Remote control 8 4 Electrical connection diagram 2014 11 01 IPOZLO8 MEDICATLANTIC 8 5 Use of specific elements 8 5 1 Centralised brake A Brakes applied press the pedal red side with your foot B Brakes released press the pedal A or C with your foot to obtain the middle position Ra C Swivelling wheel press the pedal green side with your foot A Before transporting or dismantling the bed fix the back lift and the leg lift to the mattress base frame 8 5 2 Wooden side rails Wooden gates are made to be used on beds MEDICATLANTIG with a length of 1600 mm and panels fitted reservations for their assembly These must be associated with mattresses whose specifications are given in the instructions of the bed Be sure to check the tightness of rondos guarantee the position of panels and smooth sliding of the barrier The part consisting of a wood and metal bar is always fitted at the bottom of the side rails The side rails have been tested in conformity with the test method NF EN 716 1 and 2 standard 8 5 3 Protective covers Fit the protective covers by following the instructions below _ Wrong positioning of the side rails and or absence of the protective cover is detrimental t
21. o MN MEDICAI LANTIC WINZICARE 4 Le Pas du Ch teau 85670 SAINT PAUL MONT PENIT TEL 02 51 98 55 64 FAX 02 51 98 59 07 Email info medicatlantic fr Site Internet http www winncare fr PITCHOUNE KALIN IPO2L08 Avec Ecofolio papiers tous les papiers se recyclent 2014 11 01 580049 Anglais IPO2L08 MEDICATLANTIC 1 TRANSPORT AND STORAGE oorrrrnonnnnnnnnnnnnnernnnnnnrnnnnnnnnnnnnnnnrrnnnnneeeernnensrnnnnneeeeeneessrnnnnneeeeenensesnnnnn 3 2 BED ENVIRONMENT CONDITIONS eesrmnrnnrrnnnnnrrrrnnnnnvrrnnnnnvnernnnnnrrnnnnnveeeernnnerrnnnnneeeeenennsennnnnneseee 3 3 GENERAL USE oororrrrrnnnnnnnnnnnnnnnrnnnnnrrrnnnnnnnnennnnerrnnnnneeeerenerervnnnnveeeeneeeesennnnnesseenneessnnnuneessenenessnnnnnneseee 3 3 1 Precautions for use iiiueeeeeeeeeeeeeceeeeeeceeceeeeeeeeeecceeeeeeeeceeeene 3 3 2 FISTUSE igo ceteduazesoteavaneeacnwassnanecivsudeenionesntuawadevaicuwoasasimesiesaisnducmoeaitasadsisncuuaasnsnesizdeseaaetasavaislenceueacawensesodeneuanseisesas 5 EG GT EEE EE EE EE NE 6 3 3 1 Easy Move IO AIS ee en eee en nn ea dd ce tn 6 3 3 2 Metal side rails eeceeeeeeeneseeseceeeeeeeeeeeeeececeeeeeeeeeeeeeeecceeeeeene 7 3 3 3 Wooden barriefs ae en ein 2 sance son i bis aa o Lead 1 ei on can se sombre don Leatle ie Le saanes cs 8 3 3 4 Angled lifting pole and IV stand ss 8 15 SATTE a a da a 9 3 3 6 Remote control iiiieeeeeeeeeeeecceeee
22. o patient safety and can cause malfunctioning D Pied dela 2014 11 01 22 23 MEDICATLANTIC IPO2L08 8 6 Specific technical data 8 6 1 Noise The maximum noise level measured on the bed is 48 dBa 8 6 2 Weight Bed safe working load 170 kg Patient 135 kg Mattress 20kg Accessories 15kg A1600 A9300 75 kg 8 kg Gallows safe working load Intravenous post safe working load A1700 A8400 Bed with electric knee break leg lift without boards 62 kg 8 6 3 Dimensional 160 cm Wheel 100 30 to 76 cm MEDICATLANTIC recommend the use of the XS 150 patient lift 470 mm 240 mm 526 mm 637 mm 637 mm 2014 11 01 23 23
23. ocess Measures should therefore be taken to ensure that the bed is no longer used for the purpose it was originally intended 2014 11 01 19 23 IPO2L08 MEDICATLANTIC 6 BED BOARDS 80cm WIDTH COMPATIBLE Reference PITCHOUNE wooden barriers P605 00 PITCHOUNE steel barriers P606 00 7 COMPATIBLE ACCESSORIES A gt Mattress A1700xx A5800 Charge maxi kg IV stand 2hooks Je Telescopic IV stand 2 hooks 8 Remote control lead holder NA Epoxy urinal holder INA Full length KALIN side rails NA KALIN bed ski NA A564 565 00 KALIN EPOXY sideral NA 50200 Kit for side loading bed NA Stainless steel intravenous stand on LI un Lifting pole on U shaped base fixed Only accessories and boards supplied by MEDICATLANTIC guarantee safe USE Width 76 cm minimum with a high resilience foam of 34 kg m minimum height between 74cm min and 15 to 17cm max 2014 11 01 20 23 IPO2ZL08 MEDICATLANTIC 8 SPECIFIC USE 8 1 Purpose of bed gt These beds are intended for children aged 3 to 12 years old height less than 146 cm for Home Treatment and in establishments when they are fitted with the emergency back lift lowering option CPR 8 2 General description Slatted back lift Back lift handle Slatted folding leg lift 4 Head board Mattress Foot board Gallows housings corner ka Intravenous holder mr MN rod housing vw Base a
24. oval of the bed from the establishment o Dispose of the packaging after decontamination of the inside by spraying with a Detergent Disinfectant solution o Bio cleaning operation or o Steam cleaning accessory with microfiber band of the flat surfaces and the base slats Change the washing mops regularly to prevent water accumulating Clean the parts that are difficult to access with a steam nozzle wheels hinges after opening corners etc For tubes use the steam nozzle with a microfiber cloth Never direct the nozzles onto electrical boxes or actuators o Dry hinges with compressed air o Attention Disinfect jacks electrical boxes and remote controls with a microfiber cloth soaked in disinfectant Do not rinse or wipe Check the operation of all the bed functions Repair if necessary Pack in thermoplastic film Attention Inthe event of additional precautions Contact precautions Droplets or Air apply the measures recommended by the hospital hygiene department The use of a Javel water solution of more than 5000ppm 0 5 of active chlorine should be justified by a microbiological risk and only applied for the required time Risk of ageing of some materials especially their colour The concentration of alcohol based surface disinfectant solutions should be less than 30 Note The use of the terminal disinfection process is compatible with the medical bed and its accessories A Product for external use Do not swal
25. should be at least 22 cm For greater safety the side rails can be adapted see accessories To assist patient mobility it is possible to fit a Mobility Aid System S A M TM 2014 11 01 4 23 IPO2L08 MEDICATLANTIC The cleaning instructions recommended must be complied with Only use original parts and accessories supplied by MEDICATLANTIG to guarantee safety and maintain product conformity The bed must not be modified Abnormal use of the bed may damage it or cause accidents to users in which case the warranty shall be annulled Abnormal use means failure to comply with the precautions for use maintenance instructions and other uses not related to the bed s normal purpose such as use of the bed by several people at the same time use outdoors moving the bed on a slope that is steeper than 10 etc 3 2 First use gt Remove the packaging protective devices adhesive tape packing straps and holding clamps see unpacking instructions on the pallet gt Put the bed in the designated room foreseeing an appropriate perimeter of use for the different functions variable height TR etc especially if the bed has a lifting pole or side rails Check that there is sufficient ceiling height if a lifting pole is fitted Brake the wheels The mains socket should remain accessible to enable the bed to be disconnected quickly Plug in the power lead checking that the mains comply with the standards in force and that it is suitable
26. vironment IEC 60601 Compliance level Electromagnetic environment Guide L Immunity test Severity level Electrostatic discharge 6 KV contact 6 KV contact Floors should be wood concrete or ceramic tile If floors are covered with EN 61000 4 2 8 KV air 8 kV air synthetic material the relative humidity should be at least 30 Electrical fast transients 2 KV for feeders 2 KV for feeders The quality of the main power supply must be the same as for a typical EN 61000 4 4 commercial or hospital environment 1 kV for input output lines 1 KV for input output lines Surges Differential mode 1 kV Differential mode 1 kV The quality of the main power supply must be the same as for a typical EN 61000 4 5 commercial or hospital environment Common mode 2 kV Voltage dips short lt 5 Ur for 10 ms lt 5 Ur for 10 ms The quality of the main power supply must be the same as for a typical interruptions and voltage 40 Ur for 100 ms 40 U for 100 ms commercial or hospital environment variations Don x If the user of the medical bed see references in contents wants to be 20 0e Dr 00 me 70 Jr dor 200 m able to continue to use the bed during interruptions in the main power EN 61000 4 1 1 lt 5 Ur for 5s lt 5 Ur for 55 supply it is recommended that the bed be powered by a converter or battery Power frequency Power frequency magnetic fields should be at levels characteristic of a magnetic field location