Gas Clean Filter System User Guide
Contents
1. When filters are replaced they must be treated as chemical waste and disposed of according to local policy Using the Gas Clean Filter System Installation Before you begin Be sure to choose a centralized installation location where all instruments are easily reachable and where the filter s indicating material is always visible Gas Clean Filter connecting units are available for both 1 4 inch and 1 8 inch gas lines Be sure you have selected the appropriate connecting unit to use with your lab setup Prepare your instrument Prepare your instrument by doing the following 1 Lower any heated zone temperatures on your instrument to less than 100 C 2 Set any purge flows on your instrument to 400 mL min 3 If your system pressure is higher than 7 bar 100 psi reduce the pressure 10 User Guide User Guide Using the Gas Clean Filter System Install the connecting unit The connecting unit can be mounted on a laboratory benchtop affixed to the wall with optional wall mounting bracket for the single filter connecting unit or attached to the top of the instrument Installing on a laboratory bench top You can install the connecting unit onto the laboratory bench top using optional hook and loop adhesive fabric strips or with two screws not supplied This section describes how to mount your connecting unit to a laboratory bench top using screws If you wish to use the hook and loop fabric strips refer to
2. Agilent Gas Clean Filter System User Manual fide Agilent Technologies Notices Agilent Technologies Inc 2012 No part of this manual may be reproduced in any form or by any means including electronic storage and retrieval or translation into a foreign language without prior agreement and written consent from Agilent Technologies Inc as governed by United States and international copyright laws Manual Part Number 5973 1528 Edition First edition August 2012 Printed in USA Agilent Technologies Inc 2850 Centerville Road Wilmington DE 19808 1610 USA Warranty The material contained in this document is provided as is and is subject to being changed without notice in future editions Further to the maximum extent permitted by applicable law Agilent disclaims all warranties either express or implied with regard to this manual and any information contained herein including but not limited to the implied warranties of merchantability and fitness for a particular purpose Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing use or performance of this document or of any information contained herein Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms the warranty terms in the separate agreement shall control Safety Notices CAUT
3. For CP7980 and CP 988 Upper part filter connecting unit Flush head for connecting unit Male connector 1 4 in with dust filter Male connector 1 8 in with dust filter O rings two sets Male connector stainless steel 1 4 in with dust filter Male connector stainless steel 1 8 in with dust filter Male connector stainless steel 3 mm with dust filter Male connector stainless steel 6 mm with dust filter Part No CP7981 CP7978 CP 987 CP7986 CP82117 CP7983 CP7986SS CP82117SS CP8211 7SS3 CP7986SS6 41 Using the Gas Clean Filter System 42 User Guide fee Agilent Technologies Agilent Technologies Inc 2012 5973 1528
4. Gas Clean Filter system delivers high purity gases to your analytical instruments reducing the risk of column damage sensitivity loss and instrument downtime The filters are designed for use with the GC GC MS ICP OES ICP MS LC MS and any other analysis instrument using carrier gas Six filters are available including CO oxygen moisture and organics trap charcoal The Gas Clean Filter System consists of a connecting unit and filter Figure 1 The connecting unit is equipped with inlet and outlet connectors for the gas lines and two needle like valves that automatically start the flow of gas once a filter is attached The filters are made from a heavy walled polycarbonate sealed at the base with PTFE When installed onto the connecting unit the needle like valves puncture the PTFE seals and allow gas to flow through the filter i LS EN Filter Connecting unit plimum of this unit can be gua ra nbeeed only by using Agilent Gas Clean Filters Outlet connector Inlet connector Figure 1 The components of the Gas Clean Filter System User Guide User Guide Features Filter types Using the Gas Clean Filter System The Gas Clean Filter System delivers the following features and benefits Cleaner gas longer instrument life greater sensitivity higher data accuracy and overall long term stability for your instruments and analyses A high flow connecting unit th
5. If your filter s indicator changes color from the bottom up shortly after installation see Figure 33 this indicates a leak downstream of the filter in your gas line Check for leaks between the filter and your instrument Filter material Indicating material original color Indicating material color change Figure 33 Indicating a leak downstream from the filter If the entire indicator has not changed color from the bottom up the filter is not consumed and is still usable see Figure 33 38 User Guide Using the Gas Clean Filter System Replacement Parts Table 3 Agilent Gas Clean Filter Starter Kits Description Part No Agilent Gas Clean FID Filter kit CP7995 Includes a 4 position 1 8 in connecting unit and two Charcoal filters one Oxygen filter and one Moisture filter Gas Clean Filter Kit 1 8 in CP736530 Includes a 4 position 1 4 in connecting unit and two Charcoal filters one Oxygen filter and one Moisture filter Agilent Gas Clean GC MS Filter kit 1 8 in CP1 976 Includes a 1 position connecting unit 1 8 in and two GC MS filters Agilent Gas Clean GC MS Filter kit 1 4 in CP1 977 Includes a 1 position connecting unit 1 4 in and two GC MS filters Agilent Gas Clean GC MS Filter Installation kit CP1 978 Includes 1 m copper tubing part no CP1 976 two nuts and two ferrules 1 8 in Agilent Gas Clean CO kit CP1 982 Includes 2 position 1 4 in connecting unit and CO and moisture filters Agilent Ga
6. filter see Figure 27 Filter material Indicating material Figure 27 Flow of gas through the Gas Clean Filter 32 User Guide Using the Gas Clean Filter System Replacing the Gas Clean Filter When to replace the Gas Clean Filter As gas passes through the filter material the filter picks up impurities and eventually becomes consumed over time When this occurs an increased amount of impurities reach the indicating material which causes the material to change color from the top down When 75 or more of the indicating material s color has changed this indicates that the filter is consumed see Figure 28 Filter material Indicating material color change Indicating material original color Figure 28 Indicating material color change Agilent recommends replacing the filter when 75 or more of the indicator s color has changed from the top down or within one year of installation User Guide 33 Using the Gas Clean Filter System 34 The indicator color and consumed indicator colors are different for each type of filter See Table 2 for a description of each filter indicator color and consumed color change Table 2 Indicating material color description Filter type Original color Consumed color Oxygen Filter Green Gray Moisture Filter Green Pale brown Process Moisture Filter Charcoal Filter No indicator No indicator GC MS Filter Oxygen Green Gray Moisture Green Pale brown CO Filter
7. the filter CAUTION User Guide Using the Gas Clean Filter System Wear clean lint free gloves to prevent contamination of parts with dirt and skin oils To install the filter on the connecting unit 1 Unscrew the knurled ring and remove the red plastic dust cap from the top of the connecting unit Figure 18 Optimum performance of this unit can be gua ranteed only by using Agilen t Gas Clean Filters Red plastic dust cap 35N 240438 dvd 310W34 Knurled ring Figure 18 Remove the red plastic dust cap 25 Using the Gas Clean Filter System 2 Flush the gas line of oxygen and moisture If available use the optional flush head part number CP7987 to flush the gas line Figure 19 Place the flush head over the valves and alignment pin and push down until it rests on the base of the connecting unit Install the knurled ring to hold it in place 2 Flush head Figure 19 Using the flush head to flush the gas line of oxygen and moisture 26 User Guide Using the Gas Clean Filter System If you do not have the flush head use a small object or your finger to depress the inlet valve The inlet valve has an O ring positioned at the top of the cylinder as shown in Figure 20 CR 1 NS O ring A4 Z Inlet valve Outlet valve Figure 20 Depress the inlet valve to flush the gas line of oxygen and moisture Flush the gas lines for at least several
8. the steps in the section Installing to an instrument top on page 19 Gather the following e Screws 2 at least 25 mm long and less than 5 mm od not supplied e Screw driver or power drill To install the connecting unit to the laboratory bench top 1 Remove the two black plastic caps by pushing a thin narrow object such as a paper clip from the bottom of the connecting unit into the underside of the caps until they pop out Figure 2 Figure 2 Removing the plastic caps from the connecting unit 11 Using the Gas Clean Filter System 2 Determine the best location on your laboratory bench top to install the connecting unit and filter Make sure the filter s indicator is clearly visible so that exhausted filters can be easily identified Make sure you have ample space above the bench top and connecting unit for installation and removal of the filter Make sure the gas lines reach the location and will not interfere with other activities in the chosen installation location Check that the mounting screws will not penetrate any dangerous objects such as gas and power lines 3 Place the connecting unit in the chosen location and use a screw driver or power drill to install the screws into the bench top through the two openings on the top of the connecting unit 4 Replace the black plastic caps on the connecting unit to cover the two screw hole openings Installing on a wall This option is only availa
9. ION A CAUTION notice denotes a hazard It calls attention to an operating procedure practice or the like that if not correctly performed or adhered to could result in damage to the product or loss of important data Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met A WARNING notice denotes a hazard It calls attention to an operating procedure practice or the like that if not correctly performed or adhered to could result in personal injury or death Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met Agilent Gas Clean Filter System User Manual Using the Gas Clean Filter System About the Gas Clean Filter System 4 Components 4 Features 5 Filter types 5 Safety Information 9 General gas safety 9 Cleaning 9 Recycling the product 9 Installation 10 Before you begin 10 Prepare your instrument 10 Install the connecting unit 11 Connect the gas lines 22 Install the filter 25 Install a two or four position connecting unit 30 Gas Flow ina Gas Clean Filter 32 Replacing the Gas Clean Filter 33 When to replace the Gas Clean Filter 33 To replace the Gas Clean Filter 35 Checking for Leaks 37 Indicator changes color from the top down 37 Indicator changes color from the bottom up 38 Replacement Parts 39 ee Agilent Technologies Using the Gas Clean Filter System About the Gas Clean Filter System Components The Agilent
10. White Violet See To replace the Gas Clean Filter on page 35 for more information User Guide Using the Gas Clean Filter System To replace the Gas Clean Filter To replace the Gas Clean Filter 1 Lower any heated zone temperatures on your instrument to less than 100 C 2 Set any split vent flows purge flows on your instrument to 400 mL min 3 If your system pressure is higher than 7 bar 100 psi reduce the pressure to a pressure no lower than 2 bar 80 psi to ensure easy removal of the saturated filter 4 Remove the saturated filter by unscrewing the knurled ring that secures the filter to the connecting unit Figure 27 Figure 29 Unscrew the knurled ring 5 Remove the new filter and O rings from the packaging User Guide 35 Using the Gas Clean Filter System 6 Remove the old upper set of O rings using the supplied tool and replace with new O rings Figure 30 Figure 30 Remove upper pair O rings only when replacing filter 7 Remove the two aluminum plugs from the bottom of the new filter Figure 31 Figure 31 Aluminum plugs removed from bottom of the new filter 8 Place the knurled ring over the filter then put the filter on top of the connecting unit it will only fit one way due to the alignment pin 9 Tighten the knurled ring while gently pressing down on the filter let the ring pull the filter onto the connecting unit 10 Carefully check the connections for leaks See Chec
11. an Filter System d Install the inlet and outlet fittings on the underside of the connecting unit using a 1 2 inch wrench Figure 8 Outlet fitting Inlet fitting Figure 8 Inlet and outlet fittings installed on the underside of the connecting unit 17 Using the Gas Clean Filter System 7 Install the connecting unit to the wall mounting bracket a Depending on the inlet outlet fitting orientation on the connecting unit align the connecting unit to the bracket as shown in Figure 9 Figure 9 Connecting unit orientations on wall mounting bracket b Using your hand install the hex screws from the bottom up until they are finger tight Figure 10 Figure 10 Hex screws installed c Use a 4 mm hex wrench to tighten the screws The connecting unit should be tight against the wall mounting bracket 18 User Guide User Guide Using the Gas Clean Filter System Installing to an instrument top Gather the following e Hook and loop fabric strips 5066 1110 To install the connecting unit to the top of an instrument 1 Peel the paper adhesive cover from one side of each fabric strip and apply the exposed adhesive to the underside of the connecting unit as shown in Figure 11 Underside of base plate Hook and loop fabric strips n z re 7 fy abe a Ca j ar 9 W rx x Loft vA ey sd A S p PONN ia 1 j TRA oe aa SL x pes 4 ne a p Figure 11 Hook an
12. at handles flow rates up to 20 L min for collision gas applications and supply gas for ICP MS ICP OES and LC MS A fast stabilizing and absorbent filter material Filter housing constructed with a transparent and virtually unbreakable heavy walled polycarbonate material The ability to change the Gas Clean Filter without tools or gas shut off Each Gas Clean Filter type is designed to filter out a specific impurity that may exist in the gas supply The following filter types are available Carbon Dioxide Eliminates CO from supply gas to the Total Organic Carbon TOC analyzer and improves sensitivity and accuracy in TOC analysis Oxygen Prevents oxidation of the GC column septum liner and glass wool Moisture Delivers faster stabilization times for increased GC productivity and prevents hydrolization damage to the stationary phase column liner glass wool or septum in the GC Process Moisture Prevents oxidation of GC components and is safe to use with acetylene in process GC applications Charcoal Removes organic compounds and ensures correct performance of FID detectors in the GC GC MS Delivers faster stabilization times for increased GC productivity removes oxygen moisture and hydrocarbons from the carrier gas for MS applications and provides ultimate GC column protection Table 1 on page 6 shows recommended filter connection diagrams for common instrument configurations Using the Gas Cl
13. ble for a single filter connecting unit using the optional wall mounting bracket Gather the following e Wall mounting bracket kit CP7981 e Flathead screw driver e Hex wrench 4 mm e Power drill with 5 mm od drill bit e Pencil To install the wall mounting bracket 1 Determine the location on the wall where you wish to install the wall mounting bracket Make sure you have ample space above the mounting bracket with connecting unit installed for installation removal of the filter Make sure the gas lines reach the location and will not interfere with other activities in the chosen installation location Check that the mounting screws will not penetrate any dangerous objects such as gas and power lines once installed 12 User Guide Using the Gas Clean Filter System 2 Hold the wall mounting bracket against the wall in the location that you plan to install it and use a pencil to mark the drilling location for the screw holes Figure 3 Figure 3 Mark the drilling locations on the wall using the screw holes in the wall mounting bracket as a guide Lay the mounting bracket down on a flat surface 3 Using a power drill with a 5 mm od drill bit drill about 5 cm into the wall in both locations Wipe away any excess dust and residue 4 If required install the supplied drywall anchors into the drilled holes and make sure each anchor is flush against the wall s edge Figure 4 Figure 4 Plastic drywall anchors
14. d loop fabric strips attached to the underside of the connecting unit 19 Using the Gas Clean Filter System CAUTION 20 2 Peel and remove the paper adhesive covers from the bottom of the hook and loop fabric strips Figure 12 Underside of base plate Figure 12 Peel and remove the adhesive covers from the bottom of the hook and loop fabric strips Be sure to not cover any vents or heated areas on the instrument surface User Guide Using the Gas Clean Filter System 3 Place the connecting unit onto your instrument Press firmly down so the adhesive sets properly on both the connecting unit and instrument surface Figure 13 l Figure 13 Filter base attached to the top of instrument using the supplied hook and loop fabric strips User Guide 21 Using the Gas Clean Filter System Connect the gas lines Before starting make sure you have enough gas line tubing to reach the area where you wish to mount the Gas Clean Filter For best results use pre cleaned gas lines Gather the following e Swagelok nuts included with the connecting unit assembly e Front and back ferrules included with the connecting unit assembly e Wrench 7 16 inch for 1 8 inch nuts or 9 16 inch for 1 4 inch nuts e Pre cleaned tubing e Electronic leak detector such as the Agilent G3388B Handheld Electronic Gas Leak Detector To connect the gas lines to the connecting unit 1 Place a Swagelok nut back ferrule a
15. ean Filter System Table 1 Connection diagrams for common detectors Detector Connection Diagram ECD Carrier Gas Electron Capture Detector Make Up Gas Anode Purge Gas a ECD FID Carrier Gas Flame lonization Detector Carrier Gas Make Up Gas Hydrogen gt Charcoal Filter Ai Charcoal Filter FID Carrier Gas gt GC MS Er gt Flame lonization Detector Carrier Gas differs Make UpGas Charcoal Filter from Make Up Gas Hydrogen Charcoal Filter O mt MN Charcoal Filter Carrier Gas FPD Flame Photometric Detector Hydrogen gt Charcoal Filter Flame 1 PFPD p Charcoal Filter Pulsed Flame Photometric EED Detector Flame 2 6 User Guide Using the Gas Clean Filter System Table 1 Connection diagrams for common detectors continued Detector Connection Diagram MS ITD MSD Carrier Gas gt GC MS Fier Com lon Trap Detector Mass Selective Detector MS Carrier Gas MS NPD PND Carrier Gas Nitrogen Phosphorous Detector Make Up Gas TID TSD Hydrogen gt Charcoal Filter Thermionic Detector o TSD Ai Charcoal Filter Carrier Gas Make Up Gas TCD Carrier Gas Thermal Conductivity Detector Reference Channel Make Up Gas __ ICD if necessary User Guide 7 Using the Gas Clean Filter System Table 1 Detector ICP OES ICP MS Ind
16. installed in wall User Guide 13 Using the Gas Clean Filter System 5 Hold the wall mounting bracket in place against the wall using one hand then use your other hand to install each screw into the wall or drywall anchor Figure 5 Finger tighten the screws as far as possible and then use a screwdriver to complete the screw installation The wall mounting bracket should rest flush against the wall Figure 5 Installing the wall mounting bracket 14 User Guide User Guide Using the Gas Clean Filter System 6 The connecting unit supports gas connections on the base plate edge for surface mounting and from below when mounting the connecting unit against a wall The connecting unit ships from the factory with the gas connections installed on the edge for surface mounting For wall mounting you can optionally change the gas inlet outlet locations to the bottom as follows a Using a 5 mm hex wrench remove the two plugs on the underside of the connecting unit Figure 6 Outlet fitting Inlet fitting Plugs Figure 6 Plugs and inlet outlet fittings on the connecting unit b Using a 1 2 inch wrench remove the inlet and outlet fittings on the end of the connecting unit Figure 6 15 Using the Gas Clean Filter System c Install the plugs on the end of the connecting unit using a 5 mm hex wrench Figure 7 Figure 7 Plugs installed on the end of the connecting unit 16 User Guide User Guide Using the Gas Cle
17. king for Leaks on page 37 for more information Treat used filters as chemical waste and dispose of them according to local policy 36 User Guide Checking for Leaks User Guide Using the Gas Clean Filter System During leak free operation the filter indicator should not change color for several months depending on the quality of your gas If you notice a color change in your filter s indicating material immediately following installation this indicates a gas leak in your system or poor gas quality Take note whether the color change occurred from the top down or from the bottom up and refer to one of the following sections e Indicator changes color from the top down on page 37 e Indicator changes color from the bottom up on page 38 Indicator changes color from the top down CAUTION If the indicating material changed color from the top down see Figure 32 there is a leak upstream of the filter in your gas line or your gas quality is poor Check for leaks at the cylinder regulator and fittings and check your gas quality Filter material Indicating material color change Indicating material original color Figure 32 Indicating a leak upstream from the filter When about 5 of the indicator changes color from the top down this indicates that the entire filter is consumed and needs to be replaced 37 Using the Gas Clean Filter System Indicator changes color from the bottom up
18. mal operating supply pressure and method settings Remember to keep at a minimum supply pressure of 2 bar 30 psi 29 Using the Gas Clean Filter System Install a two or four position connecting unit The steps for installing a two or four position connecting unit are similar to the steps for the single filter connecting unit Refer to the previous sections in this chapter for details Installation on page 10 Two position connecting unit When installing the two position connecting unit refer to Figure 24 for gas line and filter installation locations Oxygen Moisture Figure 24 Two position connecting unit 30 User Guide User Guide Using the Gas Clean Filter System Four position connecting unit When installing the four position connecting unit refer to Figure 25 and Figure 26 for common gas line and filter configurations Instrument Air Carrier Gas Carrier Gas Hydrogen In Out In Out In Out Charcoal Oxygen Moisture Charcoal Figure 25 Configuration 1 Four position connecting unit Instrument Air Carrier Gas Hydrogen Makeup Gas In Out In Out In Out In Out Charcoal GC MS Charcoal Charcoal Figure 26 Configuration 2 Four position connecting unit 31 Using the Gas Clean Filter System Gas Flow in a Gas Clean Filter Gas flows in an upward direction from the bottom of the filter to the top and then downward through the filter material indicating material and out the bottom of the
19. minutes to ensure the gas lines are free of oxygen and moisture User Guide 27 Using the Gas Clean Filter System 3 Remove the filter from the packaging then remove the two aluminum plugs from the bottom of the filter Figure 21 Figure 21 Remove the aluminum plugs from the filter 4 Place the knurled ring over the filter then place the filter on top of the connecting unit it will only fit one way due to the alignment pin shown in Figure 22 Be careful to not damage the connecting unit and filter components by forcing an incorrect filter orientation Alignment pin Figure 22 Alignment pin 28 User Guide User Guide Using the Gas Clean Filter System 5 Tighten the knurled ring while gently pressing down on the filter Let the knurled ring pull the filter onto the connecting unit as it is tightened Figure 23 ETT eel Knurled ring anaes Tes be Ver i RTT CCE oP a qe E Pha tt a Barres LEUR Figure 23 Tighten the knurled ring to secure the filter to the connecting unit 6 Allow the filter to pressurize for 3 minutes 7 Flush the filter Set the instrument split vent flow purge flow to 400 mL min then If using a Moisture filter in combination with an MS detector purge for at least 60 minutes This will flush the filter of dry nitrogen so that it does not appear as a background ion in your MS analysis For all other filter and detector combinations purge for 4 minutes 8 Restore to nor
20. nd front ferrule to the tubing as shown in Figure 14 Front ferrule Important Backacrrule The narrow end of the Nut back ferrule fits into the rear of the front ferrule Tubing Figure 14 Swagelok nuts and ferrules 22 User Guide User Guide Using the Gas Clean Filter System 2 Push the tubing towards the base until it rests firmly against the shoulder in the fitting and make sure that the front ferrule is touching the fitting Slide the Swagelok nut over the ferrule and thread it onto the fitting Figure 15 Nut and ferrules Fitting Tubing Figure 15 Assemble the fitting 3 Push the tube fully into the fitting then withdraw it approximately 1 to 2 mm see Figure 16 Front Nut ferrule Insert tubing fully Back ferrule Withdraw tubing 1 2 mm Tighten nut Figure 16 Insert the tubing 4 Finger tighten the nut Then use a wrench to tighten the nut 3 4 turn for 1 8 inch tubing 5 Set the gas line pressure to a value between 2 bar 30 psi and 4 bar 60 psi when installing the filter to the connecting unit Be sure to keep the pressure above 2 bar 80 psi so that air does not enter the filter 23 Using the Gas Clean Filter System 6 Check all connections for leaks using the leak detector Figure 17 shows gas lines connected to the filter connecting unit Figure 17 Properly connected gas lines single connecting unit shown 24 User Guide Install
21. s Clean CO kit CP1 983 Includes 2 position 1 8 in connecting unit and CO and moisture filters TCD Filter Kit CP738408 Includes a 2 position 1 8 in connecting unit one Oxygen filter and one Moisture filter GC Installation Kit 19199N Includes CP 736530 and many useful fittings and accessories User Guide 39 Using the Gas Clean Filter System Table 4 Connecting Units Description Part No 1 position connecting unit 1 4 in CP7980 1 position connecting unit 1 8 in CP7988 2 position connecting unit 1 4 in CP738406 2 position connecting unit 1 8 in CP738407 4 position connecting unit 1 4 in CP7989 4 position connecting unit 1 8 in CP736520 High flow connecting unit 1 4 in CP1 984 High flow connecting unit 1 8 in CP1 985 1 position stainless steel connecting unit 1 4 in CP7980P4 1 position stainless steel connecting unit 1 8 in CP 988P8 1 position stainless steel connecting unit 3 mm CP7988P3 1 position stainless steel connecting unit 6 mm CP7980P6 Table 5 Replacement Gas Clean Filters Gas Clean Filters Part No Agilent Gas Clean CO Filter CP1 969 Agilent Gas Clean Oxygen Filter CP1 970 Agilent Gas Clean Moisture Filter CP1 971 Agilent Gas Clean Process Moisture Filter CP17971P Agilent Gas Clean Charcoal Filter CP17972 Agilent Gas Clean Filter GC MS CP1 973 User Guide 40 User Guide Using the Gas Clean Filter System Table 6 Accessories and Fittings Gas Clean Filters Wall mounting bracket for connecting unit
22. uctive Coupled Plasma Optical Emission Spectrometry Mass Spectrometry Plasma Gas Nebulizer Gas ICP 0ES Inductive Coupled Plasma Optical Emission Different Optic Flush Gas Total Organic Carbon TOC Analyzer Zero Air Generator LC MS Mass Spectrometry Connection diagrams for common detectors continued Connection Diagram Feu 7 ICP MS Supply GAS ICP OES Supply Gas E ICP OES High Flow Base Plate Optic Flush Gas Carrier Gas TOC Analyzer Zero Air Generator Supply Gas LC MS High Flow Base Plate Contact your local Agilent sales representative for the filter sets applicable to your instrument configuration User Guide Safety Information User Guide General gas safety Cleaning Using the Gas Clean Filter System Wear eye protection when using compressed gas to avoid eye injury Fasten all compressed gas cylinders securely to an immovable structure or permanent wall Store and handle compressed gases in accordance with relevant safety codes Do not put gas cylinders in the path of a hot air vent including a GC oven exhaust Perform periodic leak checks on supply lines fittings and pneumatic plumbing to prevent a potentially hazardous condition Wipe away dust and residue using a lint free cloth Once the filter is consumed the filter and filter material cannot be replenished or reused Recycling the product