Home
Endress Promag 23H User Manual
Contents
1. Flange DIN 11864 2 Form A DN di G D L LK1 LK2 1 4404 316L mm mm mm mm mm mm mm letal a SHAS AAS eS 40 38 82 92 64 71 0 65 F 7 z 50 50 94 105 64 83 5 77 x 4 S 65 66 113 121 64 100 0 95 o 8 80 81 133 147 98 121 0 112 P 100 100 159 168 98 141 5 137 Fitting length 2 x L 140 mm DN 40 65 200 mm DN 80 100 eh If pigs are used for cleaning it is essential to take the inside diameters of measuring tube Page 15 and process connection di into account Coupling SMS 1145 DN di G D L LK 1 4404 316L mm mm mm mm mm mm Bele Os ae 40 35 5 Rd 60 x 1 6 92 63 71 0 50 48 5 Rd 70 x 1 6 105 65 83 5 x a d S 65 60 5 Rd 85 x 1 6 121 70 100 0 o o amp 5 80 72 0 Rd 98 x 1 6 147 75 121 0 ro S 100 97 6 Rd 132 x 1 6 168 70 141 5 Fitting length 2 x L 140 mm DN 40 65 200 mm DN 80 100 If pigs are used for cleaning it is essential to take the inside diameters of measuring tube Page 15 and process connection di into account Fitting length 2 x L 140 mm DN 40 65 200 mm DN 80 100 Coupling ISO 2853 DN di G D L LK 1 4404 316L mm mm mm mm mm mm Syl ma owe ne te 40 35 6 50 6 92 61 5 71 0 50 48 6 64 1 105 61 5 83 5 al S 65 60 3 77
2. 1 vent valve 2 siphon Endress Hauser 9 PROline Promag 23 H Orientation An optimum orientation position helps avoid gas and air accumulations and deposits in the measuring tube Promag nevertheless supplies a range of functions and accessories for correct measuring of problematic fluids Empty Pipe Detection EPD ensures the detection of partially filled measuring tubes e g in the case of degassing fluids or varying process pressures Vertical orientation This orientation is ideal for self emptying piping systems and for use in conjunction with Empty Pipe Detection FO6 2xxxxxxx 11 00 00 xx 004 Horizontal orientation The measuring electrodes should be on a horizontal plane This prevents brief insulation of the two electrodes by entrained air bubbles Achtung Empty Pipe Detection functions correctly only when the measuring device is installed horizontally and the transmitter housing is facing upward Otherwise there is no guarantee that Empty Pipe Detection will respond if the measuring tube is only partially filled or empty FO6 23Hxxxxx 1 1 00 xx xx 000 1 EPD electrode Empty Pipe Detection except for DN 2 8 2 Measurement electrodes signal acquisition 10 Endress Hauser PROline Promag 23 H Vibrations Secure the piping and the sensor if vibration is severe Information on resistance to vibration and shock can be found on Page 13
3. Ambient temperature 20 60 C Install the device at a shady location Avoid direct sunlight particularly in warm climatic regions Storage temperature 10 50 C preferably 20 C Degree of protection IP 67 NEMA 4X Shock and vibration Acceleration up to 2 g by analogy with IEC 68 2 6 resistance CIP cleaning Possible SIP cleaning Possible Electromagnetic To EN 61326 and NAMUR recommendation NE 21 compatibility EMC Process conditions Medium temperature range The permissible medium temperature depends on the sensor and the sealing material Sensor e 20 150 C 180 C in preparation for DN 2 25 e 20 150 C for DN 40 100 Seal e EPDM 20 130 C e Silicone 20 150 C e Viton 20 150 C e Kalrez 20 150 C Conductivity Minimum conductivity gt 50 uS cm for fluids in general Medium pressure range nominal pressure The permissible nominal pressure depends on the process connection and seal e 40 bar flange weld nipple with O ring seal e 16 bar all other process connections Pressure tightness liner Nominal diameter Measuring tube Resistance to partial vacuum of measuring tube lining lining Limit values for abs pressure mbar at various fluid temperatures mm inch 25 C 80 C 100 C 130 C 150 C 180 C 2 100 1 12 4 PFA 0 0 0 0 0 0
4. DN A B G D E F G N L K mm mm mm mm mm mm mm mm mm mm mm 40 122 0 86 71 0 51 0 35 3 M8 15 18 4 50 147 0 99 83 5 63 5 48 1 M8 15 18 4 65 147 0 115 100 0 76 1 59 9 M8 15 18 6 80 197 0 141 121 0 88 9 72 6 M 12 15 20 4 100 197 0 162 141 5 114 3 97 5 M 12 15 20 6 Process connections with aseptic gasket seals DN 40 100 Weld nipple for pipe DIN 11850 DN di G D L L1 LK 1 4404 316L mm mm mm mm mm mm mm BAHEA AAAA 40 38 0 43 92 42 19 71 0 X 50 50 0 55 105 42 19 83 5 pren aa xe o o A 8 65 66 0 72 121 42 21 100 0 80 81 0 87 147 42 24 121 0 g 100 100 0 106 168 42 24 141 5 L1 E Fitting length 2 x L 140 mm DN 40 65 200 mm DN 80 100 aed If pigs are used for cleaning it is essential to take the inside diameters of measuring tube Page 15 and je n process connection di into account Weld nipple for pipe ODT DN di G D L L1 LK 1 4404 316L mm mm mm mm mm mm mm Tain A e SEAS ot 40 35 3 40 92 42 19 71 0 50 48 1 55 105 42 19 83 5 8 65 59 9 66 121 42 21 100 0 80 72 6 79 147 42 24 121 0 8 100 97 5 104 168 42 24 141 5 Fitting length 2 x L 140 mm DN 40 65 200 mm DN 80 100 If pigs are used for cleaning it is essential to take the inside
5. Endress Hauser 13 PROline Promag 23 H Limiting flow The diameter of the pipe and the flow rate determine the nominal diameter of the sensor The opti mum velocity of flow is 2 3 m s The velocity of flow v moreover has to be matched to the phy sical properties of the medium e v gt 2 m s for media forming coatings e g in full fat milk etc Flow characteristics of Promag H SI units Nominal dia Recommended Factory settings meter flow rate Min max full scale value Full scale value Pulse weighting Creepage mm inch v 0 3 or 10 m s v 2 5 m s 2 pulse s v 0 04 m s 2 1 12 0 06 1 8 dm min 0 5 dm min 0 005 dm 0 01 dm min 4 5 32 0 25 7 dm min 2 dm min 0 025 dm 0 05 dm min 8 5 16 1 30 dm min 8 dm min 0 10 dm 0 1 dm min 15 1 2 4 100 dm min 25 dm min 0 20 dm 0 5 dm min 25 1 9 300 dm min 75 dm min 0 50 dm 1 dm min 40 1 1 2 25 700 dm min 200 dm min 1 50 dm 3 dm3 min 50 2 35 1100 dm min 300 dm min 2 50 dm 5 dm9 min 65 21 2 60 2000 dm min 500 dm min 5 00 dm 8 dm3 min 80 3 90 3000 dm min 750 dm min 5 00 dm 12 dm min 100 4 145 4700 dm min 1200 dm min 10 00 dm 20 dm min Flow characteristics of Promag H US units Nominal Recommended Factory settings diameter flow rate Min max full scale value Full scale value Pulse weigh
6. FO6 2xxxxxxx 1 1 00 00 xx 006 Inlet and outlet runs If possible install the sensor well clear of fittings such as valves T pieces elbows etc Com pliance with the following requirements for the inlet and outlet runs is necessary in order to ensure measuring accuracy e Inlet run gt 5x DN e Outlet run gt 2 x DN gt 5xDN FO6 2xxxxxxx 11 00 00 xx 005 Endress Hauser 11 PROline Promag 23 H Adapters Suitable adapters to E DIN EN 545 double flange junction sections can be used to install the sensor in larger diameter pipes The resultant increase in the rate of flow improves measuring accuracy with very slow moving fluids The nomogram shown here can be used to calculate the pressure loss caused by reducers and expanders The nomogram applies only to fluids of viscosity similar to water 1 Calculate the ratio of the diameters d D 2 From the nomogram read off the pressure loss as a function of flow velocity downstream from the reduction and the d D ratio mbar 100 10 05 06 07 08 09 d D F06 2xxxxxxx 05 05 xx xx 000 Pressure loss e No pressure loss with nominal diameters DN 08 and larger if the sensor is installed in a pipe of the same nominal diameter e Pressure losses for configurations incorporating adapters according to E DIN EN 545 see Adapters on Page 12 12 Endress Hauser PROline Promag 23 H Ambient conditions
7. 60 2 77 4 121 68 6 100 0 o S 80 3 72 9 90 9 147 68 6 121 0 100 4 97 4 118 9 168 68 6 141 5 If pigs are used for cleaning it is essential to take the inside diameters of measuring tube Page 15 and process connection di into account Coupling DIN 11851 DN di G D L LK 1 4404 316L mm mm mm mm mm mm clean cea ere 40 38 Rd 65 x 1 6 92 72 71 0 rr 50 50 Rd 78 x 1 6 105 74 83 5 za 5 65 66 Rd 95 x 1 6 121 78 100 0 To fo o O 5 80 81 Rd 110 x 1 4 147 83 121 0 Y g 100 100 Rd 130 x 1 4 168 92 141 5 Fitting length 2 x L 140 mm DN 40 65 200 mm DN 80 100 L 8 If pigs are used for cleaning it is essential to take the inside diameters of measuring tube Page 15 and process connection di into account Endress Hauser 25 PROline Promag 23 H Coupling DIN 11864 1 Form A DN di G D L LK 1 4404 316L mm mm mm mm mm mm GARE SEAS EAR SENS 40 38 Rd 65 x 1 6 92 71 71 0 50 50 Rd 78 x 1 6 105 71 83 5 x 5 3 65 66 Rd 95 x 1 6 121 76 100 0 oO PF ol g 8 80 81 Rd 110 x 1 4 147 82 121 0 yY4 100 100 Rd 130 x 1 4 168 90 141 5 5 L Fitting length 2 x L 140 mm DN 40 65 200 mm DN 80 100 If pigs are used for cleaning it is essential to take the inside diameters of measuring tube Page 15 and process connection di into account
8. 8 Pipe 13 x 1 5 10 54 48 5 9 60 x 42 Sy DIN 11850 DN 10 x 7 15 Pipe 19 x 1 5 16 59 48 5 9 60 x 42 co S 4 ol O Ue DIN 11850 DN 15 I 7 25 DIN Pipe 29 x 1 5 26 70 48 5 9 70 x 52 S DIN 11850 DN 25 f Fitting length 2 x L 86 mm at 8 If pigs are used for cleaning it is essential to take the inside diameters of measuring tube Page 15 and process connection di into account Coupling SMS 1145 Sensor Piping di G L Hx W 1 4404 316L Taio lain grrr DN mm Screw union mm mm mm mm inch 1 25 ANSI 1 22 1 Rd 40 x 1 6 30 8 70 x 52 es Fitting length 2 x L 86 mm e g 7 If pigs are used for cleaning it is essential to take the inside diameters of measuring tube Page 15 and oO ES x process connection di into account oe ae f L Process connections orderable only as accessories with O ring seal DN 2 25 External pipe thread NPT Sensor Piping di G L S HxW 1 4404 316L DKH GD DN mm Internal thread mm inch mm mm mm inch 2 8 NPT 3 8 10 3 8 50 15 5 60 x 42 La eZ 15 NPT 1 2 16 1 2 50 20 0 60 x 42 a m S o x 1 25 ANSI NPT 1 25 1 55 25 0 70 x 52 m S Fitting length 2 x L 86 mm i p T 22 Endress Hauser PROline Promag 23 H Int
9. Endress Hauser PROline Promag 23 H Clamp ISO 2852 Sensor Piping di G L Hx W 1 4404 316L BHW a DN inch Clamp mm mm mm mm 1 25 ANSI Pipe 25 4 x 1 65 22 6 50 5 44 5 70 x 52 ISO 1 o Fitting length 2 x L 86mm O 5 a S If pigs are used for cleaning it is essential to take the inside diameters of measuring tube Page 15 and 5 process connection di into account L Clamp DIN 32676 Sensor Piping di G L HxW 1 4404 316L B H 07nn rrr DN mm Clamp mm mm mm mm 2 8 Pipe 14x2 10 34 0 41 0 60 x 42 DIN 11850 DN 10 15 Pipe 20 x 2 16 34 0 41 0 60 x 42 5 5 DIN 11850 DN 15 5 E 5 8 25 DIN Pipe 30 x 2 26 50 5 44 5 70 x 52 x DIN 11850 DN 25 i E Fitting length 2 x L 86 mm a E a If pigs are used for cleaning it is essential to take the inside diameters of measuring tube Page 15 and process connection di into account Tri Clamp Sensor Piping di G L HxW 1 4404 316L Bee DN mm Tri Clamp mm mm mm mm 2 8 Pipe 12 7 x 1 65 9 4 25 0 28 5 60 x 42 ODT 1 2 15 Pipe 19 1 x 1 65 15 8 25 0 28 5 60 x 42 co u 6 a Bes ODT 3 4 se T 1 25 ANSI Pipe 24 5 x 1 65 22 1 50 4 28 5 70 x 52 FO6 xxHxxxxx 06 09 07 xx 020 ODT 1 Fitting length 2 x L 86 mm If pigs are used for cleaning it is essential to take the inside diameters o pro
10. the flowing medium corresponds to the moving conductor The induced voltage is proportional to the flow velocity and is detected by two measuring electrodes and transmitted to the amplifier Flow volume is computed on the basis of the pipe s diameter The constant magnetic field is generated by a switched direct current of alternating polarity FO6 xxxxxxxx 15 xx xx xx 001 UVe B L v Q A v Ue induced voltage B magnetic induction magnetic field L electrode gap v flow velocity Q volume flow A pipe cross section l current Measuring system The measuring system consists of a transmitter and a sensor Compact version transmitter and sensor form a single mechanical unit e Transmitter Promag 23 Touch Control without opening the housing four line display e Sensor Promag H DN 2 100 Input Measured variable Flow rate proportional to induced voltage Measuring range Typically v 0 01 10 m s with the specified measuring accuracy Operable flow range Over 1000 1 Endress Hauser PROline Promag 23 H Output Output signal e Current output Applied direct current 4 20 mA input from DC voltage source Terminal voltage 12 30 V DC 13 9 30 V DC Ex i e Frequency output Open collector passive galvanically isolated 30 VDC 100 mA 250 mA 20 ms Optional configurable as Frequency output Full scale frequency 500 10000 Hz
11. 120mm D 7mm Endress Hauser 15 PROline Promag 23 H Promag H DN 40 100 FO6 23Hxxxxx 06 00 xx xx 000 DN PN DI L A B Cc K M DIN ANSI DIN mm inch bar mm mm mm mm mm mm mm 40 1 1 2 16 35 3 140 340 276 64 128 M 6x4 50 2 16 48 1 140 365 288 77 153 M 8x4 65 2 1 2 16 59 9 140 365 288 77 153 M 8x4 80 3 16 72 6 200 415 313 102 203 M 12x4 100 4 16 97 5 200 415 313 102 203 M 12x4 Fitting length depends on process connections gt Page 24 ff Endress Hauser PROline Promag 23 H Front view of Promag H DN 2 25 without process connection DN 2 15 DN 25 A A A B B 4 lt 4 POE tm ite og 2 R 9 o Q Y O TE i i 8 6 6 8 5 2 8 5 DN Cc D DIN D ANSI mm mm mm 2 8 9 15 16 25 26 22 6 Process connections with O ring seals DN 2 25 Weld nipple for pipe ISO 2463 Sensor Piping di G L H x W 1 4404 316L B H Brera DN mm Pipe mm mm mm mm 2 28 13 5 x 1 6 10 3 13 5 20 3 60 x 42 15 21 3x 1 6 18 1 21 3 20 3 60 x 42 A m S 25 DIN 33 7 x2 29 7 33 7 20 3 70 x 52 ol S fa g g Fitting length 2 x L 86 mm a g Weld nipple for pipe IPS
12. 6 121 61 5 100 0 O 5 80 72 9 91 1 147 61 5 121 0 100 97 6 118 1 168 61 5 141 5 If pigs are used for cleaning it is essential to take the inside diameters of measuring tube Page 15 and process connection di into account 26 Endress Hauser PROline Promag 23 H Weight Weight data of Promag H in kg Nominal diameter Compact version mm inch DIN 2 1 12 5 2 4 5 32 5 2 8 5 16 5 3 15 1 2 5 4 25 15 5 5 40 1 1 2 6 5 50 2 9 0 65 2 1 2 9 5 80 3 19 0 100 4 18 5 Transmitter Promag compact version 3 4 kg Weight data valid for standard pressure ratings and without packaging material Materials Transmitter housing Compact housing powder coated die cast aluminium Sensor housing Stainless steel 1 4301 Wall mounting holder panel Stainless steel 1 4301 Measuring tube Stainless steel 1 4301 or 1 4306 304L Flanges e All connections 1 4404 316L e Flanges DIN ANSI JIS made of PVDF e Adhesive fitting made of PVC Ground rings e Standard 1 4435 316L e Option tantalum platinum base material titanium Grade 2 platinum coating min 12 um Alloy C 22 Electrodes e Standard 1 4435 e Option Alloy C 22 tantalum platinum rhodium 80 20 up to DN 25 only Seals e DN 2 25 O ring EPDM Viton Kalrez or gasket seal EPDM silicone Viton e DN 40 100 gasket seal EPDM silicone Endress
13. Sensor Piping di G L Hx W 1 4404 316L LS haa ecetetetetatetatetatal DN mm Pipe ODT SMS mm mm mm mm 25 8 13 5 x 2 3 9 0 13 5 20 3 60 x 42 15 21 3 x 2 65 16 0 21 3 20 3 60 x 42 _ m 1 25 ANSI 33 7 x 3 25 27 2 33 7 22 3 70 x 52 Os FO6 xxHxxxxx 06 09 07 xx 012 Fitting length 2 x L 86 mm Endress Hauser 17 PROline Promag 23 H Flange PN 40 DIN 2635 Sensor Piping di G L LK M Hx W 1 4404 316L Laid bate D Ynsaatatetatata DN mm Flange mm mm mm mm mm mm 2 8 DN 15 17 3 95 56 2 65 14 60 x 42 15 DN 15 17 3 95 56 2 65 14 60 x 42 25 DIN DN 25 28 5 115 56 2 85 14 70 x 52 Fitting length 2 x L 86 mm Fitting length to DVGW 200 mm FO6 xxHxxxxx 06 09 07 xx 014 L Flange CI 150 ANSI 16 5 Sensor Piping di G L LK M Hx W 1 4404 316L Laid betel lalate DN mm Flange inch mm mm mm mm mm mm 2 2 8 1 2 15 7 89 66 0 60 5 15 7 60 x 42 Zy H 15 1 2 16 0 89 66 0 60 5 15 7 60 x 42 7 pa 5 1 25 ANSI 1 26 7 108 71 8 79 2 15 7 70 x 52 5 I g Fitting length 2 x L 86 mm x Flange 20K JIS B2238 Sensor Piping di G L LK M Hx W 1 4404 316L Lead Sate aaa DN mm Flange mm mm mm mm mm mm zj 2 8 ND 10 10 90 67 65 15 60 x 42 ls 15 ND 15 16 95
14. 60 40 20 0 20 40 60 80 100 120 140 160 180 C FO6 xxHXxxxxx 05 xx xx xx 003 Flange material 1 4404 316L PVDF to JIS B2238 bar 30 20 rr E 40 10K PVDF o 60 40 20 0 20 40 60 80 100 120 140 160 180 C Fitted electrodes Measuring electrodes and EPD electrodes e Standard with 1 4435 Alloy C 22 tantalum platinum rhodium e DN 2 8 without EPD electrode Process connection e With O ring welding nipples ISO 2463 IPS flanges DIN ANSI JIS PVDF flanges DIN ANSI JIS external pipe thread internal pipe thread hose connection PVC adhesive fittings e With gasket seal weld nipples DIN 11850 ODT clamps ODT ISO 2852 DIN 32676 threaded fasteners DIN 11851 DIN 11864 1 ISO 2853 SMS 1145 flanges DIN 11864 2 Surface roughness e PFA liner lt 0 3 um e Electrodes 1 4485 Alloy C 22 lt 0 4 um Tantalum platinum rhodium lt 0 8 um e Process connection Promag H lt 0 8 um all data relate to parts in contact with medium Endress Hauser 29 PROline Promag 23 H Human interface Display elements e Liquid crystal display four lines with 16 characters per line e Custom configurations for presenting different measured values and status variables e 2 totalizers Operating elements Unified PROline operation concept Loc
15. 60 x 42 co 15 16 12 6 16 49 60 x 42 g ee i 2 S 15 19 16 0 19 49 70 x 52 5 I a g Fitting length 2 x L 86 mm L a Adhesive fitting Sensor Piping di G L Hx W PVC BRE ROAA a DN mm Adhesive connection inch mm mm mm mm 2 8 1 2 21 5 27 3 28 0 60 x 42 15 20 x2 20 2 27 0 38 5 60 x 42 a _ Fitting length 2x L 86 mm T The requisite ground rings can be ordered as accessories Order code DK5HR Process connections with aseptic gasket seals DN 2 25 Weld nipple for pipe DIN 11850 Sensor Piping di G L Hx W 1 4404 316L ia a O DN mm Pipe mm mm mm mm 2 8 14x2 10 14 23 3 60 x 42 15 20x2 16 20 23 3 60 x 42 Es 25 DIN 30 x 2 26 30 23 3 70 x 52 F06 xxHxxxxx 06 09 07 xx 011 Fitting length 2 x L 86 mm If pigs are used for cleaning it is essential to take the inside diameters of measuring tube Page 15 and process connection di into account Weld nipple for ODT SMS Sensor Piping di G L H x W 1 4404 316L Fated VAV ieii DN mm Pipe mm mm mm mm 2 1 8 12 7 x 1 65 9 4 12 7 16 1 60 x 42 15 19 1 x 1 65 15 8 19 1 16 1 60 x 42 o A 1 25 ANSI 24 5 x 1 65 22 1 25 4 16 1 70x52 o F06 xxHxxxxx 06 09 07 xx 013 Fitting length 2 x L 86 mm If pigs are used for cleaning it is essential to take the inside diameters of measuring tube Page 15 and process connection di into account 20
16. 67 70 15 60 x 42 x jas m 5 SE 25 DIN ND 25 26 125 67 95 19 70 x 52 5 T Fitting length 2 x L 86 mm FO6 xxHxxxxx 06 09 07 xx 016 eee See Flange PN 16 DIN 2501 Sensor Piping di G L M LK Hx W PVDF B H Geter DN mm Flange mm mm mm mm mm mm I S 2 8 DN 15 15 7 95 57 14 65 60 x 42 15 DN 15 15 7 95 57 14 65 60 x 42 al ffs 25 DIN DN 25 27 3 115 57 14 85 70 x 52 o 5 x E Fitting length 2 x L 86 mm Fitting length to DVGW 200 mm The requisite ground rings can be ordered as accessories Order code DK5HR FO6 xxHxxxxx 06 09 07 xx 029 18 Endress Hauser PROline Promag 23 H FO6 xxHxxxxx 06 09 07 xx 027 Flange CI 150 ANSI 16 5 Sensor Piping di G L M LK HxW PVDF B H Hrena DN mm Flange inch mm mm mm mm mm mm I 2 8 1 2 15 7 95 57 16 60 60 x 42 Zj 15 1 2 15 7 95 57 16 60 60 x 42 wok iF A g 1 25 ANSI ah 27 3 115 57 16 79 70 x 52 o 5 Ux kd 3 Fitting length 2 x L 86 mm 8 The requisit
17. Hauser PROline Promag 23 H Material load curves Weld nipple in 1 4404 316L with O ring to ISO 2463 IPS ISO 228 DIN 2999 bar 50 40 35 30 25 0 60 40 20 0 20 40 60 80 100 120 140 160 180 C FO6 xxHxxxxx 05 xx xx xx 000 Weld nipple in 1 4404 316L with gasket seal to DIN 11850 ODT Clamp ODT ISO 2852 DIN 32676 coupling DIN 11851 DIN 11864 1 ISO 2853 SMS 1145 flange DIN 11864 2 bar 20 15 10 60 40 20 0 20 40 60 80 100 120 140 160 180 C F06 xxHxxxxx 05 xx xx xx 00 1 Flange material 1 4404 316L PVDF Adhesive fitting PVC U to DIN 2635 and 2501 bar 40 35 30 25 20 15 10 PN16 PVDF PN16 PVC U 60 40 20 0 20 40 60 80 100 120 140 160 180 C F06 xxHxxxxx 05 xx xx xx 002 28 Endress Hauser PROline Promag 23 H Flange material 316L PVDF to ANSI B16 5 bar 30 25 20 Cl150 316L 5 Cl150 PVDF
18. Technical Information TI 051D 06 en 50097004 SZ K i 25 _ wa HEET m T FR Electromagnetic Flow Measuring System Two wire loop powered PROline promag 23 H Flow measurement in hygiene food and beverage or process applications Features and benefits Nominal diameters DN 2 100 PFA lining for cleaning temperatures up to 150 C 180 C in preparation Guaranteed product quality because suitable for CIP SIP cleaning and piggable 3A approval and EHEDG tested Measuring accuracy 0 5 Robust field housing IP 67 Touch control operation without ope ning the housing also in Ex environment Communication HART is standard Intrinsically safe Ex ia for installation in zone 1 ATEX FM CSA etc Endress Hauser The Power of Know How Transmitter supply Non Ex environment 12 30 V DC Ex environment 13 9 30 V DC Connecting to all mainstream transmit ter power supplies and input cards of process control systems Reduced installation and operation costs Application All fluids with a minimum conductivity of gt 50 uS cm can be measured beverages e g fruit juice beer wine salt solutions acids alkalis etc liza PROline Promag 23 H Function and system design Measuring principle Faraday s law of induction states that a voltage is induced in a conductor moving in a magnetic field In electromagnetic measuring
19. al operation with three optical keys E Remote operation Operation via HART Certificates and approvals Ex approval Information on the currently available Ex rated versions ATEX FM CSA etc is available on request from your E H sales outlet All information relevant to explosion protection is available in separate Ex documents that you can order as necessary Sanitary compatibility 3A authorization and EHEDG tested Seals in conformity with FDA except Kalrez seals CE mark The measuring system is in conformity with the statutory requirements of the EC Directives Endress Hauser confirms successful testing of the device by affixing to it the CE mark Other standards and guidelines EN 60529 Degrees of protection by housing IP code EN 61010 Protection Measures for Electrical Equipment for Measurement Control Regulation and Laboratory Procedures EN 61326 IEC 1326 Electromagnetic compatibility EMC requirements NAMUR NE 21 Association for Standards for Control and Regulation in the Chemical Industry Ordering information The E H service organisation can provide detailed ordering information and information on the order codes on request Accessories Various accessories which can be ordered separately from Endress Hauser are available for the transmitter and the sensor The E H service organisation can provide detailed information on request 30 End
20. cess connection di into account measuring tube Page 15 and Coupling DIN 11851 Sensor Piping di G L Hx W 1 4404 316L i e iiai DN mm Screw union mm mm mm mm 2 8 Pipe 12 x 1 10 Rd 28 x 1 8 44 60 x 42 DN 10 15 Pipe 18 x 1 or 1 5 16 Rd 34 x 1 8 44 60 x 42 DN 15 25 DIN Pipe 28 x 1 or 1 5 26 Rd 52 x 1 6 52 70x52 F06 xxHxxxxx 06 09 07 xx 017 DN 25 Fitting length 2 x L 86 mm If pigs are used for cleaning it is essential to take the inside diameters of measuring tube Page 15 and process connection di into account Endress Hauser 21 PROline Promag 23 H Fitting length 2 x L 86 mm Coupling DIN 11864 1 Sensor Piping di G L HxW 1 4404 316L Laid bee leceletetetetetetete DN mm Screw union mm mm mm mm 2 8 Pipe 13x 1 5 10 Rd 28 x 1 8 42 60 x 42 DIN 11850 DN 10 7 15 Pipe 19x 1 5 16 Rd 34 x 1 8 42 60 x 42 g DIN 11850 DN 15 i 25 DIN Pipe 29 x 1 5 26 Rd 52 x 1 6 49 70x 52 DIN 11850 DN 25 If pigs are used for cleaning it is essent process connection di into account ial to take the inside diameters of measuring tube Page 15 and Flange DIN 11864 2 Form A Sensor Piping di G L M HxW 1 4404 316L Lead ae aaa ata DN mm Flange mm mm mm mm mm mm 2
21. diameters of measuring tube Page 15 and E process connection di into account 24 Endress Hauser PROline Promag 23 H Fitting length 2 x L 140 mm DN 40 65 200 mm DN 80 100 Clamp ISO 2852 DN di G D L LK 1 4404 316L mm mm mm mm mm mm SER E ARR Aas s 40 35 6 50 5 92 68 5 71 0 50 48 6 64 0 105 68 5 83 5 3 65 60 3 LD 121 68 5 100 0 5 80 72 9 91 0 147 68 5 121 0 g 100 97 6 119 0 168 68 5 141 5 z If pigs are used for cleaning it is essential to take the inside diameters of measuring tube Page 15 and process connection di into account Fitting length 2 x L 140 mm DN 40 65 200 mm DN 80 100 Clamp DIN 32676 DN di G D L LK 1 4404 316L mm mm mm mm mm mm DA Here 40 38 50 5 92 61 5 71 0 50 50 64 0 105 61 5 83 5 8 65 66 91 0 121 68 0 100 0 8 80 81 106 0 147 68 0 121 0 100 100 119 0 168 68 0 141 5 If pigs are used for cleaning it is essential to take the inside diameters of measuring tube Page 15 and process connection di into account Fitting length 2 x L 140 mm DN 40 65 200 mm DN 80 100 Tri Clamp DN di G D L LK 1 4404 316L i aie tals beeen mm inch mm mm mm mm mm 40 1 1 2 34 8 50 4 92 68 6 71 0 50 2 47 5 63 9 105 68 6 83 5 x a 65
22. e ground rings can be ordered as accessories Order code DKSHR y i 3 Flange 10K JIS B2238 Sensor Piping di G L M LK HxW PVDF BH Jerr DN mm Flange mm mm mm mm mm mm I 2 8 ND 15 15 7 95 57 15 70 60 x 42 Zj 15 ND 15 15 7 95 57 15 70 60 x 42 _ _ _ a g 25 DIN ND 25 27 3 125 57 19 90 70 x 52 o a x Fitting length 2 x L 86 mm The requisite ground rings can be ordered as accessories Order code DK5HR 3 z i External pipe thread ISO 228 Sensor Piping di G L S HxW DIN 2999 1 4404 316L B H Krr rreran DN mm Internal thread mm inch mm mm mm inch 2 8 R 3 8 10 3 8 40 10 1 60 x 42 S a 15 R 1 2 16 1 2 40 13 2 60 x 42 o a o 1 25 ANSI R 1 25 1 40 16 5 70 x 52 O 5 F 2 Fitting length 2 x L 86 mm z Internal pipe thread ISO 228 Sensor Piping di G D L S HxW DIN 2999 1 4404 316L B H r a DN mm External thread mm inch mm mm mm mm inch S 2 8 Rp 3 8 8 9 3 8 22 45 13 60 x 42 15 Rp 1 2 16 0 1 2 27 45 14 60 x 42 ol o a 1 25 ANSI Rp 1 27 2 ile 40 49 17 70 x 52 Fitting length 2 x L 86 mm Endress Hauser 19 PROline Promag 23 H Hose connection Sensor Hose LW di LW L Hx W 1 4404 316L 5 H M N P rrr rera DN mm Inside diameter mm mm mm mm mm 22428 13 10 0 13 49
23. ernal pipe thread NPT Sensor Piping di G D L S HxW 1 4404 316L DKH GC DN mm External thread mm inch mm mm mm mm inch 5 2 8 NPT 3 8 8 9 3 8 22 45 13 60 x 42 15 NPT 1 2 16 0 1 2 27 45 14 60 x 42 al Ol j S 8 1 25 ANSI NPT 1 27 2 1 40 49 17 70 x 52 5 5 amp Fitting length 2 x L 86 mm Process connections orderable only as accessories with asceptic gasket seal Tri Clamp Sensor Piping di G L HxW 1 4404 316L DKH HF DN mm Tri Clamp mm mm mm mm 15 Pipe 25 4 x 1 65 22 1 50 4 28 5 60 x 42 ODT 1 xB Jelai FO6 xxHxxxxx 06 09 07 xx 018 Fitting length 2 x L 86 mm If pigs are used for cleaning it is essential to take the inside diameters of measuring tube Page 15 and process connection di into account Ground rings available as accessories PVDF flange PVC adhesive coupling Ground ring Sensor di B Cc D 1 4435 316L Alloy C 22 Titanium Pt Rh coated DN mm mm mm mm mm DK5HR 2 8 9 0 22 0 17 6 33 9 15 16 0 29 0 24 6 33 9 25 DIN 22 6 36 5 31 2 43 9 1 25 ANSI 26 0 39 0 34 6 43 9 F06 xxHxxxxx 06 09 07 xx 030 Endress Hauser 23 PROline Promag 23 H Front view of Promag H DN 40 100 without process connection FO6 5xHxxxxx 06 05 08 xx 001
24. fmax 12 5 Hz or Pulse output Pulse value and pulse polarity adjustable pulse width adjustable 0 01 10 s pulse frequency max 50 Hz or Status output E g for error messages Empty Pipe Detection flow direction recognition limit value configurable e Ex i version Power supply signal circuits and pulse output with intrinsically safe protection rating EEx ia IIC and EEx ia IIB only for connection to certified intrinsically safe circuits with the following maximum values Uj 30 V l 150 mA P 810 mW Effective internal inductance negligible Effective internal capacitance C lt 25 nF Pulse output Maximum values U 30 V l 10 mA P 1W Effective internal inductance negligible Effective internal capacitance negligible Signal on alarm e Current output failure response selectable e Pulse frequency output failure response selectable e Status output non conductive by fault or power supply failure Load see Page 5 Low flow cutoff Switch points for low flow cut off are selectable Galvanic isolation Outputs are galvanically isolated from sensor and from each other Endress Hauser PROline Promag 23 H Power supply Electrical connection measuring unit FO6 2xxxxxxx 04 06 xx xx 001 1 Shielded signal cable the Ex version requires the use of separate cables for transmi
25. l potential Metal process connections Potential matching usually takes place over the metallic process connection in contact with medium process connections which are directly mounted on the measuring transmitter This usually means that additional potential matching measures are unnecessary Hinweis For installation in metal pipes it is advisable to connect the ground terminal of the transmitter housing to the piping Plastic process connections For plastic process connections potential matching must be ensured between sensor and medium using additional ground rings If these ground rings are missing this can influence accu racy or destroy the measuring transmitter through the electrochemical decomposition of electro des FO6 xXHxxxxx 17 xx xx xx 001 1 allen screw process connection 2 O ring seal process connection 3 plastic washer spacer or ground ring 4 O ring seal ground ring 5 sensor When using ground rings note the following points e Depending on the option ordered plastic washers may be installed at the process connections instead of ground rings These plastic washers serve only as spacers and have no potential equalization function In addition they provide a sealing function at the interface between the sensor and process connection For this reason with process connections without ground rings these plastic washers seals must not be removed or m
26. llowing locations e Highest point of a pipeline Risk of air accumulating e Directly upstream from an open pipe outlet in a down pipe FO6 2xxxxxxx 11 00 00 xx 000 Installation of pumps Do not install the sensor on the intake side of a pump This precaution is to avoid low pressure and the consequent risk of damage to the lining of the measuring tube It might be necessary to install pulse dampers in systems incorporating reciprocating diaphragm or peristaltic pumps Information on the measuring system s resistance to vibration and shock can be found on Page 13 FO6 2xxxxxxx 1 1 00 00 xx 001 8 Endress Hauser PROline Promag 23 H Partially filled pipes Partially filled pipes with gradients necessitate a drain type configuration The Empty Pipe Detec tion EPD function offers additional protection by detecting empty or partially filled pipes Achtung Risk of solids accumulating Do not install the sensor at the lowest point in the drain It is advisable to install a cleaning valve FO6 2xxxxxxx 1 1 00 00 xx 002 Down pipes Install a siphon 2 or a vent valve 1 downstream of the sensor in down pipes longer than 5 meters This precaution is to avoid low pressure and the consequent risk of damage to the lining of the measuring tube This measures also prevent the system losing prime which could cause air inclusions gt 5m FO6 2xxxxxxx 1 1 00 00 xx 003
27. n supply voltage of 13 9 V DC Ex i required supply voltage transmitter IM A max signal transmission current failsafe mode current output 22 mA max current Hinweis The minimum load resistance R necessary for a data transfer via HART protocol by way of the current signal cable is 250 Q The minimum external supply voltage Ug therefore has to be 17 5 V DC non Ex R Q 900 800 700 600 500 400 300 200 100 o gt 10 15 17 5 20 25 30 U V F06 77xxxxxx 05 xx xx xx 000 Load at the analog current output non Ex R_ max load resistance with HART min 250 Q Ug external supply voltage non Ex Cable entry e Cable entry M20 x 1 5 8 12 mm e Threads for cable entries Pg 13 5 5 15 mm 1 2 NPT G 1 2 Cable specifications Use shielded cables Supply voltage Non Ex area 12 30 V DC with HART 17 5 30 V DC Ex area Ex i 13 9 30 V DC with HART 19 4 30 V DC Power supply failure e T DAT saves measuring system data if power supply fails e S DAT exchangeable data storage chip which stores the data of the sensor nominal diameter serial number calibration factor zero point etc Endress Hauser 5 Potential equalisation PROline Promag 23 H Perfect measurement is only ensured when the medium and the sensor have the same electrica
28. ress Hauser PROline Promag 23 H Supplementary documentation Q System Information Promag SI 028D 06 en Q Technical Information Promag 23 P TI 049D 06 en Q Operating Instructions Promag 23 BA 045D 06 en and BA 050D 06 en Q Supplementary documentation on Ex ratings AT EX FM CSA etc Endress Hauser 31 Subject to modification Endress Hauser Gmbh Co Instruments International P O Box 2222 D 79574 Weil am Rhein Germany Tel 07621 975 02 Tx 773926 Fax 07621 975 345 http Awww endress com info ii endress com AN Endress Hauser The Power of Know How la Sif TI 051D 06 en 08 01 50097004 FM SGML 6 0
29. ting Creepage inch mm v 0 3 or 10 m s v 2 5 m s 2 pulse s v 0 04 m s 1 12 2 0 015 0 5 gal min 0 1 gal min 0 001 ga 0 002 gal min 5 32 4 0 07 2 gal min 0 5 gal min 0 005 ga 0 008 gal min 5 16 8 0 25 8 gal min 2 gal min 0 02 ga 0 025 gal min 1 2 15 1 0 27 gal min 6 gal min 0 05 gal 0 10 gal min 1 25 2 5 80 gal min 18 gal min 0 20 ga 0 25 gal min 1 1 2 40 7 190 gal min 50 gal min 0 50 ga 0 75 gal min 2 50 10 300 gal min 75 gal min 0 50 ga 1 25 gal min 2 1 2 65 16 500 gal min 130 gal min 1 gal 2 0 gal min 3 80 24 800 gal min 200 gal min 2 gal 2 5 gal min 4 100 40 1250 gal min 300 gal min 2 gal 4 0 gal min 14 Endress Hauser PROline Promag 23 H Mechanical construction Design dimensions Promag H DN 2 25 gt DI K DN PN DI L K M DIN ANSI DIN mm inch bar mm mm mm mm 2 16 40 2 25 86 43 M 6x4 4 16 40 4 5 86 43 M 6x4 8 16 40 9 0 86 43 M 6x4 15 z 16 40 16 0 86 43 M 6x4 1 16 40 22 6 86 53 M 6x4 25 16 40 26 0 86 53 M 6x4 Fitting length depends on process connections Page 17 ff The permissible nominal pressure depends on the process connection and seal 40 bar flange welding adapter with O ring seal 16 bar all other process connections Wall mounting kit for DN 2 25 A 125 mm B 88mm C
30. tter supply and frequency output Terminal No 1 2 transmitter supply current output Terminal No 3 4 frequency output 2 Grounding terminal for signal cable shield 3 Service plug Outputs Terminal No Order variant 1 2 3 4 DPA A RRRASARE RENN Current output HART DOTEE A Current output HART Frequency output Mandatory A common connecting cable carries supply voltage and measuring output signal Current output passive galvanically isolated 12 30 V DC Ex i 13 9 30 V DC 4 20 mA Optional A binary output can be used as an option It can be configured as a standard impulse output a frequency output or a switching output Frequency output passive galvanically isolated max 30 V DC 100 mA Open Collector e Frequency operating mode limit frequency 500 10000 HZ fmax 12500 Hz e Pulse operating mode pulse frequency max 50 Hz e Status operating mode yes We recommend shielded signal cables as a general principle Endress Hauser PROline Promag 23 H Load The load has to be calculated as follows U V UylV _ Us V 12 V Non E R Q on Ex area R Q i A 0 022 A a U IVI UIV _ U V 13 9 V Ex area Ex i R Q MIA 0 022TA R Q max load resistance load cable resistance Us V external supply voltage of 12 30 V DC outgoing supply voltage transmitter supply unit Uy V min supply voltage of 12 V DC mi
31. ust always be installed e Ground rings can be ordered separately from E H as an accessory Also make sure that the ground rings are compatible with the electrode material Otherwise the danger exists that the electrodes could be destroyed by electrochemical corrosion You can find material data on Page 27 e Ground rings incl seals are mounted inside the process connection This has no influence on the installation length You can find the dimensions of ground rings on Page 23 Endress Hauser PROline Promag 23 H Measuring accuracy Reference operating conditions To DIN 19200 and VDI VDE 2641 e Medium temperature 28 C 2K e Ambient temperature 22 C 2K e Warm up period 30 minutes Installation e Inlet run gt 10 x DN e Outlet run gt 5 x DN e Sensor and transmitter grounded e Sensor centered relative to the pipe Max measured error Signal output 0 5 o r 4 mm s o r of reading Supply voltage fluctuations have no effect within the specified range 2 5 2 0 tS 1 0 0 5 0 a A a a a v m s F06 2xxxxxxx 05 xx xx xx 001 Max measured error in of reading Repeatability max 0 25 o r 2 mm s o r of reading Endress Hauser PROline Promag 23 H Installation conditions Installation instructions Mounting location Correct measuring is possible only if the pipe is full Avoid the fo
Download Pdf Manuals
Related Search
Related Contents
50097-193 Rev B Samsung HT-D720 Manual de utilizare MT150 Series Mechanical Test Stand User Manual MANUEL D`UTILISATION MANUAL DEL PROPIETARIO DREAM: DEKEMA Remote Access Manager Addonics Technologies ADCFASTMS User's Manual Quick Start Guide Brochure MJC 2015-2016 - La communauté de communes d`Altkirch Sikadur Arena Il transistor Copyright © All rights reserved.