da Vinci® Mitral Valve Repair

image

Contents

1. Figure 12 Port set up for Atrial Retractor as a second left hand instrument 20 da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 Atrial Retractor as second right hand instrument da Vinci S only Figure 13 e Triangulation of camera port left instrument and right instrument ports as usual for dVMVR described above e Atrial retractor should be placed one ICS below camera port and should always be closer to the camera port than to the right instrument arm in order to avoid instrument arm collisions keep 3 4 finger breadth between atrial retractor port and right instrument port e Depending on the patient s chest anatomy the lateral location of the retractor port needs to be adjusted in a flat chest port probably needs to be more lateral in barrel chest port needs to be more medial sometimes even more medial than camera port da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 21 tae ios Ss Se Figure 13 Port set up for Atrial Retractor as a second right hand instrument 22 da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 e Cannulation o Heparinize patient o Vein 5 0 Prolene purse string Seldinger technique TEE guidance bicaval view wire first then Cardiovations QuickDraw Venous Catheter 22Fr or 25Fr o Artery Cardiovations EndoReturn Arterial Cannula 21Fr or 23Fr o If using the Cardiovations EndoBalloon for cross clamping TEE guidance
2. aortic valve ascending view wire first then Cardiovations EndoClamp Aortic Cannula 100cm Docking CAUTION Once the Patient Cart is docked to the patient the OR table should not be moved e Standard da Vinci System 3 arm procedure with manual mechanical atrial retractor system used o Before roll up check for the sweet spot on the camera arm 90 angle between the camera arm joints to help ensure correct Patient Cart set up during docking The camera arm elbow should be pointed towards the patient s feet o The da Vinci Patient Cart is rolled in at a 0 30 angle to accommodate for the natural position of the mitral valve in the chest depending on patient s anatomy The center column should be lined up with the camera port in the right chest Figure 14 0 30 Figure 14 Patient Cart rolled in at a slight angle to accommodate for the natural position of the mitral valve in the patient s chest o Once the camera arm is right over the camera port cannula the Patient Cart can be locked brakes down o Then the camera arm is docked to the camera port and the endoscope inserted to assist during set up of the other arm At this point in the setup the surgeon can take responsibility for clutching and positioning the camera arm in order to free the assistant for additional setup duties o With a finger through the working port and the other hand releasing the set up joint button insert the right da Vinci
3. and CO2 insufflation is lost o The surgeon and assistant then exchange places and instrument arm is docked to the cannula in a similar fashion Dock this arm but do not engage the Atrial Retractor yet o Next the left instrument arm is docked to the cannula the black trocar removed and the appropriate instrument inserted and advanced with the clutch engaged through the port guided by the surgeon s finger through the working port and steered into the view of the endoscope Once again anticipate movement of the mediastinum toward the da Vinci instruments each time the working port is left open as the CO rapidly escapes o Right da Vinci instrument Large Needle Driver left da Vinci instrument Resano Forceps Clutch the da Vinci instruments into the center of the video image Place a 20mm Ethicon Flexipath port into the working port OO The arm set up is as follows AR left of camera e Right instrument e Atrial Retractor e Left instrument e da Vinci S System 4 arm procedure with EndoWrist Atrial Retractor used as a second right hand instrument o Before roll up check for the sweet spot on the camera arm 90 angle between the camera arm joints to help ensure correct Patient Cart set up during docking The camera elbow should be pointed towards the patient s head o The da Vinci Sis rolled in at a 0 30 angle to accommodate for the natural position of the mitral valve in the chest depending
4. e Second left hand standard da Vinci and da Vinci S e Second right hand da Vinci S only Which configuration is used on a da Vinci S System 2 left or 2 right depends on the surgeon s preference o Atrial Retractor as second left hand instrument standard da Vinci and da Vinci S Figure 12 Triangulation of camera port left instrument and right instrument port as usual for dVMVR described above Atrial Retractor should be placed one ICS above camera port and should always be closer to the camera port than towards the left instrument arm to avoid instrument arm collisions keep 3 4 finger breadth between Atrial Retractor port and left instrument port Depending on the patient s chest anatomy the lateral location of the retractor port needs to be adjusted if a flat chest port probably needs to be more lateral in barrel chest port needs to be more medial sometimes even more medial than camera port To optimize cosmesis and avoid penetration of breast tissue in women the incision should be kept away from the right breast by placing the skin incision for the retractor port as lateral as possible Then when inserting the cannula into the chest the skin is distracted medially da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 19 Atrial Retractor 3 ICS BN Camera attics Working port 4 ICS Right 5 6 ICS AON ja TE ka E ANS oF S aN lt is es 4
5. o 20mm Ethicon Flexipath Trocar used as port in working incision o ATS Medical Thoracic Port System www atsmedical com for thoracic access when using manual mechanical atrial retractor system o If using mechanical atrial retractor Cardiovations or Estech a 14Fr or 16 Fr Cook Introducer Set should be used for passing atrial retractor rod through thoracic wall o Becton Dickinson 14 GA Angiocath x 5 25 inch 5x for passing stay sutures through thoracic wall o Luer Lock caps 5x to cover ends of BD s Angiocath s and keeping CO2 insufflation initially Perfusion o Cell Saver o DLP Intracardiac pericardial Sump 20 Fr Medtronic PN 12112 20 Fr 12 length or 12012 20 Fr 15 length EndoBalloon Technique o 15Fr 17Fr 19Fr Arterial Biomedicus Cannulas for SVC Cardiovations EndoReturn Arterial Cannula 21Fr or 23Fr Cardiovations EndoClamp Aortic Cannula 100cm Cardiovations QuickDraw Venous Catheter 22Fr or 25Fr Cardiovations EndoPlege Sinus Catheter optional Cardiovations EndoVent Pulmonary Catheter optional Alternatively the Estech RAP system can be used www estechlics com O OQ 0 0 O Cross clamp Technique o Scanlan CardioVasive Chitwood Transthoracic cross clamp PN 9909 912 13 Cardiovations EndoReturn Arterial Cannula 21 Fr or 23 Fr Cardiovations QuickDraw Venous Catheter 22 25 Fr Cardiovations EndoPlege Sinus Catheter Cardiovations EndoVent Pulmonary Catheter 16 gauge 5 25 g
6. on patient s anatomy The central column should be lined up 7 10cm cephalad of the camera port in the right chest to allow more space on the left for the da Vinci arms Figure 18 da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 27 0 30 7 10cm cephalad Camera elbow towards patient s head Figure 18 Patient Cart rolled in at a slight angle and offset towards the patient s feet to provide more space for the 4 instrument arm on the left side o Once the camera arm is right over the camera port cannula the Patient Cart can be locked brakes down o Next the camera arm is docked to the camera port and the endoscope inserted to assist during set up of the other arm At this point in the setup the surgeon can take responsibility for clutching and positioning the camera arm in order to free the assistant for additional setup duties o With a finger through the working port and the other hand releasing the set up joint button dock the left da Vinci arm to the instrument cannula in the left arm port o The black trocar is removed and the appropriate instrument is inserted and advanced with the clutch engaged through the port guided by the surgeon s finger through the working port and steered into the view of the endoscope o The instrument tip is placed slightly back from the pericardium to compensate for mediastinal shift back towards the surgeon as the finger is removed from the working
7. port and CO insufflation is lost o The surgeon and assistant then exchange places and the instrument arm is docked to the cannula in a similar fashion Dock this arm but do not engage the Atrial Retractor instrument yet o Next the right instrument arm is docked to the cannula the black trocar removed and the appropriate instrument inserted and advanced with the clutch engaged through the port guided by the surgeon s finger through the working port and steered into the view of the endoscope Again anticipate movement of the mediastinum toward the da Vinci instruments each time the working port is left open as the CO rapidly escapes o Right da Vinci instrument Large Needle Driver left da Vinci instrument Resano Forceps Clutch the da Vinci instruments into the center of the video image Place a 20mm Ethicon Flexipath port into the working port O The arm set up is as follows AR right of camera e Atrial Retractor e Left instrument e Right instrument 28 da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 o In order to allow rapid da Vinci instrument changes and facilitate the patient side assistance a sterile draped Mayo stand with the requisite laparoscopic instrumentation can be positioned adjacent to and in front of the right chest ports Figure 19 assistance da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 29 4 PROCEDURE STEPS a P
8. post MediFlex Storz or similar ATS Medical suture hook www atsmedical com or alternatively blunt Nerve Hook Above listed instrumentation is also available in similar designs from other manufacturers e g Geister or Estech www estechlics com e Recommended Sutures O O O Diaphragmatic stay suture Gore Tex CV 2 TH 26 TU02B 2x alternatively Gore Tex CV 4 TH 18 4U02B Pericardial stay suture Gore Tex CV 4 TH 18 4U02B 46 long 3x alternatively 2 0 Ticron T16 3x Right atrium retraction stitch 2 0 Ticron T16 pledgeted 1x da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 Atrium closure Gore Tex CV 4 TT 17 2x alternatively Gore Tex CV 4 TH 18 4NO2B 2x Closure of aortic cardioplegia needle puncture if antegrade cardioplegia delivered directly in aortic root Gore Tex CV 4 TT 17 pledgeted 1x Leaflet repair Medtronic U Clip V50 or V60 alternatively 4 0 Prolene RB1 or 4 0 Cardionyl 4 0 Prolene BB for sliding leaflet repair Annuloplasty Ethibond SH 2 X582 single packs with soft pledgets 2 0 Ticron Y 5 without pledget multipack 2852 56 for compression stitch Medtronic U Clip V100 for annuloplasty band Other surgical supplies o Mitral Ring Sizers Edwards Cosgrove Medtronic Duran Medtronic now offers sizers that can be handled with robotic instrumentation or ATS Medical holes drilled into them and string put through
9. removal o To facilitate de airing the DLP sump is placed across the Mitral Valve o Tie 4 0 Gore Tex in each corner Large Needle Driver right Resano left run to the middle around sump sucker and maintain tension Follow instructions under d Mitral Valve Repair Section Tips for Suturing from above Figure 26 a d amp Figure 27 a f o If ventricular pacing wires are desired then this is the best time to place them Access to the inferior RV wall is good o Venous drainage is then retarded forcing blood through the lungs into the left heart where venting is accomplished by the DLP sump in the left ventricle and the EndoBalloon catheter in the ascending aorta Cardiovations EndoBalloon system or de Air via Atriotomy amp then Aortic catheter Transthoracic Cross clamp Chitwood Clamp o Following resumption of contractility the venous line is again retarded and blood advances to the left ventricle where it is vented without LV ejection o When no air is visible on TEE the retard is lessened and the DLP sump pulled o The patient side surgeon ties left atriotomy suture knots under endoscopic surveillance NOTE All Gore Tex sutures are tied with at least 12 knots da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 45 46 Conclusion of Cardio Pulmonary Bypass Cardiovations EndoBalloon system O O O Venous return is retarded to advance blood through pulmonary veins left atrium and le
10. CKING e Patient Selection o Patient types for early cases o Mitral valve exposure technique for early cases o Contraindications e OR Configuration o Overhead View e Patient Preparation o Anesthesia set up o Patient positioning o OR table considerations o Cannulation preparation e Port Placement o dVMVR Lateral Approach port placement o Camera port o Working port and instrument arm port placement o Port placement for atrial retraction o EndoWrist Atrial Retractor port set up e Cannulation e Docking o Standard da Vinci System o da Vinci S System 4 PROCEDURE STEPS Pericardiotomy Initiation of Cardio Pulmonary Bypass Left Atriotomy and Mitral Valve Exposure Mitral Valve Repair De airing and Atriotomy Closure Conclusion of Cardio Pulmonary Bypass Hemostatic Control and Closure of Chest Incisions Decannulation sa gt opa0co 5 POST OPERATIVE CARE e Post operative pain management e Internal hospital follow up and discharge e External hospital follow up 2 da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 1 INTRODUCTION da Vinci Mitral Valve Repair dVMVR using a lateral approach represents the most effective and least invasive treatment for mitral valve prolapse available today The enhanced visualization of the valve the precision and dexterity afforded by the da Vinci or da Vinci Sinstrumentation and the active participation of the assistant surgeon all increase the chances of providing repa
11. Create a working port in the same ICS as the camera port 3 4 cm lateral to it If the working port is too close to the camera port access for the assistant can be inhibited If the working port is placed more lateral then access for the assistant will be good but there may be little room to place Angiocaths for traction sutures This is really an issue only in a patient with a small chest or if the patient has not been adequately prepped and draped laterally and posteriorly In general more lateral is better because the working port trajectory will be oriented straight at the heart rather than in a posterior direction The working port for mitral valve surgery is made 3 5 cm in length if a manual mechanical retractor system is used Cardiovations or Estech This length coincides with the size of the atrial retractor blade since it has to be passed through this incision If the EndoWrist Atrial Retractor is used the working port incision can be limited to 2cm just large enough to allow a mitral valve sizer Edwards Medtronic ATS Medical etc to be introduced through the incision When making the working incision feel with a finger to be sure you are in the same ICS as the camera port There are two reasons to be in the same ICS First it provides optimal hand eye alignment for the patient side assistant facilitating surgical assistance Second if at some later point conversion to a thoracotomy approach may be necessary the came
12. a back up set space task of retractor blades should be readily available If the desired blade falls into the right pleural consider postponing retrieval until after closure rather than using cross clamp time for this da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 35 o With the retractor blade in the left atrium the assistant advances the retractor rod down to the blade and begins spinning it clockwise o The da Vinci instruments position the blade to engage the spinning retractor rod under 3D vision Figure 24 o Screw in until tight Figure 24 Retractor handle screwed in under 3D vision o As the retractor rod is pulled up the left atrium will open to provide mitral valve exposure o Only retract as much as necessary to display the valve with minimal distortion o The DLP sump sucker is then positioned in the left superior pulmonary vein e Dynamic Atrial Retraction assistance for mitral valve exposure with the EndoWrist Atrial Retractor o Engage the EndoWrist Atrial Retractor instrument on the corresponding instrument arm o Patient side assistant clutches the instrument arm with the Atrial Retractor and aims in the same direction same trajectory as the left and right da Vinci instruments o Under endoscopic vision clutch the Atrial Retractor instrument into view Aim with the tip of the retractor instrument blades towards the atriotomy o Watch the patient side monitor for any me
13. are placed in a clockwise fashion starting next to the first clip on the right trigone When the clip is fully seated and balanced the remaining release on the previous clip is deployed Next deploy the proximal release on the current clip Repeat this STEP until the final clip is placed In conclusion the furthest clip away from the initial starting point is always left unreleased until the next clip is fully placed Clip placement Technique B more clips 9 might be needed g depending on valve size a Cd da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 43 o Should a full annuloplasty ring be desired or a valve replacement become necessary sutures are placed first posteriorly and then anteriorly Figure 33 o Exteriorly the sutures are placed in suture guides Then placed through the ring then back into the suture guides o After all sutures are placed the ring is passed to the da Vinci instruments and lowered to the annulus Figure 33 Suture order for full ring or valve replacement o Sutures are then tied either intra corporeally by the console surgeon or extra corporeally with support from the surgeon s assistant under endoscopic view o The repair can be tested with standard saline test Connect a Foley catheter or similar to the tip of the testing syringe and reach that end to the da Vinci instruments for set up in the valve CAUTION When testing the valve or irrigating the left ve
14. arm through the right arm port against the finger The da Vinci arm can be advanced using the shoulder da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 23 o The black trocar is removed and the appropriate instrument is inserted and advanced with the clutch engaged through the port guided by the surgeon s finger through the working port and steered into the view of the endoscope o The instrument tip is placed slightly back from the pericardium to compensate for mediastinal shift back towards the surgeon as the finger is removed from the working port and CO insufflation is lost o The surgeon and assistant then exchange places and the left instrument arm is inserted in a similar fashion Again anticipate movement of the mediastinum toward the da Vinci instruments each time the CO rapidly escapes from the open working port o Right da Vinci instrument Large Needle Driver left da Vinci instrument Resano Forceps Clutch the da Vinci instruments into the center of the video image o Place a 20mm Ethicon Flexipath port into the working port Tailor the length if necessary to extend about 5 to 1 0 cm into the pleural space This soft flexible port is used so there is no working port interference with the right instrument arm when dissection is carried out far to the left up over the aorta It is also easy to occlude this soft port to allow insufflation of the pleural space with CO Figure 15 O vw i S
15. auge in SVC 17 Fr Arterial Biomedicus Cannulae for SVC recommended since clamp can kink SVC Or 0 0 1 20 4 da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 5 o 14Angio 5 25 x 1 for cardioplegia with small slit in end for plegia or DLP Medtronic Aortic Root Cannula with Flow Guard PN 11012L or 11014L e Supplies for postoperative care o Chest tube 19 Fr Blake Drains x 2 1 pericardial 1 chest or 28FR curved chest tube o Pain medication Papavarine 300mgs 10cc 90cc NS or Marcaine with Epi o On Q pain pump system www iflo com 6 da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 3 PATIENT SELECTION POSITIONING PORT PLACEMENT AND DOCKING Patient selection e Patient types for early cases O O O O O Relatively thin patient BMI lt 30 without skeletal deformities Age lt 70 few co morbidities No previous cardiac or thoracic surgery Patients with isolated mitral valve disease without need for concomitant tricuspid valve repair or atrial ablation Start with simple repairs like posterior prolapse P2 e Mitral valve exposure technique for early cases O Teams need to be facile in the set up of manual mechanical atrial retractors like the Cardiovations or Estech Retractor system before using the robotically controlled EndoWrist Atrial Retractor with the 4 arm It is recommended to perform at least 10dVMVR with manual mechanical retractors to develop
16. c arm site selection using a long Veress needle through a small skin incision to test the location and orientation of the planned port 16 da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 o With this set up the position of the da Vinci ports describe a triangle with the camera port so that there is sufficient space for the instruments arm to work without interference The standard port placement for the camera and da Vinci instrument ports looks as follows Figure 10 3 4 finger breadths 3 4 finger breadths Figure 10 Triangular placement of camera port working port and instrument arm ports Port Placement for Atrial Retraction e Manual Mechanical Retractor Set up o Under endoscopic camera view find the previously placed introducer sheath o Insert left atrial retractor handle through sheath until tip of the handle is visible on the endoscopic view o Peel sheath away and position the tip of the handle just inside the chest wall Secure with the Mediflex laparoscopic holding arm or similar Figure 11 o The mechanical atrial retractor should be assembled intrathoracically at this time which Saves pump time Reduces fiddle factor for the patient side assistant since the da Vinci Patient Cart is not yet docked and access to the chest is free o Under endoscopic vision reach retractor blade preferably size 2 5 x 3 5 cm through thoracotomy Use the specific retractor blade clamp that come
17. cope inserted to assist during set up of the other arm At this point in the setup the surgeon can take responsibility for clutching and positioning the camera arm in order to free the assistant for additional setup duties With a finger through the working port and the other hand releasing the set up joint button insert the right da Vinci arm Yellow against the finger through the right arm port The da Vinci arm can be advanced using the shoulder The black trocar is removed and the appropriate instrument is inserted and advanced with the clutch engaged through the port guided by the surgeon s finger through the working port and steered into the view of the endoscope The instrument tip is placed slightly back from the pericardium to compensate for mediastinal shift back towards the surgeon as the finger is removed from the working port and CO insufflation is lost The surgeon and assistant then exchange places and the GF en instrument arm is inserted in a similar fashion Dock this arm but do not engage the Atrial Retractor instrument yet Then the 3rd instrument arm BB is inserted the black trocar removed and the appropriate instrument inserted and advanced with the clutch engaged through the port guided by the surgeon s finger through the working port and steered into the view of the endoscope Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 25 Once again anticipate movement of the mediastinum towa
18. ctor is used for Dynamic Atrial Retraction specific characteristics need to be taken in consideration regarding the operating table set up OR table set up for EndoWrist Atrial Retractor as second left hand instrument standard da Vinci and da Vinci S Figure 5 0000 OR table head rest should be lowered one step OR table rotated 15 to the left in order to raise the patient s right side OR table should be in its lowest position to allow 4 arm to reach over patients head Slight Trendelenburg position can be used to lower patient s head even more and increase space above for the 4 arm if necessary da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 Head rest one step down Slight Trendelenburg can be applied to lower patients head even more Table rotated S 15 to the left Figure 5 OR table set up for Atrial retractor as second left hand instrument with standard da Vinci and da Vinci S CAUTION Set up joint set up characteristic for standard da Vinci4 arm In order to clear the patient s head and reach over to the left instrument port the 4 arm on the standard da Vinci needs to be brought into its highest position in the vertical axis This plus the OR table in the lowest position will allow for the arm to reach e OR table set up for EndoWrist Atrial Retractor as second right hand instrument da Vinci S onl Figure 6 o OR table head rest should be lowered one s
19. da Vinci Mitral Valve Repair via Lateral Approach Procedure Guide INTUITIVE SURGICAL Disclaimer The following material has been reviewed and approved by the following independent surgeons who are not Intuitive Surgical employees Douglas Murphy MD Department of Cardiothoracic Surgery St Joseph Hospital Atlanta Georgia J Michael Smith MD Department of Surgery Division of Cardiothoracic Surgery Good Samaritan Hospital Cincinnati Ohio It is based on information provided by Drs Murphy and Smith and also includes contributions from Leland Siwek MD Department of Cardiothoracic Surgery Sacred Heart Medical Center Spokane Washington Wiley Nifong MD Assoc Professor of Cardiothoracic Surgery Director of Robotic Surgery Brody School of Medicine East Carolina University Greenville North Carolina This procedure guide is provided for general information only and is not provided as formal medical training or certification Intuitive Surgical trains only on the use of the da Vinci Surgical System Intuitive Surgical does not provide clinical training nor does it provide or evaluate surgical credentialing or train in surgical procedures or techniques This material presents the opinions of and techniques used by the above surgeons and not those of Intuitive Surgical Before performing any clinical procedure utilizing the System physicians are responsible for receiving sufficient training and proctoring to ens
20. enefits of MIS to their patients short LOS minimal pain quick recovery minimal blood loss better cosmesis o With the da Vinci System the assisting surgeon has complete access to the mitral valve and can assist during complex steps of the procedure By providing the capacity to use up to four surgical instruments simultaneously two operated by the console surgeon and two operated by the assistant da Vinci Mitral Valve Repair can greatly facilitate an effective and efficient repair 96 2 received a repair with da Vinci Surgery as reported by Douglas Murphy MD compared to 60 of patients in the Fall 2005 Sternotomy database where 42 000 out of 70 000 received a replacement See Society for Thoracic Surgeons amp Duke Clinical Research Institute Executive Summary STS Fall 2005 Report Murphy D Smith M Siwek L Langford D Robinson JR Reynolds B Seshadri Kreaden U Engel A Multicenter Mitral Valve Study a Lateral Approach using the da Vinci Surgical System Pending publication Data on file with Intuitive Surgical Inc da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 3 2 INSTRUMENTS amp ACCESSORIES e OR equipment and da Vinci supplies Basic cardiac surgery room set up including O 0 O0O00000000 0 Warming blanket Bear Hugger Cautery unit set at 15 20 W for COAG CO2 insufflator set a 8 mmHg high flow TEE ultrasound system Mayo stand for da Vinci Cardiovations Supplies Ge
21. ericardiotomy NOTE In order to increase the available working space in the right chest two different approaches can be used 1 The working port can be occluded to insufflate the cavity with CO2 With the 20mm Flexipath port Ethicon it is sufficient to simply place the hand over the port requires good coordination between console surgeon and assistant at OR table With the ATS ports one may put the plastic obturator that comes with the ports back into the port to seal it for CO2 2 Going on pump just before the pericardiotomy will deflate the heart inside the pericardium and create space in the chest This prolongs the pump time slightly but facilitates placement of stay sutures as well as the following procedure steps o In order to have a good overview of the right thoracic cavity in the beginning of the case it is recommended that you start with the 30 endoscope looking up attached to the 60FOV wide angle camera head This will facilitate all stay suture placement and exposure of the anatomy for CPB control NOTE Before placing any stay sutures all the Angiocaths can be preloaded with Cardiovations suture snares in order to speed up suture passing which can be time consuming to coordinate with a laparoscopically untrained assistant The tip of the suture snare can be directed inside the thoracic cavity by having the patient side assistant guide it with his her finger through the working port o Some patien
22. ficient with endo bronchial blocker Place transesophageal echo probe Insert Cardiovations EndoPlege coronary sinus catheter for retrograde cardioplegia Insert Cardiovations EndoVent pulmonary artery vent catheter Right internal jugular IJ catheter to be prepped into field for 17Fr Biomedicus cannula for venous drainage Passed to the level of the right atrium under echocardiographic control Connected to the femoral venous cannula and then to CPB pump o Double lumen left IJ CVP internal jugular vein for central venous pressure Confirm catheter placement with TEE or angiography o Confirm cardiac anatomy with TEE OO O O Or 0 O 8 da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 Patient Positioning Figure 2 3 amp 4 Position patient at right edge of table allowing patients right arm to drop alongside the O operating table rail in order to maximally expose the right chest axilla Figure 2 Figure 2 Patient positioned on right edge of OR table CAUTION It is recommended that you do not position the right arm over the patient s head as this can lead to potential brachial plexus injury Patient s head rotated to the left side to create more space for 4 arm Figure 3 Patient s head rotated to the left O E xe Figure 3 Patient s head positioning and anesthesia set up da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 Figure 4 Sub
23. ft ventricle The left ventricular vent and EndoClamp ascending aorta vents are venting The left lung may be ventilated The EndoClamp is then released The venous retard is then released Transthoracic Cross clamp Chitwood Clamp O O O Unclamp Aorta Continue de airing via plege catheter check for air on TEE Place a pledgeted 4 0 Gore Tex purse string around cardioplegia needle 2 Large Needle Driver Once heart is beating and de aired remove the cardioplegia needle and tie the purse string extra corporeal with pump down to decompress Hemostatic control and Closure of the Chest incisions O O O The left atriotomy is inspected for bleeding with the endoscope If a manual mechanical atrial retractor has been used the anterior chest wall hole at the retractor handle site should be inspected Occasionally this area needs coagulation and rarely a suture 2 0 Gore Tex figure of 8 and tied with a knot tier Suck out thoracic cavity If the surgeons chooses to close the pericardium the use of V100 U clips 2 Large Needle Driver is recommended da Vinci instruments removed Patient Cart undocked and rolled away from patient Inspect all port locations with the endoscope for bleeding and coagulate or suture if necessary Inflate the right lung Place a single chest drain through the right instrument arm port Close the port incisions with subcuticular sutures and rarely subcutaneous sutures Decannulation o Wean f
24. ir the gold standard in surgical treatment using only 1 2 cm incisions Using da Vinci s 3 instrument arm to undertake Dynamic Atrial Retraction puts the cardiac surgeon in total control of the repair procedure la Patient Benefits Compared to other surgical approaches da Vinci Mitral Valve Repair offers patients numerous potential benefits including o Increased chance of receiving a repair the gold standard treatment Less blood loss and need for blood transfusions Reduced risk of wound infections since a large incision and rib spreading is avoided Decreased pain and scarring after surgery Shorter hospital stay Faster recovery Quicker return to normal daily activities O Or O 0 Oo O 1b Surgical Benefits o Surgeons experienced in da Vinci Mitral Valve Repair are more frequently able to successfully repair rather than replace the mitral valve because the da Vinci System provides them with extended surgical capabilities In fact da Vinci may allow surgeons to repair as many as 96 of valves compared with just 60 using sternotomy o The features of the da Vinci System enhance the cardiac surgeon s ability to repair more complex valve abnormalities Barlow s bileaflet repairs amp chordae reconstructions o The lateral approach with da Vinci MVR allows exposure of the mitral valve in its natural state without distortion of the annulus o The technique is reproducible which enables cardiac surgeons to offer the b
25. issecting the intra atrial groove is not necessary or helpful with the more lateral approach e Mitral Valve exposure with manual mechanical atrial retractor system Cardiovations or Estech O If the manual mechanical atrial retractor system has not been already assembled before system docking it needs to be done at this point with the support of the da Vinci System under 3D vision Pass the atrial retractor blade in through the working port with the specific retractor blade clamp that comes with the Cardiovations or Estech system The assistant should pay attention to the correct orientation of the retractor blade If laparoscopic alignment of the retractor rod and the blade is challenging by using the manufacturer clamp the da Vinci instruments might be used for blade manipulation The retractor blade can be inserted into the atriotomy by grasping its edges at a right angle If difficulty is encountered advancing the blade into the left atrium it is usually because the atriotomy is not large enough A stepwise extension of the atriotomy inferiorly and posteriorly will facilitate retractor blade placement and eventually prevent the inferior posterior edge of the atriotomy from being retracted anteriorly and obstructing the view of the valve NOTE The retractor blade selection is based on the left atrial size by echo and displacement of the heart to the left side Most commonly a 2 5 x 5cm or 3 5 x 5cm blade is used Nevertheless
26. itral valve in the chest depending on patient s anatomy The central column should be lined up 7 10cm caudal of the camera port in the right chest to allow more space on the left for the da Vinci arms Figure 17 0 30 7 10cm caudal Figure 17 Patient Cart rolled in at a slight angle and offset towards the patient s feet to provide more space for the 4 instrument arm on the left side 26 da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 o Once the camera arm is directly over the camera port cannula the Patient Cart can be locked brakes down o Then the camera arm is docked to the camera port and the endoscope inserted to assist during set up of the other arm At this point in the setup the surgeon can take responsibility for clutching and positioning the camera arm in order to free the assistant for additional setup duties o With a finger through the working port and the other hand releasing the set up joint button dock the right da Vinci arm to the instrument cannula in the right arm port o The black trocar is removed and the appropriate instrument is inserted and advanced with the clutch engaged through the port guided by the surgeon s finger through the working port and steered into the view of the endoscope o The instrument tip is placed slightly back from the pericardium to compensate for mediastinal shift back towards the surgeon as the finger is removed from the working port
27. ittle visual working space before going on bypass NOTE When using a da Vinci S System the System will sound an alarm since it lost the video signal input when the power to the Camera Control Units was turned off for the camera head swap Push Silence Alarm and Fault Override on the left user interface panel to override the video signal loss After the High Mag camera head is connected and powered the System should function as before da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 33 b 34 Initiation of Cardio Pulmonary Bypass Cardiovations EndoBalloon system O OO O O OO Figure 23 Endoscopic verification of correct EndoClamp position The EndoClamp is passed into the ascending aorta using echo guidance Establish the mean systemic pressure planned for the pump run Allow patient s temperature to drop to 34 C Using the endoscopic camera check for good venous drainage Initiate pulmonary vent catheter drainage Flush antegrade and retrograde catheters until blood is present Focus the endoscopic camera on the ascending aorta and use the right and left instruments to hold open the edges of the pericardium if necessary to visualize the aorta Turn CO flow to a constant0 5L min since the working port will now be constantly open there is no reason to flow based on pressure Inflate the EndoClamp while removing slack from the catheter Alternatively 30mg aden
28. l Pad or towel for patient positioning R2 pads external defibrillator electrodes Scope warmer Thermos Intuitive Surgical High Mag camera head 45 FOV Intuitive Surgical Wide Angle camera head 60 FOV Intuitive Surgical 302 endoscope Intuitive Surgical 02 endoscope e Recommended EndoWrist instrumentation O O O 00T 0 O 1 Permanent Cautery Spatula PN 4xx184 1 Valve Hook PN 4xx192 optional 2nd Valve Hook for analyzing valve before repair Carpentier technique 2 Large Needle Drivers PN 4xx006 1 Resano Forceps PN 4xx181 1 Curved Scissors PN 4xx178 1 Atrial Retractor PN 4xx204 1 DeBakey Forceps PN 4xx036 1 Long Tip Forceps PN 4xx048 or alternatively 1 Pericardial Dissector PN 4xx203 e Recommended Laparoscopic instrumentation O O 0 0 0 O O Cardiovations Forceps Straight Double Acting PN FRSD225 Cardiovations Scissors Curved PN SCC225 Cardiovations Needle Drivers Curved Locking PN NDCX223 Cardiovations Suture Snare PN HK180222 Cardiovations Knot Pusher PN KP1 or knot pusher with closed eye to avoid suture loss during knot tying Straight 5mm or long Ball Tip sucker for Cell Saver suction Extendable handle with magnet tip for needle retrieval Cardiovations Atrial Retractor set PN ARS PN ARSL with large blades or pediatric set or Estech Atrial Retractor system as back up pediatric amp adult blades OR table rail mounting arm for retractor
29. ng da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 41 o Working with Medtronic V100 U Clip Figure 29 amp 30 o Two needles loaded onto a nitinol clip with a blue and white marking o After passing the needles through tissue and band the nitinol clip can be released and will anchor the band on the annulus o In order to avoid the clip ends reaching into the valve orifice and providing a thrombus formation point post operatively it is important to place the colored needles ina specific order blue end always clockwise white end always counter clockwise Figure 29 V100 U Clip before and after release a Grasp U Clip here with Grasp Release mechanism at junction with Large Needle Driver Large Needle Driver and pull off straight to N release U Clip Figure 30 V100 U Clip release mechanism o There are two techniques for placing the annuloplasty band with Medtronic V100 U Clips Technique A Figure 31 1 Drive needle blue end through the right trigone Center the clip within the annulus bite Drive both needles through the annuloplasty band so that both slide releases are clear of the band and a balanced clip arm overlap will be achieved when released 2 Deploy only the slide release mechanism on the clip arm closest to the right trigone white end Temporarily leave the other clip arm blue end attached which will facilitate the subsequent clip placement without compromising ring stability 3 The s
30. ntricle and atrium avoid injecting any air into the left ventricle o Following repair the DLP sump vent catheter is repositioned across the mitral valve to the left ventricle apex and placed to suction with an atmospheric vent valve or needle Be sure the perfusionist applies adequate suction to this vent Figure 34 44 da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 Vent catheter placed across the repaired valve Figure 34 Sump sucker repositioning after mitral valve repair e De Airing and atriotomy closure o If amanual mechanical retractor system is used guide the retractor out of the atriotomy pull up and lock in place at the anterior thoracic wall the retractor blade can be unscrewed and extracted through the working port once atriotomy is closed this will save pump time o If the EndoWrist Atrial Retractor is used remove and extract out of the cannula With the EndoWrist Atrial Retractor move left and right da Vinci instruments out of the atrium and to the side Close the blade tips of the instrument and move slowly out of the atriotomy Straighten the wrist of the instruments and close the tips to facilitate removal from the da Vinci arm for the patient side assistant This is different to other EndoWrist instruments where you want to have the tips open before removal NOTE The EndoWrist Atrial Retractor could also be repositioned to the atrial edge and support atriotomy closure before
31. o the thoracic wall your camera port location is good Figure 8 Z 0 angle with camera port Figure 8 optimal 0 camera angle to view atrial junction da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 O If the camera head has to be tilted more than 452 toward the right costal margin to view the pulmonary veins it may be necessary to move the endoscope port up an interspace This is necessary because if the angle is more than 452 there may be a problem with interference between the endoscopic camera and the right robotic arm later on Figure 9 a ae 45 angle Right robotic arm port with camera within collision zone of port the camera arm Figure 9 Suboptimal camera angle to view atrial junction may result in collisions with the right robotic arm If a manual mechanical retractor system is used Cardiovations or Estech it is recommended that you place the port for the retractor handle at this point during the procedure o Under endoscopic camera view insert an 18 G finder needle parasternally just lateral to the RIMA in the same ICS as the camera port to identify the insertion site o Under endoscopic view insert a 16 Cook pacemaker lead introducer kit applying Seldinger technique adjacent to the finder needle da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 15 Working Port and Instrument Arm Port Placement O O NOTE Remove the endoscope
32. osine can be administered once the balloon is inflated in order to speed up cardiac arrest and avoid EndoBalloon migration Confirm position with the hemodynamic monitor echo and visualization of the aorta Give antegrade cardioplegia Begin venting the ascending aorta secure the EndoClamp and begin retrograde cardioplegia The endoscopic visualization of the proximal ascending aorta is a very valuable information source for the final positioning of the EndoClamp Correct seating of the EndoClamp can be assessed by an aortic bulge and collapsed proximal aorta Figure 23 Aortic bulge compatible with the EndoClamp Vented collapsed proximal aorta Transthoracic Cross clamp Chitwood Clamp O O A 3 0 or 4 0 pledgeted suture is placed in the anterior ascending aorta with two Large Needle Drivers For delivery of antegrade cardioplegia either a 14 Angiocath 5 25 x 1 cut a small slit in end for plegia or DLP Medtronic Aortic Root Cannula with Flow Guard PN 11012L or 11014 is used da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 O Introduce Cardioplegia cath cannula through anterior chest The console surgeon assists with guiding into aorta and holds while securing to cardioplegia line externally NOTE The DLP Medtronic Aortic root cannula can be secured in the aorta by placing two V50 U clips opposite each other through the flange and the tunica adventitia When removing
33. r inframammary fold female Hemostasis is important in this port to prevent blood from running down on the endoscope during the procedure o Depending on chest habitus either the 4t ICS short chest or 5 ICS longer chest is entered o Insert the camera port directing the tip slightly posteriorly away from the right atrium to avoid any possible cardiac injury by the cannula tip o Attach the CO line to the camera endoscope port initially later once all other ports are created it should be moved to a separate Angiocatheter to prevent fogging of the scope during periods of high CO flow o Start CO at a pressure of 6 8 mmHg and increase to higher levels if necessary and if hemodynamically tolerable o Insert normal magnifying camera wide angle with a 30 endoscope angled up and scan right thoracic cavity for adhesions o Confirm that the right lung is well deflated and that the right side of the heart can be accessed NOTE This is an important time to judge the position of the camera port as it is possible to move the port up an interspace if necessary through the same skin incision da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 13 14 o Aim the endoscope at the pulmonary veins with the camera facing the heart and not rotated If the angle of the endoscope shaft to view the pulmonary vein L atrial junction is 02 with the intercostals space Scope coming out of the chest perpendicular t
34. ra port and the working port can be connected to a larger thoracotomy incision With the left index finger through the working port the lung and diaphragm can be protected while performing the following steps 14GA 5 25 Angiocath catheters are inserted through the intercostals spaces in the anterior axilliary line AAL or mid axilliary line MAL for the retrieval of traction sutures to the diaphragm typically 6 ICS and pericardium 1x 4 5 ICS 1x 3 4 ICS Remember that the intercostal branches are running along the caudal edge of the ribs Therefore when introducing the angiocath catheters go in at a perpendicular angle to the chestwall in the midline of the intercostal space The DLP sump is inserted through a 2mm incision in the MAL 4 5 ICS Create the right instrument arm port in either the 5 or 6 ICS caudal of camera port In large males one ICS is satisfactory but two interspaces is ideal to allow easy access for the assistant to the working port Figure 10 Typical position of this port is either on the AAL or slightly medial of AAL keep 3 4 finger breadth between camera port and right instrument arm port Create the left instrument arm port in either the 3 ICS or one ICS up from the camera port Typical position of this port is 2 3cm medial of the AAL keep 3 4 finger breadth between camera port and left instrument arm port NOTE In the first several procedures it may be useful to confirm the roboti
35. ral vein o Do not cannulate yet to avoid effect of Heparin during port placement Port placement dVMVR Lateral Approach Port Placement Placing the ports very lateral in this approach allows for o Excellent mitral valve exposure including the subvalvular structures with minimal retraction and distortion o Minimal intracardiac catheter migration due to reduced distortion o Active participation of the surgical assistant during complex procedure steps due to direct access to the mitral valve structures with minimal interference from the atrial retractor endoscope or robotic instruments Camera port o The planned endoscope and working port incisions can be marked Figure 7 o The most critical part of port selection is proper placement of the camera port If this port is too medial there will not be much working space in the right chest before going on bypass and excessive retraction on the atrial septum will be needed to see the mitral valve o At least 2 3 cm lateral to the nipple in the male and as lateral as possible in the female works the best If surgeon has experience with an anterior thoracotomy the camera port is placed in the most lateral aspect of the typical 5 cm incision 12 da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 Camera Port Working Port Figure 7 Camera and Working Port sites for the lateral approach in a male o Deflate the right lung o Incise skin in 4 ICS male o
36. rd the da Vinci instruments each time the working port is left open as the CO rapidly escapes o Right da Vinci instrument Large Needle Driver left da Vinci instrument Resano Forceps Clutch the da Vinci instruments into the center of the video image o Place a 20mm Ethicon Flexipath port into the working port O The arm set up is as follows AR left of camera right instrument EndoWrist Atrial Retractor left instrument e da Vinci S System 3 arm procedure using manual mechanical atrial retractor system o Docking procedure as described above for the standard da Vinci System in a 3 arm procedure NOTE For docking of the da Vinci S System it is recommended that you insert the metal instrument arm cannulas into the patient with support of the black blunt obdurator first cannulas not attached to the da Vinci instrument arm Once all cannulas are placed move the da Vinci arm in close proximity to the cannula s proximal end and dock to the cannula using the port clutch and arm clutch e da Vinci S System 4 arm procedure with EndoWrist Atrial Retractor used as a second left hand instrument o Before roll up check the sweet spot on the camera arm 90 angle between the camera arm joints to help ensure correct Patient Cart set up during docking The camera elbow should be pointed towards the patient s feet o The da Vinci Sis rolled in at a 0 30 angle to accommodate for the natural position of the m
37. reviously placed through the chest wall is then located and positioned near the planned L atriotomy It is valuable to do this while still using the wide angle camera and before being cross clamped If you forget to do this it is possible for the assistant to find this suction catheter with a finger through the working port and deliver it to the L atriotomy using a shafted instrument but this will use up cross clamp time da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 Pericardial Stay Sutures Optional Optional Pledgeted Exposure stitch in right atrium Anterior stay suture for aorta lt a T EEEN exposure AORTA Planned left w SVC Figure 22 Layout of pericardial traction sutures and left atriotomy o Camera Head Swap The endoscopic camera head is changed from the wide angle camera head to the high mag camera head To change to the high mag camera head turn off the Camera Control Units CCUs and swap the camera head cables plus focus cable Perform black balance and white balance on the CCUs Recalibrate the scope alignment continue to use the 302 up scope with the sterile target and safe settings on the Surgeon s Console NOTE This change takes only a couple minutes if the team is prepared If you stay with the wide angle you will not be able to see as much detail as may be needed If you start with the High Mag camera you will find that there is very l
38. rom CPB o Assess Mitral Valve Repair on TEE and decannulate o Close vessel punctures and groin incision per standard suture protocol da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 5 POST OPERATIVE CARE e Post operative pain management o Pain medication Papavarine 300mgs 10cc 90cc NS or Marcaine with Epi applied directly after surgery at port sites through syringe o Mid term pain management by En Q pain pump system at camera working ports e Internal hospital follow up and discharge o Patient transferred to Intensive Care Unit Infective endocarditis prophylaxis is mandatory No specific postoperative measures are required after valve repair If the patient is in atrial fibrillation permanent anticoagulation is recommended If the patient is in sinus rhythm anticoagulation or antiplatelet drugs are advised for the immediate postoperative period approximately 8 weeks Patient discharged from hospital if Free of hemodynamic problems Free of excessive bleeding No chest wound infections Only required short mechanical ventilation time Free of major adverse cardiac events O oo0o0o e External hospital follow up o Follow up echocardiogram 30 90 days post operatively is recommended to assess the continued adequacy of the repair da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 47
39. rring Severe aortoiliac atherosclerosis Patients with previous median sternotomy Aortic valve insufficiency especially when using Endo Balloon Cardiovations Dilatation of the ascending aorta especially when using Endo Balloon Cardiovations Significant cardiac enlargement insufficient space in thoracic activity Previous history of coronary surgery History of pericarditis da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 7 OR configuration The following figure Figure 1 shows an overhead view of the recommended OR configuration for a da Vinci Mitral Valve repair The Surgeon s Console should be situated on the same side of the operating table as the assisting surgeon to allow for easier communication between assistant and console surgeon This also gives the console surgeon a direct view onto the patient side The perfusion equipment should be positioned at the left side on the foot end of the OR table in order to keep the necessary tubing short Care should be taken to allow enough space for the Patient Cart to come in at a slight angle 15 to 30 to the patient mi O Image m Processing Q Equipment Surgo t Operative Console Figure 1 Operating Room Set up for dVMVR Patient preparation e Anesthesia set up Figure 2 o Bilateral radial or brachial arterial lines mandatory if using EndoBalloon optional for cross clamp Peripheral IV Double lumen endotracheal tube or single lumen if pro
40. rument is changed to a Permanent Cautery Spatula and electrosurgical unit is set to 15W on coagulation You can go higher on the coag setting but more smoke will develop Alternatively this step can be performed with a Curved Scissors in the right hand The pericardium is opened anteriorly and the opening extended inferiorly down to the phrenic at the IVC then superiorly up over the ascending aorta in order to expose as much aorta as possible Figure 21 aorta as possible Extend Pericardiotomy to 0 5 cm above Phrenic Nerve Figure 21 Layout of the Pericardiotomy O O O O O 32 Pericardial traction sutures 4 0 Gore Tex are then placed in the edge of the pericardial flap and retrieved through the lateral Angiocaths These traction sutures are passed through the edge and the two cut ends are passed back to the da Vinci instruments and retrieved with the crochet hook through the previously placed Angiocaths Figure 22 Optional Additionally an anterior stay suture can be placed to expose the aortic root for visualization which is helpful for placing the transthoracic cross clamp or to monitor the EndoBalloon when going on bypass Figure 22 Optional For facilitating initial exposure of the atrium after atriotomy a pledgeted 4 0 Gore Tex stitch can be placed in the edge of the right atrium and brought out through an Angiocath in the 3 or 4 ICS lateral to the RIMA Figure 22 The DLP sump which has been p
41. s with the Cardiovations or Estech system O da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 17 18 Line retractor rod thread section up with blade and screw together until tight It is recommended that you practice this before the first case in a thoracic model with the patient side assistant An anterior 14GA Angiocath next to the retractor rod can be used to wash off the retractor blade after assembly in order to avoid dragging tissue particles from the body wall passage into the heart once it is opened and the retractor applied Pull the retractor up to the anterior thoracic wall and clamp slightly to hold location with the holding arm This way the atrial retractor is readily available once the arteriotomy is performed It is also possible to assemble the manual mechanical atrial retractor system with the da Vinci System and under 3 D vision at a later point during the procedure Retractor Handle Camera and 3 5cm at a gt Service Pot T oO l in 4th L Arm 3rd mr oO A R Arm 5 or 6th EE x Angio for diaphragm traction Cardiovations or Estech da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 e EndoWrist Atrial Retractor Port Set up for teams experienced in da Vinci MVR NOTE Due to the different mechanical concepts of the 4 arm on a standard da Vinci and a da Vinci S System the EndoWrist Atrial Retractor can be used in the following configurations
42. scapular roll and arm positioning 10 O O Place roll or a 6 8 gel pad size similar to a 1000ml IV soft bag under the right mid thorax caudal of scapula in the transversal axis to lift thorax and drop shoulder this helps spreading the intercostals spaces Figure 4 Patient arm off the table in a towel sling Be sure arterial lines are protected during positioning Attach a Mediflex arm or similar scope retractor holder to the left side of the table at the level between left nipple and left shoulder of the patient if manual mechanical atrial retractor system is used or as a back up for manual atrial retractor system Cardiovations or Estech when EndoWrist Atrial Retractor is used Shave if necessary Both groins should be prepared Mark the landmarks of the sternal notch xyphoid intercostals spaces and inframammary fold optional The thorax abdominal and groin region are aseptically prepared and draped Towels might be stapled to the skin to allow for extremely lateral and posterior sterile field preparation When plastic drapes are being applied to women or obese males it is recommended to cut out the drape part over the breast to make manipulation of the tissue and exposure of the infra mammary fold possible Rotate the operating table 15 to the left in order to raise the patient s right side OR table considerations For experienced teams using the EndoWrist Atrial Retractor If the EndoWrist Atrial Retra
43. shes them down with the knot tier Figure 27 b f Wis oN Figure 27 a f Extracorporeal knot tying technique CAUTION Keep suture strands organized to avoid air knots 40 da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 After knots are tied the suture is cut with robotic or laparoscopic scissors For handling and knot tying with Gore Tex suture refer to Gore Tex technical information www goremedical com e Annuloplasty o Following leaflet repair ring size can be determined by assistant passage of ring sizers into the left atrium and positioning with the da Vinci instruments Figure 28 Make sure to put a suture line on the sizer for easy retrieval in the event it is lost in the thoracic cavity o The Edwards Cosgrove ring sizers fit through the ATS port and can be handled easily by the EndoWrist Debakeys and a Large Needle Driver Medtronic also offers a newer Duran sizer that can be handled with robotic instrumentation o If the 20mm Flexipath port Ethicon is used remove it from the working port and introduce the valve sizer directly through the skin incision Take care to avoid dragging tissue particles from the body wall passage into the heart o When another ring is desired but the sizer has a non detachable handle then one can use the Edwards Cosgrove sizer and match it to the desired sizer outside the operative field Figure 28 Sizing the annuloplasty band ri
44. ssages regarding instrument insertion If the message Move instrument out of cannula appears the Atrial Retractor has not yet cleared the cannula end and needs to be inserted further o In patients with smaller chests small intra thoracic space it may be necessary to place the tips of Atrial Retractor instrument into the atriotomy in order to clear the cannula end and allow the console surgeon to take over control o Once the instrument is ready for surgeon control the console surgeon should swap to the corresponding instrument arm and take over control of the Atrial Retractor 36 da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 NOTE For details about how to operate a da Vinci Surgical System equipped with a 4 arm refer to the Users Manual for the specific da Vinci model Only a surgeon trained on the use of a 4 arm da Vinci System should use the EndoWrist Atrial Retractor instrument in dVMVR CAUTION Because the EndoWrist Atrial Retractor instrument has a long blade section gt 40mm and an opening angle of 60 versus usually 30 it moves more quickly than other EndoWrist instruments It is therefore recommended that you train with the EndoWrist Atrial Retractor in a thoracic model before using it in a first case o Before inserting the Atrial Retractor instrument into the atriotomy use either the left or the right da Vinci instrument use the da Vinci instrument opposite from the A
45. tep o OR table rotated 15 to the left in order to raise the patient s right side and drop the left leg to create space for 4 arm o OR table leg board should be flexed down to lower the patient s left leg and increase space above for the 4 arm Head rest one step down Leg board down Table rotated 15 to the left i Figure 6 OR table set up for Atrial retractor as second right hand instrument with da Vinci S da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 11 NOTE When using the robotic Atrial Retractor in a port of the right side of the camera arm 3 of the da Vinci S System may collide with the patient s left hip Therefore it is recommended that you place the patient s groin over the flex point of the OR table in order to flex the legs downward and create space for instrument arm 3 e Cannulation preparation o The right femoral side is preferred for logistical reasons but either groin can be used o The first incision is a 3 to 5 cm oblique incision in the right groin to access the femoral vessels This incision is made as small as possible and parallel to the inguinal ligament to reduce complications and promote cosmesis o Dissect out the R femoral common artery and encircle above and below the planned cannulation site Control branches per routine Purse string suture placed for introduction of arterial cannula o Expose the anterior surface of the R femo
46. the cannula after conclusion of CPB the U clips will easily pull out of the adventitia O After cooling the patient to 262C the transthoracic aortic cross clamp is inserted in the 2 or 3 intercostal space in and mid axillary line Place clamp on aorta under endoscopic vision with exposure aided by console surgeon one arm in transverse sinus and the other anterior to aorta Care should be taken not to injure the aorta pulmonary artery or left atrial appendage Clamp aorta and deliver antegrade cardioplegia Confirm aortic root distention visually with da Vinci endoscope and via TEE Retrograde cardioplegia delivered through Cardiovations EndoPlege coronary sinus catheter c Left Atriotomy and Mitral Valve Exposure e Left atriotomy O O Move the camera and instruments anterior to the pulmonary veins to visualize the left atrium A left atriotomy is performed using Resano Forceps in the left and Curved Scissors in the right Incise the atrium right at the junction of the pulmonary veins and the left atrium Assistant uses the rigid sucker ball tipped to evacuate the left atrium With the pulmonary vent on it is usually possible to easily evacuate the left atrium Extend the atriotomy inferiorly and posteriorly 3cm to 4cm in length to prevent visual interference from the posterior atriotomy edge Keeping the atriotomy a simple direct incision rather than dissecting the intra atrial groove makes closure easier further d
47. the necessary skills to expose the Mitral Valve endoscopically and have a bail out option available should the robotic EndoWrist Atrial Retractor not perform as desired Only when sufficiency with manual mechanical retractors in a robotic assisted procedure with the da Vinci Surgical System has been achieved is it recommended that you apply Dynamic Atrial Retraction with the EndoWrist Atrial Retractor using the 4 arm e Contraindications O O The standard contraindications for cardiac surgery both conventional and minimally invasive should be practiced when selecting patients for robotic procedures In these procedures requiring retrograde femoral perfusion it is recommended to screen the patients by either CT or MRI for severe atherosclerosis and pulmonary adhesions Specific attention should be paid to check on patient s intolerance to single lung ventilation The feasibility of performing a dVMVR should be evaluated on an individual basis and may not be fully appraised until the thoracic cavity has been entered and or the pericardium opened It is recommended that you consult an experienced dVMVR surgeon regarding patients with difficult mitral valve prolapse and co morbidities Experienced teams have operated on patients with the following disease states which should be considered relative contraindications per experience level Heavily calcified mitral annulus Morbid obesity chest wall thickness exceeding 110mm Right pleural sca
48. topcock closed g gipar Soft 20mm port in service port Q re CO2 moved to angiocath Figure 15 da Vinci Patient Cart docked 24 da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 e Standard da Vinci System 4 arm procedure with EndoWrist Atrial Retractor O Before roll up the check the sweet spot on the camera arm 90 angle between the camera arm joints to help ensure correct Patient Cart set up during docking The camera arm elbow should be pointed towards the patient s feet The 4 arm needs to be brought in at its highest position in the vertical axis in order to clear the patient s head and reach over to the left instrument port This plus the OR table in the lowest position will allow the arm to reach The da Vinci Patient Cart is rolled in at a 0 30 angle to accommodate for the natural position of the mitral valve in the chest depending on patient s anatomy The central column should be lined up 10cm caudal of the camera port in the right chest to allow more space on the left for the da Vinci arms Figure 16 7 10cm caudal Figure 16 Patient Cart rolled in at a slight angle and offset towards the patient s feet to provide more space for the 3rd instrument arm on the left side O O da Vinci Once the camera arm is over the camera port cannula the Patient Cart can be locked brakes down Then the camera arm is docked to the camera port and the endos
49. tractor instrument and moving to a new position d Mitral Valve Repair o Left instrument should be changed to a Resano Forceps for atraumatic handling of the valve tissue Do not use DeBakey Forceps Large Needle Driver or Long Tip Forceps as tissue damage may occur o Assess the mitral valve and subvalvular apparatus with Resano Forceps left and Valve Hook right If a Carpentier style analysis is preferred use two Valve Hook instruments e Leaflet repair and annular plication o Mitral repair techniques are standard For cutting leaflet tissue use Curved Scissors o Knot tying can be performed intra or extra corporeally with the help of the surgical assistant o For intra corporeal suture placement and knot tying it is recommended that you use Resano Forceps in the left and a Large Needle Driver in the right if tissue grasping is required Otherwise two Large Needle Drivers or one DeBakey plus one Large Needle Driver will work Do not use a Large Needle Driver on tissue da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 37 o Adistinct advantage of the da Vinci approach is the simultaneous use of the EndoWrist instruments and the patient side assistant instruments The surgeon s assistant passes suture can follow the sutures tie with a knot tier or retract This provides numerous advantages for example the facilitation of accurate placement of Gore Tex chordae e Tips for Suturing o Su
50. trial Retractor instrument to lift up slightly on the atriotomy for easier insertion o Keep the instrument nearest the Atrial Retractor instrument out of the way so you do not hinder insertion into the atriotomy o The patient side assistant might support initial atrial exposure with a rigid sucker through the working port o With the tips of the Atrial Retractor instrument closed blades overlapping move it slowly into the atriotomy until the blade portion of the instrument is sufficiently deep inside the atrium Avoid quick movements on the master since these will translate to fast tip movements of the Atrial Retractor inside the heart o Slowly open the blades of the retractor while simultaneously lifting up on the atrial septum Moving the Atrial Retractor in different directions may enhance mitral valve exposure o Once desired mitral valve exposure is achieved swap back to the corresponding da Vinci instrument arm This will lock the EndoWrist Atrial Retractor in place and return instrument control back to the console surgeon O NOTE Avoid quick movements on the master controls since this will translate to fast tip movements of the Atrial Retractor inside the heart Make sure other da Vinci instruments are out of the way of the EndoWrist Atrial Retractor to avoid internal collisions o Throughout the procedure mitral valve exposure can now be dynamically adjusted to meet surgical needs by swapping back to the Atrial Re
51. ts require a diaphragmatic tension stitch to avoid interference with the right instrument entering the thoracic cavity o Left da Vinci instrument Resano Forceps Right da Vinci instrument Large Needle Driver o After passing a suture 2 0 Gore Tex through the working port place a figure of 8 suture in the central tendon of the diaphragm o Next the console surgeon hands the needle back to the assistant before the assistant pulls the suture out the console surgeon puts the closed grip of his da Vinci instrument through the suture loop to avoid tissue rupture or pulling the stitch out Figure 20 a o The assistant pulls the needle end out makes equal lengths and passes both ends back to the da Vinci instruments Figure 20 b o The right da Vinci instrument can then bluntly push the right diaphragm down to the right o The crochet hook is passed through the Angiocath placed for this diaphragmatic traction suture to retrieve the suture ends Figure 20 c 30 da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 o With the suture ends engaged in the crochet hook the Angiocath is advanced to trap the sutures and then hook and Angiocath are withdrawn simultaneously This should be standard maneuver for all traction sutures Figure 20 d Figure 20 a b Stay suture technique da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 31 O Extend Pericardiotomy as high up on the oes Right inst
52. ture stitch Assistant Surgeon AS hands suture over to the console surgeon s CS right da Vinci instrument with the needle oriented in laparoscopic grasper in a way so that the CS can pick it up Fora single stitch the CS simply hands the needle back to the AS Fora figure of 8 stitch the CS pulls some of the suture through after the needle passes the tissue and places the second stitch Then the CS hands the needle back to the AS However before the AS pulls the suture out the CS puts the closed grip of his da Vinci instrument through the suture loop to avoid tissue rupture or pulling the stitch out Figure 25 7 oe E m e I cS builds loop with robotic instruments to protect suture Figure 25 Left and right da Vinci instruments build a suture loop NOTE To facilitate needle passing between da Vinci and non da Vinci instruments the console surgeon and assistant surgeon should orient the needle in their respective needle driver instruments so that the other surgeon can pick it up easily In addition care should be taken to orient it in a way that it passes easily through the working port without dropping it o Suture following 38 da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 Figure 26 a d Suture following technique with suture hook Should a running suture line closing of leaflet cleft atriotomy closure be required the following approach can save time and make effecti
53. ubsequent clips are placed in a clockwise fashion When the clip is fully seated and balanced the remaining release on the previous clip is deployed 4 Next deploy the proximal release on the current clip Repeat this STEP until the final clip is placed 5 In conclusion the furthest clip away from the initial starting point is always left unreleased until the next clip is fully placed 42 da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 Figure 31 Order of U Clip placement Technique A 9 3 more clips might be needed r depending on valve size N J 1 5 6 Technique B mark the middle of the band with a suture to validate correct placement at P2 Figure 32 1 Figure 32 Order of U 9 t Drive needle blue end through the right trigone Center the clip within the annulus bite Drive both needles through the annuloplasty band so that both slide releases are clear of the band and a balanced clip arm overlap will be achieved when released Deploy only the slide release mechanism on the clip arm closest to the right trigone white end Temporarily leave the other clip arm blue end attached which will facilitate the subsequent clip placement without compromising ring stability The next clip is placed in the left trigone by driving the needle white end through the left trigone Deploy only the slide release mechanism on the clip arm closest to the left trigone blue end The subsequent clips
54. ure that they have the requisite training skill and experience necessary to protect the health and safety of the patient For technical information including full cautions and warnings on using the da Vinci System please refer to the System User Manual Read all instructions carefully Failure to properly follow instructions notes cautions warnings and danger messages associated with this equipment may lead to serious injury or complications for the patient While clinical studies support the use of the da Vinci Surgical System as an effective tool for minimally invasive surgery outcomes cannot be guaranteed as surgery is patient and procedure specific 2006 Intuitive Surgical All rights reserved Intuitive Intuitive Surgical da Vinci da Vincie S InSite TilePro and EndoWrist are trademarks or registered trademarks of Intuitive Surgical All other product names are trademarks or registered trademarks of their respective holders PN 871401 Rev A 4 06 da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 1 Procedure Guide dVMVR TABLE OF CONTENTS 1 INTRODUCTION e Patient Benefits e Surgical Benefits 2 INSTRUMENTS AND ACCESSORIES e OR equipment and robotic supplies Recommended EndoWrist Instruments Recommended laparoscopic instrumentation Recommended sutures Other surgical supplies Perfusion Supplies for postoperative care 3 PATIENT SELECTION POSITIONING PORT PLACEMENT AND DO
55. ve suture placement quicker Instead of passing the needle back amp forth between the AS and CS and in amp out of the thorax the CS keeps the needle inside the body after he passes it through the tissue With one of the da Vinci instruments the CS should pull a short suture loop of the tissue edge so that the AS can pass a suture hook into that loop Figure 26 a By retrieving the suture loop with the hook to the outside of the body and keeping the suture strand tight standard suture following technique can be achieved Figure 26 b This maneuver is repeated until the running suture line is completed Figure 26 c d CAUTION Care should be taken by the AS to not over cross the suture strands o Intra corporeal knot tying Cut suture off 2 3 from where it passed through the tissue can be done with the EndoWrist Curved Scissors or by the AS with laparoscopic scissors Different types of knot techniques can then be performed with da Vinci instrumentation sliding square knot surgeon s knot etc da Vinci Mitral Valve Repair Procedure Guide PN 871401 Rev A 5 2006 39 o Extra corporal knot tying AS holds both suture strands outside the thorax Gore Tex 36 or 46 is used and cuts one needle off This suture is then fed through the loophole eye of the knot tier Figure 27 al The scrub nurse or physician assistant holds that suture strand tight while the AS ties the knots with the other strand and pu

Download Pdf Manuals

image

Related Search

Related Contents

Sony 4-199-738-13(1) User's Manual  seratec® deluxe seratec® deluxe seratec® deluxe  施主説明書 」 いま した. ` 。 丶丶 機會種部 ニ DT  PRO PATCH LIQUIDE™  Digipass manual  取扱説明書  USER'S MANUAL  Samsung Mono Multi-function Printer SCX-4521F دليل المستخدم  Instrucciones de usuario  Sony DAV-DZ230 Home Theater System User Manual  