1. 1 Press and hold the UNIT key for at least 1 5 seconds to switch from one unit of measure to another 2 Note that pressing and holding the UNIT key continuously allows for quicker scrolling Release the key when the desired unit of measure is displayed Thermcouple Type J or K Temperature Mode 1 Select the Thermocouple Temperature mode using the FUNCTION button 2 The meter will display a J or a K on the left side of the display indicating the current setting To change the thermocouple type refer to the Setup Mode section 3 Connect a Type J or Type K thermocouple to the meter s sub miniature thermocouple jack top of meter on the left 4 Hold the thermocouple in the air in the area to be tested 5 The meter will display the thermocouple temperature in the main display area 6 To change the temperature unit of measure C and F refer to the Setup Mode section Data Hold To freeze a measurement on the display press the HOLD key momentarily The meter will emit a beep the reading will hold and the HOLD display icon will switch on Press the HOLD key again to release the display and exit the Data Hold mode returning the meter to the normal operating mode 3 SDL300 EU EN V1 5 07 13 MAX MIN Readings For a given measurement session this meter can record the highest MAX and the lowest MIN readings for later recall 1 2 3 Press the MAX MIN key momentarily to access this mode of operation R
2. EC icon appears The meter is now recording the MAX and MIN readings Press the MAX MIN key again to view the current MAX reading MAX icon appears The reading on the display is now the highest reading encountered since the REC icon was switched on when the MAX MIN key was first pressed Press the MAX MIN key again to view the current MIN reading MIN icon appears The reading on the display is now the lowest reading encountered since the REC icon was switched on when the MAX MIN key was first pressed To exit the MAX MIN mode press and hold the MAX MIN key for at least 1 5 seconds The meter will beep the REC MAX MIN icons will switch off the memory will clear and the meter will return to the normal operating mode Display Backlight To turn the display backlight ON or OFF press and hold the backlight key for at least 1 5 seconds The meter will beep when switching the backlight ON or OFF unless the beeper is disabled as described in the Setup Mode section of this user guide System Reset If the meter s keys become inoperable or if the display freezes the Reset button can be used to reset the instrument 2 Use a paper clip or similar item to momentarily press the reset button located on the lower right side of the instrument under the snap off compartment cover After pressing the Reset button switch the instrument ON by pressing and holding the POWER key for at least 1 5 seconds If using the power adaptor unplu
3. EXTECH USER GUIDE INSTRUMENTS Vane Thermo Anemometer Datalogger Model SDL300 Introduction Congratulations on your purchase of the Extech SDL300 Thermo Anemometer Datalogger This meter displays and stores air velocity and temperature readings Temperature readings from the thermometer built into the vane probe or from an externally connected type K or J thermocouple temperature probe can be displayed and recorded Data is stored on an SD card for transfer to a PC In addition an RS232 port allows data streaming to a PC This instrument is shipped fully tested and calibrated and with proper use will provide years of reliable service Please visit the Extech Instruments website www extech com to check for the latest version of this User Guide Extech Instruments is an ISO 9001 certified company Safety International Safety Symbols This symbol adjacent to another symbol or terminal indicates the user must refer to the manual for further information Meter Description 1 Thermocouple input Vane input plug Air velocity reading Units of measure 2 3 4 5 HOLD Backlight key 6 7 8 MAX MIN key 3 SET and Clock key 4 PC interface jack 5 9 Reset key 6 10 Power Adaptor jack 7 11 SD card slot 8 12 ENTER and LOG key 9 13 Down Yarrow and FUNCTION key 10 14 Up A arrow and UNITS key 15 Power ON OFF key 16 Temperature reading with units of measure 17 Anemome
4. all amount of test data 2 With the meter switched OFF remove the SD Card 3 Plug the SD Card directly into a PC SD card reader If the PC does not have an SD card slot use an SD card adaptor available at most outlets where computer accessories are sold 4 Power the PC and run a spreadsheet software program Open the saved documents in the spreadsheet software program see examples of spreadsheet displays below Spreadsheet data example 238 AMTemp C B8 AMTersp C B8 AMTenp C 229 ANTenp C 22 AMTemp E BS AMTemp C WA AMTemp C BS AMTemp C WS AMTemp RE AMTeap RI AMTenp C 232 AMTemp C 232 AMTeap C 237 ANTeap E 29 1 ANTemp E 29 ANTenp E 23 9 ANTeap E 24 AMTeap E 27 AMTeap C 7 AMTemp C 23 9 AMTenp C 22 9 AMTexap C BI AMTexa C RS 232 USB PC Interface For streaming of data to a PC via the RS232 Output jack the optional 407001 USB kit RS232 to USB cable and driver CD along with the 407001 software available free at www extech com sdl300 are required 6 SDL300 EU EN V1 5 07 13 Setup Mode Basic settings at a glance To view the current configuration of the meter with regard to time date thermocouple type and datalogging sampling rate press the SET CLOCK ICON key momentarily The meter will now display the configuration in quick succession Repeat as necessary to observe all of the information Accessing the Setup mode 1 2 Press and hold the SET key f
5. cimal European and USA numerical formats differ The meter defaults to USA mode where a decimal point is used to separate units from tenths i e 20 00 European formats use a comma i e 20 00 to separate units from tenths To change this setting Ae 2 3 4 Access the dEC parameter as described in the Accessing Setup Mode section above Use the arrow keys to select USA or EUro Press ENTER to confirm setting Press and hold the SET key for at least 1 5 seconds to exit to the normal operation mode or simply wait 7 seconds for the meter to automatically switch to the normal operating mode SD Card FORMATTING 1 2 aan Access the Sd F parameter Use the arrow keys to select YES to format the card select NO to abort Note that all data on the card will be lost if formatting is attempted Press ENTER to confirm selection Press ENTER again to re confirm The meter will automatically return to the normal operating mode when formatting is complete If not press and hold the SET key for at least 1 5 seconds to exit to the normal operation mode Set the Temperature Units of Measure C or F 1 2 3 4 Access the t CF parameter Use the arrow keys to select C or F Press ENTER to confirm setting Press and hold the SET key for at least 1 5 seconds to exit to the normal operation mode or simply wait 7 seconds for the meter to automatically switch to the normal operating mode Set the Temperature T
6. eter sensor head 72mm 2 8 diameter 10 SDL300 EU EN V1 5 07 13 Electrical Specifications Ambient Temperature 23 C 5 C Air velocity Measurement Range Resolution Accuracy m S 0 4 35 0 m s 0 1 m S 2 rdg 0 2 m S Km h 1 4 126 0 km h 0 1 Km h 2 rdg 0 8 Km h mph Mile h 0 9 78 3 mile h 0 1 mph 2 rdg 0 4 mph Knots 0 8 68 0 knots 0 1 Knots 2 rdg 0 4 Knots FPM Ft min 79 6890 ft min 1 FPM 2 rdg 40 FPM Note 5 minute test limit on air speed gt 30 0m s 108km h 67mph 58 knots 591 0ft min Air temperature Measuring Range 0 to 70 C 32 to 158 F Resolution 0 1 C 0 1 F Accuracy 0 8 C 1 5 F Type K J Thermocouple Thermometer Temperature Sensor Type Resolution Range Accuracy 0 1 C 50 0 to 1300 0 C 0 4 rdg 0 5 C 100 0 to 50 1C 0 4 rdg 1 C Type K 0 1 F 58 0 to 2372 0 F 0 4 rdg 1 F i 148 0 to 58 1 F 0 4 rdg 1 8 F 0 1 C 50 0 to 1200 0 C 0 4 rdg 0 5 C i 100 0 to 50 1 C 0 4 rdg 1 C Type J OAF 58 0 to 2192 0 F 0 4 rdg 1 F 148 0 to 58 1 F 0 4 rdg 1 8 F Note Above specifications tested under an environmental RF Field Strength lower than 3 V M anda frequency lower than 30 MHz Copyright 2013 FLIR Systems Inc All rights reserved including the right of reproduction in whole or in part
7. g the adaptor and then plug it back in again to power the meter AC Power Adaptor This meter is normally powered by six 6 1 5V AA batteries However an optional AC 9V power adaptor is available To use the adaptor insert the power adaptor plug into the jack on the bottom right side of the meter under the snap off compartment cover plug the other end of the adaptor into an AC power source The meter will now be permanently powered as long as the adaptor is used and the power key will be disabled 4 SDL300 EU EN V1 5 07 13 Datalogger Types of Data Recording Manual Datalogging Manually log up to 99 readings onto an SD card via push key press Automatic Datalogging Automatically log data onto an SD memory card where the number of data points is limited only by the card size Readings are logged at a rate specified by the user SD Card Information Insert an SD card from 1G size up to 16G into the SD card slot at the bottom of the meter With regard to orientation the card should be inserted with the front of the card label side facing toward the rear of the meter If the SD card is being used for the first time it is recommended that the card be formatted and the logger s clock set to allow for accurate date time stamping during datalogging sessions Refer to the Setup Mode section for SD card formatting and time date setting instructions European and USA numerical formats differ The data on the SD card can be f
8. he same type 9 SDL300 EU EN V1 5 07 13 Specifications General Specifications Display Measurement Units Datalogger Sampling Rate Memory Card Temperature Compensation Data Hold Memory Recall Display update rate Data Output Operating Temperature Operating Humidity Auto Power OFF Power Supply Power Consumption Weight Dimension Backlit LCD LCD size 52 x 38mm 2 x 1 5 Air velocity m S meters per second Km h kilometers per hour Ft min FPM feet per minute Knots nautical miles per hour Mile h mph miles per hour Air temperature C F Type K Type J thermocouple C F AUTO 1 2 5 10 30 60 120 300 600 1800 3600 seconds Note that a one 1 second sampling rate can cause some data loss on slower computers MANUALLY Set the sampling rate to 0 SD memory card 1G to 16GB size Automatic temperature compensation for the Anemometer function and the type K J thermometer function Freeze the display reading Record and Recall the Maximum and Minimum readings Approx 1 second RS 232 USB PC computer interface 0 to 50 C 32 to 122 F 85 R H max After 10 minutes of inactivity can be disabled Six 6 1 5VDC batteries optional 9V AC adaptor Normal operation backlight amp datalogger OFF Approx 15 mA dc With backlight OFF and datalogging ON Approx 36 mA dc 347g 0 76 Ibs meter only Main instrument 182 x 73 x 47 5mm 7 1 x 2 9 x 1 9 Anemom
9. hermocouple Type J or K f 2 3 4 Access the tYPE parameter Use the arrow keys to select J or K Press ENTER to confirm setting Press and hold the SET key for at least 1 5 seconds to exit to the normal operation mode or simply wait 7 seconds for the meter to automatically switch to the normal operating mode 8 SDL300 EU EN V1 5 07 13 Battery Replacement and Disposal When the low battery X icon appears on the LCD the batteries must be replaced Several hours of accurate readings are still possible in this condition however batteries should be replaced as soon as possible 1 Remove the two 2 Phillips screws from the rear of the meter directly above the top of the tilt stand 2 Remove and safely place the battery compartment and screws where they will not be damaged or lost 3 Replace the six 6 1 5V AA batteries observing polarity 4 Replace the battery compartment cover with the two 2 Phillips screws All EU users are legally bound by the Battery Ordinance to return all used batteries to community collection points or wherever batteries accumulators are sold Disposal in household trash or refuse is prohibited Disposal Follow the valid legal stipulations in respect of the disposal of the device at the end of its lifecycle Other Battery Safety Reminders Never dispose of batteries in a fire Batteries may explode or leak Never mix battery types Always install new batteries of t
10. in any form www extech com 11 SDL300 EU EN V1 5 07 13
11. itched off However if the battery expires the clock will have to be reset after fresh batteries are installed Setting the Datalogger Sampling Time Rate i 2 Access the SP t parameter Use the arrow keys to select the desired sampling rate The available settings are 0 1 2 5 10 30 60 120 300 600 1800 and 3600 seconds Press the ENTER key to confirm the entry Press and hold the SET key for at least 1 5 seconds to exit to the normal operation mode or simply wait 7 seconds for the meter to automatically switch to the normal operating mode Enabling Disabling the Auto Power OFF Feature 1 2 Access the PoFF parameter Use the arrow keys to select ON enable or OFF disable With the Auto Power OFF feature enabled the meter will automatically switch OFF after 5 minutes of inactivity Press ENTER to confirm setting Press and hold the SET key for at least 1 5 seconds to exit to the normal operation mode or simply wait 7 seconds for the meter to automatically switch to the normal operating mode 7 SDL300 EU EN V1 5 07 13 Set the Beeper Sound ON or OFF 1 2 3 4 Access the bEEP parameter Use the arrow keys to select ON enable or OFF disable Press ENTER to confirm setting Press and hold the SET key for at least 1 5 seconds to exit to the normal operation mode or simply wait 7 seconds for the meter to automatically switch to the normal operating mode Numerical Setting comma or de
12. or at least 1 5 seconds to access the Setup menu Press the SET key momentarily to step through the available parameters The parameter type is shown on the bottom of the LCD and the current selection for that type is shown above it When a parameter is displayed that is to be changed use the arrow keys to change the setting Press the ENTER key to confirm a change Press and hold the SET key for at least 1 5 seconds to exit the Setup mode Note that the meter automatically switches out of the Setup mode if no key is pressed in 7 seconds while in the Setup mode The available Setup parameters are listed below Additional detailed information is provided below this list dAtE Set the clock time Year Month Date Hour Minute Second SP4 Set the datalogger sampling rate 1 to 3600 seconds PoFF Automatic power off management bEEP Set beeper sound ON OFF dEC Set SD card Decimal character comma for European format SdF SD memory card Format tCF Select the Temperature unit of measure to C or F tYPE Select the thermocouple type to K or J Setting the Clock Time f 2 3 4 Access the dAtE parameter Use the arrow keys to change a value Use the ENTER button to step through the selections Press and hold the SET button for at least 1 5 seconds to exit to the normal operation mode or simply wait 7 seconds for the meter to automatically switch to the normal operating mode The clock will keep accurate time even when the meter is sw
13. ormatted for either format The meter defaults to USA mode where a decimal point is used to separate units from tenths i e 20 00 The European format uses a comma i e 20 00 To change this setting refer to the Setup Mode section Manual Datalogging In the manual mode the user presses the LOG key to manually log a reading onto the SD card T 2 4 5 Set the sampling rate to 0 seconds Press and hold the LOG key for at least 1 5 seconds the lower portion of the display will show p n n memory position number 1 99 Press the LOG key momentarily to log a reading into memory The REC icon will flash each time a data point is stored the SCAN SD icon will appear when the meter accesses the card Use the A and Y keys to select one of the 99 data memory positions in which to record To exit the manual datalogging mode press and hold the LOG key for at least 1 5 seconds Automatic Datalogging In automatic datalogging mode the meter takes and stores a reading at a user specified sampling rate onto SD card The meter defaults to a sampling rate of one second To change the sampling rate refer to the Setup Mode section the sampling rate cannot be 0 for automatic datalogging 1 2 To begin an automatic Datalog session press and hold the LOG key for at least 1 5 seconds The meter will scan for an SD card and verify that it can be used to store data If a card is not inserted or if the card is defective the meter
14. ter vane 18 Vane icon faces the air flow 19 Vane handle 20 Vane connecting cable Notes Battery compartment tilt stand and tripod mount are located on the rear of the instrument Items 8 9 and 10 are located behind the snap off compartment cover on right side of meter 2 SDL300 EU EN V1 5 07 13 Operation Power ON OFF e Power the meter by pressing and holding the power key h for at least 1 5 seconds e Press and hold the power key for at least 1 5 seconds to power OFF the meter Meter Functions This meter has two modes of operation Press and hold the FUNCTION key for at least 1 5 seconds to switch from one mode to the other e Air Velocity plus Temperature mode meter briefly displays An icon e Type K or J thermocouple temperature mode meter briefly displays tP icon Air Velocity plus Temperature Mode 1 Select the Air Velocity and Temperature mode 2 Connect the vane probe to the meter s probe jack top right of meter via the probe plug 3 Hold the probe by its handle and allow the air flow to enter the vane Note that the air flow must face the vane icon 4 The meter will display the air velocity measurement upper display area and the air temperature lower display area Changing the Air Velocity Unit of Measure This meter offers five 5 units of measure selections for air velocity m s meters per second FPM feet per minute Km h kilometers per hour knots and mph miles per hour
15. will display SCAN SD indefinitely In this case switch the meter OFF and try again with a valid SD card If the SD card is valid the display will show the LOG icon or the LOG icon alternating with the temperature display and then the REC icon will flash each time that a reading is stored To pause the datalogger press the LOG key momentarily The REC icon will stop flashing To resume logging simply press the LOG key again momentarily To terminate the datalogging session press and hold the LOG key for at least 1 5 seconds When an SD card is used for the first time a folder is created on the card and named AMC01 Up to 99 spreadsheet documents each with 30 000 readings can be stored in this folder When datalogging begins a new spreadsheet document named AMC01001 xIs is created on the SD card in the AMC01 folder The data recorded will be placed in the AMC01001 xlIs document until 30 000 readings are reached If the measurement session exceeds 30 000 readings a new document will be created AMC01002 xls where another 30 000 readings can be stored This method continues for up to 99 documents after which another folder is created AMCO2 where another 99 spreadsheet documents can be stored This process continues in this same fashion with folders AMC03 through AMC10 last allowable folder 5 SDL300 EU EN V1 5 07 13 SD Data Card to PC Data Transfer 1 Complete a datalogging session as covered above For the first test simply record a sm