Home

Appendix J: SOP of Reporting and Calculation of Contaminant

image

Contents

1. SPAN buttons fail to appear see Section 11 for troubleshooting tips RANGE 500 0 PPB gt ENTR SPAN CONC RANGE 500 0 PPB gt ENTR CONC Press ENTR to change the offset amp slope values for the H2S measurements Press EXIT to leave the calibration unchanged and return to the previous menu EXIT 4 H2S X XXX EXIT returns to the main SAMPLE display EXIT _ _ gt 209 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Manufacturer Contact Information Teledyne Advanced Pollution Instrumentation 9480 Carroll Park Drive San Diego CA 92121 5201 USA Phone 858 657 9800 800 324 5190 Fax 858 657 9816 Email api sales teledyne com Website http www teledyne api com Reference Appendix K Version 1 3 09 13 06 Page 7 of 8 Instruction Manual of Model 101E UV Fluorescence H2S Analyzer Rev A2 26 September 2005 210 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix K Version 1 3 09 13 06 Page 8 of 8 Maintenance Calibration Record Sheet for API 101E H2S Analyzer Date of Calibration Calibrated by Time Items Unit Notes Ambient pressure mm Hg Sample airflow L min Zero air applied Cylinder P __ psi H2S Reading H2S ppm applied Cylinder P __ psi H2S Read
2. Pour the remaining residue into waste container designated for selentum waste When the container is full or the container has been in the lab for 90 days the liquid needs to be disposed of through EH amp S Titration 1 Before titrating a sample completely fill the buret with 0 1N HCI and drain to waste enough so the initial reading of the buret can be made Record initial reading 2 Place a small stir bar in the receiving flask and place on a stir plate beneath of the buret Stir gently 3 Add HCI from the buret and allow the color to change from emerald green to a gray to light purple 4 Titrate to the graph to light purple endpoint Note It is easy to overshoot the endpoint When it takes longer for the green color to reappear slow the addition of the acid down to one drop at a time If too much acid is added darker purple the sample has been over titrated and is lost 5 Refill the buret for the next sample Calculations Normality of acidx mL HCl added mL from blank x1 4 sample size Nitrogen in sample Use an average of the replicates of each sample for the final value Routine Maintenance and Cleaning Digestion Clean with a detergent and water solution when the unit is cool 1 Condensed sulfuric acid will accumulate in the digestion manifold If acid drops on the digestion block digestion tubes support rack or counter top in the hood the acid should immediately be wiped up with a wet paper towel 2
3. 12 Close regulator valve on the span gas cylinder Open main valve on span gas cylinder Insert the 1 4 ID tubing from the gas cylinder into the inlet of the flow meter mounting on the side of instruments rack Adjust regulator valve until vent airflow is about 5 L min read from bottom of ball of the vent monitoring flow meter This provides a little extra span gas to the analyzers and keeps the pressure inside the manifold close to the atmospheric pressure Flow Span gas for 2 minutes Press Main More Menu until GC is displayed and Press GC Press Single Wait for about 3 minutes until the CH and Residual analysis is finished If the CH or Residual reading is not in the range of span gas concentration 0 1 ppm the potentiometer needs to be adjusted i Press Setup ii Press Methane or Residual and Press Span iii Turn the front panel knob If the reading is larger than span gas concentration reduce the potentiometer value If the reading is smaller than span gas concentration increase the potentiometer value iv Press Main More Menu until GC is displayed and Press GC Press Single Wait for about 3 minutes until the CH and Residual analysis is finished Repeat steps 6 10 for both Methane and Residual until an acceptable zero reading is reached Press Main More Menu until Mod
4. 12 5 mL 125 mL of concentrated HCl using a 10 mL pipet 2 Put 1 L 10 L of deionized water in a container and add to it slowly the HCl 3 Add a stir bar to the solution and let it mix for 10 15 minutes on a stir plate 4 Bring the volume up to 1 5 L 15 L with deionized water and continue stirring Standardization of Hydrochloric Acid 1 THAM Tris hydroxymethyl aminomethane is a primary standard base that will be used to standardized the hydrochloric acid 2 Dry THAM in a desiccator it should not be subjected to heat greater than 100 C 3 Measure about 0 25 gm or less record exact weight of THAM using a weigh boat 246 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix S Version 1 3 09 13 06 Page 9 of 13 Add the THAM to three 125 mL Erlenmeyer flasks and dissolve in about 30 mL distilled 8 water by swirling the mixture Add two drops of the Methyl Red Indicator will be a yellow color Titrate the THAM solution to a red color with the HCI solution being standardized using the 25 mL buret Record the volume of acid required for each flask and calculate the normality of the HCl Average the normality of the three flasks Normality of HCl gm THAM x 1000 121 1 x mL of HCl Indicate normality and the date standardized on the label of the 0 1N HCl Sample Preparation Liquids l 2 3 4 5 Mix liquid manure
5. Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Cl API 101E check Appendix T Version 1 3 09 13 06 Page 4 of 7 Time Items Unit Notes Ambient pressure mm Hg Sample airflow L min Zero air applied Cylinder P __ psi H2S reading H2S ppb applied Cylinder P __ psi H2S reading New calibration Yes hs 0 Zero air applied H2S applied ppb Finish calibration Yes R 0 Time switch off Connect analyzer back to sampling system Note O Polytro I calibration Time Items Unit Notes Ambient pressure mm Hg Sample airflow L min Zero air applied Cylinder P __ psi H2S reading Calibration H2S ppb applied Cylinder P __ psi H2S reading Calibration New calibration Calibration Finish calibration Time switch off Connect analyzer back to sampling system Note 255 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing O Innova check Appendix T Version 1 3 09 13 06 Page 5 of 7 Time Items Unit Notes Ambient pressure mm Hg Sample airflow L min Zero air applied Cylinder P _ psi NH reading CO reading N20 reading CH reading Propane
6. Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Step 3 Perform the zero span calibration procedure Appendix K Version 1 3 09 13 06 Page 6 of 8 SAMPLE RANGE 500 0 PPB lt TST TST gt CAI SAMPLE H2S STB X XXX PPB lt TST TST gt CAL H25 XXX X Set the Display to show the H2S STB test function SETUP This function calculates the stability of the H2S x measurement H2S X XXX SETUP J ACTION rear of the instrument M P CAL H2S STB X XXX PPB Allow zero gas to enter the sample port at the Wait until H2S STB falls below 0 5 ppb This may take several minutes SO2 X XXX lt TST TST gt x SETUP M P CAL H2S STB X XXX PPB lt TST TST gt ZERO CONC SO2 X XXX EXIT M P CAL H2S STB X XXX PPB lt TST TST gt ENTR CONC ACTION rear of the instrument Allow span gas to enter the sample port at the Press ENTR to changes the OFFSET amp SLOPE values for the SO measurements Press EXIT to leave the calibration unchanged and return to the previous menu The value of H2S STB may jump significantly Wait until it falls back below 0 5 ppb The SPAN key now M P CAL H2S STB X XXX PPB This may take several appears during the minutes transition from zero to span lt TST TST gt SEAN CONC You may see both keys If either the ZERO or
7. Rapidstill Make sure the holes in the tube are completely immersed in liquid Make sure the boiler is full of water If not depress the Boiler Water switch to refill the boiler Turn the Boiler Heater on Depress the NaOH Adadition button to add 85 90 mL of NaOH per 20 mL of acid used in the digestion Measure this value according to the graduate cylinder the NaOH is contained in When the water is boiling nicely being distillation by setting the timer knob to the desired distillation time Distill for 20 minutes or until froth reaches the condenser arm The distillate is collected in the 500 mL Erlenmeyer flask Lower the flask so that the end of the outlet tube is free of contact with the liquid and continue distillation during the last minute or two to cleanse condenser and outlet tube When distillation is complete the Rapidstill will return to standby position with the boiler heater still in operation Don t let the boiler stay on switch the Boiler Heater button to the off position Once the boiler is cool depress Boiler Water switch to refill the boiler after each run If doing additional distillations turn the boiler heater back on and the machine is ready when boiling resumes 249 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix S Version 1 3 09 13 06 Page 12 of 13 11 There may be a few mL of sample left in the tube
8. Storage and Exchange Appendix O Version 1 3 09 13 06 Page 2 of 6 The measurements must be conducted with TEOM filter cartridges that are made of Teflon coated glass fiber filter paper Filters should be stored inside the sensor unit for easy access and 230 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix O Version 1 3 09 13 06 Page 3 of 6 to keep them dry and warm Do not handle new filters with your fingers Instead use the filter exchange tool and follow the procedures given in the operating manual Keep the sample pump running to facilitate filter exchange Use the two pockets on the right side of the mass transducer inside to store the next two new TEOM filters for pre conditioning and removal of excessive moisture build up prior to use System Operation after Power Failure The system resets itself when power is regained and enters the same mode as before All operating parameters are maintained in the system s battery backed up CMOS memory Setting Variables Stored in Data Logger The lt Step Screen gt key toggles the instrument between the View Storage Screen and Set Storage Screen or pressing 09 lt Enter gt can gain direct access to the Set Storage Screen from any screen The first eight lines of the Set Storage Screen contain the titles of the variables currently being stored in the data logger Program Register Codes listed in Appendix A of
9. based on routine flow auditing leak checking and mass calibration verification Since the TEOM monitor can be directly mass calibrated it can be directly quality assured using a mass standard All QA procedures should be coordinated with routine maintenance procedures to minimize down time Flow Audit A flow audit adapter is provided and the procedures are outlined in the operating manual Both the sample flow rate and total flow rate may be checked using the flow audit adapter with a capped nut for closing the flow splitter bypass line port It is recommended that the volumetric flow rates be within 7 of the set points The United States Environmental Protection Agency USEPA requires a tolerance of 10 for the total flow through the PMjo inlet If measured flows differ by more than the stated tolerances recheck all settings and perform the test again Large errors in the flow may indicate other sources of error such as a malfunctioning flow controller a system leak or improper temperature and pressure settings Leak Check The leak check procedures are included in the operating manual Section 7 6 The leak check should be performed with NO sample filter attached to the mass transducer this will prevent accidental damage from occurring to the sample filter cartridge when exposed to the high pressure drop vacuum in the sample line created during leak checks Flow rates should indicate less than 0 15 L min for the main flow and less tha
10. filter P N 30 003097 Manufacturer Contact Information Rupprecht amp Patashnick Co Inc 25 Corporate Circle Albany NY 12203 233 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Timothy Morphy Assistant Product Manager 518 452 0065 voice 518 452 0067 fax tmorphy rpco com email Peggy O Gormon Marketing Associate Phone ext 3229 email pogorman rpco2 com Reference TEOM 1400 Operation manual 234 Appendix O Version 1 3 09 13 06 Page 6 of 6 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix P Version 1 3 09 13 06 Page 1 of 2 Appendix P SOP of Barometric Pressure Sensor The purpose of barometric pressure measurements is for correction of the emission calculations The sensor has two wires a red wire for power and the black wire for the output signal Warning Always connect the sensor with the power turned off The barometric pressure sensor may be stored without any special provisions During storage place the sensor inside a bag to keep the sensor clean and store it on a shelf or hang it on a wall Barometric pressure will be monitored continuously outside of the houses using a WE100 with a range of 80 000 to 110 000 Pa and an accuracy of 300pa One barometric static pressure sensor is required for each house The barometric pressure sensor will be compared and cal
11. for 1 minute at low speed with a drill mixer Measure the liquid sample with a pipet If the sample is not able to be pipetted treat it as a solid sample At first we will try a range of sample volumes from 0 to 1 00 mL Prepare duplicates or triplicates of each sample To improve accuracy using a 25 mL buret to titrate the ammonia more than 15 mL of HCl should be used for the titration of each sample Use at least two blank tubes without sample in a run of 25 digestion tubes The blanks should contain all reagents receiving the same treatment as the samples There will be an indicator color change in the blank If not something has gone wrong Also if new boric acid or HCl is made during the day blanks are needed for the new reagents Blanks must be run everyday Samples can be weighed into digestion tubes one or more days before digesting if tubes are covered so they do not become contaminated 0 05 0 06 gm urea or 0 2 gm tryptophan can be used to check the system for complete digestion without loss of nitrogen Solids l 2 3 Weight representative sample in a tarred weigh boat and record the weight Transfer the sample to a digestion tube making sure that everything gets to the bottom of the tube Prepared duplicates or triplicates of each sample To improve accuracy using a 25 mL buret to titrate the ammonia more than 15 mL of HCl should be used for the titration of each sample For solid samples we will firs
12. methyl red and methylene blue to each receiving flask Swirl to mix You can mix the boric acid and indicator solution together prior to putting it into the receiving flask Turn on cooling tap water by going under the sink and looking way in the back In the top right corner there is a silver handle Turn it 45 to the left allowing water to flow through the condenser The distillate should be cold as it comes out The distillation coil should be cold at the bottom and somewhat warm at the top Check this flow rate periodically as the water pressure can vary If there is inadequate water flow to cool the distillation coil ammonia will be lost with steam instead of being condensed as a liquid in the receiving flask Depress Boiler Water switch and fill to the line on the flask back of the distillation unit Turn on the Boiler Heater by pressing the button When water is at a steady boil the unit is ready for operation Make sure the NaOH inlet tube located on the left side of the Rapidstill is in a 500 mL graduate cylinder that is full of NaOH Distillation Procedure 2 10 Open the safety screen pull down on the bottom lever piece and place the digestion tube containing the digestor sample in the clamping device Close the safety screen Place the 500 mL Erlenmeyer receiving flask containing the 70 mL 4 boric acid and 5 drop Indicator Solution under the Distillate Outlet tube on the front right part of the
13. operating manual The Interval variable defines the time in seconds between successive writings of data to the circular buffer Table O 1 Some popular program register codes Code Variable Units 008 Mass concentration ug m 009 Total mass ug 035 Pressure drop 039 Current main flow L min 040 Current auxiliary flow L min 041 Status condition code 057 30 min average mass concentration ug m 130 Current ambient temperature C 131 Current ambient pressure atm Setting Analog Outputs The instrument s three analog output channels are accessible from the identical 15 pin connectors on the front and back panels of the control unit Details regarding the pin assignments and voltage VDC of outputs are given in the operating manual Section 5 1 To bring up the Set Analog Output Screen press lt A O gt or select Set Analog Output from Menu Screen or press 04 lt Enter gt Setting Sampling Average Time The lt Step Screen gt key toggles the instrument to Set Hardware then press lt Enter gt Press the Data Stop to start stetting the sampling average time 231 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix O Version 1 3 09 13 06 Page 4 of 6 Nomenclature TM Total mass MR Mass rate MC Mass concentration A O Analog output atm atmosphere 746 mBar Calibration Procedures are
14. pee ee ates eee eee seep OE e e eee sS OE 0 0 00 00 00 30 01 00 01 30 02 00 02 30 03 00 Time API 101E Response Time 100 z 6 98 pa 100 R Rea ca pond a ecg a E Weare hk a ee ea S API 965 0 to 93 API 965 44 to 93 80 zL a Be tt Shao hee a eS ee o Qa i 9 2o Le eae ate ee ene en eel 60 F oO i E 4o ieee SA AE E ana Sa a AN 8 40 Aa a a 20 00 00 00 30 01 00 01 30 02 00 02 30 Time 03 00 Figure K 2 API response time Functional Check 0 After the analyzer s components have warmed up for at least 30 minutes verify that the software properly supports any hardware options that were installed Check to make sure that the analyzer is functioning within allowable operating parameters using the values listed in the manufactory data sheet with the operation manual To view the current values of these parameters press the following key sequence on the front panel 206 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix K Version 1 3 09 13 06 Page 4 of 8 SAMPLE RANGE 1000 PPB H2S XX lt TST TST gt CA SETUP RANGE DARK PMT Toggle lt TST TST gt H2S STB DARK LAMP keys to scroll through PRES H2S SLOPE list of functions SAMP FL H2S OFFS PMT HVPS NORM PMT RCELL UV LAMP TEMP LAMP BOX TEMP RATIO PMT TEMP STR LGT TIME Changing the Sample Particulate Filter The pa
15. reading CO ppm applied with Nafion tubing Cylinder P _ psi CO reading NH ppm applied Cylinder P _ psi NH reading CH ppm applied Cylinder P _ psi CH reading Propane ppm applied Cylinder P _ psi Propane reading N20 ppm applied Cylinder P _ psi N20 reading New calibration Yes INo Zero air applied Zero humidity air applied NH applied CO applied with Nafion tubing Propane applied CH applied N20 applied Download to bank Yes INo Time switch off Connect analyzer back to sampling system Note 256 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Ll VIG 200 check Appendix T Version 1 3 09 13 06 Page 6 of 7 Time Items Unit Notes Ambient pressure mm Hg Sample airflow L min Zero air applied Cylinder P _ psi Total reading CH reading Propane reading CH ppm applied Cylinder P _ psi CH reading Propane ppm applied Cylinder P _ psi Propane reading New calibration Zero air applied Total adjustment CH adjustment Propane adjustment Total reading CH Reading Propane reading CH applied Total adjustment CH adjustment Propane adjustmen
16. until the CH and Residual analysis is finished If the CH or Residual reading is not in the range of 0 1 ppm the potentiometer needs to be adjusted i Press Setup ii Press Methane or Residual and Press Zero 213 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix L Version 1 3 09 13 06 Page 3 of 5 iii Turn the front panel knob If the reading is larger than 0 reduce the potentiometer value If the reading is smaller than 0 increase the potentiometer value iv Press Main More Menu until GC is displayed and Press GC Press Single 10 Wait for about 3 minutes until the CH and Residual analysis is finished 11 Repeat steps 6 10 for both Methane and Residual until an acceptable zero reading is reached 12 Press Main More Menu until Modes is displayed 13 i Press Modes ii Press Track for Display iii Press Main More Menu until GC is displayed iv Press GC and Press Setup v Press Total vi Press Zero vii Turn the front panel knob until the Total reading in the display window reads 0 0 ppm Close regulator and remove tubing from the zero gas cylinder 14 Change the Modes display setup back to Peak Span Gas Calibration 1 2 3 oon oS 10 11
17. 8 1 8 ENTR EXIT SETUP XX t DAS RNGE PASS CLK MORE SETUP XX RANGE CONTROL MENU MODE SET UNIT EXIT Press this button to select the gt concentration units of measure PPB PPM UGM MGM Press this button to set the analyzer for SNGL DUAL or AUTO ranges Pa SETUP XX RANGE 500 0 CONC j D 0 0 EXIT ignores the new setting and To change the value of the C returns to the RANGE CONTROL reporting range span enter the gt mrad by paang the pein RANGE 500 0 Cone ENTR ea and each digit un expected value appears RANGE CONTROL MENU Step 2 Set the expected H2S span gas concentration SAMPLE RANGE 500 0 PPB H2S XXX X lt TST TST gt CAL SETUP This sequence causes the _ analyzer to prompt for the expected H S span concentration M P CAL RANGE 500 00 PPB H2S X XXX lt TST TST gt ZERO CONC EXIT The H S span l concentration values automatically default to 450 0 Conc EXIT ignores the new setting and returns to the previous display ENTR accepts the new setting and returns to the previous display To change this value to M P CAL H2S SPAN CONC 450 0 Conc the actual concentration of the span gas enter the number by pressing the key under each digit until the expected value appears ENTR EXIT 208 A Quality Assurance Project Plan for
18. A B x 280 7 mL sample 20 Liquid samples mg NH3 N L A B x 280 g dry wt sample x 20 Sludge or sediment samples mg NH3 N kg where A volume of H2SO titrated for sample mL B volume of H2SO j titrated for blank mL 244 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix S Version 1 3 09 13 06 Page 7 of 13 Rapidstill II Routine Maintenance Daily Following Use ios SOANDMN 1 Install a clean empty digestion tube Remove caustic NaOH supply tubing from vessel containing NaOH and place in vessel of purified water Turn on cooling water and fill boiler Turn on boiler heater Depress NaOH Addition until all of the caustic NaOH has been purged out of the system and into digestion tube Remove the digestion tube and dispose of the liquid properly Replace digestion tube and fill half full by depressing NaOH Addition When water is at a rolling boil turn distillation timer to 10 minutes At end of distillation cycle turn off the boiler heater and cooling water Remove digestion tube and dispose of liquid properly Clean exterior surfaces of machine using a soft cloth and a mild soap and water solution Weekly 1 2 Examine boiler for any build up of deposits Accumulated deposits may be removed by a dilute solution of hydrochloric acid or by using commercially available descaling so
19. A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Lara B Moody Hong Li Robert T Burns Hongwei Xin Richard S Gates Steven J Hoff Doug Overhults Appendix J 2008 Appendices J T pp 199 258 in A Quality Assurance Project Plan for Monitoring Gaseous and Version 1 3 Particulate Matter Emissions Copyright American Society of Agricultural and Biological Engineers 09 13 06 ASABE 913C0708e Page of 5 Appendix J SOP of Reporting and Calculation of Contaminant Concentrations Ventilation and Emissions This SOP gives instructions for calculating and reporting gas ammonia carbon dioxide hydrogen sulfide non methane hydrocarbon and dust PM2 s PMio TSP concentrations and barn emission rates Concentration Defined Concentration is broadly defined as the fraction of a constituent of interest contained within a sample Concentration of gaseous contaminants are generally reported in volumetric mass based or mixed units see next section Concentrations of particulates can be provided as number of particles in a sample or mass of particulates in a certain size range in a mass of sample gas Gaseous Contaminants Units of Measurement Concentration of gaseous contaminants may be expressed in volumetric mass based or mixed units fractions The first two are dimensionless except for any scale factor whereas mixed units fractions are typically expressed as mass of
20. Emissions from Broiler Housing Appendix J Version 1 3 09 13 06 Page 4 of 5 standard conditions are most appropriate The mass concentrations of outside and inside samples should be adjusted to standard temperature and pressure as per equation 4 In equation 5 the volume units used for volumetric flow rate denominator and mass concentration denominator must match For example non standard moist air standard moist air or standard dry air would be the consistent units to use Further typically Q e is measured see SOP for airflow measurement but Q must be computed from continuity Qiii Q Ve or Q iI Vi Ve Qe 6 where vi v are inside air and outside air specific volumes m moist air per kg dry air Specific volume may be calculated from air density kg moist air per m dry air or kg ma per kg da and humidity ratio W kg H20 per kg da by v 1 W p 7 Summary of Measurements and Calculations for Emission Rate Measurements required to compute ER include 1 Interior and exterior air state points dry bulb temperature relative humidity 2 Interior and exterior mass volumetric concentrations 3 Building static pressure differences and atmospheric pressure Intermediate calculations to compute ER include 1 Interior and exterior saturation and partial water vapor pressure humidity ratio air density specific volume from equations 4 and 7 2 Exhaust standard ventilation rate Q from FAN
21. M PhO Pump 10 LPM Tubing 1 8 ID Tee 1 4 OD Valve ball oO a oO oO Pr Cardboard box with an adjustable opening for Setra Salt LiCI and NaCl Distilled water Clean syringe 0 10 mL Wide mouth bottle changeable cover Ruler for TEOM Fitting box screwdrivers hex drivers Electrical box Coverall and sprinkler disinfect 30 ft power extension cord CFP and crossover cable Dilutor and manual H2S gas and regulator Manuals and checklist TEOM INNOVA Power tool and drill bits 253 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix T Version 1 3 09 13 06 Page 3 of 7 O Thermocouple check unit C Time Location Temp 1 Standard Temp 2 Standard Temp 3 Standard C Check air flow rate in sample lines and solenoid valve Time Temp C Location Flow rate LPM Valve SW1 SW3 Tunnel Ambient O Gal gas injection from sampling port Time Temp C Location Flow rate LPM NH Concentration ppm SwWw1 SW3 Tunnel Ambient O Leak test the TEOMs Time Temp C Location TEOM SN Main Flow rate LPM Aux Flow rate LPM 254 A Quality
22. M Total mass MR Mass rate MC Mass concentration A O Analog output atm atmosphere 746 mBar Calibration Procedures are based on routine flow auditing leak checking and mass calibration verification Since the TEOM monitor can be directly mass calibrated it can be directly quality assured using a mass standard All QA procedures should be coordinated with routine maintenance procedures to minimize down time Flow Audit A flow audit adapter is provided and the procedures are outlined in the operating manual Both the sample flow rate and total flow rate may be checked using the flow audit adapter with a capped nut for closing the flow splitter bypass line port It is recommended that the volumetric flow rates be within 7 of the set points The United States Environmental Protection Agency USEPA requires a tolerance of 10 for the total flow through the PMjo inlet If measured flows differ by more than the stated tolerances recheck all settings and perform the test again Large errors in the flow may indicate other sources of error such as a malfunctioning flow controller a system leak or improper temperature and pressure settings Leak Check The leak check procedures are included in the operating manual Section 7 6 The leak check should be performed with NO sample filter attached to the mass transducer this will prevent accidental damage from occurring to the sample filter cartridge when exposed to the high pressure drop vacu
23. Measure 83 mg of methylene blue with a weight boat using the enclosed balance 3 Dissolve the methyl red and methylene blue in a 100 mL volumetric flask with 100 mL of 95 ethanol This may take 10 20 minutes to dissolve 4 Store the solution in an amber colored bottle and label 4 Boric Acid 1 Measure 80 gm 500 gm of powdered boric acid with a large weight boat 2 Slowly add the boric acid to about 1 5 L 10 L of deionized water and allow the acid to dissolve by using a stir bar and a stir plate it might take 30 minutes to 1 hour several hours for greater amounts 3 Bring the volume up to 2 L 12 5 L with deionized water and continue stirring on the stir plate for 5 to 10 minutes 30 Sodium Hydroxide 1 Measure 750 gm 1800 gm of dry NaOH using a large weigh boat 2 Slowly add the NaOH to 1 5 to 2 L 5 L water and allow the base to dissolved by using a stir bar and a stir plate 3 This is an exothermic reaction container will get hot 4 Bring the volume up to 2 5 L 6 L with deionized water and continue stirring on a stir plate for 5 to 10 minutes Methyl Red Indicator Solution 1 Measure 20 mg of methyl red using a weigh boat 2 Dissolve the methyl red in 60 mL absolute ethanol using a 100 mL volumetric flask This may take 10 15 minutes 3 Bring the volume of the solution up to 100 mL by adding 40 mL distilled water and continue stirring for 5 10 minutes 0 1N Hydrochloric Acid 1 120 Dilution 1 Measure out
24. S calibration and measured static pressure 3 Incoming standard volumetric ventilation rate Q from equation 6 Combining equations 5 7 compute emission rate ER from ER Q C Me Vi Ve C mi mg contaminant s 8 Adjust ER time basis to reflect issues described in next section Concentration Data Extraction Averaging and Interpolation Gas Concentration Extraction and Averaging When analyzing each of the house air samples four 30 second measurement cycles by the ammonia analyzer will be performed to ensure attainment of 97 or better of the expected concentration value If fans at all three sampling locations are running the time interval of a complete sampling cycle will be 120 xx 3 360 seconds If SW3 and or TF1 are not running 202 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix J Version 1 3 09 13 06 Page 5 of 5 their sample analysis will be skipped and the sampling will go back to SW1 or fresh background air Airflow rates corresponding to the measured concentrations will be used in the calculation of the overall house emission rate Since compositions of the background air are much more stable than the house air it will be sampled only once every 2 hours As a result of the larger step change in ammonia concentration between the house air and background air a longer sampling time i e 5 minutes will be used to allow full
25. Wipe down outer surfaces of the digestion block Wipe down hood bench if acid has spilled Wipe out glass attachments at the end of the aspiration tubes Use wet paper towels 3 Wipe off metal tops of the racks and the racks themselves using wet paper towels Calibration The controller has been factory calibrated and should not require re calibration In the event re calibration is required the following procedure should be followed 1 Insert calibration resistor into the 2 pin sensor connector located on the rear of the control unit 2 Place ON OFF switch located on the controller to the ON position 3 Adjust set point indicator dial to a position where the indicator lights located on front panel of digestor are illuminated 50 of the time 250 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing 4 5 Appendix S Version 1 3 09 13 06 Page 13 of 13 Loosen setscrew on set point indicator dial and align calibration mark on dial with pointer label The digestor is now calibrated to give digestion temperature indicated on set point dial Distillation Daily following use 1 Wash digestion distillation tubes in hot water no detergent Use a scrub brush to clean the outside and inside Rinse 2 or 3 times with hot water Rinse twice with distilled water Place in drying oven or on a rack so they will dry 2 Rinse receiving flasks three times with hot water a
26. an for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix S Version 1 3 09 13 06 Page 2 of 13 Procedure for Solid Manure Wolf N 2003 Determination of manure pH p 48 49 In Peters et al eds Recommended Methods of Manure Analysis University of Wisconsin System http cecommerce uwex edu pdfs A3769 PDF accessed on May 23 2006 e Place 20 g of solid manure into a beaker e Add 40 ml distilled water e Stir and mix it well with a glass rod e Allow for 30 minutes to settle solution e Measure pH by immersing electrode into the supernatant solution e Record pH value when the meter has stabilized e Repeat above steps for replicates and report the result as pH water 1 2 Total Solids or Moisture Content Standard Method 2540 B Total Solids Dried at 103 105 C Dish Preparation ALWAYS handle dishes with gloves on 1 Place clean evaporating dishes in oven at 103 105 C for 1 hour 2 Remove dishes from oven and cool to room temperature in a desiccator This takes about 2 3 hours Testing Liquids ALWAYS handle dishes with gloves on 1 Measure and record the weight of each measuring dish using the enclosed balance 2 Place three 10 15 ml of samples about 2 3 full into separate evaporating dishes Measure the samples with a 10 mL pipet Do not get any sample on the outside of the dish If it happens wipe it off quickly Measure and record weight of each sample and dish usin
27. ants Standard Conditions The relation between C y and C y for standard conditions of 25 C and 101 325 kPa is given by C m M 24 45 C v 3 For ammonia carbon dioxide hydrogen sulfide and methane the relative molecular masses are 17 03 44 01 34 08 and 16 04 respectively From equation 3 the relation between standard concentration on mass and volumetric basis are C mnes 0 6965 C v Nus C m co2 1 8 C v co2 C m ns 1 3939 C v nzs C m cu4 0 6560 C v cus Gaseous Contaminants Effects of Temperature and Pressure A volumetric concentration measurement ppm taken at non standard conditions may be converted to mass concentration mg m by equation 2 for a given pressure p and absolute temperature T Particulate Contaminants Units of Measurement Concentration of particulate contaminants may be expressed in mixed units mass of particulate in a unit volume of sample air or as a particle count per unit volume of sample air If the latter is used the particle count is understood to be within a particle size range e g up to 2 5 um or 10 um PM2 5 and PMjo Total suspended solids may be expressed as either particle count or particle mass per unit volume of sample air In this SOP we shall always use mixed units for particulate concentrations i e mg m Ventilation Rates Units of Measurement Fan ventilation rates standard m s or sm s for each running fan are determined from buil
28. articulate Matter Emissions from Broiler Housing Appendix M Version 1 3 09 13 06 Page 4 of 6 Nomenclature TM Total mass MR Mass rate MC Mass concentration A O Analog output atm atmosphere 746 mBar Calibration Procedures are based on routine flow auditing leak checking and mass calibration verification Since the TEOM monitor can be directly mass calibrated it can be directly quality assured using a mass standard All QA procedures should be coordinated with routine maintenance procedures to minimize down time Flow Audit A flow audit adapter is provided and the procedures are outlined in the operating manual Both the sample flow rate and total flow rate may be checked using the flow audit adapter with a capped nut for closing the flow splitter bypass line port It is recommended that the volumetric flow rates be within 7 of the set points The United States Environmental Protection Agency USEPA requires a tolerance of 10 for the total flow through the PMio inlet If measured flows differ by more than the stated tolerances recheck all settings and perform the test again Large errors in the flow may indicate other sources of error such as a malfunctioning flow controller a system leak or improper temperature and pressure settings Leak Check The leak check procedures are included in the operating manual Section 7 6 The leak check should be performed with NO sample filter attached to the mass transducer this wil
29. at Using the Keypads and Software User should refer to Section 4 5 of the operating manual and become familiar with the incorporated menu driven software and keypads Filter Storage and Exchange The measurements must be conducted with TEOM filter cartridges that are made of Teflon coated glass fiber filter paper Filters should be stored inside the sensor unit for easy access and 224 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix N Version 1 3 09 13 06 Page 3 of 6 to keep them dry and warm Do not handle new filters with your fingers Instead use the filter exchange tool and follow the procedures given in the operating manual Keep the sample pump running to facilitate filter exchange Use the two pockets on the right side of the mass transducer inside to store the next two new TEOM filters for pre conditioning and removal of excessive moisture build up prior to use System Operation after Power Failure The system resets itself when power is regained and enters the same mode as before All operating parameters are maintained in the system s battery backed up CMOS memory Setting Variables Stored in Data Logger The lt Step Screen gt key toggles the instrument between the View Storage Screen and Set Storage Screen or pressing 09 lt Enter gt can gain direct access to the Set Storage Screen from any screen The first eight lines of the Set Storage S
30. becomes blank Turn off the vacuum pump by pressing the power button and disconnect the control unit from 120 VAC by unplugging the power cord Status Line on Main Screen Whenever a status code other then OK is shown on the display the instrument automatically turns on the Check Status light on the front of the control unit The information displayed on the main screen includes status condition operating mode A O 1 mode RS 232 mode protection and time For example the screen will show the following line at time XX XX OK 4 51 NU XX XX 223 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix N Version 1 3 09 13 06 Page 2 of 6 Status Condition OK Normal operation M No frequency signal T Temperature s outside of operational bounds F Flow s outside of operational bounds X Filter nearing capacity exchange filter Operating Mode 1 Temperature flow stabilization 2 Begin TM computation 3 TM computed begin MR MC computation 4 Normal operating mode S Setup mode X Stop all mode A O 1 Mode Blank Analog output 1 normal definition Analog output 1 used for status watch XX Filter loading percent RS 232 Mode N None P Print online R R amp P protocol A AK protocol G German network protocol S Storage to printer F Fast storage output Protection U Unlocked L Low lock H High lock XX XX Current time 24 hr form
31. contaminant per unit volume of sample mixture e g mgm Volumetric concentration is generally expressed in parts per million ppm or parts per billion ppb defined as the parts of contaminant per million or billion parts of air by volume 1000 ppb lppm Mass based concentration is expressed in either mg kg or ug kg with the numerator referring to the mass of contaminant and the denominator to the mass of sample air Mixed unit concentration is most convenient for this work and is generally expressed as either mg m or ug m While this is technically a mixed unit expression for concentration it is often called mass concentration and we adopt this naming convention in this SOP It is convenient because when multiplied by volumetric flow rate of air an emission rate is directly determined Conversion between volumetric C y ppm and mass C m mg m concentrations assuming the mixture behaves as an ideal gas is C v 8 309T Mp C m 1 C m 0 1204 C yMp T 2 where p sample mixture pressure kPa Air Contaminants Chapter 12 ASHRAE HOF 2005 American Society of Heating Ventilation and Air Conditioning Engineers Atlanta GA 199 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix J Version 1 3 09 13 06 Page 2 of 5 T sample mixture temperature K M relative molar mass of contaminant dimensionless Gaseous Contamin
32. creen contain the titles of the variables currently being stored in the data logger Program Register Codes listed in Appendix A of operating manual The Interval variable defines the time in seconds between successive writings of data to the circular buffer Table N 1 Some popular program register codes Code Variable Units 008 Mass concentration ug m 009 Total mass ug 035 Pressure drop 039 Current main flow L min 040 Current auxiliary flow L min 041 Status condition code 057 30 min average mass concentration ug m 130 Current ambient temperature C 131 Current ambient pressure atm Setting Analog Outputs The instrument s three analog output channels are accessible from the identical 15 pin connectors on the front and back panels of the control unit Details regarding the pin assignments and voltage VDC of outputs are given in the operating manual Section 5 1 To bring up the Set Analog Output Screen press lt A O gt or select Set Analog Output from Menu Screen or press 04 lt Enter gt Setting Sampling Average Time The lt Step Screen gt key toggles the instrument to Set Hardware then press lt Enter gt Press the Data Stop to start stetting the sampling average time 225 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix N Version 1 3 09 13 06 Page 4 of 6 Nomenclature T
33. dic maintenance and the schedule to be adopted for use in livestock barns are as follows Maintenance item Factory This Project Clean air inlet head Upon filter exchange Twice a week Replace TEOM filter cartridge 70 load Weekly Exchange in line filters 6 mon or when loaded When loaded Leak test Annually Every flock Analog board calibration Annually Every 6 months The TEOM PM bno air inlet requires regular maintenance in livestock barns The TEOM inlet itself should be cleaned weekly In order to do this one must first push the lt Data Stop gt button on the keypad Remove the dirty TEOM inlet and replace it with a clean inlet The dirty PMio inlet will be taken back to work station for cleaning and ready for next inlet swap For cleaning the dirty inlet please see Manual Appendix G Inlet Maintenance Grease as needed to maintain a slight layer of silicon grease Carefully replace the inlet unit and return to the control unit of the TEOM system and press the lt Data Stop gt button once more This will prevent skewed data from being recorded It will automatically begin recording in approximately 30 min Other Settings Flow rate through sample inlet 16 7 L min 1 m hr Main flow rate 1 L min Temperature of sample stream 50 C Particulate concentration lt 5 ug m to several g m Standard Conditions 1 atm pressure 20 C Spare Parts and Consumables Slow blow 2A 250 V fuse 2A amp 250 V in line fuse P N 04003419 Bo
34. ding static pressure difference SP Pa and the calibration equation for the fan as follows QFan at b SP where the parameters a b are different for each fan and are obtained from regression of the FANS calibration data explained elsewhere in this SOP Building ventilation rate Q is determined by summing all fans that are running at any time In the event that large spatial variations are noted building ventilation rate can be broken into 200 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix J Version 1 3 09 13 06 Page 3 of 5 representative amounts near each sampling location typically two values in the broiler housing e g Q e and Q e The relations between actual Q moist standard Q and dry standard Q volumetric flow rates are Q 1 W Q 1 W Q p p T T 4 where W humidity ratio of air kg H2O per kg dry air 0 62198 p p pv Py water vapor pressure kPa rh 100 py sat rh air relative humidity Pws saturation water vapor pressure p actual pressure kPa p standard pressure 101 325 kPa T actual absolute temperature K T standard absolute temperature K Q actual moist volumetric flow m s Q moist standard volumetric flow sm s Q dry standard volumetric flow dsm s Emission Rates Units of Measurement Emission rate ER of a gaseous or particulate contami
35. e 20 C Spare Parts and Consumables Slow blow 2A 250 V fuse 2A amp 250 V in line fuse P N 04003419 Box of 20 TEOM filter cartridges TX40 media P N 57 000397 0020 Large bypass in line filter P N 57 002758 Flow controller filter P N 30 003097 Manufacturer Contact Information Rupprecht amp Patashnick Co Inc 25 Corporate Circle Albany NY 12203 USA 221 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Timothy Morphy Assistant Product Manager Phone 518 452 0065 Fax 518 452 0067 tmorphy rpco com email Peggy O Gormon Marketing Associate Phone ext 3229 email pogorman rpco2 com Reference TEOM 1400 Operation manual 222 Appendix M Version 1 3 09 13 06 Page 6 of 6 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix N Version 1 3 09 13 06 Page 1 of 6 Appendix N SOP of Tapered Element Oscillating Microbalance TEOM for PM Introduction The TEOM instrument TEOM 1400 Ambient Particulate PM10 Monitor is a continuous PM monitoring device designated by USEPA as an equivalent method EPA Designation No EQPM 1090 079 for PMio lt 10 um aerodynamic diameter The acronym TEOM stands for Tapered Element Oscillating Microbalance an inertial measurement technique that operates on changes in the resonant frequency of an oscillating element as a function o
36. e buffer solution with a 10 mL pipet Sample Preparation of Known Sample for Calibration 150 mL samples with 100 mg L NH3 Manure and 100 mg L NH Sample 1 Extract a 13 7 mL of mixed sample place in beaker 2 Add 15 0 mL 1000 mg L NH solution 3 Add 121 3 mL distilled H20 4 Add 7 5 mL borate buffer solution Distilled H20 and 100 mg L NH Sample 1 Place 15 0 mL 1000 mg L NH solution in a digestion tube using a 10 mL pipet 2 Add 135 0 mL distilled H2O to the digestion tube by filling a 100 mL volumetric flask first then adding 35 mL with a 10 mL pipet 3 Add 7 5 mL borate buffer solution with a 10 mL pipet Preparation of Blank Sample Carry a blank sample through all steps of procedure to use in final calculations 1 Measure 75 mL distilled H20 using a 50 mL volumetric flask and a 10 mL pipet for the rest 2 Add 3 75mL borate buffer solution with a 10 mL pipet Distillation Procedure 1 Open the protective casing pull down on the bottom lever piece and place the digestion tube containing the sample in the clamping device 2 Place a 500 mL Erlenmeyer flask containing 50 mL indicating boric acid solution under the Distillate Outlet tube Make sure the holes in the tube are completely immersed in liquid 243 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix S Version 1 3 09 13 06 Page 6 of 13 3 Make sure the boiler is full of water If
37. e the download file as both a Rotem his and Excel xls file Designate the files to be saved in the corresponding folder House 1 5 or House 3 3 on the computer notebook C Documents and settings Administrator My Documents Rotem Data As soon as download is complete close the Scalenet application disconnect the communications cables and turn off the computer Repeat this process during each visit When downloading overwrite the previous file At the end of the flock email the final excel file xls to ISU for evaluation Between Flocks 1 2 3 Immediately prior to bird catch remove the scale platform from the load cell and clean and store it in a safe location Download the final flock weights and email the final excel file xls to ISU for evaluation Sometime prior to placement of the next flock check the accuracy of the scales using known weights Upon verification remove the platform and store it in a safe location Immediately after placement of the next flock of birds place the platform back on the load cell Using the Scalenet application click on the Setting icon button and reset the scales database A different screen will appear In the left column click on Growth Day to change the day to 1 click on Flock Number to change to the current flock number designation and click on Time to verify the correct time with the computer time References Rotem Poultry Scale S
38. ertial measurement technique that operates on changes in the resonant frequency of an oscillating element as a function of increases in particle mass collected on a filter attached to the element Changes in the element s resonant frequency are sampled electronically in quasi real time providing both continuous and time averaged measures of mass accumulation that are directly proportional to instantaneous and time averaged mass concentrations in air respectively The device operates at an industry standard volume controlled flow rate of 16 7 L min so that it can be outfitted with a variety of commercially available pre separator inlets suitable for measuring PM The unit is fitted with a TSP inlet head in this application Switching Instrument On Off Since the airflow is maintained at a constant volume corrected for local temperature and barometric pressure the operation of this monitor requires that the temperature and pressure sensors are connected for proper temperature and pressure readings and flow corrections Turn On Unit Supply power to the instrument by plugging in the power cord to 120 VAC and pressing the power button on the front panel of the TEOM control unit The main screen four line display will soon appear after showing the name of the instrument Turn on the pump to draw the sample streams by plugging in the power cord to 120 VAC The monitor waits at least 30 min after being powered up to compute the first mass concentrat
39. es is displayed i Press Modes ii Press Track for Display 214 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix L Version 1 3 09 13 06 Page 4 of 5 iii Press Main More Menu until GC is displayed iv Press GC and Press Setup v Press Total vi Press Span vii Turn the front panel knob until the Total reading in the display window reads 0 0 ppm 13 Close regulator and remove tubing from the zero gas cylinder 14 Change the Modes display setup back to Peak 15 Press Main More Menu until GC is displayed and Press GC Press Auto 16 On the front panel of the Labview program click the Manually control and close the manually control Manufacturer Contact Information VIG Industries Inc 4051 East La Palma Ave Suite C Anaheim California 92807 1751 Toll Free in the US 800 862 7844 Phone 714 632 8200 Fax 714 632 8201 Email Service vigindustries com Reference Manual of Model 200 Methane Non Methane Total Hydrocarbon Analyzer 215 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix L Version 1 3 09 13 06 Page 5 of 5 Maintenance Calibration Record Sheet for VIG 200 Date of Calibration Calibrated by Time Items Unit Notes Ambien
40. f 0 to 500 ppb 204 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix K Version 1 3 09 13 06 Page 2 of 8 API 101E H2S analyzers 965 and 966 y 1 0493x 8 5801 R 0 9999 API966 300 API965 Linear API965 Linear API966 y 0 996x 8 2284 R 0 9999 0 100 200 300 400 500 600 Cal_gas ppb Figure K 1 API 101E linearity results For 965 the slope of the regression line was 1 05 with an intercept of 8 58 and r value of 0 9999 For 965 the slope of the regression line was 0 996 with an intercept of 8 23 and r value of 0 9999 Over the range of concentrations tested 0 to 500 ppb H2S the API 101E demonstrated a high degree of linearity The response time of the analyzer to step changes in gas concentrations was tested Figure K 2 The analyzer was challenged with two H2S calibration span gases 44 ppb and 93 ppb H2S respectively in a No balance 2 accuracy Matheson Gas Products Inc Montgomeryville PA Response time was determined from the amount of time required for the Model 101E to reach 95 of the change in response during the zero air to 44 ppb H2S span gas 0 to 93 ppb and 44 to 93 ppb shown in Figure K 2 The response time was approximately 75 seconds Start Up After electrical and pneumatic connections are made turn on the power switch on the front panel The exhaust and PMT cooler fan
41. f increases in particle mass collected on a filter attached to the element Changes in the element s resonant frequency are sampled electronically in quasi real time providing both continuous and time averaged measures of mass accumulation that are directly proportional to instantaneous and time averaged mass concentrations in air respectively The device operates at an industry standard volume controlled flow rate of 16 7 L min so that it can be outfitted with a variety of commercially available pre separator inlets suitable for measuring PM The unit is fitted with a PM jo inlet head in this application Switching Instrument On Off Since the airflow is maintained at a constant volume corrected for local temperature and barometric pressure the operation of this monitor requires that the temperature and pressure sensors are connected for proper temperature and pressure readings and flow corrections Turn On Unit Supply power to the instrument by plugging in the power cord to 120 VAC and pressing the power button on the front panel of the TEOM control unit The main screen four line display will soon appear after showing the name of the instrument Turn on the pump to draw the sample streams by plugging in the power cord to 120 VAC The monitor waits at least 30 min after being powered up to compute the first mass concentration data Turn Off Unit Press the power button on the front panel of the TEOM control unit The four line display
42. f it and the back of the hood as possible Set temperature to 410 C using temperature dial on control unit Allow approximately one hour for unit to reach pre set temperature digestor is at pre set temperature when lights located on front of the digestor are flashing After completing sample and reagent addition to the digestion tubes carefully elevate the rack by grasping the handles provided and place in position on the digestor unit As the rack is lifted each tube will rest on its top rim Carefully lower the rack so that each tube enters its respective hole and bottoms in the base of the unit Place end plates on the flask rack These plates completely close the ends of the flask rack and allow the temperature of the upper portion of the tube to elevate providing for proper refluxing and condensation of the sulfuric acid A certain amount of foaming will occur at the beginning of the digestion The refluxing of the acid washes any carbonized material back into the digestion solution Zf foaming is uncontrollable reduce digestion temperature and or sample size Digest samples at 410 C until the acid in the flask is clear and fumes are contained within the digestion tube and not thick in appearance A time of two hours is minimal for complete digestion The clearness of the acid color and absence of particulate matter and fume density are indicators for completeness of digestion After digestion is complete remove f
43. g Appendix P Version 1 3 09 13 06 Page 2 of 2 11257 Coloma Road Gold River CA 95670 USA Phone 800 876 1172 Fax 916 638 3270 236 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix Q Version 1 3 09 13 06 Page of 1 Appendix Q SOP for Use of Rotem RSC 2 Scale System Routine Check 1 Check scales for clearance between scales and litter Optimum separation is 1 inch Routine Download 8 9 10 1 Start up the computer notebook with USB Serial Adapter connected 2 Hookup Serial Communications between notebook and Rotem Serial Cable 3 4 As soon as software establishes communications with the Rotem Scales the main screen Start up Rotem Scalenet Application will appear on the computer screen displaying the current growth day and if the scales are in operation On the application s pull down menu click on Setup Farm Name to change to the current house name i e Tyson 1 5 This changes only the software data reference not the data retained in the actual scale s database On the application s pull down menu click on Setup Set Flock Number to verify current Project Flock Designation Number i e 2 This changes only the software data reference not the data retained in the actual scales database Click on Collect icon button to start download of history file for the current flock Sav
44. g the enclosed balance Place sample into oven at 103 105 C for 24 hours Remove dishes from oven and cool to room temperature in a desiccator which takes about 2 3 hours 7 Measure and record final weight of samples and dishes using the enclosed balance Nm BB U Testing Solids ALWAYS handle dishes with gloves on 1 Measure and record the weight of each measuring dish using the enclosed balance 2 Use about gram of sample for each dish 3 Measure and record the weight of each sample and dish using the enclosed balance 240 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix S Version 1 3 09 13 06 Page 3 of 13 4 Place sample in 103 105 C oven for 24 hours 5 Remove samples from oven and cool to room temperature in a desiccator which takes about 2 3 hours 6 Measure and record the final weight of the sample and dish using the enclosed balance 7 Repeat steps 4 6 until the weight is within 4 of the last drying cycle Calculations initial weight final weight 100 total solids 1 initial weight Moisture 100 TS Ammonia Rapidstill II Standard Procedure Standard Method 4500 NH B Preliminary Distillation Step amp 4500 NH C Titrimetric Method Safety Precautions When mixing acids use the fume hood with the front pulled down Wear a lab coat safety glasses and gloves Mix small amounts of acid into la
45. gt can gain direct access to the Set Storage Screen from any screen The first eight lines of the Set Storage Screen contain the titles of the variables currently being stored in the data logger Program Register Codes listed in Appendix A of operating manual The Interval variable defines the time in seconds between successive writings of data to the circular buffer Table M 1 Some popular program register codes Code Variable Units 008 Mass concentration ug m 009 Total mass ug 035 Pressure drop 039 Current main flow L min 040 Current auxiliary flow L min 041 Status condition code 057 30 min average mass concentration ug m 130 Current ambient temperature C 131 Current ambient pressure atm Setting Analog Outputs The instrument s three analog output channels are accessible from the identical 15 pin connectors on the front and back panels of the control unit Details regarding the pin assignments and voltage VDC of outputs are given in the operating manual Section 5 1 To bring up the Set Analog Output Screen press lt A O gt or select Set Analog Output from Menu Screen or press 04 lt Enter gt Setting Sampling Average Time The lt Step Screen gt key toggles the instrument to Set Hardware then press lt Enter gt Press the Data Stop to start stetting the sampling average time 219 A Quality Assurance Project Plan for Monitoring Gaseous and P
46. handle Turn it 45 to the left allowing water to flow through condenser 2 Depress Boiler Water switch and fill to the top line on the flask see 10 under Distillation Procedure 3 Turn on the Boiler Heater by pressing the button When water is at a steady boil the unit is ready for operation 5 Make sure the NaOH inlet tube located on the left side of the Rapidstill is in a 100 mL graduated cylinder that is full of NaOH D Manure Dilution 20 1 water to manure ratio serial dilution 242 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix S Version 1 3 09 13 06 Page 5 of 13 1 Add 25 mL distilled water to a 50 mL volumetric flask 2 Mix liquid manure for 1 minute at low speed with a drill mixer 3 With a carefully broken 10 mL pipet fill the flask up to the line with manure sample to ensure there is exactly 25 mL of manure sample 4 Pour the contents of the 50 mL volumetric flask into a 500 mL volumetric flask 5 Dilute the sample to 500 mL with distilled water Sample Preparation 75 mL samples 10 1 water to dilute manure ratio 1 Mix liquid manure for 1 minute at low speed with drill mixer 2 Extract a 6 815 mL dilute manure sample place in a digestion tube 3 Add 68 15 mL distilled water by filling a 50 mL volumetric flask adding that to the digestion tube and then adding the rest 18 15 mL with a 10 mL pipet 4 Add 3 75 mL borat
47. ibrated with a calibrated portable pressure sensor at various span pressures WE100 barometric pressure transducer specifications Output 4 20mA Range 800 1100 millibars Accuracy 1 of full scale Operating Voltage 10 36VDC Current Draw Same as sensor output Warm Up Time 3 seconds minimum Operating Temperature 40 to 55 C Calibration e Calibrate the sensor every year or at a different frequency if specified by the QAPP for a particular project e Calibrate the sensor system sensor and data logger against a Fortin barometer e Follow the documented procedures for reading and adjusting the Fortin barometer e Read instantaneous measurements from data logger sensor system at the time of each Fortin barometer reading e Conduct six paired measurements over a two day period the average mean of their respective readings is taken out and if they match then the unit is calibrated e Record the results of the calibration on the sensor lab notebook form e Ifthe mean difference between the Fortin barometer and sensor data logger measurement is less than or equal to 4 hPa then the sensor passes calibration and the drift will be back corrected e Ifthe mean difference is gt 42 hPa the barometer needs to be adjusted against a Fortin barometer Manufacturer Contact Information Global Water Instrumentation Inc 235 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housin
48. in the sample stream to SO using a high temperature catalytic oxidation The chemical process is 2H S 30 gt 2H 0 2S0 The physical principle of the M101E s measurement method relies on the fluorescence that occurs when Sulfur dioxide SOz is excited by ultraviolet light with wavelengths in the range of 190 nm 230 nm This reaction is a two step process The first stage occurs when SO molecules are struck by ultraviolet photons of the appropriate wavelength 19 nm 230 nm The SO retains some excess energy that causes one of the electrons of the SO2 molecule to move to a higher energy orbital state In the case of the Model 101E a band pass filter between the source of the UV light and the affected gas limits the wavelength of the UV light to approximately 214 nanometers nm SO AV 1am 2 S02 The second stage of this reaction occurs after the SO reaches its excited state SO Because the system will seek the lowest available stable energy state the SO2 molecule quickly returns and the wavelength of this fluoresced light is also in the ultraviolet band but at a longer lower energy wavelength centered at 330 nm SO SO AV330nm The linearity of the 101E was checked Figure K 1 shows the linearity results for the two analyzer SN 965 and 966 accuracy checks For each check a linear regression was calculated from API 101E response versus the nominal H2S gas standard concentration over the range o
49. ing New Calibration Yes IN o Zero air applied H2S applied Finish calibration Yes IN o Time switch off Connect analyzer back to sampling system Notes 211 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix L Version 1 3 09 13 06 Page of 5 Appendix L SOP of VIG Model 200 Methane Non Methane Total Hydrocarbon Analyzer Introduction The methane non methane total hydrocarbon analyzer VIG 200 measures concentrations of a wide variety of hydrocarbons in gas mixtures and in air and can separate the methane component from the non methane component by using a GC column This makes VIG 200 ideal for stack monitoring laboratory or process control use The model 200 uses column technology to separate methane and non methane from total hydrocarbons THC When a small sample is fed though a column the column separates each components of the THC one at a time lightest first and heaviest last The lightest hydrocarbon is methane A small sample is sent through the column waits for the methane to come out and measures the methane the rest of the sample is sent backward through the column is measured again and is called non methane The measurement is achieved by the use of a FID flame ionization detector coupled with a very precise amplifier The FID injects the gas into a hydrogen based flame The flame he
50. ion data Turn Off Unit Press the power button on the front panel of the TEOM control unit The four line display becomes blank Turn off the vacuum pump by pressing the power button and disconnect the control unit from 120 VAC by unplugging the power cord Status Line on Main Screen Whenever a status code other then OK is shown on the display the instrument automatically turns on the Check Status light on the front of the control unit The information displayed on the main screen includes status condition operating mode A O 1 mode RS 232 mode protection and time For example the screen will show the following line at time XX XX OK 4 51 NU XxX XX 217 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix M Version 1 3 09 13 06 Page 2 of 6 Status condition OK Normal operation No frequency signal Temperature s outside of operational bounds Flow s outside of operational bounds Filter nearing capacity exchange filter x MHZ Operating mode 1 Temperature flow stabilization 2 Begin TM computation 3 TM computed begin MR MC computation 4 Normal operating mode S Setup mode X Stop all mode A O 1 mode Blank Analog output 1 normal definition Analog output 1 used for status watch XX Filter loading percent RS 232 mode N None P Print online R R amp P protocol A AK protocol G German network protocol S Storage to p
51. iquid manure thoroughly in the fume hood with the front panel pulled down e Do not immerse the pH electrode to the bottom of beaker Equipment e pH meter e 100 ml glass or plastic beakers e Glass stirring rods e Distilled water Calibration Calibrate the pH meter using standard buffer solutions pH 4 0 and 7 0 if necessary Procedure for Liquid Manure EPA SW 846 Method 9040 U S Environmental Protection Agency 1986 Test methods for evaluating solid waste USEPA report SW 846 e Thoroughly mix liquid manure manure should be at room temperature e Pour 25 ml of liquid manure into a beaker e Immerse pH electrode in liquid manure e Record pH value when the meter has stabilized e Repeat above steps for replicates Procedure for Semi Liquid Manure Wolf N 2003 Determination of manure pH p 48 49 Jn Peters et al eds Recommended Methods of Manure Analysis University of Wisconsin System http ecommerce uwex edu pdfs A3769 PDF accessed on May 23 2006 e Thoroughly mix semi liquid manure manure should be at room temperature e Place 20 ml of semi solid manure into a beaker e Add 40 ml of distilled water e Stir and mix it well with a glass rod e Allow for 30 minutes to settle solution e Measure pH by immersing electrode into the supernatant solution e Record pH value when the meter has stabilized e Repeat above steps for replicates and report the result as pH water 1 2 239 A Quality Assurance Project Pl
52. is Project Clean air inlet head Upon filter exchange Twice a week Replace TEOM filter cartridge 70 load Weekly Exchange in line filters 6 mon or when loaded When loaded Leak test Annually Every flock Analog board calibration Annually Every 6 months The TEOM PM air inlet requires regular maintenance in livestock barns The TEOM inlet itself should be cleaned weekly In order to do this one must first push the lt Data Stop gt button on the keypad Remove the dirty TEOM inlet and replace it with a clean inlet The dirty PM2 5 inlet will be taken back to work station for cleaning and ready for next inlet swap For cleaning the dirty inlet please see Manual Appendix G Inlet Maintenance Grease as needed to maintain a slight layer of silicon grease Carefully replace the inlet unit and return to the control unit of the TEOM system and press the lt Data Stop gt button once more This will prevent skewed data from being recorded It will automatically begin recording in approximately 30 min Other Settings Flow rate through sample inlet 16 7 L min 1 m hr Main flow rate 1 L min Temperature of sample stream 50 C Particulate concentration lt 5 ug m to several g m Standard Conditions 1 atm pressure 20 C Spare Parts and Consumables Slow blow 2A 250 V fuse 2A amp 250 V in line fuse P N 04003419 Box of 20 TEOM filter cartridges TX40 media P N 57 000397 0020 Large bypass in line filter P N 57 002758 Flow controller
53. l prevent accidental damage from occurring to the sample filter cartridge when exposed to the high pressure drop vacuum in the sample line created by the leak check Flow rates should indicate less than 0 15 L min for the main flow and less than 0 65 L min for the auxiliary flow with the end of the sample line closed if not systematically check plumbing for connector leaks Mass Calibration Verification The mass transducer is permanently calibrated and never requires recalibration under normal use However the mass measurement accuracy of the instrument may be verified following procedures in the operating manual R amp P offers a mass calibration verification kit to help perform this procedure Maintenance The lifetime of a TEOM filter cartridge depends on the nature and concentration of the particulate sampled and the main flow rate setting 1 2 or 3 L min The filter must be exchanged when the filter loading value as shown on the status line of the main menu 220 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix M Version 1 3 09 13 06 Page 5 of 6 approaches 70 At a flow rate of 3 L min 100 filter loading generally corresponds to a total mass accumulation of approximately 3 to 5 mg of particulate Filter lifetime at a main flow rate of 3 L min is generally 21 days at an average PMio concentration of 50 ug m Flow splitter adapters for 1 and 2 L min opera
54. lask rack containing tubes from digestor and place on heat resistant pad Wear heat protective gloves Turn heat settings on the digestion block back to zero and turn off power if there is no more digestion runs for the day With the heat protective gloves carefully remove manifolds from the digestion tubs Caution The manifolds may leak condensed acid and the flask may emit fumes Carry the manifolds to a sink Clean manifolds and racks with water and running a lot of water down the drain to dilute the acid After the tubes containing the digest have cooled to room temperature slowly add 75 mL of deionized water to each tube swirling to mix There may be some fume emissions from the tubes so be careful Dilution must be made before a cake is formed from precipitated salts but not before the digest is cool enough to contain the exothermic reaction Contents in the tubes might turn orange or red orange This is common and will not affect results This distillation should be performed after water is added If the tubes cannot be distilled gently vortex the tubes cover them with aluminum foil and place in the fume hood 248 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix S Version 1 3 09 13 06 Page 11 of 13 Setup of the Rapidstill Distillation Unit Add 70 mL of 4 Boric acid to each of the receiving flasks Add five drops of Indicator solution
55. lps to break the electrons from their orbit along with a high voltage surrounding the flame The voltage forces the electrons to jump from the flame on to the electrode giving a measurable medium The amplifier takes this information and makes it linear and in scale Installation Ho Fuel zero gas N2 Carrier Cal gas Compressor Zero Air Fuel Hydrogen 100 Ultra High Purity Grade 18 psi Combustion Air Hydrocarbon free oil free dry air 18 psi From Zero Air Generator Zero Air Zero air Ultra High Purity Grade 9 10 psi Compressed air Oil free dry Compressed air 55 psi From Air Compressor Calibration gas Methane Propane mixture balanced in air 9 10 psi Carrier gas Nitrogen Ultra High Purity Grade 30 psi 212 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix L Version 1 3 09 13 06 Page 2 of 5 Start Up Connect compressed air to the unit and set the incoming pressure to 55 psi Press Power switch Connect nitrogen to the unit at the carrier gas port and set the carrier gas pressure on the front panel to 22 psi Press any key to continue Find the Oven key If no push the More Set the desired oven temperature to 275 F by turning front panel knob Connect all required gases to the proper ports on the rear panel Go the Burner 1 and Burner 2 Press the Ignite Make sure that Auto i
56. lugging the power cord Status Line on Main Screen Whenever a status code other then OK is shown on the display the instrument automatically turns on the Check Status light on the front of the control unit The information displayed on the main screen includes status condition operating mode A O 1 mode RS 232 mode protection and time For example the screen will show the following line at time XX XX OK 4 51 NU XX XX 229 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Status Condition OK Normal operation M No frequency signal T Temperature s outside of operational bounds F Flow s outside of operational bounds X Filter nearing capacity exchange filter Operating Mode 1 Temperature flow stabilization 2 Begin TM computation 3 TM computed begin MR MC computation 4 Normal operating mode S Setup mode X Stop all mode A O 1 Mode Blank Analog output 1 normal definition Analog output used for status watch XX Filter loading percent RS 232 Mode N None P Print online R R amp P protocol A AK protocol G German network protocol S Storage to printer F Fast storage output Protection U Unlocked L Low lock H High lock XX XX Current time 24 hr format Using the Keypads and Software User should refer to Section 4 5 of the operating manual and become familiar with the incorporated menu driven software and keypads Filter
57. lutions that are compatible with glass Examine all plumbing and steam lines ensuring that the tubing is intact Examine all tubing connections Examine the connection stopper for deterioration or cracking Total Kjeldahl Nitrogen Standard Procedure Standard Method 4500 No z D and C Safety Precautions 1 2 3 4 Equipment should not be used until adequate training has been obtained and the equipment manuals have been read Read the MSDS for sulfuric acid sodium hydroxide hydrochloric acid and boric acid There are strong acids and bases both heated being used Wear long pants lab coat and close toed shoes Safety glasses and various types of gloves should also be worn This analysis should be conducted when there are other personnel in the area Reagents Required Note Unless otherwise stated prepare solutions in deionized water Label all reagents with reagent name date of mixing and your initials Concentrated Sulfuric Acid 90 98 1 Store in the cabinet labeled Corrosives underneath the fume hood 245 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix S Version 1 3 09 13 06 Page 8 of 13 2 When using place the bottle of acid in a spill tray Fishertab Kjeldahl Tablets 1 Contains 15 mg selenium in 1 5 gm K2S04 Indicator Solution 1 Measure 225 mg of methyl red with a weigh boat using the enclosed balance 2
58. means ADM Average of valid daily means It can be obtained over a month a season or a year Overall test mean Average of all valid data over the entire test 203 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix K Version 1 3 09 13 06 Page 1 of 8 Appendix K SOP of Model 101E UV Fluorescence H S Analyzer Introduction The M101E UV Fluorescence H2S Analyzer is a microprocessor controlled analyzer that determines the concentration of hydrogen sulfide H2S in a sample gas drawn through the instrument It requires that sample and calibration gases be supplied at ambient atmospheric pressure in order to establish a constant gas flow through the sample chamber The H2S in the sample gas is converted into SO which is then exposed to ultraviolet light causing the SO2 to become excited SO2 As these SO2 molecules decay back into SO2 they fluoresce The instrument measures the amount of fluorescence to determine the amount of SOz present in the sample chamber and by inference the amount of H2S present in the sample gas The M101E H3S analyzer is basically a SO analyzer with an H2S SO conversion stage inserted into the gas stream before the sample gas enters the sample chamber The H2S to SO converter receives sample gas from which the SO has been removed by a scrubber Once the naturally occurring SO is removed from the sample gas the special converter changes the H2S
59. n 0 65 L min for the auxiliary flow with the end of the sample line closed if not systematically check plumbing for connector leaks Mass Calibration Verification The mass transducer is permanently calibrated and never requires recalibration under normal use However the mass measurement accuracy of the instrument may be verified following procedures in the operating manual R amp P offers a mass calibration verification kit to help perform this procedure Maintenance The lifetime of a TEOM filter cartridge depends on the nature and concentration of the particulate sampled and the main flow rate setting 1 2 or 3 L min The filter must be exchanged when the filter loading value as shown on the status line of the main menu approaches 70 At a flow rate of 3 L min 100 filter loading generally corresponds to a total 232 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix O Version 1 3 09 13 06 Page 5 of 6 mass accumulation of approximately 3 to 5 mg of particulate Filter lifetime at a main flow rate of 3 L min is generally 21 days at an average PMjo concentration of 50 ug m Flow splitter adapters for 1 and 2 L min operation are available for use in areas with higher particulate concentrations The factory recommended schedule of periodic maintenance and the schedule to be adopted for use in livestock barns are as follows Maintenance item Factory Th
60. n a 250 volumetric flask add a stir bar and mix well on a stir plate for approximately 5 minutes aA BW NN Indicating Boric Acid Solution Prepare Monthly 1 Measure 20g H3BO3 in a weigh boat and add to a 1000 mL volumetric flask 2 Add approximately 700 mL distilled water to the flask and dissolve the H3BO3 with a stir bar and a stir plate it may take 10 to 15 minutes 3 Add 10 mL mixed indicator solution to the flask 4 Dilute to 1 L with distilled water add a stir bar and mix well on a stir plate for approximately 5 minutes Standard Sulfuric Acid Titrant 0 02N 1 Measure 2 8 mL concentrated sulfuric acid with a 10 mL pipet 2 Dilute to 1 L with distilled water in a 1000 mL volumetric flask to make 0 1N sulfuric acid Add a stir bar and mix well on a stir plate for 5 to 10 minutes 3 Measure 200 mL 0 1N sulfuric acid with a 200 mL volumetric flask 4 Pour the 200 mL of 0 1N sulfuric acid in a 1000 mL and dilute to 1 L with distilled water 5 Add a stir bar and mix well on a stir plate for approximately 5 minutes Sodium Hydroxide 6N 1 Measure 240g NaOH in a large weigh boat 2 Dilute to 1 L with distilled water in a 1000 mL volumetric flask 3 Add a stir bar to the flask and mix well on a stir plate until all the solid is dissolved this could take up to 30 minutes Distillation Preparation 1 Turn on cooling water by going under the sink and looking way in the back In the top right corner there is a silver
61. nant is expressed as a mass flow per unit time e g g NH s It may be further refined on a per live weight per animal or per animal unit basis Consideration should be given to avoid expressing ER on a time interval that is shorter than the minimum sampling frequency used to obtain the data Thus for example if several minutes are required to acquire concentration data then it is unrealistic to report ER on a mass per second basis Emission Rates Calculations Emission rate ER is the difference in mass flow between all inlets and outlets for the contaminant of interest Mass flow of contaminant entering a building can be expressed as the background volumetric concentration C y multiplied by the incoming volumetric flow rate Qi Similarly the exhaust mass flow of contaminant is the product of exhaust volumetric concentration C ve multiplied by the exhaust volumetric flow rate Qe Note that volumetric flow rates Q and Qe are typically not equal since air density differences usually exist between inlet and outlets Mass flow rate of ventilation air Mj and Mo are equal Equation 5 expresses the general definition of emission rate E Q C Me Q C mi 5 In equation 5 the volumetric flow rates Q and Q may be either actual flow rate or standardized flow rates either moist standard sm s or dry standard dsm s Moist 201 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter
62. nd twice with distilled water Place in dry oven or on a rack so they will dry 3 Install a clean empty distillation tube 4 Remove caustic NaOH supply tubing from vessel containing NaOH and place in vessel of purified water 5 Turn on cooling water and fill boiler Turn on boiler heater 6 Depress NaOH Addition until all of the caustic NaOH has been purged out of the system and into the digestion tube 7 Remove the digestion tube and dispose of liquid properly 8 Replace digestion tube and fill half full by depressing NaOH Addition 9 When water is at a rolling boil turn distillation timer to 10 minutes 10 At the end of the distillation cycle turn off the boiler heater and cooling water Remove the digestion tube and dispose of liquid properly 11 Clean exterior surfaces of machine using a soft cloth and a mild soap and water solution Weekly 1 Examine boiler for any build up of deposits Accumulated deposits may be removed by a dilute solution of hydrochloric acid or by using commercially available descaling solutions that are compatible with glass 2 Examine all plumbing and steam lines ensuring that the tubing is intact Examine all tubing connections Examine the connection stopper for deterioration or cracking 251 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix T Version 1 3 09 13 06 Page 1 of 7 Appendix T Maintenance Checklists So
63. not depress the Boiler Water Switch to refill the boiler 4 Turn the Boiler Heater on 5 Depress NaOH Addition button to add 5 mL 6N NaOH solution to the digestion tube according to the graduated cylinder the NaOH is in 6 When the water reaches boiling begin distillation by setting the timer knob to the desired distilling time 7 Distill for 20 minutes or until froth reaches the condenser arm Collect distillate in the 500 mL Erlenmeyer flask 9 Lower the flask so that the end of the outlet tube is free of contact with the liquid and continue distillation during the last minute or two to cleanse condenser and outlet tube 10 When distillation has completed the unit will return to standby position with the boiler heater still in operation DON T let the boiler stay on switch the Boiler Heater button to the off position 11 Once the boiler is cool depress Boiler Water switch to refill the boiler after each run If doing additional distillations turn the boiler heater back on and the machine is ready when boiling resumes 12 Run samples in duplicate if numbers are not in agreement up to 10 then run a third sample Co Titration 1 Fill a 50 mL buret with 0 02N H2SO4 and add a stirbar to the 500 mL flask containing the distillate 2 Titrate drop wise the distillate in the flask 3 Titrate ammonia in distillate with standard sulfuric acid titrant until indicator in distillate turns pale lavender Calculation
64. rger amounts of water not reverse Strong acids will burn holes in your clothing and will burn your skin Reagents Required Label all reagents with reagent name date of mixing and your initials Sodium Tetraborate 0 025M 1 Measure 9 5 g NayB4O7 10 H20 in a weigh boat 2 Dilute to 1 L with distilled water in a 1000 ml volumetric flask 3 Add a stir bar to the flask and mix on a stir plate until all the solid is dissolved it may take 5 to 10 minutes Borate Buffer Solution 1 Using a 500 mL volumetric flask measure 500 mL of 0 025M sodium tetraborate and put in a 1000 mL volumetric flask 2 Measure 88 mL 0 1N NaOH by filling a 50 ml volumetric flask with 0 1N NaOH then measuring 38 mL with a 10mL pipet Put into the 1000 mL volumetric flask along with the sodium tetraborate 3 Dilute to 1 L with distilled water 4 Add a stir bar and mix well on a stir plate for approximately 5 minutes 241 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix S Version 1 3 09 13 06 Page 4 of 13 Mixed Indicator Solution Prepare Monthly 1 Measure 20 mg methyl red indicator in a weigh boat and put it into a 100 mL volumetric flask Dissolve in 100 mL 95 ethyl alcohol added to the flask Measure 100 mg methylene blue and put it into a 100 mL volumetric flask Dissolve in 100 mL 95 ethyl alcohol added to the flask Mix the two 100 mL flask together i
65. rinter F Fast storage output Protection U Unlocked L Low lock H High lock XX XX Current time 24 hr format Using the Keypads and Software User should refer to Section 4 5 of the operating manual and become familiar with the incorporated menu driven software and keypads Filter Storage and Exchange The measurements must be conducted with TEOM filter cartridges that are made of Teflon coated glass fiber filter paper Filters should be stored inside the sensor unit for easy access and 218 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix M Version 1 3 09 13 06 Page 3 of 6 to keep them dry and warm Do not handle new filters with your fingers Instead use the filter exchange tool and follow the procedures given in the operating manual Keep the sample pump running to facilitate filter exchange Use the two pockets on the right side of the mass transducer inside to store the next two new TEOM filters for pre conditioning and removal of excessive moisture build up prior to use System Operation after Power Failure The system resets itself when power is regained and enters the same mode as before All operating parameters are maintained in the system s battery backed up CMOS memory Setting Variables Stored in Data Logger The lt Step Screen gt key toggles the instrument between the View Storage Screen and Set Storage Screen or pressing 09 lt Enter
66. rticulate filter should be inspected often for signs of plugging or excess dirt It should be replaced according to the service interval in Table 9 1 of manual every week even without obvious signs of dirt Filters with 1 and 5 um pore size can clog up while retaining a clean look Handle the filter and the wetted surfaces of the filter housing with gloves and tweezers Do not touch any part of the housing filter element PTFE retaining ring glass cover or O ring with bare hands 1 Turn OFF the analyzer to prevent drawing debris into the sample line 2 Open the M101E s hinged front panel and unscrew the knurled retaining ring of the filter assembly 3 Carefully remove the retaining ring glass window PTFE O ring and filter element Replace the filter element carefully centering it in the bottom of the holder 5 Re install the PTFE O ring with the notches facing up the glass cover then screw on the hold down ring and hand tighten the assembly Inspect the visible seal between the edge of the glass window and the o ring to assure proper gas tightness 6 Re start the analyzer amp 207 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix K Version 1 3 09 13 06 Page 5 of 8 Calibration Procedure Step 1 Set verify the analog output reporting range of the 101E SAMPLE RANGE 500 0 PPB lt TST TST gt CAL SAMPLE ENTER SETUP PASS 818
67. s flashing If not press the Auto key After the alarm lamp on the front panel goes out check and adjust the fuel combustion air carrier gas and compressed air to the desired value in the Table 1 on the VIG 200 user manual Warming Up the Monitor A period of 6 hours is suggested to warm up the analyzer before a calibration task is started This will reduce the time required for calibration Calibration Before calibration make sure the display and output values are the peak value Press Main More Menu until Modes is displayed Press Modes Press Peak for Display and Output On the front panel of the Labview program click the Manually control and put all four SVs solenoid valves on OFF Zero Gas Calibration 1 On NN Close regulator valve on the zero gas cylinder 2 3 Open main valve on zero gas cylinder Insert the 1 4 ID tubing from the gas cylinder into the inlet of the flow meter mounting on the side of instruments rack Adjust regulator valve until vent airflow is about 5 L min read from bottom of ball of the vent monitoring flow meter This provides a little extra zero air to the analyzers and keeps the pressure inside the manifold close to the atmospheric pressure Flow Zero gas for 2 minutes Press Main More Menu until GC is displayed and Press GC Press Single Wait for about 3 minutes
68. s should start The display should immediately display a single horizontal dash in the upper left corner of the display This will last approximately 30 seconds while the CPU loads the operating system Once the CPU has completed this activity it will begin loading the analyzer firmware and configuration data During this process a string of messages will appear on the analyzer s front panel display SAMPLE SYSTEM RESET SO2 X XXX TEST CLR SETUP Press CLR to clear initial warning messages 205 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix K Version 1 3 09 13 06 Page 3 of 8 The fault red LED will be blinking Then push the CLR key to clear the SYSTEM RESET message The M101E requires about 60 minutes warm up time before reliable H2S measurements can be taken If the warning messages persist after 60 minutes investigate their cause using the troubleshooting guidelines in Chapter 11 of the 101E User s Manual API 101E Response Time 50 45 re ore 9 Dee ieee ee ie Al 100 40 eee i pele el ae ey ee _ API 965 API 965 0 to 44 35 E eae 80 zL 030 2S 2S o Q 2 t 60 2 Hee ee aati iuieatiatiataatian ikatiatatiatiad Wkamatameda r E n D A Ir 20 S ee See een ee ieee ae es ee eee ee es 2 40 15 ee L Ss es a A ena A a p 20 Wife ease
69. stabilization of the analyzer readings Only the concentration readings at the end of the sampling cycle will be considered as valid measurements Gas Concentration Interpolation When there is only 1 valid reading per location during a 360 sec sampling cycle To account for potential concentration changes during this period linear interpolation between the two adjacent readings of the same location will be performed to determine the concentrations in between Complete Data Sets To avoid errors introduced into calculated average values as a result of partial data days that result in biased time weights only complete data days CDD that include over 75 valid data should be used for calculating average daily means ADM Similarly hourly averages should be reported only if over 75 of the data during that hour is valid and monthly averages are reported only if over 75 of the days are valid Quantity Names and Terminology Reading One data point read by LabVIEW It is not necessarily recorded by LabVIEW Data record A single data point recorded in data files by the LabView data acquisition program A data record consists of an average of 1 s readings collected over the data recording interval LabVIEW takes a reading every second and writes a data file LabVIEW also records an average of 30 readings every 30 s and records that in another data file Daily mean Average of all valid and interpolated data records during CDD Average daily
70. t Total reading CH reading Propane reading Propane applied Total adjustment CH adjustment Propane adjustment Total reading CH reading Propane reading Finish calibration Yes INo Time switch off Connect analyzer back to sampling system Note 257 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix T Version 1 3 09 13 06 Page 7 of 7 O Check all solenoid valves On Off SV NO1 NC1 NO2 NC2 NO3 NC3 NO4 NC4 O Fan current switch check Fan SW1 SW2 SW3 SW4 T1 T2 T3 T4 T5 T6 T7 T8 T9 T10 O Check airflow rate at 20 of the exhaust fans randomly selected Flow rate cfm previous calibration Flow rate cfm Fan 0 04 0 08 0 12 0 16 0 04 0 08 0 12 0 16 Diff O Replace any critical spare parts utilized during the previous flock Part Description Note CO Colect copies of calibration and event log sheets from previous flock and submit to Lara Moody for archiving C Collect litter samples and deliver to ISU for analysis 258
71. t pressure mm Hg Sample airflow L min Zero air applied Cylinder P _ psi Total Reading CH Reading Propane Reading CH ppm applied Cylinder P _ psi CH Reading Propane ppm applied Cylinder P _ psi Propane Reading New Calibration Yes INo Zero air applied Total adjustment CH adjustment Propane adjustment Total Reading CH Reading Propane Reading CH applied Total adjustment CH adjustment Propane adjustment Total Reading CH Reading Propane Reading Propane applied Total adjustment CH adjustment Propane adjustment Total Reading CH Reading Propane Reading Finish calibration Time switch off Connect analyzer back to sampling system Notes 216 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix M Version 1 3 09 13 06 Page 1 of 6 Appendix M SOP of Tapered Element Oscillating Microbalance TEOM for TSP Measurement Introduction The TEOM instrument TEOM 1400 Ambient Particulate PMjo Monitor is a continuous PM monitoring device designated by USEPA as an equivalent method EPA Designation No EQPM 1090 079 for PMio lt 10 um aerodynamic diameter The acronym TEOM stands for Tapered Element Oscillating Microbalance an in
72. t test a range of sample weights from 0 to 1 00 gm Use at least two blank tubes without sample in a run of 25 digestion tubes The blanks should contain all reagents receiving the same treatment as the samples There will be an indicator color change in the blank If not something has gone wrong Also if new Boric acid or HCl is made during the day blanks are needed for the new reagents Blanks must be run everyday Samples can be weighed into digestion tubes one or more days before digesting if tubes are covered so they do not become contaminated 247 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing 7 Appendix S Version 1 3 09 13 06 Page 10 of 13 0 05 0 06 gm urea or 0 2 gm tryptophan can be used to check the system for complete digestion without loss of nitrogen Digestion Procedure 1 10 11 Carefully insert the digestion tubes into the holes provided in the flask rack The flask rack can be used for convenient handling and transportation of the digestion tubes Add 4 Fishertab Kjeldahl tablets to each digestion tube Carefully add 20 mL of concentrated sulfuric acid using a 10 mL pipet to each tube prior to placing the tubes in the digestion block Turn the main power switch located on front of controller to ON position The digestion block must be totally contained in a chemical fume hood and must have as much space between the back o
73. tion are available for use in areas with higher particulate concentrations The factory recommended schedule of periodic maintenance and the schedule to be adopted for use in livestock barns are as follows Maintenance item Factory This Project Clean air inlet head Upon filter exchange Twice a Week Replace TEOM filter cartridge 70 load Weekly Exchange in line filters 6 mon or when loaded When loaded Leak test Annually Every flock Analog board calibration Annually Every 6 months The TEOM TSP air inlet requires regular maintenance in livestock barns The TEOM inlet itself should be cleaned weekly In order to do this one must first push the lt Data Stop gt button on the keypad Remove the dirty TEOM inlet and replace it with a clean inlet The dirty TSP inlet will be taken back to work station for cleaning and ready for next inlet swap For cleaning the dirty inlet please see Manual Appendix G Inlet Maintenance Grease as needed to maintain a slight layer of silicon grease Carefully replace the inlet unit and return to the control unit of the TEOM system and press the lt Data Stop gt button once more This will prevent skewed data from being recorded It will automatically begin recording in approximately 30 min Other Settings Flow rate through sample inlet 16 7 L min 1 m hr Main flow rate 1 L min Temperature of sample stream 50 C Particulate concentration lt 5 ug m to several g m Standard conditions 1 atm pressur
74. um in the sample line created by the leak check Flow rates should indicate less than 0 15 L min for the main flow and less than 0 65 L min for the auxiliary flow with the end of the sample line closed if not systematically check plumbing for connector leaks Mass Calibration Verification The mass transducer is permanently calibrated and never requires recalibration under normal use However the mass measurement accuracy of the instrument may be verified following procedures in the operating manual R amp P offers a mass calibration verification kit to help perform this procedure Maintenance The lifetime of a TEOM filter cartridge depends on the nature and concentration of the particulate sampled and the main flow rate setting 1 2 or 3 L min The filter must be exchanged when the filter loading value as shown on the status line of the main menu 226 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix N Version 1 3 09 13 06 Page 5 of 6 approaches 70 At a flow rate of 3 L min 100 filter loading generally corresponds to a total mass accumulation of approximately 3 to 5 mg of particulate Filter lifetime at a main flow rate of 3 L min is generally 21 days at an average PMjo concentration of 50 ug m Flow splitter adapters for 1 and 2 L min operation are available for use in areas with higher particulate concentrations The factory recommended schedule of perio
75. utheastern Broiler Emissions Project Between Flock Requirements Checklist House I D Date Incoming Flock No Clean screens and replace filters on the gas sample line intakes Check air flow rate in sample lines and solenoid valves Replace TEOM filters and exchange TEOM inlet heads and 2 5 micron cut cyclones Leak test the TEOMs after all other maintenance has been performed on them OOOdd Instrument Calibration Verification INNOVA for NH API for H2S VIG for NMH Polytron Check all solenoid valves ON OFF Perform a system check on fan operational status and monitoring Check airflow rate at 20 of the exhaust fans randomly selected Replace any critical spare parts utilized during the previous flock Collect copies of calibration and event log sheets from previous flock and submit to Lara Moody for archiving Oo 00000 Collect litter samples and deliver to ISU for analysis 252 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Checklists of Needed Items for Between Flock Audits Appendix T Version 1 3 09 13 06 Page 2 of 7 Item Quantity In stock Inclined manometer red liquid Setra 264 sensor new Humidity sensor new and calibrator 24 VDC power supply Barometric sensor calibrated 4 20 mA HOBO cable 4 Channel HOBO logger Thermometer 8 to 50 C NIST Cooler 1 2 Gal BO N
76. x of 20 TEOM filter cartridges TX40 media P N 57 000397 0020 Large bypass in line filter P N 57 002758 Flow controller filter P N 30 003097 Manufacturer Contact Information Rupprecht amp Patashnick Co Inc 25 Corporate Circle Albany NY 12203 227 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Timothy Morphy Assistant Product Manager 518 452 0065 voice 518 452 0067 fax tmorphy rpco com email Peggy O Gormon Marketing Associate Phone ext 3229 email pogorman rpco2 com Reference TEOM 1400 Operation manual 228 Appendix N Version 1 3 09 13 06 Page 6 of 6 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix O Version 1 3 09 13 06 Page 1 of 6 Appendix O SOP of Tapered Element Oscillating Microbalance TEOM for PM Introduction The TEOM instrument TEOM 1400a Ambient Particulate PMjo Monitor is a continuous PM monitoring device designated by USEPA as an equivalent method EPA Designation No EQPM 1090 079 for PMio lt 10 um aerodynamic diameter The acronym TEOM stands for Tapered Element Oscillating Microbalance an inertial measurement technique that operates on changes in the resonant frequency of an oscillating element as a function of increases in particle mass collected on a filter attached to the element Changes in the element s resonant frequenc
77. y are sampled electronically in quasi real time providing both continuous and time averaged measures of mass accumulation that are directly proportional to instantaneous and time averaged mass concentrations in air respectively The device operates at an industry standard volume controlled flow rate of 16 7 L min so that it can be outfitted with a variety of commercially available pre separator inlets suitable for measuring PM The unit is fitted with a PM2 5 inlet head in this application Switching Instrument On Off Since the airflow is maintained at a constant volume corrected for local temperature and barometric pressure the operation of this monitor requires that the temperature and pressure sensors are connected for proper temperature and pressure readings and flow corrections Turn On Unit Supply power to the instrument by plugging in the power cord to 120 VAC and pressing the power button on the front panel of the TEOM control unit The main screen four line display will soon appear after showing the name of the instrument Turn on the pump to draw the sample streams by plugging in the power cord to 120 VAC The monitor waits at least 30 min after being powered up to compute the first mass concentration data Turn Off Unit Press the power button on the front panel of the TEOM control unit The four line display becomes blank Turn off the vacuum pump by pressing the power button and disconnect the control unit from 120 VAC by unp
78. ystem RSC 2 User Manual Scalenet Communications Program for RSC 2 User Manual 237 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix R Version 1 3 09 13 06 Page of 1 Appendix R SOP for Fan Current Switch Application Figure R 1 shows the induction operated current switch CS CR9321 CR Magnetics St Louis MO application in monitoring fan operational status The CS is used in a pigtail which acts as a short extension cord The pigtail is put in line between the fan motor and its AC power source Water proof electrical tape is wrapped extensively around the plug connections to protect them from water and dust The plugs are then locked together using multiple plastic ties as a deterrent to being separated When AC current is present the CS returns a user defined amount of DC current to the MAEMU To AC power source AC power source 5Vde gt CR9321 PNP Pigtail Current Switch Water proof electrical tape covering plugs Plastic ties locking plugs together Figure R 1 CS application 238 A Quality Assurance Project Plan for Monitoring Gaseous and Particulate Matter Emissions from Broiler Housing Appendix S Version 1 3 09 13 06 Page 1 of 13 Appendix S SOP of Litter Sample Analytical Methods Determination of Manure pH Safety notes e Use lab coats close toed shoes and hand gloves e Mix liquid or semi l

Download Pdf Manuals

image

Related Search

Related Contents

大腸菌タンパク質発現系の ゴールドスタンダード pET 発現システム活用  Emerson CF770AW01 Owner's Manual    Samsung Galaxy Note 8.0 Hướng dẫn sử dụng  交通事故防止技術 - クルマ社会を問い直す会  770-412 Transfer bench instructions R3.cdr  Guía del usuario del ordenador táctil serie B  

Copyright © All rights reserved.
Failed to retrieve file