1. Super PRP Centrifuge Centrifuge SELEX User s Manual 605 1 JIWOL RI CHOWOL EUP GWANGJU SI KYEONGGI DO 464 863 KOREA TEL 82 31 769 8811 TEL 82 31 798 8817 www superprp com MOOHAN Super PRP Centrifuge 1 Precautions v The instrument must be installed on a flat stable horizontal and solid surface v Before operating the instrument check if the rotor and the rotor lid are securely fastened v Do not move the instrument when it is not completely stopped v Operate with a stable amp appropriate power inlet v Do not place dangerous materials within 30cm distance around the instrument v Only use rotors from moohan Co Ltd with appropriate tubes for centrifugation v Do not use hazardous inflammable or radioactive materials as samples v When it is necessary to use toxic or radioactive materials or pathogenic micro organisms which belong to the Risk Group II of WHO Laboratory Bio safety Manual follows national regulations v Use the emergency door open function only when the door key on the control panel is not working v The sample must be loaded weight balanced way not to cause imbalance v The density of sample material must not exceed 1 2g ml at Max RPM or RCF v Any improper handling or any usage of un qualified accessories is not able to be protected 2 Technical Specifications 416G gt Max RPM 4 000 rom gt Max RCF 2 700 xg gt Max capacity 15ml x 16 gt Run time lt 99 mi
2. ated slowly Press SOFT START STOP button instead of START button to use this function 4 2 5 Door Press the Door button to open the door The door is not opened while the instrument is running 4 2 6 START Pressing the START button the operation is started to run with setting value 4 2 7 STOP Pressing the STOP the operation is stopped to run in any case 4 3 Operation Sequence 4 3 1 Connect the AC cord to the appropriate voltage 4 3 2 Turn on the power switch located at the right side of the unit Then the latest operating values are recalled with beeping sound 4 3 3 Open the door and check the rotor 4 3 4 Input the function values according to your preferences 4 3 5 Press the START button to start the operation 4 3 6 As soon as the set time is ended the centrifugation will be ceased according to the set deceleration level If you want to interrupt before the set time press the STOP button 4 4 Miscellaneous Functions 4 4 1 In any cases needed for emergency open you can use the emergency door open tool as long as the machine is idle Insert the emergency tool into the emergency door open hole at the left upside of the unit 4 4 2 Imbalance Alarm When the rotor is imbalanced and the vibration is severe the operation is urgently stopped with the error message on the display and the alarming sound MOOHAN Super PRP Centrifuge 5 Maintenance 5 1 Outer par
3. heck supply voltage Use AVR to provide proper voltage E 6 Over speed Stop running power off and on Tuning of the firmware Download E 7 Failure of control system Stop running power off and on Tuning of the firmware Download E 8 Imbalance Check amount of samples Check weight balances of samples E 9 RPM sensor recognition error Check attached place of RPM sensor Check connection of cable for RPM sensor Change RPM sensor MOOHAN
4. here is no new setting 4 1 7 STOP Uses to stop operation 4 1 8 SOFT START amp STOP Uses to accelerate or decelerate slowly If you press the arrow button longer than 3 seconds the numbers change very rapidly and the set up is achieved faster 4 2 Function Description 4 2 1 RPM Speed display unit 1 rom Speed setting unit 10 rom Speed accuracy lt 2 at maximum speed Press the RPM RCF button Press the arrow button to change input value 4 2 2 RCF Relative Centrifugal Force The conversion value from RPM to RCF is calculated automatically based on the maximum radius and revolution Press the RPM RCF button twice Press the arrow button to change input value MOOHAN Super PRP Centrifuge 4 2 3 Time Time display 99 min 59 sec 2 digit display Time is down counted after starting centrifugation Press the TIME button The display shows the previous time value and the minute LED and value blinks Press the arrow button to change the minute value 3 Press the TIME button again to set the second value then the second LED and value blinks automatically 4 Press the arrow button to change the second value After finishing the setting your input value is saved automatically in 2 3 seconds 4 2 4 Soft Start amp Stop For the sensitive samples not to disturb contents by fast acceleration and deceleration use SOFT START STOP Centrifuge is accelerated or deceler
5. n 59 sec or continuous gt Noise level lt 60 dB gt Acceleration time lt 20 sec gt Deceleration time lt 25 sec gt Imbalance Recognition Automatic gt Motor High torque AC induction motor gt Safety Lid lock gt Available Rotors 15mlx16 50mlx4 S gt Power amp Frequency 220V 50 60Hz gt Dimension WxDxH mm 377 x 438 x 251 gt Weight 22 Kg gt Rotor S Swing rotor MOOHAN Super PRP Centrifuge 3 Installation 3 1 Delivery Checklist 3 1 1 Main body lea 3 1 2 Power cable lea 3 1 3 Emergency door open tool lea 3 4 User s manual lea 3 1 5 Rotor Optional 3 2 Unpacking the instrument 3 2 1 Open the box and lift out the instrument carefully together with the safety padding 3 2 2 Remove the safety padding and vinyl wrap 3 2 3 Place the instrument on the flat space 3 3 Installation place 3 3 1 Install the machine at the solid and flat floor or table If you place the centrifuge at the Slope the axis of rotation is possibly changed because of the rotor weight 3 3 2 Install the machine about 30cm is departed from the wall because of the air circulation 3 3 3 Install the machine at the place with appropriate temperature and humidity It also has to be maintained with the proper temperature amp humidity 3 3 4 Install the machine at the place without any kinds of corrosive gases 3 4 Supply the power 3 4 1 416G model uses 110V or 220V Check proper voltage of your instr
6. t of instrument 5 1 1 Clean the outside of the machine with a dry soft cloth If necessary dip the cloth with neutral detergents and clean contaminated parts Keep completely dry after cleaning 5 1 2 Do not use any volatile chemicals such as alcohol benzene etc 5 1 3 If any rust appears clean with neutral detergents and dry it 5 2 Inner part of instrument 5 2 1 Keep dry inside the chamber after every use of the machine 5 2 2 Clean the shaft always for avoiding an imbalance during the rotation 5 2 3 If any part is contaminated clean with neutral detergents 5 3 Rotor 5 3 1 Clean the rotor if any parts are contaminated with samples 5 3 2 Keep dry after usage 5 4 Moving or shipping of instrument 5 4 1 If you need to move or ship the instrument be cautious to protect the shaft from any physical impact 5 4 2 Remove the rotor and fill inside the chamber with proper materials to keep the shaft on place 5 4 3 It is recommended to move or ship rotors separately if no original safety padding is available 6 Trouble Shooting Error Possible Causes Actions El Failure to reach to 200RPM within Check the RPM sensor 2sec Check the connection of a RPM connector or a wire E 2 When running the door is opened Check the door latch Stop running power off and on E 3 Over temperature of motor Check the motor temperature E 4 Low voltage Check supply voltage Use AVR to provide proper voltage E5 High voltage C
7. ument and connect to adequate power outlet 3 4 2 If the power input is more than 10 of the recommended voltage or fluctuating frequently it may affect some functions of the instrument It is advised to use AVR Automatic Voltage Regulator 3 4 3 If you want to use the instrument at other voltage range please contact us 3 5 On off and the door open 3 5 1 Turn on the instrument by pressing the switch on the right side of the machine 3 5 2 Press the Door button to open the door 3 5 3 Remove the protection materials inside the chamber 3 6 Rotor coupling and disassembling 3 6 1 Mount the rotor fitting with the motor shaft 3 6 2 Grasp the rotor with one hand and rotate the rotor nut in the direction of clockwise until it is tightly assembled 3 6 3 Position the plastic lid properly and press slightly down to close 3 6 4 The rotor lid is disclosed by lifting up the knob 3 6 5 To disassemble the rotor rotate the rotor nut in a counterclockwise direction 1 1 1 1 MOOHAN Super PRP Centrifuge 4 Operation 4 1 Control Panel o HOU CRESS 4 1 1 Display FND Shows the data of each function 4 1 2 Up amp down arrow Uses to change input data 4 1 3 Door Uses to open the door 4 1 4 RPM RCF Uses to switch the RPM RCF display 4 1 5 TIME Uses to set test time up to 99 min 59 sec 00 continuous 4 1 6 START Uses to start operation The latest setting value is recalled if t