Home

2290-5

image

Contents

1. the sensitivity of the Series 2290 Power Supplies Keithley SourceMeter SMU instruments can provide sensitivity as low as 0 1fA The following diagram shows how the Model 2290 PM 200 Protection Module protects the low voltage SourceMeter SMU instrument from high voltages when the DUT breaks down The leakage current of the protection module limits the measurement resolution to picoamp levels which is five decades more sensitive than the current measurement resolution of the Series 2290 Power Supplies Keithley can provide a complete high voltage test capability with the most current measurement sensitivity available anywhere Model 263xB System SourceMeter SMU Instrument configured as an ammeter Model 2290 PM 200 Protection Module Model 2290 5 High Voltage Power Supply Reverse breakdown testing of a high voltage diode using a Keithley SourceMeter SMU instrument to measure leakage current The Model 2290 PM 200 SMU Protection Module protects the SourceMeter SMU instrument from high voltage when the diode breaks down Easily Automate a High Voltage Test System The Series 2290 s IEEE 488 interface allows the creation of an automated high voltage test system Software drivers are supplied to simplify and accelerate test system development This further enhances safety as the high voltage can be controlled from a remote location KEITHLEY A Greater Measure of Confidence A Tektronix Company Series 2290 2290 5 Specificatio
2. 40 C typical PROTECTION Arc and short circuit protected programmable voltage and current limits and current trip 1 888 KEITHLEY wis only www keithley com MONITOR OUTPUTS OUTPUT SCALE 0 to 10V for OV to full scale CURRENT RATING 10mA maximum OUTPUT IMPEDANCE 1009 ACCURACY 0 2 of full scale with a 1MQ load minimum UPDATE RATE 87 5Hz EXTERNAL VOLTAGE SET INPUT SCALE 0 to 10V for 0 to full scale INPUT IMPEDANCE 1MQ ACCURACY 0 2 of full scale UPDATE RATE 87 5Hz OUTPUT SLEW RATE 5 to 95 under full load lt 3 seconds GENERAL INPUT POWER 75 watts INPUT VOLTAGE 90V 264VAC 47 to 63Hz REAR PANEL CONNECTORS Output high voltage connector SHV male Kings type 1764 1 or equivalent BNC Connector three Input voltage monitor current monitor GPIB connector 23 pin RS 232 9 pin D sub HIGH VOLTAGE SAFETY INTERLOCK Connector 3 pin press in connector 3M part number 37103 A165 O0E MB Pin 1 5V out 7OmA maximum out Pin 2 Input High voltage output enabled 5V DC High voltage output disabled OV DC or open connection Pin 3 Chassis ground INTERFACE PROTOCOL IEEE 488 1 OPERATING ENVIRONMENT 0 to 40 C non condensing DIMENSIONS 89mm high x 206mm wide x 406mm deep 3 5 in x 8 1 in x 16 in WEIGHT 3 7kg 8 pounds SAFETY Conformance to European Union low voltage directive WARRANTY One year WARM UP TIME One hour 1 The output voltage can be programmed to volt
3. RORA RD CLEVELAND OH 44139 1891 E 440 248 0400 E Fax 440 248 6168 WE 1 888 KEITHLEY www keithley com nt BENELUX FRANCE ITALY MALAYSIA SINGAPORE 31 40 267 5506 33 01 69 86 83 60 39 049 762 3950 60 4 643 9679 01 800 8255 2835 rw www keithley nl www keithley fr www keithley it www keithley com www keithley com sg A A BRAZIL GERMANY JAPAN MEXICO TAIWAN 55 11 4058 0229 49 89 84 93 07 40 81 120 441 046 52 55 5424 7907 886 3 572 9077 7 www keithley com www keithley de www keithley jp www keithley com www keithley com tw CHINA INDIA KOREA RUSSIA UNITED KINGDOM im 86 10 8447 5556 080 30792600 82 2 6917 5000 7 495 664 7564 44 1344 39 2450 U www keithley com cn www keithley in www keithley co kr www keithley ru www keithley co uk m For further information on how to purchase or to locate a sales partner please visit www keithley com company buy Copyright 2013 Keithley Instruments Inc Printed in the U S A No 3241 02 11 14
4. Series 2290 Source voltages up to 5kV and 10kV THA current measurement resolution Low noise for precision sourcing and sensitive measurements selectable filters reduce noise to less than 3mV on the 5kV supply Safety interlock controls high voltage output IEEE 488 programmable Protection module prevents damage to low voltage instrumentation TYPICAL APPLICATIONS e High voltage device breakdown testing High voltage component testing e Insulation testing e High voltage resistivity measurements e High energy physics research Ordering Information 2290 5 5kV Power Supply 120VAC Nominal Input Voltage 2290E 5 5kV Power Supply 240VAC Nominal Input Voltage 2290J 5 5kV Power Supply 100VAC Nominal Input Voltage 2290 10 10kV Power Supply 2290 PM 200 10kV Protection Module Accessories Supplied CD with user manual software drivers and accessory information 3 Pin Interlock Connector Power cord 1 888 KEITHLEY uss only Series 2290 High Voltage Power Supplies facilitate high voltage device and material testing as well as high energy physics experimentation The Model 2290 5 5kV Power Supply provides voltage outputs up to 5000V and the Model 2290 10 10kV Power Supply offers up to 10 000V The Model 2290 5 has a 5mA current capacity and the Model 2290 10 has a 1mA current capacity These supplies measure both output voltage with 1V resolution and output current with 1uA resolution In
5. addition to the voltage and current output displays a third display shows one of four settings output voltage voltage limit current limit or current trip These values can be set precisely with the front panel keypad The settings can also be programmed over the IEEE 4888 interface or the output voltage can be set with an analog control voltage Low Noise Ensures Accurate Low Level Measurements Low output noise is essential when using sensitive measurement instruments to make leakage cur rent or high resistivity measurements The maximum output ripple for the Model 2290 10 is less than IVays Lhe maximum output ripple of the Model 2290 5 is 100mV Two selectable internal filters on the Model 2290 5 reduce the maximum output ripple to only 3mV Extremely low noise from both power supplies enables sensitive measurement instruments to make accurate current measure ments down to picoamp levels Protecting User and Instrumentation Series 2290 Power Supplies and a Model 2290 PM 200 Protection Module protect both user and instrumentation from hazardous voltages An interlock circuit built into the power supplies can be used to ensure that the output voltage is disabled if a high voltage test fixture access door is open In addition all Series 2290 power supplies have low voltage analog outputs to per mit safe monitoring of the high voltage and the output current When low voltage measurement instrumentation is used in the high
6. ages below 100V but accuracy below 100V is not specified 2 Regulation specifications apply for greater than 100V 3 Times apply with the current limit set to 105 of full scale 4 Loads are fully resistive AC POWER 100 to 240 VAC A Tektronix Company MADE IN U S A KTH EERE O i CURRENT VOLTAGE EEE 488 STANDARD PORT INTERLOCK Model 2290 10 rear panel showing the IEEE 488 interface connector RS 232 interface connector high voltage output connector analog control output connectors and interlock connector KEITHLEY A Greater Measure of Confidence A Tektronix Company Series 2290 2290 PM 200 Specifications High Voltage Power Supplies SMU Protection Module Characteristics CONDITIONS This document contains specifications and supplemental information for the Model 2290 PM 200 Protection Module Specifications are the standards against which the Model 2290 PM 200 is tested Upon leaving the factory the Model 2290 PM 200 meets these specifications Characteristics supplemental characteristics and typical values are not warranted apply at 23 C 5 C lt 70 percent relative humidity and are provided solely as useful information CHARACTERISTICS LEAKAGE lt 10pA at 200V typical DC CURRENT CARRYING CAPABILITY lt 1 5A in unclamped condition PULSED CURRENT CARRYING CAPABILITY 10A maximum at 1 duty cycle in unclamped condition MAXIMUM PROTECTION ACTIVE CURRENT Current must be limited to lt 120
7. ical values are not warranted apply at 23 C 5 C lt 70 relative humidity and are provided solely as useful information CHARACTERISTICS VOLTAGE RANGE Output Voltage Maximum Output Current 100 to 10 000 V DC 1 000 mA DC VOLTAGE SET ACCURACY 0 06 of full scale VOLTAGE DISPLAY ACCURACY Voltage set accuracy 1V typical 2V maximum VOLTAGE RESOLUTION 1V set and display VOLTAGE LIMIT RANGE 0 to 100 full scale VOLTAGE REGULATION Line 0 001 for 10 line voltage change Load 0 04 for 100 load change typical OUTPUT RIPPLE 300Hz 300kHz 0 01 of full scale Vays RISE TIME from high voltage turn on to final value under full load 3 4 lt 6 seconds to within 1V of the final value DISCHARGE TIME to OV under full load gt 1 second to under 10V DISCHARGE TIME no load 3 lt 6 seconds to lt 1 of full scale voltage with no load typical SETTLING TIME gt lt 4 seconds to within 1V of the final value RECOVERY TIME 4 120ms for 40 step change in load current typical CURRENT LIMIT RANGE 0 to 105 of full scale CURRENT SET ACCURACY 0 5 of full scale CURRENT RESOLUTION 1A CURRENT DISPLAY ACCURACY 1 typical 2uA maximum TRIP CURRENT RANGE 102A to 105 of full scale excluding stored output charge OUTPUT STORED CHARGE lt 20uC maximum TRIP RESPONSE TIME lt 10ms STABILITY 0 01 per hour lt 0 03 per eight hours TEMPERATURE DRIFT 50ppm C 0 to
8. mA in clamped condition 230V PROTECTION VOLTAGE 220VDC minimum 230VDC typical 240VDC maximum DIMENSIONS 81 28mm long x 81 53mm wide x 54 10mm high 3 20 in long x 3 21 in wide x 2 13 in high Maximum length with high voltage connectors 5 34 in 135 64mm WEIGHT 0 25kg 0 55 1b 1 888 KEITHLEY ws only www keithley com GENERAL WARRANTY 1 year SAFETY Listed to UL 61010 1 2012 3rd edition Conforms to European Union Low Voltage Directive CONNECTORS High Voltage 5kV connector male SHV 0 38 inches 9 35mm diameter Kings 1704 1 10kV connector female SHV 0 535 inches 13 59mm diameter Kings 1764 1 Low Voltage Triax female quantity of 2 SAFETY Screw ground terminal quantity of 2 KEITHLEY A Greater Measure of Confidence A Tektronix Company VY S Ta a N b a oD 00 2 O gt lt ee ae DC POWER SUPPLIES Series 2290 High Voltage Power Supplies Y a a N CS a D 00 O gt E E L rid el 91 300 0191 Fax 31 388 5433 idm idm instrumentos es Wwiw idim instrumentas es Specifications are subject to change without notice All Keithley trademarks and trade names are the property of Keithley Instruments Inc All other trademarks and trade names are the property of their respective companies KEITHLEY A Tektronix Company A Greater Measure of Confidence KEITHLEY INSTRUMENTS INC E 28775 AU
9. ns High Voltage Power Supplies 5kV Voltage Power Supply Characteristics CONDITIONS This document contains specifications and supplemental information for the Model 2290 5 High Voltage Power Supply Specifications are the standards against which the Model 2290 5 is tested Upon leaving the factory the Model 2290 5 meets these specifications Characteristics supplemental characteristics and typical values are not warranted apply at 23 C 5 C lt 70 relative humidity and are provided solely as useful information CHARACTERISTICS VOLTAGE RANGE Output Voltage Maximum Output Current Conditions 50 to 5000 V DC 5 000 mA DC No filter 50 to 3000 V DC 5 000 mA DC Filter 1 50 to 5000 V DC 4 000 mA DC Filter 2 VOLTAGE SET ACCURACY 0 01 of setting 2 5V VOLTAGE DISPLAY ACCURACY Voltage set accuracy 1V typical 2V maximum VOLTAGE RESOLUTION 1V set and display VOLTAGE LIMIT RANGE 0 to 100 full scale VOLTAGE REGULATION Line 0 001 for 10 line voltage change Load 0 005 for 100 load change typical OUTPUT RIPPLE 10kHz 100kHz 0 002 of full scale Vpus maximum No filter 1 0mV ys 1kV Filter 1 or Filter 2 2 0mV ays 3kV Filter 1 or Filter 2 3 0mV yus 5KV Filter 2 1 5 seconds for 0 to within 1V of SOOOVDC No filter 3 seconds for 0 to within 1V of 3000VDC Filter 1 4 seconds for 0 to within 1V of 5000VDC Filter 2 1 second for 5000VDC to 1VDC No filte
10. oad GENERAL INPUT POWER 55 watts 2290 5 INPUT VOLTAGE 120V 10 50 or 60Hz 2290E 5 INPUT VOLTAGE 240V 10 50 or 60Hz 2290J 5 INPUT VOLTAGE 100V 10 50 or 60Hz REAR PANEL CONNECTORS SHV male Kings type 1704 1 or equivalent Output high voltage connector GPIB connector BNC Connector two Voltage set Voltage monitor Current monitor TOGGLE SWITCH Voltage setting or Voltage monitor HIGH VOLTAGE SAFETY INTERLOCK Connector 3 pin press in connector 3M part number 37103 A165 O00E MB Pin 1 5V out 1 5mA maximum out Pin 2 Input High voltage output enabled 3 24V DC High voltage output disabled lt 1 2V DC Pin 3 Chassis ground through a 100Q resistor INTERFACE PROTOCOL IEEE 488 1 OPERATING ENVIRONMENT 0 to 40 C non condensing DIMENSIONS 89mm high x 206mm wide x 406mm deep 3 5 in x 8 1 in x 16 in WEIGHT 5 5kg 12 pounds SAFETY Conformance to European Union low voltage directive WARRANTY One year WARM UP TIME One hour 1 The output voltage can be programmed to a voltage lower than 50V however performance below 50V is not specified 2 Add 2 5VDC when Filter 1 or Filter 2 is enabled 3 Regulation specifications apply for greater than 25VDC with full load or 50VDC with no load Below these values the unit may not regulate correctly 4 Peak to peak values are within five times the RMS value 5 Add 2 5uA offset when Filter 1 or Filter 2 is enabled 6 Times apply with the cu
11. r 3 seconds for 3000VDC to 1VDC Filter 1 4 seconds for 5000VDC to 1VDC Filter 2 lt 12 seconds to lt 50VDC No filter lt 30 seconds to lt 50VDC Filter 1 or Filter 2 SETTLING TIME 7 From 0 to programmed voltage to within 99 9 of final value within 3 seconds RECOVERY TIME 120ms for 40 step change in load current typical RISE TIME full load 67 DISCHARGE TIME full load 6 7 DISCHARGE TIME no load 6 Current Limit and Trip Range Voltage Range Filter 0 4 mA to 5 25 mA No filter or Filter 1 0 4 mA to 3 25 mA oot Filter 2 0 5 mA to 5 25 mA No filter or Filter 1 0 4 mA to 3 25 mA al Filter 2 CURRENT SET ACCURACY 0 01 2 5uA CURRENT RESOLUTION 1A CURRENT DISPLAY ACCURACY 1uA typical 2uA maximum STABILITY 0 01 per hour lt 0 03 per eight hours TEMPERATURE DRIFT 50ppm C 0 to 40 C typical PROTECTION Arc and short circuit protected programmable voltage and current limits and current trip MONITOR OUTPUTS OUTPUT SCALE 0 to 10V to full scale CURRENT RATING 10mA maximum OUTPUT IMPEDANCE lt 1Q ACCURACY 0 2 of full scale with a 100kQ load minimum UPDATE RATE 8Hz 1 888 KEITHLEY ws only www keithley com EXTERNAL VOLTAGE SET INPUT SCALE 0 to 10V for 0 to full scale INPUT IMPEDANCE 1MQ ACCURACY 0 2 of full scale UPDATE RATE 16Hz OUTPUT SLEW RATE lt rise time 0 3 seconds for 0 to full range under full l
12. rrent limit set to 105 of full scale 7 Loads are fully resistive EEE 488 STD PORT KEITH LEY A Tektronix Company MADE IN U S A SET MON IMON cO O omme 0 10V 0 10V WARNING NO USER SERVICEABLE PARTS INSIDE REFER TO USER MANUAL FOR SAFETY NOTICE FOR USE 8Y QUALIFIED PERSONNEL ONLY HIGH VOLTAGE Max Voltage Rating SkV A WARNING EXO INTERLOCK Verify that the LINE VOLTAGE SELECTOR card is set so the correct AC input voltage value is visible USE CORRECT FUSE Av UNE 50 60Hz A FUSE SLOW BLOW 1A 100 120V 1 2A 220 240V Model 2290 5 rear panel showing the IEEE 488 interface connector high voltage output connector analog control output connectors and interlock connector KEITHLEY A Tektronix Company A Greater Measure of Confidence WY S Ta a 5 N b a oD 00 O gt lt ee L DC POWER SUPPLIES VY a a N CS a oD 00 O gt E oe a DC POWER SUPPLIES Series 2290 2290 10 Specifications High Voltage Power Supplies 10kV Voltage Power Supply Characteristics CONDITIONS This document contains specifications and supplemental information for the Model 2290 10 High Voltage Power Supply Specifications are the standards against which the Model 2290 10 is tested Upon leaving the factory the Model 2290 10 meets these specifications Characteristics supplemental characteristics and typ
13. unt Kit KPCI 488LPA IEEE 488 2 Interface Board for the PCI Bus KUSB 488B JEEE 488 2 USB GPIB Interface Adapter for USB port with built in 2m 6 6 ft cable 7007 05 Double Shielded Premium IEEE 488 Interface Cable 0 5m 1 6 ft 7007 1 Double Shielded Premium IEEE 488 Interface Cable 1m 3 2 ft 7007 2 Double Shielded Premium IEEE 488 Interface Cable 2m 6 5 ft 7007 3 Double Shielded Premium IEEE 488 Interface Cable 3m 10 ft 7007 4 Double Shielded Premium IEEE 488 Interface Cable 4m 13 ft SERVICES AVAILABLE Model Number EW 1 Year Factory Warranty extended one additional year from date of shipment Model Number 3Y EW _1 Year Factory Warranty extended to 3 years frcom date of shipment Model Number 5Y EW _1 Year Factory Warranty extended to 5 years from date of shipment C Model Number 3Y STD KeithleyCare 3 Year Standard Calibration Plan C Model Number 3Y DATA KeithleyCare 3 Year Calibration with Data Plan C Model Number 5Y STD KeithleyCare 5 Year Standard Calibration Plan C Model Number 5Y DATA KeithleyCare 5 Year Calibration with Data Plan Insert model number Examples 2290 5 EW C 2290 10 3Y STD 1 888 KEITHLEY wis only www keithley com High Voltage Power Supplies Make Precise Current Measurements Lower Than 1pA Use a sensitive Keithley measurement instrument such as a Keithley SourceMeter Source Measure Unit SMU instrument to measure current drawn by the device under test DUT that is below
14. voltage circuit the protection module safely clamps the voltage across the instrument to a maximum value of 200V even when a device under test DUT breaks down Thus a Series 2290 Power Supply Model 2290 PM 200 Protection Module and Keithley accesso ries provide all the elements for building a safe high voltage test environment The Model 2290 PM 200 Protection Module protects low voltage measurement equipment V Lu l a A T aa m from voltages greater than 200V A WU m A Greater Measure of Confidence A Tektronix Company N wv A a N Beem D A V 00 2 gt is _00 a DC POWER SUPPLIES Series 2290 ACCESSORIES AVAILABLE For 2290 5 2290 5 SHV 5kV SHV Female SHV Female Cable 3m 10 ft 2290 5 MHV 5kV SHV Female MHV Male Cable 3m 10 ft 2290 5 SHVBH 5kV SHV Male Bulkhead Connector 2290 5 RMK 1 Single Fixed Rack Mount Kit for 5kV Power Supply 2290 5 RMK 2 Dual Fixed Rack Mount Kit for 5kV Power Supply For 2290 10 2290 10 SHVUC 10kV SHV Male to Unterminated Cable 3m 10ft 2290 10 SHV 10kV SHV Male SHV Male Cable 3m 10 ft 2290 10 SHVBH 10kV SHV Female Bulkhead Connector 2290 10 RMK 1 Single Fixed Rack Mount Kit for 10kV Power Supply 2290 10 RMK 2 Dual Fixed Rack Mount Kit for 10kV Power Supply For both 2290 PM 200 10kV Protection Module 2290 INT CABLE 3 Pin Connector to Unterminated Interlock Cable 4299 7 Fixed Shelf Rack Mo

Download Pdf Manuals

image

Related Search

2290 5 2290 57 th ave osceola wi 22905 train route 22905 form 22905 cpt code 22905 rail 22905 glenbrook 22905 train running status 22905 county road 453 bovey mn 55709 22905 zip code 22905 cpt 22905 s vermont ave torrance ca 90502 22905 e 19th ave aurora co 80019 22905 s vermont avenue torrance ca 90502