Home

TRACE GC Ultra - Thermo Scientific

image

Contents

1. bicbNs 45 Destination nee a 46 Receiving Instruments eere enne enn enn enne 46 Installation ier er rte repe AA eee de edet uds ORAT Red a a ass 47 TN RIN 48 SECTION II Installation Chapter 3 Before YOU Begin 5 2 noeud tn 51 Verify Site Preparation ooocoonnccionconnconnnonnnonnonnoconocnnonon oran enne nennen nennt nn enn en rennen enn nnne n rra nano 5 Unpacking the Instrument sss en nre n rennen nnns 5 Gas Plumbing Basics idu decet tad En eie Ha 53 How to Use a Tubing Cutter esses eere 53 How to Use Teflon Tape on Pipe Threads sesssssssseeeeeeneen 53 How to Use Swagelok Tube Fittings sese 54 vi Site Preparation and Installation Manual Contents How to Attach a Swagelok Tee or Cross ccccseesccessesseesseesseceseeeeceseecseenseenseeneeeneeenaes 56 Chapter 4 Plumbing the Gas Supply to the GC eeeeeeenenenenenenen nnne nennen 57 Using Hdi iia 58 Using the Hydrogen Sensor esses nennen ener nnne nnns 59 Building the Gas Lines eee decada 60 Connecting the Gas Supply Cylinders sse 60 Adding Traps to the Gas Supply ccccccccsccesscesseesseesseeseceseceseeesecaeceaeseeeeeeeessecseenaeeees 61 Purging Gas Pines ioo odere a etre dera eu epe de 62 Connecting the Gas Supply to the Inlet Manifold sse 63 Connecting the Gas Supply to the Detector Manifold oo
2. Site Preparation and Installation Manual Chapter 1 Space and Load Requirements Laboratory Preparation Space and Load Requirements D NOTE We recommend the following space allowances for the area around the gas chromatograph The working area should be at least 1 5 m 5 ft wide e The working area should be 1 5 m 5 ft deep The depth includes a 75 cm 30 in bench for the gas chromatograph The area behind the instrument should be a minimum of 30 cm 12 in Provide enough space around the instrument for operators to work beside it and in front of it Keep in mind that the GC oven vents to the rear Any material exposed to the oven exhaust must withstand repeated exposure to temperatures of up to 400 C 750 F The space requirements and weights for the standard system components are given in Table 1 1 Similar information for TRACE optional instruments is provided in Table 1 2 The TRACE system should be placed on a workbench that has minimum dimensions of 0 75 x 2 m 2 5 x 6 ft The workbench must also be capable of supporting the weight of the data system computer or computing integrator 27 kg 60 Ib plus the weight of any options Be sure there is at least 16 cm 6 in of space behind the workbench for connections In addition there must be at least 92 cm 36 in of vertical clearance from the top ofthe TRACE system Therefore you should avoid placing the instrument below cabinets or shelves To kee
3. Site Preparation and Installation Manual 39 Chapter 1 Laboratory Preparation Gas and Plumbing Requirements Liquid Carbon Dioxide Liquid CO must be supplied by a high pressure cylinder with a dip tube It is your responsibility to ensure the liquid delivery connection from the CO cryogenic supply is adaptable to 1 8 in tubing The CO cryogenic valve on the TRACE GC Ultra is a 1 8 in Swagelok valve Figure 1 5 shows the proper CO tank configuration o Dip tube Correct Incorrect Figure 1 5 CO Tank Configuration quemo High pressures and extremely low temperatures make pressurized liquid CO a hazardous material High concentrations of CO are dangerous To avoid injury always follow the safety precautions and delivery system design recommended by your gas supplier 40 Site Preparation and Installation Manual Telephone Telephone Chapter 1 Laboratory Preparation We recommended you install a telephone near the instrument so if necessary you can conveniently operate the system while speaking with Thermo Fisher Scientific Technical Support Refer to Appendix B Customer Communication for a list of Thermo Fisher Scientific offices worldwide The telephone line should accept digital transmissions for direct connection to the TRACE The telephone outlet should be within 2 m 6 ft of your system Required Tools The following tools are required for the
4. For carrier or detector fuel gas 99 995 high purity with less than 1 0 ppm of total hydrocarbons after purification One full size tank that has an outlet pressure of 400 700 kPa 60 100 psi Use a regulator with a CGA 350 connection or equivalent Use water oxygen and hydrocarbon traps Air For detector fuel gas 99 995 high purity Use a regulator with a CGA 0590 connection or equivalent Air compressors are not acceptable because they do not meet pressure water and hydrocarbon requirements Nitrogen For carrier or make up gas 99 995 high purity with less than 1 0 ppm of total hydrocarbons after purification Use a regulator with a CGA 0580 connection or equivalent Argon 5 Methane For ECD make up gas 99 995 high purity Use a regulator with a CGA 350 connection or equivalent Servo Air For valve actuation and secondary cooling Commercial grade or filtered air compressor Large volume delivery at 400 500 kPa 60 80 psi free of particles oils and water Gas Regulators and Fittings Gas tanks may be equipped with either single or dual stage regulators that contain stainless steel diaphragms The regulator output pressure should be adjustable from 300 to 1000 kPa 45 145 psi Each regulator should be equipped with a 1 8 in Swagelok compression fitting or equivalent Figure 1 3 shows a dual stage gas regulator with 1 8 in fittings Site Preparation and Installation Manual 35 Chapter 1 Laboratory
5. Testing for Leaks CAUTION Once you have connected your gas supplies to the GC inlets and detectors you need to test the supply lines for leaks Turn on your gas supply and use the following information to set the gas pressures for the leak test e Set the carrier gas pressure to approximately 50 kPa 7 psi higher than the maximum pressure of the GC regulator e Set the detector gas pressures to approximately 350 kPa 50 psi Next you need to check all Swagelok fittings for leaks Use the following procedure to check the fittings for leaks 1 Use a handheld electronic leak detector Thermo Scientific GLD Pro or equivalent to check each fitting for leaks 2 Ifyou detect a leak tighten the connection and retest it 3 Repeat this process until all Swagelok connections are leak free Do not use liquid soap leak detectors to check for leaks Liquid soap leak detectors may contaminate you system A mixture of 50 H 0 50 methanol or isopropyl alcohol may be used as a liquid leak detector WARNING Never use liquid leak detectors on or around electronic pneumatic circuits Site Preparation and Installation Manual 67 Getting Connected This chapter explains the TRACE power connections and helps you mount and configure peripheral devices and data systems Chapter at a Glance hist e 69 A EE 77 Introduction If you plan to use any peripheral devices with your GC such as an a
6. 10 ft and supplying more than one GC should be 1 4 in be free of oil be free of moisture run to the back or side of the TRACE system e be spliced with Swagelok fittings Precleaned tubing Properly cleaned tubing is solvent flushed and purge dried with an inert gas such as nitrogen or helium Flush gas lines using acetone or hexane Never use a chlorinated solvent When flushing gas lines collect the solvent passing through the gas line and inspect it for discoloration or residue Continue to flush gas lines until all waste solvent is free of discoloration and residue Purge the gas line for several minutes to make sure the gas line is free of all traces of solvent DO NCT e use any tubing other than new clean metal tubing of the specified size allow any brazed or soldered plumbing joints to be made in the gas delivery system without precleaning e allow pipe threaded joints without PTFE tape as a sealant no other thread sealant is acceptable use PTFE tape on compression fitting joints Swagelok fittings or equivalent mixcomponents of one fitting brand with those of another brand for example Swagelok ferrules with Tylok body etc Site Preparation and Installation Manual 37 Chapter 1 Laboratory Preparation Gas and Plumbing Requirements Gas Lines for Cold On Column Injectors Gas chromatographs equipped with a cold on column injector require secondary cooling performed by an extra air line with a
7. P N 317 092 41 This card indicates the theoretical K Factors related to the carrier gas and the column in use Site Preparation and Installation Manual XV Preventive Maintenance Schedule PN 317 092 80 This document provides a list of recommended scheduled maintenance and a year long log book to record maintenance observations supply lists and service records Maintenance and Troubleshooting Guide PN 317 091 80 This manual contains instructions for diagnosing and resolving operational problems Standard Operating Procedures PN 317 092 00 This manual contains instructions operating procedures and test criteria for final testing of the TRACE GC Ultra Spare Parts Catalog PN 317 092 10 This catalog contains a list of spare parts for the TRACE GC Ultra SECTION Site Preparation The Site Preparation section contains the information you need to prepare for the arrival of your TRACE GC Ultra gas chromatograph In addition to laboratory and operating environment standards this section contains information about the shipping and delivery of your TRACE system Chapter 1 Laboratory Preparation gives you all the information you need to prepare your laboratory for your TRACE system Chapter 2 Instrument Arrival gives an overview of shipping and receiving procedures installation and training Laboratory Preparation This chapter gives you all the information you need to prepare your site for the arrival of you
8. 10 ul columns a 7 m test column is supplied with each GC tubing and fittings for carrier and detector gases refer to Gas Tanks on page 36 for the tubing and fitting requirements for your gas chromatograph spare septa column ferrules Swagelok fittings Thermo Scientific GLD Pro Leak Detector or equivalent or 50 50 IPA water solution data system or integrator refer to Hardware and Software Minimum Requirements on page 43 for the data system requirements It is the responsibility of the customer to replace any consumables used during installation Site Preparation and Installation Manual Hardware and Software Minimum Requirements Chapter 1 Laboratory Preparation Hardware and Software Minimum Requirements Your TRACE GC Ultra produces analog and digital data output when you perform chromatographic analysis A computer with a data system or a computing integrator can be used to process the data from the GC The following topics provide the minimum requirements for computer hardware and operating systems depending on the data system you use to process the GC data ChromQuest Data System The ChromQuest data system has the following minimum computer requirements Hardware Requirements 133 MHz Pentium processor 48 MB RAM 1 GB hard drive CD ROM mouse serial port 2 NOTE Computer compatibility should not be assumed The computer may not be included with the TRACE GC Ultra ChromQuest data system Operating
9. GC system to any cooling duct outlets The air conditioning load for a basic TRACE system is approximately 3 320 W 10 355 Btu h Table 1 5 and Table 1 6 show the approximate heat output for a TRACE standard system and TRACE optional instruments Table 1 5 Heat Output TRACE Standard System Heat Output Instrument in BTU h TRACE GC Ultra 7 855 computer 1 650 monitor 850 Total 10 355 Table 1 6 Heat Output TRACE Optional Instruments Heat Output Instrument in BTU h GC Autosamplers AI 3000 AS 3000 341 5 TriPlus 752 Printers ink jet printer plotter 200 laser printer 2 900 Other peripheral devices such as purge and trap units may be connected to the TRACE GC Ultra These peripheral devices may have additional cooling requirements Consult your local Thermo Fisher Scientific CSE for assistance Site Preparation and Installation Manual Operating Environment Arrive Altitude Humidity Chapter 1 Laboratory Preparation Oven Exhaust The GC oven exhaust vents are in the rear of the instrument Hot air 400 C 750 F exits through these vents during GC operation As stated under Space and Load Requirements in this chapter there should be at least 30 cm 12 in free space behind the instrument to allow the exhaust to dissipate Any material exposed to the oven exhaust must be able to withstand repeated exposure to temperatures of up to 400 C 750 F
10. Power Requirements in Chapter 1 Laboratory Preparation You also should have completed the gas supply plumbing procedure as described in Chapter 4 Plumbing the Gas Supply to the GC If you have not already done so connect the power cord to the GC Plug the power cord into the appropriate power outlet The power switch is located on the back side of the GC Flip the switch to the ON position If the detectors are not already installed on your TRACE GC Ultra you will need to install them now Refer to the TRACE GC Ultra Operating Manual for more information about installing detectors You will need to install a column in the GC Refer to the TRACE GC Ultra Operating Manual for more information about installing columns For system test information refer to the TRACE GC Ultra Standard Operating Procedures Manual and to the user manuals of any accessories you have connected to your TRACE GC Ultra system Site Preparation and Installation Manual 77 Chapter 5 Getting Connected System Power Up 78 Site Preparation and Installation Manual Preinstallation Checklist This appendix contains a step by step list of items you should complete before your TRACE GC Ultra arrives Please complete the checklist after reading the Site Preparation section This will ensure that your site is suitable for the installation of your new TRACE GC Ultra Please contact a Thermo Fisher Scientific Customer Support Engineer CSE if you encounter any diffic
11. See transient inch input output kilo 10 or 1024 Kelvin kilogram kilopascal length liter Local Area Network pound light emitting diode meter or milli 10 D mega 1 0 micro 107 minute milliliter millimeter 85 Glossary m z NPD Pa PCB PDD PID PN psi ROM RS 232 86 mass to charge ratio nano 1 0 Nitrogen Phosphorous Detector ohm pico 10712 pascal printed circuit board Pulsed Discharge Detector Photo Ionization Detector part number pounds per square inch random access memory radio frequency read only memory industry standard for serial communications Site Preparation and Installation Manual S S sag slow average SOP surge T TCD transient U UFM V ac V dc VGA Site Preparation and Installation Manual Glossary second See surge A gradual long term change in average RMS voltage level with typical durations greater than 2 s Standard Operating Procedures A sudden change in average RMS voltage level with typical duration between 50 us and 2 s Thermal Conductivity Detector A brief voltage surge of up to several thousand volts with a duration of less than 50 us Ultra Fast Module volt volts alternating current volts direct current Video Graphics Array width Watt 87 A Air purity 35 Air conditioning humidity 29 particulate matter 30 temperature 27 user s responsibility 27 Altitude 29
12. TRACE GC Ultra is 47 kg 105 Ibs Two people should lift the instrument onto 52 the benchtop You should already have configured your laboratory according to the space requirements in Chapter 1 Laboratory Preparation and the gas and power supplies should be accessible Remove all protecting bags if any from the detector cell Optional equipment should be placed near the GC so you can connect it easily After placing the TRACE on the bench open the oven door and remove any packing material or other debris Before continuing with the installation you should determine the following the type of detector s installed on the GC e the type of injector port s installed on the GC the type of carrier gas required for each detector the type of makeup gas required for each detector Refer to Gas and Plumbing Requirements in Chapter 1 Laboratory Preparation for a list of detector and injector gas requirements Site Preparation and Installation Manual Chapter 3 Gas Plumbing Basics Before You Begin Gas Plumbing Basics In order to perform the procedures in this section properly you must first have a working knowledge of a few basic gas plumbing operations During the gas plumbing process you will frequently perform the operations explained in the following sections How to Use a Tubing Cutter Use the following procedure to cut your gas supply tubing properly 1 Attach the tubing cutter where you want to cut the tu
13. as defined in International Standard EN 61010 1 1993 Below are definitions for the most common voltage disturbances Slow average is a gradual long term change in average root mean square RMS voltage level with typical durations greater than 2 s Sags and surges are sudden changes in average RMS voltage level with typical durations between 50 us and 2 s Transients or impulses are brief voltage surges of up to several thousand volts with durations of less than 50 us Harmonic distortion 1s a high frequency disturbance that appears as distortion of the fundamental sine wave Total harmonic distortion should be less than 3 Effects of Voltage Disturbances Constant high line voltage or surges in voltage can cause overheating and component failures Constant low line voltage or sags in voltage can cause the system to function erratically Transients even of a few microseconds duration can cause electronic devices to fail catastrophically or to degrade shortening their lives significantly Harmonic distortion can cause noise in power supply lines and degrade the performance of the instrument Therefore it is important to establish the quality of the line power in your laboratory prior to the installation of your TRACE system D NOTE The interconnected power outlets for the TRACE system require a common point to one ground connector If there are two such points with each connected to separate external grounds noise current wil
14. environment as specified in International Standard EN 61010 1 1993 Exhaust System Specialized exhaust systems generally are not required for GC detector systems However it is your responsibility to see that local exhaust codes are followed You should be aware that certain detectors require the use of hydrogen for fuel gas Be sure to read Using Hydrogen in Chapter 4 Plumbing the Gas Supply to the GC if you plan to use hydrogen with your TRACE system Testing for Leaks also in Chapter 4 contains information about leak testing to ensure that excessive amounts of flammable gases are not released into the lab atmosphere The TRACE Operating Manual also discusses leak testing in detail Laboratories meeting the heating and cooling requirements specified in this manual will sufficiently circulate lab air to prevent accumulation of gases and fumes vented during normal GC use Extensive leaks in plumbing lines to the GC may present hazards Be sure to follow all local codes regulating plumbing of gases Gas and Plumbing Requirements A You must provide the gas supplies for your gas chromatograph Be sure to order your gases and regulators far enough ahead of time to have them ready for the GC installation process WARNING All Thermo Fisher Scientific gas chromatographs are designed to use an inert gas as the carrier gas If you wish to use hydrogen as a carrier gas you must install a hydrogen sensor Contact a Thermo Fisher Scienti
15. high flow rate This line should have an inside diameter of at least 2 mm 5 64 in Cryogenic Cooling 38 If you have purchased a cryogenic cooling option for your TRACE GC Ultra you will need to provide a coolant supply Your TRACE GC Ultra is already configured for either liquid nitrogen or liquid carbon dioxide Be sure to identify which cryogenic cooling option your GC is configured for before you order cryogenic coolant Site Preparation and Installation Manual Chapter 1 Gas and Plumbing Requirements Laboratory Preparation A Liquid Nitrogen Liquid N must be supplied at pressures below 400 kPa 60 psi Plumbing to the GC should be 1 4 in copper or stainless steel tubing with insulation It is your responsibility to ensure the liquid delivery connection from the liquid N cryogenic supply is adaptable to 1 4 in tubing The liquid N cryogenic valve on the TRACE GC Ultra is a 1 4 in Swagelok fitting Subambient operation of the PTV requires liquid N cooling Figure 1 4 shows the proper configuration for an N tank 3 ead Valve gt Insulated 1 4 tubing Figure 1 4 N Tank Configuration WARNING High pressures and extremely low temperatures make liquid N a hazardous material High concentrations of N in the air can cause asphyxiation hazard To avoid injury always follow the safety precautions and delivery system design recommended by your gas supplier
16. purged the gas line to remove any debris or contaminants as described in Purging Gas Lines in this chapter Use the following procedure to connect the gas line to the detector manifold 1 Make sure the gas supply is turned off 2 Connect each detector gas line to the appropriate detector gas bulkhead with 1 8 in Swagelok fittings If the GC has been furnished with a 2x1 mm tube use the ferrule for 2x1mm 1 8 in provided in the start up kit Different detector modules have different gas plumbing requirements It is important that you connect the right gases to the right inlet fittings The inlet fittings on the detector modules are labeled Consult Table 4 1 on page 65 for the proper gas connections of the detector modules installed Repeat the process 1f you are connecting the detector gas supply to more than one bulkhead Be sure to finish all detector connections before turning on the gas supply Inside each 1 8 in detector manifold is a fritted filter If you connect a gas line smaller than 1 8 in to an detector manifold be sure the tubing does not touch the filter when inserted into the manifold To keep the gas line from touching and possibly damaging the filter extend the tubing only 5 mm past the front ferrule This will ensure the tubing does not touch the filter Refer to How to Use Swagelok Tube Fittings in Chapter 3 Before You Begin for more instructions on using Swagelok ferrules and fittings Refer to Gas and Plu
17. the tubing to the gas outlet Refer to How to Use Swagelok Tube Fittings in Chapter 3 Before You Begin for instructions on using Swagelok fittings Adding Traps to the Gas Supply You may wish to use traps such as moisture or oxygen traps in your gas supply line Traps are especially necessary if you are unsure of the purity of your gas supplies Refer to Gas and Plumbing Requirements in Chapter 1 Laboratory Preparation for more information about gas purity requirements and appropriate traps for your GC Figure 4 1 shows the proper order of traps in a gas supply line Refer to your trap s instruction manual for specific purging and installation requirements Oxygen trap Moisture Hydrocarbon trap 7 Figure 4 1 Gas Trap Configuration Site Preparation and Installation Manual 61 Chapter 4 Plumbing the Gas Supply to the GC Purging Gas Lines Use the following procedure to attach traps to your gas line 1 Determine the location for the trap in your supply line Figure 4 1 shows the recommended trap order 2 Use a tubing cutter to cut the tubing to the desired length 3 Turn the gas regulator supply handle counterclockwise to reduce line pressure then turn the gas supply on and increase the pressure to 35 kPa 5 psi for about 15 seconds to purge the line of any unwanted debris 4 Turn the gas supply off 5 Use a 1 8 inch Swagelok fitting to connect the trap inlet to the g
18. 4 Plumbing the Gas Supply to the GC Connecting Coolant to the Cryogenic System Connecting Coolant to the Cryogenic System The column oven and PTV Programmable Temperature Vaporizing injector cryogenic systems allow you to operate the GC at sub ambient temperatures The cryogenic systems can reach the following temperatures The column oven cryo system can reach 55 C with Liquid Nitrogen or 99 C with carbon dioxide e The PTV cryo system can reach 50 C with liquid nitrogen Use the following procedure to connect the coolant to your cyrogenic system Column Oven Cryo System 1 Connect the supply tubing to the CO or LN tank Refer to Figures 1 4 and 1 5 in Chapter 1 for more information about cryo supply tanks 2 Connect the other end of the supply tube to the 1 8 inch connector on the column oven cryo solenoid valve The connector protrudes from the hole on the GC rear panel labeled Sub Ambient Coolant PTV Injector Cryo System If your GC has a PTV injector use the following procedure to connect the cryo supply 1 Connect the supply tubing to the CO or LN tank Refer to Figures 1 4 and 1 5 in Chapter 1 for more information about cryo supply tanks 2 Connect the other end of the supply tube to the 1 8 inch connector on the PTV cryo solenoid valve on the upper left corner of the GC rear panel 66 Site Preparation and Installation Manual Testing for Leaks Chapter 4 Plumbing the Gas Supply to the GC
19. Ammonia carrier gases 35 Argon 5 Methane purity 35 Autosamplers heat output 28 weight 22 C Capillary columns gases 32 Carrier gases 32 ammonia 35 CGA fittings 36 Crematory minimum requirements hardware 44 operating system 44 ChromQuest minimum requirements hardware 43 operating system 43 Cold on column injectors gas lines 38 Computer heat output 28 power 26 weight 22 Coolant 38 liquid carbon dioxide 40 hazards 40 liquid nitrogen 39 hazards 39 Cryogenic cooling 38 liquid carbon dioxide 40 hazards 40 Site Preparation and Installation Manual liquid nitrogen 39 hazards 39 CSE installation 47 D Data System Chrom Card 44 ChromQuest 43 minimum requirements 43 Detectors gases 32 Doorways 20 Dust 30 E ECD gases capillary and wide bore columns 32 packed columns 33 Electrostatic discharge 30 Elevators entrance 20 Entrance 20 Equipment required 41 supplemental 42 Exhaust system 31 Ex works See Shipping origin F FID gases capillary and wide bore columns 32 packed columns 33 Floors vibration 30 Flushing gas lines 37 FPD gases capillary and wide bore columns 32 packed columns 33 Freight insurance 46 Frequency power 23 89 Index Fuel gases 32 G Gas lines 36 cold on column injectors 38 fittings 37 requirements 37 Gases carrier 32 detectors 32 fuel 32 gas lines 36 37 cold on column injectors 38 pre cleaned tubing 37 requirements 37 gas tanks 36 storage 36 make up 32 pa
20. Home Part of Thermo Fisher Scientific Thermo Scientific TRACE GC Ultra Gas Chromatograph Site Preparation and Installation Manual PN 31709190 Revision May 2010 SCIENTIFIC TRACE GC Ultra Site Preparation and Installation Manual May 2010 Edition Part Number 317 091 90 2007 2010 Thermo Fisher Scientific Inc All rights reserved Printed in Italy Published by Thermo Fisher Scientific S p A Strada Rivoltana 20090 Rodano Milan Italy Tel 39 02 95059355 Fax 39 02 95059388 Printing History First Edition released June 1998 Second Edition released November 1998 Third Edition released June 1999 Forth Edition released June 2001 Fifth Edition released January 2002 Sixth Edition released May 2003 Seventh Edition released April 2004 Eighth Edition released January 2005 Ninth Edition released September 2005 Tenth Edition released January 2007 Eleventh Editions released May 2007 Twelfth Edition released September 2007 Thirteenth Edition released April 2009 Fourteenth Edition released September 2009 Fifteenth Edition released May 2010 Disclaimer Technical Information contained in this publication is for reference purposes only and is subject to change without notice Every effort has been made to supply complete and accurate information however Thermo Fisher Scientific assumes no responsibility and will not be liable for any errors omissions damage or loss that might result from any use of
21. Hydrogen Sensor NOTE y Nds The lower limit of the hydrogen sensor is 0 596 in volume You should adjust the detection threshold to 1 in volume which is 25 of the hydrogen lower limit of explosion 4 in volume In cases where the connections begin to leak or the column breaks the sensor alerts the operator Then it automatically cuts off the gas supply and heating to the active zones and sweeps the column oven with forced air ventilation If the sensor detects anomalies or leaks during GC operation due to instrument malfunction the operator must immediately close the hydrogen supply Switch off the gas chromatograph airout the room The reliability of the sensor depends on careful maintenance After the sensor is in use you must periodically check its operating performance and calibration as recommended by the manufacturer Refer to your hydrogen sensor s instruction manual for maintenance guidelines Never use hydrogen in your TRACE system unless your GC oven has a hydrogen sensor installed Thermo Fisher Scientific CSEs are not authorized to install or repair any instrument using hydrogen as a carrier gas unless the instrument is equipped with the appropriate sensor 1 IEC 1010 1 First Edition September 1990 IEC 1010 1 Amendment 1 September 1992 IEC 1010 1 Amendment 2 June 1995 Site Preparation and Installation Manual 59 Chapter 4 Plumbing the Gas Supply to the GC Building the Gas Lines Build
22. Oven exhaust can cause severe burns Avoid working behind the instrument when the oven vents during cooling down cycles Do not expose gas tanks or bottles chemicals regulators electrical cords or other temperature sensitive items to oven exhaust The maximum operating altitude for the TRACE GC Ultra is 2 000 meters above sea level The relative humidity of the operating environment must be between 40 and 80 with no condensation Operating your TRACE GC Ultra system in low humidity will cause the accumulation and discharge of static electricity which shortens the life of electronic components Operating your GC system in high humidity causes condensation which can lead to short circuits High humidity also blocks traps on cooling fans We recommended you equip your laboratory with temperature and humidity monitors to ensure your laboratory is always within the required temperature and humidity specifications Particulate Matter The air in your laboratory must not have excessive dust smoke or other particulate matter The air should contain fewer than 1 000 000 particles per cubic meter 100 000 particles per cubic foot larger than 5 um Site Preparation and Installation Manual 29 Chapter 1 Laboratory Preparation Operating Environment Dust can clog the air traps causing a reduction in air flow around electronic components Dust will also form a layer on electronic components that will act as an insulating blanket reducin
23. Preparation Gas Tanks Gas and Plumbing Requirements 0 A Figure 1 3 Dual Stage Gas Regulator Verify with your gas supplier the type of CGA fitting on the gas bottles you order Be sure these fittings match the regulators described in Purity Requirements in this chapter If not contact your gas supplier or your area Thermo Fisher Scientific sales representative to make the appropriate changes Secondary Gas Regulators Because secondary regulation of carrier fuel and make up gases may be provided downstream of the cylinder regulator a single or dual stage regulator is acceptable but it must use a stainless steel diaphragm Gas tanks should be placed where they cannot damage cables or gas lines Use standard safety practices for securing gas tanks and bottles Gas supply lines should terminate with 1 8 in female Swagelok connectors Stored gases should be placed in an area where a wide temperature variation will not occur Don t forget to plan for gas cylinder storage when preparing your laboratory y Nus Do not place gas tanks in the path of the GC oven exhaust 36 Site Preparation and Installation Manual Chapter 1 Gas and Plumbing Requirements Laboratory Preparation Gas Lines Gas lines should be refrigeration grade precleaned copper or new stainless steel e be 1 8 in or 1 4 in diameter gas lines longer than 3 m
24. System Requirements Windows XP Vista Site Preparation and Installation Manual 43 Chrom Card Data System The Chrom Card data system has the following minimum computer requirements Hardware Requirements e 133 MHz Pentium processor 24MB of RAM 1 GB hard drive CDROM Operating System Requirements Windows XP Vista Instrument Arrival Shipping This chapter gives an overview of shipping and receiving procedures installation and training The information in this chapter will give you an idea of what to expect when your TRACE system arrives Telephone and fax numbers for the Thermo Fisher Scientific offices are listed in Appendix B Customer Communication Chapter at a Glance a BOR eee edie ae Uae ea Raa 45 DU s o v Am C 47 JC M 48 Information Instruments are shipped in the manner agreed upon at the time of sale There are two categories of shipping Origin FOB and Destination FOB CIP or CIF Origin For instruments shipped Ex Works also known as FOB Free On Board Origin damages incurred in shipment are the responsibility of the purchaser and the carrier However Thermo Fisher Scientific will assist with filing claims and billable repairs 1f necessary Site Preparation and Installation Manual 45 Chapter 2 Instrument Arrival Shipping Information Destination There are two types of Destination shipping CIP Carriage and Insuranc
25. and network connections and install drivers as necessary 6 Refer to your data system s operating manual for instructions on configuring your data system and using it to control GC functions and process information from the GC 7 Replace the top cover Connect the Autosampler You can connect a TriPlus or AI AS 3000 II autosampler to the TRACE GC Ultra Use the following procedure to connect the autosampler cable to the GC 1 Connect the RS232 cable the DIN handshake cable and the power cable supplied with the autosampler to the autosampler control module Site Preparation and Installation Manual 73 Chapter 5 Getting Connected Introduction 2 With the top cover removed connect the RS232 cable to the port on the top of the GC labeled AUTOSAMPLER Autosampler Port or 3 Snap the cable securing ring into one of the cable slots on the GC 74 Site Preparation and Installation Manual Chapter 5 Introduction Getting Connected or 4 Connect the handshake cable to the AUTOSAMPLER DIN port on the rear of the GC 5 Replace the top cover Site Preparation and Installation Manual 75 If the GC and autosampler are on the GC will automatically recognize the autosampler Refer to the autosampler operating manual for instructions on mounting the autosampler turret and tray on top of the GC and for configuration instructions Computing Integrators Setup yo
26. as supply tubing coming from the tank 6 Turn the gas supply on to 35 kPa 5 psi for about 30 seconds 7 Turn the gas supply off 8 Connect tubing to the trap outlet with a 1 8 inch Swagelok fitting Follow this procedure for all traps you wish to add to your gas supply line Purging Gas Lines You must purge the lines any time you make a cut in the tubing during the gas line assembly process This will clear them of any debris from the cut You will also need to purge the completely assembled gas lines including any traps you plan to use before you connect the your gas supply to the TRACE GC Ultra Use the following procedure to purge the gas lines 1 Turn the gas supply on and set the pressure to 35 kPa 5 psi 2 Allow the line to purge for 10 minutes 3 Turn off the gas supply 62 Site Preparation and Installation Manual Chapter 4 Connecting the Gas Supply to the Inlet Manifold Plumbing the Gas Supply to the GC Connecting the Gas Supply to the Inlet Manifold ye NOTE Aww By now you should have done the following completely built your gas line from your gas supply including any traps tees and extra tubing to allow about 40 cm 16 in of slack in the line e purged the gas line after every tube cut to remove any debris or contaminants Use the following procedure to connect the gas line to the GC inlet manifold 1 Make sure the gas supply is turned off 2 Connect the gas line to the carrier gas inlet wi
27. aust System xi entender eedem e e eee ee e a 31 Gas and Plumbing Requirements eren 31 Gas PU it st eden ates 33 guch p E 33 Purity Requirements coe eret li e Re dt 34 Gas Regulators and Fittings sssssssssseeseeseeee enne enne 35 Site Preparation and Installation Manual V Contents Secondary Gas Regulators ccccccccccessecsseeseceeeceeeeeesecsseceecseeeeeeeesseeneeseeees 36 CHENG 36 GERBER TTE 37 Gas Lines for Cold On Column Injectors ooooocciocnoonionnnonnconocononanono nono nconnnos 38 Gryogenic Coolmpg etii tte tr e E eer Rr aH Ee Ed dedito 38 Liquid Nitrogen ras 39 Liquid Carbon Drioxade euer rete nter te ie Rea 40 Pele PHONG A DERE een e orga eser EE E eer PE ERES G 41 Required T9ols iecore e re C He A il tas 41 Supplemental Equipment sess ennren nennen ener es nennen enne 42 Hardware and Software Minimum Requirement eesesssesseeeeeeeeenr enne 43 ChromQuest Data S ystet 35 2 enm ceste tUe ce ITI n MAR 62 43 Hardware Requirements sse ener enne nennen 43 Operating System Requirement ssssssssesseeeeeeeeer enne 43 Chrom Card Data System decet ne e erede sieur da estes 44 Hardware Requirements sess ener nennen 44 Operating System Requirements ssssssssseee eene enne 44 Chapter 2 Instrument Arrival eir ht tee pete in doute ical eaim Der id Ra uie 45 Shipping Information irn illa 45 uunc
28. bing and tighten the knob 2 Rotate the cutter around the tubing and tighten the knob Continue rotating and tightening the knob until you deeply score the tubing 3 Remove the cutter and snap the tubing in two How to Use Teflon Tape on Pipe Threads To ensure an inert seal around all tank and regulator pipe fittings use the following procedure 1 Wrap the tape around the pipe threads in a clockwise direction about three times You should be able to tear the tape easily to separate it from the roll 2 Make sure the tape does not extend past the threads or obstruct the pipe opening 3 Make sure the tape does not twist or bunch up when you connect the pipe pe NOTE Do not use Teflon tape on Swagelok type compression fittings Site Preparation and Installation Manual 53 Chapter 3 Before You Begin Gas Plumbing Basics How to Use Swagelok Tube Fittings The Swagelok fitting consists of four components a Swagelok nut a back ferrule a front ferrule and an inlet body It becomes a five piece connection when affixed to the tubing The two ferrules merge when the nut is tightened forming a safe and leak free seal between the tubing and body Use the following procedure to connect tubing using Swagelok fittings 1 Place a Swagelok nut over the end of the tubing 2 Place a back ferrule over the end of the tubing 3 Place a front ferrule over the end of the tubing If you are using tubing smaller than 1 8 in make sur
29. bore columns 32 Plugs international 25 USA 25 Plumbing gases cutting tubing 53 Swagelok fittings 54 Swagelok tees and crosses 56 Teflon tape 53 Pollution 31 Power 120 V ac 23 230 V ac 23 current requirements optional instruments 26 printers 26 standard system 26 frequency 23 harmonics 24 Site Preparation and Installation Manual Index quality of 23 sags and surges 24 slow average 24 technical assistance 27 transients or impulses 24 voltage disturbances 23 wall outlets 24 Power supply UPS 25 Printers heat output 28 power 26 weights 22 R Regulators 35 Required tools 41 S Sags 24 Secondary gas regulators 36 Servo air purity 35 Setting up the instrument 52 Shipping 45 container 20 damage 46 destination 46 origin 45 Site preparation preinstallation checklist 51 user s responsibility 20 verifying 51 Slow average 24 Solvent in gas lines 37 Static discharge See Electrostatic discharge Sub ambient operation 38 Supplemental equipment 42 Surges 24 Swagelok fittings 54 Swagelok tees and crosses 56 91 Index T TCD gases capillary and wide bore columns 32 packed columns 33 Technical assistance line power problems 27 See also Customer service Teflon tape 53 Telephone 41 support 41 Temperature air conditioning 27 laboratory optimum 27 See also Heat output system reliability 27 Tools 41 Training 48 Transients or impulses 24 Tubing cutter 53 U unpacking 51 UPS power supply 25 User s re
30. cifying power conditioning equipment is a complex task that is best handled by a company or consultant specializing in that field Contact your Thermo Fisher Scientific CSE for assistance in locating a power consultant Refer to Appendix B for a Thermo Fisher Scientific office in your area Operating Environment The operating environment in your laboratory is affected by such factors as temperature humidity particulate matter and electrostatic discharge It is your responsibility to provide an acceptable operating environment for your TRACE system Attention to the operating environment will ensure continued high performance of your gas chromatograph Temperature The room temperature must be maintained between 18 and 35 C 65 and 95 F and should remain consistent As the laboratory temperature increases system reliability decreases 29 NOTE All electronic components generate heat while operating This heat must be dissipated to the surrounding air for the components to operate reliably Cooling Requirements There must be a good flow of air around the system and the air conditioning must be capable of maintaining a constant temperature within the operational limits Any costs for air conditioning are more than offset by good sample throughput and reduced repair costs Site Preparation and Installation Manual 27 Chapter 1 Laboratory Preparation Aw O NOTE 28 Operating Environment Do not directly expose the
31. cked columns 33 plumbing basics 53 54 56 cutting tubing 53 Swagelok fittings 54 Swagelok tees and crosses 56 Teflon tape 53 purity 33 35 air 35 argon 5 Methane 35 helium 34 hydrogen 35 nitrogen 35 servo air 35 recommendations 32 regulators 35 secondary 36 supplies 31 33 traps 33 hydrocarbon 34 moisture 34 oxygen 34 water background 34 Ground 23 H Hardware minimum requirements 43 90 Harmonics 24 Heat output autosamplers 28 computer 28 monitor 28 optional instruments 28 printers 28 TRACE GC 28 Helium purity 34 Humidity low high 29 static discharge 29 Hydrocarbon traps 34 Hydrogen exhaust 31 purity 35 sensor 31 warning 31 l Installation 47 Customer Service Engineer 47 equipment required 41 supplemental 42 purchase option 47 Installation kit replacement of consumables 42 Instrument arrival 45 freight insurance 46 L Laboratory entrance shipping container 20 Lighting 30 Liquid carbon dioxide 40 hazards 40 Liquid nitrogen 39 hazards 39 M Minimum dimensions entrance 20 Minimum requirements Site Preparation and Installation Manual hardware 43 Moisture traps 34 N Nitrogen purity 35 NPD gases capillary and wide bore columns 32 packed columns 33 O Operating environment user s responsibility 27 Optional instruments heat output 28 power requirements 26 Overvoltage category 24 Oxygen traps 34 P Packed columns gases 33 Particulate matter 30 PID gases capillary and wide
32. dicated to the quality of our customer relationships and services This appendix also contains a one page Reader Survey Use this survey to give us feedback on this manual and help us improve the quality of our documentation How To Contact Us Use http www thermo com com cda resources resource_detail 1 12512 00 html address for products information Use http www gc gems customersupport com WebPage Share Default aspx address to contact your local Thermo Fisher Scientific office or affiliate GC GC MS Customer Support Site Preparation and Installation Manual 81 Appendix B Customer Communication How To Contact Us Reader Survey Product TRACE GC Ultra Manual Site Preparation and Installation Manual Part No A31709190 1 Please help us improve the quality of our documentation by completing and returning this survey Circle one number for each of the statements below Strongly Strongly Agree Agree Neutral Disagree Disagree The manual is well organized 1 2 3 4 5 The manual is clearly written 1 2 3 4 5 The manual contains all the information I need 1 2 3 4 5 The instructions are easy to follow 1 2 3 4 5 The instructions are complete 1 2 3 4 5 The technical information is easy to understand 1 2 3 4 5 Examples of operation are clear and useful 1 2 3 4 5 The figures are helpful 1 2 3 4 5 I was able to install the system using this manual 1 2 3 4 5 If you would like to make additional comments please do Attach additio
33. e Paid to Destination and CIF Carriage Insurance and Freight paid to Destination These are also known as FOB Destination For all types of Destination shipping Thermo Fisher Scientific will file claims against the carrier for any damages incurred in shipment Note however that Thermo Fisher Scientific will not accept liability for damage not recorded on the receiving documents Receiving Instruments O NOTE 46 Thermo Fisher Scientific instruments are shipped by electronic equipment carriers who specialize in the handling of delicate equipment Occasionally however equipment inadvertently gets damaged in transit Please take the following precautions when receiving the instruments e Check carefully for obvious damage or evidence of rough handling including triggering of Shockwatch and Tiltwatch labels e If external damage is apparent note this fact on all copies of the receiving documents and describe briefly the extent of the damage The driver should sign or initial next to your comments to signify agreement with your observations It may be necessary to photograph damaged areas for claims purposes Contact the appropriate Thermo Fisher Scientific office to report the damage Move the cartons to a protected location preferably to the installation site Leave the boxes as complete as possible and do not unpack the components unless absolutely necessary The Customer Support Engineer CSE will also check fo
34. e the tubing extends only 5 mm past the front ferrule This will prevent damage to the fritted filter inside the inlet manifold Figure 3 1 shows the proper assembly order Swagelok Front Inlet nut ferrule body Back Tubing ferrule Figure 3 1 Swagelok Ferrule and Nut Assembly 4 Push the Swagelok nut over the ferrules 5 Insert the tubing into the inlet body as far as it will go If you are using tubing smaller than 1 8 in make sure the tubing extends only 5 mm past the front ferrule when you insert it into the inlet body 6 Slide the nut over the inlet body as shown in Figure 3 2 and tighten until finger tight 54 Site Preparation and Installation Manual Chapter 3 Gas Plumbing Basics Before You Begin Inlet body 0110 Tubing ferrule and nut assembly Figure 3 2 Swagelok and Inlet Connection 7 While holding the inlet body tightly with a backup wrench tighten the nut about a 3 4 turn past finger tight as shown in Figure 3 3 Figure 3 3 Tightening Swagelok Fittings 2 NOTE You can mark the nut before tightening This will help you confirm that you have turned the nut a 3 4 turn Site Preparation and Installation Manual 55 Chapter 3 Before You Begin Gas Plumbing Basics How to Attach a Swagelok Tee or Cross To use a single gas source for more than one inlet or detector module use a Swagelok tee or cross t
35. epair of such failures is specifically excluded from the standard warranty and service contract coverage Use this chapter to ensure that your site meets all the criteria of the Preinstallation Checklist in Appendix A In addition to the information in this chapter you must also obey the building and safety rules and regulations for construction that apply in your area The entrance to your facility and the width of all hallways elevators and doorways should be at least 92 cm 37 in However you should allow additional room for maneuvering the system around corners into elevators or through doorways The TRACE GC Ultra and accessories are shipped in a container with the following dimensions 1280 cm 32 in w 80 cm 32 in A 80 cm 32 in The container and its contents weigh approximately 75 kg 166 Ibs Other modules such as the computer monitor and options are shipped in their own containers Their dimensions and weights are less than those ofthe TRACE GC Ultra container 20 1 Your instrument is shipped in a container the smallest dimension of which is 80 cm 32 in If the entrance to your laboratory will not accommodate a 80 cm 32 in container you can remove the individual modules from the container before moving them into the room If the instrument is removed from its shipping container before it is delivered to the lab site be sure that all the contents of the container remain with the instrument
36. erials are worn out broken or otherwise faulty Use the following safety precautions when using hydrogen Ensure that all hydrogen cylinders comply with the safety requirements for proper use and storage Hydrogen cylinders and delivery systems must comply with local regulations e Make sure the gas supply is turned completely off when connecting hydrogen lines e Perform a bubble test to ensure that the hydrogen lines are leak tight before using the instrument Perform this test after the pressure test described in the Testing for Leaks portion of this chapter Repeat this test to eliminate all leaks Ensure your GC column oven has a Thermo Fisher Scientific hydrogen sensor A hydrogen sensor continuously monitors the hydrogen level in the GC column oven If your GC oven does not have a hydrogen sensor already installed contact your Thermo Fisher Scientific sales representative To comply with 58 Site Preparation and Installation Manual Using Hydrogen Chapter 4 Plumbing the Gas Supply to the GC instrument safety requirements a Thermo Fisher Scientific CSE or authorized service technician should install the sensor If you plan to use a sensor other than the recommended Thermo Fisher Scientific sensor you must verify its ability to perform the functions listed above before installing it It must comply with your local safety regulations or with the IEC 61010 regulations if local regulations do not exist Using the
37. fic sales representative if you plan to use hydrogen as the carrier gas in your new TRACE GC Ultra If you don t have the hydrogen sensor you must use an inert carrier gas Site Preparation and Installation Manual 31 Chapter 1 Laboratory Preparation Gas and Plumbing Requirements Table 1 7 lists the gas recommendations for capillary column systems Table 1 7 Capillary and Wide Bore Column Gas Recommendations Detector Type Carrier Gases Fuel Gases Make Up Gas FID Helium Hydrogen Air Helium Nitrogen Nitrogen Hydrogen if necessary NPD Helium Hydrogen Air Helium Nitrogen Nitrogen Hydrogen PID Helium None Helium Nitrogen Nitrogen Hydrogen ECD Helium None Nitrogen Nitrogen Argon 5 Methane Argon FPD Helium Hydrogen Air Helium Nitrogen Nitrogen Hydrogen if necessary TCD Helium None Same as carrier Nitrogen Hydrogen Argon PDD Helium is the gas used for PDD discharge and carrier supply MS Helium must be used when the MS detector is coupled with the GC 1 Maximum delivery pressure 1000 kPa 145 psi for carrier gas 420 kPa 60 psi for detector fuel gases 2 Oxygen trap should be installed in gas lines 32 Site Preparation and Installation Manual Gas and Plumbing Requirements Gas Purity Chapter 1 Laboratory Preparation Table 1 8 lists the gas recommendations for packed column systems Table 1 8 Packed Column Gas Recommendations Detector T
38. g the transfer of heat from the components to the surrounding air Electrostatic Discharge Vibration Lighting 30 The TRACE GC Ultra with all its covers in place is not susceptible to static discharge However laboratory conditions that expose instrumentation and laboratory personnel to repeated high levels of static discharge should be considered a safety hazard Therefore we recommend you take the following precautions especially if you are operating your system at the lower end of the relative humidity specification listed above e Use a static dissipating floor covering such as tile or conductive linoleum in the room housing your instrument Use laboratory chairs covered with natural fibers or other static dissipating material e Wear laboratory coats and clothing made from natural fibers or other static dissipating material Do not place polystyrene foam cups or packing materials on the instrument Place your instruments on surfaces that are free of vibration Be aware of vibrations caused by equipment in adjoining locations Overhead lighting 1s recommended to light your work area Site Preparation and Installation Manual Exhaust System Chapter 1 Laboratory Preparation Other Environmental Factors The TRACE GC Ultra operates in an environment where normally only non conductive pollution occurs but in which temporary conductivity due to condensation must be expected This is a Pollution Degree 2
39. ime the sument indicio 51 Gas Phut bints Basis uice e he needed te t ee te tree ette ie ete remi Dua 53 Verify Site Preparation Before the TRACE system can be installed your laboratory must be in compliance with the guidelines and requirements in the Site Preparation section of this manual Use the Preinstallation Checklist Yn Appendix A to verify your laboratory conditions conform to the guidelines in Chapter 1 Laboratory Preparation before you unpack your GC and begin the installation process Unpacking the Instrument You should have already inspected the exterior of the shipping container for damage as described in Chapter 2 Instrument Arrival Carefully unpack the instrument and do the following 1 Check the contents of each box or crate against the packing list to verify the shipment is complete Site Preparation and Installation Manual 51 Chapter 3 Before You Begin Unpacking the Instrument 2 Inspect each item for damage e Ifequipment is damaged keep boxes and their equipment in their existing condition and immediately notify the carrier Submit a damage claim directly to the carrier and send a copy including any shortage claims to your authorized Thermo Fisher Scientific sales representative Do not return any equipment to the dealer or the factory without prior factory authorization 3 Place the TRACE GC Ultra on your bench allowing space to the rear of the instrument for venting Arise The
40. ing the Gas Lines Building the gas supply lines from the supply cylinders to the GC includes connecting the gas lines to the supply cylinders and adding any traps or filters to the line Connecting the Gas Supply Cylinders To properly connect the gas lines to the gas tanks you will need the following materials Are 60 1 8 in copper tubing a tubing cutter 1 8 in Swagelok nuts and front and back ferrules two 7 16 in X 3 8 in open ended wrenches Teflon tape Secure gas cylinders to an immovable structure or wall Handle and secure all gases according to local safety regulations Use the following procedure to connect regulators and tubing to the gas supply tanks 1 Make sure the initial supply valves are turned off 2 Connectthe regulator to the gas supply tank Fuel gas regulators have reverse threads Use an open ended wrench or adjustable Crescent wrench to tighten the regulator connection 3 Determine the length of tubing you need Use only enough tubing to connect the instrument to the gas supplies but allow enough slack to allow the GC to be moved at least 40 cm 16 in from other equipment This allows enough room to perform column connections and system maintenance Also be sure to account for tee connections 4 Use a tubing cutter to cut the tubing Site Preparation and Installation Manual Chapter 4 Building the Gas Lines Plumbing the Gas Supply to the GC 5 Use a Swagelok tube fitting to attach
41. installation and maintenance of your TRACE system Some of them are supplied with the instrument e two 16 X 17 mm open ended wrenches two 12 X 14 mm open ended wrenches two 8 X 10 mm open ended wrenches two 7 X 6 mm open ended wrenches two 7 16 X 3 8 in open ended wrenches 1 and 2 cross recessed Phillips screwdrivers small and large common screwdrivers wire strippers 6 in adjustable Crescent wrench elbow Channelok pliers a pen light steel rule tape Teflon tape tweezers 12 in adjustable Crescent wrench 1 8 in tubing cutter Site Preparation and Installation Manual 41 Chapter 1 Laboratory Preparation Supplemental Equipment Thermo Scientific GFM Pro Flowmeter or equivalent Allen wrench set including 1 5 mm 2 mm 2 5 mm 3 mm 4 mm 5 mm and 6 mm wrenches Supplemental Equipment The following items are supplemental equipment necessary for the installation of your TRACE system Some of them are supplied with the instrument 2 NOTE 42 carrier gas refer to Gas and Plumbing Requirements on page 31 for information about the carrier gas requirements for your gas chromatograph detector gas refer to Gas and Plumbing Requirements on page 31 for the detector gas requirements for your gas chromatograph dual stage gas cylinder pressure regulators with an output pressure range of 0 1000 kPa 0 145 psi stainless steel diaphragms and a 1 8 in Swagelok terminate nut sample syringe 5 or
42. is Manual The Glossary lists and defines terms used in this guide and the Site Preparation Help File This includes are abbreviations acronyms metric prefixes and symbols The ndex contains an alphabetical list of key terms and topics in this guide with cross references and the corresponding page numbers Conventions Used in This Manual The following symbols and typographical conventions are used throughout this manual Bold Italic Monospace Monospace Bold KEY NAME Awe NOTE Bold text indicates names of windows menus dialog boxes buttons and fields Italic text indicates cross references first references to important terms defined in the glossary and special emphasis Monospace or Courier indicates filenames and filepaths or text the user should enter with the keyboard Monospace Bold indicates messages or prompts displayed on the computer screen or on a digital display This symbol illustrates menu paths to select such as File Open Bold uppercase sans serif font indicates the name of a key on a keyboard or keypad such as ENTER This symbol alerts you to an action or procedure that if performed improperly could damage the instrument This symbol alerts you to important information related to the text in the previous paragraph Site Preparation and Installation Manual Conventions Used in This Manual WARNING eppbbpPbbpb gt Site Preparation and Installation Manua
43. l About This Manual This symbol alerts you to an action or procedure that if performed improperly could result in damage to the instrument or possible physical harm to the user This symbol may be followed by icons indicating special precautions that should be taken to avoid injury This symbol indicates electric shock hazard This symbol indicates danger from hazardous chemicals This symbol indicates danger from high temperature surfaces or substances This symbol indicates a fire hazard This symbol indicates an explosion hazard This symbol indicates a toxic hazard This symbol indicates the presence of flammable materials This symbol indicates the presence of radioactive material This symbol indicates an operation or procedure that must NOT be performed by the user A Thermo Fisher Scientific authorized Customer Support Engineer must perform this procedure Xi About This Manual gt oO e 2 oo Instrument Markings and Symbols This symbol indicates all metal objects such as watches Jewels etc must be taken off This symbol indicates an eye hazard Eye protection must be worn This symbol indicates the user must wear a protective screen when performing the procedure This symbol indicates the user must wear protective shoes when performing the procedure This symbol indicates the user must wear protective clothing when performing the procedure This symbol indicates the user mu
44. l flow through the ground system via the ground loop that is formed Wall Outlets The power cable from the TRACE GC Ultra is approximately 3 m 9 ft long 2 NOTE Use an Uninterruptible Power Source UPS to protect your data system against possible loss due to power outages At this time we do not recommend using a UPS with the GC 24 Site Preparation and Installation Manual Chapter 1 Power Requirements Laboratory Preparation Most UPS systems are not designed to provide high quality distortion free power for scientific equipment The 120 V ac systems are fitted with U S standard National Electronics Manufacturers Association NEMA 5 20P power plugs A NEMA 5 20P power plug and its corresponding outlet are rated at 20 A and 125 V ac The power cables from the personal computer monitor and printer are approximately 2 m 6 ft long The 120 V ac systems are fitted with NEMA 5 15P plugs For optional instruments the plug requirements may vary Refer to your product s user manual for specifications Figure 1 1 shows the NEMA power plugs and outlets Outlet Plug Outlet Plug NEMA 5 20P NEMA 5 15P Figure 1 1 NEMA 5 20P and 5 15P Power Plugs and Outlets 120 V ac The 230 V ac systems are fitted with Shuco German type power plugs Plug requirements are dictated by country Figure 1 2 shows the Shuco power plug and outlet Outlet Plug Shuco Plug Figure 1 2 Shuco 230 V ac Power Plugs and Outlets Site Preparatio
45. mbing Requirements in Chapter 1 Laboratory Preparation for the type of gas appropriate for each detector Site Preparation and Installation Manual Chapter 4 Connecting the Gas Supply to the Detector Manifold Plumbing the Gas Supply to the GC Table 4 1 Detector Gas Connections Connect Connect Connect Installed Connect Connect Air WEL CRU 9 Sheath Reference Detector Module Hydrogen to to Gas to Gas to Gas to AB Gas 2 Gas 1 FID AC Gas 2 Gas Gas 3 AD Gas 3 Gas 1 AA Gas 3 ECD AB Gas 2 AC Gas 3 AD Gas 3 NPD AD Gas 2 Gas 1 Gas 3 AB Gas 2 Gas 1 PID AC Gas 3 Gas 1 AD Gas 3 Gas 1 AB Gas 2 Gas 1 FPD AC Gas 2 Gas 1 Gas 3 AD Gas 3 Gas 1 TCD AB Gas 2 Gas 1 PDD Dedicated Inlet a For ECD detectors the makeup gas is N or 5 Ar CH b For PID detectors the sheath gas is N or He c FPD applications typically do not require Make up gas d For PDD detector requires Helium must have a minimum purity of 99 999954 with lt 20 ppm Ne impurity For trace analysis of fixed gases it is strongly recommended 99 9999 purity helium with lt 0 5 ppm Ne The discharge and the carrier gases must always flow through the helium purifier Site Preparation and Installation Manual 65 Chapter
46. multiple traps install the moisture trap closest the gas supply before the hydrocarbon and the oxygen trap Hydrocarbon Traps Hydrocarbon traps remove organic materials from gases If you are using multiple traps install the hydrocarbon trap after the moisture trap but before the oxygen trap Oxygen Traps Oxygen content in the carrier and gas lines should be less than ppm To achieve a level of oxygen of less than ppm install an oxygen removing trap in the carrier gas line between the gas tank and the GC If you are using multiple traps the oxygen trap should be the last trap in the series Purity Requirements The following sections describe the minimum requirements for gases used in your gas chromatograph system Always consider using UHP 99 999 pure gases when available Helium For carrier gas 99 995 high purity with less than 1 0 ppm each of water oxygen and total hydrocarbons after purification One full size tank that has an outlet pressure of 400 700 kPa 60 100 psi Use a regulator with a CGA 580 connection or equivalent Use water oxygen and hydrocarbon traps When PDD is used helium must have a minimum purity of 99 999965 with lt 20 ppm Ne impurity When MS detector is used please refer to the relevant User s Guide 34 1 99 995 gas requires the use of appropriate traps Site Preparation and Installation Manual Chapter 1 Gas and Plumbing Requirements Laboratory Preparation Hydrogen
47. n and Installation Manual 25 Chapter 1 Laboratory Preparation Power Requirements CAUTION You should never connect the TRACE GC Ultra and its peripheral devices to the A same electrical wall outlet You will run the risk of overloading the circuit Tables 1 3 and 1 4 show maximum current requirements for the TRACE GC Ultra system and optional instruments Table 1 3 Maximum Current Requirements for the TRACE Standard System 120 V Current 230 V Current Requirement Requirement Instrument in amperes in amperes TRACE GC Ultra 16 10 computer 4 2 monitor 2 Total 26 13 Table 1 4 Maximum Current Requirements for TRACE Optional Instruments 120 V Current 230 V Current Requirement Requirement Instrument in amperes in amperes GC Autosamplers AI 3000 AS 3000 0 6 0 3 TriPlus 2 1 Printers ink jet printer plotter 1 8 0 9 laser printer 7 3 5 pe NOTE Other peripheral devices such as purge and trap units may be connected to the TRACE GC Ultra These peripheral devices may have additional power requirements Consult your local Thermo Fisher Scientific CSE for assistance 26 Site Preparation and Installation Manual Chapter 1 Operating Environment Laboratory Preparation Technical Assistance Occasionally unacceptable quality in line power sources may adversely affect the operation of a TRACE system It is the user s responsibility to correct line voltage problems Spe
48. nal sheets if necessary Fax or mail this form to Thermo Fisher Scientific S p A Strada Rivoltana km 4 20090 Rodano MI ITALY Fax 39 02 95059388 82 Site Preparation and Installation Manual A A ac ADC B b B baud rate C C CIP cm CPU CSE DAC dc DS Site Preparation and Installation Manual Glossary This glossary lists and defines terms used in this guide Included are abbreviations acronyms metric prefixes and symbols ampere alternating current analog to digital converter bit byte 8 b data transmission speed in events per second Celsius Carriage and Insurance Paid To centimeter central processing unit of a computer Customer Service Engineer depth digital to analog converter direct current data system 83 Glossary 84 ECD EMC ESD F FID FOB FPD ft GC GND H h h harmonic distortion HV Hz Electron Capture Detector electromagnetic compatibility electrostatic discharge Fahrenheit Flame Ionization Detector Free on Board Flame Photometric Detector foot gram gas chromatograph electrical ground height hour A high frequency disturbance that appears as distortion of the fundamental sine wave high voltage hertz cycles per second Site Preparation and Installation Manual l IEC impulse in T O LAN lb LED Site Preparation and Installation Manual Glossary International Electrotechnical Commission
49. nstallation Manual Chapter 1 Power Requirements Laboratory Preparation Power Requirements It is your responsibility to provide an acceptable source of power You should provide a dedicated power line with a circuit breaker capable of withstanding the power ratings listed in the following sections 120 V ac Power Requirements For TRACE GC Ultra systems installed in the US and other countries using 120 V ac power the minimum power requirements are as follows e 120 V ac 6 1096 frequency of 50 60 Hz 2 Hz e three duplex outlets single phase power with a minimum power rating of 16 A at each duplex outlet e earth ground hard wired to the main panel 230 V ac Power Requirements For TRACE GC Ultra systems installed in countries using 230 V ac power the minimum power requirements are as follows e 230 V ac 10 frequency of 50 60 Hz 2 Hz e three duplex outlets with a minimum power rating of 10 A at each duplex outlet earth ground hardwired to the main panel Power Quality The quality of power supplied to your TRACE GC Ultra system must be stable and within the specifications listed in this guide The line voltage must be free of fluctuation due to slow changes in the average voltage surges sags transients and harmonics Site Preparation and Installation Manual 23 Chapter 1 Laboratory Preparation Power Requirements The TRACE GC Ultra system operates in an Overvoltage Category II environment
50. o split the gas flow Use the following procedure to connect a Swagelok tee or cross 1 Use a tubing cutter to cut gas supply tubing where you want to install the tee or cross 2 Connect the tubing to the tee or cross with a Swagelok fitting as described in How to Use Swagelok Tube Fittings on page 54 3 Measure the distance from the tee or cross to the inlets or detectors and cut tubing in the appropriate lengths 4 Connect the tubing to the tee or cross ends with Swagelok fittings 5 Install Swagelok caps on any open ends of the tee or cross that you do not plan to connect with tubing 56 Site Preparation and Installation Manual Plumbing the Gas Supply to the GC gen This chapter explains the proper gas supply connections and configuration information for your GC Use the information in this chapter to connect your inlet and detector gas supplies Chapter at a Glance Warme o O 58 Connecting the Gas Supply Cylinders c cccssoveesccestscseonscensectensseaecaasaaaecs 60 Adding Traps to the Gas SODpI cusco eret eet ee rte t sonics 61 Pomma Gas Li A E a 62 Connecting the Gas Supply to the Inlet Manifold sess 63 Connecting the Gas Supply to the Detector Manifold sess 63 Connecting Coolant to the Cryogenic Systemin ercsi eraon 66 Teine od A cing aiid An Ss cate ene 67 Before you begin plumbing gas to the GC locate the main power cable and positi
51. on it near the appropriate outlet Do not plug in the power cable at this time Follow the procedures in this section to build your gas supply lines for carrier make up and detector gases DO NOT loosen or remove caps from the GC until you have purged your gas lines and are ready to connect them Loosening or removing caps early will contaminate instruments and filters Site Preparation and Installation Manual 57 Chapter 4 Plumbing the Gas Supply to the GC Using Hydrogen WARNING Hydrogen is a dangerous gas that when mixed with air may create an explosive A mixture The use of hydrogen as a carrier gas requires the operator s extreme caution Special precautions must be taken because of the risk of explosion The gas chromatograph must be equipped with a hydrogen sensor if you use hydrogen as a carrier gas Refer to Using Hydrogen on page 58 for more information Using Hydrogen The use of hydrogen as a carrier gas or as fuel for certain flame detectors requires the operator s strict attention and compliance with special precautions due to the hazards involved Hydrogen is a dangerous gas particularly in an enclosed area when it reaches a concentration corresponding to its lower explosion level 4 in volume When mixed with air it can create an explosive mixture An explosion hazard could develop in the GC oven when hydrogen is used as a carrier gas if oven elements are not perfectly connected to each other or if the connection mat
52. oooocincninccionnonoconanannnonnoncononocononanonnnos 63 Connecting Coolant to the Cryogenic System sse nennen 66 Column Oven Cryo System sss 66 PTV Injector Cryo System sse eene nennen 66 Testing tor Leaks sacs e ere hee RR UR RR e EE ee Ue Re Rd pe eee bese 67 Chapter 5 Getting Connected meini 69 A e a De a a A sieges RN 69 Remove the TRACE C OVET a aaa ever A E S 69 Connect the Data System sse nre 70 Connect the Computer Cable sess 7 Connect the LAN Cable ice eterne tr eI e er cete 72 Connect the Autosampler nre ii 73 Computing nteprators rede AA A RE E ER pude eden Heard 76 SystemiP Ow er Up ene tote E iom A tees 77 Appendix A Preinstallation Checklist cronos ta ct 79 Appendix B jer Ide unici iia 81 How To ContactUs cue bieten te ope torte lili 81 Reader UV 5 tte e i et ee A Uer perde ete n P eR edes 82 Cc qe 83 Site Preparation and Installation Manual Vii Contents vill Site Preparation and Installation Manual About This Manual Overview This manual will help you prepare for the arrival and installation of your TRACE GC Ultra gas chromatograph Following the procedures and guidelines in this manual will help the installation process go smoothly This manual is organized as follows Section I Site Preparation contains the information you need to prepare for the arrival of your TRACE GC Ultra gas chroma
53. p the data system computer or integrator close to the electrical connections we recommend placing it on the right side of the GC Use Table 1 1 to determine the minimum space and weight requirements for a standard TRACE system Site Preparation and Installation Manual 21 Chapter 1 Laboratory Preparation Space and Load Requirements Table 1 1 Minimum Space and Weight Requirements TRACE Standard System Width Instrument TRACE GC Ultra 44 8 18 61 24 66 30 48 105 computer 42 16 5 15 6 6 5 44 5 17 5 12 27 monitor 42 16 5 40 16 43 17 3 7 keyboard 2 5 1 47 18 18 7 1 2 Use Table 1 2 to determine the space and weight requirements for optional instruments for your TRACE system Table 1 2 Space and Weight Requirements TRACE Optional Instruments Width Instrument Autosamplers AI 3000 AutoInjector 40 16 23 9 25 10 6 13 AS 3000 Autosampler 40 16 40 16 25 10 6 13 TriPlus 66 7 26 3 87 343 77 3 304 25 56 TriPlus extended version 66 7 263 122 48 77 3 30 4 25 56 Printers inkjet printer plotter 20 8 43 17 38 15 4 8 laser printer 29 12 42 17 41 16 20 44 Pe NOTE Other peripheral devices such as purge and trap units may be connected to the TRACE GC Ultra These peripheral devices may have additional bench space requirements Consult your local Thermo Fisher Scientific Customer Support Engineer CSE for assistance 22 Site Preparation and I
54. r TRACE GC Ultra system It also includes a list of tools and supplies you need to install your gas chromatograph The information in this chapter will help you plan construct and fully equip your laboratory Your laboratory must meet the requirements for power exhaust systems and environmental conditions explained in this chapter before your gas chromatograph can be installed Your laboratory preparations should be completed before the TRACE GC Ultra is unpacked Complete any work that creates dust high humidity or corrosive vapors before you begin unpacking the instruments Chapter at a Glance A 20 Space amd Load Requierente cascine tete te eec aair E e nee edens 21 Power REQUIEM oi a een eor ea dea NEE cR RS 23 Operating Epi innit e 27 Exhaust Sy syed 31 Cas and Filumbing Bequirereitte ee eet terea a 31 E C 41 rpm 41 Supplemental Egi pmetit eee te secte Per essc terere eer nao enu tone erac ea ENNS 42 Hardware and Software Minimum Requirements eese 43 Site Preparation and Installation Manual 19 Chapter 1 Laboratory Preparation pe NOTE Entrance Entrance TRACE GC Ultra systems operate reliably under controlled environment conditions Operating or maintaining a system outside the specifications outlined in this guide may cause many different types of system failures The r
55. r damage and verify the completeness of shipment This will protect you in the event of missing or damaged components Freight insurance requires that obvious damage be noted on the receiving documents Site Preparation and Installation Manual Chapter 2 Installation Instrument Arrival Installation If you have purchased the installation option you must submit a completed Preinstallation Checklist Appendix A to us before a Thermo Fisher Scientific CSE will install your system The CSE will unpack and completely install the system including optional instruments The CSE will also initialize settings verify that the system is operating according to specifications and familiarize you with the system Contact your local Thermo Fisher Scientific office for details If you have not purchased the installation option refer to Section II Installation for installation instructions Site Preparation and Installation Manual 47 Chapter 2 Instrument Arrival Training 48 Training Valuable training on Thermo Fisher Scientific instruments and software is offered worldwide Experience has shown that maximum value can be derived from a scientific instrument if there is one person who has a major responsibility for the instrument We recommend that you designate a key operator to manage the operation and maintenance of the TRACE system We also recommend that the key operator receive training at the Thermo Fisher Scientific Insti
56. sponsibilities air conditioning 27 operating environment 27 quality of power 27 spares 42 V Vibration 30 Voltage disturbances harmonics 24 sags and surges 24 slow average 24 transients or impulses 24 Voltages 120 V ac 23 230 V ac 23 USA 23 Worldwide 23 92 Workbench 22 GC connections 26 optional instruments 22 TRACE system 22 Wide bore columns Site Preparation and Installation Manual Figures Figure 1 1 Figure 1 2 Figure 1 3 Figure 1 4 Figure 1 5 Figure 3 1 Figure 3 2 Figure 3 3 Figure 4 1 Figure 5 1 Figure 5 2 NEMA 5 20P and 5 15P Power Plugs and Outlets 120 V ac oooooncnincnincnicnconncannonnoons 25 Shuco 230 V ac Power Plugs and Outlets sss 25 Dual Stage Gas Regulator esses enne enne nennen enne 36 No Tank ON SUP ANON o ooo coto eae teh e ues ROMA cU Ue Minn tad naive rd scs sitos PALA 39 CO Tank Configuratio sninn aia oa 40 Swagelok Ferrule and Nut Assembly ccccccccesceessesecseeeeeeeceeeceseeeseecseceseeeeeeneeaaees 54 Swagelok and Inlet Connection sese 55 Tightening Swagelok Fittings sese 25 Gas Trap Configuration x5 eoe eer t pe rar d pee e AR teen aber 61 Removing the TOp COVet iei ettet ert dada cdt 70 Integrator Cable Connections eese enne enne nnne ns 76 Tables Table 1 1 Table 1 2 Table 1 3 Table 1 4 Table 1 5 Table 1 6 Table 1 7 Table 1 8 Table 4 1 Minim
57. st wear gloves when performing the procedure Instrument Markings and Symbols The following table explains the symbols used on Thermo Fisher Scientific instruments Only a few of them are used on the TRACE GC Ultra gas Xii chromatograph Symbol Description Direct Current Alternating Current Both direct and alternating current Three phase alternating current Site Preparation and Installation Manual Instrument Markings and Symbols About This Manual Symbol Description Earth ground terminal Protective conductor terminal Frame or chassis terminal Equipotentiality On Supply Off Supply Equipment protected throughout by DOUBLE INSULATION or REINFORCED INSULATION Equivalent to Class II of IEC 536 Indicates that the user must refer to the manual for specific Warning or Caution information to avoid personal injury or damage to the product Caution risk of electric shock Caution hot surface Caution refer to accompanying documents In position of a bistable push control PPP puo Or Site Preparation and Installation Manual xiii About This Manual Instrument Markings and Symbols Description Out position of a bistable push control Symbol in compliance to the Directive 2002 96 EC on Waste Electrical and Electronic Equipment WEEE placed on the european market after August 13 2005
58. th 1 8 in Swagelok fittings If the GC has been furnished with a 2x1 mm tube use the ferrule for 2x1mm 1 8 in provided in the start up kit Repeat the process if you are connecting the carrier gas supply to more than one inlet Be sure to finish all inlet connections before turning on the gas supply Inside each 1 8 in inlet manifold is a fritted filter If you connect a gas line smaller than 1 8 in to an inlet be sure the tubing does not touch the filter when inserted into the manifold To keep the gas line from touching and possibly damaging the filter extend the tubing only 5 mm past the front ferrule This will ensure the tubing does not touch the filter Refer to How to Use Swagelok Tube Fittings in Chapter 3 Before You Begin for more instructions on using Swagelok ferrules and fittings Connecting the Gas Supply to the Detector Manifold Before you connect the detector gas line you should have done the following e built and connected the gas supply lines for your inlet gas supply as described in Building the Gas Lines in this chapter Site Preparation and Installation Manual 63 Chapter 4 Plumbing the Gas Supply to the GC Connecting the Gas Supply to the Detector Manifold ye NOTE A pe NOTE 64 completely built your detector gas line from your gas supply including any traps tees and extra tubing to allow about 40 cm 16 in of slack in the line as described in Building the Gas Lines in this chapter e
59. the LAN Cable 70 Site Preparation and Installation Manual Chapter 5 Introduction Getting Connected Connect the Computer Cable Perform this operation when the CPU standard board is installed into the GC a Connect the RS232 cable supplied with the data system to one of the COM ports of your computer taking note of which port you use You will need this information for software configuration b Connect the other end of the cable to the port on the top of the GC labeled COMPUTER Computer Port c Snap the cable securing ring into one of the cable slots on the GC Site Preparation and Installation Manual 71 Chapter 5 Getting Connected Introduction Connect the LAN Cable Perform this operation when the CPU LAN board is installed into the GC 2 NOTE For the LAN set up refer to the TRACE GC Ultra LAN Set up User Guide a With the top cover removed connect the LAN gray cable provided to the port on the top of the GC labeled LAN as o A lt LAN e Bs 72 Site Preparation and Installation Manual Chapter 5 Introduction Getting Connected b Connect the other end of the cable to the LAN communication port of your computer c Snap the cable securing ring into one of the cable slots on the GC 4 Turn on the GC if it is not already on the computer the monitor and the printer Check the time date and monitor settings 5 Test the printer
60. this manual or the information contained therein even if this information is properly followed and problems still arise This publication is not part of the Agreement of Sale between Thermo Fisher Scientific and the purchaser of aTRACE GC Ultra system In the event of any conflict between the provisions of this document and those contained in Thermo Fisher Scientific s Terms and Conditions the provisions of the Terms and Conditions shall govern Reference to System Configurations and Specifications supercede all previous information and are subject to change without notice Trademarks TRACE GC Ultra is a trademark of Thermo Fisher Scientific Inc and its subsidiaries Other brand and product names may be trademarks or registered trademarks of their respective companies Declaration Manufacturer Thermo Fisher Scientific Thermo Fisher Scientific is the manufacturer of the instrument described in this manual and as such is responsible for the instrument safety reliability and performance only if e installation e re calibration e changes and repairs have been carried out by authorized personnel and if e the local installation complies with local law regulations e the instrument is used according to the instructions provided and if its operation is only entrusted to qualified trained personnel Thermo Fisher Scientific is not liable for any damages derived from the non compliance with the aforementioned recommenda
61. tions About This Manu l z reet cti idas ix OvVeltVie Wen oso NE atatum uisu E UE E ix Conventions Used in This Manual essent x Instrument Markings and Symbols cccccscsseessesseceeeceseeeseecsecesecnseeeeceseeeseesaecaeceeeeeeeeeeeeaaes xii Using the TRACE GC Ultra Document Set sss enne XV SECTION I Site Preparation Chapter 1 Laboratory Pr iras 19 EMO rt it A AA AAA di 20 Space and Load Requirements sess rre 21 Power Requirements rtt ett step tend eee race bel eles a eed o Dot aereo pedes 23 120 Vac Power Requirements tee de RE THER y PEE eiue ed 23 230 V ac Power Requirements sess ener nne 23 Power Quality a tee ere A aee tere dete dps 23 Effects of Voltage Disturbances sse 24 Wall Outlets tucan PRU HE Rod Pe OI ERI ie ces trees 24 T chnical Assistarce 5 3 5 nens enirn a ee abus 27 Operating Environment eene eene nnne ron nn corn enne non rr corran naar non nennen entes 27 sud 27 Cooling Requirement seinien erie eene 27 Over Exhaust ion o oin ei boe RR e et ect E 29 ATIC ati AA A e Ad A AAA at toas 29 H midity i sss be tis mde i MS 29 Particulate Matter i ded IM MR e C bed ar ete deed 29 El ctrostatic Discharge e eta tense iiie iere des Ce e INTO ade EUR EE Sad IR S 30 Alcor 30 Light ng endete ato Ree materie ae Fei P RR E toe ad 30 Other Environmental Factors essere enne enne nnne nnne 31 Exh
62. tograph system In addition to laboratory and operating environment standards this section contains information about the shipping and delivery of your TRACE system Chapter 1 Laboratory Preparation gives you all the information you need to prepare your laboratory for the arrival of your TRACE system Chapter 2 Instrument Arrival gives an overview of shipping and receiving procedures installation and training Section II nstallation contains information to help you set up and install your TRACE GC Ultra gas chromatograph Chapter 3 Before You Begin contains unpacking instructions and provides a brief tutorial of basic gas plumbing operations you must know before you can continue the installation process Chapter 4 Plumbing the Gas Supply to the GC explains the proper gas supply connections and configuration information for your GC Chapter 5 Getting Connected explains the TRACE power connections and helps you mount and configure peripheral devices and data systems Appendix A Preinstallation Checklist contains a step by step list of items you should complete before your TRACE system arrives Appendix B Customer Communication contains contact information for Thermo Fisher Scientific offices worldwide Use the Reader Survey in this section to give us feedback on this manual and help us improve the quality of our documentation Site Preparation and Installation Manual ix About This Manual Conventions Used in Th
63. tute at your site or at one of the local Thermo Fisher Scientific offices For information on courses or enrollment contact your local Thermo Fisher Scientific office Thermo Fisher Scientific S p A Strada Rivoltana km 4 20090 Rodano MI ITALY 39 02 95059 355 Site Preparation and Installation Manual If you have purchased the installation option for your system a Thermo Fisher Scientific authorized CSE Customer Support Engineer will install your system If you have not purchased the installation option this section will help you set up your TRACE GC Ultra Included in this section are basic installation procedures you will need to be familiar with before you can install the GC Chapter 3 Before You Begin contains unpacking instructions and provides a brief tutorial of basic gas plumbing operations you must know before you can continue the installation process Chapter 4 Plumbing the Gas Supply to the GC explains the proper gas supply connections and configuration for your GC Chapter 5 Getting Connected explains the TRACE power connections and helps you mount and configure peripheral devices and data sys tems SECTION Installation Before You Begin This chapter contains unpacking instructions and a brief tutorial of basic gas plumbing operations you must know before you can continue the installation process Chapter at a Glance Verity Site Preparato ees erc a erroe Pe D eae ia a 51 Vaprck
64. ulties preparing your site for installation Ensure that entrances and hallways are at least 92 cm 37 in across Ensure that your workbench space is at least 1 5 m 4 5 ft across and can support a 75 kg 170 Ib load Ensure that the instruments can be placed on surfaces that do not vibrate Ensure that your work area has proper lighting Provide an acceptable power source for your TRACE system Provide a wall outlet configuration that meets specifications Ensure that the temperature in your laboratory is between 18 and 35 C 65 and 95 F Ensure that the relative humidity level in your laboratory is between 40 and 80 with no condensation EL Oo Pee ao E Ensure that the air in your laboratory is free of excessive dust smoke or other particulate matter Site Preparation and Installation Manual 79 Appendix A Preinstallation Checklist Ensure that your system is free of electrostatic discharge Provide an adequate exhaust system Ensure that you meet appropriate gas line requirements EJ T er ol Install a telephone near your system Fax or mail the TRACE Preinstallation Checklist to the appropriate Thermo Fisher Scientific office You can refer to Appendix B Customer Communication to select the Thermo Fisher Scientific office nearest you 80 Site Preparation and Installation Manual Customer Communication Thermo Fisher Scientific provides comprehensive technical assistance worldwide and is de
65. um Space and Weight Requirements TRACE Standard System 22 Space and Weight Requirements TRACE Optional Instruments sse 22 Maximum Current Requirements for the TRACE Standard System 26 Maximum Current Requirements for TRACE Optional Instruments 26 Heat Output TRACE Standard System ccccceecccesccesceeeeeseesseceeeeeeeeeeeeeesaeeneeseeeeaes 28 Heat Output TRACE Optional Instruments essere 28 Capillary and Wide Bore Column Gas Recommendations sss 32 Packed Column Gas Recommendations esee 33 Detector Gas Connections ce ite Rr re tee eed ete i eene be Pop gt 65
66. ur integrator and connect it to a power supply according to the manufacturer s instructions Each TRACE GC Ultra comes with shielded two conductor cables for each detector analog output and a remote ready inhibit start and stop cable handshake cable The two conductor cables consist of a blue positive wire a white negative wire and a shield wire The handshake cable has a DIN connector on one end for connection to the rear panel of the GC and leads on the other for connection to the integrator The blue positive and white negative wires of the two conductor cable connect the detector output and the integrator Figure 5 2 shows the cable connections Integrator Blue Positive Wire TRS Cable To GC GC Detector Board OV 1V 10V GND White Negative Wire Shield Wire White Negative Wire V Blue Positive Wire Two conductor cable Shield Wire Cut or insulate the shield wire connector when using an integrator or a data handling system If a recorder is used connect the shield wire to the recorder ground Figure 5 2 Integrator Cable Connections Refer to your integrator s operating manual for instructions on using the integrator to process information from the GC Chapter 5 System Power Up Getting Connected System Power Up By now your lab should be equipped with the proper power supply and outlets as described in
67. utosampler or data system you will need to unpack them and follow any setup instructions included with them Follow the instructions in the sections below to connect your peripheral devices to the GC Remove the TRACE Cover Before you can connect a data system or an autosampler you must first remove the top cover of the GC using the following procedure 1 Lift the detector cover off the GC top cover 2 Open the oven door and unscrew the two top cover fastening screws 3 Push the top cover back about cm and lift it up and off the GC Figure 5 1 illustrates this procedure Site Preparation and Installation Manual 69 Chapter 5 Getting Connected Introduction Figure 5 1 Removing the Top Cover Connect the Data System There are two types of information transfer techniques between the TRACE GC Ultra and the computer digital and analog Use the following procedure to connect your data system computer to the GC for digital transfer 1 Connect the computer monitor keyboard mouse printer if applicable and network connection if applicable according to the instructions in the accompanying documentation 2 Plug in the computer monitor and printer power cables 3 Connect your GC to the Thermo Fisher Scientific Corporation Data systems ChromQuest Chrom Card etc according to the CPU board standard or LAN installed into the GC performing the relevant cables connection Connect the Computer Cable Connect
68. xiv Site Preparation and Installation Manual Using the TRACE GC Ultra Document Set About This Manual Using the TRACE GC Ultra Document Set The TRACE GC Ultra Document Set CD Rom PN 317 095 00 includes all manuals in electronic format and serves as your library for information about the TRACE hardware and software The TRACE GC Ultra Document Set PN 317 093 00 as paper copy is also available Furthermore Thermo Fisher Scientific part numbers PN for the paper copy manuals are provided for each book title Site Preparation and Installation Manual PN 317 091 90 This manual and diskette describes how to set up a workspace for the TRACE GC Ultra and how to connect the GC to the gas supplies and peripheral devices Acceptance Package PN 317 092 20 This folder contains required shipping documents and quality report forms Getting Started PN 317 092 30 This guide contains procedures for checking configuration installing detectors and making a first analysis with the TRACE GC Ultra Operating Manual PN 317 091 70 This manual provides descriptions of the TRACE GC Ultra hardware and software and instructions for their use UFM Ultra Fast Module Device PN 317 093 98 This manual provides descriptions of the TRACE GC Ultra equipped with the UFM device and instructions for it use Quick Reference Card PN 317 092 40 This reference card contains guidelines for carrier gas use and injection procedures K Factor Quick Reference
69. ype Carrier Gases Fuel Gases Make Up Gas FID Helium Hydrogen Air Not needed Nitrogen Argon NPD Helium Hydrogen Air Nitrogen Nitrogen ECD Nitrogen None Same as carrier gas Argon 5 CH if necessary FPD Helium Hydrogen Air Not needed Nitrogen Argon TCD 2 Helium None Not needed Nitrogen Hydrogen Argon PDD Helium is the gas used for PDD discharge and carrier supply 1 Maximum delivery pressure 1000 kPa 145 psi for carrier gas 420 kPa 60 psi for detector fuel gases 2 Oxygen trap should be installed in gas lines If possible you should use only instrument or chromatographic purity grade gases in your TRACE GC Ultra Traps UHP Ultra High Purity gases should not contain impurities above ppm Impurities below 1 ppm generally do not require purification Gases with higher impurity levels may require oxygen and hydrocarbon traps A number of GC supply and accessory companies carry a variety of traps for gases See Figure 4 1 Gas Trap Configuration on page 61 for the correct trap installation sequence For more information on gas purification contact your local Thermo Fisher Scientific CSE Site Preparation and Installation Manual 33 Chapter 1 Laboratory Preparation Gas and Plumbing Requirements Moisture Traps Water in the carrier or fuel gas may damage the gas chromatograph column and contaminate the TRACE system Water content should be less than 1 ppm in all cases If you are using

Download Pdf Manuals

image

Related Search

Related Contents

医療機器の添付文書の記載要領の改正について    A9GT-80STAND User`s Manual  User-/service manual  BoConcept 2 Assembly Instruction  NI 6509 User Guide and Specifications  Tripp Lite U206-006-R User's Manual  - s3.amazonaws.com  Fire Magic REGAL II 82 User's Manual  UC-7112-LX Plus Software User`s Manual  

Copyright © All rights reserved.
Failed to retrieve file