user manual - PONSEL Instrumentation for water quality
Contents
1. 65 PAGE 9 9 CE SET PONSEL ACTEON AQUA ACTEWIN MESURE D USER MANUAL et REF PONKITAQUA 1 CASE CONTENTS The complete case comes with the following items 1 AQUA probe with its sensors connected and its protection case 2 Acteon 3000 portable terminal 9 Ballasted strainer 4 AQUA probe cable 15 20 or 30m depending on client s order 5 Actewin adapter SDI12 RS232 6 Calibration standard pH 4 01 pH 7 01 f Calibration standard for conductivity 1413uS cm 12 88 mS cm 8 Calibration standard for ORP 240mV 9 Electrolyte oxymetric standard for D O sensor 10 Sodium sulphite Zero O2 11 Case with 2 spare membranes for the dissolved oxygen sensor 12 CD ROM with ActeWin software and user manual in pdf format 13 Acteon and AQUA connection cables for adapter SDI12 RS232 14 Waterproof obturating plugs for sensors disconnected 2 PAGE 10 10 CE SET PONSEL ACTEON AQUA ACTEWIN MESURE INDEX 1 USER MANUAL REF PONKITAQUA 2 ACTEON PORTABLE TERMINAL 2 1 KEYBOARD DESCRIPTION Impulsional output Opto insulated potential free switch Backlight On Off Lighting 15 sec Navigation keys Enter Save Validate Up Down Left Right Delete Erase Escape previous screen Alphanumeric keys SDI 12 connection for AQUA probe and PC PONSEL MESURE msi
2. PAGE 11 11 DATE January 09 11 SET ACTEON AQUA ACTEWIN USER MANUAL Ded REF PONKITAQUA 2 2 POWERING UP THE DEVICE Push button CD in the upper left corner of the keyboard A homepage will show for 10 seconds before the Main Menu appears To push on any key to interrupt the homepage XS MAIN MENU gt Configuration gt Measure gt Calibration gt Results gt History 02 20 05 14 00 d 2 3 CONFIGURATION CONFIGURATION gt Languages gt Date Hour gt Units gt Parameters gt Memory gt Recording 02 20 05 14 01 ial 2 3 1 LANGUAGES LANGUAGES gt Fran ais gt English gt Deutch gt Espanol 2 taliano 02 20 05 14 03 f7 Important when the Acteon is running an energy economizer releases automatically after 20 minutes and turns off the LCD Just push on any key so as to relight the screen You come back on the last screen of the menu in progress no data is lost If a recording has been launched with the on line AQUA it carries on without interruption On the other hand in perpetual measure mode no launched recording measures stop during the screen lights out The Act on Aqua probe only consume 20 uA in this sleeping mode The up and down navigation keys allow choosing a section the ENTER key allows confirming your choice A status bar at the bottom of the screen indicates the date and time battery capacity land the 4 levels of memory occupation C2 E
3. USB2 0 DRS232 Adapter option PAGE 36 36 DATE January 09 11 SET ACTEON AQUA ACTEWIN USER MANUAL MESURE INDEX 1 REF PONKITAQUA 5 AQUA PROBE 5 1 Starting up the probe The AQUA multiparameter probe can be connected to 3 sensors enabling measuring up to 6 parameters Dissolved Oxygen Red sensor pH ORP Blue sensor Temperature Conductivity Grey sensor Pressure situated in the sleeve of the AQUA10 AQUA 50 5 2 Connecting the probe The AQUA probe can be connected to PC and data collection devices as well as to the ACTEON 3000 display recording terminal SDI 12 BUS via RS 232 The following figures show various configurations with its necessary equipment Probe connected to a computer SDI12 RS232 Adapter PONAQC AB7 Cable PONAQC AB6 AQUA Probe Cable Computer with COM port SDI12 RS232 adapter Software ACTEWIN Windows for calibration probe PAGE 37 37 SET DATE January 09 11 CE PONSEL ACTEON AQUA ACTEWIN men RE USER MANUAL INDEX 1 REF PONKITAQUA Probe connected to a laptop computer SDI12 RS232 Adapter PONAQC AB7 Cable PONAQC AB6 AQUA Probe Cable Computer with a COM port SDI12 RS232 Adapter Software ACTEWIN for probe calibration Probe connected to ACTEON 3000 terminal Cable PONAQCAB1 5 ING Ul AQUA Cable Acteon 3000 PAGE 38 38 DATE January 09 11 SET ACTEON AQUA ACTEWIN
4. USER MANUAL INDEX 1 REF PONKITAQUA MESURE SR 5 3 Preparing the probe 5 3 1 Installing the dissolved oxygen membrane The DO probe is shipped ready to use with the oxymetric electrolyte and a protective membrane already installed on the sensor If only the membrane rack is installed follow the next steps to put the oxygen sensor in operation 1 Unscrew the protective cap c and remove the membrane rack without touching the cathode Warning Don t put anything in the opening of the pressure stabiliser see Figure 1 2 Keep the sensor in vertical position and rinse the tank with some drops of electrolyte Fill the tank with the oxymetric electrolyte until a meniscus is forming on the top part of the sensor Make sure that no air bubbles are retained on the cathode 3 Take out the air of the membrane rack with one drop of electrolyte 4 Place the membrane rack in the tank 5 Screw the protective cap slowly back on the sensor end without forcing 6 Turn the sensor around and screw the cap gradually to take out the electrolyte excess 7 Check that no airbubbles are left on the cathode Metallic Protective jt HB Tightening disc care Membrane rack Membrane p L Cathode Hole of Membrane rack Pressure balance Oxygen 1 Sensor 3 Air Membrane rack Electrolyte p drop Protective cap 4 5 6 SET PAGE 39 39 ca PON SEL ACTEON AQUA ACTEWIN
5. exit ENTER next 02 21 05 19 11 M Aqu O CONDUCTIVITY 2 Range 0 200 0uS cm Probe in solut Standard Measure Tempera 74 14 F ESC pr c ENTER valid 02 21 05 19 13 IT Aqu 84 0uS cm 82 8uS cm CONDUCTIVITY 2 Calibration OK Message Enter your name correct J BURTON DEPT W04 calibration ESC cancel ENTER valid 02 21 05 19 14 diy Aqu CD CONDUCTIVITY 2 Calibration Out of margin Enter your name J BURTON DEPT W04 Message incorrect calibration ESC cancel ENTER valid 02 21 05 19 14 II Aqu CONDUCTIVITY 2 SAVING COEFFICIENTS PLEASE WAIT 02 21 05 19 14 dXX Aau Choose the measuring range in which you are going to work Identical calibration procedure for every range The AQUA sensor is placed in the air in order to reproduce the O uS or mS value As soon as the measure given by the sensor is stabilised for at least 30 seconds press Enter to validate the OpS calibration Choose the adapted conductivity calibration solution in accordance with the desired calibration range 0 200uS cm 84uS 0 20000S cm 1 413uS 0 20mS cm 12 880mS 0 200mS cm 112 800mS Dip the sensor in this solution As soon as the measure given by the sensor is stabilised for at least 30 seconds press Enter to validate the calibration value ex 84 0 uS cm At the end of the calibration process the name of the operator is asked 20 characters are available to enter a name Then p
6. oe MESURE INDEX 1 USER MANUAL REF PONKITAQUA 5 83 2 Installing the sensors A Unscrew the calibration cap and take it off of the probe to get access to the sensor connector Calibration cup Ballasted strainer U TEE Hc Probe connector cap 3 Connector cable cap B Take off the protective caps installed to the probe connectors by pulling them C Hold the sensors by the ring and click them to the probe connector Coloured drops corresponding to those on the sensors are placed on the AQUA probe compartment The red reference marks on the sensor and probe compartment help you to place the sensors correctly see compartment schedule Connector 5 pins Grey drop for Red reference mark Conductivit y temperature sensor Blue drop Connector 7 pins For pH ORP sensor Red drop for Oxy gen sensor Connector 4 pins Schedule AQUA probe compartment Sensor Pods The dissolved oxygen sensor is red and is connected to the 4 pin probe connector Protection sleeve Oxygen sensor The temperature conductivity sensor is grey and is connected to the 7 pin probe connector Protection sleeve Sensor Conductivity Temperature PAGE 40 40 e PO NSEL ACTEON AQUAIACTEWIN MESURE Z USER MANUAL S REF PONKITAQUA The pH ORP sensor is blue and is connected to the 5 pin probe connector Sleeve protection Platinium disc ORP PH gl
7. 5 to 64 C if D O and pH Redox sensors disconnected If not 5 to 45 C with all sensors connected Resolution 0 01 C Accuracy 0 1 C Linearization by mathematical model according to answer Pt100 DIN IEC 751 Scale ITS 90 This temperature sensor automatically compensates in calibration calibration with standard solutions and measurement modes the pH and conductivity parameters and generates the saturation table of dissolved oxygen this last parameter is compensated in measuring mode by an incorporated thermistor in the head of cell DO2 7 1 2 Conductivity parameter Sensor with a 4 slit electrodes cell without wall attachment effect 2 current electrodes and 2 potential electrodes 4 measurement ranges 0 200 uS cm 200 2000 uS cm 2 20 mS cm 20 200 mS cm with the same relative accuracy in each range 0 5 Resolutions 0 1 uS cm 1uS cm 0 01mS cm and 0 1 mS cm 1 standard by range Fixed ranges in calibration mode calibration on standards strong recommendation to work in automatic range in measurement mode for optimal accuracy Conductivity signal is compensated in temperature and brought back to the reference temperature of 25 C The temperature variation coefficient used for the compensation around 2 at 25 C is indexed on that of KCI N 100 1413 uS cm at 25 C very close to natural water and sea water Non linear function according to NF T 909 031 EN 27888 PAGE 52 52 NICE SET Ted PO
8. 6712 AVALANCHE sampler 2 You must be in Extended Programming to use the probe with the 6712 or Avalanche sampler code 6712 2 The sampler can automatically scan for the presence of sensors if no you have to select others functions then Hardware and select SDI12 sondes The screen will tell you one sonde identified 3 then you have to confirm the address of the sensor the software tells you 0 you have to confirm going on 0 and press Please don t go on done if the setup of the probe is not made 4 Then the screen will tell you the number of 6 data sets It S now very important to enter separately the different parameters and units respecting strictly their order 1 to 6 PARAMETER UNITS 2 CONDUCTIVITY microSiemens uS cm 4 ORP Redox milliVolts 6 DISSOLVED OXYGENE 5 Please tell if you want to store all the parameters or only selected parameters 6 Then the screen is going back on the probe address if setup is correct please go on done and press 7 The screen will tell you new hardware previous stored data will be erased please answer by YES for validation of the parameters 8 You may run a calibration only for the probes mentionned pH dissolved oxygen ORP please follow the instructions and press done afterwards 9 You may also check the right value measured for a standard well known but you will not be able to modify this value 10 The setup is now finished and you may leave the har
9. 72 17 F DO i 8 29mg l 02 21 05 18 26 17 Aqu 2 5 RESULTS Recording measures files MAIN MENU gt Configuration gt Measure gt Calibration gt Results gt History Choose menu Results using the navigation keys and press Enter 02 21 05 18 27 T Aqu RESULTS Choose section File Data Table using gt File Data Table the navigation keys and press Enter gt Period Data Table gt File Graphic gt P riod Graphic 02 21 05 18 28 iam Aqu SET PAGE 21 21 Ted PO NSEL ACTEON AQUA ACTEWIN MESURE INDEX 1 USER MANUAL REF PONKITAQUA 2 5 1 FILE DATA TABLE FILES Choose section File Data Table using gt STATION 01 the navigation keys and press Enter gt SCORFF 02 gt LAITA 01 ELLA 01 gt ISOLA 02 EVEL 01 02 21 05 18 26 qj Aqu FILES Choose the file to show using the navigation keys and press Enter gt STATION 01 gt SCORFF 02 Attention You cannot create and therefore find in this register more gt LAITA 01 than 16 recording files with the standard memory of 1MB SELLA 01 An optional Flash memory of 8 MB is available which allows you to gt ISOLA02 create 128 files very usefull in snapshots gt EVELO1 02 21 05 18 26 dX Aqu DATA TABLE READING The status bar on the bottom of the screen Level 0 49 ft indicates the precise recording starting date Conduct 250 0uS cm hour minute and second The 2 values in the pH 7 84p
10. 9 Key 7 less contrast Key 9 more contrast PAGE 33 33 DATE January 09 11 SET ACTEON AQUA ACTEWIN USER MANUAL een REF PONKITAQUA 3 PUTTING INTO SERVICE 3 1 CONNECTIONS 3 1 1 AQUA PROBE i Take the AQUA probe 1 and cable 4 Make coincide the red point of the plug A with the red point of the socket S 4 points connector then connect while pushing until the click Fix the clasp B on the steel handle 2 for the resumption of effort of the cable Attention Not to use the cable as carrier To disconnect the cable of the AQUA probe pull on the metallic body of the plug A all effort to pull on the cable is self locking 3 1 2 TERMINAL ACTEON Take ACTEON Terminal 2 and cable 4 Make coincide the arrow of the grey plug C with the arrow of the yellow socket S 12 points connector then connect while pushing until the click Fix the shackle D on the handle in Velcroc for the resumption of effort of the cable Attention Not to use the cable in carrier To disconnect the cable of the terminal Acteon pull on the grey body of the plug C all effort to pull on the cable is self locking 3 2 STARTING UP Unscrew the preservation case E from the AQUA and replace by ballasted strainer 3 SET PAGE 34 34 Ted PO NSEL ACTEON AQUA ACTEWIN MESURE INDEX 1 USER MANUAL REF PONKITAQUA To check the functioning of the set carry out a test by plunging the AQUA probe in a bowl of t
11. BURTON DEPT W04 calibration The calibration is correct validate to AQUA EEPROM the logging of the new zero and slope E S54 0s EE A PG coefficients by ENTER 21 qu E The following message appears Out of margin LEVEL 2 Values are out range Calibration Either the standard is incorrect inversion or pollution Out of margin Or the sensor is faulty i Message If the sensor seems good recalibrate with new Enter your name incorrect J BURTON DEPT W04 calibration uus u Sometimes the sensor is just at the tolerated margin EGG Cancel ENTER valid limit 10 and the measure IS all the same correct 02 21 05 19 07 JE Aqu You can disregard then and validate by ENTER The sensor will be calibrated and will deliver some correct values in Measure but this sensor must be watched because maybe at end of life example strong LEVEL 2 shift on the pH 7 or too low Nernst slope SAVING COEFFICIENTS After validation the screen saving PLEASE WAIT coefficients appears momentarily 02 21 05 19 07 d Aqu PAGE 27 27 DATE January 09 11 INDEX 1 REF PONKITAQUA SET PONSEL ACTEON AQUA ACTEWIN USER MANUAL MESURE 2 6 8 Conductivity CONDUCTIVITY SCALE gt 0 200 0uS cm 20 2 000 S cm 20 20 00mS cm 20 200 0mS cm 02 21 05 19 10 qx Aqu CONDUCTIVITY 1 Range 0 200 0uS cm Probe in air Standard OuS cm Measure 0 2uS cm Tempera 74 12 F ESC
12. Example Use the up and down arrow to choose the parameter that refers to the value range of your choice None DO Temperature ORP pH Conduct Level With the right arrow place the cursor in the field logical operation AND OR None and choose this condition with the upper or lower arrow VVVVVVV Choosing a logical operation AND OR gives the possibility to choose a second parameter on the right side on the screen Pressing the right key selects the parameter the up and down keys allow to select the parameter Pressing the right arrow gives access to the value fields The first values define the high threshold The second line defines the lower threshold Enter the interval between the sample and the duration of the impulsion This last information is variable according to the water sampler connected on the Acteon The action on the key ENTER makes appear on the right the pictogram J L in bottom of the screen PONSEL MESURE Stop recording RECORDING gt Stop recording gt Manual start gt Program time gt Pulse Option gt Snapshot 02 21 05 18 199 Aqu CONFIGURATION gt Languages gt Date Hour gt Units gt Parameters gt Memory gt Recording 02 21 05 18 20 qj Aqu PAGE 18 18 DATE January 09 11 SET ACTEON AQUA ACTEWIN USER MANUAL Ded REF PONKITAQUA To stop one of the recording modes choose section Stop recording using the
13. PROB pA ANA AA AA AA AA 36 5 3 PREPARING THE PROPRES AA AB GAAN AA 38 5 3 1 Installing the dissolved oxygen membranes disent die 3 5 3 2 STO EP SCIO S ases ace votes ico coins scious on ere AA 39 5 3 3 Installing the ballasted strainer and the calibration Cap 40 5 3 4 TOUS QPS LC CODE Rene RE 40 T MEI uui OEC SNNSIONIU 0 40 so MANTENAN TOFTE SEN GORS uenti EIE mre opt erent Sener ner cn bd cu iU DN P NA AG 41 5 5 1 DO CROP ECS CROIRE nd D ne ee ee Le oi 41 33 2 GE T IAI SONORE Re AA 41 D TOHDOFOIIE SONORE AA PA AA ANA A A E ETT 4 5 6 STORAGE CONDITIONS TOR AO UA PROBE ane tea a NINANG PNTA KANA NA NGANGA AG 42 5 6 1 OL LO OBS 15 2 NIIT Tct 42 5 6 2 TOONS OT SOI a E M 42 5 1 CALIBRATION OF THE AQUA PROBE 235245 pa PAGDADAANAN GG LAGLAGAN GA 43 5 7 1 Aa RA 43 5 8 CALIBRATION PROCEDURES a Aa UEM D MEI A 44 PAGE 8 8 PONSEL x po DATE January 09 11 CA bani ACTEON AQUA ACTEWIN id e MESURE USER MANUAL INDEX 1 REF PONKITAQUA 5 6 1 CENA 1 HC 44 5 8 2 Calibration procedures of AQUA probe sise 45 5 8 3 What to do in case of a suspicious calibration siens 50 6 CALIBRATION FR QUENCE NA DONA anna ahi uat nudo talenti cub 51 f SENSORS PRINCIPLES AND TEMPERATURE COMPENSATION METROLOGICAL STANDARDS 51 7 1 PARAM gl io M Yf Xm 51 7 1 1 Temperature parameter sensor integrated
14. and insert the two obturating plugs tight in their place on the 2 released base plates Indeed these two sensors are likely to throw out a significant quantity of KCI salt in the standard of very low conductivity especially the Ag AgCl reference by the means of its Plastogel polymer gorged with saturated KCL communication through the 1 10 mm side hole in the blue body cylindrical of pH ORP POD The pollution of the standard 84 uS cm is rather fast especially if you await the perfect thermal stabilization of the sensor which takes several minutes It is even advised to proceed the same way for the standard 1413 wS cm if one wishes a calibration of laboratory precision on this measurement range with better than 0 5 Check that in parameters the mention AUTO is apparent automatic adjustement to appropriate range To validate pass on another sensor and make Esc otherwise you desactivate the AUTO mode for a precise range Carry out the calibration and check out the quality of its calibration in Measure mode PAGE 4 4 NICE SET Ted PO NSEL ACTEON AQUA ACTEWIN em MESURE Ni USER MANUAL et REF PONKITAQUA SYMBOLS AT THE BOTTOM OF THE ACTEON SCREEN Exhaustive list of the symbols used and their meaning 10 20 03 09 06 Date Month Day Year hour minute 10 26 03 12 31 45 000357 000714 Recording file Date of measure M D Y H m s measure N Number of measures d Battery d New batteries Bat gt 6 V
15. cancel ENTER valid coefficients by ENTER 02 21 05 19 00 4 Aqu The following message appears Out of margin TEMPERATURE 2 Values are out range Either the standard is incorrect inversion or pollution Calibration l Or the sensor is faulty Out of margin ier lf the sensor seems good recalibrate with new l ae standards Enter your name calibration p l J BURTON Sometimes the sensor is just at the tolerated margin limit 5 and the measure is all the same correct ESC cancel ENTER valid You can disregard then and validate by ENTER 02 21 05 19 07 J Aqu The sensor will be calibrated and will deliver some correct values in Measure but this sensor must be watched because maybe at end of life example strong shift on the pH 7 or too low Nernst slope TEMPERATURE 2 SAVING COEFFICIENTS After validation the screen saving coefficients appears momentarily PLEASE WAIT 02 21 05 19 07 4E Aqu Very important note which relates to all measured and posted parameters in calibration mode Temperature Level Conductivity pH Redox ORP dissolved Oxygen DO In this Calibration mode the values delivered by the sensors and posted on the right of Measurement are always automatically calibrated with theoretical coefficients of zeros and slope which are beforehand recorded ex works in AQUA and cannot be changed or erased by end users Each type of sensor has specific theoret
16. capacity d Lj Acceptable batteries Bat 5 5V d Worn batteries Bat 5 V The exact battery tension is obtained in Main Menu History gt ACTEON 3000S O Watershed memory Empty Memory or OM 25 or created files lt 25 or lt 4 with 1Mo memory CO Occupation Memory gt 25 Yo or created files gt 25 Yo or gt 4 II 1 Occupation Memory gt 50 Yo or created files gt 50 or gt 8 1 II Occupation Memory gt 75 Yo or created files gt 75 or gt 12 II Full Memory OM 100 or 100 of created file 16 The 16 files can share 32 768 recordings of 6 values each for the standard memory flash Data Logger 1Mo 1024 KB In Watershed Memory mode for a memory of 1 Mo beyond recorded 16 files in Acteon the creation of a new file crushes the first created file first in first out and so on It is thus advised if needed to download on PC the totality of the 16 files starting with oldest SAFEGUARD In Watershed Memory mode or in Fixed Memory mode with the 8 Mo of flash memory data logger option you can create and record up to 128 files instead of only 16 sharing the 262 144 possible recordings of 6 values each Fixed Memory Empty Memory or OM 25 Yo or created files lt 25 or lt 4 with 1Mo memory Occupation Memory gt 25 or created files gt 25 or gt 4 Il Il Occupation Memory gt 50 or created files gt 50 or 2 8 Il Il Occupation Memory gt 75 Yo or created files gt 75 Yo or
17. correct even when recording has been launched In fact in Program time mode the recording will start by itself on any page of menus and even when the Acteon is switched off red key On Off in top on the left Idem for the stop recording Only matters that the AQUA probe is connected In Manual start mode it is imperative to start the recording by hand in Measure mode therefore Acteon switched on and to stop it by Stop recording in the menu Configuration But during the recording you can come off the Measure mode and even to switch off the Acteon with the red On Off measures and their recording continue PONSEL MESURE PULSE OPTION PAGE 17 17 DATE January 09 11 SET ACTEON AQUA ACTEWIN INDEX 1 REF PONKITAQUA USER MANUAL This mode is additional to the two previous described modes It allows generating an order besides a water sampler according to the limits tresholds high and low of one or two parameters SAMPLE TRESHOLD DO 0000 0mg l 0000 0mg l Sampl Inter 00h05m Pulse Durat 000 03s Use Up Down arrow 02 21 05 16 45 qu e SAMPLE TRESHOLD DO OR Temperature 0000 0mg l 0000 0 F 0000 0mgl v 0000 0 F Sampl Inter 00h05m Pulse Durat 000 00s Use Up Down arrow 02 21 05 16 45 di O SAMPLE TRESHOLD DO OR Temperature 0020 0mg 0090 0 F 0005 0mg l 0045 0 F Sampl Inter 00h05m Pulse Durat 000 10s Validate by ENTER 02 21 05 16 47 STO
18. gt Measure Choose the option History in the MAIN gt Calibration MENU and validate by pressing Enter gt Results gt History 02 21 05 20 02 pg Aqu HISTORY Choose the desired option using the gt Calibration log navigation keys and validate your choice Maintenance pressing Enter 5 After Sales Service gt ACTEON 3000S 02 21 05 20 03 4X Aqu MESURE 2 7 1 Calibration Log CALIBRATION LOG Level Date 01 14 05 15 17 Std 1 0 00ft Std 2 20 00ft Calibration OK J BURTON DEPT W04 02 21 05 20 03 M Aqu CALIBRATION LOG pH Date 01 14 05 15 30 Std 1 7 01pH Std 2 4 01pH Calibration OK J BURTON DEPT W04 02 21 05 20 03 f7 Aqu CALIBRATION LOG Tempera 11 No calibration 02 21 05 20 04 JM Aqu CALIBRATION LOG DO Date 01 14 05 15 40 Std 1 0 00mg l Std 2 9 21mg l Out of margin J BURTON DEPT W04 02 21 05 20 04 d Aqu CALIBRATION LOG Conduct 0 200 0uS cm No calibration 02 21 05 20 05 dE Aqu PAGE 31 31 SET ACTEON AQUA ACTEWIN USER MANUAL DATE January 09 11 INDEX 1 REF PONKITAQUA The calibration log gives you the following informations Name of the calibrated parameter Date of the last calibration Standard 1 and Standard 2 Operator name Calibration status Calibration OK Out of margin No calibration The navigation keys allow to visualise the different parameters In this example
19. housing even with new batteries and disconnecting one battery for a few seconds put it back in place no recorded files have been erased see 1 2 f the probe has not been detected and the following message is displayed for the AQUA Can t find SDI 12 line Check wires and the ACTEON terminal has been detected no formal conclusion can be drawn either the AQUA ACTEON liaison cable or the AQUA are defective n a last attempt you can reset the AQUA disconnecting it completely from the PC or the Acteon terminal Without power supply you have to wait for a few hours until the total reset of the terminal is completed During these few hours leave the probe at normal temperature 18 to 25 C it might by accident have gone through extreme temperatures or thermic shocks If after that last AQUA reset the problem remains it is advised before calling your PONSEL after sales service to check the perfect state of the three detachable probes pods sometimes sensors in short circuit broken pH glass bulb burst O2 membrane sensor connectors which have been in contact with salted water or not properly clipped can result into brutal electrical over consumption and thus lead to the termination of the SDI 12 communication thus avoiding to communicate wrong measured values and helping to identify a defective sensor
20. icon Aqu does not appear and if after activating the Measure mode the screen does not display the measured values of the 6 parameters and instead appears the message NO PROBE it means the AQUA probe has not been detected by the terminal and that the SDI 12 communication between the probe and the terminal could not be established MEASURE NO PROBE ESC for exit 04 13 05 10 09 qu In that case proceed with the following controls 1 1 Are the cable connectors correctly fixed on their respective sockets In other words are the connectors well in place and properly locked on their respective AQUA and Acteon sockets Pulling on the ends of the cable has a locking effect Also check there is no water inside the connectors otherwise rinse with demineralised water specially after use in salted water then dry for instance with an hair dryer At last check that no pin from either male connector has been torn broken or bent over in its housing 1 2 Are the batteries worn out Battery icon completely white for more than one week Replace them with new ones use a simple screw driver to open the back of the handle of the Acteon terminal Note the recording files are saved on Flash memories for 20 years without any power supply in the Acteon terminal and there are no risks of loosing them during battery change Only date and time are lost as well as recording parameters easy to reconfigure manually or which are easy to sa
21. recording mode Manual Start Pulse option was trigged by Act on 11 17 03 18 17 KI AY NG menu Configuration gt Recording gt Manual Start then gt Pulse Option Aqua probe connected or not or by Actewin on PC Programming Act on Aqua probe inevitably disconnected n The Aqua probe is connected to Act on on standby before starting in Measure mode no recording started if not in Measure mode through Manual start of recording Enter 11 18 03 08 25 MAL Aad L lt n The Manual Start mode recording was started It will stop only on another manual action However possible to auto stop in Fixed memory if 18 11 03 08 00 RecA SL the occupancy rate of the memory reaches 100 Pulse with each batch of recorded measurements values of one or two parameters will be compared with high and low threshold and according to their positions compared to these thresholds and the logical condition binding the parameters and or the Pulse exit free contact of potential with opto transistor will be generated to start a water sampling 20 11 03 08 00 Ln Aau Stop recording Manual validation in menu Configuration gt Recording gt Stop recording or possible auto stop in Fixed memory 14 11 03 09 10 PI Aut The Program Time mode was validate by Act on menu Configuration gt Recording gt Progran Time Aqua probe connected or not or by Actewin software on PC Programming Act on Aqua probe inevitably disconne
22. suspicion of one of the calibration procedures check on the following points Check that the Standard value entered by the user corresponds to the calibration procedure Check the Standard solution cleanliness calibration standard used in accordance with the procedure Check the general condition of the sensor Check the connections between the AQUA probe and the computer cable ends cleanliness of the connectors general condition of the connectors If everything has been checked and the problem persists contact the after sales service department to determine the origin of the disfunctioning defect sensor disfunctionning of the AQUA probe etc PAGE 51 51 PO NSEL PEEN y E UE es INDEX 1 REF PONKITAQUA MESURE p USER MANUAL 6 CALIBRATION FREQUENCY Period 2 years Zero Del 0 2 days 6 months 15 days Zero and slope Slope 2 years Zero and slope pH7 and pH4 Sensor ORP Redox D O Oxygen 15 days Zero 2 months Zero and slope Slope 15 days Period 7 SENSORS PRINCIPLES AND TEMPERATURE COMPENSATION METROLOGICAL STANDARDS 7 1 PARAMETERS Preliminary measures are compensated according to ISO standards and AFNOR The compensations have a maximum of precision between 5 and 30 C Never exceed water temperatures 5 50 C 7 1 1 Temperature parameter sensor integrated in the conductivity probe Principle Pt100 Ohms 1 3 DIN 3 wires assembly Scale
23. to validate the calibration ESC exit ENTER next 02 21 05 19 19 IT Aqu DO Calibration in the air Using the atmospheric Probe in air pression and temperature measures the Acteon terminal determines the value of the Saturation 9 60mg l air saturation The probe being in the air wait Measure 0 02mg l for the measure given by the probe to be Tempera 68 00 F stabilised for at least 30 seconds then press ESC prev ENTER Valid Enter to validate the calibration 02 21 05 19 21 49 Aqu At the end of the calibration process the name of the operator is asked 20 characters are available to enter a name Then press ENTER The following message appears OK In that case the calibration is correct Validate to AQUA EEPROM the logging of the new zero and slope coefs by ENTER The following message appears Out of margin In that case values are out range Either the standard is incorrect inversion or pollution Or the sensor is faulty If the sensor seems good to redo a calibration with new standards Sometimes the sensor is just at the tolerated margin limit 25 35 and the measure is all the same correct You can disregard then and validate by ENTER Strong shift on the pH 7 or too low Nernst slope After validation the screen saving coefficients appears momentarily 2 History MAIN MENU The user can get access to different kind of gt Configuration information through the option History
24. water Melting point not determined Boiling point not determined e Toxicological information owallowing cause stomach problems Contact with water causes irritation e First aid Ingestion Rinse the mouth carefully with water Contact with the skin Wash abundantly with water Contact with the eyes Wash abundantly with water If irritation persists see a specialist e Protection Carrying of protective glasses and rubber gloves is advised e Handling and storage Handling Don t eat or drink while handling the product Wash hands and face after using the product Storage Store at room temperature 15 25 C Protect from sunlight Keep out of reach from children Keep the container closed e Waste removal Chemical waste is often classified as special waste and collection is often organised locally Please contact the local authorities for details PAGE 63 63 NICE SET Ted PO NSEL ACTEON AQUA ACTEWIN em MESURE Ni USER MANUAL et REF PONKITAQUA 10 4 Conductivity standard solutions Composition Potassium chloride lt 5 Water e Physical and chemical properties Appearance uncoloured liquid Smell Odourless Solubility can be mixed with water Melting point not determined Boiling point gt 100 C e Toxicological information Swallowing a large quantity causes nausea vomiting or diarrhoea and can also lead to cardiovascular problems No mutagen or tera
25. 02 23 05 8 32 Aqu CD ATTENTION This indication of range overtaking doesn t exist in perpetual Measures and Recording and is noticeable only by the presence of a steady enough signal values 200 uS cm 10 Yo ou 2000 uS cm 10 ou 20mS cm 10 ou 200mS cm 10 These values risking to be mistaken as measures it is strongly recommended to work during the recordings with the automatic range in order to avoid a possible saturation on a no automatic range PAGE 29 29 DATE January 09 11 SET ACTEON AQUA ACTEWIN USER MANUAL PONSEL c SR MESURE INDEX 1 REF PONKITAQUA 2 6 4 pH Choose the measuring range in which you are going to work Identical calibration procedure for every Probe in range Buffer 1 7 01pH 1 Calibration with a buffer solution pH Measure 6 97pH Dip the probe in the calibration solution Tempera 73 89 F As soon as the measure given by the ESC exit ENTER next probe is stabilised for at least 30 seconds 02 21 05 19 15 qj Aqu press Enter to validate the calibration 2 Calibration with a buffer solution pH4 pH9 ou pH10 Enter the value of the used solution Dip the probe in this solution As soon as the measure given by the k probe is stabilised for at least 30 seconds Tempera 73 82 F press Enter to validate the calibration ESC exit ENTER next Nota standard solutions are worth precisely 02 21 05 19 16 Aqu pH 7 01 and pH 4 01
26. 1 0 3048 ft F Cx1 8 32 C F 32 1 8 PONSEL MESURE 2 3 4 PARAMETERS PARAMETERS gt Level gt Conduct gt pH gt ORP gt Temperature gt DO 02 21 05 16 25 Til PARAMETERS gt Level Y gt Conduct 200uS cm gt pH gt ORP gt Temperature gt DO 02 21 05 16 26 i PARAMETERS gt Level gt Conduct gt pH gt ORP gt Temperature gt DO 02 21 05 16 26 2 3 5 MEMORY MEMORY gt Watershed gt Fixed gt Reset 02 21 05 16 30 day MEMORY gt Watershed gt Fixed gt Reset 02 21 05 16 30 mm ACTEON AQUA ACTEWIN PAGE 14 14 SET DATE January 09 11 INDEX 1 REF PONKITAQUA USER MANUAL With the up and down arrows you can move through the menu and choose the parameter to confirm in measure and recording Use the right arrow or ENTER key to confirm or cancel the parameter chosen This function is realized by the y sign on the right side of the parameter When all parameters are confirmed canceled press Esc Different options are available for the conductivity sensor Automatic range 0 200 OOOUS cm maximal accuracy None and optimal linearity 200 S cm 2000 S cm 20mS cm 200mS cm Select the range with right arrow or ENTER key Then press Esc Or up down arrows for another parameter Important warning Select of preference the automatic range 0 200 000 S cm in recording see chapter 2 6 3 C
27. 1 or 2 threads so that the O2 sensor is in contact with the atmospheric pressure Wait ten minutes for the air in the cap to become saturated with water and for the temperature to stabilise Follow the measure development given by the AQUA probe When this measure is stabilised for at least 30 seconds confirm the calibration procedure with the Validation symbol The report indicates if the second calibration has worked or if it is suspect To pass on to the next screen press Next Screen 4 bis Calibration with calibration standard Enter the concentration in dissolved oxygen of the used standard mg l Pour the appropriate standard volume see Table 1 in the calibration cap and submerge the tip of the AQUA 100 probe sensors in this solution The tip of the oxygen and temperature sensors must be submerged Follow the measure development given by the AQUA probe When this measure is stabilised for at least 30 seconds confirm the calibration procedure with the Validation icon The report indicates if the second calibration has worked or if it is suspect To pass to the final screen press Next Screen 5 Gives a summary of the taken actions during the two calibration procedures and indicates if calibration of the oxygen sensor has worked or if one or both of the steps are suspect In case one of the calibrations would not have worked the screen also indicates which standard to use again To repeat a calibration step go ba
28. 12 Il Il Full Memory OM 100 or 100 of created files 16 Il Il In Fixed Memory mode for a memory of 1 Mo beyond 16 recorded files the creation of a new file is impossible you can enter the name of the 17th file and configure its recording mode but any manual or programmed attempt to launch recording will fail Short appearance of the icon Rec Record then emission of a double beep gt disappearance of Rec no saving of the file name in the Results menu which lists all recording files of Act on If you want to create an additional recording file Download the 16 files on the PC for safeguard then carry out a general reset of the memory Menu Configuration gt Memory gt Reset Or if you accept to lose old recordings switch to the Watershed Memory mode while knowing that the oldest file will be crushed and lost unless preliminary downloading it on your PC PAGE 5 5 NICE SET Ted PO NSEL ACTEON AQUA ACTEWIN em MESURE PU USER MANUAL ih REF PONKITAQUA Aqu Aqua probe connected to Act on 3000 and SDI 12 connection is working dialogue Aqua Acteon is OK Rec Record Recording of selected parameters Or on standby in the Snapshot mode Snapshot recording mode One shot at each manual pressure on Enter keypad A Manual Start Recording mode Automatic Recording of the parameters after manual start on Enter gt Ut Program Time Recording mode Automatic Recording with pro
29. 9 aai MESURE PU USER MANUAL ih REF PONKITAQUA 5 8 Calibration procedures 5 8 1 General schedule The following calibration procedures concern the sensors connected to the AQUA probe corresponding to the following parameters Temperature C or F Conductivity uS or mS pH pH unit ORP mV Pressure depth Level measuring of depth in cm m in or ft Dissolved Oxygen mg l To increase the accuracy of the calibration procedure rinse the calibration cap with water followed by a small amount of calibration solution corresponding to the sensor that is to be calibrated Throw away this rinsing volume and pour in the calibration cup the optimised calibration solution see Table 1 Submerge the sensors with precaution in this standard and screw the calibration cap on the AQUA 100 probe head Don t forget to tighten the probe body to avoid it from falling Once the sensor is submerged shake the probe in an up down movement or left right to take out eventual sensor bubbles Start the ACTEWIN software see chapter 1 and select the option AQUA followed by Calibration The screen will show the list of sensors connected to the AQUA that can be calibrated Select the sensor to calibrate and follow the instructions on the screen Screen 1 Presents the different calibration steps 2 for each sensor and indicates which standard to use Navigation keys Next To pass to t
30. ACTEON 3000 PC graphic statistic exploitation Programming logger PC ACTEON 3000 Export of recording configurations PC ACTEON 3000 PCSAQUA Calibration PC AQUA Measure AQUA PC Installing and execution of ACTEWIN software under Windows Installation of ACTEWIN software requires 2MO of free memory space Insert the installation CD in your CD ROM reader the Autorun option will start automatic installation During installation you will be asked to choose an installation language French is the standard language to register client information user name company name and a file location and name to save the software will be proposed Follow the instructions that appear on the screen to finish installation At the end an icon for the program Actewin V2 X is placed automatically on the desktop Before starting the software application connect the AQUA multiparameter probe to a PC through the communication cable Select the ACTEWIN icon on your desktop or in the program menu of your hard disk Once the program ACTEWIN is launched the software ACTEWIN on PC will initiate its communication with the software situated in the AQUA 100 probe through SDI 12 standard Acteon 3000 terminal connected to a computer SDI 12 RS232 Adapter Software ActeWin Windows for data import and programming logger Acteon 3000 Acteon Adapter cable Acteon 3000 PC with comport or USB port SDI 12 RS232 Adapter
31. H lower right corner indicate the rank of the ORP 264mV values on the screen In the example this Tempera 55 10 F means the first value of 714 The navigation DO gt 2 06mg l keys allow you to move through the memory 02 19 05 12 10 04 000001 000714 field To get back to the Menu RESULTS press twice Esc RESULTS gt File Data Table gt Period Data Table gt File Graphic gt P riod Graphic 02 21 05 18 28 dj Aqu PAGE 22 22 NICE SET Ted PO NSEL ACTEON AQUA ACTEWIN eme INDEX 1 REF PONKITAQUA MESURE mm USER MANUAL 2 5 2 PERIOD DATA TABLE PERIOD Starting and finishing date and time of events display can be chosen Use the navigation From 02 18 05 12 10 keys to enter the dates To 02 18 05 12 26 No file is selected beforehand you go directly in the period of a file if it exists otherwise beep of interdiction Validate by ENTER 02 21 05 18 29 IT Aqu DATA TABLE READING Level 0 49 ft Conduct 210 0uS cm pH 7 84pH ORP 264mV Tempera 55 10 F DO 2 06mg l 02 18 05 12 10 04 000001 000096 Using the right key allows you to display every measuring lot from the beginning up to the finishing date 2 5 3 FILE GRAPHIC FILES Choose the file to show using the navigation gt STATION 01 keys and press Enter gt SCORFF 02 gt LAITA 01 gt ELLA 01 gt ISOLA 02 gt EVEL 01 02 21 05 18 35 dial Aqu The first selected parameter shows with th
32. NSEL ACTEON AQUA ACTEWIN eme MESURE PU USER MANUAL ih REF PONKITAQUA 7 1 3 Dissolved Oxygen parameter Clark polarographic probe Platinum Ag AgCl with a PTFE membrane unsaturated KCI electrolyte Calibration in the air saturation table is automatically generated according to the temperature measured by Pt100 this calibration value is automatically compensated in altitude by a barometric sensor incorporated in the ACTEON 3000 logger terminal calibrator Fast reacting sensor requires stirring of the aqueous medium under measurement Maximum scale 0 50 mgll if super saturation Nominal 0 20 mg l Resolution 0 01 mg l Accuracy 0 2 mg l from O to 20 mg l Resolution 0 01 mg l Accuracy 0 2 mg l from O to 20 mg I Signal compensated in temperature by separate thermistor built in in the sensor Standard NF T90 106 7 1 4 pH parameter Sensor with sensitive glass electrode Combined reference saturated Ag AgCl KCI Measuring range from O to 14 pH unit Resolution 0 01 pH unit Accuracy 0 2 pH unit Answer compensated in temperature according to the NERNST law Standard NF T90 008 7 1 5 Redox Potential parameter ORP Electrode with a platinum disc Combined with reference saturated Ag AgCl KCI Nominal measurement range 1000mV in certain cases 1500 mV Resolution 1 mV Accuracy 20 mV Intrinsic answer non compensated in temperature following ISO recommendations 7 1 6 Lev
33. OD The pollution of the standard 84 uS cm is rather fast especially if you await the perfect thermal stabilization of the sensor which takes several minutes It is even advised to proceed the same way for the standard 1413 uS cm if one wishes a calibration of laboratory precision on this measurement range with better than 0 5 7 2 3 Dissolved Oxygen D O You can calibrate in the air at any temperature However it is preferable to proceed at the temperature of water measurement Immerse the AQUA sensor in the measurement water so that Pt100 sheathed stainless at the end of the conductivity POD is put at temperature as well as the D O probe itself then lift the probe out of the water drain the probe but without drying the D O cell then gauge swiftly without awaiting the temperature increase or decrease in the air The TEFLON membrane is still hydrated which improves the precision of the operation Another procedure consists in placing some water drops in the protection case of the sensors heads of the AQUA which is also used as calibration cup and screw it without exceeding the O ring not to put in pressure the oxygen sensor then immerse the end of the probe in the measurement water without exceeding the top of the preserving case await a few minutes the temperature stabilization then validate the value of measured saturation calculated with the water temperature and the barometric air pressure Nota1 The preliminary cal
34. P TE Patm en hPa SET PAGE 54 54 PO NSE L ACTEON AQUA ACTEWIN Zika sanyanggaag MESUR INDEX 1 ie USER MANUAL REF PONKITAQUA PONSEL saturation table Calibration table to the humid air mg l corresponding to the saturation table of dissolved oxygen in fresh water For Salinity 0 Depth O Altitude 0 1atm Relative humidity 100 Yo G 1L amp 1 2 1 m 3 789 x 79 s ma 78 Ca 74 s ma 3 73 Co 929 sa 723 Ca 2m Co 929 Cal ag x se s 55 a o s sm NOTA Adi 0 gt serial Atmospheric pressure 1013 hPa 760 m mm Hg e g of tT interpolation and correction of barometric pressure 998 hPa the real atmospheric pressure O2Cal 20 76 CT 1atm O2Cal 21 C 20 76 20 O2Cal 20 21 20 76 21 20 9 443 mg l O Cal 20 76 C 998 hPa O2Cal 20 76 998 1013 9 303 mg l rounded with 9 30 temperature 20 76 SET PAGE 55 55 T PO NSEL ACTEON AQUA ACTEWIN PATE January 09 11 Pi LI MESURE INDEX 1 USER MANUAL REF PONKITAQUA 7 2 4 pH You can calibrate this parameter at all temperature but it is preferable to proceed between 20 and 25 C Warning the temperature compensation of pH exploits only the intrinsic response of the probe in an unspecified aqueous medium which is delivering a potential known as of NERNST potential where the absolute temperature appears on the NERNST slope In10
35. RT nF at 20 C this slope is equal to 58 1641 mV C and introduces a compensation to deliver the real pH This compensation does not integrate the effect of the temperature variations on the various pH buffers different between them As a result you should calibrate with pH buffers stabilized at 25 C and validate 7 01 and 4 01 pH values buffer values at 25 C or calibrate at another temperature and manually define the corresponding values of the buffers before validating ex 7 03 and 4 00 at 20 C These buffer values changes according to temperature are precised here under 1 pH 7 01 2 pH 4 01 7 2 5 Redox Potential ORP It is imperative and compulsory to calibrate the ORP at 25 C values such as 240 mV and 470 mV being given at this temperature The intrinsic response of the redox probe is not corrected in temperature because according to the oxydo reducers couples in presence in water each NERNST slope of reversible reactions is proportional to the numbers of exchanged electrons and thus to the variations of oxydation steps 1 2 3 etc Thus it is not possible to carry out a multi correction of temperature Accordingly one takes this measurement in its intrinsic meaning the oxydoreduction potential being by definition what the noble metal plunged in water compared to the electrode of Ag AgCl reference delivers 2 6 Water Level At the moment this parameter is not sufficiently compensated to provide a metrologic
36. S cm in 4 ranges 7 Resolution 0 001mS cm to 0 1mS em _ Accuracy 0 5 of the reading 1 0 Range Oo units Resolution fF Accuracy J 005 ORP Sensor type JPlainumpobe o Range 999mVto 999omv Resolution mV Acuay JjJr20mv Rane jOto20omgl Resolution jOOtmgl Acuay X 02m Rane lOto10m Resolution ftom Accuracy 1 Full Scale SET PAGE 57 57 PO NSE L ACTEON AQUA ACTEWIN E do es MESUR INDEX 1 M USER MANUAL REF PONKITAQUA 8 2 2 MECHANICAL FEATURES 8 3 ACTEWIN SOFTWARE Minimum hardware PC Ill 400 MHz 64MO RAM requirements Windows 98 NT 2000 XP Required memory 2MB 0000000000000 A Format CD ROM 800MB with user manualin pdffile E COH BONE sit user manus ari file pomat Gs Actewin RS232 SDI 12 adapter with connection cables for AQUA probe and Acteon 3000 terminal PAGE 58 58 Ted PO NSEL ACTEON AQUA ACTEWIN MESURE INDEX 1 USER MANUAL REF PONKITAQUA 9 PONSEL AQUA PROBE connexion to ISCO 6712 6712FR and AVALANCHE 1 Please connect the PONSEL AQUA probe with the special cable interface ref PONAQUCAB 2 on the SDI12 rain gauge connector of the
37. SDI12 RS232 cable 1 50m PONACTW LOG CD containing ActeWin software user manual in pdf format PONACT CIMP Cable for impulsional output 5 00m water sampler Consumables EE PON PH AMP 4 Calibration standard pH 4 01 25 C 125ml PON PH AMP 7 Calibration standard pH 7 01 25 C 125ml PON PH AMP 10 Calibration standard pH 10 01 a 25 C 125ml EEE PON EH SOL 470 Calibration standard ORP 470mV 25 C 125ml PON EH SOL 240 Calibration standard ORP 240mV 25 C 125ml d p CC PON O2 ZR c Cleaning solution for electrodes pH pepsine 125ml a qq Mq2cegssedxRmu p u Wh PON ALR6 Set of 4 Alkaline batteries AA LR6 PON ANICD AA Set of 4 accumulators Ni Cd AA m SET PAGE 65 65 ca PONSEL ACTEON AQUA ACTEWIN aaa Eon er MESURE INDEX 1 USER MANUAL REF PONKITAQUA 12 FAQ Question AQUA 100 probe connected My ACTEON 3000 terminal displays NO PROBE 5 What should I do With the Act on 3000 terminal running and the AQUA 100 connected the icon Aqu must be displayed in the bottom right corner of the Act on 3000 screen Activating the Measure mode perpetual measuring of the MAIN MENU the 6 measured values of the 6 parameters are displayed every 5 seconds after the message MEASURE IN PROCESS has been displayed for a few seconds MEASURE Level 11 22ft Conduct 9 41mS cm pH 6 63pH ORP 185mV Tempera 72 17 F DO 8 29mg l 02 21 05 18 22 I Aqu If the
38. SET ACTEON AQUA d PO N S E L ACTEWIN DATE C MESURE JANUARY 09 11 INDEX 2 PONSEL Mesure Zone de Bellevue 35 Rue Michel MARION REF PONKITAQUA 56850 CAUDAN Phone 33 0 2 97 89 25 30 Fax 33 0 2 97 76 55 72 SET ACTEON AQUA ACTEWIN MULTI PARAMETER 6 SENSORS For water quality monitoring USER MANUAL AQUA PROBE AQUA10 PROBE AQUA50 PROBE ACTEON 3000 PAGE 2 2 _ SET A PO N S FE L ACTEON AQUA ACTEWIN DATE January 09 11 MESURE USER MANUAL INDEX 1 REF PONKITAQUA SET ACTEON 3000 AQUA ACTEWIN USER MANUAL WARNINGS This material is scientific handheld instrument it is important to support it with a minimum of maintenance After each use before storage in its case you must clean the sensors pH Redox cleaning solution PON SOL NET Conductivity Temperature with a brush Cleaning the level sensor but without introducing a needle or any pointed object into the holes and without using a powerful jet To store the AQUA probe don t introduce water and especially not purified water into the conservation calibration case but use some drops of conservation solution ref PON SOL CSV on a wad The use of a too important volume of water in the conservation case causes a pressurization reflecting itself on the entry of the reference of the pH sensor when the stopper is tight The pH reference is prematurely emptied of its KCI involving the fast deterioration o
39. al value i e 1 cm of water but is used in fact as an indicator of immersion for the AQUA 100 probe from O to 10 m or from O to 50 m in option A compensation of this parameter for daily variations of atmospheric pressure through the incorporated barometer of the Act on 3000 is to be implemented in the sofware by end of 2005 Without this compensation the zero can vary by several cm between day and night There is a temperature compensation on the side sensor of hydrostatic pressure but it is advised to calibrate at the temperature of the water to be measured especially for the zero in the air atmospheric pressure Thus immerse entirely the AQUA 100 in the water to be measured wait approximately one minute then lift it and action immediately Del 0 in Measures mode The slope is less sensitive to temperature variations if you still want to calibrate it use preferably a water column or a pressure standard equivalent to the full scale 10 m or 50 m PAGE 56 56 7e PO NSE ACTEON AQUA ACTEWIN MESURE INDEX 1 USER MANUAL REF PONKITAQUA 8 DESIGN FEATURES 8 1 ACTEON 3000 TERMINAL Data storage 32 768 measures batches of 6 time stamped parameters Standard version 1MB 262 144 measures batches Version 8MB 8 2 AQUA PROBE 8 2 1 SENSOR FEATURES SENSORS oo Rane 5Cto 50C Resolution jJOOUC A s Accuracy OC Rage O to 200 m
40. an dry or pre rinsed calibration cup see Table 1 and plunge the AQUA 100 probe tip in this solution The end of the temperature and oxygen sensors must be submerged Follow the measure development given by the AQUA probe When this measure is stabilised for at least 30 seconds confirm the calibration procedure with the Validation symbol The report indicates if the first calibration has worked or if it is suspect To pass on to the second calibration press Next Rinse and dry the probe and calibration cup For the second calibration two methods are proposed e Calibration by air e Calibration with a standard screen 3 Gives the two calibration options by air and by standard The first method is based on calibration of the sensor in water saturated air the second implies that the concentration of a standard in dissolved oxygen is known Choose one of the methods by clicking on the corresponding symbol PAGE 47 47 DATE January 09 11 INDEX 1 REF PONKITAQUA SET A PO NSEL ACTEON AQUA ACTEWIN USER MANUAL C y SR MESURE screen 4 Calibration by air Enter the value of the atmospherical pressure in hectopascal hPa initialised by normal atmospheric pressure 1013 Pour 5 mm of water on the bottom of the calibration cap and place the tip of the probe sensor in the cup Make sure that the oxygen and temperature sensors are not submerged in the water Screw the calibration cup lightly
41. ap water Press key l in the left upper corner of the keyboard for powering up In the configuration menu adjust date and time if necessary and check the chosen parameters p 5 Reset to zero the memory and choose fixed or rotating memory p 14 Go back to the general menu and choose the option MEASURE perpetual measuring p 18 Check the coherence between the Level raba 39 ft different parameters idis If necessary carry out a calibration Conduct 567 4yS cm of the sensors see p 24 to 28 pH 7 93pH ORP 175mV Tempera 60 07 F DO 8 23mg l 02 22 05 08 03 MA Aau Before carrying out measurements choose the recording mode Manual start and file name pages14 to 16 Once in the field proceed to perpetual measuring p 18 Make zero of the level sensor before plunging the probe p19 Pressing ENTER in this option will start the recording of data AV Go back to Configuration option at the end of the measuring session then Recording and confirm Stop recording p 18 PAGE 35 35 NICE SET Ted PO NSEL ACTEON AQUA ACTEWIN eme INDEX 1 REF PONKITAQUA MESURE mw USER MANUAL 4 ACTEWIN DATA HANDLING SOFTWARE With ACTEWIN one can connect a PC indifferently to the portable terminal ACTEON as well as to the AQUA probe using the Actewin SDI12 RS232 adapter and specific connection cables Following options are available PCSACTEON Import of data measures
42. ass bulb pH ORP sensor Before installing the sensors make sure that the sensor and AQUA connectors are absolutely clean The openings for connecting the sensors must be completely dry In case of humidity use a compressed air spray to dry the openings Installing the sensors is done as follows Hold the sensor by the ring and locate the place of the sensor using the coloured drop of the compartment Place the sensor on top of the compartment opening Insert the sensor into the opening until hearing a soft click meaning that the sensor is correctly installed 5 3 3 Installing the ballasted strainer and the calibration cap The AQUA probe is equipped with a ballasted strainer and with a calibration cup that also serves as a protection case As soon as the sensors are installed screw ballasted strainer or the calibration cup clockwise on the compartment of the probe 5 3 4 Installing the cable Remove the protective cap on the top end of the AQUA probe and place the cable connector opposite of the probe opening so that the two red marks one on the cable ring and one on the probe connector touch Insert the cable until hearing a soft click meaning the cable is correctly installed 5 4 Maintenance of the probe The maintenance of the sensor which will increase its lifetime on one hand and will reduce preparation time for future use of the AQUA on the other hand is presented in this part of the manual This parag
43. at 25 C 77 F Probe in Buffer 2 4 01pH Measure 4 08pH At the end of the calibration process the name of the operator is asked 20 characters are available to enter a name Then press ENTER The following message appears OK In that case the calibration is correct Validate to AQUA EEPROM the logging of the new zero and slope coefs by ENTER The following message appears Out of margin In that case values are out range Either the standard is incorrect inversion or pollution Or the sensor is faulty If the sensor seems good to redo a calibration with new standards Sometimes the sensor is just at the tolerated margin limit 35 and the measure is all the same correct You can disregard then and validate by ENTER Strong shift on the pH 7 or too low Nernst slope After validation the screen saving coefficients appears momentarily 2 6 5 ORP The AQUA probe is placed in the air and a short circuit is applied to the ORP sensor in order to simulate the value OmV Place the probe at free air and follow the evolution of the measure As soon as the measure given by the probe is stabilized since at least 30 seconds press the key ENTER to validate the calibration Auto 0 Measure 3mV ESC exit ENTER next 02 21 05 19 17 dt Aqu REDOX 2 Calibration with a buffer solution of 240 or 470mV according to the zone of measure Enter the value of the used calibration solution Dip the probe i
44. ck to the corresponding screen pressing Previous Confirm the calibration of the Os sensor pressing End It is possible to confirm this calibration even if one of the procedures was declared suspect Rinse and dry the sensors and the calibration cap for future use c gt Calibration of the temperature sensor The first procedure is carried out in an ice bath of 0 00 C Screen 1 Gives a summary and description of the actions to be taken during the 2 calibration procedures of the temperature sensor screen 2 Select the parameter unit initialised in C or F and enter the value of the first Standard fixed at 0 00 C 32 00 F Pour the necessary volume of a mixture of water with ice see Table 1 in the clean dry or pre rinsed calibration cap and plunge the temperature sensor tip in this liquid Wait for some minutes for the temperature sensor to reach the temperature of the iced bath When the measure given by AQUA probe is stabilised for at least 30 seconds confirm the calibration procedure with the Validation symbol The report will indicate if the first calibration is accepted or suspect To pas to the second calibration press Next PAGE 48 48 SET T PO NSEL ACTEON AQUA ACTEWIN PATE January 09 11 aai MESURE PU USER MANUAL ih REF PONKITAQUA The second calibration is carried out with a solution of a known temperature screen 3 Read the temperature of the second Standard with a the
45. cted 14 11 03 09 12 IM Aut Aou C The Aqua probe Is connected to Act on on standby waiting for the starting time of Program Time mode to start recording It is not necessary to be in Measure mode at the time of starting m mn m e 14 11 03 10 30 Bu Aut Aau Rec da a recording started at the programmed starting time and is m Stop recording 17 11 03 16 30 hd Pau Normally at the programmed end date Or possible manual interruption before this date Manual validation in menu Configuration gt Recording gt Stop Recording Or possible Auto stop in Fixed memory if memory saturated PAGE 7 7 SET rie PO NSEL ACTEON AQUA ACTEWIN eme MESURE Ni USER MANUAL et REF PONKITAQUA TABLE OF CONTENTS 1 CASE CONTENTO 9 2 ACTEON PORTABLE TERMINAL mga ANGAT FRATRE PREX DET FERRY V EcCaANE DE XE IY PR UR EVER SW RA C PAN SUCH d Rica 10 2 1 KE PORO DESC RIETI N on 10 22 JBOWBERINOCOPDE LHEJDISVIGE i NAN NAGA AN D ME 1 2 5 CONEGCGUR ATON etter ce 11 2 3 1 LAN UAE a E E TT Tm 11 2 3 2 DLE OR a 12 2 3 3 NS ieee 13 2 3 4 PARAE D AS r 14 2 3 5 MEN d ra ieee ae E E E T T E E E E I EE 14 2 3 6 RECORDING ae E E sea sett hereto EE aaa EA T E sa kaa AA 15 2 4 MEASURE PERPETUAL MEASURING vecicccccssesssasendtesasoastosannscemsessiesssusivenv
46. d Submerge the sensors in this second standard and follow the development of the measure given by AQUA 100 When this measure is stabilised for at least 30 seconds confirm the second calibration procedure with the Validation symbol The Report box indicates if the calibration procedure is confirmed or suspect Navigation keys Previous To go back to the second screen Next This option can only be chosen if the second calibration step is confirmed and allows passing to the next screen to see the final report Cancel To quit the calibration procedure Screen 4 Gives a summary of the taken actions during the calibration steps indicating if the calibration of the sensor has functioned well or if one of the procedures needs to be repeated indicates which standard to repeat Previous To go back to the third screen End To confirm the calibration of the sensor and to bring you back to the screen showing the list of sensors Note Calibration can be confirmed even if the report indicates that it is suspect Cancel To quit the calibration procedure The sensor calibration procedure is only confirmed if the two steps have been carried out and are confirmed Sensor display screen Every sensor that needs to be calibrated is represented by a symbol and its calibration status is indicated as follows The sensor has not been calibrated 4 Calibration of the sensor was successful 5 Calibration of
47. determined e Toxicological information Swallowing will cause irritation of the mucous membranes in the mouth of the pharynx and of the oesophagus and can spread to gastro intestinal irritations Contact with the skin can cause irritation Contact with the eyes can cause severe irritation Inhalation can cause irritation of the mucous membranes together with coughing e First aid Ingestion Rinse the mouth carefully with water Drink a large amount of water In serious cases see a Specialist Contact with the skin Wash abundantly with water Contact with the eyes Wash abundantly with water If irritation persists see a specialist e Protection Carrying of protective glasses rubber gloves and a mask to avoid inhalation is advised e Handling and storage Handling Don t eat or drink while handling the product Wash hands and face after using the product Storage Store at room temperature 15 25 C Protect from sunlight Keep out of reach from children Keep the container closed e Waste removal Chemical waste is often classified as special waste and collection is often organised locally Please contact the local authorities for details PAGE 62 62 NICE SET Ted PO NSEL ACTEON AQUA ACTEWIN eme MESURE PU USER MANUAL i REF PONKITAQUA 10 3 Buffer solution ORP 240 mV e Physical and chemical properties Appearance yellow liquid Smell Odourless Solubility can be mixed with
48. dware function 11 If you want to see the value of the different parameters on the screen please go to view report then data then SDI12 the readings of the different parameters appear on the screen An asterisk before the data value tells that the measurement is invalid and not working PAGE 59 59 DATE January 09 11 INDEX 1 REF PONKITAQUA CEI SET ACTEON AQUA ACTEWIN USER MANUAL Connection probe AQUA on interface cable PONSEL Galvanic insulated adapter Ton don 5029874005 Avalan aC vwar 12V DC 6 AMF This linet amen her Patani coverage aa fol Other ING and faraign patents ara pending UG Pat Ho b T25 801 amp Bai 110 A 5 b 557 508 Bodie amp 5 IL BOO 1 005 A BS 395 U K Pai NG Lore oe A ce see mas It ESCO Inc o uoaa Poa C mw USER MANUAL 10 10 1 SET PAGE 60 60 T PO NSEL ACTEON AQUA ACTEWIN PATE 22709 MESURE INDEX 1 REF PONKITAQUA SECURITY SAFETY SHEETS pH STANDARDS pH 4 Composition Potassium hydrogen phthalate lt 2 Yo Stabilisers lt 0 01 Water pH 7 Composition Phosphate compounds lt 1 Stabilisers lt 0 01 Water pH 10 Composition Carbonated compounds lt 1 Water e Physical and chemical properties Appearance slightly soapy Smell Odourless Physical condition liquid Solubility can be mixed with water Melting point not applicable Boiling point gt 100 C e Toxic
49. e current value in the upper right corner of the screen The status bar on the bottom of the screen indicates the precise recording starting date and time The 2 values on the right give the rank of the shown values In the exemple first value of 714 02 18 05 12 01 04 000001 000714 The right and left navigation keys allow moving through the memory field The crossed dot indicates the position of the current measure on the curve Date and time of this measure is shown on the bottom of the screen The counter on the right side gives the position of the memorized value 02 18 05 12 31 44 000357 000714 a aaa PAGE 23 23 Ted PO NSEL ACTEON AQUA ACTEWIN INDEX 1 REF PONKITAQUA MESURE mw USER MANUAL Conduct 276 8uS cm The up and down navigation keys allow visualising the different parameters 02 18 05 12 31 44 000357 000714 RESULTS To get back to the Results menu click gt File Data Table the Esc button twice and select the desired gt Period Data Table column using the navigation keys gt File Graphic gt Period Graphic 02 21 05 18 37 dE Aqu 2 5 4 PERIOD GRAPHIC PERIOD Start and End date and time of graphic visualisation can be chosen For that purpose From 02 18 05 12 10 use the navigation keys to enter the dates To 02 18 05 12 26 No file is selected beforehand you go directly directly in the period of a file if it exists otherwise beep of
50. e preservation cap Not to throw it keep this liquid preciously in a hermetic bottle Otherwise available on order Ref PON SOL CSV This liquid contains KCL saturated at 0 C products fongicide anti biofouling anti evaporation does not freeze over 50 C and is a pH buffer value between pH4 and 5 salinity gt 200mS cm No toxic it regenerate partially in some days the reference Plastogel Ag AgCl pH ORP in case of salt exhaustion PAGE 43 43 NICE SET Ted PO NSEL ACTEON AQUA ACTEWIN em MESURE PU USER MANUAL ih REF PONKITAQUA 5 7 Calibration of the AQUA probe 5 7 1 Recommendations e Security WARNING Certain reagents that are used for the calibration of sensors are health threatening and handling them requires specific safety measures see Appendix 1 e Calibration with calibration cup or vessel The preservation case that comes with the AQUA probe can also be used as a calibration cup in order to optimise the volumes of the used standards protecting the sensors The optimised volumes that allow a correct calibration for each sensor are noted in the table below Sensor to calibrate Vertical position pH ORP 50 ml D Oxygen 90 ml Conductivity Table 1 Optimised standard volumes Calibration of the AQUA probe can also be done with a vessel In that case several precautions need to be taken Carry out all the calibrations with the ballasted strainer to protect the sensors and ta
51. ed at 0 Place the AQUA probe at zero depth and follow its measure development When the measure is stabilised for at least 30 seconds confirm the calibration procedure by pressing the Validation symbol The report will indicate if the first calibration is validated or suspect To pass to the second calibration press Next The second calibration is carried out with a known depth other than zero screen 3 Select a parameter unit initialised in cm m in or ft and enter the depth of Standard 2 Place the AQUA probe in this depth and follow its measure development When this measure is stabilised for at least 30 seconds confirm the second calibration with the Validation symbol The report will indicate if the second calibration was successful or if it is suspect Pass to the next screen pressing Next Screen 4 Displays the taken actions during the two calibration procedures and indicates if they were accepted or if a suspicion was detected during one or both procedures This screen indicates also which Standard needs to be used for the suspect calibrations To repeat a calibration procedure go back to the corresponding screen pressing Previous Confirm the pressure sensor calibration pressing End It is possible to confirm the calibration even if one of the procedures is declared suspect Rinse and dry the probe and the calibration cup for future use 5 8 8 What to do in case of a suspicious calibration If there is
52. el parameter Absolute piezo resistive sensor scale 0 10m or 0 50 m in option Simple water level gauge with 1 Resolution 1cm Not yet compensated in atmospheric pressure will be with the next software release 7 2 TEMPERATURE RECOMMANDATIONS FOR LABORATORY CALIBRATIONS to follow with calibration of very high degree of accuracy or metrological checking of the measures in standard and buffer solution Nota If you want to perform a complete calibration with hight precision of your AQUA 100 with the aid of the calibration menu of the ACTEON 3000 you must start with the temperature sensor 7 2 1 Temperature lf you want to calibrate your temperature sensor at 0 50 C for exemple you need to disconnect the 2 other sensors pH ORP and Oxygen sensors and replace these 2 sensors by waterproof obturating plugs The combinated pH ORP POD and the Oxygen POD could be damaged by the temperature variation 0 to 50 C crack glass pH shunt or floating measure It is preferable to limit the maximum temperature at 30 C In this case you need to start with the 0 C in an ice bath ice water Attention if the probe is equipped with its 3 sensors the temperature variation has to take place progressively in order to avoid the thermal shocks During the calibration process it is advised to immerse completely the sensor heads in the calibration solution and to agitate the water to avoid heat gradients Wait 30 minutes until the t
53. emperature is stabilized 2 100 of the C In the following procedures you must always await stabilization of the temperature before validating a calibration or raise a measure to a standard PAGE 53 53 DATE January 09 11 INDEX 1 REF PONKITAQUA SET PO NSEL ACTEON AQUA ACTEWIN USER MANUAL C 7 MESURE 7 2 2 Conductivity You could calibrate the conductivity sensor in a KCI standard at all temperatures but it is preferable to proceed between 20 and 25 C Before and after immersion in a standard rinse well the sensors and entirely with distilled water and dry them especially before calibration or checking the standard of 84 uS cm this one being very easily polluted Always begin with the calibration of the lower range 200 uS cm and so on until 200 mS cm The zeros for conductivity are done in the air slit of measurement of the sensor preliminary rinsed then dried During the calibration using standard of 84yS cm it is strongly advised to first disconnect PODs sensors pH Redox and Dissolved Oxygen and insert the two obturating plugs tight in their place on the 2 released base plates Indeed these two sensors are likely to throw out a significant quantity of KCI salt in the standard of very low conductivity especially the Ag AgCl reference by the means of its Plastogel polymer gorged with saturated KCL communication through the 1 10 mm side hole in the blue body cylindrical of pH ORP P
54. f the sensor We point out that the AQUA probe should not be used in effluents being able to contain hydrocarbons colloids concentrated acids Avoid too aggressive solvents Temperature of sensors for prolonged use 50 C max If you don t use all sensors put a Waterproof obturating plug in their place PONAQUA CAP 1 DATA TRANSFER Stop the data recording on the ACTEON 3000 before transferring the data SENSORS Installation of a new pH sensor place the sensor in water during 1 hour then transfer it in a standard solution during 30 minutes in order to activate the reference Then carry out its calibration in Calibration mode and check the quality of its calibration in Measure mode placing the electrode in 2 different buffers S GOULETTE F BELLOUARD 01 09 11 Ponsel Mesure Ponsel Mesure Upgrade Name Name Name Dept Service Stamp Dept Service Stamp Dept Service Stamp Date Firm Firm Firm Written by Checked by Approved by PAGE 3 3 Ted PO NSE ACTEON AQUA ACTEWIN MESURE INDEX 1 USER MANUAL REF PONKITAQUA Installation of a new conductivity sensor if the value read in measure mode presents a significant drift pass a very fine abrasive paper on the 2 faces of the notch Concerning the calibration of conductivity during the calibration using standard of 84yS cm it is strongly advised to first disconnect PODs sensors pH Redox and Dissolved Oxygen
55. g the recognition of the probe by the Acteon terminal will appear under the screen Press the down arrow to choose the mode Measure perpetual measuring MAIN MENU gt Configuration gt Measure gt Calibration gt Results gt History 02 21 05 18 21 dX Aqu MEASURE Press ENTER the message measure in process shows for 5 seconds the measuring time MEASURE IN PROCESS 02 21 05 18 22 Aqu O PAGE 19 19 iy PO N S FE L SET DATE January 09 11 MESURE MAA Ha N INDEX 1 REF PONKITAQUA MEASURE Level 11 22 ft Display of the measuring results of the 6 Conduct 9 A1mS cm parameters are shown on the screen Every pH BE 6 63pH 5 seconds the results are updated ORP 185mV Tempera 72 17 F DO 8 29mg l 02 21 05 18 22 M Aqu 2 4 1 SNAPCHOT RECORDING MODE One shot In Measure pressing ENTER allows using this option any time a value batch can be manually recorded in a file named by the user after confirmation by the ENTER key An audible tone is heard when entering Name STATIONO1 It is the only recording mode purely manual Validate by Enter Paba are automatic after starting ention This recording mode cannot be used 02 21 05 18 23 de Aqu when the Manual start and Program time modes have been programmed as well as recording in progress under one of these modes SNAPCHOT RECORDING Data are updated every 5 seconds To enter a Level 11 22 ft new value batch
56. grammed start and end dates L Pulse Option Recording mode Associated with the 2 AUTO recording modes for water sampling PC Downloading PC Import Export Act on 3PC with Actewin software Associations of the symbols and their functional description State Functional description 11 13 03 20 03 i Date and hour batteries and memory revolving or fixed 11 13 03 20 03 Ln capacities Always on the screen 11 13 03 20 03 P Agu AQUA probe connected to Act on and correct dialogue The Manual start recording mode Auto was trigged throught the Act on 11 13 03 09 10 SAY menu Configuration 2 Recording Manual start But AQUA is not connected or by Actewin on PC Act on Programming AQUA inevitably disconnected 11 14 03 09 12 HAY Aou r ws The AQUA probe is connected to Act on on standby before starting in i Measure mode no recording started if not in Measure mode through Manual start of recording Enter The Manual start recording mode was started It will only stop with i AJ Agu Rec C2 another manual action Configuration recording stop recording The only case it will stop by itself is in fixed memory mode if the memory capacity is full 11 14 03 10 30 After Stop recording or possible auto stop if memory is full in Fixed memory mode 11 14 03 16 30 cl Plau PAGE 6 6 NICE SET Ted PO NSEL ACTEON AQUA ACTEWIN eme MESURE INDEX 1 USER MANUAL REF PONKITAQUA The
57. h procedures This screen indicates also which Standard needs to be taken for the suspect calibrations To repeat a calibration procedure go back to the corresponding screen pressing Previous Confirm ORP sensor calibration pressing End It is possible to confirm the calibration even if one of the procedures is declared suspect Rinse and dry the probe and the calibration cup for future use e Calibration of the conductivity sensor The first procedure is carried out exposing the AQUA probe to air to simulate a O uS value screen 1 Gives a summary of the two calibration procedures and a description of the actions to be taken during calibration of the ORP sensor Screen 2 The parameter unit fixed in uS mS enter the value of the first Standard fixed at 0 pS Place the AQUA probe in the air and follow the development of the measure it gives When the measure is stabilised for at least 30 seconds confirm the calibration procedure by pressing the Validation symbol The report will indicate if the first calibration is confirmed or suspect To pass to the second calibration press Next The second calibration is carried out with a conductivity buffer standard of 1413 uS or 12880 US screen 3 Select a parameter unit initialised in uS mS and enter the value of the second Standard at 1413 or 12880 LS Pour the necessary volume of the selected buffer standard see Table 1 in the clean dry or pre rinsed calibration cap a
58. hat is enough to provide humidity but not so much that it reaches the sensor surfaces especially if the liquid is pure water gt in this case premature salt depletion of the Plastogel reference Ag AgCl on the Pod sensor pH ORP Check that the conservation cap is hermetically closed to reduce the risk of evaporation Check regularly that humidity is maintained in the cap 5 6 2 Long term storage a Dissolved Oxygen sensor Disconnect the dissolved oxygen sensor take off the membrane and empty the oxymetric electrolyte Rinse the electrolyte tank with distilled water and dry with compressed air Keep the sensor dry and without the membrane or with a membrane in order to protect the cathode Place a small protection cap on the opening of the probe connector that is meant to receive the oxygen sensor to avoid humidity from coming in b Conductivity pH ORP and temperature sensors Disconnect the conductivity sensor rinse it with water and dry with compressed air Keep the sensor dry pH ORP sensor may remain connected to the AQUA probe Follow the same instructions as for short term storage using the preservation orange liquid Attention before push in the waterproof obturating plugs of sensor Pods oxygen and temperature conductivity ocrew the preservation cup completely to avoid any evaporation Preservation liquid Ref PON SOL CSV The preservation orangey liquid is provided with the probe AQUA in th
59. he next screen and to carry out the first calibration Cancel To quit the calibration procedure Screen 2 First calibration step Select the parameter unit fixed unit or initialised could be modified by following menu instructions Enter the value of Standard 1 fixed or initialised value that can be modified The measure given by AQUA is shown in real time Measure box When this measure is stabilised for at least 30 seconds confirm the first calibration procedure with the Validation symbol Inthe Report box a message will appear indicating if the calibration procedure is confirmed or suspect Navigation keys Previous To get back to the first screen Next This option can only be chosen if the first calibration step is confirmed and allows passing on to the next screen to carry out the second calibration Cancel To leave the calibration procedure Unscrew the calibration cap and take the sensors out of the first calibration standard Rinse the sensors and the calibration cup with water followed by a small quantity of the second calibration standard Screen 3 Second calibration step Select the parameter unit fixed or initialised could be modified by following menu instructions PAGE 45 45 SET T PO NSEL ACTEON AQUA ACTEWIN PATE 22909 C y MESURE INDEX 1 REF PONKITAQUA USER MANUAL Enter the value of Standard 2 fixed or initialised value that can be modifie
60. ibration of zero O2 in water added with sodium sulphite is completely independent of the temperature of this water Nota 2 With normal atmospheric pressure and null salinity the PONSEL saturation table according to the temperature thus calibration of the dissolved oxygen is shifted approximately 0 5 mg l on average compared to the saturation table of a not salted pure water This to hold account of the difference state of the hydration of PTFE membrane between the air saturated with water and water saturated with oxygen in balance with this air According to this PTFE hydration state naturally hydrophobic the diffusion of oxygen through this membrane is very slightly higher in the air whereas the pO2 partial pressures are equal in the gas phase and the dissolved phase and that consequently the response of the manometric probe of Clark should not present a difference in theory The table presented below holds account for this correction Act on and ActeWin have this table in memory and during the calibration carry out the interpolation of temperature then Act on automatically applies the barometric correction built in barometer whereas it is introduced manually in hPa for ActeWin on PC In all the cases calculations providing the calibration value are automatic and the user does not need to know the table and the function of correction but they are given here for information purposes O2Cal t C Patm O2Cal t C 1atm Patm
61. ical coefficients The real calibration coefficients with physicochemical standards liquid or air are always calculated with these theoretical coefficients and with the buffer values which avoids risking to divide by zero or uncontrolled divergence especially when proceding with previous erroneous calibrations because of pollution or buffers inversions After zero and slope calibrations it is necessary to return in Measure mode to check the calibration quality the value you read a few seconds later in Measure mode must coincide with better than 0 5 with the value of the standard or the buffer in which the sensor soaks Then the sensor is correctly calibrated with the real coefficients of zero and slope Now if you return again in Calibration mode for this parameter do not be surprised to still find in the standards the same initial values as if the sensor had not been calibrated because again calibration in this mode is done with the theoretical coefficients which never change These theoretical coefficients were statistically calculated by average of the characteristics of an important batch of representative sensors of a normal production Calibrated with these theoretical coefficients the sensors should ideally deliver correct values but the variability of their initial characteristics shift of the zero and sensitivity like their ageing clogging and poisoning cause more or less important variation of the reading on the sa
62. in the conductivity probe eese nenne enne 51 UN D Conductivity Parameter 9 51 Z3 Dissolved Oxygen parameter 222200000aaaaaaaaaaa naaa 52 7 1 4 DH DOFQWELIOES sob ET Cee due dtd OTe ey re UNI ne reer ee ener ore eer MIU ter III PIE 52 7 1 5 Redox Porenaalparomeler ORP Ta GAD detre states uode bc E ua Po dti sque Vds i b NGANGA 52 7 1 6 LOGAN a I EO AA TEE 52 7 2 TEMPERATURE RECOMMANDATIONS FOR LABORATORY CALIBRATIONS 52 7 2 1 Temperature RR 52 722 O9 7T ITTe AIT ARR 53 US PIONEER DO ane a a nat nt Dee 53 7 2 4 DE EEE 55 7 2 5 Kedox POLATIGL ORT ies a aaa a NG NONA ANAN NAA Gas 55 7 2 6 WTL TO M A TN N AAN NAREN NS EEN AAN 55 8 DESIGN FEATURES ea ea ee ee D ae 56 8 1 ACTEON S000 TERNINAL GAO AA ANGKAN a a a E 56 Ba 1080 3 2148 Rc moem 56 8 2 1 SENSOR FEATURES asana ama kA AA AA AA 56 8 2 2 MECHANICAL FEATURES naa PANA BG GAGA BAG 57 8 3 ACTEWIN SOFTWARE ccnn 57 9 PONSEL AQUA PROBE connexion to ISCO 6712 6712FR and AVALANCHE aaa 58 To SECURED SABE NO ana EE AG UA 60 O FPE BANA E E E 60 10 2 BUFFER STANDARD ORP 470 MN T an nou canon een een nn 0 08 61 10 3 BUFFER SOLUTION ORP 240 WEY anse aan GF en 62 104 CONDUCTIVITY STANDARD SOLUTIONS a DANDAN AATTEND ti oies 63 Ta PARTIS REFERENCE a E a ee MU EU IM tan ire 64 FAO C
63. interdiction Validate by ENTER 02 21 05 18 38 qx Aqu The first selected parameter shows with the current value in the upper right corner of the screen The status bar on the bottom of the screen indicates the recording start date as well as the time on the second precise The two values on the right give the order of the shown values In the example 1 value of a 02 18 05 12 10 04 000001 000096 total of 96 Conduct 252 4uS cm The up and down navigation keys allow visualising the different parameters 02 18 05 12 25 04 000075 000096 SET PAGE 24 24 Ted PO NSEL ACTEON AQUA ACTEWIN MESURE INDEX 1 USER MANUAL REF PONKITAQUA 2 6 CALIBRATION MAIN MENU Choose the option Calibration using the Configuration navigation keys and press enter gt Measure gt Calibration gt Results gt History 02 21 05 18 40 Hy Aqu This menu gives access to the calibration of CALIBRATION each parameter gt Temperature gt Level IMPORTANT Start with the temperature gt Conductivity sensor because this parameter is used in the calibration of conductivity and dissolved oxygen 02 21 05 18 40 JE Aqu 2 6 1 Temperature Temperature factory calibrated original with higth pr cision This calbration is valid for more than one year Therefore recalibrate the temperature only with very TEMPERATURE 1 stable and homogeneous temperature standards Solution 1 Water Ice Reference thermome
64. ke off the stainless steel ballested tip Fix the body of the probe to avoid falling for the majority of the vessels have convex bottoms Numerous calibrations integrate different factors from other sensors like temperature so make sure that all the sensors are submerged at the time of a calibration procedure Also see that the calibration standard volume reaches the connectors e Optimise the effectiveness of a calibration To optimise the accuracy of the calibration procedure rinse the probe with a small volume of calibration standard planned for the calibration and collect this volume in the calibration cap to rinse Don t throw away old calibration solutions because they can be used for this rinsing Prepare a bowl of water at room temperature to rinse the sensors and the calibration cup after every calibration procedure Prepare absorbent paper to dry the sensors and the preservation case every time you rinse the sensors with water or the calibration standard This is to avoid contamination of the calibration standard and will increase the accuracy of the calibration f one or more sensors are disconnected from the AQUA probe make sure that the electric connections are watertight by placing the small protection caps that were designed for this purpose Submerge the sensor head completely in the standard before validating the calibration procedure PAGE 44 44 SET T PO NSEL ACTEON AQUA ACTEWIN PATE 2290
65. l Aut Manual start abbreviation of Automatic appears at the gt Program time bottom of the screen gt Pulse Option The recording will be automatic to every gt Snapshot interval of time selected and will start stop at the programmed times 02 21 05 16 42 dXJAut CD MAIN MENU gt Configuration gt Measure gt Calibration gt Results gt History To record values sufficient time before the programmed starting plug the AQUA probe on the Acteon the Aqu icon AQUA appears on the bottom of the screen after a while of some seconds meaning that the dialog is in progress As long as the probe is connected this 02 22 05 18 30 M Aut Aqu symbol stays displayed Next go to the main menu MAIN MENU Either gt Configuration gt Measure gt Calibration gt Results gt History 02 22 05 19 00 Aut Aqu Rec MEASURE Level 3 22 ft Conduct 15 80mS cm pH 9 03pH ORP 72mV Tempera 59 71 F DO 7 25mg l 02 22 05 19 004 Aut Aqu Rec CO You stay in the Main Menu and at the programmed hour the recording starts The Rec icon appears then on the right bottom of the screen The registration will only stop at the date of end programmed but it is possible to stop it before 3 Configuration Recording Valid Stop recording Or Choose the mode Measure perpetual measuring and at the programmed hour the recording releases It is the recommended mode because you can visualize that measures are
66. libration 5 5 2 Conductivity pH OHP sensor Cleaning of the sensor becomes essential as soon as sediments or pollution appear on the glass surfaces on the platinum point on the conductivity cell or when the response of one of those elements slows down 1 Disconnect the sensor from the AQUA probe by holding the probe by its ring while pulling and proceed cleaning the glass bulb the platinum point and the conductivity cell with clear water and a soft cloth or a cotton bud After that check the measuring of every parameter before starting measuring again 2 If the results of one of the sensors pH ORP or conductivity are not satisfactory carry out the following procedures a Soak the sensor for 15 minutes in a solution of water with a detergent Bleach 10 or Mercryl b Clean with precaution the glass bulb the platinum point and the conductivity cell with a cotton bud soaked beforehand in a solution of water with some drops of dishwasher liquid c Rinse the sensor with clean water clean it with a wet cotton bud and rinse again in a bowl of clean water d Check the functioning of the sensor and carry out calibration of the pH ORP and conductivity parameters Before placing back the sensor on the AQUA probe make sure that the connectors are perfectly dry If necessary take off humidity traces using compressed air 3 Sand the ORP sensor platinium disc with a 500 abrasive wetted with a measuring standard 5 5 8 Tempera
67. me standard between the measurement in Calibration mode theoretical calibration and the standard value This variation makes it possible for the user to check the state of the sensor and the possible drift of its response in time During their calculation these real sensor coefficients are compared with the theoretical coefficients and a message of suspicion can appear if the variation exceeds a certain predetermined percentage Out of margin tolerated spread from 5 to 50 according to the parameter PAGE 26 26 Ted PO NSEL ACTEON AQUA ACTEWIN MESURE INDEX 1 USER MANUAL REF PONKITAQUA 2 6 2 Level LEVEL 1 While the water level sensor is in the air Water at depth level press Enter to validate the Zero of 0 00 ft Measure 0 05 ft ESC exit ENTER next 02 21 05 19 03 dij Aqu e LEVEL 2 Water at depth level of 32 01 ft Measure 31 67 ft On this screen the software gives the possibility to calibrate the sensor in a known depth Enter the value of the second calibration and place the sensor at the indicated level As soon as the value given by the sensor is stabilised for at least 30 ESC prev ENTER Oo seconds press Enter to validate 02 21 05 19 07 4E Aqu LEVEL 2 At the end of the calibration process the name Calibration of the operator is asked 20 characters are OK Message available to enter a name Enter your name correct The following message appears OK J
68. mpty or 2596 CO 225 96 250 a gt 75 O 100 Full Battery capacity New batteries d Bat gt 6 V Acceptable batteries d E From 6 5 to 4 5 V Worn batteries d Bat 4 5 V Change the 4 batteries Three other symbols will appear according to the actions taken Language can be chosen with the up and down arrow keys the ENTER key allows confirming your choice Next press Esc PAGE 12 12 SET DATE January 09 11 ACTEON AQUA ACTEWIN USER MANUAL MESURE INDEX 1 REF PONKITAQUA 2 3 2 DATE HOUR DATE HOUR gt Time set up Press ENTER key for time set up gt Date Format 02 20 05 14 05 dj DATE HOUR The cursor is initially placed on the month Date 02 21 05 enter date and time and confirm by pressing Hour 16 22 the ENTER key Month Day Year Validate by ENTER 02 20 05 14 05 477 CONFIGURATION gt Languages gt Date Hour gt Units gt Parameters gt Memory gt Recording 02 21 05 16 22 d Check date and time modifications on the bottom of the CONFIGURATION menu DATE HOUR gt Time set up If you wish to change the date format go gt Date format down with the navigator s key Down Press ENTER key for date format 02 21 05 16 22 q DATE FORMAT Day Month Year a gasa the format gt Month Day Year ARE gt Year Month Day International ISO 8601 standard Navigate until the date format wished Press ENTER key for date format selected
69. n this solution As soon as the measure given by the probe is stabilized since at least 30 seconds press Enter to validate the calibration Probe in buffer 240mV Measure 245mV ESC prev ENTER valid 02 21 05 19 17 quj Aqu At the end of the calibration process the name of the operator is asked 20 characters are available to enter a name Then press ENTER The following message appears OK In that case the calibration is correct Validate to AQUA EEPROM the logging of the new zero and slope coefs by ENTER The following message appears Out of margin In that case values are out range Either the standard is incorrect inversion or pollution Or the sensor is faulty If the sensor seems good to redo a calibration with new standards Sometimes the sensor is just at the tolerated margin limit 25 and the measure is all the same correct You can disregard then and validate by ENTER Strong shift on the pH 7 or too low Nernst slope After validation the screen saving coefficients appears momentarily PAGE 30 30 NICE SET Ted PO NSEL ACTEON AQUA ACTEWIN eme INDEX 1 REF PONKITAQUA MESURE mw USER MANUAL 2 6 6 DO Dissolved Oxygen DO 1 The probe is placed in a solution of water and Solution Water sulfit sulphite at DO 0 00mgll As soon as the measure given by the probe DO standard Omg is stabilised for at least 30 seconds press Measure 0 02mg l Enter
70. navigation keys and press Enter The symbol Rec on the status bar will disappear as well as all icons associated with recordings AL Aut The manual Stop recording is recommended systematicaly before connecting the Acteon on the PC either to transfer recording files on the hard disk or to reprogram the Acteon by PC Indeed if it subsists the Rec icon associated to Al ou Aut ou JL a no connection message will appear on the PC ActeWin software Acteon Connection impossible Check connections There is absolute refusal of link SDI 12 because all recording in progress by the Acteon has priority Press Esc two times to get back to the main menu Very important disconnecting the AQUA from the Acteon during a recording in progress is not equivalent to a Stop recording Aqu and Rec remain displayed If you try then to connect the Acteon on the PC impossible to establish a dialog beforehand it is necessary to validate a Stop recording Nota You can disconnect and reconnect the AQUA probe on Acteon in the process of recording there will only be a dead gap in the recording corresponding to the disconnection period 2 4 MEASURE PERPETUAL MEASURING Before choosing the mode Measure or perpetual measuring connect the AQUA probe to the female 12 point connector located at the bottom of the case An arrow indicates the connecting correspondence between the plug and the female connector The symbol Aqu meanin
71. nd submerge the conductivity sensor tip in this solution Follow the measure development given by the AQUA probe When this measure is stabilised for at least 30 seconds confirm the second calibration with the Validation symbol The report will indicate if the second calibration was successful or if it is suspect Pass to the next screen pressing Next Screen 4 Displays the taken actions during the two calibration procedures and indicates if they were accepted or if a suspicion was detected during one or both procedures This screen indicates also which Standard needs to be used for the suspect calibrations To repeat a calibration procedure go back to the corresponding screen pressing Previous Confirm conductivity sensor calibration pressing End It is possible to confirm the calibration even if one of the procedures is declared suspect Rinse abundantly and dry the probe and calibration cup for future use f gt Calibration of the pressure sensor Depth Placing the AQUA probe at a zero depth carries out the first procedure screen 1 Gives a summary of the two calibration procedures as well as a description of the actions to be taken during calibration of the pressure sensor SET PAGE 50 50 T PO NSEL ACTEON AQUA ACTEWIN PATE 22909 ig LI MESURE INDEX 1 USER MANUAL REF PONKITAQUA screen 2 Select the parameter unit initialised in cm m in or ft the value of the first Standard initialis
72. nected this symbol stays displayed Next go to the gt History main menu and press the down arrow to choose the mode Measure perpetual measuring 02 21 05 16 38 MA Aqu Press ENTER after the message Measure in process the values of the 6 parameters appear and are refreshed all 5 s You can remain the time that you want in measure without recording MEASURE To launch recording press once again ENTER which starts Level 10 22 ft recording Conduct 9 41mS cm The Rec icon Record appears in bottom and will remain as long as pH 7 63pH the recording is not stopped The abbreviation AY remain on the ORP 282MV bottom of the screen to indicate that the Automatic recording of values Tempera 10 74 F having started in manual mode The recording cannot be stopped DO 9 23mg from this screen 02 21 05 16 39 A1 Aqu Rec To put an end to this recording Esc gt Configuration Recording Valid Stop recording AJ and Rec disappear PAGE 16 16 DATE January 09 11 INDEX 1 REF PONKITAQUA SET ACTEON AQUA ACTEWIN USER MANUAL PONSEL MESURE b PROGRAM TIME PROGRAM TIME Fill in the fields start date time end date time time interval as well as the 14 character alphanumeric filename From 02 22 05 19h00 To 02 25 05 a 23h35 Interval 00h00m10s Name STATION 03 Validate by ENTER 02 21 05 16 42 d RECORDING gt Stop recording After programming the symbo
73. neesasessinegenasnseabosssendendaweastwsneodeumeevssssous es 18 2 4 1 SNAPCHOT RECORDING MODE One shot 1 0 aaa nananana 19 2 4 2 FAST ADJUSTMENT OF THE ZERO FOR THE LEVEL SENSOR 00111111 19 2 5 RESUL o T M 20 2 3 4 PIE DATA E UP E 21 2 3 2 las P BEND WEN Uo ee 22 2 3 3 OR AG E A EE E E SINAPO AALAGA 22 2 5 4 PT a 23 RO a ME SU LE PR Re ea P EOE 24 2 6 1 DCT 1 cera ae AA 24 2 0 2 T 26 2 6 3 Bog C 27 2 6 4 SNR T AA et trrerecrrrrer a 29 2 6 5 QT HH 29 2 6 6 DO Dissolved Oh a RON yer 30 Ar O EEE ee err 30 2 7 1 CANDAL H 31 2 Maintenance RR 32 la AFTER SALES SERVICE DERE 32 2 7 4 LUI NO UIRTLL UR 32 3 PUTTING INTO SERVIGBE ceiuxtestnisstni satum ADA LA AA dock 33 3 1 CONNECETONS a a O 33 3 1 1 AQUA PROBE 33 3 1 2 TEV VET CT EFON nce 33 3 4 SIR NIRE AUI SP Re m 33 4 ACTEWIN DATA HANDLING SOFTWARE 111111 aa 35 5 POUS PROBE GA PII NI NER DM IM dc den eee TIN E 36 5 1 STARTING UP THE PROP PAANAN c 36 3g COMIC ING IE
74. ological information Swallowing large amounts will cause nausea vomiting and diarrhoea Contact with the skin can cause irritation and dermatitis in case of long lasting exposure Contact with the eyes can cause irritation and conjunctivitis No mutagen or teratogen effects known e First aid Ingestion Rinse the mouth carefully with water Drink a large amount of water In serious cases see a Specialist Skin contact Wash abundantly with water and soap Eye contact Wash abundantly with water If the irritation persists see a specialist e Protection Carrying of protective glasses and rubber gloves is advised e Handling and storage Handling Don t eat or drink while handling the product Wash hands and face after using the product Storage Store at room temperature 15 25 C Protect from direct sunlight Keep out of reach from children Keep the container closed e Waste removal Chemical waste is often classified as special waste and collection is often organised locally Please contact the local authorities for details PAGE 61 61 NICE SET Ted PO NSEL ACTEON AQUA ACTEWIN eme INDEX 1 REF PONKITAQUA MESURE mw USER MANUAL 10 2 Buffer standard ORP 470 mV Composition Sulphuric acid lt 5 Yo Water e Physical and chemical properties Appearance clear yellow liquid Smell Odourless Solubility can be mixed with water Melting point not determined Boiling point not
75. onductivity page 22 Check always beforehand menu Configuration gt Parameters gt gt Conduct Auto then Esc In this example the parameters pH Temperature and Oxygen are confirmed the others canceled Alone these 3 parameters will be measured displayed and recorded There are three memory possibilities Watershed memory Once the memory is full the first data will be erased to create memory space for writing current data according to the first in first out method When choosing this option the symbol CD will appear on the lower right side of the screen Fixed memory Once the memory is full recording will stop When choosing this option the symbol Ewi appear on the lower right side of the screen Nota filling up by 1 4 slices see chapter 2 2 page 6 Reset Allows complete erasing of the memory All data recording files are irreversibly destroyed at the same time the register RESULTS becomes empty Validate this option by the ENTER key For the option Reset a user confirmation is required Are you sure No Yes No The MEMORY menu screen comes back Yes The message Erasing Flash appears for 4 seconds on the screen then goes back to MEMORY PAGE 15 15 Ted PO NSEL ACTEON AQUA ACTEWIN MESURE INDEX 1 USER MANUAL REF PONKITAQUA 2 3 6 RECORDING RECORDING 2 AUTOMATIC RECORDING MODES FOR PARAMETERS gt Stop recording gt Man
76. opment of the measure given by the AQUA probe When this measure is stabilised for at least 30 seconds confirm the calibration procedure with the Validation symbol The report indicates if the second calibration has worked or if it is suspect To pass to the second calibration press Next Rinse the probe with water and dry it Rinse abundantly and dry the calibration cap for the next calibration screen 4 Gives a summary of the taken actions during the two calibration procedures and indicates if the calibration has functioned well or if one or both calibrations are suspect If one of the calibrations didn t function well this screen indicates also which buffer standard to use again To repeat a calibration procedure go back to the corresponding screen pressing Previous Confirm calibration of the pH sensor pressing End A calibration procedure can be validated even if one of the procedures is declared suspect Rinse the probe and the calibration cup with water and dry it for future use b gt Calibration of the oxygen sensor The first procedure is carried out with a mix of water and sulphite to simulate a 0 00 mg l value screen 1 Gives a summary of the two calibration procedures and a description of the action to take during calibration of the oxygen sensor Screen 2 The parameter unit is fixed in mg l and the value of the fist measure at 0 00 Pour the necessary quantity of the sulphite water mixture in the cle
77. press ENTER Conduct 9 41mS cm For recording in the same file press ENTER pH 6 63pH successively two times preserving thus the same ORP 185mV file name Recording is confirmed by an audible Tempera 12 77 F beep The indications J and Rec are displayed at DO 9 29mg l the bottom of the screen To leave this mode and 02 21 05 18 23 JM Aqu Rec get back to the Main Menu press Esc Icons Jand Rec disappear then To get back to the Main Menu press Esc 2 4 2 FAST ADJUSTMENT OF THE ZERO FOR THE LEVEL SENSOR MEASURE In the continuous measuring mode the zero Level 0 05 ft value of the level sensor can be fixed before Conduct 9 41mS cm putting the probe in the water pH 6 63pH For this purpose sensor in air press successively Del ORP 185mV Tempera 72 17 F DO 8 29mgp 0 02 21 05 18 25 JE Aqu inen Paver Warning A zero adjustment is impossible in the recording mode PAGE 20 20 NICE SET Ted PO NSE ACTEON AQUA ACTEWIN em MESURE USER MANUAL INDEX 1 REF PONKITAQUA MEASURE During calibration this screen remains in operation for 5 to 10 seconds according to the measuring cycle The 6 parameters are displayed after calibration Warning A zero adjustment is forbidden in recording mode Level offset Calibration in process 02 21 05 18 25 il Aqu MEASURE Level 0 00 ft Conduct i 9 41mS cm pH 6 63pH The level sensor now indicates 0 feet ORP 185mV Tempera
78. raph also explains storing conditions for shorter or longer periods of time up to several months Before starting the installation of a sensor or cap on the sensor port make sure that the two connectors that will touch are perfectly dry to prevent water from coming in these connectors If humidity is detected dry the connector using compressed air If the connector is corroded send the connector back to the after sales services of Ponsel NEOTEK division The connector that links the AQUA probe to a cable must remain permanently covered to protect it against humidity Traces of humidity can be taken away using compressed air When the probe is in communication a cable must have been connected beforehand Likewise when the probe is not in communication a cap that is placed in the connector opening replaces the cable This cap can be easily removed by pulling it lightly PAGE 41 41 SET T PO NSEL ACTEON AQUA ACTEWIN PATE January 09 11 aai MESURE PU USER MANUAL ih REF PONKITAQUA 5 5 Maintenance of the sensors 5 5 1 Dissolved Oxygen sensor When the sensor is used for a longer period of time change the oxymetric electrolyte and the membrane at least once a month Furthermore the electrolyte and membrane must be replaced in the following cases Presence of bubbles under the membrane Presence of electrolyte sediment white on the membrane The sensor shows unstable measurements a doubtful O ca
79. ress ENTER The following message appears OK The calibration is correct validate to AQUA EEPROM the logging of the new zero and slope coefficients by ENTER The following message appears Out of margin Values are out range Either the standard is incorrect inversion or pollution Or the sensor is faulty If the sensor seems good recalibrate with new standards Sometimes the sensor is just at the tolerated margin limit 50 100 and the measure is all the same correct You can disregard then and validate by ENTER The sensor will be calibrated and will deliver some correct values in Measure but this sensor must be watched because maybe at end of life example strong shift on the pH 7 or too low Nernst slope After validation the screen saving coefficients appears momentarily PAGE 28 28 Ted PO NSEL ACTEON AQUA ACTEWIN MESURE INDEX 1 USER MANUAL REF PONKITAQUA IMPORTANT REMARK If in the fields Measure shows off uS cm or mS cm it means that the conductivity standard solution has a too elevated value for the selected range signal saturated Choose a weaker value standard appropriated to the range under calibration CONDUCTIVITY 2 Range 0 2000 S cm In this overflow case the standard is for example 12 Probe in solut 880uS cm too strong a value for this range Standard 1413uS cm Measure US cm Tempera 75 39 F ESC pr c ENTER valid
80. ription of the actions to take during calibration of the ORP sensor Screen 2 The parameter unit fixed in mV enter the value of the first Standard fixed at 0 mV Place the AQUA probe in the air and follow its measure developments When the measure is stabilised for at least 30 seconds confirm the calibration procedure by pressing the Validation symbol The report will indicate if the first calibration is validated or suspect To pass to the second calibration press Next The second calibration is carried out with an ORP buffer standard of 240 or 470 mV Screen 3 The parameter unit fixed at mV enter the value of the second Standard initialised at 240 or 470 mV Pour the necessary volume of the selected buffer standard see Table 1 in the clean dry or pre rinsed calibration cap and submerge the ORP sensor tip in this solution Follow the measure development given by the AQUA 100 probe When this measure is stabilised for at least 30 seconds confirm the second calibration with the Validation symbol The report will indicate if the second calibration was successful or if it is suspect Pass to the next screen by pressing Next SET PAGE 49 49 T PO NSEL ACTEON AQUA ACTEWIN PATE January 09 11 Pi Li MESURE INDEX 1 USER MANUAL REF PONKITAQUA Screen 4 Displays the taken actions during the two calibration procedures and indicates if they were accepted or if a suspicion was detected during one or bot
81. rmometer of which the functioning has been checked beforehand Select the unit initialised in C and enter the temperature of Standard 2 Pour an appropriate volume of Standard 2 see Table 1 in the clean dry or pre rinsed calibration cup and submerge the temperature sensor tip in this liquid Wait some minutes for the sensor to reach the temperature of Standard 2 When the measure given by the AQUA probe is stabilised for at least 30 seconds confirm the second calibration by pressing the Validation symbol The report will indicate if the second calibration has worked or if it is suspect Pass to the next screen pressing Next Screen 4 Displays the taken actions during the two calibration procedures and indicates if they were accepted or if a suspicion was detected during one or both procedures This screen indicates also which Standard needs to be used for the suspect calibrations To repeat a calibration procedure go back to the corresponding screen pressing Previous Confirm calibration of the temperature sensor pressing End It is possible to confirm the calibration even if one of the procedures is declared suspect Rinse and dry the probe and the calibration cup for future use d 5 Calibration of the ORP sensor The first procedure is carried out while exposing the AQUA probe to air A short circuit is applied to the ORP sensor to simulate a O mV value Screen 1 Summary of the two calibration procedures and desc
82. temperature sensor calibration needs to be done In this example Dissolved Oxygen sensor calibration is soiled of suspicion and needs to be redone standard mistake or sensor in question In this example conductivity sensor calibration on the range 0 200 0uS cm needs to be done PAGE 32 32 Ted PO NSEL ACTEON AQUA ACTEWIN MESURE INDEX 1 USER MANUAL REF PONKITAQUA 2 7 2 Maintenance MAINTENANCE In this history the user can make notes ELECTROLYTE CHANGE concerning maintenance performed 126 DO SENSOR 01 14 05 alphanumeric characters are available for a description of the actions taken 02 21 05 20 07 TM Aau 2 7 3 AFTER SALES SERVICE INFORMATION AFT SALES The after sales informations reserved for FAULTY CONNECTOR factory interventions Only one access code REPLACEMENT 01 02 05 gives access to the information registered in this history 02 21 05 20 08 J Aqu 2 7 4 ACTEON 3000S ACTEON 3000S This screen gives information concerning Hard V2 0 Soft A4 03 hardware characteristics and Acteon software Flash 1024k 0 10m version It shows installed memory capacity 1024kB until 8MB the variant of the depth Bat 6 1V Bar 1012hP sensor 0 10m 0 50 m as well value of battery voltage values of the internal barometric pressure sensor 02 21 05 20 09 M Aqu a CONTRAST BRIGHTNESS CONTROL In this last page the contrast of the LCD screen can be adjusted using the keys number 7 and
83. ter with 1 100 of degree C r solution Standard 32 00 F Measure 31 65 F The probe is placed in an ice bath to simulate a 32 F value or 0 C if units SI selected Agitate continuously the probe in the middle As soon as the measure given by the probe is stabilised for at least 10 minutes press Enter to validate the calibration ESC exit ENTER next 02 21 05 18 45 f7 Aqu Water must be constantly stirred because of o E Q TEMPERATURE 2 E Calibration in water of a stable temperature with Solution 2 Water Se a value measured by a precision thermometer S between 68 F and 104 F that it to say Standard 68 00 F 8 between 20 C et 40 C depending of the Measure 68 21 F 1 preferential measure zone S Dip the probe in the water while agitating ESC exit ENTER valid continuously in the middle and as soon as the 26 02 03 18 55 fim Aau measure given by the probe is stabilised for at least 10 minutes press Enter to validate the calibration PAGE 25 25 Ted PO NSEL ACTEON AQUA ACTEWIN MESURE INDEX 1 USER MANUAL REF PONKITAQUA TEMPERATURE 2 At the end of the calibration process the name Calibration of the operator is asked 20 characters are OK Message available to enter a name Enter you name correct The following message appears OK J BURTON calibration The calibration is correct validate to AQUA EEPROM the logging of the new zero and slope ESC
84. the sensor is suspect Note A calibration procedure can be confirmed even if it is suspect 5 8 2 Calibration procedures of AQUA probe a Calibration of the pH sensor The first procedure is carried out with a standard buffer of pH 7 and 20 C Screen 1 Gives a summary of the two calibration procedures with a description of the actions to be taken during calibration of the pH sensor Screen 2 The parameter unit being fixed pH enter the value of the first buffer initialised at 7 01 Pour the necessary volume for buffer pH7 see table 1 in the clean dry or pre rinsed calibration cup and submerge the pH sensor tip in this liquid Follow the development of the measure given by the AQUA probe When this measure is stabilised for at least 30 seconds confirm the calibration procedure with the Validation symbol The report indicates if the first calibration has worked or if it is suspect To pass to the second calibration press Next PAGE 46 46 SET T PO NSEL ACTEON AQUA ACTEWIN PATE 22709 aai MESURE PU USER MANUAL ih REF PONKITAQUA The second calibration is carried out with a buffer standard pH 4 pH 9 or pH 10 screen 3 The parameter unit being fixed pH enter the value of the second initialised buffer 4 01 9 00 or 10 01 Pour the necessary volume for buffer standard see Table 1 in a clean dry or pre rinsed calibration cup and submerge the pH sensor tip in this liquid Follow the devel
85. togen effects known e First aid Ingestion Rinse the mouth carefully with water Contact with the skin Wash abundantly with water and soap Contact with the eyes Wash abundantly with water If irritation persists see a specialist e Protection Carrying of protective glasses and rubber gloves advised e Handling and storage Handling Don t eat or drink while handling the product Wash hands and face after using the product Storage Store at room temperature 15 25 C Protect from sunlight Keep out of reach from children Keep the container closed e Waste removal Chemical waste is often classified as special waste and collection is often organised locally Please contact the local authorities for details PAGE 64 64 7e PO NSE ACTEON AQUA ACTEWIN MESURE INDEX 1 USER MANUAL REF PONKITAQUA 11 PARTS REFERENCES Eee PONACTEON ACTEON 3000 delivered in a bag PONAQUA AQUA 100 multiparameter probe with 10m cable PONACTEWIN ACTEWIN software with SDI12 RS232 adapter and cables rn eee PONAQU CAUSB USB gt RS232 Adapter PONAQU CAB2 AQUA 100 lt ISCO interface cable O YY PONVM3 Travelling bag PONVM4 AA eee YY Stainless steel ballast 200g for AQUA Stainless steel ballast 500g for AQUA Ch PONACTW ADPT Actewin SDI12 RS232 adapter PONACTW CAQU AQUA lt Adapter SDI12 RS232 cable 1 50m PONACTW CACT Acteon lt Adapter
86. ture sensor This sensor doesn t need any specific maintenance and can be simply cleaned with a solution of water with some drops of dishwasher liquid as soon as sediment traces appear PAGE 42 42 NICE SET Ted PO NSEL ACTEON AQUA ACTEWIN em MESURE PU USER MANUAL ih REF PONKITAQUA 5 6 Storage conditions for AQUA probe 5 6 1 Short term storage For short term storage leave the sensors connected to the AQUA probe under a humid atmosphere according to one of the following methods 1 Put approximately 5 mm of initial preservation orangey liquid in the preservation cap and place it on the probe head The sensors should not be submerged 2 Place a humid sponge or cotton disc thickness 5 mm impregnated with preservation orangey liquid to the bottom of the preservation cap and screw it on the probe head In both cases make sure that the preservation cap is hermetically closed to avoid any evaporation Noted If you no longer have preservation orangey liquid Ref PON SOL CSV use the oxymetric liquid electrolyte KCL Ref PON 02 EL Especially no distiled or pure water or then 5 drops to the bottom of the cap only to maintain a climate of humidity on sensor heads The communication cable of the probe must be removed and replaced by a small protection cap delivered with probe to avoid humidity from entering the connector opening Storage rules are thus as follows gt Use a quantity of liquid t
87. ual start Manual start automatic recording mode with manual start and stop gt Program time gt Pulse Option Program time automatic recording mode with departure and stopping gt Snapshot dates programmed in advance Pulse Option is only an optional complement for Manual start and 02 21 05 16 35 J Program time modes for waters samples Snapshot is not an automatic mode but measures logging bit by bit purely manual which is not configured in advance see 2 4 1 If you validate Snapshot in this menu a simple information appears Press ENTER in measure mode a MANUAL START MANUAL START Type an interval time between a minimum of 5 seconds and a maximum of 23h59m59s Interval 00h00m05s then type a file name of 14 alphanumeric Name STATION 02 Validate by ENTER a 02 21 05 16 37 i RECORDING After programming the symbol A4 appears on gt Stop recording the bottom of the screen gt Manual start gt Program time AJ is the abbreviation of Automatic Manual gt Pulse Option start gt Snapshot The recording will be well automatic to every interval of time selected but the start is 02 21 05 16 37 dE AY launched by hand MAIN MENU To record values gt Configuration Plug the AQUA probe on the Acteon the Aqu icon AQUA Measure appears on the bottom of the screen after a while of some Calibration seconds meaning that the dialog is in progress As long as the Results probe is con
88. ve on your PC hard disk through Actewin __ PONSEL SET en rie tlk ACTEON AQUA ACTEWIN Mu MESURE INDEX 1 USER MANUAL REF PONKITAQUA 1 3 Is the Acteon still in recording mode most frequent case If Rec icon is still displayed on the screen with or without other associated icons like Aut AJ SL the Act on terminal is still in automatic recording mode whether the AQUA probe is connected or not and whatever the starting mode which has been initially chosen manual or programmed In this case go to MAIN MENU gt Configuration gt Recording Stop recording Check that the Aqu icon reappears with the probe connected and that the 6 measured values are refreshed every 5 seconds 1 4 The previous checks have not been successful and the icon Aqu doesn t appear on the screen the AQUA probe cannot be detected from the Acteon Try to connect the AQUA to a PC using ACTEWIN and the SDI12 RS232 connexion to the RS232 port preferably COMI using the appropriate cable which has been supplied If the probe is detected activating the AQUA 100 menu it s very likely that the first Acteon AQUA cable is defective and the AQUA probe is not to blame The Acteon terminal is usually not to blame it can also be tested connecting it to the PC and activating the Import or Recording Programming menu in Actewin If very unusual the ACTEON does not talk with the PC under Actewin you should reset the ACTEON opening the batteries
89. y 02 21 05 16 22 dj Then press Esc If you select for ex the International ISO 8601 standard the date displayed down on the status bar will become PAGE 13 13 Ted PO NSEL ACTEON AQUA ACTEWIN MESURE INDEX 1 USER MANUAL REF PONKITAQUA 2 3 3 UNITS UNITS gt Sl m C gt Anglo saxon ft F y You can choose the units for measures SI meters and Centigrade Celsius degrees Anglo saxon Feet and Fahrenheit degrees Navigate until the units wished Press ENTER key for units selected gt y 02 21 05 16 23 d Then press Esc MEASURE If you chose and validated the Anglo Saxon units Level 11 22 ft Feet and Fahrenheit degrees Conduct 9 41mS cm pH 6 63pH All Depth and Level values will be displayed and ORP 185mV recorded in ft all temperature values in F for Tempera 72 77 F calibrations and calibration history DO 8 29mgll In Measures mode you will see displayed the 02 21 05 16 24 fil Aau following values on the screen MEASURE If you chose and validated the SI units Level 3 42 m Meters and Celsius degrees 9 41mS via l 6 63pH uu All Depth and Level values will be displayed and ORP 185mV recorded in m all temperature values in C too for Tempera 22 65 C calibrations and calibration history as well DO EP 8 29mg l In Measures mode you will see displayed the 02 21 05 16 24 qx Aau following values on the screen Nota 1 ft 0 3048 m 1 m