1. Cycle start faults cya sh 54 acest ok bhweds Faults at end of autocalibration Faults in sniffer mode 0000 Helium measurement problem I O interface problem C 10 C 20 C 30 C 40 C 50 C 60 C 70 D 10 D 20 D 30 D 40 D 50 D 60 D 70 D 80 D 90 D 100 D 110 D 120 2 3 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 04 September 97 Contents User s Manual ASM 180 TD TD ASM 181 TD Maintenance Analyzer cell maintenance E 10 Sheets Greasing the hybrid turbomolecular pump E 20 Greasing the molecular drag pump E 30 Replacement Recalibration of the detector internal calibrated leak 2200 E 40 Autocalibration of the detector with an external calibrated leak 259 eeisdagends sans Ges E 50 MD4E membrane pump maintenance E 60 CP20 Partial maintenance E 70 I O interface board fuse replacement E 80 Components Preventive maintenance components delivered with the detector 0 0 0 0000000005 F 10 Monitoring and display F 20 Power and electrical supply 2 0 F 30 Automatism and electronic circuits F 40 Measurement a42ec sd eqacddodaaanues F 50 PUMPINg ucts chee dard 46 ar Meher teat Wear oe Bora epee F 60 VANES a a a ee eee eee ee ee ee ee F 70 PD
2. 10 11 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 03 May 97 SYMPTOM The detector does not switch to Fine Leak mode for example after 5 min fora 1 4 line length lt 100 m The helium background noise does not decrease for example the helium signal remains at the 107 scale The minimum detectable leak is limited to the value of the helium background noise CAUSE Gross Leak on installation The carrier gas contains a significant helium concentration NO Gross Leak on installation C 70 In the event of problems when using the I gas line test option REMEDY Inject the carrier gas 40 SCCM and test the installation page 10 The leaks at the connections or the line are displayed on the GL measurement scale if the detector is in GL mode or Inlet Pressure display if the detector is in roughing mode e Vary the carrier gas flow while remaining in FL mode If the helium signal rises with the carrier gas flow the carrier gas contains helium e Purge the carrier gas injection system again to remove any pos sible trace of residual air If the problem persists the test can be performed in the background noise limit or change the carrier gas Vary the carrier gas flow while remaining in FL mode If the helium signal rises or remains practically constant when the car rier gas flow falls there is a gross l
3. Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 1 Edition 03 May 97 C 10 Starting up the detector In G 10 the user will find a view of the operator interface It can be used to identify the operational parts of the control panel and remote control unit Before starting up the detector check that the I O plug connector is present see sheet B 70 Power up Set the circuit breaker switch to The roughing pump is started The cycle control button green indicator light flashes around 20s Once the primary pressure MD4E or CP20 fo Mo threshold has been reached the by pass valve a opens the indicator light comes on and 900 the molecular roughing MDP is started The molecular pump is in the acceleration phase When the by pass pressure threshold is detected the valve closes The molecular pump reaches its nominal rotational speed in 2 to 3 min Alcatel High Vacuum Technology ASM 180 TD TD ASM 181 TD User s Manual 2 C 10 Starting up the detector Once the exhaust pressure threshold has been reached after approximately 25 s the P indicator light comes on and the secondary pump TMP is started The secondary pump is in the acceleration phase It reaches its nominal rotational speed in 2 to 3 min The filament on phase is started The filament key indicator light flashes for approx 8 seconds and becomes steady
4. Remove the rear cap from the pump 4 allen head screws This cap is directly accessible from underneath the leak detector The grease syringe is equipped with a black clip A and a red clip B These clips are used as stops fo control the amount of grease injected into the bearing Remove black clip A Push the grease syringe in through the screw hole 1 until it comes up against a stop Inject grease pushing in the plunger until it stops at clip B Remove the red clip B and distribute the grease between the injection points smooth holes directly opposite each other 2 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 2 E 20 Greasing the hybrid turbomolecular pump PTM 5154 Distributing the grease Set the SW1 switch on the CD2 board to the breakin R in the bearings position and run the detector for 10 minutes with it in this position Then reset SW1 to its normal N position SW1 switch R Running in N Normal 2 2 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 04 September 97 Greasing the molecular drag pump MDP 5011 The frequencies of preventive maintenance are listed in D 10 Components P N Grease syringe eceeeeee 056993 Accessing the bearings Remove the clips and use the extractor to take out the plug see F10 item
5. NO CP20 pump overheating NO REMEDY Contact customer service Contact customer service Contact customer service Contact customer service Check the cooling system Contact customer service Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 3 D 30 Problem with the roughing pump SYMPTOM CAUSE REMEDY CP20 pump has stopped during cycle ASM 180 TD ASM 181 TD Power supply voltage is less than the minimum allowed the pump turns into safe Check the power supply voltage stop the detector and restart the CP20 pump NO Contact customer service MDP The by pass indicator light does never come on PI3 inlet gauge is defective When inlet port is Patm no inlet pressure display Change PI3 inlet gauge See D 70 NO Supervisor electronic board Contact customer service Wrong setting of supervisor board NO Contact customer service 2 3 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 04 September 97 D 30 Problem with the roughing pump SYMPTOM CAUSE REMEDY The MDP fault bet MDP indicator light y is on Uae Set it on N See E 30 CD2 board SW1 switch is set on R NO Check the roughing pump MD4E ASM 180 TD pressure lt 3
6. 5 0E 12 test by releasing READY MAN CYCLE The filament is off temporarily while he LDS probe is being roughed down The corresponding fault is displayed After a few seconds the sniffer test is operational The measurement is displayed according to the selected unit see page 5 Return to the basic display See page 9 for the available operations during a Manual cycle test 8 27 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 03 May 97 Operations available during manual mode test cycle Display the time elapsed since start of cycle Vo A Press W Print the current measurement Press W A simultaneously Switch off the filament see C20 Press Switch on the filament see C20 Press Modify the FIXED manual reject setpoint Press on IF Adjust the displayed setpoint with the key M and A Press on IF to valid and to return to display during the manual cycle test C 50 Alphanumeric Control and Display Panel ACDP operation Manual test basic display MAN FL 1 0 E 09 INLET P 1 0 E 03 MAN FL 1 0 E 09 CHRONO 275 MAN FL 1 0 E 09 CHRONO 375 FIL DEF SPECTRO R aos m Peete Shorea MAN FL 1 0 E INLET P 1 0 E 03 MAN FL 1 ae REJECT 1 0 mm ore Return to manual test basic display Timer time elapsed since start of cycle Instanta
7. ASM 181 TD ltem MUSE CP 20 eee CP 20 Partial m 2 bal 2 Or Ori PUMP 20 defraud teary tiiti pump controller aintenance kit CP 20 including bearings 6001 CP ings 2 x 27 ng 3 x 96 grease tube 10 g 1 plastic box F 60 103499 2 3 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 04 September 97 oe f ltem Description TMP 5154 Standard seal TMP 5154 Shock absorber MDP 5011 Greasing syringe for TMP 5154 and MDP 5011 D N O F 60 P N 798023 055232 795600 056993 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 3 3 Edition 04 September 97 F 70 Valves ASM 180 TD Valve block Calibration Roughing Venting Item Description a Roughing valve b Detection valve VAT NW 25 valve operator 5 2 VAT 180 260V valve coil z 3 Seal kit for VAT valve not pictured 4 Grease tube 10g for VAT valve i not pictured 5 Bacosol 24V DC 23W valve exhaust 104655 6 Minisol valve 2 2 NO 24V DC 101304 roughing system vent with coil 7 Bacosol 24V DC 23W valve by pass 104655 8 Calibration block 100973 9 Bacosol 24V DC 23W valve air inlet 104655 10 Minisol valve 2 2 NF LDS without coil 038101 4 11 Coil for 24V DC minisol valve 067040 LDS valve 12 Vickers silencer 075990 connected to the TMP rear 13 Valve block
8. Fuse on the P0318 board is defective NO Contact customer service Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 1 Edition 03 May 97 D 80 Cycle start faults SYMPTOM CAUSE REMEDY The cycle key is disabled green indicator light on Filament off filament key indicator light is off Press the filament key the indicator light comes on Otherwise see D 60 NO LDS mode selected EZA LDS key indicator is ON Release the EZ LDS key to switch to vacuum test mode NO W Contact customer service The cycle key is ES disabled indicator lights off I O plug connector This is not connected to the rear of the detector Switch off the detector and connect the I O plug connector Start up again see B 70 NO Contact customer service Note The cycle key green indicator light only indicates that autocalibration is valided The fact that the indicator light is off does not inhibit the use of the cycle key Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 1 Edition 03 May 97 D 90 Faults at end of autocalibration Note if a printer or a micro computer is connected to the RS232 interface configured in printer mode a default ticket is emitted which provides diagnostic assistance see B 30 SYMPTOM CAUSE Electrical zero is P
9. Test No The measured Helium signal is less than the Reject level and the Warning level The measured Helium signal is less than the Reject level but greater than the Warning level WL lt Measured S lt RL The measured Helium signal is greater than the Reject level Switching to test mode is not possible within the roughing time part refused for very gross leak The test has not been validated the cycle is interrupted READY AUTO CYCLE The test is not counted DEF code X Explanation 6 FIL OFF in roughing 7 FIL OFF in test 9 CYCLE OFF in roughing 10 CYCLE OFF in test 1 Pressure RISE in test Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 13 27 Ticket printing examples AUTO mode Alphanumeric Control and Display Panel ACDP operation FILE 1 CUSTOMER reference S level mbar I s T timer sec S rejected 2 8E 09 S alarm 9 9E 10 FILE 2 S level mbar s T timer sec S rejected 1 0E 08 S alarm 8 0E 09 T roughing 9 T roughing 10 T test 6 T test a PRINTER ON PRINTER ON lt lt number TEST number SERIAL t 26JUNE 96 26JUNE 96 CUSTOMER reference FILE 2 TEST Nr 5 AUTO MODE FL PASS signal 9 0E 10 26JUNE 96 16 6 37 CUSTOMER reference TEST Nr 6 AUTO MODE FL PASS signal 1 5E 09 Warning level exceeded 26JUNE 96 16 10 08 Modifiable and validate
10. a calibration certificate To install this new calibrated internal leak e Switch off the detector and unplug it from the main power e Install the leak DN 16 connector heat sensor Before the front cover is closed the new value of the leak must be entered This is carried out using the three thumbwheels located on the supervisor board placed in the front cover e Set on the thumbwheels the helium value of the calibrated leak at 20 C marked on the leak label as in the following example 1 5 107 mbarl s or 1 5x107 atm cm3 s O srr Set S4 to l S3 to 5 S2 to7 The internal calibrated leak value must be between 1 0E 8 and 8 0E 6 mbarl s e Check that the switch S1 is set to OFF adjacent red indicator light off internal autocalibration enabled 2 3 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 03 May 97 Intensive use of the detector E 40 Replacement Recalibration of the detector internal calibrated leak e Close the front cover 4 screws e Connect the detector to the main power e Switch it on an autocalibration is performed automatically at the end of the startup sequence lt is recommended to repeat an autocalibration after 1 hour of operation w
11. choice can be set the volume to O switch off and then on the filament see page 9 briefly expose the detector to a helium source in order to higher up the audio signal frequency and then let it lower down until it stops Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 23 27 C 50 Alphanumeric Control and Display Panel ACDP operation Default ACDP A board P0192 specific to configuration he ACDP option is located on the righthand side inside the front cover of the unit This board controls the ACDP panel and the associated printer output see B 40 Setting ranges ACDP RS232 board t contains switches which are PO192 used to configure the ACDP operator interface See g Switch Function Position Action i Display and ON saul printing language OFF French 4 destination 2 Not used none 3 Audio signal ON by ACDP 2 monitoring OFF by standard 1 control panel AUDIO zone 4 Background setpoint ON enabled i OFF disabled Factory configuration 1 without ACDP option 2 with ACDP option 24 27 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 03 May 97 C 50 Alphanumeric Control and Display Panel ACDP operation Test paramet
12. the bearing surfaces and the support disc 9 Put a new membrane in place On top of it position the tension disc Pour a drop of low thread braking fluid loctite on the screw threads and secure the disc see page 2 The pump itself must reach a limit pressure lt 3mbar If necessary plan a gross leak test with alcohol Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 3 3 Edition 04 September 97 E 70 Partial maintenance of the CP 20 pump The frequency of the preventive maintenance tasks is listed in section D 10 Components P N Partial maintenance kit 103499 including Ball bearing 6001 CP per 2 O ring 200 02700 G2 T47501 FPM72 inlet side ball bearing per 2 O ring 300 09600 G2 T47501 FPM72 stator Grease tube D101 ultrathermique 200 Plastic box caubere 6532 rectangular Tools required Allen wrench 3mm 3 Amm 4 and 5mm 5 Open end wrench or ring spanner 10 PJ Phillips screwdriver F4 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 3 Access to the CP 20 pump E 70 Partial maintenance of the CP 20 pump On the ASM 180 TD the dismantling of the pump from the frame is not necessary to perform a partial maintenance Disconnect the detector power supply from mains safety precaution ASM 180 TD ASM 181 TD 5 Open the rear cover E Remove th
13. 1 2 10 10 10 mA mbar AUTOCAL The purpose of this sheet is to identify the activated keys or iV dec j parts of the operator interface while the detector is in operation OD REMOTE CONTROL UNIT He mbar l s P mbar r 1 Atm w w 5 2 2 5 2 3 5 2 w4 5 2 5 w gt G gt m a ALCATEL Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 1 Edition 03 May 97 G 20 Recording curve 1V dec The purpose of this sheet is to show the logarithmic response curve of the analogue output located on the control panel OC Analog output chart Volts Signal He FL Signal He GL Volts Signal He FL Signal He GL Volts Signal He FL Signal He GL 0 10 3 07E 12 3 07E 10 2 25 1 64E 08 1 64E 06 4 40 2 50E 06 2 50E 04 0 15 1 12E 11 1 12E 09 2 30 1 85E 08 1 85E 06 4 45 2 81E 06 2 81E 04 0 20 2 08E 11 2 08E 09 2 35 2 08E 08 2 08E 06 4 50 3 15E 06 3 15E 04 0 25 3 20E 11 3 20E 09 2 40 2 35E 08 2 35E 06 4 55 3 53E 06 3 53E 04 0 30 4 51E 11 4 51E 09 2 45 2 64E 08 2 64E 06 4 60 3 97E 06 3 97E 04 0 35 6 03E 11 6 03E 09 2 50 2 98E 08 2 98E 06 4 65 4 45E 06 4 45E 04 0 40 7 80E 11 7 80E 09 2 55 3 35E 08 3 35E 06 4 70 5 00E 06 5 00E 04 0 45 9 84E 11 9 84E 09 2 6
14. 2 Sniffer mode LDS Gross leak mode Fine Leak mode Cycle in progress Filament on Helium signal gt Reject setpoint Analog output O 10 VDC inlet pressure O 8 VDC analogue output Helium signal Internal ground Common external ground Common external ground 2 2 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 04 September 97 B 30 Controlling the detector with a micro computer RS 232 The RS232 interface is used to control the detector with a micro computer Preparing the RS 232 Use a Sub D9 pin female connector link cable Pins used Reception data RD TD Transmission data Data set ready DSR DTR Data terminal ready GND Ground Oll O2 O3 O4 O5 O 06 O7 O8 O9 O Ready to send RTS CTS Clear to send Connection cable Micro computer 7 and 8 Connections are necessary only if RTS and LTS are used in a software created by the user Connecting the detector to a micro computer Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 8 RS 232 transmission parameters Main board B 30 Controlling the detector with a micro computer RS 232 At the first startup the user will find the default configuration E Transmission speed 9600 baud E Data length 8 bits E Parity NONE E Stop bit 1 The transmission spee
15. 3 Once the extractor is in place pull it vertically Using the grease The grease syringe is equipped syringe with a black clip A and a red clip B These clips are used as stops fo control the amount of grease injected into the bearing Greasing the front Remove black clip A bearing Push the grease syringe in through the screw hole 1 until it comes up against a stop Inject grease pushing in the plunger until it stops at clip B Greasing the rear Remove the red clip B and distribute the grease between bearing the injection points smooth holes directly opposite each other 2 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 2 E 30 Greasing the molecular drag pump MDP 5011 Distributing the grease Set the SW1 switch on the CD2 board located in the rear in the bearings cover to the breakin R position and run the detector for 10 minutes with it in this position Then reset SVV1 to its normal N position SW1 switch R Running in N Normal 2 2 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 03 May 97 Purpose of the calibrated internal leak Frequency of internal leak recalibration How to recalibrate the internal leak Removal of the internal calibrated leak E 40 Replacement Recalibration of the detector internal calibrated leak The freque
16. 39 39 686 3855 Fax 39 39 667 125 Japan Alcatel ITS JAPAN High Vacuum Technology Division Ebisu MT Building 5 Floor 15 7 Higashi 3 Chome Shibuya ku TOKYO 150 0011 Tel 81 3 5468 1621 Fax 81 3 5468 1603 Korea Alcatel Vacuum Technology Korea 5th Floor Sunghyun B D 10 5 Karak Dong Songpa ku Seoul SOUTH KOREA Tel 82 2 409 6277 Fax 82 2 409 6279 United Kingdom Alcatel Vacuum Technology U K Ltd 10 Lancaster Court Coronation Road HIGH WYCOMBE HP 12 3TD Tel 44 1 494 448 444 Fax 44 1 494 448 222 USA Alcatel Vacuum Products 67 Sharp Street HINGHAM MA 02043 Tel 1 781 331 4200 Fax 1 781 331 4230 High Vacuum Technology y ALCATEL Alcatel Vacuum Technology France 98 avenue de Brogny B P 2069 74009 ANNECY Cedex France Tel 33 0 4 50 65 77 77 Fax 33 0 4 50 65 77 89 Telex 385 153 F Edition 04 September 97 Contents User s Manual ASM 180 TD TD ASM 181 TD Introduction M The ASM 180 series 2 02 A 10 ASM 180 TD Detector operating principle A 20 ASM 180 TD ASM 181 TD Detector operating principle A 21 Analyzer cell operating principle 0 2 A 30 Testing methods 0 000200005 A 40 Operator interface 2 2 2 000000 008 A 50 COIS erea pense eae a A 60 POSES OUSS a nih Soma owed Sees a Ses A 70 Technical characteristics 2 ee A 80 Dimensions 4 426 u cms dade he ws deere
17. 45 2 25E 09 2 25E 07 3 60 3 92E 07 3 92E 05 6 50 3 16E 04 3 16E 02 1 50 2 56E 09 2 56E 07 3 65 4 40E 07 4 A0E 05 6 60 3 98E 04 3 98E 02 1 55 2 91E 09 2 91E 07 3 70 4 95E 07 4 95E 05 6 70 5 01E 04 5 01E 02 1 60 3 30E 09 3 30E 07 3 75 5 55E 07 5 55E 05 6 80 6 31E 04 6 31E 02 1 65 3 75E 09 3 75E 07 3 80 6 24E 07 6 24E 05 6 90 7 94E 04 7 94E 02 1 70 4 25E 09 4 25E 07 3 85 7 00E 07 7 00E 05 7 00 1 00E 03 1 00E 01 1 75 4 82E 09 4 82E 07 3 90 7 86E 07 7 86E 05 7 10 1 26E 03 1 26E 0 1 80 5 46E 09 5 46E 07 3 95 8 83E 07 8 83E 05 7 20 1 58E 03 1 58E 0 1 85 6 18E 09 6 18E 07 4 00 9 91E 07 9 91E 05 7 30 2 00E 03 2 00E 0 1 90 6 99E 09 6 99E 07 4 05 1 11E 06 1 11E 04 7 AO 2 51E 03 2 51E 0 1 95 7 91E 09 7 9 1E 07 4 10 1 25E 06 1 25E 04 7 50 3 16E 03 3 16E 0 2 00 8 93E 09 8 93E 07 4 15 1 40E 06 1 40E 04 7 60 3 98E 03 3 98E 0 2 05 1 01E 08 1 01E 06 4 20 1 57E 06 1 57E 04 7 70 5 01E 03 5 01E 0 2 10 1 14E 08 1 14E 06 4 25 1 77E 06 1 77E 04 7 80 6 31E 03 6 31E 0 2 15 1 29E 08 1 29E 06 4 30 1 98E 06 1 98E 04 7 90 7 94E 03 7 SAE O 2 20 1 45E 08 1 45E 06 4 35 2 23E 06 2 23E 04 8 00 1 00E 02 1 00E 00 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 He signal mbar l s 1 0E 0 1 0E 1 1 0E 2 1 0E 3 1 0E 4 1 0E 5 1 0E 6 1 0E 7 1 0E 8 1 0E 9 1 0E 10 1 0E 11 1 0E 12 0 1 2 3 4 Output voltage V
18. A3 006 page 26 View of ACDP panel A3 cece cece eeee page 27 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 27 C 50 Alphanumeric Control and Display Panel ACDP operation Purpose of the ACDP This option is intended for industrial inspection option It is used to e display messages concerning the status of the detector e display the measurement in digital form e configure and control the unit in two operating modes manual or automatic e print inspection tickets using an external printer connected to the associated interface see B 40 e To monitor the audio signal according to the selected contiguration Manual operation e The operator retains control of the test cycle start and stop using the cycle key I or the C key on the ACDP e The Red and Green indicator lights rejected part accepted part come on according to the value of the Helium signal in relation to the programmed manual reject level e The choice of test mode gross leak fine leak or sniffer is up fo the user e The detector can be used as if it did not include this option see standard detector operation in C 20 e The audio signal is working on a FIXED or FLOATING mode according to the selected configuration Automatic operation The test cycle is automated according to the programmable test parameters e It is possible to program up to 10 references of different pa
19. ASM 180 TD The connections are made according to the recommendations on page 7 Close the valves 6 and 9 Make sure that the DN4O inlet port is blocked Activate the atmospheric pressure key EES Press the yellow button above the ON switch Wait until the detector enters Fine Leak mode and the analyzer cell pressure is sufficiently low for example no more than 3 green indicator lights on Initially the helium background noise increases briefly and then decreases and becomes stable Note The inlet pressure displayed on the remote control unit is not the pressure at the gas line circuit separated by a membrane inside the detector see detector mimic diagram It is the pressure at the level of the DNAO inlet port However this pressure varies as a function of the pressure in the gas line 8 1 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 03 May 97 C 70 Use of the I gas line option For ASM 180 TD and ASM 180 TD Inject the carrier gas Gradually open the carrier gas supply until the maximum flow allowed is obtained The detector must remain in Fine Leak test mode The inlet pressure and the cell pressure increase as the carrier gas flow increases If the detector switches to Gross Leak mode reduce the carrier gas flow The length of time it takes for the pressure to stabilize in the gas line depends on the length of the
20. Minisol valve 2 2 NC LDS without coil 038101 Minisol valve 2 2 NO 24V DC 101304 roughing system vent with coil Non injection TD plug 104426 Roughing valve a Detection valve b By pass valve c Grease tube 10g for VAT valve not pictured Seal kit for VAT valve not pictured i VAT 180 260V valve coil VAT NW 25 valve operator K Valve block ASM 180 TD not equiped Valve block ASM 180 TD not equiped Vickers silencer 075990 Pipes F80 CP 20 pipe q CP 20 pipe plug Diaphragm NVV 25 without seal j Flexible tube NW 16 lenght 250 mm 068369 Inlet adaptor NW 25 NW40 180 TD Inlet adaptor NW 25 NW40 180 TD Inlet filter NVV 25 without seal 072857 PVC tube 4 x 2 x Rilsan tube 6 Silencer 1 4 101552 Stainless steel tube NWV 25 NW 40 101539 Connections and Seals F90 Calibration block O ring 8 x 1 9 082195 Oring 082113 block valve blank off 16 9 x 2 7 Oring NW 16 079237 Oring NW 25 079238 Oring NW 40 082129 O ring NW 63 082140 Contact customer service Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 3 4 Cover F100 Options and accessories F110 Components summary Description Heyco stopper per unit Lifting ring per unit Stopper for frame per unit ASM 180 TD TD wheel per unit pivoting ASM 181 TD wheel per unit Rear wheel fixed Front wheel pivoting with brake ACDP control assembly 529C 24V bulb for l option switch not p
21. Selection of Serial Nr or Test Nr in part modification menu page 19 At the end of each validated test a serial number for the tested port incremented by one over the previous one is proposed Possibility to manually modify it with the A and key Systematic validation by key IF Note if a cycle is started using the key before validation this cycle is stopped and is not taken in account Return to the basic display The test number TEST Nr or Serial number SERIAL Nr is incremented for the next test The display of the result the possible emission of an audio signal and the printing of a test ticket depend on the programmed test parameters for the selected part see part modification menu page 15 12 27 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 03 May 97 Result message and audio signals at end of automatic test Audio signal None None Fixed signal Fixed signal Error message at end of unvalidated automatic test Two fixed audio signals C 50 Alphanumeric Control and Display Panel ACDP operation A Measured Helium Signal mbar s FILE 1 Accepted part N Accepted part ww Accepted part Test result display PASS 2 0E 10 PASS 3 0E 9 FAIL 1 0E 06 FAIL ROUGHING RETEST DEF 11 M Accepted part see page 15 on Rejected part gt
22. Unpacking B 10 Precautions and unpacking Before switching on the unit the user should read the safety instructions supplied with the detector and be sure to follow them When the equipment is received unpack it carefully do not discard the packaging until you have made sure that the unit has not been damaged during transport The following are supplied with your unit an instruction manual a maintenance kit the calibration certificate of the internal calibrated leak If one of these parts is missing contact ALCATEL immediately Check the packaging tilt indicator of the detector Before opening check the name of the model and the serial number After opening check the colour of the hydrating bags packed in the detector casing red in the event of humidity Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 3 B 10 Precautions and unpacking Handling the leak ASM 180 compact version ASM 181 console version detector with a hoist and slings Two lifting rings are supplied Four lifting rings are supplied with the leak detector with the leak detector Plugs are also supplied to They must be located on the replace the rings during upper part of the leak normal use of leak detector detector frame after having removed the work surface of the leak detector fixed by one screw on each side A A ALEA m SLE SRA E O T ea EA 1 Work surfa
23. When the external autocalibration is completed the test cycle can be interrupted and the external leak removed The detector is ready for use If the internal calibrated leak will be removed for a long period of time it is recommended to replace it with a DN 16 blank off to prevent dust from entering the lines Note To return to the internal autocalibration simply set Switch S1 on the monitor board to OFF and enter the value of the interna calibrated leak at 20 C on the thumbwheels S2 S3 and S4 see E 40 4 4 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 04 September 97 Remove the membrane pump from the detector Open the intake chamber MD4E membrane pump maintenance The frequency of the preventive maintenance tasks is listed in section D 10 Components P N Membrane kit eeee0 062968 A membrane kit is included in the maintenance kit supplied with the detector see F 10 Tools required 10 P317 and 20mm 21 thin spanner Phillips screwdriver M 5mm allen wrench 22 Switch off the detector and disconnect from the main power F Open the rear cover Disconnect the membrane pump inlet port Disconnect the power supply cable D Unfasten the 3 nuts 2 fixing feet and 1 angle bracket which secure it on the frame and remove the pump oO Unfasten the connectors 1 between the pump
24. a vacuum vessel Its characteristics allow it to be evacuated SPRAY method SNIFFING method BOMBING method Leak rate measurement from Minimum detectable leak of Leak not located 10 19 to 10 mbar and 106 mbar and possibility of possibility of locating the leak locating the leak Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 4 Spray method Response time A 40 Testing methods This involves removing air from the test part connecting it to the analyzer and then spraying helium over the outer surface Global test Spray test The part is placed under a Areas liable to leak are cover into which helium is sprayed with helium injected The leak cannot be The leak can be located located The detector measures the flow of helium penetrating the part When spraying starts the leak signal is not displayed instan taneously on the analyzer there is a response time which depends on the volume V being tested and the helium pumping speed S of the system at the opening of the part according to the following relation T za T in seconds V in litres S in s T is the time required for the signal to reach 63 of the final value 2 4 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 03 May 97 Sniffer method Testing methods A 40 The test part is pressurized with helium The detector
25. can still be used After an initial autocalibration fault signal it is recommended fo run a second autocalibration A repeated calibration fault is an indication that the cell is polluted and requires maintenance An internal calibrated leak can deliver helium for years In order to guarantee the reliability of measurements in most applications ALCATEL recommends recalibration of the internal calibrated leak at least every two years The value of the leak decreases over time according to a ratio indicated on the internal calibrated leak label and on the calibration certificate e g 2 every year The calibrated leak is recalibrated in approved centers using reference leaks see E 40 for the internal calibrated leak replacement procedure Alcatel High Vacuum Technology ASM 180 TD TD ASM 181 TD User s Manual 5 5 Edition 03 May 97 C 40 Switching off the detector The unit can be switched off at any time by setting the circuit breaker switch to O To keep the connection lines clean it is recommended to leave them under vacuum when the detector is shut off To maintain vacuum the inlet vent key must be off Alcatel High Vacuum Technology ASM 180 TD TD ASM 181 TD User s Manual 1 Edition 03 May 97 Contents C 50 Alphanumeric Control and Display Panel ACDP operation The present instructions only apply to units which are equipped with the ACDP option Purp
26. date Tl hh mn ss lt CR gt Adjustment of time STB xx lt CR gt Timer for CP 20 stand by mode reduced rotational speed O1 to 60 min default value is O1 min 4 8 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 04 September 97 B 30 Controlling the detector with a micro computer RS 232 Software Remote mode The detector sends back the requested data ASCII Code O6 ctrl F A lt Ck gt R lt CrR gt The detector sends back its status in code form Detecto Detecto r not in Cycle r in roughing phase or he filament is off T 1 0E 7 lt Ck gt Detecto back the measured helium signal TG 1 0E 7 lt CR gt Detecto r in FL test mode it sends r in GL test mode it sends Printer mode The detector sends tickets Residual helium signal when cycle is started Elapsed time for GL crossover a gt Elapsed time for FL crossover Helium signal before switching to FL mode Cycle duration Helium signal 7 at the end of the cycle Test ticket back the measured helium signal Sniffing test ticket MANUAL CYCLE C Elapsed time H M S S Signal mbar 1 s CYCLE START 13 NOV 1996 10 48 13 C 00 00 00 S 4 4E 10 S 1 5E 09 ALCATEL ASM180 series LEAK RATE 1 5E 9 UNITS mbar 1 s 13 NOV 1996 10 48 25 Autozero ticket ALCATEL ASM180 series ELECTRICAL Z
27. detected Place a voltmeter between erminal TP7 ground and erminal TP3 acceleration voltage Voltmeter rating 400 V Using a screwdriver adjust he potentiometer R31 to adjust the voltage as a function of the tracer gas used Right side of the supervisor board Tracer gas Mass Acceleration voltage Helium 4 4 150 2V Helium 3 3 198 2V Hydrogen 2 290 2V Set the circuit breaker switch to O For mass 2 or 3 detection autocalibration is performed only with an external calibrated leak refer to section E 50 Autocalibration with external calibrated leak For helium 4 detection standard to perform autocalibration with a calibrated leak inside the detector it is necessary to proceed as for a newly installed leak refer to section E 40 Replacement recalibration of the detector internal calibrated leak to perform autocalibration with an external calibrated leak refer to section E 50 Autocalibration with external calibrated leak Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 3 3 Edition 03 May 97 Contents Purpose of the option C70 Use of the I gas line option For ASM 180 TD and ASM 180 TD A Purpose of the I option 008 page 1 Operating principle 000 page 2 Choice of carrier gas e cece eeees page 4 Installation preparation 0e00 page 5 Installation connection 0eee
28. gas line f a mass flow controller 7 is used the maximum carrier gas flow can be defined quickly before connection to the installation by connecting the injection system directly to the detector The maximum flow is of the order of 40 to 60 SCCM or 0 6 to 1 atm cm s for the ASM 180TDI Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 9 11 Calibrate the installation Test the installation Stop the test Deactivate the gas line test mode Switch off the detector q 4 C 70 Use of the I gas line option For ASM 180 TD and ASM 180 TD Spray the reference leak 5 for a defined period e g 5 seconds Note the time required to obtain a signal on the detector any leak on the gas line will give a response lt this reference time the ratio read on the detector Reference leak value Helium signal value this ratio depends on the detector and the carrier gas flow Value between 10 and 20 Spray the various test points and according to the reference time defined above wait fo go to the next point It is recommended to start on the detector side and to test progressively by moving further away increasing response times Close the carrier gas injection Stop the test cycle by pressing rei Deactivate the gas line test mode by pressing the yellow button Set the circuit breaker switch to O
29. l s Hybrid turbomolecular pump air 130 l s Measurement range 2 1011 to 10 mbar s Electronic response time lt O 1s 8 decade log recording output 1 V dec Setpoint setting Fine leak 10 11 to 102 mbar s Setpoint setting Gross leak 108 to 10 mbarl s Inlet pressure display 103 to 103 mbar Triode pressure display Spectro 10 to 103 mbar Emission current display 0 2 to 2 mA Cell sensitivity 3 104 A mbar He pumping speed at detector inlet port AA l s Air pumping speed at spectrometer 110 I s He pumping speed at spectrometer 30 l s TMP exhaust pressure safety limit 6 mbar Startup time 3 min Cycle time inlet port blanked off GL FL mode Power voltage 100 115 200 220 230 240 V 2 As Power frequency 50 60 Hz single phase Power consumption 1 2 kVA 5 kVA 1 6 kVA Ambient operating temperature 10 to 40 C Weight 73 kg 160lb kg 210lb 1155 kg 340lb Noise level at 1m alarm not operational 54 dB 65 dB 65 dB Inert gas purge absolute pressure 1 40 1 bar abs No Purge flow rate 1 102 mbarl s No Purge Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 1 Edition 04 September 97 Dimensions mm inch A 90 ASM 180 TD ei 3 8 Inlet port 210 DN 40 8 3 220 8 7 A hid H
30. required between two immediate printings 10 27 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 03 May 97 C 50 Alphanumeric Control and Display Panel ACDP operation Sniffer ticket FILE 1 lt x Selected part CUSTOMIZED REF Customer reference MANUAL CYCLE C Elapsed time sec P Pressure mbar S Signal mbar s Measurement unit ppm or mbar s CYCLE START 3 FEB 95 8 53 39 C 0s P 1 0E 03 S 1 0E 12 C 14s Intermediate measurement P 1 0E 03 gt requested by keeeping W and S 1 OE 03 wy d It in LDS test pressed simultaneously C z OE 03 lt Intermediate measurement with m test long distance sniffer blocked STOP CYCLE S 9 5E 02 in LDS test Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 11 27 C 50 Alphanumeric Control and Display Panel ACDP operation Working in automatic mode Vacuum test Start a test C If SERIAL Nr and ticket test and printer are selected SERIAL v 23 Basic display FILE 1 READY AUTO CYCLE AUTO ROUGHING o AUTO FL 1 8E 10 TS If Nr TEST is selected Display of the test result Indicates alternately Previous test result see page 13 Auto cycle standby message Next test No Selected part reference Roughing time count down Test time count down
31. service MD4E or CP20 The dry roughing pump is defective NO Contact customer service 2 2 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 03 May 97 D 120 I O interface problem SYMPTOM CAUSE REMEDY No 24V on the I O I O interface board fuse Change the fuse jumper plug defective see E 80 pins 24 25 NO ST1 and ST2 straps incorrectly positionned on the I O interface board regarding 24V function Set up correct configuration see E 80 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 1 Edition 04 September 97 Chapter E User s Manual ASM 180 TD TD ASM 181 TD Maintenance sheets Analyzer cell maintenance 000 E 10 Greasing the hybrid turbomolecular pump E 20 m Greasing the molecular drag pump E 30 Replacement Recalibration of the detector internal calibrated leak 2 2 2 ee E 40 Autocalibration of the detector with an external CGUIDIGIE JEGK 94 os iedanewsn dees hs oa dee E 50 m MD4E membrane pump maintenance E 60 m CP20 Partial maintenance E 70 I O interface board fuse replacement E 80 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 1 Edition 03 May 97 Special precautions E 10 Analyzer cell maintenance The frequen
32. the carrier gas must have a helium content which is less than a few ppb 10 If 0 999 999 999 concentration nitrogen is considered too expensive nitrogen obtained from a tank or a source of liquid nitrogen can be used Any gas free of helium can be used as a carrier gas e g Argon However for safety reasons the method is not applicable to process gases which are toxic reactive explosive or flammable In addition the detector is not designed to pump chemically reactive gases 4 11 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 03 May 97 Installation preparation Equipment required in addition to the detector Flexible connection components 3 Helium spray equipment 10 Carrier gas source 8 A carrier gas flow adjustment device 7 A reference leak 5 C 70 Use of the I gas line option For ASM 180 TD and ASM 180 TD E g flexible stainless steel tubes of a diameter not greater thanlO mm so as not to increase the response time and connection accessories compatible with the installation under test E g helium cylinder with pressure relief valve tube and spray gun E g helium tree nitrogen cylinder and pressure relief valve This source must be compatible with the cleanliness or purity requirements within the installation at the time of the test Th
33. the dust Prepare special metal seal for the analyzer cell using the seal former see F 10 item 8 or use an elastomer seal optional see F 110 Wg The ends of the seal must only Fibs cross once no twist Check that the ends cross near one of the six screws holes one end on either side of hole Place the prepared seal on a flat surface protected from contamination 2 8 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 03 May 97 Cleaning the base of the cell E 10 Analyzer cell maintenance Unscrew the 6 screws and carefully extract the flange from the body pull directly upwards There may be traces of metallic deposits in the internal duct in the cell base 2 to the right of the filament in which case clean off with abrasive paper 180 grade Vacuum out any residue and complete the cleaning with alcohol Clean the special metal seal channel with alcohol 1 Block the opening immediately with the seal former Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 3 8 Removing the filament Removing the electron collector Replacing the filament and the electron collector E 10 Analyzer cell maintenance Remove the filament by unscrewing the retaining screw and loosening the 2 connection screws Remove the retaining screws Install the electron collector 1 on the ionization
34. 0 3 77E 08 3 77E 06 4 75 5 61E 06 5 61E 04 0 50 1 22E 10 1 22E 08 2 65 4 25E 08 4 25E 06 4 80 6 29E 06 6 29E 04 0 55 1 49E 10 1 49E 08 2 70 4 78E 08 4 78E 06 4 85 7 06E 06 7 O6E O4 0 60 1 80E 10 1 80E 08 2 75 5 38E 08 5 38E 06 4 90 7 93E 06 7 93E 04 0 65 2 16E 10 2 16E 08 2 80 6 05E 08 6 05E 06 4 95 8 89E 06 8 89E 04 0 70 2 57E 10 2 57E 08 2 85 6 81E 08 6 81E 06 5 00 9 98E 06 9 98E 04 0 75 3 04E 10 3 04E 08 2 90 7 66E 08 7 66E 06 5 10 1 26E 05 1 26E 03 0 80 3 57E 10 3 57E 08 2 95 8 61E 08 8 61E 06 5 20 1 58E 05 1 58E 03 0 85 4 18E 10 4 18E 08 3 00 9 68E 08 9 68E 06 5 30 1 99E 05 1 99E 03 0 90 4 88E 10 4 88E 08 3 05 1 09E 07 1 09E 05 5 40 2 51E 05 2 51E 03 0 95 5 67E 10 5 67E 08 3 10 1 22E 07 1 22E 05 5 50 3 16E 05 3 16E 03 1 00 6 56E 10 6 56E 08 3 15 1 38E 07 1 38E 05 5 60 3 98E 05 3 98E 03 1 05 7 58E 10 7 58E 08 3 20 1 55E 07 1 55E 05 5 70 5 01E 05 5 01E 03 1 10 8 74E 10 8 74E 08 3 25 1 74E 07 1 74E 05 5 80 6 31E 05 6 3 1E 03 1 15 _1 00E 09 1 00E 07 3 30 1 95E 07 1 95E 05 5 90 7 94E 05 7 IAE 03 1 20 1 15E 09 1 15E 07 3 35 2 19E 07 2 19E 05 6 00 1 00E 04 1 00E 02 1 25 1 32E 09 1 32E 07 3 40 2 46E 07 2 46E 05 6 10 1 26E 04 1 26E 02 1 30 1 51E 09 1 51E 07 3 45 2 77E 07 2 77E 05 6 20 1 58E 04 1 58E 02 1 35 1 73E 09 1 73E 07 3 50 3 11E 07 3 11E 05 6 30 2 00E 04 2 00E 02 1 40 1 97E 09 1 97E 07 3 55 3 49E 07 3 49E 05 6 40 2 51E 04 2 51E 02 1
35. 18 9 19 HF IS ars ope X i Se Xa oes Ls 12 11 1 Detector inlet port 2 Roughing molecular drag pump 2 Inlet vent valve MDP 3 Vent filter connector 3 Detection valve 4 Roughing valve 4 Calibrated leak module 5 Inlet pressure gauge PI3C 5 Connector for long distance 7 Bypass valve sniffer 8 Exhaust pressure gauge PI1 6 Sniffer valve 9 Exhaust valve 17 Hybrid turbomolecular pump 10 Roughing pump vent valve PTM 5154 11 Roughing membrane pump 8 Analyzer cell MD4E 9 Connector for inert gas purge Pumping capacities 4 m h roughing membrane pump MD4E 10 l s molecular drag pump MDP Helium pumpi Test capacities Short test cyc ng speed at inlet port 4 4 s e Quick response time Autocalibration with integrated calibrated leak Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 4 A 20 ASM 180 TD Detector operating principle Note Only operational parts are represented Operation in vacuum test mode 3 stages Primary roughing Molecular roughing eH let 12 11 12 11 Note Valve 7 closes at about 6 mbar 2 Gross leak test mode 3 Fine leak test mode Gl FL 2 4 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual E
36. 450 17 7 430 ASM 16 9 180 td a 1 1 wW wW Ww ww 410 161 425 16 7 600 23 6 96 A0 1 6 m 3 8 ASM 180 TD a 210 Inlet port 8 3 DN 40 f 260 10 2 T I r y 5 O 450 455 430 17 7 17 9 ASM 16 9 h 180 td 1 1 il w wW p 455 17 9 R p 465 18 3 600 23 6 J Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 2 A 90 Dimensions mm inch ASM 181 TD 200 7 79 3 5m 11 feet eae standard lenght 50 2 0 300 100 Inlet port TEETE aa A EEE 1 2 680 A A 26 8 7A 50 2 0 885 34 8 TA Ny 0 80mm diam casters without brakes 2 2 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 04 September 97 Chapter B User s Manual ASM 180 TD TD ASM 181 TD Installation Precautions and unpacking B 10 L Controlling the detector with the I O interface B 20 L Controlling the detector with a microcomputer R8232 053435 24564 He ows B 30 Connecting an external printer 2 2 0 2 B 40 Connecting a neutral gas purge ASM 180TD only B 50 L Connecting the leak detector to the installation via the hardware interlace B 60 L Before starting up the detector B 70 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 1 Edition 04 September 97
37. 5 083446 4 4 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 04 September 97 Components summary F 120 Description P N Maintenance kit delivered Maintenance kit ASM 180 TD 090201 with the detector F10 Maintenance kit ASM 180 TD 104434 Monitoring and display F20 Ball and socket joint 090172 Clamping Oring 082116 Elapsed time counter 037861 Holder 090211 Remote control unit 101298 Remote control unit 3 5 m 101496 or remote control unit 7 m 104286 or remote control unit 25 m 104287 Power and electrical supply Breaker switch F30 100 115V 8A 180TD 101779 200 220 240V 4A 180TD 101781 100 115V 16A 180TD 181TD 101780 200 220 240V 8A 180TD 181TD 101779 Fuse 100 115V T3 15A 1801TD TD 181TD4 060855 200 220 115V T6 3A 1801D TD 181TD 060860 Jumper plug Sub D 25 pins 101824 Automatism and 8 Q loud speaker 060097 electronic circuits F40 Control panel assembly 101299 Fan 101094 Fan protective grid 056067 POO90 CD2 MDP 5011 power supply board 072402 POO90 CD2 TMP 5154 power supply board 072402 PO191E1 Distribution board ASM 181TD 100436 PO195 1 0 Interface board 101404 PO316E1 Supervisor board PO318 Booster board 104153 Contact customer service Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 4 Measurement F50 Pumping F60 Components summary Descript
38. 50 Alphanumeric Control and Display Panel ACDP operation MANUAL Mode FIL ON 1 0E 10 READY MAN CYCLE CYCLE MODE MANUAL lt i FILE 1 CHANGE NO lt m MANU REJECT SETP FIXED PRINTER NO LDS TEST UNITS mbar l s FIL ON 1 0E 10 READY MAN CYCLE Automatic mode overleaf Basic display va A ate used to change operating mode A Yes is used to access to part modification basic menu see page 15 The manual reject setpoint monitors the switching of the Red light indicator of ACDP panel in relation to the fixed reject setpoint only the emission of an audio signal when controled by ADCP The closing of the reject setpoint contact on I O interface see B 20 when the helium signal value is greater than the setpoint for further details on Fixed Floating see page 22 Enable or disable the printer interface in Manual mode Selection of the measurement unit in sniffer mode Return to the basic display Manual mode Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 5 27 C 50 Alphanumeric Control and Display Panel ACDP operation Main menu continued Manual mode AUTOMATIC mode overleaf FILE 1 Basic display READY AUTO CYCLE aE y Auto A CYCLE MODE A Y are used to change Manual Y AUTO operating mode Fal Yes A FILE 1 A Yes is used to access the No W CHANGE NO bas
39. 5011 power supply board 072402 10 PO191E1 Distribution board ASM 181 TD 100436 Contact customer service Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 1 Edition 03 May 97 Measurement Item Description Complete VHS analyser cell with lead seal without magnet Electron multiplier amplifier VHS Filament Electron collector set of 5 lead gasket 10 meter Accessories kit analyzer cell Co A3 F 50 P N 072493 072494 053146 068842 083478 090294 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 2 F 50 Measurement ltem Description P N 7 Calibrated leak Fe1407 with thermal probe 101302 8 Aluminium PI gauge 795706 9 Spare filament for PI set of 5 068835 10 Aluminium PI3C gauge 786434 2 2 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 04 September 97 Pumping MD4E membrane pump ASM 180 TD Item Description 1 MD4AE membrane pump 100 V 50 60 HZ 220 240 V 50 60 HZ 120 V 50 60 HZ 200 V 50 60 HZ 2 Seals kit for MD4E membrane pump 3 a MDAE Shock absorber per unit b MD4E Shock absorber per unit F 60 P N 062980 062981 062982 062984 062968 101554 101555 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 3 Pumping CP20 pump ASM 180 TD
40. 5AT Voltage 100 115 V Basra 6 30AT Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 2 Edition 04 September 97 B 60 1 Connect the remote control unit 2 Connect the 1 O interface see B 20 3 Connect the RS232 see B 30 4 Connect an external printer see B 40 5 Connect the LDS probe 6 Connect to atmospheric pressure 7 Connect accessories Connect the unit to the main power Connecting the leak detector to the installation via the hardware interface 4 The 1 O interface connector should never be connected or disconnected with the unit on If the detector is not controlled by the I O interface the jumper plug must be connected It the detector is controlled by the I O interface install the interface cable to the Sub D 25 pin connector on the detector f the detector is to be connected to a micro computer connect the cabled RS232 ACDP option required see A 60 Connect the printer output using an RS232 cable quick connector When a neutral gas is used the filter is unscrewed and replaced by the connection to the selected gas supply source 1 4 BSP connector Pressure 1 02 atm absolute Additional plugs 3 power plugs allow to connect accessories such as recorders gauges Maximum current 4A Specific 4A Fuses provided Check that the voltage marked on the unit identification plate corr
41. 7 Cell principle Magnetic deflection spectrometry Leak flow rate Vacuum operation Separation of He ions from noise Total pressure A 30 Analyzer cell operating principle The mass spectrometry analyzer cell is used for helium partial pressure measurements The molecules of the gas being analyzed are bombarded by an electron beam from a heated tungsten filament 1 in an ionization chamber 3 A large proportion of the molecules are transformed into ions These ionized particles are accelerated by an electrical field the acceleration voltage A magnetic field deflects the ion beam by a radius propotional to the mass of the ions The acceleration voltage directs the Helium ions to the target at the entrance of an amplifier an electron multiplier based system developed and patented by ALCATEL The stream of Helium ions is proportional to the partial pressure of helium in the installation and its measurement is used to find the value of the flow rate of the detected leak lt is essential for the total pressure in the analyzer cell to be less than 104 mbar so that the paths of the electrons and ions are not disturbed by residual molecules In order to separate the helium ions from the noise due to dispersed ions a braking electrode 6 placed in front of the target eliminates secondary low energy ions The top of the cell contains an auxiliary electrode which collects ions that have a higher mass than that
42. ASM 180 TD not equiped j Contact customer service Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 2 Edition 04 September 97 F 70 Valves ASM 180 TD ASM 181 TD ltem on o oO a K 5 6 7 8 9 10 1 2 YS Valve block A Description Roughing valve Detection valve By pass valve VAT NW 25 valve operator VAT 180 260V valve coil Seal kit for VAT valve not pictured Grease tube 10g for VAT valve not pictured Bacosol 24V DC 23W valve exhaust Calibration block Bacosol 24V DC 23W valve air inlet Minisol valve 2 2 NF LDS without coil Coil for 24V DC minisol valve Non injection TD plug Valve block ASM 180 TD not equiped Vickers silencer Contact customer service P N 104655 100973 104655 038101 067040 104426 075990 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 2 2 LDS valve connected to the TMP rear Edition 04 September 97 F 80 Pipes z5 without seal without seal ON a o gt ie Item Description P N 1 Inlet filter NVV 25 without seal 072857 2 Diaphragm NW 25 without seal j 3 Silencer 1 4 101552 4 Inlet adaptor NW 25 NVWV40 180 TD i Inlet adaptor NVV 25 NW40 180 TD i 5 Rilsan tube 6 i 6 PVC tube 4 x 2 7 Flexible tube NW 16 lenght 250 mm 068369 8 Stainless steel tube NVV 25 NW 40 101539 9 CP 20 pipe 10 CP 20 pip
43. ASM 180 TD TD ASM 181 TD Helium Leak Detector AA a p ni ae Sa Qi I yj Tela Ta a E 3 7 i i oe B E ww J gt EY wai i T FOA SALES AND SERVICE PLEASE CALL PTO SALES woot 868 392 0800 Bout 626 394 2154 wwe_pilbaaldd com SALES y View our inventory ALCATEL Edition 04 July 97 A very wide range of helium leak detectors You have just purchased an ALCATEL leak detector This product is part of a very wide range of products resulting from 30 years of experience The applications of helium leak testing are extremely varied ranging from high tech installation maintenance to high speed testing of industrial products Each product of the ALCATEL Integrable or turnkey solutions for detector range is designed automated leak testing to meet the specific needs of each application unit portability high sensitivity pumping capacity pumping type automation and integration in an industrial process Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 1 France Alcatel Vacuum Technology France 98 avenue de Brogny BP 2069 74009 ANNECY Cedex Tel 33 4 50 65 77 77 Fax 33 4 50 65 77 89 Germany Alcatel HOCHVAKUUMTECHNIK GmbH Am Kreuzeck 10 D 97877 WERTHEIM Tel 49 9342 9610 0 Fax 49 9342 9610 30 ltaly Alcatel Vacuum Systems S p A Via Trento 30 20059 VIMERCATE MI Tel
44. At the end of an automatic test ACDP Display panel Detector ACCEPTED 2 0E 10 RELAI 225 ACD READY AUTO CYCLE TEST 3 AUTO MODE FL signal 1 1E 69 NEXT NUMBER 4 RELAI 22S ACD 4 4 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 03 May 97 B 50 Connecting a neutral gas purge ASM 180 TD only Use Used to accelerate the cleanup of the helium background noise after detecting a significant leak Make high sensitivity testing easier due to the reduction and stabilization of the helium background noise Neutral gas The neutral gas supplied must have a helium concentration supply less than or equal to 1 ppm Supply pressure 1 4 0 1 bar absolute 20 psia Connection A quick connector is located to the left at the rear of the detector near the LDS connector The corresponding male connector to be fitted on the gas inlet tube is supplied in a plastic bag with the detector Note The neutral gas purge connector is different than the inlet hole connector The latter can also be connected to a neutral gas source to purge the inlet and anything connected fo it at the end of a cycle The supply pressure of the gas for the inlet vent must be atmospheric pressure 1 01 atm absolute 14 psia Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 1 B 60 Connecting the leak detecto
45. D 60 Spectro fault CAUSE REMEDY Filament key Press filament key in OFF position once to ON position NO Switch off the detector and connect the I O jumper plug Start up again see B 70 I O jumper plug This is not connected to the rear of the detector NO Contact customer service Filament On the cell JAEGER connector open circuit between pins 1 and 5 Replace the filament see E 10 NO Eliminate the short circuit check conductivity between pins 1 and 5 of the Jaeger connector see E 10 Short circuit in the cell On the cell JAEGER connector conductivity between 6 and the other pins of the connector NO Contact customer service Spectro pressure gt 104 mbar NO Wait for a few seconds until the vacuum improves in the cell and reset the filament Eliminate the leak check particulary the seal of the analyzer cell Analyzer cell Gross leak visible with the alcohol test NO WWWWWContact customer service Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 1 Edition 03 May 97 D 70 Inlet pressure problem SYMPTOM CAUSE REMEDY No inlet PI3 gauge Replace the PI3 inlet pressure is defective pressure gauge display NO Contact customer service No pressure drop at Replace the fuse start of a cycle fuse 250V 5x20 T1 25A see F 40
46. ERO O K 13 NOV 1996 10 47 59 MANUAL CYCLE C Elapsed time H M S S Signal mbar 1 s LDS START 13 NOV 1996 10 48 29 C 00 00 00 S 3 6E 10 STOP LDS C 00 00 13 2 3E 05 ALCATEL ASM180 series LDS LEAK RATE 1 5E 9 UNITS mbar 1 s 13 NOV 1996 10 48 42 Autocalibration ticket ALCATEL ASM180 series CALIBRATION COMPLETED Calibrated leak value 7 7E 08 mbar 1 s 13 NOV 1996 10 47 34 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 5 8 Printer mode continued B 30 Controlling the detector with a micro computer RS 232 Default ticket DEFAULT CODE 200 13 NOV 1996 10 49 06 List of defaults 3 digit code FAMILY code 1 DEFAULT code 2 ERER INIT DEFEC IOE EA EE EANAN ke RER EE 011 RAM test defect 012 Real time clock defect 013 EPROM Checksum defect EROPA ONVMANDIDE E OLE 100 Time Out Expired 101 Unknown command 102 Uncomplete command line 103 Invalid character BD es es SPECTRO DEFOD A as Ba PR Re Ta oh Eta at 200 Spectro parameter Unit 201 Incompatible reference leak value 202 Background level too high 204 Helium Peak Adjustment defect 205 Emission current adjustment limit exceeded 206 Calibration Interupted 208 Electronic Zero Init 209 Filament emission defect 210 Triode spectro pressure safety activated 211 Amplifier Zero adjustment Init FF FF
47. FF F F F 25 SPUN PISS YG TE MDE ee he SEN ios 301 Exhaust pressure gt 10 mbar 302 TMP in acceleration mode 303 TMP defect 304 LDS flow too high 305 LDS probe clugged FF F 6 8 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 04 September 97 B 30 Controlling the detector with a micro computer RS 232 Example of Terminal operation under Window 3 11 communication with a PC s soon as the connections are done and the Terminal function opened under Window the main two parameters to be configured are Emulation and communication Terminal Untitled ia Settings Phone Number Terminal Emulation Terminal Preferences Function Keys Terminal eater do Binary Transfers Communications Modem Commands Printer Echo Timer Mode Show Function Keys TTY Generic DEC VT 100 ANSI em O10 300 8600 1200 2400 4800 9600 19200 Data Bits Stop Bits os Os O7 8 er Ors O2 Parity Flow Control Connector None Xon Xoff a Odd Hardware oe Even None Mark Space o Parity Check o Carrier Detect Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 7 8 B 30 Controlling the detector with a micro computer RS 232 Function keys may be programmed to allow to send commands to the detector without the use of the
48. Ge Ga SX Ca F 10 Preventive maintenance components delivered with the detector Description Filament assembly Tube including Collector Stainless steel screw CS M2x4 Screw CHC M4x80 Allen wrench 5 mm PIT spare gauge Screwdriver Tube including Aluminium gasket Lead gasket L 1 m Fuse slow blow 5x20 0 5A Fuse slow blow DITD 5x20 1 25A Fuse slow blow DITD 5x20 3 15A Fuse slow blow DITD 5x20 6 3A Fuse slow blow DITD 5x20 10A Block Seal former Straight FACOM 5 amp 6 mm Allen keys Membrane kit Quantity eee eee i A M M Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 2 Maintenance kit delivered with the detector ASM 180 TD ASM 180 TD i ASM 181 TD P N 104434 F 10 Preventive maintenance components delivered with the detector ltem N S Ge Ga SX Ca Description Filament assembly Tube including Collector Stainless steel screw CS M2x4 Screw CHC M4x80 Allen wrench 5 mm PIT spare gauge Screwdriver Tube including Aluminium gasket lead gasket L 1 m Fuse slow blow 5x20 0 5A Fuse slow blow 5x20 1 25A Fuse slow blow 5x20 3 15A Fuse slow blow 5x20 5x20 6 3A Fuse slow blow 5x20 5x20 10A ock Seal former raight FACOM 5 amp 6 mm Allen keys uv W Quantity Root ND OND eee i i OO Ma 2 2 Alcatel Vacuum Technology France ASM 180
49. M ee No pressure drop at start of cycle P mbar display 0 Inlet pressure Mets suec e5 55i02 4250n 80654064 R084 45 No change to FL mode P lt 2x 102 mbar oaoa naana aaa ee ISW SERVI ota nee e Soa eee eee eee Bedok eae wd High background NOSE 05 s Bes eas os NIG ANS Sw See SRW RS CP 20 pump has stopped during cycle ASM 180 TD 181 TD Problems during LDS test cycle The E LDS indicator light does not come on 2 ee The LDS helium signal lt 5x106 mbarl s 000 ee Spectro indicator lights off filament off oaaao aaa 000000000 Autocalibration problems Electrical zero check failed filament off 0 000 00 000000 Autocalibration failed filament on 0 0 000 000 0c ee eee Problems on I O interface board No 24V on ie jumper OG so ee eae eo Ree ASSESS Sede eee SEES D 20 D 80 D 70 D 70 D 110 D 110 D 110 D 30 D 100 D 100 D 100 D 90 D 90 D 120 2 2 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 04 September 97 D 30 Problem with the roughing pump SYMPTOM CAUSE Circuit breaker switch e Difference in voltage between input and output e impossible to keep in position 1 No noise from the MD4E ASM 180 TD NO Pump motor does not rotate NO Problem on CP20 electrical module No noise from the CP20 ASM 180 TD ASM 181 TD
50. PASS available in the MANUAL mode only allows to reduce the number of menus displayed because it has no automatic test parameter 16 alphanumerical characters introduced by connecting the detector to a PC microcomputer allow to customize the 10 part references see page 18 Total automatic cycle time Start End ri Roughing Gross leak test Fine leak H lium signal measurement test Reject setpoint and warning setpoint comparaison l Test ticket printout ae Test count Roughing time p Test time A Checking of backgound level If the Helium signal when not in a cycle and at the time the cycle start is activated is greater than the background setpoint the test cycle is cancelled and an empty cycle is requested to clean the detector optional selected with a switch inside the detector see page 24 The background setpoint can be useful for high sensitivity tests reject setpoint in the 10 to 1010 mbarl s range If the detector has not changed to test mode after the roughing time the cycle is stopped and the part rejected for Gross Leak Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 15 27 Test time Test ticket Test reference Helium signal measurement Reject setpoint Warning setpoint C 50 Alphanumeric Control and Display Panel ACDP operation When the detector is changed to test mode the test time is counte
51. SS fama a ean ka eee ed Bees F 80 Connections and seals 05 F 90 Oe oer oft oe FE Reseed See Y F 100 Options and accessories 2 2 eee F 110 Components summary 00000 F 120 Appendix View of the operator interface G 10 Recording curve 2 0 0 ee eee G 20 MANUAL REFERENCE 104435 EDITION 04 September 97 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 3 3 Edition 04 September 97 Chapter A User s Manual ASM 180 TD TD ASM 181 TD Introduction m The ASM 180 series 2 2 A 10 ASM 180 TD Detector operating principle A 20 m ASM 180 TD ASM 181 TD Detector operating principle 0 A 21 m Analyzer cell operating principle A 30 Hl Testing methods 00000 0005 A 40 m Operator interface 0 ee A 50 we CG eG Lae ee a ees eee s A 60 ACCESSOS Heda esa pue paee trga Ee ga A 70 F Technical characteristics ooon A 80 L Dimensions n naaa aaa A 90 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 1 Edition 04 September 97 A 10 The ASM 180 series The main characteristics of this series of products are very high sensitivity 2x1011 mbarl s a range of pumping capacities to meet different requirements sturdy design adapted to severe industrial environments usertriendly The ASM 180 series includes dif
52. TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 04 September 97 Monitoring and display F 20 Item Description P N 1 Remote control unit 101299 2 Elapsed time counter 037861 3 Remote control unit 3 5 m 101496 or remote control unit Z m 104286 or remote control unit 25 m 104287 4 Holder 090211 Ball ket joint 172 and ASM 180 TD jae ea Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 F 30 Power and electrical supply AOO DONITE CT DANI TT Item Description 1 Jumper plug Sub D 25 pins 2 Breaker switch 100 115V 8A 180TD 200 220 240V 4A 180TD 100 115V 16A 180TD 181TD 200 220 240V 8A 180TD 181TD eee Fuse 100 115V T6 3A 180TD TD 1817D 200 220 115V T3 15A 180TD TD 1817D ASM 180 TD TD ASM 181 TD aT P N 101824 101779 101781 101780 101779 060855 060860 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 1 Edition 04 September 97 F 40 Automatism and electronic circuits A AAT ANNAN pi AAA AUN ANION AMAA MRA VAAIA Description P N 1 PO316E1 Supervisor board i 2 Control panel assembly 101299 3 Fan 101094 4 Fan protective grid 5 8 Q loud speaker 060097 6 P0090 CD2 TMP5154 power supply board 072402 7 P0318 Booster board 104153 8 PO195 1 0 Interface board 101404 9 P0090 CD2 MDP
53. The purpose of this sheet is to show the key or the part of the display in action during use of the option Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 27 27 ACDP PANEL 1 LCD display 2 x 16 character lines 2 Yellow indicator light signalling the activation of the autocalibration process 3 Red indicator light part rejected 4 Green indicator light part accepted F key used to access the various functions Shift down key Y ee Shift up key A parameters C key cycle control Edition 03 May 97 C 60 Configuring the unit according to the gas to be detected The following instructions only apply to units which are equipped with the 3 Mass option and for a change of tracer gas Introduction The unit equipped with the 3 Mass option does not have to the unit ony external differences in relation to the standard unit The modifications are inside the unit analysis cell magnet and electronic supervisor board The functions are the same as the standard detector The tracer gases which Gas Atomic mass Alcatel does not supply a can be used calibrated internal leak in Helium 4 4 Helium 3 and Hydrogen Helium 3 3 The calibration is made Hydrogen 2 with an external calibrated leak Initialize the acceleration While the unit is switched off voltage as a function of P the front cover of the unit and tilt it forwards to access the mass to be detected the supervisor bo
54. alidated connect the external calibrated leak to the detector inlet port and run a test cycle with the key Wait a few minutes for the Helium signal to stabilize Adjust the displayed Helium signal value manually as a function of the external calibrated leak using the filament current fine adjustment potentiometer Remember to take into account the effect of the temperature on the value of the external calibrated leak in accordance with the information given on the label In the event of problems contact customer service Alcatel High Vacuum Technology ASM 180 TD TD ASM 181 TD User s Manual 3 5 4 Autocalibration procedure C 30 Detector autocalibration Throughout the autocalibration cycle the autocal key red indicator light flashes e Residual check e Autocalibration circuit roughing e Connection with the analyzer cell The Helium signal measurement is displayed e Comparison of the measured signal with the calibrated leak value after waiting a few seconds for the signal to stabilize If the difference is less than 10 the autocalibration is stopped and validated If the difference is greater than 10 the autocalibration is continued e Helium peak detection The automatic control system varies the P E S mina Mash w 5 5 a w 2 2 5 os A 2 w w w3 BES 5 5 2 0 acceleration voltage in the analyzer cell see A 30 This varies the path of the Helium ions until the m
55. ard Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 3 Configure the switch S1 Switch on the detector while resetting the autocalibration parameters C 60 Configuring the unit according to the gas to be detected If the detection can be performed on Helium 3 or Hydrogen set the switch 1 to ON external autoca libration If the detection must be performed on Helium 4 the switch 1 can if required be set to OFF autocalibration with an internal calibrated leak or ON external autocalibration Teee ees mm JOG O OFFEELION S1 ou Supervisor board Set the circuit breaker switch to I While the green cycle key indicator light is flashing first seconds of the commissioning cycle press the autocal key MEA If the cycle key indicator light stops flashing before you press the autocal key 4 switch off the unit and repeat the operation The red autocal key indicator light comes on 2 3 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 03 May 97 Initialize the acceleration voltage Switch off the detector Autocalibrate the detector C 60 Configuring the unit according to the gas to be
56. aximum Helium signal is obtained The Helium signal display oscillates during this stage e Sensitivity adjustment The filament current in the analyzer cell see A 30 is automatically adjusted so that the detector displays a correct leak value internal calibrated leak value corrected as a function of temperature Stages 6 and 7 can be performed several times if necessary until the correct display is obtained A 5 Alcatel High Vacuum Technology ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 03 May 97 5 In the event of a calibration fault Internal calibrated leak service life C 30 Detector autocalibration e End of cycle The sequence has been performed correctly the detector is calibrated in Fine Leak mode the green indicator light comes on If a fault is encountered the sequence is stopped and the red indicator light comes on Faults which stop the autocalibration sequence are zero impossible background signal too high sensitivity adjustment impossible voluntary stop by operator 1 If the current cycle is stopped voluntarily this is considered as a fault Following a fault the calibration parameters stored previously are restored The autocal key red indicator light remains continuously lit a defect ticket is emitted on the RS232 interface in printer mode see B 30 The cycle key green indicator light is lit The unit
57. bration and auto zerotickets Note During the detector startup process the RS 232 interface sends data regarding the EPROM edition LOOAO index Factory default configuration Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 3 8 Protocol selection Common protocol commands ASCII Code 05 09 26 B 30 Controlling the detector with a micro computer RS 232 The selection of the protocol is made from the micro computer with following commands L Local Hardware protocol R Remote Software protocol P Printer Printer protocol Language selection F French E English D German List the commands space Commands A Cell autocalibration start C Test cycle start under vacuum B LDS test cycle start S Test cycle stop vacuum or LDS U gt GL mode selection same as key m u gt GL mode selection cancelled V gt Air vent same as key gt No air vent T gt Manual adjustment of helium peak ON t Manual adjustment of helium peak OFF Q gt Manual adjustment of emission current ON q gt Manual adjustment of emission current OFF ctrl E Switches the filament on off ctrl I Returns to the factory default configuration values of emission current and helium peak calibration ctrl Z Returns to the default zero value helium signal Increase selected parameter Decrease selected parameter DA dd mm yy lt CR gt Adjustment of
58. bration with integrated calibrated leak Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 4 A 21 ASM 180 TD ASM 181 TD Detector operating principle Note Only operational parts are represented Operation in vacuum test mode 3 stages 1 Primary roughing 12 20 2q Gross leak test mode GL b Gross leak test mode FL 1 mbar lt Inlet Pressure lt 6 mbar Inlet Pressure lt 6 mbar 20 2 4 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 04 September 97 ASM 180 TD ASM 181 TD Detector operating principle Operation in 3 Fine leak test mode FL vacuum test mode 3 stages 18 l continued Operation in 18 sniffing mode LDS Sniffer probe A 21 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 3 4 A 21 ASM 180 TD ASM 181 TD Detector operating principle Operation in internal calibration mode M Roughing of calibrated 2 Calibration leak 3 Venting of calibrated leak The leak is returned to atmospheric pressure 4 4 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 03 May 9
59. calibration Autocalibration system C 30 Detector autocalibration Used to ensure that the detector analyzer cell detects the helium properly test of the Helium ion path so that they reach their target see A 30 Analyzer cell principle the Helium leak value displayed corresponds to the real value he detector uses an internal calibrated leak equipped with temperature dependent compensation system he value of the leak is approximately 1 x 107 mbar l s ectrovalves are used to connect the calibrated leak to the nalyzer cell The electrovalves are controlled and the two calibration parameters are set entirely automatically by pressing the button om 4 0 sl eck Autocal circuit layout So x 12 AET ee 20 ASM 180 TD ASM 180 TD ASM 181 TD Alcatel High Vacuum Technology ASM 180 TD TD ASM 181 TD User s Manual fi Running an autocalibration D At start up 2 During operation w a C 30 Detector autocalibration At detector startup as soon as the analyzer cell spectro is operational an autocalibration is performed automatically During operation the calibration is checked as follows Electrical zero check Switch off the filament indicator light off check that the helium signal on the fin
60. ce fixing screws 2 Location of the lifting rings In the event of any damage contact the shipper and if necessary notify ALCATEL 2 3 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 04 September 97 Storage Installation B 10 Precautions and unpacking For prolonged storage factors such as temperature humidity saline atmosphere etc may damage the detector elements In this case it may have operating problems Before starting up after storage for over six months it is recommended to change all the seals contact customer service The seal kits must be kept away from heat and light direct sunlight and ultraviolet light in order to prevent hardening of the elastomers The pertormances of the detector pumping speed accuracy and reliability depend on the ambient temperature the vacuum connections the frequency and quality of maintenance the helium calibration Position the unit so there is no possible risk of the unit falling or tilting Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 3 3 Edition 03 May 97 Connect the jumper plug if the I O interface is not used Prepare the connector wiring The controls inputs B 20 Controlling the detector with the I O interface This makes it possible to control the detector using a PLC In the absence of external c
61. chamber by partially tightening the two screws Install a new filament 2 by inserting the two wires into the connectors do not tighten the latter 4 8 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 03 May 97 Adjusting the components E 10 Analyzer cell maintenance The correct condition and adjustment of these components are determining factors in maintaining the specifications of the detector The insulator is perfectly seated on the two faces of the support The filament is centered in the electron collector the same number of turns on each side of the collector opening the X dimension in the front view diagram Continue to adjust the collector so that the filament is vertically centered in the collector opening and is parallel to the opening the Y dimension in the front view diagram Tighten the 2 collector retaining screws completely If necessary fine adjust the Front view parallelism between the axis of fi the filament and the axis of the TM aam aa ar collector oblong opening by pocannan Ey adjusting the squareness of the Ih ai collector Work on the area marked ss in the diagram Section AA T I Check that the filament is correctly aligned on the axis 4 indicated by the section AA in the diagram If necessary adjust one of the branches of the filaments two con
62. cy of preventive maintenance is listed in D 10 Components P N Filament cccccceccccccce 053146 Electron collector pack of 5 4 068842 Special seal wire 10 meter roll 083478 DOSS aeeaiei Maintenance kit see F 10 Disconnect the detector from the main power The VHS amplifier and the analyzer cell are very sensitive to any form of contamination and particularly to dust When assembling to avoid gettering due to dust or finger prints you are advised to work in a clean room on lent free paper with unpowdered vinyl gloves clean room gloves to dust each part with filtered dry air to block all the openings in the vacuum lines and the VHS preamplitier e Every time the VHS preamplifier and the analyzer cell are disassembled their special seals must be replaced Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 8 Dismantle the VHS amplifier Prepare the new special metal seal E 10 Analyzer cell maintenance e Disconnect electrically the VHS amplitier and the spectro cell 2 connectors e Remove VHS amplifier trom the cell by unscrewing the 2 cHc screws using the 6 allen wrench supplied in the maintenance case e Position carefully head down the VHS amplifier on a clean support dust free e t is advisable to keep the electron miltiplier of the VHS lying down during all the removal in order fo protect it from
63. d SERIAL Nr at the end of each test SERIAL Nr 3826 AUTO MODE FL PASS signal 7 2E 09 26JUNE 96 15 18 20 lt TEST Nr automatically incremented S FILE 2 SERIAL Nr 4217 AUTO MODE FAIL signal 3 2E 06 26JUNE 96 15 20 31 CUSTOMER reference TEST Nr 7 AUTO MODE FAIL T Roughing 26JUNE 96 16 12 05 C 50 Parameters ticket lt Selected part Customer refe rence Test ticket to be printed est reference Test ticket Accepted part Accepted part with warning level Rejected part Rejected part on account of roughing timer Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 03 May 97 Purpose of the part modification menu File Automatic test parameters Background setpoint Roughing time C 50 Alphanumeric Control and Display Panel ACDP operation This menu is used to set up the automatic test parameters for 10 part references memorized as file 1 to 10 select the part to be tested among the 10 memorized modity and check the date and time which are displayed and printed on the control tickets As indicated above 10 part references can be memorized with their associated automatic test parameters Access to these files is allowed whichever test mode of the detector AUTO or MANUAL has been selected An eleventh reference called FILE BY
64. d can o Switch be modified by modifying Boud 1 2 3 4 the S6 switches of the main 110 off off off nu 150 on off off nu 300 off on off nu 1200 off off on nu 1800 on off on nu 2400 on on off nu A800 off on on nu nu pin not used 9600 on on on nu board in the detector f D g e i u 1a ha i ic a Et a E CT A iz ate Board located inside the detector front cover 2 8 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 04 September 97 Data exchange protocol B 30 Controlling the detector with a micro computer RS 232 Users of PC type microcomputers can communicate easily with the detector using the Terminal program in Windows Three protocols are proposed for communications E Hardware local mode The detector sends a continuous data stream reflecting its status in the form of a string of 50 lt CR gt characters e g CYCLE OFF FILAMENT ON lt CR gt FL TEST 1 2 E 8 lt CR gt E Software remote mode This protocol is adapted to the use of the ALSTAT software Optional There is no continuous emission The detector answers the requests sent from the terminal E Printer Printer mode This protocol allows to connect a printer directly to the RS 232 interface The detector sends test default autocali
65. d down After the test time the Helium signal is measured instantaneously and the result analysed according to the next parameters A test ticket is printed if a printer is connected see B 40 and if this parameter is activated When automatic tests are performed for a selected part reference each individual test is identified by a number called TEST REFERENCE Two kinds of test references are available TEST Nr test number automatically incremented after each cycle SERIAL Nr serial number incremented after each cycle but manually modifiable and validated at the end of each test see page 12 A Measured Helium Signal FILE 1 mbar s t t c p a J5 8 a E i 13 8 S E 4 lt K lt E 2 3 4 5 Test No If the measured Helium signal is Reject setpoint the part is rejected If the measured Helium signal is Warning setpoint and lt Reject setpoint the part is accepted but a warning signals that the measurement is approaching the Reject level If the measured Helium signal is lt Warning setpoint the part is accepted Note The test parameters can be programmed independently for the 10 different file available 16 27 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 03 May 97 Basic part modification menu Yes A No Simultaneously press C 50 Alphanumeric Cont
66. d of AUTOCALIBRATION Two audio signals are emitted for a short time with a fixed high frequency when autocalibration failed During MANUAL test Two operation modes are available according to the type of MANUAL REJECT SETPOINT selected in the main menu see page 5 FIXED or FLOATING reject setpoint FIXED reject setpoint An audio signal is emitted when the Helium signal is higher than the setpoint The frequency of the audio signal depends on the value of the Helium signal the higher the helium signal the higher the audio signal frequency The fixed reject setpoint can be moditied while the leak detector is in manual test mode using key F see page 9 This operation mode allows an accurate audio leak detection based on the fixed reject setpoint 22 2 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 03 May 97 FLOATING reject setpoint C 50 Alphanumeric Control and Display Panel ACDP operation e A modulated audio signal is AUTOMATICALLY emitted according to the Helium signal fluctuations When the Helium signal increases the audio signal switches on There is no manual setting in this operation mode This operation mode provides a usefull audio assistance for pin pointing leaks when using the sniffing or spray method Note To quickly shut down an audio signal which is hindering in case of increasing background level the
67. dition 03 May 97 Edition 03 May 97 A 20 ASM 180 TD Detector operating principle Operation in sniffing mode LDS Sniffer probe Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 3 4 A 20 ASM 180 TD Detector operating principle Operation in internal calibration mode M Roughing of calibrated 2 Calibration leak 3 Venting of calibrated leak The leak is returned to atmospheric pressure 4 4 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 04 September 97 A 21 ASM 180 TD ASM 181 TD Detector operating principle Vacuum circuit ee ice 12 20 1 Detector inlet port 3 Detection valve 2 Inlet vent valve 4 Calibrated leak module 3 Vent filter connector 5 Connector for long distance 4 Roughing valve sniffer 5 Inlet pressure gauge PI3C 6 Sniffer valve 7 Bypass valve 7 Hybrid turbomolecular pump 8 Exhaust pressure gauge PI1 PIM 5154 9 Exhaust valve 8 Analyzer cell 12 Roughing molecular drag pump 20 Dry primary roughing pump MDP CP20 Pumping capacities 25 m h 15 cfm roughing dry primary pump CP20 10 l s molecular drag pump MDP Helium pumping speed at inlet port 4 4 s Test capacities Short test cycle Quick response time Autocali
68. ductors located above the insulator to correct this alignment Tighten the filament electrical connections Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 5 8 Installing the analyzer cell metal seal Electrical connections JAEGER connector 1 Filament 2 Triode electrode 3 Braking electrode 4 lonisation chamber 5 Filament 6 Ground E 10 Analyzer cell maintenance Place the metal seal in the cell body seal seat Check that the point where the 2 ends of the seal cross is located near a retaining screw whith one end on either side or use an elastomer seal Check that no electrical conductor is located outside the area delimited by the guide piece Install the cell taking care to lower it into the duct without touching the sides Install the 6 screws with their respective washers Q 6 Tighten the 6 screws in the 3 1 5 P Q y aS sequence shown at right to lo am a torque of 0 8 m daN Ne ZF SP oe aw 6 2 4 6 8 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 03 May 97 E 10 Analyzer cell maintenance Re installing Prepare the metal seal for the VHS preamplifier the preamplifier using the seal former Install the seal and center it in the cell flange seal seat placing the point where the two ends cross near a retain
69. e leak display is at O 3 to 4 bars of the bargraph if this is not the case perform an electrical reset by pressing the key The red indicator light flashes for a moment In the event of failure the red indicator light remains lit contact customer service Autocalibration Check that the detector is not in test mode and that the indicator light is lit green indicator of the cycle key lit Press the key The red indicator light flashes The Helium signal oscillates and stabilizes on the fine leak bargraph The green indicator light comes on to indicate that the autocalibration has ben completed see details of the autocalibration cycle in para 4 It is recommended to run an autocalibration after 1 hour of operation and then on a regular basis once every day In the majority of applications autocalibration is used to make sure that the detector analyzer cell is operating correctly 2 5 Alcatel High Vacuum Technology ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 03 May 97 3 Calibration range Use of an external leak C 30 Detector autocalibration If very accurate leak measurement is required in a measurement range other than 1 108 1 106 it is recommended to use an external calibrated leak the value of which is close to the required value Run an autocalibration as shown in section 2 Once the autocalibration has been completed and v
70. e plug d Contact customer service Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Ll Edition 04 September 97 Connections and Seals Seals D F 90 ltem Description P N 1 Oring NW 16 079237 Oring NW 25 079238 Oring 082113 block valve blank off 16 9 x 2 7 Oring NW 40 082129 Oring NW 63 082140 Calibration block O ring 8 x 1 9 082195 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual yl Edition 04 September 97 Cover F 100 ltem Description P N 1 Lifting ring per unit 076192 2 Stopper for frame per unit 075940 3 Heyco stopper per unit 082922 4 ASM 180 Compact version wheel 101816 per unit pivoting ASM 181 Consol version wheel per unit 5 Rear wheel fixed 101528 6 Front wheel pivoting with brake 101529 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 1 Edition 04 September 97 F110 Options and accessories Metal seals Q Analyser cell 2 onip eee pressure ri 6 turbomolecular PI 3C Oa Cell elastomer seal ltem Description P N 1 Seal NW 25 helicoflex 100745 2 Seal NW 25 helicoflex with colaret 079934 3 Seal NW 40 helicoflex 101492 4 Seal DI 128 7 helicoflex for TMP 5154 079089 5 TMP 5154 NW 40 metal 798024 6 Seals kit for TMP 5154 metal 7 Cell elastomer seal 102823 8 Complete analyzer cell VHS
71. e quickest method to adjust the gas flow is the mass flow controller Mass Flow Controler As an alternative a manual microflow valve DN16 can be used According to usual connection procedures stop valves 6 and filters may be inserted used fo calibrate the installation response time for the furthest point from the detector ratio of actual leak helium signal read on the detector ALCATEL offers reference leaks specially designed for this application without reservoir with 1 4 VCR connectors Different values of leaks are available mbar s 1x109 Part No 103371 1x108 Part No 103372 1x107 Part No 103373 1x10 Part No 103374 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 5 11 Installation connection Principle Precautions C 70 Use of the I gas line option For ASM 180 TD and ASM 180 TD e The detector DNAO inlet port 1 must be blocked e Connect the gas line under test 4 to the detector s 1 4 VCR connector 2 via flexible connection components 3 e Connect the reference leak 5 e Connect the carrier gas flow control accessories composed of a mass flow controller 7 or manual microflow rate valve and stop valves 6 if necessary e Connect the carrier gas source via a flexible tube 8 e A laminar flow must be maintained in the entire line under test to obtain the expected result the response time is increased if a significant vol
72. e side cover unscrewing the 4 fixing screws unscrewing the fixing screws hinge system provided fo Disconnect the DN 25 pumping line near the CP 20 pump 2 3 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 04 September 97 E 70 Partial maintenance of the CP 20 pump Disassembly m Disassemble the intake stator by removing its 4 attachment 5 screws 1 Em Remove the Oring 2 and the 2 bearing O rings 3 on the intake stator 3 E Remove the 2 screws 4 at the end of the shafts followed by the 2 washers 5 and extract the 2 bearings 6 Reassembly m Clean the shaft ends Q m Check the D101 mark on the new bearings Fit them with the mark facing outwards secure using the washers 5 and M5 12 FHe allen head screws 4 m Clean and lightly grease the new O rings and the bearing housings on the intake stator fit 2 seals 3 m Clean the groove of the O ring on the pump body and fit a new seal 2 E Position the intake stator on the pump body and secure it using 4 M6 70 CHc allen head screws 1 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 3 3 Edition 04 September 97 Purpose of the interface board Accessing the interface board 24 V output E 80 I O interface board fuse replacement Fuse specifications T 1 25A 250V 5x 20m
73. eak on the installation Test the installation page 10 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 11 11 Edition 04 September 97 Chapter D User s Manual ASM 180 TD TD ASM 181 TD Maintenance m Toble of preventive maintenance intervals D 10 m General troubleshooting guide D 20 m Problem with the roughing pump D 30 Ne disslay 2 24264 vase Saonee eons ewe D 40 Problem with the secondary pump D 50 Spectro fault 4 o9 aeabu 44 heads dd bade5 ws D 60 Inlet pressure problem 00 D 70 m Cycle start fauls 2 0 ee D 80 m Faults at end of autocalibration D 90 mE iue inh anier Modes esse exp par ee ees D 100 m Helium measurement problem D 110 i O interface problem 00 D 120 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 1 Edition 04 September 97 FREQUENCY 4000 H 5000 H 8000 H 10000 H 24000 H Every 2 years Table D 10 of preventive maintenance intervals MAINTENANCE OPERATIONS TO BE PERFORMED Clean the vacuum lines the valves and the gauges with alcohol Dust the electronic boards and the fan Clean the analyzer cell with alcohol Replace the filament and the electron collector Replace the 2 special seals on the analyzer cell and the VHS preamplifier This cleaning may also be necessary i
74. easured helium flow signal is shown on the gross leak measurement display When the LDS probe is placed in the ambient air the He signal displayed is approximately 5 106 to 1 10 equivalent to the natural concentration of helium in the air Check that the helium signal decreases when the probe hole is blocked with your finger Given the detector configuration the measured flow corresponds to the helium concentration e g Display of 5 10 corresponds to a measured leak of 5 10 mbarl s of He and to a measured He concentration of 5 10 or Sppm The inlet pressure displayed on the remote control unit does not affect operation it is an independent circuit see A 20 The cycle key is not used Alcatel High Vacuum Technology ASM 180 TD TD ASM 181 TD User s Manual 3 5 Audio alarm 90 dB Display the alarm setpoint by pressing Set the audio FO alarm setpoint Feng Adjust the audio eum volume C 20 Detector operation The audio alarm is triggered when the leak rate is greater than the reject setpoint The frequency of the audio signal depends on the leak rate measured by the unit the higher the rate the higher the signal frequency The threshold is then shown on the He mbar l s P mbar measurement displays Gross leak or RE Ea ee Fine leak depending on the test mode used Adjust the setpoint using a screwdriver to turn the potentiometer Release the key Adjust the audio
75. eees page 6 Test procedure 5564 6 gecdveccseeasededans page 8 In the event of a problem 26 page 11 Used to perform spray testing on long lines typical diameter 1 4 with a reduced response time due to the transfer of the helium by a carrier gas injected in the viscous state This option allows the unit to detect leaks of the order of 10 mbarl s in a considerably reduced time in relation to the conventional vacuum test The test is thus quicker and more reliable Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 11 Operating principle Test principle C 70 Use of the I gas line option For ASM 180 TD and ASM 180 TD 1 Detector inlet port 6 Manual valve 2 Gas line inlet port 7 Mass flow controller N2 VCR connector 8 Flexible connector 3 St steel flexible connector 9 Pressure controller 4 Rigid line under test 10 Helium spray 5 Reference leak The detector is connected at the 1 4 VRC connection to one end of the line under test The carrier gas is injected at the other end of the line The line is pumped by the detector and the carrier gas is injected to obtain a laminar flow a few mbar absolute pressure Helium is sprayed around the line In the event of a leak the helium which enters the line is transported to the detector by the carrier gas The sensitivity of the test depends on the heliu
76. er 97 A 60 Factory options Metal seals These reduce the Helium background noise Metal seal locations Inlet port Pie Analyser J cell 1 Valve block Li Inlet pressure bales lecul gauge turbomolecular PI 3C E a PI 3C Elastomer cell seal Used for easier maintenance operations on the analysis cell mass spectrometer This seal replaces the lead seal and can be reused Spare elastomer seal part number 102823 In the event of a high helium concentration in the room in which the test is being conducted the use of this type of seal may generate an increase in the residual signal of the unit Control panel protection A Plexiglas cover equipped with a key is used to lock the access to the detector setting parameters for non qualified operators 3 masses For use of one of the three following tracer gases Helium 4 Helium 3 or Hydrogen 2 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 4 Alphanumeric Control and Display Panel ACDP Automatic test chamber This option includes the ACDP option A 60 Factory options Designed for industrial control it is used to display the measurement in digital form automate the unit test cycle sort tested parts and print the test results on an optional external printer It consists of A user interface mounted on t
77. ers Parameters Setting range Default configuration Reject setpoint 1 0 1010 to 1 0 1077 5 102 Alarm setpoint 1 0 1010 to 1 0 10 1 108 Background setpoint 1 0 1011 to 1 0 10 1 108 Roughing time 1 to 255 s 9s Test time to 255 s Os Test ticket YES NO WES Test reference SERIAL Nr TEST Nr TEST Nr TEST Nr 0 653535 O SERIAL Nr O 65535 0 MAN Reject setpoint FIXED FLOATING FIXED FIXED manual setpoint 1 0 1011 to 1 0 10 5 108 FLOAT manual setpoint FL 2 0 1010 to 1 0 10 Automatic FLOAT manual setpoint GL 2 0 108 to 1 0 101 Automatic MANUAL test Ticket YES NO NO LDS measurement mbar l s or ppm mbar l s Note The default parameters are valid for part reference 1 the values of the parameters are at random for parts 2 to 10 Supervisor board The four following straps allow to configure the audio signal monitoring Se SE SI 22 ST 23 ST 26 Standard audio signal ON OFF ON ON control 1 ACDP audio signal OFF ON OFF OFF control 2 Factory configuration for detectors without ACDP option 1 and 2 with ACDP option 5123 ME e 126 ij THI Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 25 27 Edition 03 May 97 C 50 Alphanumeric Control and Display Panel ACDP opera
78. esponds to that of the electrical source Power CONSUMPTION cceeeeeeeeeeteees 1 6 kVA Fuse Voltage 200 220 240 V 3 SAT Valtore IOC NS We scrsiounnccee 6 30AT Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 2 2 ASM 181 TD console version Edition 04 September 97 Check that the jumper plug is present Without the jumper plug the unit will not operate B 70 Before starting up the detector If the detector is not controlled by the I O interface the jumper plug supplied with the unit must be connected at the rear of the unit ct version 180 Com a dooocoo0oo0o000000 OV wpe O Jumper plug The jumper plug must be connected or disconnected with the power off Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 1 Edition 04 September 97 Chapter C User s Manual ASM 180 TD TD ASM 181 TD Operation Starting up the detector 002 C 10 Detector operation 00000 C 20 Detector autocalibration 00 C 30 1 Switching off the detector C 40 Alphanumeric Control and Display Panel ACDP OPi Oee ae ee ee ee ee ee ee E C 50 L Configuring the unit according to the gas to be o ae o soo aie e wed Riek SNES SERS Ed C 60 Use of the I gas line option ASM 180 TD TD only 0 C 70
79. f there is general contamination which creates insulating surface deposits Check and clean the MD4E membrane pump Replace the membrane Re grease the TMP 5154 hybrid pump Re grease the MDP 5011 pump Check the electronics Calibrate Calibrate the pressure gauges Adjustments Calibration Replace the long distance sniffer filter if necessary Partial maintenance CP 20 pump Replace the ball bearings and the seals of the TMP 5154 Hybrid Pump Replace the ball cau and the seals of the MDP 5011 Molecular Drag Pump Complete maintenance CP 20 Pump Recalibration of internal calibrated leak a F oO ee N lt ASM 181 TD SEE SHEET E 10 E 60 E 20 E 30 Call customer service Call customer service Call customer service Ca Ca Ca E70 customer service customer service customer service E 40 Service intervals The service intervals given are for applications and work rates which conform to the normal operating conditions If the machine is operating under more difficult conditions they can be shortened Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 1 Edition 04 September 97 D 20 General troubleshooting guide A These checks must be performed with the detector isolated from all installations and supplied with the correct electrical power It is assumed that the connection and the electrica
80. ferent models compact detectors 180 console detectors with work surface 181 conventional detectors equipped with oil sealed vacuum pumps oiHtree D dry detectors Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 2 A 10 The ASM 180 series The versions according the detector types PUMPING y O COMPACT VERSION CONSOLE VERSION Standard Rotary vane pump PPM 2021 ASM 180 T ASM 181 T 1 Hybrid pump PTM 5154 Enhanced conventionnal roughing ASM 181 T 2 Rotary vane pumps PPM 2021 with 40 m3 h 1 Hybrid pump PTM 5154 roughing option 2 Rotary vane pumps PPM 2021 1 Turbomolecular pump ATP 100 ASM 181 T2 1 Hybrid pump PTM 5154 Dry 1 Primary membrane pump Type MD4E Molecular drag pump MDP 501 1 ASM 180 TD 1 Hybrid pump PTM 5154 Dry 1 Dry pump Type CP20 1 Molecular drag pump MDP 501 1 1 Hybrid pump PTM 5154 ASM 180 TD ASM 181 TD Enhanced dry roughing 2 Dry pump Type CP20 ASM 181 TD Molecular drag pump MDP 501 1 with 50 m3 h 1 Hybrid pump PTM 5154 roughing option 1 Dry pump Type CP20 1 Molecular drag pump MDP 501 1 1 Turbomolecular pump ATP 100 1 Hybrid pump PTM 5154 ASM 181 T2 D 2 2 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 03 May 97 A 20 ASM 180 TD Detector operating principle Vacuum circuit 1
81. he front panel of the detector an RS232 interface at the rear used to connect an external printer This is used for the automatic bombing testing of small components When the chamber cover is closed the test cycle is initiated via a contact Three aluminium alloy models are available a hemispheric chamber diam 72 mm depth 31 mm a cylindrical chamber maximum diam 85 mm and maximum depth 68 mm a cylindrical chamber maximum diam 160 mm and maximum depth 200 mm 2 4 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 04 September 97 A 60 Factory options Remote control unit Remote control unit with with different cable a 7 m 21 feet cable instead of 3 5 m 11 feet or lenghts c 25m 76 feet cable instead of 3 5 m 11 feet Stainless steel cover Designed for use of the unit UCT in clean rooms Ultra Clean for compact versions Technology The front and rear covers and frame are made of stainless steel An adapter can be attached to the side of the unit for connection to an exhaust system diameter 100 mm Part No 102867 proposed as accessories Test of gas line I Used to perform spray for compact versions testing on long lines typical diameter 1 4 with a reduced response time due to the transfer of the helium by a carrier gas injected in viscous flow In this case
82. hen the temperature has stabilized inside the unit In the case of intensive use of the detector it is recommended to have a spare internal calibrated leak If this is not possible the detector can still be used and auto calibrated using an external calibrated leak see E 50 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 3 3 Edition 03 May 97 Disconnecting the detector Correcting the value of the external calibrated leak E 50 Autocalibration of the detector with an external calibrated leak Turn off the detector and unplug it from the main power Open the front cover lt is recommended to correct this value as a function of the ambient temperature the leak is assumed to have a stable temperature and the time elapsed since its calibration date marked on the leak label E g External calibrated leak of 1 1 x 107 mbar l s helium at 20 C calibrated 1st February 1994 Ambient temperature 25 C Temperature coefficient 3 per C Annual loss 2 The leak value to be entered on Ist February 1995 is Value Temperature Time Value at 20 C correction correction at 25 C 7 3 2 m 7 1 1x10 7 x l 100 x 1 700 1 2x10 i i 3 2 25 C 20 C yearyr The value of the external calibrated leak entered must be between 1 0x10 8 and 8 0x10 inclusive Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 4 Ente
83. ic part modification menu E see page 15 FILE 1 Return to the basic display READY AUTO CYCLE automatic mode 6 2 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 03 May 97 C 50 Alphanumeric Control and Display Panel ACDP operation Basic display Working in manual mode FI ON 1 0E 10 Vacuum test READY MAN CYCLE Start a cycle C or MAN ROUGHING INLET P 2 0E 02 MAN GL 5 OE 06 INLET P 5 OE 02 MAN FL 5 OE 08 INLET P 5 OE 03 Stop the cycle C or FIL ON 5 0E 12 READY MAN CYCLE Indicates alternately filament status Helium signal selected part reference manual cycle standby message The test part is roughed down The test part is connected to the Gross leak measurement system The test part is connected to the Fine leak measurement system Return to the basic display Note in manual mode the normal detector controls are entirely available see C 20 See page 9 for the available operations during a Manual cycle test Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 7 27 Alphanumeric Control and Display C 50 Panel ACDP operation Basic display Working in manual mode FIL ON 1 0E 10 Sniffer test LDS READY MAN CYCLE E Select i FIL DEF 1 0 E 12 SPECTRO RESET LDS T 1 2 E 05 mbar s Stop the sniffer l FIL ON
84. ictured Cell elastomer seal Complete analyzer cell VHS with elastomer seal without magnet Option switch not pictured LDS spare filter set of 5 LDS spare needle LDS probe 5 meter tube LDS spare metal tube 3 Masses magnet Metal seal cajon for option Minisol valve 2 2 NF with coil for Injection valve for lt l gt option a Roughing valve for I option b Protective cover of control panel RS 232 board for ACDP Seal NW 25 helicoflex Seal NW 25 helicoflex with colaret Seal NW 40 helicoflex Seal DI 128 7 helicoflex for TMP 5154 Seals kit for TMP 5154 metal Spray gun Spray gun spare nozzle TDI TD l Block O ring TMP 5154 NW 40 metal VCR Cajon connector Contact customer service F 120 P N 082922 076192 075940 101816 101528 101529 102827 102823 102826 068843 072606 072301 067838 101584 101303 100348 100745 079934 101492 079089 083465 083446 082111 798024 101583 4 4 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 03 May 97 Chapter G User s Manual ASM 180 TD TD ASM 181 TD Appendix m View of the operator interface 0 G 10 L Recording curve Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 1 Edition 03 May 97 G 10 View of the operator interface Senn PAMEE SPECTRO coe 02
85. icult to determine It depends on the pressurisation time pressurisation pressure internal volume of the part dwell time before vacuum test and size of the leak 4 4 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 03 May 97 A 50 Operator interface CONTROL PANEL The counter operates when Detector pumping system Detection parameters the unit is switched on status display analyzer cell details on C 10 elecronic ionization cell current pressure display SPECTRO o2 1 2 wW w0 eeeeceeaaee eeaeaeee mA mbar AUTOCAL le 1 V dec J Q Analog output Alarm setpoint Detector to record the leak and volume autocalibration signal control setting start up details on C 20 details on C 20 Emission current fine Mode selection adjustment Sniffer LDS Gross leak Automatic inlet vent Filament reset Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 2 A 50 Operator interface REMOTE CONTROL UNIT Leak flow Detector Helium signal inlet diselay He mbar l s P mbar pressure T i A display Measurement scale in Fine leak s mode 2 e Measurement scale in i Gross leak or LDS vt mode T Test mode Gross leak Fine leak LDS Filament Cycle ON OFF ON OFF 2 2 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 04 Septemb
86. ing screw one on either side Position the VHS preamplifier carrefully on the cell Tighten the 2 screws progressively and alternately Torque 1 m daN Connect the VHS preamplifier to the cell electrically 2 connectors Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 7 8 Check the flanges for leaks SYMPTOM AND CAUSE FILAMENT DOES NOT LIGHT YES The filament indicator does not come on or the alarm indicator light is on NO FILAMENT LIGHT YES The filament indicator is flashing Spray helium to leak check the flanges of the analyzer cell and preamplifier Eliminate any leaks as describe above Gross leak E 10 Analyzer cell maintenance The detector must be on for a few minutes before leak REMEDY Eliminate the leak Check the torque of the screws on the analyzer cell and VHS preamplifier flanges If necessary install the new metal seals cell seal and VHS preamplitier seal Run an autocalibration see C 30 8 8 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 04 September 97 Accessing the bearings Using the grease syringe Greasing the front bearing Greasing the rear bearing E 20 Greasing the hybrid turbomolecular pump PTM 5154 The frequency of preventive maintenance is listed in D 10 Components P N Grease syringe eeeeeeee 056993
87. ing stages Position the pump vertically Remove the 4 CHC screws 2 and remove the cover from the casing Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 3 MD4E membrane pump maintenance The intake chamber lt components Sang 9 Securing screw 6 Tension disc TD Membrane 8 Support disc 9 Cp Dismantle the intake Remove the O ring 3 chamber Mark the position of the safety valve seals 4 and remove them Remove the top cover 5 if N necessary use the hole on F the edge of the part Clean the soiled parts with alcohol or a solvent Reassemble the intake Put the top cover in place Position the safety valve seals as chamber shown above Close the chamber If the membrane is damaged replace it page 3 2 3 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 04 September 97 Change the membrane Access the membrane Remove the membrane 8 Clean Install a new membrane Reassemble the intake chamber Repeat the operation on the other 3 intake chambers Check that the pump is operating correctly MD4E membrane pump maintenance PJ Remove the securing screw 6 attached screw Remove the tension disc 7 using if necessary a tool screwdriver allen wrench to detech the disc from the membrane using a seal plate if necessary
88. ion Accessories kit analyzer cell Aluminium PI gauge Aluminium PI3C gauge Calibrated leak Fel 407 with thermal probe Complete VHS analyzer cell with lead seal without magnet Electron collector set of 5 Electron multiplier amplifier VHS Filament Lead gasket 10 meter Spare filament for PII set of 5 CP 20 pum CP 20 pump controller Greasing syringe for TMP 5154 and MDP 5011 MDAE membrane pump 100 V 50 60 HZ 220 240 V 50 60 HZ 120 V 50 60 HZ 200 V 50 60 HZ MDP 5011 MDAE Shock absorber type a per unit MDAE Shock absorber type b per unit Partial maintenance kit CP 20 including 2 ball bearings 6001 CP grease tube 10 g 2 Orings 2 x 27 1 Oring 3 x 96 1 plastic box Seals kit for MD4E membrane pump TMP 5154 Standard seal TMP 5154 Shock absorber Contact customer service O F 120 P N 090294 795706 786434 101302 072493 068842 072494 053146 083478 068835 056993 062980 062981 062982 062984 795600 101554 101555 103499 062968 798023 055232 2 4 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 04 September 97 Components summary F 120 Description P N Valves F70 Bacosol 24V DC 23W valve air inlet 104655 Bacosol 24V DC 23W valve by pass 104655 Bacosol 24V DC 23W valve exhaust 104655 Calibration block 100973 Coil for 24V DC minisol valve 067040
89. keyboard as shown in the following example Key Name Command d d fd 77 7 7 71 m o gt len a e LOCAL mode ALSTAT mode Send DATA 7 On eo mw Then communication can be settled Terminal Untitled File Edit Settings Phone Transfers Help Information send by the detector KIE ee Start vac CYCLE Stop CYCLE fart AUTOCALIBR ALSTAT mode Level 1 Start LDS CYCLEfFilament ON OFF LOCAL mode SendDATA 09 41 48 AM Function keys 8 8 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 04 September 97 Purpose Type Interface configuration B 40 Connecting an external printer ACDP option required The ACDP option see A 60 is used to connect an external printer directly to the detector and print test tickets calibration tickets or test parameter readings stored inside the detector see C 50 This type of function guarantees the traceability of leak testing Operations Any printer equipped with an RS232C type serial link is suitable It should have a minimum buffer memory of 2K The tickets printed using the ACDP option contain a maximum of 25 characters per line The external printer should get electrical power from a source external to the detector E Transmission speed 9600 baud E Data length 8 bits E Parity NONE E Siop bit 1 TD Transmission data Pin used GND Gro
90. l continuities have been checked beforehand Note the troubleshooting guide follows a chronological order and a methodology which is the result of the experience of Alcatel CIT Customer Service It is therefore recommended to follow this order so as to locate faults effectively Symptoms detector inlet port blanked off Problems at start up The roughing pump does not start no noise o ee D 30 The by pass indicator light does not come on aoaaa ee D 30 The MDP fault indicator light is on ee D 30 No display on the control panel MDP part 2 2 0220 00 00200 D 30 No display on the remote control and the control panel 2 0 0 2 D 40 The TURBO P indicator light does not come on after 2 min D 50 The TURBO f acceleration indicator light does not come on D 50 The TURBO fault indicator light is on ooann aaa aaa D 50 The filament on indicator light does not come on D 60 The filament on indicator light is flashing gt oaa D 60 The spectro alarm indicator light comes on 0 2 aaa D 60 No inlet pressure display P mbar display 0 00004 D 70 Autocalibration failed 5 d40d5 a4 a4 ee wh Gawd ood VEEN reas D 90 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 2 General troubleshooting guide Symptoms detector inlet port blanked off Problems during vacuum test cycle The cycle key is disabled BB
91. l unit has priority over the control panel keys This key remains activated whatever the display This note is valid for all the commands of the standard detector control panel and remote control Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 3 27 C 50 Alphanumeric Control and Display Panel ACDP operation Unit start up see also C 10 START UP The time is counted down until the XX min YY sec analyzer cell filament comes on AUTOCALIBRATION IN PROGRESS The control panel and remote CAL COMPLETED control unit keys are not functional during this startup phase If MANUAL mode selected lf AUTOMATIC mode selected FIL ON 1 0E 10 FILE 1 Basic display READY MAN CYCLE READY AUTO CYCLE Basic display The basic display gives information which depends on the operating mode automatic or manual in which the detector was previously configured Using the basic display the user can e start a test cycle sy or ie key e access the various configuration menus Pressing only the F_ key makes it possible to return to the basic display without modifying any configuration parameters e perform an autocalibration See the following pages for further details 4 27 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 03 May 97 Access to the various configuration menus Main menu ppm A mbar s C
92. logy France ASM 180 TD TD ASM 181 TD User s Manual 19 27 Date and time modification Increase A Decrease ww C 50 Alphanumeric Control and Display Panel ACDP operation Basic part modification menu page 17 DATE 13 JUL 95 TIME 18 50 38 After A and F are pressed at the same time a blank screen is displayed for a few seconds a Ea id Date selection a j Eci Month selection DATE 13 JUL 95 TIME Year selection DATE 13 JUL 95 ie TIME 18 our selection DATE 13 JUL 95 TIME 18 50 Minute selection DATE 13 JUL 95 Second selection TIME 18 50 38 i DATE 13 JUL 95 The seconds are running TIME 18 50 38 Return to the basic part modification menu page 17 20 27 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 03 May 97 C 50 Alphanumeric Control and Display Panel ACDP operation Detector autocalibration Basic display FIL ON 1 0E 10 READY MAN CYCLE CAL COMPLETED or AUTOCALIBRATION IN PROGRESS AUTOCAL FAILED ESCAPE F FIL ON 1 0E 10 READY MAN CYCLE To be taken in account by the ACDP the autocalibration has to be started from basic display Autocalibration cycle started see C 30 Note electrical autoreset is not taken in account by ACDP Display of result In case of fault 2 audio signals are emitted Autocalibra
93. m It controls the I O interface which is used to link the detector to an external control system On ASM 180 TD and ASM 180 TD it is located in the bottom part of the detector protected by a stainless steel cover On ASM 181 TD it is located on the rear inside the frame ASM 180 TD and ASM 180 TD Switch off the detector and disconnect all the connectors at the rear Remove the rear cover and unfasten the 3 attachment screws of the stainless steel bottom cover located at the same level as the interface connectors on the frame Open the front cover and unfasten the 4th screw located near the hybrid turbomolecular pump oblong hole on the frame At the bottom part of the detector release the O interface board and the stainless steel protective cover on which it is attached ASM 181 TD Remove the rear and side cover The I O board is accessible at the rear inside of the frame near the Sub D connectors A 24 V DC ouput is provided on this board pins 24 and 25 of the O interface This output can or cannot be selected using straps located on this board ST1 and ST2 If these straps are present the 24 V output is available on the pins 24 and 25 of J3 If there are no straps an external power supply coming from an automatic control system for example should be provided see B 20 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 2 Change
94. m Technology ASM 180 TD TD ASM 181 TD User s Manual J5 C 20 Detector operation Starting up evacuation of the line and the part ne ese Loa a Start a cycle by S The pressure drop is shown on pressing on the key B the display unit According to the characteristics of the test part and therefore the pressure reached the unit is placed in gross leak or fine leak test mode Gross Leak mode 6 mbar gt P gt 2 10 2 mbar Fine Leak mode P lt 2 10 2 mbar He mbar s P mbar w2 wl Atm Working in Gross Leak t is sometimes preferable mode to work in Gross Leak mode in order to reduce It is possible to preset cycle times the gross leak mode by pressing the key 2 5 Alcatel High Vacuum Technology ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 04 September 97 Working in Sniffer mode LDS Select the LDS function Check the probe operation Measured flow concentration LDS mode specificities C 20 Detector operation Connect the th im probe to the quick z sa connector o ae SM 181 Conse gt The filament emission goes off for a few seconds during the probe roughing phase The test is operational when the emission presence indicator light is lit He mbar I s P mbar w Atm 2 w w 5 w 5 w 5 5 2 w w 5 2 w wi 2 1 The sniffer test mode indicator light is lit on the unit The m
95. m content of the carrier gas which must be as low as possible 2 11 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 03 May 97 Detector operation The gas line test option is an addition to the basic detector functions The ASM 180TD I is optimized for the 1 4 gas ine test in terms of response ime and sensitivity The ASM 180TD I provides reduced response time for gas lines diameters higher han 1 4 C 70 Use of the I gas line option For ASM 180 TD and ASM 180 TD ASM 180 TDI lt I PH 23 18 14 a 5 3 MDF i Y 4 19 16 is amp 11 ASM 180 TD I gt a 18 144 AY 35 IAF r x x Dm S 16 As Detector inlet port 4 Roughing valve 3 Detection valve 1 Roughing membrane pump MD4E 2 Roughing molecular pump MDP 12 20 17 Hybrid turbomolecular pump 18 Analyzer cell 20 Dry roughing pump CP20 21 l gas line inlet port 22 gas line membrane 23 l gas line valves Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 3 11 Choice of carrier gas C 70 Use of the I gas line option For ASM 180 TD and ASM 180 TD the most commonly used carrier gas is nitrogen In order to be able to identify leaks of approximately 10 mbar s
96. mbar or CP20 ASM 180TD pressure lt 3 107 mbar Exhaust pressure of the MDP is not correst it must be lt 5mbar NO Leak on lines or valves Contact customer service NO Contact customer service A At cycle start it is normal for the MDP 5011 rotational speed to slow temporarily CD2 board Fuse or board defective Check the fuse on CD2 board fuse DITF 5x20 0 5A No display on jam the control panel MDP part Gillam NO Contact customer service Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 3 3 Edition 03 May 97 D 40 No display SYMPTOM CAUSE Main fuse is out of order REMEDY No display on the remote control and the front control panel but the primary pump Change the fuse located under the main switch See F 30 starts on NO Contact customer service Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 1 Edition 04 September 97 SYMPTOM The P TURBO indicator light ano 6 does not come el on after more ka 7 than 2 min Sa D 50 Problem with the secondary pump CAUSE REMEDY PI gauge Replace the gauge detective NO TMP exhaut valve The valve does not open The coil is defective or the plunger is stuck Contact customer service NO Roughing p
97. n a few seconds Contact customer service Change the LDS probe filter see F 110 Contact customer service Contact customer service Recondition Contact customer service Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 1 Edition 04 September 97 SYMPTOM No change to Fine Leak mode P lt 2 10 2mbar Low sensitivity D110 Helium measurement problem CAUSE GL key engaged The indicator light is on NO High triode gauge pressure Analyzer cell degassing time too short NO Cell contamination The cell is dirty NO REMEDY Release the key Contact customer service Wait for a spectro pressure lt 10 4 mbar Clean and recondition see E 10 Contact customer service Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 2 SYMPTOM High background noise D110 Helium measurement problem CAUSE REMEDY Leak inside the detector Eliminate the leak Leak visible during a Helium leak check NO Vacuum line contamination Clean with alcohol or The lines are dirty or greasy replace them if necessary NO Wait for a spectro High triode gauge pressure pressure lt 10 4 mbar Analyzer cell degassing time too short NO Cell contamination Clean and recondition The cell is dirty and polluted see E 10 NO Ultimate pressure of Contact customer
98. ncy of preventive maintenance is listed in D 10 Components P N Internal calibrated leak 101302 It enables the auto calibration of the detector Autocalibration is triggered at startup or when the button is pressed on the control panel see C 30 In order to ensure the reliability of the helium test ALCATEL recommends to recalibrating the internal calibrated leak at least every 2 years from the calibration date marked on the leak label and its calibration certificate Recalibration is generally performed using a comparative method with a reference standard This work can only be performed in ALCATEL or other approved service centers For this therefore it is necessary to remove the internal calibrated leak from the detector Switch off the detector and disconnect it from the main power Open the front cover of the detector attached with 4 screws Disconnect the heat sensor connector 3 pin connector Disconnect the DN 16 flange and remove the calibrated leak Caution Do not separate the heat sensor from the calibrated leak Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 3 Installation of a new internal calibrated leak Entering the new internal calibrated leak value E 40 Replacement Recalibration of the detector internal calibrated leak A recalibrated leak is returned to you with a new value
99. neous printout of ticket if the external printer is connected to the printer interface see B40 and activated see page 5 sample printout see page 10 Vacuum test only The filament indicator light goes off Vacuum test only The filament indicator light is lit fase after a few seconds Only if FIXED reject is selected and when leak detector is in test mode The FIXED setpoint is memorised until next modification It is active whichever the test status GLFL or LDS Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 9 27 Ticket printing examples manual mode Vacuum test ticket Alphanumeric Control and Display Panel ACDP operation FILE 1 CUSTOMIZED REF MANUAL CYCLE C Elapsed time sec P Pressure mbar S Signal mbar s CYCLE START 19 NOV 96 16 09 42 C 0s P 1 0E 03 S 7 5E 10 GL MODE C 5s P 1 5E 00 S 2 2E 06 S 1 1E 09 in FL test Selected part Alcatel reference Selected part 16 characters customer reference During test Request for immediate printing by keeping y and A pressed simultaneously C 50 FILE 1 MANUAL CYCLE C Elapsed Time sec P Pressure mbar S Signal mbar s CYCLE START 19 NOV 96 16 11 46 GL MODE C 5s P 7 4E 00 S 1 0E 06 FL MODE C 6s P 1 0E 01 S 2 4E 08 STOP CYCLE C 22s P 1 0E 02 S 1 0E 10 in FL test N B A minimum of 1 second is
100. ng on the detector again Running a test E 50 Autocalibration of the detector with an external calibrated leak Close the detector cover Connect the detector to the main power Switch on the detector Connect the external calibrated leak directly to the detector inlet if it is equipped with a valve the valve should be open Start a test cycle Allow the signal to stabilize for a few minutes Note external autocalibration can be performed either in GL or FL mode provided that the residual helium signal of the detector background noise is at least one decade less than the corrected value of the external calibrated leak Internal autocalibration is performed systematically in FL mode internal calibrated leak of approximately 1x107 mbar s maximum Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 3 4 Running the calibration AUTOCAL E 50 Autocalibration of the detector with an external calibrated leak When the helium signal has stabilized in test mode press the key on the control panel AUTOCAL The red indicator light comes on and flashes wet the external autocalibration is performed _ The result appears on the key indicator lights in the same way as for an internal autocalibration see C 30 If the autocalibration fails the red indicator light of the key comes on the previously saved settings are retained and the use of the detector is not disabled
101. of helium This electrode the triode electrode 4 is used to measure the total pressure inside the analyzer Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Analyzer cell operating principle N 5 magnetic field p i 4 He Electron a d lt multiplier ba as bd c Say 200V 28 Target adjustable 5 2 leak signal A 39 gt __ He leak signa Helium lt j p a peak p gt X sov fier p amP mm Electronic 1 a current sa IRA 0 2 2 0 ma 1 adjustable Sensitivity P H 1 Filament 2 Electron collector 3 lonization chamber 4 Triode electrode 5 Diaphragm 6 Braking electrode Electron beam Heavy ions He Helium ions Light ions Edition 03 May 97 A 40 Testing methods Leak detection is used to detect micro openings porosities etc in test parts The detection of these passages involves the use of a light gas which is capable of infiltrating the smallest passages quickly Helium The detector samples and measures the helium flow rate entering the test part via the leak s The testing method is selected according to the test part and the measurement accuracy required The part can be connected to an evacuation line The part is sealed lts characteristics allow it to be pressurised with helium or a mixture containing helium lts characteristics allow it to be placed in
102. once the filament is emitting The panel then displays the filament current and the pressure in the analyzer cell The detector checks the cell calibration The autocalibration key red indicator light flashes for a few seconds see autocalibration sheet C 30 for details If no problems are encountered the unit is considered to be calibrated the green indicator light comes on The detector is ready to be used The cycle control is enabled when the green indicator light on the cycle control key comes on the autocalibration is validated 2 2 Alcatel High Vacuum Technology ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 04 September 97 C 20 Detector operation In G 10 the user will find a view of the operator interface It can be used to identify the operational parts of the control panel and remote control unit The following pages contain Working in vacuum fest mode Pages l 2 Working in Gross leak mode Page 2 Working in sniffer mode Page 3 Setting the audio alarm setpoint Page 4 Saving the filament Page 5 Inlet port venting at the end of the test Page 5 Recording the Helium signal Page 5 Working in Make sure that the parts can withstand the vacuum test mode difference in internal external pressure to which they are subjected Connect the test part Alcatel High Vacuu
103. ontrol the jumper plug supplied with the detector must be kept in place in order to use the operator interface contacts 23 24 connected OT OO 0 OC E A 440 0000000 aA SO OO 6G oo 6 Of Sp S Os O 5 sere A Ogg t l The common points of the i l I controls are UER pins 21 and It is recommended to use a shielded ca t which is grounded on the connector cap 23 Interface Contact open the detector is controlled by the I O interface the operator can not access the keys E m on the control panel or the filament key on the remote control unit Contact closed the unit is controlled by the operator interface 22 Calibration Falling edge Autocalibration sequence start 8 Cycle Falling edge Cycle start 20 Filament Closed Filament on 18 GL mode Closed Gross leak mode selection 17 LDS mode Closed DS mode selection 16 Inlet vent Closed Automatic vent mode selection Note if contacts 22 and 8 are kept closed to ground the cycle and autocal keys on the operator interface are inactive Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 2 The signals outputs Dry contacts Direct current 60V 60W or 2A max Alternative current AOV 125VA or 2A max Recorder output Note B 20 Controlling the detector with the I O interface Contact closed lez 3 A 50 ZQ tO 11 PER 19 15 l4 15 15 24
104. orm maintenance on the cell see E 10 Check the consistency of the display with an external calibrated leak If the calibration is accurate enough it may be possible to continue using the detector if necessary e Perform maintenance on the cell see E 10 Contact customer service 2 2 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 04 September 97 SYMPTOM The a indicator light on the remote control unit does not come on LDS key engaged The LDS en si is less fha 5x106 mbar V s LDS probe in ambient air Spectro indicator lights No He a Filament off D 100 Faults in sniffer mode CAUSE Circuit breaker switch The cycle key yellow indicator light is on or the FL ramp is on NO LDS probe filter blocked e Blocking the end of the LDS le effect on the helium signal e Changing complete probe corrects problem NO LDS tube pinched or blocked e Blocking the end of the LDS le effect on the helium signal e Changing complete probe corrects problem NO Hole in LDS tube probe with your finger has litt probe with your finger has litt e Normal operation can be restored when the probe is disconnected e Changing complete probe corrects problem NO REMEDY Press the key to stop the current cycle The LDS is started up i
105. ose of the ACDP option eeeeeees page 2 Use of the configuration keys ceceeeeee page 3 SIGMPUPs 66568540309 5054 nrs iE Ea ERE ES page 4 Access to the various configuration menus page 5 Manual mode ein meni Automatic mode Working in manual mode Vacuum test page 7 Start a cycle Stop a cycle Working in manual mode LDS test page 8 Start a test Stop the test Operations available during the manual test cycle page 9 Display the elapsed time Print the current measurement Switch off the filament Switch on the filament Example of test ticket in manual mode Working in automatic mode Vacuum test only page 12 Start an auto test Result messages at end of auto test Error messages at end of auto test Examples of auto test tickets Part modification menu ee cece ceeees page 15 Purpose of part modification menu Automatic test parameters Basic part modification menu Test part selection change Test part reference customization via RS 232 Automatic test parameter modification Time and date modification Unit autocalibration 0 cece eee eee eee page 21 Monitoring of the audio signal with the ACDP page 22 Default unit configuration cee cee eeees page 24 Internal switches Automatic test parameters Setting ranges Summary functional diagram
106. r the manual test only the manual reject setpoint is taken into account see page 5 Function only possible if the detector is connected to a micro computer PC in terminal emulation through the RS ACDP interface B 40 E g Relai 22S ACD In the event of an error repeat CTRL D 18 27 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 03 May 97 Part parameter modification Increase amp Decrease V Yes A No v SERIAL A TEST VW Basic part modification menu page 17 CHANGE TEST PARAMETERS YES REJECT LEVEL 1 0 E 01 BACKGROUND SETPOINT ROUGHING TIME sec 255 TEST TIME sec 255 TEST TICKET NO TEST REFERENCE SERIAL TEST 15 RESET NO PRINT FILE NO IE Return to next display of basic part modification menu page 17 C 50 Alphanumeric Control and Display Panel ACDP operation See pages 15 and 16 for the definition of the test parameters Above this setpoint the part is rejected ls used to prevent drift of the tested parts or an increase in the unit background noise When it is activated the back ground setpoint can be adjusted Roughing time selection Test time selection Ticket printout after each test see page 14 A Yes permit to reset the test number Part test parameter printout see page 14 Alcatel Vacuum Techno
107. r to the installation via the hardware interface ASM 180 TD TD compact versions DN 40 INLET PORT O Connect the remote control unit Sub D 25 pts plug Edition 04 September 97 2 Connect the i The 1 0 interface connector should never be 1 O interface connected or disconnected with the unit on see B 20 If the detector is not It the detector is controlled A N controlled by by the I O interface install the I O interface the interface cable to the the jumper plug Sub D 25 pin connector on must be connected the detector 3 Connect the RS232 If the detector is to be connected to a micro computer see B 30 connect the cabled RS232 4 Connect an ACDP option required see A 60 external printer Connect the printer output using an RS232 cable see B 40 5 Connect the LDS probe 6 Connect to atmospheric pressure 7 Connect the inert gas purge Connect the unit to the main power quick connector When a neutral gas is used the filter is unscrewed and replaced by the connection to the selected gas supply source 1 4 BSP connector Pressure 1 070 atm absolute quick connector ASM 180 TD only The helium concentration of neutral gas must be lt 1 ppm Pressure 1 4 0 1 bar absolute 20 psia Check that the voltage marked on the unit identification plate corresponds to that of the electrical source ASM 180 TD 1 2 kVA _ ASM 180 TD 1 5 kVA Fuse Voltage 200 220 240 V 3 1
108. reamplifier check failed filament off red autocal indicator light on FL ramp too high NO Autocalibration system configuration Helium signal displayed on the e Incorrect position of switch S1 and status of indicator light CR1 e Temperature sensor disconnected NO Sensitivity fault e Long autocal cycle e Spectro filament current red during autocal cycle i see page 2 2 e Incorrect thumbwheel values REMEDY Contact customer service Contact customer service Correct the faults inside the detector see E 40 e Check the consistency of the display with an external calibrated leak If the calibration is accurate enough it may be possible to continue using the detector if necessary e Perform maintenance on the cell see E 10 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 2 D 90 Faults at end of autocalibration SYMPTOME CAUSE Helium background fault e Autocal cycle interrupted e High background out of test relative to internal calibrated leak NO Peak fault e weak or non existent oscillations of the helium signal during the autocal cycle NO REMEDY Check the consistency of the display with an external calibrated leak If the calibration is accurate enough it may be possible to continue using the detector if necessary e Perf
109. ring the external calibrated leak value E 50 Autocalibration of the detector with an external calibrated leak eh ara Rereye gt ieye al EE memen wo BERRE i Nae AG EE i m ao oa ta ie 9 This value is entered on the p Jtn WM thumbwheels 2 SMN 2 3 S4 located on the m supervisor board e Set on the thumbwheels the helium value of the corrected calibrated leak of 5 re Set as example S4 to 1 oo Io 2 S2 to 7 for a value of 1 2 x 107 mbar l s Set the switch S1 on the main board to ON The adjacent red indicator light comes on in the external autocalibration position uly Ns OFFOMON v n a as In this position the internal autocalibration in particular the control of the internal autocalibration system valves is disabled the automatic internal autocalibration is no longer performed at detector startup Only the external autocalibration is authorized it is started by pressing the key with the detector in test mode see next pages 2 4 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 03 May 97 Switchi
110. rinter Connection components A 70 Accessories Stainless steel cover option required This is connected to the I O interface and frees the operator s hands The test cycle is initiated by pressing on the pedal The unit equipped with the ACDP option can issue test tickets and autocalibration reports fo guarantee measurement traceability Refer to the B 40 and C 50 section concerning the use of the ACDP option e e D NAnanaaranrenn e gt D Part No Part No 120V 60Hz 100V 50 60 z 220V 50Hz eel flexible hose L 250 mm DN 40 eel flexible hose L 500 mm DN 40 eel flexible hose L 1000 mm DN 40 eel symmetrical T DN 40 eel symmetrical cross DN 40 reducing nipple DN40 DN25 centering ring with viton seal DN 40 102867 Part No 100913 103593 103594 102873 Part No 068373 068374 068375 068564 068571 068253 068230 AAN ely H NOS e AY Ws For any other accessories contact our sales department 2 2 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 04 September 97 A 80 Technical characteristics ASM 180 TD ASM 180 TD ASM 181 TD with 50 m3 h Standard Standard Standard roughing option Roughing primary pump 4 m3 h 25 m3 h 25 m3 h 2x 25 m3 h 2 4 cfm 15 cfm 15 cfm 2 x 15 cfm 10 l s 10 s 10 l s 10
111. rol and Display Panel ACDP operation Basic menu page 5 or 6 FILE 1 CHANGE YES y FILE SELECTION 10 parts in automatic and 10 1 parts 1 file by pass in manual F mode can be defined page 18 CHANGE TEST A Yes used to modify the part PARAMETERS NO parameters page 19 y DATE 13 JUL 95 ls used to modify the date and time TIME 18 50 38 page 20 F FILE 1 Return to the main menu CHANGE NO Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 7 27 Selected part change Select the part using the keys A and M Example amp Part reference customization From the PC send CTRL D decimal code 4 Send 16 characters defining the reference assigned to the part Delete the customized reference send CTRL N decimal code 1 4 Delete all the customized references send CTRL O decimal code 15 Alphanumeric Control and Display C 50 Panel ACDP operation Basic part modification menu page 17 FILE SELECTION 1 FILE SELECTION 2 FILE 2 FILE 2 RELAI 22S ACD FILE SELECTION RELAI 22S ACD FILE SELECTION 2 F Y Return to next display of basic part modification menu page 17 lt is possible to select 10 different test parts each with its own test parameters When the manual test mode is activated it is possible to select an 11th imaginary file by pass part without any parameters specifically fo
112. rts each with its own test parameters e At the end of each automatic cycle the test result is displayed the part is sorted as Accepted or rejected according to the programmed reject level e An audio signal could be activated according to the result of the test and the selected configuration The ACDP panel operating instructions are given in the following pages We advise you to read them carefully in order to become familiar with its operation At the end of C 50 pages 26 and 27 the user will find a view of the front panel of the option and a flow chart showing all the available functions 2 27 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 03 May 97 Use of the configuration keys C 50 Alphanumeric Control and Display Panel ACDP operation This key is used to validate the selections made and to access the various configuration menus It is activated in the basic display see page 4 These keys are used to select a response or adjust values They are activated whenever the display shows a 3 To adjust values such as reject level hold down the v or A keys fo scroll through the figures more quickly This key allows to start a test cycle in manual or automatic mode It also allows to stop the test cycle in manual mode It is activated when the display shows READY AUTO CYCLE or READY MAN CYCLE Note The cycle key of the remote contro
113. signal volume with the Audio section knob When this knob is at the minimum position O position the audio signal is cut off A 5 Alcatel High Vacuum Technology ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 04 September 97 Saving the filament w Press the key on the remote control unit B Enable the inlet vent Disable m the inlet vent press E Record the helium signal C 20 Detector operation To save the filament it is possible to switch it off when it is not to be used for a period of time The indicator light then goes off The filament is switched on again by pressing on the key a second time The indicator light flashes for approximately 8 s before lighting up the filament is now operating Note The filament can only be switched off when the detector is in test mode When the inlet vent indicator light is lit at the end of EA the cycle the inlet vent valve is open It is possible to disable the opening of this valve by releasing the key The indicator light goes out This function is important to prevent the installation from returning to atmospheric pressure by mistake This output supplies a voltage of O to 8V see recording curve in G 20 The response curve is logarithmic 1 volt per decade Alcatel High Vacuum Technology ASM 180 TD TD ASM 181 TD User s Manual 5 5 Edition 03 May 97 Purpose of auto
114. the detector is equipped with an additional 1 4 VCR connector specitic to this option and a luminous button to activate the function Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 3 4 A 60 Factory options 50 m3 h roughing In order to reduce the roughing time when testing large for console version volumes a second CP 20 rotary vane pump can be added to the roughing system Vacuum circuit of the ASM 181 TD equipped with the 50 m3 h 2x1 5cfm roughing option 3 13 ve N N O D 20 Apart from the roughing capacity and the weight 185 kg 406 Ib with the option the characteristics and the use of the leak detector remain the same A A Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 03 May 97 ALSTAT statistical software kit Cart for compact versions Long Distance Sniffer LDS probe Spray probe A 70 Accessories To be used when the detector is connected to a PC compatible computer Part No 785911 Part No 072654 This is used for long distance sniffing tube length 5m Part No 072301 Helium spray probe less tubing Part No 083465 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 2 Exhaust line adapter for compact versions Cycle control pedal P
115. the fuse E 80 I O interface board fuse replacement In case of the use of the detector with the jumper plug STI and ST2 have to be present 24V 24V f r wu yuo e 3 4 T1 25A M24 24 L E E 100nF c31 Z ST2 100pF o 2 1 M24 The fuse located on this interface board T1 25A 250V protects the 24 V supply when the straps ST and ST2 are present 2 2 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 04 September 97 Chapter F User s Manual ASM 180 TD TD ASM 181 TD Maintenance components Preventive maintenance components delivered with the detector 202 0000000 F 10 Monitoring and display F 20 Power and electrical supply 00 F 30 Automatism and electronic circuits F 40 MECCUIEMIEN 035604 65 bans lt 8d kaa e F 50 c Pumping 00 c eee eee F 60 VIVES sas eri Gea E w EEGs e F 70 SE 0S ok eae etree EE yn ee tees Bes F 80 H Connections and seals 0 F 90 b Covers nance e a a o4 F 100 Options and accessories 2 0 2 02 ee F 110 Components SUMMON ss soos ooum 34 eh owe y F 120 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 1 Edition 03 May 97 Maintenance kit delivered with the detector ASM 180 TD ASM 180 TDi P N 090201 ltem N S
116. the same way as for standard RS 232 link see B30 ct version ASM 180 com gt oF Reception data RD __ TD Transmission data Data set ready DSR DTR Data terminal ready GND Ground k Ol C2 O3 A CS WARY Ready to send RTS CTS Clear to send Connecting cable Micro computer Detector 7 and 8 Connections are necessary only if RTS and LTS are used in a software created by user Link configuration Speed 9600 bands Data length 8 bits Te shape ae tg ios a dea is None Stop bit ieee bee ee 4 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 3 4 Example of customized reference of the tested part via ACDP option see procedure in C50 Example of the acquisition on a PC of a test result controlled by the ACDP option Copy of test ticket B 40 Connecting an external printer ACDP option required e Connect the PC to RS 232 printer interface as explained above e Set the Terminal function under Window see B30 e As soon as the ACDP panel proposes the choice of a part Basic part modification menu see C50 page 17 e Send from the PC CTRL D and then the 16 characters to identity the part see C50 page 18 PART SELECTION NUMBER 1 FILE SELECTION RELAI 22S ACD e PC connected to RS 232 printer interface See on C50 for the operating mode of ACDP option
117. tion MAIN MENU Start up WORKING CONFIGURATION aaa er FIL ON 1 0E 10 FILE 1 E ond os a ee Basic display READY MANUAL CYCLE READY auTOcycie euuiutleal eo E F or g o manual E ee Closin testcycle i Cycle mode selection test chamber Beas d PART MODIFICATION MENU Y Manual Auvo A General screen The purpose of this sheet is to give a guide for the use of ee WER i flow chart the option s menus i ay Fj Yes N Vv File selection F a If manual cycle IF auto cycle E Manual reject setpoint If ref prototype A Floating Fixed W If ref 1t010 F Change test parameters Printer FV NO Ys A z A Yes No VW z F Reject level i A v LDS test units F A PPM mbar l s W Warning level i A v i Background level A Option v F Roughing time A v F Test time A v F Printer A Yes No W F f an L Test reference Test reset F Reset A Serial Tet WHFiW NO Yes A 3 i ____ a E Print file F a Prit 1 Yv NO Yes ASFA A malig Date time display A F gt Change time date Return to the main menu F Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 26 27 Edition 03 May 97 C 50 Alphanumeric Control and Display Panel ACDP operation View of the ACDP panel
118. tion ticket printout if printer connected see B 40 examples of tickets AUTOCALIBRATION REPORT STARTED 13 DEC 95 9 13 5 COMPLETED 13 DEC 95 9 13 40 DETECTOR STARTED 13 DEC 95 8 05 30 END OF AUTOCAL REPORT AUTOCALIBRATION REPORT STARTED 13 DEC 95 9 13 48 FAILED 13 DEC 95 9 13 51 DETECTOR STARTED 13 DEC 95 8 05 30 END OF AUTOCAL REPORT Return to the basic display Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 21 27 C 50 Alphanumeric Control and Display Panel ACDP operation Audio signal operation DETECTOR and monitoring CONFIGURATION AUDIO SIGNAL MONITORING SELECTION BY Te ens eater oe BY STANDARD CONTROL INTERNAL SWITCH FACTORY CONFIGURATION PANEL lt RS232 and l fase e WHEN OPTION IS SEE C 20 See page 24 and 25 La Lo SEE HERE BELOW Adjust the audio volume The audio signal volume is adjusted with the knob located in the AUDIO section of the standard control panel When this knob is at the minimum position O the audio signal is cut off This button is the only one to be activated in the AUDIO section of the standard control panel End of AUTOMATIC test An audio signal is emitted for a short time with a fixed frequency when the tested part is REFUSED e Two audio signals are emitted for a short time with a fixed high frequency when the part has to be restested because of a fault En
119. ume is between the carrier gas supply and the detector e t is advisable to place the detector as close to the zone liable to leak as possible e Purge the injection system with the carrier gas in order to eliminate the air e t is better to stop the gas line test function in case of autocalibration Note It is not necessary fo connect a neutral gas purge to the detector the carrier gas acts in the same way as the purge 6 11 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 03 May 97 Edition 03 May 97 Multiple line test C 70 Use of the I gas line option For ASM 180 TD and ASM 180 TD Shut off the line under test as much as possible with the valves and fittings available The zone under test is limited to the line through which the carried gas flows to the detector It is therefore necessary to prevent the flow of carrier gas through the lines not under test using the valves V1 V2 and V3 To test the line open A y B V2 C V3 Vi V3 ii 3 STI Aa AN a close connect V2 and V3 V1 and V3 2 V2 and V1 3 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 7 11 Test procedure Close the carrier gas supply Start up the detector A Activate the gas line test function Run a cycle C 70 Use of the I gas line option For ASM 180 TD and
120. ump Contact customer service limit pressure MD4E ASM 180 TD pressure lt 3 mbar or CP20 ASM 180TD pressure lt 3x10 mbar pump rotating with its inlet blanked off NO Eliminate the cause of the leak Leak in the line Gross leak visible with the alcohol test NO Contact customer service Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 2 D 50 Problem with the secondary pump SYMPTOM CAUSE REMEDY The CD2 Converter circuit Replace it je 3 The board fuse is faulty indicator light does not come on NO Contact customer service The CD2 converter circuit Set it to N TURBO board switch see E 20 is set to R fault i indicator light VA IS On TMP overheating due to an Contact customer service external cause The fan is stopped NO TMP internal overheating Contact customer service The TMP thermistor is triggered Temperature gt 60 C NO TMP faulty Contact customer service Abnormal noise from the pump The pump does not rotate seized NO Contact customer service 2 2 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 04 September 97 SYMPTOM The filament PFS indicator light i does not come on The filament indicator light is flashing The spectro alarm indicator light comes on
121. und Detector male connector Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 4 Connecting the printer External printer option Connecting configuration Detector side female o 080 O80 A eee o sito 9 pin 70602 Wy oa OJ S60 20 pin Printer side serial port B 40 Connecting an external printer ACDP option required The connection is made directly to the printer RS232 interface port Interface port only valid with the ACDP option jon mpact versio ASM 180 5 l e When a detector is ordered Alcatel offers an external printer which includes the ACDP option Alphanumeric Control and Display Panel a thermal printer with 112 mm wide paper and electrical power supply adapter the detector printer connection cable Printer offered SEIKO DPU 414 4O B printer Thermal paper SEIKO TP 41 1 28CL width 112mm reel diameter 48mm 2 4 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 04 September 97 B 40 Connecting an external printer ACDP option required Connecting a PC The connection of a PC micro computer allows to customize micro computer to the the reference of the parts to be tested under of the control of RS 232 printer interface ACDP option see C50 Wiring and transmission are done in
122. via an LDS Long Distance Sniffer probe collects the helium esca ping from the part Global test Sniffer probe The part is placed under a cover containing a sniffer probe The leak cannot be located The helium from the leak accumulates over time inside the cover The detector measures the concentration Local sniffing test Sniffer ornhe The sniffer probe is moved over areas likely to contain leaks The leak can be located The signal supplied by the analyzer is not a direct measurement of the leak The sniffer probe only collects part of the helium escaping from the part depending on the distance separating the leak from the tip of the probe Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 3 4 Bombing method A 40 Testing methods This method is used for sealed objects that cannot be connected directly to the detector semiconductors waterproof watches etc Fae Part Detector Chamber 1 Chamber 2 he part is placed in a vessel containing pressurised helium he helium penetrates the part if it has a leak The part is then removed from the vessel and placed in another vacuum vessel which is connected to the detector The helium escapes from the part through the leak and produces a signal This signal is not a direct measurement of the leak as the helium pressure inside the part is diff
123. wd A 90 Installation Precautions and unpacking B 10 Controlling the detector with the I O interface B 20 Controlling the detector with a microcomputer RS232 00 0 B 30 Connecting an external printer 0 2 0 B 40 Connecting a neutral gas purge ASM 180TD only B 50 Connecting the leak detector to the installation via the hardware interface B 60 Before starting up the detector B 70 MANUAL REFERENCE 104435 EDITION 04 September 97 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 3 Operation Maintenance Troubleshooting MANUAL REFERENCE 104435 EDITION 04 September 97 Contents User s Manual ASM 180 TD TD ASM 181 TD Starting up the detector o oo naaa aaa Detector operation 0 000005 Detector autocalibration Switching off the detector Alphanumeric Control and Display Panel ACDP OUSICQNON ssa 2634904006448 6 2054054 Configuring the unit according to the gas to be detected 2 eee Use of the I gas line option ASM 180 TD TD only 24 54 2455085 Table of preventive maintenance intervals General troubleshooting guide Problem with the roughing pump No display gous caciuteia boven Seeasd Problem with the secondary pump Spectro fault 258 4 oa aes dusts e ee Stee 4 Inlet pressure problem
124. with elestomer seal without magnet Contact customer service Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 1 4 Control panel protection Alphanumeric Control and Display Panel ACDP Options and accessories Item Description Q Protective cover of control panel 10 ACDP control assembly 11 RS 232 board for ACDP Contact customer service F110 NCCE y a GT OO A a P N 100348 2 4 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual Edition 04 September 97 Edition 04 September 97 Options and accessories l option F110 ASM 180 TD Item Description VCR Cajon connector Metal seal Cajon for l gt option Injection valve Roughing valve Minisol valve 2 2 NF for lt l option with coil TDI TD Block O ring option switch not pictured 529C 24V bulb for l option switch not pictured P N 101583 101584 101303 082111 102826 102827 Alcatel Vacuum Technology France ASM 180 TD TD ASM 181 TD User s Manual 3 4 LDS probe Spray gun 3 masses magnet F110 Options and accessories ltem Description 16 LDS probe 5 meter tube 17 LDS spare needle 18 LDS spare filter set of 5 19 LDS spare metal tube 20 Spray gun 21 Spray gun spare nozzle 22 3 Masses magnet Contact customer service P N 072301 072606 068843 067838 08346