Home
samaritan® PAD Saver EVO® User Instructions
Contents
1. PAD Language is described on the next page and how to change the language of Saver EVO is available on page 21 Metronome can be enabled or disabled on your samaritan PAD except for 500P and 450P models where it is automatically enabled for the CPR assist functionality it offers This indicates the software version that is installed on your samaritan PAD 15 e Hea rtSine Saver EVO User Instructions Inventor Innovator Lifesaver To Change the PAD Language Settings 1 Click the Change Language button The PAD Language Selection window appears PAD Language Selection HeartSine Inventor Innovator Lifesaver BIT TL U OVAL a Sal VE Language Selection Digital 2 Select the language from the drop down menu the choice of languages may be limited to those that are programmed into the device and or the country in which the device was purchased 3 Click the Reprogram button You will be asked to confirm the language change Yes No 4 If you clicked Yes you must enter your password to continue 5 Check that the confirmation window shown below is displayed If this message is not displayed after the programming has completed please contact your authorized HeartSine distributor or HeartSine directly ak HeartSine HeartSine Inventor Innovator Lifesaver Speech Prompts Successfully Changed Download the Spanish American PAD user manual from the Heartsine Technologies Ltd website O
2. HeartSine Saver EVO User Instructions inventor Innovator Lifesaver Run Diagnostic Tests This button allows the user to run some basic tests to check that sa their samaritan PAD is functioning The list of available diagnostic tests is accessed from a drop down menu as shown below Saver EVO PAD Diagnostic Tests window E Saver EVO140 HeartSine Inventor Innovator Lifesaver Pads placement icon Shock key icon Stand dear icon CPR icon Audible beep On Off key operation Capacitor charge up Welcome to Saver EVO 14 HeartSine Saver EVO User Instructions inventor Innovator Lifesaver dOh Edit PAD Configuration This allows you to view and modify various features of the configuration on your samaritan PAD device Saver EVO Edit PAD Configuration Window Memory Capacity Speaker Volume Current Language Metronome Settings Software Version HeartSine Inventor Innovator Lifesaver Memory Capacity Z1Y Speaker Volume Current Language US English Metronome Settings Enable Disable Software Version 6 0 0 The amount of memory that has been used on your samaritan PAD If your memory is full you should save page 8 and or erase the events on your samaritan PAD page 11 View your current speaker volume and set your speaker volume to low medium high or maximum Displays the current language programmed onto your samaritan PAD How to change the
3. O start screen OSG e HeartSine Saver EVO User Instructions Inventor Innovator Lifesaver Retrieving Event Data Information from the samaritan PAD To retrieve defibrillation event data from the samaritan PAD ensure that your samaritan PAD is connected to the PC and then click the PAD Data button This will search the memory of the PAD and display all recorded events as shown below Saver EVO PAD event log list screen GS Saver EVO 1 4 0 HeartS ne Please select an event to download inventor Innovator Litesaver Date Time Duration Ai O05 Aug 2014 09 42 58 00 10 09 2 O05 Aug 2014 10 07 31 3 O05 Aug 2014 10 12 19 Welcome to Saver EVO Events shown in green text with a symbol are events that have already been saved to the PC Save All Clicking on the Save All button will cause all the events listed in the PAD event log to be saved to the PC in a single file N B Events saved by this method cannot be retrieved for viewing by this version of Saver EVO For retrieval of events from this file contact HeartSine Technologies customer support To save events individually for retrieval later please refer to the description of the Save function in Viewing Events on a samaritan PAD on page 8 5 r View Clicking on this button will cause the selected event to be displayed Erase Clicking on this button will erase a l events from the PAD event log See page 9 for more detailed advice on erasing d
4. al Number of Analyses 2 Click the Save button Total Number of Shocks 1 to save the data in a Time from power on to first shock 00 00 24 984 PDF document at a location of your choice Save Close If you are using a samaritan PAD 450P the following additional CPR compression data will also displayed No CPR 0 00 Percentage of CPR rate equal to 0 CPM Slow CPR 49 79 Percentage of CPR rate below 100 CPM Good CPR Speed 49 38 Percentage of CPR rate between 100 and 120 CPM Fast CPR 0 83 Percentage of CPR rate above 120 CPM Average CPM rate calculated for CPR episode Percentage of time CPR was administered when advised Average Rate CPM 95 77 Compression Fraction 100 00 CPM refers to Compression per Minute 10 6 HeartSine Saver EVO User Instructions inventor Innovator Lifesaver Erasing Memory on a samaritan PAD AE To erase the memory of a samaritan PAD click the Erase button As acaution Saver EVO will ask you to enter your password prior to the memory being erased When the correct password is entered the memory of the samaritan PAD will be erased For information on how to change the default password please see Changing the Password and Language in Saver EVO on page 21 Caution HeartSine recommend that you download the complete PAD memory prior to erasing it This information should be stored safely for future reference It is recommended t
5. are download page Yes No Check for Automatic Updates Choose Yes or No accordingly if you want to download a newer version of Saver EVO 23
6. ata O kk Back Click on this button to return to the Saver EVO PAD access screen e Hea rtSine Saver EVO User Instructions Inventor Innovator Lifesaver Viewing Events on a samaritan PAD To view an evenit highlight the event in the list and then click the View button This will display the defibrillation event with the ECG trace and the event log in a single viewing window The Saver EVO PAD event viewer window HeartSine a a Inventor Innovator Lifesaver Save Print Print Preview Summary Back 29 Jan 2014 10 18 08 GMT Standard Time 00 011 AED On 100 216 Adult Patient 1 01 314 Pads On 1 01 619 Pads Off 1 03 008 Pads On 1 14 323 SHOCK 1 1503 14 604 Start CPR 03 19 015 Stop CPR 3 34 016 Start CPR 39 030 Stop CPR 205 54 019 Start CPR 59 028 Stop CPR 14 029 Start CPR 19 031 Stop CPR 34 032 Start CPR 39 047 Stop CPR 112 54 036 Start CPR 14 59 044 Stop CPR 115 14 045 Start CPR 17 19 048 Stop CPR 17 29 132 AED Off You can now choose to save this displayed event to the PC by clicking the Save button Events saved in this way can be retrieved and viewed in Saver EVO This provides options to enter information to be able to identify the patient from the list of saved events on the PC The details about the patient can consist of alohanumeric characters and each field can be completed or left blank if desired H
7. eartSine Saver EVO User Instructions Inventor Innovator Lifesaver The Saver EVO PAD Patient details window Enter Patient Details lt HeartSine inventor Innovator Lifesaver Record Type Subject Surname Forename D O B Select Date Patient ID Event Date and Time 09 Aug 2009 01 17 56 GMT Daylight Time nn Tena N B Once the defibrillation event data is saved to your PC this information cannot be changed The Saver EVO event View window also provides the following buttons Print Click on this button to print the event currently displayed in the view window 4 Print Preview Clicking on this button will cause a preview of the printed data to be displayed onscreen 4 Summary Click this button to view and save details of arrhythmia and CPR episodes See page 10 H Back Click on this button to return to the Saver EVO PAD event log list screen Hea rtSine Saver EVO User Instructions Inventor Innovator Lifesaver Saver EVO event summary The summary will display the following event information for all samaritan PAD devices amp Summary L 3 Date Hearts Ne 21 May 2014 17 41 10 GMT Daylight Time Inventor Innovator Lifesaver Information Serial Number 13E00600042 Patient ID Non_cpr_stats Record ID Subject0002 Episode Summary Episode Start Time 00 00 25 857 Total Length of Episode 00 3 34 998 5 Tot
8. h the samaritan PAD 6 Saver EVO PAD access screen 6 Retrieving Event Data Information from the samaritan PAD r Saver EVO PAD event log list screen 7 Viewing Events on a samaritan PAD 8 The Saver EVO PAD event viewer window 8 The Saver EVO PAD Patient details window 9 Saver EVO event summary 10 Erasing Memory on a samaritan PAD 11 Changing samaritan PAD Settings 12 Saver EVO Change PAD Settings Window 12 saver EVO Check Set PAD Time window 13 Saver EVO PAD Diagnostic Tests window 14 Saver EVO Edit PAD Configuration Window 15 To Change the PAD Language Settings 16 Viewing Self Test Status 17 Using Saver EVO with Saved Events 18 Export and Import 19 To Export Events 19 To Import Events 19 To Delete Saved Events 20 Changing the Password and Language in Saver EVO 21 Changing the password 22 Language Selection 22 Check for Automatic Updates 23 63 HeartSine Saver EVO User Instructions inventor Innovator Lifesaver Saver EVO User Instructions Saver EVO software allows you to retrieve defibrillation event data from your samaritan PAD Saver EVO will also allow the management of event data that has been saved to the PC with respect to printing importing and exporting event data to other Saver and Saver EVO users Download from Website 1 Go to http www heartsine com support system updates and click the Download link under Saver EVO Software Run SaverEVO_setup exe to install Saver EVO This will be identifiable by
9. hat when event data is retrieved from the samaritan PAD and saved to the PC that the events should be erased from the memory of the PAD This will ensure that the complete memory is available for recording the data from another defibrillation event 11 e HeartSine Saver EVO User Instructions Inventor Innovator Lifesaver Changing samaritan PAD Settings The Change PAD settings button allows users to configure some features of the samaritan PAD as follows e Check set PAD time e Run diagnostic tests e Edit PAD Configuration Clicking the Change PAD Settings button will access the following screen Saver EVO Change PAD Settings Window HeartSine Inventor Innovator Lifesaver Check Set PAD Time a Edit PAD Configuration Welcome to Saver EVO 12 E Hea rtSine Saver EVO User Instructions Inventor Innovator Lifesaver e Check Set PAD Time set the time on the samaritan PAD to the same time as the PC to which it is connected Saver EVO Check Set PAD Time window Manage PAD Clock HeartSine Inventor Innovator Lifesaver Local Time Time hh mm ss 16 74 46 Date Tue 08 Jul 14 PAD Time Time hh mm ss 15 24 52 Date Tue 08 Jul 14 Synchronize Close Clicking the Synchronise button will set the time on the samaritan PAD to the same time as the PC to which it is connected Click the Close button to return to the Saver EVO Change PAD Settings window 13 e
10. oftware screens not the language of the PAD voice prompts See page 16 for instructions on changing the PAD language Soran SEE Current Password H d rt S In Enter current password sseeeee Inventor Innovator Lifesaver Saver EVO 1 4 0 New Password Copyright Heartsine Technologies Enter new password For support please contact Confirm new password support heartsine com Language Selection UDI 72 7S5AVEREVOHO 130 55 7 1 4 08 Please Select Language English Spanish Check for Automatic Updates Enable Disable After clicking Apply the following screen will appear and Saver EVO will have to be restarted for the language changes to take effect ahe Information amp HeartSine Inventor Innovator Lifesaver Saver EVO will have to be restarted for language changes to take effect Would you like to dose Saver EVO now 22 HeartSine Saver EVO User Instructions Inventor Innovator Lifesaver Check for Automatic Updates If the Enable box is selected then each time Saver EVO is started it will attempt to connect with the HeartSine website to check for more recent versions of Saver EVO If an update is available you will be prompted with the following window when Saver EVO is launched amp Information HeartSine Inventor Innovator Lifesaver A more recent version of SaverEVOQ is now available Do you want to be directed to the HeartSine Technologies Ltd Softw
11. ptionally you can also download the User Manual 6 Also check that the language has been set on the PAD by inserting a Pad Pak turning the device on and listening to the first messages e Adult patient Child patient e Call for medical assistance 7 When you are sure that the correct language has been installed on your device you may then switch the device off 16 6 HeartSine Saver EVO User Instructions inventor Innovator Lifesaver Viewing Self Test Status Click the Self Test Status button to view details of self tests that have run on the samaritan PAD The following details are displayed e The Date and Time of each self test e The Result of each self test This may be o Passed o Warning o Failed e The Temperature of the device at the time of the self test e A Description of the Result of the self test This may be o For Passed results Self Test Passed o For Warning results Temperature Fault Configuration Fault RTC Fault Undefined Fault Code o For Failed results Software Integrity Fault Memory Fault Capacitor Fault If any Failed results are logged for your device you should contact HeartSine support at support heartsine com Click the Save button to save the data in a PDF document at a location of your choice 17 Hea rtSine Saver EVO User Instructions Inventor Innovator Lifesaver Using Saver EVO with Saved Event
12. s The Manage Saved Events button on the main screen of Saver EVO allows users to d review event data that has been downloaded from the samaritan PAD and saved to the computer Clicking this button will open a window as follows o sevo TTT e HeartSine e inventor Innovator Lifesaver Record Name PatientID Date Patient Name Assunto0001 06 Mar 2014 14 04 3 H008 03 Aug 2009 03 01 1 Assunto0002 05 Aug 2014 09 42 5 Saver EVO Welcome to Saver EVO Highlighting an event in the list and clicking View will open the saved event to view the ECG data and event log Any saved patient details will be displayed across the top of the screen and an ordered list of sub events will be displayed on the left of the screen The sub events available vary according to the model of samaritan PAD being used HeartSine id Inventor Innovator Lifesaver Print Print Preview Summary Back 31 Jul 2014 20 13 47 GMT Daylight Time 00 005 AED On 00 246 Adult Patient 31 753 Pads On 49 554 SHOCK 1 1503 49 783 Start CPR 09 014 Good Speed 21 014 Good Speed 33 015 Good Speed 45 015 Good Speed 57 016 Good Speed 09 016 Good Speed 21 017 Good Speed 33 017 Good Speed 45 018 Good Speed 54 018 Stop CPR 09 013 Start CPR 10 859 AED Off f bl 5 amp 5 l l 5 amp 5 BB HS 18 63 HeartSine Saver EVO User In
13. samaritan PAD Saver EVO User Instructions Saver EVO 63 HeartSine Saver EVO User Instructions inventor Innovator Lifesaver The information in this document is subject to change without notice and does not represent a commitment on behalf of HeartSine Technologies Inc No part of this manual may be reproduced or transmitted in any form or by any means electrical or mechanical including photocopying and recording for any purpose without the express written permission of HeartSine Technologies Inc Copyright 2015 HeartSine Technologies Ltd All rights reserved samaritan is a registered trade mark of HeartSine Technologies Ltd Saver EVO and SCOPE are registered trademarks of HeartSine Technologies Inc All other trademarks and registered trademarks are the property of their respective owners Saver EVO 2015 C HeartSine Technoloqies Canberra House 203 Airport Road West Belfast BT3 9ED Northern Ireland Telephone 44 28 9093 9400 e Fax 44 28 9093 9401 HeartSine Technologies Inc 121 Friends Lane Suite 400 Newtown PA 18940 Toll Free 1 866 478 7463 e Tel 1 215 860 8100 e Fax 1 215 860 8192 Website www heartsine com e E mail support heartsine com English H013 001 400 2 63 HeartSine Saver EVO User Instructions inventor Innovator Lifesaver Contents Saver EVO User Instructions 4 Download from Website 4 Quick Instructions 4 Using Saver EVO 5 Using Saver EVO wit
14. samaritan PAD Opens the menu for functions specific to the PAD See page 6 for details Manage Saved Events Opens the menu for functions specific to events saved to the PC See Page 18 for details Exit Saver EVO Exits the Saver EVO software Change Saver EVO Settings Change the Saver EVO password and select language options Also allows the user to view Saver EVO revision information and Saver EVO s UDI ila H e E HeartSine Saver EVO User Instructions inventor Innovator Lifesaver Using Saver EVO with the samaritan PAD To use Saver EVO with the samaritan PAD you need to have the USB data cable that was supplied with your samaritan PAD Use the USB data cable to connect the samaritan PAD to the USB port on your PC Open the Saver EVO software Clicking the samaritan PAD button will open the window shown below Saver EVO PAD access screen HeartSine Inventor Innovator Lifesaver Change PAD Settings E self Test Status Q PAD Data Click on this button to retrieve event data from the PAD See page 6 for details Change PAD Settings Click on this button to run diagnostic tests or change the configuration volume language or time on the samaritan PAD See page 9 for details Self Test Status Click this button to view and save details of Self Tests that have run on the samaritan PAD See page 17 for details Back Click on this button to return to the Saver EV
15. structions inventor Innovator Lifesaver Export and Import To send or receive copies of saved events use the Export and Import functions of Saver EVO To Export Events A 1 From the Manage Saved events screen highlight the event s that you wish to export and then click the Export button 2 The PC will ask for a location on the computer where you wish to save the export file and also lets you choose a name for the exported file Click the Save button to save the file in the chosen location From the Manage Saved events screen highlight the event s that you wish to export and then click the Export button 5 The PC will ask for a location on the computer where you wish to save the export file and also lets you choose a name for the exported file 6 Click the Save button to save the file in the chosen location It is then possible to attach this file to an e mail using your normal e mail software and send the exported file to another user who has the Saver EVO software Event data from Saver EVO can only be viewed by another Saver EVO user To Import Events To import events and view these on your PC the file that has been sent to you from another user of Saver or Saver EVO needs to be saved to a location on your PC 1 From the Saver EVO Manage Saved Events screen click the Import button 2 Browse to the location where the file was saved and click Open 3 The records will be added to the list of Saver EVO files on
16. tan FAD Welcome to Saver EVO This will open the following window displaying the Saver EVO version and UDI Saver EVO UDI The Saver EVO Unique Device Identification UDI will be displayed in this window HeartSine Inventor Innovator Lifesaver Saver EVO 1 4 0 Copyright Heartsine Technologies For support please contact support heartsine com UDI 72 7SAVEREVOHO130 55 71 4 08 Current Password Enter current password New Password Enter new password Confirm new password Apply Language Selection Please Select Language English Check for Automatic Updates Enable Disable 21 HeartSine Saver EVO User Instructions inventor Innovator Lifesaver Changing the password Enter the current password and click Confirm This will allow you to enter and confirm your new password Click Apply to make the change A confirmation that the password was successfully changed will appear Click OK and again click OK in the Change Password window to exit this function Note that reinstalling Saver EVO will reset the password to the default in the software This action however will not remove any event data that has previously been saved Language Selection To choose the language of the Saver EVO software select the language from the drop down list and click Apply Note that this screen makes it possible to change the language used in the Saver EVO s
17. this icon Follow the on screen installation instructions Read the user licence agreement In order to install this software you must accept the terms of this agreement To accept the terms select accept the agreement and click the Next button Saver EVO will be installed in its default location with its standard configuration Click the Next button to accept These default settings should not be altered unless you are an experienced computer user If asked about device driver certification click the Continue anyway button On the final installation screen click Finish A new Saver EVO icon will be created on the desktop Quick Instructions To retrieve recorded defibrillation event data from the samaritan PAD G le le ay 1 Connect the relevant device to the PC 2 Open the Saver EVO software 3 Click the button labelled samaritan PAD 4 Click the button labelled PAD Data 5 Click the required recorded event in the list 6 Click the button labelled View 7 Click the button labelled Save and enter the patient information as required to save the event on the PC 8 If required click the button labelled Print to print the event e HeartSine Saver EVO User Instructions inventor Innovator Lifesaver Using Saver EVO The Saver EVO start screen is shown below HeartSine Inventor Innovator Lifesaver samaritan PAD Welcome to Saver EVO
18. your computer These can then be viewed by highlighting the event in the list and clicking the View buiton 19 6 HeartSine Saver EVO User Instructions inventor Innovator Lifesaver To Delete Saved Events Saved events can be deleted from your computer memory by Saver EVO Events can be selected by highlighting the event s to be deleted multiple events can be selected xX When you have confirmed that only the events you wish to delete have been selected click the Delete button As a precaution Saver EVO will ask you to enter your password prior to the events being deleted When the correct password is entered the selected events will be deleted from your computer Caution Please ensure that only the events you want to delete have been selected prior to deleting Once deleted from your computer s memory events cannot be regenerated and all information will be lost 20 HeartSine Saver EVO User Instructions inventor Innovator Lifesaver Changing the Password and Language in Saver EVO Saver EVO has some functions that can only be changed or completed by entering a password This allows administrators of the software to restrict access to certain capabilities to authorised users Saver EVO is shipped with the password set to the default oassword JOR To change this password to one of your own choice click the button on the Saver EVO start screen HeartSine Inventor Innovator Lifesaver samari
Download Pdf Manuals