1. Remove the mirror cover it is held in place with magnets 2 Loosen the clamping screws on the laser module and then rotate the laser to position beam vertically Installation and set up manual 208684 57 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 5 Aligning the unit ORTHOPANTOMOGRAPH OP30 Focal trough 1 Remove the lower cover from the lower shelf 6 screws iy Loosen the appropriate screw s and rotate the light and or adjust it in the backwards forewards direction until it is correctly positioned Rotation adjustment s Back front adjustment Check the positions of the lights again and then replace all the covers and adjusting knob 58 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 5 Aligning the unit 5 5 Calibrating the CCD sensor NOTE The CCD sensor is calibrated at the factory and thus does not normally need to be calibrated again However if the error message Please calibrate the device appe
2. Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 3 Installing the unit Express Share Express Share allows the unit to be used by several PCs within a network APC must reserve the unit before it can be used 1 PC In Tools OP30 Setup window select the Settings page 2 PC Click the Use Express Share check box to select it and then from the drop down list select a unique Workstation identifier number between 1 and 8 for the PC being configured Additional workstation information for example user name location etc can be entered into the field next to the work station identifier number NOTE Each PC in the system MUST have a different Workstation identifier Express Share Use Express Share Workstation identifier a ROOM Device Autorelease timeout seconds 300 The Device Autorelease timeout is the length of time that the unit will remain reserved and usable by a PC After the set time elapses the PC automatically releases the unit so that it can be used by another PC in the network the unit can be reserved in advance from another PC The default setting is 300 seconds This can be changed by keying in a new value Installation and set up manual 208684 39 Reviewed
3. However if problems occur due to firewalls etc use the Manual alternative instead Installation and set up manual 208684 29 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 3 Installing the unit ORTHOPANTOMOGRAPH OP30 The Remote alternative is explained as part of the section Connection using IP based connection method The Manual alternative is explained in section Manually changing the IP address IMPORTANT NOTE The unit can have any IP address but the unit and PC must always be in the same subnet For example IP for PC 194 9 227 250 and IP for the unit 194 9 227 251 and the subnet mask is fixed to 255 255 255 0 30 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 ORTHOPANTOMOGRAPH OP30 3 Installing the unit Connection using the Direct Connection method 1 Standalone configuration Connect one end of the Ethernet cable to the unit the connector at the rear of the column and the other end to the PC Network configuration Connect one end of the Ethernet cable to the unit the connector at the rear of the column and the other end to the network 2 Switch the Unit on 3 PC Start the PC Then op
4. 3 Installing the unit ORTHOPANTOMOGRAPH OP30 5 Stand alone configuration PC Key in the desired IP address of the unit for example 194 9 227 250 into the IP address field Network configuration PC Key in the new IP address obtained from the network administrator into the IP address field IP based Device IP address Enable changing IP address 194 9 297 250 Send to Device IMPORTANT NOTE The unit can have any IP address but the unit and PC must always be in the same subnet The subnet mask of the unit is fixed to 255 255 255 0 NOTE With the network configuration the default IP address of the unit must be changed to the IP address obtained from your local IP administrator With standalone configuration you can avoid changing the IP address of the unit by setting the IP address of the PC to the same sub net as the default value of the unit After changing the PC IP address in this way and entering the default or current IP address of the unit into the IP address field as instructed above you can proceed to step 12 to finish the configuration 6 Touch the service key on the main control panel of the unit 34 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All righ
5. 6 Support the hinged column section before loosening the screws that hold it to prevent it falling and causing injury or damage Loosen the two nuts M10 inside the column that hold the hinged lower column section in the vertical transportation position Lower the hinged column section down so that it is horizontal Secure it in position with the two screws supplied M10 x 16 the fixing holes are on the underside rear of the column Retighten the two nuts M10 inside the column 14 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 3 Installing the unit 7 Lift the top of the column and slide the packing piece that supported it underneath so that you can access the underside of the column Attach the wall mounting assembly to the back of the column with the screws nuts and washers supplied M10 NOTE The 4 x M10 washers must be positioned BETWEEN the wall mounting assembly and the rear of the column Top of column The wall mounting assembly is adjustable so that the unit can be adjusted in the backwards forwards direction Make sure that the wall mounting assembly is in the mid position of its adjustment range To adjust loosen the two
6. Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 1 Introduction ORTHOPANTOMOGRAPH OP30 Exposure warning light Midsagittal light Mirror Frankfort light and light positioning knob Focal trough positioning knob Patient support Focal trough light Patient support handles Head support OCONDOBRWNDN 4 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 1 Introduction Unit controls A Side control panel Lights key switches the patient positioning lights on and off Up key drives the unit up Down key drives the unit down Return key drive the unit to the patient in out position PIO AUN B Main control panel Program selection keys P1 Standard Pan P2 Pediatric Pan P3 TMJ BW Bitewing Patient size selection keys Child Juvenile Adult Large adult 7 Manual M mode selection key 8 kV and mA selection keys manual mode only 9 Test key operates the unit without x rays 10 Exposure values 11 Service key 12 Patient Name key visible only if patient name display is supported by the d
7. interface panel in position and then remove the screws 4 that hold it to the inside of the hinged column section Carefully remove the main power and interface panel from inside of the hinged column section and place it next to the unit Make sure that none of the cables that are connected to it are damaged or accidently Hinged column section 7 Dn Las gt ae DN geen DETA Sate al ramien re M Screwsx4 64 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 6 Using the exhibition stand 4 Remove the hinged column section from the bottom of the column 2 x M10 screws and nuts Hinged column section 5 Position the assembled exhibition stand near the bottom of the column Place some protective material on the floor and then turn the exhibition stand over onto its back 6 Lift the bottom of the column until the fixing holes on the sides of the column are level with the fixing holes on the sides of the exhibition stand Place some packing material under the column to hold it in position 7 Slide the exhibition stand towards the bottom of the column until the fixing holes at the sides of the exhibition stand line up with the fixing hol
8. of these connection methods allow the unit to be used in either a Standalone or a Network Configuration Direct Connection Direct Connection is a connection method based on the unit serial number for establishing a connection between the unit and the PC Direct connection is not based on IP protocol but instead utilizes a proprietary Ethernet protocol NOTE In some configurations the PC firewall or other network devices may prevent direct connection from operating Direct connection requires traffic to be allowed between the PC and unit through the network devices and the PC firewall in the inbound outbound direction for Ethernet protocol ID Ox6000 If there are difficulties with the direct connection use the IP based connection method instead IP based connection IP based connection uses IP Protocol to establish a connection between the unit and the PC It requires the IP addresses of both the unit and the PC to be configured correctly IMPORTANT NOTE The default IP address of the unit is 194 9 227 251 and the subnet mask is fixed to 255 255 255 0 The unit offers two alternatives for configuring its IP address Remote and Manual With the Remote alternative the unit receives its P address from the PC With the Manual alternative the user sets the IP address of the unit using the Graphical User Interface GUI of the unit It is recommended that Remote alternative is used when configuring the IP address of the unit
9. supply before making the adjustment e Ifthe unit is to be used with a 3rd party imaging application software not produced or supplied by Instrumentarium Dental the 3rd party imaging software must comply with all applicable local laws and regulations on patient information software This includes for example the Medical Device Directive 93 42 EEC and or the FDA During unit operation e Read and familiarize yourself with the warnings and precautions in the unit User Manual e When checking the alignment of the radiation beam protect yourself from radiation e After the unit has been installed and set up advise the people who will operate the unit to familiarize themselves with the warnings and precautions in the User Manual 2 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 1 Introduction 1 4 Unit description Main parts Column Upper shelf Rotating unit Emergency stop button Press to stop rotate to release On off switch rear of column PC with MDD compliant dental imaging software Motorized carriage NOOB WON Installation and set up manual 208684 3 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen
10. the unit back to the original position and then attach the unit to the floor with the fixing hardware 4 Check once more that the unit is vertical and then tighten all the fixing screws and wall bracket fastenings 3 4 Installing the guide pin 1 From the upper shelf unscrew the guide pin and two holding screws from the base plate DO NOT remove the base plate or loosen the nuts that attach it to the unit SESEEREA Guide pin in SU transportation Installation and set up manual 208684 19 Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 3 Installing the unit ORTHOPANTOMOGRAPH OP30 2 Flip the guide pin assembly around so the plain section of the guide pin is downwards and the nut is uppermost Then insert the guide pin assembly into the hole in the base plate Plain s K i downwards IMPORTANT NOTE If the guide pin does not fully enter the hole in the base plate the top of bearing is NOT level with the top of the base plate the rotating unit must be moved slightly To do this loosen the transportation screws see step 4 for information on the transportat
11. 0 Touch the OK key to accept the IP address and then the Arrow key to return to the main screen PC Click the OK button to close the settings window NOTE If the remote IP address setting fails use the MANUAL method See section Manually changing the IP address 36 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 3 Installing the unit Manually changing the IP address The IP address of the unit can be changed manually NOTE Always set the IP address of the PC before changing the IP address of the unit 1 Touch the service key on the main control panel of the unit 2 The configuration screen will appear It shows i S the unit serial number and the firmware Unit OP30 2 version numbers SN 101300284 FW 2 05 Core 37 Graphics 3 Also there are three function keys IP Settings Installation and QA Select IP Settings NOTE see section 4 1 Checking alignment of the beam and CCD sensor for information about Installation see section 4 4 Quality Assurance test for information about QA 3 The IP configuration screen with two configuration mode function keys Manual and Current P address Remote will appear 194 9
12. 2 Make sure that all of the cables are connected Front of exhibition stand 10 CAUTION HEAVY OBJECT A minimum of two persons are required for the following task Lift the unit and exhibition stand to the vertical position NOTE When lifting the unit to the vertical position hold the unit by the column NOT the rotating unit CAUTION Take steps to ensure that the unit and exhibition stand do not slip when being lifted Installation and set up manual 208684 67 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 6 Using the exhibition stand ORTHOPANTOMOGRAPH OP30 11 Remove all the packaging material from the C arm and then remove the cover from the upper shelf 4 x M4 screws we ig Place a spirit level first widthwise and thenlengthwise on the upper shelf to check that the unit is level If the unit is not level level it by turning the levelling feet in the base of the exhibition stand Levelling feet x 4 Insert cover plugs after the unit has been levelled 68 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out
13. 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 6 Using the exhibition stand 12 OPTIONAL STEP If the unit has to be fixed to the wall follow the instructions in section 3 2 Attaching the unit to the wall step 2 onwards 13 Attach the covers to the exhibition stand Side covers 4 x M4x12 washers Front cover 4 x M4x8 14 Now follow the instructions in section 3 4 Installing the guide pin to complete the installation Installation and set up manual 208684 69 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08
14. 227 251 Choose IP Configuration Mode Select Manual Installation and set up manual 208684 37 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 3 Installing the unit ORTHOPANTOMOGRAPH OP30 4 The Manual IP address configuration screen m will appear IP Adress 194 9 227 253 Key in the new IP address Has When the new IP address has been keyed in 4 5 6 fe q the OK tick key will become active 7 8 9 o Touch the OK tick key to accept the change 6 Z and return to the main screen v4 Touch Arrow key to discard the change and f i return to the main screen 5 PC Start the PC Then open CLINIVIEW software or the dental imaging software that was installed previously and select Settings CLINIVIEW software Select Tools OP30 Settings and then the Settings page 6 PC In the Device Connection section click the IP based radio button to select it Then click the Enable changing IP address check box to activate the IP address field IP based Device IP address Enable changing IP address 194 9 227 253 7 PC Key in the new IP address into the IP address field 8 PC Click the OK button to close the settings window 38 Installation and set up manual 208684
15. Checking the alignment ORTHOPANTOMOGRAPH OP30 6 Protect yourself from radiation Press and hold down the exposure switch to take an exposure 7 PC Use the measuring tools in the dental imaging software to examine the dimensions of the pins and ball that appear on the image Magnify the image so that you can accurately measure the dimensions If the image geometry is correct the center ball must be round with a difference of no more than 0 2 mm between the horizontal diameter and the vertical diameter Also the distances from the side of the center ball to the sides of the pins must be equal with a difference of no more than 1 0 mm between the two distances Horizontal diameter Vertical diameter If the center ball is NOT ROUND and or the distances between the pins and the center ball are NOT WITHIN THE TOLERANCE then the rotating unit RU or the chin rest must be adjusted see 5 3 Adjusting the image geometry IMPORTANT NOTE In some cases you may have to adjust the position of both the rotating unit AND the chin rest 44 22 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 ORTHOPANTOMOGRAPH OP30 4 Checking the alignment 4 3 Checking the positioning lights 1 Position the chin rest to the O position Te ZAN 2 Place the laser alignment tool supplied in the N chin rest 3 Press t
16. Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 3 Installing the unit ORTHOPANTOMOGRAPH OP30 3 PC Click OK to connect the PC to the unit NOTE An automatic technique will automatically locate the unit within the local area network and connect the PC 4 PC Repeat the above process for all the other PCs in the network Make sure that you define the IP address of the unit in each PC and that the workstation identifier selected is not used in more than one PC 5 Check the installation by starting image capture using the imaging software If the Use Express Share was selected the Workstation identifier of the PC 1 8 being used will appear on the unit display 40 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 4 Checking the alignment 4 Checking the alignment and finishing the installation 4 1 Checking the alignment of the beam and the CCD sensor 1 PC Prepare the unit for an exposure NOTE If the dental imaging software prompts an error message Pl
17. See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ENGLISH ORTHOPANTOMOGAPH OP30 Digital Panoramic X ray Unit Installation and Set up Manual 208684 rev 4 2014 11 ae For type OP30 2 Ser No 101300001 and later Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 4 D508777 Savinen Antti Juhani 2015 03 11 17 52 Approved Ukkonen Juha Pekka 2015 03 13 13 08 Reviewed 60 07 71 0 Y0 9 LOZ NO Pau JUBWNDOP Sly JO SN EIS 94 SUILUA ap 0 W JS S Wid 99S pamasa syfu ily AO dnog x3qoled Ag S LOZ 1ubuAdoy See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved Copyright Document code 208684 rev 4 2014 11 Copyright 2014 by PaloDEx Group Oy All rights reserved Documentation trademark and the software are copyrighted with all rights reserved Under the copyright laws the documentation may not be copied photocopied reproduced translated or reduced to any electronic medium or machine readable form in whole or part without the prior written permission of Instrumentarium Dental This is the original English language version of the manual This installation manual is for units with serial number Ser No 101300001 and later Instrumenta
18. age on the left hand side using two 2 screws If the user does not want to have the exposure switch positioned on the unit attach the exposure switch holder within the range of the exposure switch spiral cable to where requested Make sure that the exposure switch is positioned so that the user can protect them self from radiation and can see the patient during an exposure 2 Connect the exposure switch cable to the connector at the bottom rear of the column and clamp the cable in place using one of the cable clamps Place the exposure switch on the exposure switch holder Installation and set up manual 208684 23 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 Copyright 2015 by PaloDEx Group Oy All rights reserved See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 3 Installing the unit ORTHOPANTOMOGRAPH OP30 3 8 Installing the remote exposure switch A wall mountable remote exposure switch with integrated exposure warning light is supplied with the unit If a remote exposure switch needs to be used install it as follows 1 Connect one end of the remote exposure switch extension cable to connector J2 inside the remote exposure switch 2 Attach the remote exposure switch to the wall in the required position 2 fixing screws Make sure that the cable from J2 is positioned in the small cut out in the bottom o
19. an exposure Installation and set up manual 208684 25 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 3 Installing the unit ORTHOPANTOMOGRAPH OP30 3 9 Connecting the unit to the power supply The unit must be connected to its own dedicated power supply e 100 to 120 VAC or e 220 to 240 VAC No other equipment should be connected to the same power supply Extension cables MUST NOT be used to connect the unit to the power supply 1 The power supply must be equipped with a circuit breaker e 100 120 VAC min 16 A 250 VAC e 220 240 VAC min 10 A 250 VAC 2 Connect one end of the power lead to the back of the unit and the other end to the power supply Observe any national and local electrical requirements with regards the connection of dental x ray equipment to a power supply 26 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 3 Installing the unit 3 10 Preparing the PC NOTE The PC and any other external device s connected to the s
20. andard Standard Standard column leg column leg column leg 965 1740 mm 2290 15 mm 2275 15mm 38 68 5 90 2 0 6 89 6 0 6 Short Short Short column leg column leg column leg 875 1650 mm 2200 15 mm 2185 15 mm 34 5 65 86 6 0 6 86 0 6 at 225 mm 8 9 1060 mm 41 7 830 mm 32 7 790 mm 37 790 mm The height of the unit on the exhibition 31 stand is the same as the height of the unit with the standard column leg Exhibition stand 2340 mm 92 1 10 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 2 Pre installation requirements 2 6 Fixing hardware and installation and setup tools The following tools and hardware are required to install and set up the unit NOTE Fixing hardware and tools are NOT supplied with the unit Fixing hardware for the unit WALL attachment requires 10mm 3 8 diameter fixing hardware two sets FLOOR attachment requires 8mm 5 16 diameter fixing hardware two sets NOTE DO NOT use smaller diameter fixing hardware The type and length of hardware to be used will depend on the wall material and floor material to which the unit i
21. ani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 3 Installing the unit 3 Installing the unit 3 1 Preparing the unit for installation 1 Transport the boxed unit to the location where the unit is to be installed 2 Remove the straps that hold the box to the pallet Ay 3 Lift off the top of the box and then the sides Installation and set up manual 208684 13 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 3 Installing the unit ORTHOPANTOMOGRAPH OP30 4 Remove all the accessories the accessory box and as much packaging material as possible NOTE DO NOT remove the packaging material that protects the upper shelf and the rotating unit NOTE If the unit is to be attached to the exhibition stand see section 6 The exhibition stand and the separate instructions supplied with the exhibition stand 5 Fold down the end of the box that is nearest the hinged section of the column and then slide the unit over the edge of the pallet so that you can access the underside of the hinged section of the column
22. ars on startup calibrate the CCD sensor as described below Also if you adjusted the alignment of the sensor described in section 5 1 and or adjusted the alignment of the beam described in section 5 2 during installation then the CCD sensor must be calibrated 1 PC Prepare the unit for an exposure 2 Remove the chin rest and make sure that there is nothing between the sensor and the tubehead 3 Touch the service key on the main control panel of the unit to open the configuration screen 4 Select Installation Installation 5 The personal identification number PIN screen will appear eee Key in 1204 and touch the OK tick key Image Calibration Then select Image Calibration Installation and set up manual 208684 59 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 5 Aligning the unit ORTHOPANTOMOGRAPH OP30 6 Protect yourself from radiation Press and hold down the exposure switch to take an exposure 7 PC Use the dental imaging software to examine the image It should be evenly grey all over with no granularity nor horizontal lines stripes If the image is not evenly gray repeat the calibration until it is If artefacts appear on the image make sure that there is nothing between the se
23. ating unit RU must be must be adjusted 1 Loosen the two nuts that hold the guide pin assembly in place Turn the adjusting nut counter clockwise to move the rotating unit towards the column and clockwise to move it away from the column A move RU TELE EEELELELE 444 towards column f ty B move RU away from column ois Mi is a it r 1 E gt en TIN 4 r i RN ns A ig ki assembly Iii E Guide pin A Move rotating unit towards the column B Move rotating unit away from the column 54 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 5 Aligning the unit 2 Re tighten the guide pin assembly nuts after the adjustment 3 Take another panoramic exposure of the Ball and Pin Phantom to confirm that the center ball is round Also verify that the distances from center ball to left and right pins are equal if not proceed to the next section Distances between left and right pins not equal Distances between center ball and left and right pins not equal If the center ball is round but the distances between the pins and the ball are DIFFERENT then the chin rest must be must be adjusted 1 Loosen the two screws tha
24. cetuas iare E EEEa 23 3 8 Installing the remote exposure SWItCN cccccceec eee eeeeeeeeeees 24 3 9 Connecting the unit to the power SUDDIY cccceeeeeee eee eees 26 3 10 Preparing the PC os cecewnesionesecaseesnaceensseseszencseseceseesvensieesecadaceaeeees 2 3 11 Connecting the unit to the PC cc cecceccseecsseseeeseeeeeeseees 28 Installation and set up manual 208684 V Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved Contents ORTHOPANTOMOGRAPH OP30 4 Checking the alignment and finishing the installation 41 4 1 Checking the alignment of the beam and the CCD sensor 41 4 2 Checking the image geometry Ball and Pin Phantom test 43 4 3 Checking the positioning lights cc cceccsecceeeseeeeeeseeeeeeeaes 45 4 4 Quality Assurance TEST cece ceeceeeceeeeeeceeeseeeseeeeeeseeeaeeeaes 46 4 5 Completing the Installation ccccccceccsseceeeceeeceeceeeeeeeseeeaes 50 ANGNIN the UNIT sssrinin iinan a aaa 51 5 1 Aligning the CCD S NSOM ccccccceecsseceeecuecaeecaeecueeeeesaeenaees 51 5 2 Adjusting the beam cccecceccseceeeceecaeeseeceesaeeceecaesaseeeesensaees 52 5 3 Adjusting the image geometry ccccecsecceeceesseeseeeaees
25. cial steps regarding EMC need to be taken when installing the unit For more information refer to the EMC declaration in the User Manual e The unit must be installed at least 1 85 m 73 away from any non medical electrical or electromechanical equipment e The unit must be connected to its own dedicated power supply e 100 to 120 VAC or e 220 to 240 VAC e Extension cables MUST NOT be used with the unit 8 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 2 Pre installation requirements 2 2 The PC e The PC to be used with the unit must be installed in a location that meets all local and national safety requirements with regards the connection of a PC to an x ray device e The connection of the unit to the PC must meet EN60601 1 requirements The PC must be positioned at least 1 85 m 73 away from the unit e The use of ACCESSORY equipment not complying with the equivalent safety requirements of this equipment may lead to a reduced level of safety of the resulting system Consideration relating to the choice shall include use of the accessory in the PATIENT VICINITY evidence that the safety certification of the ACCESSORY has been performed i
26. ease calibrate the device on startup perform CCD re calibration as described in section 5 5 Calibrating the CCD sensor Remove the chin rest Touch the service key on the main control panel of the unit to open the configuration screen Select Installation The PIN number screen will appear This is z s a security feature to prevent the user from Enter PIN Code accidentally accessing the installation service commands XXXX a23 L Key in 1204 and touch the OK tick key Installation and set up manual 208684 41 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 4 Checking the alignment ORTHOPANTOMOGRAPH OP30 6 The service screen will appear Still Mode Select Still Mode 7 Protect yourself from radiation Press and hold down the exposure switch to take an exposure NOTE The rotating unit doesn t move while taking exposures in Still Mode 8 PC Use the dental imaging software to examine the image that appears If the beam and CCD sensor are correctly positioned the image will show a vertical black line the beam in the middle of a vertical light gray area the CCD aperture Correct Both the beam black line and the CCD aperture light gray area must be vertical and the b
27. ed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 3 Installing the unit Connection using the IP based method 1 Standalone configuration Connect one end of the Ethernet cable to the unit the connector at the rear of the column and the other end to the PC Network configuration Connect one end of the Ethernet cable to the unit the connector at the rear of the column and the other end to the network 2 Switch the Unit on 3 PC Start the PC Then open CLINIVIEW software or the dental imaging software that was installed previously and select Settings CLINIVIEW software Select Tools OP30 Settings and then the Settings tab 4 PC In the Device Connection section click the IP based radio button to select it and then click the Enable changing IP address check box to activate the IP address field OP30 Setup IP based Device IP address Enable changing IP address send to Device Installation and set up manual 208684 33 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved
28. eeeeeaees 54 5 4 Aligning the positioning lights cccceccecceecseeseeeeeeeeeeeeeeees 57 5 5 Calibrating the CCD Sensor ccccccssccseeceeeceeseecueeeeesaeesaees 59 5 6 Advanced Quality Assurance and Advanced Still Mode 60 6 The exhibition SCAN hos tise ne cipe eeee ceteeeeece eae csemeeweenceeecnenees 64 6 1 Attaching the unit to the exhibition stand cccccecceeeeeeees 64 VI Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 1 Introduction 1 Introduction 1 1 About this manual This manual explains how to install and set up the ORTHOPANOTMOGRAPH OP30 digital panoramic dental x ray unit the unit You will need to operate the unit during the alignment checking procedure Make sure that you know how to operate the unit before starting to install and set up the unit It is also essential that person installing and setting up the unit should have attended the manufacturer s service training course 1 2 Associated documentation The ORTHOPANOTMOGRAPH OP30 Digital Panoramic X ray Unit User manual The ORTHOPANOTMOGRAPH OP30 Digital Panoramic X ray Unit Service manual not supplied in the delivery The insta
29. en CLINIVIEW software or the dental imaging software that was installed previously and select Settings CLINIVIEW software Select Tools OP30 Settings and then the Settings tab 4 PC In the Device Connection section click the Direct Connection radio button and then key in the complete device Serial Number it is on the type label into the Device serial number field OP30 Setup d lS Device Connection Direct Connection Direct Connection settings Device seral number 100120719 Computer network connection Broadcom Netxtreme 5 4x Gigabit Controller Select the correct Nertwork Interface Card NIC from the Computer network connection drop down list Installation and set up manual 208684 31 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 3 Installing the unit ORTHOPANTOMOGRAPH OP30 NOTE If there is more than one NIC in the PC make sure that you select the one that is connected to the device Click OK to save the settings 5 The unit will automatically connect to the PC within 5 15 seconds 32 Installation and set up manual 208684 Review
30. ental imaging software 13 Dose Area Product DAP 14 Unit status indicator 15 Segmented panoramic program selection key O oOo Installation and set up manual 208684 5 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 1 Introduction ORTHOPANTOMOGRAPH OP30 1 5 Disposal of the transportation packaging The unit is transported in a corrugated cardboard box that is attached to a wooden pallet Inside the box preformed expanded polystyrene packaging material protects the unit from shock and damage Packaging material Weight kg Ib Cardboard box and lid 21 0 46 3 Wooden pallet 7 4 16 3 Expanded polystyrene 7 6 16 8 Total weigh 36 0 79 4 When disposing of the transportation packaging materials make sure that you follow all local national and international regulations regarding the correct and safe disposal and or recycling of these materials In some countries the above transportation pack aging materials may be considered to be non environmentally friendly If this is the case the transportation packaging materials must be dis posed of in accordance with all local national and international regulations regarding the disposal of non environmentally friendly materials NOTE It is advisable to keep some transpo
31. es at the sides of the unit column Then slide the main power and interface panel into the Exhibition stand Make sure that you do not damage the cables when doing this Installation and set up manual 208684 65 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 6 Using the exhibition stand ORTHOPANTOMOGRAPH OP30 Attach the top of the exhibition stand to the bottom of the unit column with the 2 x M10 screws and nuts Do not fully tighten the nuts yet 2 x M10 screws and nuts 8 While the unit and stand are still on the floor attach the 2 x M10 screws to the rear of the exhibition stand Tighten these screws and then tighten the 2 x M10 nuts on the inner sides of the unit Bottom of unit column 2 x M10 screws 3 Top of exhibition stand The 2 x M10 screws and nuts 66 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 6 Using the exhibition stand 9 Attach the Main power and interface panel to the inside rear of the exhibition base 4 x M4x1
32. f the remote exposure switch housing Also make sure that the cable does not press against the circuit board when the housing is attached to the wall _ Exposure warning light Exposure button Exposure switch connector Cable to unit from J2 Route the remote exposure switch extension cable to the rear of the unit and then secure the cable in position under one of the cable clamps after connecting the cable to the unit This will ensure that the cable cannot be accidentally disconnected 24 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 DaUe i i ly Approved Ukkonen Juha Pekka 2015 03 13 13 08 4 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 3 Installing the unit 4 Ifthe user wants to use the exposure switch with the remote exposure switch option connect the exposure switch to the appropriate connector on the remote exposure switch Exposures can be taken either with the exposure switch or with the button on the remote exposure switch 5 Attach the exposure switch holder within the extension range of exposure switch spiral cable to the requested position Exposure switch cable Make sure that when the exposure switch is used the user can protect them self from radiation and at the same time see the patient during
33. he lights key 4 Check that the lights hit the corresponding grooves on the laser alignment tool Frankfort light Focal through light Midsagittal light ne If any of the lights are not positioned correctly their positions must be adjusted See 5 4 Aligning the positioning lights Installation and set up manual 208684 45 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 4 Checking the alignment ORTHOPANTOMOGRAPH OP30 4 4 Quality Assurance test Additional Quality Assurance QA exposures can be taken after installation and periodically to ensure that the image quality remains stabile The QA exposures can be taken using special QA program To take QA images you need an additional Cu Copper filter and Digital Test Tool Prufkorper test tool that need to be ordered separately There are different thicknesses of the Cu filters select the correct one depending on the national requirements The unit provides two different QA program selections QA1 and QA2 QA1 is intended for use with 0 8mm thick Cu filter in accordance to IEC61223 3 4 QAZ2 is intended for
34. ings and precautions c cece eeccecceeeeeeceeceeeeeeseeeaeeneeseeeees 2 1 4 Unit description ee cecceceeceeceeeeeeeeeeteeseesnececeeeaeeeesaeseeseegereeees 3 1 5 Disposal of the transportation packaQINg ccsccceeeeeeeeeeeees 6 2 Pre installation reQuireMent ccccecceeceeeeeeeeeeseeeeeeneneenseeeeesaes T ZW ITNE OUNT ea E E E EE E EEN T ZZ VS Wr O A ness vasewsaneniatanautysassncnyenseneanmtineessasemaseenees 9 2 3 The dental imaging software cccccccccecceeeceecceeeeeeeeeeseeeaeeeees 9 2 4 Space FEQUIFEMENHS cccceccecceeceeceeeeeeeeteeeeeeeeceeeceseueeseneeeaeees 9 2 5 Unit and fixing hole dimensions cccccc cece eeeceeeeeeeeeeeeeeeeees 10 2 6 Fixing hardware and installation and setup tools 00 11 3 Installing the UNit ccceeeseeeecceeeeeseneeeeceeseneceseeeeenseeeseeseneeneseess 13 3 1 Preparing the unit for installation c cece cecceeeseeeeeeeeeeeeees 13 3 2 Attaching the unit to the wall ccccecceecsseceeceeeeeeseesaeenens 16 3 3 Attaching the unit to the floor ssceiees ccceercsazseesacenssnceaceyeeneceusa secees 18 3 4 Installing the guide PIN ccccccecceccceeseeceeeaeeceecaseeeeseesaeenees 19 3 5 Attaching the covers to the COIUMN ccceccecceeceeeeeeeeeeeeees 22 3 6 Checking the bite rod cccccccceccceeeeeeceeceeeeeesaeeceeeeeeeaeesaees 23 3 7 EXPOSUTO SWIC erecta saju
35. ion screws and then slide the rotating unit gently backwards or forwards until the guide pin drops into place 20 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 3 Installing the unit 3 Secure the guide pin in position with the two holding screws 4 Remove the four 4 transportation screws painted RED and four 4 packing pieces that hold the rotating unit to the upper shelf a rE EE Wage ee a e E a a screws es Holdin Now rotate the rotating unit 180 counterclockwise so that the control panel is on the left hand side the unit patient in out position Packing pieces x 4 screws x 4 Installation and set up manual 208684 21 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 3 Installing the unit ORTHOPANTOMOGRAPH OP30 Unit relocation If the unit ever needs to be moved or relocated replace the packing pieces and transportation screws and then remove the guide pin assembly and place it back into its transportation
36. ity Assurance test phases 2 5 If Advanced Still Mode exposure is to be taken prepare the unit and related measurement equipment for a still mode exposure 3 Touch the Service key on the main control panel of the unit to open the configuration screen 4 Select Installation Installation and set up manual 208684 61 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 5 Aligning the unit ORTHOPANTOMOGRAPH OP30 5 The PIN number screen will appear This is j R s3 a security feature to prevent the user from Enter PIN Code accidentally accessing the installation service AAAXK 2e commands Key in 1204 and touch the OK tick key Select Advanced QA 7 The Advanced QA program screen appears Select the needed program Advanced QA QA or Advanced Still mode Still and adjust the exposure values using the plus and minus keys o sev 66kv 96mA oe O 9 6 mA 4 10 0 10 05 54 mGycm ri 54 mGycm 62 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx G
37. kV 9 6 mA 10 0 s 54 mGycm 9 Protect yourself from radiation Press and hold down the exposure switch to take an exposure 48 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 4 Checking the alignment 10 PC Visually evaluate the result using the dental imaging software Subjects to be evaluated 1 Exposed area Smoothness of the exposed area Non exposed area surrounds the whole image 2 Line Pair Resolution minimum 2 5 LP mm must be distinguishable 3 Low contrast resolution Minimum two low contrast resolution holes must be visible 4 Overall image geometry Roundness of the center ball and equal distances to the pins see 4 2 Checking the image geometry Ball and Pin Phantom test NOTE Some adjustments may be required to be made to the default image processing settings in the dental imaging software when evaluating the QA image Write down the used settings so that the same settings can be used when taking the control images later on NOTE The QA images are recommended to be saved to a dedicated place or patient card This way it is easy to use previous QA images as a reference during e g monthly Quality Assurance ins
38. llation manual for CLINIVIEW software or for other dental imaging software to be used with the unit The user manual for CLINIVIEW software or for other dental imaging software to be used with the unit Installation and set up manual 208684 1 Reviewed Savinen Antti Juhani 2015 03 11 17 52 DaUe i i ly Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 1 Introduction ORTHOPANTOMOGRAPH OP30 1 3 Warnings and precautions During installation e When installing the unit always observe local and national safety regulation concerning the installation and use of dental x ray equipment e Failure to install the unit in an approved location according to these installation instructions may cause the device to be dangerous to both patient and operator e The aperture plate in the collimator is made of lead Pb which is a toxic material Do not touch it with your bare hands e Be careful when working on the unit when the covers have been removed Some mechanical parts have sharp edges that can cause injuries if mishandled Some internal components become hot during operation Take suitable precautions to avoid burning yourself Some electric components are high voltage and can be lethal if touched Whenever you are making adjustments to the unit disconnect it from the main power
39. n accordance to the appropriate IEC 601 1 and or IEC 601 1 1 harmonized national standard 2 3 The dental imaging software The dental imaging software for example CLINIVIEW software that is installed in the PC used with the unit must be MDD compliant 2 4 Space requirements When installing the unit make sure that there is enough space at the front and sides of the unit to allow patients to enter and exit the unit easily Patients in wheelchairs will require more space than standing patients the unit is positioned at least 1 85 m 73 away from any non medical electrical or electromechanical equipment the unit is positioned so that the operator when protected from radiation can see and hear the patient during an exposure Installation and set up manual 208684 9 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 2 Pre installation requirements ORTHOPANTOMOGRAPH OP30 2 5 Unit and fixing hole dimensions TE S TEs i 200 10 mm 7 8 0 4 Standard i column leg I 3 180 10 mm 7 1 0 4 gt 2340 mm a 140 10 mm 5 5 0 4 w 92 1 rN gt N 440 10 mm 5 5 0 4 Short mr Centre to hole column leg Hr _ 2250 mm i 88 6 eg ACLS H 1 St
40. ni 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 ORTHOPANTOMOGRAPH OP30 3 Installing the unit 3 Drill two holes with diameters suitable for the 10 mm 3 8 fixing hardware being used and then attach the unit to the wall with the appropriate fixing hardware Do not fully tighten the screws yet 4 Remove the packaging material that protects the rotating unit and vertical carriage and remove the cover from the upper shelf 4 x M4 screws Place a spirit level widthwise on the upper shelf and level the unit in the left right direction Make small adjustments by loosening the wall bar fastenings and moving the top of the column to the left or the right to level the unit Make large adjustments by moving the base of the column left or right 6 When the unit is level in the left right direction place the spirit level lengthwise on the upper shelf and level the unit in the forwards backwards direction Make large adjustments by moving the base of the column backwards or forwards Installation and set up manual 208684 17 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDE
41. nit ORTHOPANTOMOGRAPH OP30 3 11 Connecting the unit to the PC The unit supports 10Base 1T 100Base TX Ethernet connections The PC must be connected to the unit in accordance with IEC 60601 1 1 Connection options Standalone With the standalone configuration the unit is connected to a PC and the unit then reserves the whole network interface card The unit and PC then function as a single standalone system with no access to other networks If it is necessary for the PC to be connected to another network the PC will require two network interface cards Network With the network configuration the unit is connected to the PC via a network device With this configuration the PC is connected to the unit and to other networks with a single network card The unit is also accessible from other PC s in the same network same subnet For the network configuration you may need a new IP address for the unit from your local system administrator Network device 28 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 3 Installing the unit Connection methods The unit can be connected to the PC in two ways Direct connection or IP based connection Both
42. nsor and the tubehead Incorrect image with an Correct image no artifacts example of an artifact 8 Touch the Arrow key to exit Image Calibration 60 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 5 Aligning the unit 5 6 Advanced Quality Assurance and Advanced Still Mode Advanced Quality Assurance and Advanced Still Mode programs are similar to the normal QA and Still Mode programs with the exception that in the Advanced programs the user can adjust the exposure values These programs can be used in special cases Advanced QA program is meant to be used similarly as the normal QA program described in section 4 4 Quality Assurance test Advanced Still Mode can be used for example for radiation measurements without the device movements The user can set the duration kV and mA values for the exposure This program can be used for example in commissioning inspections as it enables service functions without service control through PC Taking Advanced Quality Assurance and Still exposures 1 PC Prepare the unit for an exposure 2 If Advanced Quality Assurance exposure is to be taken to evaluate image quality prepare the unit as described in 4 4 Qual
43. ottom edge of the beam black line must be between 0 1 and 2 mm from the bottom edge of the image If the beam black line is too far to the right or left or too high or too low or if the CCD aperture gray area is tilted to one side the CCD sensor and or the beam must be adjusted See 5 1 Aligning the CCD sensor and 5 2 Adjusting the beam 9 Touch the Arrow key to exit the Still Mode and return to the main screen 42 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 4 Checking the alignment 4 2 Checking the image geometry Ball and Pin Phantom test 1 PC Prepare the unit for an exposure 2 Reattach the chin rest Standard 3 Setup the unit to take a Standard Panoramic panoramic Juvenile 70 kV 7 5 mA exposure Refer to the unit User manual for information on how to do this Juvenile 70 kV 4 Set the chinrest to the zero 0 position Installation and set up manual 208684 43 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 4
44. pections Installation and set up manual 208684 49 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 4 Checking the alignment ORTHOPANTOMOGRAPH OP30 NOTE The expected Line Pair resolution is 3 1 LP mm or better and all four low contrast resolution holes are possible to be distinguishable The line pair and low contrast resolutions distinguishable from the image depend not only on the unit but also on other parts of the system for example the dental imaging software being used 4 5 Completing the Installation Check that the unit functions correctly Take a test exposure to verify that the unit operates correctly and the image is transferred successfully to the PC Protect yourself from radiation when taking the exposure Check the up down keys drive the unit up and down Check that the light key switches the positioning lights on and off Check that when the emergency switch is pressed the unit stops operating Remote exposure switch If the remote exposure switch was installed check that the remote exposure switch button works initiates an exposure Also check that the remote exposure switch warning light comes on and warning sound is audible during the exposure Finally reattach all the cove
45. position step 1 BEFORE removing the unit from the wall or the exhibition stand 3 9 Attaching the covers to the column 1 Attach the covers to the bottom of the column Attach the Front Cover A to the front of the column with the special long screws 4 they are inserted from the rear Do not tighten them yet Special long screws x4 A Attach the Side Covers B to the sides of the column Note that the front edge of the Side Covers is positioned between the column and the Front Cover The rear edge is attached to the back of the column with two screws Tighten all the screws 22 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 3 Installing the unit 3 6 Checking the bite rod 1 Check that the bite rod can be inserted into the chin rest easily and that it stays in position when released If it is too tight or too loose adjust the tightness with the set screw slot blade screwdriver 0 5 x 3mm Bite rod Set screw 3 Exposure switch NOTE If the exposure switch is to be used with the remote option refer to the section 3 8 Installing the remote exposure switch 1 Attach the exposure switch holder to the rear of the vertical carri
46. r i clockwise clockwise Move Move CCD CCD to right to left l ma D k Adjust and recheck the position of the CCD sensor until the beam is vertical and centered in the middle of the CCD aperture 3 After the adjustment tighten the four nuts A that hold the CCD sensor in place 4 Touch the Arrow key to exit the Still Mode and return to the main screen 5 Now calibrate the sensor see section 5 5 Calibrating the CCD sensor 5 2 Adjusting the beam If the beam was not correctly aligned after carrying out the alignment check 4 1 Checking the alignment of the beam and the CCD sensor the beam must be adjusted and new still mode images taken to confirm that it is correctly positioned NOTE When carrying out the beam alignment the unit must be in the Still mode If the unit is not already in the still mode activate it through the Installation screen see section 4 1 Checking the alignment of the beam and the CCD sensor steps 1 6 52 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D5087777 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 4 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights rese
47. rium Dental reserves the right to make changes in specification and features shown herein or discontinue the product described at any time without notice or obligation Contact your Instrumentarium Dental representative for the most current information ORTHOPANTOMOGRAPH and CLINIVIEW are either registered trademarks or trademarks of Instrumentarium Dental in the United States and or other countries 0537 Manufactured by Instrumentarium Dental PaloDEx Group Oy Nahkelantie 160 Fl 04300 Tuusula FINLAND Tel 358 10 270 2000 www instrumentariumdental com Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 Copyright 2015 by PaloDEx Group Oy All rights reserved See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Contents Reviewed Savinen Antti Juhani 2015 03 11 17 52 Approved Ukkonen Juha Pekka 2015 03 13 13 08 ORTHOPANTOMOGRAPH OP30 Installation and set up manual 208684 D508777 4 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 Contents Contents Ws NEE OCC UO M aapea EE EEEE EOE ENOTE E EOE EEEREN 1 1 1 About this manual c ccc ceccecceeceeceeeeece eee eeeeeeeeeeeseeeeeseeseeeeeees 1 1 2 Associated documentation ccccecceeceeeceeceeseeeeeeeeeeeeeeeeeaeeaes 1 1 3 Warn
48. roup Oy All rights reserved ORTHOPANTOMOGRAPH OP30 5 Aligning the unit 8 Protect yourself from radiation Press and hold down the exposure switch to take an exposure NOTE The rotating unit doesn t move while taking exposures in Still Mode Advanced QA only 9 PC Visually evaluate the result using the dental imaging software For QA image evaluation see 4 4 Quality Assurance test Installation and set up manual 208684 63 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 6 Using the exhibition stand ORTHOPANTOMOGRAPH OP30 6 The exhibition stand 6 1 Attaching the unit to the exhibition stand The exhibition stand is used if the unit cannot be attached to a wall or needs to be freestanding for some other reason The exhibition stand can be used also for installations in areas prone to earthquakes if it is not possible to attach the unit to the floor and wall CAUTION HEAVY OBJECT A minimum of two persons are required for the following task 1 Unpack the unit as described in section 3 1 Preparing the unit for installation steps 1 4 2 Assemble the Exhibition stand refer to the Exhibition stand assembly disassembly instructions supplied with the Exhibition stand 3 Hold the main power and
49. rs that were removed 50 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 5 Aligning the unit 5 Aligning the unit 5 1 Aligning the CCD sensor lf the sensor was not correctly aligned after carrying out the alignment check 4 1 Checking the alignment of the beam and the CCD sensor the sensor must be repositioned and new still mode images taken to confirm that it is correctly positioned NOTE When carrying out the CCD sensor alignment the unit must be in the Still Mode If the unit is not already in the still mode activate it through the Installation screen see section 4 1 Checking the alignment of the beam and the CCD sensor steps 1 6 1 Remove the inner sensor cover from the rotating unit It is held in place with four 4 screws 2 To adjust the CCD sensor loosen the four nuts A that hold the CCD sensor in place and then adjust its position with the two screws B on the left hand side Installation and set up manual 208684 51 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 5 Aligning the unit ORTHOPANTOMOGRAPH OP30 TurnCCD lt gt Turn CCD a counte
50. rtation packaging so that units can be repacked and easily and safely moved to new locations if required 6 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 2 Pre installation requirements 2 Pre installation requirements 2 1 The Unit e The unit is supplied in one box on a pallet The dimensions of the box with pallet are Length Width Height 191 cm 85 cm 103 cm 76 34 41 e The assembled unit weighs 120 kg 265 Ib Make sure that the floor where the unit is to be installed can support this weight e The unit MUST be permanently attached to the wall It is strongly recommended that the base of the column is permanently attached to the floor If it is not possible use the exhibition stand See section 6 The exhibition stand e Ifthe area where the unit is to be installed is subject to earthquakes the unit MUST be permanently attached to the wall AND the base of the column MUST be permanently attached to the floor If the base of the column cannot be permanently attached to the floor use the exhibition stand Refer to section 6 The exhibition stand e The fixing hardware not supplied for attaching the unit to the wall floor and the fixing hardware no
51. rved ORTHOPANTOMOGRAPH OP30 5 Aligning the unit 1 Remove the inner tubehead cover from the rotating unit it is held in place with four 4 screws 2 Loosen the four locking nuts that hold the collimator in place 3 Adjust the beam position with the height adjusting screw If the beam is too low turn the screw counter clockwise to move it up If the beam is too high turn the screw clockwise to move the beam down P Move beam up Move beam down S Y Height adjusting screw gt N Correct a i m Too igh f N Too N low m F Locking nuts x 4 Recheck the beam position If not in the correct position readjust When the position is correct tighten the locking nuts 4 Touch the Arrow key to exit the Still Mode and return to the main screen 5 Now calibrate the sensor see section 5 5 Calibrating the CCD sensor Installation and set up manual 208684 93 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 5 Aligning the unit ORTHOPANTOMOGRAPH OP30 5 3 Adjusting the image geometry IMPORTANT NOTE In some cases you may have to adjust the position of both the rotating unit AND the chin rest Center ball not round If the center ball is NOT ROUND then the position of the rot
52. s to be attached Make sure that you use the correct fixing hardware for the wall material and the floor material NOTE Fixing hardware is not required if the exhibition stand is used without wall attachment See section 6 The exhibition stand and the separate instructions supplied with the exhibition stand Fixing hardware for the remote exposure switch WALL attachment requires 3mm 1 8 diameter fixing hardware two sets NOTE DO NOT use smaller diameter fixing hardware Installation and set up manual 208684 11 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 2 Pre installation requirements ORTHOPANTOMOGRAPH OP30 Installation tools Electric drill Spanners wrenches 7 10 17 two mm AF Hexagon socket wrenches Allen keys 1 5 8mm Slot blade screwdrivers 0 5 x 3mm 0 6 x 3 5 Torx screwdrivers T20 T25 T30 Spirit level Pliers and wire cutters Scissors Knife Transportation wheels Optional Calibration and setup tools Ball and Pin Phantom Laser alignment tool Image quality test tools Digital test tool and copper filter where use of this calibration tool is required To be ordered separately 12 Installation and set up manual 208684 Reviewed Savinen Antti Juh
53. screws that hold the bracket part of the wall mounting assembly attached to the column to the support bar to be attached to the wall and position it to the mid position Retighten the screws Installation and set up manual 208684 15 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 3 Installing the unit ORTHOPANTOMOGRAPH OP30 3 2 Attaching the unit to the wall CAUTION HEAVY OBJECT A minimum of two persons are required for the following task 1 Transport the unit to the place where it is to be installed NOTE The transportation wheels optional are attached to the base of the column with two M10 screws The transportation wheels can be used to wheel the unit to the place where it is to be used 2 Mark the position of the two fixing holes on the wall according to the dimensions given in the section 2 5 Unit and fixing hole dimensions Alternatively lift the unit into the position where it is to be used and then using the wall bar as a template mark the positions where the holes must be drilled Hold a spirit level on the side of the column and then on the front to help to position the unit so that it is as vertical as possible 16 Installation and set up manual 208684 Reviewed Savinen Antti Juha
54. t hold the chin rest in place on the underside of the patient support Installation and set up manual 208684 55 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 5 Aligning the unit ORTHOPANTOMOGRAPH OP30 2 Ifthe right hand distance in picture C is greater that the left hand distance move the chin rest to the left If the left hand distance in picture D is greater that the right hand distance move the chin rest to the right Move chin rest to left Move chin rest to right 3 Take another panoramic exposure of the Ball and Pin Phantom to confirm that the unit is now correctly aligned Readjust if required 4 After the adjustment has been done re tighten the two screws that hold the chin rest in place 56 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 5 Aligning the unit 5 4 Aligning the positioning lights Midsagittal and frankfort e 1 Remove the knob 1 screw that adjusts the a position of the frankfort light
55. t supplied for attaching the remote exposure switch to the wall must be the correct type for the wall floor materials See section 2 6 Fixing hardware for information on fixing hardware Installation and set up manual 208684 T Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 2 Pre installation requirements ORTHOPANTOMOGRAPH OP30 e Make sure that the fixing hardware to be used to attach the unit to the wall and the wall itself can withstand a continuous pull out force of at least 300 kg 660 Ib e Ifthe wall is made of thin material you may have to use a reinforcing plate on the rear side of the wall to hold the fixing hardware e Ifthe wall is not strong enough to support the unit or if you are not sure if the wall will support the unit use the exhibition stand e The place where the unit is to be installed and the position from where the user will take exposures must be correctly shielded from the radiation that is generated when the unit is operated Since radiation safety requirements vary from country to country and state to state it is the responsibility of the installer to ensure that all safety requirements are met e Do not install the unit in environments where there are corrosive or explosive vapours e Spe
56. ts reserved ORTHOPANTOMOGRAPH OP30 3 Installing the unit 7 The configuration screen will appear It shows the unit serial number and the firmware version numbers Unit OP30 2 SN 101300284 FW 2 05 I Core 37 Graphics 3 Also there are three function keys IP Settings Installation and QA Select IP Settings NOTE see section 4 1 Checking alignment of the beam and the CCD sensor for information about Installation see section 4 4 Quality Assurance test for information about QA 8 The IP configuration screen with two configuration mode function keys Manual and Current IP address Remote will appear 194 9 2297 251 Choose IP Configuration Mode Installation and set up manual 208684 35 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 3 Installing the unit ORTHOPANTOMOGRAPH OP30 9 The IP address screen will appear which shows the current IP address of the unit Current IP address 194 9 227 251 10 PC Press the Send to Device button IP based Device IP address Enable changing IP address 194 9 22 7 250 Send to Device 11 The new IP address will appear on the display Current IP address 194 9 227 251 Received IP address 194 9 227 25
57. use with 1 8mm thick Cu filter 46 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved ORTHOPANTOMOGRAPH OP30 4 Checking the alignment Taking a QA exposure 1 PC Prepare the unit for an exposure 2 Attach the Ball and Pin Phantom to the Chin rest 3 Then attach the Digital Test Tool Prufkorper test tool to the Ball and Pin Phantom 4 Attach the additional filter Cu filter to the center line of the tubehead inner cover as shown below The filter has suction cups Additional Filter 5 Set the chinrest to the zero 0 position 6 Touch the service key on the main control panel of the unit to open the configuration screen 7 Select QA Installation and set up manual 208684 47 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 4 Checking the alignment ORTHOPANTOMOGRAPH OP30 8 The QA program screen appears Select from two preset programs which to use for the QA exposure QA1 for use with 0 8mm thick Cu filter QA2 for use with 1 8mm thick Cu filter 66
58. x Group Oy All rights reserved 3 Installing the unit ORTHOPANTOMOGRAPH OP30 Make small adjustments by loosening the two screws that hold the bracket to the support bar and pulling or pushing the top of the column forwards or backwards Bracket N Support bar 3 3 Attaching the unit to the floor It is recommended that the base of the unit is attached to the floor If the unit is being installed in an area prone to earthquakes OR if you think that the wall bracket by itself will not hold the unit rigidly enough to the wall then the base of the column MUST be fixed to the floor NOTE If the base of the column needs to be fixed to the floor but it is not permitted or possible use the exhibition stand See section 6 The exhibition stand and the separate instructions supplied with the exhibition stand 1 Make sure that the unit is level and then mark the positions of the two fixing holes on the floor 18 Installation and set up manual 208684 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 ORTHOPANTOMOGRAPH OP30 3 Installing the unit 2 Loosen the two screws that hold the bracket to the support bar and rotate the unit to one side so that you can access the mark for the floor hole Drill a hole suitable for the 8 mm 5 76 fixing hardware that is to be used Q Rotate the unit to the other side and drill the second hole 3 Turn
59. ystem must meet the IEC 60950 standard minimum requirements Devices that do not meet the IEC 60950 standard must not be connected to the system as they may pose a threat to operational safety 1 Position the PC to be used with the unit at least 1 85 m 73 away from the unit NOTE If the unit and PC are to be part of a larger dental system or connected to local network make sure that all the other system components and devices are installed connected and configured correctly Refer to the documentation supplied with the other components and devices for information on how to do this 2 Switch the PC on and install CLINIVIEW software or other dental imaging software that will be used with the device For information on how to do this refer to the documentation supplied with the dental imaging software you are installing IMPORTANT NOTE After the dental imaging software has been installed a Select device window will appear Make sure that you select OP30 NOTE If a disk for updating the driver is supplied with the unit install the driver update after you have installed the dental imaging software Installation and set up manual 208684 2 Reviewed Savinen Antti Juhani 2015 03 11 17 52 D508777 4 Approved Ukkonen Juha Pekka 2015 03 13 13 08 See PDM system to determine the status of this document Printed out 2015 04 10 14 40 09 Copyright 2015 by PaloDEx Group Oy All rights reserved 3 Installing the u