Home

User Manual - Meena Medical Inc.

image

Contents

1. ERROR Accompanied by a number Thermometer has detected an internal malfunction Remove and reinstall battery To perform a Self Test Press and hold probe cover detector flush against outer case Figure 4 While holding press and release temperature switch once to initiate thermometer self test ERROR m La to verify the display and thermometer s electronics are in working order If error message appears again record the number and contact qualified service personnel The display messages indicates the ambient room temperature is outside the thermometer s operating range lo A Ambient temperature is below 15 6 C 60 F hi A Ambient temperature is above 43 3 C 110 F r L ror 104 m 9 1 LEM um The display messages indicates the patient s temperature is outside the thermometer measurement range lo Measured temperature is below 25 C 77 F hi Measured temperature is above 43 3 C 110 F 14 INFORMATION DISPLAYS MAINTENANCE CORE CHECK Thermometer Technical Service Manuals are available from ALARIS Medical Systems They include routine service schedules interconnect diagrams component parts lists and descriptions test procedures and or references and other technical information to assist service personnel in repair and maintenance of each instrument s repairable components Maintenance procedures are intended to be
2. ALARIS MEDICAL SYSTEMS IVAC Tympanic Thermometer System Model 2090 Directions for Use THIS PAGE LEFT BLANK TABLE CONTENTS INTRODUCTION ABOUT THE SYSTEM A E apo na td di dos FEATURE SUMMARY ee A dede T a rar d Thermometer Featiles aid DEFINITION Sia a RE a deu GETTING STARTED WARNINGS AND CAUTIONS eres DeflriltiOlls suu e tasa a a c reddes i c i UR ae ee Warnings m bad Mer dot ad A ptr ik need t e e E INSTRUMENT metet A eae Pe oA da dung SETUP PROCEDURE hearre oe beh el tater P M IL UNN gt 000 0 OPERATING INSTRUCTIONS PROBE COVER DISPENSE LOAD INSTRUCTIONS 2 222 2 2 45 254 4 4 44 30 07 0 1222 Probe Cover DISPENSER occi da ad Load Probe Cover Onto Thermometer 2 2 2 22 242 2 2 2 2 TAKINGA TEMPERATURE 3 22 5282 ile emet Ae EE EN E ER 1 INFORMATION DISPLAYS CORE CHECK THERMOMETER LCD DISPLAY 13 MAINTENANCE SPEGIFIGATIONS uil RUNS nn Se dete add td deed 15 THERMOMETER CLEANING ccoo nes 16 SERVIGE INFO
3. ED INCLUDING ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE MAINTENANCE 19 THIS PAGE LEFT BLANK 20 MAINTENANCE ALARIS MEDICAL SYSTEMS ALARIS Medical Systems Inc 10221 Wateridge Circle San Diego CA 92121 USA Mail P O Box 85335 San Digo California 92186 5335 USA Authorized EU Representative ALARIS Medical UK Ltd The Crescent Jays Close Basingstoke Hampshire RG22 4BS UK ALARIS ALARIS Medical Systems IVAC and CORE CHECK are registered trademarks of ALARIS Medical Systems Inc All other trademarks belong to their respective owners U S Patent Nos 4 907 895 6 332 090 CA Patented Brevet Nos 2037775 2037940 DE D B P No 446788 GB Patent No 446788 JP Patent Nos 2603003 2603004 Other Patents Pending 148281 000 Copyright 2003 ALARIS Medical Systems Inc All rights reserved 0086
4. RMATION 2 docete t eee eh bea aye Rae Me a ye BERE e eo Ed ias 18 C stomer ServiGe ooo tee t ebd e e eter er RE n ome Heo fal er ete etos ae 18 Technical Support 22 22 28 se ER e Le Dee CIAR atado Mere ae 18 225 9 2 ar OO Tuo Pe e oor adu bres tes Eo Ami 18 WARRANTY atenta entente cual somos SA d AE brem eR berita eM Ace d gue 19 TABLE OF CONTENTS i SNOILONULSNI NOILVINYOSNI NOILONGOWLNI m J 2 gt A m ADNVNALNIVIN GENERAL CONTACT INFORMATION Customer Advocacy For clinical and technical questions feedback and troubleshooting assistance Phone toll free within the United States and Canada 800 854 7128 Ext 7812 E Mail CustomerFeedback alarismed com Technical Support and Service Center Addresses AE ALARIS Medical Systems Middle East Office PO Box 5527 Dubai United Arab Emirates Tel 971 4 28 22 842 Fax 971 4 28 22 914 AU ALARIS Medical Australia Pty Ltd 8 167 Prospect Highway Seven Hills NSW 2147 Australia Tel 61 2 9838 0255 Fax 61 2 9674 4444 BE ALARIS Medical Belgium B V Otto De Mentockplein 19 1853 Strombeek Bever Belgium Tel 32 2 267 38 99 Fax 32 2 267 99 21 CA ALARIS Medical Canada Ltd 5975 Whittle Road Suite 120 Mississauga Ontario 1427 3N1 Ca
5. Terminals C F Switch Battery Cover Figure 4 Probe Cover Detector Figure 5 CORE Information Display I LI LI LIE REPLACE BATT LOW COVER CAL ERROR RETURN TO BASE 8 GETTING STARTED OPERATING INSTRUCTIONS The following instructions assist dispensing and loading probe covers onto the thermometer Probe Cover Dispense Load Instructions Probe Cover Dispenser Remove thermometer and dispenser from the home base Press and slide dispenser door up until it locks in open position nstall a package of P90 probe covers into the dispenser with arrow pointing up toward opened sliding door Slide door down until locked in closed position leaving the tab outside the door Firmly grasp tab and pull it out of the dispenser U m gt z o NOTE Use only ALARIS Medical Systems P90 probe covers with the CORE CHECK Thermometer The size and thermal characteristics of the probe cover affect thermometer s performance Inaccurate readings may occur unless recommended probe covers are used zZ M A O zZ Load Probe Cover Onto Thermometer Dispense a probe cover by pushing dispenser slide completely in and slowly releasing Center probe tip onto probe cover film then press firmly to engage base of probe This will actuate the probe cover detector The clear film will stretch smoothly over the entire probe Failure to
6. ature measurement BATTERY lt lt COMPARTMENT 000000000000 00000000000 00000000000 DVAI Figure 2 Probe Cover Dispenser THERMOMETER RECEPTACLE SLIDER PROBE COVER LEVEL INDICATOR 00000000000 000000000000 PROBE TIP INFORMATION DISPLAY Displays temperature for approximately four minutes or until next measurement is taken GREEN INDICATOR LIGHT Flashes once to indicate temperature measurement complete 6 GETTING STARTED Setup Procedure 1 Carefully unpack the CORE CHECK Tympanic Thermometer System and ensure the thermometer dispenser and home base are undamaged Packaging will include a battery battery Set C F switch mode compartment cover extra probe tips and a probe cover wall Install battery bracket 2 Set Temperature mode Set thermometer to Celsius C or Fahrenheit F by sliding the recessed switch located on left side of the battery compartment Use a small screwdriver or pen tip to slide the switch into position F toward the outside C toward the center 3 Install Battery Place a 9 volt alkaline battery into the compartment as shown in Figure 3 and snap connector onto battery terminals 4 Attach battery compartment cover Orient the cover s arrow as shown in Figure 3 Slide cover over opening until cover locks in place If necessary to remove the cover begin by fi
7. cal Systems be liable for any incidental indirect or consequential damages in connection with the purchase or use of any ALARIS Medical Systems product This warranty shall apply solely to the original purchaser This warranty shall not apply to any subsequent owner or holder of the product Furthermore this warranty shall not apply to and ALARIS Medical Systems shall not be responsible for any loss or damage arising in connection with the purchase or use of any ALARIS Medical Systems product which has been A repaired by anyone other than an authorised ALARIS Medical Systems service representative altered in any way so as to affect ALARIS Medical Systems judgement the stability or reliability of the product or has had the product s serial or lot number altered effaced or removed C subjected to misuse or negligence or accident or gt zZ par m 2 gt zZ e m D improperly maintained or used in any manner other than in accordance with the written instructions furnished by ALARIS Medical Systems This warranty in lieu of all other warranties express or implied and of all other obligations or liabilities of ALARIS Medical Systems and ALARIS Medical Systems neither assumes nor authorises any representative or other person to assume for it any other liability in connection with the sale of ALARIS Medical Systems products ALARIS MEDICAL SYSTEMS DISCLAIMS ALL OTHER WARRANTIES EXPRESS OR IMPLI
8. d every ten thousand readings or every six months whichever comes first The CAL message automatically flashes after ten thousand readings Calibration must be performed using the ALARIS Medical Systems CORE CHECK Thermometer Calibrator Model 9000 Refer to the calibrator s operating instructions for calibration procedure If thermometer is dropped inspect the lens and probe tip for damage The lens and probe tip can be replaced without opening the instrument Refer to the Technical Manual for procedure Calibrate the CORE CHECK Thermometer if lens replacement was required The CORE CHECK Thermometer can be cleaned with a moist cloth warm water and a mild detergent such as dish washing liquid or with a variety of other agents See Figure 7 for appropriate cleaning methods and acceptable cleaning agents for the different thermometer components Do not immerse the CORE CHECK Thermometer Frequently check the lens for cleanliness Incorrect readings can result from debris or fingerprints Lens should appear clear shiny and unobstructed from protective probe tip 16 MAINTENANCE Thermometer Cleaning Continued If a specific cleaning agent or disinfectant is currently used ALARIS Medical Systems recommends examining chemical ingredients prior to use on the thermometer If an uncertain cleaning agent or disinfectant is currently being used on the instrument call ALARIS Medical Systems PROBE LENS In Probe Fi
9. d to reduce potential injury observe the following precautions Warnings and Cautions Definitions A warning is an alert to potential serious outcomes death injury or serious adverse events to the patient or user CAUTION A caution is an alert to take special care for the safe and effective use of the device Cautions CAUTION Do not autoclave or immerse the CORE CHECK Thermometer as damage will occur a m 2 gt A 4 m CAUTION Operating the CORE CHECK Thermometer near equipment which radiates high energy frequencies for example electrosurgical cauterizing equipment may affect temperature readings CAUTION If unusually low or high temperature readings are obtained remove the probe cover verify the probe lens is clean shiny dry undamaged and free of debris Attach a new probe cover and repeat measurement CAUTION Keep lens clean Lens should be cleaned with a dry lint free scrub or tissue Replace lens if it remains cloudy CAUTION Damage may occur if alcohol or acetone is used GETTING STARTED 5 Instrument Features PROBE LENS Figure 1 CORE CHECK Tympanic Thermometer System PROBE COVER DETECTOR pucr Assures a new temperature is not displayed unless probe cover Is replaced after each measurement Use P90 probe COVers BACKLIGHT BUTTON Lights the display for use in low light conditions TEMPERATURE SWITCH Initiates temper
10. ew Zeland Tel 64 9 273 3901 Fax 64 9 273 3098 SE ALARIS Medical Nordic AB Hammarbacken 4B 191 46 Sollentuna Sverige Tel 46 8 544 43 200 Fax 46 8 544 43 225 US ALARIS Medical Systems Inc 10221 Wateridge Circle San Diego CA 92121 USA Tel 1 800 854 7128 Fax 1 800 854 6179 ZA ALARIS Medical S A Pty Ltd Unit 2 Oude Molen Business Park Oude Molen Road Ndabeni Cape Town 7405 South Africa Tel 27 0860 597 572 Fax 27 21 5107567 INTRODUCTION AE About the System This manual is a guide for using the CORE CHECK Tympanic Thermometer System Model 2090 The CORE CHECK Thermometer is designed to measure infrared energy emitted from the tympanic membrane Infrared energy is generated by the vibration and movement of molecules in all matter As temperature increases so does molecular activity causing an increase in the radiation of infrared energy zZ par A O Oo e zZ Infrared energy emitted from the tympanic membrane in the ear canal provides an excellent measurement of the body s core temperature The blood supply for the tympanic membrane is common with the hypothalamus the area of the brain that regulates body temperature The CORE CHECK Thermometer is positioned in the external auditory canal The thermometer senses infrared radiation emitted from the surface of the tympanic membrane and surrounding tissue This reading is quickly processed throu
11. fully load the probe cover will result in a REPLACE COVER message NOTE If probe cover backing is not fully seated against the base of the probe a REPLACE COVER message will appear on the display when the measurement is initiated If this occurs replace the cover OPERATING INSTRUCTIONS 9 Taking A Temperature The most important step in obtaining accurate temperatures is correct placement of the probe tip with a P90 disposable probe cover in the patient s ear Center the probe tip far enough into the ear canal to firmly seal the ear canal opening and point in the direction of the tympanic membrane Gently restrain the head if necessary Press and hold the temperature switch until the green light flashes and temperature reading is displayed Correct positioning of the probe tip is necessary for accurate readings See Figure 6 below Figure 6 Positioning of Thermometer GENTLY PUSH TO SEAL EAR CANAL CAUTION Evaluate condition of the patient s ear and adjacent skin The thermometer should not be used on patients with ear drainage infected or chafed skin adjacent to the ear or on patients with incisions adjacent to the ear Such usage may increase the risk of cross contamination Operating the CORE CHECK Thermometer near equipment which radiates high energy frequencies for example electrosurgical cauterizing equipment may affect temperature readings CAUTION If an un
12. gh the CORE CHECK Thermometer technology to provide an accurate measurement of core body temperature The CORE CHECK Tympanic Thermometer System includes the thermometer Probe Cover Dispenser Model 2992 Home Base Model 2993 and Probe Cover Bracket Model 2994 The thermometer uses ALARIS Medical Systems Disposable Probe Covers Model P90 For additional information contact your ALARIS Medical Systems representative INTRODUCTION 1 Feature Summary Thermometer Features Accurate non invasive determination of core temperature Temperature measurement within 3 seconds Anti theft feature Durable by design and construction User selectable Celsius or Fahrenheit temperature readings Display backlight for low light conditions Powered by a single 9 volt battery Home base holds thermometer and dispenser Disposable probe covers and probe cover detectors reduce the risk of cross contamination Dispenser conveniently stores and dispenses probe covers Thermometer performs a self test Disposable Probe Cover PROBE COVER BRACKET Attaches to the wall and holds a box of five probe cover packets HOME BASE Holds thermometer and dispenser also resets the anti theft feature THERMOMETER DISPENSER Holds and dispenses P90 probe covers Combines with thermometer to provide a hand held unit DISPOSABLE PROBE COVERS Reduce the risk of cross SS contamination and eas
13. gure 7 Cleaning of Thermometer Tip Wipe lens with a lint free swab If soil or debris remain use a cotton tip swab slightly dampened with alcohol followed with a lint PROBE TIP free wipe to remove any THERMOMETER alcohol film Allow 30 to 45 DISPENSER HOME minutes for evaporation BASE before using thermometer WALL MOUNTING BRACKET not shown Wipe surfaces using a DISPLAY SCREEN cloth dampened with mild DO NOT USE alcohol on detergent or a less than Screen 4096 bleach solution Wipe screen using a cloth See table below for other dampened with a mild detergent solution or a less than 4096 bleach solution See table below for other cleaning agents cleaning agents 000000000000 00000000000 CAUTION Only use Isopropanol Alcohol on the lens as directed Do not use Alcohol on other parts of the system Only use a cloth dampened with one of the following cleaning agents to clean the system Avoid contact with the lens Cleaning Agents Chemical Base Ammonium Quaternaries Bleach 40 Chloride Chlorine gt 2 pa m z gt 2 m Cidex Waviside Glutaraldehyde Mild Detergent Alkali MAINTENANCE 17 Service Information NOTE Ifthe instrument shows evidence of damage in transit notify the carrier s agent immediately Do not return damaged equipment to the factory before the carrier s agent has authorized repairs If the instrument fails t
14. icates the battery should be replaced CORE Displays with each temperature measurement REPLACE COVER Replace Probe Cover A temperature reading was initiated without a probe cover a used probe cover or an improperly attached probe cover Check lens for cleanliness attach a new probe cover and reinitiate the measurement RETURN TO BASE Anti Theft Feature A limited number of measurements can be taken until thermometer displays a continuous RETURN TO BASE message Place thermometer into home base to reset the anti theft feature RETURN TO BASE Displays as a flashing message when the limit of temperature measurements have been met Return to base to reset Anti theft feature has been reset Thermometer can be used immediately CORE rir me LI LI LIE REPLACE BATT LOW COVER CAL Ennon RETURN TO BASE REPLACE COVER RETURN TO BASE RETURN TO BASE INFORMATION DISPLAYS 13 z 56 y 55 23 vo z CORE CHECK Thermometer LCD Display Screens Continued BATT LOW Battery Low Battery is low and requires CORE replacement nr LOW REPLACE Replace Battery Battery must be replaced Figure 3 REPLACE BATT CAL Recalibrate Instrument Indicates calibration of the thermometer should be checked Refer to Maintenance section of this manual for information concerning calibration
15. ily loads onto thermometer probe tip 2 INTRODUCTION Definitions ENSURE LENS IS CLEAN Cleaning Advisory A BEFORE USE USE ONLY 9V ALKALINE Thermometer will not work if taken from building Q lt OPEN PRESS TO SLIDE External Auditory Canal Infrared Energy Molecular Activity Tympanic Membrane CE 0086 MM YYYY N Alkaline battery used Anti Theft notification Instructions for probe cover dispenser A narrow tube like passage through which sound enters the ear also named an external auditory canal Relating to the range of invisible radiation wavelengths from 750 nanometers Generating using or sensitive to infrared radiation Temperature is a measure of molecular motion Pressure is a measure of molecular activity The higher the pressure the faster the molecules are moving and greater energy equals greater temperature The membrane in the ear that vibrates to sound Conformance to the Medical Device Directive 93 42 EEC Manufactured Date expressed in four digits as the year and two digits as the month Attention Refer to accompanying documentation Indicates degree of protection fluid ingress INTRODUCTION 3 2 2 O O 2 THIS PAGE LEFT BLANK 4 INTRODUCTION GETTING STARTED To ensure proper performance of the CORE CHECK Thermometer an
16. nada Tel 1 905 507 1131 Fax 1 905 507 6664 CN ALARIS Medical Systems Office Suite 401 No 88 Chang Su Shanghai PC 200040 China Tel 56 8621 62488591 Fax 56 8621 62482567 DE ALARIS Medical Deutschland GmbH Pascalstr 2 52499 Baesweiler Deutschland Tel 49 2401 6040 Fax 49 2401 604 121 ES ALARIS Medical Espa a S L Avenida Valdelaparra 27 Edificio Alcor 28108 Alcobendas Madrid Espa a Tel 34 91 657 20 31 Fax 34 91 657 20 42 FR ALARIS Medical France S A 95 rue P reire 78105 St Germain en Laye Cedex France T l 33 1 39 10 50 00 Fax 33 1 30 61 22 23 GB Manufacturer s ALARIS Medical UK Ltd The Crescent Jays Close Basingstoke Hampshire RG22 4 5 United Kingdom Tel 44 0800 389 6972 Fax 44 1256 388 411 HU ALARIS Medical Hungary D brentei ter 1 H 1013 Budapest Magyar Tel 36 14 88 0232 Fax 36 12 01 5987 IT ALARIS Medical Italia S P A Via Ticino 4 50019 Sesto Fiorentino Firenze Italia T l 39 055 34 00 23 Fax 39 055 34 00 24 NL ALARIS Medical Holland B V Kantorenpand Hoefse Wing Printerweg 5 3821 AP Amersfoort Nederland Tel 31 33 455 51 00 Fax 31 33 455 51 01 NO ALARIS Medical Norway A S Solbr veien 10A 1383 ASKER Norge Tel 47 66 98 76 00 Fax 47 66 98 76 01NO NZ ALARIS Medical NZ Ltd Unit 14 13 Highbrook Drive East Tamaki Auckland N
17. o respond as described in this manual and the cause cannot be determined do not use the instrument Contact qualified ALARIS Medical Systems service personnel Customer Service Information or assistance may be obtained by calling one of the following Customer Service numbers United States 800 482 4822 UK Phone 44 1256 388229 Fax 01256 330 860 Technical Support Technical support service information applications and manuals may be obtained by contacting an ALARIS Medical Systems representative United States 800 854 7128 extension 6003 UK phone 44 1256 388229 When submitting any request for service include a description of difficulty experienced instrument settings administration set lot number solution s used message displayed at time of difficulty Product Return If it is necessary to return the instrument for service obtain a return authorization number prior to shipment Carefully package the instrument preferably in the original packaging reference the return authorization information and return it to the appropriate service or distribution center ALARIS Medical Systems does not assume any responsibility for loss of or damage to returned instruments while in transit 18 MAINTENANCE INTERNATIONAL WARRANTY ALARIS Medical Systems Inc hereinafter referred to as ALARIS Medical Systems warrants that A Each new infusion instrument pump controller
18. or peripheral instrument is free from defects in material and workmanship under normal use and service for a period of two 2 years from the date of delivery by ALARIS Medical Systems to the original purchaser Each new accessory is free from defects in material and workmanship under normal use service for a period of ninety 90 days from the date of delivery by ALARIS Medical Systems to the original purchaser C Each Mains Cable Battery Flow Sensor ECD and non disposable probe is free from defects in material and workmanship under normal use and service for a period of ninety 90 days from the date of delivery by ALARIS Medical Systems to the original purchaser D Each new Thermometer is free from defects in material and workmanship under normal use and service for a period of one 1 year from the date of delivery by ALARIS Medical Systems to the original purchaser If any product requires repair during the applicable warranty period the purchaser should communicate directly with its local ALARIS Medical Systems service centre to determine the appropriate service facility Except as provided otherwise in this warranty repair or replacement will be carried out at ALARIS Medical Systems expense The product requiring service should be returned promptly properly packaged and postage prepaid by purchaser Loss or damage in return shipment to ALARIS Medical Systems shall be at purchaser s risk In no event shall ALARIS Medi
19. performed only by qualified personnel Specifications Measurement Range 25 C to 43 3 C 77 F to 110 F Accuracy Response Time Battery Approximately 8 000 readings on one 9 volt alkaline battery under normal use Size 7 5 x 1 5 x 3 0 inches 19 05 x 3 81 x 7 62 centimeters Weight 8 ounces with battery 266 grams Case Material Durable impact resistant plastic gt Disposable ALARIS Medical Systems 90 Probe Covers Humidity 15 to 90 Noncondensing Storage Temperature 35 C to 60 C 32 F to 140 F Anti Theft Feature Display Messages 0 1 C 0 2 F Meets ASTM Specifications No E1112 86 at ambient temperature range of 18 3 to 32 2 C 65 90 F Less than 3 seconds 40 measurements are available before return to base s displayed Low Replace Return to Base hi A high ambient lo A low ambient lo low temperature hi high temperature and numbered error messages for internal malfunctions MAINTENANCE 15 Thermometer Cleaning The LCD display is tested before each measurement by CAUTION lighting all of its segments for a brief period It is recommended Do not autoclave or immerse the to inspect the display during this time to ensure all segments CORE CHECK Thermometer as are operating properly The display test ends when the damage will occur temperature or Io is displayed ALARIS Medical Systems recommends the CORE CHECK Thermometer be calibrate
20. rmly pressing center of cover with thumb causing a corner to lift then using the other hand push cover at loosened corner and slide off DO NOT PRY COVER WITH SHARP OBJECT m 2 gt A 4 m 5 Self Test Press and hold probe cover detector flush against outer case Figure 4 While holding press and release temperature switch once to initiate thermometer self test to verify both the display and thermometer s internal electronics are in working order display is in working order if all segments momentarily appear Figure 5 If all segments do not appear do not use thermometer and refer to Service Information section of this manual thermometer s internal electronics in working order if a temperature measurement or a lo message appears on the display numbered ERROR message appears on the display the thermometer has detected an internal malfunction Refer to ERROR message in the Information Display section of this manual 6 After a successful self test routine the thermometer is ready for temperature measurement 7 Home Base The home base should be mounted to a convenient location using either adhesive tape or screws supplied with the instrument Always store thermometer and dispenser in the home base when system is not in use which resets the Anti Theft feature GETTING STARTED 7 Setup Procedure Continued Figure 3 Battery Battery Compartment
21. usually low or high temperature reading is obtained remove probe cover verify the probe lens is clean shiny dry undamaged and free of debris Attach a new probe cover and repeat measurement 10 OPERATIONAL INSTRUCTIONS Taking A Temperature Continued Hold thermometer in place until temperature reading is displayed Approximately three seconds The green indicator light below the display will flash briefly NOTE Removal of the thermometer from the ear before temperature displayed will result in a low temperature reading 2 Remove thermometer from patient s ear and pull off i 9 probe cover used probe cover must be replaced by a new 4 one before another measurement be taken 2413 gt oz A 05 AN m C No Record the displayed temperature Another temperature reading can be taken immediately after replacing the probe cover The temperature will remain displayed for approximately four minutes unless another measurement is taken Store in home base OPERATING INSTRUCTIONS 11 THIS PAGE LEFT BLANK 12 GETTING STARTED The following examples LCD display screens modes INFORMATION DISPLAYS CORE CHECK Thermometer LCD Display Screens Self Test Displays with initiation of each measurement Inspect display to ensure all segments are operating as shown NOTE If this self test display continues to flash it ind