Home

ozone monitor introduction

image

Contents

1. 1 Zero air source 2 Certified Transfer Standard 3 Sampling manifold inert material such as PTFE or FEP only 4 Sampling lines inert material such as PTFE or FEP only Zero air can be generated either from compressed cylinders or from scrubbed ambient air If ambient air is used contaminants such as ozone and nitric oxide must be removed Detailed procedures for generating zero air are in the EPA s Technical Assistance Document for the Calibration of Ambient Ozone Monitors Model 205 Dual Beam Ozone Monitor Manual Rev D 23 Instrument Preparation Prior to calibration follow the steps below 1 2 Turn on the Model 205 Ozone Monitor and allow it to stabilize for a minimum of one hour Connect the instrument to the manifold on the ozone calibration setup If a particle filter will be used in normal operation the calibration must be performed through the filter The manifold must be vented to atmosphere so that pressure does not build up in the calibration setup Connection of the Model 205 directly to a pressurized output of any device can damage the ozone monitor Verify that the flow rate into the manifold is greater than the total flow required by the ozone monitor and any other flow demand drawing from the manifold Calibration Setup Preparation As indicated in the EPA Technical Assistance Document there are several tests that should be performed prior to calibration to ensure the accuracy of the measurements T
2. Svc Menu A amp B Lmp The A amp B menu allows one to output the calculated ozone values for both detection cells cells A and B If A amp B is chosen the following submenu appears Output A amp B No Yes If Yes is chosen as an example the serial data line will appear as follows 65 4 67 6 66 6 35 3 980 6 1545 1 3876 2 3143 0 1875 15 10 01 18 31 27 where the current value measured in detection cell A is 65 4 ppbv the current value measured in detection cell B is 67 6 ppbv and the 2 point running mean for both cells is 66 6 ppbv The 2 point running mean is the only ozone concentration output if No is selected The main purpose of this option is for diagnostics to make sure that both cells are performing correctly and with good precision but it also provides unaveraged data for applications that require the fastest possible response time The remaining data in the line are the temperature internal pressure volumetric flow rate analog inputs if this option is turned on date and time To Test the Lamp Voltage and Precision A diagnostic lamp test is provided in the Lmp submenu of the service menu Before the lamp test begins Pump On or Off is displayed Turning the pump off during the lamp test is recommended because it eliminates any noise related to the air flow in order to better focus the test on the lamp stability Comparing the results from testing with the pump off and with the pump on can help determine the
3. and begin making measurements again select and click on the left arrow To Log Data Select the Dat submenu from the Main Menu using the Select button The display will now show Data Menu Xmt Log End To start logging data rotate the Select switch to move the cursor to Log and click to select the logging mode You will then be asked whether you want to overwrite the data stored in the logger Overwrite Data No Yes lt If you select yes and start logging all data previously stored in the logger will be irretrievably lost If you have data in the logger that you want to keep be sure to download it before starting logging If you are ready to start logging then select Yes by moving the cursor under Yes and clicking Either selection will return you to the Main Menu To start data acquisition select lt and click Model 205 Dual Beam Ozone Monitor Manual Rev D The Ozone Monitor will then alternately display 1 the ozone mixing ratio and log number and 2 the ozone concentration internal temperature and internal pressure For example the display might read O3 56 7 ppbv T 305 6 P 730 4 where the ozone value is the most current measurement of ozone and T and P are the cell temperature and pressure in units of K and Torr in this case After 5 seconds midway between the next 10 s measurement cycle as an example the display will be replaced by O3 56 7 ppbv Log 193 0 where O3 is the ozone value most
4. data logger Up to 16 383 data lines containing log number ozone mixing ratio internal temperature internal pressure volumetric flow rate date and time may be stored in internal memory corresponding to an operational time of 9 1 hours with no averaging Averaging times of 10 s 1 min 5 min and 1 hr also may be selected from the menu thereby allowing the instrument to operate for 1 9 days 1 6 weeks 1 9 months and 1 87 years respectively before filling the memory The maximum number of data lines is halved if the three analog inputs are logged along with the other data Model 205 Dual Beam Ozone Monitor Manual Rev D The following diagram summarizes the complete menu D T 10 32 21 14 10 2008 Model 205 Dual Beam Ozone Monitor Manual Rev D Selecting the Menu The menu is accessed using the Select button on the front panel of the instrument To reach the menu hold in the Select button until the display shows Menu Then release the button The panel will now display Menu Dat Avg Cfg Sve lt where Dat Avg Cfg and Lmp are submenus that may be selected A blinking cursor will show across the D of the Dat submenu The Select button may be rotated clockwise or counterclockwise to move the cursor under the first letter of one of the other submenus To select a particular submenu move the cursor under the first letter of a submenu and click press in the Select button To exit the Main Menu
5. other instruments to be transmitted to a computer simultaneously with those of ozone and the time and date stamp these inputs may also be logged in the instrument s internal memory as described below Examples of external measurements that are commonly made along with ozone are external temperature pressure and relative humidity but the outputs of any instrument may be input to the Ozone Monitor The analog inputs may range from O to 2 5000 volts and are measured with an accuracy of approximately 30 0001 volt An input voltage greater than 5 0 volts or less than 0 3 volts may damage the printed circuit board If the Ozone Monitor has been set to the log data mode the output serial data line will be preceded by the log number e g 2893 67 4 35 3 980 6 851 1 3876 2 3143 0 1875 15 10 01 18 31 27 where 2893 is the log number In addition to data lines messages are written to the serial port when logging is begun or ended when transmission of data from the logger is begun and ended when data collection is interrupted e g due to a power failure and when the averaging time is changed Model 205 Dual Beam Ozone Monitor Manual Rev D Data Averaging and Data Logging Using the Menu When first turned on the instrument will start making measurements at a rate of once every 2 s unless a different averaging time was previously chosen Internally generated data along with up to three external voltages may be logged in the internal
6. recently written to the logger and the log number is 193 If averaging has been selected then the above display will be replaced by Avg 03 56 7 ppbv Log 193 4 Again 193 refers to the most recent log number The 4 in 193 4 refers to the number of 10 s data points that have been measured so far for inclusion in the next average to be displayed and logged If 2 s or 10 s averaging is used this number will always be 0 If 1 min averaging is used this number will increment from 0 to 5 for 5 min averaging the number will increment from 0 to 29 and for 1 hr averaging it will increment from 0 to 359 This number is displayed so that the user will Know how many more 10 s measurements need to be made before a new average is displayed and logged If there is a power failure while the instrument is in the logging mode logging will resume after power is restored A note of Data Interrupt will be written to the logger prior to writing the first new data line In the case of a power failure as many as 10 data lines may be lost because the microprocessor writes to the logger memory in groups of 10 lines All data residing only in the volatile memory of the microprocessor are lost when power is interrupted Also the start time for logging of additional data following a Model 205 Dual Beam Ozone Monitor Manual Rev D power interruption will be accurate only to the nearest minute or nearest hour when Avg 1 hr The instrument can
7. should be performed at least every 12 months or within the calibration frequency any time major disassembly of components is performed or any time the zero or span checks give results outside of the acceptable limits Instrument Preparation 1 2 3 Turn on the Model 205 Ozone Monitor and allow it to stabilize for a minimum of one hour Enter the calibration menu Main Menu Cfg Cal O3 and set the zero Z value to 0 and the slope S value to 1 00 Connect the ozone monitor to the manifold on the ozone calibration setup If a particle filter will be used in normal operation the calibration must be performed through the filter The manifold must be vented to atmosphere so that pressure does not build up in the calibration setup Connection of the Model 205 directly to a pressurized output of any device can damage the ozone monitor Verify that the flow rate into the manifold is greater than the total flow required by the ozone monitor plus any other flow demand drawing from the manifold such as a UV photometer or ozone transfer standard Measurement of Zero Air 1 2 3 Verify that the zero air supply is on and the ozone generator is off The same zero air supply used in the ozone generator must be used in the ozone generator Allow the Model 205 to sample zero air until the response is stable Record the average zero air response Measurement of Ozone Standards 1 Generate an ozone concentration slightly less th
8. 0 Air pump is not drawing sufficient flow Instrument always Solenoid valve cable is reads close to zero for not properly connected ozone concentration to circuit board As a first check hold your finger over the air inlet to determine whether air is being drawn in If there is flow measure the flow rate by removing the bottom cover and attaching a high conductance flow meter to the exit port of the pump Air flow should be greater than 0 7 L min If flow is lower check for leaks If there are no leaks replace air pump Reattach solenoid valve cable to circuit board Model 205 Dual Beam Ozone Monitor Manual Rev D 31 a NH HH rp Fa zH bo x anl P i LE a ES lia 5 FIGURE 3 1 Top view of Ozone Monitor with cover removed Model 205 Dual Beam Ozone Monitor Manual Rev D 22 FIGURE 3 2 Bottom view of Ozone Monitor with cover removed Model 205 Dual Beam Ozone Monitor Manual Rev D 33 15 U18 RH rie LCS x A30 ES 2B Technologies Inc DB NOx Ver 2087 DII 22 eve p OT C NN nee M L sa TEES TE FIGURE 3 3 Printed Circuit Board Model 205 Dual Beam Ozone Monitor Manual Rev D 34 Air Inlet 1 Use PTFE PFA or RS 232 Other Inert 12 V Power In Tubing Only BNC Analog External Analog Output 0 2 5 V Inputs 0 2 5 V e g T P RH FIGURE 3 4 Back of Instrument Model 205 Dual Beam Ozone Monitor Manu
9. Ozone Monitor Dual Beam 2B Technologies Inc OPERATION MANUAL Model 205 Copyright 2001 2011 2B Technologies Inc All rights reserved Ozone Monitor Manual Rev D TABLE OF CONTENTS Ozone Monitor Manual Rev D IDENTIFICATION RECORDS Record the following information for future reference Unit serial number Warranty start date date of receipt Model 205 Dual Beam Ozone Monitor Manual Rev D iii PRINTING HISTORY New editions are complete revisions of the manual and incorporate all previous update pages and write in instructions This manual will be revised as necessary Revisions can be in the form of new editions update pages or write in instructions Rcs ccs ces teh ei dr ns asa teehee sie October 2005 Revision se Ma September 2008 ad Eq IE September 2010 R VISION S used onesdd A A Eed pee February 2011 TRADEMARKS amp PATENTS 2B Technologies 2B Tech 2B and Ozone Monitor are trademarks of 2B Technologies Inc CONFIDENTIALITY The information contained in this manual may be confidential and proprietary and is the property of 2B Technologies Inc Information disclosed herein shall not be used to manufacture construct or otherwise reproduce the goods disclosed herein The information disclosed herein shall not be disclosed to others or made public in any manner without the expressed written consent of 2B Technologies Inc Copyright 2001 2011 2B Technologies I
10. PA under Title 40 of the Code of Federal Regulations Part 50 Appendix D 40 CFR Part 50 2 An Ozone Transfer Standard is a system a portable ozone monitor and or a portable ozone generator which can produce accurate ozone concentration standards which are quantitatively related to a primary ozone standard An example of an ozone transfer standard is the 2B Technologies Model 306 Ozone Calibration Source Ozone transfer standards must be certified before use in accordance with the procedures prescribed by the U S Environmental Protection Agency EPA under Title 40 of the Code of Federal Regulations Part 50 Appendix D 40 CFR Part 50 Equipment Required The equipment that is needed to carry out the calibration is commercially available or it can be assembled by the user Calibration using a primary ozone standard involves the generation of ozone concentrations that are simultaneously measured by a primary ozone standard and the instrument undergoing calibration This procedure requires the following equipment Zero air source Ozone generator Sampling manifold inert material such as PTFE or FEP only Sampling lines inert materials such as PTFE or FEP only UV Photometer Cle ps To Use of a certified transfer standard for calibration involves the generation of ozone concentrations using the calibrated ozone generator that are measured by the instrument undergoing calibration This procedure requires the following equipment
11. accommodate multiple data interruptions due to power failures For example one can purposely switch the instrument off move to another location and restart logging simply by turning the instrument back on To Stop Logging Data Hold the Select button down to obtain the Menu Go to the Dat submenu by clicking on Dat Choose and click on the End function This will end data logging You may now transmit the data to a computer by clicking on Xmt see below Alternatively you may return to the Menu by clicking on lt The stored data will reside in memory even when new measurements are being made and can be transmitted using the Xmt function as often as you like However all stored data are lost once logging is started again using the Log function Thus you should always transmit your data to a computer before restarting logging If you fail to End logging prior to transmitting the data using the Xmt function the instrument will automatically execute the End function for you prior to transmitting the data To Transmit Logged Data to a Computer Using the Serial Port Connect the serial port of the instrument to the serial port of your computer using the cable provided If your computer does not have a serial port you can use the USB port by means of a serial to USB adapter Such adapters are available in most computer stores or can be supplied by 2B Tech Enable a data acquisition program on the computer such as Microsoft Hyperterminal ava
12. al Rev D 35 6 PARTS LIST The following list includes those parts that are user serviceable Part Number SCRBINT ZEROEXT OZLAMPAS OZVLV2 OZDSP OZPUMP205 OZCELL DEW SERCABL 110ADP PWRWIR 12VADP TEFTYG SILTUB FILTERHS FILTERPK Description Ozone scrubber internal Ozone scrubber external Lamp and connector Solenoid valve LCD display and cable Air pump Absorption cell DewLine Serial port cable to computer 110 V AC adapter Bare wire power cable 12 V DC cigarette lighter adapter Teflon lined Tygon tubing Silicone tubing PVDF Filter Housing 25 mm with 4 Fittings Package of Ten 25 mm PTFE Particle Filters 5 6 u pore size Model 205 Dual Beam Ozone Monitor Manual Rev D 36 7 Service Log New O New Pump New Scrubber main backup Lamp ust Date Calibrated Cleaned Model 205 Dual Beam Ozone Monitor Manual Rev D 37 Date Calibrated Cleaned New O4 New Pump New Other Scrubber main backu Lam Model 205 Dual Beam Ozone Monitor Manual Rev D 38
13. alibration setup or the ozone monitor being checked The most likely problems in the ozone monitor are leaks a malfunctioning ozone scrubber a contaminated valve or contamination in the optical setup See the Troubleshooting section of the manual 5 TROUBLESHOOTING If the instrument fails to operate correctly common problems can be identified and corrected using Table If the problem cannot be easily corrected please contact Customer Service at 2B Tech via our web ticketing software at www twobtech com techsupport Alternatively you can email us at techsupport twobtech com or call us at 1 303 273 0559 If we mutually determine that the instrument cannot be repaired onsite we will provide you with a Return Authorization number and a short form to be filled out and returned to our Service Department along with the instrument The figures following Table provide a guided tour of the Model 205 Ozone Monitor so that critical components and connectors may be easily identified A list of serviceable parts is provided in Section 4 at the end of this manual Table I Troubleshooting the Ozone Monitor for performance problems Instrument does not Power not connected Check external power turn on properly or circuit connection for reverse breaker open polarity or a short and wait a few minutes for the thermal circuit breaker to reset Power cable not Remove top cover and connected to circuit disconnect and board
14. an the concentration range of interest and allow the ozone generator to warm up for at least 5 minutes The same zero air supply used for making zero air measurements must be used in the ozone generator 2 Allow the Model 205 Ozone Monitor to sample the ozone concentration standard until a stable response is measured Model 205 Dual Beam Ozone Monitor Manual Rev D 26 3 Record the average response of the ozone monitor as well as either the average response of the UV photometer the transfer standard 4 Generate several other ozone concentration standards At least 5 ozone concentration standards are recommended over the range of interest 5 For each ozone concentration standard record the response of the ozone monitor as well as either the response of the UV photometer or the transfer standard Calibration Curve 1 Plot the Model 205 Monitor responses x axis versus the corresponding standard ozone concentrations y axis 2 Fit the data to a straight line y mx b using the linear regression technique to determine the calibration relationships 3 Determine if any points deviate significantly from the line which is an indication of an error in determining the calibration curve The error may be due to the calibration setup or the ozone monitor being calibrated The most likely problems in the ozone monitor are leaks a malfunctioning ozone scrubber a contaminated valve or contamination in the optical setup See the T
15. as entsprechende Kapitel in der Bedienungsanleitung lesen sollte WARNHINWEIS Wenn das Ger t nicht wie durch die Firma 2B Technologies Inc USA vorgeschrieben und im Handbuch beschrieben betrieben wird k nnen die im Ger t eingebauten Schutzvorrichtungen beeintr chtigt werden DUTCH A OPGELET ledere handeling binnenin het toestel kan beschadiging veroorzaken Om iedere mogelijk gevaarlijke shock te vermijden moet de aansluiting met het net verbroken worden v r het openen van het toestel OPGELET Het symbool A geeft aan dat de gebruiker de instructies in de handleiding moet raadplegen OPGELET Indien het toestel niet gebruikt wordt volgens de richtlijnen van 2B Technologies Inc USA gelden de veiligheidsvoorzieningen niet meer vii Model 205 Dual Beam Ozone Monitor Manual Rev D 1 OZONE MONITOR INTRODUCTION The 2B Technologies Dual Beam Ozone Monitor is designed to enable accurate measurements of atmospheric ozone over a wide dynamic range extending from a limit of detection of 1 part per billion by volume ppbv to an upper limit of 100 parts per million ppmv based on the well established technique of absorption of ultraviolet light at 254 nm The Ozone Monitor is light weight 4 7 Ib 2 1 kg and has a low power consumption 5 watt relative to conventional instruments and is therefore well suited for applications such as e vertical profiling using balloons kites remotely piloted aircraft a
16. bv or 296 ACCUTACY REA ae ie Dati dba higher of 1 0 ppbv or 296 Model 205 Dual Beam Ozone Monitor Manual Rev D 2 OPERATION Please read all the following information before attempting to install the Ozone Monitor For assistance please call 2B Technologies at 303 273 0559 NOTE Save the shipping carton and packing materials that came with the Ozone Monitor If the Ozone Monitor must be returned to the factory pack it in the original carton Any repairs as a result of damage incurred during shipping will be charged Shipping Box Contents Open the shipping box and verify that it contains the following Ozone monitor 110 220 V AC power adapter Cigarette lighter adapter Bare wire 12 V DC battery adapter Serial port cable Zeroing cartridge Ozone Monitor manual on CD Ozone Monitor birth certificate Quality control data sheet and graph Three external jacks for analog inputs DOS LO Or os If anything is missing or obviously damaged contact 2B Technologies immediately Operation of the Ozone Monitor To operate the Ozone Monitor connect it to an external power source and turn the instrument on by flipping the front panel switch The instrument requires a 12 V DC source which can be supplied by 1 the 110 220 V AC power adapter 0 42 amp or higher 2 a cigarette lighter adapter plugged into a 12 V DC source such as found in an automobile or many light aircraft or 3 a 12 V battery The source can be in the ra
17. ckup pump is tried again and three carats appear on the display If the power to the instrument is cycled the carat symbols will disappear and the instrument will attempt to use the main pump again This feature allows the user to obtain the full lifetime of an air pump without significant loss of data Once the main pump fails the user typically will have up to several months of operation to replace the main pump The volumetric flow rate is continuously measured output over the serial port as part of the data line and stored in the internal data logger 3 MAINTENANCE The Ozone Monitor is designed to be nearly maintenance free The only components that require routine maintenance are the ozone scrubber which should be changed at least once every six months of operation and the air pump which has rated lifetime of 5000 hours of operation as discussed below Other user serviceable components include the lamp clock battery and solenoid valves which are easily replaced should they fail In addition the inlet filter user supplied should be changed as recommended by the filter manufacturer The miniature air pumps have a lifetime specified by the manufacturer to be in the range 3 000 to 7 000 hours with a mean of about 5000 hours depending on environmental factors This corresponds to about 7 months of continuous operation The pumps selected for use are based on several years of experience with different pump manufacturers and mode
18. cross section extinction coefficient as used in other commercial instruments A new ozone measurement is made every 2 s for both cells based on updated values of J and Z These two values are Model 205 Dual Beam Ozone Monitor Manual Rev D averaged and then output as both serial data and an analog voltage between 0 and 2 5 V The data may also be stored in the instruments internal memory and or on a flash memory card if the instrument has the flash memory option The logarithm of equation 1 is calculated in the microprocessor of the instrument with sufficient accuracy to provide five orders of dynamic range ozone mixing ratios are measured up to 100 ppmv The shorter path length of the 2B Ozone Monitor also contributes to the wide dynamic range which is limited at the high end by the absorption beginning to become optically thick base 10 optical absorbance 0 2 The pressure and temperature within the absorption cells are measured so that the ozone concentration can be expressed as a mixing ratio in parts per billion by volume ppbv The instrument displays and records the cell temperature and pressure in addition to the ozone mixing ratio The cell pressure is displayed and logged in units of either Torr or mbar and the cell temperature in units of either C or K In principle the measurement of ozone by UV absorbance requires no external calibration it is an absolute method However non linearity of the photodiode response and ele
19. ctronics can result in a small measurement error Therefore each instrument is compared with a NIST traceable standard ozone spectrophotometer in the laboratory over a wide range of ozone mixing ratios typically 0 300 ppbv for atmospheric applications These results are used to calibrate the Ozone Monitor with respect to an offset and slope gain or sensitivity The corrections for offset and slope are recorded in the instrument Birth Certificate and on a calibration sticker that can be viewed by removing the top cover of the instrument These calibration parameters are entered into the microprocessor memory prior to shipment The user may change the calibration parameters from the front panel if desired It is recommended that the Ozone Monitor be recalibrated at least once annually and preferably more frequently The offset may drift due to temperature change or chemical contamination of the absorption cell As discussed below an accurate offset correction can be measured from time to time using the ozone scrubber supplied with the instrument The user may change the slope and offset calibration parameters by entering the Menu Model 205 Dual Beam Ozone Monitor Manual Rev D OZONE MONITOR SPECIFICATIONS Power Requirements 11 14 V DC nominally 420 mA at 12 V 5 0 watt Dimension S NR PH 3 5 x 8 3 x 11 6 Weight c acssstu tapete on nn 4 7 lbs 2 1 kg Weight with case removed 1 6 Ib 0 7 kg PI CISION E RUE higher of 1 0 pp
20. d exclusive remedies In no event shall 2B Technologies be liable for direct indirect special incidental or consequential damages including loss of profits whether based on contract tort or any other legal theory The Ozone Monitor manual is believed to be accurate at the time of publication and no responsibility is taken for any errors that may be present In no event shall 2B Technologies be liable for incidental or consequential damages in connection with or arising from the use of the Ozone Monitor manual and its accompanying related materials Warranty is valid only for the country designated on the 2B Technologies quote or invoice Model 205 Dual Beam Ozone Monitor Manual Rev D vi Warnings ENGLISH A WARNING Any operation requiring access to the inside of the equipment could result in injury To avoid potentially dangerous shock disconnect from power supply before opening the equipment WARNING This symbol N on the instrument indicates that the user should refer to the manual for operating instructions WARNING If this instrument is used in a manner not specified by 2B Technologies Inc USA the protection provided by the instrument may be impaired FRAN AIS Aa TTENTION Chaque op ration l int rieur de l appareil peut causer du pr judice Afin d viter un shock qui pourrait tre dangereux disconnectez l appareil du r seau avant de l ouvrir ATTENTION Le symbol A indique que l utilisateur do
21. hese tests include Setup check Ozone loss test Linearity check Intercomparison test Setup Check A visual inspection of the calibration setup should be performed before calibration to verify that the setup is in proper order All plumbing connections should be checked and verified to follow the manufacturer s instructions Any obvious leaks should be fixed and the manifold and sampling lines should be checked for general cleanliness For more information refer to the manufacturer s User Manual for the primary ozone standard or ozone transfer standard Ozone Loss Test Some ozone may be lost in the calibration setup due to reaction with the walls of the manifold and sampling lines Any significant loss of ozone must be measured and be subsequently applied to correct the calibration Model 205 Dual Beam Ozone Monitor Manual Rev D 24 measurements For more information refer to the manufacturer s User Manual for the primary ozone standard or ozone transfer standard Linearity Check Since the Model 205 is inherently linear over several orders of magnitude a linearity check provides a test that the instrument is operating properly Instrument linearity can be checked by comparison to an ozone standard see Calibration Procedure Calibration Curve or by dilution of an ozone measurement To check the instrument linearity by dilution of an ozone measurement generate and measure a concentration of ozone near the upper range of ozo
22. ilable on most Windows platforms usually in Start All Programs Accessories Communications Hyper Terminal or Terra Term Pro which can be downloaded at http hp vector co jp authors VA002416 teraterm html As mentioned earlier the disadvantage of Hyperterminal is that it has a 500 line buffer limit However both programs may used to log an unlimited number of data lines to a file on your computer For more details see our Tech Note 007 here http www twobtech com tech_notes TN007 paf Model 205 Dual Beam Ozone Monitor Manual Rev D 13 Hold down the Select button to obtain the Main Menu Go to the Dat submenu by clicking on Dat Next click on Xmt The message Logged Data will be written to the serial port followed by a carriage return and all of the lines of logged data After all data are transmitted the message End Logged Data and a carriage return are written After transmission is complete you can return to any position in the menu or resume ozone measurements The logged data continues to be available for transmission until a new data log is started To Average Data Hold down the Select button to obtain the Menu Select and click on Avg to obtain the Avg menu Avg Menu 2s 10s 1m 5m th Use single clicks to move the cursor to 2s 10s 1m 5m or 1h for averaging times of 2 s no averaging 10s 1 min 5 min or 1 hr averaging respectively Then click on the averaging time you want to use To return to
23. instrument is subject to some drift and variation in response making it necessary to periodically check the calibration Dynamic calibration is a multipoint check where gas samples of known concentration are sampled by the instrument in order to determine a calibration relationship For more information on calibration of ozone monitors refer to the Code of Federal Regulations Title 40 Part 50 Appendix D and the EPA s Technical Assistance Document for the Calibration of Ambient Ozone Monitors Calibration is the process of adjusting the gain and offset of the Model 205 Ozone Monitor against some recognized standard The reliability of the data collected from any analytical instrument depends on the accuracy of the calibration which is largely dependent upon its analytical traceability to a reference material or reference instrument calibration Because of the instability of ozone the certification of ozone concentrations in a compressed gas cylinder is impossible due to loss of ozone over time When ozone concentration standards are required the ozone must be generated and certified on site Ozone standards can be classified into two basic types 1 A Primary Ozone Standard is the combination of an ozone generator and an ozone monitor based on UV absorbance a UV photometer that has Model 205 Dual Beam Ozone Monitor Manual Rev D 22 been setup in accordance with the procedures prescribed by the U S Environmental Protection Agency E
24. it consulter le manuel d instructions ATTENTION Si l instrument n est pas utilis suivant les instructions de 2B Technologies Inc USA les dispositions de s curit de l appareil ne sont plus valables ITALIANO A ATTENZIONE Qualsiasi intervento debba essere effettuato sullo strumento pu essere potenzialmente pericoloso a causa della corrente elettrica Il cavo di alimentazione deve essere staccato dallo strumento prima della sua apertura ATTENZIONE Il simbolo A sullo strumento avverte l utilizzatore di consultare il Manuale di Istruzioni alla sezione specifica ATTENZIONE Se questo strumento viene utilizzato in maniera non conforme alle specifiche di 2B Technologies Inc USA le protezioni di cui esso dotato potrebbero essere alterate ESPANOL Aa TENCION Cualquier operaci n que requiera acceso al interior del equipo puede causar una lesi n Para evitar peligros potenciales desconectarlo de la alimentaci n a red antes de abrir el equipo ATENCION Este s mbolo A en el instrumento indica que el usuario deber a referirse al manual para instrucciones de funcionamiento ATENCION Si este instrumento se usa de una forma no especificada por 2B Technologies Inc USA puede desactivarse la protecci n suministrada por el instrumento DEUTSCH L wa RNHINWEIS Vor dem Offnen des Ger tes Netzstecker ziehen WARNHINWEIS Dieses Su dem Ger t weist darauf hin dab der Anwender zuerst d
25. l 205 Dual Beam Ozone Monitor Manual Rev D 29 Select switch does not work Serial port does not work Required calibration parameters are outside the adjustable range 9 ppbv offset and or 49 slope when calibrated using a standard ozone source or reliable ozone instrument Cable not properly connected between select switch and circuit board Cable not properly connected between serial port 9 pin connector and circuit board Wrong serial cable used Wrong baud rate selected Ozone scrubber is contaminated Flow path is contaminated Solenoid valve is contaminated and not opening and closing properly Remove top cover and reconnect select switch cable to circuit board Remove top cover and reconnect serial port cable to circuit board A straight through serial cable is provided Some data collection devices require a cross over cable in which pins 1 and 3 are exchanged between the two ends of the cable Use a cross over cable or additional connector that switches pins 1 and 3 Make sure that the baud rate chosen in the menu matches the baud rate setting of your data acquisition program Replace ozone scrubber Be sure to use an inlet filter to remove particulate matter Clean flow path with methanol following the Cleaning Procedure Remove solenoid valve rinse with methanol dry with zero air and replace Model 205 Dual Beam Ozone Monitor Manual Rev D 3
26. lect I O to give the input output menu for example 1 0 Menu Bdr Ext LCD Selecting Bdr and clicking allows you to change the baud rate used for data transmission The choices are 2400 4800 and 19200 bps The submenu appears as 2400 4800 19200 lt After clicking on a chosen baud rate the display returns to the I O menu To Turn the Display Light On and OFF Selecting LCD from the 1 0 submenu allows you to turn the light of the front panel display on and off To conserve power use the Select switch to set the LCD submenu to OFF To Turn the Analog Inputs On and Off and Change the Analog Output Scaling Factor Model 205 Dual Beam Ozone Monitor Manual Rev D From the I O submenu select the Ext submenu which appears as Ext Menu V_IN V_Out lt The two submenus allow one to turn the analog inputs on and off V_IN and to change the scaling factor for the analog voltage output V Out Analog Inputs To turn the three analog inputs on or off select the V_IN submenu If the analog voltage inputs are turned on you will see VIN Menu Voltage InZON lt You may the use the Select switch to toggle between ON and OFF If no analog inputs are being used it is advantageous to turn the analog inputs off in order to increase the number of data lines that can be logged in the internal data logger and to reduce clutter in the output data lines sent via the serial port Analog Output An analog output is provided via a BNC connect
27. ls An external long life air pump is available as an upgrade A wide range of Technical Notes are provided on the 2B Tech website The complete list with links may be found at clickable link www twobtech com tech notes htm These Tech Notes are continuously updated and new ones created Tech Notes specific to maintenance and maintenance and repairs may be downloaded from clickable links from this pdf TN 017 Procedure for Replacing Lamp Housing in Model 202 and Model 205 Ozone Monitors Model 205 Dual Beam Ozone Monitor Manual Rev D 21 TN 018 Cleaning Procedure for Model 202 Ozone Monitor TN 021 Procedure for Replacing Air Pumps in Model 202 and Model 205 Ozone Monitors The above are clickable links from the pdf version of the manual If you are reading this from a hard copy the links are www twobtech com TNxxx pdf where xxx is the Tech Note number Also please note that all 2B Tech instrument manuals are posted online at www twobtech com downloads htm For your convenience a Service Log which may be printed is provided at the end of this manual for recording calibrations replacement of pumps lamps etc Records of repairs made at 2B Tech are maintained in a database at 2B Technologies as well That database also includes detailed information about the construction and initial calibration of your instrument including digital of photos of the interior its interior 4 CALIBRATION Every analytical
28. ly using the 2B Technologies Display and Graphing Software free download from http twobtech com software htm Other terminal emulation software such as HyperTerminal a program provided with Windows or Tera Term Pro may be used as well The ozone mixing ratio ppbv internal cell temperature K or C cell pressure Torr or mbar volumetric flow rate cc min values of three external analog inputs in volts if activated from the menu date and time are sent as comma delimited ASCII text to the serial port 2400 4800 or 19 200 baud 8 bits no parity 1 stop bit every 2 seconds 10 seconds 1 minute 5 minutes or 1 hour depending on the averaging time selected from the microprocessor menu Time is provided in 24 hour military format and the date is given in European style day month year Model 205 Dual Beam Ozone Monitor Manual Rev D A typical data line might read 67 4 35 3 980 6 1545 1 3876 2 3143 0 1875 15 10 01 18 31 27 where Ozone 67 4 ppbv Cell temperature 35 3 C may be expressed in K if chosen from menu Cell pressure 980 6 mbar may be expressed in Torr if chosen from menu Volumetric flow rate 2 1545 cc min Analog input A 1 3876 volts Analog input B 2 3143 volts Analog input C 0 1875 volts Date October 15 2001 Time 2 6 31 27 pm The three external inputs are omitted from the data line if they are turned off using the menu as described below The analog inputs allow measurements made by
29. n of air flow or loss of catalytic activity by the internal ozone scrubber but it also can be adjusted Once the zero of the instrument is corrected the slope may be adjusted so that the instrument readout agrees with a standard ozone source or with the readout from another instrument whose calibration is considered to be accurate To change the calibration parameters select Cfg from the Main Menu Cfg Menu D T Cal I O Unt Now use the rotary select switch to select and click on Cal The following submenu with the values of the current calibration parameters will then appear Cal Menu Zz 2 Sz1 01 Here Z is the offset applied in this case 2 ppbv and S is the slope applied in this case 1 01 The value of Z is added to the measured ozone value and the value of S is then multiplied by the measured ozone value For example if the instrument reads an average of 3 ppbv with the external scrubber in place the value of Z should be set to 3 If after correction for the zero the instrument consistently reads 296 low the value of S should be set to 1 02 When the Cal Menu first appears the Z will be underlined with a cursor You may rotate the Select switch to choose the calibration parameter S or Z A single click on S or Z will select that parameter for change and activate a blinking cursor Once S or Z is selected its value can be changed by rotating the Select switch to the left or right After choosing the desired value a si
30. nc All rights reserved Model 205 Dual Beam Ozone Monitor Manual Rev D WARRANTY STATEMENT 2B Technologies Inc warrants its products against defects in materials and workmanship 2B Technologies will at its option repair or replace products which prove to be defective The warranty set forth is exclusive and no other warranty whether written or oral is expressed or implied 2B Technologies specifically disclaims the implied warranties of merchantability and fitness for a particular purpose Warranty Periods The warranty period is one 1 year from date of receipt by the purchaser but in no event more than thirteen 13 months from original invoice date from 2B Technologies Inc Warranty Service Warranty Service is provided to customers via web ticket email and phone support Monday Friday from 9 00 a m to 5 00 p m Mountain Time USA The preferred method of contacting us is through our web ticketing software at www twobtech com techsupport This way all technical staff at 2B Tech will be alerted of your problem and be able to respond When you receive an email reply please click on the Ticket link provided to continue to communicate with us directly over the internet The web ticket approach to customer service allows us to better track your problem and be certain that you get a timely response We at 2B Tech pride ourselves on the excellent customer service we provide You may also contact us by email at techsuppor
31. nd other aircraft where space and weight are highly limited e long term monitoring at remote locations where power is highly limited e urban arrays of ground based detectors e personal exposure monitoring for studies of health effects of air pollutants Advantages of the Model 205 Dual Beam Ozone Monitor over the 2B Technologies single beam Model 205 Ozone Monitor is a factor of five faster response time measurements made every 2 s for the Model 205 vs every 10 s for the Model 205 and greater stability against zero drift When data are averaged for 10 s the Model 205 provides better precision as well Theory of Operation Absorption of UV light has long been used for measurements of atmospheric ozone with high precision and accuracy The ozone molecule has an absorption maximum at 254 nm coincident with the principal emission wavelength of a low pressure mercury lamp Fortunately few molecules found at significant concentrations in the atmosphere absorb at this wavelength However interferences such as organic compounds containing aromatic rings can occur in highly polluted air Mercury vapor can be a significant interference inside buildings where mercury spills have occurred in the past and in the vicinity of certain mining operations Figure 1 is a schematic diagram of the ozone monitor Ozone is measured based on the attenuation of light passing through two separate 15 cm long absorption cells fitted with quartz windows A single lo
32. ne monitor 80 of full scale is recommended Additional ozone concentrations should be generated by accurately diluting the ozone flow with zero air and each concentration should be measured once the instrument reaches a stable response The accuracy of the linearity test relies on the accuracy of the flow meters used to perform the dilution The percent of non linearity is calculated from the formula Rae F F 2 eue E x100 3 where R Dilution ratio F Ozone generator flow Fa Diluent zero air flow E Linearity error in percent C Measured concentration of original concentration C Measured concentration of diluted concentration The linearity error should not be greater than 596 If the error is greater than 596 the accuracy of the flow dilution should be checked before assuming that the ozone monitor is not linear Note that the inherent linearity of the Model 205 is better than the error calculated in this linearity check due to the uncertainty introduced by the flow measurements Intercomparison Test Model 205 Dual Beam Ozone Monitor Manual Rev D 25 Comparison of the calibration setup with other ozone standards is a good check of the overall accuracy of the setup If measurements from another ozone standard are found to deviate from the calibration setup greater than the instrument specifications one of the calibration setups is not accurate CALIBRATION PROCEDURE A multipoint calibration
33. nge 11 14 V DC without any detrimental effects on the measurement When using a battery be certain to attach the Model 205 Dual Beam Ozone Monitor Manual Rev D positive red and negative black wires correctly A circuit breaker and diode are installed on the circuit board in case of an electrical short or incorrect battery attachment If activated the breaker will reset itself after a few minutes Lead acid batteries are available from numerous manufacturers in a wide range of sizes and amp hour ratings The larger of these such as those for automobiles or boats will supply power for up to several days Battery packs in the correct voltage range may be constructed from nickel cadmium rechargeable or lithium light weight but not rechargeable batteries for operation for a few hours Battery options available through 2B Technologies may be found on our webpage www twobtech com Once turned on the instrument will display the version number of the software installed on the microprocessor followed by a display of the time and date After a few seconds the instrument will start displaying readings for ozone and the temperature and pressure of the absorption cell The first dozen readings requiring about two minutes will be spurious with large positive and negative swings due to the rapid warmup of the lamp and electronics Also ozone readings may be inaccurate during the 10 20 minutes required for the lamp photodiode and internal tem
34. ngle click turns off the blinking cursor and allows you to scroll to the other parameter or to lt to exit the submenu Once the values of Z and S are set clicking on lt will return the display to the Cfg menu and another click on lt will return to the Main Menu The calibration parameters reside in non volatile memory and are not affected by power failures To Set the Time and Date Model 205 Dual Beam Ozone Monitor Manual Rev D From the Main Menu select the Cfg submenu Next select the D T submenu The display will read for example D T 14 32 21 17 10 2008 meaning that it is 21 seconds after 2 32 p m on October 17 2008 military time and European date To change a number in the date and time rotate the Select switch to underline the numeral you want to change A single click then causes a blinking cursor to cover that numeral The number can then be changed by rotating the Select switch Once the number is correct click on the Select switch to turn off the blinking cursor You may now rotate the Select switch to choose another numeral to change Once the time and date is correct clicking on lt will set the internal clock to that time and return the display to the Cfg menu As in setting a digital watch the seconds should be set in advance of the real time since the clock starts to run again only when the set time is entered in this case by clicking on lt To Change the Output Baud Rate From the Cfg submenu se
35. or at the back of the instrument for those who want to record their ozone concentration data with a chart recorder or external logger The full scale of the analog output is 2 5 V To change the analog output voltage scaling factor select V_Out from the Ext Menu VOUT Menu 1V 000200 ppb lt In this example the output scaling factor is set as 1 Volt 200 ppb Since the maximum output voltage is 2 5 V the maximum output concentration in this case is 500 ppb and 1 ppb will provide an output of 5 mV You can use the select switch to change the scaling factor to the value of your choice by selecting and changing the individual digits in the scaling factor To Change the Units for Internal Temperature and Pressure From the Cfg menu select Unt to give the following submenu for example Units Menu T C P mbar Rotating the Select switch will cycle the cursor between temperature T and pressure P Temperature units may be selected as either Kelvin K or Model 205 Dual Beam Ozone Monitor Manual Rev D 17 Celsius C by first clicking to obtain the blinking cursor and then rotating the Select switch to obtain the desired units Pressure units may be selected as either torr or mbar A click on lt returns the display to the Cfg menu To Output Data from Both Detection Cells The instrument has a service menu for diagnostics The service menu is accesses by choosing Sve from the main Menu The following submenu choices will then appear
36. perature of the absorption cell to stabilize Inlet tubing may be attached to the inch nylon Swagelok fitting on the back of the instrument The inlet tubing should be made of PTFE Teflon PFA or some other inert material that does not destroy ozone and that does not desorb plasticizers and other organics that can contaminate the flow path The length of tubing should be kept as short as possible not more than a few feet to minimize ozone destruction Tygon polypropylene which may look like Teflon and metal tubing should not be used Teflon lined Tygon tubing which is used inside the instrument provides the flexibility of Tygon with the inertness of Teflon A Teflon inlet filter is highly recommended to prevent internal contamination of the tubing and absorption cell by particulate matter The filter should be tested for ozone loss by measuring ambient ozone with and without the filter attached Filters and filter holders are available through 2B Technologies See our website www twobtech com If the instrument is being flown the inlet should not point into the wind because the resulting pressure fluctuations will cause a noisy signal Although the instrument compensates for temperature drift if strong temperature fluctuations are expected as in vertical profiling applications using balloons the instrument should be placed in a thermally insulated box in order to slow the rate of temperature change Model 205 Dual Beam O
37. reconnect power cable to circuit board Instrument turns on Burned out air pump Remove top cover and then powers off unplug air pump Turn Model 205 Dual Beam Ozone Monitor Manual Rev D 28 Display is blank or nonsense Cell temperature reads low by several 10 s of degrees Readings are noisy with standard deviations greater than 2 0 ppbv Analog output is constant or does not track front display Bad connection of display to circuit board Absent or loose connection of temperature probe cable to circuit board Lamp output is weak Excessive vibration Flow path contaminated Cable not properly connected between analog output BNC and circuit board Wrong scaling factor selected In menu instrument on if it remains running then the air pump motor is burned out and shorting Replace air pump Remove top cover and reconnect display to circuit board Check solder connections to display Remove top cover and reattach connector to circuit board Remove top cover and check lamp connection to circuit board Run Lamp Test from menu If photodiode voltage is less than 0 7 V replace lamp Provide additional vibration insulation for the instrument such as a foam pad Clean flow path with methanol according to the Cleaning Procedure Remove top cover and reconnect cable between analog output and circuit board Check and reset analog output scaling factor in the Menu Mode
38. roubleshooting section of the manual 4 The slope of the line is the gain factor S and the intercept is the offset Z that need to be applied to the ozone monitor response to calibrate it to the primary ozone standard If the intercept is outside of the range from 10 to 10 or the slope is outside of the range from 0 90 to 1 10 this is an indication of a problem in the calibration setup or the ozone monitor being calibrated The most likely problems in the ozone monitor are leaks a malfunctioning ozone scrubber a contaminated valve or contamination in the optical setup See the Troubleshooting section of the manual 5 Enter the calibration menu Main Menu Cfg Cal O3 in the instrument software and set the calibration parameters PERIODIC ZERO AND SPAN CHECKS To ensure the quality of the ozone monitor data periodic zero and span checks can be performed by following the steps below 1 Azero check is performed by sampling zero air with the Model 205 following the Measurement of Zero Air section above 2 Aspan check is performed by sampling an ozone concentration at the high end of the concentration range of interest following the Measurement of Ozone Standards section above 3 Average measurements from the zero check or span check should be within the instrument specifications If the measurements are not Model 205 Dual Beam Ozone Monitor Manual Rev D 27 within specifications this is an indication of problem in the c
39. source of noise If the noise is greater with the pump on Model 205 Dual Beam Ozone Monitor Manual Rev D 18 this is a good indication that the flow path needs to be cleaned and the ozone scrubber replaced During the lamp test the instrument measures effective ozone concentrations in the two detection cells without switching the solenoid valve on and off These are electronic zeros and should after a few readings settle down to a few ppbv If either of the values are outside the range 9 to 9 the instrument may not be operating correctly The display also gives a standard deviation of the electronic zeros For best results the standard deviations should be not greater than 2 5 Note that because less averaging is involved this is not equal to the overall precision of the instrument it is a diagnostic test of lamp fluctuations The lamp test also measures and outputs the voltage levels of detector A and detector B For best performance both voltages should be in the range 0 7 2 0 volts For detector voltages less than about 0 7 volts the data may be noisy due to insufficient light intensity to make precise measurements If the voltage is zero the lamp is not ignited and may have burned out For voltages above 2 5 volts the A D converter is saturated and the measured ozone value will always be zero This could happen if the instrument is very hot so that the lamp output is too bright To exit the Lmp test mode hold in the Selec
40. t switch and release to return to the Main Menu To Read the Number of Hours of Ozone Monitor Use The instrument keeps track of the total number of hours of use This is helpful for determining when the instrument should be serviced a pump replaced etc To read the number of hours of operation choose Menu Cfg I O Hrs Accessing the Serial Menu Measurements and logging tasks can be accessed via the serial port or the USB using a terminal emulator such as Tera Term Pro or HyperTerminal running on an attached computer Commands can be sent using the terminal emulator set with the properties listed in the section of this manual entitled Collecting Data over the Serial Port in Real Time Listed below are the lower case letters that are commands for performing certain operations while the instrument continues to measure Model 205 Dual Beam Ozone Monitor Manual Rev D l t e h m Start logging and write over existing logged data Transmit logged data End logging Output serial data line header Serial menu If the letter m is sent as a command menu gt will be displayed in the terminal emulator window When the serial menu is accessed the instrument is no longer making measurements it is waiting for the next command to be entered The following is the list of menu items accessible from this point t D N yaTaQac lt crtctd x Start logging and write over existing logged data Transmit logged data End logging Ou
41. t twobtech com or by phone at 1 303 273 0559 In either case a web ticket will be created and future communications with you will be through though that ticket Initial support involves trouble shooting and determination of parts to be shipped from 2B Technologies to the customer in order to return the product to operation within stated specifications If such support is not efficient and effective the product may be returned to 2B Technologies for repair or replacement Prior to returning the product a Repair Authorization Number RA must be obtained from the 2B Technologies Service Department We will provide you with a simple Repair Authorization Form to fill out to return with the instrument Model 205 Dual Beam Ozone Monitor Manual Rev D Shipping 2B Technologies will pay freight charges for replacement or repaired products shipped to the customer site Customers shall pay freight charges for all products returning to 2B Technologies Conditions The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance adjustment calibration or operation by customer Maintenance adjustment calibration or operation must be performed in accordance with instructions stated in the Ozone Monitor manual Usage of maintenance materials purchased from suppliers other than 2B Technologies will void this warranty Limitation of Remedies and Liability The remedies provided herein are the Customer s sole an
42. the Main Menu click on To exit the Main Menu and start acquiring data click on lt again While in averaging mode the current 2 s measurement is displayed alternately with the average value as discussed above Averaged data may be logged thereby greatly extending the length of time that the data logger can be used To Set the Calibration Parameters The instrument is calibrated at the factory where slope and offset parameters are entered into the instrument s memory These preset calibration parameters are given in the instrument s Birth Certificate and recorded on the calibration sticker viewable with the top cover removed However the calibration parameters may be changed by the user For example it may be desirable to provide a positive offset by a known amount e g 10 ppbv if the analog output is being used for external data logging since the analog output does not go negative below zero ppbv Because of noise and or an inherent offset some measured values will be below zero at very low ozone mixing ratios or while zeroing the instrument with an external scrubber Also the instrument zero Model 205 Dual Beam Ozone Monitor Manual Rev D may drift by a few ppbv over time For this reason frequent zeroing of the instrument using an external ozone scrubber to determine the offset is recommended Any change in the slope gain of the instrument is likely due to a serious problem such as contamination an air leak obstructio
43. tput serial data line header Display list of possible averaging times and the number that must be entered to change to the desired averaging time Display current zero calibration setting and wait for new setting followed by a carriage return Display current slope calibration setting and wait for new setting followed by a carriage return Clock menu display current date and time and wait for d or t to be entered From clock menu d Asks to enter date in DDMMYY format t Asks to enter time in HHMMSS format Perform lamp test with pump on Perform lamp test with pump off Output raw ozone for both A and B channels before averaging Stop outputting raw ozone for both A and B channels Turn the LCD backlight on Turn the LCD backlight off Output the serial header also available during measurements Output the help menu Exit menu and return to measuring Backup Pump The Model 205 Ozone Monitor has two air pumps installed One serves as the main air pump and the other as a backup pump If the air flow rate drops below a critical value the main pump is powered off and the backup air pump is powered on When Model 205 Dual Beam Ozone Monitor Manual Rev D 20 the backup pump is activated the carat symbol appears in the left hand corner of the display If the second pump fails the backup pump is turned off and the main pump is powered again At this point two carat symbols appear on the display If the flow is still out of range the ba
44. w pressure mercury lamp is located on one side of the absorption cells and photodiodes are located on the opposite side of the absorption cells The photodiodes have built in Model 205 Dual Beam Ozone Monitor Manual Rev D interference filters centered on 254 nm the principal wavelength of light emitted by the mercury lamp An air pump draws sample air into the instrument at a flow rate of approximately 1 5 L min A pair of solenoid valves switch in unison so as to alternately send ozone scrubbed air and unscrubbed air through the two absorption cells Thus the intensity of light passing through ozone scrubbed air l is measured in Cell 1 while the intensity of light pass through unscrubbed air id measured in Cell 2 Every 2 seconds the solenoid valves switch changing which cell receives ozone scrubbed air and which cell receives unscrubbed air O Scrubber All paths Zero Cycle gt Valve 1 Valve 2 Air Inlet Figure 1 Schematic diagram of the Model 205 Dual Beam Ozone Monitor Ozone concentration is calculated for each cell from the measurements of J and according to the Beer Lambert Law 1 fI C In al e where is the path length 15 cm and o is the absorption cross section for ozone at 254 nm 1 15 x 10 cm molecule or 308 atm cm which is known with an accuracy of approximately 196 The 2B Technologies instrument uses the same absorption
45. zone Monitor Manual Rev D Measurement of the Zero Offset The electronic zero of the instrument may be measured by attaching an ozone destruction cartridge to the air inlet for a period of 5 10 minutes For an accurate measurement the instrument must have been turned on long enough for the internal temperature to stabilize The observed offset which can amount to a few ppbv can be corrected by changing this calibration parameter from the front panel as described below or by correcting the data at a later time Collecting Data from the Analog Output The data may be logged in real time using a data logger attached to the BNC analog output The range of the analog output is 0 2 5 V The output is scaled according to a sensitivity you define in the menu For example you may define 1V 100 ppb In that case the maximum output is 250 ppb There is a small positive offset typically 2 mV in the analog output but this offset varies from instrument to instrument The offset can be measured by simultaneously observing the panel display and measuring the analog output with a voltmeter Collecting Data over the Serial Port in Real Time To transmit data to a computer over the serial port in real time connect the Ozone Monitor to the serial port of the computer using the 9 pin cable provided Note that this is a straight through female female serial cable A cross over cable will not work Start your data acquisition software preferab

Download Pdf Manuals

image

Related Search

Related Contents

Limpiador Jabonoso MPL  Renesas Starter Kit for SH7124 User`s Manual  Rad-8®  Sono sempre di più le persone che si rivolgono al  man_ADS Online_eng_under prod.indd  Sony SRS-A202 User's Manual  Inhaltsverzeichnis  Problem Set 4: Hardware, Software, and the Internet  Manuel d`utilisation de la poussette Nuna Pepp  KWB Multifire - Kammerer Heizung  

Copyright © All rights reserved.
Failed to retrieve file