1. herpesvirus 5 min HIV 1 HBV 10 min Hexanios is a detergent disinfectant used to pre disinfect clean and disinfect HEXANIOS reusable instruments instruments and small items of medical equipment replacing chlorine and chlorhexidine cetrimide Concentrated solution to be diluted to 0 5 before use Staff protection 25 ml 1 stroke of dosing pump Apron gloves mask and goggles should always be worn when using this product for 5 water Avoid contact with skin and eyes in case of contact rinse immediately and or thoroughly with clean water and seek medical advice 50 ml 2 strokes of dosing pump PROCEDURE for 10 water depending on the number of Step 1 Pre disinfection instruments 1 Prepare a soaking bath solution according to the volume of instruments to be pre disinfected 5 or 101 2 Immerse the instruments completely in the soaking bath and let them soak for 15 minutes 3 Rinse with clean water 4 Throwthe used solution away Do not re use this solution prepare a fresh bath for each load of instruments you treat Step 2 Cleaning and disinfection 5 Prepare a soaking bath according to the volume of instruments to be disinfected 5 or 101 6 Completely immerse the instruments in the soaking bath brush them carefully and let them soak for 15 minutes 7 Rinse thoroughly with clean water and dry with a clean cloth 8 This solution can be kept for a maximum of 24 hours In case of visible contam
2. Important After sterilization is completed and before any instruments are sent to the storage room confirm that the TST strip has changed color as shown in the above diagram if it has not the complete load must be considered unsterile and sent for repackaging and re sterilization STERILIZATION GUIDELINES 23 Code MSTECONRS18 Control Strip TST T Steam Time 134 18 min adesive 24 STERILIZATION GUIDELINES Before After SILILI 9 7 1 2 External indicator Tape These indicators are used only to identify packages that have been through a sterilization process Autoclave indicator tape consists of yellow indicator stripes and pressure sensitive adhesive When the steam sterilization is completed the color of the indicator strips will turn from yellow to dark brown or black Take a piece of indicator tape and before placing it on the package write the date of sterilization and the initials of the person who checked and wrapped the package on it Figure 10 External indicator Tape 9 7 1 3 Checking the package is undamaged at the end of the cycle upon unloading The packages should not be damp or wet and the paper should be intact Inspect each package for tears holes Torn packages are unsterile and must be sent for repackaging and re sterilization 9 8 Autoclave with vacuum pump 9 8 1 Checking the airtight seal before sterilization Bowie Dick test The Bowie Dick test is performed to as
3. 1 3 Checking the package is undamaged at the end of the cycle upon unloading 10 11 12 13 14 14 14 15 15 16 17 17 18 19 19 20 20 21 21 21 22 22 22 23 23 23 24 24 4 STERILIZATION GUIDELINES 9 8 Autoclave with vacuum pump 9 8 1 Checking the airtight seal before sterilization Bowie Dick test 9 8 2 Reading the graph for a vacuum pump autoclave 9 9 Traceability 9 10 Storage 9 10 1 Expiry 9 10 2 Checking 9 10 3 Stock management 10 Emergency sterilization 10 1 Operating with instruments sterilized using the emergency procedure 10 2 Emergency readiness 11 Sterilization procedure for linen 11 1 Organizing a laundry 11 2 Collection 11 3 Procedure for washing linen 11 3 1 Washing by hand 11 3 2 Machine washing 11 4 Packaging sterilization and storage 12 References ANNEX I Instructions for folding a surgical gown ANNEX II Summary procedure for pre disinfecting and disinfecting instruments ANNEX III HEXANIOS PROTOCOL Technical File ANNEX IV SURFANIOS PROTOCOL Technical File ANNEX V Top loading vertical autoclave without vacuum pump e g TBM 90 litre ANNEX VI Front loading horizontal autoclave with vacuum pump with or without steam generator e g Matachana SC 500 ANNEX VII Operating instructions for the TBM 90 litre autoclave ANNEX VIII Operationing instructions for the 45 litre autoclave ANNEX IX Operating instructions for the 39 litre autocla
4. different colour one bucket with the diluted Surfanios solution one bucket for clean cold water 1 clean cloth or mop Technique Work from the cleanest to the dirtiest surface and from top to bottom Surfaces and furniture Immerse the cloth mop in the bucket of Surfanios wring it out to remove excess solution and wipe the furniture If necessary rinse the cloth mop in the bucket of clean water and wring it out before immersing it again in the Surfanios solution 2 5 ml litre 20 ml 1 stroke of the dosing pump 5 I can 38 STERILIZATION GUIDELINES Floors Immerse the cloth mop in the bucket of Surfanios wring it out to remove excess solu tion and wipe the floor If necessary rinse the cloth mop in the bucket of clean water and wring it out before immersing it again in the Surfanios solution Use the Surfanios solution on every surface in the room Do not rinse Leave to dry for 15 minutes without entering the room except in an emergency After cleaning 1 keep the remaining Surfanios solution for the next cleaning operation 2 empty the water bucket 3 rinse the bucket and mop with Surfanios solution 4 let them dry before using them again At the end of each day 1 clean and rinse all material with Surfanios 2 letit dry 3 throw away the rest of the Surfanios solution Once every month all surfaces should be cleaned with an ammoniated detergen
5. from the tub rinse it with copious amounts of running water until the soap is totally washed out 4 Drain the dirty water in the sink into the waste water system 5 Disinfect the linen in a non metal bucket containing a 0 1 active chlorine solution for 15 minutes 6 Rinse copiously at least three times to wash out the chlorine solution 7 Drain the chlorine solution in the sink into the waste water system 8 Wear clean gloves and a clean apron to wring out the linen and put it into the clean plastic bag used for transporting the linen to the drying area 9 Hang the linen to dry on a clothesline in an area that is protected from rain insects and dust or put it in the clothes dryer if one is available 10 Transport the dry linen to the ironing and folding area of the laundry 11 3 2 Machine washing 1 Pre wash linen soiled with organic matter or blood twice at low temperature below 25 C for 5 7 minutes before running the full cycle Put all linen through a full wash cycle Low temperature pre wash for 5 to 7 minutes Wash at 60 C for 15 minutes if possible thermal disinfection Rinse Disinfect for 2 to 3 minutes in 0 1 chlorine solution Rinse three times with clean water Spin At the end of the wash cycle put on clean gloves and apron to empty the machine into the bin used for this purpose 10 Hang the linen to dry on a clothesline in an area that is protected from rain and dust 11 Do not leave damp t
6. gown 8 Fold the closest part of the gown keeping 3 centimetres away from the collar 9 Fold the opposite part of the gown keeping 3 centimetres away from the collar 10 Fold twice to obtain a square Jeannette de Vries ICRC Jeannette de Vries ICRC Jeannette de Vries ICRC Jeannette de Vries ICRC STERILIZATION GUIDELINES 33 34 STERILIZATION GUIDELINES ANNEX II SUMMARY PROCEDURE FOR PRE DISINFECTING AND DISINFECTING INSTRUMENTS 1 Pre disinfection In the operating theatre Immerse the instruments completely in the Hexanios soaking bath and soak for 15 minutes Use the timer Rinse with clean water This solution must be changed after each use i e after each operation In the ward Use the same procedure for instruments before bringing them to the sterilization department Use the timer ONE patient ONE bath 2 Cleaning and disinfection Immerse the instruments completely in the soaking bath brush them thoroughly and soak in the solution for 15 minutes 3 Rinsing Rinse thoroughly with clean water 4 Drying The instruments must be dry before packaging Wipe them with a clean lint free linen cloth 5 Packaging Non woven paper Instruments are packed in wire baskets wrapped in two layers of non woven paper stronger than crepe paper Steam indicator tape Close each package with this tape ATST strip has to be put in the big surgical wire basket
7. medical staff The procedures for cleaning and disinfecting these materials should therefore include protective measures for those who handle them Correct handling and safe processing of surgical materials are part of the standard precautions that should be taken in any medical facility including those supported by the ICRC In 2004 a workshop was held for experienced ICRC operating theatre OT nurses and experts from M decins Sans Fronti res to discuss the main problems relating to sterilization in the context of humanitarian operations This workshop led to the drafting of the first edition of the Sterilization Guidelines Ten years later it was time to review the Guidelines in light of recent scientific discoveries new sterilization procedures and technological developments While the systems and procedures outlined in the Guidelines may be difficult to fully implement in all ICRC supported medical facilities they represent good practice and set the standard to achieve in order to ensure a basic level of safety This will sometimes involve an investment in infrastructure equipment and training all of which requires planning and budgeting The Guidelines are intended for use by OT nurses working in ICRC supported medical facilities While they do not cover all aspects of infection control and sterilization in detail for more information see the publications selected as resource material under Reference they should help OT nurs
8. s micro organisms and pathogens If these instruments are not scrupulously cleaned and sterilized before being used on another patient pathogens can be passed from one patient to another The population of viable organisms on materials instruments linen etc is referred to as the bioburden or initial contamination Hospital acquired or nosocomial infections are infections that patients acquire during their stay or treatment at a hospital or health centre Such infections are easily transmitted during daily hospital practice e ifthe instruments used during surgery are not free of micro organisms if staff do not wash their hands properly before and after treating patients ifthe material used to treat wounds is not free of micro organisms if contaminated waste is not disposed of properly see the ICRC publication Medical Waste Management if food is contaminated Cross infections are infections transmitted from one patient to another through contaminated equipment instruments and materials Cross infection is one of the major causes of nosocomial diseases It has a huge impact on the patient extended stay in hospital leading to more nursing care more operations and more rehabilitation increased health complications that can ultimately result in death For health care providers the extra costs can lead to a rise of up to 10 per cent in the hospital budget It is essential for health staff to unde
9. soiled gloves 10 Besure not to leave any gloves paper or other foreign objects in the linen bags 11 The empty plastic bag is discarded with the waste to be incinerated Never re use it 12 The cloth bag is washed with the linen before being re used After each operation linen bins are disinfected with a surface disinfectant Surfanios from Laboratoires Anios before a new plastic bag is inserted At the end of the day these bins are washed with soap and water and disin fected with the surface disinfectant 11 3 Procedure for washing linen Linen coming from the operating theatre and soiled linen from the wards 1 All staff must wear heavy duty long rubber gloves a waterproof apron and eye protection when handling contaminated linen 2 Take the bags of dirty linen and empty them into the wash sink 3 Cloth bags are washed and disinfected with the linen plastic bags are thrown away with the waste 4 Linen soiled with organic matter or blood should be left to soak for 30 minutes in a sink filled with enough cold water to immerse the linen completely 5 Stir with a stick to loosen the debris 6 Remove the linen from the sink rinse it and follow the procedure below 11 3 1 Washing by hand 1 Immerse the linen completely in a sink filled with soapy water heated if possible to body temperature after the detergent washing powder has completely dissolved 2 Scrub the linen with a brush 3 Remove the linen
10. to monitor time steam and temperature inside the packages This is not needed for small sets like dressing suture and delivery sets A second strip wrapped in a double layer of crepe paper is placed in the middle of the autoclave load Crepe paper Use crepe paper for individual instruments diathermy devices dressing sets and compresses in groups of 10 20 compresses Steam indicator tape close each package with this tape STERILIZATION GUIDELINES 35 6 Autoclave sterilization Loading Do not put linen and instruments in the same load Sterilize separately Cycle Instruments and linen 134 C for 18 minutes 7 Traceability Complete the sterilization log date cycle number load TST signatures and comments wet OK or not OK 8 Expiry date Write the sterilization date on the steam indicator tape with a marker Items double wrapped in crepe paper or non woven sheets can be stored fora maximum of four weeks before use 36 STERILIZATION GUIDELINES ANNEX III HEXANIOS PROTOCOL TECHNICAL FILE Code Microbiological Properties Hexanios MASDCLEMIES Detergent Disinfectant for instruments and Active against Standards Contact time equipment 5L and pump 0 25 Bacteria EN 1040 EN 13727 5 min MSRA EN 13727 NET 72 190 T 72 300 A baumannii 15 min BMR EN 13697 15 min Mycobacteria Mycobacterium tuberculosis tb 15 min Yeast EN 1275 5 min EN 13624 5 min Viruses BDV surrogate of HCV
11. viruses and bacteria Protecting clean instruments from the risk of re contamination during handling and after sterilization in order to keep them sterile and protected until they are used A bacterium virus or other micro organism that can cause disease A medical device is considered sterile and referred to as sterile if the possibility that it has been contaminated by viable micro organisms is equal to or less than one in a million CEN and ISO standards A process that destroys all micro organisms including bacterial spores on the surface of an object or in a fluid Tuberculosis World Health Organization 6 STERILIZATION GUIDELINES 2 INTRODUCTION All hospitals need to sterilize their equipment and supplies Even the smallest hospital requires sterile surgical instruments for minor procedures and sterile dressing materials If a hospital has a full surgical unit the sterilization of surgical instruments and linen together with dressing materials for the wards and operating theatre plays a key role in infection control In medical facilities supported by the ICRC care must therefore be taken to provide safe sterile surgical material and safe blood for transfusion This is all the more important when the ICRC is working in a general hospital setting where elective or semi elective surgery is a routine part of the workload Processing potentially infected materials also presents a hazard to
12. 1 Two types of paper Non woven paper Instruments are packed in wire baskets wrapped in two layers of non woven paper stronger than crepe paper Baskets should be wrapped with single sheets twice Crepe Paper Use crepe paper for single instruments diathermy devices dressing sets and compresses in groups of 10 20 compresses as drums are no longer recommended Procedure The paper used must be at least 3 or 4 times the size of the total material to be packed in both width and length Once the paper has been autoclaved it cannot be re autoclaved The filters in the paper are closed by the first sterilization and are therefore not re autoclavable Use the American square or envelope folding technique for both layers Close the package carefully using sterilization tape Jeannette de Vries ICRC Figure6 Baskets wrapped in two layers of non woven paper 9 5 2 American folding technique 8 Figure 7 American folding technique 9 6 Saturated steam sterilization the autoclave 9 6 1 General information The sterilization of materials using steam and pressure is a reliable method of destroying all forms of microbial life However the autoclave must be properly used and understood to be effective The technique steps recommended by the manufacturer must be followed exactly otherwise the sterilization process will be incomplete see Annexes V VI VII and VIII Anautoclave is basically a large pressure cooker it
13. STERILIZATION GUIDELINES International Committee of the Red Cross 19 avenue de la Paix 1202 Geneva Switzerland T 41 22 7346001 F 41 22 7332057 E mail shop icrc org www icrc org ICRC December 2014 Cover photos Julie Barnet ICRC Reda instruments Jeannette de Vries ICRC STERILIZATION GUIDELINES STERILIZATION GUIDELINES TABLE OF CONTENTS 1 Terms and abbreviations 2 Introduction 3 Acknowledgements 4 General information 5 Principles of contamination 5 1 Methods of reducing the bioburden 6 Classification of instruments Spaulding Classification 7 Organizing the sterilization department 7 1 The flow of instruments through the central sterile supplies department 7 1 1 The cleaning or dirty zone 7 1 2 The packaging or clean zone 7 13 The autoclaving or sterile zone 7 14 The storage and distribution zone 8 Water and electricity needs 9 The sterilization cycle 9 1 Collection 9 2 Pre disinfection 9 3 Cleaning disinfection and drying 9 4 Checking to ensure instruments are and remain intact 9 5 Packaging 9 5 1 Two types of paper 9 5 2 American folding technique 9 6 Saturated steam sterilization the autoclave 9 6 1 General information 9 6 2 The steam sterilization cycle 9 6 3 Sterilization cycle temperature 9 6 4 Safety 9 7 Control 9 7 1 Checking the sterilization cycle 9 7 1 1 Chemical indicators TST STRIP 9 7 1 2 External indicator Tape 9 7
14. air pockets 15 Let the pressure go back up to 2 1 bars 134 C 134 C 2 1 bars in relative pressure 3 1 bars in absolute pressure and 0 985 bars in atmospheric pressure Sterilization 16 As soon as the pressure reaches 2 bars the regulating valve starts to release steam Time 18 minutes from now using the timer 17 Reduce the heat lower the burner flame or position 1 or 2 on the electric heater to keep just a light jet of steam coming out 18 Regularly check the pressure while the load is being sterilized and adjust the heat if necessary Expelling steam and water 19 When the sterilization process is finished after 18 minutes switch off the burner or set the switch to 0 to cut the heat otherwise the element will burn out the timer incorporated into the autoclave acts only as reminder it does not switch the electricity off The heating indicator light goes out at position 0 20 Open the YELLOW valve Be careful with the jet of boiling steam and water use a flexible tube to vent the steam and water safely away preferably into a clean 20 I jerry can resistant to boiling water in order to reuse this distilled clean water 21 Close the YELLOW valve when the pressure drops to 0 1 bar Drying 22 Fill the outer tank with cool water 16 litres Note The cooler the water the more effective the drying 23 Open the BIDE valve The cool water flows into the coil inside the autoclave making th
15. atures of the sterilization process Corrosion will result in serious damage to the instruments and may even render them useless and dangerous for the patient This is why every instrument should be checked carefully before being packed If it is damaged or no longer functioning alert the senior nurse so that the instrument can be exchanged Apply oil if necessary but use only specific silicone free instrument oil which allows steam to penetrate during sterilization The instrument should be rinsed to remove excess oil and dried Jeannette de Vries ICRC Figure 5 Checking the instruments Julie Barnet ICRC STERILIZATION GUIDELINES 19 20 STERILIZATION GUIDELINES Reda instruments Code XSTEBASK201 Basket Sterilizing 201x145x50 mm plus lid XSTEBASK402 Basket Sterilizing 380x240x90 mm plus lid XSTEBASK502 Basket Sterilizing 480x240x90 mm plus lid XSTEBASK302 Basket Sterilizing 300x200x35 mm plus lid MSTESTSHNW120 Sheet Sterilisation 60 gr m2 120x120 cm non woven Jeannette de Vries ICRC 9 5 Packaging Aim Before sterilization packaging protects the clean instruments from the risk of contamination Packaging must allow the packages to be opened easily and aseptically Packaging must be permeable to steam water repellent resistant to tears and impermeable to micro organisms e After sterilization packaging must keep the instruments sterile until they are used 9 5
16. ction cleaning disinfection packaging sterilizing distributing and storage In the field this task may not exist or may be carried out by a specialized department that may not even be in the same premises as the health facility In ICRC supported hospitals that have a surgical department the preferred option is to set up a central sterile supplies department Within this department all surfaces walls floors ceilings shelves should be washable All rooms should be well ventilated and lit In a challenging environment dust flies damaged infrastructure or extreme heat cold a balance has to be struck between adequate light and ventilation and the potential for contamination of clean material An uninterrupted supply of water and electricity is essential STERILIZATION GUIDELINES 13 14 STERILIZATION GUIDELINES 7 1 The flow of instruments through the central sterile supplies department Cleaning Dirty zone 7 1 1 The cleaning or dirty zone This is the area in which reusable instruments are collected registered cleaned and dried These activities can all be carried out in the same room provided that all steps are carried out in one direction only and that nothing ever moves back towards the dirty zone dirty and clean items must never cross The dirty zone should be laid out in a way that facilitates this flow In some settings a wall will separate the dirty zone from the clean zone although th
17. e steam condense onto the coil the pressure reading falls because a vacuum is created inside 24 Wait for 10 minutes until the tank is empty 25 Refill the outer tank with cool water and wait for 10 minutes 26 Close the BEWE valve 27 Break the vacuum by opening the RED valve the process takes about 2 minutes 28 Open the autoclave 29 If the towel placed over the load is wet remove it 30 Wait for about 30 minutes with the lid half open for everything to cool Drying will be completed by the heat remaining in the body of the autoclave 31 Remove the basket s and place it them on a clean towel 32 Take out the sterilized equipment and close the sliding windows of the drums immediately Checking the cycle 33 Check that the packages wrapped in paper are intact Discard any wet packages they will need to be re processed and re sterilized 34 Put the lid back on the autoclave leave it slightly open Storage and distribution 35 Transfer the sterile packages and drums to the storage area use a clean container for transport Remarks The chimney on top of the lid must be filled with cotton wool or gauze compresses to filter the air reintroduced into the autoclave after drying Change the cotton every 50 cycles or once a month If you use an electric power source this autoclave can be switched to normal posi tion 3 4 500 W medium position 2 3 000 W or low power position 1 1 500 W In the case of l
18. e two have to be connected via a doorway or a wicket to facilitate the transfer of instruments All items must be checked to ensure they have been properly cleaned before they leave the dirty zone and before being dried and packaged Any items that are not properly clean must be recleaned before they can leave the dirty zone gt The dirty zone should be easily accessible from the operating obstetric and surgical emergency departments to facilitate the reception of reusable instruments It must also be located next to the clean zone Packaging Clean zone 7 1 2 The packaging or clean zone This is the area for checking inspecting instruments reassembling instrument sets high level disinfection packaging and storage of packages ready to be sterilized Damaged instruments should be set aside for the attention of the senior nurse The clean zone should be separated from the dirty zone by a wall with a door or a wicket connecting them to facilitate the transfer of material gt It should be located between the dirty zone and the sterile zone Autoclaving Sterile zone 7 1 3 The autoclaving or sterile zone This is the area in which the instruments ready to be sterilized by steam sterilization in an autoclave are registered The appropriate heating source for the autoclave is defined with the technical resource person trained hospital technician The choice between electricity gas kerosene will depend on the type
19. effective use of steam sterilizers autoclaves the decontamination of cultures and other materials the preparation of sterile supplies and the safe operation of autoclaves Cleaning disinfection and sterilization are key to preventing the spread of infections In spite of this many health care facilities either lack the means to perform basic infection prevention and control tasks or their personnel receive insufficient training Every hospital must have the means to sterilize equipment and supplies Even the smallest health facility will need sterile surgical instruments for minor surgical procedures and sterile dressing material The sterilization department must be able to provide the operating theatre and wards with a sufficient quantity of sterile items for routine and emergency work In the health facilities the ICRC supports it has a responsibility to provide safe and sterile surgical material As stated in ICRC Assistance Policy Doctrine 49 biosecurity is of prime importance The main micro organisms of concern are HIV TB hepatitis B and C staphylococcus and bacterial spores Biosecurity is an important hospital function and calls for trained essential staff The aim of sterilization is to treat medical supplies in such a way that they can be used safely without unnecessarily endangering the patient or the user To ensure that equipment and materials are safely sterilized it is essential for the person using them to be well trai
20. es Infection control in health care settings guidelines for MSF projects Paris FORGHO 2012 ICRC Guidelines Sterilization of Surgical Material for Hospitals Supported by ICRC internal publication ICRC Geneva June 2005 ICRC Medical Waste Management Geneva 2011 PPT La sterilisation sur les missions Pascale Choquenet hospital pharmacist MSF France FORGHO Bordeaux WHO Prevention of hospital acquired infections 2nd edition 2008 Centers for Disease Control Guideline for Disinfection and Sterilization in Healthcare Facilities 2008 H pitaux universitaires de Gen ve Sterilization Department Mr Ney Herv The European Forum for Hospital Sterile Supply EFHSS http www efhss com World Forum for Hospital Sterile Supply WFHSS http www wfhss com G F Gardner M M Peel Sterilization Disinfection and Infection Control third edition Harcourt Australia 2001 Jeannette de Vries ICRC o Jeannette de Vries ICRC ANNEX I INSTRUCTIONS FOR FOLDING A SURGICAL GOWN 1 Clear the table 2 Lay the inside of the gown face down 3 Cross the 2 sleeves on the upper part of the gown chest 4 Tie the 2 thick outside fasteners together loosely 5 Fold over the closest shoulder so that the seam is visible all the way down the gown 6 Adjust the opposite shoulder so that the seam is visible all the way down the gown 7 Ensure that the 2 top and 2 inside fasteners are towards the bottom of the
21. es in various hospitals facing different challenges to set up safe and simple systems for processing the entire range of surgical materials requiring sterilization The Guidelines can be applied in situations where resources are limited and may be used to provide essential training for OT nurses and for anyone involved in sterilization 3 ACKNOWLEDGEMENTS would like to express my deepest gratitude to all the people who helped and supported me throughout the process of reviewing the Sterilization Guidelines My special thanks go to the author of the first edition of the Guidelines Jenny Hayward Karlsson on whose excellent work the present update is based also take this opportunity to thank those who helped review and correct the new Guidelines without whose professional support the undertaking would have been impossible am particularly indebted to Julie Barnet who initiated the project and wrote the first draft am also very grateful to Mr Herv Ney head of the Sterilization Department at the University Hospital in Geneva for his technical advice and support which were of great help A special mention goes to Catherine O Shea who provided continuous support and expert advice throughout the review process To all of you extend my most sincere thanks Jeannette de Vries STERILIZATION GUIDELINES 7 8 STERILIZATION GUIDELINES 4 GENERAL INFORMATION This document establishes guidelines for the
22. evel disinfection elimination of all micro organisms except high numbers of bacterial spores Objects that are in contact with intact skin such as bedpans blood pressure cuffs and bedside tables Low level disinfection is required These items are less likely to spread infections except when contaminated with pathogens by the hands of health care personnel They require rigorous cleaning washing and disinfection and hands must be washed after handling them Figure 2 Spaulding Classification 7 ORGANIZING THE STERILIZATION DEPARTMENT Before an instrument can be used a whole range of measures must be taken to prevent it from being a danger to the patient Sterilization is only one link in the chain of activities required to make an instrument sterile at the moment of use Each link involves specific methods and techniques that prevent contamination of instruments materials and living tissue This is called the aseptic technique and is designed to prevent contamination of instruments materials and living tissue It includes for example covering the patient during the operation with sterile drapes cleaning the operating table and cleaning and decontaminating the patient s skin with iodine before starting the operation The aseptic technique must also be followed in the sterilization department Hospitals in the developed world employ various methods of organizing the entire instrument sterilization process collection pre disinfe
23. extiles or fabrics in the machine overnight 12 Be careful not to overload the machine follow the manufacturer s recommendations 10 09 NO RE NU After each surgical operation the linen bins must be disinfected with a surface disinfectant Surfanios before a fresh plastic bag is inserted At the end of the day ALL bins must be washed with soap and water and disinfected with the surface disinfectant 11 4 Packaging sterilization and storage If linen is to be sterilized pack it into two layers of crepe paper after it has been ironed and folded The sterilization process is the same as for instruments including the TST strip 134 C for 18 minutes Remember that instruments and linen should never be put in the same load The expiry date for the sterile linen is four weeks using the storage techniques recommended for instruments see section 9 10 Linen that does not need to be sterilized e g bed sheets can be stored ironed and folded in a clean and dry cupboard until it is sent to the wards STERILIZATION GUIDELINES 31 32 STERILIZATION GUIDELINES 12 REFERENCES Swissmedic Bonnes pratiques de retraitement des dispositifs m dicaux pour les cabinets m dicaux et les cabinets dentaires ainsi que de petits st rilisateurs a la vapeur d eau satur e version 1 April 2010 M decins Sans Fronti res Sterilization and laundry guidelines for MSF projects Paris M decins Sans Fronti r
24. g the sterilization Each autoclave must have its own log Figure 13 Example of a traceability log 9 10 Storage The aim is to keep the packaging intact and the equipment sterile until it is used The packages should be stored in such a way as to allow inventory to be properly managed and expiry dates to be monitored The storage room should be clean closed and well ventilated It should not be a room that has many people passing through STERILIZATION GUIDELINES 27 Sterilized packages should be stored away from dust humidity and sunlight To further ensure the integrity of sterilized packages it is recommended that they be stored in plastic storage boxes with lids readily available on the local market For the mobile surgical team dust covers are included in the kits 9 10 1 Expiry The length of time materials remain sterile depends on the quality of the entire sterilization process the quality of the packaging and the quality of storage If the storage conditions described above are respected items double wrapped in crepe paper or non woven sheets can be stored for a maximum of four weeks before use In order to avoid humidity the instruments should not be placed in the plas tic boxes or dust covers right after the sterilization process They need to be cooled before storage preferably on a wire rack taking care not to tear the packaging 9 10 2 Checking Expiry dates should be checked once a wee
25. h covering the load Put the basket back with its content inside the body of the autoclave while it is still hot Replace the lid leaving it slightly open This enables the accumulated heat to dry the equipment Wait 30 minutes or more depending on the climate for everything to cool and dry Take out the sterilized equipment and close the vents of the drums immediately Checking the cycle 24 25 26 Check the integrity of the packages wrapped in paper Discard any damp or wet packages they will need reprocessing Write on the paper tape of the packs and sets date of sterilization Replace the lid on the autoclave Storage and distribution 27 Transfer the sterile packs and sets to the storage area use a clean container for transport Remarks Never pour cold water into or onto an autoclave while it is still warm to avoid cracking This autoclave requires a powerful heating source electric plate minimum 1500 W or safe gas or kerosene burner Place the autoclave on a stable surface make sure that the burner or fire is stable If using a burner use the wind screen If steam is escaping from around the lid grease the lid with Vaseline every week Check regularly at least once a month the air exhaust tube by blowing air through Clean with a small diameter wire If there is any build up of foreign material or corrosion inside discard and replace the tube MISSION The International Commit
26. heater position 3 for electric heater maximum strength for oil stove or gas burner 7 As soon as a continuous jet of steam escapes from the purple valve close the valve valve handle perpendicular to the pipe 8 When the pressure reaches 1 bar on the gauge open the purple valve and let the pressure fall to about 0 2 bar then close the valve Repeat this operation at least once to eliminate all air pockets Sterilization 9 As soon as the selected pressure is reached the control valve starts to release steam Start timing the sterilization process from now using the timer Reduce the heat turn electric heater to position 1 lower flame on burner to maintain a slight release of steam Water and steam evacuation 10 When the sterilization time is up stop the heating process for electricity set the switch to 0 otherwise the element will burn out for gas stop the flow for oil turn the stove off Open the yellow valve and let pressure fall to about 0 1 bar then close the yellow valve NOTE Opening the yellow valve releases a jet of boiling water and steam use a flexible tube to drain the steam and water safely using a suitable container Drying equipment 11 Fill the upper external tank with cold water up to the gate about 16 litres 12 Open the blue valve The steam inside condenses on the coil which is filled with cold water to create a vacuum the gauge falls Wait about 10 minutes then close the blue valve and b
27. ination dirty solution renew the solution immediately STERILIZATION GUIDELINES 37 ANNEX IV SURFANIOS PROTOCOL TECHNICAL FILE Code MASDCLEMS5 Detergent disinfectant for surfaces 5 amp pump Microbiological Properties Surfanios Active against Standards Contact time 0 25 Bacteria EN 1040 EN 127 EN 13727 5 min BMR EN 1276 5 min EN 13697 30 min 15 min BMR EN 13697 15 min Mycobacteria Mycobacterium terrae 60 min 30 min EN 14348 Mycobacterium terrae 60 min 30 min EN 14563 Yeast EN 1275 15 min EN 13697 30 min 15 min Moulds EN 1275 15 min EN 13697 30 min EN 1650 120 min Viruses HIV 1 PRV surrogate of HBV 5 min BDV surrogate of HCV and herpesvirus Vaccinia virus 60 min 15 min Surfanios is a detergent disinfectant used for surfaces that replaces chlorine and chlorhexidine cetrimide It is used to disinfect all surfaces floors walls and medical equipment e g mayo table operating table operating lamp SURFANIOS Concentrated solution to be It is non corrosive and compatible with any kind of covering flooring diluted at 0 25 before use Staff protection 20 ml for 8 of cold water Apron gloves mask and goggles should always be worn when using this product Avoid contact with skin and eyes in case of contact rinse immediately and thoroughly with clean water and ask for medical advice Minimum necessary equipment 2 buckets of a
28. ing up of surgical sets easier and quicker pre disinfection easily available and inexpensive on local markets Aim To ensure staff safety and minimize the risk of spreading micro organisms in the environment Medical devices should be collected and pre disinfected as soon as possible after use In the operating theatre instruments must be sent for cleaning after each operation On the wards they should be collected at least twice a day This is to avoid blood and secretions drying onto the instruments 18 STERILIZATION GUIDELINES 9 2 Pre disinfection Aim To reduce of the number of micro organisms on the surface of the equipment thus reducing the risk of contamination and ensuring the safety of staff handling the equipment Procedure HEXANIOS 0 5 Laboratoires Anios Microbiological properties see Annex III Instructions for use concentrated solution to be diluted at 0 5 before use gt 1 stroke 25 ml of the dosing pump for 5 water gt 2 strokes 50 ml of the dosing pump for 10 water How Prepare a soaking bath solution according to the volume of medical instruments 5 or 10 1 Immerse the instruments completely in the soaking bath and leave them to soak for 15 minutes using a timer Rinse with clean water t2 S Code E XLABALAR60M T J Alarm clock analog 1 mn 60 mn pce Figure4 Alarm clock Instruments and equipment should be pre disinfected and cleaned as soon as possible after u
29. is unable to reach the chemical indicator If the sterilizer fails the Bowie Dick test do not use it until it has been inspected by the sterilizer maintenance personnel and passes the Bowie Dick test Figure 11 Bowie Dick test The Bowie Dick test does not replace the TST strips In an autoclave with a vacuum pump the Bowie Dick test must be performed daily TST strips must be used in every load as described above and recorded in the log 9 8 2 Reading the graph for a vacuum pump autoclave Pression absolue mbar Programme 310 instruments en corraners Ooie Heure d but Heure tn Dur e 2670772005 142701 160229 03528 Fin 34 Fin du programme soldes Figure 12 Vacuum pump graph 4 World forum for hospital sterile supply STERILIZATION GUIDELINES 25 Code MSTECONRBW Bowie Dick Test 134 C Mod PB amp D134 single use 26 STERILIZATION GUIDELINES Julie Barnet ICRC Check if the time exposure is correct Check that the three vacuum peaks are visible Check that the sterilization plateau is parallel for temperature and pressure Once all these checks have been performed the load can be cleared meaning that it has been correctly sterilized following the right procedure 9 9 Traceability Each sterilizing cycle must be recorded in a log The log should contain the following data e date contents of the load cycle number the TST name of the person performin
30. k and recorded in a logbook 9 10 3 Stockmanagement One important rule first in first out i e items due to expire first should be used first During the weekly inspection packages should be checked to ensure that items expiring sooner are at the front and those with later expiry dates are at the back 28 STERILIZATION GUIDELINES 10 EMERGENCY STERILIZATION Emergency sterilization should be employed as a last resort ONLY if surgical instruments cannot be sterilized because of an autoclave failure It is a temporary solution ICRC senior health staff i e the operating theatre nurse health delegate in conjunction with any other relevant support department logistics WatHab should immediately put into action a plan to correct the problem The ICRC Health Unit in Geneva should be informed and will provide guidance on how to continue 10 1 Operating with instruments sterilized using the emergency procedure In the event of an autoclave failure the surgeon must be informed straight away It is ultimately the surgeon s decision whether or not to use instruments or material sterilized by means of the emergency procedure If s he decides that a life saving procedure must go ahead despite the risk to the patient this should be noted in the patient s file 10 2 Emergency readiness Where autoclave capacity is limited or to ensure readiness for an autoclave failure a supply of disposable sterile surgical linen and co
31. mpresses should be available to enable life saving procedures to be carried out Semi elective or elective surgery should not be performed under these conditions Remember that instruments that have undergone emergency sterilisation high level disinfection are not considered sterile Only a limited number of instruments are concerned Non autoclavable semi critical thermo sensitive instruments that are likely to be reused e g laryngoscope 11 STERILIZATION PROCEDURE FOR LINEN Aim To collect linen used in the operating theatre and from the wards for transportation to the washing area of the laundry while ensuring staff safety and minimizing the risk of spreading micro organisms in the environment The aim of washing linen is to remove soiling via the chemical action of soap combined with the mechanical action of brushing followed by rinsing Linen to be sterilized before use needs to be carefully cleaned before going into the autoclave The procedure is the same as for the sterilization of instruments 11 1 Organizing a laundry The process of cleaning soiled linen is a chain of activities where the procedures must be carried out from the dirtiest to the cleanest zone with no possibility of moving back or crossing clean with dirty linen The clean and dirty zones should be separated Dirty Zone Reception area Soaking area Washing area All linen must be washed as quickly as possible after delivery to the la
32. ned and for the autoclave to be in perfect condition While the systems and procedures outlined in these guidelines may be difficult to achieve in all health facilities in which the ICRC is present they nevertheless represent best practices ensuring a basic level of safety This should be an achievable goal in any ICRC supported health facility provided investment in essential infrastructure is planned and budgeted All changes and adaptations to the 2005 edition of these guidelines are based on WHO CEN and ISO norms relating to patient safety and control of health care associated infections Their aim is to ensure good quality care and safety for ICRC patients and health staff Health staff in charge of sterilization need to have a thorough knowledge of the nature of the organisms that cause infection the measures taken to limit the spread of disease how disease causing organisms can be destroyed how to use sterilization equipment safely and successfully the design and operation of sterilization equipment all types of loads and packaging systems and their behaviour when exposed to steam the methods for checking the performance of sterilizers STERILIZATION GUIDELINES 9 10 STERILIZATION GUIDELINES 5 PRINCIPLES OF CONTAMINATION A body penetrated by disease causing micro organisms is said to be contaminated Instruments used during an operation are also considered to be contaminated by the patient
33. ng at the end of sterilization Electric overheating safety cut out optional Control box Door for kerosene gas Code XSTEAUTO90 AUTOCLAVE COMBINED 90 elec kero incl kero burner 42 STERILIZATION GUIDELINES 4 Place the items to be sterilized in the basket s separating them so that the steam can easily circulate between them leave enough space for a hand to slide between the packages 5 Avoid overloading maximum 75 of capacity 6 Put a clean absorbent cloth towel over the load to absorb condensation drops falling from the inside of the lid 7 Close the lid tightening the diametrically opposite screws Note Make sure that all the nuts are on top of the screws 8 Choose the pressure required by moving weights 2 notch 2 1 bars 134 C 134 C 2 1 bars in relative pressure 3 1 bars in absolute pressure and 0 985 bars in atmospheric pressure 9 Fill the chimney with cotton wool or compresses Heating 10 Put on full heat if using a burner start it on full heat e if using electric heating turn the red button to position 3 11 As soon as a continuous jet of steam comes out of the GREEN valve close this valve valve perpendicular to pipe Purging 12 When the pressure reaches 1 bar on the pressure gauge first red line open the GREEN valve 13 Let the pressure fall to about 0 2 then close the valve 14 Repeat steps 12 and 13 twice more to eliminate all
34. oclave designed for hospitals Provided with flexible power source electricity gas kerosene and an incorporated drying system by condensation specially designed for use without electricity Da ewe ONE MATERIAL REQUIRED Clean water low mineralization preferably distilled as cold as possible Heating power source electricity or kerosene gas burner Clean thick towel Timer in case the timer integrated into the autoclave fails Cotton wool or compresses to fill the chimney Preparations for use 1 Close the YELLOW RED and BRUE valves valve perpendicular to pipe 2 Leave the GREEN valve purge open parallel to pipe 3 Pour water into the autoclave water must cover the element properlyf e 8litres of clean water if heating by kerosene or gas burner 16 litres of clean water if heating by electricity 6 The removable lower tank can be used as a measure when filled up to the line 16 litres i Clean water contains no solids e g filtered 10 11 12 13 14 15 16 17 18 19 Air vent Graduated water tank 8 and 161 BIWE valve Condensation coil Removable collection tank Air filter filled with cotton wool REX valve Heating indicator and electric switch Heating element 1 500 3 0000 4 5000W Timer Pressure gauge Isolation valve not used Safety valve Weights left 1 bar right 2 bars GFEER valve bleeding Yellow valve for emptyi
35. of autoclave the level of the health facility and the availability of electricity fuel this should be discussed with the WatHab engineer and if one is available an ICRC biomedical engineer gt It is important to ensure that there is enough space on all sides of the autoclave for safe use and maintenance TBM 90 autoclaves should be at least 0 60 m from other equipment and adjacent walls A steam exit has to be provided in line with the manufacturer s specifications this is usually created through an exterior wall with access to the exterior area so that people will not be scalded by escaping steam This access must be restricted to qualified personnel to prevent scalding Storage and distribution zone 7 1 4 Thestorage and distribution zone This is the area in which sterile packages are stored until distributed It is important to have a storage place in which instruments and sterile items are protected from the risk of recontamination i e dust insects damage to packages humidity light and in which their sterile status is maintained until they are used gt It should be located next to the autoclaving zone in a separate room STERILIZATION GUIDELINES 15 16 STERILIZATION GUIDELINES 8 WATER AND ELECTRICITY NEEDS The average amount of water needed in the sterilization room exclusively not including activities in the operating theatre is set out below Pre disinfection disinfection rin
36. operates using steam under pressure as the sterilizing agent The high pressure heats the steam to high temperatures thus increasing its heat content and killing power Steam is able to penetrate objects with cooler temperatures because once the steam contacts a cooler surface it immediately condenses to water This creates negative pressure at the point of condensation and draws more steam to the area Condensation continues as long as the condensing surface is cooler than the steam once both are the same temperature a saturated steam environment is formed Achieving a high and even moisture content in the steam air environment is important for effective autoclaving Steam also efficiently kills cells and coagulates proteins Moist heat is thought to kill micro organisms by causing coagulation of essential proteins The micro organism dies because of an accumulation of irreversible damage to all its metabolic functions STERILIZATION GUIDELINES 21 22 STERILIZATION GUIDELINES Atmospheric 9 6 2 The steam sterilization cycle The steam sterilization cycle consists of various phases Pressure gt TIME Pre vacuum Cycle Phases Start door seals jacket warms chamber Purge steam enters chamber while air is purged through the chamber drain Conditioning positive pressure and negative vacuum pulses continue to heat load and purge air Heat up steam pressure build
37. ow power a supplementary heating source is required e g gas butane propane burner STERILIZATION WITH 150 LITRE AUTOCLAVE The instructions for the 150 I autoclave with vacuum pump are the same as for the 90 I model except when it comes to drying Drying 22 Operate the BIDE valve of the vacuum pump to reach 0 7 0 8 bar then release the BEUE valve STERILIZATION GUIDELINES 43 44 STERILIZATION GUIDELINES ANNEX VIII OPERATIONING INSTRUCTIONS FOR THE 45 LITRE AUTOCLAVE Code MOBILE 45 LITRE AUTOCLAVE XSTEAUTO45 Autoclave combined 45 elec kero incl Kero ane vertical Hybrid model electricity oil gas with drying system V45 version two cycles 1 User manual see appendix V on generalities about sterilization by steam under pressure Preparation 1 Pour water into the autoclave from the top about 16 litres for electric heater 8 litres for oil or gas burner The outer tank can serve as a measure graduation marks at 8 and 16 litres 2 Load the device with the material to be sterilized 3 Close the unit and tighten the nuts crosswise tighten one nut then the nut diametrically opposite 4 Select the operating pressure using the black valve Opened valve valve handle parallel to the pipe 1 1 bars 121 C closed valve 2 1 bars 134 C 5 Close the yellow red green and blue valves Leave the purple valve purge in the open position Purge 6 Setup the
38. reak the vacuum by opening the red valve STERILIZATION GUIDELINES 45 ANNEX IX OPERATING INSTRUCTIONS FOR THE 39 LITRE AUTOCLAVE MOBILE 39 LITRE AUTOCLAVE Code XSTEAUTO39 AUTOCLAVE UNICEF 39 pressure cooker Autoclave for use in health centres or in hospitals as a back up system Direct heat source can be electricity gas or kerosene burner wood or charcoal MATERIAL On N Clean water low mineralised preferably distilled Heater and fuel gas burner multi fuel burner electric plate eee Me e Clean thick towel 3 gt t lt lt Timer Vaseline if necessary Preparation 1 Pour 3 litres of clean water into the bottom of the autoclave 2 Place the items to be sterilized in the basket separating them so that the steam can easily circulate between them leave space to allow a hand to slide between the packages Avoid overloading max 75 of capacity 3 Put a clean absorbent cloth towel over the load to absorb drops of condensation falling from the inside of the lid onto the load 4 Put the lid back making sure the marks arrows line up and that the flexible metal tube attached to the lid connects with the conduit provided for this purpose in the basket Lubricate the lid with Vaseline to maintain quality of rubber seal 5 Close the lid and tighten screws crosswise tighten one screw then the screw diametrically opposite Heating 6 Place the autoclave on
39. rstand the mechanisms by which infection is transmitted and correct microbicidal processes so that ICRC supported health facilities can provide good quality health care to those who need it 1 WHO Hospital Infection Prevalence Survey June 1987 STERILIZATION GUIDELINES 11 5 1 Methods of reducing the bioburden Sterilization Dirt and soils Microorganisms Spores Figure 1 Methods of reducing the bioburden Cleaning The process of removing dirt and soils but not killing micro organisms and spores Disinfection The process of destroying all pathogenic micro organisms but not bacterial spores Sterilization The process of destroying all forms of microbial life on inanimate surfaces including bacterial spores 12 STERILIZATION GUIDELINES 6 CLASSIFICATION OF INSTRUMENTS SPAULDING CLASSIFICATION In 1968 Dr E H Spaulding classified medical surgical instruments as critical semi critical and non critical based on their potential to spread infections The classification helps to decide how to proceed with the instruments Critical items Semi critical items Non critical items Items that penetrate soft tissue bone or the vascular system or through which blood flows such as implanted medical devices which should be sterile when used Objects that touch mucous membranes or non intact skin such as endoscopes and respiratory therapy equipment and which require high l
40. s to selected exposure temperature and pressure Exposure timing begins for selected exposure time and temperature Exhaust chamber drain opens and ejector water creates vacuum in chamber to exhaust steam Drying ejector water controls vacuum in chamber for selected dry time Air in chamber returns to atmospheric pressure Cycle complete door can be opened Exposure Time Drying Time Figure 8 The steam sterilization cycle 9 6 3 Sterilization cycle temperature All surgical instruments and linen used for major surgery Choice 1 134 C at 2 1 bars for 18 minutes 134 C 2 1 bars in relative pressure 3 1 bars in absolute pressure and 0 985 bars in atmospheric pressure 9 6 4 Safety Because autoclaves use steam heat and pressure the risk of individual exposure and accident is high Personnel should wear proper personal protective equipment i e heat resistant gloves eye protection and a lab coat particularly when unloading the autoclave Do not assume that the temperature and pressure have fallen when opening the chamber Even if the pressure gauge shows 0 open the chamber carefully start by opening the door just a crack to allow steam to dissipate don t fling the door open as steam might billow out and scald you After opening the door leave the items inside for five minutes before handling This will reduce the risk of boil over and scalding Regularly inspect the autoclave
41. se to prevent the formation of a biofilm Throw the used solution away Prepare a fresh bath for each load Jeannette de Vries ICRC 9 3 Cleaning disinfection and drying Aim Elimination of blood secretions and debris via the chemical action of a detergent Hexanios combined with the mechanical action of brushing and rinsing All items must be carefully cleaned before sterilization Instruments and equipment must be thoroughly cleaned and disinfected otherwise the sterilization process will not be effective Procedure HEXANIOS 0 5 Instructions for use concentrated solution to be diluted at 0 5 before use gt 1 stroke 25 ml of the dosing pump for 5 water gt 2 strokes 50 ml of the dosing pump for 10 water How Prepare a soaking bath solution according to the volume of medical instruments 5 or 101 Immerse the instruments completely in the soaking bath scrub them thoroughly and leave them to soak in the solution for 15 minutes Rinse thoroughly with clear water discard the rinsing water and dry with a clean cotton cloth This solution can be kept for a maximum of 24 hours Prepare a fresh bath immediately if it is visibly contaminated dirty 9 4 Checking to ensure instruments are and remain intact Medical instruments are expensive precision instruments Small deposits of blood may easily develop into serious corrosion rust This corrosion is aggravated by the moisture and high temper
42. sess air removal efficiency in steam sterilizers used in pre vacuum sterilization cycles It detects air leaks and inadequate air removal and consists of folded 100 cotton surgical towels that are clean and preconditioned It is mandatory for pre vacuum sterilizers and has to be carried out ona daily basis in an empty autoclave before the first load A commercially available Bowie Dick test sheet is placed in the centre of the package The test package is placed horizontally in the front bottom section of the sterilizer rack near the door and over the drain in an otherwise empty chamber The test is run at 134 C for 3 5 minutes The test is performed each day the vacuum type steam sterilizer is used before the first load is processed Air that is not removed from the chamber will interfere with steam contact Smaller disposable test packages or process challenge devices have been devised to replace the stack of folded surgical towels for testing the efficacy of the vacuum system in a pre vacuum sterilizer These devices are designed to simulate the product to be sterilized and to constitute a defined challenge to the Julie Barnet ICRC sterilization process They should be representative of the load and simulate the greatest challenge to the load Sterilizer vacuum performance is acceptable if the sheet inside the test package shows a uniform color change Entrapped air will cause a spot to appear on the test sheet because the steam
43. sing and cleaning of instruments e 1001 operation entire instrument load for any kind of surgery e 301 delivery all instruments for a non surgical delivery 301 5 dressings or 5 sutures Autoclave e 481 TBM90 I autoclave cycle 8 161 45 autoclave cycle High level disinfection e 351 procedure Cleaning disinfection of the room and staff hand hygiene e 301 day minimum For info Water supply needed in the operating department 100 surgical operation Important The water for the autoclave drying cycle must be cold lt 25 C if possible Water hardness should be soft to avoid limescale e Water consumption temperature and pressure for larger autoclaves are according to manufacturer specifications e g Matachana Getinge Cisa Consult with the biomedical engineer in Geneva before installing these types of autoclaves 2 See P Perrin H E L P Public health course in the management of humanitarian aid ICRC Geneva 2001 Julie Barnet ICRC STERILIZATION GUIDELINES 17 9 THE STERILIZATION CYCLE Preventing the spread of infection Move in one direction only Never move backwards Never cross clean and soiled instruments 9 1 Collection Use perforated baskets placed in closed containers Julie Barnet ICRC Julie Barnet ICRC Figure 3 Perforated baskets Why safe collection less handling of soiled instruments easy and safe transport to the sterilization room less mix
44. t e g Ajax Colgate Palmolive to remove the fat layer that is formed by the successive application of Surfanios 5 Non ionic and anionic detergents chelating agent bacteriostatic and fungistatic agents colour ing and citrus fragrance STERILIZATION GUIDELINES 39 ANNEX V TOP LOADING VERTICAL AUTOCLAVE WITHOUT VACUUM PUMP E G TBM 90 LITRE Loading Unloading Make sure that the load is homogeneous linen or Wait until the end of the sterilization cycle instruments because these items do not respond the same way to steam It is harder for steam to penetrate linen condensation is different as is drying Let the autoclave cool down and the drying finish before opening the autoclave Do not crowd the packages steam must be able to circulate USENET IOS freely Remove the crepe paper that was placed under the lid Make sure that the packages do not touch the walls of the Empty the autoclave baskets place the packages on a clean autoclave chamber dry surface Place packages in the autoclave chamber filling the lower Check to make sure the wrappers are intact and dry basket first small packages can be placed vertically but instrument sets packaged in wire baskets should be placed horizontally Unwrap the TST strip and check that it has changed colour Do not overload the lower basket the load should not extend beyond the top of the basket otherwise the upper basket might res
45. t on the load below and tear the wrappers Crepe paper sterilization sheets are fragile Load the second basket the same way For each basket place the TST strip packaged in a double layer of crepe paper in the middle of the load 40 STERILIZATION GUIDELINES ANNEX VI FRONT LOADING HORIZONTAL AUTOCLAVE WITH VACUUM PUMP WITH OR WITHOUT STEAM GENERATOR E G MATACHANA SC 500 Loading Unloading Jo minimize condensation make sure that the load is homoge Wait until the load has finished neous either linen or instruments O Meie nurs irie eive Fill the chamber two thirds full Put on heat resistant gloves fae crowd the packages steam must be able to circulate Empty the autoclave place the packages on a clean dry sur face Place the packages on the autoclave shelf wire tray or bas O ere kets the packages must not touch the walls of the autoclave e DE EME sim Aaa Ie chamber Unwrap the TST strip and check that it has changed colour Separately wrap the TST indicator strip in a double layer of Remove the sterilization cycle graph or the printed ticket from crepe paper and place it in the middle of the load the printer check to make sure it matches the reference graph write the date and cycle number on the paper and ar chive it STERILIZATION GUIDELINES 41 ANNEX VII OPERATING INSTRUCTIONS FORTHE TBM 90 LITRE AUTOCLAVE STERILIZATION WITH 90 LITRE AUTOCLAVE TBM Aut
46. tee of the Red Cross ICRC is an impartial neutral and independent organization whose exclusively humanitarian mission is to protect the lives and dignity of victims of armed conflict and other situations of violence and to provide them with assistance The ICRC also endeavours to prevent suffering by promoting and strengthening humanitarian law and universal humanitarian principles Established in 1863 the ICRC is at the origin of the Geneva Conventions and the International Red Cross and Red Crescent Movement It directs and coordinates the international activities conducted by the Movement in armed conflicts and other situations of violence ICRC 4218 002 12 2014
47. the heating source 7 Open the purge pressure valve set to vertical position and put on full heat 8 As soon as a continuous jet of steam comes out close the purge valve Purging 9 Let the pressure rise until 2 1 bar 134 C then open the purge valve 10 Let the pressure fall to 0 2 bar then close the valve 11 Repeat steps 9 and 10 twice more to eliminate all air pockets 12 Let the pressure go back up to 2 1 bar 134 C again Sterilization 13 As soon as the pressure reaches 2 1 bar the regulating valve starts to release steam continually The heat may be adjusted to maintain a constant temperature pressure 14 Time 18 minutes from now using the timer Check that the pressure remains correct and constant during sterilization Adjust the heat if necessary Expelling steam and water 15 When the sterilization process is finished after 30 40 minutes cut the heat open the purge valve and let the steam escape 16 When the pressure has fallen to 0 open the lid 17 Remove the basket and place it on a clean towel 18 Empty the clean water in a clean container for later reuse Caution Be careful as the autoclave is still very hot 8 Incase of difficulty opening the lid reheat for a few minutes without taking into account the pressure stop the heating source and open the lid immediately 46 STERILIZATION GUIDELINES Drying period 19 20 21 22 23 If wet remove the clot
48. to make sure it is working properly Have a trained hospital technician follow a maintenance schedule Medical Equipment Procedures in ICRC Hospital Care Projects R Ayer May 2013 3 The ICRC follows the Swiss Ordonnance sur la pr vention de la maladie de Creutzfeldt Jakob lors des interventions m dico chirurgicales Statute on the prevention of Creutzfeld Jakob disease during medical and surgical activities of 20 November 2002 OMCL RS 818 101 21 9 7 Control Aim For every load it must be possible to prove that the autoclave was working properly on that date and that the sterilization cycle achieved the necessary standard To that end ISO and the CEN recommend recording the physical conditions or parameters in the sterilizer during the process 9 7 1 Checking the sterilization cycle 9 7 1 1 Chemical indicators TST STRIP These indicators undergo a chemical change from yellow to dark blue when exposed to Time Steam Temperature TST gt A TST strip must be put in the big surgical wire basket to monitor time steam and temperature inside the packages This is not needed for small sets like dressing suture and delivery sets A second strip wrapped in a double layer of crepe paper must be placed in the middle of the autoclave load to monitor steam and temperature inside the autoclave Figure 9 TST strip Tip Write the date of sterilization and the autoclave and cycle number on the TST strip
49. undry STERILIZATION GUIDELINES 29 30 STERILIZATION GUIDELINES 11 2 Collection In the operating theatre All used linen must be collected as soon as possible at the end of each operation by the operating theatre or auxiliary staff In the wards All used linen must be collected at least once a day from each ward including changing rooms by the nursing or auxiliary staff In practice two bags must be available one for wet linen or linen soiled with organic matter e g blood faeces mucus wound exudate preferably a leak resistant container or plastic bag one for dry soiled but visibly clean linen preferably a cloth bag Steps 1 Check the linen especially from the operating theatre for sharps 2 Place wet linen or linen soiled with organic matter sheets drapes etc preferably in a plastic bag inside a leak resistant container bin with a lid 3 Place dry soiled but visibly clean linen i e sheets and work clothes in a cloth bag 4 Replace the bags when they are full Close the plastic bags with a knot 5 Transport the bags on trolleys to the dirty area of the laundry 6 Donot overfill the bags so that they cannot be properly closed fill two thirds full 7 Do not mix linen soiled with organic matter and used unsoiled linen 8 Keep the linen bin closed with the lid when it is not in use 9 Be sure not to soil the outside of the bags and bins by handling them with
50. ve 24 24 25 26 26 27 27 27 28 28 28 29 29 30 30 30 31 31 32 33 34 36 37 39 40 41 44 45 STERILIZATION GUIDELINES 1 TERMS AND ABBREVIATIONS Free from disease causing contaminants The number of viable micro organisms in or on an object entering a sterilization process A thin usually resistant layer of micro organisms e g bacteria that forms on and coats various surfaces European Committee for Standardization Visual inspection and manual testing to ensure instruments are intact and perform properly Consists of removing organic and non organic substances through the physical chemical action of a detergent combined with a brushing and rinsing action The presence of a minor and unwanted constituent contaminant on a material or in a physical body The use of disinfectant solution to destroy or prevent the growth of organisms capable of infection The elimination of vegetative forms of bacteria and viruses by immersing pre disinfected and cleaned instruments in a disinfectant solution used for semi critical materials that cannot tolerate steam sterilization Human Immunodeficiency Virus International Committee of the Red Cross International Standardization Organization Any medical equipment used on a patient for the purpose of preventing detecting diagnosing or treating a pathology Destroying or reducing the infectivity of microbes such as