WFM6100/WFM7000/WFM7100 Waveform
Contents
1. Show Measure Quick Start User Manual Timing a Studio Timing a Studio Timing a studio involves adjusting the references going to different sources so that their output feeds have the same timing when they arrive a common point such as a production switcher For digital systems the timing typically only needs to be close because most switchers have some tolerance to timing errors For analog composite systems the timing may need to be matched within a small part of a subcarrier cycle to prevent hue shifts when switching between sources The waveform monitor supports multiple methods and techniques of timing a studio All of these methods require an external reference to the waveform monitors Using the Timing Display Method The Tektronix proprietary Timing Display patent pending provides an easy to use measure of the timing of an input relative to the external reference The rectangular display automatically scales to match the input signal For progressive signals the display represents one field for interlace signals the display represents one frame and for composite inputs the display represents one color frame The cross hair in the center represents zero offset and the circle represents the timing of the input signal Lines of advance or delay are shown as vertical displacement while timing errors of less than one line are shown as horizontal displacement If the input is at the same time as the reference then the c
2. Print What Sc pn Event Log Test Page 1890 045 Print What Page Properties Page Properties Portrait Screen Lv Close 10 20 50 100 200 500 Print Page Limit ee L jJ 525i 59 94 SDI Input A Ref Internal Thu May 04 14 44 53 C High 120 IRE O 7 C Low 24 IRE Y High 103 0 Stylus OY Low 1 0 C Area 0 OY Area 0 Print What Page Limit Page Properties Select Printer Print Screen 10 59 How to Operate Your Instrument 9 Touch the Print soft key to print the nt V ee iG a a selected item on ge Properties Select Printe Creating a Printer Before you can print anything you must create a printer for the Select Printer dialog To create a printer 1 Press the MAIN button to display the Main menu ware 2 Touch the Print soft key to display the S zA Configure Print submenu Mount User Menu 3 Touch Select Printer to display the Select Printer dialog box 56 Quick Start User Manual How to Operate Your Instrument 4 Touch the Create soft key to adda printer to the list Local USB Printer 5 Scroll through the Printer Make and Model listing to select an appropriate printer 6 Select a USB Device identifier Use different identifiers for different printers USB Printer 1 7 Touch the Keyboard soft keys to enter a name for the new printer and then touch ca TTT TT Tr PEPE EEE b
3. Quick Start User Manual How to Operate Your Instrument To Measure Waveforms with Cursors Cursors enable you to measure time or voltage on a waveform Cursors are available in tiles set to Waveform Eye or Jitter mode To display and adjust cursors Tue Apr 25 180738 1 Choose a tile that is currently displaying a waveform eye pattern or Jitter display V Gain x1 00 H Mag x1 00 Overlay Color Space Components Filter Sweep Settings Cursors YPbPr YPbPr Flat 1 Line 2 Touch the Cursors soft key to display Close ae Show Volt Show Time the Cursors submenu Cursors Independent To Center 3 To display cursors touch the soft key for the type of cursors you want displayed so that the box is checked 4 Touch Track Independent to specify how cursor pairs move with respect to each other Select Track if you want both cursors V1 V2 or T1 T2 to move together as a single unit Select Independent to move each cursor by itself Quick Start User Manual 39 How to Operate Your Instrument 5 To move a cursor touch the cursor SC Wed Apr 26 095827 readout on the display and turn the large S ES knob to move the selected cursor on the waveform The active cursor readout appears in yellow with a filled in knob icon 6 Repeat the process to adjust the other cursor if cursors are set to independent 7 Read the cursor measurement in the Cursors readout NOTE If the display
4. OY Gain x1 00 H Mag x1 00 Close Colors amp Readouts amp Diagnostics Closed Caption Alarm Setup Utilities Config Intensity Brightups 3 Touch any soft key to get help on that Wed Apr 26 18 40 26 soft key rna 4 Use the navigation keys at the bottom of the help window to scroll through the help text 5 Touch Close to remove the Help window 6 Touch the icon in the Status Bar to turn off the context sensitive help Back Page Up Scroll Up Scroll Donn Page Doun Quick Start User Manual 49 How to Operate Your Instrument Displaying help when there is no Status Bar displayed is slightly different Instead of selecting the icon you select the Help soft key To display online help when no Status Bar is visible Exit Ballistics Digital Settings True Peak 1 Touch the Help soft key to enable Help eee a eee lbh mode 8dBFs 18 dBFS 0 dBFS Scal Scal L scale S Hegt offer naan 70 dB 0 dB Peak Error Peak held segment OFF hold time Ol hold time 2 Sec 2 Sec Peak held segment ON Samples Samples fee Clip O epee an E P Silence Duration Over level Duration level for Silence O for Over 70 dBFS 5 Sec 0 dBFS 1 Sec gti Liss AGC Off e Use Validity Bit FAO Liss AGC On Ignore Validity Bit 2 Touch any soft key to get help on that soft ke Exit Ballistics y Digital Settings True Peak 3 Use the navigation keys at the bottom eee ia ae GL of the help win
5. Place Err Field Length Err F Line Length Err Fl AP CRC Statistics RGB Gamut E Cmpst Gamut Error L G ID Embd PPPP VITC 09 31 41 26 2 Video Session Data Collect R ing i 0 d 03 59 55 un ime Auto 525i 59 94 one Ancillary Data Stuck Bits F2 AP CRC Err Secs Err Fields Err Fields 14394 862502 100 0000 31 7 AP CRC Error 0 EDH Error 0 0 Standby Help Embd PPPP VITC 09 39 05 07 2 H Mag x1 00 Video Session SDI A Data Collect R ing i 0 d 04 07 19 un Time Auto 525i 59 94 None Ancillary Data Stuck Bits F2 AP CRC Err Secs Status Err Fields Err Fields 889111 100 0000 889111 AP CRC E EDH Err 0 25 How to Operate Your Instrument To Set Display and Measurement Parameters You adjust display parameters and measurement settings using soft key menus Soft key menus appear at the bottom of the display and for measurements the settings apply to the active tile For example suppose you displayed a WFM measurement in three tiles The Sweep setting for each tile is set independently so even though the soft key menu for each tile would be identical 1 Select the tile whose parameters you want to change by touching the tile not necessary in 1 View mode 2 Select the parameter you want to change from the soft key menu 3 Some settings are accessed from the Settings soft key Depending on the setting you might need to ma
6. Data outside the nominal allowed values m Red Data with illegal values ANC Data Display Pressing the MAIN button calls up the Main pop up menu From the Main pop up menu you can select the ANC Display display under the Measure heading The ANC Data Display shows the raw undecoded data words in the selected ancillary data packets The ancillary data is shown only in hexadecimal format Since the display is updated at approximately two second intervals rapid data changes cannot be detected Use the large knob and the Select button to enter the DID and or SDID values for the type of ancillary data packet to be monitored You can also touch the readouts on the screen to select either DID or SDID for control using the large knob DID and SDID values have the following characteristics m DID and SDID Values range between 0x01 and Oxff m For DID values in the range 0x00 Ox F Type 2 packets the SDID values are used m For DID values in the range 0x80 OxFF Type 1 packets there are no SDID values used the SDID soft key is grayed out Quick Start User Manual Tue May 02 17 53 02 525i 59 94 ID SDI Input A Audio In Dolby 1 AES Al 2 Ref Internal ANC Data Display DID 0 SDID 0 DC Field Line Stream Status Checksum Should be Format 000 O16 032 048 064 080 096 112 128 144 160 176 T1 How to Operate Your Instrument Status Displays Status displays are text displays that show sig
7. E E C Anc Parity Error g oO E E SMPTE 352M Missing oO go oO O O 9 Touch the Exit Alarm Setup soft key to close the alarm setup screen and return to the instrument display Quick Start User Manual 91 Monitoring the SDI Physical Layer To Take Eye Measurements After you have configured the instrument for Eye measurements you can use the cursors to take manual waveform measurements or you can use the automatic eye measurements function Option Phy only To Set Up for Eye Measurements 1 Configure the instrument to monitor the SDI physical layer See page 86 To Configure the Physical Layer Settings 2 Connect a serial video signal to the instrument using a 75 cable of two meters or less in length Use high quality low loss coaxial cable such as Belden 1694A 3 Press the MAIN front panel button 4 Touch the Eye soft key to display the Eye soft key menu The procedures on the following pages describe how to take manual measurements of the Eye waveform If Option Phy is installed the instrument can also take automatic eye measurements NOTE If you use the measurement cursors to measure the eye waveform you may measure a different result than that shown in the automatic eye measurements readout This is because the automatic amplitude measurement uses histograms in order to be minimally affected by overshoot ringing and noise on the signal Similarly the automatic rise time and fall time measurements use histogr
8. Features m A high resolution XGA display with the FlexVu interface and high quality waveform processing CaptureVu capability that can capture a full frame of video data either user initiated or triggered by an alarm condition m MyMenu capability that lets you create custom menus that simplify instrument operation m Available measurement options that offer industry leading capability for SDI signal measurement and video data analysis m Available audio options that offer industry leading audio monitoring capability especially for Dolby audio formats m Available composite video option that offers analog monitoring capability m A wide range of options and upgrades that let user create instrument configurations best suited to their current needs with the ability to add features later as their needs change All the key features of the WFM7100 except CaptureVu measurement options and Dolby audio support All the key features of the WFM7100 except support for HD digital video m High resolution XGA display with the FlexVu interface and high quality waveform processing providing outstanding picture and signal displays and exceptional flexibility in creating multi view displays yi Quick Start User Manual Preface Extensive fault detection alarm generation error logging and status reporting including quality statistics on both video and audio content Up to four independent picture displays of the
9. Headphone Jack For listening to audio g USB port For attaching peripherals such as a printer or USB nO Flash Drive The USB port is USB 1 1 and 2 0 USB compliant full speed 12 Mb s rate canes Elements of the Display Areas of the Display Three major areas make up the display Status bar They are the Status Bar area the Main Display area and the Menu area The Main Display area is where the waveform monitor displays measurement results signal traces and status information The Status Bar appears at the top of the display The Status Bar provides a number of text and icon elements to give you an easily viewed guide to the status of the instrument and the monitored signal The Menu area is where the waveform monitor displays menus The menus provide access to settings that select measurement displays measurement settings instrument configuration parameters and more Lum rror 2111 126 0 EDH Error 0 0 0 0000 Overlay Color Space Components Filter Sweep ag YPbPr YPbPr Flat 1 Line 1890 016 Main display area Menu area Quick Start User Manual 13 How to Operate Your Instrument Making Selections from a Menu A menu consist of a group of soft keys There are several ways to respond to soft keys How you make a selection depends on the type and number of choices available for the soft key The types of menu keys are command toggle option knob value popup option and check box soft keys Command So
10. Options 2 Eye 85 Phy 85 power cords 1 Other display elements of 69 Overlay Waveform menu 62 p Parade Waveform menu 62 Part numbers manuals 1 Phase display 116 117 Phase Pair Audio menu 69 Phase Style Audio menu 68 Phase Normal V Vector menu 64 Picture button 64 display 64 menu 65 PICTURE button 12 Pix Mon Output connector 23 Power AC requirements 4 Connecting 4 Connector 17 Switch none required 4 Power cord options 1 Power connecting 4 Power up Preset changing 35 36 Presets copying from a flash drive 32 copying to a flash drive 32 using 30 PRESETS button 12 Printing creating a printer 56 how to 54 Product description vi R Readout Format DataList display 76 Readouts amp Brightups soft key Config menu 48 Index Rear panel Layout and description 16 Recall Presets 31 Reference signals supported 18 Rel to Rear Offset Timing display 73 Rel to Saved Offset Timing display 73 Relative to Timing display 73 Remote Application 125 Remote connector 23 Remote Control Port 52 Remote Web Interface 52 124 RGB Color Space submenu 62 RGB gamut Diamond display 107 S Safe Area Picture menu 65 Safety Summary iii Save Presets 31 Save Offset Timing display 73 SDI gt Composite Color Space submenu 62 SDI input connectors 18 SDI Status display 85 SELECT button 12 selecting audio input 114 Settings Picture menu
11. PPPP SDI Input A Ref Internal man 255 255 248 0 128 181 216 1 O j OV Gain x1 00 Y H Mag x1 00 Config Mode Network Instrument View Hw Sw SNMP Setup y Manual Setup Name Version vna we STIs ASS LEEST in 525i 59 94 Wed Apr 26 19 37 08 ID SDI Input A Hide Menu Embd PPPE Ref Internal a O 000 000 000 000 x 000 000 000 000 000 000 000 000 es ee nstrument View Hw Sw SNMP Setup Manual Setup Name Version uw PiWwiiom SS LEEST ind Quick Start User Manual How to Operate Your Instrument Change the settings based on your requirements m SNMP Traps This setting allows you to turn on or off the capability of the waveform monitor to generate SNMP traps m SNMP Public Comm String This setting specifies the string that SNMP commands must include to have read only access The default public community string is public If you enter a new string the waveform monitor will strip any leading or trailing spaces in the string for example my stuff will become my stuff The public community string cannot be blank If you input a blank string the instrument will set it to the default string m SNMP Private Comm String This setting specifies the string that SNMP commands must include to have read write access The default private community string is private If you enter a new string the waveform
12. 1 Push the PRESETS button Tue May 09 1751 08 2 Touch one of the soft keys in the Presets menu Unless you have renamed them they will be labeled A1 A5 OV Gain x1 00 H Mag x1 00 Group Settings Al A2 A3 A4 AS Quick Start User Manual 31 How to Operate Your Instrument 3 Ifa the preset you want is not in the Tue May 09 1753112 displayed group Touch the Group soft key and select the appropriate group from the submenu The front panel setup will switch to the saved preset corresponding to the soft key pressed OV Gain x1 00 H Mag x1 00 Close A B G D Settings Al Group I I I eee To Copy a Preset to or from a USB Flash Drive 1 Insert a flash drive into the USB port on 7 the front panel N USB 1890 013 2 Touch the PRESETS button 5251 59 94 Tue May 09 1751 08 Ref ee 3 Touch the Settings soft key OV Gain x1 00 H Mag x1 00 Group Settings Al A2 A3 A4 A5 4 Touch the Copy Preset s soft key Close Save Recall Rename Copy Power up Preset Settings Settings Preset Preset Preset s Preset s Latest Lz 32 Quick Start User Manual 5 Verify that the USB flash drive is recognized a description of the drive will appear after USB Storage Device If it is not recognized touch the Reset soft key to mount the flash drive Internal m Ifthe flash drive is still not recognized the flash drive may not be compatible try using another flash drive 6 Touch
13. 68 AES V BIT Indicates that the Validity bit is set high for one or more data samples In the AES EBU standard a set validity bit indicates that the sample is not suitable for conversion to audio By default the level meter bars and Lissajous display treat the affected samples as zero samples Audio Menu The Audio menu enables you to select the audio input source and specify whether or not the signal phase is displayed Aux Display If the signal phase is displayed you can also specify the style of the phase display and the pair of inputs that is shown in the phase display To display the Audio m Press the AUDIO button to display the menu To select the source for the Audio display select Audio Input in the menu and select a source The number of available sources depends on which audio option is installed Choosing any given input will make that audio the monitored signal regardless of which video input is active Alternatively you can choose Follows video to enable the mapping that allows changing the audio source as the video input selection is changed NOTE Use the Audio Settings menu to select the mapping of input to bar the meter type and the Follows video mapping of audio to video The Audio menu also provides two pictorial displays that allow you to monitor the signal Audio phase displays include phase correlation meters a flexible Lissajous display for any pair of channels and a multiple channel Surround Sound
14. A one line area that is visible if Luma errors are present An icon that is visible when alarms of types other than those in the four readouts just listed occur Touching this soft key causes the menu at the bottom of the display to appear or disappear Touching this icon enables disables the context sensitive help mode While enabled the icon color changes to green and touching any soft key displays help for that soft key Freeze appears when a Capture Trace or Capture Buffer action has been performed A readout showing the selected time format A 16 character string indicating embedded audio channel status Each character shows the status of a specific channel for not present and P for present The instrument name set in MAIN gt Config gt Utilities gt Communications gt Instrument Name Readout of the date and time set in MAIN gt Config gt Utilities gt Set Clock How to Operate Your Instrument Rear Panel Connectors The following figures shows the waveform rear panel with optional connectors An instrument without the Eye or Phy options has the SDI inputs located along the bottom right portion of the rear panel An instrument with an Eye or Phy option has the SDI inputs arranged vertically AIN 1 2 B 1 0 1 2 OO aes AN34 B 1 0 3 4 O LIL ILILJ 8 AINS6 B1056 OOOO okeee o AIN78 B1 07 8 oo0o o REMOTE A 00000 00000 00000 0000 Eole 00000 EJ EXT DISPLAY A ETHERNET S
15. On this display you can use the voltage and time measurement cursors and their readouts to monitor amplitude and timing measurements on the Eye waveform A jitter thermometer and readout displays the jitter magnitude You can set the high pass filter to allow measurement of different jitter types timing and alignment The jitter thermometer graphically relates the jitter measurement to the alarm limits m Jitter Display On this display you can view additional time domain information such as that listed below that is useful in separating sources of jitter whether they are within a single circuit on a circuit board or from various pieces of equipment in a system Jitter components that are synchronous or nearly synchronous to video line or frame These appear as stationary or near stationary artifacts in line or field sweeps The wave shape of the jitter as modified by the high pass filter setting m SDI Status Display This display shows both the Jitter1 and Jitter2 thermometers described above Additionally a Cable Loss thermometer indicates the signal loss due to cable length Use the Eye Settings menu to configure the instrument for the cable type and allowed length With Option Phy installed the SDI Status Display also shows the eye amplitude eye rise time eye fall time and eye rise fall delta measurement values To Configure the Physical Layer Settings Before you use Eye display to monitor the SDI physical layer you should c
16. Power Cords The WFM6100 WFM7000 WFM7100 Waveform Monitors are shipped with one of the following power cord options Power cords for use in North America are UL listed and CSA certified Cords for use in areas other than North America are approved by at least one authority acceptable in the country to which the product is shipped International Power Plugs re m Opt AO North America power E Opt r A1 Universal EURO power F m Opt A2 United Kingdom power m Opt A3 Australia power mr m Opt A4 240 V North America power rme E Opt Ad Switzerland power m Opt A6 Japan power mre m Opt A10 China power m Opt A99 No power cord or AC adapter Optional Accessories m WFM7F02 Portable cabinet with handle feet tilt bail and front panel cover Quick Start User Manual 1 Before Installation m WFM7F05 Dual rackmount for WFM6100 WFM7000 WFM7100 1700 Series WFM601 Series WFM700 Series 760A and 764 Each half of the rack can be ordered as either Option O or Option N Option N is used for WFM700 WFM6100 WFM 7000 and WFM7100 monitors Option O is used for 1700 Series WFM601 Series 760A and 764 instruments m WFM6100 WFM7000 WFM7100 Waveform Monitors Service Manual Tektronix part number 071 0915 XX Options A number of hardware options are available for the WFM6100 WFM7000 amp WFM7100 Waveform Monitors The following table lists the options that a
17. cursor V1 is selected by default Quick Start User Manual Monitoring the SDI Physical Layer O 2 492 0 mV A 984 0 mV O yi 492 0 mV 3 A 996 0 mV Close Cursor Show Volt Show Time Cursors Independent To Center 93 Monitoring the SDI Physical Layer 94 4 Use the general purpose knob to position voltage cursor V1 at the peak of the overshoot at the top horizontal part of the waveform Select cursor V2 and use the general purpose knob to position the voltage cursor V2 at the bottom of the top line of the waveform Include any ringing the oscillation following overshoot in the measurement Essentially you are measuring the thickness of the top line of the waveform The cursor Delta readout displays the amplitude of the aberration Perform the same voltage cursor measurement on the bottom line thickness including any undershoot and ringing NOTE Aberrations at the top or bottom line of the waveform should not exceed 10 of the signal amplitude Automatic equalizer circuits in receivers may be sensitive to larger aberrations 5251 59 94 Wed May 17 17 11 30 SDI Input A Ref Internal O yi 492 0 mY O HMag x1 00 A 40 0 mV cn NR A S OA ELENEN O yi 492 0 mY a A 40 0 mV To Manually Measure Rise Time Using Variable Gain 1 Perform the initial setup for Eye measurements See page 92 To Set Up for Eye Measurements Press the MAG GAIN front panel butt
18. the balanced outputs to an unbalanced input you must ground the unused lead You can ground either lead NOTE Note that grounding the unused lead does not attenuate the output but it does halve the clipping level Therefore you must attenuate the output by at least 6 dB to avoid clipping The output signal level in unbalanced mode is double the signal level in balanced mode Units that have both analog and digital capability can have AES or embedded inputs converted to analog and then routed to the six balanced outputs Embedded audio can be output to the AES B connector when it is configured as an output AES A can also be routed to the AES B connector Decoded Dolby can also be routed to the analog output connector EXT DISPLAY Connector Pin Assignment This is the display output The display resolution is 1024 x 768 The output is compatible with standard analog PC monitors either CRT or LCD based The EXT DISPLAY connector is a 15 pin D type connector with socket contacts 5 1 Pin Pin name D000 gt 1 Red Video o s E 2 Green Video 15 11 1890 006 3 Blue Video 4 Not connected 5 Ground 6 Red Ground T Green Ground 8 Blue Ground 9 5 V for monitor EEPROM 10 Not Connected 11 Not Connected 12 ID Bit 13 Horizontal Sync 14 Vertical Sync 15 ID Clock Quick Start User Manual How to Operate Your Instrument Pix Mon Connector Pin Assignment This is the video picture output The output is compatible with standard analog P
19. the cursor is enabled in the Main gt Config gt Readouts amp Brightups menu using the Pix Brightup Line setting 525i 59 94 Wed Apr 26 17 43 09 SDI Input A Ref Internal Line F1 114 Active Word 1492 Cb HB Quick Start User Manual How to Operate Your Instrument To Adjust Headphone Volume and Source 1 Press the AUDIO button AR R mie E A Hide menu Embd PPPP PPPP PPPP PPPP NOTE The headphone icon appears at the bottom of the Audio display It appears _True Peak Input EMB A on the left if the Aux Display is off The T T headphone icon appears on the right if the Aux display is set to Phase or Surround i2 Ride eo Aux Display Phase Style Phase Pair FlexPhase Ch A FlexPhase Ch B Audio Input g Phase Soundstage Bars 1 amp 2 Audio Bar 1 Audio Bar 2 Follows Video 2 Touch the headphone icon The headphone icon changes color to yellow when it is selected WARNING To avoid damaging your hearing before you put on headphones always turn the volume down to the minimum and then turn it up slowly 3 Adjust the headphone volume using the large knob The color of the headphone bar changes to give an indication of the volume level 4 To change which meter bars are used as the source for the headphones touch either meter bar in a pair to select that pair as the source 5 Touch the headphone icon after you are finished adjusting the volume to disassociate it fro
20. 0 OY C Area 0 m The adjustable thresholds are OY Area 0 indicated by the dark blue horizontal graticule lines m The thresholds are defined in terms of percent of full scale m The range for the Upper limit is 90 r to 108 Default Limits yyy E SSS The range to the Lower Limit is 6 aie to 5 110 Quick Start User Manual Checking Gamut Another useful Arrowhead function is a measure of how well the active video signal is using the dynamic range of the video channel A properly adjusted signal should be centered in the arrowhead graticule and have transitions that approach all the limits this won t be true of dark scenes or scenes with low color content Automating Gamut Checks You can use alarms to automatically monitor for out of gamut conditions 1 Press MAIN to display the Main menu ZED Thu May 04 15 49 24 Ref Bae 2 Touch CONFIG to display the SDI Status Configuration menu Jitter F P E Jitter2 Vector Arrowhead ANC Display Error Log P P Picture Aux Status Audio Session SDI Status Dolby Status Lightning USB Configure Print Config Standby Help Mount User Menu 3 Touch Alarm Setup to display the alarm Close TE Colors amp Readouts amp 7 Diagnostics Closed Caption Alarm Setup Utilities setup display Config Intensity Brightups 4 On the alarm setup display touch Video Exit Content A a E Video Content AES and Embedded Video Format Embedded Audio Specific SDI
21. 12 From the Main menu touch the USB Unmount soft key and them remove the USB flash drive from the instrument To Recall the Factory Preset 1 Push the PRESET button 2 Touch the Settings soft key to display the submenu 3 From the Settings submenu touch the Recall Preset soft key 4 From the menu in the center of the display touch the Factory soft key Fri May 19 10 43 39 USB WFM7100 preset 525i 59 94 ID SDI Input A E60 22 Ref Internal VITC 14 21 19 14 2 Close Recall O V Gain x1 00 H Mag x1 00 Save Recall Rename Copy Power up Preset Settings Preset Preset Preset s Preset s Latest tay Quick Start User Manual How to Operate Your Instrument To Change the Power up Preset 1 Push the PRESET button 3 Thu May 18 18 41 44 T Hide menu TC 2 Touch the Settings soft key to display the submenu OV Gain x1 00 H Gain x1 00 Group Settings Al A2 A3 A4 A5 3 From the Settings submenu touch the Close Save Recall Rename Copy Power up Preset Power up Preset soft key Settings Settings Preset Preset Preset s Preset s Latest La 4 From the menu in the center of the S251 50 94 Thu May 18 1531 22 ID opr display choose any available preset for sie the instrument to use at power up EA bk 5 r g Po E NOTE The Latest preset is the instrument setup at the time the instrument was last shutdown using the Shutdown command MAIN gt
22. 142548282 configured differently OH Mag x1 00 k 1 00 OH Mag x1 00 True Peak Input EMB A in phase oO H Mag 8 Quick Start User Manual How to Operate Your Instrument Controlling the Waveform Monitor There are three ways to control the waveform monitor You can use the buttons on the front panel you can use the touchable settings on the display and you can use the menus that appear on the display including the Main pop up menu Use the front panel buttons to select frequently used displays to select modes LINE SEL that affect multiple displays change values using the large knob or change the horizontal and vertical position of a waveform ea re 6 oo i 1890 040 Many values can be set from the display by Wed Apr 26 0958 27 selecting a touchable readout in the display For example you can change the position of a cursor by touching the cursor readout and changing its position using the large knob Ovi 0 0m Ti 0 00 us OY Gain x1 00 H Mag x1 00 T2 62 50 us A 700 0m 4 62 50 us Overlay Color Space Components Filter Sweep Settings Cursors YPbPr YPbPr Flat 1 Line 3 Overlay Color Space Components Filter Sweep ems YPbPr YPbPr Flat Fes Pork Virtually every display has its own menu The display menu is used to change selections or set values for parameters that apply to the display Many menus have multiple levels Quick Start User Manual 9 How to Operat
23. 45 connector 1 8 1890 006 Quick Start User Manual To Select a Measurement How to Operate Your Instrument 1 After you have selected a tile you can choose what to display in it The Display is independent for each tile 1 Select a tile by touching it 2 Push the button corresponding to the measurement that you want to display in the selected tile 3 Select the measurement you want to display from the Main pop up menu 4 Repeat steps 1 through 3 until you have selected measurement displays for all tiles that you want to define 5 To display the same measurements in more than one tile select each tile in sequence and then choose the same measurement for each one The display at right shows the display with WFM selected for three tiles with different sweep settings NOTE The Audio and DataList displays can be displayed in only one tile at a time Quick Start User Manual 525i 59 94 SDI Input A Ref Internal Waveform Vector Picture 525i 59 94 SDI Input A Ref Internal Settinds Color Space Components Filter Sweep C g Parade YPbPr YPbPr Flat 1 Line Mon Apr 24 13 35 59 ma B 2 H Mag x1 00 aput EMB A Format 352M Payload SAY Place Err Lightning FL AP CRC Statistics Status RGB Gamut Error Cmpst Gamut Error G tE 60 1 2 3 4 5 i 7 8 i USB Configure Hohl Mount oe User Menu Mon Apr 24 13 43 23 Input Signal Format 352M Payload
24. 65 Tile 10 Show Instruction Dialogs soft key 61 Signal connection Line termination requirements 6 SNMP settings 51 Software Upgrade procedure 127 Sound Stage Phase Style 117 Specifications 133 137 Index Standard accessories 1 documents 1 power cord 1 Status display 78 Status bar Elements of 15 STATUS button 12 Surround Audio menu 68 Surround display 118 elements of 120 examples of 121 Surround Filter Audio menu 119 Sweep submenu 1 Field 62 1 Line 62 2 Field 62 2 Line 62 gt Termination Requirements for 6 Tile settings 10 Timing Chroma Luma delay 102 display 72 97 Relative to 73 Save Offset 73 Timing a studio Router inputs 99 Task description 97 Timing display method 97 Traces DataList display 76 138 Transmission Signal Identifier Aux Status display 79 Triggering Capture 42 U Upgrade kits 3 system software 127 USB port speed 13 supported standards 13 USER button 12 User menu how to create 58 Utilities soft key Config menu 48 V V Chip Rating Aux Status display 79 Vector display 63 102 display elements of 63 menu 64 VECTOR button 12 63 Vector menu Bar Targets 64 VERT knob 13 Vertical Offset Timing display 72 Video Data DataList display 76 Video input connectors 18 Video Session Main menu 81 View soft keys 11 Views button 11 W Waveform Components 62 display 61 display ch
25. 75 lines advanced Horizontal Offset 7 778 us delayed Relative to P Rear Panel Circle will be in the center when input Rel to Rel to Save Rear Panel Saved Offset Offset 1890 020 More complex timing relationships display multiple circles See page 73 Timing Displays for Simple Versus Complex Timing Rear Panel is the default setting in which offset is shown as zero when the input and reference are at the same timing at the rear panel of the waveform monitor If you select Saved Offset you can save the timing from one signal and then display the timing relative to that saved offset Quick Start User Manual How to Operate Your Instrument Timing Menu The Timing menu enables you to save a timing setting to compare with another signal and specify the zero point of the timing display To save the timing of the current input as an offset to the timing display use the Save Offset soft key The current timing becomes the zero point for the saved offset mode of the timing display This applies to both the cross hair target in the middle of the display and the numeric readouts NOTE You cannot save the timing offset if either the input or reference is missing or unlocked You also cannot save a reference when in internal mode Saving an offset in these conditions would lead to misleading results so it is not allowed by the instrument Save Offset allows you to measure the timing between inputs or to match multiple sig
26. Audio AUDIO button 65 display elements 65 menu 68 AUDIO button 12 audio input selecting 114 Audio Input Audio menu 68 audio level checking 116 audio phase checking 116 Audio Session Main menu 82 Aux Display Audio menu 68 B Bar Targets Vector menu 64 Basic operation 7 Bowtie display 75 brightups 105 C Cable Loss thermometer 85 Capture Buffer 40 42 clearing 43 copying to a flash drive 44 restoring from a flash drive 45 Trace 40 42 triggering 42 Using the capture display 40 CAPTURE button 12 Capture Display On Capture menu 41 CC 608 Service Picture menu 65 CC Services Aux Status display 79 Index Chroma Components submenu 62 Chroma Luma delay 102 Clear Capture buffer 43 Close Caption soft key Config menu 48 Closed Caption Type Aux Status display 79 Color Space RGB 62 SDI gt Composite 62 YPbPr 62 YRGB 62 Colors amp Intensity soft key Config menu 48 Compass Rose On Vector menu 64 Component Composite Vector menu 64 Components Waveform menu 62 Components submenu Chroma 62 Flat 62 Luma 62 Composite gamut Diamond display 109 Composite Input connectors 18 Config menu 48 Connecting AC power 4 Connecting signals Line termination 6 Connector Compatibility 6 Connector s AES 20 Analog Input Output 20 Ethernet 24 Pix 23 Power 17 Remote 23 Video input 18 XGA 22 Cursors Displaying 39 How to u
27. EEE D EEE E ESSE EEE EEG EEE EERE EES 39 Tean ure PEDI Aan nee E ere E ee E ee er i eer eee eee eer en 40 To Set Line Select MOE rcucanaanmiscounardainschunantaneddodau aang daw aaspiudtnet tain awnstignetaseantaat a aeinannriieiacansn 46 To Adjust Headphone Volume and SOurce cccc cece nce cence eee e eee e eee ee sense ede EE dea EE ESSE Se EE EE eEE EE EES 47 To Configure Your Instrument ueesssnnnnnnueersrrrrnrrrrrrrrrrrrrrrerrrrrrrrrrrerrrrrrrrrrrrrrrrrrrrrrrrrrrrrrrrrrr 48 TUSE ONDO A D oe a E E ee ee re eer 49 To Connect to a Network nuuussnnnuensnrruenrrrrurrrrrrrer rriren rnrn t rnnr DEE EE E EEE EE DEE EEE SSE EEEE EEE EEE EEE EES 51 TO Conheci Directly 0 a PO sruscrssioienoer eiger tineia a E iae i rani aea 54 10 Prnt screens and Event LoS oesiasoiitaninaasia eani A a E A EER aa 54 Torea a User NE E EEE E E AE E 58 Waveform Display a nh a anedier a ee eee 61 VOCIONE ISDIAY scans siccednce mune E EE E A E N E E E A 63 FEMS SIO A EEIEIE A AE O IE E A E O EEE 64 PRUGIO DIS OI e TEE E A E E E deers E TE E E E E E TEE T E E E E AT 65 Lightining DiSplay ecsssscninrensrrsrinriassdernidniin o be Sn stati Sten edo nda A E O a ain liege 69 Diamond DIS OAV srerrererisnipa damaien iee a e A i EEN R e a e E inie 70 PARMOMVENG AVIS DIAN eearri A a eran E TE EE EE E EEA E amen ante E E E EEA 71 BALLEBEI assent O I Sou tts a toes eave ease pe EIE acon Saat PA Sanwa A 72 Quick Start User Manual Table of Contents B
28. Picture Menu The Picture menu enables you to specify the closed caption service type and set safe areas From the Settings submenu you can set the aspect ratio SD only and chose whether to display only the active picture portion of the signal or the full frame picture With Active Picture selected only the active video portion of the signal is displayed and the aspect ratio is correct With Full Frame selected elements of the signal outside the active video are visible for SDI signals When full frame is selected you can see user data embedded audio and elements in the vertical interval To display the Picture menu press the PICTURE button on the front panel NOTE You can see signal elements outside the active video only when the PICTURE display is set to FULL SDI inputs only You will not be able to see sync signal elements on Composite signals Audio Display Pressing the AUDIO button brings up the Audio Display The Audio display provides level meters and a phase or surround display for monitoring audio signals The Audio display always shows the level meters and correlation meters When you choose to display the Aux Display you can choose to display a Phase display also known as Lissajous or a Surround display The left portion of the Audio tile displays the level meters and the right portion shows the Phase or Surround display You must have one of the following options installed to monitor audio signals Option DS monitor
29. SD and SMPTE 292 HD monitoring alarm thresholds back to the factory default values touch the Set to SMPTE259 292 Limits soft key Setting the SMPTE limits also sets both clock recovery circuits to measure both timing and alignment jitter This can be seen on the SDI Status display 3 To set individual threshold limits touch the Show Limits soft key in the Settings submenu to enable displaying of the limit values Which threshold limits are displayed depends on the installed options 4 Touch the limit value to set and use the general purpose knob to change the value NOTE You can use the SELECT front panel button to sequentially step through and select the threshold limit values 5 Touch the Close Settings soft key Setting the Equalizer Bypass 1 Touch the Settings soft key See page 86 Displaying the Eye Measurement Menu Quick Start User Manual Set to SMPTE Limits z Ferie Y Ampi Min 720 MY O Ampl Max 880 mV Rise Min 400 ps Rise Max 1500 ps Fall Min 400 ps O Fall Max 1500 ps Delta Max 500 ps Monitoring the SDI Physical Layer Ae aa PEPA SIPPIE atter Amp I7 0 20 UI O Cable Len Max 200 m Cable Loss Max 2 0 dB Launch Ampl Min 720 mV O Launch Ampl Max 880 mV Close E Settings Settings l pza Show Cable Type 8281 Set to C SMPTE Limits Lav rie Bypass E Show Limits ia Measure Equalizer L eum 89 Monitoring the SDI P
30. SDI stream If missing Absent will display otherwise the TSID is displayed as a hexadecimal number Wed Apr 05 08 38 28 ID Audio In Dolby 1 AES Al 2 Dolby Status Option DDE display shown Not Supported with Dolby E source Ref Internal Dolby Audio Status Program 1 Program Desc Text Tektronix Quick Start User Manual Dolby Format Channel Mode Program Config Dolby Source Dolby Data Rate Bitstream Mode Dolby E Frame Rate Dialogue Level Line Mode Cmpr RF Mode Cmpr RF Overmoad Prot Center Mix Lvl Surrnd Mix Lvl Mixing Level Room Type SMPTE Timecode Copyright Bit Original Bitstream DC Filter Lowpass Filter LFE Lowpass Filter Surround 3 dB Atten Surround Phase Shift Preferred Stereo Dmix Lt Rt Center Mix Lvl Lt Rt Surrnd Mix Lvl Lo Ro Center Mix Lvl Lo Ro Surrnd Mix Lvl Surround EX Mode Headphone Mode A D Converter Type 19 80 How to Operate Your Instrument Dolby Status Option DDE display shown with Dolby D source Dolby Status Option DD display shown with Dolby D source 525i 59 94 SDI Input A Ref Internal Dolby Format Channel Mode Dolby Source Dolby Data Rate Dolby Sample Rate Bitstream Mode Dialogue Level Line Mode Cmpr RF Mode Cmpr RF Overmod Prot Dolby Surrnd Mode Mixing Level Room Type SMPTE Timecode 525i 59 94 SDI Input A Ref Internal Dolby Format Channel Mode Dolby Source Dolby D
31. Standby gt Shutdown It is not the instrument setup the last time power was removed from the instrument OY Gain x1 00 H Gain x1 00 Save Recall Rename Copy Power up Preset Settings Preset Preset Preset s Preset s Latest Ea Quick Start User Manual 35 How to Operate Your Instrument To Rename a Preset or Group 1 Push the PRESET button 2 Touch the Settings soft key to display the submenu 3 From the Settings submenu touch the Rename Preset s soft key 4 From the menu in the center of the display choose the Source location if necessary 36 y Gain x1 00 O H Gain x1 00 Thu May 18 16 41 44 Ea ID Embd PPPP y Group Settings Al A2 A3 A4 A5 A Close Save Recall Rename Copy Power up Preset Settings Settings Preset Preset Preset s Preset s Latest Internal m Quick Start User Manual How to Operate Your Instrument 5 Touch the soft key for the preset or group name that you want to change Internal 6 Touch the Keyboard soft key and type in anew name for the selected preset or group Press Return to save the new name NOTE Preset names are limited to 11 characters Group names cannot exceed 8 characters Internal Quick Start User Manual 37 How to Operate Your Instrument 38 Press the Rename soft key to apply the new name Press the Exit soft key when you are finished renaming presets Internal
32. appear directly above the C level indicator The white tic mark will move right or left depending on the relative balance between the three channels A short bar to the left of the white tic mark indicates a positive correlation between L and C The bar will grow as the correlation decreases Like the L R PSI the bar will continue to grow at a 45 angle for negative signal correlations The bar to the right of the white tic mark behaves in the same way depending on the C R correlation This PSI indicator uses the same color coding as the other PSI indicators 120 Quick Start User Manual Usage Notes The following displays are examples of how the surround sound waveform appears for some typical signal types 1 Uncorrelated signals same level in the L C R Ls and Rs channels 2 Sine wave test tone same level in the L C R Ls and Rs channels All signals are in phase creating phantom sources between the adjacent channels Quick Start User Manual Selecting and Monitoring Audio 121 Selecting and Monitoring Audio 3 Same as 2 with the exception that Channel L is out of phase 4 Surround sound program with strong center channel presence Note that the center bar is above the left and right levels This indicates that the center level is greater and will be more audible than the ambient sound 122 Quick Start User Manual Selecting and Monitoring Audio 5 Surround sound program with weak center channel pr
33. cancel this operation Items that may be copied will be shown in Yellow Jitter2 HPF P P Navigate through the menus until you find the menu item you wish to opy then press the button at the top center of the screen Close Conf User incenedan Remove 1 Clear All eel act On Dialogs User Button User Buttons Quick Start User Manual Once Config User is enabled a icon appears in the Status Bar Only yellow colored soft keys can be added to your custom menu To adda soft key to your custom menu touch the icon The icon will change color to blue Once the icon is blue the next yellow soft key you touch will be added to your User menu For example if you wanted to add the Diamond display to your User menu since it is not available from the front panel you would touch the Diamond soft key lf the soft key you want to add to your User menu is not currently visible you must navigate to the menu where the soft key is located and then touch the icon to add the soft key to the User menu Quick Start User Manual 525i 59 94 Thu May 04 15 57 52 SDI Input A Ref Internal SDI Status Jitterl F EE Jitter2 PEs How to Operate Your Instrument 59 How to Operate Your Instrument 60 6 Once you touch a soft key that you want 8 in your User menu the waveform monitor displays the current custom menu To add your selection to the User menu touch one of the soft keys selecting an existi
34. center pin The BNC connectors on the waveform monitor are designed to work with the 50 Q standard large diameter center pins Do not use connectors or terminators with the smaller center pins They could cause intermittent connections 6 Quick Start User Manual How to Operate Your Instrument How to Operate Your Instrument Overview The waveform monitor uses a flexible tiled ee eae display The waveform monitor can display Ref Internal VITC 140410912 four tiles at one time Each tile can display erek pt Em a different measurement effectively creating four independent instruments To enable the tiles to function independently most of the controls affect only one tile at a time The tile that is currently being controlled is considered the active tile and it is indicated ee ee on the display by a light blue outline around the tile The waveform monitor can also display one si 5008 Dane tile at a time For example you might want Ref Internal VITC Missing O O to display just an Eye diagram to view the rise time measurement To select a display ae Psi QAO Mgr LOBUL ase Se EEEE EAE to be shown in full screen mode press the k geen ETE DISPLAY button and select the icon that represents the tile you want displayed in full screen mode Line F1 21 Active O Vi 0 0 mv Eye Type i weep Eye 10 Hz i Line Settings Quick Start User Manual T How to Operate Your Instrument The waveform monito
35. cord See page 1 Power Cords The typical power draw is 50 Watts for a base unit Refer to the Specifications and Performance Verification document for additional information on power and environmental requirements Connect the supplied power cord to the rear panel power connector There is no power switch on the waveform monitor so the instrument will turn on as soon as you apply power Installing the Monitor in a Video System The waveform monitor can operate almost anywhere in the distribution system The following diagrams for serial digital systems and for the analog composite inputs Quick Start User Manual To install for monitoring the video bit stream of a serial receiver 1 Route the incoming serial signal into one of the waveform monitor SDI inputs NOTE See the Specifications on the Documents CD for maximum allowed cable lengths Quick Start User Manual Installation Serial Source O Serial source O Serial output Serial output Waveform Monitor rear panel fo O 1890 014 SDIB SDIA Option Eye Phy Serial source Serial source Serial output Serial output Waveform Monitor rear panel o 0110 Bole Ce e O lt 00 CC SDIA SDIB Option SD HD Installation To install for monitoring composite signals Composite source 1 Connect your sources to the CM
36. display the Utilities menu 12 Touch System Upgrade This displays a dialog box that requires you to confirm that you want to upgrade the system The waveform monitor display goes to black and will display a screen stating that it is waiting for data transfers from the network Quick Start User Manual To Upgrade the Waveform Monitor Software 255 255 248 0 28 181 216 1 OV Gain x1 00 H Mag x1 00 Config Mode Network Instrument View Hw Sw SNMP Setup Manual Setup Name wRrewiio Stirs Version uw yiitu jpa Close View Hw Sw View Instmt l P Set Clock Display Levels Communications System Upgrade Utilities Version Options kuway we Wl jpa 129 To Upgrade the Waveform Monitor Software Run the Upgrade c C temp WF M6K7K_Firmware_upgrade transfer exe 13 On the PC double click the transfer exe H28 184 216 128 OY AASeSS OF target instrument file to launch the transfer program This displays the window shown at right 14 Type the network address for example 192 168 221 2 or DNS name of the waveform monitor you want to update and then press Enter This starts the software upgrade process 15 The transfer utility asks you to enter c C temp WF M6K7K_Firmware_upgrade transfer exe another IP address when the software Please enter DNS name or address of target instrument upgrade completes Opened TCP connect ion to 128 181 216 128 Reading Firmware Data 238
37. entry box like this SN ad itp f122 181 216 120 Jo http 123 123 123 123 NOTE Many Web browsers do not Tektronix WFM6100 WEM7100 Remote Interface correctly interpret IP addresses with leading Run the Remote Application zeros If the IP address shown in the Get a snapshot of Complete Instrument display BMP i i i i Get a snapshot of Graticles amp Waveform only BMP Configuration menu contains leading ene Get a snapshot of Java amp Waveform only BMP as in 124 161 038 151 remove the 0 when Get a snapshot of Waveform only BMP entering it into the address line of the browser Get a snapshot of ava onk BMP For example 124 161 038 151 should be Instrument Options entered as 124 161 38 151 Event Log Diagnostics Log Diagnostics Summary Log amp al Local intranet 124 Quick Start User Manual Do any of the following steps as desired 3 Click Run the Remote Application to start the Java applet see next procedure 4 Click one of the Get a snapshot links to download a bitmap image 5 Click Instrument Options to display which options the instrument is equipped with 6 Click Event Log to display the Error Log as an HTML table 7 Click Diagnostic Log to display the diagnostics log as an HTML table 8 Click Diagnostics Summary Log to display a brief summary of the diagnostics log as an HTML table Using the Remote Application Remote Communications Run the Remote Ap
38. is in Parade mode do not try to measure from one element to Ovi 0 0 mV OTi 0 00 us the next Stay within single component for OY Gain x1 00 H Mag xi 00 O T2 62 50 us accurate measurements A 700 0 mv 4 62 50 us Overlay Color Space Components Filter Sweep Settings Cursors YPbPr YPbPr Flat i Line Other Usage Tips If you use other functions such as Line Select while cursors are active the knob will be assigned to those other functions Touch one of the cursor readouts to transfer the knob control back to cursors You can display independent cursors in all four tiles at the same time although you will only be able to change the cursors in the active tile Cursors track the live trace so they may not be correctly registered on a captured trace Press the SELECT button to change the cursor assigned to the knob If both voltage and time cursors are displayed repeatedly pressing Select will cycle the knob assignment through all the displayed cursors If only one set of cursors is displayed pressing the Select button will cycle the knob assignment between cursor 1 and cursor 2 To Capture the Display Use the Capture function to capture the current image for traces waveform gamut displays and vector if selected Capture is useful for comparing sources or for capturing transient events The frozen image is shown in a different color to distinguish it from the live image Use Capture Trace Capture Buffer to freeze
39. monitored input that can show different safe area graticules and decoded closed caption services Proprietary Split Diamond and Arrowhead displays that help quickly detect and correct gamut problems Sophisticated audio level and phase displays including an exclusive display for surround sound audio which helps reduce time and effort in creating and verifying audio content A proprietary Timing display that graphically shows the timing relationship between the input signal and an external reference signal or a saved offset This easy to interpret display along with numeric readouts of vertical and horizontal timing relationships reduces time and effort in verifying and correcting signal timing User defined menus MyMenu give easy access to frequently used operations and simplify instrument operations Up to 20 instrument presets organized in four groups of five presets per group for easy access with user definable labels for quick identification A front panel USB 2 0 port for easily storing and recalling captured data and instrument presets A convenient front panel headphone port for quickly verifying sound quality Simple touch screen controls let you quickly check the stereo pairs shown on the audio bar display Network access and control through a simple Web page interface a more complete Java based remote control interface or by SNMP Ground closure remote control port with an alarm output and support for recalling instrument pr
40. pixels in error exceeds the Y Area setting Y Low threshold Shows the level below which the Luminance value is considered too small and out of gamut Signals below this level will trigger an error alarm if the number of pixels in error exceeds the Y Area setting OY High 103 0 y Low 1 0 o OY Area 0 Y Area Specifies the percentage of the total image pixels up to 10 that can be outside the current Luma limits yet not be reported as an error Default Linfits Y C High threshold Shows the level EBU 103 fuma a above which the Luma Chroma value is considered too large and out of gamut Signals above this level will trigger an error alarm if the number of pixels in error exceeds the Y C Area setting Y C Low threshold Shows the level below which the Luma Chroma value is considered too small and out of gamut Signals below this level will trigger an error alarm if the number of pixels in error exceeds the Y C Area setting Y C Area Specifies the percentage of the total image pixels up to 10 that can be outside the current gamut limits yet not be reported as an error Threshold indicators indicates the threshold settings using blue dashed lines Quick Start User Manual How to Operate Your Instrument Y C High 120 IRE D Y C Low 24 IRE Y C Area 0 1890 024 11 How to Operate Your Instrument lf the signal appears outside the boundary shown by th
41. recognized the flash drive may not be compatible try using another flash drive 5 Select the desired Capture from the list 6 Touch Restore to copy the file to the waveform monitor 7 When the operation is complete touch Exit to remove the dialog Quick Start User Manual 1890 013 5 Live i Capture Settings Ea Ea j Display On Display On Close Copy Buffer Trigger Types Settings To USB USB testcapturel cap Capture Buffer Restore Buffer From USB 45 How to Operate Your Instrument To Set Line Select Mode 46 To Toggle Line Select Mode 1 Select the tile containing the display for which you want to set Line Select Mode NOTE Line Select Mode can only be active on one tile at a time 2 Push LINE SEL to toggle Line Select Mode on or off When on the tile displays the selected line information only The Line selection menu appears at the bottom of the selected tile NOTE Ifthe LINE SEL button is lit but the Line Select menu is not visible pressing the LINE SEL button will display the menu it will not turn off Line Select mode 3 Press the appropriate soft key to select the appropriate field The selections available depend on input signal Standard and format 4 Turn the GENERAL knob to select the line you want to view NOTE If other Waveform display tiles are in Field Sweep or Picture mode the selected line will be highlighted with a cursor For Picture displays
42. set an address that is compatible with the setting of your PC Configure the SNMP setup if used See page 51 To Connect to a Network Ethernet Ethernet Lap Top PC connection connection A HUB ri IP address L JL 192 168 1 1 APP Waveform Monitor IP adress 192 168 1 2 1890 015 Ethernet connection The waveform monitor can print screens and event logs to printers connected by a USB cable 1 Connect your printer to the USB port on the front panel If you want to print a screen set up the monitor display as you want to print it No setup is required to print an event log 1890 013 Quick Start User Manual 3 Press the MAIN button to display the Main menu 4 Touch the Print soft key 5 Touch the Print What soft key and select what you want to print 6 Touch the Page Properties soft key and set the page properties as required 7 If you are printing an error log you may want to limit the number of pages printed error logs can be long Select the Page Limit soft key and chose a page limit 8 Touch Select Printer and chose which printer you will use See the following procedure Creating a Printer for instructions on how to add a printer to the list Quick Start User Manual How to Operate Your Instrument 1890 044 USB Configure Standby User Menu Close Page Properties Print What
43. temporally Since Cb and Cr have half the data rate of Y the samples appear twice as long Use the Select Channels soft key to turn off individual components of the display Data Mode SD Shows the data in the same sequence as it occurs in the serial domain You see a sample of Y then of Cb then of Y Y prime and then of Cr Then the sequence repeats The Y sample is the co sited sample and the Y sample is the isolated luma sample Data Mode HD Splits the serial data into a Y channel and a multiplexed Cb Cr channel The intent is to show the data in the same orientation as it occurs in the serial domain even though it is displayed as 8 or 10 bit values Traces In Video mode you can select which traces to view Readout Format Select the format of the data readout Line Word Touch either readout to use the large knob to select the line or word to be viewed Press the SELECT knob to toggle control between Line and Word ID Audio In Dolby 1 AES Al 2 Y Line F1 21 Active 3 Word 1439 Y Active Video Traces Readout Fmt Readout Fmt Readout Fmt YCbCr Hex Decimal Binary 1890 025 Different fields of the data stream are shown in different colors m Green Active video data m Gray Data in horizontal or vertical blanking intervals Quick Start User Manual How to Operate Your Instrument m White EAV and SAV packets other reserved values such as the XYZ word E Yellow
44. the selected measurement was displayed in the tile As the procedure below indicates Gain Horizontal Magnification and Sweep do not apply to all display types 28 To set Gain 1 Select the tile whose gain you want to adjust This will restore the menu for the tile if it is not currently displayed Push the MAG GAIN button If MAG GAIN is not available for the tile you have selected a message is displayed to let you know that MAG GAIN is not available To set a fixed gain touch the V Gain readout and use the large knob to change the gain You can also touch one of the fixed gain soft keys Tosetthe VAR iable Gain touch the Var Gain Enable soft key so it is checked Set the gain that you want using the large knob 525i 59 94 Mon Apr 24 16 09 34 ID SDI Input A Embd PPPP Ref Internal VITC 12 05 07 19 2 V Gain x1 00 _ H Mag x1 00 Default Vert Gain Fl Fixed Gain Fixed Gain Fixed Gain Fixed Gain Mag Gain Enable x1 00 x2 00 x5 00 x10 00 Quick Start User Manual To set Horizontal Magnification 1 Select the tile whose horizontal magnification you want to adjust This will restore the menu for the tile if it is not currently displayed Push the MAG GAIN button If MAG GAIN is not available for the tile you have selected a message is displayed to let you know that MAG GAIN is not available Touch the Vert Gain Horz Mag soft key to highlight Horz Mag To set the horizon
45. timing relationship between input and reference is also compatible with waveform mode That is if you have zero timing on the Timing display and change from internal to external reference the displayed waveform will not shift position m When timing analog composite signals first use the Timing display to get within one clock cycle then adjust system phase with the Vector display See page 63 Vector Display m If you have an input and reference combination that requires multiple timing indicator circles then it can be misleading to compare timing offsets between multiple inputs Because the timing display chooses the smallest of the possible timing offsets if a large timing difference exists between two inputs then they may not be matched This problem will also occur using traditional timing methods unless one uses something similar to the SMPTE318 10 field flag to identify a specific submultiple of the reference Quick Start User Manual 101 Checking Chroma Luma Delay Lightning Display Checking Chroma Luma Delay Lightning Display The Lightning display can be used for interchannel timing measurement If the color difference signal is not coincident with luma the transitions between the color dots will deviate from the center mark of a delay scale The amount deviated represents the relative signal delay between luma and color difference signal 1 Connect a signal containing color bar information 2 Press the IN OUT button to
46. to Rel to Save Rear Panel Saved Offset Offset Quick Start User Manual 73 How to Operate Your Instrument If you time input rates that are not 1080p 23 98 Mon May 15 15 17 03 integer multiples of the reference rates Ref NTSC EEA the instrument cannot measure the E ener eee timing deterministically so it displays the relationships as several circles Each circle represents a possible timing offset measurement relative to the reference cross hair Display emphasis is given the circle that times closest to zero offset and the numerical readouts track this pair Nondeterministic cases include timing of slow E rate inputs against fast frame rate references ee or when timing video against film rates Vertical Offset 45 lines delayed Horizontal Offset 7 792 us delayed Circle will be in the center when input Rel to Rel to Save Rear Panel Saved Offset Offset 74 Quick Start User Manual How to Operate Your Instrument Bowtie Display Pressing the MAIN button calls up the Main pop up menu From the Main pop up menu you can select the Bowtie display under the Measure heading The Bowtie display evaluates the relative amplitude and timing between the three video channels The Bowtie display requires a special test signal to be sent from the monitored source The Tektronix TG2000 Signal Generation Platform can produce an SD Bowtie test signal with 20 ns time marks which aid in signal evaluation The signal is a 50
47. view of video and audio digital data stream and ANC data extraction Description Adds support for Composite Analog Monitoring 2 Composite Analog Inputs Adds support for Standard Definition Monitoring 2 SD SDI Inputs Adds support for High Definition inputs 2 HD SDI Inputs Description Quick Start User Manual Video Options DS AD WFM6100 Video Options CPS SD Audio Options DS AD DD DDE Eye Jitter and Data Analysis Options EYE PHY DAT Before Installation Description Adds support for Digital Audio Monitoring in embedded and AES EBU formats Adds support for Analog Audio monitoring and Digital Audio monitoring embedded and AES EBU inputs Description Adds support for Composite Analog Monitoring 2 Composite Analog Inputs Adds support for Standard Definition Monitoring 2 SD SDI Inputs Description Adds support for Digital Audio Monitoring in embedded and AES EBU formats Adds support for Analog Audio monitoring and Digital Audio monitoring embedded and AES EBU inputs Adds support for Dolby Digital AC 3 decode Digital Audio embedded and AES EBU inputs and for Analog Audio Monitoring Adds support for Dolby E decode Dolby Digital AC 3 decode Digital Audio embedded and AES EBU inputs and for Analog Audio Monitoring Supports digital audio formats with up to eight channels Description Adds eye pattern and basic jitter measurement The Eye option enables the display of Eye
48. 0 Hz i Line 1890 026 83 How to Operate Your Instrument Jitter Display Pressing the MAIN button calls up the Main pop up menu From the Main pop up menu you can select the Jitter display under the Eye Phy heading The Jitter display presents a Unit Interval versus time waveform of the input serial signal This trace is composed of multiple samples of the SD or HD serial bit stream 84 To view the Jitter display you must have Option Phy installed The Jitter display can be set to display measurements and limits 1 Jitter gauge Augment the jitter readouts with an easy to read visual indicator Limits Touch any limit readout and use the knob to adjust the limit Select Show Limits from the Settings menu to display these values V Gain amp H Mag Adjust the vertical gain and horizontal magnification with these labels Eye and Jitter measurements Select Show Measure from the Settings menu to display these values Line When in Line Select mode use the knob to select which line is being viewed Line F1 21 Active Jitter Amp 2 0 20 UI Cable Len Max 200 m Cable Loss Max 2 0 dB 2 Launch Ampl Min 720 mV Launch Ampl Max 880 mV YY Gain x1 00 H Mag x1 00 Jitter HPF Sweep Settings Cursors 10 Hz 1 Line 1890 027 Quick Start User Manual How to Operate Your Instrument SDI Status Display The SDI Status display shows the current status of the SDI signal This display
49. 0 kHz sine wave on CH 1 luminance and 502 kHz sine waves on CH 2 Pb and CH 3 Pr For HD the TG2000 can produce an test signal with 1 ns markers The signal is a 2 5 MHz sine wave on CH 1 luminance and 2 502 MHz sine waves on CH 2 Pb and CH 3 Pr Elements of the Bowtie display res Tre Ip ieeaee 7 SDI Input B Hide menu Embd PPPP PPPP PPPP PPPP Ref Internal 1 V Gain The vertical gain Touch the readout and use the knob to adjust 2 H Gain Horizontal Gain Touch the readout and use the knob to adjust O V1 0 0 mV P H Mag x1 00 V2 700 0 mV A 700 0 mV 1890 028 Quick Start User Manual 75 How to Operate Your Instrument DataList Status Display Pressing the MAIN button calls up the Main pop up menu From the Main pop up menu you can select the DataList display under the Measure heading The Data Display mode shows the actual SD or HD input data without any interpolation The DataList display shows the signal data in two formats as a trace and as data The trace form is shown on the left portion of the display and the data form is show on the right portion of the display 76 Display Elements 1 Video Data The waveform monitor provides two modes for representing signal data in the trace portion Video and Data Video Mode Shows the data like the video YPbPr display in waveform mode but without interpolation The Y Cb and Cr traces are offset vertically to separate them but are aligned
50. 24875 bytes done CAUTION Do not close the transfer exe ATEN Pamp Aele a 242 L bytes done window until the transfer utility asks for Initializing Application Upgrade another IP address initislicing Filesystem Upgrade Erasing Flash done Programming Flash done Nerifying Flash Programming done done When the software upgrade completes the waveform monitor will reboot Please enter DNS name or address of target instrument CC 16 To update another waveform monitor begin at step 5 and skip step 12 If you won t be updating another waveform monitor press Enter to exit the transfer utility Verifying the Upgrade 130 After you have completed the software upgrade the instrument will reboot and come up in the touchscreen calibration mode Follow the instructions that appear on screen to complete the upgrade NOTE See the Readme txt file that is included with the software upgrade package to determine whether all of the current Configuration menu and instrument mode settings are preserved through upgrading Quick Start User Manual To Upgrade the Waveform Monitor Software To verify the software upgrade check the software version in the View HW SW Version display in the Config gt Utilities menu 1 Press the MAIN button S251 50 04 Thu May 04 15 49 24 D perp Ref fate 2 Touch the Config soft key SDI Status Jitter Vector Arrowhead Picture Bowtie Dolby Status Lightning 3 Touch the U
51. C monitors either CRT or LCD based The PIX MON connector is a 15 pin D type connector with socket contacts 5 1 Pin Pin name aa 1 Red Video ol sooo j O 2 Green Video 15 11 l 1890 005 3 Blue Video 4 Not connected 5 Ground 6 Red Ground T Green Ground 8 Blue Ground 9 Not Connected 10 Not Connected 11 Not Connected 12 Not Connected 13 Horizontal Sync 14 Vertical Sync 15 Not Connected Remote Connector Pin Assignment The REMOTE connector interface uses ground closures for remote control and indicating to external equipment when alarms have occurred The input of LTC is through the REMOTE connector The REMOTE connector is a 15 pin D type connector with socket contacts 5 1 Pin Pin name of R O 1 Ground 5 s 2 Time Code Positive IN 1890 004 3 Time Code Negative IN 4 Ground 5 Ground Close Output Has weak pull up with a diode to 5 V so you can get logic swing without pull up OUT 6 Preset Recall 1 Grounding this pin is will select the associated preset IN 7 Preset Recall 2 Grounding this pin is will select the associated preset IN 8 Preset Recall 3 Grounding this pin is will select the associated preset IN 9 Preset Recall 4 Grounding this pin is will select the associated preset IN Quick Start User Manual 23 How to Operate Your Instrument 24 Ethernet Connector The waveform monitor provides a I 10 100 BaseT Ethernet interface The Ethernet connector is a standard RJ
52. DIA SDIB SDI OUT 100 240VAC 10 50 60 Hz 140 WATTS MAXIMUM 1890 042 Options SD HD O TeS AIN1 2 B101 2A AES 100 240VAC 10 50 60 Hz 140 WATTS MAXIMUM AIN56 B 1 0 5 6 AIN78 B1 07 8 CLOCK OUT Ee rea EIO Q 1 r EXT DISPLAY A ETHERNET SDI B SDI OUT 1890 041 Options Eye Phy 16 Quick Start User Manual Power Requirements m A single phase power source with one current carrying conductor at or near earth ground the neutral conductor m The power source frequency must be 50 or 60 Hz and a operating voltage range must be from 100 to 240 VAC continuous m Systems with both current carrying conductors live with respect to ground such as phase to phase in multiphase systems are not recommended as power sources 1890 012 NOTE Only the line conductor is fused for over current protection The fuse is internal Video Input Connectors SDI inputs are active inputs Reference and Composite inputs are passive inputs compensated for 75 Q 1 Ref Loop A synchronization input The input signal can be analog black burst analog composite video or analog L_perioop tri level for HD 1890 010 Quick Start User Manual How to Operate Your Instrument How to Operate Your Instrument 18 2 SDI A The Digital A component serial digital input 3 SDI B The Digital B component serial digital input 4 SDI Out Digital ve
53. E gt Give people information they can truly use See how Application Solutions Serial Data Defense Electronics Design amp Analysis Resources Careers International myTek Product Resources Manuals Technical Information Application Notes Technical Briefs Primers Use and Learn Glossary of Video Terms and Acronyms Classroom amp Lab Resources Online Tutorials Product Support Ask the Tektronix Video Experts My Product Support Contact Information File Edit View Favorites 3 Q sex amp P Search jy Folders 4 Tools Help Trusted sites a 3 Links v EJ co Address B C temp WFM6K7K_Firmware_upgrade Name Size Type File and Folder Tasks Y firmware pkg 43 465KB PKG File E transfer exe 49KB Application Other Places y B ReadMe txt 1KB Text Document v lt gt 3 objects 47 3 MB ig My Computer Quick Start User Manual Check and Record the IP Address 4 Press CONFIG to display the Configuration menu 5 Select Utilities to display the Utilities menu 6 Select Communications to display the Communications menu 7 Select Network Setup to display the network settings 8 Write down the number listed for IP Address in the menu Enter System Upgrade Mode 9 On the waveform monitor press the MAIN button to display the Main menu 10 Touch Config to display the Configuration menu 11 Touch Utilities to
54. E ERA RAS 1 2 8 17 I ACCES OEE a E E E EE ern Pe 1 OD IONS oaa A E E EE NAE AOE E ES E T E E EE E EAEE EAT 2 ae Eo a EE E Greece E T EE sees ene og E E N EN EEEE A E E A AT 4 Connecting Power and Powering On OfF nnueeessnannnnnueennrrrrnnnrrrsrrrrrrrrrrrrrrrrrrrrrrerrrrrrrrrrrrrrrrrrrre 4 installing the Monitor in a Video System a5 cc cosssinndvrcncanieehalinnseniadeubaacehsie A 4 How to Operate Your Instrument essnnneesnnnuuenrrnunrrrrnurrrrrrrrrrrrrrr rrr rrr err rrr renr rrn EEE HESS SEES EE EDGES EEE EEE EES T VENIO N aarre A E EE E E A EE E AA 7 Front Panel Controls and COnnectors ccc cece cece cence een e eee en nee ened E HEED EEE EEE D EEE E EE SDE EEE EERE EERE EEE EEE ES 11 ETERS IS Orn D seca E E E E wens eres EE AEE EEA EEEE 13 ROG eee i ReC ni 0 ci a ee ee 16 To Select a Measurement rosa Goracca4 nacuete bain laiansatsniandaonstrtonedaaeataianadeneasaielenierieadorridasera 25 To Set Display and Measurement Parameters c ccc cece cence cence eee ee ee ee eee eaeeeeeee esate senses eaeteneanenegs 26 TO Selec AMONO Video NAG UNG iipcca scene duierces wna E E be acarn aves E ES 26 To Set Gain Horizontal Magnification and SWEE 00 cece eee e cece eee tee e eee ee eee eee e nates eens ete een aE EE EEE EES 28 NOS PC SOS scr e rc ice se oe victoire Aedes oe OE EE NE E ree ere ee heed EEEN 30 To Measure Waveforms with Cursors cc ccc ce eee e nee e eee ene eee enn EEE ee EEE EE EE
55. Embd PPPP Page l of 3 Status Additional Information Detected 525i 59 94 525i 59 94 SDI Input A Ref Internal 1B 1B Field Status Format 000 0l6 048 064 Wed May 10 16 42 14 ide veru 2 ARIB TR B 22 Display SDID Line Checksum 080 096 1lI2 128 ID Embd PPPP DC Stream Should be 160 176 192 208 224 240 ARIB ARIB ARIB ARIB ARIB ARIB B 39 B 37 B 35 TR B 23 1 TR B 23 2 TR B 22 Quick Start User Manual How to Operate Your Instrument Auxiliary Data Status Displays detailed 1080 50 08 gt information about the auxiliary data Ref Internal embedded in the video signal Auxiliary Data Status Tue May 02 09 27 47 Closed Caption Type m Closed Caption Type Displays the format and transport service of the detected closed caption data None Detected is displayed when there is no closed caption data present V Chip Rating CC Services Ancillary Data Present Transmission Signal Identifier V Chip Rating Displays V Chip rating None is displayed when there is no V Chip rating data present m CC Services When closed caption data is detected the closed caption service s being used are listed Ancillary Data Present Present is displayed when ancillary data is detected None is displayed when there is no ancillary data present Transmission Signal Identifier Indicates whether the TSID is present in the
56. HD Line Num Error Video Format Error l Luma Gamut Error L Cmpst Gamut Error RGB Gamut Error r SD AP CRC Error SD FF CRC Error HDM CRE Error r HD C CRC Error r HD Y ANC CSUM Error r HD C ANC CSUM Error Buffer i Copy Buffer Trigger Types Settings To USB 1 wispy wit wiopimy wt Live Capture Capture Settings oe P Clear p Display On Display On Stop Buffer Live Capture Capture ettings ear Setting p c p Display On Display On Buffer 525i 59 94 Wed Apr 26 14 41 45 SDI Input A Ref Internal Hide Menu OV Gain x1 00 H Mag x1 00 Live Capture Settings j Clear Display On Display On 43 How to Operate Your Instrument 44 Copying the Capture Buffer to a USB Flash Drive 1 Insert a flash drive into the USB port on the front panel Touch the CAPTURE button to display the Capture menu Touch the Settings soft key to display the submenu Touch the Copy Buffer To USB soft key This displays the Copy Capture Buffer dialog window Verify that the USB flash drive is recognized a description of the drive will appear after USB Storage Device If it is not recognized touch the Reset soft key to mount the flash drive lf the flash drive is still not recognized the flash drive may not be compatible try using another flash drive Touch the Keyboard soft key to display a keyboard you can use to enter a name for the file Enter a name for the file and touch R
57. Input Dolby Specific Composite Input Closed Captions Metadata General ARIB Specific General Audio Physical Layer Quick Start User Manual 111 Checking Gamut 5 Select each alarm that you want to be checked 6 Touch Save and Close to save your configuration 7 Check that Enable Alarms is set to On in the Alarms menu before leaving the Alarms menu NOTE For information on Alarms press the icon while the CONFIG menu is displayed and then select Alarm Setup 8 Touch Exit Alarm Setup to exit the Alarm Setup menu Adjusting Gamut Limits 1 Press the MAIN button to display the Main menu 2 Select the Gamut display whose limits you want to set either Diamond or Arrowhead You can also reset thresholds to defaults 112 Save and Cancel and Close Close VIDEO CONTENT Txt Icon Log Beep SNMP GC RGB Gamut E E E Composite Gamut O L O Luma Gamut BS O O Exit Enable Alarm Setup Set All Alarms to This Mask Alarms Video Content AES and Embedded Video Format Embedded Audio Specific SDI Input Dolby Specific Composite Input Closed Captions Metadata General ARIB Specific General Audio Physical Layer 525i 59 94 Fri May 05 14 59 21 ID SDI Input A Hide menu Embd PPPP Ref Internal Z Waveform Diamond DataList Alarm Status Aux Status Line F1 21 Active Picture Timing Audio Session SDI Status Dolby Status Lightning Configure Standby Help User Menu Quick Start
58. Jitter Picture Audio Session SDI Status Bowtie Dolby Status Lightning USB Configure Print Config Standby Help Mount User Menu 4 Touch Save Offset from the Timing 1050588 SETTER men U Ref Internal Hide Menu 5 Touch the Rel to Saved Offset soft Input Timing Relative to Reference key 1 The timing of the first input the master signal is at the center of the Vertical Offset screen Horizontal Offset Relative to Rear Panel Circle will be in the center when input Rel to Rel to Save Rear Panel Saved Offset Offset 100 Quick Start User Manual Timing a Studio 6 Remove the master input signal from the 1080 50 08 See aa o waveform monitor and in its place apply Ref Internal Hide menu another input to the router Input Timing Relative to Reference 7 Adjust the sync source or studio timing of that signal so that it matches the saved Vertical Offset timing When the timing matches the circle and cross in the display will both be at the center of the Timing display Horizontal Offset Relative to Rear Panel Circle will be in the center when input Rel to Rel to Save Rear Panel Saved Offset Offset 1 1 You cannot save the timing offset if either the input or reference is missing or unlocked You also cannot save a reference when in internal mode Saving an offset in these conditions would lead to misleading results so it is not allowed by the instrument Usage Notes m This
59. Lightning Display Pressing the MAIN button calls up the Main pop up menu From the Main pop up menu you can select the Lightning display under the Measure heading The Lightning display is a measurement method that allows for the evaluation of the luma signal gain and for making chroma luma gain comparisons It can also provide simple indication of inter channel timing errors indicated by a bowing in the trace between the green magenta transition The display is created by plotting luminance versus B Y in the upper half of the display and inverted luminance versus R Y in the lower half of the display The bright dot in the center of the screen is the luminance blanking level The points above and below this show the plots of the different color components based on their signal amplitude The Lightning display requires a color bar test signal be used However since most programs have leaders of color bars for setup you can use the Lightning display to check a program during the leader section Display Elements 1 V Gain The vertical gain 525i 59 94 Tue May 09 12 46 32 ID E tide menu fila 2 ia r ara 2 H Gain Horizontal Gain a Line All 158 Active 3 Line The line selected for display Appears only in Line Select mode NOTE To change the settings for either Vertical or Horizontal gain touch the gain readouts and then adjust their values using the large knob y Gain x1 00 H Gain x1 00 Lightning Menu
60. OWUG DEDE Y enda duc ele dit dul EE temo Matawan aaue malar A EE 15 Datalist Status DIS CIAY oranana cites Meir tel eeheye aaa bela AE ae outa betel a Seeded Dalen ewe hes 76 ANG Data DIS Play osioiden tava naa aaa A S seeds ae tua nal ithaca Shaan 77 Staus DISIN are an E E ee Re A AE A S 78 VG IG DIAN a aware Maier ratuh eiteaw arena doneuuaeenntae a a irene auernn E E 82 JINEN DISDA ipsa ckaneiiens suas eh ease a scalar nine adh luadaauish olansei a a amelamaanededinduareaaseotoutou nares 84 SDI Status DISDIAY nctd ta onic cast A at eatecnatl ca tasd aes EAA ianeud A AA 85 Monitoring the SDI Physical Lave s2 csac cuciesnas ieaaucaatecanudeadanao aaueddan sees nash taneekaeetemeanataaea nleemacns tazedaanas 86 To Configure the Physical Layer Settings cccccccce cence eee e eee ee seen ee eee eee eee ee esas tee eateeeaeteeaateeeaaengas 86 Oakes EVE MCASUICMENM Satie tuavnntnn tae dicavnaSetaede dee r tanning e RE aie nianeeamedamde leds 92 TNO Ay SCO 4s accor tet qondes a tenon a a m aa tea a a tat cataas tensa a a a seeders 97 Using the Timing Display Method 0 02490 ace ssicaheew arbi iieuntian ubaaniiandelndiseeeentatxancddeemiaresi hoeaniernana 97 Timing Multiple Inputs 1O S ROUTEL c u c 0cdneetnt ds cde al desaenenenahedaduuenenniment eile dtiwmen ald aeasmennbadadudrmmnonee 99 Checking Chroma Luma Delay Lightning Display ccccceccceeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeesaeeeeeeeeenneneas 102 GH fetes N GaN Sener onic amr sear t
61. Operate Your Instrument To Determine Input Signal Status At A Glance The status bar shows the status of the instrument and of the monitored signal The status bar appears in almost every display The elements of the status bar are detailed below 525i 59 94 SDI Input A Ref Internal Element 1 10 11 12 Quick Start User Manual Display element Video Format Video Input Video Reference EDH CRC Error RGB Gamut Error Composite Gamut Error Luma Gamut Errors Alarm Error Indicator Show Menu Hide Menu Online Help Enable Freeze Status Timecode readout Audio Channel Status ID Date and Time Wed May 10 16 29 52 ID Embd PPPP 1890 055 Indicates the format of the signal on the selected video input or whether the signal is missing or unlocked Description Text indicating the selected input Possible inputs are SDI A SDI B Cmpst Input A Cmpst Input B depending on model and installed options Indicates if the current input is not in Auto mode by displaying the format in yellow text Text indicating the source of the current reference Also indicates format and whether the reference is missing or unlocked A one line area that is visible if EDH errors are present SD inputs or CRC errors are present HD A one line area that is visible if RGB Gamut errors are present A one line area that is visible if Composite Gamut Arrowhead errors are present
62. PST A or CMPST B loop through inputs on the rear panel Waveform Monitor rear panel Composite output o OO O Q O Composite video loop OOJO receiver or termination cp O a APIO O 1890 038 CMPST A LOOP Option CPS Line Termination The waveform monitor uses passive loop through analog and reference inputs Accordingly the loop through inputs must be terminated externally It is important that this external termination meet accuracy and return loss requirements lf the waveform monitor is installed to monitor an operating link the destination receiver and the connecting cable serve as the termination This monitoring connection is best because it checks the performance of the entire path The return loss of the waveform monitor is sufficiently high that in most cases the destination receiver sets the system return loss In cases where the waveform monitor is placed at the end of a link a BNC termination must be installed on one side of the loop through analog or reference connector The termination must be 75 and DC coupled good return loss extends to DC An appropriate termination would be Tektronix part number 011 0102 00 It is a 75 Q End of Line termination Compatibility of BNC Center Pins Most video equipment BNC connectors whether 50 or 75 Q use a 50 Q standard center pin Some laboratory 75 Q BNC connectors use a smaller diameter
63. Patterns in 3 Eye 10 Eye SD or 20 Eye HD modes cable length measurements including source signal level and cable loss and Jitter readout Includes the capabilities of Eye option plus advanced physical layer measurements jitter waveform and automated eye measurements Adds data analysis capabilities Allows for logic level view of video and audio digital data stream and ANC data extraction There are three post sale upgrade kits available for installation in WFM6100 WFM7000 and WFM7100 Waveform Monitors that you already own Some options are customer installable and others can be installed only at a Tektronix Service Facility The following table lists the options that are available when ordering an instrument and specifies the models for which the option can be ordered Option WFM61UP WFM70UP WFM71UP Quick Start User Manual Description Upgrade kit for WFM6100 Upgrade by ordering WFM61UP with appropriate options CPS SD DS AD DD DDE EYE PHY DAT Upgrade kit for WFM7000 Upgrade by ordering WFM70UP with appropriate options CPS SD HD DS AD Upgrade kit for WFM7100 Upgrade by ordering WFM71UP with appropriate options CPS SD HD DS AD DD DDE EYE PHY DAT Installation Installation A The waveform monitor is shipped in a wrap around chassis that covers the instrument bottom and two sides A cover is installed on the chassis and the rear panel is made up of the module rear panels You can op
64. Pin assignments and pin names are listed in the following figure and table CAUTION Use care when connecting the Analog Audio Output Refer to the instrument Specifications to ensure that the Audio Load and Output Power meet specifications Exceeding Analog Audio Output Power may result in damage to the instrument 62 43 esos Option AD Pin Description 1 ANALOG_INPUT_A1_P Balanced differential analog audio input Ch 1 line A positive 2 ANALOG_INPUT_B1_P Balanced differential analog audio input Ch 1 line B positive 3 ANALOG_INPUT_A2_P Balanced differential analog audio input Ch 2 line A positive 4 ANALOG_INPUT_B2_P Balanced differential analog audio input Ch 2 line B positive 5 ANALOG_INPUT_A3_P Balanced differential analog audio input Ch 3 line A positive 6 ANALOG_INPUT_B3_P Balanced differential analog audio input Ch 3 line B positive 7 ANALOG_INPUT_A4 P Balanced differential analog audio input Ch 4 line A positive 8 ANALOG_INPUT_B4 P Balanced differential analog audio input Ch 4 line B positive 9 ANALOG_INPUT_A5 P Balanced differential analog audio input Ch 5 line A positive 20 Quick Start User Manual 10 ANALOG_INPUT_B5_P 11 ANALOG_INPUT_A6_P 12 ANALOG_INPUT_B6_P 13 GND 14 ANALOG_OUTPUT_1_P 15 ANALOG_OUTPUT 2 P 16 ANALOG_OUTPUT_ 3 _P 17 ANALOG_OUTPUT 4 P 18 ANALOG_OUTPUT_5_P 19 ANALOG_OUTPUT 6 _P 20 ANALOG_OUTPUT_7_P 21 ANALOG_OUTPUT 8 P 22 ANALOG_INPUT_A1_N 23 ANALOG_INPUT_B1_N 24 ANALOG
65. The Lightning menu allows you to set the bar targets to either 75 or 100 Quick Start User Manual 69 How to Operate Your Instrument Diamond Display Pressing the MAIN button calls up the Main pop up menu From the Main pop up menu you can select the Diamond display under the Gamut heading The Diamond display shows Gamut violations of the SDI input if translated to RGB color space Display Elements 1 High threshold Shows the level above 4 which RGB components are considered spi fap it A arnal too large and out of gamut Signals above this level will trigger an error alarm if the number of pixels in error exceeds the Area setting High 721 mv Low 21 mV 2 Low threshold Shows the level below DO Area 0 which RGB components are considered too small and out of gamut Signals below this level will trigger an error alarm if the number of pixels in error exceeds the Area setting 3 Area Specifies the percentage of the total image pixels up to 10 that can be outside the current gamut limits yet not be reported as a gamut error Er Default Limits CL a 4 Threshold indicators indicates the ek RGB EBU 103 RGB threshold settings using blue dashed 1890 023 lines NOTE To change the threshold settings touch the threshold readouts and then adjust their values using the large knob If the signal appears outside the boundary shown by the blue dashed lines then the signal is violating the li
66. The bar target setting either 75 or 100 Elements of the Vector menu for Composite input signals m Normal SCH Phase Selecting SCH Phase adds one or more radial traces to the vector display that correspond to the phase of a subcarrier that is synchronous with the horizontal sync pulse m Phase Normal Phase V Present only on Vector displays with PAL input format The Phase Normal mode shows both phases of the vector that results from the phase alternating on lines PAL The Phase mode inverts the phase on every other line to undo the PAL sequence and thus shows all lines in one phase orientation Bar Target Setting The bar target setting either 75 or 100 Picture Display Pressing the PICTURE button calls up the Picture display which lets you see the picture generated by the video signal You can choose to display the picture with or without a frame Display Characteristics 10801 50 94 Mon May 01 172019 SDI Input B Embd PPPP PPPP Ref Internal m In full screen mode shown there is no cropping E In Active picture mode pictures are decimated horizontally or vertically to achieve the correct aspect ratio This decimation may cause some artifacts This behavior may be evident on a sweep or convergence signal m Also see the online help for the Active Picture Full Frame for related information Settinas CC 608 Service Safe g None Area 64 Quick Start User Manual How to Operate Your Instrument
67. User Manual Checking Gamut 3 With either Diamond or Arrowhead 52515994 Fri May 05 15 01 00 displayed touch the readout in the Ref Internal display for the limit you want to adjust 4 Use the large knob to adjust the limit High 721 mV value O Low 21 mV O Area 0 Default Limits Default Limits Tek RGB EBU 103 RGB Quick Start User Manual 113 Selecting and Monitoring Audio Selecting and Monitoring Audio The waveform monitor provides several methods to monitor audio signals You can measure levels monitor phase and display phase correlation You can specify meter ballistics and scales set the Test and Peak Program indicator levels and specify how phase is displayed NOTE With audio option DA the waveform monitor can monitor AES analog and Embedded audio Option DG can only monitor AES and embedded audio selecting the Audio Input You select the audio input to be monitored from the front panel AUDIO button pop up menu NOTE The Audio display can appear in only one tile at a time To select the audio input to monitor 1 Select a tile 2 Press the AUDIO button to open the Audio display 114 Quick Start User Manual 3 Select Audio Input and then select one of the audio input options shown NOTE The available selections depend on the installed audio option Selecting Follows Video selects the Video to audio mapping set in the Video to Audio Map submenu Quick Start Use
68. WFM6100 WFM7000 WFM7100 Waveform Monitors Quick Start User Manual www tektronix com Tektronix 071 1890 01 S Copyright Tektronix All rights reserved Licensed software products are owned by Tektronix or its subsidiaries or suppliers and are protected by national copyright laws and international treaty provisions Tektronix products are covered by U S and foreign patents issued and pending Information in this publication supersedes that in all previously published material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of Tektronix Inc Manufactured under license from Dolby Laboratories Dolby Pro Logic and the double D symbol are trademarks of Dolby Laboratories Contacting Tektronix Tektronix Inc 14200 SW Karl Braun Drive P O Box 500 Beaverton OR 97077 USA For product information sales service and technical support n North America call 1 800 833 9200 Worldwide visit www tektronix com to find contacts in your area Warranty 2 Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one 1 year from the date of shipment If any such product proves defective during this warranty period Tektronix at its option either will repair the defective product without charge for parts and labor or will provide a replacement in exchange for the defective product Parts modules and replacement products used b
69. Waveform Diamond Alarm Status Video Session Eye errors Zo Vector ANC Display ARIB Error Log Jitter m Arrowhead use to detect and correct composite gamut errors without g p DI A Data Collect Running r R Time l d 01 15 21 employing a composite encoder Audio Bowtie Dolby Status ee gia eee uto 525i 59 94 one SAY Place Err Ancillary Data None e e Field Length Err Tektronix Lightning Line Length Err Stuck Bits Fl AP CRC 589Dh F2 AP CRC 589Dh Statistics Status Err Secs Err Fields Err Fields RGB Gamut Error 68 3913 0 Cmpst Gamut Error 1559 92966 95 Luma Gamut Error 66 3532 3 4 FF CRC Error 0 0 0 0000 AP CRC Error 0 0 0000 Configure Standby Help User Menu 1890 047 Picture Timing Aux Status Audio Session SDI Status 106 Quick Start User Manual Example of an Arrowhead display Checking RGB Gamut The Diamond display intuitively shows how the R G and B signal video signals relate making it a good tool for detecting gamut errors The waveform monitor converts the Y P and P components recovered from the serial signal to R G and B to form the Diamond display To predictably display all three components on a monitor they must lie between peak white 700 mV and black 0 V For a signal to be in gamut all signal vectors must lie within the G B and G R diamonds Conversely if a signal vector extends outside the diamond it is out of gamut The direction of an excursion out of gamut indi
70. _INPUT_A2_N 25 ANALOG_INPUT_B2_N 26 ANALOG_INPUT_A3_N 27 ANALOG_INPUT_B3_N 28 ANALOG_INPUT_A4 N 29 ANALOG_INPUT_B4_N 30 ANALOG_INPUT_AS_N 31 ANALOG_INPUT_BS_N 32 ANALOG_INPUT_A6_N 33 ANALOG_INPUT_B6_N 34 GND 35 ANALOG _OUTPUT_1_N 36 ANALOG_OUTPUT_2_N 37 ANALOG_OUTPUT_3_N 38 ANALOG_OUTPUT_4_N 39 ANALOG_OUTPUT_5_N 40 ANALOG_OUTPUT_6_N 41 ANALOG_OUTPUT_7_N 42 ANALOG_OUTPUT_8 N 43 62 Quick Start User Manual How to Operate Your Instrument Balanced differential analog audio input Ch 5 line B positive Balanced differential analog audio input Ch 6 line A positive Balanced differential analog audio input Ch 6 line B positive Balanced differential analog audio output Balanced differential analog audio output Balanced differential analog audio output Balanced differential analog audio output Balanced differential analog audio output Balanced differential analog audio output Balanced differential analog audio output Balanced differential analog audio output Ch Ch Ch Ch Ch Ch Ch Ch Balanced differential analog audio input Ch Balanced differential analog audio input Ch Balanced differential analog audio input Ch Balanced differential analog audio input Ch Balanced differential analog audio input Ch Balanced differential analog audio input Ch Balanced differential analog audio input Balanced differential analog audio input Balanced differential analog audio input Ba
71. a display you want to keep or show to someone else In this application setting the Capture Display On setting and disabling the Live Display On setting might be best For comparing sources the enabling both the Live Display On and Capture Display On modes is useful You can capture waveforms two ways as a trace and as a buffer Buffer and Trace modes are different in the following ways Trace mode takes a snapshot of the display so it cannot change size or mode after the freeze m In Buffer mode a frame of the input signal is captured This allows it to be redisplayed as the display mode and size is changed For example you can perform a buffer capture in waveform mode and then change to vector to see the same data in the new display Buffer mode also resizes the capture image as you go from four tile to full screen mode m Trace mode capture works in both tile and full screen modes but frozen images do not change from tile to full or visa versa The frozen image is still available if you go back to tile mode 40 Quick Start User Manual How to Operate Your Instrument Since the Buffer method captures a whole frame of the input it can be redisplayed for different display types Vector Diamond or Arrowhead Buffer mode supports a triggering capability which initiates a capture when specific errors occur Buffer mode does not work on physical layer displays such as Eye and Jitter nor can it be used with composite signals Trace mode works w
72. a picture AUDIO Changes the selected tile to display the Audio level and status screen STATUS Displays the last Status display shown in the active tile and displays a menu allowing you to change the selected tile to display a status screen a session display or the error log MAIN Displays the Main menu and Main pop up menu providing access to primary and secondary menus Some displays are selectable only in the pop up menu USER Displays a user defined menu CAPTURE Displays menu that enables screen capture of live display MEASURE Displays the last Measure display shown in the active tile and displays a menu allowing display and configuration of the Eye Jitter DataList ANC Display Timing Bowtie and Lightning displays PRESETS Displays menu enabling save and recall of instrument settings MAG GAIN Displays the Mag Gain menu with controls for the Vertical Magnification and Horizontal Gain LINE SEL Enables disables Line Select mode displays Line Select menu if Line Select is enabled but the Line Select menu is not visible 12 Quick Start User Manual How to Operate Your Instrument IN OUT Displays a menu for configuring inputs and outputs Large Knob Used to select or adjust values for example the cursor position VERT knob Used to adjust the vertical position of the signal in the display HORIZ knob Used to adjust the horizontal position of the signal in the display Front Panel Connectors Connector Function
73. ams to find the centers of the distributions of 20 and 80 crossings The difference between manual and automatic measurements is usually insignificant for clean symmetrical signals To Manually Measure Eye Amplitude 1 Perform the initial setup for Eye measurements See page 92 To Set Up for Eye Measurements 2 Touch the Cursors soft key The Close eee instrument displays the cursors ape Show Volt spo a o To Center Submenu 3 Select the Show Volt soft key to enable display of the voltage cursors Voltage cursor V1 is selected by default 92 Quick Start User Manual 4 Use the general purpose knob to position cursor V1 at the top part of the waveform ignoring any overshoot or undershoot on the rising or falling edges 5 Press the SELECT front panel button to select voltage cursor V2 6 Use the general purpose knob to position the second voltage cursor at the bottom of the waveform ignoring any overshoot or undershoot on the rising or falling edges 7 The cursor Delta readout displays the amplitude of the eye waveform NOTE Signal source amplitudes outside of 10 of 800 mV p p can degrade receiver performance To Manually Measure Aberrations 1 Perform the initial setup for Eye measurements See page 92 Jo Set Up for Eye Measurements 2 Touch the Cursors soft key The instrument displays the cursors submenu 3 Select the Show Volt soft key to enable display of the voltage cursors Voltage
74. aracteristics of 64 Display elements of 61 Filter 62 Filter Flat 62 Filter Low Pass 62 Filter Luma 62 menu 62 Overlay 62 Parade 62 web browser accessing remote interface 124 WFM button 12 61 X X Y Phase Style 117 XGA Output connector 22 Y YPbPr Color Space submenu 62 YRGB Color Space submenu 62 Quick Start User Manual
75. as SDI Status You can use the Main menu to perform the following tasks Selecting one of the available display modes to view in the active tile m Printing the Error Log or the on screen display Configuring instrument settings Configuring the custom User menu m Accessing the Shutdown and Quiet modes from the Standby menu m Mount and unmount USB flash drives m Display the Main Help page Quick Start User Manual How to Operate Your Instrument Front Panel Controls and Connectors The waveform monitor can display either one or four views tiles at a time Each view can display a different measurement or display type effectively creating four independent instruments To enable the tiles to function independently most of the controls only affect one tile at a time FlexVu Configuring the Display The FlexVu feature allows you to view either one or four different displays tiles of the currently selected input signal Listed below are two examples of using the four tile display mode to monitor signals 1 To switch between 4 Views and 1 View De e e e display mode press the DISPLAY a ee button This displays the Views menu at the bottom of the screen Each tile is represented by a icon soft key in the menu 2 To display a tile full screen touch the icon that represents the tile you want displayed in full screen mode In a full display the displayed tile is always selected 3 To return
76. ata Rate Dolby Sample Rate Bitstream Mode Dialogue Level Dolby Surrnd Mode Mixing Level Room Type SMPTE Timecode Wed Apr 05 08 52 01 Dolby Audio Status Copyright Bit Original Bitstream Wed Apr 05 08 56 53 Dolby Audio Status Copyright Bit Original Bitstream Quick Start User Manual ID Audio In Dolby 2 Emb 1 amp 2 Stream Auto ID Audio In Dolby 2 Emb 1 amp 2 Stream Auto How to Operate Your Instrument 1080i 59 94 SDI Input B Ref Internal Video Session Shows immediate status in the upper portion of the display and shows Statistical overview of the entire session on the lower portion of the display Statistics are controlled by the Run Stop and Reset soft keys Tue May 02 09 31 38 ID Embd PPPP PPPP Video Session SDI B Data Collect Run Time Input Signal Format 352M Payload SAV Place Err Field Length Err Line Length Err Line Number Err Running 0 d 00 05 08 Auto 1080i 59 94 None Y and C Present Ancillary Data Y Stuck Bits C Stuck Bits Statistics Status Err Secs Err Fields Err Fields RGB Gamut Error 0 0 Cmpst Gamut Error 309 18443 0 0000 100 0000 Error Log Shows error over time For errors to appear Alarms must be enabled for logging in the Main gt Config gt Alarm Setup menu The soft keys in the menu are used to navigate through the log The error log is limited to 10 000 entries Q
77. aveform Monitor Software Installing the Software 128 To install new software in a waveform monitor Download the software package 1 To download the latest version of the waveform monitor software navigate to the Tektronix Web site at http www tek com products video_test From there follow the link Software and Firmware under Product Resources to locate the waveform monitor software upgrade package and download it Click on the downloaded self extracting archive to extract the following files into a directory of your choice Files extracted are transfer exe firmware pkg and readme txt Tektronix Products gt Video Test Home Microsoft Internet Explorer File Edit View Favorites Tools Help Q x amp x a E P search E DER Address g http www tek com products video_test Tektronix Enabling Innovation Home Products Support Buy Contact Us Investors Tektronix Products Video Test Home Search Video Test Home gt Find by Product Category Test amp Measurement Instruments Oscilloscopes Video Audio Signal Monitors MPEG Analyzers Real Time Spectrum Analyzers MPEG Generators MPEG Monitors Content Verification Find by Industry Content Creation and Post Production Content Delivery R amp D and Manufacturing Communications Network Management amp Diagnostics The Tektronix Video Studio See and Solve
78. cates which signal is excessive Errors in green amplitude affect both diamonds equally blue amplitude errors affect only the top diamond and red errors affect only the bottom diamond Quick Start User Manual Checking Gamut Wed May 03 19 33 30 Y C High 120 IRE OY High 103 0 OY C Low 24 IRE r 09 OY Low 1 0 Y C Area 0 OY Area 0 2 Default Limits EBU 103 Luma X Y plot Diamond plot 700 mV Legal G and B space Green 0V Blue 700 mV 0V Red 700 mV Green Legal G and R space 700 mV 1890 030 107 Checking Gamut 108 To use the Diamond display to check RGB 525i 50 94 Wed May 03 19 46 41 Gamut Ref RERE 1 Perform the Setting up for Gamut Check procedure selecting Diamond in step 6 O High 721 mV Low 21 mV 2 Compare the signal to the display to determine out of gamut components noting the following Area 0 m The intensity of a vector indicates its duration A momentary out of gamut condition appears as a faint trace Long duration violations show as a bright trace See the following step for some sample out of gamut signals 3 Consider the following examples when evaluating the signal for out of gamut components G Example A A gt R OK pan G gt 700 mV Ge B OK i Example B R OK G OK B gt 700 mV Example C R OK G OK 350 mV B lt 0 mV 1890 031 Usage Note
79. display that shows audio levels weighted or un weighted total sound volume phantom source locations and the dominant sound position To add a phase display to the audio tile select Aux Display in the Audio menu From the Aux Display submenu you select whether a phase display is shown or not To choose the format of the Aux Display select between the following two entries in the submenu m Phase The Phase display is a plot of one channel against another on an orthogonal pair of axes The Phase display can be displayed in two styles Soundstage style which plots the two channels at 45 degree angles with the mono combination appearing on the vertical axis like a left right image in a studio and X Y style which plots left channel data on the vertical axis and right channel data on the horizontal axis emulating the X Y display of an oscilloscope Surround The Surround display enables you to monitor relative loudness of the individual elements rendered in a surround sound listening environment See page 120 Surround Display Elements Quick Start User Manual How to Operate Your Instrument To choose the pair of inputs that is displayed in the Phase display select Phase Pair and select which meter bar pair you want displayed in the Phase display If you select Bars Flexible Pair for the Phase Pair you can use the FlexPhase Ch A and FlexPhase B soft keys to specify any combination of meter bars to be displayed in the Phase display
80. display the IN OUT menu 3 Select the input corresponding to the connected signal 4 Select a tile 102 C Ger fraa T no f 1890 040 Settings Digital Input A 525i 59 94 SDI Input A Ref Internal Input EMB A L Tektronix Settings Overlay Color Space Components Fi YPbPr YPbPr F Digital Input B Wed May 03 18 55 57 Linear E 50 0 50 SDI A ID Embd PPPP OH Mag x1 00 Video Session Data Collect i un ime Auto 525i 59 94 F t 352M Payload None SAY Place Err Field Length Err Li Length Err Fl AP CRC 589Dh Statistics RGB Gamut Error Cmpst Gamut Error L Ancillary Data Stuck Bits F2 AP CRC Status Err Secs 4 3413 0 0000 0 0000 66 3532 0 0 0 0 Sweep Cursors i Line Quick Start User Manual 5 Press the MAIN button to display the Main pop up menu 6 From the Main pop up menu touch Lightning to display the signal in a Lightning display 7 Press the DISPLAY button 8 Touch the view soft key that corresponds to the tile containing the Lightning display This will display the Lightning display in full screen mode 9 Set Bar Targets to match your input signal Quick Start User Manual 525i 59 94 SDI Input A Ref Internal Input EMB A Waveform Vector Picture Audio Tektronix 525i 59 94 SDI Input A Ref Internal True Peak Input EMB dBFS 0 Tektro
81. dow to scroll through the sders Pers eders Scal Scal Scale S P butto help text io E ofe E e Sagan 70 dB 0 dB 10 dB sensitive help 4 Touch Close to close the Help window e ae el a eN 2 Sec 2 Sec Peak held segment ON 5 Touch the Press any button to get context sensitive help soft key to disable Help mode Samples Sambles Back Page Up Scroll Up sero Down Page Doun 50 Quick Start User Manual To Connect to a Network The following topics cover configuring the IP settings so that you can use the waveform monitor over a network and configuring SNMP which is required if you are using commands to control the waveform monitor Connection and IP Settings How to Operate Your Instrument To allow network access to the instrument you need set the IP address Network addresses can be assigned either automatically or manually If your network does not use DHCP you will have to manually enter the address for the instrument To get an address talk to your LAN administrator Set up the monitor as described in the steps that follow NOTE You may want to connect directly between your PC and the WFM7100 See page 54 To Connect Directly to a PC 1 Press the MAIN button to display the Main menu Touch the Config soft key to display the Configuration submenu Touch the Utilities soft key Touch the Communications soft key Touch the Config Mode soft key Set the Config M
82. e Your Instrument 10 All displays are accessible from the Main 525i 59 94 Fri May 05 1459 21 ID pop up menu Some displays are available Ref Internal c vT only from the Main pop up menu The Main Line F1 21 Active menu at the bottom of the display is used to gain access to functions not tied to a specific display such as printing or network settings To display the Main pop up menu press OA Waveform Diamond Alarm Status MAIN and then touch any soft key in the PTOA e pop up menu to activate the desired display Picture Aux Status Audio Session SDI Status Bowtie Dolby Status Lightning USB Configure Print Config Standby Help Mount User Menu Tile Settings Each tile maintains its own settings independent of the other tiles such as GAIN SWEEP and display type For instance when you switch a tile to a different display the GAIN and SWEEP settings will be changed to what they were the last time the selected display was shown in the tile Display type is also independent for each tile NOTE Only one tile can contain an Audio display or DataList display and only one tile can have the Line Select mode enabled Perform the following steps to configure the display View All Available Measurements Using the Main Menu Press the front panel MAIN button to open the Main menu and display a pop up menu showing all available measurements Note that this pop up menu is the only place you can access some measurements such
83. e blue dashed lines then the signal is violating the limits that are set lf alarms have been enabled then gamut violations will be reported when the signals falls outside the boundaries shown by the blue lines if the number of pixels in error exceeds the Area threshold Arrowhead Menu The Arrowhead menu enables you to reset the threshold limits to a default setting Default Limits EBU 103 Luma Resets the luma thresholds to the values recommended in the EBU R103 2000 specification NOTE If you need to reset the Y C limits to their factory default values you must perform a factory reset To do this press the PRESETS button touch Settings touch Recall Preset and select Factory from the Recall menu Timing Display The Timing display is a Tektronix proprietary display that simplifies measuring the timing difference between two signals Display Elements 1 12 Input Signal Indicator A single circle representing timing of input signal relative to the reference Reference Indicator A cross hair indictor centered in the display represents the reference signal Vertical Offset The timing difference between the reference and input signal Horizontal Offset The timing difference between the reference and input signal Relative to Indicates the chosen zero point for the timing display 525i 59 94 M ID SDI Input A Embd PPPP PPPP PPPP PPPP Ref NTSC nput Tinjing Relatie to Ref rence Vertical Offset 9
84. e cursor Delta time readout displays the 20 80 rise time measurement To Take Automatic Eye Measurements Option Phy Only The following procedure describes how to take automatic measurements of the Eye waveform when Option Phy is installed 1 Perform the initial setup for Eye measurements See page 92 To Set Up for Eye Measurements 2 Display the Eye measurement in full screen mode NOTE Eye measurements are shown on the Eye and Jitter displays only in full screen mode However the measurements are displayed on the SDI Status display 3 Touch the Settings soft key to display the Settings submenu 4 Touch the Show Measure soft key The measurements are displayed on the left side of the screen NOTE If you use the measurement cursors to measure the eye waveform you may measure a different result than that shown in the automatic eye measurements readout This is because the automatic amplitude measurement uses histograms in order to be minimally affected by overshoot ringing and noise on the signal Similarly the automatic rise time and fall time measurements use histograms to find the centers of the distributions of 20 and 80 crossings The difference between manual and automatic measurements is usually insignificant for clean symmetrical signals Close E Show el Show Limits Set to Bypass Cable Type ow Limi Settings Settings Measure SMPTE Limits Equalizer oe eat Lew t amp j L keli tj
85. e decoded and all data and other errors are ignored This means that if an AES input is selected nothing recognizable is present on the input or if embedded audio is selected the VIDEO input is unrecognizable AES PARITY The incoming subframe does not have even parity as specified by the digital audio standards The data sample is unreliable and is ignored The level meters and Lissajous display treat the sample as a zero sample AES CRC ERROR The CRC code in the AES channel status packet is incorrect Sometimes the CRC code is set to zero indicating that the signal is missing when this is the case this message is not displayed MUTE The number of consecutive all zero samples is at or exceed the Samples for Mute setting SILENCE The signal has been at or below the specified Silence Level for a time exceeding the Duration for Silence setting DISABLED Disabled can appear for three reasons m Analog audio is selected as the bar source which causes bars 7 and 8 to display DISABLED because there are only 6 analog inputs m Embedded audio is selected as the bar source when Composite video is selected as the video input There is no embedded audio in composite so all of the bars will display DISABLED m lf the currently selected audio source s Bar to Input Map does not have all of the bars mapped to inputs any bars that are unmapped will display DISABLED Quick Start User Manual 67 How to Operate Your Instrument
86. e to rear panel mode this three microsecond conversion delay is accounted for in the displayed offset In the Relative to Saved Offset mode it has no effect Timing Multiple Inputs to a Router You can also use the Relative to function to set the offset between a master and a reference signals as the zero point reference for time down applications The Relative to box indicates the chosen zero point for the timing display m Rear Panel In this mode default the offset is zero when the measured signal input and reference are at the same timing at the reference at the instrument rear panel This setting was used in the Timing Display Method procedure m Saved Offset In this mode you can save the timing from a master signal relative to the reference as the zero point offset Then route other inputs and measure relative to this saved offset To time the inputs to a router 1 Apply the reference signal to the Ref Loop J o L REF LOOP 1890 010 2 Select the master input to the router and connect it to the Composite or SDI input 0 o J as required SDI A SDI B SDI OUT 1890 011 LOOP CMPSTA LOOP CMPSTB co oOo Se o oS oO Quick Start User Manual 99 Timing a Studio 3 Press the MAIN button to display the 525i 59 94 Thu May 04 15 49 24 ID l Main pop up menu Select the Timing Ref Internal display from the pop up menu SDI Status Jitter P P Error Log
87. ei are Jitter2 HPF 10Hz Be IS 2 Cable Loss Thermometer and P P CUA mn Eye Falltime readouts The Cable Loss Thermometer Eye Rise Fall works just like the Jitter Thermometer using color to indicate the relationship of the measurement to the Cable Loss limit which is user adjustable from the Eye or Jitter display 3 Eye Measurements With Option Phy installed a set of eye measurements are displayed below the Cable Loss measurements Quick Start User Manual 85 Monitoring the SDI Physical Layer Monitoring the SDI Physical Layer With Options Eye and Phy the waveform monitor supports the checking verification and monitoring of the SDI physical layer in the following ways m Option Eye If present this option adds support for eye pattern monitoring of the physical layer of video signals WFM6100 instruments display eye patterns for only SD signals WFM7100 instruments display eye patterns for both HD and SD signals when the options are installed Option PHY If present this option includes the Option Eye features and adds automatic eye measurements and a jitter waveform to aid the monitoring of the SDI physical layer WFM6100 instruments display jitter waveforms for only SD signals WFM7100 instruments display jitter waveforms for both HD and SD signals when the options are installed With Options Eye and PHY installed you can use the following display modes to monitor the SDI physical layer m Eye Display
88. er OMe Patan ORCI ear ernetr A ows Me eae ter ets Pee LTE a RT ees een ee 105 Selling Uptor Gamut CHECKS asiana na n E AA dapat tana tn atiass daneudadaadnanstean inne 105 Checking RGB Gaim cn iairawicnsoutu craeates a eo xtid ea veut e a ated amass A ound RA 107 Checking Composte Cie 101 eee ene en eee eee eek eer a et Semen ne ene oC nent keene ere een ere er terre 109 Checking LIma ANNI Shot endo Scan ses enner es areus te a teat ot Dna Sew senna sacle AEE 110 Automating Gamut CHECKS axran inei a duties cadences ai a a eae r fue a Send ned elses Sanita 111 Adjusting ANIM FLING te cee se teat tee nna a e a E eee mae 112 Selecting and MONNGTING AUGIO sssr eraran ANA EESE EEN EA EAE ENAT KERA EASE NENDE 114 Selecino lo COs MDU rrano eeren a S N a EO 114 Checking Audio Level amp Phase nnnnnnnnnnnnnnnnrrrrrnurnsrsrrrrrrrrrrsrrrrrrrrrrrsrrrrrrrrrresrrrrrrrrrerrrrrrr 116 CHECKING SUMOUNG SOUNG eor a a a a a a a aT 118 Remote COMMUNICAUONS seemeniana a a A a ea a a AEA 124 Using the Remote Web Interface nnnennnnnnnnnnnnnnnrrrrrnrrrrrrrrrrrrrrerrrrrrrrrrrssrrrrrrrrrresrrrrrrrrresene 124 Using the Remote Application annnnnannnnnnnnnnnnnnrnnnrnnrerrrunrrrrsrrrrerrrnerrrrrrrrrrrrerrrrrrrrrerrrrern 125 To Upgrade the Waveform Monitor Software cccc cece cece nee eee ene ents ene e rrr rrrrrrrrrrrrrrrrrrrrrrrrrrrrrrnn gt 127 PC SySiemn ReOGUIRGMEIS 12055 lt aemia onde ae chastieencaree R L aly ata eeansesemad
89. er bars for current audio levels which are displayed according to the selected meter ballistics Each bar displays three colors Green Indicates portions of the level meters below the test level m Yellow Indicates portions of the level meters between the test and peak program level m Red Indicates portions of the level meters above the peak program level Quick Start User Manual Check Audio Phase Select Aux Display From the Aux Display submenu select Phase Touch Phase Style and choose Soundstage or X Y orientation for the Lissajous signal Touch Phase Pair and select the Phase Pair to be displayed Set the Audio Input to match the signal that you are checking Check for Phase correlation of the signals noting the following m Phase correlation meters are located under the appropriate bars and one is duplicated under the phase display E For correlated signals the diamond will be green and it will move to the right side m For uncorrelated signals the indicator will be yellow and tend to Stay in the middle m For anticorrelated signals one goes up when the other goes down the indicator will be red and will move to the left side Usage Notes Selecting and Monitoring Audio 525i 59 94 Thu May 18 09 52 53 ID SDI Input B Audio In Dolby 3 Emb 3 amp 4 Ref Internal VITC 05 13 14 23 2 True Peak Dolby 3 Emb 3 4 D 3 2 L Listen Mode Full dBFS 0 Antes Gominae Aux D
90. er mark See the following table H 11 You can use the V Gain and H Gain settings touch the readouts and use the knob to adjust to expand the scale of the Lightning display to make it easier to read the delay scale The center mark of the nine marks spanning at each green magenta transition is the zero error point Alignment to a mark towards black means the color difference signal lags with respect to luma Alignment to a mark towards white means the color difference signal leads the luma signal The upper half of the display measures the Pb to Y timing the bottom half measures the Pr to Y timing Deflection SD 0 marks 0 ns 1 mark 20 ns 2 marks 40 ns 3 marks 1 74 ns 4 marks 148 ns 1 Luma sample 2 Chroma sample 104 Green to Magenta transition O y Gain x1 00 H Gain x1 00 1890 029 Delay scale Green to Magenta transition O Line Fi 158 Active 1890 054 Delay scale HD 0 ns 2ns 5 ns 13 5 ns 2 ns Quick Start User Manual Checking Gamut Checking Gamut Signals that are legal and valid in one signal representation may not be legal in another representation Specifically signals that are legal in the Digital YCbCr representation may not be legal if transcoded to RGB or encoded to NTSC PAL Any signal that fails this test is considered out of gamut for these color spaces The waveform monitor supports multiple displays and alarms to allow detecting out of gamut sig
91. erate the waveform monitor in the instrument chassis be sure the top cover is on or installed in an approved portable cabinet or rack adapter You can also install the waveform monitor in a custom installation such as a console To install the waveform monitor in a cabinet or rack follow the instructions that are included with the optional accessory kits that are available for each type of installation CAUTION Do not install the waveform monitor in any cabinet except those that are listed in Optional Accessories Attempting to do so can damage the waveform monitor and the cabinet If you install the waveform monitor in a custom application such as a console be sure to provide adequate airflow Do not block the ventilating holes CAUTION Failure to provide adequate airflow to the waveform monitor could cause the instrument to shut down If the airflow is blocked and the instrument does not shut down the instrument could be seriously damaged Connecting Power and Powering On Off The waveform monitor operates from a single phase power source with the neutral conductor at or near earth ground The line conductor is fused for over current protection A protective ground connection through the grounding conductor in the power cord is essential for safe operation AC Power Requirements The waveform monitor operates from an AC line frequency of 50 or 60 Hz over the range of 100 240 Volts without the need for configuration except the power
92. erence Formats 720p 720p 720p 1080p 1080sf 1080p 1080sf 1080i Input Format NTSC 23 98 29 97 59 94 23 98 23 98 29 97 29 97 59 94 483i 59 94 Hz X X X 525 2 720p 23 98 Hz X X X X X 720p 23 97 Hz X X X 720p 59 94 Hz X X X X X 1080p 23 98 Hz X X X X X 1080sf 23 98 Hz X X X X X 1080p 29 97 Hz X X X 1080sf 29 97 Hz X X x 1035i 59 94 Hz3 X X X 1080i 59 94 Hz X X X 1 Available with Opt CPS 2 Available with WFM7100 Opt SD or with WFM6100 Opt SD 3 Available with WFM7000 and WFM7100 Opt HD only Reference Formats 720p 1080p 1080sf 1080p 1080sf 1080 Input Format 24 720p 30 720p 60 24 24 30 30 60 720p 24 Hz X X X X 720p 30 Hz 1 X X 720p 60 Hz 1 X X X X 1080p 24 Hz 1 X X X X 1080sf 24 Hz1 X X X X 1080p 30 Hz 1 X X 1080sf 30 Hz X X 1035i 60 Hz 1 X X X X 1080i 60 Hz 1 X X X X 1 Available with WFM7000 and WFM7100 Opt HD only Quick Start User Manual 19 How to Operate Your Instrument AES A B Connectors 1 These BNC connectors support AES p audio inputs 0 0 A1 2 In AIN12 B101 A3 4 In ASS In A7 8 In aes AINS4 B1034 B1 2 I O 1 B3 4 1 0 1 Bo 6 0 1 AIN5 6 B 1 0 5 6 AIN78 B1 07 8 1890 007 Options DS AD DD DDE 1 The AES B connectors can be configured to output embedded audio channels decoded Dolby or the AES A inputs Analog Input Output Connector The Analog I O connector is used to receive and send analog signals The Analog I O connector is a 62 pin D subminiature connector
93. esence Note that the center bar is below the left and right levels This indicates that the center level is weaker and will be less audible than the ambient sound 6 Monaural signal in channels Ls and Rs creating a phantom source in the center as ina 3 1 surround sound system Quick Start User Manual 123 Remote Communications Remote Communications Remote communications that are covered here include those that are Over an Ethernet network through a Web browser E Over an Ethernet network through a Java applet Using the Remote Web Interface You can connect to the waveform monitor and perform simple operations without installing any software and using just a Web browser To use a Web browser to connect to a waveform monitor the waveform monitor must be connected to an IP network via Ethernet You will need to set the IP Config Mode IP Address Subnet Mask and possibly the Gateway Address depending on your network configuration Using only a Web browser you can save screen captures download the error log download the diagnostics log and view the instrument options To connect to the waveform monitor using a Web browser 1 Setup the waveform monitor to connect to a network if you have not already done so See page 51 To Connect to a Network 2 Start your Web browser and type the Fie Edt View Favorites Tools Help ae network address of the waveform Ou O x 2 OD Osean O B Links monitor into the URL
94. esets Support for printing the display and event log on USB printers WFM6100 and WFM7100 only CaptureVu capability that can capture a full frame of video data and can compare live to captured data on the waveform vector gamut and picture displays Data can be captured using the touch screen menu or triggered by up to 14 alarm conditions Captured data can be stored on a USB flash drive for transfer to another instrument or to a PC for more in depth analysis Up to four independent displays of the signal characteristics of an SDI input including Eye diagrams and measurements jitter readouts and bar display jitter waveform display and SDI status screen Two independent jitter measurements with separate high pass filters that offer simultaneous display of peak to peak amplitudes of timing and alignment jitter Cable loss measurement with source level and cable length estimates based on user specified cable type Extensive support for monitoring Dolby audio content including auto sensing Dolby formats decoding Dolby content and automatically configuring the appropriate audio level and phase displays and showing decoded Dolby metadata Quick Start User Manual vii Preface Documentation Item WFM6100 WFM7000 amp WFM7100 Series Waveform Monitors Quick Start User Manual WFM6100 WFM7000 amp WFM7100 Series Online Help WFM6100 WFM7000 amp WFM 7100 Series Waveform Monitors Performance Verification and Speci
95. etaetaeteninetaterasa mene 127 Bere YOUIBCOIN ss anea sass genus canons a stan sented send ypoaay aA aa tases 127 Mstaling Me SOM WANG ace Vacusecctestettestnases eect tessa a aioe satendaciasee baer nete mas hune tala dente 128 VOT IEV ING CO GAC orains oecumn ane a cta oe nea em tane asa eon tauaiemta aa e 130 STUE UNO esse saz assed aia cee Me cps cc se org esas area caine oe eae eka ean Nee eae ea mance ne E E 133 Index Quick Start User Manual General Safety Summary General Safety Summary Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it To avoid potential hazards use this product only as specified Only qualified personnel should perform service procedures While using this product you may need to access other parts of a larger system Read the safety sections of the other component manuals for warnings and cautions related to operating the system To Avoid Fire or Personal Injury Use Proper Power Cord Use only the power cord specified for this product and certified for the country of use Connect and Disconnect Properly Do not connect or disconnect probes or test leads while they are connected to a voltage source Connect and Disconnect Properly De energize the circuit under test before connecting or disconnecting the current probe Connect and Disconnect Properly Connect the probe output to the measurement instrument before connecting the probe to t
96. eturn to save the name ji 1890 013 Live Capture Capture Settings Display On Display On Buffer Close l Copy Buffer Restore Buffer Trigger Types Settings To USB From USB emere Ban oA ae gt Er Buffer Copy Buffer Rere Buffer l Trigger Types Settings To USB From USB Ard cAed RA wispy wit dete ds ede ted de fw ea ee ee Ee ee ee C3 AAA e Gl MESCA Ea EA Quick Start User Manual 8 Touch the Copy soft key to save the buffer capture to the USB Flash Drive The waveform monitor will indicate its progress saving the file to the flash drive with the Progress Status readout While the copy is occurring the Capture menu will flash to indicate the copy is in process The waveform monitor will display Operation Complete when the copy is finished 9 When you are ready to remove the USB flash drive from the instrument press MAIN 10 From the Main menu touch the USB Unmount soft key and them remove the USB flash drive from the instrument How to Operate Your Instrument Restoring a Buffer Capture from a USB Flash Drive 1 Insert a flash drive into the USB port on the front panel 2 Touch the CAPTURE button to display the Capture menu 3 Touch the Settings soft key to display the submenu 4 Ifthe USB Flash Drive is not recognized a description of the drive will appear after USB Storage Device touch the Reset soft key to mount the flash drive mf the flash drive is still not
97. fications WVR amp WFM Series Programmer Manual WFM6100 WFM7000 amp WFM7100 Series Waveform Monitors Service Manual Purpose Installation and high level operational overview This manual Context sensitive operation and user interface help Procedure for checking performance and list of specifications Programmers command reference for controlling the instrument Optional manual supporting module level servicing of the instrument Conventions Used in this Manual The following icons are used throughout this manual viii Sequence Step Connect power Network XGA aR USB Quick Start User Manual Before Installation Before Installation Check Package Contents Unpack the instrument and check that you received all items listed as Standard Accessories Check the Tektronix Web site www tektronix com for the most current information Standard Accessories Documents The following documents are standard accessories m WFM6100 WFM7000 amp WFM7100 Waveform Monitors Quick Start User Manual Tektronix part number 071 1890 XX m WFM6100 WFM7000 amp WFM7100 Waveform Monitors Release Notes Tektronix part number 071 1895 XX m WFM6100 WFM7000 amp WFM7100 Waveform Monitors Documents CD Tektronix part number 063 3922 XX m WFM6100 WFM7000 amp WFM7100 Waveform Monitors Specifications and Performance Verification Technical Reference Tektronix part number 071 1897 XX
98. formation about the environmental impact of the product Product End of Life Handling Observe the following guidelines when recycling an instrument or component Equipment Recycling Production of this equipment required the extraction and use of natural resources The equipment may contain substances that could be harmful to the environment or human health if improperly handled at the product s end of life In order to avoid release of such substances into the environment and to reduce the use of natural resources we encourage you to recycle this product in an appropriate system that will ensure that most of the materials are reused or recycled appropriately The symbol shown below indicates that this product complies with the European Union s requirements according to Directive 2002 96 EC on waste electrical and electronic equipment WEEE For information about recycling options check the Support Service section of the Tektronix Web site www tektronix com Mercury Notification This product uses an LCD backlight lamp that contains mercury Disposal may be regulated due to environmental considerations Please contact your local authorities or within the United States the Electronics Industries Alliance www eiae org for disposal or recycling information Restriction of Hazardous Substances This product has been classified as Monitoring and Control equipment and is outside the scope of the 2002 95 EC RoHS Directive This product is kno
99. ft Keys perform an action a ah aig tse ace A Faas Chem Ofivet such as invoking another menu zi we For example in most menus you select the Settings soft key to display a submenu of parameters that set for the displays Toggle Soft Keys select between two or more values For example in the Waveform menu there is a soft key that you use to specify whether the waveform is displayed in Overlay or Parade mode Option Soft Keys select a single item trom Fixed Gain Fixed Gain Fixed Gain Fixed Gain aveltee NOUR AEGNE x1 00 x2 00 5 00 10 00 For example in the Mag Gain menu Option menu keys are used to select the amount of vertical gain Knob value Menu Keys connect the value Saige to the knob enabling the value to be changed O VBI CC Line by turning the knob 1 For example in the Closed Caption submenu you can specify the line used for Closed Captioning by selecting the VBI CC Line soft key and turning the knob Note that when the soft key is connected to the knob the knob icon is filled in You can disconnect the soft key from the knob by touching the soft key again Check box Menu Keys select between i two settings normally On Off or Components Enabled Disabled For example in the Waveform Components menu you can select the components that appear in the display You can select all components to be displayed or only one component 14 Quick Start User Manual How to
100. gt E EEE EER Jae 1 8 Touch OK to save the configuration Local USB Printer moan USB Printer 1 HP LaserJet 6 series pcl 6 Quick Start User Manual 57 How to Operate Your Instrument To Create a User Menu The waveform monitor enables you to create a user defined menu for your requirements The MyMenu custom menu saves you time by placing frequently used controls on an easily accessed menu 58 To create a User menu 1 Press the MAIN button to enter the Main menu 2 Select Configure User Menu 3 Select Conf User On from the menu A check mark will appear in the check box when the function is enabled 4 Touch OK to begin configuring your custom menu NOTE The menu configuration process will display a series of instructions to guide you through the process of creating your menu 525i 59 94 Thu May 04 15 49 24 SDI Input A Ref Internal SDI Status Jitter P P Vector Arrowhead ANC Display Error Log Picture Aux Status Audio Session SDI Status Bowtie Dolby Status Lightning USB Configure Print Config Standby Help Mount User Menu 525i 59 94 Thu May 04 15 52 26 SDI Input A Ref Internal SDI Status his feature allows you to select defined menu entries to place within he User menu By selecting the Ok button you will start the User menu configuration Jitterl HPF 10Hz P P At any point you can navigate back to the main menu and deselect he Conf User On menu entry to
101. he circuit under test Connect the probe reference lead to the circuit under test before connecting the probe input Disconnect the probe input and the probe reference lead from the circuit under test before disconnecting the probe from the measurement instrument Ground the Product This product is grounded through the grounding conductor of the power cord To avoid electric shock the grounding conductor must be connected to earth ground Before making connections to the input or output terminals of the product ensure that the product is properly grounded Ground the Product This product is indirectly grounded through the grounding conductor of the mainframe power cord To avoid electric shock the grounding conductor must be connected to earth ground Before making connections to the input or output terminals of the product ensure that the product is properly grounded Observe All Terminal Ratings To avoid fire or shock hazard observe all ratings and markings on the product Consult the product manual for further ratings information before making connections to the product Connect the probe reference lead to earth ground only Do not apply a potential to any terminal including the common terminal that exceeds the maximum rating of that terminal Power Disconnect The power cord disconnects the product from the power source Do not block the power cord it must remain accessible to the user at all times Do Not Operate Without Covers Do n
102. hysical Layer 90 2 Touch the Bypass Equalizer soft key to toggle the setting on or off m On Bypass the equalizer when the waveform monitor is connected to a signal with a short length of cable to minimize the jitter contributed by the internal equalizer This setting is generally used only for very low jitter signals m Off Enables the equalizer which enables operation with typical cable lengths and reduces jitter due to cable effects Use this setting for most signals CAUTION If the equalizer is left in bypass mode the waveform monitor will not work with cables more than a few meters long This applies to all SDI displays 3 Touch the Close Settings soft key To Configure Physical Layer Alarms 1 Press the Main front panel button 2 Touch the Config soft key 3 Touch the Alarm Setup soft key to open the alarm setup controls Bypass Equalizer Close Colors amp Readouts amp Diagnostics Closed Caption Alarm Setup Utilities Config Intensity Brightups Exit Enable Alarm Setup Set All Alarms to This Mask Alarms Video Content Video Format SDI Input Composite Input General General Audio AES and Embedded Embedded Audio Specific Dolby Specific Closed Captions Metadata ARIB Specific Physical Layer Quick Start User Manual Monitoring the SDI Physical Layer 4 Touch the Enable Alarms soft key to turn on or off all limit alarms When Enable Enable Alarms is turned off no alarms Alarm
103. ignal in line or field sweep You can choose the mode in which SDI signal elements are displayed RGB YRGB or YPbPr and you can apply filters to the signal You can also display an SDI input as though it were a composite signal You can also control whether EAV and SAV is included in the display in the Settings submenu Waveform Display Elements 1 Is colored white when vertical gain is X1 otherwise yellow color indicates V Gain is x2 or greater or set to variable 525i 59 94 SDI Input A Ref Interna In l O Line F1 96 Active 2 Is colored white when Horiz Mag is X1 2 Word 1492 Cb HB otherwise yellow color indicates H Mag m is x10 or greater or set to Best View 3 Lists the currently displayed waveform color space Dashes indicate components not displayed 4 List the current sweep rate for tile 5 Lists the currently selected field and line when in Line Select mode B V Gain x1 00 H Mag x1 00 Qa D ae i Color Space Components Filter Sweep Settings Cursors Parade YPbPr YPbPF Flat 1 Line 1890 017 Quick Start User Manual 61 How to Operate Your Instrument 62 Waveform Menu and Settings The Waveform menu enables you to choose the display style and display mode used in the active tile SDI inputs only select a filter to apply to the input signal and set the sweep mode To display the Waveform menu m Press WFM on the front panel To specify the color space used to di
104. in use If you need to measure launch amplitude or physical length you must set the cable type to the actual cable used Setting the Eye Type 1 Touch the Eye Type soft key See page 86 Displaying the Eye Measurement Menu Select one of the following Eye The Eye display shows the SDI input signal directly Equalized Eye The Eye display shows the SDI input signal after it has passed through the internal cable equalizer and comparator Touch the Close Eye Type soft key Setting the Jitter High Pass Filter 1 Touch the Jitter HPF soft key and select a high pass filter value See page 86 Displaying the Eye Measurement Menu NOTE Timing sets the filter value to 10 Hz Alignment sets the filter value to 1 kHz for SD or 100 kHz for HD 2 Touch the Close Jitter HPF soft key Close l Eye Type 3 Eyes Eye Equalized Eye Settings Eye Type Eye L 1m L Latha Close Timing Alignment Jitter HPF 100 KHz A 99 556 ns ek share aya Byes 4 EERE PEINE Se PEERAA Eye Type 3 Eyes Jitter HPF Sweep Cursors Eye 10 Hz 1 Line Setting the SMPTE 259 292 Alarm Thresholds 1 88 Touch the Settings soft key See page 86 Displaying the Eye Measurement Menu Close Show Set to Bypass Cable Type a E Show Limits i j Settings Settings Measure SMPTE Limits Equalizer 8281 L 7 J Le w t amp j L ww Quick Start User Manual 2 To reset both the SMPTE 259
105. ings Surround Bars L amp R Linear Dolby 1 Close Aux Display Phase Pair Dominance Linear A Weighted Audio Settings Surround Filter Surround Bars L amp R Ind On o they i 525i 59 94 Thu May 18 09 59 48 ID SDI Input B Audio In Dolby 3 Emb 3 amp 4 Ref Internal VITC 05 20 09 03 2 True Peak Dolby 3 Emb 3 4 D 3 2 L Listen Mode Full BFS LR PSI L C C R PSI Phantom source grid 50 center of L or C and R Dominance indicator R Rs PSI j Center volume indicator T _ Otal Volume indicator polygon L Ls PSI pa Level indicator Loudness grid 10 dB per I a marker Ls Rs PSI 119 Selecting and Monitoring Audio Surround Display Elements m The Level Indicators indicates the audio level balance among the left L right R left surround Ls and right surround Rs channels as cyan lines on the Loudness Grid m The Loudness Grid the ruled scale radiating from the center on which the level indicators register their audio level and balance The scale has tic marks at 10 dB intervals An additional tic mark indicates the 18 dB level The 18 dB and 20 dB levels are typically used for aligning audio levels Total Volume Indicator the cyan polygon formed by connecting the level indicator end points showing the total sound volume of the L R Ls and Rs channels Each connecting line will bend away from the center if the two signals have a positive correlation wil
106. ion of the two audio signals Displays phase correlation between the two channels of each appropriate pair Located under the bars for the same channels Duplicated under the Phase display Has configuration response time Diamond shaped markers between the level bars that indicate the configurable limits set up for the display Above the Test level the bar displays in a yellow color Above the Peak level the bar displays in a Red color Test level is also known as Reference level or Line up level Quick Start User Manual How to Operate Your Instrument Element Description Level meter scale and units By default the zero dB mark is at digital full scale and units are in dB relative to full scale ABFS You may also set the 0 dB mark to either the Peak Program level or the Test level Above bar Warning Messages The waveform monitor displays warning messages above the level meter bars The warning messages that can appear above the bars are shown below CLIP The number of consecutive samples is at or exceeds the Samples for Clip setting OVER The signal has been at or above the specified Over Level for a time exceeding the Duration for Over setting In bar Warning Messages The waveform monitor displays warning messages within the level meter bars The warning messages that can appear are shown below in order of priority UNLOCKED The instrument is not locked to an incoming signal on the indicated input channel Data cannot b
107. ircle will be centered on the cross hair The timing offset is also shown numerically as lines and microseconds of advance or delay in the boxes at the right side of the display For input and reference signals with closely related frame rates there is only one timing relationship so a single circle is shown on the display to indicate the timing offset of the input signal For input and reference combinations with more complex relationships multiple circles are displayed to indicate all the possible interpretations of the timing offset with the one that is closest to zero shown with emphasis The numerical readouts will correspond to the timing indicator circle with the emphasis The Relative to box indicates the chosen zero point for the timing display The default is Rear Panel In this mode the offset is zero when the input and reference are at the same timing at the rear panel of the waveform monitor The other choice is Saved Offset In this mode you can save the timing from one signal and then display the timing relative to that saved offset To use the Timing display to time a signal to a reference 1 2 3 Select a tile in which to time the active za Reed Dita eee input Apply the input signal to be timed to the appropriate input terminate it properly if necessary and then select it from the IN OUT menu Apply the house reference signal to the external reference input Press IN OUT and then select Ex
108. isplay Phase Style Phase Pair FlexPhase Ch A FlexPhase Ch B Audio Input g Phase X Y Bars L amp R Audio Bar L Audio Bar R Dolby 3 525i 59 94 Thu May 18 09 51 53 ID SDI Input B Audio In Dolby 3 Emb 3 amp 4 Ref Internal VITC 05 12 14 24 2 True Peak Dolby 3 Emb 3 4 D 3 2 L Listen Mode Full dBFS i L in phase a Aux Display Phase Style Phase Pair FlexPhase Ch A FlexPhase Ch B Audio Input g HEEJ Soundstage Bars L amp R Audio Bar L Audio Bar R Dolby 3 The phase or Lissajous display is a plot of one channel against another on an orthogonal pair of axes Soundstage plots the two channels at 45 degree angles with the mono combination appearing on the vertical axis more like a left right image in a studio X Y plots left channel data on the vertical axis and right channel data on the horizontal axis emulating the X Y display of an oscilloscope The response time of the correlation meters can be set from the CONFIG menu The choices follow Quick Start User Manual 117 Selecting and Monitoring Audio Speed setting default O CO lIN Oo o w rh gt NIe ea la _ gt M a d CO O o nNn Ds ou amp ow NM gt lo Checking Surround Sound Response averaging time s 0 04 0 04 0 08 0 12 0 28 0 52 1 0 1 5 2 0 2 9 3 0 3 9 4 0 4 5 5 0 5 5 6 0 6 5 7 0 1 9 The Audio display can also display the s
109. isplay presents a voltage versus time waveform of the input serial signal This trace is composed of multiple samples of the SD or HD serial bit stream 82 Tue May 02 13 11 43 Audio Session Audio Input Embedded A Data Collect Running Signal Loss Run Time 0 d 00 19 05 Analog Output 2 1 3 amp 4 3 3 amp 6 5 6 AES B Output eL e 6 5 Channel Clip s Over s Mute s Silence s Peak dBFS 99 0 High dBFS 99 0 Active bits 0 Smpl Rate Changed since reset Depending on the options you have purchased the waveform monitor supports checking verification and monitoring of the physical layer of SDI signals connected to its SDI inputs in the following ways m Option EYE If present the Eye option adds support for eye pattern monitoring of the physical layer of video signals The Eye display includes readouts for P P jitter cable length and cable loss and source level m Option PHY The Phy option adds support for eye pattern monitoring of the physical layer of video signals and includes support for jitter waveform display Additional measurements included with this option are amplitude rise time fall time and rise fall time difference Quick Start User Manual The Eye display can be set to display eye measurements and limits 1 Jitter Thermometer The Jitter Thermometer is a visual indicator of the jitter amplitude The Jitter Thermometer uses a colored bar to indicate the relationship of
110. ith all the waveform displays Use this method to capture only the displayed trace from a video frame Captures made using this method do not include the associated decoded data To capture a display eae n Wed Apr 26 14 28 47 1 Press the front panel CAPTURE button This displays the Capture menu 2 Press the Settings soft key O V Gain x1 00 H Mag x1 00 Live Capture Capture Settings j Clear 5 Display On Display On Buffer 3 Touch the Buffer Trace soft key to Clos Copy Buffer Restore Buffer set the type of Capture Touch Close Settings Trager Types to USB From USB Settings to remove the submenu 4 To display the capture be sure the Live Capture Capture ettings Capture Display On soft key is checked a Display On Display On Buffer NOTE The Live Display On setting is specific to the tile selected and can be set independently for each tile Quick Start User Manual 41 How to Operate Your Instrument 42 5 To capture the trace or buffer touch the Wed Apr 26 14 41 45 Capture Trace or Capture Buffer soft rn key m Only one capture can be active at a time After a capture has been performed the Freeze Active icon appears in the Status Bar the Capture soft key is no longer selectable and the CAPTURE button on the front panel is illuminated m When Freeze Active is displayed another capture cannot be made until the current capture is cleared O V Gain x1 00 H Mag x1 00 Live Captu
111. ke selections from a submenu Some settings provide multiple choices within the submenu For example when selecting the components that are shown in the WFM display you choose from all the available components within one submenu NOTE The selections in a submenu can change depending on settings To Select Among Video Inputs f Overlay Color Space Components Filter Sweep ess YPbPr YPbPr Flat 1 Line Saet Close Chrm Offset Settings Overlay Color Space Components Sane Stri YPbPr omame ee Close Settings uveray Color Space Maca J L st we a TOEN YRGB RGB SDI gt Cmpst J L acj l L ia Close Settings Overlay Color Space Components Y Pb Pr YPbPr I L t vI jJ I L anj I n w j I You can connect serial digital component signals and select them for display in a tile Depending on the model and options that you ordered you can connect standard definition component signals high definition component signals and analog composite signals To select a component input 1 Connect any digital component video signals to the A and B SDI inputs on the rear panel NOTE The A and B inputs are separate inputs and cannot be used as a loop through path 26 SDI A SDI B SDI OUT CLOCK OUT 1890 008 Option Eye Phy Quick Start User Manual EHOW SDI A SDI B SDI OUT 1890 011 Option SD HD 2 Connect any a
112. l bend towards the center if the signals have a negative correlation and will not bend if the signals are uncorrelated m The Center Volume indicator displays the sound volume of the center channel as a vertical yellow bar between the L and R channels and connects the ends of the L C and R audio level indicators with straight lines m Channel Pair Phantom Sound Indicators PSls located on each side of the Surround Sound display indicate the location of potential phantom sound sources formed by adjacent channels The white tic marks on these moving bar indicators show the phantom source locations The bar length indicates the correlation between the adjacent channels A short to medium length green bar indicates positive correlation between the channels forming a localized phantom sound source located at the white tic mark The bar grows to full length and changes color to yellow as the correlation moves to zero indicating a wide non localized sound image The bar changes color to red for adjacent channels with significant negative correlation For negative correlations the ends of the PSI for the L and R channels will continue to grow at a 45 angle while the other PSIs remain at full length Center Pairs Phantom Sound Indicator a fifth PSI at the top of the display shows potential phantom sources formed by the LC channel pair and the CR channel pair If the L R and C channels all have the same signal level the white tic mark on the bar will
113. lanced differential analog audio input Balanced differential analog audio input Balanced differential analog audio input Balanced differential analog audio output Balanced differential analog audio output Balanced differential analog audio output Balanced differential analog audio output Balanced differential analog audio output Balanced differential analog audio output Balanced differential analog audio output Balanced differential analog audio output No connection Ch Ch Ch Ch Ch Ch Ch Ch Ch Ch Ch Ch Ch Ch 1 positive 2 positive 3 positive 4 positive 5 positive 6 positive 7 positive 8 positive 1 line A negative 1 line B negative 2 line A negative 2 line B negative 3 line A negative 3 line B negative 4 line A negative 4 line B negative 5 line A negative 5 line B negative 6 line A negative 6 line B negative 1 negative 2 negative 3 negative 4 negative 5 negative 6 negative 7 negative 8 negative 21 How to Operate Your Instrument Connecting Signals When connecting audio signals to the Analog Input connector you can use either balanced or unbalanced signals If you connect unbalanced signals to the inputs you do not have to ground the unused lead grounding the unused lead can reduce noise When connecting the Analog Output connectors you can connect them as balanced or unbalanced However if you connect
114. ll be in the center when input Rel to Rel to Save Rear Panel Saved Offset Offset NOTE As you adjust timing the circle representing the input timing may jump occasionally This is because the color frame detection circuit can be temporarily disrupted as the signal shifts The jump is often a multiple of the field time The circle will settle back to the correct location in a second or so Usage Notes m The resolution of the timing display for Composite and SD signals is one 27 MHz clock cycle or 37 ns For HD signals it is one clock at 74 25 MHz which equates to about 13 5 ns To get the greater accuracy needed for a composite signal first use the timing display to get close then use a vector display for the final burst phase alignment Since the waveform monitor can display both the timing display and a vector display simultaneously each in its own tile this process can still be easy and quick 98 Quick Start User Manual Timing a Studio m For composite signals the definition of time aligned is obvious but for an SDI input relative to a analog reference the situation is more complex For the timing display the definition of zero offset on an SDI input uses the methodology described in SMPTE RP168 This method specifies the SDI signal will be converted to analog The converted analog signal is then compared to the analog reference For the conversion a D A converter with delay of about three microseconds is used m Inthe relativ
115. m the large knob Quick Start User Manual 47 How to Operate Your Instrument To Configure Your Instrument The Configuration menu is used to change the settings of the waveform monitor that are changed only occasionally or settings that are not specific to a tile such as printer settings To display the Configuration menu 525i 59 04 Wed Apr 26 181238 1 Press the MAIN button to display the Main menu 2 Touch the Config soft key to display the Configuration menu Waveform Vector Picture Lightning OV Gain x1 00 H Mag x1 00 USB Configure Print Config Standby Help Unmount User Menu To Select Adjust a Parameter Close a Colors amp Readouts amp p Diagnostics Closed Caption Alarm Setup Utilities Config Intensity Brightups 3 Touch the soft key for the parameter you would like to adjust 4 Adjust parameters as you would any other setting The following table lists the parameters that can be set from the Config menu Soft key Description Diagnostics Run diagnostic tests and calibration procedures Calibrate the touch screen For details on how to run the diagnostic and calibration procedures see the service manual Colors amp Intensity Set colors for trace and graticule and set intensity for elements of the display Readouts amp Brightups Enable the display of brightups on gamut errors enable the display of the line select cursor on the Picture display specify whether Capture applies t
116. mits that are set lf alarms have been enabled then gamut violations will be reported when the signals falls outside the boundaries shown by the blue lines if the number of pixels in error exceeds the Area threshold Diamond Menu The Diamond menu enables you to reset the threshold limits to two default settings Default Limits Tek RGB Resets limits to factory defaults Default Limits EBU 103 RGB Resets the limit thresholds to the values recommended in the EBU R103 2000 specification 70 Quick Start User Manual Arrowhead Display Pressing the MAIN button calls up the Main pop up menu From the Main pop up menu you can select the Arrowhead display under the Gamut heading The Arrowhead display is used to show out ot gamut conditions in composite color space without requiring a composite encoder The Arrowhead display plots luma on the vertical axis with blanking at the lower left corner of the arrow The magnitude of the chroma subcarrier at each luma level is plotted on the horizontal axis with zero subcarrier at the left edge of the arrow The upper sloping line forms a graticule indicating total luma subcarrier amplitudes The lower sloping graticule indicates lumatsubcarrier extending toward sync tip maximum transmitter power Display Elements 1 Y High threshold Shows the level above which the Luminance value is considered too large and out of gamut Signals above this level will trigger an error alarm if the number of
117. monitor will strip any leading or trailing spaces in the string for example my stuff will become my stuff The private community string cannot be blank If you input a blank string the instrument will set it to the default string m Trap Destination 1 4 Address These settings allow you to set IP addresses to which SNMP traps are sent Remote SNMP Mode This setting enables you to allow or deny remote control access to the waveform monitor over a network You can still access the waveform monitor using the Web Interface when this setting is Disabled but you will not be able to view or change instrument settings using SNMP management software If this setting is Read Only you can view instrument settings but not change them If this setting is Enabled you can view and change instrument settings using SNMP management software 10 Press the Close soft keys to close the submenus Quick Start User Manual 53 How to Operate Your Instrument To Connect Directly to a PC A common configuration is to connect the monitor directly to a PC To Print Screens and Event Logs 54 1 Connect your waveform monitor to a PC through a HUB as shown the actual IP addresses will depend on your network A standard Cross Over cable can be substituted for the HUB Set up the monitor as though you are going to connect to a network See page 51 Jo Connect to a Network Choose Manual IP mode and set the IP address manually Be sure to
118. nal status You can view current alarms and errors those occurring now and within the last few seconds a history of errors and alarms up to 10 000 entries video or audio error statistics and Dolby audio status You can display a different Status display type in all four tiles at once Pressing the STATUS button calls up the last Status display shown in the selected tile and displays the Status menu 78 Status Display Types Alarm Status Shows the status of all alarms the instrument is set to monitor The status of an alarm is displayed in different colors m Gray The alarm is not enabled for reporting m Red The alarm is occurring now m Yellow The alarm is not occurring now but has occurred within the last five seconds m Green The alarm is not occurring and has not occurred within the last five seconds ARIB Display Decodes several types of ARIB data You select the type of ARIB format to decode from the ARIB menu 525i 59 94 SDI Input A Ref Internal Alarm SDI Input Missing SDI Input Signal Lock Reference Missing Ref Lock Ref Fmt Mismatch RGB Gamut Error Composite Gamut Error Luma Gamut Error Video Fmt Change Video Fmt Mismatch Vid Ref Mismatch Video Not HD Line Length Error Field Length Error EAV Place Error SAV Place Error Y Anc Parity Error C Anc Parity Error AP CRC Error FF CRC Error EDH Error Jitter Level Jitter2 Level Wed May 10 16 38 19 o B Alarm Status ID l
119. nalog composite signals to the A or B Composite inputs at the rear panel On Option CPS equipped instruments only 3 For the composite inputs terminate the loop through input properly at the rear panel for any inputs that are not routed to another device 4 Press the IN OUT button to display the 525i 59 94 SDI Input A In Out menu Ref Internal Input EMB A 5 Press the soft key for the input you want to display Digital Digital Analog Analog Int Ref soe Input A Input B Input A Input B Quick Start User Manual How to Operate Your Instrument Mon Apr 24 15 59 48 ID Embd PPPP VITC 11 55 22 01 EA H Mag x1 00 Video Session Input SDI A Data Collect Signal Run Time Format Auto 525i 59 94 352M Payload Non Ancillary Data None Stuck Bits Fl AP CRC F2 AP CRC Statistics Status Err Secs Err Fields Err Fields RGB Gamut Error 23023 1379567 100 0000 Cmpst Gamut Error 10121 0 7336 Luma Gamut E 23023 1379568 100 0000 FF CRC Error 0 0 0000 AP CRC Error o 0 0000 EDH Error 0 0 0000 21 How to Operate Your Instrument To Set Gain Horizontal Magnification and Sweep Each tile maintains its own settings independent of the other tiles These settings include Gain Horizontal Magnification and Sweep For instance when you switch a tile to a different measurement the Gain Horizontal Magnification and Sweep settings will be changed to what they were the last time
120. nals The flexible tile display allows you to simultaneously view several of the gamut measurements to learn which is most appropriate for a given application The displays are Diamond for checking that SDI signals conform to legal RGB gamut space which appears as a Split Diamond on the WFM6100 WFM7000 WFM7100 monitors and separates the upper and lower diamonds to show excursions below black m Arrowhead for checking if an SDI signal is legal for composite color space Composite Waveform modes for checking both SDI and composite signals for legality in composite color space Diamond and Arrowhead have adjustable thresholds If the signal goes outside the area defined by the thresholds the signal is out of gamut If these limits are exceeded then the waveform monitor can generate alarms if configured to do so For composite waveforms the legal limit is simply the maximum level allowed for the combination of luma and chroma This limit depends on application For example a recording on a tape may be able to handle a larger signal than if driving into a transmitter The waveform monitor also provides brightups for checking gamut Brightups or contrast patterns appear on areas of the Picture display that are out of gamut as defined by the different threshold settings Brightups can be displayed for RGB gamut errors Composite gamut errors and Luma gamut errors To set thresholds for brightups adjust the thresholds in the Diamond and Arro
121. nals To select the definition for the zero timing offset select one of the following m Rel to Rear Panel which means that the timing offset will be shown as zero when the two signals are timed down at the rear of the waveform monitor Rel to Saved Offset which means that the timing will be shown as zero offset when the input signal matches the timing of the signal that was present when the offset was saved using the Save Offset menu entry This selection changes both the numeric readouts and the target in the middle of the timing display Timing Displays for Simple Versus Complex Timing The number of circles with each representing Mon May 15 15 18 49 Embd PPPP PPPP PPPP PPPP SDI Input A a time offset varies with the complexity Ref NTSC heres of the timing between the reference and input signals If you are timing input signals with rates that are integer multiples of the reference rates the instrument can measure the timing deterministically and displays the relationship as one circle offset relative crosshair reference as is shown Such cases include timing an NTSC input multiplier of 1 or a 525 SDI input whose frame time is 33 36 ms multiplier of 2 against an NTSC reference which has a frame time of 66 73 ms Input Timing Relative to Reference Vertical Offset 7 lines advanced Horizontal Offset 7 778 us delayed Relative to Rear Panel Circle will be in the center when input Rel
122. ng handles and BNCs Weight Net Shipping Quick Start User Manual Description 100 to 240 VAC 10 50 60 Hz 50 to 110 VA at 110 or 240 VAC Depends on options installed 50 Watts typical for base unit 75 Q BNC internally terminated 800 mV p p 10 1 8 V to 2 2 V DC peak AC 6 0 V to 6 0 V DC peak AC BNC 75 Q terminated unbalanced 0 2 V to 2 V p p 0 C to 40 C 20 C to 60 C 3 000 meters 9 842 feet 12 192 meters 40 000 feet 2 Indoor use only mm in 133 4 5 25 215 9 8 5 460 4 18 125 kg Ibs 5 5 12 9 1 20 133 Specifications 134 Quick Start User Manual Index Symbols and Numbers 1 Field Sweep submenu 62 1 Line Sweep submenu 62 2 Field Sweep submenu 62 2 Line Sweep submenu 62 A Accessories documents 1 Optional 1 power cords 1 standard 1 4 Active Picture Picture Settings 65 Active tile 11 AES connectors 20 Alarm Setup soft key Config menu 48 Alarm Status Main menu 78 Analog audio connecting 22 Analog Input Output connector 20 ANC Data display 77 Ancillary Data Present Aux Status display 79 Applications Checking Chroma Luma delay 102 Checking Gamut 105 Monitoring Audio 114 Monitoring SDI Physical Layer 86 Remote communications 124 Timing a studio 97 ARIB status Main menu 78 Arrowhead composite gamut 109 display 71 display elements of 71 menu 72 Quick Start User Manual Arrowhead display 105
123. ng soft key will overwrite it The item you add will be placed at the location of the soft key you touch Once the item has been added to the User menu the Main menu is displayed To exit the Configure User menu mode select Configure User Menu from the Main menu and clear Conf User On To display your custom menu press CUSTOM on the front panel Once you have completed your User menu save it in a Preset CAUTION User menus are saved as part of a Preset Therefore if you recall a Preset your existing User menu will be overwritten by the recalled Preset Even if the recalled Preset does not have a User menu the existing User menu will be overwritten 525i 59 94 SDI Input A Ref Internal Jitter HPF 10Hz EE Jitter2 HPF EPs Thu May 04 16 34 50 ID Embd PPPP ade SDI Status Cable Loss SD Approx Cable SD Source Level SD Cable Type Belden 8281 Eye Amplitude Eye Risetime Eye Falltime Eye Rise Fall O C S S S l 525i 59 94 SDI Input A Ref Internal Jitter HPF EE Jitter2 HPF P P Thu May 04 16 49 10 ID Embd PPPP SDI Status Cable Loss SD Approx Cable SD Source Level SD Cable Type Belden 8281 Eye Amplitude Eye Risetime Eye Falltime Eye Rise Fall Audio Input Audio Inputs Eye Jitter SDI Status Video Session Audio Session Follows Video and Outputs Quick Start User Manual How to Operate Y
124. nix Checking Chroma Luma Delay Lightning Display Wed May 03 19 00 19 ice orf Linear Run Time Dolby Status uto 525i 59 94 ID Embd PPPP Audio Session SDI Status DIA Data Collect Running 1 d 01 15 21 one SAY Place Err Ancillary Data 5 5 Field Length Err Lightning Line Length Err Stuck Bits Fl AP CRC 589Dh F2 AP CRC Statistics Status Err Secs Err Fields Err Fields RGB Gamut Error 68 3913 4 0324 Cmpst Gamut Error 1559 92966 95 8027 Luma Gamut Error 6 Configure User Menu Input Signal Format 6 3532 3 6397 FF CRC Error 0 0 0000 AP CRC Error 0 0 0000 Standb Help ID Embd PPPP 0 0 O H Gain x1 00 Bars Video Session SDI A Data Collect Running Run Time 1 d 01 18 28 Auto 525i 59 94 352M Payload None SAY Place Err Ancillary Data None Field Length Err Line Length Err Stuck Bits meme Fl AP CRC 589Dh F2 AP CRC 589Dh Statistics Status Err Secs Err Fields Err Fields RGB Gamut Error 68 3913 3 6145 Cmpst Gamut Error 1746 1041385 96 2377 Luma Gamut Error Hi1 View 2 66 3532 3 2626 FF CRC Error 0 0 0 0000 AP CRC Error 0 0 0000 mmi View _ BR 1 view ele Le 4 Bars 75 Composite 103 Checking Chroma Luma Delay Lightning Display 10 Determine where transitions intersect the delay scales and derive the timing error in nanoseconds as deflected from cent
125. o all tiles or the selected tile and define custom Safe Areas Closed Caption Specify closed caption parameters Alarm Setup Specify what type of alarms are monitored and how the alarms are reported Utilities View instrument options and software versions set time and date control LCD and front panel button illumination levels specify network communications parameters and initiate a system software upgrade 48 Quick Start User Manual How to Operate Your Instrument To Use Online Help The waveform monitor online help is both a quick reference to instrument operation and the reference for the details of instrument operation There are two types of help the Main menu Help and context sensitive help The Main menu Help is accessed from the Main menu by touching the Help soft key The Main menu Help contains task oriented guides to using the waveform monitor The context sensitive Help is accessed by touching the icon in the status bar and then selecting the soft key for which you want help When the online help is enabled selecting any soft key on the display will display a help topic To makes changes to instrument settings you must turn off the context sensitive help To display online help Wed Apr 26 183539 1 Display the menu or submenu for which you want to view the online help 2 Press the in the Status Bar When Help mode is enabled the icon changes color to green If the Help mode is not enabled the icon is white
126. ode to Manual or DHCP depending on your network setup Selecting the configuration mode will close the Config Mode submenu Quick Start User Manual Wed Apr 26 18 12 38 525i 59 94 SDI Input A Ref Internal Waveform Vector Picture Lightning OV Gain x1 00 H Mag x1 00 USB Configure Print Config Standby Help Unmount User Menu Close View Hw Sw View Instmt Utilities Version Options Close Config Mode Network Instrument Communications Manual Setup Name wpemiio LS d Close Config Mode Manual DHCP View Hw Sw Config Mode 5 Manual Pe pees LE L EE eee Version wPrewiio Stoney ALES LEEST ALES LLES ET E 51 How to Operate Your Instrument 52 T 8 Touch the Network Setup soft key m f you plan to control the waveform monitor over the network with SNMP commands select Enabled under Remote Control Port m lf you plan to access the waveform monitor over the network using a web browser select Enabled under Remote Web Interface If you can t use DHCP you will have to set the IP Address subnet mask and gateway address network parameters in the Network Setup menu see your LAN administrator for the required values Be sure to use compatible addresses between the PC and the monitor If you will use SNMP commands with the waveform monitor touch the SNMP Setup soft key 525i 59 94 Wed Apr 26 19 28 18 ID Embd
127. of the R Y and B Y color signals Display Elements for SDI Inputs 1 Line When LINE SEL is enabled touch to select the line being viewed 2 Gain x1 00 Touch to select and use the O Line F1 96 Active knob to adjust the Gain setting Gain x1 00 Bars 75 Component Compass z 10 Axis On Rose On 1890 018 Display Elements for Composite Inputs 1 Line When LINE SEL is enabled touch NTSC Tue May 02 10 16 36 ID the readout to select and use the knob to a wera A Metis ech AES A 1 Ref Internal select line being viewed eu Line All 116 Active 2 Gain x1 00 Touch to select and use the knob to adjust the Gain setting 3 Phase Touch to select and use the knob to adjust the Phase setting NOTE You can toggle between adjusting the Phase and Gain by pressing the SELECT button Phase 0 0 O Gain x1 00 Vector Bars 75 1890 019 Quick Start User Manual 63 How to Operate Your Instrument Vector Menu The Vector menu enables you to specify a number of display parameters Elements of the menu for SDI input signals m IQ Axis On When checked adds graticule lines in Component Vector mode for the axis and the Q axis Compass Rose On When checked adds the compass rose graticule for component signals in the Vector mode display Component Composite Toggles the vector scaling and graticules between the Component and Composite signal representations m Bar Target Setting
128. on Default Vert Gain TA Fixed Gain Fixed Gain Fixed Gain Mag Gain state 8 x1 00 x2 00 x5 00 Quick Start User Manual Monitoring the SDI Physical Layer 3 Touch the Vert Var Gain Enable soft key 525i 59 94 Wed May 17 17 17 47 F to enable variable gain adjustment and Ref Internal VITE Mie use the general purpose knob to set the oul 2 034 Ul waveform height to 10 major graticule divisions 4 Use the VERT knob to position the top and bottom of the waveform on graticule lines y Gain x1 03 H Mag x1 00 5 Press the MAIN front panel button 6 Touch the Eye soft key 7 Touch the Cursors soft key The instrument displays the cursors submenu 8 Select the Show Timing soft key to enable display of the voltage cursors Timing cursor T1 is selected by default 9 Use the general purpose knob to position Wed May 17 17 06 04 timing cursor T1 at the crossing of the ra pa VITC Miss rising edge of the Eye waveform and the graticule line two divisions above the bottom of the waveform 10 Select timing cursor T2 11 Use the general purpose knob to position the timing cursor T2 at the crossing of the rising edge of the Eye waveform and the graticule line two divisions below the top of the waveform F 71 0 263 ns Oy Gain x1 02 H Mag x1 00 O72 0 254 ns A hg 517 ns RI AD Aan jik AER aiies aag ming EEN ik KA aoe Quick Start User Manual 95 Monitoring the SDI Physical Layer 96 12 Th
129. onfigure the eye and physical layer settings as described on the following pages The eye and physical layer settings are initially set to FACTORY defaults See page 36 Jo Rename a Preset or Group Displaying the Eye Measurement Menu The following procedures require that you have enabled the Eye measurement and are displaying the Eye soft key menu items 1 Press the Main front panel button 86 Quick Start User Manual Monitoring the SDI Physical Layer 2 Touch the Eye soft key The instrument displays the Eye soft key menu items SDI Status Eye Type Jitter HPF Sweep Eye 10 Hz 1 Line Cursors Setting the Cable Type 1 Touch the Settings soft key See Close Show mime Sette fm ByPass Cable Type page 86 Displaying the Eye Settings Settings Measure SMPTE Limits Equalizer 8281 pz pawie Measurement Menu 2 Touch the Cable Type soft key 3 Select the best match for the type of Close cable connecting the SDI signal you are CanEs measuring IMG 1000 4 Touch the Close Cable Type and Close Settings soft keys Quick Start User Manual 87 Monitoring the SDI Physical Layer Usage Note Most serial receivers specify cable accommodation for 8281 or 1694A Even if you use a different cable you can use the 8281 or 1694A setting and evaluate the loss as if it were that cable type This allows you to compare the specification for the receiver regardless of the actual cable type
130. ot operate this product with covers or panels removed Do Not Operate With Suspected Failures If you suspect that there is damage to this product have it inspected by qualified service personnel Avoid Exposed Circuitry Do not touch exposed connections and components when power is present Quick Start User Manual li General Safety Summary Do Not Operate in Wet Damp Conditions Do Not Operate in an Explosive Atmosphere Keep Product Surfaces Clean and Dry Provide Proper Ventilation Refer to the manual s installation instructions for details on installing the product so it has proper ventilation Terms in this Manual These terms may appear in this manual A WARNING Warning statements identify conditions or practices that could result in injury or loss of life A CAUTION Caution statements identify conditions or practices that could result in damage to this product or other property Symbols and Terms on the Product These terms may appear on the product DANGER indicates an injury hazard immediately accessible as you read the marking m WARNING indicates an injury hazard not immediately accessible as you read the marking CAUTION indicates a hazard to property including the product The following symbols may appear on the product A CAUTION Protective Ground Refer to Manual Earth Terminal iV Quick Start User Manual Environmental Considerations Environmental Considerations This section provides in
131. our Instrument Usage Notes Until you press the icon so it is colored blue touching a soft key may or may not initiate an action For example if the Main menu is displayed touching the Audio soft key will display the Audio menu From the Audio menu touching Audio Settings will display the Audio Settings submenu But touching several of the soft keys in the Audio Settings submenu will not display a lower level submenu though some will This indicates that you cannot select anything below that level You do not have to remove a soft key from the User menu to replace it If you select an existing soft key in the custom menu it will be overwritten by the soft key you add The first time you create a custom menu instruction dialogs will appear to explain how to create a User menu You can turn off the instructions by checking the Don t show this message again checkbox in the dialog You can also turn off the instructions and turn them back on using the Show Instruction Dialogs soft key in the Configure User Menu submenu You can remove a single button from the User menu by using the Remove 1 User Button soft key in the Configure User Menu submenu You can remove all buttons from the User menu by using the Remove 1 User Button soft key in the Configure User Menu submenu Waveform Display The WFM button displays the Waveform WFM display which is the familiar voltage versus time display used to view a waveform You can view the input s
132. ow that represent the knobs on the waveform monitor front panel m Toclose the Remote Application click on the close window button at the top right corner of the Applet window Quick Start User Manual To Upgrade the Waveform Monitor Software To Upgrade the Waveform Monitor Software System Upgrade CONFIG gt Utilities gt System Upgrade is used to upgrade the system software of the waveform monitor The upgrade is performed across an Ethernet network using a PC to transfer the new software to the waveform monitor PC System Requirements The data transfer utility for upgrading the waveform monitor runs on these PC systems Windows 98 Windows NT 4 0 Windows ME Windows 2000 and Windows XP The data transfer utility also requires Winsock 2 0 an Ethernet interface and 128 MB of free disk space be present on the PC system Before You Begin Read the entire procedure before starting the upgrade The time required to complete the upgrade is about 25 minutes CAUTION Once the instrument has started to erase internal flash DO NOT remove power from the instrument If you do so the instrument flash will be corrupted The instrument will have to be sent to a Tektronix factory service center to have the system software restored NOTE If power to the instrument is lost before it begins erasing the internal flash you can restart the software upgrade after the instrument reboots Quick Start User Manual 127 To Upgrade the W
133. plication Get a sat Get a sna Get a sna Get a sna Get a sna tions Instrument C nastics Summary Lo To further extend your control from a remote PC launch the Remote Application from the Web browser The Remote Application reproduces the display and front panel of the waveform monitor on your PC display Running the Remote Application gives you a virtual front panel that gives you complete remote control over the instrument Note that the Web interface operation is optimized for screen resolutions of 1024 x 768 or higher Operation on computers with lower resolution requires scroll bar use to see the entire interface Remote Application Requirements If you install Java on your PC you can launch a Java applet to control the waveform monitor The Java Applet requires m Version 1 41 or later of the Java Run Time Environment JRE from Sun Microsystems installed on your PC To download the JRE Java Run Time Environment plug in from Sun Microsystems download the appropriate code from the URL http java sun com 2se 1 4 Once you download the executable file launch the installation software and follow the instructions Microsoft Internet Explorer 5 0 or later running on a Windows PC m Netscape on Apple Linux or UNIX machines Quick Start User Manual 125 Remote Communications 126 Launching the Remote Application When you launch the Remote Application it is downloaded f
134. r Manual NTSC Cmpst Input A Ref Internal True Peak Close Follows Video Input Dolby 2 Dolby 1 Audio Settings Aux Display Surround Fri May 05 15 05 58 Input AES A AES A AES B Dolby 3 Dolby 4 Phase Pair Dominance Bars 1 amp 2 Ind On Selecting and Monitoring Audio ID Audio Input AES A Linear C zim Embedded Analog A Analog B 12 Surround Filter Audio Input Linear AES A 115 Selecting and Monitoring Audio Checking Audio Level amp Phase 116 The Audio display level meters are vertical bar graphs where the height of the bar indicates the amplitude of the audio program in the corresponding input channel You can change the input and turn on and off the Phase display from the audio pop up menu Other meter characteristics such as ballistics scale units and program test levels are configured through the Audio Settings menu The waveform monitor with optional audio can display phase of one input pair with a Lissajous pattern display and it can display relative phase of all four pairs of inputs with correlation meters It can also display a representation of the surround sound field Check Audio Level 1 Select a tile 525i 59 94 Thu May 18 09 59 48 ID SDI Input B Audio In Dolby 3 Emb 3 amp 4 Ref Internal VITC 05 20 09 03 2 2 Press the AUDIO button to open the Audio display rs Peak Dolby 3 Emb 3 4 D 3 2 L Listen Mode Full 3 Check the level met
135. r can display S251 59 94 aS EEE e een measurements in multiple tiles Rel tere VITC 141112092 simultaneously For example you can ce eee Inputs SDA Data Gollect Running m Running display four Status screens at one time Two or ee al ee 9 displays that cannot appear in more than Reb Gamut Ener Feld Length Err amin one tile at a time are the Audio and DataList 2 a eee Statistics Status j R Err Fields Err Fields displays They can be displayed in only one E ma Sao ipa osa tile at a time FE URC Enar oF ee EDH Error 0 0 0 0000 Changed since reset Audio Session Error Status Log Page 4 of 4 Audio Input Embedded A Data Collect Running Error Status Timecode Date Time Signal Loss Run Time 0 d 00 59 45 Video Input SDI B 19 Apr 06 17 06 46 Audio Input Source Embedded B 19 Apr 06 17 06 46 Analog Output 1 amp 2 1 amp 4 3 4 amp 6 5 6 7 amp 8 7 8 SDI Fmt Auto 1080i 59 94 19 Apr 06 17 06 46 AES B Output 1 2 1 4 3 4 6 5 6 7 3 7 3 Video Input SDI A 13 11 29 13 2 19 Apr 06 17 08 50 Audio Input Source Embedded A 13 11 19 Apr 06 17 08 50 Channel SDI Fmt Auto 525i 59 94 13 11 29 13 2 19 Apr 06 17 08 50 Clip s Over s Mutejs Silence s Peak dBFS y i i 99 0 99 0 99 0 High dBFS 99 0 99 0 99 0 Active bits 2 o o o Smpl Rate Changed since reset You can display four waveform displays at 52515094 hee ete ae the same time with each waveform display pei VITC
136. re Settings j Clear Display On Display On Triggering a Capture Based on Error Events You can set up the waveform monitor to initiate a Capture when certain errors occur This triggered Capture mode requires that the capture type be set to Buffer To initiate a capture based on an error event 5251994 Wed Apr 26 14 28 47 1 Select the tile you want to capture 2 Press the CAPTURE button This displays the Capture menu 3 Press the Settings soft key V Gain x1 00 H Mag x1 00 Live Capture Capture Settings j Clear 5 Display On Display On Buffer Quick Start User Manual 4 Touch Buffer Trace so that Buffer is selected Buffer is displayed in white text 5 Touch the Trigger Types soft key to select the type of error that will trigger the capture Select the trigger type s you want to initiate a capture 6 Touch OK to save your selections 7 Touch Close Settings to remove the submenu 8 To initiate the trigger capture process touch the Run Stop soft key The waveform monitor will perform a buffer capture when one of the selected trigger events Occurs Clearing the Capture Buffer 1 Press the CAPTURE button to display the Capture menu 2 Touch the Clear soft key Quick Start User Manual How to Operate Your Instrument 525i 59 94 Wed Apr 26 17 18 25 ID SDI Input A Ref Internal Hide Menu Embd PPPP Select Trigger Types F Lock Error C EAV Error C SAV Error r
137. re available when ordering an instrument and specifies the models for which the option can be ordered WFM7100 Video Options CPS SD HD Audio Options DS AD DD DDE Eye Jitter and Data Analysis Options EYE PHY DAT WFM7000 Video Options CPS SD HD Audio Options Description Adds support for Composite Analog Monitoring 2 Composite Analog Inputs Adds support for Standard Definition Monitoring 2 SD SDI Inputs Adds support for High Definition inputs 2 HD SDI Inputs Description Adds support for Digital Audio Monitoring in embedded and AES EBU formats Adds support for Analog Audio monitoring and Digital Audio monitoring embedded and AES EBU inputs Adds support for Dolby Digital AC 3 decode Digital Audio embedded and AES EBU inputs and for Analog Audio Monitoring Adds support for Dolby E decode Dolby Digital AC 3 decode Digital Audio embedded and AES EBU inputs and for Analog Audio Monitoring Supports digital audio formats with up to eight channels Description Adds eye pattern and basic jitter measurement The Eye option enables the display of Eye Patterns in 3 Eye 10 Eye SD or 20 Eye HD modes cable length measurements including source signal level and cable loss and Jitter readout Includes the capabilities of Eye option plus advanced physical layer measurements jitter waveform and automated eye measurements Adds data analysis capabilities Allows for logic level
138. reases the time or difficulty of servicing the product THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY OTHER WARRANTIES EXPRESS OR IMPLIED TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE TEKTRONIX RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT SPECIAL INCIDENTAL OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES Table of Contents Table of Contents General Safety SUMMAIY 6 05505 4ton pac cvactagsearaeneun dessus eran scanatos edu auenutnaneoracs suador bas aisantaoreercssanntocnanatar il Environmental OMS ICSF AOS cies erie ere esseeea eens eeerte rises etn e rs reese earn ueeso E EE V AEE S E A E E E EE A E E E E E A PEE E E E a awneninar vi aE AUS EE A EE E A T E A EE E E A eecmapaes vi PC SOM STA O ag ea E E A S deci tees EET TEE ou O T E EEA ERE A viii Conventions Used in this Manual iccincrenniuaimeawaeacaute seamen n aa E E i viii Before Installation ns ctucterssoriucin ctvaiy or earane sonst wrdhsrasecnd abe taltgnnonsepiperarinsnada tannins ana r rnrn err rrr ErP EArEE PEENE PP PEEEEE EEEE EE EEEn rerni 1 Check Package ONC DS asrsiiscraienrciiorinia na EER E aa Ei 1 Standard ACCS SSO NES assioni EE A A AN A
139. rom the waveform monitor and launched There is no software installation required to use the applet other than the Java Run Time Environment noted previously 1 Launch the Web browser See File Edit View Help r page 124 Using the Remote Web O dla pal O Interface Address http 128 181 216 128 Eco Tektronix WFM6100 WFM7100 Remote Interface 2 Once the Remote Interface appears l Run the Remote Application select Run the Remote Application Get a snapshot of Complete Instrument display BMP The Java applet WI be lau nched Get a snapshot of Graticles amp Waveform only BMP Get a snapshot of Java amp Waveform only BMP Get a snapshot of Waveform only BMP Get a snapshot of Java only BMP Instrument Options Event Log Diagnostics Log Diagnostics Summary Log al Local intranet 3 To control the waveform monitor click m vot a EE maltona on the buttons in the Applet Control Dr Input E E window or click on soft keys or labels in the Applet window as though you were touching the display of the waveform monitor ID Embd PPPP PPPP PPPP PPPP 4 For front panel functions click the virtual EES front panel controls to remotely press ome eoe ll oie 5O tae ii Mount Cong ser Menu standby Hep Operating Notes m When adjusting variable settings such as vertical and horizontal position click on the up and down arrows in the Applet Control wind
140. rsion of RGB YPbPr analog pix monitor output Gamut and or line select bright ups optionally displayable Can also be selected to be a loop through of the active SDI input 5 Clock Out Recovered clock output Options Eye Phy only CMPST A and CMPST B composite inputs Supported Reference Signals SDI A SDI B O O r O gt m lt D U lt Q wm ro La co Ss o 2s 4 H JOIO SDI A SDI B SDI OUT 1890 011 Option SD HD LOOP CMPSTA OKO LOOP CMPSTB 1890 009 Option CPS The following tables indicate the supported combinations of input format and reference format An X indicates a tested combination Other combinations of input signal and reference signal within each table may work but are not warranted Operation with an input from one table and a reference from another is not supported for example a 1080i 50 input will not work with 1080i 60 reference Reference Formats 1080sf 25 1080i 50 Input Format PAL PAL 50 Hz X 576i 50 Hz 2 625 2 X 720p 25 Hz X 720p 50 Hz X 720p 25 720p 50 Quick Start User Manual How to Operate Your Instrument Reference Formats Input Format PAL 720p 25 720p 50 1080p 25 1080sf25 1080i 50 1080p 25 Hz3 X X X 1080sf 25 Hz 3 X X X 1080i 50 Hz 3 X X X 1 Available with Opt CPS 2 Available with WFM7100 Opt SD or with WFM6100 Opt SD 3 Available with WFM7000 and WFM7100 Opt HD only Ref
141. s To choose a filter select one of the following filters from the Filter menu m Flat Display with the full available bandwidth Low Pass Display only the low frequency portion of the signal The Components submenu allows you to select filters to be applied to Composite video signals This is useful to isolate a specific characteristic of the input For example to measure amplitude you may want to use a Luma or Low pass filter to remove the high frequency components To choose a filter select one of the following filters from the Filter submenu m Flat Display with the full available bandwidth m Luma Display only the low frequency portion of the signal m Chroma Display only the portion of the signal with frequencies near the color subcarrier For Composite inputs only The Sweep soft key is used to specify whether lines or fields are shown in the Waveform display To set the sweep select on of the following choices from the Sweepsubmenu m 1 Line One horizontal line is displayed Use the Line Select function to choose one line out of a field or frame m 2Line Overlay layout mode only Two consecutive horizontal lines are displayed m 1 Field All lines for one video field are displayed m 2 Field Overlay layout mode only All lines for two video fields are displayed Quick Start User Manual How to Operate Your Instrument Vector Display The VECTOR button calls up the Vector display which provides plots
142. s On the Diamond display monochrome signals appear as vertical lines Nonlinear component processing such as from a gamma corrector that alters white balance can cause deviations along the vertical axis For example the signal from a camera shooting a monochrome chip chart should appear on the Diamond display as a straight vertical line if it is not then the camera is not balanced To isolate gamut brightups try the following m Use LINE SEL mode to select individual lines m Use the PICTURE display to examine the monitor signal with brightups m Use the Area setting to specify a percentage of the screen for which to ignore for gamut violations Quick Start User Manual Checking Gamut Checking Composite Gamut The Arrowhead gamut display plots luminance Y against chrominance C to check if the composite signal adheres to standard gamut NTSC and PAL Arrowhead displays 75 Color bars and indicates the values of the graticule lines The arrowhead shape of the graticule results from overlaying the standard limits for luminance and luminance plus peak chrominance NTSC display PAL display 100 110 120 131 700 930 950 i Alarm level thresholds Alarm level thresholds 230 300 mV C Ampl C Ampl 1890 032 1 Perform the Setting up for Gamut Checks Wed May 03 193330 procedure See page 105 Setting up for a Gamut Checks 2 Compare the signal to the display to determine out of gamut composi
143. s are reported this function is useful as a global mute in the event of multiple alarms that occur at the same time 5 To set the same alarm notification types for any alarm error condition touch the Set All Alarms to This Mask soft Set All Alarms to This Mask key The instrument displays the alarm notification types Select one or more alarm notification types For a description of the alarm notification types access the Online help for the Alarm Setup soft key display the Config menu touch the Online Help soft key and then touch the Alarm Setup soft key 6 To set individual limit alarms for specific error conditions touch the soft key for 7 the appropriate measurement category EE The instrument displays the available oat Spee error conditions that can trigger an alarm ee osea captonspeasota and the alarm notification types 7 Ras Physical Layer AES and Embedded 7 Select the alarm notification types for Save and Cancel and one or more error conditions SDI Input Missing oO oO Oo B C SDI Input Unlocked E g E go 8 Touch the Save and Close soft key to emai am an wa save the alarm settings and return to FF CRC Alarm eM aA the previous screen Touch the Cancel peaa 0 0 0 0 0 Y Chan CRC Error m O oO oO E and Close to cancel changes to the C Chan CRC Error milala lel ee alarm settings and return to the previous We slashed O OU UB Uo C Anc Checksum Error oO oO oO oO oO screen Y Anc Parity Error g o g
144. s digital audio both embedded and AES EBU inputs m Option AD monitors analog audio formats and all the digital audio formats included in Option DS Option DD monitors and decodes Dolby Digital audio AC 3 plus the analog and digital audio formats included in Option AD E Option DDE monitors and decodes Dolby E plus all the audio formats included in Option DD Quick Start User Manual 65 How to Operate Your Instrument Display Elements Level meters C2 A Meter ballistics readout Gtr eak Peak held lt n t indicator Peak level indicator Test level indicator Level meter scale and units Level meter labels Phase correlation meters Element Descriptions Element Level meters Aux Display Audio source Level meter labels Level meter ballistics readouts Axes for phase display Phase correlation meters Test level and Peak program level indicators 66 Phase display Audio source Input EMB A in phase Selected phase pair Axes for phase display Headphone volume adjust and source 1890 021 indicator 1 2 Correlation meter for selected pair In bar warning messages Description Indicate the amplitude of selected audio channels Shows either a plot of one channel versus the other in a pair or plot of surround channels The selected audio input Identifies the signal in each meter bar Displays the selected dynamic response characteristic Shows the orientat
145. se 39 135 Index D DataList display 76 display colors 76 Default Limits EBU 103 RGB Arrowhead menu 72 Diamond menu 70 Default Limits Tek RGB Diamond menu 70 Diagnostics soft key Config menu 48 Diamond menu 70 Diamond display 70 105 checking RGB Gamut 107 construction of the diamond graticule 107 RGB gamut 107 Display ANC Data 77 Arrowhead 71 Bowtie 75 changing views 11 DataList 76 Diamond 70 Eye 82 Jitter 84 Lightning 69 resolution 22 Status 78 Timing 72 DISPLAY button 12 Display connector See XGA Output connector Documentation conventions used in viii purposes of viii Dolby Status Main menu 79 E Error Log Main menu 81 Ethernet connector 24 Eye display 82 F Factory defaults recalling 34 136 Features List of key vi Filter Waveform menu 62 Flat Components submenu 62 Waveform menu 62 FlexPhase Audio menu 69 FlexVu 11 Follows video Audio menu 68 front panel connectors 13 controls 12 Full frame Picture Settings 65 G Gain How to set 28 Gamut Arrowhead display 109 checking 105 checking RGB gamut 107 Diamond display 107 display elements of 70 Gamut checks automating 111 Ground closure connector 23 H Headphone volume adjusting 47 HORIZ knob 13 Horizontal Offset Timing display 72 How to capture the display 40 configure your instrument 48 connect to a network 51 54 create a User menu 58 measure wa
146. shows both the Jitter1 and Jitter2 thermometers Additionally a Cable Loss thermometer indicates the signal loss due to cable length Use the Settings menu in the Eye display to configure the instrument for the cable type and length of your installation With Option Phy installed the SDI Status Display also shows the eye amplitude eye rise time eye fall time and eye rise fall delta measurement values To view the SDI Status display the instrument must have Option Eye or Phy installed There are two Jitter Thermometers in the SDI Status display because there are two independent clock recovery circuits in the instrument and each circuit can be set to a different bandwidth Note that the Eye and the Jitter displays share these two circuits The upper two tiles use clock recovery circuit one and the lower two tiles use clock recovery circuit two This means that any combination of Eye and Jitter displays in the top two tiles will have the same bandwidth but any combination of eye and jitter in the left or right two tiles can have any two bandwidths from the list of possible choices Elements of the SDI Status Display ZER ETT TR Ref ret y 17 f 2 1 Jitter1 and Jitter2 thermometers and PRSS readouts The Jitter Thermometer uses a colored bar to indicate the relationship of the jitter measurement to the Jitter Jitter HPF 10Hz Cable Loss SD Amplitude limit which is user adjustable is Se etre from the Eye or Jitter display rina tds s
147. splay the signal use the Color Space menu to select among the following the choices only available while displaying SDI inputs m YPbPr Displays the input as Luminance Y and color difference Pb Pr components m YRGB Displays the input as Luminance Y Red R Green G and Blue B components m RGB Displays the input as Red R Green G and Blue B components m SDI gt Composite Displays the SDI input as if it has been encoded into composite The sync and burst in this mode are synthetic and convey no information about signal quality When viewing 525 line SDI input as a composite waveform while using line select mode both burst phases may appear when you would expect to see only one This is because the line selection in SDI Mode is an odd even selection while composite signals are normally viewed with a one of four or one of eight line selection To choose how the signal components are displayed in the active tile use the display mode setting to select m Parade has all the components shown one beside the other like cars in a parade m Overlay has all the components drawn at the same location so that they appear one on top of the other The Waveform menu Filter selection allows you to select filters to be applied to SDI video signals This is useful to isolate a specific characteristic of the input For example to measure amplitude you may want to use a Low pass filter to remove the high frequency component
148. t Ref to set the reference to external mode Connected to the house reference Quick Start User Manual 97 Timing a Studio 5 Press MAIN to display the Main pop up TEOR Mon May 15 1518 49 D menu and select the Timing display for Ref NTSC VITC Missing the tile selected in step 1 Input Timing Relative to Reference 6 If only one circle is displayed adjust the timing offset of the black generator Vertical Offset to match the timing to the external 75 lines advanced reference Adjust for a perfect coincidence of the circle around the Horizontal Offset reference target circle turns green at 7 778 us delayed coincidence and null values of the vertical and horizontal timing readouts Relative to Rear Panel Circle will be in the center when input opi ie oe 7 If multiple circles are displayed the 1080p 23 98 Mon May 15 1517 03 D timing is complex and you must choose Ref NTSC e VITC Missing which circle you want The measurement that is closest to zero offset is displayed with emphasis and appears in the Input Timing Relative to Reference Vertical Offset readouts 45 lines delayed NOTE For additional information on complex timing displays see the previous Horizontal Offset discussion for information on this display and 7 792 us delayed its elements See page 73 Timing Displays for Simple Versus Complex Timing Relative to Rear Panel 8 Repeat step 6 or 7 for any other signals Circle wi
149. tal magnification touch the H Mag readout and use the large knob to change the magnification You can also touch one of the Fixed Mag soft keys Select Mag Best View to optimize the magnification to show the blanking interval for two line and two field displays Line the blanking up on the center graticule and then press Mag Best View to expand NOTE You can also change the V Gain and H Mag values by touching the readouts and using the large knob to adjust their values How to Operate Your Instrument Mon Apr 24 17 24 53 ID 9 Embd PPPP VITC 13 20 22 11 2 V Gain x1 00 _ H Mag x1 00 Default METS Fixed Mag Fixed Mag Fixed Mag Fixed Mag Mag Gain Horz Mag Best View x1 00 x10 00 x20 00 x50 00 NOTE If either V Gain or H Mag is selected you can switch to adjusting the other value by pressing the SELECT button Quick Start User Manual 29 How to Operate Your Instrument To set Sweep Mon Apr 24 B 1 Select the tile whose sweep you want to adjust This will restore the menu for the tile if it is not currently displayed Touch the Sweep soft key Select the desired sweep from the submenu The choices available in the WFM display depend on the Overlay Parade mode setting Overlay Choices available are 1 Line 2 Lines 1 Field and 2 Fields O V Gain x1 00 _ H Magzx1 00 Parade Choices available are 1 Line ID Embd PPPP VITC 13 57 11 01 2 f Close or 1 Field Se
150. te O Y C High 120 IRE components noting the following ig m O sou OY High 103 0 Y C Low 24 IRE m Signals exceeding the luminance OY Low 1 0 amplitude gamut extend above the OY Area 0 top horizontal limit top electronic graticule line O Y4C Area 0 m Signals exceeding the luminance plus or minus peak chrominance amplitude gamut extend beyond the upper and lower diagonal limits m The bottom horizontal line shows the npemi a ee minimum allowed luminance level of 7 5 IRE for NTSC and 0 mV for PAL Quick Start User Manual 109 Checking Gamut Usage Notes m Adjusting the threshold level limits See page 112 Adjusting Gamut Limits To automate this check See page 111 Automating Gamut Checks Then enable alarms in Main gt Config gt Alarm Setup Checking Luma Gamut Luma limit thresholds can be configured for identifying luma exceeding threshold limits They apply to both the incoming SDI and to the arrowhead representation of the SDI input as a composite signal The thresholds are affected by your choice of set up or no set up for the Arrowhead display 1 Perform the Setting up for Gamut Checks TEDN Wed May 03 193350 procedure See page 105 Setting up for E Gamut Checks Select Arrowhead as the display 2 Compare the signal to display to Y C High 120 IRE determine out of gamut luminance Note i the following OY High 103 0 gt Y C Low 24 IRE OY Low 1
151. the Source soft key Internal is the default Source location 7 If you want to copy a Preset from the instrument to the USB flash drive touch the Preset shown in the Source Presets list that you wish to copy or touch All to copy all of the Presets or select a group to copy 8 If you want to copy a Preset from the USB flash drive to the instrument touch the source location if more than source is listed Then touch the Preset in the source Presets list that you want to copy to the instrument m Once you touch a Preset label the Source Location and Source Presets labels will change to Destination Location and Destination Presets 9 Select a destination location from the list either Internal or USB depending on the direction you want to copy the Preset m fyou are copying to the USB flash drive you must copy the presets to a folder w Tternal You can create a new folder in which to save the Preset using the New USB Folder soft key Jitte E NOTE The only folders accessible from Copy Presets are the subfolders within the root level folders that the instrument automatically creates Quick Start User Manual How to Operate Your Instrument 33 How to Operate Your Instrument 34 10 Touch the Copy soft key to save the Preset Once the copy is complete the instrument is ready to copy another Preset 11 When you are ready to remove the USB flash drive from the instrument press MAIN
152. the jitter measurement to the Jitter Amplitude limit which is user adjustable 2 Limits Touch any limit readout and use the knob to adjust the limit Select Show Limits from the Settings menu to display these values Option Eye has includes some of these limits Option Phy includes additional limits 3 V Gain amp H Mag Adjust the vertical gain and horizontal magnification with these touchable readouts 4 Eye parameter measurements Option Phy adds on screen display of eye and jitter measurements Option Eye has includes some of these measurements Option Phy includes additional measurements 5 Line Select When in Line Select mode use the knob to select which line is being viewed Quick Start User Manual 525i 59 94 SDI Input A Ref Internal a tae je O Ampl Max 880 imV O Rise Min 400 ps Y Rise Max 1500 ps Fall Min 400 ps O Fall Max 1500 ps Y Delta Max 500 ps SArApI Min 720 mV How to Operate Your Instrument ID Audio In Dolby 1 AES Al 2 DO Line Fi 21 Active el Atr pase s r Ra ee ny gt Ate UH Be DA ees 2 Jitter Amp 2 020 UI Cable Len Max 200 m 7 Cable Loss Max 2 0 dB Launch Ampl Min 720 mV Y Launch Ampl Max 880 mY Vi 0 0 mY 3 O V Gain x1 00 H Mag x1 00 V2 700 0 mv Eye Type Settings Eye fay A 700 0 m EE AAEN ke A rhe bey wae we EOIN w RAT ute i Begs pore Sa ee hARcn js E TERN 3 Eyes Jitter HPF Sweep Cursors 1
153. tilities soft key 4 Touch the Hw Sw Version soft key 525i 59 94 Fri May 12 115019 D Kat akarnal Fide menu ia ee 5 Check the software version listed under the copyright Verify that the Tee Pa Ionut EMB A version number matches the version you installed i Y 192 View Hw Sw View Instmt Set Clock Display Levels Communications System Upgrade Version Options SITIO uw iiu pwa To verify the functional performance of your instrument perform the Incoming Inspection procedure in the WFM6100 WFM7000 and WFM7100 Waveform Monitors Specifications and Performance Verification Technical Reference Tektronix part number 071 1897 XX The manual is available as a PDF file on the Documents CD and can also be downloaded from the Tektronix Web site free of charge Quick Start User Manual 131 To Upgrade the Waveform Monitor Software 132 Quick Start User Manual Specifications Specifications This section provides an abbreviated list of the instrument specifications Characteristic Input Voltage Input Power Frequency Power Consumption typical Serial Digital Video Interface Input Type Input Level Composite Video Interface Maximum Operating Amplitude Absolute Maximum Input Voltage Digital Audio AES Input Characteristics Temperature Operating Non operating Altitude Operating Non operating Pollution Degree Physical Characteristics Dimensions Height Width Depth front to back includi
154. to four tile view press the Display button and select the 4 Views icon NOTE Only one tile can contain an Audio display or DataList display and only one tile can have the Line Select mode enabled If you try to open a second Audio or DataList display a pop up message will appear informing you that a second display cannot be opened and the active tile will shift to the existing display Quick Start User Manual 11 How to Operate Your Instrument Front Panel Controls The waveform monitor uses a touch panel display front panel knobs and buttons to control the waveform monitor N 7 J S WFM CAPTURE Maccan i PICTURE PRESETS IN OUT 4 AUDIO VERT STATUS a g USER VA WA MENUS n t j A Tektronix WJUUUUUUUUULUUUL 1890 001 Button Function SELECT Toggles the selection between two or more items sequences the selected item through a group of items The SELECT button is lit when it is available for use DISPLAY Displays the Views menu allowing you to change the display between one and four tile view and select which tile is displayed in full screen mode WFM Changes the selected tile to display the input signal as a waveform VECTOR Changes the selected tile to display the input signal on a vectorscope screen PICTURE Changes the selected tile to display the input signal as
155. ttings mae aa raat Sweep eke 1 Field For Eye displays the choices available are 1 Line or 1 Field For Jitter displays the choices available are are 1 Line 2 Lines 1 Field and 2 Fields To Use Presets A Preset is a saved version of the instrument settings in a particular configuration Using Presets saves you time because you only need to press a couple of buttons to change a range of instrument settings to meet a specific measurement need Presets are created and recalled using the PRESETS button on the front panel 30 NOTE In addition to instrument settings User menus are stored in Presets This allows you to create Presets that contain User menus tailored to specific applications Quick Start User Manual How to Operate Your Instrument To Save a Preset 1 Setup the instrument as you want it Mon Apr 24 1833 08 VITC 142859022 2 Push the PRESETS button 3 Touch the Settings soft key 4 Touch the Save Preset soft key 5 Select a preset number such as A2 or rent ef C3 in the grid w 6 When prompted to rename select Yes to create a descriptive 11 character name i for the Preset or select No to save the aA iii d N At b Adde mele preset using the default label such as A2 O V Gain x1 00 H Mag x1 00 Save Recall Rename Copy Power up Preset Settings Preset Preset Preset s Preset s Latest uy The setup will be stored for later recall To Recall an Existing Preset
156. uick Start User Manual Luma Gamut Error Y Chan CRC Error C Chan CRC Error Y Anc Checksum Error C Anc Checksum Error Changed since resct Tue May 02 13 13 07 Error Status Log 525i 59 94 SDI Input A Ref Internal Error Status Log Server State Running Audio Input Source AES A Detected Dolby Format None Video Input SDI A SDI Fmt None Detect invalid format Cmpst Fmt Auto None Ref Input Internal Ref Fmt Auto None Audio Input Source none Audio Input Source Embedded A SDI Fmt Auto None SDI Fmt Auto 525i 59 94 Prey Next First Last Reset Stop Timecode 00 00 00 00 0 0000 0 0000 0 0000 0 0000 0 0000 ID Embd PPPP Page 1 of 1 Date 02 May 06 12 52 02 02 May 06 12 52 12 02 May 06 12 52 12 02 May 06 12 52 38 02 May 06 12 52 39 02 May 06 12 52 39 02 May 06 12 52 39 02 May 06 12 52 40 02 May 06 12 52 40 02 May 06 12 52 40 02 May 06 12 52 40 02 May 06 12 52 41 Time 81 How to Operate Your Instrument Audio Session Shows immediate status in the upper portion of the display and shows Statistical overview of the entire session on the lower portion of the display Statistics are controlled by the Run Stop and Reset soft keys Eye Display Pressing the MAIN button calls up the Main pop up menu From the Main pop up menu you can select the Eye display under the Eye Phy heading The Eye pattern d
157. urround sound listening environment To Check Surround Sound 1 Select a tile 2 Press the AUDIO button to open the Audio display 3 Touch Aux Display and then touch Surround 4 Set the Audio Input to match the signal that you are checking 118 Close l Aux Display Surround Audio Settings Aux Display Off iii vwo fr Embedded Analog A Analog B Close Follows Video AES A Input Dolby 1 Dolby 2 Dolby 4 L R Aux Display Phase Pair Dominance Surround Filter Audio Input Audio Settings Surround Bars L amp R Ind On Linear Dolby 1 Quick Start User Manual 5 Select the Dominance Ind On checkbox to display the Dominance Indicator or not as desired 6 Touch the Surround Filter soft key to set the Loudness Filter type preferred A Weighed biases the response towards that of human hearing 7 Use the level bar display to monitor level control 8 Use the surround display to monitor relative loudness of the individual elements rendered in a surround sound listening environment Check the surround sound display for performance parameters and indicators shown right The indicators are described under usage notes below NOTE The Audio Surround Sound display is courtesy of Radio Technische Werkstaetten GmbH amp Co KG RTW of Cologne Germany Quick Start User Manual Selecting and Monitoring Audio Aux Display Phase Pair Dominance Surround Filter Audio Input Audio Sett
158. veforms with cursors 39 operate your waveform monitor 7 print 54 select a measurement 25 select an input 26 set gain and sweep 28 set line select mode 46 set measurement parameters 26 set to factory default settings 30 use online help 49 use presets 30 IN OUT button 13 Inputs How to select 26 Installation 4 BNC connector compatibility 6 Connecting power 4 In a serial video system 4 Line termination requirements 6 IP settings 51 IQ Axis On Vector menu 64 J Java applet accessing remote application 125 Jitter display 84 thermometer 85 L large knob 13 Lightning display 69 102 limits Gamut 105 Line Word DataList display 76 Quick Start User Manual LINE SEL button 12 Line Select mode how to set 46 Live Display On Capture menu 41 Luma Components submenu 62 Luma Chroma delay 102 MAG GAIN button 12 MAIN button 12 Main menu 10 Manuals conventions used in viii purposes of viii Measure display elements of 72 How to with cursors 39 MEASURE button 12 Measurement How to select 25 How to set parameters 26 Menus Audio 68 Config 48 Main 10 Picture 65 Vector 64 Waveform 62 MyMenu how to create 58 N Network Connect to 51 54 IP settings 51 SNMP settings 51 Normal SCH Phase Vector menu 64 O Online help Displaying 49 50 How to use 49 Operation Basic 7 Optional accessories 1 Quick Start User Manual
159. whead displays Setting up for Gamut Checks 1 Connect a signal containing color bar information For set up purposes the use of a color bar signal is suggested but these displays are useful on all signals 2 Press the IN OUT button to display the IN OUT menu 3 Select the input corresponding to the Settings cae T signal connected Quick Start User Manual 105 Checking Gamut 525i 59 Wed May 03 18 55 57 4 Select a tile 525i 59 94 SDI Input A TERS J Ref Internal IRBS True Peak Input EMB A Linear C 0 mY Teme OH Mag x1 00 Video Session Input SDI A Data Collect Running Signal Run Time 1 d 01 10 59 Format Auto 525i 59 94 352M Payload None SAY Place Err Ancillary Data Field Length Err Tek tr on X Line Length Err Stuck Bits Fl AP CRC 589Dh F2 AP CRC Statistics Status Err Secs Err Fields Err Fields RGB Gamut Error 68 3913 4 8095 Cmpst Gamut Error 1297 77286 94 9938 Luma Gamut Error 66 3532 4 3413 FF CRC Error 0 0 0 0000 0 AP CRC Error 0 Overlay Color Space Components Filter Sweep Settings YPbPr YPbPr Flat i Line 5 Press the MAIN button to display the 525i 59 94 Wed May 08 19 0019 D P Sur input A Embd PPPP Main menu Refi Internal vide menu fll 2 0 0000 True Peal 6 From the Main menu select one of the two gamut displays E Diamond use to detect isolate and correct RGB component gamut i
160. wn to contain lead cadmium mercury and hexavalent chromium Quick Start User Manual V Preface Preface The WFM6100 WFM7000 amp WFM7100 Waveform Monitors offer the monitoring capabilities needed in the production post production distribution and transmission of video content in either high definition HD digital formats Standard definition SD digital formats or composite analog formats With the installation of available measurement options the WFM7100 and WFM6100 waveform monitor offers measurement capabilities important in the design installation and maintenance of digital video systems including Eye diagrams with automated measurements jitter and data measurements With available audio monitoring options you can expand the capabilities of any model to monitor both digital and analog audio content The WFM7100 and WFM 7100 also have options for monitoring Dolby audio formats These products combine the best of traditional waveform monitors with the measurement accuracy repeatability and stability achievable with fully digital technology The products that make up the WFM6100 WFM7000 amp WFM7100 family are Model Description WFM7100 HD and SD capable with high performance monitoring and measurement capabilities WFM 7000 HD and SD capable with basic monitoring capabilities WFM6100 SD capable with high performance monitoring and measurement capabilities Key Features The key features for all models include
161. y Tektronix for warranty work may be new or reconditioned to like new performance All replaced parts modules and products become the property of Tektronix In order to obtain service under this warranty Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix with shipping charges prepaid Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located Customer shall be responsible for paying all shipping charges duties taxes and any other charges for products returned to any other locations This warranty shall not apply to any defect failure or damage caused by improper use or improper or inadequate maintenance and care Tektronix shall not be obligated to furnish service under this warranty a to repair damage resulting from attempts by personnel other than Tektronix representatives to install repair or service the product b to repair damage resulting from improper use or connection to incompatible equipment c to repair any damage or malfunction caused by the use of non Tektronix supplies or d to service a product that has been modified or integrated with other products when the effect of such modification or integration inc