Contents

1. EMA 20 Start up manual GALVANIC 6000 7 Once the circuit is established move the active electrode consistently and making circles around its axis following the client s face and the direction of the facial muscles without separating it from the skin Do not go near the eyes or the temples Concentrate on the areas with more dilated pores and an abundance of grease 8 The session must not last more than three or four minutes 9 Finally slowly slide the electrode towards the neck below the ear while slowly reducing the intensity of current down to 0 miliamps with the other hand 10 Afterwards disconnect GALVANIC 6000 and remove the electrodes 11 Later apply the appropriate beauty treatment to reestablish the pH Important notes Do not perform descaling more than once a month If the cutis is extremely greasy it can be carried out a maximum of twice a month More sessions could cause the so called reactional fat secretion the skin feeling excessively cleaned off its grease the sebaceous glands will produce a hypersecretion of sebum causing the oppposite effect to that desired Always end descaling with a pH balancer to avoid a reactional fat secretion BEAUTY TREATMENT OF BLACKHEADS AND IMPURITIES OF THE SKIN This trouble of the sebaceous follicles is characterised by a regional eruption in adolescence possibly caused by a state of sensitivity of the pilosebaceous system due to many possible causes Superfi
2. 89 336 CEE 73 23 CEE MARZO 2003 MARCH 2003 MARS 2003 Nombre del Fabricante Manufacturer s name Nom du fabricant Direcci n del Fabricante Manufacturer s address Adresse du fabricant Sant Mart de l Erm 22 E 08970 Sant Joan Desp BARCELONA SPAIN Electrocosm tico Electrocosmetic Electro cosm tique Declara que el producto Declares that the product D clare que le produit Nombre del producto GALVANIC 6000 Name of the product Nom du produit Marca EMA Brand Marque Cumple con la normativa Complies with the norm Est conforme a la norme CLASE I TIPO BF UNE EN 60601 Seguridad El ctrica Electrical Security S curit lectrique Compatibilidad Electromagn tica Electromagnetic Compatibility Compatibilit lectromagn tique Electroest tica amp Mobiliario CE BIOINSIDE INVESTIGACION Y TECNOLOGIA S L EN 61000 3 2 EN 61000 3 3 EN 50081 1 EN 55011 EN 5002 1 EN 61000 4 2 EN 61000 4 2 EN 61000 4 3 ENV 50204 EN 61000 4 4 EN 61000 4 5 EN 61000 4 6 EN 61000 4 8 EN 61000 4 11 EMA 35 Start up manual GALVANIC 6000 HOW TO REPORT A BREAKDOWN At the expense of the CLIENT Chente sess 20 ca ta EA RA Le EN att ced Pace Identity caidnumber ei sia aa oale Aia Meurthe Mint id CI a ia Tel DEVICE titan dain Serial number Date of Purchase Sold by Date is levine en
3. Always start a treatment with a current inferior to the maximum tolerated until the resistance of the skin has stabilized by itself The current is then progressively increased to the working level Exceding an excessively high current can result in a burn Alkaline burns under the electrodes are very ugly unpleasant and take a long time to heal They can leave a mark for life Worse still are the burns of the bones An excess of current going through the bone can carbonise it A bone burn is very painful and heals slowely EMA 18 Start up manual GALVANIC 6000 At the end of the beauty treatment As we move towards the end of the treatment the current must be gradually reduced to allow the completion of the chemical effects and prevent any chemical waste from remaining on the skin Alternatively reverse the current for the last moments of the treatment to neutralise the chemical waste Finally if the client notices that the treatment produces an unbearable irritation in the skin stop the treatment and reverse the current to neutralise the cause of the irritation EMA 19 Start up manual GALVANIC 6000 FACIAL DESCALING TREATMENT We will begin giving specific instructions for a descaling facial treatment Descaling is advised for excessively greasy skins ca Essentially descaling is an ionisation allowing to 5 KA get rid of the excess of embedded fat This deep A AA cleansing of the skin comes from
4. GALVANIC 6000 USER S MANUAL Start up manual GALVANIC 6000 TABLE OF CONTENTS IMEFO CUCA AAA YE SEY Mareen 5 Safety Standard 5y M cZn r y kiya yi ata ca a ata aha la a a a eta kik a ya ku ia 6 General POINTS R ie en ke NS tien 7 Technical fGatures armar n a a naa IT 8 Protection system against current breaks RS 9 Description OF the CONO Si A mesrine rites 10 ACCESOS A tr a a ananas nn A 12 ADPICAUONGS a aaa 14 Galvanic and microgalvanic Current iran ma aaa pa aaa aaa kk kk versie sens 14 Ionisation TECHNIQUES icc irradia 14 Work intensity and duration of the treatment 17 Safety in the galvanic beauty treatment 18 Limits in the current intensity sia a an A aa Dra 0 ua at ades 18 At the end of the beauty treatment cccccccesseeseceeseeeeeseereeseeseetensaeeanaaers 19 Facial descaling treatment ccccsccecestecseseensseeeseeeenseesseueaseneensneraenenteens 20 Important notes yA a xy say ice ase o atitea 0 agata din DE e E de kw 21 Beauty treatment of blackheads and impurities of the skin 21 Fa cele at O e alo 29 Facial dehydration beauty treatment 23 Dehydration process i ea titi aaa ct a nada a a iai a Ba kik da a ada ahs 23 Natural hydration process seca za aia ba ii 24 Factors contributing to dehydration smantana aa ata air uita aa 24 Treatment of dehydrated skin sine di pia oa a a a ea mia dale ut 25 Preventiv treatment se kii kilsa a aa 25
5. insolubility in cosmetic solvents and its action is therefore limited to its superficial keratolytic effect However there is a method allowing the passage of the sulphur through the folicular canal up to the altered sebaceous gland We will proceed as follows 1 Clean the skin with a lotion not containing any mineral oil 2 Apply vapour with antiseptic aromatic substances rosemary sage etc 3 Apply ionisation as described for descaling but using an electrolyte solution with a predominance of sodium thiosulphate since the thiosulphate anion will penetrate through the negative pole with the client holding the positive pole in his her hand The intensity is comprised between 1 and 2 milliamps 4 Take care not to pass the electrode over the inflamed areas papules pustules The client will notice the characteristic smell of sulphur liberated by ionisation The session must last from 3 to 5 minutes and should take place twice a month at the most When treating impurities ionisation replaces descaling 5 Afterwards extraction of blackheads is carried out as is normal If you EMA 22 Start up manual GALVANIC 6000 want and to complete the effect of the sulphur a face mask can be applied 6 Finally disinfect the area and apply a lotion or azulene cream LIFTING Another treatment where facial ionisation can be used is the facelift cosmetic or electronic For this type of treatment ionising products will be us
6. Dermohydrating refreshing treatment nea eee rara rea anna 26 C ndu ctive excipients Le tia me St am gt kasek k Ke a a a ten ce duel 27 GALVANIC 1000 appropriate device nare ea anna nana eee sana aaa 27 Go ontralindiCcati l58 aa aa ai 28 Products sen i aaa Jr a cine eg Vereen eRe DO na Tan a e Va at 29 Ionising products positive DO Buscadas nent 29 TONISING products n gative polenta dede 29 Product on the passive electrode kk kk kk kk kk kk kk kk kk kk kk kk kk kk ka 29 Bipolar products acre ce Sl sonner e ka ek aw hay at nourrie 30 Start UD v na na E E A ES 31 1 a e La 5 1 y POROS nao un nsc ed concerne samir sonne Cent 31 Possible technical problemS mini daa 33 Guarantee anc SATA AAA A AAA 34 Declaration of conform Yi A 35 Mantitacturers notes deia ks Kl ti KEKE Cela gt le d al 36 deles il 87 Notes and commentaire tina 38 EMA 2 Start up manual GALVANIC 6000 EMA Start up manual GALVANIC 6000 INTRODUCTION Thank you for your trust in us and in the device you just acquired With this device you bring the most advanced technology in electro aesthetics into your center and you introduce a tool that will become soon indispensable to obtain good results This device provides you with an advanced option of cleansing and facial treatment User friendly this device has been conceived to fully satisfy the requirements of the most demanding users This device has been su
7. a voltage of 125 V please inform your supplier at the time of order The inappropriate use not included in this instruction manual from the beauty professional can have undesirable results The manufacturer is not responsible for the results derived from an improper use of the machine EMA 39 Start up manual GALVANIC 6000 SUGGESTIONS Your opinion is important to us You can send your comments and suggestions to get a better service day after day You can write at the following address EMA BIOINSIDE INVESTIGACION Y TECNOLOGIA S L REGISTRATION No B60829793 MAIN OFFICES AND MANUFACTURING Sant Mart de l Erm 22 Poligono Industrial Fontsanta 08970 Sant Joan Despi Barcelona Spain Tel 34 93 477 42 62 Fax 34 93 477 42 90 info ema es BARCELONA DELEGATION C Gerona 156 Bajos 08037 Barcelona Spain Tel 34 93 459 18 54 Fax 34 93 459 11 32 barcelona ema es National number 902 31 51 71 International number 34 93 477 42 63 EMA 40 Start up manual GALVANIC 6000 NOTES AND COMMENTS EMA 41 Start up manual GALVANIC 6000 EMA BIOINSIDE INVESTIGACION Y TECNOLOGIA S L Sant Marti de l Erm 22 Poligono Industrial Fontsanta 08970 Sant Joan Desp Barcelona Spain Tel 34 93 477 42 62 Fax 34 93 477 42 90 info ema es EMA 42
8. phase of treatment Note It is recommendable that the second step of the treatment be of the same or less duration than the first The interval of five minutes between the first step and the second is given as an orientation Consult the manufacturer of the product for any doubt START UP As prior precautions the beauty professional must ensure that the intensity controls are on zero and the electrodes are well cleaned and disinfected Avoid areas of continuity on the client s skin The treatment must focus on the area to treat the ionisation substance desired the polarity to apply or change of polarity intensity and time Prepare the solution before starting the treatment 1 Check that the device is connected 2 Activate the on switch 1 3 Put a gauze filter paper or sponge or apply the product directly on the skin 4 Put the static electrode on top or adjust the electrode holder handles with their corresponding roller accessory if you wish to use the microgalvanic To this end the electrode holders must be connected to the relevant outlets 6 7 and the microgalvanic switch 8 must be on 5 Connect the polarity selected on the relevant switch 9 6 Turn the intensity control of the galvanic current 4 and increase slowly up to the value chosen Maintain the same treatment time and change the polarity half way through the treatment if it is convenient 7 At the end of the treatment slowly reduce the intensity
9. time 17 At the end of the session wash carefully covers and electrodes with soap and water WORK INTENSITY AND DURATION OF THE TREATMENT The manufacturer of the ionising product must indicate the work intensity and the duration of the treatment However here are a few indications 1 In the first place the intensity to apply must be calculated according to the surface of the electrodes that we are going to use and always starting from the basis that the minimum intensity will correspond to 0 02 milliamps per cm of active electrode and the maximum 0 05 milliamps per cm of active electrode so that if an electrode has a surface of 100cm the minimum work intensity with this electrode will be 0 02 by 100 2 milliamps and the maximum 0 05 by 100 5 milliamps These minima and maxima depend on the resistance to the passing of the current specific to every individual 2 Another consideration is the following An ionisation treatment using a current of 1 milliamp for 10 minutes lets penetrate half the number of ions introduced in a 10 minute treatment of 2 milliamps with the same product Although this seems to indicate that the greater the intensity applied the easier it will be to ionise the product we must take into account that there is a limit in the minimum and maximum previously mentioned 3 When using small electrodes for example during facial treatments the current is ten times lower and minimum intensities will be
10. used in case of descaling and facelift With this type of electrodes roller pencil ball etc work intensities vary between1 and 2 5 mA EMA 17 Start up manual GALVANIC 6000 4 The treatment does not lose its efficiency if we work at a lower intensity for a longer time that is to say if we reduce the flow of ions and leave it applied for longer 15 minute treatment with an intensity of 2 milliamps produces the same effect than a 30 minute treatment of 1 milliamp Logically it is necessary to take into account that these considerations are different according to each type of ion since the ionic velocity characteristic of each ion plays a very important role Safety in the galvanic beauty treatment Galvanic beauty treatment is very easy to perform but in the hands of an untrained or incompetent operator it can cause intolerable and permanent damage to the client It is very important to take all the precautions for an effective comfortable and safe treatment Limits in the current intensity 0 3 milliamperes per square centimetre of contact electrode 2 milliamperes per square inch The tota of the current for Facial treatment must not exceed 3 milliamperes Body treatment must not exceed 8 milliamperes With the proper flow of current the client should feel a tickling sensation under the electrodes GALVANIC 6000 has an automatic control circuit for the current limiting it to these values
11. Ras Stamp and Signature REPAIRS UNDER WARRANTY 1 The costs and risks derived from the transport of the device under warranty are to be assumed by the client 2 Those accessories subject to wear and tear and affected by lack of maintenance and cleaning active and passive plates clables electrodes electrode probes probes etc 3 The device under warranty must be send in its original packing 4 Ifthe technician does not find any manufacturing defaults or only natural consequences of wear and tear the costs derived from the revision run at the buyer s expenses even though the device is still under warranty 5 The warranty conditions are fully detailed in this instruction manual EMA 36 Start up manual GALVANIC 6000 WARRANTY KR amp N gt 10 11 12 13 14 15 16 The buyer will verify the correct state of the merchandide within 24 hours after reception The warranty does not cover any anomalies and or damages caused by transport The device remains under warranty until 24 months after purchase Assistence under warranty can be provided from the moment on the manufacturer receives a copies of the installation certificate signed by the client and the salesperson In lack of this document all the repairs and interventions even performed in period of warranty will run at the client s expenses they will not be for free Those accessories subject to natural wear and tear act
12. al movement b The products used in galvanic treatments can have properties applicable in the field of aesthetics As ionising products they can be introduced by ionisation through the positive or negative pole depending on the nature of the product supplied by the manufacturer To avoid mistakes always put the product in the positive red electrode Ionising products through the positive pole Es e If the product is ionised through the positive electrode the inverter 15 must be in the normal position product on the red active electrode Ionising products through the negative pole e If the product is ionised through the negative electrode the inverter 15 must be in the inverted position product on the red active electrode EMA 29 Start up manual GALVANIC 6000 Product on the passive electrode the electrode without product The passive electrode is in this case that which is connected to the black cable There is no need for a treatment product on this electrode since the surface of the body under this electrode does not receive treatment Nevertheless it is necessary to put some solution on this electrode so that the current circulates NEITHER TAP WATER NOR SALT WATER MUST BE PUT IN THIS ELECTRODE These options could produce chemical burns Unfortunately distilled water is not a valid option either since it conducts the current poorly The ideal is to put a chemically neutral aqueous s
13. bject to technical and application tests to guarantee its quality and results Before starting up the device please READ CAREFULLY this MANUAL Failing to do so could turn out to be dangerous and provoke damage to third parties We invite you to contact us or our distributor for any doubt or query We would be delighted to assist you Yours sincerely The Management EMA 4 Start up manual GALVANIC 6000 SAFETY STANDARDS This device complies with the safety standards in force at the time of manufacturing However the following safety recommendations should be observed 1 Before handling the device read carefully this manual 2 The client s electrical safety is ensured by a double network insulation However do not connect to a socket without an earth connection The device must be connected to an earth plate with an earth conductor Never connect to a system without a differential switch 3 Do not use on customers with a pacemaker or any other active implant because of the risk of interference with this device 4 This appliance produces direct galvanic currents always flowing between the point of application and the metal return electrode The heart should not be found on the path of the galvanic current 5 There should be no metal elements e g osseous prosthesis and or cables in the treatment area 6 Before placing the accessories set the power button to zero 7 This device must be used follo
14. cial hyperkeratosis closing the pilosebaceous orifice e Hyperplasia of the stratum corneum surrounding the pilosebaceous follicle e Inflammatory process originated by the intrinsic composition of the blackhead Excessive production of sebum Presence of impurities Presence of staphylococcus albus Bacterial lipases of the previous microorganisms that develop irritating fatty acids e Endocrine factors Sick follicles are characterised by EMA 21 Start up manual GALVANIC 6000 e Folicular canals dilated and full of keratinolipids and microorganisms e Presence of impurities in the skin e Presence of hair making the cleansing of the follicle difficult In this type of follicles two different forms of blackheads can appear e Those that can be expelled to the exterior by mechanical means due to the fact that the mouth of the follicle is dilated pimples e Those that are contained inside the follic and can not be expelled to the exterior sebaceous microcysts Although the doctor must be the one to diagnose and treat the imperfections of the skin especially with respect to the hormonal and antibiotic medication it is well known that beauticians collaborate very effectively in maintaining the skin clean There is a common point between the different cabin treatments the use of keratolytic substances and or regulating the sebaceous gland Among the substances we can mention sulphur used for centuries Its drawback is its
15. ction 3 Plug 4 Potency control for galvanic currents 5 Visual Indicator of the galvanic currents coming through EMA Start up manual GALVANIC 6000 6 Positive outlet red cable for galvanic currents is 2 7 Negative outlet black cable for galvanic currents 8 Line reversing switch for galvanic or microgalvanic current i 9 Choice of galvanic currents a Microgalvanic current m b Galvanic current Indicator of the irregularities in the operation of the device The flickering red light shows that the power control has not been set to zero To carry on with the treatment the beautician must reset the power button EMA 10 Start up manual GALVANIC 6000 REF 5111014 510056 5151060 5112650 5111042 5111038 ACCESSORIES ACCESSORIES DESCRIPTION QTY Indirect aluminium electrode 1 Network cable sisi rn 1 Red Black banana cable c ceeace 1 Aluminium mushroom electrode for descaling 2 Coil microgalvanic movements kez 1 Microgalvanic roller electrode EMA Start up manual GALVANIC 6000 REF DESCRIPTION Un 5111014 Indirect aluminium electrode 1 5151060 Red Black banana cable 1 5111042 Coil micro galvanic 1 movements 5111038 Microgalvanic roller 1 electrode 5112650 Aluminium mushroom 2 electrode for descaling 510056 Network cable 1 EMA 12 Start up ma
16. ed for a cosmetic facelift or electrical current for an electronic one The electrodes vary also needle ball roller etc each one with a specific application FACIAL DEHYDRATION TREATMENT Dehydration process Someone once said that our ageing is due to the inability of living tissues to retain water Indeed with the passing of time the water content in our tissues and organs decreases causing a lack of turgescence and a minor capacity to eliminate toxins and catabolites this is ageing Logically the skin is not excluded from this process and also suffers the consequences of time passing this circumstance being aggravated by its condition as barrier of separation between our body and the hostile and dehydrating external environment However our skin has certain mechanisms that tend to limit the loss of internal water by evaporation and transpiration without which the oss of water through the skin would be massive and fatal These mechanisms are various e Cutaneous lipids Due to their hydrophobia they exercise a physical chemical barrier to the flow of water through the epicutaneous emulsion in both directions e Keratinous layer The special disposition of the horn cells and the characteristics of its membranes as well as the maintenance of the keratin at the most of its isoelectric point help to maintain a water content of between 12 and 15 in this external epidermic area e Natural hydration factor It is formed b
17. example in descaling treatments In these cases an appropriate cleansing milk must be used oil in water emulsion with few fats and electrolytes 3 When preparing the ionising product if it comes in ampoule format break the two ends and pour the content into an appropriate bowl Dilute with distilled water In case of a gel or an electrolytic solution use the quantity indicated by the manufacturer With your fingertips moisten the areas to be treated with the ionising solution particularly for facial treatments 4 If there are injuries erosions spots etc on the application area these parts must be covered with little pieces of paraffin paper before placing the electrodes 5 Place some pieces of double gauze previously cut according to the shape and size of the active electrodes and soaked in the ionising solution on the areas to treat Once the gauze impregnated with the ionizsing solution are placed put the active electrodes on top The electrodes are EMA 14 Start up manual GALVANIC 6000 to be protected with spongy tissue covers 6 These active plates covers will have previously been soaked in distilled water Parasite ions are not introduced with distilled water so the current intensity will only introduce cosmetological ions soaking the gauze In addition since the distilled water is hot it will favour peripheral blood flow and avoid an unpleasant contact with the skin On the other hand the conduction
18. g treatment When an important dehydration takes place because of the unfavourable factors already mentioned we have to act in a more vigorous and remedial way The occlusive and hydrating techniques of the epidermic surface are not enough we have to act at the level of the water reservoir that is to say of subcutaneous tissue We have seen how the skin in senescence is poor in collagen amino acids and EMA 26 Start up manual GALVANIC 6000 mucopolysaccharides and particularly in hyaluronic acid Supplying these elements in depth is made difficult by their molecular complexity it is then when the ionisation techniques can help us to achieve their penetration Indeed using the technique of marked molecules it has been shown that after the ionisation these elements are found in the depths of the skin many of them even forming part of complex molecules collagen fibres and mucopolysaccharides originated from much simpler elements hyaluramine amino acids etc Due to their polyanionic nature many of these molecules get abundantly hydrated causing a greater nutrition of the dermis and favouring the elasticity and the young and hydrated aspect of the skin Conductive excipients Once the ionising active principles have been chosen they must be incorporated to an excipient conducting the electrical current Aqueous solutions Polyethylene glycol creams Carboxymethyl cellulose gels Carbomer gels O A Emulsions
19. ish to include possible improvements in our appliances compel us to reserve the right to modify the characteristics of our devices without prior notice 3 All devices are provided for a mains voltage of 230 volts If you require a voltage of 125V notify it to your provider at the time of the order 4 AS MANUFACTURERS we offer a 24 month guarantee on our devices against any manufacturing fault In order for this guarantee to be efficient the following technical specifications must be taken into account 4 1 Minimum specifications in the electrical installation e The device must always be connected to the electrical current by means of a voltage regulator with ground point whose output voltage must be included between 110 and 120 volts in the case of installations with this voltage particularly in Latin American countries or 220 and 230 volts in European countries e The socket where the device is connected must be in perfect conditions 4 2 This guarantee becomes invalid if A The device is used improperly B The electrical installation or the regulator connected to the device are defective C The device is exposed to overloads short circuits electrical discharges floods or other similar unforeseeable circumstances D The device is handled or repaired by staff non authorised by EMA EMA 34 Start up manual GALVANIC 6000 DECLARACION DE CONFORMIDAD DECLARATION OF CONFORMITY DECLARATION DE CONFORMIT
20. ive and passive plates cables electrodes etc are not included in the warranty Any defect encountered and reported within 30 days after purchase will automatically lead to the replacement of the defectuous article after the reception of the latter Each repair and or replacement will require the presentation of the warranty certificate that comes with the device at the moment of purchase The device must be send to the distributor in its original packing The warranty does not cover those parts damaged because of a negligent use or lack of maintenance consequence of the client s failure to contemplate the instructions Regardless the fact wether it concerns an incorrect maintenance or other circumstances that are not the consequence of manufacturing defaults The manufacturer declines any responsability for damages caused directly or indirectly to persons or objects as a consequence of the failure to comply with the prescriptions indicated in the user s manual and in particular the warnings regarding the use and maintenance of the device The warranty does never under no circumstances cover indemnisations for damages of any nature caused to persons or objects The warranty will be nullified and costs derived from assisting the client will be charged to the latter in the following cases a When the buyer cannot show the certificate of warranty b When the device has been given other uses than the ones it has been designed for c When the serial
21. lied during the day e Indirect action Works on the epidermis protecting it from evaporation by means of covering or occlusive preparations They are unctuous like vaseline type onctions or emulsions of external oil phase with an abundance of non absorbable mineral oils such as paraffin and lipophilic substances that halt dehydration Due to its greasiness it is advised to apply this type of preparation at night although it is not inconvenient to use it during the day on some extremely dry skins Face masks are also mentioned in this chapter since they form an occlusion over the skin in a way that by impeding transpiration the water vapour reverts to the keratinous layer hydrating it at the expense of the water lost by evaporation However this type of cosmetic must not be used for more than 15 days except if there is a risk of excessive maceration of the horn cells with a loss of the protective capacity of the keratine Incorporating hydrating substances in these masks is recommended for example amino acids Sugars urea etc In a same formula both effects can be obtained which is why products perfectly adapted to the needs of each type of skin can be obtained Although principles with a direct action such as water and moisturizers are desirable in every case it is different with oils and other fat substances of indirect action more indicated for the treatment of dry skins than of greasy complexions Dermohydrating refreshin
22. ly make sure that the mains voltage correspond to that of the device If the pilot light is on but the device is not working check with this start up manual that the operation process is the right one If the problem persists consult the technical department 90 OF TECHNICAL PROBLEMS COME FROM A LACK OF ATTENTION WHEN HANDLING THE BUTTONS AND FROM THE NATURAL WEAR OF THE ACCESSORIES For this reason please check all these details before contacting your nearest technical service 2 I discovered that the device does not work properly since yesterday However there has been a storm Storms and rains can sometimes affect national hydroelectric systems If you discover problems with your electrical devices after a storm especially the most sensitive it is advised to connect a current stabiliser to the device This electrical device facilitates the apparatus proper operation by purifying the inconstant electrical signal received by the user and creating a continuous stable signal Sensitive devices can then work properly since they receive the same amount of mains voltage at all time EMA 33 Start up manual GALVANIC 6000 GUARANTEE AND SAFETY 1 The manufacturer is not responsible for an improper use of this device nor for the consequences derived from it Any use not included in this instruction manual can be dangerous We therefore ask you to consult your nearest authorised distribution centre for any doubt 2 Our w
23. nt Before connecting them to the appliance fix the banana extremity to the metallic or rubber plates Very important before connecting the cables to the machine check that it supports the voltage of the current available 125 or 220 volts and that the power control is set to the minimum and to which terminal the positive and negative polarities correspond 10 Connect the cables to the appliance following the instructions of the corresponding technique and turn on the machine with the operating handle EMA 15 Start up manual GALVANIC 6000 11 12 13 14 Slowly turn the control of each outlet to the right until reaching the intensity indicated by the technique Sometimes a burning or tickling sensation may occur with a moderate intensity which can be the result of an inadequate moistening of the sheaths a bad positioning of the plates on the skin or a direct contact between a metallic and the skin Intermittently and during the treatment check the work intensity of the appliance outlet by outlet since the resistance that the body to the passage of the current often reduces and the amperage could go up more than is required This increase in the intensity can also take place after a few minutes of direct current that is when the cosmetologic ions have begun to orientate themselves and move towards the opposite electrode which facilitates the passing of the current reducing the resistance and increasing
24. ntive and repairing dermohydrating ones Preventive treatment Apart from avoiding where possible very dry environments and excessive exposure to the sun as well as contact with detergents and cosmetics the beautician will recommend the use of cosmetic products preventing the lack of water in the skin These hydrating cosmetic products can act in different ways e Direct action On the superficial layers of the skin by means of hydrating preparations containing hydrophilic substances capable of strenghtening the natural hydration factors amino acids such as glycine hydroxyproline arginine etc sugars sodium salt from carboxylic pyrolidin acid sodium or glyceryl lactates urea etc Hygroscopic or moistening substances are often added whose role consists in retaining the water in the stratum corneum the longest possible Many have been tested but the glycerine has been discarded the most effective are Sorbital and mucopolysaccharides EMA 25 Start up manual GALVANIC 6000 among them hyaluronic acid which thanks to its ability to retain water and its affinity with cations helps to maintain the water content in the active principles appropriate for cutaneous dehydration treatment Using zinc and titanium oxyde is also advocated for their covering capacity and protection against solar radiation The excipient of these hydrating creams with direct action is generally an emulsion of external aqueous phase and is app
25. nual GALVANIC 6000 APPLICATIONS These are a few of the many treatments that can be done with the GALVANIC 6000 device e Disinfection of the skin with high frequency after the extraction of impurities e Reaffirming of a double chin and of the facial muscles with the help of passive gymnastics currents e Anti wrinkle treatments lifting of eyes nutrition hydration etc by means of the ionisation of the suitable product and the use of micro galvanic currents e Descaling of the fat of problematic skins with the help of micro galvanic currents thanks to the application of a descaling product and of a metallic electrode in the shape of a mushroom EMA 13 Start up manual GALVANIC 6000 1 GALVANIC AND MICROGALVANIC CURRENT GALVANIC CURRENT Galvanic current is a type of uninterrupted electrical current IONISATION TECHNIQUES In all ionisation techniques we must always consider the following points 1 The person that will receive the treatment must remove all metal objects in direct contact with the skin watches bracelets rings necklaces etc This goes for braces too The beautician should also verify wether the client carries any metallic anticonceptive instrument 2 If the skin to treat is very greasy the grease will have to be removed with a piece of cotton soaked in a neutral soap This will not be necessary in cases where a previous cleansing of the skin has been carried out as occcurs for
26. numbers have been whipes out altered or eroded d When the device has been tampered with even when it concerns only a part or component The costs and risks derived from transporting the device are at the buyer s expenses If the technical team does not encounter any manufacturing defaults all the costs derived from the revision will run at the client s expenses even in period of warranty The warranty conditions con not be modified by any oral or written agreement Only the manufacturer can perform the assistance under warranty The use of a tension stabilizer is compulsory if the tension of the electrical subministration varies 10 from the nominal value The warranty does not cover alterations suffered by the device because of tension pikes or because of the incorrect state of the electrical installation EMA 37 Start up manual GALVANIC 6000 DEVICE DATA MODEL Date of Manufacturing Serial Number Verifyer Date of buying Distributor s stamp EMA 38 Start up manual GALVANIC 6000 MANUFACTURER S NOTE Our wish to offer to the aesthetic sector equipments that keep adapting to the requirements of this young profession leads us to conceive new equipments and to try and improve those already existing For this reason we are compelled to reserve the right to modify these specifications without prior notice All equipments are intended for a mains voltage of 230 V If you require it for
27. of electricity is favoured precisely by this same water heating 7 The sheaths covering the indifferent plates or attracting electrodes have to be soaked with a conducting solution This solution may be tap water due to the salts that it contains However sometimes because of the chemical treatment that the waters of the network undergo they can contain ions that are not too convenient and that may produce alterations or allergies in the skin For this reason it is preferable to use a conductive solution to soak the sponges of these plates or indifferent electrodes for example sodium chloride at 9 x 1 000 i e a physiological solution buffered or not Ringer s solution or better still a special solution composed of salts of weak acids and bases duly buffered 8 When placing the plates active and indifferent electrodes with their corresponding soaked bands must be well fastened over the surface of the skin and in the areas indicated according to the treatment to carry out For a correct fixing elastic bands intended for that purpose must be used perfect adaptation to the relief of the surface of the skin must be ensured making sure that no metallic part is in contact with the skin and that no corner of the electrodes is folded and that there is no space between the skin and the plate 9 Choose a colour of cable for each type of electrode for example as is established conventionally black for the active and red for the indiffere
28. olution remaining neutral even after the current has passed typically a product that is chemically a BUFFER This type of product can be found in gel or solutions Without such products the passive electrode the black one could produce chemical burns The use of a buffer solution is unnecessary logically when working with bipolar products If you do not have the appropriate product for the passive electrode you can work as follows e Put active product ON BOTH ELECTRODES e CARRY OUT THE TREATMENT AS IF IT WERE A BIPOLAR PRODUCT This second option prolongs the time of treatment but only the parts of the body covered by BOTH plates receive the treatment NB The procedure described is not the only one but it is the easiest to use and the least prone to mistakes Bipolar products There are products in which both positive and negative ions can be used In this case the areas of the body under the positive red electrode and the black negative electrode receive the effects of the treatment The procedure to observe is the following e Put product on both electrodes e Start the first phase of treatment with the inverter on position normal e Lower the intensity until reaching zero and wait 5 minutes before starting the second phase see note e Put the inverter 15 in the inverted position EMA 30 Start up manual GALVANIC 6000 e Rise the intensity again to the value selected and start the second
29. r and apply a hydrating cream appropriate to the type of skin being treated This dermohydrating repair treatment must follow the following pattern e Jn cabin One session every 5 or 8 days 1 Frequency One or two treatments per year one of which without fail after the summer e At home Maintenance on alternate days at night alternating with nutritive creams and without forgetting the day protective hydrating base With a galvanic pencil 5 minutes of ionization on alternate days will be enough Apply a mask of amino acids every two weeks In animal experimentation with rat skin the effect of 15 ionisation sessions with a collagen gel amino acids and mucopolysaccharides can be observed Microphoto A cut of skin ofa witness animal without treatment Microphoto B cut of skin of the treated animal We can observe a better turgescence of the tissues and the formation of new fibres EMA 28 Start up manual GALVANIC 6000 CONTRAINDICATIONS Pregnancy even recent Postpartum quarantine Troubles of the skin Cardiovascular anomalies Mammary pathology Precaution in cases of hepatitis diabetes external and internal scars Do not use on people with pacemakers PRODUCTS a The function of the products used to impregnate the plates in passive gymnastics is to improve the electrical contact plate client The product should be in gel format for a more effective transmission of the electric
30. s 140 x 220 x 75 mm Ground connection the device has a ground connection through its network cable Only connect to installation with that kind of connection e Safety the device remains insulated from the network through a coil transformer This symbol aims at dissuading the use of the device by non qualified personnel or newly introduced to its use warning of the possible adverse effects that an improper use could provoke on the client This device provides a special protection rate against electrical discharges particularly regarding the authorised leakage current and the reliability of the earth connection EMA ih Start up manual GALVANIC 6000 PROTECTION SYSTEM AGAINST CURRENT BREAKS This device includes a protection system against current breaks In case of an interruption in the power supply the device will stop Once the current has been restored the protection circuit will be activated The device will then block the outlets interrupting the treatment When this occurs pilot light 5 Will light up In order to continue the treatment put all power controls to zero 4 Once the protection circuit has been disactivated the pilot light 5 will go out and the green light indicating that the device can function 5 will light up again EMA 8 Start up manual GALVANIC 6000 DESCRIPTION OF THE CONTROLS Description of the controls 1 Main on off switch 2 Conne
31. tant especially the summer dry atmosphere high mountain exposure to the sun and wind The dry environment of houses and offices heated in winter or air conditioned is also an important factor of cutaneous dehydration 2 Chemical The skin dries easily with the contact of organic solvents alkaline soaps and detergents normally used EMA 24 Start up manual GALVANIC 6000 a The use of inappropriate cosmetics to clean the skin can lead to the elimination of the hydrolipid film the modification of the permeability of the stratum corneum or the disappearance of natural hydration factors In this case elimination of water is accelerated and the epidermis dries 3 Age With age sebaceous productions decrease the cutaneous biological reactions slow down lowering the rate of natural hydrating factors and the fibroblasts produce less collagen and mucopolysaccharides In short the keratinous layer loses its water retention capacity with age and on the other hand the water coming from inside becomes less and less because of the degeneration of the water reservoir of the subcutaneous tissue 4 Diet Due to pathological factors or reductions Treatment of dehydrated skins This type of skin is mainly characterized by a rough aspect for cracking or scaling easily for being fragile and or irritable not being very elastic and presenting a sensation of tautness As for beauty treatment it is necessary to differentiate the preve
32. the saponification of fatty acids important components of the retained sebum by means of sodium magnesium and potassium ions as well SD GN as anions above all carbonic containing the ionising solution It also enables to reduce the diameter of excessively dilated pores Hovv to proceed 1 Put the ionising solution in an appropriate container 2 Cover the metallic bar or indifferent electrode with cotton soaked in a conductive solution drain it and give it to the client vvho vvill hold it firmiy in his her hand or place the indifferent plate on the arm with a conducting solution or tap water 3 Moisten the clients face with cotton soaked in the ionising solution especially in the areas with a greater quantity of grease and open pores 4 Dampen the active mobile electrode with the descaling solution If using the ball electrode cover it with cotton 5 Once the appliance has been checked in the way described previously connect the cable of the indifferent electrode that is to say the one the client will be holding in his her hand to the positive pole of the apparatus and the cable of the active electrode to the negative pole 6 Place the active electrode on the client s forehead connect the device and turn the potentiometer control to the right until the milliammeter indicates an intensity included between 1 and 2 milliamps without any trouble for the client If the alter tolerates it it can be raised up to 3 milliamps
33. the intensity However the appliances that are available nowadays like GALVANIC 1000 have anticipated this circumstance in such a way that once the work intensity of each of the outlets has been fixed a special mechanism prevents the intensity from varying even though the resistance of the human body decreases This is due to a possible variation of the tension and if we remember the formula T tension intensity R resistance 13 We can perfectly check that when the resistance varies if there is an automatic compensable variation of the voltage the intensity can be maintained constant This however is within some limits and there are some fluctuations of tens of milliamps that can be insignificant Therefore in these cases with these type of appliances constant watch can now be spread out more 14 At the end of the session or when the technique indicates that a change of polarity is necessary before cutting the passing of the current or before working the inverter the intensity of each outlet must be gradually reduced to zero EMA 16 Start up manual GALVANIC 6000 15 Once the treatment is over with the controls to zero the appliance can be disconnected and the electrodes removed 16 Once the electrodes have been removed a slight reddening of the skin may appear in the treated area This is completely normal and shows that there has been a real ionisation that will disappear after a short
34. tioned that ionisation produces a certain effect of hardening in areas close to the anode which is used for firming treatment This phenomenon comes from the change of permeability of the skin as a result of the movement of ions and fluids In the description of descaling and acne treatments the negative ions have been considered as the active ones If on the other hand the active ions of the product used were the positive ones the inverter at the front of the device 8 would have to be pressed EMA 32 Start up manual GALVANIC 6000 POSSIBLE TECHNICAL PROBLEMS 1 The device is not working Why The cable of the accessories may be cut in one of its points or at the extremity of connexion to the device For a proper operation of the device replace the broken part of the cable The device blocks any current output when it detects a leak To extend the lifetime of these accessories it is recommended to place the apparatus at a 909 angle from the working couch to avoid any unnecessary suffering of the cables Also check that the plug is connected properly and that tension reaches the appliance you can check the presence of current with another device Verify the rear connector and the reliability of the mains cable using it to connect another appliance It thus indicates that the problem does not come form the mains cable Once these verifications are made the protection fuse may be examined to see if it has not meted Final
35. to zero and turn the switch 4 to its off position Remove the electrodes and clean them 8 Turn off the main switch 1 IMPORTANT e To avoid mistakes when starting a new session always make sure that the inverter 8 is always off and on its normal position when finishing a treatment e Thoroughly clean all electrodes and covered used in each treatment with soap and water at the end of the session EMA 31 Start up manual GALVANIC 6000 In some cases a certain reddening of the skin can be observed when removing the electrodes This phenomenon known as endosmosis is caused by the movement of fluid particles of the tissues that are concentrated in front of the cathode This reddening is perfectly normal and disappears after a short period of time Likewise due to the movement of the ions the client can feel a sensation of heat in the treated area We must try to reduce this heat production using the whole surface of the electrodes When only a small part of the total surface of the plates is used the heat generated concentrates in that area and there is a risk of burn However when using the whole surface of the electrodes the heat produced is spread throughout the surface of the skin in contact with the plate To make sure that this contact is perfect and that the electrical conduction of the current is optimum the covers and cotton wool that cover and protect the electrodes must always be moistened It has to be men
36. wing the instructions of qualified personnel for example graduated beauticians EMA 5 Start up manual GALVANIC 6000 GENERAL POINTS GALVANIC 6000 is a device that uses galvanic and microgalvanic currents The use of these currents for the penetration of ionic cosmetic substances is traditional in aesthetic devices Thanks to its compact and multifunctional design this device allows to perform a wide range of facial treatments inside a booth or at the client s house needing one device only Because of its small size it allows you to save space in small beauty booths and to carry it along in a Suitcase in case you would like to take it with you WARNING This device is not a medical or sanitary product and it is not aimed at diagnosing controlling monitoring or relieving any disease injury or deficiency Always use cosmetic products complying with the law in force In case of doubts concerning the application on a specific client consult a specialist EMA 6 Start up manual GALVANIC 6000 TECHNICAL FEATURES e This device combines the following functions in one unit galvanic and microgalvanic currents Independent voltage control Galvanic and microgalvanic current indicator User friendly Provided with the necessary accessories for basic functions possibility of adding additional accessories e e Supply voltage 230 V 110V on request Input 125 W Weight 1 4 Kg Dimension
37. with under 10 fat Gelicrem O A emulsion with a gelified aqueous phase GALVANIC 6000 Appropriate device The unit of galvanic and microgalvanic current of the GALVANIC 1000 is the perfect tool for the beauty professional wanting to carry out a deep facial rehydration treatment Technique 1 Clean the face with a makeup remover low in fats 2 Apply water vapour with ozone intermittently for 10 minutes 3 Perform a soft peeling If there are any blackheads extract them a few days preceding the treatment to avoid irritating the skin 4 Moisten the skin with an isotonic conducting solution to increase conductivity 5 Extend the deep hydrating repair cream chosen over the area to treat 6 Slowly slide the roller electrode connected to the negative pole over EMA 27 Start up manual GALVANIC 6000 the treatment area for 15 to 20 minutes with a work intensity included between 1 and 2 miliamps If it is necessary to put product over the skin again stop the supply of the current momentarily 7 Remove the surplus cosmetological product 8 Spray a decongestive lotion made from marigold or chamomile letting it dry over the skin Massage with essential oils since the massage will facilitate the pentration of ions The muscular stimulation can also be reached using passive gymnastics current following the facial muscles 9 Apply a mask with amino acids for 15 or 20 minutes 10 Remove the mask with lukewarm wate
38. y a little known mixture of EMA 23 Start up manual GALVANIC 6000 various substances produced by the cutaneous biological reactions whose mission of retaining water is due to its hydrophylia These substances include amino acids sugars organic acids mineral ions urea etc Fibroblasts These cells situated in the dermis make the collagen and the mucopolysaccharides composed by fibres and the fundamental substance of the subcutaneous tissue For their marked hydrophilia they are an important water reserve with 65 to 70 water Natural hydration process We have seen how the dehydration process of the skin takes place Now let s see what the natural hydration process of the skin is Internal factor Water continually reaches the epidermis from the deepest layers of the skin External factor A percentage of the water lost by sweating is reabsorbed the epidermic water with keratinous structures and components of the natural hydration factor Water absorbed by capillarity in the keratinous layer microstructure called imbibition water that performs a lubricating and plastic role with respect to the keratin and confers a uniform appearance to the epidermis smooth and elastic United water remains relatively fixed in the epidermic structures but the same does not happen with the absorbed water easily evaporated by action of various external factors Factors contributing to dehydration 1 Climatic They are very impor

Download Pdf Manuals

image

Related Search

Related Contents

  Istruzioni per l'uso  DR 2800 - STEP Systems  Manual de instrucciones www.knick.de SE 630 (X) MS    Carrier Compressor and Condensing Unit User's Manual  Kenmore Elite 200 gal. Water Filter Brochure  casiers touts formats, tous modes de fonctionnement  Delta R4700-FLWS Installation Guide  製品カタログPDF - OralStudio  