Home

Honeywell 9782 Series pH/ORP Analyzer/Controller Operator Manual

image

Contents

1. DIAGNOSTICS lt gt Step Screen Action 1 ages os Prepare containers of the two standard reference CALIBRATION solutions MAINTENANCE I O SETUP DIAGNOSTICS 2 P a Go to CALIBRATION BUFFER CAL pH SAMPLE CAL AUTO BUFFER CAL AUTO BUFFER SETUP CAL DIAGNOSTICS DIAGNOSTICS 3 Sue eee e Put the unit in Hold mode BUFFER ATTENTION Remove the electrode from the process WAIT FOR STABLE SSS Se Rinse the electrode thoroughly with distilled or de ionized HOLD INEXT water Immerse it in the buffer solution with a pH closest to the process Go to the next screen 4 The display will show the pH of the buffer solution as STANDARDIZE measured by the electrode system pH VALUE 4 00 Once the reading is stable change the value to match the actual pH of the buffer solution at its current temperature Press ENTER to standardize the instrument adjust instrument zero 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 85 Manual Calibration ORP CAL SAMPLE CAL AUTO BUFFER CAL AUTO BUFFER SETUP CAL DIAGNOSTICS SLOPE FAIL Step Screen Action 5a See ee en If the standardization is successful the completion screen COMPLETED will be displayed and the entered buffer value will be STD
2. Panel Mountin into existing 7075 7076 7077 and 7078 series cutouts t 6 43 0 70 163 32 1 78 S 1 06 025 L 26 92 0 64 25 max L 6 27 040 6 35 159 26 1 02 Customer Panel Panel Mountin 6 43 0 70 163 32 1 78 rd Customer Panel 1 00 025 L 25 40 0 64 12 min 37 max 6 27 040 3 04 9 40 159 26 1 02 Figure 3 2 2 Mounting and Dimensions Non Backlit Display Option 24 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Unpacking Preparation and Mounting 7 21 040 le 6 12 030 Case Outline OSes 2 4 20 Tapped Holes 155 57 0 76 6 00 in rear of case for Square 152 40 mounting bracket d GE 1 12217 A LI 87 D a 2 41 9 _ Dia holes 4 61 21 22 22 for lead wires Terminal Boards accomodates ER conduit by customer 5 516 047 140 11 1 19 i __inches 0 LEGEND Milimeters minspace 687 L 276 _ 4 ween FA NE horizontal 14 70 05 cutouts o 5 516 047 125 M 140 11 SR 3 18R max 2 76 permissibl 70 05 l A 937_ 23 81 Panel Cutout BEEN f o a i min
3. Action Go to CONFIGURATION DISCRETE CONTROL DAT ATTENTION Any value entered for one of these control related parameters will take effect as soon as it is entered UNLESS THE 9782 IS IN HOLD MODE Use this screen to select the DAT control output to be configured The input type relay assignment and appropriate unit of measure will already be filled in based on the process variable assigned to this DAT control point during I O setup This is read only information Enter the setpoint at which the output should be turned on Enter the PB Limit This is the value for which the output s on duration will be maximum Specify the length of the cycle period in seconds for the time proportioned control The cycle period specified for the last DAT control output configured will apply to all DAT outputs Take the unit out of Hold mode 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 49 UO Setup and System Configuration 5 6 Configuring Alarms Purpose If you configured input and action for one or more alarms during I O setup described in 5 3 configure each alarm s setpoint deadband and delay as described below Procedure The procedure for configuring parameters for an alarm is described in Table 5 7 Table 5 7 Procedure for Configuring Alarms Step Screen MENU CONFIGURATION CALIBRATION MAINTENANCE I O SETUP T PREV
4. Option SW2 1 SW2 2 SW2 3 SW2 4 SW2 5 SW2 6 SW2 7 SW2 8 None OFF OFF OFF OFF OFF OFF OFF OFF Single Parameter OFF OFF OFF OFF OFF OFF OFF OFF Current or Voltage Output Auxiliary Relays OFF OFF ON OFF OFF OFF Two Analog ON OFF OFF OFF OFF OFF Outputs Three Analog ON ON OFF OFF OFF OFF Outputs 148 9782 Series pH ORP Analyzer Coniroller Operator s Manual 3 00 Honeywell Sensing and Control Honeywell 11 West Spring Street Freeport IL 61032 70 82 25 73 Rev 5 0300 Printed in USA www honeywell com sensing
5. Press ENTER to standardize the instrument adjust instrument zero 92 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Manual Calibration Step Screen Action 10 SE The completion screen will be displayed COMPLETED STD BUFFER VALUE Exit the menus SAVED DIAGNOSTICS HOLD NEXT 11 Turn off the voltage source and turn off power to the Analyzer Controller Do not open the case until power is disconnected 12 Reconnect field wiring removed in Step 4 13 Replace the safety cover and secure it with the screw 14 Close the case and power up the unit Do not apply power until case is closed 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 93 Manual Calibration 8 6 Calibrating pH Electrodes Using Automatic Buffer Recognition 8 6 1 Introduction Analyzer Controller stores information on multiple buffers Automatic buffer recognition is an advanced feature which simplifies the task of calibrating by the buffering method Note that if the system is configured to include a Durafet electrode a separate temperature compensator is required to use the automatic buffer recognition feature The 9782 Multi Function Analyzer Controller contains in its permanent memory information on several commonly used buffer solutions including the pH versus temperature characteristics of each By command the instrument will automatically sele
6. E Proportional Control Tuning E 1 Overview Adjust effective proportional band width The only adjustment in proportional only control is the effective proportional band width For startup enter the desired control setpoints for discrete control types or the value representing 0 in CAT control Set the proportional band limits value for 100 output to the extreme pH or ORP values anticipated before treatment of the batch Treatment takes too long but no unacceptable overshoot After observing treatment performance some improvement may be possible If treatment takes too long and there has been no unacceptable overshoot beyond the setpoint make one or more the following changes e Increase the reagent delivery capacity e Increase the reagent concentration CAUTION More concentrated reagents freeze at higher temperatures e Decrease the proportional band width by adjusting the proportional band limit closer to the setpoint Unacceptable overshoot If unacceptable overshoot occurs make one or more of the following changes e Increase mixing in the treatment tank e Reduce the reagent delivery capacity e Reduce the reagent concentration CAUTION Sulfuric acid at lower concentrations is more corrosive to metals e Widen the proportional band by adjusting the proportional band limit further from the setpoint 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 145 Proportional Control Tuning 146
7. ATTENTION This equipment contains devices that can be damaged by electrostatic discharge ESD As solid state technology advances and as solid state devices get smaller and smaller they become more and more sensitive to ESD The damage incurred may not cause the device to fail completely but may cause early failure Therefore it is imperative that assemblies containing static sensitive devices be carried in conductive plastic bags When adjusting or performing any work on such assemblies grounded work stations and wrist straps must be used If soldering irons are used they must also be grounded A grounded work station is any conductive or metallic surface connected to an earth ground such as a water pipe with a 1 2 to 1 megohm resistor in series with the ground connection The purpose of the resistor is to current limit an electrostatic discharge and to prevent any shock hazard to the operator The steps indicated above must be followed to prevent damage and or degradation which may be induced by ESD to static sensitive devices ATTENTION The preamplifier power supply at terminals V V and SC provides a nominal unregulated 10 and 10 volts for an external preamp If the 31022283 preamp is used it will perform correctly from this supply although its terminals are designated 16 16 and SC 110 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Maintenance 10 2 Output Calibration Introducti
8. Table 5 9 Procedure for Configuring AutoClean PREV NEXT Step Screen Action 1 EE Go to CONFIGURATION ADVANCED FEATURES CALIBRATION AUTO CLEAN MAINTENANCE I O SETUP DIAGNOSTICS T PRE T NEXT CONFIGURATION CAT RETRANSMISSION ALARMS DISCRETE CONTROL ADVANCED FEATURES DIAGNOSTICS PREV NEXT ADVANCED FEATURES CLOCK SETUP AUTO CLEAN DIAGNOSTICS PREV NEXT 2 AUTO CLEAN Timer setup is used to specify the frequency of the rinses TIMER SETUP ON i e f several rinses are to be performed in a single day MINUTES 30 a specify the number of days as zero CURRENT DAY gt DIAGNOSTICS b specify the number of hours and minutes in the l HOLD interval between rinses e f rinses are to occur at intervals of days a specify the number of days in the interval b specify the time of day hours and minutes the rinse should occur In our example the rinse will occur every third day at 4 30 PM e If you plan to use AutoClean Cal but initiate rinses manually using the real time AutoClean Cal Time Remaining display a specify the number of days as zero b specify the number of hours and minutes in the interval between rinses as zero also IF YOU ENTER ALL ZEROS NO AUTOMATIC RINSES WILL BE PERFORMED Specify where today falls in the AutoClean cycle For example if the interval for rinses is 3 days and you want
9. the first rinse to be tomorrow specify that today is day 2 3 AUTO CERAN Enter the duration of the rinse in seconds GE 180 Enter the number of seconds the Analyzer Controller RESUME DELAY Si should wait after the rinse before resuming normal DIAGNOSTICS operation 54 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 UO Setup and System Configuration Table 5 10 Procedure for Configuring One Point AutoCal Step Screen Action 1 BEE E Go to CONFIGURATION AUTOCAL CLEAN CALIBRATION MAINTENANCE I O SETUP CONFIGURATION CAT RETRANSMISSION ALARMS DISCRETE CONTROL ADVANCED FEATURES DIAGNOSTICS ADVANCED FEATURES CLOCK SETUP AUTO CAL 1PT DIAGNOSTICS NEXT AUTO CAL 1PT ifi i i i i a 2 Een Buffer specification is described in Table 8 6 FUNCTION REQUIRES AUTO BUFFER SETUP IN Go to the next screen CALIBRATION MENU DIAGNOSTICS HOLD NEXT AUTO CAL 1PT i i 3 E EA Specify whether or not output values should be held during the automatic operations Alarms are always held and for that reason HOLD ACTIVE will appear on the alarm stripe DIAGNOSTICS l NEXT Go to the next screen 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 55 UO Setup and System Configuration Step Screen Action 4 AUTO CAL 1PT TIMER SETUP DAYS 3 HOURS 16 MI
10. ONOFF combined input HYPOCHLORITE 6 Nes i SH ANALYZER REDOX Bi HYPOCHLORITE ae TT CYANDE REDOX WASTE CYANIDE TO CYANATE CYANATE TO CARBONATE SETTLNG SLUDGE AND NEUTRALIZATION Figure B 1 Cyanide Treatment System 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 135 Cyanide Waste Treatment B 2 First Stage of Cyanide Destruction Raise pH and oxidize cyanide Sodium hydroxide caustic is used to raise the effluent to about 11 pH which will promote the oxidation reaction and ensure complete treatment The oxidizing agent is usually sodium hypochlorite NaOCl The reaction for the first stage is given below using the NaOCl and with cyanide expressed in ionic form CN The result is sodium cyanate NaCNO and chloride ion CT NaOCI CN NaCNO CT This first stage reaction is analyzed and controlled by independent control loops caustic addition by pH control and oxidizing agent addition by ORP control redox potential or ORP oxidation reduction potential Often an ON OFF type of control using solenoid valves or metering pumps can be used The pH controller simply calls for more caustic whenever pH falls below 11 The ORP controller calls for additional hypochlorite whenever ORP potential falls below about 450 mV The metal ORP electrode is positive with respect to the reference electrode Titration curve The ORP titration curve
11. Output Range Adjustable to any pH ORP TEMP range within the display range Output Signals Proportional to selected output range 9782P __ 00 None 9782P __ VC 0 1 V or 0 10 Vdc or isolated 4 20 mA de 600 ohms maximum load 9782P __ C2 Two isolated 4 20 mA dc 9782P __ C3 Three isolated 4 20 mA dc Standard Alarm Control Relays Two SPDT Form C general purpose relays rated 0 6 amps at 120 240 Vac 0 6 amps at 110 Vdc 2 amps at 30 Vdc Maximum switching power for ac 125 VA resistive for dc 60 W resistive Optional Alarm Control Relays 9782P __ RO Two hermetically sealed Form C rated at 3 amps at 120 Vac 28 Vdc 0 1 V or 0 10 V or isolated 4 20 mA 14 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Specifications and Model Number Control Settings On off period 1 to 1000 seconds On off percent on time 0 to 100 1 resolution Setpoint and proportional band limit ranges 19 99 pH 1999 mV 10 C to 100 C 14 F to 284 F one count resolution DAT cycle period 1 second to 200 seconds PFT maximum frequency 1 to 200 pulses minute PFT pulse width 50 ms compatible with electronic pulse type metering pumps Temperature Compensation Conventional compensation for changing electrode output Nernst response plus selectable solution temperature compensation for high purity water Power Requirements 108 132 216 264 Vac 47 63 Hz 15 VA Memory retained by EEPR
12. setpoint default for Relays 3 and 4 TEMP 1 to 100 C default 1 for Relays 1 and 2 N A for Relays 3 and 4 configuration alarm delay 0 to 100 seconds default 0 configuration cycle timer period 1 to 1000 seconds default 60 configuration configuration cycle timer on time 0 to 100 at 1 resolution default 100 configuration frequency configuration configuration PFT control maximum pulse DAT control cycle period 24 hour clock set mode 1 to 200 pulses second default 60 1 to 200 seconds default 15 OFF ON W RESET ON W O RESET configuration 24 hour clock set hour 0 23 default 00 configuration 24 hour clock set minute 0 59 default OX AutoCal AutoClean timer days 0 to 28 default 0 configuration AutoCal AutoClean timer hours 0 to 23 default 1 configuration AutoCal AutoClean timer minutes 0 to 59 default 0 configuration configuration AutoCal AutoClean rinse duration 1 to 1999 seconds default 60 configuration AutoCal AutoClean output hold YES NO 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 37 UO Setup and System Configuration PARAMETER configuration AutoCal AutoClean buffer flow duration CHOICES or RANGE Default is underlined 1 to 1999 seconds default 200 configuration AutoCal Clean
13. AutoClean and AutoCal Theory and Piping 7 AutoClean and AutoCal Theory and Piping 7 1 Overview Introduction Periodic calibration of pH electrodes is necessary for best system performance because electrode outputs change over time One point calibration standardization is a zero adjustment to compensate for electrode drift Two point calibration standardization and slope adjustment includes a span adjustment to match the gain of the Analyzer Controller to the electrode response Standardization and slope each depend on the electrode measuring a reference solution buffer of known pH The AutoClean and AutoCal features periodically rinse and calibrate pH electrodes automatically as described in this section To take advantage of these features you must e Select them during I O setup see Section 5 e Set the clock and specify the frequency and duration of the cleaning and calibration operations see Section 5 e Configure the system to use automatic buffer recognition see Subsection 8 6 2 e Wire the relays assigned to these operations during system setup to operate the necessary valves See Section 6 for relay wiring e Install piping and valves as diagrammed in this section If manual calibration will be used instead see Section 8 What s in this section The topics in this section are listed below Topic See Page 7 1 Overview 77 7 2 AutoClean Sequence and Piping 78 7 3 A
14. cycle period and on time e PFT Pulse Frequency Type Proportional Only Control using one or two relays The pulse output is generated by repeated relay contact opening and closure The frequency of the pulse is proportional to the deviation from the configurable setpoint The proportional band limit and maximum pulse frequency rate are also configurable This type of output is used to control such devices as pulse type electronic metering pumps e DAT Duration Adjusting Type Proportional Only Control using one or two relays Also known as time proportioning control DAT cycles a relay output on and off varying the on time in proportion to the deviation from setpoint This type of output is used to control devices such as solenoid valves PFT and DAT can use two relays to control two reagents Figure 1 2 diagrams this application 3 00 100 base S addition acid d addition 0 i i i PB1 SP1 SP2 PB2 0 7 14 PH Figure 1 2 Dual DAT or PFT Control Using Two Setpoints and Relays 9782 Series pH ORP Analyzer Controller Operator s Manual 9 Introduction Optional analog outputs for control Current and voltage outputs are available as options These analog outputs can be used to retransmit a process variable using an output signal that is directly proportional to the input This directly proportional analog output can also be used for Current Adjusting Type CAT control Despite
15. mounting assembly Do not connect red wire when using Durafet electrodes For simultaneous pH and ORP see Appendix 3 Analyzer is factory set for 120 V or 240 V operation 9782P 01 00 Preamp Input No Outputs PE Protective Earth Ground Gase as Ee see directions supplied with Analyzer To Preamp x 4 Auxiliary power cable is for output option Note 2 7 N circuit cards Do not attempt to connect to CH input circuit card TH R ORE 5 9782P 03 only Durafet cable receptacle ET W OTH 6 RE SC jumper resistor must remain in place SC k Nc Note 4 peep IEEE pH and ORP 16 B Ovi LA 16 G V S PA OUT O or N oi W BEEN LT Q2 COM Relay 2 olor Code Terminals for A7 os NO Note 1 connection of D A NC Relay 4 K black Y yellow remote O5 COM G ay 7 be O6 NO Note 1 W white A gray preamplifier Lt 28 EN wt Wire eae G green v violet 9 SS H dei 2 A 50 60 Hz R red P pink Note 3 B blue S slate PE EH Power amp N brown T tan Relay Terminals O orange Figure 6 3 Terminal Connections 68 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Input and Output Wiring Case as viewed from front QRE TH I 4 20 mA 600 N max Osc 10V 2000 N min V VCom
16. 1 5 2 Using Relays Relays assigned automatically by priority Every 9782 model includes two internal relays Two more relays are available as an option Many features use relays During system setup the first stage of configuration you choose the features to be used Based on your selections the software will automatically assign relays based on the priority scheme described below You will not be permitted to select features requiring more that the available number of relays Relay priority scheme During I O setup see 5 3 the 9782 s software will assign relays for the features listed below in the order shown The 9782 implements this scheme for you automatically However it is presented here to make you aware of which features use relays and of the trade offs involved when selecting features 1 Using an alternate electrode DualCal has top priority Relay 4 will automatically be assigned to switch between electrodes If the unit has only two relays DualCal can still be used but switchover of electrode input must be done manually see 6 4 2 2 Automatic functions AutoClean and AutoCal have second priority Manual calibration described in Section 8 does not require use of any relays e AutoClean always uses Relay 1 to activate the solenoid valve used to direct washing fluid to the electrode see Section 7 for piping e AutoCal with one reference solution standardization for zero adjustment uses Relay 2 to activate the sol
17. 4 The display will show the pH of the process as measured SEET by the electrode system PROCESS VALUE Collect a beaker of the process sample from a point near NOTE SET PROCESS pH 20 ROVAN Das the electrode mounting and measure its pH value with a calibrated portable instrument For high purity water see special instructions below DIAGNOSTICS lt Once the reading is stable change the displayed value to match the value on the portable meter Press ENTER to standardize the instrument adjust instrument zero 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 87 Manual Calibration Step Screen Action SAMPLE CAL H i H i 5a BEEN If the standardization is successful the completion screen COMPLETED will be displayed STANDARDIZATION VABEB SAVED After successful standardization take the unit out of Hold DIAGNOSTICS mode and return to the calibration menu HOLD NEXT 5b A aie If the attempted standardization is greater than the SAMPLE CAL allowable 5 pH offset the operation will not be AUTO BUFFER CAL i i H i Ge EOE ae USE TOP successful The display will return to the calibration menu CAL DIAGNOSTICS and an error message will be displayed Check the erT LE electrode and replace if necessary Special instructions for high purity water applications For a high purity water application do not remove the sample
18. 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Microprocessor Board Switch Settings F Microprocessor Board Switch Settings F 1 Overview Introduction The information in this appendix will not be needed for normal set up and operation of the 9782 Analyzer Controller It is included here for use during troubleshooting with the telephone assistance of the Honeywell Technical Assistance Center Location of switches Two 8 position DIP switches to be set by the user are mounted on the microprocessor board far left board The switch locations are shown in Figure F 1 witch 1 Switch 2 res SCH Ss CJ 5 On Oe O3 CA O5 Ce 7 CB Oe a CH 7 DOI Figure F 1 Microprocessor Board DIP Switches Note that S1 switch 030480 CRI CRS and CR12 CR16 188056 are not used as of June 1993 SW1 all positions set to off Regardless of the options used all eight switches in SW1 are set to off SW2 settings are dependent on options The settings of the eight switches in SW2 depend on the options used The appropriate settings are shown in Table F 1 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 147 Microprocessor Board Switch Settings Table F 1 Microprocessor Board Switch Settings
19. BUFFER SETUP CAL DIAGNOSTICS 3 S oa eee Put the unit in Hold mode ORP SOLUTION ATTENTION Remove the electrode from the process WAIT FOR STABLE SR Rinse the electrode thoroughly with distilled or de ionized HOLD NEXT water Immerse it in the reference solution Go to the next screen 4 The display will show the oxidation reduction potential of Grae the reference solution as measured by the electrode ORP VALUE 86 0 system Once the reading is stable change the value to match the DIAGNOSTICS actual oxidation reduction potential of the reference solution at its current temperature Press ENTER to standardize the instrument adjust instrument zero 5 EE The completion screen will be displayed Take the unit out of Hold mode and return to the calibration menu 90 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Manual Calibration 8 5 ORP Calibration Using Voltage Input Calibrates Analyzer Controller only The procedure described in this sub section calibrates the Analyzer Controller only It does not involve compensating for electrode drift Instead a known millivolt signal is applied to the 9782 input terminals in place of the signal from the electrode and the 9782 is adjusted so that its reading matches the known input ATTENTION This procedure can only be used when measuring ORP only with a 9782P 02 not simultaneous measurement with glass pH Materials The materials required to calibra
20. C Chrome Waste Treatment C 1 Use of Chromates Corrosion inhibition Chromates are used as corrosion inhibitors in cooling towers and in metal finishing operations including bright dip conversion coating and chrome plating Necessity for removal of chromium ion from wastewater The wastewater form rinse tanks dumps and cooling tower blowdown contains toxic soluble chromium ion CT which must be removed typically to a level less than 0 5 ppm before discharge Technique for chrome removal The technique most often used for this chrome removal is a two stage chemical treatment process The first stage lowers the pH and adds the reducing agent to convert the chrome from soluble Cr to Cr The second stage neutralizes the wastewater forming insoluble chromium hydroxide which can then be removed Consistent treatment and stable control in this type of process requires well mixed reaction tanks with enough volume for adequate retention time see Figure C 1 Retention time is calculated by dividing the filled or usable tank volume by the waste flowrate Typically it is ten minutes or more acD l EE Ex E NOTE The seperate pH and ORP u gt 6 measurements and control shown 5 E in the first stage may be handled po CAUSTIC 5 with a single 9782 Analyzer with Soi 16 E combined input SE Z BISULFITE pH ANALYZER REDOX CONTROLLER CH
21. H Side View Wall Mounting Panel Mounting Be 4 A d 6 63 i 168 40 6 43 0 70 f 163 32 1 78 3 31 84 07 t t 1 21 025 I lt gt 4 Heed 30 73 0 64 195 07 1 52 3 04 9 40 159 26 1 02 Figure 3 3 2 Mounting and Dimensions Backlit Display Option 26 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 4 Power Wiring 4 1 Overview Introduction Power Wiring This section contains instructions for installing ac power wiring for the Analyzer Controller in preparation for configuring the software and performing I O setup as described in Section 5 We recommend that you wait to install input and output wiring see Section 6 until after I O setup During I O setup the software will determine for you which relay to use for each feature What s in this section The topics in this section are listed below Topic See Page 4 1 Overview 27 4 2 General Wiring Practices 28 4 3 Power Wiring Considerations 29 4 4 Installing Power Wiring 30 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 27 A A Power Wiring 4 2 General Wiring Practices WARNING Wiring should be performed by qualified personnel only Safety precaution WARNING A disconnect switch must be installed to break all current carrying conductors Turn off power before working on conductors Failure to observe this precaution may result in serio
22. INPUT 2 N A DIAGNOSTICS NEXT Action Go to I O SETUP The first screen shows the present use of available relays This information is read only In this example the 9782 has four relays Relay 1 and Relay 2 are used for AutoClean and one point AutoCal standardization Relay 3 is used to trigger an annunciator if process variable pH goes over the alarm setpoint Relay 4 is used for a low alarm on the temperature value If the 9782 has only two relays the screen will show Relay 1 twice This permits display of alarm assignments even in the case where two alarms are assigned to Relay 1 Go to the next screen The next screen shows the present assignments for the optional current or voltage outputs if available This information is read only Go to the next screen I O setup is performed only once before field connections are made to the Analyzer Controller Go to the next screen Specify the type of input s to be used NOTE Set INPUT1 to GLASS if using Durafet Il electrode with Cap Adapter For regular Durafet with external or internal preamp input set to DURAFET If you specify any inputs other than the defaults the unit will immediately perform a cold reset returning all configuration and calibration values to the factory defaults A cold reset will also return some I O setup values to factory settings If you change the input type be sure to cycle through all the other I
23. Manual 3 00 Maintenance temperature compensation coefficient experimentally by measuring the pH of the process liquid at several different temperatures as described in Table 10 10 Once you have determined the appropriate value for your application enter it in memory as described in Table 10 11 Table 10 9 Solution Temperature Compensation Coefficients for Selected Treatment Types at Specified Temperatures Temperature Range Treatment Type Degrees Celsius All Volatile Pure Water Treatment AVT Amine Phosphate Oxygenated AVT Treatment 0 to 25 ong om 0035 0 037 0 037 15 to 25 0 017 0 034 0 033 0 034 0 035 25 to 50 o014 ons 0027 0 028 0 029 15 to 40 0 016 0 031 0 030 0 032 0 032 0 to 50 ann 0 033 om 0033 om Temperature range typical for most power plant sampling panels Determining coefficient experimentally Use the procedure in Table 10 10 for determining a solution temperature coefficient experimentally Table 10 10 Procedure for Determining a Solution Temperature Coefficient Experimentally Step Action 1 Using the procedure in Table 10 11 enter a value of zero for solution temperature coefficient 2 Immerse the electrode with temperature compensator in a sample of the process fluid 3 Warm or cool the sample to several different temperatures within the operating range Hold the temperature constant at each point for at lea
24. O setup screens and check that the settings are appropriate for your application Go to the next screen 40 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 UO Setup and System Configuration CONTROL ALARM ACTION RLY1 RLY2 RLY3 RLY4 pH ei ORP TEMP ALAD PFTI ALM3 ALM4 HIGH N A LOW _ HIGH DIAGNOSTICS CG Step Screen Action 6 EEN Specify the automatic functions to be used SELECT ACAL 2PT Note that if AutoCal is used AutoClean is always also used too DIAGNOSTICS EXT Go to the next screen 7 DISCRETE CONTROL If discrete control using relays is planned enter the control S type ON OFF with cycle timers Pulse Frequency Type or SES Ka Duration Adjusting Type DIAGNOSTICS Go to the next screen NEXT 8 ANALOG OUTPUTS If the hardware supports analog outputs specify which E i process variables will be retransmitted pH ORP or Geer ORP TEMP or used for Current Adjusting Type control using an OUTPUT3 TEMP output signal that is directly proportional to the input signal DIAGNOSTICS PCAT for pH and OCAT for ORP e Any output can be used for retransmission e Only Outputs 1 and 2 can be used for CAT control e A unit having only one output can use it for retransmission CAT control of a single variable or DualCAT control of two reagents Go
25. Operator s Manual 121 Maintenance 10 10 Determining and Entering a Solution Temperature Coefficient 10 10 1 Introduction Automatic temperature compensation Corrective adjustment or compensation is needed for the effects of temperature on pH measurement This is done automatically by the Analyzer Controller which offers two types of automatic temperature compensation They are referred to as conventional electrode temperature compensation Nernstian and solution temperature compensation Neither type of compensation applies to ORP measurement Conventional electrode temperature compensation Nernst Response Conventional temperature compensation is included in most pH instruments Compensation is made for the fact that the measuring electrode response is proportional to temperature as well as pH The electrode output is greater at higher temperature The Analyzer Controller uses information from a temperature sensor in the electrode assembly to adjust the measured value for this effect Conventional compensation is always active and requires no setup Note that the Durafet electrode preamplifier performs the basic temperature compensation Solution temperature compensation This type of compensation adjusts for the fact the actual pH of a solution will change if its temperature changes The change is so slight for normal applications it is usually ignored but for high purity water measurements it is significant By means of a sp
26. Output Wiring Thermistor Thermistor Reference Shield Note 5 pH electrode Terminals for connection of direct electrode input Note 2 Thermistor Thermistor Reference Shield Note 5 pH electrode Terminals for connection of direct electrode input Note 2 PE Protective Earth Ground Case as viewed from front Case as viewed from front Relay Terminals option PCA NO com l Relay 3 NC Note 4 RE NO TH COM Relay 4 SC Nc Note 4 Ov PA NSi Relay 2 NO Note 1 Neo Relay 1 NO Note 1 L1 K 7120V 240V N 15A 50 60 Hz Note 3 a PE Power amp Relay Terminals 9782P 02 R2 Direct GlassElectrode Input No Outputs Auxiliary Relay Option Installed Relay Terminals option PCA NO COM Relay 3 Nc JNote4 Se Relay 4 NC Note 4 4 20 mA 600_ 11 max 10V 2000 min VCom 1V 200 Nmin Relay 2 COM H NO Note 1 NC es 1 D SC Note 1 L1 K 120V 240V L2 N W 115A 50 60 Hz Note 3 lkololNIoNA Et Voltage amp Current L PE Output Terminals Power amp Relay Terminals 9782 02 RO Direct Electrode Input Isolated Current amp Voltage Outputs Auxiliary Relay Option Installed NOTES 1 Relay contacts max ratings Resistive Load Current 0 6 A at 120 240 Vac 0 6 A at 110 Vac 2 A at 30 Vdc Voltage 240 Vac Switching Power 125 VA ac 60 W dc See electrode mounting diagram Connections are shown for single pH measurement For OR
27. Series pH ORP Analyzer Controller Operator s Manual 43 UO Setup and System Configuration Table 5 3 Procedure for Configuring CAT Control and or Retransmission of Process Variables Step Screen Action 1 OER SS Sak oe Go to CONFIGURATION CAT RETRANSMISSION CALIBRATION MAINTENANCE I O SETUP DIAGNOSTICS PRE T NEXT EE ATTENTION Any value entered for one of these Ae Ds parameters will take effect as soon as it is entered AUTO CAL V CLEAN UNLESS THE 9782 IS IN HOLD MODE DIAGNOSTICS 2 EE What configurable items are displayed depends on what CAT 2 analog output type was specified during I O setup OUTPUT 3 e A retransmitted variable will be indicated here as DIAGNOSTICS OUTPUT n e CAT based on pH PCAT or ORP OCAT will be CAT RETRANSMISSION indicated here as CAT n E e DualCAT 2CAT is only available if the unit has only one analog output so this choice is always displayed alone DIAGNOSTICS Select the output to be configured I CAT RETRANSMISSION DE ENABLED DIAGNOSTICS 3 EE SC For each output to which a retransmitted process variable 0 1 pH was assigned during I O setup the PV will be displayed 100 7 i i A Ee e and the appropriate unit of measure will be filled in This FAILOVER VAL 9 pH is read only information DIAGNOSTICS HOLDI lt l Specify the range by e
28. adjusted to match the known pH value for the buffer stored in the 9782 memory see Note above The instrument will be a ey standardized instrument zero adjusted If you see a stable reading before 50 seconds have elapsed you can avoid waiting by pressing ENTER 5a ten Can Stee reset If the standardization is successful the completion screen COMP LETED will be displayed STD BUFFER VALUE at After successful standardization go the to the next screen DIAGNOSTICS described in Step 6 HOLD NEXT 5b Reet If the attempted standardization is greater than the SAMPLE allowable 5 pH offset the operation will not be Ee successful The display will return to the calibration menu CAL DIAGNOSTICS and an error message will be displayed OFFSET gt SpPH l l Check the electrode and replace if necessary 6 AV OBUP PERT CSL Leave the unit in Hold mode Rinse the electrode thoroughly with distilled or de ionized water Immerse it in the second buffer solution Go to the next screen 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 97 Manual Calibration Step Screen Action 7 BEE Go to the next screen SLOPE VALUE IS ENTERED DIAGNOSTICS HOLD NEXT 8 BUFFER VALUE HHH HH The display will show the pH of the buffer solution Within fifty seconds the reading will be automatically adjusted to match the known pH value for the buffer s
29. alarm Use the maintenance menu to specify relay action see 10 12 If set to de energize on alarm this means that when an alarm occurs or the discrete control point becomes active the relay coil will be de energized The NC contacts will then be closed and the NO contacts will be open Conversely during normal non alarm operation or when the control point is not active the NC contacts will be open and the NO contacts will be closed If de energize on alarm is selected a power loss will force all relays to the same position as an alarm condition The relay contact maximum ratings are in Table 6 3 Table 6 3 Relay Contact Maximum Ratings Current Standard 0 6 amps at 120 240 Vdc 0 6 amps at 110 Vdc 2 amps at 30 Vdc Current Optional 3 120 Vac 28 Vdc Voltage Standard 120 240 Vac Switching Power Standard ac 125 VA resistive dc 60 watts resistive ATTENTION Alarm circuits are not internally fused in the Analyzer Controller Provision for fuses in external circuits is recommended Procedure Follow the procedure in Table 6 4 to install relay output wiring If the RC arc suppression circuits in the Analyzer Controller interfere with other devices at your site refer to 6 4 2 for instructions for disconnecting the suppression circuits WARNING While the unit is powered a potentially lethal shock hazard exists inside the case Do not open the case while the unit is powered More than
30. and remove the input wiring from the preamp at the 9782 input terminals 5 Feeding the test wiring through the conduit hole in the case connect a voltage supply to the PR and SE input terminals see Figure 6 4 e To apply a signal in the range 0 to 1999 mV connect the plus to R and the minus to E e To apply a signal in the range 1 to 1999 mV connect the plus to E and the minus to R 6 Close the case and power up the unit Do not apply power until the case is closed 7 ENT Go to CALIBRATION ORP CAL CONFIGURATION CALIBRATION MAINTENANCE I O SETUP DIAGNOSTICS pH BUFFER CAL ORP CAL SAMPLE CAL AUTO BUFFER CAL AUTO BUFFER SETUP CAL DIAGNOSTICS 8 GRE CAD Put the unit in Hold mode PUT ELECTRODE IN ORP SOLUTION ATTENTION Ignore the instructions to put the electrode in the WAIT FOR STABIE H ne reference solution Instead apply an appropriate DIAGNOSTICS millivolt signal between 1999 and 1999 mV to the HOLD NExT input terminals To obtain a negative value you must reverse the input to the unit as described in Step 5 Go to the next screen 9 The display will show the an oxidation reduction SE eee potential in millivolts The value should match the input ORP VALUE 68 0 signal Once the reading is stable if it does not match the input DIAGNOSTICS signal change the value to match the voltage being applied to the input terminal
31. by telephone at the numbers listed below Organization Phone Number United States and Canada Honeywell 1 800 423 9883 Tech Support 1 888 423 9883 Q amp A Faxback TACFACS 1 800 525 7439 Service Asia Pacific Honeywell Asia Pacific 852 2829 8298 Hong Kong Europe Honeywell PACE Brussels Belgium 32 2 728 2111 Latin America Honeywell Sunrise Florida U S A 854 845 2600 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual Symbol Definitions The following table lists those symbols used in this document to denote certain conditions Symbol Definition This CAUTION symbol on the equipment refers the user to the Product Manual for additional information This symbol appears next to required information in the manual WARNING PERSONAL INJURY Risk of electrical shock This symbol warns the user of a potential shock hazard where HAZARDOUS LIVE voltages greater than 30 Vrms 42 4 Vpeak or 60 Vdc may be accessible Failure to comply with these instructions could result in death or serious injury gt P Protective Earth PE terminal Provided for connection of the protective earth green or green yellow supply system conductor iv 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Contents Te INFRODUCTION nan eee Sanane tentent in geesde 1 1 1 DESCHptiOn sn RE AE 1 1 2 attente dE Mn dr an let deet 2 1 3 Operating the Analyzer Controller ss 4 1 4 Ment Eerarchye s i
32. in Figure B 2 shows the entire millivolt range if cyanide is treated as a batch For continuous treatment operation is maintained in the oxidized positive region of the curve near the 450 mV setpoint The ORP setpoint can vary between installations depending upon pH the oxidizing agent the presence of various metals in solution and the type of reference electrode used Determine the exact setpoint empirically at that potential where all the cyanide has been oxidized without excess hypochlorite feed This point can be verified with a sensitive colorimetric test kit or similar check for cyanide pH 10 5 REDOX POTENTIAL mV VOLUME OF HYPOCHLORITE ADDED Figure B 2 First Stage Cyanide Oxidation Typical Titration Curve 136 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Cyanide Waste Treatment Importance of pH control As shown in Figure B 2 pH has a direct effect on the ORP potential and must be closely controlled to achieve consistent ORP control especially if hypochlorite is used as the oxidizing agent Hypochlorite raises pH which lowers the ORP potential which in turn calls for additional hypochlorite a runaway situation To avoid this situation use close pH control and locate the ORP electrode at a distance from the hypochlorite addition point Reliable measurement with gold electrode For this application a gold ORP electrode gives a more reliable measurement than does a plat
33. label to perform keys the action 4 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Introduction Selecting a parameter for edit To select a parameter for edit e display the screen containing the parameter e use the UP or DOWN keys to highlight the parameter name e press the ENTER key to highlight the displayed current value Editing a parameter assignment from a list of available choices To edit a parameter having a text string as an assigned value e select the parameter and highlight its current value as described above e use the UP or DOWN keys to display the other valid choices e when your choice is displayed press ENTER Editing a parameter numeric value To edit a parameter having a numeric value e select the parameter and highlight its current value as described above the lt and gt soft keys will be displayed During calibration the lt and soft keys will be displayed once the UP or DOWN key has been pressed e use the lt or soft keys to move the cursor to the digit to be changed e use the UP and DOWN keys to increment or decrement the value e when all digits necessary have been changed press ENTER Entering and retrieving data During I O setup configuration and calibration while the unit is either retrieving data from memory or entering a new data value into memory a time period will elapse until this action is complete For the duration of this time period which can
34. limit PARAMETER standard CAT 1 or CAT 2 setpoint standard CAT 1 or CAT 2 DualCAT pH CAT1 or CAT 2 UO Setup and System Configuration CHOICES or RANGE Default is underlined pH 2 00 to 14 00 default 0 ORP 1999 to 1999 mV default 0 TEMP 10 to 140 1 C default 0 PH 2 00 to 14 00 default 14 00 ORP 1999 to 1999 mV default 1999 TEMP 10 to 140 1 C default 140 2 00 to 14 00 default 0 configuration DualCAT pH CAT1 or CAT 2 proportional band limit configuration configuration configuration configuration configuration DualCAT pH switchover value alarm setpoint value cycle timer setpoint PFT control setpoint DAT control setpoint 2 00 to 14 00 default 14 2 00 to 14 00 default 0 for Relays 2 and 4 14 00 for Relays 1 and 3 pH 19 99 to 19 99 default 0 for Relays 2 and 4 14 00 for Relays 1 and 3 ORP 1999 to 1999 mV default 0 for Relays 2 and 4 1999 for Relays 1 and 3 TEMP 1 to 100 C default 0 for Relays 2 and 4 140 for Relays 1 and 3 configuration configuration configuration alarm deadband value cycle timer deadband upper limit PFT control proportional band limit pH and ORP only DAT control proportional band limit pH and ORP only pH 0 01 to 19 99 default 0 01 for Relays 1 and 2 setpoint default for Relays 3 and 4 ORP 1999 to 1999 mV default 1 for Relays 1 and 2
35. messages displayed on the alarm stripe are related to normal operation of the Analyzer Controller These messages shown in Table 9 1 alert the operator that a special feature is in use or that a special mode is active Table 9 1 System Status Messages MESSAGE AUTOSEQUENCE WHEN DISPLAYED This message indicates that AutoClean or AutoCal operations are active To see how much time is remaining use the DISPLAY key to access the special AutoClean Cal screen available during these operations HOLD ACTIVE This message indicates that the unit is in Hold mode maintaining a constant output and alarm status and disabling solution temperature compensation if used The unit can be put in Hold mode and taken out of Hold using the soft key available on the appropriate screens During AutoClean and AutoCal operations the alarms are always held Hold of outputs is configurable Whether or not outputs are being held when an automatic sequence is active the message will be displayed The Hold is released at the end of the operation RETRIEVING DATA ENTERING DATA This message is displayed during I O setup configuration and calibration while the unit processes data from its memory This message is displayed during I O setup configuration and calibration while the unit processes data entered by the user 102 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Diagnost
36. of the case The panel may be up to 3 8 in thick Cutouts for adjacent 9782 Analyzer Controllers may be no closer than 0 687 in horizontally and 0 937 in vertically resulting in a nominal edge to edge bezel separation of 0 078 in Note that if the panel already has a cutout for a 7082 instrument the same cutout can be used for the 9782 If the panel has a cutout for a 7070 7075 7076 7077 or 7078 instrument a Reducer Panel Kit is available for mounting the 9782 The part number is in Section 11 Wall and pipe mounting A bracket is available as an option for wall or pipe mounting the unit as shown in Figures 3 2 1 3 2 2 3 3 1 and 3 3 2 The mounting kit part numbers are in Section 11 ATTENTION Pipe mounting is not recommended if the pipe is subject to severe vibration Excessive vibration may affect system performance 22 9782 Series pH ORP Analyzer Coniroller Operator s Manual 3 00 Unpacking Preparation and Mounting 7 00 040 R le 6 12 030 Case Outline E 2 14 20 Tapped Holes 155 57 0 76 6 00 in rear of case for Square 152 40 mounting bracket L J O 81 OOO 4 122 17 Ei je 2 41 1 37 87 Dia holes 4 61 21 34 92 22 22 for lead wires Terminal Boards L
37. offset back to 0 0 and the percent theoretical slope back to 100 use the procedure described in 10 3 Procedure To view the zero offset and percent theoretical slope being used for the measurement system follow the procedure in Table 8 9 Note that percent theoretical slope applies only to pH calibration not ORP Table 8 9 Procedure for Viewing System Calibration Values Step Screen Action ENU 1 eg Go to CALIBRATION CAL DIAGNOSTICS CALIBRATION MAINTENANCE I O SETUP DIAGNOSTICS BUFFER CAL SAMPLE CAL AUTO BUFFER CAL AUTO BUFFER SETUP DUAL CAL ELECTRODE CAL DIAGNOSTICS DIAGNOSTICS 2 View the values currently being used This is read only CAL DIAGNOSTICS information ZERO OFFSET SS SS SLOPE VS NOTE The display shown is for pH only measurement For ORP only measurement only zero offset is displayed ee For simultaneous pH and ORP measurement an T NEXT additional zero offset for ORP is displayed below the pH offset and slope 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 99 Manual Calibration 8 8 Recommendations for Successful Measurement and Calibration Selection and care of electrode system essential Successful measurements and calibration depend upon selection and care of the electrode system Always prepare electrodes and their mountings in accordance with the instructions supplied with them ob
38. one switch may be required to disconnect power 72 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 A 3 00 Step Input and Output Wiring Table 6 4 Procedure for Wiring Relays Action If you have not already noted the relay assignments made during I IO setup go to I O SETUP on the main menu to view the display showing relay output use Note the assignments shown You must wire the unit to match these assignments in order for the unit to work as expected See Section 5 Turn off the power to the Analyzer Controller More than one switch may be required to remove power With the power off open the case e Grasp the bottom center portion of the front bezel and pull it downward and toward you slightly to disengage the bottom of the bezel from the edge of the case e Lift the bezel gently to disengage it from the top edge of the case e Swing the bezel to the left The bezel and display assembly is mounted on pivot arms The safety cover with its warning label will be visible With the power off remove the safety cover by removing the single screw holding it in place Wire the relays in accordance with the system setup assignments Refer to the wiring diagram provided with the external device receiving the output and to Figure 6 2 through Figure 6 6 WARNING An external disconnect switch is required for any hazardous voltage connections to the relay ou
39. procedure in Table 4 1 to install AC power wiring A Turn power off at mains before installing AC power wiring Table 4 1 Procedure for Installing AC Power Wiring Step Action 1 Check the tag on the outside of the case to be sure that the voltage rating of the unit matches the input voltage at your site ATTENTION The unit may be damaged if you apply power with the wrong voltage 2 Open the case e Grasp the bottom center portion of the front bezel and pull it downward and toward you slightly to disengage the bottom of the bezel from the edge of the case e Lift the bezel gently to disengage it from the top edge of the case e Swing the bezel to the left The bezel and display assembly is mounted on pivot arms The safety cover with its warning label will be visible 3 Remove the safety cover by removing the single screw holding it in place 4 Install a fused disconnect switch in the power line which will be connected to the Analyzer Controller elf a 230 240 Vac line is to be connected use a 0 125 amp fuse elf a 110 120 Vac line is to be connected use a 0 25 amp fuse 5 To ensure that the unit meets the immunity levels specified by EMC directive 89 336 EEC install power line filter supplied with Analyzer Controller Wind three turns of each wire 14 AWG maximum through the filter core for the two power and protective earth leads as illustrated in Figure 4 1 The wound filter cores shall be located external to the
40. to the next screen 9 CONTROL ALARM INPUTS If you specified a type of discrete control earlier use this ns y screen to specify the process variables inputs on which PFTI Qu the control action will be based ALM3 ORP ALMA a ue If you have any relays left for alarms use this screen to NEX specify what will be alarmed Whether relays energize or de energize on alarm is configurable using the Maintenance Menu see Section 10 Go to the next screen 10 URT O If the unit has only two relays and none are being used for AutoClean AutoCal or discrete control then a screen is RLY1 RLY2 available for assigning alarms to relays A parallel screen A1 2 ALM3 P is never displayed for units with four relays because the DIAGNOSTICS EXT mapping of an alarm to an available relay is fixed Go to the next screen 11 This screen shows the control or alarm input and alarm number if applicable associated with each relay using earlier screens Use this screen to configure the action high or low associated with each alarm Note that relays assignments for discrete control AutoClean and AutoCal cannot be changed here Go to the next screen 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 41 UO Setup and System Configuration PRESS ENTER TO RETURN TO MAIN MENU DIAGNOSTICS NEXT Step Screen
41. when process flow is blocked Minimize liquid volume in system Keep pipe sizes small and couplings close to minimize the liquid volume in the system Smaller volumes require less buffer solution and less time to rinse and to calibrate 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 81 AutoClean and AutoCal Theory and Piping Items outside this area provided by user a LL E buffer buffer SC 1 solution 2 9782 gt lt gt lt eh Selten deeg _ to process oer Ho At ei L to drain me gt ry gt gt lt _inse water S1 Two Point Calibration Alarms men Relay 1 Activated Rinse Duration ae ie Duration Relay 2 Activated Slope Duration Relay 3 Activated Same Period as Standardize Resume Delay Time seconds T an T nee I Interval Between 5 Cleaning Calibration days or hours Two Point AutoCal Operation One point AutoCal and AutoClean operations omit steps using Relay 3 and Relays 2 amp 3 respectively Figure 7 3 Two Point AutoCal Operation 82 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Manual Calibration 8 Manual Calibration 8 1 Overview Introduction Calibration of pH or ORP measuring instruments 1s necessary because similar electrodes may produce slight
42. 1 Shield Note 4 NO G green V violet pH electrode L1 K 20v 240v L2 N W 15A R red P pink 50 60 Hz B blue S slate Note 3 Terminals for N brown T tan connection of PE Power amp O orange direct electrode Relay Terminals input Note 2 9782 02 00 Direct Electrode Input No Outputs PE Protective Earth Ground Figure 6 4 Terminal Connections 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 69 Input and Output Wiring Relay Terminals Case as viewed from front option PCA To Preamp Ge NO Relay 3 nes 2 CG H S ch Note 4 TH R QRE ROH TT NO Relay 4 ET W OTH HE NC Note 4 Seah DE EN 4 20 mA 600 CLmax EI OV 10V 2000 Nmin 16 G Qv VCom OUT O or N etc 1V 200 N min D NC Relay 2 Se COM Note 1 Terminals for NO connection of GER k Relay 1 remote NO Note 1 preamplifier L1 K 120V 240V L2N W T 35 a O p p 50 60 Hz Note 3 NOTES Voltage amp Current uji PE Power amp 1 Relay contacts max ratings Resistive Load Current 0 6 A at 120 240 Vac 0 6 A at 110 Vac 2 A at 30 Vdc Voltage 240 Vac Switching Power 125VA ac 60 W dc 9782P 01 RO Preamp Input See diagram supplied with electrode Isolated Current and Voltage Outputs mounting assembly Do not connect red Auxil
43. 1 Procedure for Calibrating Outputs ss 112 Table 10 2 Procedure for Restore Factory Calibration Values 114 Table 10 3 Procedure for Specifying a Tagname or Other Display String 115 Table 10 4 Procedure for Viewing Product Information and Changing the Stored Model Number 116 Table 10 5 Procedure for Adjusting the Screen Contrast ss 118 Table 10 6 Procedure for Entering a Password 119 Table 10 7 Procedure for Resetting All Configuration to Factory Settings 120 Table 10 8 Procedure for Switching Between Electrodes with DualCal 121 Table 10 9 Solution Temperature Compensation Coefficients for Selected Treatment Types at Specified Temp ratures ested haat cota site eee eo ee a ee ae ae 123 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual ix Table 10 10 Procedure for Determining a Solution Temperature Coefficient Experimentally 123 Table 10 11 Procedure for Entering a Solution Temperature Coefficient 124 Table 10 12 Procedure for Turning on Temperature Display 125 Table 10 13 Procedure for Specifying Relay Activation 126 Table 10 14 Procedure for Replacing the Fuse ss 128 Table 1 Lt accueil EE Sen 130 Table F 1 Microprocessor Board Switch Settings 0 0 0 eeceeseesseesseceseeeseecseecssecesecssaecaecaecnseesseessaeenaeens 148 x 9782 Series pH ORP Analyzer Coniroller Operator s Manual 3 00
44. 2 75 accomodates conduit 69 85 by customer 5 516 047 La 77n 44 o 140 11 119 LEGEND inches Se millimeters min space 2 76 between P re horizontal 168 70 05 cutouts T 5 516 47 125 ry 140 11 Sg 18 R max u 2 76 permissi 70 05 Y Y 937 23 81 A Panel Cutout Ge min space between vertical cutouts 7 43 070 x G 7 68 060 gt 188 72 1 78 195 07 1 52 h f O il 3 06 O Ql 77 80 at IR GH EE Jh Lot JI KR el 1 00 IPS Pipe 25 40 by customer Note 1 Do not exceed 80 Ib in torque Horizontal Rear Pipe Mounting when tightening fastners Figure 3 2 1 Mounting and Dimensions Non Backlit Display Option 3 00 9782 Series pH ORP Analyzer Coniroller Operator s Manual 23 Unpacking Preparation and Mounting Vertical Rear Pipe Mounting LEGEND millimeters inches 3 06 7 43 070 188 72 1 78 ES g E 1 00 IPS Pipe 25 40 by customer 77 80 Y CO Note 1 Do not exceed 80 Ib in torque when tightening fastners OC DO e J 7 68 060 195 07 1 52 Top View le 7 43 070 188 72 1 78 __ 1 50 AT 38 10 3 00 T LAN Four holes in bracket or 1 4 dia mounting bolts bolts by customer R H Side View Wall Mounting 4 4 A 6 63 168 40 IT 3 31 84 07 L A w 7 68 060 195 07 1 52
45. 9782 case within 25 cm 6 Connect the power wiring to terminals L1 and L2 N as shown in Figure 4 1 Connect the hot side of the line to terminal L1 and the grounded side of the line to terminal L2 N If the L2 N side of the line cannot be grounded add an isolating transformer electrically between the power supply and the instrument and ground the L2 N terminal WARNING The ground terminal green screw must be connected to a reliable earth ground for proper operation and to comply with OSHA and other safety codes If metal conduit is used connect a bonding wire between conduits Do not rely upon the conductive coating of the instrument case to provide this connection Failure to observe this precaution may result in serious personal injury 7 Replace the safety cover and secure it with the screw 8 Close the case and power up the unit Do not apply power until the case is closed gt Et 30 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Power Wiring PP I Power Terminations Li L2 N Se E DJL PE PE Protective Earth Ground Figure 4 1 AC Power Terminals 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 31 Power Wiring 32 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 UO Setup and System Configuration 5 I O Setup and System Configuration 5 1 Over
46. AY specify which electrode is being used if 10 9 calibration values for two electrodes are stored DualCal enter a temperature coefficient if solution temperature compensation feature is used turn Durafet temperature display on and off RELAY ACTIVATION UO SETUP 3 00 9782 Series pH ORP Analyzer Coniroller Operator s Manual specify whether relays should be activated or de activated on alarm or when discrete control output is on choose features make relay and analog output assignments Introduction 1 5 Planning 1 5 1 Overview Configuration system minimizes decisions The 9782 was designed for easy configuration using menu displays and the keys on the front panel Numeric values such as setpoints and range limits are easy to enter using the UP and DOWN keys and function keys see 1 3 Other configuration parameter assignments are selected by scrolling through a list of available choices using the UP and DOWN keys During configuration only those parameters and choices relevant to features supported by your Analyzer Controller s hardware will be available for configuration For example if the model purchased does not contain hardware for the optional analog outputs then none of the screens relating to output configuration will be applicable Although configuration is a simple process to make best use of the 9782 Analyzer Controller and its features a little advance planning is recommended
47. Action 12 NEW RELAY SETTINGS The screen will display the new relay assignments reflecting the choices you made earlier This information RLY1 RLY2 RLY3 RLY4 S pH pH ORP TEMP IS read only ALM1 PFT1 ALM3 ALM4 HIGH N A LOW HIGH j j j SR Make a note ofit so that you can wire the relay terminals NEXT correctly see Section 6 Go to the next screen 13 NEW The screen will display the new current or voltage output PORT ee assignments reflecting the choices you made earlier This OUIPUIZ RP information is read only OUTPUT3 TEMP Make a note of it so that you can wire the output terminals DIAGNOSTICS NEXT correctly see Section 6 Go to the next screen 14 Pressing ENTER will do a warm reset and will also store ENTER TO CONTINUE these changes in memory overwriting any previous I O setup DIAGNOSTICS I NEXT 15 Press ENTER again to complete the operation and return the display to the main menu For Durafet II electrodes with a standard preamp input if you want to display process temperature or take advantage of the Durafet electrode s ability to compensate for pH changes in a solution due to temperature changes then a separate temperature input is required It is not specified here because temperature input uses special input terminals However the temperature input must be wired as shown in section 6 In addition to use the solution temperature compensation feature a solution temperature coefficient must be entered as descr
48. Analyzer Controller Operator s Manual 67 Input and Output Wiring No as viewed from front To Preamp Option Card St Note 2 iti position K zl OUT 1 Output Board H R QRE 31 OUT2 LET I s a Connections ae OTH 4 20 mA 600 2 max Q4 OUT 2 4 20 mA C K NSSE I Q5 our 3 mA 10V 2000 0 min DS our 3 e B V VCom 16 G Q v HV 200 9 min OUT O or N PA om NC Oil gf AR Lie a Terminals for 3 3 erminals or 2A NC Note 1 Oil connection of Ge COM Gs one Relay 1 remote e N E ifi A JE L1 K 1 Note 1 Gi preamplifier D Ga Cam w 120V 240V e Microprocessor La EE CH 3 J 15A Gal Sam e 50 60 Hz CS ae PE Note 3 Note 5 Voltage amp Current Output Connections Output Terminals Power amp for 9782P 1 C2 or Relay Terminals 9782 11 C3 9782P 01 VC Preamp Input 9782P 03 VC Direct DURAFET Electrode Input Isolated Current and Voltage Outputs Not for Durafet Electrodes with Cap Adapter NOTES Relay contacts max ratings Resistive Load Current 0 6 A at 120 240 Vac 0 6 A at 110 Vdc 2A at 30 Vde Voltage 240 Vac Switching Power 125VA ac 60 W dc 2 See Diagram supplied with electrode
49. BUFFER VALUE saved SAVED DIAGNOSTICS After successful standardization go the to the next screen HOLD NEXT described in Step 6 5b e If the attempted standardization is greater than the SAMPLE CAL allowable 5 pH offset the operation will not be Se cope oe successful The display will return to the calibration menu CAL DIAGNOSTICS and an error message will be displayed OFF SET gt 5pH l l Check the electrode and replace if necessary 6 Le Gee Leave the unit in Hold mode gt 2pH UNITS AWAY FROM STD BUFFER Go to the next screen DIAGNOSTICS HOLD NEXT 7 EE Rinse the electrode thoroughly with distilled or de ionized BUFFER water ATTENTION Fe aie Immerse it in the second buffer solution DIAGNOSTICS HOLD NEXT Go to the next screen 8 The display will show the pH of the buffer solution as SLOPE measured by the electrode system pH VALUE 9 00 Once the reading is stable change the value to match the actual pH of the buffer solution at its current temperature DIAGNOSTICS el Press ENTER to make the slope adjustment adjust instrument gain 9a gee E If the slope adjustment is successful the completion screen will be displayed SLOPE BUFFER VALUE ve Take the unit out of Hold mode and return to the DIAGNOSTICS calibration menu HOLD INEXT 9b BEE If the slope adjustment is not successful the display will return to the menu automatically and an error message will be displayed Possible causes are e The difference in pH between the tw
50. ECURITY RESET UNIT DIAGNOSTICS 116 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Maintenance 9782X XXXX XXXXXXXXX ATTENTION RESET UNIT FOR EDITS TO BECOME ACTIVE DIAGNOSTICS RESET Step Screen tion CER 2 PRODUCT UINEOEMETEON The software version is read only information VERSION A1 0 MODEL NUMBER If the model number displayed no longer matches the PR D Ee hardware because you have added or changed an option DIAGNOSTICS card use the specified function key to edit the model SRE number See Section 2 for available model number selections 3 FACTORY PASSWORD Note that a special password is required to change the 0000 model number This is not the same as the security ENTER PASSWORD password described in 10 7 Enter it when prompted DIAGNOSTICS 4 MODED NUVRPR ATTENTION Changing the model number stored in memory will do a cold reset of the unit returning ALL configuration values to the factory settings shown in Table 5 1 During a cold reset the standardization offset is also returned to 0 0 and the percent theoretical slope is set back to 100 0 eliminating any changes made during manual or automatic calibration To reset the unit use the specified function key To cancel the edit and avoid resetting the unit press the MENU key If you do not have the factory pa
51. Figures Figure 1 1 9782 Front Panel 2 282 Zeen 3 Figure 1 2 Dual DAT or PFT Control Using Two Setpoints and Relays eee eeeesseceseceseceeeeeseceteeeeaeees 9 Figure 1 3 DualCAT Control Using a Single Analog Output 10 Figure 3 1 Sample Nameplates iranien einen ech il ced tienda ACEN 21 Figure 3 2 1 Mounting and Dimensions Non Backlit Display Option 20 0 0 cee eeeesseceseceseeeneeenseceseeenees 23 Figure 3 3 1 Mounting and Dimensions Backlit Display Option 25 Pigure 4 1 gt AC Power Terminals eae nm gerett sated dE ees pane eg ahaa desi 31 Figure 6 1 Dual Reagent CAT Typical Wiring Using Relay 1 to Select Reagent Feeder 0sissessneeneee 66 Figure 6 2 Terminal Connections smart ar met in tirs nn Re nent 67 Figure 6 3 Terminal Connections summer NEEN EES ENKEN 68 Figure 6 4 Terminal Connections sise 69 Figure 6 5 Terminal Connections sosser eie eae Ea ea a TAa Ea ra Ea aE SA 70 Peure G p TermnalhtC onnectons vz eege ios oth n e aa a a a ner 71 Figure 6 7 Backup Electrode Installation Using DualCal Alternate Electrode Calibration 0 74 Figure 6 8 Two Tank Batch Treatment with DualCal Alternate Electrode Calibration eee 75 Figure 7 1 Automatic Electrode Wash Setup 78 Figure 7 2 Rinse and One Pont Calibration ss 81 Figure 7 3 Two Point AutoCal Operaton ss 82 Figure 10 1 Location of Power Fuses tegen dE ee eene ee 127 Figure A 1 Resistance vs Temperature Characteristics of the Temperat
52. Honeywell 9782 Series pH ORP Analyzer Controller Operator s Manual 70 82 25 73 Rev 5 3 00 Sensing and Control Copyright Notices and Trademarks Printed in U S A Copyright 2000 by Honeywell Revision 5 March 30 2000 WARRANTY REMEDY Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship Contact your local sales office for warranty information If warranted goods are returned to Honeywell during the period of coverage Honeywell will repair or replace without charge those items it finds defective The foregoing is Buyer s sole remedy and is in lieu of all other warranties expressed or implied including those of merchantability and fitness for a particular purpose Specifications may change without notice The information we supply is believed to be accurate and reliable as of this printing However we assume no responsibility for its use While we provide application assistance personally through our literature and the Honeywell web site it is up to the customer to determine the suitability of the product in the application CE CONFORMITY This product is in conformance with the protection requirements of the following European Council Directives 89 336 EEC the Electromagnetic Compatibility Directive and 73 23 EEC the Low Voltage Directive Conformance of this product with any other CE Mark Directive s shall not be assumed ATTENTION The emissi
53. KEYBOARD TEST KEYBOARD TEST OUTPUT TESTS OUTPUT CALIBRATION CALIBRATION RESET DIAGNOSTICS 4 EE Rare ULES Press any key except MENU to display the name of PRESS A KEY the key TO TEST MENU TO ABORT Use MENU to exit DIAGNOSTICS I T 106 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 9 5 2 Output Tests Procedure Diagnostics and Messages Use the procedure in Table 9 5 to test the relays and analog outputs Testing the analog outputs requires ammeters or voltmeters as appropriate d A WARNING While the unit is powered a potentially lethal shock hazard exists inside the case Do not open the case while the unit is powered Do not access the output terminal as described below while the unit is powered More than one disconnect switch may be required to remove power CAUTION Consider the effects on your process before testing relays with controlled devices active Table 9 5 Procedure for Testing Relays and Analog Outputs Step Screen AA Action Turn off the power to the Analyzer Controller More than one switch may be required to remove the power With the power off open the case e Grasp the bottom center portion of the front bezel and pull it downward and toward you slightly to disengage the bottom of the bezel from the edge of the case e Lift the bezel gently to disengage it from the top edge of the case e Swing the bezel to the left The bezel and display
54. NEXT normal operation here as the duration of the flow of slope buffer solution past the electrode read only information Enter the number of seconds the Analyzer Controller should wait after the AutoCal operation before resuming 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 59 UO Setup and System Configuration 60 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Input and Output Wiring 6 Input and Output Wiring 6 1 Overview Introduction This section contains instructions for installing input and output wiring for the Analyzer Controller We recommend that you wait to install input and output wiring until after I O setup see Section 5 During I O setup the software will determine for you which relay to use for each feature What s in this section The topics in this section are listed below Topic See Page 6 1 Overview 61 6 2 General Wiring Practices 62 6 3 Wiring Analog Inputs and Outputs 64 6 4 Wiring Relays 72 d A WARNING While the unit is powered a potentially lethal shock hazard exists inside the case Do not open the case while the unit is powered Disconnect power before installing I O wiring More than one switch may be required to disconnect power 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 61 Input and Output Wiring 6 2 General Wiring Practices d A WARNING
55. NUTES 30 CURRENT DAY 2 DIAGNOSTICS NEXT Timer setup is used to specify the frequency of the AutoClean rinses that are included in the AutoCal schedule e f several rinses are to be performed in a single day a specify the number of days as zero b specify the number of hours and minutes in the interval between rinses 1 hour is the minimum e f rinses are to occur at intervals of days a specify the number of days in the interval b specify the time of day hours and minutes the rinse should occur In our example the rinse will occur every third day at 4 30 PM e If you plan to use AutoCal but initiate operations manually using the real time AUTOCAL CLEAN display a specify the number of days as zero b specify the number of hours and minutes in the interval between rinses as zero also IF YOU ENTER ALL ZEROS NO AUTOMATIC RINSES AND NO AUTOMATIC CALIBRATIONS WILL BE PERFORMED Specify where today falls in the AutoCal AutoClean cycle For example if the interval for rinses is 3 days and you want the first rinse to be tomorrow specify that today is day 2 Go to the next screen AUTO CAL 1PT RINSE SETUP DURATION 180s BETWEEN STD 2 DIAGNOSTICS NEXT Enter the duration of the AutoClean rinse in seconds 60 is the minimum Specify the frequency of AutoCal standardization operations by entering the number of rinses between standardization operati
56. OM when power is off Fuse Rating 120 V operation 0 25 amp 250 V 240 V operation 0 125 amp 250 V Case Gray Noryl plastic waterproof and corrosion resistant Interior conductive coating to provide effective RFI EMI shielding Case Dimensions 156 x 156 x 178 mm 6 1 8 x 6 1 8 x 7 in panel cutout 141 x 141 mm 5 53 x 5 53 in Weight 1 8 kg 4 Ib Mounting Panel mounting hardware supplied Surface and pipe mounting hardware available See Figure 3 2 or the Parts List 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 15 Specifications and Model Number 2 2 Model Number Breakdown Introduction The model number breakdown is presented in the tables that follow The basic model number consists of a key number Appended to this key number are characters that identify the features in various categories The meaning of the characters in each category is presented in a table identified below Key Number TableI Table Il Table WI Table IV 9782P Model Number Table I INPUTS Description Model No pH ORP Electrodes pH from external pre amp or Cap Adapter direct glass antimony pH ORP electrode internal preamp 01 02 direct Durafet pH electrode internal preamp Direct Glass from HPW 7000 Hi pHurity Water Assembly 03 04 Model Number Table II AVAILABLE OUTPUTS in addition to two standard relays Description no analog outputs nor additional
57. P Redox only measurement connect metal electrode to R and reference electrode to E For simultaneous pH and ORP measurement remove RE SC jumper and connect reference electrode to RE of upper terminal block Connect metal electrode to R and connect pH electrode to E of lower terminal block 3 Analyzer is factory set for 120 V or 240 V operation See directions supplied with Analyzer 4 Relay 3 amp 4 contact ratings 3 A 120 Vac 28 Vdc resistive 5 On some electrodes the coaxial shield carries the reference No connection is then made to the Shield terminal 6 RE SC jumper resistor must remain in place except for simultaneous pH and ORP measurements D Color Code K black Y yellow W white A gray G green V violet R red P pink B blue S slate N brown T tan O orange 3 00 Figure 6 6 Terminal Connections 9782 Series pH ORP Analyzer Controller Operator s Manual 71 Input and Output Wiring 6 4 Wiring Relays 6 4 1 Installation Introduction Every 9782 model includes two relay outputs that can be used with special features such as AutoClean and AutoCal or to perform Pulse Frequency Type or Duration Adjusting Type control or for alarm annunication In additon to these two standard relays two more are available as an option The Analyzer Controller can be programmed for these relays to de energize on alarm or to energize on
58. Page 10 1 Overview 109 10 2 Calibrating Outputs 111 10 3 Restoring Factory Calibration Settings 114 10 4 Specifying a Tagname or Other Display String 115 10 5 Viewing Product Information and Changing Model Number Stored in Memory 116 10 6 Adjusting the Screen Contrast 118 10 7 Entering a Password for Security 119 10 8 Resetting All Configuration and Calibration Values to Factory Settings 120 10 9 Switching Between Electrodes with DualCal 121 10 10 Determining and Entering A Solution Temperature Coefficient 122 10 11 Turning on Durafet Temperature Display 125 10 12 Specifying Relay Activation 126 10 13 Replacing the Fuse 127 d A WARNING While the unit is powered a potentially lethal shock hazard exists inside the case Do not open the case while the unit is powered 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 109 Maintenance ATTENTION Most of the operations described in this section are performed with the unit off line that is not connected to the process However some of the operations described in this section are performed with the Analyzer Controller on line Unless you put the unit in Hold mode before starting a maintenance function with the unit on line your process may be affected as soon as you initiate the maintenance function For example if you change the relay activation from ENERGIZE on alarm to DE ENERGIZE on alarm the change will be implemented as soon as you press the ENTER key
59. Performance under reference operating conditions Accuracy 0 02 pH or 2 mV display Drift Negligible Repeatability 0 1 of span 1 count Reference Operating Conditions 25 1 C 10 40 RH 120 Vac Operating Influences under normal operating conditions Effect on accuracy of span Temperature 0 05 per C RH lt 1 Line Voltage lt 0 1 per volt Power Loss Memory retention by EEPROM no batteries 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 13 Specifications and Model Number Operating Conditions Ambient Temperature Normal 0 C to 60 C 32 F to 140 F extreme 10 C to 60 C 4 F to 140 F storage 30 C to 70 C 22 F to 158 F Line Voltage Normal 120 10 Vac extreme 100 Vac to 132 Vac 47 Hz to 63 Hz Normal 240 10 Vac to 264 Vac extreme 200 Vac to 264 Vac 50 Hz to 60 Hz RH 90 maximum non condensing 40 C 104 F max Installation Category II Pollution Degree 2 Altitude lt 20000 M 6562 ft Output Adjustable to within 0 01 mA repeatable to within 0 1 of span Optional Output Two or three 4 20 mA outputs are available via an optional plug in circuit card and program module Isolated from ground inputs and each other Maximum isolation voltage 240 Vac will withstand 1500 V hi pot test Output adjustable to within 0 01 mA Repeatability 0 1 of span 600 ohm maximum load resistance
60. ROME a Kul gt d POLYELECTROLYTE REDUCTION NEUTRALIZATION SETTLING SLUDGE Figure C 1 Chrome Treatment System 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 139 Chrome Waste Treatment C 2 First Stage of Chrome Removal Lower pH and add reducing agent Sulfuric acid is used to lower the pH to about 2 5 which promotes the reduction reaction and ensures complete treatment The reducing agent may be sulfur dioxide sodium sulfite sodium bisulfite sodium metabisulfite sodium hydrosulfite or ferrous sulfate The reaction is given below The chromate is expressed as chromic acid CrO with a 6 charge on the chromium and the reducing agent is expressed as sulfurous acid HSC which is generated by sulfites at low pH The result is chromium sulfate Cr SO 3 with a 3 charge on the chromium The reaction is expressed as 2CrO 3H SO gt Cr SO 34 0 This first stage reaction is analyzed and controlled by independent control loops acid addition by pH control reducing agent addition by redox potential or ORP oxidation reduction potential control Often an ON OFF type of control using solenoid valves or metering pumps can be used The pH controller simply calls for additional acid whenever the pH rises above 2 5 The ORP controller calls for additional reducing agent whenever the ORP potential rises above about 250 mV The metal ORP electrode is positive with respect to the reference elect
61. T NEXT CONFIGURATION CAT RETRANSMISSION ALARMS DISCRETE CONTROL AUTO CAL CLEAN ALARMS 1 LARM 2 3 LARM 4 5S DIAGNOSTICS ALARMe INPUT ALARM DELAY SETPOINT DEADBAND pH HIGH 58 OpH 1pH T DIAGNOST S HOLD T re Action Go to CONFIGURATION ALARMS ALARM e ATTENTION Any value entered for one of these alarm related parameters will take effect as soon as it is entered UNLESS THE 9782 IS IN HOLD MODE Enter a delay value to prevent brief upsets from activating an alarm This configured delay value will apply to all alarms Select the appropriate alarm for configuration The process variable and action configured for the alarm during UO setup will be displayed The delay time in seconds configured under ALARMS will also be displayed Enter the setpoint value at which the alarm will become active Specify the deadband Once the alarm is active it will not deactivate until the PV differs from the setpoint by at least the value of the deadband limit Take the unit out of Hold mode 50 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 5 7 Setting the Clock Purpose UO Setup and System Configuration A display showing the current time can be accessed when the unit is online The 24 hour clock must be set before you can use the AutoClean feature described in Subsection 5 8 Procedure Fol
62. T TEMP DISPLAY MAINTENANCE I O SETUP DIAGNOSTICS PREV NEXT MAINTANENCE OFF LINE FUNCTIONS INSTRUMENT SETUP ELECTRODE FUNCTIONS RELAY ACTIVATION DIAGNOSTICS ELECTRODE FUNCTIONS SELECT ELECTRODE SOLUTION TEMP COMP DURAFET TEMP DISPLAY DIAGNOSTICS l l 2 BURA T TEME DL SPTAN The display will indicate whether temperature display is DISPLAY ON NO currently turned on Use this display to toggle display of temperature ON and DIAGNOSTICS OFF I T 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 125 Maintenance 10 12 Specifying Relay Activation Introduction All 9782 models are equipped with two relays As an option two additional relays are available The relay action used for alarm and or control action for all relays in the Analyzer Controller can be specified as energized or de energized on alarm or when discrete control point is ON Procedure Use the procedure in Table 10 13 to specify relay action Table 10 13 Procedure for Specifying Relay Activation Step Screen Action 1 ea Go to MAINTENANCE RELAY ACTIVATION CONFIGURATION CALIBRATION MAINTENANCE I O SETUP DIAGNOSTICS PREV NEXT MAINTANENCE OFF LINE FUNCTIONS INSTRUMENT SETUP ELECTRODE FUNCTIONS RELAY ACTIVATION DIAGNOSTICS 2 EE The display will indicate whether relay activation in ca
63. V 1V 200 min PA NC COM Ree 2 Thermistor NO Note 1 Thermistor NC Relay 1 Reference COM Note 1 Shield Note 4 NO pH electrode e e 120w 240v A 15A 50 60 Hz Terminals for E Note 3 L PE connection of Voltage amp Current Power amp direct electrode Output Terminals Relay Terminals input Note 2 NOTES 9782P 02 VC 1 Relay contacts max ratings Resistive Load Current 0 6 A at 120 240 Vac 0 6 A at 110 Vac Isolated Current amp Voltage Outputs Du Direct Glass Electrode Input Voltage 240 Vac Switching Power 125 VA ac 60 W dc See electrode mounting diagram Connections are shown for single pH measurement For ORP Redox only measurement connect metal electrode to R and reference electrode to E For simultaneous pH and ORP measurement remove RE SC jumper and connect reference electrode to RE of upper terminal block Connect metal electrode to R and connect pH electrode to E of lower terminal block Analyzer is factory set for 120 V or 240 V operation See directions supplied with Analyzer On some electrodes the coaxial shield carries the reference No connection is then made to the Shield terminal D wo Case as viewed from front A 5 RE SC jumper resistor must remain in place except for simultaneous pH and ORP measurements Color Code NC COM eek K black Y yellow Thermistor NO ote 7 Thermistor NC Relay 1 W white A gray Reference COM Note
64. Wiring should be performed by qualified personnel only Safety precaution d A WARNING A disconnect switch must be installed to break all current carrying conductors Turn off power before working on conductors Failure to observe this precaution may result in serious personal injury d A WARNING An external disconnect switch is required for any hazardous voltage connections to the relay outputs Avoid damage to components ATTENTION This equipment contains devices that can be damaged by electrostatic discharge ESD As solid state technology advances and as solid state devices get smaller and smaller they become more and more sensitive to ESD The damage incurred may not cause the device to fail completely but may cause early failure Therefore it is imperative that assemblies containing static sensitive devices be carried in conductive plastic bags When adjusting or performing any work on such assemblies grounded work stations and wrist straps must be used If soldering irons are used they must also be grounded A grounded work station is any conductive or metallic surface connected to an earth ground such as a water pipe with a 1 2 to 1 megohm resistor in series with the ground connection The purpose of the resistor is to current limit an electrostatic discharge and to prevent any shock hazard to the operator The steps indicated above must be followed to prevent damage and or degradation which may be induced by ESD to
65. agnetic electrostatic and radio frequency interference pickup of sufficient magnitude to overload input filters The best instrument performance is obtained by keeping the interfering signals out of the instruments altogether by using proper wiring practices References Refer to the following when wiring the unit e IEEE STD 518 Guide for the Installation of Electrical Equipment to Minimize Electrical Noise Inputs to Controllers from External Sources e Appropriate wiring diagram supplied with electrode mounting or preamplifier module 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 63 Input and Output Wiring 6 3 Wiring Analog Inputs and Outputs 6 3 1 Installation Introduction Every 9782 Analyzer Controller requires one or more analog input signals to provide pH and or ORP values for the system Temperature input can also be provided for display and for use by the solution temperature compensation algorithm see 10 10 In addition some 9782 models provide from one to three analog output signals current and or voltage that can be used to retransmit process variables or to provide Current Adjusting Type control see 1 5 3 Wiring these analog inputs and outputs is described here Wiring relay outputs is described in 6 4 Procedure Follow the procedure in Table 6 2 to install analog I O wiring If DualCAT control will be used also refer to the special instructions in 6 3 2 d A WARNING Wh
66. ally only have to access this function once in the life of your Analyzer Controller Purpose of configuration menu items The configuration menu contains the following items for accessing screens used to configure the features described below e CAT RETRANSMISSION If the hardware supports analog outputs configure parameters related to use of these outputs as described in 5 4 Output 1 and or Output 2 can be used for Current Adjusting Type CAT control Any one two or three outputs can be used for retransmission of process variables If the unit contains only one analog output it can be used for DualCAT pH only this feature controls two reagents using a single output channel e ALARMS If one or more relays are available for annunciation of alarms configure each alarm setpoint deadband and delay as described in 5 6 Action is configured with I O SETUP e DISCRETE CONTROL One or more relays can be used to achieve control of process variables using On Off Duration Adjusting Type DAT or Pulse Frequency Type PFT control strategies Configure the related parameter values as described in 5 5 e ADVANCED FEATURES This menu item provides access to configuration screens for the following features Clock Setup A display showing the current time can be accessed when the unit is online The 24 hour clock must be set before you can use the AutoClean feature described below Set the clock as descibed in Section 5 7 Auto Clea
67. assembly is mounted on pivot arms The safety cover with its warning label will be visible With the power off remove the safety cover by removing the single screw holding it in place Label and remove the field wiring connected to the analog output terminals Feed the test wiring through the conduit hole in the case and connect the appropriate type of meter to each of the analog output terminals to be tested Be sure to observe correct polarity Close the case and power up the unit Do not connect power until the case is closed 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 107 Diagnostics and Messages AA AA Step Screen Action 7 eae EON Go to MAINTENANCE OFF LINE FUNCTIONS CALIBRATION OUTPUT TESTS MAINTENANCE I O SETUP DIAGNOSTICS PREV NEXT MAINTANENCE OFF LINE FUNCTIONS INSTRUMENT SETUP ELECTRODE FUNCTIONS RELAY ACTIVATION DIAGNOSTICS OFF LINE FUNCTIONS DISPLAY TEST KEYBOARD TEST OUTPUT TESTS OUTPUT CALIBRATION CALIBRATION RESET DIAGNOSTICS I I 8 a ae ae eee Select the combination of output values to be supplied to 0 0 0 AAAA the terminals or go to the next window for additional 25 0 0 A H 50 25 0 S z S A choices 75 50 25 AAAA ae DIAGNOSTICS When one of the rows of output combinations has been cs selected observe the meter readin
68. ating Outputs Step Screen Action 1 Turn off the power to the Analyzer Controller A More than one switch may be required to disconnect power 2 With the power off open the case e Grasp the bottom center portion of the front bezel and pull it downward and toward you slightly to disengage the bottom of the bezel from the edge of the case e Lift the bezel gently to disengage it from the top edge of the case e Swing the bezel to the left The bezel and display assembly is mounted on pivot arms The safety cover with its warning label will be visible 3 With the power off remove the safety shield by removing the single screw holding it in place 4 Label and remove the field wiring connected to the analog output terminals 5 Feed the test wiring through conduit hole in the case and connect the appropriate type of meter to the terminals for the output to be calibrated Be sure to observe correct polarity 6 Close the case and power up the unit 7 oe ae Go to MAINTENANCE OFF LINE FUNCTIONS CALIBRATION OUTPUT CALIBRATION MAINTENANCE I O SETUP DIAGNOSTICS PREV NEXT MAINTANENCE OFF LINE FUNCTIONS INSTRUMENT SETUP ELECTRODE FUNCTIONS RELAY ACTIVATION DIAGNOSTICS OFF LINE FUNCTIONS DISPLAY TEST KEYBOARD TEST OUTPUT TESTS OUTPUT CALIBRATION CALIBRATION RESET DIAGNOSTICS T 112 9782 Series pH ORP A
69. be as great as 2 minutes for a large step change in values a message will appear on the alarm stripe If data is being retrieved from memory the message will be Retrieving Data and if a new value is being stored in memory the displayed message will be Entering Data Placing the unit in Hold mode To put the Analyzer Controller in Hold mode maintaining a constant output and alarm condition and disabling solution temperature compensation if used use the HOLD soft key available on the appropriate screens HOLD ACTIVE will be displayed on the alarm stripe Use of Hold mode is recommended e Before starting manual calibration e Alarms are always held during an Automatic clean calibration The user selects whether outputs are held during these automatic functions In either case the alarm stripe will display HOLD ACTIVE e Before making configuration changes that should not take effect immediately To take the unit out of Hold mode press HOLD again 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 5 Introduction Controlling an AutoClean and or AutoCal operation manually To initiate an Auto Clean and or AutoCal operation manually press the DISPLAY key until the AutoClean Cal screen is displayed Next press the START soft key This display can also be used to see how much time is remaining during an automatic operation When AutoClean and or AutoCal are active the alarm stripe will
70. ble timer initiated AutoClean you must either turn off the 24 hour clock which preserves your timer setup or set the timer to all zeros see Step 2 in Table 5 9 Unless the Time Remaing display is on screen you will not see the AUTO CLEAN message during a scheduled cleaning Operator initiated operation To operate in this mode you must have the AutoClean Time Remaining display on screen Initiate the AutoClean sequence by pressing the START soft key Once the START soft key is pushed the time display will go to zero and the message AUTO CLEAN will appear on the alarm stripe The sequence will stop cleaning if the STOP soft key is pressed or if you leave this display Once stopped the Analyzer Controller will wait for the configured resume delay time to allow the cell measurement to stabilize before coming back on line During the resume the AUTO CLEAN message will flash If the PAUSE soft key is used while the AutoClean Time Remaing display is on screen the AutoClean sequence will stop temporarily If the PAUSE soft key is not pushed again to resume the sequence then the sequence will automatically resume in 20 minutes 52 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 UO Setup and System Configuration Once the AutoClean sequence has been completed including resume delay the AUTO CLEAN message disappears and the time remaining clock will be displayed again showing the time remaining
71. ct one of these buffers and use its values in the calibration process Automatic checks are included to ensure that reasonable and correct values are entered The procedure for using the automatic buffer recognition feature in an actual calibration is provided in 8 6 3 Automatic buffer recognition is also used in the AutoCal automatic calibration procedure as described in Section 7 Three default buffers The three default buffer nominal values at 25 C are 4 01 7 00 and 9 00 pH While 4 01 will always be used for the acid buffer a choice of neutral and alkaline buffers may be selected as described in 8 6 2 Representative values are in Table 8 6 Table 8 6 Representative Buffer Values at Various Temperatures Nominal C pH 5 10 15 20 25 30 35 40 50 60 4 4 00 4 00 4 00 4 00 4 00 4 01 4 02 4 03 4 06 4 09 6 86 6 95 6 92 6 90 6 88 6 86 6 85 6 84 6 84 6 83 6 84 7 7 08 7 06 7 03 7 01 7 00 6 99 6 98 6 98 6 97 6 98 9 9 21 9 15 9 10 9 05 9 00 8 96 8 92 8 89 8 83 8 78 9 18 9 39 9 33 9 28 9 23 9 18 9 14 9 11 9 07 9 02 8 96 10 00 10 24 10 17 10 11 10 05 10 00 9 96 9 92 9 88 9 82 9 76 94 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Manual Calibration 8 6 2 Choosing Neutral and Alkaline Buffers NOTE If the defaults 7 00 and 9 00 are acceptable skip this sub section and go to 8 6 3 The procedure for specifying neutral and alkaline buffers is described in Tabl
72. d Changing Model Number Stored in Memory Introduction During I O setup see Section 5 the software will display prompts only for those features the hardware can support To do this the software uses the model number of the unit stored in memory at the factory Each combination of features has a unique model number see Section 2 Therefore if you add an option card to the 9782 it is very important that you change the model number stored in memory using the procedure described below ATTENTION A special password is required to change the stored model number If the password was not supplied with the new option card call the Honeywell Technical Assistance Center Before calling for technical support follow the procedure in Table 10 4 to determine the version of the software in the Analyzer Controller The support team will ask for the this information Procedure Use the procedure in Table 10 4 to see the software version and model number and to change the model number if required Table 10 4 Procedure for Viewing Product Information and Changing the Stored Model Number Step Screen Action 1 DE Go to MAINTENANCE INSTRUMENT SETUP CALIBRATION PRODUCT INFO MAINTENANCE I O SETUP DIAGNOSTICS L PREV NEXT MAINTANENCE OFF LINE FUNCTIONS INSTRUMENT SETUP ELECTRODE FUNCTIONS RELAY ACTIVATION DIAGNOSTICS INSTRUMENT SETUP INSTURMENT TAGGING PRODUCT INFO SCREEN CONTRAST S
73. d can be any number between 1 and 9999 When the password is zero the operator will not be prompted to enter a password ATTENTION The configurable password entered here is not the same as the special password needed to edit the model number stored in memory as described in 10 5 Procedure Follow the procedure in Table 10 6 to enter a password Table 10 6 Procedure for Entering a Password Step Screen Action 1 en Go to MAINTENANCE INSTRUMENT SETUP CALIBRATION SECURITY MAINTENANCE I O SETUP DIAGNOSTICS PREV NEXT MAINTANENCE OFF LINE FUNCTIONS INSTRUMENT SETUP ELECTRODE FUNCTIONS RELAY ACTIVATION DIAGNOSTICS INSTRUMENT SETUP INSTURMENT TAGGING PRODUCT INFO SCREEN CONTRAST SECURITY RESET UNIT DIAGNOSTICS I I 2 PASSWORD SECURITY To enable the security function select YES ENABLE SECURITY YES e P PASSWORD 000 The current password will be displayed It can be edited using this screen DIAGNOSTICS I 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 119 Maintenance 10 8 Resetting All Configuration and Calibration Values to Factory Settings Introduction A cold reset function is available to return ALL configuration values to the factory settings shown in Table 5 1 During a cold reset the standardization offset is also returned to 0 0 and the percent theoretical slope i
74. dard types is appropriate select it from DIAGNOSTICS the list of available types The coefficient for that type will E be displayed automatically SOLUTION TEMP COMP If one of the standard types in not appropriate select OTHER from the list of available types then enter the appropriate coefficient TYPE OTHER pH VALUE 10 C xxxxx DIAGNOSTICS HOLD 1 T 124 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Maintenance 10 11 Turning on Temperature Display Durafet Electrodes without Cap Adapter Only Introduction Temperature is one of the process variables that can be displayed To turn on the display of temperature follow the procedure in Table 10 12 ATTENTION A Durafet pH electrode has a temperature compensator for normal Nernstian temperature compensation This integral temperature compensator can not be used to display temperature For temperature to be displayed when a Durafet electrode type is selected temperature display must be activated see Table 10 12 and a separate temperature input must be connected to the TH and SC terminals If a separate temperature input is not provided display and alarming of temperature is not possible AutoCal and Automatic Buffer Recognition will assume 25 C Table 10 12 Procedure for Turning on Temperature Display Step Screen Action 1 SOLEO ae Go to MAINTENANCE ELECTRODE FUNCTIONS CALIBRATION DURAFE
75. display AUTOSEQUENCE and the AutoClean Cal display will show the remaining time If necessary pause the operation using the PAUSE soft key To resume the operation press PAUSE again To cancel the operation use the STOP soft key 1 4 Menu Hierarchy Menus for every task For your convenience menus are provided for configuration calibration maintenance and I O setup tasks The menu hierarchy is shown in Table 1 2 Not every menu item applies to every system Table 1 2 9782 pH Analyzer Controller Menu Hierarchy a ae i i CONFIGURATION set up Analyzer Controller to implement your alarm and control strategy CAT RETRANSMISSION configure parameters for using analog 5 4 outputs for control and retransmission of process values ALARMS specify alarm setpoints deadbands and 5 6 delays DISCRETE CONTROL specify setpoints range limits etc to use 5 5 relays for control ADVANCED FEATURES set up automatic cleaning and calibration 5 7 and 5 8 of electrodes and sets up clock I CALIBRATION select buffers for automatic buffer 8 recognition and perform manual calibration pH BUFFER CAL calibrate pH system manually using one or 8 2 two reference solutions ORP CAL calibrate ORP system manually using a 8 4 or 8 5 reference solution or voltage input SAMPLE CAL calibrate pH system manually using a 8 3 standard meter AUTO BUFFER CAL calibrate pH
76. djusting Type DAT or Pulse Frequency Type PFT Control Purpose During I O setup you had an opportunity to specify whether to use one or more relays to achieve control of a process variable with a discrete control strategy On Off DAT or PFT If one of these control types was selected you will be able to configure the related parameter values For a discussion of tuning proportional control applicable to PFT and DAT control see Appendix E Procedure The procedure for configuring On Off control is shown in Table 5 4 The procedure for configuring PFT control is shown in Table 5 5 The procedure for configuring DAT control is shown in Table 5 6 46 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 UO Setup and System Configuration Table 5 4 Procedure for Configuring On Off Control Step Screen Action 1 Ge Go to CONFIGURATION DISCRETE CONTROL CALIBRATION ON OFF MAINTENANCE I O SETUP DIAGNOSTICS PRE T NEXT ee eas pean ee ATTENTION Any value entered for one of these control EE related parameters will take effect as soon as it is entered AUTON GAL IGN UNLESS THE 9782 IS INHOLD MODE DIAGNOSTICS I DISCRETE CONTROL ON OF F PET DAT DIAGNOSTICS 2 BEER Use this screen to specify the length of the cycle period for CYCLE TMR 1 all timers CYCLE TMR 2 E To configure an individual timer associat
77. e amp Current Output Connections Output Terminals Power amp for 9782P C2 or Relay Terminals EK Beef 9782P 01 VC NOTES Durafet Il Cap Adapter Input 1 Relay contacts max ratings Resistive Load Isolated Current and Voltage Outputs Current 0 6 A at 120 240 Vac 0 6 A at 110 Vdc 2Aat30 Vdc Voltage 240 Vac Switching Power 125 VA ac 60 W dc 2 See Diagram supplied with electrode mounting assembly 3 Analyzer is factory set for 120 V or 240 V operation see directions supplied with Analyzer 4 ii leis f il From Cap Adapter Case as viewed from front EE with Durafet Il input circuit card Electrode P RE SC jumper resistor must remain in place SE SS 6 9782 Unit does not support RTD temperature input FE ORE Wires color coded W K are not used W STH Note 4 K NO Sc Note 6 Ps B O V G Q v KZ n Ei NG Color Cod Oa COM Relay 2 olor Code oa Ae Note 1 K black Y yellow Ge NO Note 1 W white A gray 7 or Ee ec G green V violet 9 15 S K 4 J 50 60 Hz R red P pink Note 3 B blue S slate L PE Power amp N brown T tan Relay Terminals O orange 9782P 01 00 Durafet Il Cap Adapter Input No Outputs PE Protective Earth Ground Figure 6 2 Terminal Connections 3 00 9782 Series pH ORP
78. e 8 7 The buffers selected will be used along with 4 01 whenever the automatic buffer recognition function is executed by operator command or by AutoCal Table 8 7 Buffer Specification Procedure Step Screen Action 1 RNU Go to CALIBRATION AUTO BUFFER SETUP CONFIGURATION CALIBRATION MAINTENANCE I O SETUP DIAGNOSTICS BUFFER ORP CAL SAMPLE CAL AUTO BUFFER CAL AUTO BUFFER SETUP CAL DIAGNOSTICS DIAGNOSTICS 2 AUTO BUFFER SETUP Cae Geen Select one neutral buffer and one alkaline buffer ACID FIXED 4 01 3 NEUTRAL 7 00 Go back to the calibration menu You choices have been ALKALINE 9 00 stored DIAGNOSTICS l 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 95 Manual Calibration 8 6 3 Manual Calibration Using Automatic Buffer Recognition Materials and procedure To perform manual calibration using the automatic buffer recognition feature follow the instructions in Table 8 8 Materials required are e Two standard buffer reference solutions out of the three acid neutral alkaline in Auto Buffer Setup described in 8 6 2 e A container for each large enough to immerse the electrode to measuring depth e Distilled or de ionized water to rinse the electrode NOTE A Durafet pH electrode has a temperature compensator for normal Nernstian temperature compensation This integral temperatur
79. e 9782 stores the new calibration value using automatic buffer recognition Diagnostics detect excessive instability or offset prevent erroneous calibrating and can activate an alarm depending on configuration If the diagnostic fails an error message is always displayed on the alarm stripe see Section 9 If an unacceptable value is obtained it will be rejected and the previous value will be retained for uninterrupted operation All valves are deactivated to resume measurement of the sample 6 A delay period 1 to 1999 seconds can be configured to permit the measurement to stabilize on the process sample At the end of the delay period normal alarm control and output operation resumes The HOLD and AUTOSEQUENCE messages are cleared Note that the operator can make the operation pause using the soft key on the special AutoClean AutoCal display If the operator does not remove the pause by pressing the PAUSE key again the Analyzer Controller will resume normal operation after 20 minutes 80 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 AutoClean and AutoCal Theory and Piping Eee Items outside this area buffer provided by user solution 9782 to process H electrode 7 4 l L to drain B l ett Ze 4 process sample N gt e rinse water S1 g oe Figure 7 2 Rinse a
80. e Page 11 1 Overview 129 11 2 Part Numbers 130 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 129 Accessories and Replacement Parts List 11 2 Part Numbers Introduction Part numbers for field replaceable parts and accessories are provided in Table 11 1 Table 11 1 Part Numbers Kit Part Number 51198177 501 51198063 501 51198064 501 Description miscellaneous parts for all 9782 controller analyzers output card single current or voltage output output card single current or voltage output with two hermetically sealed relays Parts Included in Kit 120 volt fuse 2490 volt fuse case gasket upper card retainer bracket lower card retainer bracket screws for retainer brackets card installation instructions card installation instructions Quantity l ch ND sch 51198065 501 output card two general purpose relays and no analog outputs card installation instructions 51198066 501 replacement output card two current outputs card installation instructions 51198066 502 51198067 501 51198067 502 new output card two current outputs replacement output card three current outputs new output card three current outputs card PROM installation instructions card installation instructions card PROM installation instructions 51198068 501 Case and gasket case no bezel case
81. e compensator cannot be used to display temperature When a Durafet electrode type is selected temperature display must be activated see 10 11 and a separate temperature compensator must be connected to the TH and SC terminals If a compensator is not used temperature measurement is not possible AutoCal and Automatic Buffer Recognition will assume 25 C 96 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Manual Calibration Table 8 8 Standardization Procedure Using Automatic Buffer Recognition PUT ELECTRODE IN BUFFER NOTE BUFFER MUST BE gt 2pH UNITS AWAY FROM STD BUFFER DIAGNOSTICS HOLD NEXT Step Screen Action 1 E Prepare containers of two standard reference solutions CALIBRATION from Auto Buffer Setup MAINTENANCE I O SETUP DIAGNOSTICS 2 EE Go to CALIBRATION AUTO BUFFER CAL SAMPLE AUTO BUFFER CAL AUTO BUFFER SETUP CAL DIAGNOSTICS DIAGNOSTICS I 3 EE Put the unit in Hold mode FER Se Remove the electrode from the process STD VALUE IS ENTERED See Rinse the electrode thoroughly with distilled or de ionized HOLD L NEXT water Immerse it in the buffer solution with a pH closest to the process Go to the next screen 4 BUFFER VALUE 3 0 The display will show the pH ofthe buffer solution Within fifty seconds the reading will be automatically
82. e off line function INSTALLING FACTORY CALIBRATION menu DEFAULTS MENU TO EXIT DIAGNOSTICS 114 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Maintenance 10 4 Specifying a Tagname or Other Display String Introduction The real time displays of process values show the instrument s tagname or other configurable fixed nineteen character string at the top of the screen Procedure Use the procedure in Table 10 3 to specify the text to be displayed at the top of the process value displays Table 10 3 Procedure for Specifying a Tagname or Other Display String Step Screen Action 1 a ae Go to MAINTENANCE INSTRUMENT SETUP CALIBRATION INSTRUMENT TAGGING MAINTENANCE I O SETUP DIAGNOSTICS PREV NEXT MAINTANENCE OFF LINE FUNCTIONS INSTRUMENT SETUP ELECTRODE FUNCTIONS RELAY ACTIVATION DIAGNOSTICS INSTRUMENT SETUP INSTRUMENT TAGGING PRODUCT INFO SCREEN CONTRAST SECURITY RESET UNIT DIAGNOSTICS UPDATE TAGE DATA EDIT 19 CHAR STRING XXXXXXXXXXXXXXXXXXK DIAGNOSTICS T lt gt The text string currently stored in memory will be displayed If desired edit the string nineteen characters maximum Press ENTER to change text 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 115 Maintenance 10 5 Viewing Product Information an
83. e tel nn 43 5 5 Configuring On Off Duration Adjusting Type DAT or Pulse Frequency Type PFT Control euer Eed aan ed tres 46 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual v 5 6 Confisurine E EE 50 5 7 Setting the Clock aie sires ina ra adh es lee co SEET 51 5 8 Configuring AutoCal and Auto Jean 52 6 INPUT AND OUTPUT WIRING exeseugeueersguereeugerierdeuekiereeeEuEEEKUNeSEEeKEERKEEEEEOE NEEN 61 6 1 OVELVICW fees Seat ee ae eee eae Sn eae ne nee ee ee i eet ately 61 6 2 G n ral Wiring Practices fist din tt tnt nt title es tte 62 6 3 Wiring Analog Inputs and Outputs ss 64 eh Installauon gees leese ege EE EE 64 6 3 2 Special Information for Dua AT 66 6 4 Wiring Relays orire iini iieii ieni R el en EE ita T2 GL Installation ss sn tnea o EE SNE 72 6 4 2 Special Information for DualCal ss 74 6 4 3 Disconnecting RC Arc Suppression Circuits 76 7 AUTOCLEAN AND AUTOCAL THEORY AND PIPING nn rrrssnnnnns 77 7 1 OVervE W oa E nn tn nn a ee aa ele ee eaten ene 77 7 2 AutoClean Sequence and Piping ses 78 7 3 AutoCal Sequence and Pup 9 2 cose 28020 sea Cd ec bs este ect site niet a enr nds nie 80 8 MANUAL CALIBRATION iiisscisticciiictiscnrcntiacisnassuassatuduansuiesuaildedsiuniduaedcestatpennnonnisunae 83 8 1 A EE 83 8 2 Buffering Method of Calibrating pH Electrodes A 85 8 3 Grab Sample Method of Calibrating pH Electrodes AAA 87 8 4 ORP Calibration Using Reference Solutio
84. ecial algorithm the Analyzer Controller applies the needed correction for this effect once a value for an appropriate solution temperature coefficient has been entered using the procedure in Table 10 11 A separate temperature input is required for solution temperature compensation when Durafet electrode without a Cap Adapter is used For Durafet electrodes with a Cap Adapter no separate temperature input is required Using this compensation algorithm the Analyzer Controller measures pH at any temperature and displays the pH referenced to 25 C During the Hold mode the algorithm is bypassed so that the display is allowed to show uncompensated pH of a buffer solution for calibration Buffer solutions are provided with their own pH temperature data on the label 10 10 2 Determining Solution Temperature Coefficient Introduction Recommended coefficients for some applications are already stored in the Analyzer Controller s memory If your application is one of the following specify it as described in Table 10 11 e power plant steam and condensate with ammonia and or morpholine 15 C to 40 C e boiler phosphate treatment 15 C to 40 C e untreated pure water samples 15 C to 40 C For these applications at a temperature range other than 15 C to 40 C select an appropriate solution temperature compensation coefficient from Table 10 9 or determine the solution 122 9782 Series pH ORP Analyzer Controller Operator s
85. ed to displaying process alarm messages status messages and system diagnostic messages whenever any are active Dedicated function membrane keys to the right of the display on the front panel are used to e cycle through the displays of real time values e access configuration calibration and maintenance menus e change and enter setpoints and other parameters In addition three variable function keys are below the display The function of each of these soft keys depends on the purpose of the screen currently on display Soft key labels are always displayed at the bottom of the screen immediately above the relevant function key Easy to configure Input 3 00 Menu driven configuration is fast and easy Only configuration parameters related to features supported by the hardware and relevant to the specific application are displayed Input to the 9782 can be from any Honeywell pH electrode preamplifier system or Durafet II electrode with Cap Adapter at distances up to 914 4 m 3000 ft If the Analyzer Controller is equipped with an optional internal preamplifier input can come directly from any glass pH or ORP electrode located within 3 66 m 12 ft of the analyzer or directly from a Durafet Il pH electrode within 15 24 m 50 ft 9782 Series pH ORP Analyzer Controller Operator s Manual 1 Introduction Outputs Optional isolated 0 1 V 0 10 V and 4 20 mA outputs are available Use these analog outputs for retransmissio
86. ed with one of the DIAGNOSTICS relays select it from the menu 3 CN Ce RE The input type relay assignment and appropriate relay OUTPUT oeh output action will already be filled in based on the process SETPOINT 9pH variable assigned to this On Off control output during UO dees setup This is read only information DIAGNOSTICS HOLD lt 7 Enter the setpoint at which the output should be turned on If desired specify a deadband limit If the PV varies from the setpoint while the output is ON but by less than the value of the deadband limit the output will remain on Enter the percent of the cycle period that the relay should stay on when the setpoint is reached Take the unit out of Hold mode 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 47 UO Setup and System Configuration Table 5 5 Procedure for Configuring PFT Control Step Screen MENU CONFIGURATION CALIBRATION MAINTENANCE I O SETUP CONFIGURATION CAT RETRANSMISSION ALARMS DISCRETE CONTROL AUTO CAL CLEAN DIAGNOSTICS DISCRETE CONTROL ON OFF PET DAT DIAGNOSTICS PFT CONTROL PFT1 PFT2 DIAGNOSTICS HOLD NEXT PFT 1 INPUT pH OUTPUT RLY1 SETPOINT 9pH PB LIMIT 14pH PB FREQ RATE 60Hz DIAGNOSTICS HOLD lt gt Action Go to CONFIGURATION DISCRETE CONTROL PFT ATTENTION Any value entered for
87. eiis ge sa iene ne nn tnt 139 C 2 First Stage of Chrome Removal siennes 140 C 3 Second Stage of Chrome Removal sise 141 CA Batch Treatments entre Ge Ad UE a dre ts 142 C 5 ORP Potential a Measure of Status ee 142 D SIMULTANEOUS PH AND ORP MEASUREMENT sn 143 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual vii E PROPORTIONAL CONTROL TUNING nn rrrrennemmennnnnmmmnnennennennte 145 EI OVERVIEW E E EIAS EE E ce 145 F MICROPROCESSOR BOARD SWITCH SETTINGS cccccssecccseeceeseeeseceeeeees 147 F 1 EE 147 viii 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Tables Table J l Key F unctonss oera reai t rta aa EE ee CA 4 Table 1 2 9782 pH Analyzer Controller Menu Hierarchy 6 Table 1 3 Overview of Analyzer Controller Installation Tasks AA 11 Table 3 1 Procedure for Unpacking and Preparing the 9782 20 Table 4 1 Procedure for Installing AC Power Wiring ss 30 Table 5 1 Menu Choices and System Defaults A 35 Table 5 2 Procedure for Performing I O Setup ss 40 Table 5 3 Procedure for Configuring CAT Control and or Retransmission of Process Variables 44 Table 5 4 Procedure for Configuring On Off Control 47 Table 5 5 Procedure for Configuring PFT Control ss 48 Table 5 6 Procedure for Configuring DAT Control 49 Table 5 7 Procedure for Configuring Alarms ss 50 Table 5 8 Procedure for Setting the Clock sise 51 Table 5 9 Procedure for Configur
88. electrodes manually using 8 6 3 automatic buffer recognition AUTO BUFFER SETUP select buffers for automatic buffer 8 6 2 recognition CAL DIAGNOSTICS display standardization and slope values 8 7 6 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Introduction Purpose Described in Section MAINTENANCE do occasional system tasks 9 and 10 OFF LINE FUNCTIONS see functions listed below see below test display 9 5 1 test keyboard 9 5 1 OUTPUT TESTS OUTPUT CALIBRATION DISPLAY TEST KEYBOARD TEST 9 5 2 test relays and analog outputs calibrate outputs electrically 10 2 CALIBRATION RESET set standardization offset and slope values 10 3 back to zero INSTRUMENT SETUP see functions listed below see below INSTRUMENT TAGGING configure tagname or other text to be 10 4 displayed when no alarms are active PRODUCT INFO view software version and stored model 10 5 number if a special technical support password is known alter model number in memory in case of hardware upgrades SCREEN CONTRAST adjust display contrast 10 6 SECURITY assign password to limit access to 10 7 configuration calibration and maintenance functions RESET UNIT reset all configuration and calibration 10 8 values back to factory settings ELECTRODE FUNCTIONS see functions listed below see below SELECT ELECTRODE SOLUTION TEMP COMP DURAFET TEMP DISPL
89. ended Connect the ac mains through a fused disconnect switch To ensure that the unit meets the immunity levels specified by EMC directive 89 336 EEC install power line filter kit p n 079163 Wind three turns of each wire 14 AWG maximum through the filter core for the two power and protective earth leads as illustrated in Figure 4 1 The wound filter cores shall be located external to the 9782 case within 25 cm 28 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Power Wiring Conform to code Instrument wiring should conform to regulations of the National Electrical Code 4 3 Power Wiring Considerations Recommended wire size Observe all applicable electrical codes when making power connections Unless locally applicable codes dictate otherwise use 14 gauge 2 081 mm wire for ac power including protective earth Power supply voltage and frequency within specs The power supply voltage and frequency must be within the limits stated in the specifications in Section 2 Power for external preamp The preamplifier power supply at terminals V V and SC provides a nominal unregulated 10 volts and 10 volts for an external preamp If the 31022283 preamp is used it will perform correctly from this supply although its terminals are designated 16 16 and SC 3 00 9782 Series pH ORP Analyzer Coniroller Operator s Manual 29 Power Wiring 4 4 Installing Power Wiring Procedure Follow the
90. enoid value used to direct the reference solution to the electrode see Section 7 for piping 8 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 3 4 5 Introduction e Using a second reference solution for AutoCal permitting slope adjustment to match instrument gain to the electrode output response requires Relay 3 to activate the solenoid valve for the second reference solution Use of relays for one of the discrete control strategies see 1 5 3 has third priority e Up to four relays can be used for On Off cycle timers e One or two relays can be used for Duration Adjusting Type DAT control e One or two relays can be used for Pulse Frequency Type PFT control Use of a relay for DualCat switchover has next priority permitting two reagents to be fed using a single analog output for Current Adjusting Type control Lowest priority is given to use of relays for activation of an annunciator for process alarms and or system errors Note that any problems found by the Analyzer Controller s self diagnostics will always be displayed on the alarm stripe 1 5 3 Deciding on a Control Strategy All 9782 models can use relays for control All 9782 models have at least two relays Two more relays are optional These relays can be used to control process variables using one of the three types of discrete control available e On Off Control using one to four cycle timers with configurable setpoint deadband
91. erator can make the operation pause using the soft key on the special AutoClean display If the operator does not remove the pause by pressing the PAUSE soft key again the Analyzer Controller will resume normal operation after 20 minutes rinse cleaning ies i solution SC Y gt to process fe i ltems outside this area provided by user SK process sample Figure 7 1 Automatic Electrode Wash Setup 78 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 AutoClean and AutoCal Theory and Piping Select valves and fittings with appropriate pressure ratings Make the process connections as shown in Figure 7 1 Be sure that valves and fittings S1 have sufficient pressure ratings to withstand pressure peaks which will occur when process flow is blocked Minimize liquid volume in system Keep pipe sizes small and couplings close to minimize the liquid volume in the system Smaller volumes require less time to rinse 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 79 AutoClean and AutoCal Theory and Piping 7 3 AutoCal Sequence and Piping Introduction AutoCal can include one point calibration standardization to adjust zero to compensate for electrode drift or two point calibration standardization and slope adjustment to also adjust span to match the gain of the Analyzer Controller to the e
92. ettling period then begins or the batch is pumped to a separate settling tank or pond C 5 ORP Potential a Measure of Status Sulfite is reducng ion An oxidation reduction reaction involves the transfer of the electrons from the reducing agent to the ion being reduced In the chrome removal application sulfur in the sulfite ion donates electrons to reduce the chromium simultaneously the chromium oxidizes the sulfur The ORP potential is a measure of the status of the oxidation reduction reaction the platinum or gold electrode detects the solution s ability to accept or donate electrons Sulfite SO a reducing ion donates electrons which makes the electrode more negative The chromium an oxidizing ion Cr accepts electrons and makes the electrode more positive The net electrode potential is related to the ratio of concentrations of reducing and oxidizing ions in the solution Potential cannot be used as monitor of effluent This electrode potential is extremely sensitive in measuring the degree of chrome treatment in the reaction tank However it cannot be related to a definite concentration of chrome and therefore cannot be used as a final effluent monitor of chrome concentration Importance of clean electrode Reliable ORP measurements require a very clean metal electrode surface Clean the electrodes routinely with a soft cloth dilute acids and or cleaning agents to promote fast response Control at low pH levels in the first s
93. ettling tank or pond B 5 ORP Potential a Measure of Status of Reaction Cyanide is reducing ion An oxidation reduction reaction involves the transfer of electrons from the ion being oxidized to the oxidizing agent In cyanide destruction chlorine or hypochlorite accepts electrons from the cyanide oxidizing it while simultaneously the hypochlorite is reduced to chloride ORP potential is a measure of the status of the oxidation reduction reaction 1 e the gold electrode detects the solution s ability to accept or donate electrons The hypochlorite an oxidizing ion accepts electrons which makes the electrode more positive The cyanide a reducing ion provides electrons and makes the electrode more negative The net electrode potential is related to the ratio of concentrations of reducing and oxidizing ions in the solution Potential cannot be used as monitor of effluent This electrode potential is extremely sensitive in measuring the degree of treatment in the reaction tank However it cannot be related to a definite concentration of a cyanide or cyanate therefore it cannot be used as a monitor of final effluent concentration Importance of clean electrode Reliable ORP measurement requires a very clean metal electrode surface Routinely clean the electrodes with a soft cloth dilute acids and or cleaning agents to promote fast response 138 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Chrome Waste Treatment
94. ff Line Diagnostics Introduction The following off line diagnostics can be initiated by the operator e display test see 9 5 1 e keyboard test see 9 5 1 e output tests see 9 5 2 9 5 1 Display and Keyboard Test Procedure The procedure for initiating the display and keyboard test is described in Table 9 4 Table 9 4 Procedure For Initiating Display and Keyboard Tests Step Screen Action 1 SE Go to MAINTENANCE OFF LINE FUNCTIONS CONFIGURATION CALIBRATION DISPLAY TEST MAINTENANCE I O SETUP DIAGNOSTICS PREV NEXT MAINTANENCE OFF LINE FUNCTIONS INSTRUMENT SETUP ELECTRODE FUNCTIONS RELAY ACTIVATION DIAGNOSTICS OFF LINE FUNCTIONS DISPLAY TEST KEYBOARD TEST OUTPUT TESTS OUTPUT CALIBRATION CALIBRATION RESET DIAGNOSTICS I I 2 Dee Be TEST Pressing the specified function key will display a test pattern on the first line of the display Press the key repeatedly to test the rest of the display DIAGNOSTICS After confirming that each line of the display is DIESTI functional press the MENU key to return to the off line functions menu DISPLAY TEST F1 TOGGLE MENU ABORT DIAGNOSTICS 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 105 Diagnostics and Messages Step Screen Action e jrs Go to MAINTENANCE OFF LINE FUNCTIONS
95. from the process for measurement Bring the portable instrument to the sampling site and measure a continuously flowing sample that has not been exposed to air This prevents lowering the sample pH by absorption of carbon dioxide from the air 88 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Manual Calibration 8 4 ORP Calibration Using Reference Solution Recommended to adjust for changes in electrode potential over time An ORP measuring system can be checked by measuring a solution having a known oxidation reduction potential then adjusting the 9782 to match Although a reference solution provides only an approximation of ORP potential the system can be adjusted periodically to compensate for changes in electrode potiential over time Materials The materials required to use the ORP standardization method are e A solution with a known oxidation reduction potential See Instructions for preparing solution below e A container for the solution large enough to immerse the electrode to measuring depth e Distilled or de ionized water to rinse the electrode Instructions for preparing solution To prepare an ORP standardization solution dissolve 0 1 g of quinhydrone powder available as Honeywell part number 31103015 for a 2 oz bottle in 5 cc of acetone or methyl alcohol methanol Add this to not more than 500 cc of a standard pH reference solution buffer about 1 part saturated quinhydrone
96. gasket glass ORP 51198172 501 pH input card card installation instructions 51198173 501 bezel assembly assembly 511981 76 501 Non backlit LCD display assembly 51198176 502 Backlit LCD display assembly 51198056 501 internal preamp for use with preamp mounting rivets installation instructions 51198057 501 51205501 501 51205502 501 internal preamp for use with Durafet Power supply PC Microprocessor PC preamp mounting rivets installation instructions card installation instructions card installation instructions 130 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Accessories and Replacement Parts List Kit Part Number Description Parts Included in Kit Quantity 51205503 501 Backplane PC card 1 installation instructions 1 51198059 501 relay standard general purpose used relay 1 on all 9782 models 51198060 501 relay optional hermetically sealed relay 1 51198061 501 relay optional general purpose relay 1 079163 power line filter kit filter cores 4 installation instructions 1 51198055 501 panel mounting kit bracket 1 screws 2 washers 2 084711 pipe wall mounting kit bracket 1 U bolt for 1 in pipe 2 U bolt for 2 in pipe 2 mounting plate 1 nuts 4 washers 4 lock washers 4 056349 reducer panel kit filler plate 1 31103015 quinhydrone powder 2 oz bottle 1 31103001 pH buffer s
97. gs to check that the analog output values match those selected Also listen for the click of an energized relay If the results of the analog output test are not satisfactory recalibrate the output s zero and span using the procedure in 10 2 If the results of the relay output test are not satisfactory consult Honeywell for assistance 9 ree a ee Select one of these combinations or use the specified 100 75 50 AAAN function key to return to the previous display 0 100 75 ANAA 0 0 100 AANA 0 0 O ANNN DIAGNOSTICS PREV 10 If only one meter is available so that only one output can be tested at a time then for each analog output repeat Steps 1 through 9 including powering down the unit before changing the connections to the output terminals 11 When output testing has been completed re install field wiring removed in Step 4 Disconnect power before opening case 12 Replace the safety cover and secure it with the screw 13 Close the case and power up the unit Do not apply power until case is closed 108 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Maintenance 10 Maintenance 10 1 Overview Introduction This section provides instructions for a variety of maintenance tasks A ETS Maintenance should be performed by qualified personnel only What s in this section The topics in this section are listed below Topic See
98. iagnostics and System Error Messages SYSTEM ERROR MESSAGE A D CONVERTER ERROR AUTOCAL SLOPE FAIL AUTOCAL STAND FAIL WHEN TESTED background at calibration at calibration background CONDITION faulty analog to digital converter failed slope value limit failed standardize value limit CORRECTIVE ACTION Have instrument serviced Check electrode and buffer solutions AutoCal only Check electrode and buffer solutions AutoCal only readings are out of range CLOCK RESET clock reset usually by power Reset day and time then turn loss clock off if not required RAM ERROR IN CPU at reset faulty RAM in microprocessor Have instrument serviced ELECTRODE ERROR background sensor readings are out of Check electrode connections range and presence of sample EPROM ERROR at reset faulty EPROM Have instrument serviced RAM DATA at reset data in non volatile memory Power down check power CORRUPTED has been changed by a fault supply and wiring for high voltage spikes Have instrument serviced if necessary RAM ERROR at reset faulty RAM Have instrument serviced TEMP COMP ERROR background faulty thermistor pH only Check electrode and connections ZERO ADJUST ERROR background sensor zero adjustment Check electrode then have instrument serviced if necessary 104 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Diagnostics and Messages 9 5 O
99. iary Relay Option Installed wire when using Durafet electrodes i For simul H RP Not for Durafet Electrodes with Cap Adapter Le a ango Output Terminals Relay Terminals N 3 Analyzer is factory set for 120 V or 240 V operation See directions supplied with Analyzer 4 Relay 3 and 4 contact ratings 3 A at Relay Terminals 120 Vac 28 Vdc resistive Case as viewed from front option PCA 5 RE SC jumper resistor must remain in To Preamp AS place except for simultaneous pH and Note 2 e NO es 3 ORP measurements 1 COM y xO NC_ Note 4 TH R ORE EE No ET W OTH NG CoM Relay 4 Nose Gen 1 NC JNotei SC K SC NC 16 B OV 16 G Qv OUT O or N OPA 1 NC Re Sen 2 COM Jeans Terminals for 7 2 NO Note 1 connection of 5 Se pe Color Code remote 6 NO Note 1 preamplifier Z L1 K 120V 240V K black Y yellow LAN W Juan aah E ee CH U e 50 60 Hz EE SES G green V violet Jee R red P pink Power amp Relay Terminals B blue S slate N brown T tan 9782 01 RO Preamp Input No Outputs Auxiliary Relay Option Installed O orange Not for Durafet Electrodes with Cap Adapter PE Protective Earth Ground Figure 6 5 Terminal Connections 70 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Input and
100. ibed in section 10 For Durafet II electrodes with a Cap Adapter a separate temperature sensor is not required 42 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 UO Setup and System Configuration 5 4 Configuring Current Adjusting Type CAT Control and or Retransmission of Process Variables Purpose If the hardware supports analog outputs during I O setup you had an opportunity to specify which process variables will be retransmitted or used as the basis for DualCAT or CAT control The DualCAT feature permits control of pH with two reagents alternately using a single output as described in Section 1 Before using an analog output for retransmission or control you must configure the related parameter values Regardless of whether you plan to retransmit a process value to a recorder or control a device based on that value the output signal will be directly proportional to the input However the configuration screens are different for the different output uses For a discussion of tuning proportional control see Appendix E Procedure The procedure for configuring CAT control and or retransmission of process variables is shown in Table 5 3 Hold mode To place the unit in Hold mode before configuring CAT control go to an online display screen and press the HOLD soft key To take the unit out of Hold mode after configuration go back to an online screen and press the HOLD soft key again 3 00 9782
101. ics and Messages 9 3 Process Alarm Messages Alarms dependent on configuration Alarms are configurable as described in Section 5 When an alarm is active it is always displayed on the alarm stripe All available alarm messages are shown in Table 9 2 Table 9 2 Process Alarm Messages MESSAGE WHEN DISPLAYED HIGH pH HIGH TEMP HIGH ORP Process value has exceeded high alarm or fallen below LOW pH low alarm the configured alarm setpoint LOW TEMP LOW ORP HI pH DIAG Process value has exceeded high alarm or fallen below HI TEMP DIAG low alarm the configured setpoint HI ORP DIAG LO pH DIAG OR LO TEMP DIAG LO ORP DIAG System error is active see 9 4 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 103 Diagnostics and Messages 9 4 On Line Diagnostics and System Error Messages Self tests at 0 5 second intervals On line diagnostics are self tests which are performed automatically as a background task every 1 2 second The system error messages displayed on the alarm stripe if a test is failed are listed in Table 9 3 Output affected by error In an error condition the Analyzer will continue to perform all the functions it is capable of except that the output signal s will go to 0 or on single parameter output models to the preset failover value if entered Output signal manipulation does not occur with calibration failure or clock reset diagnostics Table 9 3 On Line D
102. ile the unit is powered a potentially lethal shock hazard exists inside the case Do not open the case while the unit is powered Step Table 6 2 Procedure for Installing Analog UO Wiring Action Go to I O SETUP to view the displays showing analog input relay and analog output use Note the assignments shown You must wire the unit to match these assignments in order for the unit to work as expected See Section 5 Turn off the power to the Analyzer Controller More than one switch may be required to remove power With the power off open the case e Grasp the bottom center portion of the front bezel and pull it downward and toward you slightly to disengage the bottom of the bezel from the edge of the case e Lift the bezel gently to disengage it from the top edge of the case e Swing the bezel to the left The bezel and display assembly is mounted on pivot arms The safety cover with its warning label will be visible With the power off remove the safety cover by removing the single screw holding it in place 64 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Step Input and Output Wiring Action 5 Connect the inputs from the electrode to the terminals in accordance with the system setup assignments Refer to the wiring diagram provided with the electrode and or preamp and to Figure 6 1 through Figure 6 6 6 If a Durafet pH electrode is being used and y
103. ing AutoClean seen 54 Table 5 10 Procedure for Configuring One Point AutoCal ss 55 Table 5 11 Procedure for Configuring Two Point AutoCal ss 57 Table 6 1 Recommended Maximum Wire ze 63 Table 6 2 Procedure for Installing Analog I O Wiring ss 64 Table 6 3 Relay Contact Maximum Ratings sense 72 Table 6 4 Procedure for Wiring Relays sf him tit ele tte EE 73 Table 6 5 Procedure for Disconnecting RC Arc Suppression Circuits 76 Table 8 1 Procedure for Standardization and Slope Adjustment 85 Table 8 2 Procedure for Calibrating Using Grab Sample 87 Table 8 3 Oxidation Reduction Potential of Reference Solutions at Specified Temperature 89 Table 8 4 Procedure for Calibrating ORP System Using Reference Solution 90 Table 8 5 Procedure for Calibrating ORP Analyzer Controller Using Voltage Input 91 Table 8 6 Representative Buffer Values at Various Temperatures 94 Table 8 7 Buffer Specification Procedure ss 95 Table 8 8 Standardization Procedure Using Automatic Buffer Recognition 97 Table 8 9 Procedure for Viewing System Calibration Values 99 Table 9 1 System Status Messages iii 102 Table 9 2 Process EE 103 Table 9 3 On Line Diagnostics and System Error Messages 104 Table 9 4 Procedure For Initiating Display and Keyboard Tests 105 Table 9 5 Procedure for Testing Relays and Analog Outputs ss 107 Table 10
104. inum electrode because platinum may catalyze additional reactions at its surface and is more subject to coating than gold Note that the solubility of gold in cyanide solutions does not present a problem as it is in contact primarily with cyanate In fact a slight loss of gold serves to keep the electrode clean B 3 Second Stage of Cyanide Destruction Neutralize and further oxidize cyanate The waste water is neutralized in order to promote additional oxidation and to meet the discharge pH limits Typically sulfuric acid is added to lower the pH to about 8 5 At this pH the second oxidation occurs more rapidly WARNING Failure to comply with these instructions could result in death or serious injury An interlock must be provided to prevent the addition of acid before the positive oxidation of ALL cyanide Failure to observe this precaution can result in the generation of highly toxic hydrogen cyanide Additional chlorine or sodium hypochlorite NaOCl can be added in proportion to that added in the first stage or by separate ORP control to complete the oxidation to sodium bicarbonate NaHCO in the following reaction 2NaCNO 3NaOCl H O gt 2NaHCO N 3NaCl ORP control in the second stage is very similar to that in the first stage except that the control point is near 600 mV Control of pH in the second stage is more difficult than in the first stage because the control point is closer to the sensitive neutral area Prop
105. it must be compatible ATTENTION In Figure 6 1 Relay 1 is used for illustrative purposes only When wiring your unit use the relay assigned by the software during I O setup as described in Section 5 A Power amp Relay Terminals Voltage amp Current Output Terminals S 1 S 4 20 mA amp Related K Q 3 Control Outputs A Y S t Acid Feeder S i z 10 tp Feed ve S e _ Base Feeder er E BE Protective Earth Ground Figure 6 1 Dual Reagent CAT Typical Wiring Using Relay 1 to Select Reagent Feeder 66 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Input and Output Wiring From Cap Case as viewed from front Adapter with A Option Card x Durafet Il Electrode osition 2 i S i Output Board R ORE ao a eae R B LS a OMT fa Connections W RE 1 4 20 mA 600 max ae 4 20 mA Note6 K i 6 T 3 B ep E min OU S Z x 1V 200 2 min Q1 N EN 2 Gel Relay 2 C2 GR Note 1 SE A Qal Zu Ka St gs NO Relay 1 LS oa on 120V 240V Ga Microprocessor LC O KL W 15A BI E Ser Ze ri E 50 60 Hz DR PE Note 3 Voltag
106. its Sodium hydroxide or lime is used to raise the pH to 7 5 to 8 5 in the following reaction Cr2 SO 6NaOH gt 3Na SO 2Cr OH pH control point close to neutral point Control of pH in the second stage is more difficult than in the first because the control point is in the sensitive area closer to the neutral point Although this reaction is fast for stability a retention time of at least 10 minutes is usually needed for continuous treatment Proportional pH control is often used in this stage Remove suspended chromium hydroxide Subsequently a settling tank and or filter will remove the suspended chromium hydroxide Flocculating agents are helpful in this separation 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 141 Chrome Waste Treatment CA Batch Treatment Sequence of steps Continuous treatment for chrome removal is shown in Figure C 1 However all of the reactions can be achieved with semi automatic batch control Only a single tank with a pH controller and an ORP controller are required The steps of the treatment are sequenced and the pH setpoint is changed to give the same results as for the continuous treatment Acid is added to lower pH to 2 5 then reducing agent is added to lower ORP potential to 250 mV After waiting a few minutes to ensure a complete reaction and possible test for Cr the sodium hydroxide is added to raise pH to 8 as in the second stage of the continuous treatment The s
107. l or pipe mounted see 3 4 ATTENTION Pipe mounting is not recommended if the pipe is subject to severe vibration Excessive vibration may affect system performance 6 If extremely hot or cold objects are near the installation location provide radiant heat shielding for the instrument 20 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Unpacking Preparation and Mounting Honeywell York PA U S A ATTENTION UNIT SET FOR 240 VAC 9782P 01 VC E0010 00 S N 9751Y712345670001 USA WARRANTY ASSISTANCE 1 800 423 9883 120 240 VAC 50 60 Hz 15VA Contact Rating 1 amp 2 0 6 Amps at 120 240 VAC 0 6 Amps at 110 VDC and 2 Amps at 30 VDC If installed 3 amp 4 3 Amps at 120 VAC 28 VDC CE Directions 70 82 25 73 Figure 3 1 Sample Nameplate 3 3 Installing Optional Internal Preamplifier Instructions elsewhere If an optional internal preamp is to be installed in the 9782 follow the directions provided with the preamp kit 3 00 9782 Series pH ORP Analyzer Coniroller Operator s Manual 21 Unpacking Preparation and Mounting 3 4 Mounting Introduction The 9782 Analyzer Controller can be panel wall or pipe mounted Panel mounting Figures 3 2 1 and 3 3 1 show panel cutout dimensions They also illustrate how the instrument is attached to a panel by clamping the edges of the cutout between the case flange and the supplied U bracket which is fastened to the rear
108. lectrode response Standardization and slope each depend on the electrode measuring a reference solution buffer of known pH Rinse and one point calibration sequence The AutoCal operation automatically occurs at the configured intervals It always includes AutoClean rinsing of the pH electrode in addition to any other AutoClean sequences that are configured to occur between standardization operations The sequence is described below 1 All alarm action is held at existing levels The output s can be held or be active depending on configuration Even if the outputs are not held HOLD ACTIVE is displayed on the alarm stripe because alarms are always held Also AUTOSEQUENCE is displayed Pressing the DISPLAY key will call up a special display that shows how much time is left in the operation 2 In preparation for the calibration Relay 1 activates 3 way solenoid valve S1 see Figure 7 2 to direct rinsing fluid to the electrodes for the configured rinse duration 1 to 1999 seconds If the measured sample is normally returned to the process but quantities of rinsing fluid cannot be tolerated there use an additional 3 way solenoid valve S4 It is activated simultaneously with S1 to divert the discharge to drain 3 Relay 2 activates solenoid valve S2 for the preset buffer time 1 to 1999 seconds to direct buffer solution past the electrodes by gravity 4 After a stable reading is reached or the set maximum buffer time elapses th
109. lectrodes Automatic buffer recognition The instrument automatically recognizes six commonly used buffer solutions and the temperature variation characteristic of each This simplifies manual calibration and permits automatic calibration 2 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Introduction Password protection A password up to four digits can be configured If the security feature is enabled the password will be required to access configuration calibration and maintenance software functions Extensive diagnostics The 9782 Analyzer Controller performs extensive self diagnostics as a background task during normal operation If a problem is detected a message is displayed on the alarm stripe to alert the operator In addition the operator can initiate keypad and display tests using Maintenance Menu functions Watertight corrosion resistant case The 9782 is enclosed in a watertight and corrosion resistant industrial case designed for panel pipe or wall mounting It has an EMI RFI shielded plastic case rloneyyell Figure 1 1 9782 Front Panel 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual Introduction 1 3 Operating the Analyzer Controller Front panel keys used for all operator tasks As shown in Figure 1 1 five keys with dedicated functions are on the front panel In addition three soft keys vary their function according to the needs of the scree
110. low the procedure in Table 5 8 to set the 24 hour clock Table 5 8 Procedure for Setting the Clock Step Screen Action 1 BEER Go to CONFIGURATION ADVANCED FEATURES CALIBRATION CLOCK SETUP MAINTENANCE I O SETUP DIAGNOSTICS T PRE NEXT CONFIGURATION CAT RETRANSMISSION ALARMS DISCRETE CONTROL ADVANCED FEATURES DIAGNOSTICS PREV NEXT ADVANCED FEATURES CLOCK SETUP AUTO CLEAN DIAGNOSTICS 2 ee ae Set the clock For example if you will start the clock at CLOCK ON W RESET 1 30 PM enter 13 hours 30 minutes HOURS 13 Ee 38 Start the clock and specify what should happen to the DIAGNOSTICS clock if the power fails NEXT e ON W RESET starts the clock now Following power failure the clock will be reset to 00 00 and the diagnostic message RESET CLOCK will be displayed e ON W O RESET starts the clock now Following power failure the clock will remain off and the diagnostic message RESET CLOCK will be displayed Display of RESET CLOCK will continue until you either reset the clock or turn it off 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 51 UO Setup and System Configuration 5 8 Configuring AutoCal and AutoClean Purpose Periodic calibration of electrodes is necessary for best system performance because electrode outputs change over time One point calibration standardizatio
111. ly different potentials in the same solution requiring a corrective adjustment at the measuring instrument Also electrode outputs change over a period of time making periodic re calibration necessary for best performance Determine re calibration intervals based on operating experience This section contains instructions for calibrating electrodes manually To calibrate electrode automatically using the the AutoCal feature follow the configuration instructions in Section 5 and install piping and valves as shown in Section 7 What s in this section The topics in this section are listed below Topic See Page 8 1 Overview 83 8 2 Buffering Method of Calibrating pH Electrodes 85 8 3 Grab Sample Method of Calibrating pH Electrodes 87 8 4 ORP Calibration Using Reference Solution 89 8 5 ORP Calibration Using Voltage Input 91 8 6 Calibrating pH Electrodes Using Automatic Buffer Recognition 94 8 7 Viewing Zero Offset and Percent Theoretical Slope 99 8 8 Recommendations for Successful Measurement and Calibration 100 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 83 Manual Calibration pH calibration pH instrument calibration consists of standardization and slope adjustments Standardization is a zero adjustment to compensate for electrode drift Slope adjustment is a span adjustment to match the gain of the instrument to the electrode output response The 9782 supports
112. millivolt signal to the Analyzer Controller instead of input from the electrode then adjust the 9782 reading to match the actual millivolt input as described in 8 5 IMPORTANT In order to calibrate the instrument if the simultaneous pH ORP technique is to be used all three electrode connections must be made according to the drawing in Appendix D IMPORTANT If you plan to measure pH glass and ORP simultaneously then before doing an ORP calibration the first time you must do a complete pH buffer calibration standardization and slope Once the first pH and ORP calibrations have been performed there is no further restriction on the order of pH and ORP calibration Multiple ORP calibrations can be performed between pH calibrations 84 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Manual Calibration 8 2 Buffering Method of Calibrating pH Electrodes Recommended for most applications This technique is recommended for best accuracy in most applications Materials and procedure To perform manual standardization and slope adjustment follow the instructions in Table 8 1 Materials required are Two standard buffer reference solutions that are at least 2 pH different from one another A container for each large enough to immerse the electrode to measuring depth Distilled or de ionized water to rinse the electrode Table 8 1 Procedure for Standardization and Slope Adjustment
113. n 89 8 5 ORP Calibration Using Voltage Input 91 8 6 Calibrating pH Electrodes Using Automatic Buffer Recognition 94 8 6 1 Introductions acc iene ses en SEENEN ae ea 94 8 6 2 Choosing Neutral and Alkaline Buffers AAA 95 8 6 3 Manual Calibration Using Automatic Buffer Recognition 96 8 7 Viewing Zero Offset and Percent Theoretical Slope cece ceeceseeeseeceeeeeeeeeeeeeeeeeeseeseeeeees 99 8 8 Recommendations for Successful Measurement and Calibration cee eeeeeeeeseeereeeneee 100 9 DIAGNOSTICS AND MESSAGES nn nnsennmmmmnnennnnnmnnnennnnnenneenne 101 9 1 ON TE 101 9 2 Systems tats Message hentai Webel dei dt SEA 102 9 3 Process Alarm Messages stats mettent Ne EENS 103 9 4 On Line Diagnostics and System Error Messages scesseesseceseeeseeenseceseceseecaecaeenaeenseees 104 9 5 Off Line Diagnostics ss cn nn NEE NEE et beanies 105 9 5 1 Display and Keyboard Test 105 OD GER e KEE 107 vi 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 10 MAINTENANCE nil leriiitesenasdtigentetegein au tete aa tets tien gere tn ad ei ten ad nca 109 10 1 OVERVIEW mn dant te a tn date nee en den laine de Pes tendons 109 10 2 Output Calibration ea feet e re nn dE ne ft ain 111 10 3 Restoring Factory Calibration Settings 0 cece eecseeeesceseeeseeeeeeeseneesaeeeaeecaaeesaeeeaeeeaaeenaees 114 10 4 Specifying a Tagname or Other Display String 115 10 5 Viewing Product Information and Changing Model Number S
114. n AutoClean is a feature that lets the Analyzer Controller periodically rinse the electrodes automatically Configure AutoClean as described in Section 5 8 34 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 UO Setup and System Configuration AutoCal Ipt This is a feature used in conjunction with AutoClean that lets the Analyzer Controller periodically standardize the electrodes automatically Configure AutoCal 1pt as described in Section 5 8 AutoCal 2pt This is a feature used in conjunction with AutoClean that lets the Analyzer Controller periodically standardize and slope the electrodes automatically Configure AutoCal 2pt as described in Section 5 8 Illustrations The procedures in this section contain illustrations of display screens These pictures are intended to show the types of information displayed on each screen However the screens you will actually see on your Analyzer Controller may be different from the pictures in the manual The displays will be customized for the hardware and application System defaults Table 5 1 contains the menu choices available or ranges for numeric values and system defaults for I O assignments and configuration parameters as well as for values entered during maintenance functions described in Section 10 Table 5 1 Menu Choices and System Defaults PARAMETER CHOICES or RANGE Default is underlined UO setup input 1 9782P 03 DURAFET without Cap Ada
115. n is a zero adjustment to compensate for electrode drift Two point calibration standardization and slope adjustment includes a span adjustment to match the gain of the Analyzer Controller to the electrode response Standardization and slope each depend on the electrode measuring a reference solution buffer of known pH The AutoCal and AutoClean features periodically rinse and calibrate electrodes automatically see Section 7 If these features were selected during I O setup described in 5 3 configure control of AutoCal and AutoClean as described below If manual calibration will be used instead see Section 8 AutoCal and AutoClean configuration consists of setting the clock and specifying the frequency and duration of the cleaning and calibration operations ATTENTION For Auto Clean Cal to operate the 24 hour clock must be set as described in 5 7 The piping and valves required by AutoCal and AutoClean are shown in Section 7 Timer initiated operation With the 24 hour clock turned ON and the AutoClean feature enabled and the appropriate piping in place this mode performs unassisted cell cleaning The AutoClean Time Remaining on line display provides an up to date time stamp of the next occurrence of an automatic cell cleaning Once the timer reaches zero the cell cleaning will occur Upon completion the timer will be reset and the Time Remaining display will again provide a countdown to the next cleaning To temporarily disa
116. n for special applications and proportional control tuning tips appendices 12 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Specifications and Model Number 2 Specifications and Model Number 2 1 Specifications Display LCD dot matrix display 128 by 64 dpi Backlit display solid state LED Displays pH ORP temperature time alarm conditions alarm setpoints calibration output limits and diagnostics Display Ranges pH 2 00 pH to 14 00 pH resolution 0 01 pH ORP 1638 to 1638 mV resolution 1 mV Temperature 10 to 140 C resolution 1 C Display Scan Mode Key used to cycle through displays of measured parameters and autosequence operation status Alarm messages status messages and diagnostic messages available on all displays Keypad Monoplanar front panel with 8 keys Push button entry with tactile feedback Auto Buffer Recognition Recognizes 4 01 6 86 7 00 9 00 9 18 10 00 pH standard buffer solutions with temperature correction from 0 C to 60 C 32 F to 140 F AutoCal Automatic Calibration and AutoClean Electrode Washing Schedule Settings Clock Cycle 28 days with day hour minute resolution Rinse Time 60 1999 seconds Buffer Duration 60 1999 seconds Resume Process Time Delay 1 1999 seconds Calibration Interval Adjustable number of rinse cycles between one point calibrations or number of one point calibrations between two point calibrations
117. n of process variables or for Current Adjusting Type control using an output signal that is directly proportional to the input see 1 5 3 Relays Two 2A SPDT alarm control relays are standard with an additional two 3A relays either general purpose or hermetically sealed available as an option These relays can be used to e take advantage of special features such as automatic cleaning of electrodes e control process variables see 1 5 3 e annunciate alarms 1 2 Features Storage of calibration data for two electrodes The DualCal feature allows storage of calibration data for two electrode assemblies The second electrode can be used as a manual backup or in a two tank batch control setup Standard and solution temperature compensation In addition to standard temperature compensation for electrode response to temperature the 9782 Analyzer Controller also features solution temperature compensation to adjust for the solution s change in pH as temperature changes These temperature induced changes are significant in analysis of high purity water Automatic electrode washing and calibration The standard AutoClean and AutoCal features periodically rinse and calibrate electrodes automatically The schedule and duration of the operations are configurable AutoCal can be configured for one or two point calibrations Internal relays actuate external solenoid valves to control the flow of the buffer solutions and rinse water to the e
118. n on display Use of the keys 1s described in Table 1 1 Table 1 1 Key Functions DISPLAY When process values are on display Use DISPLAY to cycle through available real time displays When the main menu Configuration Calibration Maintenance UO Setup is on display Use DISPLAY to return to displaying process values MENU When process values are on display Use MENU to access the menus If the security feature is enabled the display will prompt for entry of the password before access to the menus is permitted Enter the password using the procedure for editing a parameter numeric value as described on the next page When any menu or configuration screen is on display Use MENU to go up a level in the display hierarchy Frequently this means returning to the main menu UP When a menu or configuration screen is on display Use UP to highlight a different item When changing a numerical value Use UP key to increment the value of the digit at the cursor DOWN When a menu or configuration screen is on display Use DOWN to highlight a different item When changing a numerical value Use DOWN to decrement the value of the digit at the cursor ENTER When a menu item is highlighted Use ENTER to select it When editing a parameter Use ENTER to save the new value F1 F2 F3 When a soft key label is displayed below the alarm function stripe Use the function key directly below the
119. nalyzer Controller Operator s Manual 3 00 Step Screen Action Maintenance 10 OUTPUT1 SPAN OUTPUT1 ZERO OUTPUT2 SPAN OUTPUT2 ZERO OUTPUT3 SPAN OUTPUT3 ZERO The first display will reflect the quantity of analog outputs DIAGNOSTICS SPAN CAL OUTPUT1 US E UP DOWN ARROW EYS TO ADJUST ENTER TO SAVE Use the UP and DOWN keys to correct the reading on the DIAGNOSTICS QUT PUT CALIBRATION SAVED ENU TO EXIT This screen confirms that the value was saved DIAGNOSTICS supported by the hardware Select the value to be calibrated Go to the next screen test meter To save the correction press ENTER AA AA To calibrate additional values repeat Steps 1 through 10 including powering down the unit before changing the connections to the output terminals When output calibration has been completed re install field wiring removed in Step 4 Disconnect power before Opening case 13 AA Replace the safety cover and secure it with the screw Close the case and power up the unit Do not apply power until the case is closed 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 113 Maintenance 10 3 Restoring Factory Calibration Settings Introduction A pH measuring system can be calibrated using one and two point manual or automatic calibratio
120. nd One Point Calibration Rinse and two point calibration sequence With this function rinse and one point standardization operations are performed as described previously according to the configured schedule If two point calibration is to be performed periodically then after the configured number of standardization operations Steps 4a and 4b shown below are also performed before Step 5 above to make the slope adjustment 4a Relay 3 activates solenoid valve S3 for the configured buffer time to direct the second buffer flow to the electrodes 4b After stability is reached or the set maximum buffer time elapses the instrument calculates and stores a new slope value using automatic buffer recognition Diagnostics detect excessive instability or offset prevent erroneous calibrating and can activate an alarm depending on configuration If the diagnostic fails an error message is always displayed on the alarm stripe see Section 9 If an unacceptable value is obtained it will be rejected and the previous value will be retained for uninterrupted operation Select piping and valves based on chemical resistance and pressure ratings Make the process connections as shown in Figure 7 2 or Figure 7 3 Materials and components should be carefully selected for chemical resistance to process and buffer solutions at anticipated temperatures Be sure that valves and fittings have sufficient pressure ratings to withstand pressure peaks which will occur
121. ns One point calibration also know as standardization is a Zero adjustment to compensate for electrode drift Two point calibration also includes a slope adjustment of span matching gain of the instrument to the electrode output response A standardization operation can also be performed on an ORP measuring system The calibration reset function described here undoes standardization and slope adjustments This function has no effect on output calibration as described earlier in this section Procedure Use the procedure in Table 10 2 to restore factory calibration values returning the standardization offset to 0 0 and the percent theoretical slope to 100 Table 10 2 Procedure for Restore Factory Calibration Values Step Screen Action 1 eh Go to MAINTENANCE OFF LINE FUNCTIONS CONFIGURATION CALIBRATION CALIBRATION RESET MAINTENANCE I O SETUP DIAGNOSTICS PREV NEXT MAINTANENCE OFF LINE FUNCTIONS INSTRUMENT SETUP ELECTRODE FUNCTIONS RELAY ACTIVATION DIAGNOSTICS OFF LINE FUNCTIONS DISPLAY TEST KEYBOARD TEST OUTPUT TESTS OUTPUT CALIBRATION CALIBRATION RESET DIAGNOSTICS 2 EE Pressing the specified function key at this point will return ALL CALIBRATIONS the standardization offset to 0 0 and the percent theoretical WILL BE RESET T FACTORY EEN slope to 100 0 DIAGNOSTICS FCAL 3 Wait for the display to go back to th
122. ntering the values to be represented by outputs of 0 and 100 If reverse acting output is required enter the high range value at the 0 prompt and the low range value at the 100 prompt For output options VC or RO the FAILOVER feature is available If failover is enabled here then when the unit s self diagnostics find a malfunction the output will go the failover value entered on this screen When FAILOVER is ENABLED or DISABLED the 0 and 100 values are changed back to factory defaults 44 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 UO Setup and System Configuration Step Screen Action 4 If DualCAT is used specify the range for control of each DUALCAT pH reagent CAT1 SP 7 pH CAT1 PB 2 PH Enter the value at which the output switchover must occur CAT2 SP 7 pH CAT2 PB 14 pH SWITCHOVRRLY 7 pH DIAGNOSTICS Te 7 gt 5 ae LL be For each output to which a CAT based on pH or ORP was CAT2 SP 7 pH assigned during I O setup the PV will be displayed and Merete mets the appropriate unit of measure will be filled in This is read only information DIAGNOSTICS El Specify the setpoint and proportional band limit 6 When all output parameters have been configured take the unit out of Hold mode 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 45 UO Setup and System Configuration 5 5 Configuring On Off Duration A
123. o solutions was less than 2 Use a different solution e The glass electrode is cracked Replace it e The attempted slope adjustment is too great Replace the electrode 86 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Manual Calibration 8 3 Grab Sample Method of Calibrating pH Electrodes Recommended where pH is stable or for high purity water applications This method is recommended only where the pH is stable and changes very slowly It is also recommended for high purity water measurement applications Special instructions for high purity water applications are provided below Materials and procedure To use the grab sample method follow the instructions in Table 8 2 Materials required are e A clean beaker for collecting the sample e A calibrated portable instrument for measuring pH of the sample e Distilled or de ionized water to rinse the electrode Table 8 2 Procedure for Calibrating Using Grab Sample Step Screen Action 1 END Prepare the calibration meter CONFIGURATION CALIBRATION MAINTENANCE I O SETUP DIAGNOSTICS 2 PUTTEN CRE Go to CALIBRATION SAMPLE CAL ORP CAL SAMPLE CAL AUTO BUFFER CAL AUTO BUFFER SETUP CAL DIAGNOSTICS 3 EE Put the unit in Hold mode ATTENTION DO NOT REMOVE ELECTRODE FROM DO NOT remove the electrode from the process PROCESS STAGNOSTICS Go to the next screen HOLD NEXT
124. ocess ORP value TEMPCAT retransmission of process temperature value N A default for Outputs 2 and 3 pH TEMP ORP p D pH or self diagnostic T D temperature or self diagnostic O D ORP or self diagnostic N A UO setup alarm associated with relay two relay units only one to one correspondence between relay and alarm is fixed when four alarms are available ALM1 alarm 1 default for Relay 1 ALM2 alarm 2 default for Relay 2 ALMS alarm 3 ALM1 2 alarm 1 and or 2 ALMA alarm 4 N A UO setup relay alarm action configuration retransmission 0 configuration retransmission 100 configuration CAT retransmission failover usage only available with VC or RO output options HIGH high alarm default for Relay 1 and Relay 3 LOW low alarm default for Relay 2 and Relay 4 N A pH 2 00 to 14 00 default 0 ORP 1999 to 1999 mV default 0 TEMP 10 to 140 1 C default 0 PH 2 00 to 14 00 default 14 00 ORP 1999 to 1999 mV default 1999 TEMP 10 to 140 1 C default 140 DISABLED ENABLED configuration CAT retransmission failover value only available with VC or RO output options and when Failover ENABLED pH 2 00 to 14 00 default 14 00 ORP 1999 to 1999 mV default 1999 36 9782 Series pH ORP Analyzer Controller Operator s Manual configuration setpoint configuration configuration proportional band
125. olution 4 01 pH 25 C 16 fluid oz bottle 1 31103002 pH buffer solution 6 86 pH 25 C 16 fluid oz bottle 1 31103003 pH buffer solution 9 18 pH 25 C 16 fluid oz bottle 1 _Preamp to instrument cables for 7758 06 7773 70 7774 10 50 70 and 90 and all 7777 mountings 31075723 20 ft length with quick disconnect and cable 1 tinned leads 51309677 001 50 ft length with quick disconnect and cable 1 tinned leads 51309677 002 100 ft length with quick disconnect cable 1 and tinned leads 51309677 003 200 ft length with quick disconnect cable 1 and tinned leads pre amp to instrument cables for 7758 01 05 7773 10 12 50 52 60 62 and 7774 30 60 80 mountings 834023 six conductor cable specify length cable 1 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 131 Accessories and Replacement Parts List 132 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Temperature Measurement Characteristics 3 00 40 000 20 000 10 000 5 000 4 000 2 000 1 000 800 600 Resistance in Ohms 400 200 A Temperature Measurement Characteristics 0 10 20 30 40 50 60 70 80 90 100 110 Temperature in Degreees Celsius 120 130 140 Figure A 1 Resistance vs Temperature Characteristics of the Temperature Compensator 9782 Series pH ORP Analyzer Controller O
126. on The 9782 is available with one or more optional analog outputs The output signals can be adjusted to trim the high and low output current or voltage values over a range of 0 4 of span to compensate for component tolerance variations Required equipment Output calibration involves connecting a meter to the Analyzer Controller s output terminals The meter required for output calibration depends on the type of outputs e current outputs current meter capable of resolving 0 01 mA over the range 0 to 20 mA dc e voltage outputs a 250 ohm 0 05 shunt and a volt meter capable of measuring 1 to 5 Vdc within 1 mV A screwdriver to fit the terminal block screws and the screw securing the safety cover is also required Procedure To calibrate outputs follow the procedure described in Table 10 1 The output terminals are inside the case as shown in Figures 6 1 through 6 6 d A WARNING While the unit is powered a potentially lethal shock hazard exists inside the case Do not open the case while the unit is powered Do not access the output terminal as described below while the unit is powered d A WARNING A disconnect switch must be installed to break all current carrying conductors Turn off power before working on conductors Failure to observe this precaution may result in serious personal injury 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 111 Maintenance Table 10 1 Procedure for Calibr
127. on limits of EN 50081 2 are designed to provide reasonable protection against harmful interference when this equipment is operated in an industrial environment Operation of this equipment in a residential area may cause harmful interference This equipment generates uses and can radiate radio frequency energy and may cause interference to radio and television reception when the equipment is used closer than 30 m to the antenna e In special cases when highly susceptible apparatus is used in close proximity the user may have to employ additional mitigating measures to further reduce the electromagnetic emissions of this equipment Sensing and Control Honeywell 11 West Spring Street Freeport Illinois 61032 AutoCal AutoClean DualCal and Durafet are trademarks of Honeywell Noryl is a trademark of GE Company Other brand or product names are trademarks of their respective owners 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 About This Document Abstract This manual contains instructions for installation and operation of the 9782 Series pH ORP Analyzer Controller Contacts World Wide Web The following lists Honeywell s World Wide Web sites that will be of interest to our customers Honeywell Organization WWW Address URL Corporate http www honeywell com Sensing and Control http www honeywell com sensing International http www honeywell com Business global asp Telephone Contact us
128. one of these control related parameters will take effect as soon as it is entered UNLESS THE 9782 IS IN HOLD MODE Use this screen to select the PFT control output to be configured The input type relay assignment and appropriate unit of measure will already be filled in based on the process variable assigned to this PFT control output during I O setup This is read only information Enter the setpoint at which the output should be turned on Enter the PB Limit This is the value for which the output s pulse output rate will be at its maximum frequency Specify the maximum frequency rate representing 100 output Be sure this rate is compatible with the controlled device The rate specified for the last PFT control output configured will apply to all PFT outputs Take the unit out of Hold mode 48 9782 Series pH ORP Analyzer Coniroller Operator s Manual 3 00 Table 5 6 UO Setup and System Configuration Procedure for Configuring DAT Control Step Screen MENU CONFIGURATION CALIBRATION MAINTENANCE I O SETUP CONFIGURATION CAT RETRANSMISSION ALARMS DISCRETE CONTROL AUTO CAL CLEAN DIAGNOSTICS DISCRETE CONTROL ON OFF PET DAT DIAGNOSTICS DAT CONTROL DATI DAT2 DIAGNOSTICS HOLD NEXT DATI INPUT pH OUTPUT RLY1 SETPOINT 9pH PB LIMIT 14pH CYCLE PERIOD _120Hz DIAGNOSTICS HOLD lt
129. ons In our example timer setup was configured for rinses to be done every third day If during rinse setup two rinses are scheduled to occur between standardization operations then the one point AutoCal standardization will be performed every sixth day Go to the next screen AUTO CAL 1PT STANDARDIZE SETUP DURATION 70s DELAY 90s DIAGNOSTICS NEXT Enter the length of time in seconds 60 is the minimum The standardization reference solution should flow past the electrode Enter the number of seconds the Analyzer Controller should wait after the AutoCal operation before resuming normal operation 56 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Table 5 11 Procedure for Configuring Two Point AutoCal UO Setup and System Configuration Step Screen MENU CONFIGURATION CALIBRATION MAINTENANCE I O SETUP CONFIGURATION CAT RETRANSMISSION ALARMS DISCRETE CONTROL ADVANCED FEATURES DIAGNOSTICS Go to CONFIGURATION AUTOCAL CLEAN Action ADVANCED FEATURES CLOCK SETUP AUTO CAL 2PT DIAGNOSTICS NEXT AUTO CAL 2PT ATTENTIO FUNCTION REQUIRES AUTO BUFFER SETUP IN CALIBRATION MENU DIAGNOSTICS NEXT AUTO CAL 2PT HOLD OUTPUTS YES DIAGNOSTICS NEXT Buffer specification is described in Table 8 6 Go to the next sc
130. ortional type pH control is often used Removal of suspended metal hydroxides Following the second stage a settling tank and or a filter can be used to remove suspended metal hydroxides However further treatment may be required to lower concentrations of some metals below their hydroxide solubilities 3 00 9782 Series pH ORP Analyzer Coniroller Operator s Manual 137 Cyanide Waste Treatment BA Batch Treatment AN Sequence of steps Continuous treatment is shown in Figure B 1 However all of the reactions can be achieved with semi automatic batch control Only a single tank with a pH controller and an ORP controller are required The steps are sequenced and the pH and ORP setpoints are changed to give the same results as for the continuous treatment Caustic is added to raise pH to 11 then hypochlorite is added to raise the ORP potential to about 450 mV while more caustic is added as required to maintain 11 pH WARNING Failure to comply with these instructions could result in death or serious injury An interlock must be provided to prevent the addition of acid before the positive oxidation of ALL cyanide Failure to observe this precaution can result in the generation of highly toxic hydrogen cyanide Then the acid can be added to neutralize the batch and further oxidation will complete the cyanate to carbonate conversion A settling period can then be used to remove solids or the batch can be pumped directly to another s
131. ou want to display temperature and or use solution temperature compensation described in 10 10 then connect a separate temperature input to the TH and SC terminals on the Analyzer Controller 7 If analog outputs are to be used connect the outputs from the Analyzer Controller terminals in accordance with the system setup assignments Refer to the wiring diagrams provided with the field devices receiving the signals and to Figure 6 1 through Figure 6 6 8 If the relay outputs are to be used leave the unit open and powered down See 6 4 9 If the relays will not be used replace the safety cover and secure it with the screw 10 Close the case and power up the unit Do not apply power with the case open The Durafet pH electrode has a built in temperature compensator for normal Nernstian temperature compensation However this temperature compensator can not provide input for temperature display nor for solution temperature compensation 9782 Series pH ORP Analyzer Coniroller Operator s Manual 65 Input and Output Wiring 6 3 2 Special Information for DualCAT Current adjusting type control Two devices can be controlled using Current Adjusting Type Control and a single analog output This DualCAT control is an option selectable during system setup see Section 5 If DualCAT is used refer to Figure 6 1 for acid and base feeder connections Note that the unused feeder receives an open circuit at its input with which
132. perator s Manual 133 Temperature Measurement Characteristics 134 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Cyanide Waste Treatment B Cyanide Waste Treatment B 1 Introduction Uses of cyanide solutions Cyanide solutions are used in plating baths for zinc cadmium copper brass silver and gold The toxic rinse waters and dumps from these operations require destruction of the cyanide typically to a level below 0 1 ppm before its discharge Technique for cyanide destruction The technique most often used for cyanide destruction is a one or two stage chemical treatment process The first stage raises the pH and oxidizes the cyanide to less toxic cyanate When required the second stage neutralizes and further oxidizes the cyanate to harmless carbonate and nitrogen The neutralization also allows the metals to be precipitated and separated from the effluent Consistent treatment and stable control in this type of process requires well mixed reaction tanks with enough volume for adequate retention time See Figure B 1 Retention time is calculated by dividing the filled or usable tank volume by the waste flowrate Typically it is 10 minutes or more Se CAUSTIC PROPORTIONAL RECORDER NOTE The seperate pH and ORP tt iat eis aca measurements and control shown Ko in the first stage may be handled CC Sir with a single 9782 Analyzer with 5
133. processes and for pretreatment in continuous neutralization systems Proportional only control does not include any reset action As a result changes in reagent demand that occur in continuous processes are not accommodated and control will not reach the setpoint In a batch process reagent is added until a setpoint is reached and the batch is not emptied until this occurs If you decide to use proportional only control select the type based on the input requirements of the device to be controlled See Appendix E for a discussion of tuning proportional only control 10 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Introduction 1 6 Overview of Installation and Setup Tasks Setup tasks described in this manual This manual contains instructions for all installation and operation tasks relating to the Analyzer Controller Instructions for installing and using the electrodes are provided in the manuals supplied with the electrodes Table 1 3 provides an overview of the Analyzer Controller installation tasks as well as providing references to the relevant sections of the manual Table 1 3 Overview of Analyzer Controller Installation Tasks 3 00 Sequence Task Section 1 Consider the environmental and electrical specs when 2 selecting a site to install the Analyzer Controller 2 Unpack inspect and mount the unit 3 3 Install any optional hardware 3 4 Check tag
134. pter 9782P 01 02 or 04 DURAFET not for Durafet with Cap Adapter GLASS including Durafet with Cap Adapter ANTIMONY ORP when ORP only is measured UO setup input 2 9782P 03 R2 or 9782 03 R0 ALTELTRD alternate electrode N A 9782P 03 009782P 03 VC 9782P 03 C2 9782P 03 C3 N A 9782P 01 R2 9782P 01 R0 9782P 02 R2 9782 02 R0 ORP when both pH and ORP are measured ALTELTRD when pH or ORP is measured N A when pH only or ORP only is measured or Durafet is used 9782P 01 00 9782P 01 VC 9782P 01 C2 9782P 01 C3 9782P 02 00 9782P 02 VC 9782P 02 C2 9782P 02 C3 ORP N A UO setup automatic functions ACLN ONLY AutoClean only ACAL 1PT AutoClean plus AutoCal one point ACAL 2PT AutoClean plus AutoCal two point 4 relays only N A 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 35 UO Setup and System Configuration PARAMETER UO setup discrete control type UO setup analog output types UO setup value alarmed or controlled CHOICES or RANGE Default is underlined two relays only 1PFT pulse frequency type 1DAT duration adjusting type 10 F on off 2PFT 2DAT 20 F ALMA alarm ALM2 ALM3 N A four relays all of the above plus 30 F 4 O F 2CAT DualCAT PCAT CAT control based on pH measured value OCAT CAT control based on ORP measured value pHCAT retransmission of process pH value default for Output 1 ORPCAT retransmission of pr
135. ption card to the 9782 it is very important that you change the model number stored in memory see Section 10 During I O setup the software will display prompts only for those features the hardware can support The model number of the unit was stored in the 9782 memory at the factory Each combination of features has a unique model number see Section 2 Therefore the processor can use this information to determine which prompts and selections are appropriate Entering the new number when you modify the hardware will let the processor know about the new hardware capabilities As a result you will be able to configure the unit to take advantage of features requiring that hardware Procedure The procedure for I O setup configuration is shown in Table 5 2 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 39 UO Setup and System Configuration Table 5 2 Procedure for Performing I O Setup Step Screen MENU CONFIGURATION CALIBRATION MAINTENANCE I O SETUP DIAGNOSTICS l NEXT RELAY SETTINGS RLY1 RLY2 RLY3 RLY4 N A N A pH TEMP ACLN ACAL HIGH LOW N A N A ALM3 ALMA DIAGNOSTICS PREV T NEXT ANALOG OUTPUTS OUTPUT 1 pH OUTPUT 2 pH OUTPUT 3 TEMP DIAGNOSTICS PREV NEXT ATTENTION ANY RECONFIGURATION CAN AFFECT OTHER PARAMETERS DIAGNOSTICS PREV NEXT SELECT INPUT TYPE INPUT 1 DURAFET
136. r Table 6 5 Procedure for Disconnecting RC Arc Suppression Circuits Action Turn off the power to the Analyzer Controller Disconnect and tag all wiring from the terminal boards With the power off open the case e Grasp the bottom center portion of the front bezel and pull it downward and toward you slightly to disengage the bottom of the bezel from the edge of the case e Lift the bezel gently to disengage it from the top edge of the case e Swing the bezel to the left The bezel and display assembly is mounted on pivot arms The safety cover with its warning label will be visible With the power off remove the safety cover by removing the single screw holding it in place cards half way out of the case Remove the circuit card retainer clamp at the top of the case Then slide all the circuit If the auxiliary relays are installed card p n 056382 or p n 056442 also remove 330 On the right most vertical printed circuit card power supply card p n 056305 remove 330 ohm resistors R29 and R30 for Relay 1 and R31 and R32 for Relay 2 from the upper corner of the card ohm resistors R39 R40 R44 and R45 from those optional assemblies Reconnect all wiring removed in Step 4 Replace the safety cover and secure it with the screw AA Close the case and power up the unit Do not apply power until the case is closed 76 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00
137. r rinses is 3 days and you want the first rinse to be tomorrow specify that today is day 2 Go to the next screen AUTO CAL 2PT RINSE SETUP DURATION 180s BETWEEN STD 2 DIAGNOSTICS NEXT Enter the duration of the AutoClean rinse in seconds 60 is the minimum Specify the frequency of AutoCal standardization operations by entering the number of rinses between standardization operations In our example timer setup was configured for rinses to be done every third day If during rinse setup two rinses are scheduled to occur between standardization operations then the AutoCal standardization will be performed every sixth day Go to the next screen 58 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 UO Setup and System Configuration Step Screen Action 6 ee ee Enter the length of time in seconds 60 is the minimum DURATION 70s Each reference solution should flow past the electrode BETWEEN SLOPE 5 Specify the frequency of AutoCal slope adjustments by DIAGNOSTICS entering the number of standardization operations NEXT between slope adjustments In our example slope adjustments will be made every 30 days 5 x the 6 day interval in Step 5 for standardizations Go to the next screen 7 AUTO CAT PRE The length of time specified in Step 6 will be displayed SLOPE SETUP SLOPE INTERVAL 70s DELAY 90s DIAGNOSTICS
138. reen Specify whether or not output values should be held during the automatic operations Alarms are always held and for that reason HOLD ACTIVE will appear on the alarm stripe Go to the next screen 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 57 UO Setup and System Configuration Step Screen Action 4 AUTO CAL 2PT TIMER SETUP DAYS 3 HOURS 16 MINUTES 30 CURRENT DAY 2 DIAGNOSTICS NEXT Timer setup is used to specify the frequency of the AutoClean rinses that are included in the AutoCal schedule e f several rinses are to be performed in a single day a specify the number of days as zero b specify the number of hours and minutes in the interval between rinses 1 hour is the minimum e f rinses are to occur at intervals of days a specify the number of days in the interval b specify the time of day hours and minutes the rinse should occur In our example the rinse will occur every third day at 4 30 PM e If you plan to use AutoCal but initiate operations manually using the real time AUTOCAL CLEAN display a specify the number of days as zero b specify the number of hours and minutes in the interval between rinses as zero also IF YOU ENTER ALL ZEROS NO AUTOMATIC RINSES AND NO AUTOMATIC CALIBRATIONS WILL BE PERFORMED Specify where today falls in the AutoCal AutoClean cycle For example if the interval fo
139. relays Model No 00 one analog output 4 to 20 mA or0 to 1 V or Oto 10 V two analog outputs both 4 to 20 mA VC C2 three analog outputs all 4 to 20 mA two additional relays hermetically sealed and one analog output 4 to 20 mA or0to1 VorOto 10 V C3 RO 16 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 3 00 Specifications and Model Number Model Number Table III OPTIONS Description Model No User s Manual English E Pipe Mounting Kit none arene one kit ses Tagging none ee linen LE stainless steel S Power factory set for 110 120 Vac 0 factory set for 220 240 Vac 1 I i Future sent Model Number Table IV BACKLIT DISPLAY OPTION Description Model No Backlit display option No 00 Nes 9782 Series pH ORP Analyzer Controller Operator s Manual 17 Specifications and Model Number 18 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 3 Unpacking Preparation and Mounting 3 1 Overview Introduction This section contains instructions for unpacking preparing and mounting the Unpacking Preparation and Mounting Analyzer Controller Instructions for wiring are provided in Sections 4 and 6 Software configuration is described in Section 5 What s in this section The topics in this section are listed belo
140. resume delay 1 to 1999 seconds default 60 configuration AutoCal AutoClean number between operations maintenance enable security maintenance password maintenance solution temperature compensation type maintenance solution temperature compensation value maintenance Durafet temperature display maintenance electrode selection maintenance relay activation on alarm control 1 to 1999 occurences default 1 NO YES 0000 to 9999 default 0000 PURE WATER AMMONIA PHOSPHATE MORPHALINE OTHER N A for PURE WATER 0 16 for AMMONIA 0 32 for PHOSPHATE 0 32 for MORPHALINE 0 32 for N A 0 ON OFF PRIMARY ALTERNATE ENERGIZE DE ENERGIZE 38 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 UO Setup and System Configuration 5 3 Performing I O Setup Purpose Before you can configure your Analyzer Controller to implement your control strategy you must perform I O setup In I O setup you choose the features to be used based on the hardware capability of the 9782 model being configured Using your selections the software will display the relay and analog output assignments appropriate for your application The choices made during I O setup determine which prompts and menu choices are available during configuration For a discussion of the features available and of the priority scheme used to assign relays see Section 1 ATTENTION If you add an o
141. rode Titration curve 140 The ORP titration curve in Figure C 2 shows the entire millivolt range if Cr chrome is treated as a batch With continuous treatment operation is maintained in the fully reduced portion of the curve near the 250 mV setpoint The ORP setpoint can vary between installations depending on pH reducing agent presence of additional contaminants and dissolved oxygen and the type of reference electrode used Determine the exact setpoint empirically This ORP setpoint should be at a potential where all of the Cr has been reduced without excess sulfite consumption which can release sulfur dioxide gas This point can be verified with a sensitive colorimetric test kit or similar check 9782 Series pH ORP Analyzer Coniroller Operator s Manual 3 00 Chrome Waste Treatment REDOX POTENTIAL mV VOLUME OF BISULFATE ADDED Figure C 2 Chrome Reduction Typical Titration Curve Chrome reduction is slow enough that 10 to 15 minutes may be required for a complete reaction and this time increases if pH is controlled at higher levels The pH also has a direct effect on the ORP potential as shown in Figure C 2 Therefore pH must be controlled to achieve consistent ORP control C 3 Second Stage of Chrome Removal Neutralize the wastewater In this stage the wastewater is neutralized to precipitate the Cr as insoluble chromium hydroxide Cr OH 3 Another reason is to meet the discharge pH lim
142. s product information will be displayed RESE Finally the display will return to showing process values 120 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Maintenance 10 9 Switching Between Electrodes with DualCal Introduction The DualCal feature allows storage of calibration data for two electrode assemblies The second can be used as a backup or in a two tank batch control setup During I O setup see Section 5 the Analyzer Controller can be configured to use this feature If the feature was enabled during TO setup use the procedure in Table 10 8 to select the electrode for which the input and calibration factors are to be used Table 10 8 Procedure for Switching Between Electrodes with DualCal Step Screen Action 1 SrA Go to MAINTENANCE ELECTRODE FUNCTIONS CONFIGURATION CALIBRATION SELECT ELECTRODE MAINTENANCE I O SETUP DIAGNOSTICS PREV NEXT MAINTANENCE OFF LINE FUNCTIONS INSTRUMENT SETUP ELECTRODE FUNCTIONS RELAY ACTIVATION DIAGNOSTICS ELECTRODE FUNCTIONS SELECT ELECTRODE SOLUTION TEMP COMP DURAFET TEMP DISPLAY ei AGNOSTICS I 2 SEERCT ELECTRODE The display will show the electrode currently being used Select the PRIMARY or ALTERNATE electrode ELECTRODE PRIMARY DIAGNOSTICS HOLD 3 00 9782 Series pH ORP Analyzer Controller
143. s set back to 100 0 eliminating any changes made during manual or automatic calibration All alarms will be cleared and outputs will go to zero Some I O setup values will also return to factory settings This cold reset contrasts with the warm reset that occurs following loss of power and at the end of system setup see Section 5 During a warm reset alarms are not cleared and outputs are maintained at their last value prior to the loss of power or initiation of the cold reset A warm reset has no effect on configuration or calibration values Procedure To do a cold reset returning all configuration and calibration values to their factory settings follow the procedure in Table 10 7 Table 10 7 Procedure for Resetting All Configuration to Factory Settings Step Screen Action MENU ee Go to MAINTENANCE INSTRUMENT SETUP RESET CALIBRATION UNIT MAINTENANCE I O SETUP DIAGNOSTICS PREV NEXT MAINTANENCE OFF LINE FUNCTIONS INSTRUMENT SETUP ELECTRODE FUNCTIONS RELAY ACTIVATION DIAGNOSTICS INSTRUMENT SETUP INSTURMENT TAGGING PRODUCT INFO SCREEN CONTRAST SECURITY RESET UNIT DIAGNOSTICS 2 UNIT RESET ATTENTION Pressing the specified function key at this point will ALL PARAMETERS perform a cold reset as described above WILL BE RESET TO FACTORY DEFAULTS As the reset is completed the display will go blank then DIAGNOSTICS for a few second
144. se of alarm or control output ON state is currently ENERGIZE SELECT ENERGIZE or DE ENERGIZE DIAGNOSTICS Use this display to specify the desired relay action 126 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Maintenance 10 13 Replacing the Fuse Introduction Figure 10 1 shows the location of the power fuse on the power supply card in the right most position inside the case See Section 11 for the part number of the kit containing fuses 2 A Figure 10 1 Location of Power Fuse d A WARNING Disconnect power before opening the instrument case to replace the fuse A potentially lethal shock hazard exists inside the case if the unit is opened while powered More than one switch may be required to disconnect power ATTENTION To avoid potentially serious damage to expensive components a blown fuse should not be replaced until the cause has been determined by a qualified service technician Procedure To replace the fuse follow the procedure in Table 10 14 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 127 Maintenance Table 10 14 Procedure for Replacing the Fuse Step Action 1 Determine why the fuse blew and correct the problem 2 Check the fuse s packaging in miscellaneous parts kit 51198177 501 to be sure that the fuse is correct for the voltage at your site 3 Turn off the power to the Analyzer Con
145. serving temperature pressure and flow limitations Note the following recommendations e Rinse electrodes thoroughly between buffer solutions e Always use HOLD or otherwise deactivate control or alarm circuits before removing electrodes from the process e Standardize with a buffer solution which is at about the same temperature and pH as the sample solution e Inspect and if necessary clean and or rejuvenate the electrode system periodically according to experience and conditions 100 9782 Series pH ORP Analyzer Coniroller Operator s Manual 3 00 Diagnostics and Messages 9 Diagnostics and Messages 9 1 Overview Introduction This section contains information on status and alarm messages as well as on diagnostics and system error messages All these messages are displayed on the alarm stripe in reverse video with the background lit and the letters black If more than one message is active the display will cycle through all the messages then repeat What s in this section The topics in this section are listed below Topic See Page 9 1 Overview 101 9 2 System Status Message 102 9 3 Process Alarm Messages 103 9 4 On Line Diagnostics and System Error Messages 104 9 5 Off Line Diagnostics 105 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 101 Diagnostics and Messages 9 2 System Status Message Related to normal operation System status
146. space between vertical cutouts 7 64 070 7 68 060 194 06 1 78 195 07 1 52 La F EE Zut O O 7780 Eee me H a eee Kot S 1 00 IPS Pipe 25 40 by customer Note 1 Do not exceed 80 Ib in torque when tightening fastners Horizontal Rear Pipe Mounting Figure 3 3 1 Mounting and Dimensions Backlit Display Option 3 00 9782 Series pH ORP Analyzer Coniroller Operator s Manual 25 Unpacking Preparation and Mounting Customer Panel Vertical Rear Pipe Mounting _ inches DEER LEGEND millimeters 3 06 7 64 070 77 80 194 06 1 78 re 3 f gt OC DO hy V 5 F jo d 7 68 060 195 07 1 52 l d K 1 00 IPS Pipe J 25 40 by customer odo Note 1 Do not exceed 80 Ib in torque when tightening fastners Kg Top View Panel Mounting 7 64 070 into existing 7075 7076 7077 and 7078 series cutouts ra 194 06 1 78 __ 1 50 T Sep 38 10 E E 6 43 0 70 70 20 163 32 1 78 d eed 1 27 025 Li 32 26 0 64 Four holes in bracket or 25 max 6 27 040 1 4 dia mounting bolts 6 35 159 26 1 02 Customer Panel bolts by customer R
147. ssinn nie ne mi an nn el un fin Pair A n eg 6 1 5 M sense Ut mnt tn nn der Ant Ronnie ter etre 8 L OVEN E Wain a a a dE 8 15 2 Using Relays site Drame ARE ege rs nn er ne 8 1 5 3 Deciding on a Control Strategy ss 9 1 6 Overview of Installation and Setup Tasks 11 2 SPECIFICATIONS AND MODEL NUMBER cccccssseessseeeeeeeseeeeeeeesseneeeeenes 13 2 1 e DEED ALL ONS ee ee cscs uel die rires 13 2 2 Model Number Breakdown sise 16 3 UNPACKING PREPARATION AND MOUNTING nn sssrrnnenmnre 19 3 1 huewe Sed hath ean nn aan Sina SE een ee ate 19 3 2 Unpacking and Preparing etleche Glee 20 3 3 Installing Optional Internal Preamplifier ss 21 3 4 Meute ss Rate deed E a 22 Ai POWER WIRING eege ees Eed deed GEET aides 27 4 1 OVNI We EE ren EE SE ESA Ee 27 4 2 General Wiring Practices iieiaeie ee e ne dE 28 4 3 Power Wiring C OnSiGeratioms meriete es Goad ches safe sel Seed be dire tie deteste dE 29 4 4 Installing Power WADING snanar oE a NET 30 5 VO SETUP AND SYSTEM CONFIGURATION ccccccsssseeeeesseeeeeeeseneeeeeeseeneeees 33 5 1 OVEIVICW ee EE dE Ee enol an SES ante een laa ER 33 5 2 VO Setup and Configuration Tasks ss 34 5 3 Perlormin EE 39 5 4 Configuring Current Adjusting Type CAT Control and or Retransmission of Process Map bl gz usant nine ei el fe better te tell tent ledges tit
148. ssword call the Technical Assistance Center The telephone number is in the front of this manual 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 117 Maintenance 10 6 Adjusting the Screen Contrast Introduction The display contrast can be adjusted to suit the viewing conditions at the installation site Procedure To adjust the display contrast follow the procedure in Table 10 5 Table 10 5 Procedure for Adjusting the Screen Contrast Step Screen Action 1 pied se Ts Go to MAINTENANCE INSTRUMENT SETUP SCREEN CALIBRATION CONTRAST MAINTENANCE I O SETUP DIAGNOSTICS PREV NEXT MAINTANENCE OFF LINE FUNCTIONS INSTRUMENT SETUP ELECTRODE FUNCTIONS RELAY ACTIVATION DIAGNOSTICS INSTRUMENT SETUP INSTURMENT TAGGING PRODUCT INFO SCREEN CONTRAST SECURITY RESET UNIT DIAGNOSTICS SCREEN CONTRAST 2 CONTRAST 0 Use the UP and DOWN keys to change the contrast USE UP DOWN ARROW KEYS TO ADJUST When the displayed contrast is acceptable use the MENU ee key to return to the instrument setup menu The new DIAGNOSTICS contrast setting will be saved automatically 118 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Maintenance 10 7 Entering a Password for Security Introduction TO setup configuration calibration and maintenance functions can be password protected The passwor
149. st 10 minutes to be sure of a stable reading and uniform temperature throughout 4 Record the pH and temperature indicated by the instrument at each of the test temperatures 5 Plot the data to produce a graph of pH versus temperature and calculate the slope near the region of interest in pH units per 10 C Enter that value for solution temperature coefficient 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 123 Maintenance 10 10 3 Entering Solution Temperature Coefficient Introduction To use the solution temperature compensation algorithm correctly a solution temperature coefficient must be selected from among those stored in memory at the factory or entered as described below Procedure Use the procedure in Table 10 11 to enter a solution temperature coefficient for pH measurement into memory Table 10 11 Procedure for Entering a Solution Temperature Coefficient Step Screen Action d E Go to MAINTENANCE ELECTRODE FUNCTIONS CALIBRATION SOLUTION TEMPERATURE COMP MAINTENANCE I O SETUP DIAGNOSTICS PREV NEXT MAINTANENCE OFF LINE FUNCTIONS INSTRUMENT SETUP ELECTRODE FUNCTIONS RELAY ACTIVATION DIAGNOSTICS ELECTRODE FUNCTIONS SELECT ELECTRODE SOLUTION TEMP COMP DURAFET TEMP DISPLAY DIAGNOSTICS 2 POHUTTON EMEC COME The display will show the application type currently being TYPE PURE WATER used If one of the stan
150. static sensitive devices Immunity Compliance In applications where either the power input or output wiring are subject to electromagnetic AN disturbances shielding techniques will be required Grounded metal conduit with conductive conduit fittings is recommended Conform to code Instrument wiring should conform to regulations of the National Electrical Code 62 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Input and Output Wiring Recommended maximum wire size Table 6 1 Recommended Maximum Wire Size Gage Number mm Description 14 2 081 power relays and PE protective earth 18 0 823 inputs 18 0 823 isolated outputs Cables for locations without interference For steel metal conduit or open wiring away from any sources of interference use recommended Honeywell cables listed in the parts list in Section 11 Shielded wiring for locations with interference In applications where plastic conduit or open wire trays are used shielded 6 conductor 22 gage 0 326 mm or heavier signal input wiring is required Avoiding interference Instrument wiring is considered Level 1 per section 6 3 of IEEE STD 518 for plant facilities layout and instrumentation application Level 1 wiring must not be run close to higher level signals such as power lines or drive signals for phase fired SCR systems etc Unprotected input wiring in high electrical noise environments is subject to electrom
151. tage of treatment has also been found to help maintain clean ORP electrodes 142 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Simultaneous pH and ORP Measurement D Simultaneous pH and ORP Measurement de CS 316441 022283 Preamplifier note 2 ee Terminal NI eg W S n eH A lo O O Qscl x Cvt B 4 16 O REF OVH 0O 77 16 O OOUT Cra IN 4 Use OTH a ET E Note 3 O NOTES FE 1 Remove RE SC jumper from 2 9782 For power and output 2 2 S connections see appropriate amp T S o 7082 installation diagram a EK 2 Add SC ET jumper and 316441 S e ets terminal to preamp 5 EE 3 6 Conducter cable Part 834023 Belden 8446 Red and black 18 AWG 0 823 mm Others 22AWG 0 325 mm Color Gode K black yellow W white A gray G green _ V violet R red P pink B blue S slate N brown T tan O orange Figure D 1 Installation Diagram Simultaneous pH and ORP Measurements with 9782 pH ORP Analyzer and 7773 60 0 40 0 Electrode Mounting Plus ORP Electrode 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 143 Simultaneous pH and ORP Measurement 144 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Proportional Control Tuning
152. te the Analyzer Controller using a voltage input are e A source of a known millivolt signal e A screwdriver to fit the Analyzer Controller input terminal screws and the screw securing the safety cover Procedure To calibrate the Analyzer Controller follow the instructions in Table 8 5 A A WARNING This procedure should be performed by qualified personnel only Disconnect the power before opening the instrument case A potentially lethal shock hazard exists inside the case if the unit is opened while pwered More than one switch may be required to dosconnect power Table 8 5 Procedure for Calibrating ORP Analyzer Controller Using Voltage Input Step Screen Action 1 Turn off the power to the Analyzer Controller More A than one switch may be required to disconnect power 2 With the power off open the case e Grasp the bottom center portion of the front bezel and pull it downward and toward you slightly to disengage the bottom of the bezel from the edge of the case e Lift the bezel gently to disengage it from the top edge of the case e Swing the bezel to the left The bezel and display assembly is mounted on pivot arms The safety cover with its warning label will be visible 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 91 Manual Calibration Step Screen Action 3 With the power off remove the safety cover by removing the single screw holding it in place 4 Label
153. the name of this control type voltage outputs can also be used DualCAT is a feature that permits control of two reagents with a single analog output An internal relay is used to make the switchover This scheme is diagrammed in Figure 1 3 When an output is used to retransmit a process variable the values representing 0 and 100 outputs are configurable In CAT control the process values for the setpoint and proportional band limit are configurable In addition if the DualCAT feature is used a failover value can be configured If the feature is enabled then this value will be used as the output if the Analyzer Controller s diagnostics detect a problem with the system To control two reagents with two output signals simply configure the output ranges appropriately No switching relay is needed if two outputs are used 100 5 a base acid 5 ddition additio O 0 A 4 A CATIPB CATISP CAT2SP CAT2PB 0 2 4 6 8 10 12 14 SWITCHOVRRLY pH Figure 1 3 DualCAT Control Using a Single Analog Output Selecting a strategy When considering whether to use On Off control or one of the proportional only strategies compare the requirements of your application with the benefits of each control type Proportional control will usually provide faster batch treatment with less overshoot than On Off control However it is important to note that proportional only control is recommended only for batch
154. to 100 parts buffer solution The oxidation potential of this solution is listed below for several temperatures The polarity sign shown is that of the measuring element with respect to the reference element These solutions are unstable and should be used within eight hours of preparation All mV values in Table 8 3 have a 30 mV tolerance Table 8 3 Oxidation Reduction Potential of Reference Solutions at Specified Temperature pH Buffer Solution Temperature Honeywell Part Number 20 C 25 C 30 C 4 01 25 C 31103001 267 mV 263 mV 259 mV 6 86 25 C 31103002 100 mV 94 mV 88 mV 7 00 25 C not available from Honeywell 92 mV 86 mV 80 mV 9 00 25 C not available from Honeywell 26 mV 32 mV 39 mV 9 18 25 C 31103003 36 mV 43 mV 49 mV 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 89 Manual Calibration Procedure To calibrate the system follow the instructions in Table 8 4 Table 8 4 Procedure for Calibrating ORP System Using Reference Solution COMPLETED STD BUFFER VALUE SAVED DIAGNOSTICS HOLD INEXT Step Screen Action 1 e Prepare container of a solution with known oxidation CALIBRATION reduction potential MAINTENANCE I O SETUP DIAGNOSTICS 2 eo CAN Go to CALIBRATION ORP CAL SAMPLE CAL AUTO BUFFER CAL AUTO
155. to confirm that unit is compatible with the voltage at 4 your site 5 Install power wiring 6 Perform I O setup and configure the software to implement your control strategy 7 Wire inputs and outputs to match I O setup in Step 6 8 If AutoClean and AutoCal will be used install necessary piping and valves 9 If AutoCal will not be used calibrate the system manually 10 Specify whether relays should be energized or de energized on alarm 11 If solution temperature compensation will be used enter a solution temperature coefficient If desired specify that temperature will be one of the displayed real time values 12 Configure a tagname or other text string to be displayed at the 10 top of the screens showing real time process values 13 To limit access to I O setup configuration calibration and 10 maintenance functions enter a password 14 Display model number stored in memory and software version 10 number Note these for future reference If you call for technical assistance you will need this information 9782 Series pH ORP Analyzer Coniroller Operator s Manual 11 Introduction The manual also contains e information about diagnostics status messages and system error messages Section 9 e instructions for returning all parameter values to the factory settings Section 10 e instructions for calibrating the outputs and changing the fuse Section 10 e parts list Section 11 e supplementary informatio
156. tored in Memory 116 10 6 Adjusting theScreem Contrast eege de eebe E e E ee Ae dates 118 10 7 Entering a Password for Security ss 119 10 8 Resetting All Configuration and Calibration Values to Factory Settings 120 10 9 Switching Between Electrodes with DualCal ss 121 10 10 Determining and Entering a Solution Temperature Coefficient eseeeeeeeeereereereereeererrsre 122 TOTO cD ever e 122 10 10 2 Determining Solution Temperature Coefficient 122 10 10 3 Entering Solution Temperature Coefficient 124 10 11 Turning on Temperature Display Durafet Electrodes without Cap Adapter Only 125 10 12 Specifying Relay Activation 126 10 15 Replacing the ugeet gegen Eed ege 127 11 ACCESSORIES AND REPLACEMENT PARTS LIST seeccsssseeeeseeeeeeeenes 129 11 1 CQVETVIEW Lean Ee lee ee on M ei een tn Ph Rite Malle tete l 129 11 2 Part N mbers en geste Gee elie ets et tre 130 A TEMPERATURE MEASUREMENT CHARACTERISTICS seeccesseseeeesesees 133 B CYANIDE WASTE TREATMENT scccccssssseeeeessseeeeeeesseeeeessseneeeesseneeeeennes 135 B 1 Introduction E 135 B 2 First Stage of Cyanide Destruction ss 136 B 3 Second Stage of Cyanide Destruction 137 BA Ereegnes deeg 138 B 5 ORP Potential a Measure of Status of Reaction 138 ES CHROME WASTE TREATMENT gtosgerskeetesbeegageegket ieegkekd rk geereegegeregegeeeeeg 139 Cl Use of Chromates
157. tored in the 9782 memory see Note above The instrument gain slope er will be adjusted If you see a stable reading before 50 seconds have elapsed you can avoid waiting by pressing ENTER 9a ted Aa tee If the slope adjustment is successful the completion screen will be displayed SLOPE BUFFER VALUE ger Take the unit out of Hold mode and return to the DIAGNOSTICS calibration menu HOLD NEXT 9b dee If the slope adjustment is not successful the display will CALIBRATION return to the menu automatically and an error message e will be displayed Possible causes are DIAGNOSTICS e The difference in pH between the two solutions was less than 2 Use a different solution e The glass electrode is cracked Replace it e The attempted slope adjustment is too great Replace the electrode 98 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Manual Calibration 8 7 Viewing Zero Offset and Percent Theoretical Slope Introduction The manual standardization and slope adjustments described in this section as well as the automatic calibration procedure described in Section 7 change the zero offset standardization and the percent theoretical slope used by the system These values can be viewed as read only information It is good practice to observe these values after calibration Monitoring the values over time will help you predict when the electrode will need to be replaced To set the zero
158. tputs 6 Replace the safety cover and secure it with the screw 7 Close the case and power up the unit Do not apply power until the case is closed 9782 Series pH ORP Analyzer Controller Operator s Manual 73 Input and Output Wiring 6 4 2 Special Information for DualCal Backup alternate electrode for critical process The DualCal feature allows precalibration of an alternate electrode system for backup purposes in applications where interruptions in measurements must be avoided Because the 9782 retains the calibration data in memory the spare electrode can be manually put in place and quickly switched on In the meantime the spare electrode can be stored in the process or in a storage solution see Figure 6 7 One Analyzer Controller in two tank batch treatment system DualCal may also be used for two tank batch treatment systems where a single Analyzer Controller can be used to select electrode preamp systems in each tank see Figure 6 8 External switch determines which electrode signal reaches the Analyzer Controller A 5 PDT switch supplied by the customer is used to switch between the two electrode systems This 5 PDT switch can be switched manually by the operator or can be switched by the Analyzer Controller internal Relay 4 assigned during I O setup see Section 5 If the use of DualCal is specified during I O setup and internal Relay 4 is used then switching between electrodes can be done q
159. troller More than one switch may be A required to remove power 4 With the power off open the case e Grasp the bottom center portion of the front bezel and pull it downward and toward you slightly to disengage the bottom of the bezel from the edge of the case e Lift the bezel gently to disengage it from the top edge of the case e Swing the bezel to the left The bezel and display assembly is mounted on pivot arms The safety cover with its warning label will be visible 5 With the power off remove the safety cover by removing the single screw holding it in place 6 If more room is needed remove the circuit card retainer bracket at the top of the case and pull the circuit card assembly partially out of the instrument case It may be necessary to temporarily disconnect leads from some terminals to do so 7 Remove the old fuse and insert the new one 8 If any leads were disconnected in Step 6 reconnect them 9 Replace the safety cover and secure it with the screw i OG 10 Close the case and power up the unit Do not apply power until the case is closed 128 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Accessories and Replacement Parts List 11 Accessories and Replacement Parts List 11 1 Overview Introduction This section provides part numbers for field replaceable parts and for accessories What s in this section The topics in this section are listed below Topic Se
160. two methods of calibration e With the buffering method described in 8 2 you use your electrode system to measure two reference solutions buffers having known pH values then adjust the 9782 so that its readings match the actual pH of each The two reference solutions must have a pH difference of at least 2 With the grab sample method described in 8 3 you measure your process both with your electrode system and with a separate accurately calibrated meter then adjust the 9782 so that its reading matches the meter The 9782 includes AutoCal an automatic buffer calibration that eliminates the need to go out to the instrument to manually calibrate the electrodes This feature is described in Section 7 Configuring the unit to use AutoCal is described in Section 5 ORP calibration ORP calibration consists of adjusting the reading of the 9782 to match a known value There are two types of ORP calibration supported To calibrate the system to compensate for changes in electrode potentials over time the ORP electrode is placed in a reference solution of known ORP value and the 9782 reading is adjusted to match this value as described in 8 4 Instructions for preparing standard solutions are also provided in 8 4 These solutions are stable for only short periods of time less than 8 hours and are only approximations of ORP potentials To calibrate the 9782 only not the whole system including electrodes apply a known
161. ues A cold reset can be initiated using a maintenance menu function as described in Section 10 It will also be initiated during I O setup if the input type is changed 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 33 UO Setup and System Configuration 5 2 I O Setup and Configuration Tasks Ease of configuration Configuring the 9782 software is fast and easy A menu is provided for every configuration task You will be permitted to configure only those parameters relevant to your application and supported by the Analyzer Controller model you purchased In fact configuration screens will contain only prompts and menu choices that apply to your application Accessing the I O setup and configuration menus Use the MENU key to access the main menu which includes CONFIGURATION and I O SETUP choices If the security feature has been enabled as described in Section 10 the password must be entered in response to the prompt before you can proceed Out of the box units do not require entry of a password to access the menu system Purpose of I O setup TVO setup is the process of choosing the features to be used so that the software can make appropriate I O assignments I O setup see 5 3 must precede configuration The choices made during I O setup determine which prompts and menu choices are available during configuration Even though I O setup is performed first I O SETUP is last on the main menu because you usu
162. uickly at the Analyzer Controller keypad using the maintenance menu see 10 9 Electrode selection signal Primary electrodes 5PDT relay Selected ee EE pH signal y Backup N electrodes A process sample Figure 6 7 Backup Electrode Installation Using DualCal Alternate Electrode Calibration 74 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 3 00 Input and Output Wiring wastewater Electrode selection signal Selected pH signal a Ge control signal 5PDT relay ld wastewater sy g Figure 6 8 Two Tank Batch Treatment with DualCal Alternate Electrode Calibration 9782 Series pH ORP Analyzer Controller Operator s Manual 75 Input and Output Wiring 6 4 3 Disconnecting RC Arc Suppression Circuits Introduction Relay contacts are protected by RC arc suppression circuits Alternating current leakage through these circuits may be sufficient to trigger very light ac loads such as neon bulbs PLC logic inputs continuity testers etc If this leakage cannot be tolerated the suppression circuits may be disconnected as described below Procedure Follow the procedure in Table 6 5 to disconnect the RC arc suppression circuits Step i A WARNING 1 More than one switch may be required to remove powe
163. until the next scheduled timer initiated cleaning ATTENTION Once an AUTOSEQUENCE is initiated you cannot go into the menu displays until sequence is completed If you do sequence will stop Hold mode During AutoClean and AutoCal operations alarms are always held This means that no alarms will become active during the automatic operation If one or more relays are being used for alarm annunciation they will go to the non alarm state The behavior of analog outputs and discrete control relays during automatic operations is configurable The outputs can either be held at their current value or state when the operation starts or they can continue to respond to input changes during the automatic operation ATTENTION The alarm stripe will display HOLD ACTIVE At the completion of the automatic operation the hold will be released and normal operation will resume after the configured delay Procedure If only AutoClean was selected during I O setup follow the procedure in Table 5 9 to use the displayed screens for configuration when AUTO CAL CLEAN is selected from the menu If one point AutoCal standardization only was selected during I O setup follow the procedure in Table 5 10 If two point AutoCal standardization and slope adjustment was selected during I O setup follow the procedure in Table 5 11 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 53 UO Setup and System Configuration
164. ure Compensator 133 Figure B 1 Cyanide Treatment System 135 Figure B 2 First Stage Cyanide Oxidation Typical Titration Curve 136 Figure C 1 Chrome Treatment System diese oud eanobaldede SEENEN EEN EEN 139 Figure C 2 Chrome Reduction Typical Titration Curve ss 141 Figure D 1 Installation Diagram Simultaneous pH and ORP Measurements with 9782 pH ORP Analyzer and 7773 60 0 40 O0 Electrode Mounting Plus ORP Electrode 0 cccesesceseeeeeeeeeeees 143 Figure F 1 Microprocessor Board DIP Switches A 147 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual xi xii 9782 Series pH ORP Analyzer Controller Operator s Manual 3 00 Introduction 1 Introduction 1 1 Description Multi function instrument The Honeywell 9782 pH ORP Analyzer Controller Figure 1 1 is a microprocessor based instrument for analysis of pH or ORP Oxidation Reduction Potential also referred to as Redox in industrial processes The pH ORP and temperature can be measured simultaneously depending on configuration see Section 5 The units of measure are pH mV and C Resolution is 0 01 pH 1 mV ORP and 1 C User interface An easy to read display provides instant access to process values The displays of process values show the instrument s tagname or other configurable text at the top of the screen Every display includes an alarm stripe a line dedicat
165. us personal injury WARNING An external disconnect switch is required for any hazardous voltage connections to the relay outputs Avoid damage to components ATTENTION This equipment contains devices that can be damaged by electrostatic discharge ESD As solid state technology advances and as solid state devices get smaller and smaller they become more and more sensitive to ESD The damage incurred may not cause the device to fail completely but may cause early failure Therefore it is imperative that assemblies containing static sensitive devices be carried in conductive plastic bags When adjusting or performing any work on such assemblies grounded work stations and wrist straps must be used If soldering irons are used they must also be grounded A grounded work station is any conductive or metallic surface connected to an earth ground such as a water pipe with a 1 2 to 1 megohm resistor in series with the ground connection The purpose of the resistor is to current limit an electrostatic discharge and to prevent any shock hazard to the operator The steps indicated above must be followed to prevent damage and or degradation which may be induced by ESD to static sensitive devices Wiring for immunity compliance In applications where either the power input or output wiring are subject to electromagnetic disturbances shielding techniques will be required Grounded metal conduit with conductive conduit fittings is recomm
166. utoCal Sequence and Piping 80 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 77 AutoClean and AutoCal Theory and Piping 7 2 AutoClean Sequence and Piping Rinse sequence The AutoClean operation occurs at the configured intervals The sequence is described below 4 All alarm action is held at existing levels The output s can be held or be active depending on configuration Even if the outputs are not held HOLD ACTIVE is displayed on the alarm stripe because alarms are always held Also AUTOSEQUENCE is displayed Pressing the DISPLAY key will call up a special display that shows how much time is left in the operation Relay 1 activates 3 way solenoid valve S1 see Figure 7 1 to direct rinsing fluid to the electrodes for the configured rinse duration 1 to 1999 seconds If the measured sample is normally returned to the process but quantities of rinsing fluid cannot be tolerated there use an additional 3 way solenoid valve S4 It is activated simultaneously with S1 to divert the discharge to drain At the end of the configured rinse time Relay 1 de activates the solenoid valve S1 and S4 if used After the configured delay period 1 to 1999 seconds the Analyzer Controller resumes sampling the process Note that even with S4 for diversion one system volume of washing fluid will pass to the process at this point The HOLD and AUTOSEQUENCE messages are cleared Note that the op
167. view Introduction This section provides instructions for performing I O setup and for configuring the Analyzer Controller s software to implement your control strategy What s in this section The topics in this section are listed below Topic See Page 5 1 Overview 33 5 2 WO Setup and Configuration Tasks 34 5 3 Performing I O Setup 39 5 4 Configuring Current Adjusting Type Control and or Retransmission 43 of Process Variables 5 5 Configuring Duration Adjusting Type or Pulse Frequency Type Control 46 5 6 Configuring Alarms 50 5 7 Setting the Clock 51 5 8 Configuring AutoCal and AutoClean 52 Powering up the Analyzer Controller Every time the Analyzer Controller is powered up the unit performs a warm reset During a warm reset alarms are not cleared and outputs are maintained at their last value prior to the loss of power or initiation of the warm reset A warm reset has no effect on configuration or calibration values In addition to this warm reset a cold reset function is available to return ALL configuration values and some I O setup values to the factory settings shown in Table 5 1 During a cold reset the standardization offset is also returned to 0 0 and the percent theoretical slope is set back to 100 0 eliminating any changes made during manual or automatic calibration All alarms will be cleared Analog outputs will go to zero momentarily then go to the factory default val
168. w Topic See Page 3 1 Overview 19 3 2 Unpacking and Preparing 20 3 3 Installing the Optional Internal Preamplifier 21 8 4 Mounting 22 3 00 9782 Series pH ORP Analyzer Controller Operator s Manual 19 Unpacking Preparation and Mounting 3 2 Unpacking and Preparing Procedure Table 3 1 contains procedure for unpacking and preparing the 9782 Table 3 1 Procedure for Unpacking and Preparing the 9782 Step Action ATTENTION For prolonged storage or for shipment the instrument should be kept in its shipping container Do not remove shipping clamps or covers Store in a suitable environment only see specifications in Section 2 1 Carefully remove the instrument from the shipping container 2 Compare the contents of the shipping container with the packing list e Notify the carrier and Honeywell immediately if there is equipment damage or shortage e Do not return goods without contacting Honeywell in advance 3 Remove any shipping ties or packing material Follow the instructions on any attached tags and then remove such tags 4 All 9782 Analyzer Controllers are calibrated and tested at the factory prior to shipment Examine the model number on the nameplate Figure 3 1 to verify that the instrument has the correct optional features See Section 2 for model number breakdown 5 Select an installation location that meets the specifications in Section 2 The 9782 can be panel wal

Download Pdf Manuals

image

Related Search

Related Contents

    Panasonic PT-DZ21KE User Manual  Kingston Technology HyperX KHX1866C9S3K2/4GX memory module  Fujitsu ESPRIMO Mobile D Series D9500 + E22W-1 + LX850 + MLK Office SBE + Port replicator  LG Electronics M198WA Computer Monitor User Manual  取扱説明書 T-1  Acnodes RPS 1200 User's Manual  User Manual  December 2004 Newsletter - Greater Cincinnati BMW Club  

Copyright © All rights reserved.
Failed to retrieve file