Contents
1. properly 4 The Vacuum power on the sample a Vacuum regulator set to minimum a Turn the vacuum regulator clockwise and check the value of the side is very low or absent b Protection filter blocked or damaged vacuum on the gauge c Connection tubes blocked kinked or b Replace the filter disconnected c Replace or reconnect the tubes check the bottle connections d Shut off valve blocked or damaged d Empty the bottle or disconnect the tube from the bottle and e Pump motor damaged unblock the shut off valve The unit twill only work in the upright position e Refer to authorised service personnel 5 The float in the overflow valve If the cap has been washed ensure that Insert the float into its proper position doesn t close the float is not partially detached 6 The float in the overflow valve The float it s covered by dirty material Unscrewed the cap wash the overflow valve in clean water and doesn t close replace Foam inside the bottle Fill the bottle to 1 3 capacity with water Faults 1 2 3 4 5 6 7 None of the remedies has achieved the Contact Benchmark Scientific for Service Assistance desired results If the overfill valve system fails continued aspiration will cause liquid to contact the bacteriological filter and air flow will be blocked Should this back up security also fail there is a risk of liquid entering the vacuum pump system Do not attempt to continue using the Aspirator in the event of liquid entering the
2. to improper use or connection to an electrical system not complying with current regulations 9 Internal electrical and mechanical parts are not designed to be repaired by customers or end users Do not open the pump unit do not mishandle the electric mechanical parts Always contact Benchmark Scientific for technical assistance 10 Using the device in environmental conditions different than those indicated in this manual may be dangerous to the user and may cause damage to the instrument TECHNICAL SPECIFICATIONS ACCURIS ASPIRE LABORATORY ASPIRATOR 115V 50 60Hz or 230V 50 60Hz POWER CONSUMPTION 105VA FUSE F 1 x 4A 250V MAXIMUM SUCTION PRESSURE without jar 75kPa 0 75 Bar Adjustable from 75kPa 0 75bar to 10kPa 0 10 bar 15 l min 4 8 Lbs 2 2 Ko 250mm x 190mm x 160 mm SICILICONE TUBE SIZE 6 x 10 mm ACCURANCY OF VACUUM INDICATOR 5 AMBIENT OPERATING CONDITIONS Room temperature 5 to 35 C Room humidity percentage 30 to 75 RH Altitude 0 to 2000m STORAGE AND TRASPORT CONDITIONS Room temperature 40 to 70 C Room humidity percentage 10 to 100 RH MODEL VOLTAGE MAXIMUM SUCTION FLOW without bottle filter WEIGHT DIMENSIONS SYMBOLS Class II isolation equipment Warning consult the instruction manual Keep in a cool dry place Storage transport temperature 40 70 C Type B equipment Alternate Current Mains Frequency Please note technical specifications may change without notice
3. Guidance and manufacturer s declaration Electromagnetic Emissions The ACCURIS ASPIRE LABORATORY ASPIRATOR is intended for use in the electromagnetic environment specified below The customers or the user of the ACCURIS ASPIRE LABORATORY ASPIRATOR should assure that it s used in such an environment Power disturbance Group 1 THE ACCURIS LABORATORY ASPIRATOR CISPR11 RF emissions are very low and will not cause interference in proximity to any Electronic appliances Irradiated Conducted emissions CISPR11 Class B THE ACCURIS LABORATORY ASPIRATOR can Harmonic emissions IEC EN 61000 3 2 Class A be used in all environments including domestic Voltage fluctuations flicker emissions Complies and those connected directly to the public mains IEC EN 61000 3 3 distribution that supplies power to environments used for domestic scopes Guidance and manufacturer s declaration Electromagnetic Immunity THE ACCURIS LABORATORY ASPIRATOR is intended for use in the electromagnetic environment specified below Users of this product should assure that it s used in such an environment Electrostatic discharge ESD 6kV on contact Floors should be wood concrete or ceramic IEC EN 61000 4 2 8kV in air tile If floors are covered with synthetic material the relative humidity should be at least 30 IEC EN 61000 4 4 commercial environment IEC EN 61000 4 5 commercial environment Loss of voltage brief voltage interruptions and 5 Ur for 0 5
4. Rev 0 11 12 13 ross T PRU NES 1 T S ASPIRE Laboratory Aspirator USER MANUAL THE ACCURIS ASPIRE LABORATORY ASPIRATOR is a portable instrument that operates on 115V 60Hz or 230V 50Hz power input It is designed for the aspiration of liquid samples and suitable for removal of tissue culture media and supernatant from laboratory plates tubes and vessels The ASPIRE is housed in a heat resistant and electrically insulated plastic casing incorporates a vacuum regulator on the front panel amd also includes a polycarbonate autoclavable bottle and lid with overflow protection valve silicone tubing hydrophobic filter and handheld vacuum pipette GENERAL WARNING A READ THE INSTRUCTION MANUAL IN ITS ENTIRETY AND CAREFULLY BEFORE USE AN THE INSTRUMENT SHOULD ONLY BE USED BY PROPERLY TRAINED LABORATORY PERSONNEL MN THE INSTRUMENT S PUMP UNIT MUST NOT BE DISASSEMBLED FOR A SERVICE ALWAYS CONTACT BENCHMARK SCIENTIFIC IMPORTANT SAFETY INSTRUCTIONS 1 Check the condition of the unit before each use The housing collection bottle and all components should be carefully inspected for visual damage 2 Check the main power cable and do not connect to power if damage is apparent 3 Before connecting to power check that the electric input data indicated on the data label and the type of plug used correspond to that of the electrical outlet and electrical supply 4 If the plug supplied with the appliance is incompatible with
5. cycle Mains power quality should be that of a variations 40 Ur for 05 cycle typical commercial environment IEC EN 61000 4 11 70 Ut for 25 cycle lt 5 Ut for 5 sec Magnetic field 3A m The power frequency magnetic field should be a measured in the intended installation location to assure that it s sufficiently low porenetonsee eraoianestateretitessumoting IEC EN 61000 4 6 for appliances that aren t life supporting IEC EN 61000 4 3 for appliances that aren t life equipment Note Ur is the value of the power supply voltage ACCESSORIES SUPPLIED DESCRIPTION COMPLETE COLLECTION BOTTLE 2000ml TUBING SET 3 pieces HANDHELD VACUUM PIPETTE 2 ADAPTERS PIPETTE TIP AND FINE POINT METAL TUBE ANTIBACTERIAL FILTER Antibacterial Filter The filter is produced with PTFE hydrophobic material witch prevents fluids entering the pump unit and also offers protection to prevent bacterial contamination of the pump unit When the filter is wet it s not possible to use the unit therefore the filter should be changed immediately In case of possible contamination or discoloration of the filter change the filter immediately Do not use the instrument without the filter properly installed CLEANING OF ACCESSORIES To clean the plastic housing of the device wear disposable protective gloves and clean with denatured alcohol or hypochlorite solutions Washing and or cleaning the autoclavable collection bottle can be carri
6. ed out as follows e Wear protective gloves and apron glasses and face mask if necessary to avoid contact with contaminating substances Disconnect the bottle from the device Disconnect all tubes from the jar and the filter Empty and properly dispose of the contents in the collection bottle Separate all parts of the cover overflow valve o ring Thoroughly rinse all parts in cold water then soak in warm water temperature not to exceed 60 C Wash thoroughly and if necessary use a non abrasive brush to remove and deposits Rinse in running warm water and dry all parts with a soft cloth non abrasive The collection bottle and lid can be autoclaved by placing the parts into an autoclave and running one sterilization stem cycle at 121 C 1 bar relative pressure making sure that the jar is positioned upside down Mechanical resistance of the jar is guaranteed up to 30 cycles of sterilization and cleaning at the indicated conditions EN ISO 10079 1 Beyond this limit the physical mechanical characteristics of the plastic may decrease and replacement of the part is therefore recommended After sterilization and cooling at ambient temperature make sure that the parts are not damaged Assemble the bottle as follows e Place the overflow valve into its seat in the cover under VACUUM connector e Insert floating valve keeping the o ring towards the opening of the cage e Place the o ring into its seat around the cover e After completing assembling
7. nt or misuse
8. operations always make sue that cover seals correctly to avoid vacuum or liquid leakage The aspiration tubes can be sterilized in an autoclave using a sterilization cycle at 120 C DO NOT WASH STERILIZE OR AUTOCLAVE THE ANTIBACTERIAL FILTER MAINTENANCE The ACCURIS ASPIRE LABORATORY ASPIRATOR does not require any routine maintenance or lubrication It is however necessary to inspect the unit before each use Unpack the instrument and always check the plastic parts for any damage that may have occurred during prior use Connect cable to an appropriate electrical outlet with correct voltage and turn the power switch to on Close the aspiration outlet by blocking with a finger and with suction regulator set to maximum check that the vacuum gauge reaches at least 75 kPa 0 75 bar maximum Rotate the knob from right to left and check the aspiration regulating control The vacuum indicator should go down to 25 kPa 0 25 bar Check that no loud noises are present A fuse F 1x4A 250V is located in the plug to protect the instrument In the event of a fuse blowing replace with another or the correct type Fault type 1 The vacuum system doesn t work Cable is damaged the cable External power source failure Check the external power source 2 No aspiration Bottle Cap not screwed down properly Unscrewed the cap and re attach carefully 3 No aspiration Silicone cap seal o ring not seated Unscrew the cap and insert the seal properly
9. t Optional Put some water disinfecting liquid ie 10 bleach to water and or antifoam agent into the collection bottle prior to aspiration of samples It may be filled to 1 3 capacity with this liquid Adding liquid prior to use will provide easier cleaning may help prevent growth of biological contaminants and will also increase the speed to reach full vacuum level e Push the green power switch on the front of the instrument to the on I position to start suction e During operation the collection bottle must remain in a vertical position to prevent the overflow protection valve from cutting off suction Should this happen switch off the device and disconnect the tube from the jar cover from VACUUM outlet e Once finished push switch on O position and unplug e After usage it is recommended to remove the tubing and accessories and clean with water and mild detergent or autoclave AN WARNING NEVER USE THE DEVICE WITHOUT THE COLLECTION JAR AND FILTER PROPERLY INSTALLED REQUIREMENTS FOR RETURNING AND REPAIRING Accuris Brand Products carry a 24 month warranty from the date of purchase covering manufacturing defects In the event of a service related issue contact Benchmark Scientific s Service Department at 908 769 5555 Benchmark Scientific cannot be held liable for accidental or indirect damages should the device be modified repaired without authorization or should any of its component be damaged due to accide
10. the power supply outlet contact qualified staff for replacement of the plug with a suitable type The use of simple or multiple and or extension adapters is not generally recommended 5 Always follow general safety requirements for electrical appliances e Use original components and accessories provided by Benchmark Scientific to guarantee performance and safety Do not operate the aspirator without the bacteriological filter in place between the collection bottle and vacuum Never immerse the aspirator pump unit in water or other liquids Install the aspirator on a stable and flat surface Position the instrument in a way that the air vents on the back and sides are not obstructed Don t use in the presence of flammable gases Don t touch the device with wet hands and always prevent the appliance from coming into contact with liquids Keep out of reach of children and unauthorized personnel Always use care when removing the power cord from the instrument Always use the instrument at a safe distance from heat sources 6 For repairs contact Benchmark Scientific technical services at 908 769 5555 and always use original spare parts and accessories Failure to comply with the above can jeopardise the safety of the user of the device 7 This laboratory instrument must be used exclusively for its intended applications as described in this manual 8 Any different use could be dangerous the manufacturer will not be responsible for damage due
11. vacuum pump system Contact Benchmark Scientific for Service Assistance Benchmark Scientific will provide upon request electric diagrams components list descriptions setting instructions and any other information to assist with quick repair IN THE EVENT OF DAMAGED PARTS OR ABNORMAL OPERATION PLEASE CONTACT BENCHMARK TECHNICAL SERVICE THE MANUFACTURER CANNOT HONOR THE WARRANTEE IF THERE HAS BEEN AN UNAUTHORIZED ATTEMPT AT REPAIRING THE INTERNAL COMPONENTS INSTRUCTIONS e Place the Aspire Aspirator on a flat horizontal surface e Connect the end of the short silicon tube with antibacterial filter to the suction connector e The other tube already connected to the filter has to be connected to the port on the bottle cap labelled VACUUM This is the port with the overflow prevention valve directly below When the collection bottle is 90 full the overflow prevention valve will be activated to block liquid from traveling to the filter and pump system J WARNING Ensure that the filter is installed correctly and that the side marked IN is toward the collection bottle If the filter is installed incorrectly liquid may come through the filter and cause damage to the pump system Connect the long silicone tube to the SAMPLE port on the bottle lid Connect the other end of the long silicon tube to the hand held vacuum pipette Connect the power cord to the device then connect the power cord to an appropriate electrical outle
Download Pdf Manuals
Related Search
Related Contents
Life Fitness Lifecycle 9500HR Service Manual Bétonnez vos chantiers ! TENSIÓMETRO ANÁLOGO KICTeam KW3-CRB15 cleaning media Asrock X58 Extreme Manuel utilisateur série SI Manual Boletin tramit telematica certif inst electricas Kenroy Home 91562RBRZ Instructions / Assembly - Figshare 4 - Tyco Copyright © All rights reserved.