1. as large as possible but instead will be printed on the sheet with 512x512 imager pixels Incorrect DICOM context ID is now permitted Certain systems for example Aloka Toshiba and Siemens ultrasound have been found to use Presentation Context Ds in a manner which is incompatible with Horizon when sending color print jobs via DICOM A grayscale film box is created and then a color image box is sent within its scope but the N SET of the image actually uses two context IDs one for the Color PMM and one for Grayscale PMM This caused a failure in the Horizon software since the color image box being sent doesn t really exist as identified The actual problem is that the FilmSession and FilmBox are N CREATED in one context and the N SET of the images is done in another context Horizon now correctly prints in this situation Obscure PostScript Interface Improvements These improvements are required for proper operation with Vital Images Toshiba GE Advantage Windows and Siemens Magicview workstations They may also be required for other workstations which do not use Codonics Windows print drivers PostScript PPD files now contain proper release version string Obscure e Windows XP driver installer now properly queries for a Windows XP CD Previously the installer queried for a Windows 2000 CD The PostScript driver uninstaller for Windows NT 2000 and XP now correctly removes all Horizon PostScript driver files
2. number of 002 22 or higher on the back of the unit Support for this hardware was present in version 1 5 2 but many customers will upgrade directly to 1 6 0 from 1 5 0 or 1 4 1 Corrected banding problem with test prints The Mixed Patterns test print had horizontal and vertical eray wedges that contained visible banding This has been corrected Corrected problem with dropped LPR images Previously in certain rare circumstances image files transferred via LPR were processed before the transfer was completed This caused a blank cell with the caption Image Processing Failed to be printed instead of the image Obscure Cassette barcode reader window cleaning position corrected Previously the barcode reader was not positioned correctly Obscure Improved safety interlock for cassette barcode reader The laser is now immediately turned off if the scanning mirror movement is not detected Corrected problem with losing date and time Previously under rare circumstances it was possible for the system s real time clock to reset to an incorrect date and time Obscure Automatic downgrading of Control Panel firmware supported Previously the automated software installation process supported upgrading but not downgrading Obscure Eliminated border fill for A4 and 14x17 DirectVista Paper Previously border fill could be turned on for this media which printed off the side of the sheet causing buildup on the print hea
3. when using the Start gt Settings gt Printers menu Previously the driver could not be uninstalled which interfered with installing newer versions of the driver Windows Network Protocol SMB print interface is now supported for Windows 98 and ME This eliminates the need for third party LPR clients for PostScript printing with these operating systems PostScript processing now properly uses Job Settings for pagesize imagable area definition Previously if the pagesize was not specified within the PostScript job the sheet would be always be rendered at Letter size regardless of the media size selected via Job Settings PostScript processing no longer requires the pagesize to match the pagesize of the selected media Previously jobs that had a pagesize mismatch would be rejected Note Image data that is rendered outside the imagable area of the selected media will be cropped PostScript jobs which are created without Codonics Windows drivers and are sent to logical device 7 will have all unmarked space cropped to allow for maximum image size Previously such jobs would be printed with a white border User Interface Improvements Changed Image Settings menu name to Default User Settings for more consistent nomenclature Informational pop up boxes can now be dismissed by pressing any valid key on the Control Panel Previously the user was forced to wait for this type of pop up box to disappear befor
4. Horizon Imager Software Version 1 6 0 Release Notes Differences Since Versions 1 5 0 and 1 5 2 Summary Horizon Version 1 6 0 Software offers many product improvements as detailed below Important Note Once Version 1 6 0 Software is installed onto a Horizon Imager it will not operate properly with any earlier version such as Version 1 4 1 Do not install an earlier software version after installing 1 6 0 New Feature List Added capability for OEM specific system Job Settings files This release adds a system Job Settings file called SYNGOFILM which forces the following settings Gamma 1 0 Contrast 0 LUT Linear Saturation 0 MCM 0 Decimation warn On Support for multiple IP address The most efficient way to take advantage of Horizon s multiple media features is to define multiple virtual printers on a host system with the same IP address but different media settings or DICOM AE titles Some host systems will not allow multiple printers to be set up with the same IP address A single Horizon imager can now be configured with multiple IP addresses to overcome this limitation See Product Bulletin 104 903 001 004 for details Verbose logging for print jobs As jobs are added to the print queue all information on image sheet and job parameters are displayed in the log files Improved color registration on ChromaVista Paper prints The three color planes on ChromaVista prints are no
5. calibration were previously confusing because they did not accurately represent the internal values The default values are now always displayed as 0 00 Improved Multi Modality test print The Multi Modality test print has been updated with more modern medical images Color test image improved Cassette barcode reading reliability improved Software for reading and decoding the cassette barcode has been made more reliable resulting in fewer failed reads Ribbon patch detection improved The software for detecting each color patch on a Chroma Vista ribbon has been made more robust This improvement helps to prevent skipping of patches resulting in incorrect colors especially with A4 paper End of ribbon handling improved Previously if the end of the Chroma Vista ribbon was reached and the user opened the top cover and then closed it without replacing the ribbon the imager would try to printing using the spent ribbon Obscure Improved centering of printable area on A4 ChromaVista and DirectVista Paper Previously images printed on A4 paper were shifted approximately 0 100 to the right Job Settings file names properly truncated Remotely sent Job Settings file names gt 16 characters are now properly truncated to 16 characters Previously this would cause a software error requiring the system to be rebooted Currently Job Settings file names gt 64 characters still cause the imager to lock up requiring a
6. d Defects Improved Faster startup and shutdown times Startup time is approximately 20 seconds faster than with 1 5 2 software Shutdown time is approximately 12 seconds faster Faster printing Time to first print for a typical 14x17 film job is approximately 11 seconds faster than with 1 5 2 software DICOM Interface Improvements DICOM format STANDARD 1 1 is now handled correctly Previously images using this format were padded with a border causing undesirable downscaling Allow a DICOM photometric interpretation of RGB for Grayscale Image Boxes The GE Entegra system which is based on ISG Cedera Hardcopy Server software specifies a photometric interpretation of RGB for grayscale images In this case the value is disregarded and the default value of MONOCHROME is substituted Obscure gt O CODONICS DICOM photometric interpretation of MONOCHROME is now handled correctly Grayscale images that specify this parameter are now properly printed with reverse polarity Obscure DICOM odd image sizes are now properly handled Previously the imager would not print DICOM images with an odd number of pixels Obscure Images are now left unscaled when the DICOM scaling type is specified as none Previously a scaling type of none caused the default scaling setting to be used For example a 512x512 image sent with a scaling type of none will not be printed
7. d Obscure Disconnected smart card reader no longer causes the imager to continuously reboot Imager will now report an invalid smart card in such circumstances Obscure Loading system configurations from a zip disk no longer causes a double reboot Added automatic shutdown when internal temperature limits exceeded CPU temperature above 70 degrees C or below 0 degrees C will cause a controlled shutdown completing a print in progress CPU temperature above 74 degrees C will cause an immediate shutdown Eliminated extraneous pick tire movement Previously the pick tires would spin in cassette 1 after the last sheet of 14x17 film was picked Obscure Corrected defects with supply cassette selection Previously the imager did not continue with the current supply cassette after a reboot but instead would select the first available cassette Now it will continue with the same cassette as long as it is still loaded Also if a cassette is empty and another cassette with the same media is available the imager will correctly switch to the next cassette even if the previous empty cassette is replaced before the next sheet is printed Obscure Software support for improved side edge sensors Newer imagers have improved side edge sensors which increase the range of side edge detection This eliminated a problem where the position of the outer black film border was incorrect in some cases Such imagers can be identified a CNFG
8. dvance the ribbon too far causing unnecessary ribbon consumption Automatic color print detection currently checks only the first image on a sheet An empty VMF canvas a sheet with no cells will result in an error and cause the imager to reboot Only special custom applications use VMF Variable Multi Formatting Neither DICOM nor PostScript use VMF SA 1000 not fully supported Extraneous error messages on SA 1000 Network Framegrabber related to Horizon imager status Captions not supported with landscape printing Color saturation adjustment not yet supported Installed captions appear on color test print Workaround is to turn captions off in the Default User Settings menu Chroma Vista print jam clearing problems A Chroma Vista print which has jammed while rewinding the sheet for the next color pass reports a second jam after the first one has occurred Also such a jammed print cannot be purged If such a jam occurs the work around to purge the print is as follows gt O CODONICS 1 After the jam has occurred the top cover will be opened and the Control Panel will display the following error message Clear media jam from the printing area 2 Remove the receive trays and clear the partially printed sheet 3 Insert the receive trays 4 Close the top cover 5 The imager will now attempt to automatically resume printing a pop up box will read Printing resumed Press the PAUSE key immediately a
9. e proceeding with Control Panel operations When a Control Panel data entry operation is cancelled by pressing the MENU key followed by the CANCEL key the cancelled operation is now confirmed by a pop up box Obscure Please wait pop up box now displays on Control Panel when requesting a print head calibration during a print Obscure Pop up box confirms now reset operations on Control Panel for example resetting the default media Smart card removal now causes Control Panel beep regardless of fault tone or volume settings MENU key is now available from help screens Corrected problem where Control Panel keys would beep even through key beep had been turned off Imager no longer reboots unnecessarily when network addresses menu is entered but no changes made Help text added to MCM menu Corrected problem where key beeps were disabled after entering a feature key Obscure Improved key repeat operation when holding down Control Panel keys It is now easier to use Other miscellaneous user interface clean up O CODONICS Improved DirectVista Film density accuracy Sensitometry tables have been improved to allow more accurate mapping of image data to printed output densities This also eliminates contouring effects on certain test patterns such as smooth gray wedges Eliminated incorrect displayed values for densitometer calibration defaults The displayed values for the default densitometer
10. fter this happens 6 Remove the ChromaVista paper cassette 7 Purge All Print Jobs from the Control Panel Utilities menu Note this purges all queued jobs 8 Replace the ChromaVista paper cassette Dmin and Dmax incorrectly applied to ChromaVista and DirectVista Paper This applies to Job Settings or remotely configured User Settings does not apply to Default User Settings configured from the Control Panel Workaround is to set Dmin to 0 and Dmax to 300 for all paper prints Software cannot be downgraded from 1 6 0 to an earlier version For 1 5 0 or 1 5 2 work around is to use the 1 5 0 KDU Firmware Downgrade Disk part number SP 00147 There is no work around for versions earlier than 1 5 0 Automatic file system cleanup is not performed when captions are installed For optimal performance work around is to uninstall captions reboot imager and reinstall captions approximately once per month A4 paper size unavailable in Windows 2000 PostScript driver Work around is to select Letter for the paper size and Scaled for the sheet layout Macintosh formatted Job Settings Files to not work correctly Workaround is to save Job Settings Files in MS DOS or Unix format before sending them to Horizon System Job Settings file 1417 dvf does not work Using a system Job Settings file name or DICOM AE title of 1417 dvf does not direct jobs to 14x17 DirectVista blue film as indicated in the Use
11. long FTP filenames and for filenames containing spaces or special characters FIP connections now properly support file names containing the characters amp Also filename length maximum has been increased to 255 characters Corrected several memory leaks and performance leaks Previously system memory was slowly consumed and performance slowly degraded eventually causing the imager to stop queuing jobs corrected by rebooting after several hundred prints This has been corrected A memory watchdog has been added to monitor system memory and take corrective measures if required gt O CODONICS e An automatic file system cleanup now occurs every three weeks Since a reboot is required for this operation the user is reminded via a system alert to reboot when appropriate Correct several log file inefficiencies which were slowing system performance Viewable custom User Settings Custom User Settings can now be viewed by logging into the imager and typing the command stat user where user is the user ID used to create the User Settings Obscure Corrected rare system lockup Previously in rare circumstances it was possible for an imager to lock up requiring the power to be cycled Obscure Cassette barcode properly read after booting with top cover open and subsequently closing top cover Previously this operation would require the cassette to be removed and reinserte
12. r s Manual Workaround is to use 1417 dvfb CODONICS Get it all with just one call 1 800 444 1198 Copyright 2002 by Codonics Inc All registered and unregistered trademarks are the property of their respective owners Specifications subject to change without notice Printed in the U S A Part No 901 080 001 Rev A We bring the future into focus 17991 Englewood Drive Middleburg Heights OH 44130 USA 440 243 1198 440 243 1334 Fax Email info codonics com www codonics com
13. reboot Obscure Miscellaneous reliability improvements A variety of additional minor changes were made to the imager software to improve overall operation and reliability Known Issues PostScript Interface Issues Windows Network Protocol SMB not yet supported for PostScript printing from Windows NT 2000 or XP Workaround is to print via LPR interface e Imagable area used by PostScript RIP does not always match imagable area on printed sheet In certain cases this prevents the entire printable area from being utilized There is currently no driver uninstaller for Windows NT 4 0 or XP DICOM Interface Issues e DICOM hosts that leave associations open will cause the Horizon Activity light to flash even though there are no jobs queued e Unsupported DICOM image compression results in an out of memory error at the host e DICOM file format not yet implemented for non DICOM interfaces such as LPR and FIP Using a Magnification Type of NONE with the specialBracket device causes blank cells with the caption Image Processing Failed to be printed instead of the image Work around is to set the DICOM Magnification Type to something other than NONE FTP server does not support passive mode connections Workaround is to disable passive mode in FIP client Gamma Contrast bracket prints say TCR instead of Saturation Unnecessary ribbon consumption Some systems will occasionally a
14. w more precisely aligned producing consistently sharper looking prints Release Notes Blue and Clear DirectVista Film calibration now works correctly Previously an imager that had been calibrated first with blue film would print too dark on clear film and an imager that had been calibrated first with clear film would print too light with blue film see Product Bulletin 103 903 001 003 Print counter Imager now stores the total number of ChromaVista and DirectVista sheets printed This information is currently accessible only by a factory technician Invalid network gateway handling corrected Previously when upgrading a system that had an invalid network gateway configured e g did not match the network portion of the IP address the system would reset all network settings to default values Now it resets only the gateway to the default value Obscure Prevent shutdown when print head not parked To safely move the imager and avoid permanent damage the print head needs to be in the parked position all the way back This is done automatically by performing a soft power off using the power key on the Control Panel If the receive trays are removed or the top cover is open a safety interlock will prevent the print head from moving to the parked position If a soft power off is initiated and the print head cannot park the power off will be cancelled the imager will beep and a fault message will be displayed Support for