1. Installation and User Guide IntegratlR NIR Integrating Sphere The information in this publication is provided for reference only All information contained in this publication is believed to be correct and complete PIKE Technologies Inc shall not be liable for errors contained herein nor for incidental or consequential damages in connection with the furnishing performance or use of this material All product specifications as well as the information contained in this publication are subject to change without notice This publication may contain or reference information and products protected by copyrights or patents and does not convey any license under the patent rights of PIKE Technologies Inc nor the rights of others PIKE Technologies Inc does not assume any liability arising out of any infringements of patents or other rights of third parties This document contains confidential or proprietary information of PIKE Technologies Inc Neither this document nor the information herein is to be reproduced distributed used or disclosed either in whole or in part except as specifically authorized by PIKE Technologies Inc PIKE Technologies Inc makes no warranty of any kind with regard to this material including but not limited to the implied warranties of merchantability and fitness for a particular purpose Copyright 1991 2016 by PIKE Technologies Inc Madison WI 53719 Printed in the United States of America All
2. intensities shape of the NIR beam the beam splitter characteristics vary from bench to bench thus the single beam spectra could be slightly different PN 350 048601 01 Page 5 Figure 2 NIR single beam energy spectrum Figure 3 100 transmission spectrum for noise evaluation PN 350 048601 01 Page 6 NIR Sampling Sampling with the NIR IntegratlR is very easy because generally no sample preparation is required To obtain a spectrum place the diffuse gold reference over the sample port and collect a background spectrum Replace the reference with the sample and collect the sample spectrum IMPORTANT NOTE After use please replace protective cap to avoid damage or contamination to gold surface Samples for near infrared need to be presented to the spectrometer in an optically uniform and reproducible fashion to obtain the same spectrum every time Parameters that have significant effect on the samples are particle size and distribution crystallinity phase melted or solidified oils for example have very different spectra compression temperature especially water containing samples and other physical or chemical differences The round plastic cap enclosed in the kit with the IntegratlR serves as a light shield to protect the detector from external light getting into the sphere The InGaAs detector is very sensitive and could pick up light from the fluorescent room light When running thin samp
3. or with a camel hair brush Under no circumstances must the mirrors be rubbed with paper products such as Kleenex since this may produce scratching of the mirror coating Options Part Number Description 048 2999 Sample Vials Pack of 25 048 3000 Diffuse Gold Reference 044 3010 21 mm Glass Vial Holder 048 0151 Beaker Adapter 048 1050 Rotating Sample Holder for inhomogeneous sample 048 3070 Extended Range NIR Wavelength Standard 10 000 4000 cm Figure 6 NIR IntegratiR with Beaker Adapter Figure 7 NIR IngetratiR with Rotating Sample Holder PN 350 048601 01 Page 10 PIICE TECHNOLOGIES Grete Cretu 6125 Cottonwood Drive Madison WI 53719 5120 608 274 2721 TEL 608 274 0103 FAX info piketech com www piketech com
4. world rights reserved No part of ti publication may be stored in a retrieval system transmitted or reproduced in any way including but not limited to photocopy photograph magnetic or other record without the prior written permission of PIKE Technologies Inc Address Comments to PIKE Technologies Inc 6125 Cottonwood Drive Madison WI 53719 Phone 608 274 2721 Fax 608 274 0103 E mail info piketech com Web Site www piketech com Jan 1 2016 Contents PN 350 048601 01 Introduction IntegratIR series Near Infrared Integrating Sphere module is a compact sampling accessory for measuring the composition of a wide variety of solid and paste materials Using near infrared chemometrics qualitative product identification or quantitative analysis may be performed on pharmaceutical nutraceutical chemical polymer textile food agricultural or other samples The IntegratI module features a 2 high reflectivity gold coated integrating sphere and a high speed low noise Indium Gallium Arsenide detector with transfer optics and interface electronics The IntegratiR fits in the sample compartment of the FTIR spectrometer and connects to the external detector port of the spectrometer The 0 5 diameter borosilicate glass window serves as a sampling port on the top of the integrating sphere The window is bonded and sealed to protect the sphere from corrosive materials and contamination The optical design of the NI
5. R IntegratIR directs the beam through the bottom of the sphere at a near normal angle and is focused at the sample position at the top of the sphere Figure 1 Optical diagram of the NIR IntegratiR PN 350 048601 01 Page 1 Unpacking Your Accessory In order for you to quickly verify receipt of your accessory we have included a packing list Please inspect the package carefully Packing NIR IntegratiR User Manual NIR IntegratiR Accessory Diffuse Gold Reference PN 350 048601 PN 048 60XX PN 048 3000 Quantity 1 Quantity 1 Quantity 1 Light Shield Vials Vial Holder Quantity 1 PN 048 2999 PN 044 3010 Quantity 1 Pack of 25 Quantity 1 Spoon Camel Hair Brush Spatula Quantity 1 Each Hex Wrenches Quantity 1 Set PN 350 048601 01 2 Installation Before installation make sure the spectrometer is working in the near infrared region without the IntegratiR Please consult the manufacturer of the FT IR spectrometer regarding the conversion of the bench to near infrared This may involve replacing the beam splitter and source to optimize for the 10 000 to 4000 cm wavelength region of the IntegratiR It is advisable to perform an alignment of the interferometer before inserting the IntegratiR into the sample compartment Each IntegratiR unit is custom manufactured to interface with a specific model FTIR Spectrometer Operation with a different model may require a factory conversion to be done b
6. allow the best signal to noise however the signal PN 350 048601 01 Page 3 magnitude cannot exceed the A D converter required signal magnitude The interferogram signal therefore should be optimized to the maximum allowable using the aperture selection on the spectrometer In the near infrared high resolution is very rarely used thus the default aperture could be in the fully open position and may need to be reduced Accessory Adjustment The accessory has been aligned and tested to ensure that it performs to specifications and additional alignment may not be necessary In order to optimize the beam positioning in your instrument small adjustments may be required This should be preferably done by trained service personnel The alignment procedure is as follows 1 With no sample on the sampling window and covering the window with the enclosed Shielding Cap establish the magnitude of the interferogram In most instruments with proper alignment it should be a fraction of a volt size because all the light is exiting the window The remaining small interferogram is caused by the residual small fraction of the incoming light reflected back from the window itself On most instruments the visible part of the light is observable by placing a semi transparent paper such as vellum or scotch tape over the window 2 Should the illuminated spot not be centered part of the light will be reflected back from the sphere causing an unwanted backgrou
7. les it is advisable to cover the sample with the Light Shield Normal room light fluorescent or incandescent will not damage the detector but may end up adding noise or spurious peaks to the spectrum Polymers For plastic sheets painted test panels and other flat samples the reproducible sample positioning has to be assured If the sample does not lay flat under its own weight a small weight could be placed on the sample to press it against the window to aid in the accurate and repeatable sampling PN 350 048601 01 Page 7 Powders and Pastes Other materials such as powders and pastes can also be directly placed on the window The choice whether you use a vial or a microscope slide to place the sample on depends on how easy it is to clean the sample off the window The window is sealed using epoxy compound thus itis relatively chemically inert and can be cleaned preferably with alcohol Transparent and Semi Transparent Films In all cases the accessory is measuring the reflected light For transparent materials films produce very small signal In certain cases a reflective mirror can aid in returning the near infrared light to the sphere detector This double transmission called transflectance is a convenient way of measuring transparent or semi transparent samples that have a reproducible thickness or pathlength Use the Gold Reference placed on the window for background spectrum Samples in Vials Samples liquids solids powder
8. nd To align the beam the front panel of the IntegratIR should be removed 4 slotted screws hold the front panel NOTE Please be careful the inside of the IntegratlR contains sensitive optical and electronic components Avoid touching the gold surfaces of the mirrors 3 Left right and forward backward alignment of the illuminated spot can be aligned by a 3 32 hex wrench The IntegratiR should come pre aligned thus only a very small adjustment may be needed During alignment the spectrometer should be in monitor mode The goal of the alignment is to PN 350 048601 01 Page a minimize the stray light detector signal while making small adjustments such as 1 16 of a turn at a time with the adjustment screws 4 Replace the front panel by tightening the four slotted screws Detector Gain Adjust service only Forward Backward Adjust Left Right Adjust Performance Verification With the accessory installed and the appropriate aperture set place the diffuse gold reference onto the sample port located at the top of the sphere Using a 1 minute data collect time collect a background followed immediately by collecting a sample spectrum to obtain a 100 transmission spectrum The peak to peak noise between 6200 6000 cm should be less than 0 08 T Typical single beam and 100 spectrum are shown in Figures 2 and 3 respectively The spectral results using the IntegratiR depend to some degree on the spectrometer Source
9. s can be placed in the enclosed glass vials and directly measured using the vial holder ring to position the vial reproducibly Other glass containers can also be used for samples but you have to assure that the glass is uniform reproducible and as thin as possible to position the sample surface close to the integrating sphere Sample Mount The Sample Mount is the blue plate on top of the IntegratlR It can be centered to hold the Vial Holder or the Gold Reference Sample Mount can be rotated to get direct access to the window area for cleaning or for larger samples Sample Mount can be removed by pulling on the black knurled knob straight upward If the removal requires more force please hold the top plate down while pulling the knob PN 350 048601 01 Page 8 Typical Spectra The NIR IntegratiR offers a simple method of measuring a wide range of samples Example spectra are shown Figure 5 NIR spectrum of aspirin powder collected using the NIR IntegratiR PN 350 048601 01 Page 9 Precautions In order to provide the maximum signal in the infrared with the minimum spectral interference the mirrors used in this device are gold coated Since the coatings are soft care must be taken to avoid damage Normally these mirrors will not need cleaning since they are contained within the housing of the accessory If they do need cleaning they may be gently wiped with a lint free abrasive free cloth such as lens tissue
10. y PIKE Technologies Place the accessory in the sample compartment of the spectrometer The baseplate of the accessory assures the alignment of the accessory relative to the beam coming from the spectrometer A captive thumb screw on the baseplate of the IntegratiR attaches the accessory to the sample compartment baseplate of the bench and prevents it from rotating from the nominal position Connect the detector cable to the spectrometer CAUTION Make sure power to the spectrometer bench is off before connecting the detector cable For ThermoFisher instruments use the cable attached to the 15 pin D connector in the sample compartment of the spectrometer Select the appropriate source optical path and detector Please consult the software manual provided by the spectrometer manufacturer concerning how to perform the detector and source selection To connect the IntegratlR please turn off the spectrometer After the accessory detector cable is connected to the instrument and the accessory power up the instrument Spectrometer Settings Set the collection parameters as follows Wavelength Range 10 000 4000 cm Velocity Select appropriate for InGaAs detector for your FTIR spectrometer Aperture Appropriate J stop Resolution 4or8cm ScanTime 30 605 Place the Diffuse Gold Reference on the sample window to maximize the light level in the sphere The gain of the IntegratIR is set during manufacturing to