Home

Keysight-Agilent E3645A Datasheet

image

Contents

1. offset Activation Time lt 0 5 0 5 V lt 1 5 msec OVP gt 3 V lt 10 msec OVP lt 3 V Temperature Coefficient per C output offset Voltage Current lt 0 01 3 mV lt 0 02 5 mV for output 2 of E3646 47 48 49A lt 0 02 3 mA 1 Accuracy specifications are valid after a 1 hour warm up with no load and calibration at 25 C Maximum time required for the output voltage to change from 1 to 99 or vice versa following the receipt of VOLTage or APPLy command via direct GPIB or RS 232 interface 3 Average time for output to start and drop after an OVP condition occurred Stability constant load amp temperature output offset 8 hrs lt 0 02 2 mV lt 0 1 1 mA Voltage Current Remote Sensing 1V Max voltage drop in each load lead AC Input 47 Hz 63 Hz Warranty 100 Vac 10 Opt 0E9 115 Vac 10 Std 230 Vac 10 0E3 Designed to comply with UL3111 1 certified to CSA 22 2 No 1010 1 conforms to IEC 1010 1 complies with EMC directive 89 336 EEC Group 1 Class A Product Regulation Agilent E364xA Series Power Supplies Rackmount Kits E3640A 30 Watt Single Power Supply E3641A 30 Watt Single Power Supply E3642A 50 Watt Single Power Supply E3643A 50 Watt Single Power Supply E3644A 80 Watt Single Power Supply E3645A 80 Watt Single Power Supply E3646A 60 Watt Dual Power Supply E3647A 60 Watt Dual Power Supply E3648A 100 Watt Dual Powe
2. Test Equipment Solutions Datasheet Test Equipment Solutions Ltd specialise in the second user sale rental and distribution of quality test amp measurement T amp M equipment We stock all major equipment types such as spectrum analyzers signal generators oscilloscopes power meters logic analysers etc from all the major suppliers such as Agilent Tektronix Anritsu and Rohde amp Schwarz We are focused at the professional end of the marketplace primarily working with customers for whom high performance quality and service are key whilst realising the cost savings that second user equipment offers As such we fully test amp refurbish equipment in our in house traceable Lab Items are supplied with manuals accessories and typically a full no quibble 2 year warranty Our staff have extensive backgrounds in T amp M totalling over 150 years of combined experience which enables us to deliver industry leading service and support We endeavour to be customer focused in every way right down to the detail such as offering free delivery on sales covering the cost of warranty returns BOTH ways plus supplying a loan unit if available and supplying a free business tool with every order As well as the headline benefit of cost saving second user offers shorter lead times higher reliability and multivendor solutions Rental of course is ideal for shorterterm needs and offers fast delivery flexibility try before you buy zero capital expendi
3. ftware drivers are available for Agilent VEE National Instruments LabView and LabWindows With these drivers integration of the E364xA into your system can never be any easier The drivers are supported under Microsoft Windows 98 and NT Front Panel Operation An easy to use rotary knob and self guiding keypads allow you to set the output at your desired resolution without any effort Also both voltage and current levels can be set to a maximum resolution of 10 mV 1 mA from the front panel In addition you can store and recall for up to five complete power supply setups using the internal non volatile memory The output on off button sets the output to zero If you own a dual output model you can view two voltages or currents that are displayed simultaneously Versatile Power Agilent E364xA power supplies give you the flexibility to select from dual output ranges Output load is protected against overvoltage which can be easily monitored and adjusted from the front panel and remote interface Remote sensing Is available in the rear terminal to eliminate errors caused by voltage drops on the load leads These power supplies offer new versatile binding posts on the front panel and screw type terminals on the rear panel New front panel binding posts allow you to use safety test leads as well as conventional banana clips and stripped wires An optional rackmount kit is available The Agilent E364xA Series emplo
4. ing Price With the output power of 30 to 100 W the Agilent E364xA Series programmable DC power supplies provide great performance at a great price All ten models deliver clean power excellent regulation fast transient response and built in GPIB and RS 232 interfaces They are designed to meet the needs of R amp D design verification production testing QA verifications and other demanding applications with Agilent Technologies s quality and reliability Steady Output With 0 01 percent load and line regulation Agilent E364xA power supplies are able to maintain a steady output when power line and load changes occur They also specify normal mode voltage noise and low common mode current noise The low normal mode noise specification assures clean power for precision circuitry applications and the low common mode current provides isolation from power line current injection Agilent E364xA power supplies specify less than 90 msec of voltage settling time at any output load condition Remote Interface Agilent E364xA power supplies support any PC with a GPIB IEEE 488 card or RS 232 interface Every model ships standard with both GPIB and RS 232 The easy to use SCPI Standard Commands for Programmable Instruments allow fast and simple programming procedures Besides the user manual provides Sufficient information on programming for all end users from beginners to veterans Broad Support VXI plug and play so
5. nd open Agilent Open simplifies the process of connecting and programming test systems to help engineers design validate and manufacture electronic products Agilent offers open connectivity for a broad range of system ready instruments open industry software PC standard 1 0 and global support which are combined to more easily integrate test system development LAI is the US registered trademark of the LXI Consortium www agilent com For more information on Agilent Technologies products applications or services please contact your local Agilent office The complete list is available at www agilent com find contactus Phone or Fax United States tel 800 829 4444 fax 800 829 4433 Canada tel 877 894 4414 fax 800 746 4866 China tel 800 810 0189 fax 800 820 2816 Europe tel 31 20 547 2111 Japan tel 81 426 56 7832 fax 81 426 56 7840 Korea tel 080 769 0800 faxf 080 769 0900 Latin America tel 305 269 7500 Taiwan tel 0800 047 866 fax 0800 286 331 Other Asia Pacific Countries tel 65 6375 8100 fax 65 6755 0042 Email tm_ap agilent com Revised 1 06 Product specifications and descriptions in this document subject to change without notice Agilent Technologies Inc 2002 2007 Printed in USA October 26 2007 5968 7355EN Agilent Technologies
6. r 0to8V 8A or 0 to 35 V 2 2 A or 0 C to 40 C 0 to 20 V 1 5 A Oto 60 V 0 5 A 0 to 20 V 2 5 A Oto 60 V 0 8 A Oto 20V 4A 0 to 60 V 1 3A Net Weight 53kg 5 2 kg 6 3 kg 6 2 kg 6 6 kg 6 7 kg Dimensions without bumper 212 6 mm W x 88 5 mm H x 348 3 mm D Model Number E3646A E3647A E3648A E3649A DC Output Rating Two Two Two Two 0 C to 40 C 0to8V 3A or 0 to 35 V 0 8 A or 0to8V 5A or 0 to 35 V 1 4 A or 0 to 20 V 1 5 A 0 to 60 V 0 5 A 0 to 20 V 2 5 A 0 to 60 V 0 8 A Dimensions without bumper 212 8 mm W x 433 0 mm H x 348 3 mm D Load and Line Regulation of output offset Voltage Current lt 0 01 3 mV lt 0 01 250 pA Ripple and Noise 20 Hz to 20 MHz Normal Mode Voltage Normal Mode Current Common Mode Current lt 5 mVpp 0 5 mVrms for 84 720 M models lt 8 mVpp 1 mVrms for35 V 60 V models lt 4mArms lt 1 5 pArms Accuracy 12 Months 25 C 5 C output offset Programming Voltage Current Readback Voltage Current Resolution lt 0 05 10 mV lt 0 1 25 mV for output 2 of E3646 47 48 49 A lt 0 2 10 mA 005 5 mV lt 0 1 25 mV for output 2 of E3646 47 48 49 A lt 0 15 5 mA lt 0 15 10 mA for output 2 of E3646 47 48 49 A Readback lt 2mV 1 mA Transient Response Less than 50 usec for output to recover to within 15 mV following a change in output current from full load to half load or vice versa OVP Accuracy output
7. r Supply E3649A 100 Watt Dual Power Supply Accessories Included User s Guide Quick Reference Guide and AC power cord Power Options Opt 0E3 230 Vac 10 Opt OEM 1150 Vac 10 Opt 0E9 100 Vac 10 Other Options Opt 1CM Rackmount kit Single Output P N 5063 9240 Dual Output P N 5063 9243 Opt 0L2 Extra Manual Opt UK6 Commercial calibration with test result data E3600A 100 Test lead kit Agilent E3640A 41A 42A 43A 44A 45A To rackmount two instruments side by side Lock link Kit P N 5061 9694 Flange Kit P N 5063 9212 To rackmount one or two instruments in a sliding support shelf Support Shelf P N 5063 9255 Slide Kit P N 1494 0015 required for support shelf For a single instrument also order Filter Panel P N 5002 3999 Agilent E3646A 47A 48A 49A To rackmount two instruments side by side Lock link Kit P N b061 9694 Flange Kit P N 5063 9214 To rackmount two instruments in a sliding support shelf Support Shelf P7N 5063 9256 Silde Kit PAN 1494 0015 Rackmounting with 1CM or lock link flange kit requires Agilent or customer support rails Agilent Support Rails E3663AC pel Agilent Email Updates www agilent com find emailupdates Get the latest information on the products and applica tions you select o Agilent Direct www agilent com find agilentdirect Quickly choose and use your test equipment solutions with confidence Agilent f Open aa www agilent com fi
8. ture lower risk and off balance sheet accounting Both second user and rental improve the key business measure of Return On Capital Employed We are based near Heathrow Airport in the UK from where we supply test equipment worldwide Our facility incorporates Sales Support Admin Logistics and our own in house Lab All products supplied by Test Equipment Solutions include No quibble parts amp labour warranty we provide transport for UK mainland addresses Free loan equipment during warranty repair if available Full electrical mechanical and safety refurbishment in our in house Lab Certificate of Conformance calibration available on request Manuals and accessories required for normal operation Free insured delivery to your UK mainland address sales Support from our team of seasoned Test amp Measurement engineers 1509001 quality assurance Test equipment Solutions Ltd Unit 8 Elder Way Waterside Drive Langley Berkshire SL3 6EP T 44 0 1753 596000 F 44 0 1753 596001 Email info TestEquipmentHQ com Web www TestEquipmentHQ com Agilent A E3640A E3649 SOK Agilent Technologies Reliable Power Repeatable Results Single and dual outputs e Dual range output e 30 Wto 100 W output power Front and rear output terminals e Over voltage protection e Remote Sensing e GPIB and RS 232 standard e Save and recall functions Great Performance Outstand
9. ys a cooling fan with automatic speed control to reduce the acoustic noise LabView and LabWindows are registered trademarks of National Instruments Microsoft Windows 98 and Windows NT are US registered trademerks of Microsoft Corporation Set the overvoltage protection and enable the overvoltage protection circuit Highly visible vacuum y n Rotary knob for quick flourescent display l and analog like control Ye Agilent E2840A 0 v 00 0 209 170 _ of voltage and current F Tough handle for easy carriage and unit Selectable dual range prop up provides flexibility and convenience ar Versatile binding posts e offer flexibility to use safety test leads banana plugs or stripped wires Store and recall up to five different operating Electronic calibration from the front panel statuses in non volatile storage locations 1 year warranty protects your investment G ES SW ABN Cooling fan with oi Ge aspas ns pao anima o service trend personal We automatic speed y FP control for low FUBEl 2501 acoustic Noise Built in GPIB and I es a lh ao RS 232 interfaces Pir rere mae Rear output and WARNING For continued fire protection use specified line fuse E y sensing terminals Agilent E3640A E3649A Programmable DC Power Supply Specifications 0 W 50 W 80 W Maximum Power 3 of Outputs DC Output Rating 0 to 8 V 3 A or 0to35V 0 8Aor 0to8V 5Aor 0 to 35 V 1 4A o

Download Pdf Manuals

image

Related Search

Related Contents

  "53 jours" : pour lecteurs chevronnés  Pesquisa aplicada ao jogo Tetris  01 hdt liquido para frenos lf3 _dot-3  Dell D630 User's Manual  Lighting - Martin Atomic 3000 Strobe  DES IMAGES POUR L`ÉGALITÉ - République et Canton du Jura  2001 Catalog - Pages 241-280  Thanks for downloading this user manual. This document was  Bischoff & Bischoff Mode d`emploi BB-80  

Copyright © All rights reserved.
Failed to retrieve file