1. First you close the sterilizing valve 9 and afterwards you open the release valve for chamber pressure 25 carefully In order that the escaping steam is cooled down before entering the drain open the cold water valve 29 so that a temperature of max 50 C is reached This procedure is continued until the pressure of the chamber 1 is 0 bar to be seen at the pressure gauge 19 for the sterilizing chamber The closing of the sterilizing valve 9 has the effect that the steam flow between outer chamber 2 and sterilizing chamber 1 is interrupted The separation of both chambers 1 and 2 allows that the sterile goods can be removed from the sterilizing chamber while the operating pressure in the outer chamber is maintained for further operation Open the sterilizer lid After pressure reduction indication 0 bar at the pressure gauge 19 for the sterilizing chamber the sterilizer lid 3 may be opened and the sterile goods be removed if the temperature has decreased to approx 60 C to be seen at the pointer thermometer For that purpose the function safety valve 23 is opened with the operation lever 24 operation lever 24 must be brought to the right side in its horizontal limit position The eventually remaining overpressure now escapes via the exhaust tube of the function safety valve 23 Attention It is still possible that steam escapes at the exhaust tube of this valve By means of the operation handles 3a
2. yellow 17 Control lamp water shortage red 19 Pressure gauge for sterilizing chamber 1 20 Pressure gauge for steam jacket 2 21 Adjusting device for sterilizing pressure with regulation 23 Function safety valve with exhaust tube 24 Operating lever for 23 25 Release valve for chamber pressure for sterilizing chamber 26 Automatic aerator and deaerator 27 Return valve 28 Cooler 28a Drain 29 Inlet valve for cooling water 30 Safety valve with exhaust tube 31 Emptying valve for 2 32 Apparatus feet with fixation holes 33 Gasket for 1 3 34 Indicating thermometer for 1 Remark All valves are in closed position when they are moved clockwise seen on the operation handle until the limit stop D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 31 D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 32 D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 33 D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 34 D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 35 D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 36
3. If the radiators 10 are not switched off due to a failure of the regulator for sterilization pressure 21 and an inadmissible high steam pressure is established in the outer chamber the safety valve 30 reacts upon an overpressure of more than 2 5 bar It opens the way outside removes safely the surplus pressure via a tube thus preventing dangerous overpressure in the outer chamber In case of reaction of the safety valve switch off the apparatus by the ON OFF switch 13 and examine the apparatus D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 26 9 5 Table Programme Goods to be Jacket Sterilizing Sterilizing Sterilizing Method of sterilized pressure pressure temperature time cooling Instruments Quick steam M Glas and 1 3 bar 1 4 bar 1 2 bar 121 C 30 Min reduction or Plastic goods Self cooling Instruments Quick steam A2 Glas and 2 5 bar 2 6 bar 1 2 bar 121 C 30 Min reduction or Plastic goods Self cooling Instruments Quick steam B Glas 2 5 bar 2 6 bar 2 4 bar 134 C 10 Min reduction or Self cooling Goods which may be sterilized In programme A 1 2 bar 121 C only unwrapped solids with a destroy temperature S 121 C may be sterilized In programme A 2 4 bar 121 only unwrapped solids with a destroy temperature lt 121 C may be sterilized In programme B 2 4 bar 134 C only unwrapped solids like instruments and other similar
4. Programme A2 121 C jacket pressure 2 4 bar Adjustment Loading of the sterilizer Close sterilizing lid Heating up time Sterilization End of programme Open the sterilizer lid Removal of sterile goods Preparations for further sterilizations Switching off of sterilizer D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 8 Programme B 134 C jacket pressure 2 4 bar 22 8 1 Adjustment 22 8 2 Loading of the sterilizer 22 8 3 Close sterilizing lid 22 8 4 Sterilization 134 C 23 8 5 End of programme 24 8 6 Open the sterilizer lid 24 8 7 Removal of sterile goods 24 8 8 Preparations for further sterilizations 25 8 9 Switching off of sterilizer 25 9 Savety Devils snra ii iva iu 26 9 1 Water shortage in the outer chamber 26 9 2 Function safetv valve 26 9 3 Sterilizer lid 26 9 4 Overpressure safetv device 26 9 5 Table 27 10 Maintenance of Sterilizer esse 28 TI Sterili et Qasket i i ienet 28 12 Search TOF EIFOES cei teda cs 29 13 Short Operating Instruktions 30 tai T 31 D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 3 Description Name and address of manufacturer KSG Sterilisatoren GmbH Tel BuchhoferstraBe 5 Fax D 82140 Olching E Mail Order No XXXXXX Aukl No Aukl 002 067 Appar
5. e g after the last load of the day the ON OFF switch 13 has to be switched off light goes out and the inlet valve for cold water 29 has to be closed D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 16 7 2 7 2 1 7 2 2 7 2 3 Programme A2 121 C jacket pressure 2 4 bar Adjustment With max jacket pressure adjustment 2 4 bar The saturated steam necessarv for sterilization is produced in the steam jacket 2 bv means of two electric radiators 10 To start the heating process please proceed as follows Control the water level at the water level indication 6 and add water if necessarv Control that filling valve 5 and sterilization valve 9 are closed Adjust the sterilizing pressure on the adjusting device 21 as shown in table 9 5 so that the working pressure is approx 0 1 0 2 bar higher then the required pressure 2 4 bar Press ON OFF switch 13 to ON position control lamp lights green to start the jacket heating procedure Now the initiated heating process lasts approx 30 min when there is maximum water level and maximum required pressure approx 2 4 bar until the necessarv steam pressure is reached As long as the operating pressure is not reached heating is in operation This can be seen by the lighting of the yellow control lamp heating 16 The process of jacket preheating is finished when the pressure gauge for the steam jacket 20 indicates the pressure preselected on the a
6. the lid 3 is turned in clockwise direction until the limit stop seen from the top and then opened Removal of sterile goods Now the possibly still hot sterile goods can be removed Attention It is still possible that steam escapes from the sterilizing chamber D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 24 8 8 8 9 Preparations for further sterilizations When a sterilization has been ended by quick pressure reduction the sterilizer can quickly be prepared for the next charge as the steam pressure in the outer jacket 2 is maintained Control the water level at the water level indication 6 and add water if necessary Switching off of sterilizer For final switching off of the sterilizer e g after the last load of the day the ON OFF switch 13 has to be switched off light goes out and the inlet valve for cold water 29 has to be closed D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 25 9 1 9 2 9 3 9 4 Savetv Devices Water shortage in the outer chamber A sensor 11 is built in above the radiators so that thev cannot burn through in case of extreme shortage of destilled resp demineralized water The sensor consists of a small radiator with integrated temperature sensor In case of water shortage the temperature sensor heated up bv the small radiator gives a signal to the water shortage protected switch 11a Via a contactor 10a this switch durably separates the rad
7. is produced in the outer jacket The sterilizing process interrupted by the water shortage must be repeated as the goods have to be considered as not sterile D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 11 5 2 5 3 Preparations for further sterilizations When a sterilization has been ended by quick pressure reduction the sterilizer can quickly be prepared for the next charge as the steam pressure in the outer jacket 2 is maintained Controll the water level at the water level indication 6 and add water if necessary Switching off of sterilizer For final switching off of the sterilizer e g after the last load of the day the ON OFF switch 13 has to be switched off light goes out and the inlet valve for cold water 29 has to be closed D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 12 6 2 6 2 1 6 2 2 6 2 3 6 3 Programme A1 121 C jacket pressure 1 2 bar Adjustment The saturated steam necessary for sterilization is produced in the steam jacket 2 by means of two electric radiators 10 To start the heating process please proceed as follows Control the water level at the water level indication 6 and add water if necessary Control that filling valve 5 and sterilization valve 9 are closed Adjust the sterilizing pressure on the adjusting device 21 as shown in table 9 5 so that the working pressure jacket pressure is approx 0 1 0 2 bar higher than the requ
8. is the condition of a medical product which is free from active micro organisms EN 556 D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 5 2 1 2 2 2 3 2 4 2 5 2 6 2 7 2 8 Installation Before setting into operation please read the operating instructions thoroughiv Take care for correct current supplv see tvpe plate Take care for electrical connection which corresponds to the prescriptions valid on site Attention Install main switch on site Prepare cold water connection with a locallv installed hand valve for cooling of steam ds care that under the sterilizer there is a water outlet e g gully for the waste pipes of the sterilizer To guarantee the stability of the sterilizer the apparatus feeet 32 have to be fixed to the ground If the sterilizer is not used close the local hand valve for the cold water D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 6 3 Technical Data Typ Chamber Chamber Chamber Current Capacity Operating water Operating water Diameter Height Volume to NW NW to HW EE weav asa ssw pesnea II 50 70 2 137 dm 19 0 A 12 5 kW 42 21 ges B T B ii Operating pressure pe 2 5 bar Operating temperature 134 C Nominal voltage 230 400 V Current Alternating current Nominal frequency 50 60 cs D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 7 lt Feeding water For operation of the sterilizer either des
9. limit stop observation of sterilizer lid from the top At last close the function safety valve 23 with the operating lever 24 This process is executed properly when the lever 24 is situated in the intended groove of the link of the operating handle Attention Only by this procedure it is secured that the chamber lid and the function safety valve are closed orderly and pressure in the sterilizing chamber can be produced after opening the sterilization valve D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 13 6 4 6 4 1 6 4 2 6 4 3 6 4 4 6 4 5 6 4 6 Sterilization 121 C Before the sterilizing process controll the adjustment of the adjusting device for sterilizing pressure 21 Compare the pressure of the outer chamber at the pressure gauge 20 for the jacket 2 when the controll lamp 16 Heating has gone out with the adjusted value acc to table 9 5 for the goods to be sterilized and readjust the device 21 if necessary Controll the reducing valve for chamber pressure 25 for the sterilizing chamber 1 and close it if still open If the operating pressure necessary for sterilization is reached in the outer chamber to be seen at the pressure gauge 20 for the steam jacket 2 open the sterilization valve 9 Now the heating up time starts Steam is streaming from the outer chamber 2 into the sterilizing chamber 1 This can be observed at the pressure gauge 19 for the sterilizing
10. unit is switched off at the on off switch thus switching off the electrical heating After the self radiation the sterilizing valve 9 must be closed again By quick reduction of the pressure from the sterilizing chamber 1 First you close the sterilizing valve 9 and afterwards you open the release valve for chamber pressure 25 carefully In order that the escaping steam is cooled down before entering the drain open the cold water valve 29 so that a temperature of max 50 C is reached This procedure is continued until the pressure of the chamber 1 is 0 bar to be seen at the pressure gauge 19 for the sterilizing chamber The closing of the sterilizing valve 9 has the effect that the steam flow between outer chamber 2 and sterilizing chamber 1 is interrupted The separation of both chambers 1 and 2 allows that the sterile goods can be removed from the sterilizing chamber while the operating pressure in the outer chamber is maintained for further operation Open the sterilizer lid After pressure reduction indication 0 bar at the pressure gauge 19 for the sterilizing chamber the sterilizer lid 3 may be opened and the sterile goods be removed if the temperature has decreased to approx 60 C to be seen at the pointer thermometer For that purpose the function safety valve 23 is opened with the operation lever 24 operation lever 24 must be brought to the right side in its horizontal limit position The e
11. utensils may be sterilized D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 10 11 Maintenance of Sterilizer In case of external contamination of unit and inner chamber clean these parts moistly When the switchboard is cleaned separate it first from mains supply The inner chamber must always be kept clean If necessary wash it out let off the liquid via the emptying valve 25 and rinse If the water of the outer chamber becomes turbid to be seen on the level glass 6 the water can be let off via the emptying valve 31 In case of calcareous water the apparatus has to be decalcified once a year These intervals can be extended when the outer chamber is emptied regularly Sterilizer gasket In course of time the sterilizer gasket 33 is subject to a natural wear and tear and has to be exchanged if necessary e g if it is deformed or damaged or if during sterilization steam escapes at the gasket The gasket has to be greased in regular intervals with silicone fat or similar D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 28 12 Search for Errors Temperature indication does not work gt stream obstructed gt condensate trap defective No pressure in the outer chamber gt no water in the outer chamber gt no current fuse defective gt water shortage protection disengaged Chamber gasket leaky gt gasket defective gt dirt on the gasket gt gasket has to be grease
12. OPERATING INSTRUCTIONS for vertical sterilizer KSG 40 60 2 KSG 50 70 2 KSG 50 80 2 double walled execution without sterilization timer and contact thermometer only for for sterilization of unwrapped and not porous hospital utensils as instruments glas and plastic goods KSG Sterilisatoren GmbH Modifications reserved 16 02 06 BuchhoferstraBe 5 62XEXMXX XXX 82140 Olching Serial No Federal Republic of Germanv Phone 49 0 8142 2957 0 Wiring diagram 01 040602 001 Fax 49 0 8142 40384 Piping diagram 1417 2 E mail service ksg sterilisatoren de D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 1 1 1 2 5 1 5 2 5 3 6 1 6 2 6 3 6 4 6 5 6 6 6 7 6 8 6 9 7 1 7 2 7 3 7 4 7 5 7 6 7 7 7 8 7 9 7 10 Contents Description Intended use of the unit Intended aim Installation suse Technical Data od coed vests aa General Operating Instructions Control of feeding water Preparations for further sterilizations Switching off of sterilizer Programme A1 121 C jacket pressure 1 2 bar Adjustment Loading of the sterilizer Close sterilizing lid Sterilization 121 C End of programme Open the sterilizer lid Removal of sterile goods Preparations for further sterilizations Switching off of sterilizer
13. X 40 60 2 ED 62xexmxxxxx PDF 18 7 5 Sterilization The process during programme step Sterilization is the same as for programme step Heating up i e the operator maintains the chamber temperature while observing the pressure gauge 19 and the thermometer 34 bv opening and closing of the sterilizing valve 9 between 121 and 124 It is the task of the operating personal to secure bv a suitable timer that the sterile goods are subject to the sterilizing temperature for the complete time adjusted acc to table 9 5 For controlling the actual temperature the operating personal uses the thermometer 34 Measurement of the time can be started when the pointer of the thermometer 34 has reached a chamber temperature of min 121 C It must be interrupted when the temperature is below 121 C and can be started again after having reached this temperature Attention When the temperature has sunk below the adjusted value the sterilizing timer has to be stopped and only be continued after having reached the temperature again D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 19 7 7 7 8 End of programme After the end of the sterilizing time the sterilizing process must be ended The sterile goods can be removed after pressure reduction in the sterilizing chamber This pressure reduction can be reached by different methods acc to requirements see table 9 5 By self radiation of the complete sterilizer i e the
14. amber pressure 25 carefully by simultaneous opening of the inlet valve for cold water so that a rapid pressure reduction of the jacket 2 via the sterilizing chamber 1 is reached Open the sterilization valve 9 Thus steam streams out of the jacket 2 into the sterilizing chamber 1 Open carefully the reduction valve for chamber pressure 25 Afterwards open the cold water valve 29 correspondingly so that the escaping steam is cooled before entering the drain When the pressure of the outer chamber 2 is 0 bar the pointer of the pressure gauge 20 for the outer jacket 2 indicates 0 you open the function safety valve 23 with the operation levers 24 Afterwards close the filling valve 5 and fill in distilled and demineralized water in the filling funnel 4 until the water level has reached max marking After the end of the filling process close the valves 5 9 25 29 and the function safety valve 23 by putting the operating lever 24 from horizontal to vertical pos Afterwards the water shortage protection switch 11a has to be set into operation by pressing the reset button which is situated behind the protection cover at the front side of the water shortage protection switch 11a The red controll lamp water shortage 17 goes out After having actuated the reset button the heating elements are switched on again and the steam pressure adjusted at the regulator for sterilizing pressure 21
15. atus No XXXX Year of construction XXXX Operator Full mark of steam sterilizer KSG 40 60 2 Aukl 002 067 double walled electrically heated KSG 50 70 2 Aukl 002 069 double walled electrically heated KSG 50 80 2 Aukl 002 077 double walled electrically heated D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 08142 29 57 0 08142 40 38 4 service ksg sterilisatoren de 1 1 1 2 Intended use of the unit This execution of sterilizer can only be used for sterilization of unwrapped non porous solid goods like instruments glass rubber and plastic at a sterilization temperature of 134 C or 121 C As the sterilizer works according to the principle of gravitation no porous goods i e textiles no hollow bodies i e tubes pipes and no wrapped goods i e in shrink or paper foil in fabrics or cassettes can be sterilized Hollow receptacles may preferably be sterilized opening downward Solutions independent from the size of the receptacles may not be sterilized in this sterilizer In no case inflammable materials may be sterilized No living beings may be sterilized The device is not been suitable for the sterilisation destruction by the lab goods which were submitted to a chemical and or biological treatment Such treated goods may not be sterilized Intended aim The aim of sterilization is the achievement of the sterility of a sterilizer load charge acc to the sterilizing process Hint Sterile
16. chamber Simultaneously the temperature for the sterilizing chamber is also increasing to be seen at the pointer thermometer 34 for the sterilizing chamber Remark a When steam flows from steam jacket 2 into sterilizing chamber 1 the pressure in the outer chamber decreases to be seen at the pressure gauge 20 for the outer chamber Via the adjusting device for sterilizing pressure 21 the radiators are switched on automatically to be seen on the yellow control lamp heating 16 b The air in the sterilizing chamber is displaced by the entering steam and is led off via the automatic aerator and deaerator 26 At a temperature of 100 C and higher this component is closed thus rendering possible that the steam pressure of the sterilizing chamber is reduced Attention The sterilizer lid 3 is hot during operation 1 When the indication of both pressure gauges 19 and 20 is identical i e the pressure in the sterilizing chamber corresponds to the pressure adjusted acc to table 9 5 the same refers to the temperature the sterilizing time can be measured It is the task of the operating personal to secure by a suitable timer that the goods to be sterilized are subject to the sterilizing temperture for the complete time adjusted acc to table 9 5 For controlling the actual temperature use the pointer thermometer 34 When the temperature is lower than the adjusted temperature the measurement of the sterilizing ti
17. d D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 29 13 Short Operating Instruktions for vertical autoclave KSG 40 60 2 50 70 2 50 80 2 without sterilization timer Filling of the apparatus with distilled water until at last middle of water level indication Preheat steam jacket up to the adjusted pressure Fill unit with goods to be sterilized and close the lid Open sterilization valve 9 and wait until the sterilizing time has run down Close the sterilization valve 9 Open the reduction valve for chamber pressure 25 Open the cold water inlet valve 29 Close the cold water inlet valve 29 when the chamber 1 is pressureless Open function safetv valve 23 turn the lid into opposite direction and lift it off if necessarv close reducing valve for chamber pressure 25 D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 30 14 Legend 1 Sterilizing chamber 2 Outer chamber steam jacket 3 Sterilizer lid with swivel bow Operation handles for 3 4 Filling funnel for distilled demineralized water 5 Filling valve 5a Flap trap for 5 6 Water level with protective covering 7 Water level valve top 8 Water level valve bottom with drain valve 9 Sterilization valve 10 Plunger type radiators 2x 11 Water shortage sensor with micro radiator 11a Water shortage protected switch resettable 12 Switch box electric control case 13 ON OFF switch main switch 16 Control lamp heating
18. djusting device for sterilizing pressure and the yellow control lamp heating 16 goes out for the first time Now the jacket is ready for sterilization Loading of the sterilizer Please classify the sterilizing goods acc to sterilizing temperature and time see table 9 5 If necessary scheduled put the sterilizing goods into the provided baskets accessories Place the sterilizing goods into the sterilizing chamber 1 and put them on the ground plate resp on the trays accessories The chamber can be filled up until its upper edge Remark When selecting the goods to be sterilized and the arrangement in the chamber please take into consideration that during the process described above the radiation temperature approx 134 C of the outer jacket 2 is much higher than the real sterilizing temperature approx 124 C in the sterilizing chamber 1 Take care that the sterile goods doe not contact the walls of the chamber D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 17 7 3 7 4 7 4 1 7 4 2 7 4 3 7 4 4 7 4 5 Close sterilizing lid To that end turn the lid by the operating handles 3a counter clockwise until the limit stop and fix it in this axial position Then tilt the lid downwards until it lies on the gasket 33 Afterwards please turn the lid by means of the operating handles 3a in clockwise direction until the limit stop observation of sterilizer lid from the top At last close the functio
19. e sterilizing chamber is reduced Attention The sterilizer lid 3 is hot during operation When the indication of both pressure gauges 19 and 20 is identical i e the pressure in the sterilizing chamber corresponds to the pressure adjusted acc to table 9 5 the same refers to the temperature the sterilizing time can be measured It is the task of the operating personal to secure by a suitable timer that the goods to be sterilized are subject to the sterilizing temperture for the complete time adjusted acc to table 9 5 For controlling the actual temperature use the pointer thermometer 34 When the temperature is lower than the adjusted temperature the measurement of the sterilizing time has to be stopped and can only be continued when the sterilizing temperature is reached again D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 23 8 5 8 5 1 8 5 2 8 7 End of programme After the end of the sterilizing time the sterilizing process must be ended The sterile goods can be removed after pressure reduction in the sterilizing chamber This pressure reduction can be reached by different methods acc to requirements see table 9 5 By self radiation of the complete sterilizer i e the unit is switched off at the on off switch thus switching off the electrical heating After the self radiation the sterilizing valve 9 must be closed again By quick reduction of the pressure from the sterilizing chamber 1
20. iators 10 and the small radiator in the water shortage sensor 11 from electric mains At the same time the red control lamp water shortage 17 lights up When the water shortage has been removed by refilling of water the reset button has to be pressed at the water shortage protection switch Thus the red lamp 17 water shortage extinguishes Sterilization can be started again Function safety valve When the sterilizer lid is orderly closed this valve 23 prevents by its operation lever 24 that the lid can be opened when the sterilizing chamber is under pressure If anyone tries to open the chamber nevertheless the lever 24 has to be operated first This would mean that via the activated valve 23 possibly existing pressure would be reduced and be led away via the exhaust pipe Sterilizer lid The sterilizer lid is opened and closed by a tongue and groove system Thus an opening of the lid 3 is not possible when there is pressure inside Should the lid be opened due to an accumulation of misfunctions depending on the apparatus or on wrong operation when there is still a remaining pressure inside the chamber this inner pressure presses the lid into the mechanical locking position before it can be opened completely Due to the arrangement of the tongue and groove system the remaining pressure between chamber and lid can escape but a dangerous flying open of the lid is surely prevented Overpressure safety device
21. intained for further operation Open the sterilizer lid After pressure reduction indication 0 bar at the pressure gauge 19 for the sterilizing chamber the sterilizer lid 3 may be opened and the sterile goods be removed if the temperature has decreased to approx 60 C to be seen at the pointer thermometer For that purpose the function safety valve 23 is opened with the operation lever 24 operation lever 24 must be brought to the right side in its horizontal limit position The eventually remaining overpressure now escapes via the exhaust tube of the function safety valve 23 Attention It is still possible that steam escapes at the exhaust tube of this valve By means of the operation handles 3a the lid 3 is turned in clockwise direction until the limit stop seen from the top and then opened Removal of sterile goods Now the possibly still hot sterile goods can be removed Attention It is still possible that steam escapes from the sterilizing chamber D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 15 6 8 6 9 Preparations for further sterilizations When a sterilization has been ended by quick pressure reduction the sterilizer can quickly be prepared for the next charge as the steam pressure in the outer jacket 2 is maintained Controll the water level at the water level indication 6 and add water if necessary Switching off of sterilizer For final switching off of the sterilizer
22. ired pressure 1 2 bar Press ON OFF switch 13 to ON position control lamp lights green to start the jacket heating procedure Now the initiated heating process lasts approx 20 min when there is maxium water level and maximum required pressure approx 1 2 bar until the necessary steam pressure is reached As long as the operating pressure is not reached heating is in operation This can be seen by the lighting of the yellow control lamp heating 16 The process of jacket preheating is finished when the pressure gauge for the steam jacket 20 indicates the pressure preselected on the adjusting device for sterilizing pressure and the yellow control lamp heating 16 goes out for the first time Now the jacket is ready for sterilization Loading of the sterilizer Please classify the sterilizing goods acc to sterilizing temperature and time see table 9 5 If necessary scheduled put the sterilizing goods into the provided baskets accessories Place the sterilizing goods into the sterilizing chamber 1 and put them on the ground plate resp on the trays accessories The chamber can be filled up until its upper edge Close sterilizing lid To that end turn the lid by the operating handles 3a counter clockwise until the limit stop and fix it in this axial position Then tilt the lid downwards until it lies on the gasket 33 Afterwards please turn the lid by means of the operating handles 3a in clockwise direction until the
23. lization Loading of the sterilizer Please classify the sterilizing goods acc to sterilizing temperature and time see table 9 5 If necessary scheduled put the sterilizing goods into the provided baskets accessories Place the sterilizing goods into the sterilizing chamber 1 and put them on the ground plate resp on the trays accessories The chamber can be filled up until its upper edge Close sterilizing lid To that end turn the lid by the operating handles 3a counter clockwise until the limit stop and fix it in this axial position Then tilt the lid downwards until it lies on the gasket 33 Afterwards please turn the lid by means of the operating handles 3a in clockwise direction until the limit stop observation of sterilizer lid from the top At last close the function safety valve 23 with the operating lever 24 This process is executed properly when the lever 24 is situated in the intended groove of the link of the operating handle Attention Only by this procedure it is secured that the chamber lid and the function safety valve are closed orderly and pressure in the sterilizing chamber can be produced after opening the sterilization valve D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 22 8 4 8 4 1 8 4 2 8 4 3 8 4 4 8 4 5 8 4 6 Sterilization 134 C Before the sterilizing process control the adjustment of the adjusting device for sterilizing pressure 21 Compare the press
24. me has to be stopped and can only be continued when the sterilizing temperature is reached again D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 14 6 7 End of programme After the end of the sterilizing time the sterilizing process must be ended The sterile goods can be removed after pressure reduction in the sterilizing chamber This pressure reduction can be reached by different methods acc to requirements see table 9 5 By self radiation of the complete sterilizer i e the unit is switched off at the on off switch thus switching off the electrical heating After the self radiation the sterilizing valve 9 must be closed again By quick reduction of the pressure from the sterilizing chamber 1 First you close the sterilizing valve 9 and afterwards you open the release valve for chamber pressure 25 carefully In order that the escaping steam is cooled down before entering the drain open the cold water valve 29 so that a temperature of max 50 C is reached This procedure is continued until the pressure of the chamber 1 is 0 bar to be seen at the pressure gauge 19 for the sterillizing chamber The closing of the sterilizing valve 9 has the effect that the steam flow between outer chamber 2 and sterilizing chamber 1 is interrupted The separation of both chambers 1 and 2 allows that the sterile goods can be removed from the sterilizing chamber while the operating pressure in the outer chamber is ma
25. n safety valve 23 with the operating lever 24 This process is executed properly when the lever 24 is situated in the intended groove of the link of the operating handle Attention Only by this procedure it is secured that the chamber lid and the function safety valve are closed orderly and pressure in the sterilizing chamber can be produced after opening the sterilization valve Heating up time At this sterilizing method you can start the sterilizing process when the pressure in the outer chamber 2 is higher than the pressure necessary for sterilization To initiate the heating up time open the sterilizing valve 9 carefully Now the steam is streaming from the outer jacket into the sterilizing chamber Observe the pointer of the pressure gauge 19 for the sterilizing chamber 1 When it has reached approx 1 2 1 3 bar close the sterilizing valve Open it again when the pointer has dropped down to approx 1 1 bar to close it again when the pointer has reached approx 1 2 1 3 bar From the movement of this pointer in the thermometer 34 for the sterilizing chamber 1 you can see the temperature increase in the sterilizing chamber By the opening and closing of the sterilizing valve 9 as described above a constant chamber temperature of approx 121 C 124 C should be reached The heating up time is ended when the pointer of the thermometer 34 has reached the sterilizing temperature D e VERT2 62XEXMXX XX
26. ter chamber 2 is 0 bar pointer of pressure gauge 20 for the outer jacket must indicate 0 you open the function safety valve 23 with the operation levers 24 Afterwards close the filling valve 5 and fill in distilled and demineralized water in the filling funnel 4 until the water level has reached max marking After the end of the filling process close the valves 5 9 25 and 29 When you have reset the unit into operation by actuation of the on off switch 13 the adjusted steam pressure is produced in the outer jacket D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 10 Filling in of jacket water when the min marking has been passed the outer chamber 2 is heated up and under pressure and the water shortage protection switch 11a has reacted The filling of jacket water can only be executed when the outer chamber 2 is without pressure If the red controll lamp water shortage is illuminated at the electric switch board 12 the water level in the steam jacket has fallen below the admissible value during operation Before filling the water into the outer chamber it must be put into a pressureless condition The outer chamber becomes pressureless by self radiation e g by switching off the sterilizer at the on off switch 13 so that the heating elements 10 are switched off If the sterilizer lid is orderly closed you can additionally open the sterilization valve 9 and the release valve for ch
27. terile goods doe not contact the walls of the chamber D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 21 8 2 8 2 1 8 2 2 8 2 3 8 3 Programme B 134 C jacket pressure 2 4 bar Adjustment The saturated steam necessarv for sterilization is produced in the steam jacket 2 bv means of two electric radiators 10 To start the heating process please proceed as follows Control the water level at the water level indication 6 and add water if necessarv Control that filling valve 5 and sterilization valve 9 are closed Adjust the sterilizing pressure on the adjusting device 21 as shown in table 9 5 so that the working pressure is approx 0 1 0 2 bar higher then the required pressure 2 4 bar Press ON OFF switch 13 to ON position control lamp lights green to start the jacket heating procedure Now the initiated heating process lasts approx 30 min when there is maximum water level and maximum required pressure approx 2 4 bar until the necessarv steam pressure is reached As long as the operating pressure is not reached heating is in operation This can be seen by the lighting of the yellow control lamp heating 16 The process of jacket preheating is finished when the pressure gauge for the steam jacket 20 indicates the pressure preselected on the adjusting device for sterilizing pressure and the yellow control lamp heating 16 goes out for the first time Now the jacket is ready for steri
28. tilled demineralized or completelv desalted water is necessarv Indications concerning the filling quantitv of operating water for steam jacket in case of emptv chamber until min water level NW up to max level HW can be taken from paragraph 2 Technical data Two radiators 10 Water shortage sensor 11 with micro heating element 0 5 kW and water shortage protected switch 11a resettable Regulator and adjustor for steam jacket pressure 21 max 2 5 bar D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 8 5 1 Operating Instructions For setting the sterilizer into operation please proceed as follows Control of feeding water Level of feeding water destilled or demineralized water at the water level glass 6 has to be between the markings for minimum and maximum water level Should this not be the case the outer chamber 2 has to be filled with a corresponding quantity of water This can only be done when the outer chamber 2 is without pressure resp cold i e the pressure gauge 20 for the steam jacket 2 indicates 0 bar Filling in of jacket water for first filling or after having emptied the jacket completely or in case of cold pressureless outer chamber Open the sterilization valve 9 Open the function safety valve 23 by means of the operating lever 24 if the sterilizer lid 3 is closed Open the filling valve 5 Fill in destilled or demineralized water into the filling funnel 4
29. until water level 6 has reached the maximum marking Close the filling valve 5 Close the sterilization valve 9 D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 9 5 1 2 Filling in of jacket water when the minimum marking is reached and the jacket is heated up and under pressure between two sterilizing processes The filling of jacket water can only be executed when the outer chamber 2 is without pressure If you observe that the water level has reached the minimum marking the outer chamber 2 must be put into pressureless condition before filling jacket water into the jacket The outer chamber 2 becomes pressureless by self radiation e g by switching off the sterilizer at the on off switch 13 so that the heating elements 10 are switched off If the sterilizer lid 3 is orderly closed you can additionally open the sterilization valve 9 and the release valve for chamber pressure 25 carefully by simultaneous opening of the inlet valve for the cold water 29 so that a rapid pressure reduction of the jacket 2 via the sterilizing chamber 1 is reached Open the sterilization valve 9 Thus steam streams out of the jacket 2 into the sterilizing chamber 1 Open carefully the reduction valve for chamber pressure 25 Now steam escapes to the drain Afterwards open the cold water valve 29 correspondingly so that the escaping steam is cooled before entering the drain When the pressure of the ou
30. ure of the outer chamber at the pressure gauge 20 for the jacket 2 when the control lamp 16 Heating has gone out with the adjusted value acc to table 9 5 for the goods to be sterilized and readjust the device 21 if necessarv Control the reducing valve for chamber pressure 25 for the sterilizing chamber 1 and close it if still open If the operating pressure necessarv for sterilization is reached in the outer chamber to be seen at the pressure gauge 20 for the steam jacket 2 open the sterilization valve 9 Now the heating up time starts Steam is streaming from the outer chamber 2 into the sterilizing chamber 1 This can be observed at the pressure gauge 19 for the sterilizing chamber Simultaneously the temperature for the sterilizing chamber is also increasing to be seen at the pointer thermometer 34 for the sterilizing chamber Remark a When steam flows from steam jacket 2 into sterilizing chamber 1 the pressure in the outer chamber decreases to be seen at the pressure gauge 20 for the outer chamber Via the adjusting device for sterilizing pressure 21 the radiators are switched on automatically to be seen on the yellow control lamp heating 16 b The air in the sterilizing chamber is displaced by the entering steam and is led off via the automatic aerator and deaerator 26 At a temperature of 100 C and higher this component is closed thus rendering possible that the steam pressure of th
31. ventually remaining overpressure now escapes via the exhaust tube of the function safety valve 23 Attention It is still possible that steam escapes at the exhaust tube of this valve By means of the operation handles 3a the lid 3 is turned in clockwise direction until the limit stop seen from the top and then opened Removal of sterile goods Now the possibly still hot sterile goods can be removed Attention It is still possible that steam escapes from the sterilizing chamber D e VERT2 62XEXMXX XXX 40 60 2 ED 62xexmxxxxx PDF 20 7 9 7 10 Preparations for further sterilizations When a sterilization has been ended by quick pressure reduction the sterilizer can quickly be prepared for the next charge as the steam pressure in the outer jacket 2 is maintained Control the water level at the water level indication 6 and add water if necessary Switching off of sterilizer For final switching off of the sterilizer e g after the last load of the day the ON OFF switch 13 has to be switched off light goes out and the inlet valve for cold water 29 has to be closed Remark When selecting the goods to be sterilized and the arrangement in the chamber please take into consideration that during the process described above the radiation temperature approx 134 C of the outer jacket 2 is much higher than the real sterilizing temperature approx 124 C in the sterilizing chamber 1 Take care that the s