Home
Thermo Scientific KingFisher Viral NA Kit
Contents
1. Thermo Scientific KingFisher Viral NA Kit Instruction Manual Rev 1 2 Cat no N11998 Copyright 2012 Thermo Fisher Scientific Inc First edition published in 2010 All rights reserved Reproduction of the accompanying user documentation in whole or in part is prohibited BindIt KingFisher Microtiter Multidrop NanoDrop and Versette are registered trademarks of Thermo Fisher Scientific Virkon is a registered trademark of E I du Pont de Nemours and Company or its affiliates All other trademarks and registered trademarks are the property of their respective holders Thermo Fisher Scientific reserves the right to change its products and services at any time to incorporate technological developments This manual is subject to change without prior notice as part of continuous product development Although this manual has been prepared with every precaution to ensure accuracy Thermo Fisher Scientific assumes no liability for any errors or omissions nor for any damages resulting from the application or use of this information This instruction manual supersedes all previous editions The Product will operate substantially in conformance with Thermo Fisher Scientific s published specifications THERMO FISHER SCIENTIFIC DISCLAIMS ALL OTHER WARRANTIES WHETHER EXPRESSED OR IMPLIED ORAL OR WRITTEN WITH RESPECT TO THE PRODUCTS INCLUDING WITHOUT LIMITATION ALL IMPLIED WARRANTIES OF PRODUCT
2. Cat No Product 5400000 KingFisher magnetic particle processor 5400050 KingFisher mL magnetic particle processor 5400100 KingFisher Duo magnetic particle processor 5400630 KingFisher Flex magnetic particle processor with 96 deep well head 5400640 KingFisher Flex magnetic particle processor with 24 deep well head Discontinued KingFisher 96 magnetic particle processor Table 1 3 Thermo Scientific KingFisher Flex consumables Cat No Product Package size 97002514 KingFisher Flex 96 tip comb for PCR magnet 80 pcs 97002524 KingFisher Flex 96 tip comb for KF magnet 100 pcs 97002534 KingFisher Flex 96 tip comb for deep well magnet 100 pcs 97002610 KingFisher Flex 24 deep well tip comb and plate 50 pcs 97002540 KingFisher Flex 96 KF plate 200 ul 48 pcs 95040450 Microtiter deep well 96 plate non sterile 50 pcs 95040460 Microtiter deep well 96 plate sterile 50 pcs 95040470 KingFisher Flex 24 deep well plate 50 pcs 95040480 KingFisher Flex 24 deep well plate sterile 50 pcs 8 Thermo Scientific KingFisher Viral NA Kit Instruction Manual Thermo Fisher Scientific Table 1 4 Thermo Scientific KingFisher Duo consumables Kit Content Cat No Product Package size 97003500 KingFisher Duo 12 tip comb for Microtiter deep well 96 plate 50 pcs 97003510 KingFisher Duo 6 tip combs and KingFisher 24 deep well plate 48 pcs 12 pcs of 24 deep well plates each including 4 tips combs 97003520 KingFisher Duo elu
3. 50 Safety instructions Harmful by inhalation in contact with skin and if swallowed Contact with acids liberates very toxic gas Harmful to aquatic organisms may cause long term adverse effects in the aquatic environment Keep away from food drink and animal feed Binding Buffer Sodium perchlorate lt 30 and ethanol lt 55 Flammable Harmful if swallowed Wash Buffer 1 Sodium perchlorate lt 18 and ethanol lt 24 Flammable Wash Buffer 2 Ethanol lt 80 Highly flammable Keep container tightly closed Keep away from sources of ignition Carrier RNA Buffer Guanidine thiocyanate lt 50 Harmful by inhalation in contact with skin and if swallowed Contact with acids liberates very toxic gas Harmful to aquatic organisms may cause long term adverse effects in the aquatic environment Keep away from food drink and animal feed Proteinase K Thermo Fisher Scientific Lyophilized Proteinase K Irritating to eyes respiratory system and skin May cause sensitization by inhalation Do not inhale dust Avoid contact with skin In case of contact with eyes rinse immediately with plenty of water and seek medical advice Wear suitable protective clothing and gloves Thermo Scientific KingFisher Viral NA Kit Instruction Manual 17 Safety Information 18 Thermo Scientific KingFisher Viral NA Kit Instruction Manual Thermo Fisher Scientific Storage conditions Preparation of the
4. QUALITY CONDITION DESCRIPTION MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE THERMO FISHER SCIENTIFIC DOES NOT WARRANT THAT THE PRODUCTS ARE ERROR FREE OR WILL ACCOMPLISH ANY PARTICULAR RESULT THERMO FISHER SCIENTIFIC HEREBY EXPRESSLY DISCLAIMS ANY WARRANTY REGARDING RESULTS OBTAINED THROUGH THE USE OF THE PRODUCTS INCLUDING WITHOUT LIMITATION ANY CLAIM OF INACCURATE INVALID OR INCOMPLETE RESULTS Thermo Fisher Scientific and its affiliates shall have no liability to an End User arising out of the use or inability to use the product including without limitation for any loss of use or profits business interruption or any consequential incidental special or other indirect damages of any kind regardless of how caused and regardless of whether an action in contract tort strict product liability or otherwise Contents Chapter 1 Chapter 2 Chapter 3 Chapter 4 Chapter 5 Thermo Fisher Scientific Kit COMPO naa cconcscoccsecctstcessssescseectcsnsetinnenecnenstnebsnusincbsneetaneesnsdeerbanstes 7 Product Description s scsscsscsessecseeseeseseesesseesnesteeneaeentens 11 latrod ction misa ei ee t deneubdoredosen tioned test 11 Intended sei miN a E A EA ERS 11 Principle and procedure s ssssssesssssssssssssesresresresresressessesresses 11 Kit sp cifications sonrie aana a aE EEEE 12 KingFisher magnetic particle processors sssssssssssssrssssesreessees 13 Safety Information sssss
5. Available elution volumes of Thermo Scientific KingFisher instruments KingFisher instrument Elution volume KingFisher 20 200 pl KingFisher mL 50 1000 ul KingFisher Duo with 12 pin magnet head 30 130 pl KingFisher Duo with 6 pin magnet head 200 5000 pl KingFisher Flex with 96 deep well head 50 150 pl elution in a KingFisher Flex 96 KF plate KingFisher Flex with 96 deep well head 50 1000 ul elution in a Microtiter deep well 96 plate KingFisher Flex with 96 well head 20 250 pl elution in a KingFisher Flex 96 KF plate KingFisher Flex with 24 deep well head 200 5000 pl To gain a maximal yield of purified nucleic acids avoid the lowest permitted volumes of Elution Buffer in the KingFisher instruments The Elution Buffer should cover the KingFisher Magnetic Beads completely and 36 Thermo Scientific KingFisher Viral NA Kit Instruction Manual Thermo Fisher Scientific General Information Decontamination and disinfection of sample material any possible magnetic bead pellet should be completely resuspended In addition the volume of the Elution Buffer should be adequate for efficient mixing of the beads in order to obtain a maximal release of the purified nucleic acids from the beads Elution can be conducted at room temperature However it is strongly recommended that both the lysis and elution steps are carried out at 56 C in the KingFisher Flex and KingFisher Duo in accordance with the BindIt Software 3 2 protocol
6. Proteinase K working solution Preparation of the Carrier RNA working solution Thermo Fisher Scientific Chapter 4 Storage Conditions and Preparation of the Working Solutions All buffers and reagents included in the KingFisher Viral NA Kit can be stored at room temperature 20 25 C and are stable for up to one year from the manufacturing date The buffers are ready for use To prepare the Proteinase K working solution for the KingFisher Viral NA Kit add 1 ml of Proteinase K Buffer to the vial of the lyophilized Proteinase K The Proteinase K working solution should be stored at 20 C in aliquots Repeated freezing and thawing should be avoided To prepare the Carrier RNA working solution add 500 pl of Carrier RNA Buffer to the vial of the Carrier RNA The Carrier RNA working solution should be stored at 20 C in aliquots Repeated freezing and thawing should be avoided Thermo Scientific KingFisher Viral NA Kit Instruction Manual 19 Storage Conditions and Preparation of the Working Solutions Preparation of the Carrier RNA working solution 20 Thermo Scientific KingFisher Viral NA Kit Instruction Manual Thermo Fisher Scientific Chapter 5 Protocols and Pipetting Instructions Before beginning the nucleic acid purification protocol carefully read through the Thermo Scientific KingFisher Flex User Manual Cat No N07669 the Thermo Scientific KingFisher Duo User Manual Cat No N12420 or the Thermo Scientifi
7. When using the KingFisher instrument without a heating option the Lysis and Elution Buffers can be prewarmed to 55 C and dispensed into the Lysis or Elution tubes The Elution Buffer should be dispensed into the Elution tubes during an additional pause in the protocol before the elution step If some KingFisher Magnetic Beads remain in the Elution Buffer centrifuge the Elution plate briefly or place the Elution plate or the tube strip on a magnet for a few minutes to collect the residual beads at the bottom of the well or tube and transfer the supernatant to a new tube Decontamination he sample material and the reagents and plastics that and disinfection of have been in contact with the sample material should l terial be decontaminated in order to minimize the risk of samp e materia contamination For this purpose a decontaminant such as Virkon should be used in accordance with the manufacturer s instructions Accordingly the appropriate treatment and or disposal of waste should be taken care of Thermo Fisher Scientific Thermo Scientific KingFisher Viral NA Kit Instruction Manual 37 General Information Decontamination and disinfection of sample material 38 Thermo Scientific KingFisher Viral NA Kit Instruction Manual Thermo Fisher Scientific Table A 1 Troubleshooting guide Problem Low DNA yield Appendix A Troubleshooting Possible cause and actions There should be an adequate volume of the Elution Buffer to cover
8. acid purificati on When using the KingFisher Viral NA Kit for the first f 200 ul of time prepare the working solutions for Proteinase K and rom pi 0 Carrier RNA For more instructions refer to Chapter 4 cell free body fluid Storage Conditions and Preparation of the Working Solutions 1 Take four empty Microtiter deep well 96 plates and two empty Thermo Scientific KingFisher Flex 96 KF plates 2 Prepare four Microtiter deep well 96 plates and one KingFisher Flex 96 KF plate as indicated in the table below Plate Plate Plate Content Sample reagent number type name volume per well 1 Microtiter Sample Sample 200 pl deep well 96 plate Lysis Buffer 200 pl Proteinase K 10 ul working solution Carrier RNA working solution 44 2 Microtiter Wash1 Wash Buffer 1 500 pl deep well 96 plate 3 Microtiter Wash2 Wash Buffer 2 500 ul deep well 96 plate 4 Microtiter Wash3 Wash Buffer 3 550 ul deep well 96 plate 5 KingFisher Elution Elution Buffer 100 pl Flex 96 KF plate 22 Thermo Scientific KingFisher Viral NA Kit Instruction Manual Thermo Fisher Scientific Protocols and Pipetting Instructions Instructions for KingFisher Flex with 96 deep well plates for nucleic acid purification from 200 ul of cell free body fluid 3 Place a Thermo Scientific KingFisher Flex 96 tip comb for deep well magnets on a Tip Plate that is an empty KingFisher Flex 96 KF plate 4 Start the KF_ViralNA_Flex96 protocol with the KingFisher Fle
9. of a chaotropic salt The following effective wash steps dispose of proteins cell debris and any residual contaminants while the nucleic acids bound to the KingFisher Magnetic Beads are transferred through sequential wash steps Three different Wash Buffers are used followed by an air drying step which considerably improves the purity of the nucleic acids High quality nucleic acids are eluted into the Elution Buffer and are ready for subsequent downstream processes such as enzymatic reactions Heat incubation is included in the lysis and elution steps of the purification protocol when using the KingFisher Flex or KingFisher Duo The KingFisher Viral NA Kit is designed for rapid automated preparation of highly pure viral nucleic acids from cell free body fluids using Thermo Scientific KingFisher magnetic particle processors When excluding a dispense step requiring the addition of Binding Buffer and KingFisher Magnetic Beads the approximate processing time is 40 minutes for the purification of 96 samples in the KingFisher Flex and 12 samples in the KingFisher Duo and 45 minutes for the purification of 15 samples in the KingFisher mL The obtained nucleic acids can be used directly in various downstream applications Suitable sample materials for purification are cell free body fluids such as serum and plasma The procedure is optimized for the sample volume of 200 pl The sample type as well as the handling and storage of the sample af
10. protocol can be found in Chapter 4 Using the software in the BindIt Software for KingFisher Instruments version 3 2 User Manual 5 Add the KingFisher Magnetic Beads and Binding Buffer to the Sample tubes during the dispense step When the KingFisher mL pauses at the dispense step after the lysis step approximately after 20 minutes remove the tube strip tray from the instrument and add the KingFisher Magnetic Beads and Binding Buffer separately to the Sample tubes as indicated below Resuspend the KingFisher Magnetic Beads well e g by vortexing before removing them from the bottle Tube Tube name Content Reagent volume A Sample KingFisher 30 ul Magnetic Beads Binding Buffer 550 ul 6 Place the tube strip tray back into the instrument and press Start After the pause the protocol will continue to the end 7 After the run is completed remove the tube strips and store the purified nucleic acids Thermo Scientific KingFisher Viral NA Kit Instruction Manual 31 Protocols and Pipetting Instructions Instructions for KingFisher mL for nucleic acid purification from 200 ul of cell free body fluid When the protocol is completed remove the tubes and turn off the instrument Store the purified nucleic acids accordingly The purified nucleic acids are ready for use in downstream applications To increase the nucleic acid yield the Lysis Buffer can be prewarmed to 55 C and the Sample tubes can be prepared just before t
11. the KingFisher Magnetic Beads completely during the elution step Do not let the KingFisher Magnetic Beads dry as this might result in lower elution efficiency Efficient lysis of the sample material at 56 C or preheating the Lysis Buffer to 55 C increases the nucleic acid yield Heating during the elution step or using a preheated Elution Buffer enhances the release of the nucleic acids from the KingFisher Magnetic Beads and the yield of the purified nucleic acids Prolonged storage of the sample material may reduce the nucleic acid yield Use only Thermo Scientific KingFisher plates or tubes with the KingFisher instruments Use of products from other manufacturers may cause unsuitable mixing and affect the yield of purified nucleic acids Low purity Prolonged storage of the sample material may reduce the quality of the nucleic acids Insufficient washing causes impurities in the Elution Buffer Wash Buffers 1 and 2 contain ethanol Carryover of the buffers may cause unsatisfactory performance in downstream applications Carryover of the KingFisher Magnetic Beads to the Elution Buffer may affect the A A ratio Make sure that the KingFisher Magnetic Beads do not affect the measurement by centrifuging the samples or placing them on a magnet for a few minutes to collect the residual beads at the bottom of the well Carryover of the KingFisher Magnetic Beads does not affect most downstream processes Continued T
12. 0 ul of cell free body fluid Plate Plate Plate Content Reagent volume number type name per well 1 Microtiter Sample KingFisher 30 pl deep well Magnetic Beads 96 plate Binding Buffer 550 pl 6 Place the Sample plate back into the instrument and press Start After the pause the protocol will continue to the end 7 After the run is completed remove the plates and store the purified nucleic acids When the protocol is completed remove the plates according to the instructions on the KingFisher Flex display and turn off the instrument Store the purified nucleic acids accordingly The purified nucleic acids are ready for use in downstream applications The final nucleic acid concentration in the Elution Buffer may increase if the purified nucleic acids are eluted into a smaller volume of the buffer than is recommended but this can slightly reduce the overall nucleic acid yield 24 Thermo Scientific KingFisher Viral NA Kit Instruction Manual Thermo Fisher Scientific Protocols and Pipetting Instructions Instructions for KingFisher Flex with 96 deep well plates for nucleic acid purification from 200 ul of cell free body fluid Summary of Table 5 1 Summary of plate contents plate contents Thermo Fisher Scientific Plate Plate Plate Content Sample reagent number type name volume per well 1 Microtiter Sample Sample 200 ul deep well 96 plate Lysis Buffer 200 ul Proteinase K 10 ul working solution Carrier RNA 4ul worki
13. Scientific Versette Thermo Scientific BindIt Software 3 2 can be used for running ready made and optimized protocols for the Thermo Scientific KingFisher Kits It is also possible to transfer the defined protocol onto the onboard software and run it directly from the instrument The KingFisher instruments offer a rapid and automated solution for complicated and time consuming purification processes without risk of carryover or cross contamination resulting in high purity nucleic acids The KingFisher instrument family comprises four systems covering working volumes from 20 to 5000 ul Each system consists of an instrument specially designed plastic consumables and the easy to use BindIt Software 3 2 The KingFisher Viral NA Kit is optimized and ready for use with KingFisher instruments The KingFisher magnetic particle processors are intended for professional research use by trained personnel Detailed information and user instructions for the KingFisher instruments can be found in their respective user manuals Thermo Scientific KingFisher Viral NA Kit Instruction Manual 13 Product Description KingFisher magnetic particle processors Table 2 1 Overview of Thermo Scientific KingFisher systems KingFisher Flex KingFisher Duo 96 format 24 format 12 format 6 format Processing volume 20 1000 pl 200 5000 pl 30 1000 pl 200 5000 pl Capacity Up to 96 samples Up to 24 samples Up to 12 samples Up to 6 samples per run sample per run
14. ample volume Up to 24 samples per run 200 ul with the KingFisher Viral NA Kit Magnetic head 15 format 24 format Plates KingFisher mL tube special tube strip with KingFisher plate 100 or 200 pl 20 100 ul 1x 5 tubes 50 1000 pl or 20 200 ul Tip combs KingFisher mL tip comb 1 x 5 format KingFisher tip comb 1 x 12 format Heating No heating available No heating available temperature Thermo Fisher Scientific The BindIt Software 3 2 protocols optimized for the KingFisher Viral NA Kit are available for the KingFisher Flex the KingFisher Duo and the KingFisher mL instruments BindIt Software 3 2 protocols for the Thermo Scientific KingFisher and the Thermo Scientific KingFisher 96 are available on request For more information contact your local authorized distributor Thermo Scientific KingFisher Viral NA Kit Instruction Manual 15 Product Description KingFisher magnetic particle processors 16 Thermo Scientific KingFisher Viral NA Kit Instruction Manual Thermo Fisher Scientific Chapter 3 Safety Information The following components of the KingFisher Viral NA Kit contain hazardous contents Table 3 1 Wear a laboratory coat disposable gloves and goggles and follow the safety instructions given in the kit instruction manual It is recommended that Good Laboratory Practice GLP is followed to guarantee reliable analyses Table 3 1 Safety precautions Reagent Lysis Buffer Hazardous contents Guanidine thiocyanate lt
15. ble B 1 Thermo Scientific KingFisher Viral NA Kits Cat No Product Package size 97040196 KingFisher Viral NA Kit 1x 96 Table B 2 Thermo Scientific KingFisher Flex consumables Cat No Product Package size 97002514 KingFisher Flex 96 tip comb for PCR magnet 80 pcs 97002524 KingFisher Flex 96 tip comb for KF magnet 100 pes 97002534 KingFisher Flex 96 tip comb for deep well magnet 100 pcs 97002610 KingFisher Flex 24 deep well tip comb and plate 50 pcs 97002540 KingFisher Flex 96 KF plate 200 ul 48 pcs 95040450 Microtiter deep well 96 plate non sterile 50 pcs 95040460 Microtiter deep well 96 plate sterile 50 pcs 95040470 KingFisher Flex 24 deep well plate 50 pcs 95040480 KingFisher Flex 24 deep well plate sterile 50 pcs Table B 3 Thermo Scientific KingFisher Duo consumables Cat No Product Package size 97003500 KingFisher Duo 12 tip comb for Microtiter deep well 96 plate 50 pcs 97003510 KingFisher Duo 6 tip combs and KingFisher 24 deep well plate 48 pcs 12 pcs of 24 deep well plates each including 4 tips combs 97003520 KingFisher Duo elution strip 40 pcs 97003530 KingFisher Duo Combi pack for Microtiter deep well 96 plate 1 box tips combs plates and elution strips for 96 samples Thermo Fisher Scientific Thermo Scientific KingFisher Viral NA Kit Instruction Manual Ordering Information Table B 4 Thermo Scientific KingFisher mL consumables Cat No Product Package size 97002111 Kin
16. c KingFisher mL User Manual Cat No 1508260 and the Thermo Scientific BindIt Software for KingFisher Instruments version 3 2 User Manual Cat No N07974 BindIt Software 3 2 protocols for the KingFisher and the KingFisher 96 can be obtained on request Handl ing of a homogeneous distribution of the KingFisher Magnetic Kin gFisher Beads in the container is essential before the beads are Magnetic Beads transferred to the wells or tubes in order to ensure a high consistency between the wells or tubes To gain complete resuspension of the beads shake the container vigorously or vortex briefly The KingFisher Magnetic Beads have a tendency to sediment relatively quickly in the Binding Buffer Once a premixture of the beads and the Binding Buffer has been made the mixture should be used immediately to avoid the risk of transferring variable amounts of the beads to the respective wells or tubes Thermo Fisher Scientific Thermo Scientific KingFisher Viral NA Kit Instruction Manual 21 Protocols and Pipetting Instructions Instructions for KingFisher Flex with 96 deep well plates for nucleic acid purification from 200 ul of cell free body fluid Instructions for These instructions are for the nucleic acid purification KingFisher Flex from 200 pl of cell free body fluid using the KingFisher with 96 deep well plates for nucleic Viral NA Kit Cat No 97040196 and the KingFisher Flex with Thermo Scientific Microtiter deep well 96 plates
17. e A Sample Content Sample Lysis Buffer Proteinase K working solution Carrier RNA working solution Reagent Sample volume per well 200 ul 200 pl 10 pl 4ul Dispense step add 30 pl of KingFisher Magnetic Beads and 550 ul of Binding Buffer per well to row A during the pause in the Bindlt protocol B Tip 12 tip comb Empty c Empty Empty Empty D Empty Empty Empty E Empty Empty Empty F Wash 1 Wash Buffer 1 500 pl G Wash 2 Wash Buffer 2 500 pl H Wash 3 Wash Buffer 3 550 pl Elution strip Elution Elution Buffer 100 pl Thermo Fisher Scientific Thermo Scientific KingFisher Viral NA Kit Instruction Manual 29 Protocols and Pipetting Instructions Instructions for KingFisher mL for nucleic acid purification from 200 ul of cell free body fluid Instructions for KingFisher mL for nucleic acid purification from 200 pl of cell free body fluid These instructions are for the nucleic acid purification from 200 ul of cell free body fluid using the KingFisher Viral NA Kit Cat No 97040196 and the KingFisher mL When using the KingFisher Viral NA Kit for the first time prepare the working solutions for Proteinase K and Carrier RNA For more instructions refer to Chapter 4 Storage Conditions and Preparation of the Working Solutions A tube strip tray in the KingFisher mL may contain up to 15 separate Thermo Scientific KingFisher mL tube strips and one sample processing uses one
18. fect the yield of the purified nucleic acids The KingFisher Viral NA Kit can be processed at room temperature but to ensure a good yield of purified nucleic acids it is advisable to use heating during the lysis and elution steps or in case of the KingFisher mL heat the Lysis and Elution Buffers The KingFisher Magnetic Beads are highly reactive superparamagnetic beads 12 Thermo Scientific KingFisher Viral NA Kit Instruction Manual Thermo Fisher Scientific KingFisher magnetic particle processors Thermo Fisher Scientific Product Description KingFisher magnetic particle processors The KingFisher magnetic particle processors are designed for the automated transfer and processing of magnetic particles in microplate format The patented technology of the Thermo Scientific KingFisher systems is based on the use of magnetic rods covered with disposable specially designed tip combs and plates or tubes Use only Thermo Scientific KingFisher plastic consumables Use of products from other manufacturers may cause unsuitable mixing or even instability in the KingFisher instrument The instrument functions without any dispensing or aspiration parts or devices Samples and reagents including magnetic particles are dispensed into the plates according to the corresponding instructions Dispensing can be done manually or partially automatically using automatic dispensers for example the Thermo Scientific Multidrop Combi and or the Thermo
19. gFisher mL tip comb 800 pcs 97002121 KingFisher mL tube 20 x 45 pcs 97002131 KingFisher mL combi tubes and tip combs for 60 samples 60 97002141 KingFisher mL combi tubes and tip combs for 240 samples 240 Table B 5 Thermo Scientific KingFisher consumables Cat No Product Package size 97002070 KingFisher tip comb 50 pcs 97002080 KingFisher plate 100 ul 50 pcs 97002084 KingFisher plate 200 ul 50 pcs 97002090 KingFisher plastics 100 pl 8 pack 8 plates and 8 tip combs 1 box 97002094 KingFisher plastics 200 pl 8 pack 8 plates and 8 tip combs 1 box 42 Thermo Scientific KingFisher Viral NA Kit Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Oy Ratastie 2 P O Box 100 Fl 01621 Vantaa Finland www thermoscientific com N11998
20. he KingFisher Magnetic Beads well e g by vortexing before transferring them from the bottle Plate name and type Row Rowname Content Reagent Sample volume per well Viral NA plate A Sample KingFisher 30 pl Magnetic Beads Microtiter deep well 96 plate Binding Buffer 550 ul 7 Place the Viral NA plate back into the instrument and press OK After the pause the protocol will continue to the end 8 After the run is completed remove the plate and store the purified nucleic acids When the protocol is completed remove the plate and elution strip according to the instructions on the KingFisher Duo display and turn off the instrument Store the purified nucleic acids accordingly The purified nucleic acids are ready for use in downstream applications The final nucleic acid concentration in the Elution Buffer may increase if the purified nucleic acids are eluted into a smaller volume of the buffer than is recommended but this can slightly reduce the overall nucleic acid yield 28 Thermo Scientific KingFisher Viral NA Kit Instruction Manual Thermo Fisher Scientific Summary of plate and elution strip contents Plate name and type Viral NA plate Microtiter deep well 96 plate Protocols and Pipetting Instructions Instructions for KingFisher Duo with 12 pin magnet head and 96 deep well plates for nucleic acid purification from 200 ul of cell free body fluids Table 5 2 Summary of plate and elution strip contents Row Row nam
21. he beginning of the purification protocol The Elution Buffer can also be prewarmed to 55 C and dispensed into the Elution tubes during an additional pause step in the protocol before the elution step If some KingFisher Magnetic Beads remain in the elution place the tubes on a magnet for a few minutes to collect the residual beads at the bottom of the tube The final nucleic acid concentration in the Elution Buffer may increase if the purified nucleic acids are eluted into a smaller volume of the buffer than is recommended but this can slightly reduce the overall nucleic acid yield Summary Of Table 5 3 Summary of tube contents tube contents Tube Tube name Content Sample reagent volume A Sample Sample 200 ul Lysis Buffer 200 ul Proteinase K 10 pl working solution Carrier RNA 4ul working solution Dispense step add 30 pl of KingFisher Magnetic Beads per tube and 550 ul of Binding Buffer per tube into the Sample tubes A tubes during the pause in the Bindlt protocol B Wash 1 Wash Buffer 1 500 ul C Wash 2 Wash Buffer 2 500 pl D Wash 3 Wash Buffer 3 550 pl E Elution Elution Buffer 100 pl 32 Thermo Scientific KingFisher Viral NA Kit Instruction Manual Thermo Fisher Scientific Quantification of purified viral RNA and DNA Thermo Fisher Scientific Protocols and Pipetting Instructions Quantification of purified viral RNA and DNA The amount of viral nucleic acids in cell free body fluids is usually very smal
22. hermo Fisher Scientific Thermo Scientific KingFisher Viral NA Kit Instruction Manual 39 Troubleshooting Problem Magnetic particles remaining in the lysed sample or elution well Continued Possible cause and actions Starting material that is too viscose prevents efficient collection of the KingFisher Magnetic Beads from the lysed sample The magnetic rods will not be able to collect all the beads unless the viscose samples are diluted before the beginning of the purification process The samples can for example be diluted into 1 x PBS Improper lysis may also cause problems collecting the KingFisher Magnetic Beads If the KingFisher Magnetic Beads are inefficiently collected from the Elution Buffer the addition of a small amount of detergent e g 0 02 Tween 20 may improve the results Carryover of the KingFisher Magnetic Beads to the Elution Buffer may affect the A A ratio Refer to Low purity on page 39 KingFisher Magnetic Beads that occasionally remain attached to the tip combs at the end of the process do not affect the nucleic acid yield as the nucleic acids have already been released from the KingFisher Magnetic Beads into the Elution Buffer If the KingFisher magnetic particle processor does not work properly refer to the relevant user manual of the KingFisher instrument in use 40 Thermo Scientific KingFisher Viral NA Kit Instruction Manual Thermo Fisher Scientific Appendix B Ordering Information Ta
23. l and the yields of purified nucleic acids are low In addition the Carrier RNA is co purified and affects the analyses of the yield with a spectrophotometer To determine the yield of viral nucleic acids it is recommended to use amplification techniques Thermo Scientific KingFisher Viral NA Kit Instruction Manual 33 Quantification of purified viral RNA and DNA 34 Thermo Scientific KingFisher Viral NA Kit Instruction Manual Thermo Fisher Scientific Chapter 6 General Information Reagent specificity A reagent must not be used with any kit other than that and volumes for which it is intended It is strongly recommended that the volume of reagents in each well or tube is kept within the limits specified in the KingFisher Flex User Manual KingFisher Duo User Manual or KingFisher mL User Manual to avoid spillover and to keep performance at the most efficient level Handl ing of The KingFisher Magnetic Beads have a tendency to magn etic beads sediment relatively quickly in the Binding Buffer Once a premixture of the beads and Binding Buffer has been made the mixture should be used immediately to avoid the risk of transferring variable amounts of the beads to the respective wells or tubes The amount of beads in the wells or tubes affects the yield of the purified nucleic acids Bindi ng and The binding between the nucleic acids and the KingFisher wash steps Magnetic Beads is strong in the presence of a chaotropic salt but chaotropic sa
24. lts are not present in the Wash Buffer 3 and accordingly the binding is weak Avoid strong mixing speeds and releasing the KingFisher Magnetic Beads into the Wash Buffer 3 in order to minimize the loss of nucleic acids during the wash step A short wash and a slow mixing speed without releasing the beads into the buffer are recommended Thermo Fisher Scientific Thermo Scientific KingFisher Viral NA Kit Instruction Manual 35 General Information Elution step Elution step Carryover of ethanol to the Elution Buffer causes impurities in the Elution Buffer and may affect some downstream applications To remove traces of ethanol make sure that there is a wash step e g washing without releasing the beads or the drying step before the elution step is long enough There is a delicate balance between complete removal of the ethanol and loss of nucleic acids The volume of the Elution Buffer can be modified depending on the user requirements concerning the purified nucleic acid concentration The final nucleic acid concentration in the Elution Buffer may increase if the purified nucleic acids are eluted into a smaller volume of the buffer but this can slightly reduce the overall nucleic acid yield The modifications of the elution step must be done in BindIt Software 3 2 and according to the volume ranges suitable for the KingFisher instrument The table below indicates the available elution volumes of the KingFisher instruments Table 6 1
25. nd wash steps ccceecseseeeseseeeeeesseseeeeeeseeeeeeeseeeeees 35 Elution step stint tcieeoiae asain 36 Decontamination and disinfection of sample material 37 TroublesShooting scsssscsssscsessesessessessesesseaesssseesessesseaesseaeeeseeaees 39 Ordering Information sssssssssssssssessessessesscssesasssesseeseesseees 41 6 Thermo Scientific KingFisher Viral NA Kit Instruction Manual Thermo Fisher Scientific Chapter 1 Kit Content Table 1 1 Thermo Scientific KingFisher Viral NA Kit Item KingFisher Viral NA Kit Cat No 97040196 Package size 1x 96 samples KingFisher Magnetic Beads 3 2 ml Proteinase K 1 via Proteinase K Buffer 1 8 ml Carrier RNA 1via Carrier RNA Buffer 500 ul Lysis Buffer 25m Binding Buffer 65m Wash Buffer 1 55m Wash Buffer 2 55m Wash Buffer 3 65m Elution Buffer 11m The KingFisher Viral NA Kit Cat No 97040196 is intended for the purification of DNA and RNA from samples using the Thermo Scientific KingFisher Flex with a 96 deep well head or the Thermo Scientific KingFisher Duo with a 12 pin head or Thermo Scientific KingFisher mL and a sample volume of 200 ul Thermo Fisher Scientific Thermo Scientific KingFisher Viral NA Kit Instruction Manual 7 Kit Content Table 1 2 Thermo Scientific KingFisher magnetic particle processors Equipment and reagents to be supplied by the user e Magnetic particle processor
26. ng solution Dispense step add 30 pl of KingFisher Magnetic Beads per well and 550 ul of Binding Buffer per well to the Sample plate Plate 1 during the pause in the Bindlt protocol 2 Microtiter Wash1 Wash Buffer 1 500 ul deep well 96 plate 3 Microtiter Wash2 Wash Buffer 2 500 ul deep well 96 plate 4 Microtiter Wash3 Wash Buffer 3 550 ul deep well 96 plate 5 KingFisher Elution Elution Buffer 100 pl Flex 96 KF plate 6 KingFisher Tip Plate Flex 96 KF plate Thermo Scientific KingFisher Viral NA Kit Instruction Manual 25 Protocols and Pipetting Instructions Instructions for KingFisher Duo with 12 pin magnet head and 96 deep well plates for nucleic acid purification from 200 yl of cell free body fluids Instructions for KingFisher Duo with 12 pin magnet head and 96 deep well plates for nucleic acid purification from 200 pl of cell free body fluids Plate name and type Viral NA plate Microtiter deep well 96 plate These instructions are for the nucleic acid purification from 200 ul of cell free body fluid using the KingFisher Viral NA Kit Cat No 97040196 and the KingFisher Duo with a 12 pin magnet head and Microtiter deep well 96 plates When using the KingFisher Viral NA Kit for the first time prepare the working solutions for Proteinase K and Carrier RNA For more instructions refer to Chapter 4 Storage Conditions and Preparation of the Working Solutions 1 Take one empty Micro
27. per run per run volume 200 pl with the KingFisher Viral NA Kit Magnetic head 96 interchangeable 24 format for 12 pin magnethead 6 pin magnet head formats for PCR KingFisher Flex 24 for Microtiter deep for KingFisher Flex plate KingFisher Flex 96 KF plate Microtiter deep well 96 plate deep well plate well 96 plate 24 deep well plate Plates KingFisher Flex 96 KingFisher Flex 24 Microtiter deep well KingFisher Flex 24 KF plate 20 200 ul deep well plate 96 plate 50 1000 ul deep well plate 96 well PCR plate 200 5000 ul KingFisher Duo 200 5000 ul skirted 20 100 ul elution strip Microtiter deep well 30 130 ul 96 plate 50 1000 ul Tip combs KingFisher Flex 96 KingFisher Flex 24 KingFisher Duo KingFisher Duo tip comb for PCR tip comb for deep 12 tip comb 6 tip comb magnets KingFisher well magnets Flex tip comb for KF magnets KingFisher Flex 96 tip comb for deep well magnets Heating Heating block temperature from 5 C Heating block temperature from 10 C temperature above ambient room temperature to 75 C elution strip 4 C to 75 C in to 115 C 14 Thermo Scientific KingFisher Viral NA Kit Instruction Manual room temperature Thermo Fisher Scientific Product Description KingFisher magnetic particle processors Table 2 1 Overview of Thermo Scientific KingFisher systems KingFisher mL KingFisher Processing volume 50 1000 pl 20 200 ul Capacity Up to 15 samples per run s
28. refore suitable for direct use in many different downstream applications such as PCR polymerase chain reaction restriction analysis and in several other enzymatic reactions The KingFisher Viral NA Kit is developed for the purification of viral RNA and DNA from cell free body fluids such as serum or plasma using paramagnetic particles The purification process requires no phenol chloroform extraction or alcohol precipitation and needs very little hands on time The reagents and specific plastic consumables are designed to work with the KingFisher Flex KingFisher Duo or KingFisher mL magnetic particle processors as part of an integrated system The KingFisher Viral NA Kit is only intended for research use not for clinical or diagnostic use The user is responsible for validating the performance of the KingFisher instrument and the KingFisher Viral NA Kit for any particular use because the performance of the kits has not been validated for any specific organism The KingFisher Viral NA Kit uses magnetic particle technology for viral RNA and DNA purification The Thermo Scientific KingFisher technology combines the speed and efficiency of nucleic acid purification with easy handling of magnetic particles Nucleic acids bind to the surface of the Thermo Scientific KingFisher Magnetic Thermo Scientific KingFisher Viral NA Kit Instruction Manual 11 Product Description Kit specifications Kit specifications Beads in the presence
29. ssssssssssssessessessessessessessssseesseseseesaeees 17 Storage Conditions and Preparation of the Working Solutions sssssssssssssssssssssssssssssssssssessssesssesesaees Storage conditons s sasise inaa aE a nEn Preparation of the Proteinase K working solution 0 0 0 0 Preparation of the Carrier RNA working solution Protocols and Pipetting Instructions Handling of KingFisher Magnetic Beads Instructions for KingFisher Flex with 96 deep well plates for nucleic acid purification from 200 ul of cell free body fluid 22 Summary of plate contents 0 seseseseeeseseeeseseetenenaees 25 Instructions for KingFisher Duo with 12 pin magnet head and 96 deep well plates for nucleic acid purification from 200 ul of cell free body fluids eee piatan LO Summary of plate and elution strip contents eee 29 Instructions for KingFisher mL for nucleic acid purification from 200 ul of cell free body fluid oo cceee eee eseeeeeeeeees 30 Summary of tube contents cccssseesseseseseeeseseeeeeseeeeneesees 32 Quantification of purified viral RNA and DNA ee 33 Thermo Scientific KingFisher Viral NA Kit Instruction Manual 5 Contents Chapter 6 Appendix A Appendix B General Informmation csssessecssescssesssessssessseeeseeeseeesseeenteenseeens 35 Reagent specificity and volumes cece eeeeeeeeeeeeeeeeeeeeeeees 35 Handling of magnetic beads oo ee eee eeeeeeeeeeeeeeeeeeeeeeeeees 35 Binding a
30. th the KingFisher Duo and load the plate and elution strip Switch on the KingFisher Duo and make sure that you are using the KingFisher Duo 12 pin magnet head and heating block Connect the PC with BindIt Software 3 2 to the KingFisher Duo Start the KF_viralNA_Duo protocol Insert the Viral NA plate and elution strip into the instrument as indicated on the KingFisher Duo display and press OK Make sure that the elution strip is placed in the correct direction into the elution block Ensure that the perforated end is facing towards the user When the KingFisher Duo is to be run as a standalone instrument transfer the KF_viralNA_Duo protocol to the KingFisher Duo The instructions for transferring the protocol can be found in the BindIt Software for KingFisher Instruments version 3 2 User Manual 6 Add the KingFisher Magnetic Beads and Binding Buffer to row A during the dispense step When the KingFisher Duo pauses at the dispense step after the lysis step approximately 15 minutes after starting Thermo Scientific KingFisher Viral NA Kit Instruction Manual 27 Protocols and Pipetting Instructions Instructions for KingFisher Duo with 12 pin magnet head and 96 deep well plates for nucleic acid purification from 200 yl of cell free body fluids the protocol run remove the Viral NA plate from the instrument and separately add the KingFisher Magnetic Beads and Binding Buffer to row A in the Viral NA plate as indicated below Resuspend t
31. tion strip 40 pcs 97003530 KingFisher Duo Combi pack for Microtiter deep well 96 plate 1 box tips combs plates and elution strips for 96 samples Table 1 5 Thermo Scientific KingFisher mL consumables Cat No Product Package size 97002111 KingFisher mL tip comb 800 pcs 97002121 KingFisher mL tube 20 x 45 pcs 97002131 KingFisher mL combi tubes and tip combs for 60 samples 60 97002141 KingFisher mL combi tubes and tip combs for 240 samples 240 Table 1 6 Thermo Scientific KingFisher consumables Cat No Product Package size 97002070 KingFisher tip comb 50 pcs 97002080 KingFisher plate 100 pl 50 pcs 97002084 KingFisher plate 200 pl 50 pcs 97002090 KingFisher plastics 100 pl 8 pack 8 plates and 8 tip combs 1 box 97002094 KingFisher plastics 200 pl 8 pack 8 plates and 8 tip combs 1 box Thermo Fisher Scientific Thermo Scientific KingFisher Viral NA Kit Instruction Manual 9 Kit Content 10 Thermo Scientific KingFisher Viral NA Kit Instruction Manual Thermo Fisher Scientific Introduction Intended use Principle and procedure Thermo Fisher Scientific Chapter 2 Product Description The KingFisher Viral NA Kit is designed for rapid automated purification of viral RNA and DNA from cell free body fluids using KingFisher instruments The nucleic acids NA purified using the KingFisher Viral NA Kit are of high quality and free of proteins nucleases and other contaminants or inhibitors They are the
32. titer deep well 96 plate and one Thermo Scientific KingFisher Duo elution strip 2 Prepare the Viral NA plate Microtiter deep well 96 plate Add the following reagents to the rows Note that row B is reserved for the tip comb and should be left empty Note that rows C D and E are left empty Row Row name Content Reagent Sample volume per well A Sample Sample 200 ul Lysis Buffer 200 ul Proteinase K 10 pl working solution Carrier RNA 4ul working solution B Tip 12 tip comb Empty c Empty Empty Empty D Empty Empty Empty E Empty Empty Empty F Wash 1 Wash Buffer 1 500 ul G Wash 2 Wash Buffer 2 500 ul H Wash 3 Wash Buffer 3 550 ul 26 Thermo Scientific KingFisher Viral NA Kit Instruction Manual Thermo Fisher Scientific Protocols and Pipetting Instructions Instructions for KingFisher Duo with 12 pin magnet head and 96 deep well plates for nucleic acid purification from 200 ul of cell free body fluids Thermo Fisher Scientific 3 Fill the KingFisher Duo elution strip as follows Make sure that the elution strip is placed in the correct direction into the elution block Ensure that the perforated end is facing towards the user and the Elution Buffer is pipetted into the correct wells Elution strip Content Reagent volume per well KingFisher Duo elution strip Elution Buffer 100 pl 4 Place a Thermo Scientific KingFisher Duo 12 tip comb into row B on a Viral NA plate 5 Start the KF_viralNA_Duo protocol wi
33. tube strip with five tubes The orientation of the tube strip is fixed Note that the tube strips have to be positioned so that the slip ends are facing left One tip comb with five tips is used to process five samples at a time 1 Place empty KingFisher mL tubes on a tube strip tray 2 Prepare the tubes that is starting from the first tube at the slip end of a tube strip Add the following reagents to the tubes Tube Tube name Content Reagent volume A Sample Sample 200 ul Lysis Buffer 200 ul Proteinase K 10 pl working solution Carrier RNA 4ul working solution B Wash 1 Wash Buffer 1 500 ul C Wash 2 Wash Buffer 2 500 pl D Wash 3 Wash Buffer 3 550 ul E Elution Elution Buffer 100 pl 3 Prepare the KingFisher mL for the run 30 Thermo Scientific KingFisher Viral NA Kit Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Protocols and Pipetting Instructions Instructions for KingFisher mL for nucleic acid purification from 200 ul of cell free body fluid Switch on the KingFisher mL and insert the tray into the instrument Insert the tip combs into their slots and close the front lid 4 Start the KF_ViralNA_mL protocol with the KingFisher mL Connect the PC with BindIt Software 3 2 to the KingFisher mL Start the KF_ViralNA_mL protocol When the KingFisher mL is to be run as a standalone instrument transfer the KF_ViralNA_mL protocol to the KingFisher mL The instructions for transferring the
34. x 96 and load the plates Switch on the KingFisher Flex and make sure that you are using the KingFisher Flex 96 deep well head and heating block Connect the PC with BindIt Software 3 2 to the KingFisher Flex Start the KF_ViralNA_Flex96 protocol Insert the Tip Plate and the filled plates into the instrument as indicated on the KingFisher Flex display After all the plates have been loaded into the instrument the protocol will begin When the KingFisher Flex is to be run as a standalone instrument transfer the KF_ViralNA_Flex96 protocol to the KingFisher Flex The instructions for transferring the protocol can be found in Chapter 4 Using the software in the BindIt Software for KingFisher Instruments version 3 2 User Manual 5 Add the KingFisher Magnetic Beads and Binding Buffer to the Sample plate during the dispense step When the KingFisher Flex pauses at the dispense step after the lysis step approximately 15 minutes after starting the protocol run remove the Sample plate from the instrument and separately add the KingFisher Magnetic Beads and Binding Buffer to the Sample plate as indicated below Resuspend the KingFisher Magnetic Beads well e g by vortexing before transferring them from the bottle Thermo Fisher Scientific Thermo Scientific KingFisher Viral NA Kit Instruction Manual 23 Protocols and Pipetting Instructions Instructions for KingFisher Flex with 96 deep well plates for nucleic acid purification from 20