1. Oxygen Concentration 90 3 for all flow settings Oxygen Concentration Indicator Green Light Normal Operation Yellow Light Warning or Caution less than 85 3 Red Light Flashing Abnormal Operation less than 70 5 Maximum System Pressure 15 psig 103 5 kPa Oxygen Outlet Pressure 5 0 psig 34 5 kPa nominal Electrical Power AC Operation DC Operation Power Cartridge AC Power Supply DC Power Supply 100 240 VAC 50 60 Hz 12V 195 W Hrs 200 Watts Max 150 Watts Max Operating Temperature 50 F to 104 F 10 C to 40 C Operating Humidity 10 95 at 82 4 F 28 C dew point Nominal Sound Level 48 dBA at 3 0 LPM in Continuous Flow Mode 40 dBA at 3 0 LPM in Pulse Flow Mode Dimensions H x W x D 19 3 inches x 12 3 inches x 7 1 inches 49 0 cm x 31 2 cm x 18 0 cm Weight 17 9 pounds 8 1 kg with Power Cartridge 14 5 pounds 6 6 kg without Power Cartridge Continuous Flow Indication Expressed in liters per minute LPM Audible Alarm Indicators Back Up Alarm Power Loss of power Hot Power Cartridge Low Battery Warm Power Cartridge Low Oxygen Output O flow outside normal limits No inspiration detected in Pulse Mode Unit Malfunction 9V Internal Battery Filters Device Classification Air Inlet HEPA IEC Class Type B Applied Part IPXO Transport Storage Requirements Temperature 4 F 20 C to 140 F
2. Contact your dealer provider Check tubing or replace Clean Air Inlet Filter Humidifier Check humidifier attachment and tubing ECLIPSE not ON Turn ECLIPSE ON Tubing or cannula is not properly connected or kinked Power Cartridge is discharged or no external power is present Check tubing cannula and connections Install charged Power Cartridge or attach external power Malfunction Restriction in tubing or humidifier Contact your dealer provider Repair or replace tubing or humidifier Filter restricted Malfunction Restriction in humidifier or tubing Clean or replace air inlet filter Contact your dealer provider Repair or replace humidifier or tubing Filter Blocked Clean or replace air inlet filter Malfunction Tubing cannula longer than 7 feet 2 1m Contact your dealer provider Attach 7 foot 2 1m tubing cannula Humidifier attached Remove humidifier No inspiration detected Power Cartridge needs calibration Contact your dealer provider Fully discharge the Power Cartridge and recharge ECLIPSE Model 1000 User Manual 31 OTER Accessories Use of certain oxygen administration accessories not specified for use with this oxygen system may impair the performance ONLY use recommended oxygen delivery accessories ECLIPSE Model 1000 Cart 7 lt DC Power Supply Model 6105 AC Power Supply M
3. 60 C Humidity up to 95 Non Condensing Back pressure Effect on Flow Indicator None Altitude 0 13 123 feet 0 4 000 meters 38 ECLIPSE Model 1000 User Manual Model 1000 User Manual Pulse Flow Mode Principle of Operation Pulse Flow Mode is designed to deliver 16 ml boluses of oxygen per incremental flow setting for each patient inspiration detected Bolus volume is held constant regardless of breath rate within a given setting Delivery of the bolus is triggered by detection of the end of inspiration followed by the start of a subsequent inspiration Sensitivity is dealer provider adjustable Based on this inspiration trigger signal a pulse of oxygen is released to the patient by an electronically controlled flow valve There is no mixing of oxygen with room air Range of Operation Settings Increments of 1 6 times x the bolus of 16 ml 16 to 96 ml per breath Accuracy of Delivered Bolus 15 Breathing Frequency range 6 to 25 breaths per minute average sustainable Minimum time between breaths 1 25 seconds maximum 3 consecutive breaths Response to missing breaths ECLIPSE switches to Continuous Flow Mode after 30 seconds without breath detection Specifications are subject to change without notice ECLIPSE Model 1000 User Manual 39 OTER Summary of All Warnings and Cautions FEDERAL LAW RESTRICTS THIS DEVICE TO SALE B
4. turn the system back ON to resume oxygen production 24 ECLIPSE Model 1000 User Manual Model 1000 User Manual Operating Your ECLIPSE STEP 1 LOCATING YOUR ECLIPSE m Place the ECLIPSE in a well ventilated place so that the air inlet and exhaust have clear access m Position the ECLIPSE so that any alarm can be heard Check to see if the air inlet filter is in place before using If the air inlet filter needs to be cleaned wash with soapy water and allow drying before using If necessary replace with a new air inlet filter m Plug unit into AC Power DC Power or make sure there is a charged Power Cartridge installed ECLIPSE Model 1000 User Manual 25 OTER Operating Instructions continued STEP 2 TURN ON THE UNIT AND ALLOW TO WARM UP m Press and hold the ON OFF Button for one 1 second to turn ON your ECLIPSE A brief self test will happen Verify that all the indicators are lit and a 3 second buzzer sounds indicating the system is in working order Warm up Period Oxygen Monitoring The ECLIPSE Oxygen System has an oxygen monitoring system with a yellow alert and red alarm visual indicator and buzzer The buzzer will activate in the event of an internal malfunction or power outage external power loss with no Power Cartridge installed The red yellow and green all OK indicators will initially light up and then the red and yellow indicators sequentially go out when the oxy
5. ALERT Yellow Indicator Low and Medium Priority Alerts When steadily illuminated this indicates a low priority awareness J condition or Caution Continue to use your system and refer to the ECLIPSE Model 1000 User Manual 13 OTER Understanding the Controls and Front Panel continued Troubleshooting section for the proper response A flashing yellow indicates a medium priority alert Prompt response is needed to deal with a condition Refer to the Troubleshooting section ALARM Red Indicator High Priority Alarms Indicates a high priority alarm condition Immediate response is necessary to deal with the condition Refer to the Troubleshooting section and contact your dealer provider Buzzer An audible alarm or buzzer is used to alert you of the condition of the unit either a warning or failure and to confirm a valid key press Flow Rate Setting Indicator Green This is the main focus on your control panel Your dealer provider will set up your prescribed flow rates for your Continuous Flow Mode LPM and your Pulse Flow settings if pulse flow has been prescribed DO NOT change these settings unless it is under the order of your physician Each time you turn the unit on the last flow mode and flow setting will be used Power Cartridge Status Gauge This indicator displays the charge remaining in the Power Cartridge When the Power Cartridge is fully charged all 10 bars are illuminated Each bar represents app
6. ECLIPSE m Store the ECLIPSE unit in a cool dry location ECLIPSE Model 1000 User Manual 29 Alarm Conditions and Troubleshooting The table shows possible alarm conditions and suggests appropriate responses If any other alarms occur call your authorized SeQual dealer provider Green Yellow Red Audible Condition Indicator Indicator Indicator Alarm What to do uw Warming Up Please Wait On None Wait The system typically takes a 5 3 5 minutes to reach specified lt 2 GE performance aA 5 2 0 or flow setting Off None Nothing The system is operating o properly at the specified flow rate Low or Warm Power Blinking Off Beeps Plug into external power or replace Cartridge with a charged Power Cartridge O2 Concentration lt 85 Blinking Off None Continue Use Check or clean air intake filter If condition persists contact dealer provider O Concentration lt 70 Blinking Beeps Continue Use Check or clean air intake filter If condition persists z contact dealer provider E Z Flow Rate Error Blinking Off Beeps Check tubing and or humidifier Blocked Flow for obstruction Check clean F air intake filter If condition a persists past 10 minutes H contact dealer provider No Inspiration Detected Off Beeps Make sure tubing does not gt in Pulse Flow Mode only exceed 7 ft 2 1m when using Pulse Flow Mode Audible alarm will be
7. System and its accessories Thoroughly read all the information in this manual in its entirety before operating your ECLIPSE Contact your dealer provider for proper training in the use of this device YOUR PHYSICIAN HAS PRESCRIBED SUPPLEMENTAL OXYGEN AS PART OF A TREATMENT PLAN LONG TERM OXYGEN THERAPY IS NOW WIDELY ACCEPTED AS THE STANDARD TREATMENT FOR CHRONIC HYPOXIA CAUSED BY CHRONIC OBSTRUCTIVE PULMONARY DISEASE OXYGEN THERAPY HAS BECOME MORE VERSATILE IN RECENT YEARS DUE TO THE DEVELOPMENT OF OXYGEN CONCENTRATORS THE ECLIPSE MAKES IT EASIER FOR PHYSICIANS TO SELECT A MODE OF OXYGEN THERAPY THAT SUITS THE PATIENT S LIFESTYLE THE ECLIPSE MAY BE USED IN HOME INSTITUTION AND MOTOR VEHICLES USE OF THE ECLIPSE IN HOSPITALS SHOULD INCLUDE THE USE OFA HOSPITAL GRADE POWER CORD YOUR PHYSICIAN HAS PRESCRIBED A SPECIFIC OXYGEN FLOW RATE SETTING TO MEET YOUR INDIVIDUAL NEEDS THIS FLOW RATE IS DETERMINED BY THE AMOUNT OF OXYGEN YOU NEED RECOMMENDED OXYGEN FLOW RATES SHOULD BE ADJUSTED ONLY UNDER THE ADVICE OF A PHYSICIAN 10 ECLIPSE Model 1000 User Manual Model 1000 User Manual Understanding the Controls and Front Panel Description of the ECLIPSE The key features of the ECLIPSE Oxygen System and control panel are important items to become familiar with Control Panel Oxygen Outlet Port Service Port Simplified User Not for Patient Use Instructions Under Base Air Inlet Filter Externa
8. and make sure the air inlet and exhaust airflow are not blocked This may impact performance WARNING DO NOT leave the ECLIPSE or DC Power Supply plugged into the vehicle without the engine running or attempt to start the vehicle while the DC Power Supply is connected to the vehicle This may drain the vehicle s battery A buzzer in the DC Power Supply is intended to warn user when low vehicle power is present Immediately disconnect the DC Power Supply from the vehicle for low voltage alarm conditions If the vehicle power source vehicle accessory connector drops below 11 5 Volts brownout the ECLIPSE will revert to Power Cartridge operation if present The External Power Present Indicator may remain on TYPICAL FLOW RATES AVAILABLE AT EACH FUSE SETTING 10 AMP 15 AMP Setting Continuous Pulse 0 5 1 0 Yes 1 5 Yes 2 0 Yes 2 5 Yes 3 0 Yes 3 5 Yes 4 0 Yes 4 5 Yes 5 0 Yes 5 5 Yes 6 0 Yes Actual performance may vary dependent upon usage conditions and vehicle 20 ECLIPSE Model 1000 User Manual Model 1000 User Manual When operating in the 10 AMP switch position flow rate settings above 2 0 LPM continuous flow will automatically switch to 2 0 LPM continuous flow when the DC Power Supply is connected 3 RECHARGEABLE POWER CARTRIDGE The ECLIPSE system can also be powered by a rechargeable Power Cartridge Installing the Power Cartridge Align
9. and eyes Keep the Power Cartridge away from children ECLIPSE Model 1000 User Manual 41 OTE WARNING Use of a damaged Power Cartridge may cause personal injury WARNING Exposing the Power Cartridge to water or other liquids may cause personal injury ONJ Protect the ECLIPSE and AC DC Power Supplies from fluid spills or drips to avoid possible shock hazards DO NOT use any cord if damaged Improper use of the Power Cord may result in a fire or electrical shock hazard DO NOT touch the Power Cartridge cavity connectors with the Power Cartridge removed while the unit is connected to an external power source This may result in electrical shock WARNING Secure the ECLIPSE and the DC Power Supply in your vehicle and make sure the air inlet and exhaust airflow are not blocked This may impact performance DO NOT leave the ECLIPSE or DC Power Supply plugged into the vehicle without the engine running or attempt to start the vehicle while the DC Power Supply is connected to the vehicle This may drain the vehicle s battery If the vehicle power source vehicle accessory connector drops below 11 5 Volts brownout the ECLIPSE will revert to Power Cartridge operation if present The External Power Present Indicator may remain on WARNING When operating in Continuous Flow Mode to ensure adequate flow DO NOT use a tubing cannula extension more than 50
10. in a small enclosed space such as a bathroom or closet with cannula leading out of the enclosed space 8 ECLIPSE Model 1000 User Manual Model 1000 User Manual Power Cartridge Safety WARNING This Power Cartridge may explode and cause potential injury if exposed to or disposed of in a fire DO NOT short circuit the Power Cartridge s metal contacts with metallic objects such as keys or coins It may cause sparks or excessive heat DO NOT disassemble puncture or crush the Power Cartridge Power Cartridge electrolytes may be toxic if swallowed and can be harmful to skin and eyes Keep the Power Cartridge away from children Use of a damaged Power Cartridge may cause personal injury WARNING Exposing the Power Cartridge to water or other liquids may cause personal injury CAUTION DO NOT expose the Power Cartridge to temperatures above 140 F 60 C such as in a vehicle parked in the sun or on a hot day CAUTION Charging the Power Cartridge below 50 F 10 C or above 104 F 40 C may degrade the performance CAUTION DO NOT drop the Power Cartridge or expose it to mechanical shock CAUTION Only use the Power Cartridge for its intended purpose ECLIPSE Model 1000 User Manual 9 E Introduction to Your ECLIPSE Oxygen System This User Manual will inform you about the use and care of the ECLIPSE Oxygen
11. quilt etc DONOT overfill the humidifier if used Fill with water only to the level shown by the manufacturer m When using ECLIPSE in a vehicle make sure it is properly secured or restrained ECLIPSE Model 1000 User Manual 7 OTER Safety Guidelines continued Operating Safety continued WARNING DO NOT remove the cover no serviceable parts inside Qualified service personnel should only remove the cover WARNING DO NOT operate device or accessories on wet surfaces or in standing water and DO NOT submerse or expose to water The ECLIPSE enclosure does not provide protection against the harmful effects of liquid ingress Electrical shock or damage to the unit may result WARNING Protect electrical cords from sharp edges Electrical shock and serious physical injury may result WARNING Use only the SeQual AC Power Supply and DC Power Supply supplied with your ECLIPSE Oxygen System Use of a non SeQual Power Supply may be hazardous cause damage to your oxygen system and will void the warranty CAUTION Availability of an alternate source of oxygen is recommended in case of power outage or mechanical failure CAUTION DONOTexpose unit to temperature outside of specified operating or storage temperatures as this may damage the unit CAUTION Locate oxygen tubing and Power Supply cords to prevent tripping hazards CAUTION DO NOT place ECLIPSE
12. the Power Cartridge so that it will slide into the Power Compartment Push the Power Cartridge into the Power Compartment until it is secure and flush with the back of the system When the Power Cartridge is properly inserted and the ECLIPSE is ON the Power Cartridge Status Gauge will appear on the Control Panel If you DO NOT have a Power Cartridge installed the Status Gauge will not light up To remove the Power Cartridge push up on the latch on the Power Cartridge The Power Cartridge will slide out so that you can grasp it for removal Power Cartridge Capacity The capacity of the Power Cartridge is lit up on the control panel A variety of factors such as flow rate and Pulse or Continuous Flow Mode operation determine the duration of operating time The following table provides estimates of amount of time that the ECLIPSE can operate on a fully charged Power Cartridge given certain flow rates and operating conditions Power Cartridge ECLIPSE Model 1000 User Manual 21 OTE Operating Instructions continued TYPICAL NEW POWER CARTRIDGE OPERATION TIME Flow Rate Setting Continuous Flow Mode LPM Pulse Flow Mode Setting 0 5 4 4 hours 1 0 3 7 hours 4 4 hours 2 0 2 4 hours 3 6 hours 3 0 1 3 hours 3 0 hours 4 0 a 2 6 hours 5 0 2 3 hours 6 0 2 1 hours Operating time will degrade with Power Cartridge use and age CAUTION Store your Power Cartridge in a cool dry locat
13. you operate and maintain the unit If you have any further questions or problems call your SeQual dealer provider or authorized service representative Important DO NOT attempt to operate the ECLIPSE System without first reading the Safety Guidelines section of this manual Follow all operating instructions Observe all warnings on the product and in the operating instructions To reduce the risk of fire bodily injuries and damage to the oxygen system observe all safety precautions CON Protect the ECLIPSE and AC DC Power Supplies from fluid spills or drips to avoid possible shock hazards CAUTION Always check to see that the Air Inlet and the Exhaust Vent are not blocked and the air inlet filter is clean before using your system Air Inlet and Exhaust Vent Air Inlet Filter ECLIPSE Model 1000 User Manual 17 Ls Operating Instructions continued Powering Your ECLIPSE System The ECLIPSE is a lightweight portable oxygen system capable of being operated directly from three different power sources 1 an AC Power Supply 2 a DC Power Supply or 3 a rechargeable Power Cartridge 3 z 3 RECHARGEABLE POWER CARTRIDGE 1 AC POWER SUPPLY 2 DC POWER SUPPLY The ECLIPSE charges the optional Power Cartridge when external power AC or DC is present and when the Power Cartridge temperature is below a safe charging temperature When external power is disconnected the sys
14. 15 2m long 42 ECLIPSE Model 1000 User Manual Model 1000 User Manual WARNING Operating your ECLIPSE outside the recommended operating environment may impact performance cause damage and will void the warranty Pulse Flow settings should be determined for each patient individually Settings from Continuous Flow applications may not be applicable to Pulse Flow Mode When operating in Pulse Flow Mode to ensure adequate breath detection the length of the tubing cannula must not exceed 7 feet 2 1m WARNING As with all conserving devices the ECLIPSE may not be able to detect some respiratory efforts WARNING DO NOT change from prescribed flow rate setting without consulting with a qualified clinician AVAILABILITY OF AN ALTERNATE SOURCE OF OXYGEN IS RECOMMENDED IN CASE OF POWER OUTAGE OR MECHANICAL FAILURE CONSULT YOUR PHYSICIAN FOR THE TYPE OF BACK UP SYSTEM REQUIRED DO NOT IGNORE ALARMS WARNING Use of certain oxygen administration accessories not specified for use with this oxygen system may impair the performance ONLY use recommended oxygen delivery accessories EZOM The tube fitting may be tight DO NOT use oil or grease to lubricate ECLIPSE Model 1000 User Manual 43 OTE If tubing becomes disconnected during operation ECLIPSE does not alarm Check flow at the cannula to verify oxygen delivery DO NOT tamper with
15. ESENCE OF FLAMMABLE ANESTHETICS CON AS WITH ANY ELECTRICALLY POWERED DEVICE THE USER MAY EXPERIENCE PERIODS OF NON OPERATION AS THE RESULT OF ELECTRICAL POWER INTERRUPTION OR THE NEED TO HAVE THE OXYGEN CONCENTRATOR SERVICED BY A TECHNICIAN AN OXYGEN CONCENTRATOR IS NOT APPROPRIATE FOR ANY PATIENT WHO WOULD EXPERIENCE ADVERSE HEALTH CONSEQUENCES AS THE RESULT OF SUCH TEMPORARY INTERRUPTION ECLIPSE Model 1000 User Manual 5 OTER Safety Guidelines This section of the manual provides warnings and cautions that apply to hazards or unsafe practices that could result in personal injury or property damage Oxygen Therapy Safety Read Instruction Manual DO NOT operate this oxygen system without first reading and understanding this manual If you are unable to understand the warnings and instructions contact your authorized SeQual dealer provider before attempting to use this oxygen system be hazardous therefore it is advisable to seek medical advice before using the oxygen system i Caution or Attention In certain circumstances oxygen therapy can DO NOT deviate from prescribed flow rates Adjust the flow only to your prescribed flow rate unless otherwise instructed by your physician WARNING No Smoking or Naked Flames The use of oxygen therapy requires that special care be taken to reduce the risk of fire Oxygen enhances S the combustion of flammable material For safety conce
16. N DO NOT connect the ECLIPSE to an extension cord or electrical outlet controlled by a switch CAUTION Store your Power Cartridge in a cool dry location This will help to assure the longevity of your Power Cartridge CAUTION DOT UN Regulations require the removal of the Power Cartridge for international travel CAUTION There are no serviceable parts inside the Power Cartridge DO NOT open CAUTION Replace the Power Cartridge with only SeQual approved Power Cartridge CAUTION The Power Cartridge used in this device may present a risk of fire or chemical burn if mistreated DO NOT disassemble incinerate or heat above 140 F 60 C CAUTION Return your Power Cartridge to your dealer provider for proper disposal 46 ECLIPSE Model 1000 User Manual Model 1000 User Manual CAUTION Keep your Power Cartridge away from children CAUTION Store the Power Cartridge in a cool dry place when not in use CAUTION DONOT leave your ECLIPSE or your Power Cartridge in your vehicle or in a trunk during a hot day CAUTION The ECLIPSE system can only work with a SeQual Power Cartridge Use of another Power Cartridge or Battery may damage the unit present a risk of fire or explosion and will void the warranty CAUTION To ensure there is adequate flow to deliver oxygen the length of the tubing must not ex
17. SE outside the recommended operating environment may impact performance cause damage and will void the warranty CAUTION Locate the unit in a well ventilated room that provides adequate air flow and check if air is being prevented from entering or exiting the system by furniture draperies bedding clothing or pets ECLIPSE Model 1000 User Manual 15 as Recommended Operating Environments continued The Proper Location Select a location that avoids fumes and pollutants and allows the oxygen system to draw in room air through the air inlet filter on the back of the cabinet and exhaust air out the vent at the bottom of the case without being restricted Locate the unit such that the alarms can be heard Route oxygen tubing such that it does not kink or is subject to occlusion DO NOT cover unit with a blanket towel quilt or other covering as the unit may overheat DO NOT locate near flammable materials or cleaning agents or in the direct path of any heat source such as a heat register or car heater Keep the oxygen system at least five 5 feet 1 5 m away from hot sparking objects or open sources of flame 16 ECLIPSE Model 1000 User Manual Model 1000 User Manual Operating Instructions Before Operating Your SeQual dealer provider or authorized service representative will demonstrate how to operate the oxygen system This manual serves as a supplemental reference to help
18. SS SEQuaL The Smart Choice Model 1000 User Manual Y e S 4 9 be Contents TEOL ETETEN Important Safety Instructions 2 Symbols Used in User Manual and on Oxygen System 2 Indications For Use 5 Contraindications 5 Safety Guidelines 6 Oxygen Therapy Safety 6 Operating Safety 7 Power Cartridge Safety 9 Introduction To Your ECLIPSE Oxygen System 10 Understanding the Controls and Front Panel 11 Description of the ECLIPSE 11 Control Panel Features 12 User Controls 13 Front Panel Icons 13 System Status Indicators 13 Recommended Operating Environments 15 The Proper Location 16 Operating Instructions 17 Before Operating 17 Powering Your ECLIPSE System 18 Operating Your ECLIPSE System 25 Alarm Conditions and Troubleshooting 30 Accessories 32 Cleaning Care and Routine Maintenance 35 Technical Data 38 Summary of All Warnings and Cautions 40 Electromagnetic Compatibility 48 Disposal of Equipment and Accessories 49 OTER Important Safety Instructions Important Safety Instructions are Shown Under the Following Headings SeQual Technologies Inc SeQual recommends that this User Manual be given to each patient receiving an ECLIPSE Oxygen System This User Manual describes all the necessary information in order for a patient or caregiver to operate the ECLIPSE WARNING Important safety information for hazards that migh
19. Supply or the DC Power Supply if adequate power is available The Power Cartridge will take longer to recharge when connected to the DC Power Supply and will depend on the flow setting and the amperage switch setting Power Compartment The Power Compartment is available when Power Cartridge operation is not required When using the Power Compartment your ECLIPSE will ONLY operate on the AC and DC Power Supplies ECLIPSE Model 1000 User Manual 33 OTER Accessories continued AC Power Supply Model 6104 Your AC Power Supply enables the ECLIPSE to be connected to a grounded AC electrical outlet with the country specific power cord Please contact your dealer provider for the country specific power cord DC Power Supply Model 6105 Your DC Power Supply enables the ECLIPSE to be operated in an automobile or other vehicle with an accessory outlet Contact your dealer provider to set up the DC Power Supply for proper use in your own automobile The indicator on the AC and DC Power Supplies will be either green or red Green indicates normal operation and red indicates abnormal operation DO NOT expose the AC and DC Power Supplies to water Electrical shock or serious injury may result Cart Your ECLIPSE Oxygen System can be easily transported while attached to the Cart The cart has wheels and a telescoping handle for easy mobility DO NOT drip spill or expose the ECLIPSE to water during portable use The ECLIPSE encl
20. Y OR ON THE ORDER OF A PHYSICIAN THE ECLIPSE IS NOT INTENDED FOR LIFE SUPPORTING OR LIFE SUSTAINING APPLICATIONS NOR DOES IT PROVIDE ANY PATIENT MONITORING CAPABILITIES IN CERTAIN CIRCUMSTANCES THE USE OF NON PRESCRIBED OXYGEN CAN BE HAZARDOUS THIS DEVICE SHOULD ONLY BE USED WHEN PRESCRIBED BY A PHYSICIAN FOR ADULT USE ONLY NOT QUALIFIED FOR USE ON PEDIATRIC PATIENTS WARNING NOT FOR USE IN THE PRESENCE OF FLAMMABLE ANESTHETICS AS WITH ANY ELECTRICALLY POWERED DEVICE THE USER MAY EXPERIENCE PERIODS OF NON OPERATION AS THE RESULT OF ELECTRICAL POWER INTERRUPTION OR THE NEED TO HAVE THE OXYGEN CONCENTRATOR SERVICED BY A TECHNICIAN AN OXYGEN CONCENTRATOR IS NOT APPROPRIATE FOR ANY PATIENT WHO WOULD EXPERIENCE ADVERSE HEALTH CONSEQUENCES AS THE RESULT OF SUCH TEMPORARY INTERRUPTION No Smoking or Naked Flames The use of oxygen therapy requires that special care be taken to reduce the risk of fire Oxygen enhances the combustion of flammable material For safety concerns all possible sources of ignition must be kept away from the oxygen system and preferably out of the room in which it is being used Smoking in the proximity of an operating oxygen concentrator is dangerous and can permanently damage the device and void the warranty 40 ECLIPSE Model 1000 User Manual Model 1000 User Manual WARNING Use No Oil or Grease A spontaneous and viole
21. ae will appear on the control panel CAUTION The AC Power Cord used with the ECLIPSE Oxygen System must meet the requirements of the country where the product is used CAUTION Use with grounded 3 pin receptacle DO NOT remove ground pin from cord Use only with SeQual supplied cords CAUTION DO NOT connect the ECLIPSE to an extension cord or electrical outlet controlled by a switch 2 DC POWER SUPPLY A DC Power Supply allows the system to operate from DC accessory outlets such as those found in motor vehicles You need to determine the Fuse rating in your motor vehicle either 15 AMP or 20 AMP Consult your vehicle s Operator s Manual or look at your fuse box in the vehicle Set the switch on the DC Power Supply to the appropriate one Use the 10 AMP position on the DC Power Supply for a 15 AMP vehicle fuse and use the 15 AMP position for a 20 AMP or greater vehicle fuse Start your vehicle Insert the DC cord in the proper receptacle on the side of the ECLIPSE Next insert the DC Power Supply plug into the DC accessory outlet in your vehicle When the system is properly plugged in and receiving power from the DC source a green indicator light on the Power Supply and the External Power Present indicator a on the control panel will appear ECLIPSE Model 1000 User Manual 19 OTER Operating Instructions continued WARNING Secure the ECLIPSE and the DC Power Supply in your vehicle
22. ceed 50 feet 15 2m when using Continuous Flow Mode and must not exceed 7 feet 2 1m when operating in Pulse Flow Mode for adequate breath detection CAUTION When operating the ECLIPSE on the cart DO NOT bang drop or bump the unit while in transport CAUTION The filter should be free of liquid water and dry to the touch before reinstallation as excess moisture may impair the proper operation of the oxygen system CAUTION When the filter is removed the Service Port is exposed DO NOT attach any cables to this Service Port It is for Service ONLY CAUTION DO NOT poke objects through the vent holes ECLIPSE Model 1000 User Manual 47 Electromagnetic Compatibility This equipment has been tested and found to comply with the limits for medical devices to the IEC60601 1 2 2001 09 Electromagnetic Compatibility standard These limits are designed to provide reasonable protection against harmful interference in a typical medical installation This equipment generates uses and can radiate radio frequency energy and if not installed according with the instructions may cause harmful interference to other devices in the vicinity However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to other devices which can be determined by turning the equipment off and on the user is encouraged to try t
23. d clean it at least once a week 1 Remove the filter 2 Wash the filter in warm water using a mild detergent solution 3 Rinse the filter thoroughly and squeeze out the excess water 4 Allow the filter to air dry 5 Reinsert the filter in the cabinet ECLIPSE Model 1000 User Manual 35 OTE CAUTION The filter should be free of liquid water and dry to the touch before reinstallation as excess moisture may impair the proper operation of the oxygen system CAUTION When the filter is removed the Service Port is exposed DO NOT attach any cables to this Service Port It is for Service ONLY CAUTION DO NOT poke objects through the vent holes The air inlet filter should be replaced as needed If the oxygen system is used in a dusty environment the filter may need to be replaced more often You should not operate the ECLIPSE without a filter installed for longer than 30 minutes Cabinet AC Power Supply DC Power Supply and Power Cord Turn OFF the ECLIPSE and disconnect from AC or DC power before any cleaning or disinfection activity Use a damp not soaking wet cloth or sponge and a mild detergent solution to clean the cabinet and power supplies To disinfect the ECLIPSE use Lysol Brand II disinfectant Proceed as directed by the manufacturer Unplug Power Cords AC or DC Power Supplies before cleaning the exterior cabinet DO NOT use denatured alcohol or apply liqu
24. disassemble crush or heat the Power Cartridge above 140 F 60 C The Power Cartridge may present a risk of fire or explosion and will void the warranty WARNING DO NOT expose the AC and DC Power Supplies to water Electrical shock or serious injury may result WARNING DO NOT disassemble the ECLIPSE or attempt to perform any routine maintenance other than what is described in the Patient Cleaning and Care section below Disassembling the oxygen system may create a hazard of electrical shock and void your warranty Contact your dealer provider for any maintenance required Unplug Power Cords AC or DC Power Supplies before cleaning the exterior cabinet DO NOT use denatured alcohol or apply liquid spray or aerosol cleaners WARNING DO NOT tamper with or try to repair the Power Cartridge No serviceable parts inside WARNING Exposing the Power Cartridge to water or other liquids may cause personal injury WARNING DO NOT drip spill or expose the ECLIPSE to water during portable use The ECLIPSE enclosure does not protect against the harmful effects of liquid ingress Electrical shock or damage to the unit may result 44 ECLIPSE Model 1000 User Manual Model 1000 User Manual CAUTION Availability of an alternate source of oxygen is recommended in case of power outage or mechanical failure CAUTION DONOT expose unit to temperature o
25. ep after 15 seconds if no inspiration is detected If no inspiration detected 15 seconds after that the ECLIPSE will automatically switch to Continuous Flow Mode u Power Cartridge Shutdown Off Off Blinking On Seek external power source 5 or hot Power Cartridge with If Power Cartridge malfunction E no external power present persists contact dealer provider a A Loss of External Power Off Off Blinking On Seek an alternative oxygen without charged Power supply until external power is z 3 r i Cartridge installed restored or install a charged 2 Power Cartridge 5 System Malfunction Off Off On On Contact dealer provider Use N back up oxygen supply AVAILABILITY OF AN ALTERNATE SOURCE OF OXYGEN IS RECOMMENDED IN CASE OF POWER OUTAGE OR MECHANICAL FAILURE CONSULT YOUR PHYSICIAN FOR THE TYPE OF BACK UP SYSTEM REQUIRED DO NOT IGNORE ALARMS 30 ECLIPSE Model 1000 User Manual Model 1000 User Manual Symptom ECLIPSE does not power on when ON OFF button is pressed No Oxygen Low Oxygen Concentration Low Oxygen Flow No Oxygen delivered in Pulse Mode Power Cartridge Status Gauge never indicates fully charged Probable Cause No Power Cartridge installed Remedy Install charged Power Cartridge or plug into external power Power Cartridge is discharged Plug into external power Malfunction Blocked or kinked tubing Filter Blocked
26. gen purity is within specification After initially turning ON the unit allow up to five 5 minutes for the unit to reach its specified performance The control panel will indicate when the unit is done warming up When the green ON indicator is lit the system is ready for use Green light Yellow light Red light ready alert alarm UUWUUWUUU0 4 26 ECLIPSE Model 1000 User Manual Model 1000 User Manual STEP 3 CONNECT THE TUBING TO THE OXYGEN OUTLET AND THE CANNULA PER THE INSTRUCTIONS FROM YOUR DEALER PROVIDER m Replace the cannula and tubing on a regular basis Consult your dealer provider for cannula and tubing cleaning and replacement information STEP 4 SELECT FLOW MODE m Press the Flow Mode button to select the flow delivery mode Continuous Flow Mode Operation e When operating in Continuous Flow Mode a continuous supply of oxygen will flow through your tubing and nasal cannula When operating in Continuous Flow Mode to ensure adequate flow DO NOT use a tubing cannula extension more than 50 15 2m long ECLIPSE Model 1000 User Manual 27 OTE Pulse Flow Mode Operation e Pulse Flow Mode should only be used under the direction of a qualified clinician or physician e When operating in Pulse Flow Mode an alarm will beep after 15 seconds if inspiration is not detected If another 15 seconds elapses and no inspiration is detected t
27. he unit will switch to Continuous Flow Mode at the last Continuous Flow Mode rate setting used Pulse Flow settings should be determined for each patient individually Settings from Continuous Flow applications may not be applicable to Pulse Flow Mode When operating in Pulse Flow Mode to ensure adequate breath detection the length of the tubing cannula must not exceed 7 feet 2 1m WARNING As with all conserving devices the ECLIPSE may not be able to detect some respiratory efforts STEP 5 ADJUST FLOW RATE TO PRESCRIBED LEVEL Using the Flow Rate Setting buttons adjust the flow rate to the prescribed setting supplied by your physician DO NOT change from prescribed flow rate setting without consulting with a qualified clinician or physician 28 ECLIPSE Model 1000 User Manual Model 1000 User Manual STEP 6 BEGIN USE m Breathe normally through the cannula m You will get longer operating time on your Power Cartridge if you operate your ECLIPSE in Pulse Flow STEP 7 CART OPERATION m The Cart is designed for use on pedestrian surfaces m Align the locator holes on the bottom of the unit to the tabs on the cart m Align the screw to the insert on the back of the ECLIPSE and hand tighten the knob on the cart Push the button on the cart handle to adjust the height STEP 8 TURN OFF THE UNIT m Press and hold the ON OFF Button for two 2 seconds to turn OFF your
28. id spray or aerosol cleaners Humidifier If used clean your humidifier daily to reduce possibility of contamination Replace humidifier monthly or as recommended by the manufacturer or your dealer provider Cart Before cleaning the cart remove the ECLIPSE Use a damp not soaking wet cloth or sponge and mild detergent solution to clean the Cart 36 ECLIPSE Model 1000 User Manual Model 1000 User Manual Power Cartridge Cleaning Care and Maintenance The Power Cartridge in the ECLIPSE Oxygen System requires special care to assure a longer life and higher performance The SeQual Power Cartridge is the only approved Power Cartridge to be used Once a month completely discharge your Power Cartridge and fully recharge to calibrate the Power Cartridge Use a damp cloth or sponge and mild detergent to clean the Power Cartridge case and latch DO NOT tamper with or try to repair the Power Cartridge No serviceable parts inside Power Cartridge Disposal Your Power Cartridge is rechargeable and can be recycled Always return it to your dealer provider for proper disposal Power Cartridge Storage Your Power Cartridge should be stored in a cool dry location Exposing the Power Cartridge to water or other liquids may cause personal injury ECLIPSE Model 1000 User Manual 37 Technical Data Flow Rate Settings 1 to 6 setting in Pulse Flow Mode 0 5 to 3 0 LPM in Continuous Flow Mode
29. ion This will help to assure the longevity of your Power Cartridge CAUTION DOT UN Regulations require the removal of the Power Cartridge for international travel CAUTION There are no serviceable parts inside the Power Cartridge DO NOT open CAUTION Replace the Power Cartridge with only SeQual approved Power Cartridge CAUTION The Power Cartridge used in this device may present a risk of fire or chemical burn if mistreated DO NOT disassemble incinerate or heat above 140 F 60 C CAUTION Return your Power Cartridge to your dealer provider for proper disposal CAUTION Keep your Power Cartridge away from children 22 ECLIPSE Model 1000 User Manual Model 1000 User Manual Typical Power Cartridge Recharge Time The typical time to recharge your Power Cartridge to achieve 80 capacity from a fully discharged Power Cartridge is 1 4 hours to 5 0 hours dependent upon the flow setting If the Power Cartridge is too warm charging will not begin until it sufficiently cools The Power Cartridge may not charge or the charge time may increase when operating from the DC Power Supply set in the 10 AMP position INITIAL POWER CARTRIDGE OPERATION The new Power Cartridge packaged with your ECLIPSE Oxygen System is not fully charged Before using your ECLIPSE Oxygen System for the first time you must fully discharge the Power Cartridge 1 Turn y
30. is dangerous and can permanently damage the device and void the warranty Use No Oil or Grease Indicates important information about the SS possibility of spontaneous and violent ignition that may occur if oil grease or other petroleum based substances come into contact with oxygen under pressure PXO Drip Proof Equipment IPX0 The ECLIPSE enclosure does not provide protection against the harmful effects of liquid ingress IPXO per IEC 60529 PX1 Drip Proof Equipment IPX1 The ECLIPSE AC and DC Power Supplies provide protection against the harmful effects of liquid ingress IPX1 per IEC 60529 a IEC Icon for DC Power Input Type B Equipment Equipment providing a particular degree of IK protection against electric shock particularly regarding Allowable LEAKAGE CURRENT Reliability of the protective earth connection A Amperes ECLIPSE Model 1000 User Manual 3 OTE Symbols continued Off e QqDo Bb Alternating Current Audible Alarm An audible indicator is provided to announce a reduction in oxygen concentration or when the power has been interrupted or lost ON OFF Turns the unit ON and OFF but does not directly disconnect the mains power Increase Flow Setting Increases the flow setting in 0 5 increments each time the button is pushed Decrease Flow Setting Decreases the flow setting in 0 5 increments each time the b
31. l Power Cart Insert Screw Receptacle Exhaust Vent Power Cartridge Latch Main Rating Label FRONT Under Base Power Cartridge BACK INPUT OUTPUT CONNECTIONS Oxygen Outlet Port Extension tubing or a cannula is attached to this port Air Inlet Filter Ambient air is drawn into the unit through the air inlet located on the top back of the unit This filter prevents dust and debris from entering the ECLIPSE and should be cleaned regularly Service Port This connection allows service personnel to diagnose and monitor ECLIPSE performance It is not for patient use External Power Receptacle Either the ECLIPSE AC or DC Power Supply output cords may be connected to this receptacle Exhaust Vent Exhaust air from the ECLIPSE is expelled from this vent ECLIPSE Model 1000 User Manual 11 Understanding the Controls and Front Panel continued Control Panel Features The ECLIPSE displays important operating information using the icons numbers and visual indicators on the control panel This section will help you understand these features No Smoking Icon ON OFF Button ON OFF Indicator Green Alert Indicator Yellow Flow Rate Setting 2 Digits Green Alarm Indicator Red Increase Flow Rate Button Decrease Flow Rate Button Power Cartridge Status Gauge Green Ex
32. nt ignition may occur if oil grease or other petroleum substances come into contact SS with oxygen under pressure Keep these substances away from the oxygen system tubing and connections and any other oxygen source DO NOT use any petroleum based or other lubricants WARNING DO NOT remove the cover no serviceable parts inside Qualified service personnel should only remove the cover WARNING DO NOT operate device or accessories on wet surfaces or in standing water and DO NOT submerse or expose to water The ECLIPSE enclosure does not provide protection against the harmful effects of liquid ingress Electrical shock or damage to the unit may result WARNING Protect electrical cords from sharp edges Electrical shock and serious physical injury may result WARNING Use only the SeQual AC Power Supply and DC Power Supply supplied with your ECLIPSE Oxygen System Use of a non SeQual Power Supply may be hazardous cause damage to your oxygen system and will void the warranty WARNING This Power Cartridge may explode and cause potential injury if exposed to or disposed of in a fire WARNING DO NOT short circuit the Power Cartridge s metal contacts with metallic objects such as keys or coins It may cause sparks or excessive heat DO NOT disassemble puncture or crush the Power Cartridge Power Cartridge electrolytes may be toxic if swallowed and can be harmful to skin
33. o correct the interference by one or more of the following measures m Reorient or relocate the receiving device E Increase the separation distance between the equipment m Connect the equipment into an outlet on a circuit different from that which the other device s are connected m Consult the manufacturer or service technician for help Guidance and Manufacturer s declaration electromagnetic emissions The SeQual Technologies ECLIPSE Oxygen System is intended for use in the electromagnetic environment specified below The customer or the user of the SeQual Technologies ECLIPSE Oxygen System should assure that it is used in such an environment Emissions Test Compliance Electromagnetic Environment Guidance RF emissions CISPR 11 Group 1 The SeQual Technologies ECLIPSE Oxygen System uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B The SeQual Technologies ECLIPSE Oxygen System is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings Voltage fluctuations Complies used for domestic purposes flicker emissions IEC 61000 3 3 Harmonic emissions Class A IEC 61000 3 2 48 ECLIPSE Model 1000 User Manual Model 1000 User Manual Disposal of Equipmen
34. odel 6104 Power Cartridge Model 2400 There are many different types of oxygen tubing cannula and humidifiers The following accessories are recommended to be used with your ECLIPSE Oxygen System Refer to the Maintenance section of this manual for accessory maintenance instructions Humidifier Salter Labs Bubble Humidifier Part Number 7600 or equivalent If your physician has prescribed a humidifier follow the manufacturer s instructions for use Attach the humidifier to the oxygen outlet port of the oxygen system The O symbol marks the oxygen outlet port on the concentrator Use of humidifiers not specified for this oxygen system may impair performance and void the warranty DO NOT use a humidifier in Pulse Flow Mode The ECLIPSE will not detect an inspiration and will default to Continuous Flow Mode Humidifier Adapter SeQual Part Number 3509 If your physician has prescribed a humidifier you may need to use the SeQual Humidifier Adapter Follow the instructions for use Attach the Humidifier Adapter to the oxygen outlet port of the oxygen system and then to the humidifier Attach cannula or extension tubing to humidifier outlet 32 ECLIPSE Model 1000 User Manual Model 1000 User Manual Oxygen Tubing Salter Labs Series 2000 or equivalent tubing internal diameter of the tubing should be no less than 3 16 0 48 cm Connect the oxygen tubing to the outlet port of the humidifier or directly to the oxygen o
35. osure does not protect against the harmful effects of liquid ingress Electrical shock or damage to the unit may result CAUTION When operating the ECLIPSE on the cart DO NOT bang drop or bump the unit while in transport 34 ECLIPSE Model 1000 User Manual Model 1000 User Manual Cleaning Care and Routine Maintenance Routine Maintenance A trained technician will carry out all routine maintenance at least once a year or as needed Your dealer provider will perform inspections and maintenance of the compressor filters alarms 9V battery Power Cartridge and other internal parts as needed DO NOT disassemble the ECLIPSE or attempt to perform any routine maintenance other than what is described in the Patient Cleaning and Care section below Disassembling the oxygen system may create a hazard of electrical shock and void your warranty Contact your dealer provider for any maintenance required Patient Cleaning and Care Cannula Replacement Replace your tubing and cannula on a regular basis as recommended by your dealer provider Your physician or dealer provider will provide you with cleaning disinfection and replacement information Air Inlet Filter Ambient air is drawn into the unit through the air inlet located on the top back of the unit Cleaning the air inlet filter is the most important activity you will perform to keep your oxygen system performing properly Check this air filter daily an
36. our ECLIPSE ON and allow the Power Cartridge to completely discharge Unit will sound alow Power Cartridge alert until the unit completely shuts down This could continue for up to approximately 20 minutes 2 Connect the AC Power Supply and plug in the AC Power Cord to completely charge the Power Cartridge This may require up to five hours of uninterrupted charging This completes the calibration of the Power Cartridge ECLIPSE Model 1000 User Manual 23 OTER Operating Instructions continued WARNING DO NOT tamper with disassemble crush or heat the Power Cartridge above 140 F 60 C The Power Cartridge may present a risk of fire or explosion and will void the warranty CAUTION Store the Power Cartridge in a cool dry place when not in use CAUTION DO NOT leave your ECLIPSE or your Power Cartridge in your vehicle or in the trunk during a hot day CAUTION The ECLIPSE system can only work with a SeQual Power Cartridge Use of another Power Cartridge or Battery may damage the unit present a risk of fire or explosion and will void the warranty In the event of a power interruption the ECLIPSE Oxygen System will automatically switch to the Power Cartridge operation When power is restored the Power Cartridge will automatically start recharging If the Power Cartridge is not present during a power interruption the ECLIPSE Oxygen System will stop operation When power is restored
37. rns all possible sources of ignition must be kept away from the oxygen system and preferably out of the room in which it is being used Smoking in the WARNING proximity of an operating oxygen concentrator is dangerous and can permanently damage the device and void the warranty Use No Oil or Grease A spontaneous and violent ignition may occur if oil grease or other petroleum substances come into contact with oxygen under pressure Keep these substances away from the oxygen system tubing and connections and any other oxygen source DO NOT use any petroleum based or other lubricants 6 ECLIPSE Model 1000 User Manual Model 1000 User Manual Operating Safety m Keep oxygen system and cord away from hot surfaces DO NOT use the oxygen system if it has a damaged cord or plug contact your dealer provider m Your oxygen system should be located in a well ventilated area to allow for adequate air intake and to avoid the intake of airborne pollutants and fumes ONLY use this device with accessories designed or recommended by SeQual Technologies Inc m f your oxygen system has been dropped damaged or exposed to water contact your dealer provider for examination or repair DONOT drop or insert any object into any opening DONOT block the air inlet or the exhaust vent of the ECLIPSE when it is on a soft surface such as a bed or couch or car seat DONOT cover unit with a blanket towel
38. roximately 10 of the total Power Cartridge charge When the Power Cartridge is being charged the charge indicator bars will blink in a waterfall fashion If the Power Cartridge is not installed or if it is improperly installed the Status Gauge will not light External Power Present Indicator When the ECLIPSE system is properly plugged in using the AC or DC Power Supply this indicator will appear on the Control Panel 14 ECLIPSE Model 1000 User Manual Model 1000 User Manual Recommended Operating Environments The following information recommends certain boundaries that will allow the ECLIPSE Oxygen System to operate at its full potential Operating Temperature 50 F to 104 F 10 C to 40 C Operating Humidity 10 95 at an 82 4 F 28 C dew point Transport Storage Temperature 4 F to 140 F 20 C to 60 C Humidity Up to 95 Non Condensing Electrical Use no extension cords or electrical outlets controlled by a switch Altitude O 13 123 feet 0 4000 meters Placement Do Not block air inlet or bottom exhaust vent keep at least 3 inches away from walls draperies furniture etc Environment Smoke pollutant and fume free Operating Time Up to 24 hours per day when connected to external power When moving the ECLIPSE from an extreme environment allow time for your ECLIPSE to acclimate to the recommend operating environment WARNING Operating your ECLIP
39. t and Accessories Follow local governing ordinances and recycling plans regarding disposal of device components DO NOT dispose of the Power Cartridge Power Cartridges should be returned to your dealer provider MANUFACTURED BY EUROPEAN UNION AUTHORIZED AGENT SeQual Technologies Inc MeddiQuest Limited 11436 Sorrento Valley Road Business amp Technology Centre San Diego CA 92121 USA Bessemer Drive 858 202 3100 Stevenage SG1 2DX 800 826 4610 United Kingdom www sequal com Tel 44 1763 222505 Fax 44 1763 222049 DEALER PROVIDER NAME AND ADDRESS PHYSICIAN NAME AND ADDRESS SeQual and ECLIPSE are trademarks of SeQual Technologies Inc San Diego CA 92121 Salter Labs is a registered mark of Salter Labs Arvin CA 92303 Lysol is a registered trademark of Reckitt Benckiser UK ECLIPSE Model 1000 User Manual 49 ECLIPSE Model 1000 User Manual PN 6112 03 06
40. t cause serious injury CAUTION Important information for preventing damage to the ECLIPSE Oxygen System DO NOT Information to which you should pay special attention Important Clarifying information specific instructions commentary sidelights or interesting points Symbols Used in User Manual and on Oxygen System Refer to this section for an explanation of the symbols and the warnings that accompany them Read Instructions This symbol when used alone or in conjunction Ci with any of the following symbols indicates the need to consult the operating instructions provided with the product A potential risk exists if the operating instructions are not followed symbol indicates important information that requires the user to take f Attention or Caution Consult accompanying documents This special precautions when certain conditions are present 2 ECLIPSE Model 1000 User Manual Model 1000 User Manual No Smoking or Naked Flames The use of oxygen therapy requires that special care be taken to reduce the risk of fire Materials that burn in air and even some materials that will not burn in air are easily ignited and burn rapidly in high concentrations of oxygen For safety concerns all possible sources of ignition must be kept away from the oxygen concentrator and preferably out of the room in which itis being used Smoking in the proximity of an operating oxygen concentrator
41. tem will automatically switch over to the Power Cartridge if installed When external power is restored by plugging into AC or DC outlets the system will accept power from the external source and replenish the Power Cartridge while operating if adequate power is available Connect the AC or DC Power Supply by aligning the arrow on the plug locking ring with the white dot on the receptacle of the unit Push until the locking ring twists slightly counter clockwise then snaps back into alignment with the dot on the receptacle Disconnect by grasping the plug twist counter clockwise and pull Connect the AC Power Cord to a grounded power outlet 18 ECLIPSE Model 1000 User Manual Model 1000 User Manual DO NOT use any cord if damaged Improper use of the Power Cord may result in a fire or electrical shock hazard DO NOT touch the Power Cartridge cavity connectors with the Power Cartridge removed while the unit is connected to an external power source This may result in electrical shock 1 AC POWER SUPPLY The ECLIPSE System includes a universal AC Power Supply for use at home or where standard AC power is available To plug into AC make sure the AC Power Supply cord is securely inserted in the power receptacle on the side of the unit and plug the AC Power Supply into an AC outlet When the unit is plugged in properly a green indicator on the AC Power Supply will light and the External Power Present Indicator
42. ternal Power Present Indicator Green Pulse Flow Mode Indicator Green Flow Mode Button 12 ECLIPSE Model 1000 User Manual Model 1000 User Manual User Controls This section describes the control buttons and function for your ECLIPSE Oxygen System FRONT PANEL ICONS ON OFF Button and Indicator This button turns the system ON or OFF The Green Indicator is lit when the unit is ON Increase or Decrease Flow Rate Button Use these buttons to set the flow rate to your prescribed level Flow Mode Button and Indicator The ECLIPSE contains a button to toggle between Continuous Flow and Pulse Flow mode Pulse Flow mode allows a significant increase in the amount of time the system can operate from the Power Cartridge When the Pulse Flow Mode is activated the green Flow Mode indicator goes on and a pulse of oxygen is delivered upon inhalation When the pulse of oxygen is delivered the indicator briefly goes off Only change your flow settings at the recommendation of your physician SYSTEM STATUS INDICATORS The ECLIPSE control panel displays important operating information This section will help you understand these features Green Yellow and Red indicators on the front panel indicate the operating condition of the system For more information on the indicators see the Alarm Conditions and Troubleshooting table ly OA ON OFF Green Indicator System OK Indicates that the ECLIPSE is ON and operating properly
43. utlet port if you do not use a humidifier Connect the other end of the tube to the nasal cannula if not already attached Tubing not specified for use with this oxygen system may impair the performance BGR The tube fitting may be tight DO NOT use oil or grease to lubricate WARNING lf tubing becomes disconnected during operation ECLIPSE does not alarm Check flow at the cannula to verify oxygen delivery CAUTION To ensure there is adequate flow to deliver oxygen the length of the tubing must not exceed 50 feet 15 2m when using Continuous Flow Mode and must not exceed 7 feet 2 1m when operating in Pulse Flow Mode for adequate breath detection Cannula Salter Labs Part Number 1600 Series or equivalent Your physician will have prescribed a cannula to deliver oxygen In most cases they are already attached to the oxygen tubing If not follow the instructions included with the cannula to attach it to the oxygen tubing Use of cannula not specified for use with this oxygen system may impair the performance Power Options You have various options to provide power to your ECLIPSE A Power Cartridge AC Power Supply and DC Power Supply Power Cartridge Model 2400 The rechargeable Power Cartridge supplied with your ECLIPSE is capable of providing you with battery operation depending on your prescribed flow rate Your Power Cartridge will be recharged when the ECLIPSE is connected to the AC Power
44. utside of specified operating or storage temperatures as this may damage the unit CAUTION Locate oxygen tubing and Power Supply cords to prevent tripping hazards CAUTION DONOT place ECLIPSE in a small enclosed space such as a bathroom or closet with cannula leading out of the enclosed space CAUTION DO NOT expose the Power Cartridge to temperatures above 140 F 60 C such as in a vehicle parked in the sun or on a hot day CAUTION Charging the Power Cartridge below 50 F 10 C or above 104 F 40 C may degrade the performance CAUTION DO NOT drop the Power Cartridge or expose it to mechanical shock CAUTION Only use the Power Cartridge for its intended purpose CAUTION Locate the unit in a well ventilated room that provides adequate air flow and check if air is being prevented from entering or exiting the system by furniture draperies bedding clothing or pets CAUTION Always check to see that the Air Inlet and the Exhaust Vent are not blocked and the air inlet filter is clean before using your system ECLIPSE Model 1000 User Manual 45 OTET CAUTION The AC power cord used with the ECLIPSE Oxygen System must meet the requirements of the country where the product is used CAUTION Use with grounded 3 pin receptacle DO NOT remove ground pin from cord Use only with SeQual supplied cords CAUTIO
45. utton is pushed Flow Mode Activates either Continuous Mode or Pulse Flow Mode Operation External Power Present Indicator Indicates the presence of external power to operate the concentrator Power Cartridge Status Shows the amount of charge remaining in the Power Cartridge Power Cartridge Symbol The symbol on the Power Cartridge pack Oxygen Output The port in which the oxygen flows out and where the tubing is attached 4 ECLIPSE Model 1000 User Manual Model 1000 User Manual Indications for Use THE ECLIPSE OXYGEN SYSTEM IS INDICATED FOR THE ADMINISTRATION OF SUPPLEMENTAL OXYGEN THE DEVICE IS NOT INTENDED FOR LIFE SUPPORT NOR DOES IT PROVIDE ANY PATIENT MONITORING CAPABILITIES YOUR PHYSICIAN HAS PRESCRIBED A SPECIFIC OXYGEN FLOW RATE SETTING TO MEET YOUR INDIVIDUAL NEEDS RECOMMENDED OXYGEN FLOW RATES SHOULD BE ADJUSTED ONLY UNDER THE ADVICE OF A PHYSICIAN FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN Contraindications THE ECLIPSE IS NOT INTENDED FOR LIFE SUPPORTING OR LIFE SUSTAINING APPLICATIONS NOR DOES IT PROVIDE ANY PATIENT MONITORING CAPABILITIES IN CERTAIN CIRCUMSTANCES THE USE OF NON PRESCRIBED OXYGEN CAN BE HAZARDOUS THIS DEVICE SHOULD ONLY BE USED WHEN PRESCRIBED BY A PHYSICIAN WARNING FOR ADULT USE ONLY NOT QUALIFIED FOR USE ON PEDIATRIC PATIENTS WARNING NOT FOR USE IN THE PR
Download Pdf Manuals
Related Search
Related Contents
CLH 7482 BLH 7502 / BLH 7602 / BLH 7603 BPH 7502 Printing Barcodes User Guide Manuel d`utilisation Générateur de courant, Essence - BC Omega Manual - Thompson Center Arms SAS to SAS/SATA RAID Subsystem User Manual Sony MZ-E40 User's Manual Sharkoon Side Panel closed Informações Essenciais para Aerospray Hematology" Slide Stainer Coby CVE196 headphone Copyright © All rights reserved.