Home
Invacare®ScanBed 750TM
Contents
1. 1421555 Dacapo Basic 38 kg m cold foam incontinence Cover 83x183x12 1421556 Dacapo Basic Heavy user 3 coated 50 50 38 kg m cold foam incontinence cover 105x200x15 484429 Dacapo Basic Heavy user 3 coated 50 50 38 kg m cold foam incontinence cover 120x200x15 1451262 Dacapo Organic 50 kg m Organic foam egg tray shape incontinence cover 85x200x12 1421572 Dacapo Organic 50 kg m Organic foam egg tray shape incontinence cover 90x200x12 1421573 Dacapo Combi 2 coated 50 38 kg m Visco elastic cold foam incontinence cover 85x200x14 1421489 Dacapo Combi 2 coated 50 38 kg m Visco elastic cold foam incontinence cover 90x200xI4 1421490 Dacapo Square 2 coated 50 38 kg m Organic foam cold foam incontinence cover 83x200x14 1421515 Dacapo Square 2 coated 50 38 kg m Organic foam cold foam incontinence cover 88x200x14 1421516 Dacapo Top 50 kg m Visco elastic foam incontinence cover 85x200x6 1421587 Dacapo Top 50 kg m Visco elastic foam incontinence cover 90x200x6 1421588 It is possible to buy custom fit insertion mattresses for extended beds LOW SLUNG SET Low slung set with 75 mm braking castotr ovde ome rur ar ch ee Ee EE 021343 D0 SIDE RAIL COVERS Scala Cover for Britt III IV and Lilli II wooden side rail e RR RII 1449412
2. Invacare ScanBed 750 User Manual CE Quality Declaration Congratulations with your new bed Invacare ScanBed 750 from Invacare EC H ng A S Your new bed is C marked in accordance with directive 93 42 EEC concerning medical devices Invacare ScanBed 750 is developed and constructed with consideration for the user and others handling or assisting at the bed Furthermore the bed is developed in accordance with the European Standard EN 1970 Invacare ScanBed 750 is supervised and quality controlled throughout the entire production process and the finished bed is inspected by our finished goods control Identification label and QA mark are located on the bed confirming that the finished goods control has approved the bed Please read the entire user s manual before using the bed Invacare EC Hong A S is certified according to ISO 9001 and ISO 13485 For and on behalf of Invacare EC Hong A S Brian Mundeling R amp D Manager DK KVALITETSDEKLARATION Tillykke med Deres nye seng Invacare ScanBed 750 fra Invacare EC Hong A S Deres nye seng er CE market og lever op til alle krav i henhold til direktiv 93 42 E F om medicinske anordninger Sengen er udviklet og konstrueret under st rst mulig hensyntagen til brugeren samt alle andre der enten h ndterer sengen eller hj lper til ved sengen Invacare ScanBed 750 er udviklet under hensyntagen til de sikkerhedsm ssige krav i den Europ iske
3. Scala Cover for Britt Ill wooden side rail Short ces 144941 I Scala Cover for Britt Il and Lilli Ill wooden side rail extended eene 1493565 Scala Cover for Britt III IV and Lilli Ill wooden side rail net een ee nena 1449413 Scala Cover for Britt Il and Lilli Ill wooden side rail net extended eee 1449415 Scala Cover for metal side rail Baeic eee hs 1449423 Scala Cover for metal side rail increased lees 1449419 Scala Cover for metal side rail extended ee ee ee ee ss 1449421 Safety Cover for Britt Ill IV and Lilli lll wooden side rail 1511890 Safety Cover for Britt Ill wooden side rail short 1513592 Safety Cover for Britt Il and Lilli Ill wooden side rail extended 2 0 cece eee eee 1511892 20 9 Maintenance and check ups Service and maintenance of the SB750 necessary instruction or training must only be performed by personnel who have received the With normal daily use service must be carried out according to the service schedule after 2 years use and thereafter every second year When moving the bed service must be carried out according to the service schedule When using Flex or Karin bed ends which are made of a flexible type of wood it will be necessary to tighten the hexagon head screws at intervals other than described in the service schedule Note The top frame must be supported
4. 8 25 ERI DI TTTT PPPIBMPTI 26 I6 IDISDOSAls 606505 ee eee ee teow 343 9 9 9 EE 26 User part Congratulations on your choice of Invacare EC H ng s B750 nursing bed e The bed has been developed for patients over 2 years of age requiring care in a nursing home or home care nursing e The bed combines excellent stability and ergonomic design with easy disassembly and operation e The unique backrest return of the bed provides the patient with excellent comfort and the carer with ideal conditions for an ergonomically correct working position In order to optimise the patient s comfort Invacare EC Heng A S recommends using a 2 cm mattress General information The SB750 bed fulfils all requirements regarding maximum distances However if the bed is used for the care of patients with small body dimensions it must be especially noted that there is a risk of such a patient slipping through the openings between the side rails or through the opening between the side rail and the mattress support The bed must not be used by patients under 12 years of age or by patients with body size equivalent to an average 2 year old or smaller The bed in combination with side rails must not be used for persons under 45 kg or under 150 cm s height or for persons that are restless spasms or confused unless e a professional risk assessment has taken place and been accepted or e acorrectly installed safety cover on the side
5. accessories 3 Run the bed to the marked height level so that the cams on the base frame s guide way and the shear arms are opposite each other 4 Disconnect the power supply 5 Remove the locking cam from the control unit 6 Disconnect the plugs of backrest thigh and leg motors and check that the cables are not caught in the base frame 7 Remove the 2x2 finger screws in the side bars 8 Pull the head top frame out of the inserts in the side bars to stop and lower the section onto the floor 9 Release the 2 lock fittings snap locks on the foot top frame by pulling out the locking pin and turning it to the right 10 Lower the foot section of the top frame and lift it off the base frame Remove the head section of the top frame from the base frame l Connect the base frame to the mains again and lower it completely 15 Assembly Make sure that the castors have their brakes on 2 Connect the base frame to the mains 3 Run the bed to the marked height level so that the cams see picture 3 page 15 on the base frame s guide way and the shear arms are opposite each other 4 Lift the foot end of the top frame so that the two 2 lock fittings snap locks fit onto the pins on the shearing arms 4 1 Make sure that the locking pin has clicked into place and is 4 2 locked 4 2 Then move the foot end of the top frame into horizontal position 5 Place the head end of the top frame in front of the base
6. is CE gecertificeerd en goedgekeurd conform richtlijn 93 42 EEC betreffende medische hulpmiddelen Invacare ScanBed 750 is ontwikkeld en geconstrueerd met inachtneming van de behoeften van de gebruiker en verzorgers Invacare ScanBed 750 is ontwikkeld in overeenstemming met de Europese Standaard EN 1970 Gedurende het gehele productieproces is het ScanBed 750 bed gecontroleerd op kwaliteitsaspecten en het complete bed is na productie zorgvuldig gecontroleerd Het productlabel en de QA markering zijn op het bed bevestigd om aan te tonen dat de bedden zijn gecontroleerd en goed zijn bevonden door de afdeling productcontrole Wij verzoeken u vriendelijk de gehele gebruikershandleiding te lezen voordat u het bed in gebruik neemt Invacare EC H ng A S is ISO 9001 en ISO 13485 gecertificeerd IT DICHIARAZIONE DI QUALIT Complimenti per aver scelto il letto Invacare ScanBed 750 prodotto da Invacare EC H ng A S Il vostro nuovo letto marcato CE ai sensi della Direttiva 93 42 EEC relativa ai dispositivi medici II letto stato progettato e costruito con un occhio di riguardo per gli utilizzatori e per i loro assistenti Invacare ScanBed 750 stato realizzato nel rispetto della Normativa Europea EN 1970 Il letto stato oggetto di accurate verifiche qualitative durante l intero processo produttivo una volta completato stato controllato dal nostro servizio prodotti finiti Un apposita targhetta e la marcatura QA sono s
7. 5539 extended foldable with loose finger screws 50 55390 D0 L R BRITT IV wooden side rail high 2 5 cm beech cee cee ence e 427875 0101 BRITT IV wooden side rail high 2 5 cm cher 1427875 0102 BRITT IV wooden side rail high 2 5 cm white pigmented 0 ccc eee eee eee 1427875 0105 BRITT IIl wooden side rail beech Ae A r gegen eg e kaserner 020434 0 I BRITT Ill wooden side rail short beech cee renere rn vn 021342 0 I BRITT Uvdal 020434 02 BRITT Ill wooden side rail short white pigmented llli 021342 05 BRITT Ill wooden side rail white pigmented avuuuvvvvvv cee renere RII 020434 05 BRITT Il wooden side rail extended beech auuuvuv anne vr revner rever erre hn 019372 0 I BRITT Il wooden side rail extended cherry 0 2 0 0 cece eee 019372 02 BRITT Il wooden side rail extended white pigmented 0 cece eee eee eee eens 019372 05 LILLI III wooden side rail DECC Lerner MANDAR eben al MU MESS TT 1433812 0101 LILLI III wooden side rail cherry sanser TRES RESP PENNA ES PS 1433812 0102 LILLI III wooden side rail extended Deech 1434044 0101 Covering tubes for mattress support exvtenslon 0 cece eee ene hh hh 017639 DO SUPPORT HANDLES Support handie x sus senten od n RIP PR URS i a geg jc E 021964 DO Support hafidle 22s 3d aa a dan EPOR tenant UX5U CIT 24 nea e begin aded EE S 021963 DO SUDDOFE handl
8. Standard EN 1970 Hver seng har gennem hele produktionsforl bet v ret overvaget og kontrolleret og den f rdige seng er blevet inspiceret af vores f rdigvarekontrol Typeskilt og QA m rke p s ttes sengen som dokumentation for at f rdigvare kontrollen har godkendt sengen F r ibrugtagning af Deres seng skal De genneml se brugermanualen grundigt Invacare EC H ng A S er certificeret i henhold til ISO 9001 og ISO 13485 DE QUALITETSDEKLARATION Herzlichen Gl ckwunsch zu Ihrem neuen Pflegebett Invacare ScanBed 750 von Invacare EC H ng A S Ihr neuer Pflegebett ist gem der Richtlinie 93 42 EWG f r Medizinprodukte CE gekennzeichnet Der Invacare ScanBed 750 wurde unter Ber cksichtigung der Bed rfnisse von Benutzer und Pflegepersonal entwickelt und in bereinstimmung mit der Europ ischen Norm EN 1970 konzipiert und hergestellt W hrend des gesamten Herstellungsprozesses unterliegt der Invacare ScanBed 750 einer st ndigen Qualit tskontrolle und wird im Anschluss nochmals durch unsere Endkontrolle gepr ft Das Typenschild und QA Kennzeichen sind am Bett angebracht um die Abnahme durch unsere Qualit tskontrolle zu best tigen Bitte lesen die gesamte Bedienungsanleitung bevor Sie den Bett in Gebrauch nehmen Invacare EC H ng A S ist zertifiziert nach ISO 9001 und ISO 13485 NL KWALITEITSGARANTIE Gefeliciteerd met de aanschaf van uw nieuwe bed ScanBed 750 van Invacare EC Hong A S Invacare ScanBed 750
9. c interference increase the distance between the bed and the products or switch them off This medical bed can be used together with medical electrical equipment connected to the heart intracardially or the blood vessels intravascularly provided that following points are respected The medical bed should be equipped with means for potential equalization connection marked out by a symbol shown in the back of this manual Medical electrical equipment should not be fixed on the bed s metallic accessories such as side rails lifting pole drip rod bed ends ect In addition the medical electrical equipment power supply cord should be kept clear of the accessories or any other moving part of the bed Invacare EC Heng A S accepts no liability for any use change or assembly of the product other than as stated in this User s Manual If the functions of the bed change please send the bed immediately for a check up according to the maintenance chart chapter I I 4 Receiving the SB750 A To avoid condensation the bed should not be used until it has reached a temperature of 10 50 C A Check whether the bed shows any signs of damage If the bed is damaged see terms of delivery If the bed has been delivered in parts it must be assembled pen dp Locked pin Cf chapter 6 Disassembly Assembly of the SB750 snap lock Check that the connection between mattress support and base frame is properly locked cf the pic
10. d I Braking Step on the pedal 2 Releasing the brake Step on the release pedal Operating castors with central braking system When the bed is positioned correctly it must be locked I Braking If the brake is in neutral step on the red pedal 2 Releasing the brake Step on the green pedal until the brake is in neutral position Operating the steerable castor The SB750 with central braking system may be equipped with a steerable castor The steerable castor is operated by means of the central braking pedal Activating the steering If the brake is in neutral step on the green pedal 2 Deactivating the steering Step on the red pedal until the brake is in neutral position Set off from the castors might under special conditions appear at different types of absorbing floor covering including untreated or badly treated floors In matters of doubt Invacare recommends to place a suitable kind of protection between the castors and the floor Operating the BRITT II III IV wooden side rail Locking pins The bed end may be equipped to lock the side rail at half height as E well as in the top position Normally the side rail can only be locked in top position SS MENER Eege IN T US 3 Up Pull up the top wooden side rail bar until che locking pin locks with an audible click l A Down Lift the top wooden side rail bar and press the two locking RN pem pins together Lower the si
11. de rail Lift Release Lower Y Operating the steel side rail Up Lift the top steel bar of the side rail towards the end with the l locking mechanism o a Down Press the release button and pull the top bar of the steel side rail away from the locking mechanism 4 There is risk of entrapment of fingers on assembly and operation of the side rails Ki Please ensure that the side rail locking system has engaged properly by pulling pushing the upper bar Adjusting the height of the lifting pole Loosen the cord as shown in fig A The lifting handle can now be adjusted to the desired height Press the cord together as shown in fig B and check that the cord is locked in the cord lock by pulling the handle Position the lifting pole in such a way that the handle is above the bed If the lifting pole is used A while the handle has been turned away from the bed the bed can tip when the handle is used Technical part Invacare EC Hong A S is certified according to DS EN ISO 9001 and ISO 13485 which ensures that our customers are always supplied with Invacare EC Hang A S products of uniform quality Throughout the entire production process our materials and products are quality controlled by the operators Furthermore there is a final test when the product is fully assembled The operator carrying out the final test which comprises of checking all moveabl
12. ding accessories Bed end SANNE III FLEX III KARIN III Wooden side rail BRITT II BRITT III Wooden side rail BRITT IV Lifting pole 6 5 kg 18 Disposal This product has been supplied from an environmentally aware manufacturer that complies with the Waste Electrical and Electronic Equipment WEEE Directive 2002 96 CE This product may contain substances that could be harmful to the environment if disposed of in places land fills that are not appropriate according to legislation The crossed out wheelie bin symbol is placed on this product to encourage you to recycle wherever possible Please be environmentally responsible and recycle this product through your recycling facility at its end of life 26 Yes you can Customer Sales and Service Denmark INVACARE A S Sdr Ringvej 37 DK 2605 Brendby Phone 45 36 90 00 00 Fax 45 36 90 00 OI Sweden amp Finland INVACARE AB Fagerstagatan 9 Box 66 S 163 91 Spanga Phone 46 8 761 70 90 Fa 46 8761 81 08 Norway amp Iceland INVACARE AS Grensesvingen 9 Postbox 6230 Etterstad N 0603 Oslo Phone 47 22 57 95 00 Fax 47 22 57 95 0l Spain INVACARE S A C Areny s n Poligon Industrial de Celr E 17460 Celr Girona Phone 34 972 49 32 00 Fax 34 972 49 32 20 Switzerland INVACARE AG Benkenstrasse 260 CH 4108 Witterswil Phone 41 61 487 70 80 Fax 41 61 487 70 8l Invacare EC Hong AIS Ident no 1417427 Version 14 Dat
13. during service inspections to prevent accidental lowering We recommend a safety test comprising the motors performance and mechanical state once a year Motors hand control and control unit These parts are serviced by exchanging the faulty part 10 Beds equipped with accumulator back up A bed with accumulator back up is equipped with an additional box next to the control unit Preventive maintenance The accumulators have to be exchanged after 4 years Depending on how the back up is used the exchange may have to take place earlier Frequent sudden discharges reduce the back up s life We recommend the accumulator be tested at least once a year Accumulators are not damaged by continuous connection to the mains Exchanging the accumulators Accumulators must be exchanged as sets and replaced with accumulators of the same type or mechanically and electrically compatible accumulators The accumulators are delivered as a pair under order no 818323 I2V 1 2 Ah Accumulators must be new and charged at least every 6 months for maintenance purposes Accumulators in a set must have identical production codes Prior to insertion make sure that the accumulator set is correctly connected cf the drawing in the accumulator box and that none of the connections are loose Warning Old or faulty accumulators may generate an explosive gas mixture during charging The accumulator box is provided with vents to ensure adequate ve
14. e a 2 sop snina saddest SUC PENES ZOXOO CW PD 1417510 1015 SUDDOFE Dandle sager denen DS p 1417511 1015 Support handle secco ses rede 4095 IIl eos ui E EHE EREMO UTER EU eee 1417512 1015 LIFTING POLE Eid ND 50 57600 D0 LIMITATION CROSSBAR Limitation crossbar prevents the leg section from being lowered below the level of the top frame 1510820 1015 Article Dimensions Order no DRIP Drip TOG cst Gueee a 8 d tosh earns eee ami d ed E DUST 50 60030 D0 Drip rod for mounting in lifting pole bel 50 609 10 00 BUMPER WHEELS Bumper wheels ONG sopran 75 mm 50 59600 00 Bumper wheels Ole s uuu s soe erede quur fe ii des ada Eee ad vede Ee reus iru 100 mm 50 59610 00 Bumper wheels 010 222 xcd aes ar p i ae a Spitz den qr diana ie rd dd ecdesiae dis 150 mm 50 59640 00 Bumper wheels one vertical 2 0 cece ee ee eee hr 100 mm 50 59630 D0 MATTRESSES Dacapo Standard 38 kg m cold foam cotton cover 85x200x10 1421530 Dacapo Standard 38 kg m cold foam cotton cover 90x200x10 1421531 Dacapo Standard 38 kg m cold foam cotton cover 105x200x10 1421532 Dacapo Standard 38 kg m cold foam cotton Cover 120x200x10 1421533 Dacapo Basic 38 kg m cold foam incontinence Cover 88x200x12 142155 Dacapo Basic 38 kg m cold foam incontinence Cover 83x200x12 1421552 Dacapo Basic 38 kg m cold foam incontinence Cover 88x183x12
15. e 01 2011 Belgium amp Luxemburg INVACARE N V Autobaan 22 B 8210 Loppem Br gge Phone 32 50 83 10 10 Fax 32 5083 1011 Netherlands INVACARE BV Celsiusstraat 46 INL 6716 BZ Ede Phone 31 318 695 757 Fax 31 318 695 758 Germany INVACARE Aquatec GmbH Alemannenstrasse 10 D 88316 Isny Phone 49 75 627 000 Fax 49 75 62 7 00 66 Portugal INVACARE Lda Rua Estrada Velha 949 P 4465 703 Leca do Balio Phone 351 225 1059 46 47 Fax 351 225 1057 39 Australia INVACARE Australia Pty Ltd I Lenton Place North Rocks NSW 2151 Phone 61 2 8839 5333 Fax 61 2 8839 5353 3rd party certified according to EN ISO 9001 EN ISO 13485 France INVACARE Poirier S A S Route de St Roch F 37230 Fondettes Phone 33 2 47 62 64 66 Fax 33 247 42 12 24 Italy INVACARE Mecc San s r l Via dei Pini 62 36016 Thiene VI Phone 39 0445 38 00 59 Fax 39 0445 38 00 34 United Kingdom amp Ireland INVACARE Ltd Pencoed Technology Park Pencoed UK Bridgend CF35 5HZ Phone 44 656 776 200 Fax 44 656 776 201 Austria MOBITEC Austria GmbH Herzog Odilostrasse 10 A 5310 Mondsee Phone 43 6232 5535 0 Fax 43 6232 5535 4 New Zealand INVACARE NZ 4 Westfield Place Mt Wellington Auckland Phone 64 9 9 7 3939 Fax 649 917 3957 Manufacturer INVACARE EC Hong A S Ostergade 3 DK 4270 Hong www invacarebeds dk
16. e parts motors gas springs and castors issues a QA number on the product thereby confirming its quality If the product does not correspond to the quality demands of Invacare EC H ng A S it will be discarded If contrary to our expectations a problem should arise in connection with the delivered product please contact your Invacare supplier 3 Information Please read the entire technical information section carefully before using or servicing the bed All indications of right and left are based on a patient lying on his back in the bed Please note that there may be sections in this User s Manual which are not relevant for your bed The SB750 is CE marked in accordance with directive 93 42 EEC concerning Medical Apparatus The SB750 is tested without remarks by HMI according to EN 1970 SB750 is tested and approved in accordance with NF Norme Francaise SB750 is EMC tested and approved in accordance with EN 60601 1 2 93 by UL Demko The motors and control unit of the SB750 are approved according to EN 60601 1996 03 SB750 has undergone a risk analysis according to EN 14971 Max patient weight 185 kg The hand control control unit and motors are protected according to IP 66 A lock cam must be used on the control unit if not Invacare EC Hong cannot guarantee the IP protection Electromagnetic interference between the bed and other electrical products can occur To reduce or eliminate such electromagneti
17. e the side rail to the bed end until the spring support snaps into place and prevents the side rail from falling out of the bed end Repeat points 3 to 5 when mounting the other end of the side rail Dismounting the BRITT II III IV side rail Raise the bed about 1 3 Lower the side rail Dismount one end at a time Lift the lower wooden bar so that the spring support is visible Press the spring support down with a wrench or a screwdriver C Hold the lower wooden bar Lower the side rail until the metal bow is free of the bed end rail AWN Repeat points 2 to 4 when dismounting the other end of the side rail Fitting the lifting pole Lifting pole tube with plastic stopper Remove the plastic stopper in the lifting pole tube in which the lifting pole is to be mounted Mount the lifting pole the finger screw must be used Finger screw S a Fitting the bumper wheel Mount the bushing in the bumper wheel and tighten it with a screw under the fitting Bumper wheels may not be mounted on SANNE bed ends Mounting the limitation crossbar under the leg section Raise the leg section of the mattress support and the lifter arm and mount the limitation crossbar Tighten both screws The limitation crossbar is The screws are tightened Correctly mounted placed here limitation crossbar 6 Disassembly assembly of the SB750 Disassembly Make sure that the castors have the brakes on 2 Dismount all
18. efective motor No relay sounds Defective hand control Piston rod does not move No motor sound Defective control unit Motor does not run 13 Lubrication plan We recommend lubricating the B750 according to the following instructions Lifting arm bearings lubricate with oil 2 Motor gas spring bearings lubricate with oil 3 Gliders and guide ways lubricate with grease 4 Shear arms and rollers shaft lubricate with oil grease Lubricate with medically clean oil e g KEW WO 50 order no 813239 and grease order no 1497607 23 14 Cleaning The SB750 cannot withstand cleaning in washing machines nor flushing with a water jet Use ordinary disinfectioning detergents Dry the bed after cleaning Never use acids alkalines or solvents Please make sure the power plug is removed from the socket outlet The back rest and the leg thigh sections can be lowered onto the mattress support which facilitates cleaning and access to the control unit Important details for cleaning hand control control unit and motors IP 66 Always clean with water and brush Water may be pressurised but never use high pressure cleaning or steam cleaning directly on the electrical parts 15 Technical specifications i A O 81 2534 E S g
19. frame and lower it down so that the plastic gliders slide into the guide way 6 Lift the head end of the top frame and push it into the inserts 6 until it stops 7 Screw the 2x2 finger screws into the side bars 8 Connect the plugs of the backrest thigh and leg motors to the control unit see section 7 regarding wiring 9 Mount the accessories 7 Wiring Mount the wires through the cable bow and further to the control unit These photos show a bed with 2 motors for the mattress support Control the wiring by operating all the motors of the bed to their outer positions When all wires have been correctly mounted there is no squeezing risk of the wires ACP Box optional The hand control can be locked by activating the switch on the ACP box located in the head end of the bed To check if the locking function has been activated depress a button on the hand control Open Locked Removal of safety lock Replacement of hand control fitted to the ACP Box The white locking ring on the ACP box is depressed after which the hand control cable can be removed 8 Order numbers for accessories Article Dimensions Order no WOODEN BED END SANNE Ill beech 455 er ni he rode e e ds Bo CI uo cocer RO bU ERU PR PR uad 020429 01 SANNE III beech JON anca rea a alesii d oct a du pal e dl a 020740 01 SANNE lll beech 521212225 mE me ni POSH tese pe eae TUSCE ENTE S 021953 0 I SANNE III beech ace ae
20. la on backrest fourth lamella on leg section mattress frame Mounting the bed end fittings If fittings have not been mounted in the factory Mount the 2 small fittings at the head end and tighten them with the Allen key 2 Mount the large fittings and the welded fitting at the foot end and tighten them with the Allen key Mounting the PIGGY and ODA bed ends Press the bed end into the U profiles Mounting the FLEX III KARIN III bed ends Fasten the bed end to the bed end fittings with the bolts and tighten them with the Allen key Mounting the SANNE III bed end Lift the bed end and place on the fittings at both sides Push down to click into place The locking pin must then be rotated and locked into postion Make sure that the bed end is locked Dismounting the SANNE III bed end Pull the locking pin out and turn it halfway Repeat at the other end Pull the bed end up and out Mounting dismounting the steel side rail Mounting Referring to labelling on the side rail Tighten the side rail with 2 finger screws Dismounting Loosen the finger screws and remove the steel side rail Mounting the BRITT II III IV side rail Raise the bed about 1 3 2 Remove the protective tape from the locking pins at the ends of the side rail 3 Mount one end at a time Lift the lower wooden bar and guide the metal bow to the bed end rail A 4 Press the locking pin of the side rail with a finger B 5 Guid
21. ntilation of the box The vents must not be blocked or covered as this may result in pressure built up and the risk of explosion Waste disposal Old accumulators may be returned to Invacare alternatively they are disposed of like car batteries 2 I Maintenance chart Service and maintenance of the SB750 must only be performed by personnel who have received the necessary instruction or training S N located on mattress support Date Initials Circlips cotter pins and plastic fixing ring checked Screws tightened Weldings checked Side rail locking and moving system tested Castor fittings tightened Castor brakes checked Height adjustment motor checked Back rest motor checked Thigh section motor checked Leg section gas spring motor checked Lead in connections correctly suspended and undamaged Plugs undamaged Lubrication according to the lubrication plan NB The wooden side rails gliding system must not be lubricated with oil otherwise the wooden bars will move sluggishly A service contract can be made in the countries where Invacare has its own sales company In certain countries Invacare offers courses in service and maintenance of the SB750 Spare parts lists and additional user manuals are available from Invacare 22 12 Trouble shooting the electrical system Note Make sure that the mains are connected Motor plug not connected Sounds from the relay l D
22. o final da cama confirmando e garantindo a conformidade com o nosso controlo de qualidade Por Favor leia atentamente este manual de utilizador antes de utilizar a sua cama A Invacare EC Heng A S est certificada em conformidade com ISO 9001 e ISO 13485 Table of contents User Dare 64 oe ee ee e e e se eo n n eee 1399 9 9 9 9 91 9 9 9 9 9 9 9 9 9 8 9599 8 6 I General information 5 222292 3299549 99 eee 9 e ER 6 2 Operation of the SB750 een eee 7 Technical Dart ou 2939929222992 5232094039 0 3 93 99 4 3 799 9 10 S INONMA s 252a 94099 EE 10 4 Receiving the SB750 II 5 Fittingtheaccessories E 6 Disassembly assembly of the SB750 15 Te VYING 4 sek a rreren 2 eee es bee eee rmm 17 8 Order numbers for accessories eee eee 19 9 Maintenance and check ups 21 10 Beds equipped with accumulator back up 21 II Maintenance Cal 6s aoo oce EE 22 I2 Trouble shooting the electrical system 23 13 LUDFCAtION Dialis 3 9 0 909 915 3 at 0 0 aj 19 49 D EEN 23 I4 CICAMING 625525 e 43929 33 23 9 2 23 39 3359 3 8 89 3 8 79 09 98 13 8 978098 80908 9080 8 24 15 Technical specifications 24 16 Electrical dtd oo 399239919 409 9999093 199 9 99 90909 9 9 909090809 90909
23. rails are used When using side rails and the according safety cover it is essential to ensure correct fitting otherwise there is danger of entrapment suffocation between mattress support side rail and bed end If the bed is used by restless spasms or confused persons an ACP box can be used for blocking the hand control functions and or a limitation cross bar installed under the leg section of the mattress support This cross bar will prevent the leg section from being lowered below the level of the mattress support thereby reducing the opening between leg section and bed end Ensure the bed is adjusted to it s lowest position before leaving the bed unattended thereby the effect of fall down entrapment accidents is reduced Invacare EC Hang A S accepts no liability for any use change or assembly of the product other than as stated in this User s Manual When entering or exiting the bed the height adjustment should be used Furthermore the backrest can be used as support yellow button on the hand control However the thigh and leg section must be horizontal otherwise there is risk of overloading of the mattress support 2 Operation of the SB750 The bed can be equipped with this type of hand control Soft Control e Height adjustment of the bed UP DOWN Sitting position irrespective of present position Horizontal mattress support irrespective of present position Out of bed button Raises the back
24. re bears quio Kn pne piro CERE 021954 0 I SANNE III cherry beech 2 20 e eee OO CM 24405045505 oe dese eee senses 021956 07 SANNE III increased and lowered beech BOC nois aeu Ata natal on di A seus 145 1258 0101 SANNE III increased and lowered beech DOCH geleet meee 1451257 0101 SANNE III increased and lowered beech EE eee a CEU E e ME PE 484403 0101 SANNE III increased and lowered beech L20 CM eon ae eee Gee a at ae eae ae 484404 0101 To be used with the LILLI III wooden side rail FLEX DESC cues 46555040504 eee5eeeery sens nen OO CM p P 02041 7 01 FLEX IIl beech NEE DU Morem 020737 01 PIGGY beech U profiles llle BOC M P 020431 01 PIGGY beech U profiles llllllllllssne JON 2222392599592 992997929195 5 9 3 018940 0 I ODA beech U profiles cc BS ON EC 020430 0 I ODA beech U profiles 0 0 0 cee eee eee SU Cl se 016609 01 Bed ends are available in other versions than the above mentioned contact Invacare Side rails Scala Basic Metal side rail foldable leen 1432784 1015 Scala Basic Plus Metal side rail foldable with plastic insert 0 cece elles 1432787 1015 Scala Medium Metal side rail foldable ccc cee eee eee eh 1432781 1015 Scala Decubi Metal side rail foldable lille 1432793 1015 Metal side rail Model
25. rest and brings the thigh section to horizontal Raises the thigh section and brings gt Kate the backrest to horizontal Adjustment of leg section UP DOWN not on beds with 3 sectioned mattress support and manual adjustment of the leg section Or with this type of hand control HB 70 UP DOWN Sitting position Backrest Leg section electric not on beds with 3 sectioned mattress support and manual adjustment of the leg section Thigh section AA AA AN Height adjustment of the bed Operating the leg section on beds with gas springs Lift the handle with one hand while adjusting the leg section to the desired position with the other hand Release the handle fn Leg section manual Release handle Operating the leg section on SB750 120 cm width with Rastofix To raise the leg section Lift the leg section to desired height NB Please ensure that both Rastofix brackets are correctly locked in place To lower the leg section Lift the leg section slightly and lower to horizontal accidental lowering of the mattress support A person can be seriously injured under the bed during height adjustment Always leave the bed in the lowest position Otherwise there is a risk of squeezing due to Operating the braking castors Operating castors without central braking system When the bed is positioned correctly at least one castor at the head end and one castor at the foot end must be locke
26. screws for fastening the fitting below the side bar by means of the Allen key Pull out the fitting so that it extends 16 cm insert the 2 screws again and retighten them Loosen the 2 screws below the mattress handle on the leg section pull the handle completely up and retighten the screws Insert the mattress support extension Emergency release of the mattress support parts An emergency release of the mattress support is necessary in the case of e g a power or motor failure An emergency release of the height adjustment is NOT possible Remove the plug from the mains before emergency release of the mattress support In an emergency the mattress sections are released by pulling out the cotter pin from the motor gas spring in question A minimum of 2 persons are required to release a mattress section Both persons hold the mattress section in locked position One of them pulls out the cotter pin Both slowly lower the mattress section until it is completely down 5 Fitting the accessories The tools necessary for mounting dismounting the accessories of the SB750 are an Allen key and an adjustable wrench The Allen key is kept on the inside of the cross bar at the head end of the top frame The Allen key is used for mounting dismounting all fittings Mounting the mattress retainer Mattress retainer is To be placed between third and To be placed between third and squeezed down over the fourth lamel
27. t C N 00 O CO LO te T3 all Ta D 136 ci J B at EES ERED ES san o a o ERA All measurements are stated in cm All angles are stated 90x200 200 213 i tor so in degrees All measurements and angles are stated without tolerances Invacare reserves the right to change 85x183 183 196 a5 95 43 the stated measurements and angles without warning mo o e a fo e an 0 as res ve reas 20 20 re oa Transportation and storage conditions Temperature between 10 and 50 C Humidity between 10 and 80 RH Pressure between 700 and 1 100 hPa 14 22 F 1 100 hPa 24 16 Electrical data Voltage supply Max current input Intermittent periodic use of motors Max accumulator capacity Protection class Output Sound level Double insulated class Il type B Alternating current V Direct current Max load SWL Patient mattress siderail lifting pole other equipment The patient is not separated from the ground and the chassis Potential equalization 230 V AC 50 Hz 1 5 A 10 max 6 min h 1 2 Ah IP 66 70 VA 48 dB A gt PD B 220 kg The control unit has no contact breaker disconnecting the plug is the only means of interrupting the power supply 25 17 Weight Measurement Weight Top frame head with mattress support Top frame foot with mattress support Base frame Complete bed exclu
28. tate apposte sul letto a conferma dell avvenuta verifica e accettazione del prodotto da parte del servizio qualit Prima di utilizzare il letto vi invitiamo a leggere integralmente il manuale d uso Invacare EC Hang A S un azienda certificata ai sensi della Normativa ISO 9001 e ISO 13485 ES DECLARACI N DE CALIDAD Enhorabuena por su nueva cama Invacare ScanBed 750 de Invacare EC Hong A S Su nueva cama cuenta con el marcaje CE de acuerdo a la directiva 93 42 EOF que hace referencia a los aparatos m dicos Invacare ScanBed 750 ha sido disefiada y fabricada teniendo en cuenta a los usuarios y las personas que los asisten La cama Invacare ScanBed 750 ha sido disefiada de acuerdo a la normativa Europea EN 1970 El proceso de producci n del modelo Invacare ScanBed 750 ha sido supervisado en su totalidad y su calidad inspeccionada por nuestro control de producto acabado La cama cuenta con una placa y marca CE que confirma que la misma ha sido inspeccionada por el control de producto acabado Por favor lea el manual antes de utilizar la cama Invacare EC Hong A S es una empresa certificada ISO 9001 e ISO 13485 FR D CLARATION DE QUALIT F licitations Vous avez choisi votre nouveau lit Invacare ScanBed 750 de Invacare EC Hong A S Votre nouveau lit est marqu CE conform ment la directive 93 42 EEC concernant les dispositifs m dicaux Invacare ScanBed 750 a t d velopp et construit en consid rant s
29. ture Check that all plugs of the motors and hand control are Unlocked pin correctly connected to the control unit according to the snap lock printed symbols A The bed must have the brakes engaged during nursing of a patient in the bed Disconnect the plug from the mains before moving the bed The cable must be kept clear off the floor and the castors during transport Connect the SB750 to the mains For ergonomic working position Run the bed to approx 2 height level Cf chapter 2 Operation of the SB750 Mount the accessories Bed ends side rails lifting pole and bumper wheel Cf chapter 5 Fitting the accessories Parts of a standard bed et SY ULIS A a Accessories Brake for beds provided with brake castors Castor Base frame 9 24 V motor backrest Shear arms Top frame head end Control unit Allen key Backrest Seat section 24 V motor leg section or gas spring leg section Thigh section 24 V motor thigh section Leg section Top frame foot end Brake pedal for beds provided with central brake Battery box Fittings Bed end Wooden side rail Steel side rail Finger screw Lifting pole Bumper wheel Operating the built in mattress support extension The mattress support extension is mounted using the Allen key The Allen key is kept on the inside of the cross bar at the head end l 2 3 Remove the wooden side rails cf chapter 5 Remove the 2
30. yst matiquement les besoins de l utilisateur et des tierces personnes lors de la manipulation du lit ou de son utilisation Invacare ScanBed 750 a t d velopp conform ment au Standard Europ en NF EN 1970 et NF EN 60601 2 38 Invacare ScanBed 750 a t supervis et contr l tout au long du process de fabrication et le lit achev a t inspect par le contr le des produits finis La plaque d identification et la marque QA sont plac es sur le lit attestant que le contr le des produits finis a approuv le lit Le lit r pond aux exigences de l analyse de risques de la norme NF EN 14971 Nous vous remercions de lire le Manuel de l Utilisateur dans son int gralit avant d utiliser le lit Invacare EC Hong A S est certifi e ISO 9001 et ISO 13485 PT DECLARA O DE QUALIDADE Parab ns pela sua nova cama Invacare ScanBed 750 da Invacare EC H ng A S A cama tem a marca CE em conformidade com a directiva 93 42 EEC referente a aparelhos m dicos A cama foi concebida e desenhada tendo em considera o o seu utilizador e o seu s assistente s que o ajudar o a manipular a cama A cama foi concebida em conformidade com o Standard Europeu EN 1970 Durante todo o processo de fabrico e produc o a cama Invacare ScanBed 750 foi supervisionada e a sua qualidade controlada sendo o produto final inspeccionado e testado pelo nosso controlo de qualidade As etiquetas e a marca QA s o colocadas na cama ap s a aprova
Download Pdf Manuals
Related Search
Related Contents
Toshiba Satellite L50-B-176 Soleus Air KFHHP-18-OD User's Manual FT-1208 Copyright © All rights reserved.
Failed to retrieve file