Home

Fluke 740 Series Documenting Multifunction Calibrators

image

Contents

1. e Provides important information for installation and maintenance Tag IDs measured values range and span data product information and diagnostics e Reduces operation costs by making it easier to manage and fully utilize smart instrument networks The fast one handed tool for loop checks This feature rich solution combines the best features of a push button and knob base loop calibrator and offers Large display and simple quick click interface for easy one handed operation simultaneous mA and readout for quick easy interpretation of readings mA accuracy of 0 015 superior to other loop calibrators in this price range 1 uA resolution for mA source simulate and measure Push button 25 steps for fast easy linearity checks O 100 span check for fast confirmation of zero and span selectable slow ramp fast ramp and step ramp to provide smooth outputs for valve slewing and loop functional tests 24 volt internal loop supply so you can power and read a transmitter at the same time without carrying a DMM Measures V dc to 28 V single 9 V battery that is easily changed O 20 mA or 4 20 mA default start up modes Innovative output adjustment dial with 1 uA and 100 uA resolution HART mode connects 250 Q resistor in series with 24 V loop for compatibility with HART communicators Select and schedule Tag IDs for calibration Examine and analyze the detailed calibration histories o
2. T TRANSAT gt Visit us at Transcat com 35 Vantage Point Drive Rochester NY 14624 Call 1 800 800 5001 cat d a a E ele FIELD COMMUNICATIONS PRO GEN Whether you re calibrating instruments troubleshooting a problem or running routine maintenance Fluke 740 Series process calibrators can help you get the job done faster It does so many different tasks so quickly and so well it s the only process calibrator you need to carry e Multifunctional Calibrate temperature pressure voltage current resistance and frequency since it both measures and sources you can troubleshoot and calibrate all with one rugged tool e Powerful yet easy to use The easy to follow menu driven display guides you through any task Pro grammable calibration routines enable you to create and run auto mated as found as left procedures to ensure fast consistent calibrations e Records and documents results To support your ISO 9000 or regula tory standards the Fluke 741B 743B and 744 capture your calibration results eliminating the need to juggle a pen and pad in the field The RS 232 interface in the Fluke 743B and 744 lets you transfer the results to a PC thus saving the time of having to manually transcribe them when you return to the shop e Supports popular instrumentation management software The 743B and 744 work with the new Fluke DPC TRACK software and with
3. AMS from Emerson Process Management formerly Fisher Rosemount Exit to Windows Login to one of four user levels Complete user s manual on line Honeywell Loveland DocuMint BLUE MOUNTAIN QUALITY RESOURCES Prime Technologies On Time Support Process Track All trademarks are the property of their respective holders To see the Fluke 740 Series calibrators in action call 1 800 44 FLUKE U S 31 40 2 675 200 Europe 1425 446 5500 other countries or your local Fluke representative for a demo Visit our Web site www fluke com Processtools DC current measurement Accuracy of reading of full scale Range full scale DC voltage measurement Accuracy of reading of full scale Range full scale 1 year years 110 000 mV 110 000 V 300 00 V Temperature coefficient 0 001 reading 0 0015 f s C from 10 C to 18 C and 28 C to 50 C Input impedance 5 MQ Common mode error 0 008 f s Common Mode Volt Maximum input voltage 300 V rms 1 10000 V 11 0000 V AC voltage measurement Accuracy of reading counts 1 year years 500 Hz to 1 kHz 1 to 5 kHz Ranges 1 1000 11 000 110 00 300 0 V Specifications apply for 10 to 100 of range Input impedance 5 MQ and lt 100 pF Input coupling AC only Temperature coefficient 10 of spec C from 10 C to 18 C and 28 C to 50 C Maximum input voltage 300 V rms Minimum input voltage 0 5 V
4. above 1 kHz 20 to 40 Hz 40 to 500 Hz 110 000 mV 0 01 0 005 0 015 0 005 1 10000 V 0 01 0 005 0 015 0 005 15 0000 V 0 01 0 005 0 015 0 005 Temperature coefficient 0 001 output 0 001 f s C from 10 C to 18 C and 28 C to 50 C Maximum output current 10 mA Loading 0 001 f s 1 uV mA Common mode error 0 008 f s Common Mode Volt Maximum input voltage 30 V dc DC current output Accuracy of output of full scale Range full scale 1year ___ 2 years 22 000 mA 0 01 0 015 0 02 0 015 Current sink sim 0 02 0 03 0 02 0 03 ulate transmitter Specification applies from 2 to 22 mA below 2 mA typical accuracy is 0 15 of full scale Maximum burden voltage 24 V Temperature coefficient 0 003 output 0 003 f s C from 10 C to 18 C and 28 C to 50 C Common mode error 0 008 f s Common Mode Volt Maximum input voltage 30 V dc Temperature coefficient 0 001 reading 0 002 f s C from 10 C to 18 C and 28 C to 50 C Common mode error 0 01 f s Common Mode Volt Maximum input voltage 30 V dc Resistance measurement Accuracy of reading ohms Range full scale 1 year years Temperature coefficient 0 01 f s 2 mQ C from 10 C to 18 C and 28 C to 50 C Common mode error 0 005 f s Common Mode Volt Maximum input voltage 30 V dc Continuity Continuous tone lt 25 Q No tone g
5. calibrators Included with DPC TRACK software Software media Instruction Manual Serial Port Cable DB9 to DB25 9 pin to 25 pin Adapter FLUKE 700 Pxx Pressure Modules Included with each Fluke Pressure Module BP ISO Adapter s except with P29 P31 Instruction Sheet NIST traceable calibration report and data one year warranty Accessories Fluke 70OPMP Pressure Pump 100 psi 7 bar Fluke Z7OOLTP Low Pressure Test Pump Fluke 700PTP Pneumatic Test Pump 360 psi 25 bar Fluke 7OOHTP Hydraulic Test Pump 10 000 psi 700 bar Fluke 7OOHTH Hydraulic Test Hose Fluke Z7OOPRV Pressure Relief Valve Kit for HTP Fluke 700 IV Current Shunt for mA mA applications Fluke Z7OOPCK Pressure Calibration Kit Fluke 7OOBCW Bar Code Wand Fluke 7JOOTC1 TC Mini Plug Kit 9 types Fluke 7JOOTC2 TC Mini Plug Kit JKTERS SOT IR Infrared Probe SOPK IR Infrared Probe BE9005 Battery Eliminator BC7217 Battery Charger BP7217 NiCd Battery Pack BP7235 NiMH Battery Pack C700 Hard Carrying Case C781 soft Carrying Case C789 soft Carrying Case C75 Test Lead Case Fluke Keeping your world up and running TRANSGAT _ gt vstesattnscatcom 35 Vantage Point Drive Rochester NY 14624 Call 1 800 800 5001 2001 Fluke Corporation All rights reserved Printed in U S A 10 2001 1263323 A ENG N Rev F
6. fast HART communication e Supports popular models of HART transmitters with more device specific command support than any other HART field calibrator e Fully complies with the Data Link Layer of the HART protocol including multiple masters burst mode and multi drop configurations Is easy to update as additional instruments are added and new HART versions are released e Is based on the 743B the most rugged reliable multifunction field calibrator available Is backed by the service and support of the Fluke organization a member of the HART Communica tions Foundation The 744 is designed to take on nearly all the day to day tasks you now perform with a separate communi cator In fact it offers the communication capabilities of the 275 HART communicator except for the DD interpreter which can read command set libraries from any HART supplier This is not necessary for daily HART maintenance It s easy to calibrate and maintain HART instrumentation with one powerful tool FLUKE 744 DOCUMENTING PROCESS CALIBRATOR Miniplug Measure mA and Loop Power with built in 250 Q resistor With the 744 DPC you can e Generate precision electrical temperature or pressure signals for analog stimulus or sensor simulation e Simultaneously measure electrical temperature or pressure signals from transmitter output e Determine type manufacturer model tag ID by interrogating HART devices e Read HART PV fu
7. popular programs from Honeywell Loveland Cornerstone Sand Cove Beamex Blue Mountain and others It allows you to create procedures instructions and action lists to deliver fast easy documentation e Truly hand held Small enough to fit easily into a tool bag and to use in tight spaces Runs an entire shift on a rechargeable NiCd or NiMH battery pack e Rugged and reliable Overmolded urethane case stands up to rough handling in industrial environments Calibrators offer one or two year calibration cycles and three year warranty Bright white display lets you read your results in any kind of light Soft keys provide one touch access to enhanced functions such as task lists automated procedures scaling min max stepping and ramping and review memory Three operating modes Measure Source or simultaneous Measure Source enable technicians to troubleshoot calibrate or maintain instrumentation with just one tool Integrated HART communication capability lets you program and control HART instrumentation ie 744 only Multi lingual interface displays instructions in English French German Spanish and Italian AutoStep allows technicians to set the calibrator for a delayed start and a specific sequence of steps so it can run unattended as a continu ously varying test source User entered values enable users to capture readings measured or sourced by other devices Custom un
8. 10 9 KHz 30 of output 0 5 f s 11 to 50 kHz Maximum input voltage 30 V dc Temperature Measurement and Simulation Specifications Temperature RTDs Accuracy C Meas Source 10 Q Cu 42 o O 10 Q Cu 427 100 to O C O to 260 C 100 Q Pt 3916 200 to 190 C 190 to O C O to 630 C 100 Q Pt 3926 200 to O C O to 630 C 100 Q Pt 385 200 to O C O to 400 C 400 to 800 C 200 Q Pt 385 200 to 0 C O to 400 C 400 to 630 C 500 Q Pt 385 200 to O C O to 400 C 400 to 630 C 0 3 C 0 3 C 0 5 C 0 3 C 0 5 C 0 3 C 0 5 C 0 8 C 0 3 C 0 5 C 0 8 C 0 3 C 0 5 C 0 8 C 0 4 C 0 4 C 0 8 C 0 4 C 0 8 C 0 5 C 0 8 C 1 0 C 0 5 C 0 8 C 1 0 C 0 5 C 0 8 C 1 0 C 0 3 C 0 1 C 0 2 C 0 1 C 0 2 C 0 1 C 0 2 C 0 4 C 0 1 C 0 2 C 0 4 C 0 1 C 0 2 C 0 4 C 2 C 2 C 1 C 1 C 2 C 2 C 1 C 1 C 0 4 C 0 2 C 0 4 C 0 2 C 0 4 C 0 2 C 0 4 C 0 5 C 0 2 C 0 4 C 0 5 C 0 2 C 0 4 C 0 5 C 1000 Q Pt 385 200 to 0 C O to 400 C 400 to 630 C 120 Q Ni 672 80 to 260 C 04 C For 2 wire and 3 wire measurements add 0 4 C Sensor inaccuracies not included Resolution 0 1 C except 1 C for 10 Q Cu Temperature coefficient 0 02 C C from 10 C to 18 C and 28 C to 50 C Maximum input voltage
9. 30 V dc Maximum input current for RTD Source function 10 Q RTD 8mAdc 100 120 Q RTDs 8 mA dc 200 500 1000 Q RTDs 1mAdc 741B 743B 744 Addresses pulsed transmitters and PLCs with pulses as short as 1 ms For 741B serial number 7935XXXX and greater For 743B serial number 7940XXXX and greater Otherwise 3 mA and 100 ms 0 3 C 0 5 C 0 8 C 0 5 C 0 8 C 1 0 C 0 120 0 2 C 0 4 C 0 2 C 0 4 C 0 5 C 0 1 C How to compare calibrators based on specifications Analyzing specifications can be complex To get a true picture of calibrator performance you should be aware of the key components of a specification and how to interpret them Specifications must be carefully considered when compar ing calibrators from different vendors The most important components of a process calibrator specification are e Reference uncertainty Performance of a calibrator al 2a C F3 al the time it is verified by the manufacturer This specification does not include the effects of time and temperature two of the largest components of calibrator error e Temperature Fluke process calibrator specs reflect perfor mance from 18 to 28 C Compensation factors are provided to permit specified use of the calibrators over a wide 10 to 50 C range Allowance for traceability Fluke specs are not relative specs but total specs including an allowance for uncertainty of standards tha
10. 740 Series are 1 and 2 years Typical 90 day measurement and source accuracy can be estimated by dividing the one year of reading or Oo of output specifications by 2 Floor specifications expressed as of f s or counts or ohms remain constant Power Internal battery pack NiCd 7 2 V 1700 mAh NiMH 744 only 7 2 V 3500 mAh Battery Life Typical usage gt 8 hours Dimensions 130 x 236 x 61 mm 5 1 x 9 3 x 2 4 in Weight 1 4 kg 3 lbs 1 oz Side Port Connections e Pressure module connector e RS 232 connector 743B and 744 to interface to your PC e Connection for optional battery eliminator Safety Complies with CAN CSA C22 2 No 1010 1 92 ANSI ISA 82 01 1994 UL3111 and EN610 1 1993 Data Storage Capacity Fluke 741B 1 day of calibration results Fluke 743B and 744 1 week of calibration results FLUKE 741B Documenting Process Calibrator FLUKE 743B Documenting Process Calibrator FLUKE 744 Documenting Process Calibrator HART Included with the Fluke 740 Series Type TL24 Industrial Test Leads two sets AC20 Test Clips 2 sets TP20 Test Probes 1 set Battery Pack Battery Charger serial port cable except 741B HART communications cable 744 only DPC TRACK Sample with free PC communication utility software except 741B Instruction Manual NIST traceable calibration report and data three year warranty FLUKE 700SW DPC TRACK Software for use with Model 743B or 744
11. IST Monograph 175 for E N J K T B R S thermocouples IPTS 68 per IEC 584 1 for E J K T B R S thermocouples IPTS 68 per DIN 43710 for L U thermocouples Temperature coefficient 0 05 C C from 10 C to 18 C and 28 C to 50 C Common mode error 0 01 C Common Mode Volt Maximum input voltage 30 V dc 1 0 C 0 5 C 0 6 C 1 5 C 0 8 C 0 9 C 0 6 C 0 5 C 0 3 C 0 9 C 0 8 C 0 4 C 0 6 C 0 3 C 0 5 C 0 9 C 0 4 C 0 8 C 0 3 C 0 2 C 0 2 C 0 4 C 0 3 C 0 3 C 0 6 C 0 3 C 0 5 C 0 9 C 0 4 C 0 8 C 0 3 C 0 2 C 0 2 C 0 4 C 0 3 C 0 3 C 0 7 C 0 3 C 0 5 C 0 7 C 1 0 C 0 4 C 0 8 C 1 0 C 0 4 C 0 3 C 0 3 C 0 3 C 0 6 C 0 4 C 0 4 C 0 4 C 1 7 C 0 6 C 0 3 C 2 5 C 0 9 C 0 4 C 0 9 C 0 4 C 0 3 C 1 4 C 0 6 C 0 4 C 0 6 C 0 3 C 0 9 C 0 4 C 0 4 C 0 3 C 0 6 C 0 4 C 1 3 C 1 0 C 0 9 C 2 0 C 1 5 C 1 3 C 1 0 C 0 8 C 0 8 C 1 5 C 1 2 C 1 2 C 2 3 C 1 5 C 1 0 C 2 8 C 2 2 C 1 5 C 1 2 C 1 1 C 0 9 C 1 8 C 1 7 C 1 4 C Note When simulating temperature in As Found As Left procedures steps may be either linear by temperature or linear by mV potential The Fluke family of 29 pressure modules Covers virtually any pressure application includ
12. cess plants begin to take advantage of smart transmitters the need for a new generation of calibrators has emerged calibrators that can communicate via industry standard digital protocols The 744 combines HART communication capability with the popular 740 Series Documenting Process Calibrators to deliver an integrated communi cating calibrator This rugged reliable tool is ideal for calibrating maintaining and troubleshooting HART instrumentation The 744 offers e Integrated HART communication functions permit ting you to monitor control and calibrate HART instrumentation e Handling of fast pulsed RTD transmitters and PLCs with pulses as short as 1 mS e NiMH battery with 3500 mA hour life and gas gauge FIELD COMMUNICATIONS PRO wn ay Nau Capability 741B 743B 744 Source Measure Automated procedures Results capture pe ew e Uses all Fluke pressure modules e Transmitter mode tee ee e Barcode entry Sf Serialinterface lele Datalogging lele HART communications e Pulsed RTD simulation to 1 ms NiMH battery with Gas Gauge The first field calibrator for HART instruments that s both powerful and easy to use The 744 offers the most complete HART implementa tion of any process calibrator The 744 e Requires no external box or second tool for every day HART calibration and maintenance e Offers
13. f your Tag IDs and instruments Import test parameters Manage your database with system utilities Export calibration results as delimited ASCII data Fluke 7OOSW DPC TRACK software includes an instrumentation data base that makes it easy to manage your instrumentation create and schedule tests load and unload the 743 or 744 print a variety of standard reports and manage calibration data DPC TRACK is available in English French German Spanish and Italian Print standard reports automati cally The software assembles pre formatted reports from your database files saving time and reducing errors Reports include calibration certificates instruments due for calibration inventory characteristics calibration histories calibration procedures and traceability to instruments touched from your existing system Load selected procedures to the calibrator and unload calibration results to your PC Print a selection of pre formatted reports Tags Instruments Test Equipment Examine and document the calibration histories of your test equipment Manage your inventory of Tag IDs and instruments Documentation of results The scheduling of calibrations creation of procedures and documentation of your calibration results are facilitated by a number of instrumentation management software packages Cal Station and Base Station Fluke DPC TRACK Cornerstone The Open Platform for Instr
14. fication includes common mode pressure Modules are C rated Metric adapter s 1 4 NPT female to male BSP ISO 1 4 19 tapered thread included with all modules except P29 P30 and P31 Effective October 1996 all modules include a NIST traceable certificate and test data 11 Data log function except 741B Measure functions Voltage current resistance frequency temperature pressure Reading rate 1 2 5 10 20 30 or 60 readings per minute Maximum record length 8000 readings 7980 for 30 or 60 readings per minute Ramp function Source functions Voltage current resistance frequency temperature Rate 4 steps second Trip detect Continuity or voltage Continuity detection not available when sourcing current Loop power function Voltage Selectable 24 V or 28 V Accuracy 5 Maximum current 22 mA short circuit protected Maximum input voltage 30 V dc Note 2509 series resistance is automatically supplied whenever loop power is enabled on 744 HART modem interface 744 only Maximum input voltage 30 V dc Environmental specifications All calibrator specifications apply from 18 C to 28 C unless stated otherwise Operating temperature 10 C to 50 C 20 C typical except for frequency and ac voltage measurement Storage temperature 20 C to 60 C Operating altitude 2800 m above mean sea level 9186 ft 90 day specifications The standard specification intervals for the
15. ing gage differential dual compound absolute and vacuum e Display pressure readings in any of 10 different pressure units you specify in the calibrator setup e Rugged urethane molded cases protect the modules from rough handling and harsh conditions e Features internal temperature compensation from O to 50 C for full accuracy performance e Includes NIST traceable calibration certificate e Modules can be calibrated locally helping to control costs Pressure module specifications all specifications in of full span Specifications reflect a confidence interval of 95 Reference High Low Max over Range Range approx uncertainty Stability Temperature Total side side Fitting pressure Resolution Resolution 23 3 C 1 year O to 50 C uncertainty media media material x nominal Differential FLUKE 700P00 1 in H 0 0 001 0 25 kPa 0 0002 FLUKE 700P01 10 in H 0 0 01 2 5 kPa 0 002 FLUKE 700P02 1 psi 0 0001 6900 Pa 0 7 FLUKE 700P22 1 psi 0 0001 6900 Pa 0 7 FLUKE 700P03 5 psi 0 0001 34 kPa 0 001 FLUKE 700P23 5 psi 0 0001 34 kPa 0 001 FLUKE 700P04 15 psi 0 001 103 kPa 0 01 FLUKE 700P24 15 psi 0 001 103 kPa 0 01 FLUKE 700P05 30 psi 0 001 207 kPa 0 01 FLUKE 700P06 100 psi 0 01 690 kPa 0 07 FLUKE 700P27 300 psi 0 01 2070 kPa 0 1 FLUKE 700P07 500 psi O0 01 3400 kPa O0 1 FLUKE 700P08 1000 psi 0 1 6900 kPa O0 7 FLUKE 700P09 1500 psi O 1 10 MPa 0 001 Absolu
16. its allow readings to be scaled and displayed in any user defined units Limit switch calibration procedures perform fast automated calibration of one and two point limit switches for voltage current temperature and pressure Differential pressure flow instrument calibration routines use a Square root function to directly calibrate DP flow instruments Built in algebraic calculator with four functions plus square root stores recalls and performs calculations required for setting up instruments or evaluating data in the field Use it to set the source function to a calculated value There s no need to carry a pencil and paper or a separate calculator Programmable measurement delay inside automated procedures permits calibrating instruments that respond slowly The Fluke 740 Calibrators offered in three models let you choose the right set of capabilities for your needs e The Fluke 741B offers simultaneous source and measure capabilities for all common process parameters Create and execute automated proce dures and automatically capture the results e The Fluke 743B offers all the capability of the 741B plus adds a serial interface for two way communica tion with popular PC based instrumentation management applications e The Fluke 744 offers all of the capabilities of the 743B plus the ability to maintain and calibrate selected HART transmitters without a second tool As more and more pro
17. lenges of maximizing productivity while minimizing maintenance costs Smart digital transmitters offer superior performance and reliability while saving time and effort in maintenance and calibration Manufacturers of field instruments have helped accelerate the changeover by offering smart transmitters at prices nearly as low as analog units As digital instruments using the HART protocol quickly become the standard communicators and calibrators are becoming essential everyday tools What is HART HART the Highway Addressable Remote Transducer protocol uses a 1200 baud Frequency Shift Keying FSK signal to superimpose digital information on the conventional 4 20 mA analog signal Why use the HART protocol HART is an industry standard developed to define the communications protocol between intelligent field devices and a control system HART is the most widely used digital communication protocol in the process industry More than five million HART field instruments are installed in more than 100 000 plants worldwide The HART protocol e Is supported by all of the major suppliers of process field instruments supported by the HART Communi cation Foundation an industry wide non profit organization See the Web site HYPERLINK http www hartcomm org for information on the HART standard e Preserves present control strategies e Allows traditional 4 20 mA signals and digital communication to share the same two wire loops
18. nction and smart transmitter digital output while measuring analog mA output e Read and write HART configuration functions to make field adjustments to PV range points damping and other top level configuration settings RO pa LD Q SARRO AYN R als Sy KN i j QY Change sensor configuration on supported temperature transmitters Re label smart transmitters by reading and writing HART tag and message fields Clone additional transmitters by reading and storing basic HART configurations Perform automated HART sensor trim and output trim for selected devices in conjunction with As Found As Left tests Perform loop test with simultaneous analog and digital mA readout Address new fast pulsed excitation smart transmit ters and PLCs Fluke 744 The first HART calibrator that is easy to use Today s 744 supports device specific instructions for a variety of popular instruments Additional support may be added periodically with a simple software update available on diskette or via download for a modest upgrade fee HART operating modes supported e For Point to Point operation the most commonly used mode connects the 744 to a single HART device in a 4 20 mA loop e In Multi Drop mode several HART instruments can be bussed together The 744 searches for each identifies addresses in use and allows you to select the instrument for calibration and related operations e In Burst Mode the HART ins
19. t provide traceability to national standards Confidence level Fluke uses a conservative 95 confidence level when setting specifications increasing your confidence that your calibrator will remain in spec for its stated calibration interval e Time Fluke 740 Series calibrators are delivered with both one year and two year specs to limit your calibration support costs You choose your cal interval based For more information refer to the application note Understanding Specifications For Process Calibrators upon the performance you need 10 Temperature Thermocouples Accuracy C mease sowe Type and range E 250 to 200 C 200 to 100 C 100 to 600 C 600 to 1000 C N 200 to 100 C 100 to 900 C 900 to 1300 C J 210 to 100 C 100 to 800 C 800 to 1200 C L 200 to 100 C 100 to 800 C 800 to 900 C K 200 to 100 C 100 to 400 C 400 to 1200 C 1200 to 1372 C T 250 to 200 C 200 to O C O to 400 C U 200 to O C O to 600 C B 600 to 800 C 800 to 1000 C 1000 to 1820 C R 20 to O C O to 100 C 100 to 1767 C S 20 to O C O to 200 C 200 to 1400 C 1400 to 1767 C C O to 800 C 800 to 1200 C 1200 to 1800 C 1800 to 2316 C Sensor inaccuracies not included Accuracy with external cold junction for internal junction add 0 2 C Resolution 0 1 C Temperature scale ITS 90 or IPTS 68 selectable Compensation ITS 90 per N
20. t 400 Q Frequency measurement ya eT 1 00 to 109 99 Hz 0 05 Hz 0 05 Hz 100 10 1090 9 1 100 to 10 999 kHz 0 005 kHz 0 005 kHz 11 00 to 50 00 kHz 0 05 kHz 0 05 kHz For frequencies lt 109 99 Hz specification applies for signals with slew rates gt 5 V ms Minimum amplitude for Hz measurement Squarewaves 1 Hz to 1 kHz 300 mV p p 1 kHz to 30 kHz 1 4 V p p gt 30 kHz 2 8 V p p Maximum input 1 Hz to 1 kHz 300 V rms gt 1 kHz 30 V rms Input impedance 5 MQ DC voltage output basen Accuracy of output of full scale Range full scale 1year ___ 2years Resistance sourcing Range full scale 11 000 Q 110 00 Q 0 01 40 mQ 1 1000 KQ 0 02 0 5 Q 0 03 0 5 Q 11 000 KQ 0 03 5 Q 0 04 5 Q Temperature coefficient 0 01 f s C from 10 C to 18 C and 28 C to 50 C Maximum and minimum current through source resistance Accuracy of output ohms 0 01 20 mQ 0 02 20 mQ 0 02 40 mQ Maximum Minimum 11 Q range 3 mA dc 0 1 mA dc 110 Q range 3 mA dc 0 1 mA dc 1 1 KQ range 3 mA dc 0 01 mA dc 11 kQ range 1 mA dc 0 01 mA dc Common mode error 0 008 f s Common Mode Volt Maximum input voltage 30 V dc Frequency sourcing ae a 1year 0 00 to 10 99 Hz 0 01 Hz 0 002 kHz 0 005 kHz Waveforms Squarewave w 50 duty cycle sinewave Amplitude 0 1 to 10 V p p Amplitude accuracy 3 of output 0 5 of f s 1 to 1099 Hz 10 of output 0 5 of f s 1 1 to
21. te not compatible with Fluke 701 or 702 FLUKE 7OO0PA3 5 psi 0 0001 34 kPa 0 001 FLUKE 7O0PA4 15 psi O 001 103 kPa 0 01 FLUKE 7OOPAS 30 psi 0 001 207 kPa O0 01 FLUKE 7OOPA6 100 psi O0 01 690 kPa 0 07 Vacuum not compatible with Fluke 701 or 702 FLUKE 700PV3 5 psi 0 0001 34 kPa 0 001 0 040 0 015 0 015 0 070 316 SS Dry 316 SS 3x FLUKE 700PV4 15 psi O 001 103 kPa 0 01 0 040 0 015 0 015 0 070 316 SS Dry 316 SS 3x FLUKE 7OOPD2 1 psi O 0001 6900 Pa 0 7 FLUKE 700PD3 5 psi O0 0001 34 kPa 0 001 FLUKE 700PD4 15 psi O0 001 103 kPa 0 01 FLUKE 7OOPD5 15 30 psi O 001 100 207 kPa 0 01 FLUKE 7OOPD6 15 100 psi O 01 100 690 kPa 0 07 FLUKE 7OOPD7 15 200 psi 0 01 100 1380 kPa O 1 FLUKE 700P29 3000 psi O 1 20 7 M Pa O 001 0 050 0 010 0 020 0 080 C276 N A C276 FLUKE 700P30 5000 psi O 1 34 M Pa 0 001 0 050 0 010 0 020 0 080 C276 N A C276 FLUKE 700P31 10000 psi 1 69 M Pa 0 007 0 050 0 010 0 020 0 080 C276 N A C276 1 Total uncertainty one year for temperature range O C to 50 C Total uncertainty 1 0 of full span for temperature range 10 C to O C For POO module only compensated temperature range is 15 to 35 C 2 Dry indicates dry air or non corrosive gas as compatible media 316 SS indicates media compatible with Type 316 Stainless Steel C276 indicates media compatible with Hastelloy C276 Use of pressure zero is required prior to measurement or source Maximum overpressure speci
22. trument transmits bursts of data without waiting to be interrogated by a master unit The 744 can take transmitters out of burst mode during test or calibration then later restore them to burst mode Versatile HART protocol support The 744 supports the commands contained in HART protocol version 5 7 With 2 MB of memory the 744 supports a substantial set of HART instructions e Universal commands provide functions that are implemented in all field devices for example read manufacturer and device type read primary variable PV or read current output and percent of span e Common practice commands provide functions that are common to many but not all field devices for example read multiple variables set damping time or perform loop test e Device specific commands provide functions that are unique to a particular field device for example sensor trim The 744 supports these devices Manufacturer Pressure Instruments Temperature Instruments ABB Kent Taylor GOOT 658T ABB Contrans P Hartmann amp Braun AS 800 Series Endress amp Hauser CERABAR M CERABAR S DELTABAR S Foxboro Eckardt TI RTT20 IAP IDP IGP o Honeywell ST3000 Moore Products O Rosemount 3044C 644 3144 Siemens SITRANS P DS SITRANS P ES LD301 TT3O1 YTA 110 and 310 Sensor Trim not supported Why use smart instrumentation Like most process plants your organization is probably facing the dual chal