User Manual CuroCell® AUTO420

image

Contents

1. Follow the operation and maintenance instructions carefully in order to prevent physical injury as well as damage to the product or other equipment 9 2 Labelling Labels on the mattress include product name item number date of manufacture size EAN number and care and washing instructions The mattress also has labels for the top and bottom and head and foot ends as well as space for your own marking such as ward number and name 14 WO 3 Symbol key RE Item number a a Read the user manual pa Do not dispose of with household waste follow recycling instructions Counteracts shearing Machine wash at 95 C E Drip dry o Class Il equipment Double insulated lemme The mattress should be used with the user lying in the lengthwise direction XXXX XX XX Year Month Day _ GRO c 4nna C Manufacturer SELA anna 3 Q mg O 5 oa S O c xo with MDD 93 42 EEC Category gt x S S lt Zi Recommended user weight K Da Heel function Do not machine lt 1 Tumble dry Chlorine Do not dry clean IP__ Xi gt Do not iron lt R gt gJ rating 15 CE marked in accordance User information category IP code ingress protection Foot placement ye Y ue Recycling Lo Placed directly on the bed base Machine wash at 70 C Rol Do not tumble dry Wipe cle
2. Care of sweden Supporting Life User Manual 2015 09 15 CuroCell AUTO420 Alternating pressure mattress Item no user manual 95 001617 7 331345047959 CE Contents Important safety instructions Introduction Product description Assembly Installation Operation Cleaning and reconditioning Storage Maintenance Troubleshooting Technical description Spare parts Options and accessories Warranty Other information oOo wo nO A W 11 12 12 13 14 16 17 18 19 IMPORTANT SAFETY INSTRUCTIONS Read all instructions prior to use WARNING To minimise the risk of fire electric shock or personal injury 1 The product must only be used for its intended purpose as described in this manual The product may not be combined assembled or repaired with other parts such as the pump and mattress accessories or spare parts other than those described in this manual or other documentation from Care of Sweden 2 Do not use the product in wet rooms or where there is a risk of the pump coming into contact with water or other liquids Never handle a product that has come into contact with water liquid Pull the plug out of the electric socket immediately and send the product to an authorised service provider for service 3 DO NOT USE in the vicinity of or in contact with naked flames or hot surfaces such as fires burning cigarettes or hot lamps Although the product has u
3. being changed inadvertently 6 Power off on Turns the power to the control panel on or off NOTE Check that the power cord is properly connected and that the power switch on the side of the pump is in position 1 on 7 Alarm off Q In the event of problems with alarms see section 8 for troubleshooting A Power supply alarm If the pump loses power an audio visual alarm is triggered Turn off the alarm by pressing the Alarm off button B Low pressure alarm A If the mattress pressure is too low an audio visual alarm is triggered Turn off the alarm by pressing the Alarm off button It is reactivated after about five minutes if the pressure has not returned to normal 3 Assembly Installation When unpacking check that no parts are damaged If any damage is found contact Care of Sweden before using the product 1 Place the mattress on the bed base Secure the mattress to the bed using the straps on the underside of the mattress ee Note en Ge e Make up the bed with a sheet to improve user comfort e Check that the CPR valve is closed e If an adjustable bed is used the mattress should be secured only to the moving parts of the bed e Check the cells and push buttons to ensure they are correctly assembled pump on a firm level surface a a 4 2 Hang the pump on the foot end of the bed or place the Place the power cord in the cable holder by opening the fasteners inserting the
4. Care of Sweden or one of its authorised service providers 2 Do not get the pump wet or submerge it in any liquid eT RRB BRE BREE eee De Wipe the pump clean with a damp cloth and mild detergent such as washing up liquid If necessary use an alcohol based disinfectant cleaner such as DAX surface cleaner or similar Perform Virkon or Desisoft can also be used If using a different agent choose one that does not affect the plastic of the pump housing MATTRESS COVER The cover can be wiped clean with a mild detergent such as washing up liquid and water or with an alcohol based disinfectant cleaner such as DAX surface cleaner or similar Perform Virkon or Desisoft can also be used Excrement and blood stains must be removed as soon as possible using cold water Follow local instructions and detergent usage instructions carefully 57 Q ll Bet RY Multi layer covers should be separated before washing The parts should be washed with similar colours Chlorine and phenol based cleaners could adversely affect the PU surface and should zi be avoided If chlorine is used we recommend a maximum concentration of 1 INNER COVER AND COMFORT FABRIC accessories Te O Ill Bx K Clean the affected area using a mild cleaner such as washing up liquid and water or an alcohol based disinfectant cleaner such as DAX surface cleaner or similar Perform Virkon or Desisoft can also be used 5 2 Reconditioning PUMP See section 5 1 f
5. Haesler Ed Cambridge Media Osborne Park Western Australia 2014 1 3 Other information Action plan for pressure ulcers For the best possible results when using this product we recommend a structured and planned approach Examples that can be applied to the use of our products can be found in the Action plan for pressure ulcers guidelines The Mattress Guide This Guide provides guidance to personnel and prescribers when choosing a mattress from Care of Sweden CuroCell AUTO420 is a mattress in function group C This information can be obtained from our website www careofsweden se or ordered from Customer Service see contact information on the last page 2 Product description 2 1 Pump and mattress 1 Mattress 2 CPR Rapid deflation 3 Pump connector 4 Pump Mattress without cover k p gi 7 1 Tube connector 3 Fuses two 6 Hooks 2 Control panel 4 Off On 0 1 7 Air filter 5 Power cord connector 2 3 Pump functions control panel i O care of sweden www careofsweden se 1 Automatic settings gt This pump is equipped with a function that automatically sets the pressure based on the user s weight posture and position There are three ways to use this function 1 The system adjusts the settings automatically when the system has been started 2 Press the button and the system readjusts the automatic settings The system itself will start this fu
6. an Temperature range 10 Spare parts Item no CC 906010 CC 906011 CC 906012 CC 906013 CC 906014 CC 906015 CC 906016 CC 906017 CC 906018 CC 906019 CC 901410 CC 901412 CC 901414 CC 901416 CC 901418 CC 901420 CC 901422 CC 901424 CC 901426 CC 901428 CC 901514 CC 901512 CC 901510 CC 901516 CC 901518 CC 901524 CC 901522 CC 901520 CC 901526 CC 901528 CC 901544 CC 901542 CC 901540 CC 901546 CC 901548 CC 400905 CC 400906 CC 8217 Description Top part Olivia grey Top part Olivia grey Top part Olivia grey Top part Olivia grey Top part Olivia grey Top part Olivia grey Top part Olivia grey Top part Olivia grey Top part Olivia grey Top part Olivia grey Head cell PU Head cell PU Head cell PU Head cell PU Head cell PU Heel cell PU Heel cell PU Heel cell PU Heel cell PU Heel cell PU Safety cell PU for mattress width 80 cm Safety cell PU for mattress width 85 cm Safety cell PU for mattress width 90 cm Safety cell PU for mattress width 105 cm Safety cell PU for mattress width 120 cm Safety cell PU for mattress width 80 cm Safety cell PU for mattress width 85 cm Safety cell PU for mattress width 90 cm Safety cell PU for mattress width 105 cm Safety cell PU for mattress width 120 cm Safety cell PU for mattress width 80 cm Safety cell PU for mattress width 85 cm Safety cell PU for mattress width 90 cm Safety cell PU for mattress w
7. cable and closing the fasteners If the cable holder is not used Place the power cord so there is no risk of tripping over it running over it with the bed wheels or pinching it when raising or lowering the bed ee te 3 Connect the air tubes from the mattress to the side of the ee pump A click is heard or felt when they are correctly i Ey connected pee a _ 4 Check that the power switch on the side is set to 0 off h METT y f S Plug the power cord into an approved socket 220 240 V ex r i CPR U For 80 85 90x200 210 For 105 120x200 210 4 Operation f Note Always read the user manual prior to use 4 1 Start 1 Set the power switch on the side of the pump to 1 on 2 The mattress will start to inflate It takes about 30 minutes to fully inflate the mattress 3 You can start using the mattress when the top left light is lit only the light under the checkmark NOTE Each time the systems starts up it always begins by fully inflating the mattress and automatically adjusting the settings See section 2 3 for more information After automatically adjusting the settings the system switches to alternating mode by default If necessary the user may lie down on the mattress immediately without it being inflated The undermattress bears the weight during the inflation phase If the undermattress is an air filled type it must be fully inflated 4 2 Programme selection D Two different p
8. eral health The mattress should be used with the user lying in the lengthwise direction m on the mattress with the user s feet positioned according to the instructions on the mattress The mattress must always be cleaned between users NOTE 1 For certain users such as amputees the recommended height measurement may not be met Users in these groups may however require other settings as the entire surface is not under load To check functionality see sections 4 2 4 3 and 4 4 2 When used together with positioning pillows read the pillow user manual carefully for correct positioning of the user 3 Whenever this product is used with an evacuation sheet it is the responsibility of the attending personnel to ensure that evacuation can be performed safely 4 Be careful when using bed rails or other protection on the bed so that the mattress is not pinched or damaged 5 The mattress must not be lifted with a person lying on it or used for any other type of transport other than necessary evacuation using an evacuation sheet 6 The mattress is not suitable for use during X ray examinations This is because of the risk of image blurring and artefacts that may lead to diagnostic errors PEP BBR BPR RRR RRR RRR Ree eee 1 National Pressure Ulcer Advisory Panel European Pressure Ulcer Advisory Panel and Pan Pacific Pressure Injury Alliance Prevention and Treatment of Pressure Ulcers Quick Reference Guide Emily
9. ess the light for the low pressure alarm lights and a buzzer sounds See section 8 for troubleshooting Deactivate the alarm by pressing the Alarm off button on the pump control panel If the problem is not resolved within about five minutes the alarm will be reactivated The pump is also equipped with a power loss alarm If the pump loses power an audio visual alarm is triggered Turn off the alarm by pressing the Alarm off button WARNING If the air pressure remains low search for damage to the mattress and the tubes Repair when required or contact Care of Sweden or one of its authorised service providers 4 7 Transport function If the user needs to be moved in the bed the tubes can be connected together detached from the pump The mattress becomes static and is in so called transport mode We recommend using this function for short periods only 4 8 Heel function S e Be Bac SING D The mattress has built in heel relief The last six cells at the foot end are designed to relieve sensitive heels 10 5 Cleaning and reconditioning It is important to follow the instructions below before the product is used for another user Consult your hygiene manager or Care of Sweden for help and instructions if you are unsure 5 1 Cleaning PUMP Pee eee eee ee ee ee ee ee ee eee eee ee ee eee eee eee ee eee WARNING 1 Do not open the pump housing risk of electric shock Service and maintenance must be performed by
10. idth 105 cm Safety cell PU for mattress width 120 cm Pump for mattress width 80 85 90 cm Pump for mattress width 105 120 cm Fuse 2 pack Size 80x200 cm 80x210 cm 85x200 cm 85x210 cm 90x200 cm 90x210 cm 105x200 cm 105x210 cm 120x200 cm 120x210 cm 80x15 cm 85x15 cm 90x15 cm 105x15 cm 120x15 cm 80x13 cm 85x13 cm 90x13 cm 105x13 cm 120x13 cm 75x60 cm 80x60 cm 85x60 cm 100x60 cm 115x60 cm 75x80 cm 80x80 cm 85x80 cm 100x80 cm 115x80 cm 75x70 cm 80x70 cm 85x70 cm 100x70 cm 115x70 cm If other spare parts are required please contact Care of Sweden 16 11 Options and accessories 11 1 Options Comfort fabric For users who are very sensitive to alternating pressure a comfort fabric can be used which further improves the comfort offered by the mattress Sheet holder Helps keep the sheet in place A clip for attaching to the cover six per mattress 11 2 Accessories Comfort fabric CuroCell Item no Size CC 907210 80x200 cm CC 907211 80x210 cm CC 907212 85x200 cm CC 907213 85x210 cm CC 907214 90x200 cm CC 907215 90x210 cm CC 907216 105x200 cm CC 907217 105x210 cm CC 907218 120x200 cm CC 907219 120x210 cm Sturdy transport bag Item no Size 58 900909 80 120 cm Sheet holder each Item no Size CC 9075 Top part LF low friction Top part with increased sliding function can be used for heavy immobile patients 11 3 Other accessories To ensure the best level of co
11. mfort we recommend the use of accessories from Care of Sweden since our products are designed to work well together If you require any accessories or for more information about our products please contact Care of Sweden or visit our website www careofsweden com 17 12 Warranty 12 1 Scope This CuroCell system is covered by a 2 year warranty for manufacturing defects The warranty does not apply to normal wear and tear or product impairment or damage stemming from negligence or improper handling care 12 2 Service and maintenance The CuroCell alternating pressure mattress is a medical device in accordance with MDD 93 42 EEC that is used to prevent and help treat pressure ulcers For this reason we recommend that the pump be regularly serviced and inspected to maintain functionality and performance See below for information on Annual Service Like other technical devices a pump that is properly cared for works better and lasts longer At an Annual Service the pump is thoroughly checked and serviced by our service department Filters and consumables are cleaned or replaced as required For more information please contact Care of Sweden Service and maintenance must always be performed by Care of Sweden or one of its authorised service providers Only use spare parts that have been approved by Care of Sweden The use of non approved spare parts invalidates the warranty Faults and defects regarded as warranty faults a
12. nction if any significant changes occur during use Status lights 1 All three lights lit Automatic settings adjustment under way 2 Only the light under the checkmark left is lit The mattress is ready to use normal mode 3 The light on the right is lit The pump will automatically adjust its settings within five minutes Once the settings have been automatically adjusted the system will revert to the previously used programme However when started it will switch to alternating mode by default To stop the automatic setting adjustments press the Co button 2 Comfort setting The pressure can be increased or decreased one level depending on the user s comfort requirements 3 Programmes Two programmes are available Press the button to switch between programmes 1 Alternating mode Here the pressure alternates between the cells 2 Static mode Here the pressure will be same in all cells This is also called low pressure 4 Maximum pressure Nursing mode With this function the entire mattress is inflated and provides firm support This function automatically reverts to the previous setting after about 30 minutes NOTE The automatic settings are overridden when using this programme 5 Panel lock Press the Panel lock button to lock or unlock the panel The green light is lit when the panel is locked The panel locks automatically after five minutes of inactivity This is to prevent the settings
13. ndergone fire testing it could become damaged if it comes into contact with naked flames 4 The product must be placed and used so that it is not pinched or damaged NB Be particularly aware of pinching when using bed rails Never use the product if the pump power cord or plug is defective the pump casing is damaged or the pump is not functioning properly Contact a service provider for examination and repair 6 Regularly check product functionality with a hand check see section 4 4 7 Continual monitoring is required when the product is used for individuals needing special supervision such as children 8 The mattress is protected by a hygiene cover avoid combining the mattress with other fully covering plastic protectors The cover s vapour permeability may be affected 9 The hygiene cover does not allow liquid or air to penetrate but is vapour permeable vapour is wicked away Make sure that the patient is positioned correctly to avoid the risk of suffocation 10 Avoid damaging the hygiene cover when using sharp objects 11 If the cover is equipped with side handles these are for lifting the mattress All other use takes place under your own liability and is not covered by the product warranty Le Ui 1 Introduction CuroCell air mattresses work according to the alternating pressure principle a technique that is thoroughly tried and tested and has been used for a long time Alternating pressure means that
14. or instructions MATTRESS 1 Disconnect the tubes from the pump and deflate the mattress Also open the CPR valve 11 Remove the cover For more information on cleaning covers see section 5 1 Move the mattress to a clean area that is suitable for cleaning Wipe clean and disinfect the cells all tubes and the CPR valve with an alcohol based disinfectant cleaner such as DAX surface cleaner or similar Perform Virkon or Desisoft can also be used Allow to dry Note The upper and lower layer of air cells can be divided using a quick connector making it easier to clean the cells Wipe the work surfaces clean using a suitable detergent and or disinfectant 6 Put the mattress together If the tubes have come loose from any of the cells they must be BERBER ERR reattached according to the illustration in section 2 1 NOTE Check the hygiene cover each time the product is cleaned If damaged it must be replaced or repaired Also check the pump tube connectors and power cord during cleaning Damaged parts must be replaced or repaired 6 Storage It is advisable to store the mattress and pump in the product bag accessory Handle the stored product with care Do not place any heavy objects on top of it 7 Maintenance 7 1 General 1 2 Check the power cord and plug for damage and scratches Check the mattress protection to make sure it is undamaged and that covers and cells are correctly assembled Check tha
15. re rectified free of charge A warranty claim must be submitted to Care of Sweden before the product is returned 12 3 Life expectance This product s expected life is estimated to be 5 years 18 13 Other information 13 1 Disassembly and recycling Except for certain parts of the pumps energy recovery is possible for almost all material in CuroCell products through incineration at waste incineration facilities Pump A used CuroCell pump must not be disassembled Instead it should be taken to a recycling centre The product is sorted as electronic waste Mattress A used CuroCell mattress must be taken to a recycling centre The product is sorted as combustible waste Care of Sweden complies with its manufacturer responsibility through its affiliation with the REPA register and El kretsen AB For more information please contact Care of Sweden 13 2 Reconditioning Service Repairs For service and repairs contact Care of Sweden or one of its authorised service providers Describe the fault problem before returning the product Telephone number Monday to Friday 8am 4 30pm 46 0 771 106 600 Fax 46 0 325 128 40 E mail service careofsweden se 13 2 Returns and warranty claims Contact Care of Sweden before returning the product The return postage will be paid by Care of Sweden if the fault is covered by the product warranty otherwise by the customer 19 we Manufactured by Care of s
16. rly connected Check the mattress in accordance with Alarm triggered above See sections 4 3 and 4 4 for more Check that the mattress is secured to the bed base with the straps underneath two at the head end and two on each of the long sides This is normal if the alternating pressure programme is selected as the air supply alternates between cells every cycle one cycle ten minutes Check that the pump is correctly attached to the bed Resonance may occur in parts of the bed Remove the pump and listen to see whether there s any difference This can be resolved by placing the pump on a level surface Or by putting for instance a towel between the pump and the bed If your questions are not answered by the above information please contact Care of Sweden or one of its authorised service providers 13 9 Technical description 9 1 Technical specification SPECIFICATION uroCell AUTO420 20V 240 V 50 60 Hz Voltage pe fBtorages isrc sore Shipping t5 C 70 C S po Humidity Operation 10 90 non condensing ee a po ee o y y Fusetype PAL 250M 0 mi R920 KATE S ee O S rE po Storage 10 90 non condensing Classification a CASS II Type BF PAT MATTRESS SPECIFICATION Model o CuroCele auTO420 SS Dimensions WxtxH 80 85 90 105 120x200 210x20 cm Weight keom S NB Care of Sweden reserves the right to modify the product specification at any time BER PRR RRR WARNING
17. rogrammes are available alternating pressure and static low pressure See section 2 3 for detailed information We recommend alternating mode as the default setting There is another programme called Maximum Pressure am This function is used in conjunction with nursing duties changing the user s position and getting the user in or out of the bed 4 3 Comfort setting fine tuning l The air pressure in the mattress can be adjusted up to three levels NOTE When only parts of the mattress are under load such as in the case of amputees it may be necessary to raise the setting using the comfort setting 4 4 Hand check checking the set pressure The hand check is used to ensure that the mattress is working properly This must be performed regularly we recommend once per shift Place a hand between the cells under the user s pelvic region Check that the user doesn t bottom out and rest on the underlying mattress We recommend a margin of one or two fingers If the margin is too small initiate automatic settings adjustment Wait ten minutes and perform a new hand check 4 5 CPR In case of an emergency where CPR cardiopulmonary resuscitation is necessary Quickly pull the CPR strap then loosen the quick connectors on the side of the pump for the fastest possible deflation NOTE Make sure that the CPR strap is properly secured after use 4 6 Alarms In the event of unusually low air pressure in the mattr
18. t the air flow alternates between the different outputs on the side of the pump Check that no tubes or connectors are damaged or pinched Contact Care of Sweden if any spare parts are required 7 2 Changing fuses A iif 1 Remove the plug from the wall socket when replacing fuses B Wiggs Ong ee a M D lt 2 Use a screwdriver to remove the fuse cover Pull out 3 Fra ae 7 N _ y f g ee a j the fuse holder 3 Insert a new fuse of the same type T1AL 250 V and screw the protective cover back into place There are two fuses 12 7 3 Changing the air filter T MNA CY J ae J J i 8 Troubleshooting 1 Loosen the small round protective plate on the back of the pump 2 Remove the filter from the holder 3 Insert the new filter in the holder and replace the protective plate If the pump is used in a very dirty environment the filter should be checked regularly Problem Solution Check that the power cord is connected to a socket Pump does not start Alarm triggered mattress losing air User bottoms out Mattress not stationary Some cells less inflated Noisy pump vibrations can be felt Check the fuses Check that the quick connectors for the tubes on the side of the pump are properly connected Check that the CPR valve is closed Open the mattress and check for any leaks from air cells or tubes Check that all tube connectors are fine and prope
19. the pressure is alternated regularly between the cells in order to simulate the body s own movements Periods are created in which the pressure is lower or completely absent This provides good conditions for the skin to recover and to prevent injury LJ Always read the user manual prior to use 1 1 General information This system is a medical device with CE marking in accordance with MDD 93 42 EEC In accordance with Swedish Medical Products Agency L kemedelsverket regulations in respect of medical devices the manufacturer is obliged to report accidents and near accidents We would therefore be grateful if any accidents and near accidents involving our products could be reported immediately to Care of Sweden This system has been tested and approved according to the following standards IEC 600601 1 IEC 60601 1 2 SS EN ISO 12182 SS EN ISO 14971 SS EN ISO 10993 EN 597 1 EN 597 2 EN 14126 1 2 Intended use CuroCell AUTO420 is an alternating pressure mattress that can be used in all types of medical environments to prevent and help treat pressure ulcers up to category 4 The mattress is intended to be used for people with a recommended minimum height of 120 cm and a recommended user weight of up to 250 kg CuroCell AUTO420 is a combi mattress placed directly on the bed base We recommend changing position regularly The time interval must be assessed by the attending personnel depending on the user s status diagnosis and gen
20. weden Supporting Life Contact Telephone 46 0 771 106 600 Fax 46 0 325 128 40 E mail info careofsweden se Internet www careofsweden com Postal address Care of Sweden AB Box 146 SE 514 23 Tranemo SWEDEN Visiting address Fabriksgatan 5A SE 514 33 Tranemo SWEDEN Deliveries address Byns vag 4A SE 514 33 Tranemo SWEDEN Care of Sweden and CuroCell are trademarks of Care of Sweden AB Care of Sweden AB 2015 20