Assistina 301 plus
Contents
1. The built in air filter in the connecting hose is only fine filter It cannot clean unfiltered and possibly contaminated compressor air Dirty fine filters must be replaced 13 Commencing operation Initial filling of the system Pull the drip guard off the maintenance attachment and store it safely Hold a cloth or tissue in front of the maintenance attachment Press the program button about 15 times for 2 seconds with a pause of 2 seconds between each actuation am If no oil can be seen on the cloth or tissue and no cleaning fluid can be detected by its smell alcohol the supply lines are not yet completely full Press the program button 10 times for 2 seconds 14 8 Test Do not connect any transmission instrument turbine air motor or dental scaler during the test If the adaptor is fitted on the maintenance attachment remove it Press the program button once Place a fleece inlay in the for 2 seconds waste disposal tray Close the cover and waste disposal tray Check that the fleece inlay is damp to ensure correct function If itis not damp repeat the process to 3 times Dispose of saturated fleece inlays as with cloths for surface disinfection 15 9 Function check Function display A Continuously monitor the function displau gt Press the program button for 2 seconds A gre
2. Turbines with gt 2 hole or 3 hole Borden connection and gt spray air to drive air connection Screw the turbine adaptor to the turbine Insert the turbine adaptor onto the motor attachment Turbines with standard 4 hole connection Screw the turbine adapter onto the respective turbine 19 Maintenance Push the appropriate adaptor onto the maintenance attachment up to the limit stop With its spray exit nozzle directed downwards push the turbine with quick release coupling onto the adaptor until it engages Close the cover and the waste disposal tray Test run the turbine When the turbine has been wiped over it is ready to be sterilized Press the program button once for about 2 seconds Open the cover at the end of the maintenance cucle Press the red release button of the maintenance attachment while at the same time removing the turbine from the adaptor 20 Maintenance for air motors and air driven dental scalers The maintenance program will be completed in about 35 seconds Use the turbine adaptor REF 02685000 with standard 4 hole connection gt Set the rotary ring of the air motor to clockwise or anticlockwise running no maintenance will take place in the Zero setting gt Carry out maintenance as for turbines see Page 20 21 Maintenance of the FG chucking system of FG contra angle handpieces A Clean the chuck at
3. least once a week Remove the rotary instrument Use the T1 adaptor 02693000 With its spray exit nozzle directed downwards push the T1 adaptor onto the maintenance attachment Initially keep the cover open Push the instrument onto the adaptor from underneath Support the instrument at the edge of the cover Press the program button once for about 2 seconds Press the instrument against the adaptor for about 10 seconds Close the cover Allow the remainder of the maintenance cycle to complete 22 11 Cleaning disinfection recommends external cleaning of the Assistina once a week or as required Use W amp H cleaning fluid to clean the outside of the case as required Squeeze the cover slightly Remove the cover from its fixing Use W amp H cleaning fluid to clean the inside of the case Clean the waste disposal tray in a thermo washer disinfector 23 Use only original W amp H accessories spare parts 1e W amp Haccessorles accessories approved byW amp H Optional accessories 02676400 Drip guard for maintenance attachment 02679000 Adaptor for removable contra angle handpiece heads 02680600 Seal 02685000 4 hole turbine adaptor 02689300 Funnel white 02689400 Funnel green 02690400 Adaptor for all W amp H products with Roto Quick system 02691000 Adaptor for Sirona T1 Classic transmis
4. with ISO coupling esee 10 nennen entente nnne 18 Maintenance C E 19 20 Maintenance of air motors and air driven dental scalers 22 nennen nennen 21 Maintenance of transmission instruments T1 TITAN T1 Turbines and SIRONA Motor 51 21 Maintenance of the FG chucking system of FG contra angle 2 22 11 Cleaning anu PIED CIS 23 12 WE amp H accessorles ettet te n eet tu edd ae SR 24 25 13 Seri C ees cuter suentetavenentet 26 30 Technical datar M 31 iiie aaa ata ed dic i t ced DO E EL EE aod 32 WSH COMPACTS NE 33 symbols A WARNING Risk of injury W amp H Service ATTENTION To prevent damage occurring gt Data Matrix Code for product identification e g in hygienic maintenance process General explanations with no danger to persons or property A Highly flammable Avoid sources of ignition do not smoke 1 Introduction For your safety and the safety of your patients This User Manual explains how to handle yo
5. PEOPLE HAVE PRIORITY Instructions for use CWSH OSSISuno 301 plus Contents WEHsumbolS x3 meh em ee Rn eoe e etiim E TD 3 15 UlnitrOd UICCIODIG nee teo ert RO rte mI PUR nen spe E ee E tute uot ec eue PLUR ERE CES NER PLUR ERN AR d 4 5 e Unpatcking ne ni eni p pt pr B ente a e d e ERU ER eel oe a er e ce aea 6 3 Pack ge contents 4 2 2 0 A 4 Safety precautiornis iiie a ntes tete reti e t eeu nik fes Du eee 8 5 Description of the front side tee etre En sohn eoe n Ro RR RE eia Re Rug 9 6 Description of the rear slde iuueni pat Shes ne e EO ia re he ERE RE ARE RE CH Ran PREND et doe 10 f COMMENCING operation First USE 5 ttt rite nne ANG GR 11 Commencing operation Filling with service oil cleaning 8 12 Commencing operation Connecting the compressed 13 Commencing operation Initial filling of the system 14 15 9 Function check Function display function check of the mechanical drive 16 10 Maintenance edet ed c Ded re id d Ded Lr re a PD edens 1 Maintenance of transmission instruments
6. c a dec com au 33 Manufacturer W amp H Dentalwerk B rmoos GmbH Ignaz Glaser Stra e 53 5111 B rmoos Austria t 43 6224 6236 0 f 43 6224 6236 55 office wh com wh com Form Nr 50458 AEN Rev 015 09 06 2015 Subject to alterations
7. channels 1 Maintenance of transmission instruments with ISO coupling The maintenance program will be completed in about 35 seconds Pull the cover down Press the red release button and pull off the turbine adaptor Keep the adaptor in the recess With its spray exit nozzle directed downwards push the transmission instrument onto the maintenance attachment up to the limit stop Close the cover and the waste disposal tray Test run the transmission instrument When the transmission instrument has been wiped over it is ready to be sterilized Press the program button once for about 2 seconds Open the cover at the end of the maintenance cucle Press the red release button of the maintenance attachment while at the same time removing the transmission instrument 18 Maintenance for turbines The maintenance program will be completed in about 30 seconds Use the turbine adaptor REF 02685000 with standard 4 hole connection Pull the cover down 02685000 W amp H turbines Screw together the turbine adaptor and the adaptor Adapters not suitable for use a with turbine hose 04213200 02690400 02014500 02685000 Other turbines 02685000 E Wer Screw the turbine adaptor to the relevant quick release bbe coupling e g adapter fe 02692000 04713200 e Mr Fe
8. d 6 Description of the rear side Suction filter Quick release coupling for L SS C SS gt compressed air 10 2 Commencing operation First use Service oil must be at room temperature when it is poured in Never reverse funnels white to white green to green and store them in a dust free environment Check the filling level indicator gt Use only clean filtered compressed air Unscrew the two caps and remove the two transport seals malfunction The W amp H cleaning fluid is aldehyde free and nontoxic It is biodegradable in waste water W amp H service oil represents absolutely no danger to health correct disposal of residues is required 11 Commencing operation Filling with W amp H service oil cleaning fluid Unscrew the cap from the white storage container Insert the white funnel Add 250 ml of cleaning fluid Add 250 ml of service oil Close the storage containers for the service oil and cleaning fluid Unscrew the cap from the green storage container Insert the green funnel 12 Commencing operation Connecting the compressed air Fit the suction filter so that it is leak tight 02697000 Push the quick release coupling of the connecting hose into the coupling socket on the Assistina Connect the compressed air line using the 1 8 screw connection
9. e have been followed As manufacturer W amp H is liable for material or manufacturing defects within a warranty period of 12 months from the date of purchase We accept noresponsibility for damage caused by incorrect handling or by repairs carried out by third parties not authorized to do so by W amp H Claims under warranty accompanied by proof of purchase must be sent to the vendor or to an authorized W amp H service point The provision of service under warranty extends neither the warranty period nor any other guarantee period W amp H contacts Find you nearest W amp H service partner at http wh com Simply go to the menu option Service for full details Alternatively please contact W amp H UK LIMITED 6 Stroud Wood Business Centre Park Street St Albans Herts AL2 2NJ United Kingdom t 44 172 874990 f 44 1727 874628 E Mail technical uk wh com W amp H Impex Inc 6490 Hawthorne Drive Windsor Ontario 1J9 Canada t 1800 2656277 1519 9446739 f 1 519 9746121 E Mail service ca wh com W amp H Impex Inc 14300 Henn Rd Dearborn MI 48126 USA t 1800 2656277 1519 9446739 f 1 519 9746121 E Mail service us wh com W amp H DENTAL MECHANIK ASIA PTE LTD 18 Boon Lay Way 10 94 Tradehub 21 609966 Singapore t 65 63168370 f 65 63168371 E Mail service sg wh com A DEC AUSTRALIA CO INC Unit 8 5 9 Ricketty Street Mascot NWS 2020 Australia t 61 2 83324000 f 61 2 83324099 E Mail a de
10. en ball for the cleaning fluid appears in the right hand window Release the program button green ball for the service oil appears in the left hand window After a few seconds both balls will drop down again Correct operation of the maintenance process is indicated by these two function displays If ever the balls disappear during a program cucle this indicates a malfunction Function check of the mechanical drive N 7 gt a transmission instrument with a rotary instrument inserted onto the Assistina maintenance attachment gt Press the program button for 2 seconds I The maintenance cycle begins and the rotary instrument turns If the rotary instrument does not turn there is a malfunction 16 10 Maintenance First carry out the external disinfection and cleaning Carry out daily maintenance midday and evening prior to sterilization Use the adaptor REF 02679000 for maintenance of transmission instruments in those cases where the manufacturer has allowed for removal of the treatment heads With sterilization after every patient Press the program button once With maintenance twice a day Press the program button twice on each occasion accurate dosage of W amp H cleaning fluid service oil is provided for these maintenance processes A Additional pressing of the program button can cause too much cleaning fluid to enter the spray air and coolant
11. ions have been observed The Assistina must be used in compliance with this User Manual gt The Assistina has no parts which are repairable by the user Assembly modifications or repairs may be carried out only by skilled persons authorized by W amp H 2 Unpacking Have the User Manual at hand Take out the white and green funnels Remove the turbine adaptor Lift out the Assistina in the fibre board box Open the box at the tucked in closure and remove the air hose from the cover Remove the initial accessories Take out the service oil cleaning LY fluid suction filter and the optional accessories from the side compartments The W amp H packaging is environmentally friendly and can be disposed of through specialist recycling organizations However we recommend that you keep the original packaging V M M VV V VOM 3 Package contents W amp H Assistina cleaning and maintenance unit REF 00030125 00030140 Suction filter complete 4 hole turbine adaptor REF 02685000 Funnel white REF 02689300 Funnel green REF 02689400 Air connection hose 2 m REF 02692000 complete Initial provision of 5 layers of fleece REF 02690300 2 spare seals REF 02680600 and 2 spare O rings REF 02695200 User Manual with warranty card 4 Safety precautions A It is essential that you comply with the following instructions Place the unit only in adequately ventilated premise
12. orrect maintenance Once a year gt Carry out a function check Page 16 gt Check the fine filter in the air connection hose for contamination 26 Service Cam The O rings can be exchanged without the need for special tools Grease free fingers will facilitate this exchange gt Firmly squeeze the O ring between thumb and forefinger to form a loop gt Pull off the O ring gt Push on new O ring No 02695700 2 replacement O rings No 02695700 are attached here 22 Service Renewing the filter pack Do not touch the vents of the suction filter risk of damage Use the W amp H Assistina for maintenance only when the suction filter with a fully serviceable filter pack is fitted gt Renew the filter pack in the suction filter after approx 2 000 maintenance cycles corresponds to the consumption of a 500 ml bottle of W amp H service oil Pull out the suction filter Take the filter pack out horizontally towards the rear of its packaging Press the filter pack on the frame in the filter case Refit the suction filter so that it is leak tight Press out the filter pack using a suitable implement Ensure that it is placed in the correct waste disposal 28 Service Returns gt If you have any questions please contact your local dental dealers or one of our service points see Page 33 Always use the original packaging for
13. return transport 2 transport seals No 02680600 are attached here 29 Service Pull out the suction filter horizontally Disconnect the air connection hose Empty the cleaning fluid and service oil from the storage containers Do not empty the containers simultaneously The cleaning fluid and service oil must not be mixed Fit a seal in each filler hole Screw on the two caps Before emptying the first container fit the transport seal on the second container Fit the drip guard No 02676400 on the maintenance attachment If its packaging no longer available Ask your dealer for W amp H packaging or use secure substitute packaging 30 14 Technical data Height Width Depth with filter Weight Air consumption Noise Flow pressure Capacity 223 mm 190 mm 415 mm 2 70 kg approx 60 l min 63 dB A at 50 cm distance 4 bar to 10 bar pressure regulation provided by integral automatic pressure regulator 250 ml service oil sufficient for approx 3 500 maintenance cycles 250 ml cleaning fluid sufficient for approx 2 500 maintenance cycles 31 This W amp H product has been manufactured with great care by highly qualified specialists A wide variety of tests and controls guarantee faultless operation Please note that claims under warranty can only be validated when all the directions in the Instructions for us
14. s Keep your hands out of the inside of the unit during maintenance operations Fit the transmission instrument turbine turbine adaptor air motor or dental scaler only when parts are no longer moving Damage to the drive driver assembly transmission instruments Always close the cover and waste disposal tray before starting up the unit Keep cleaning fluid service oil away from the eyes Do not swallow cleaning fluid Keep service oil away from the skin Cleaning solution and its residues are flammable Flash point 25 C Avoid sources of ignition Do not smoke Use only suitable tools Comply with the manufacturer s instructions for the maintenance of dental transmission instruments turbines air motors and air driven dental scalers Improper use Improper use in addition to incorrect assembly installation modification or repairs ofthe Assistina or non compliance with our instructions invalidates all claims under warranty and any other claims 5 Description of the front side Filler hole Filler hole storage container storage container for service oil for cleaning fluid Recess for adaptor Button to cleaning press 2 sec maintenance process wsvee 4 _ Fill level of cleaning fluid Fill level of service oil Indicator window Cover for service oil Indicator window for cleaning fluid Waste disposal tray service oil W amp H cleaning flui
15. sion instruments 02692000 Adaptor for Sirona T1 T2 T3 turbines 02693000 Adaptor for cleaning FG chucking system and others 05204600 Entran contra angle head 02692000 Airconnection hose 2 m with screw connection 6 x 4 mm 02699901 Coverforlong transmission instruments 04213200 Adaptor for Multiflex System Multiflex is a registered trademark of KaVo Medizintechnik GmbH Germany 07014500 Adaptor Assistina Borden 24 W amp H accessories 02625200 02695200 02620500 02625910 02680200 02680500 02690300 02694500 Air filter for air connection hose 02692000 0 for maintenance attachment Filter pack 500 mI service oil 1000 cleaning fluid Pack of W amp H service oil and W amp H cleaning fluid Initial provision of 5 layers of fleece Fleece pack of 20 25 13 Service Regular testing of Assistina Regular testing of function and safety is necessary With each application Check the function display Once a week gt Check the filling level displays f a gt Check the O rings on the maintenance attachment gt Change the O rings in the event of visible defects or every two months gt Replace any missing O ring immediately Firmly squeeze the O ring between thumb and forefinger and pull it off Push on a new O ring Only 0 in good condition and correctly positioned will ensure consistently safe and c
16. ur W amp H product We must however warn of possible dangerous situations Your safety the safety of your team and of course the safety of your patients is of paramount importance to us A For this reason it is essential to comply with the safety precautions on Page 8 Intended use Unit for cleaning the spray channels and lubricating the moving parts of dental transmission instruments turbines air motors and air driven dental scalers Misuse can damage the Assistina and involves risk and danger to user and third parties Care and maintenance of the articles specified with W amp H Assistina does not provide disinfection or sterilization User qualifications The W amp H Assistina unit shall be used only by those personnel who have received professional and practical training In the development and design ofthe W amp H Assistina we assumed the target group of users to be dental surgeons andtheir assistants 4 Introduction ce Production according to EU Directive This medical product has been designed and manufactured according to EU Directive 93 42 EEC and applies to the unit gt Assistina inthe condition as supplied by us This declaration does not apply to unforeseen internal or external attachments or the like Responsibility of the manufacturer W amp H Dentalwerk B rmoos accepts responsibility for consequences affecting the safety reliability and performance of the Assistina only when the following direct