Alaris Syring e Modu le v 9
Contents
1. Verify appropriate pressure sensing disc is in use and functioning properly Verify syringe plunger movement is not impeded If syringe is not empty and above actions do not correct alarm contact qualified service personnel gt D mi O wo lt m Ce D O Q g lt D gt D cf D c T gt a amp lt Channel Error Errors Meaning Error detected Operation stops on affected module Response To silence alarm and continue operation of unaffected modules press CONFIRM soft key Replace module with an operational instrument as required Service by qualified personnel is required Syringe Calibration Required Error on infusing module indicating calibration is required Infusion stops on affected module CALIBRATE scrolls in Message Display To silence alarm and continue operation of unaffected modules press CONFIRM soft key Replace instrument with an operational instrument as required Service by qualified personnel is required Syringe Driver Head Error Noninfusing module with plunger grippers open senses excessive pressure being applied downward on Drive Head OCCLUSION scrolls in Message Display To silence alarm and continue normal operation press CONFIRM soft key P00000067
2. ALL option is disabled for the VTBI a syringe volume estimate will be dispalyed after the VTBI soft key is selected Selecting Pressure Limits Pressure Sensing Disc Installed 1 Press CHANNEL SELECT key 2 Press OPTIONS key 3 Press Pressure Limit soft key 4 To enter a new pressure limit value press Change Value soft key and use numeric keys If Auto Pressure feature is enabled press Auto Pressure soft key Note If Auto Pressure is selected and current pressure is 100 mmHg or less system adds 30 mmHg to current pressure to create a new alarm limit e Greater than 100 mmHg system adds 30 to current pressure to create a new alarm limit 5 Press CONFIRM soft key Pressure Sensing Disc Not Installed Press CHANNEL SELECT key 2 Press OPTIONS key 3 Press Pressure Limit soft key 4 Press appropriate soft key Low Med High 5 Press CONFIRM soft key gt D mi v U lt O fo Q D lt To Alarms and Troubleshooting Alarms Channel Disconnected Meaning Channel s disconnected while in operation or there is a communication problem Response To silence alarm and clear message from screen press CONFIRM soft key Reattach channel if desired ensuring it is securely clicked into place at Channel Release Latch If alarm is still present replace module with an operational instrument Check Syringe Plunger grippers op
3. Cool Gray 7 U Eol Text PMS 294U QUICK REFERENCE GUIDE fwanninas PAN e To ensure proper operation of the Alaris System you must be familiar with related features disposables administration sets set up and programming e This guide is not intended to be comprehensive instructions for the setup and operation of the Alaris System For complete instructions along with Warnings and Cautions refer to Alaris System User Manual v9 Programming an Infusion using Guardrails Suite MX Protection The programming user interface is identical for the Alaris Pump Module and the Alaris Syringe Module Set Priming Loading and Removal Manual Priming of Non Dedicated Sets 1 Ensure set is not connected to patient and that device is level with patient 2 If used hold pressure sensing disc between two fingers and invert disc so patient side is up 3 Slowly prine set while gently massaging disc to ensure disc does not become over or under filled and all air is expelled 4 When priming is complete clamp set 5 Load syringe into device and install pressure sensing disc for proper placement information refer to the Alaris User Manual 6 Program infusion 7 Release clamp and press Auto START CareFusion San Diego CA 800 854 7128 2003 2014 CareFusion Corporation or one of its affiliates All rights reserved 8 Attach set to patient Note Starting infusion before attaching s
4. acing device Apply upward pressure until disc snaps into disc housing Note At the start of an infusion program the system prompts to select and confirm syringe type and size Ensure syringe manufacturer and size correctly identify the installed approved syringe Mismatches may cause under infusion or over infusion to the patient Drive Head Gripper Control Drive Head Release closed position Plunger Grippers closed position Barrel Flange Grippers Syringe Syringe Barrel Barrel Clamp Clamp A To prevent free flow close set tubing clamp before loading or unloading syringe Removal 1 Close tubing clamp 2 Open plunger grippers and syringe barrel clamp silence safety alarm 3 Remove syringe apply downward pressure to remove disc from housing Near End of Infusion When infusion is near completion Near End of Infusion is enabled e Main display alternates between NEAR END and remaining VTBI e Yellow standby status indicator flashes e An audio prompt sounds which requires being silenced just once and will not reoccur following initial silencing All Mode Enabled If the ALL option is enabled for the VTBI entire contents of the syringe will be delivered unless your facility has established a KVO volume Note KVO is not available for intermittent infusions All Mode Disabled If the
5. ened during Securely lock plunger grippers infusion and then closed Infusion press CHANNEL SELECT key stops on affected module and reselect syringe Syringe barrel clamp opened Securely lock syringe barrel during infusion and then closed clamp and press RESTART key Infusion stops on affected module Syringe plunger not captured Check for potential siphoning while in idle state System alarms Ensure administration set clamp after 30 seconds to indicate roller slide is in closed position potential siphoning condition Securely lock plunger grippers over syringe plunger Occlusion Increased back pressure Clear occlusion Press sensed while infusing Infusion stops on affected module RESTART key or press CHANNEL SELECT key and then START soft key Pressure Disc Installed Pressure sensing disc installed during an infusion Infusion stops on affected module Press CONFIRM soft key and restart infusion Pressure Disc Removed Alarms and Troubleshooting Alarms Continued Meaning Pressure sensing disc removed Infusion stops on affected module Response Reinsert pressure sensing disc and press RESTART key Syringe Empty Syringe is empty If syringe is not empty other possibilities are e Pressure sensing disc inappropriate defective Syringe plunger travel impeded e Pressure transducer defective Set up new infusion or press CHANNEL OFF key
6. et to patient minimizes any potential bolus that could be released from pressure build up in set due to normal priming and syringe loading Device Priming 1 Ensure set is not connected to patient and that device is level with patient 2 Load and select syringe type and size 3 Program infusion parameters but do not start infusion Press OPTIONS Select Prime Set with Syringe If used hold pressure sensing disc between two fingers and invert disc so patient side is up 7 Press and hold PRIME while gently massaging disc to ensure disc does not become over or under filled and all air is expelled 8 Release PRIME when fluid flow is observed at end of set 9 Install pressure sensing disc Bu 10 Press EXIT 11 Press START 12 Attached set to patient Note Infusions programmed to infuse at low flow rates lt 1 0mL hr may encounter a delay upon start up When manual priming is complete use Prime Set with Syringe feature to decrease potential start up delays Loading Infusions 1 Open syringe barrel clamp until it clears syringe chamber 2 Twist gripper control clockwise and raise drive head to fully extended position 3 Insert syringe barrel flange between barrel flange grippers 4 Lock syringe in place by closing barrel clamp 5 Twist gripper control clockwise lower drive head lock plunger in place with plunger grippers 6 Insert pressure disc if used with membrane f