Home
PDF - Scientific Industries, Inc.
Contents
1. REVERSE mode toggles the shaker between operating in only one direction and switching directions at regular intervals To initiate the REVERSE mode press the NORMAL REVERSE button The REV INT indicator will light Fig 8 Press the RUN STOP button to begin the mixing action ORBITAL Please note that if you set acceleration to lower values in combination with short reversing or cycle A times the unit will limit the maximum speed setting to values that are achievable ie reversing time of Fig 8 5 seconds with acceleration deceleration of 10 RPM second unit will only allow maximum speed of 25 RPM 2 5 seconds of acceleration at 10 RPM second and 2 5 seconds of deceleration at 10 RPM second Also note that for safety reasons when pressing the RUN STOP button to end a cycle instead of ORBITAL letting the timer stop the unit maximum deceleration 100RPM second is used GENIE HRS MIN In between shaking cycles the platform can move freely To lock the movement of the platform for ease of loading vessels press and hold the RUN STOP button and simultaneously press the NORMAL REVERSE button The shaking platform will lock See Fig 9 for display readout Initiating a Fig 9 a shaking cycle will exit this mode INTERVAL Orbital Genie can be programmed to run for an arbitrary length of time 1 minute to 99 hours pause for a length of time also 1 minute to 99 hours and repeat This repeating can
2. go on indefinitely or stop when the timer runs down to zero To initiate the INTERVAL mode press the INTERVAL button The REV INT indicator will light Fig 8 The right led will display On The left led will begin flashing and display the time that the unit will run between pauses Set the time you would like 1 min to 99 hrs and press the interval button again The right led will display OFF The left led will flash and display the time that the unit will pause between running Set the time you would like 1 min to 99 hrs and press the interval button again Press the RUN STOP button to begin the mixing action SPECIFICATIONS The ORBITAL GENIE Shaker is classified as Installation Category 2 Environmental 0 C 38 C 32 F 100 F 95 Humidity max MODEL POWER REQUIRED AMPS MODEL POWER REQUIRED AMPS SI 1500 120V f 230V British Plug SI 1600 120V 230V Swiss Plug SI 1700 120V 230V Swiss Plug SI 1501 230V No Plug 230V Swiss Plug SI 1601 230V No Plug 230V Australian Plug 230V Australian Plug 230V Australian Plug 100V 100V 100V SI 1701 230V No Plug SI 1502 230V Euro Plug SI 1602 230V Euro Plug SI 1702 230V Euro Plug SI 1503 230V British Plug SI 1603 230V British Plug NNN AH A A AS AS St an i 1ondadaono O h h ee i i ee ee oOo OOO G o Weight 21 Kg 46 lbs Base Dimensions DxW 590 x 500mm 23 x 20in Orbit 35mm diameter Care and Handling Your ORBITAL GENIE Shaker should be given
3. is compressed against the vessels Fig 4 To release the ratcheting clamps pull the release arm and slide away Fig 5 ORBITAL GENIE for Standard Flask Clamps SI 1600 The shaker is supplied with a platform for accepting various size flasks or beakers Fig 2 Flask clamps attach to the platform with captive mounting screws Flask clamps are sold separately Position clamps evenly distributed on the platform Uneven loading may result in excessive vibrating of the shaker and support table Should this occur reduce the mixing speed or turn the shaker off and redistribute the load of the vessels ORBITAL GENIE with Adhering Mat SI 1700 The shaker is supplied with an adhering rubber mat for use with bottles and dishes that cannot be held with conventional clamps Fig 3 Patented material is naturally adhering Rinse clean to renew adhering properties Flasks clamps can be used on this unit by removing the adhering mat OPERATION Power on the ORBITAL GENIE rocker switch on lower right side of housing The software version number will briefly be shown on the display After 5 seconds the display will show time and speed settings TIME Time is displayed in the left window Press the UP DOWN arrow button beneath the TIME window to set the desired cycle time from 1 minute to 99 hours 59 minutes Press the RUN STOP button to begin the shaking action The shaker will ramp up to speed and run until the set time has elapsed Upon time
4. models 220 240 VAC for the SI 1501 SI 1502 SI 1503 SI 1504 SI 1506 SI 1601 SI 1602 SI 1603 SI 1604 SI 1606 SI 1701 SI 1702 SI 1703 SI 1704 and SI 1706 models 100 VAC for the SI 1506 SI 1606 and SI 1706 models Illuminated numbers will appear in the TIME and SPEED windows indicating that the unit is ON and ready for use The work surface on which the shaker is placed must be level stable and non slip ORBITAL GENIE with Universal Ratcheting Clamps SI 1500 The shaker is supplied with two ratcheting and one non ratcheting cushioned clamps Fig 1 shown with additional ratcheting clamp The non ratcheting clamp acts as a stationary stop for securing the vessels It should be located at the far end or middle of the platform rods and locked in place with the thumbscrews Position the two ratcheting clamps on the platform rods with sufficient space to fit the vessels Place the vessels so that they do not touch each other and are seated firmly against the non ratcheting clamp Position vessels evenly distributed on the platform Uneven loading may result in excessive vibrating of the shaker and support table Should this occur reduce the speed or turn the shaker off and redistribute the vessels The clamping platform can accommodate 250mL 2 liter vessels with a 26 lbs load capacity It is very important to regularly monitor shaking cycles to ensure that the vessels remain firmly secure Press the ratcheting clamps until the foam
5. 5 EB 1500 500 TPP0OO06 PP 1500 400 HWP0076 NTP0004 BRPOO1 SCP0076 WAP0031 WAP0038 HWP0075 OM 1500 208 WAP0037 SCP0077 SCP0078 O 18 16 A F 15 Description 500 mL Flask Clamp 1 L Flask Clamp 2 L Flask Clamp Adhering Mat Double Sided Adhesive Universal Ratcheting Clamp 20 x 2 inch Screw Driver Motor Stepper Motor Keypad Orbital PC Board Gasket Enclosure V2 Nylon Spacer No 10 Hex Nut Bearing 6 32 x 3 8 long screw Washer Special Flat Washer Ribbed Bushing Mounting Cup Nylon Washer Shoulder Screw 1 4x 1 1 4 Shoulder Screw 1 4x 1 1 2 LTP0071 Rev B ORBITAL GENIE BENCHTOP ORBITAL SHAKER OPERATING INSTRUCTIONS Models SI 1500 through SI 1706 E gt B Scientific Industries Inc Scientific Industries Inc 80 Orville Drive Suite 102 Bohemia NY 11716 U S A Phone 631 567 4700 Fax 631 567 5896 Toll Free 888 850 6208 Customer Service custsvc scientificindustries com www scientificindustries com The ORBITAL GENIE is a large capacity multi vessel orbital shaker Speed and time are digitally controlled for reproducibility and accuracy with true speed display OPERATING INSTRUCTIONS We recommend you retain the original packaging for 90 days in case you need to return the product for any reason to your distributor or Scientific Industries Plug the line cord into a properly grounded electrical outlet 120 VAC for the SI 1500 SI 1600 and SI 1700
6. Parts Assembly List To order parts for the ORBITAL GENIE Shaker Contact your local distributor or visit www scientificindustries com Please specify Part No quantity and electric voltage 35 SX Se e B IDLER COUNTER BALANCE ASSEMBLY Item No Part No 1 OK 1500 901 2 OM 1500 200 3 ESP0006 ESP0007 4A 318 0510 02 4B OK 0246 901 4C OK 0256 901 4D OK 0266 901 4E OK 0276 901 4F OK 0286 901 4G ECP0021 5 EPP0023 6 OM 1500 203 7 OM 1500 202 8 OM 1500 207 9 OM 1500 206 10 OM 1500 204 11 OM 1500 215 12 OM 1500 205 13 SI 1575 14 OM 1500 214 15 OM 1500 218 16A SI 1610 16B SI 1611 16C SI 1612 Sl 1600 SI 1700 4 PARTS S I 1500 4 PARTS IDLER COUNTER BALANCE ASSEMBLY Description Feet 4 Bottom Plate Switch 100 120V 230V 120V Line Cord 230V Line Cord No Plug 230V Line Cord Euro Plug 230V Line Cord British Plug 230V Line Cord Swiss Plug 100V Line Cord Australian Plug plug only Power Supply Label Bracket Mounting Plate Counter Balance Motor Counter Balance Idler Top Platform Threaded Spacer Rod Ratchet Non ratcheting Clamp Non slip Rubber Pad Flask Clamp Top Platform 100mL Flask Clamp 125 150mL Flask Clamp 250mL Flask Clamp Orbital Genie is a registered trademark of Scientific Industries Inc Scientific Industries Inc 2014 a Part No Sl 1613 Sl 1614 Sl 1615 Sl 1617 SI 1616 SI 1510 SCP0074 EMP0010 EMP0009 EL 1500 57
7. elapse an alarm will beep every 15 seconds and the TIME set value will blink Beeping will end after 5 minutes but the TIME display will continue to blink Press any button to exit the cycle end state NON TIMED CONTINUOUS Press and hold the DOWN arrow button beneath the TIME window until the illuminated display reaches 00 00 Press the RUN STOP button to begin the shaking action The timer will count up from 00 00 to 99 59 and then wrap back to 00 00 SPEED Set the speed control by pressing the UP DOWN arrow beneath the SPEED window to any value between 10 RPM and 300 RPM REVERSING and ACCELERATION DECELERATION To modify the ACCELERATION DECELERATION ramp speed and REVERSING TIME press and hold the NORMAL REVERSE button for 2 seconds The TIME and SPEED displays will flash Fig 7 The left display is reversing time The right display is acceleration deceleration rate Use the UP DOWN arrows on the left side to change the time between reversing minutes and seconds from one second to 99 minutes and the arrows on the right side to change the acceleration deceleration speed value from 10 to 100 RPM SEC The new value will be applied immediately and memorized upon key release This mode flashing will be exited automatically after 10 seconds or by pressing the NORMAL REVERSE button once By modifying these values you can change the nature of mixing action from smooth vortexing to sloshing aeration or anything in between
8. the care normally required for any electrical appliance Avoid wetting or unnecessary exposure to fumes The finish can be washed with water after unplugging and soap or detergents using a cloth or sponge Keep the unit clean by immediately blotting any spills Replacement parts are available through your laboratory equipment distributor or directly from Scientific Industries Refer to Parts Assembly List PF cannon Do not lift your ORBITAL GENIE Shaker by the clamping platform CAUTION Failure to follow operating instructions can compromise the user s safety CAUTION Unplug from power before cleaning Do not immerse