Guardian 1000 Digital Airflow Monitor Installation Guide

image

Contents

1. Install Elbow Locknut and Washer Figure 1 Hole Plug and Mounting Screws Location Airflow Sensor Display Digital 1000 Monitor Figure 2 Digital 1000 Airflow Monitor Details Labconco Instruction Sheet 3909400 Rev F ECO F288 2 of 7 Accessory Light Enclosure Hose Cover Monitor Figure 3 Digital 1000 Airflow Monitor Installation Labconco Instruction Sheet 3909400 Rev F ECO F288 3 of 7 Digital 1000 Calibration 1 Calibrate the airflow monitor according to the instruction manual that comes with the kit To successfully calibrate it will be necessary to change the face velocity by adjusting the airflow exhaust volume The exhaust volume can be adjusted with the speed control on the FilterMate or by using an adjustable damper on the exhaust blower Typical calibration conditions are set at face velocity air sample differences of at least 20 feet per minute The airflow monitor is factory set to be calibrated with a difference of at least 50 fpm and can be changed by changing the lower higher air sample difference The following suggested in flow face velocity speeds are recommended to successfully calibrate Typical low air alarms are set 10 20 fpm below operational speeds Follow Step 2 below and review the Labconco 1000 Alarm User s Manual that comes with the airflow monitor Low Air Alarm Enclosure Operating Low Calibration High Calibration Set Point In flow Speed Set P
2. MENU adjust the red low alarm to the desired setting See the chart below for range and suggested settings Then adjust the yellow CAUTION or WARNING to the desired setting See the chart below for range and suggested settings Then adjust the CAUTION or WARNING air reset to 3 fpm This sets the alarm condition CAUTION WARNING LOW ALARM Low Calibration High Calibration YELLOW LED Setting RED LED Set Point Set Point Setting Sash Open or Enclosure Range Suggested Range Suggested fpm Operating user defined Set Point user defined Set Point 8 Sash 10 Sash Inflow Speed fpm fpm fpm fpm Height Height Sash Closed 43 53 SI 40 50 48 27 23 60 fpm 63 73 67 60 70 64 37 30 80 fpm 83 93 83 80 90 80 47 37 100 fpm Because of airflow fluctuations in a typical laboratory environment Labconco suggests setting the RED LED low alarm set point to 20 below the enclosure s operating speed 5 To complete the CAL CONFIGURATION be sure to enter DONE If needed refer to the Configuration procedure provided on the following page for additional details 6 To start the calibration mode enter CALIBRATION mode on the display from the SETUP menu 7 Follow the instructions on the display and simulate the low exhaust volume by fully opening the sash You may measure the average face velocity for the low calibration set point or utilize the c
3. See Figure 2 only to reference internal assembly of the airflow monitor Locate the elbow locknut and washer and install it in the 1 19 dia hole per Figure 1 and Figure 2 The enclosure baffle pivots down to install the elbow washer and locknut 3 Cut the 1 hose supplied with the kit to 10 5 approximate length and install it between the airflow sensor and the elbow 4 Secure the Guardian Digital alarm to the enclosure with double stick tape as shown in Figure 3 The airway passage between the alarm module and the enclosure is now complete 5 Locate the metal hose cover and install with double stick tape per Figure 3 Omit this step on the Protector Work Stations 6 Locate the power supply transformer One end should already be connected to the two pin connector labeled 15 VDC on the back of the alarm module and through the strain relief bushing If disconnected then reconnect to power the airflow monitor Plug the 115V power supply into a standard 115V duplex receptacle the back of the accessory FilterMate portable exhauster or the back of the accessory light For 230V plug into a standard receptacle with your specific outlet plug It is recommended that the airflow monitor be connected directly to the FilterMate switched auxiliary outlet so the airflow monitor is powered at the same time 7 Installation is now complete Labconco Instruction Sheet 3909400 Rev F ECO F288 1 of 7 Remove Large 1 19 Dia Hole Plug
4. calibration will take about 5 seconds Be sure to enter DONE after successfully completing the low and high calibration set points Once calibration is completed go to RUN and hit ENTER The value should read close to the high calibration set point To lower the face velocity to the operating point simply alter the exhaust volume with the speed control on the FilterMate or exhaust damper Then recheck the face velocity with an anemometer to confirm the display on the digital airflow monitor Labconco Instruction Sheet 3909400 Rev F ECO F288 4 of 7 Digital 1000 Alternate Calibration Procedure Constant Volume Conditions 1 To successfully calibrate it will be necessary to change the face velocity by opening and closing the enclosure s sash The airflow monitor is factory set to be calibrated with a difference of at least 50 fpm and can be changed by adjusting the lower higher air sample difference The in flow face velocity speeds provided in the chart below are suggested to successfully calibrate the Digital 1000 2 Before proceeding with calibration it will first be necessary to configure the airflow monitor Go to the setup and then CAL CONFIG MENU and adjust the lower higher air sample difference to 10 fpm This will allow you to successfully calibrate with minimum difference values of 10 fpm While in CAL CONFIG MENU change the sensor difference from 10 to 3 4 While in CAL CONFIG
5. in the user s area concerning the dumping of waste materials in or upon water land or air and to comply with such regulations Labconco Corporation is held harmless with respect to user s compliance with such regulations Contacting Labconco Corporation If you have questions that are not addressed in this manual or if you need technical assistance contact Labconco s Customer Service Department or Labconco s Product Service Department at 1 800 821 5525 or 1 816 333 8811 between the hours of 7 00 a m and 6 00 p m Central Standard Time Labconco Instruction Sheet 3909400 Rev F ECO F288 7 of 7
6. MW _ aboratory environment o on ponent lt 7 LABCONCO LABCONCO CORPORATION 8811 Prospect Ave Kansas City MO 64132 816 333 8811 Fax 816 363 0130 800 821 5525 Guardian Digital Kit No 3908800 02 or 3908801 03 Guardian Digital 1000 Operation The Guardian Digital Airflow Monitor consists of the airflow sensor the Alarm Unit and the 15 VDC power supply For 115V operation the alarm unit is powered by plugging the power supply into the factory prepared digital airflow monitor socket For 230V operation the Alarm Unit is powered by plugging the power supply into a building outlet The alarm has Enter and buttons to program the monitor There is also a green LED SAFE yellow LED CAUTION and red LED LOW with audible alarm for airflow conditions The audible alarm can be permanently muted if desired The Guardian Digital 1000 Airflow Monitor displays a face velocity value provides an RS232 communications port to a PC or building computer system can be configured for external input connections such as night setback or external alarm and provides up to three output relays that can be configured For complete detailed information please refer to the separate Labconco 1000 Alarm User s Manual provided with the enclosure Installation Procedure 1 The enclosure comes prepared to except the Guardian Digital airflow monitor system 2 First remove the large 1 19 dia gray hole plug See Figure 1
7. alculated value provided in the chart above The average face velocity for the low set point is accurately measured by dividing the opening of the enclosure into equal area grids consisting of at least 9 data collection points and measuring the velocity at the center of each grid with a calibrated thermo anemometer Enter the low value on the display Be careful not to block the opening The low exhaust volume calibration will take about 5 seconds 8 Now simulate the high exhaust volume by fully closing the sash to its normal operating position Measure the average face velocity for the high calibration set point to confirm that the source of constant air volume is providing the desired face velocity for the enclosure The average face velocity for the high set point is accurately measured by dividing the opening of the enclosure into equal area grids consisting of at least 3 data collection points and measuring the velocity at the center of each grid with a calibrated Labconco Instruction Sheet 3909400 Rev F ECO F288 5 of 7 thermo anemometer Enter the high value set point on the display The high value must be at least 10 fpm greater than the low value The high exhaust volume calibration will take about 5 seconds 9 Be sure to enter DONE after successfully completing the low and high calibration set points 10 Once calibration is completed go to RUN and hit ENTER The value should read close to the high calibratio
8. for a range of air velocity settings that may be used 5 To complete the CAL CONFIG be sure to enter DONE You are returned to the Main Menu Labconco Instruction Sheet 3909400 Rev F ECO F288 6 of 7 The information contained in this manual and the accompanying products are copyrighted and all rights reserved by Labconco Corporation Labconco Corporation reserves the right to make periodic design changes without obligation to notify any person or entity of such change Warranty Labconco provides a warranty on all parts and factory workmanship The warranty includes areas of defective material and workmanship provided such defect results from normal and proper use of the equipment The warranty for all Labconco products will expire one year from date of installation or two years from date of shipment from Labconco whichever is sooner except the following e Purifier Logic Biological Safety Cabinets and PuriCare Lab Animal Research Stations carry a three year warranty from date of installation or four years from date of shipment from Labconco whichever is sooner SteamScrubber amp FlaskScrubber Glassware Washers carry a two year warranty from date of installation or three years from date of shipment from Labconco whichever is sooner Blood Drawing Chairs carry a ten year warranty Carts carry a lifetime warranty Glassware is not warranted from breakage when dropped or mishandled This limited war
9. n set point 11 With the sash fully open the Digital 1000 monitor should go into RED LED low air alarm if successfully calibrated Digital 1000 Alternate Configuration Procedure for Constant Volume Conditions Note Enter Button stores information and Buttons allow for scrolling 1 Push the ENTER Button on the face of the alarm until the SET UP Menu is displayed 2 Scroll to SET UP and hit ENTER 3 The PASSWORD MENU displays The Password is 0000 Press the ENTER button repeatedly until the CAL CONFIG MENU is displayed 4 In the CAL CONFIG MENU set the following CALIBRATION CONFIGURATION MENU SETTINGS DISPLAY UNITS FPM LOW AIR ALARM RED LED 48 FPM 60 fpm Operating inflow 64 FPM 80 fpm Operating inflow 80 FPM 100 fpm Operating inflow LOW AIR CUTOFF OFF WARNING AIR ALARM YELLOW LED 51 FPM 60 fpm Operating inflow 67 FPM 80 fpm Operating inflow 83 FPM 100 fpm Operating inflow WARNING AIR RESET 3 FPM HIGH AIR ALARM OFF LOWER AIR SAMPLE FLUCTUATIONS OFF HIGHER AIR FLUCTUATIONS 3 LOWER HIGHER AIR SAMPLE DIFFERENCE 3 WARN TO ALARM AIR TIME 10 SECONDS ALARM TO WARN AIR TIME 3 SECONDS SHOW AIR FLOW ON SHOW TIME LINE OFF DISPLAYS BAR GRAPH OFF AUDIBLE ALARM ENABLED SENSOR DIFFERENCE 2 Suggested Air Alarm Settings Refer to the chart on the previous page
10. oint Set Point 40 50 fpm 60 fpm 40 60 fpm 100 120 fpm 60 70 fpm 80 fpm 50 90 fpm 100 150 fpm 80 90 fpm 100 fpm 50 110 fpm 100 170 fpm Go to setup and then CAL CONFIG MENU and change the lower higher air sample difference to 20 fpm This will allow you to successfully calibrate with values of a minimum of 20 fpm difference 3 While in CAL CONFIG MENU change the sensor difference from 10 to 3 While in CAL CONFIG MENU adjust the red low air alarm to the desired setting such as 55 fpm Then adjust the yellow CAUTION or WARNING to 59 fpm Then adjust the CAUTION or WARNING air reset to 3 fpm This sets the alarm condition 5 To complete the CAL CONFIGURATION be sure to enter DONE 10 11 To start the calibration mode use the Labconco 1000 Manual and enter CALIBRATION mode on the display from the SETUP menu Follow the instructions on the display and alter the low exhaust volume with the speed control on the FilterMate or exhaust damper Measure the average face velocity and enter the low value on the display Be careful not to block the opening The low exhaust volume calibration will take about 5 seconds Now alter the high exhaust volume with the speed control on the FilterMate or exhaust damper Measure the average face velocity and enter the high value on the display The high value must be at least 20 fpm greater than the low value The high exhaust volume
11. ranty covers parts and labor but not transportation and insurance charges In the event of a warranty claim contact Labconco Corporation or the dealer who sold you the product If the cause is determined to be a manufacturing fault the dealer or Labconco Corporation will repair or replace all defective parts to restore the unit to operation Under no circumstances shall Labconco Corporation be liable for indirect consequential or special damages of any kind This statement may be altered by a specific published amendment No individual has authorization to alter the provisions of this warranty policy or its amendments Lamps and filters are not covered by this warranty Damage due to corrosion or accidental breakage is not covered Returned or Damaged Goods Do not return goods without the prior authorization from Labconco Unauthorized returns will not be accepted If your shipment was damaged in transit you must file a claim directly with the freight carrier Labconco Corporation and its dealers are not responsible for shipping damages The United States Interstate Commerce Commission rules require that claims be filed with the delivery carrier within fifteen 15 days of delivery Limitation of Liability The disposal and or emission of substances used in connection with this equipment may be governed by various federal state or local regulations All users of this equipment are required to become familiar with any regulations that apply