Home
User manual P
Contents
1. __________________ _ ____ select Cal Fle __________________________ _ ____ 21 3 You can remove one or several logs by clicking on Remove Analysis Logs 4 Once you have loaded all curves you would like to analyze select the calibration field to use for each by clicking on the drop down menu in C 5 The software will calculate and display the Thermal Indicators of Set D i i Initial Set E Mix 18 Redaimed Water with 1 CC loz ewtDelvo 7 Window of Finishability El Mix 20 Reclaimed Water with 2 CC loz cwt 150 Calmetrix Manual O Cameri Manua 6 Tbase is the temperature in the mix the beginning of hydration F P Cal Set pre fills this field with the initial temperature when the test was started You can modify this temperature manually e g if you know the real mix temperature to fine tune 7 The upper right hand side graph F displays the calculated initial set final set and the Finishability time window 22 Calmetrix 1000 User Manual 2014 Calmetrix Inc All rights reserved How to use Method 2 Live Analysis This method is often used to determine saw cutting times but it can also be applied to the live determination of Setting times Just like in Method 1 you will have to create a model for a given mix to link its saw cut times or setting times measured in physical testing to its calorimetric curves To
2. Ensure that the mixing procedure is sufficient to disperse binder properly If the water cement w c ratio is pre defined perform a pre test mix with blank cement paste or mortar no admixtures and another pre test with the highest desired dosage of admixtures Adjust w c ratio if necessary If the water demand of the cement admixture combination to be tested is unknown perform a pre test to determine required water to binder ratio binder includes cement and any supplementary material used as a cement replacement Preparation 4 14 We recommend you start logging data in your P Cal at least one hour before inserting the first sample This allows the unit to reach a stable baseline temperature Pre dilute admixtures to ensure accurate weighing or use high accuracy micro pipettes for accurate volumetric dosing Use the same mixing sequences and times as far as possible within a test series Prepare concrete according to standard or sample concrete directly from production in batch plant mixer or ready mix truck Be sure to enter mix data in the Mix Configuration area of the P Cal Logger tab for future reference Calmetrix 1000 User Manual 2014 Calmetrix Inc All rights reserved 8 Fill your 100mm x 200mm or other standard sample container completely Make sure the container is properly closed to avoid any spillage Insert the container into the P Cal and close it by securing the three latches G VISUALIZI
3. Manual 2014 Calmetrix Inc All rights reserved H INFERRING SETTING TIMES F P CAL SET F P Cal Set has two ways to infer setting tines Method 1 Post testing analysis of calorimetric curves to infer setting times in lieu of physical testing such as with a penetrometer needle This feature is useful if you have multiple samples in a large test matrix and need approximations of setting times without incurring the cist and time for physical testing Method 2 Live analysis while conducting a calorimetry test to infer setting times or saw cut times and get notified when mix reaches its initial of final setting time or saw cut time This method can be applied to get an indicator of when to start performing finishing work How to use Method 1 Post testing analysis F P Cal Set can help you infer initial and final setting times within close approximation by applying the Fractions Method and generate Thermal Indicators of Set TIS This method is based on the assumption that for a given mix design there is a correlation between the heat evolution and setting times Initial set is reached when the temperature level in a sample placed in the P Cal reaches a certain fraction of the peak Each mix has its very own fraction values corresponding to initial and final set Once the fraction values are determined for a given mix design they can be used to predict setting times when the same mix is used again in the future To be able to
4. in the corresponding fields This data will be stored along with the sample test data and will appear when you visualize curves in the F P Cal Reports or other software modules 6 Make sure you enter the mixing time when the water meets the cement in the Mix Date Time field This time will be used as the point of origin so as to make side by side comparisons of curves from tests performed at different times Calmetrix 1000 User Manual 2014 Calmetrix Inc All rights reserved Plot area d The type and serial number of the equipment currently connected is displayed on top of the chart area Right click on the x or y axis to change scaling from Automatic to Manual In Manual mode you can adjust the min and max using the cursor f Click on a graph legend color on the right of the screen to change the appearance color line type etc g Click on the Zoom and Motion tools on top f to Zoom in out or move the graph 11 This area is used to operate the P Cal s logging functions Displays current temperature as measured both inside the P Cal and for the external sensor located underneath the instrument used for Compressive Strength estomates To change units from F to C click on the SETTINGS tab on top select the Measurement Units field Click on Apply New Settings on the bottom of the SETTINGS tab to validate your changes The Logging light turns green
5. predict setting times for a given mix design you will first have to create a model which establishes the relationship between the temperature curve and initial and final set as measured by a standard method for example ASTM C403 First build a model To build a new Model for a given mixture proceed as follows 1 Goto F P Cal Set and select the Modeling tab Click on New Model a 19 20 Click on Add Modeling Log b A new window will pop up letting you browse to the containing the calorimetry curve for the mix for which you would like to build a model You can load up to 8 curves for the same mix Select one or several curves for this mix and click on Accept The selected curve s will appear on the screen as shown below For each curve enter the measured Initial Setting time in the Initial TIS field and the Final Setting time in the Final TIS field c These values should be measured using the standard you normally use to determine setting times e g ASTM C403 Now select the curves that you want to use to build your Fractions Model for this mix by checking or un checking the Valid box d The average fraction values of all Valid mixes will display in the Initial Set Fraction and Final Set Fraction fields to the right e Save your model by clicking on Save f You will be asked to provide a name for your model Once saved your model will be stored in the Calibration folder and can l
6. the P Cal other than the external probe in a fog room as the equipment may suffer irreparable damage Step 6 Measure setting time of concrete or optionally use F P Cal Set to estimate the initial setting time of the concrete tested for compressive strength Step 7 Test the samples prepared for compressive strength measurement which are in the curing bath the fog room or curing at ambient temperature and make sure to record the curing time for each measurement for example measure compressive strength at 16 h 1 day 2 days 4 days 7 days 28 days 29 30 Modeling and Strength Prediction In this section you will build a model that correlates degree of hydration as measured by the calorimetry curve to compressive strength development IMPORTANT NOTE We recommend that you have at least 96 hours of calorimetry data and at least four physical measurement of compressive strength before building a model Step 1 Connect the computer to your P Cal unit Start CalCommander and select the F P Cal Logger tab Click on Read data from logger and wait for the data to load on the screen When finished click on Save log data to and save to your location of choice click on P Cal Strength and select the Modeling tab Step 2 Go to the P Cal Strength tab and click on the Modeling tab Click on Add Modeling Logs a then browse for and select the calorimetry data previously recorded and saved in Step 1
7. use Method 2 you will use the Live Modeling and Live Analysis in the F P Cal Set software tab The Live Analysis tab can only be used when the calorimeter is connected to the computer and is logging HEE initial Set Window of Finishability 23 Each mix has specific hydration characteristics that influence the initial and final saw curing times as per the curve below Window Initial saw cut time 24 In order to adjust to these mix specific characteristics users can create a simple model for each mix and store it in a database so that the saved model can be called up every time the same mix is used Calmetrix 1000 User Manual 2014 Calmetrix Inc All rights reserved 1 Click on New Model to create a model for a new mix or click on Load Model to modify an existing model 2 Input parameters for the initial saw cut time in section A The parameters are defined as the minimum temperature increment A1 in a given time period A2 For example if you input 1 5 C in A1 and 00 10 10 minutes in A2 initial saw cut is reached at the time t defined by the first instance when Te 2 1 5 where Tris the temperature reading at time t If you input 0 in the Time Increment field the model is such that the initial saw cut time is reached as soon as the temperature increment from the start of the experiment reaches the value specified in A1 3 Proceed a
8. when your P Cal is logging data Log Conflict lights up when more than one is connected to your computer and the software is unable to determine proper start times for each equipment Low Battery will light up when the battery needs to be replaced Refer to Section H Troubleshooting Bad Sensor Contact indicates a bad connection between the sensor and the logger located within your P Cal equipment Refer to Section H in this User Manual to troubleshoot bad connections Click on Get Latest Sensor Readings to update the temperature and battery fileds Copy Paste is not accessible on single channel instruments like P Cal 1000 When disconnecting your P Cal your configuration data will not appear automatically upon reconnection Click on Read Configuration from Logger to display Configuration data Click on Update Configuration Data to save changes made to the Configuration Data after logging has started Click on Start Logging to begin logging temperatures The sensor in your P Cal will start logging Note it is not necessary to keep your computer connected after logging has been activated All data will be stored in the P Cal s internal memory Reconnect your computer at a later time to retrieve data Stop Logging at the end of your test After Stop Logging has been selected you cannot continue logging data on your current test Data will be saved automatically in your default Logs folder Read Data from Log
9. CTA T gt or d mo _ Te os Tus Ton quee pen Lam dnm Ee pM Fae fee FS calmetrix P GAL 1000 USER MANUAL Calmetrix P Cal 1000 User Manual CALMETRIX INC P CAL 1000 PORTABLE HIGH PERFORMANCE CALORIMETER FOR CONCRETE AND CEMENT USER MANUAL Calmetrix 1000 User Manual 2014 Calmetrix Inc All rights reserved IMPORTANT SAFETY RULES CAUTION Always make sure to hold your P Cal firmly with both hands when moving it especially when it contains a sample as it could fall on your feet and cause an injury DISCLAIMER Calmetrix does not assume any liability as related but not limited to the accuracy relevance or interpretation of results while using P Cal COPYRIGHT This User Manual is subject to copyright and may not be reproduced in whole or part without permission Calmetrix 1000 User Manual 2014 Calmetrix Inc All rights reserved TABLE OF CONTENTS A BEFORE YOU START USING THE EQUIPMENT Page GENERAL USE a Rota ties ai pera ed ae Page 8 os HANDLING INSTRUCTIONS sia fena ade r tees 8 D OPERATING THE EQUIPMENT CALCOMMANDER SOFTWARE Page 9 E OPERATING YOUR UNIT FOR THE FIRST TIME Page 13 F SAMPLE PREPARATION AND RECOMMENDED TEST PROCEDUR
10. E Page 14 VISUALIZING RESULTS F P CAL REPORT GENERATOR Page 15 INFERRING SETTING TIMES WITH F P CAL SET Page 19 PREDICTING COMPRESSIVE STRENGTH WITH P CAL STRENGTH Page 28 Je TROUBLESHOOTING ei a ent vacua oed ud e DR pA SU ned Page 34 Calmetrix 1000 User Manual 2014 Calmetrix Inc All rights reserved A BEFORE YOU START USING THE EQUIPMENT What is in your box 1 P Cal 1000 equipment 2 USB Cable to connect the P Cal to your computer s USB port 3 Quick Set up guide 4 USB Memory Stick with Software and Manuals CalCommander Software Installation 1 Insert the USB Memory Stick in your compuetr s USB port and browse to the Software folder Look up and double click on the CalCommSetUp exe file and follow the on screen installation instructions 2 Restart your Computer 3 Connect the P Cal to your computer using the USB cable that is provided with the equipment When connecting your P Cal for the first time your Windows Operating System will automatically install the necessary drivers Click on the CalCommander Icon on your Desktop to open CalCommander 4 Proceed to the Settings tab to find the Module ID numbers for software registration NOTE Your P Cal includes one license for each of the following software modules e F P Cal Logger to operate the equipment e F P Cal Reports to retrieve past re
11. L 1000 EQUIPMENT Questions Find us at 1 617 203 2090 support calmetrix com http www calmetrix com calmetrix The Calmetrix Team v 100274 1 53 35 calmetrix Calmetrix Inc PO Box 70 Arlington MA 02476 0001 USA T 1 617 203 2090 info calmetrix com vv http www calmetrix com 2010 2014 Calmetrix Inc All rignts reserved
12. NG RESULTS F P CAL REPORT GENERATOR With CalCommander open on your computer click on the F P Cal Reports Tab on top A new window will appear as in the picture below LL 15 Add Logs to plot select C Users Public Documents Calmetrix CalCommander Logs default storage of completed log files or any other location where you store completed log files Select the log file you would like to display A new window will appear Date modified Type _ Admix Cement incompatibility acallog 2 23 2007 11 54 ACALLOG File Recent Places Ash admix incompatibility acallog 3 15 2007 1013 AM ACALLOG File _ Diet sodas full dose acallog 3 18 2009 3 15 PM ACALLOG File Different Cements acallog 3 14 2009 2 02 PM ACALLOG File Desktop _ Linear retarder dosage acallog 3 15 2007 10 09 AM ACALLOG File Reclaimer and linear retarder dose acallog 3 24 2009 1 05 File Retarding Effect of Color acallog 2 23 2007 11 55 PM ACALLOG File Libraries A Computer Select the log file you would like to load and click OK Repeat the steps above to add more files from other log files You can add up to 8 files on any Report Generator graph Click Accept and the files you selected will appear in the Report Generator window The F P Cal Report Generator can read files that were generated by the following eq
13. above Step 3 Enter values of compressive strength b and the corresponding curing time c from the physical test data performed in the Data Generation part Also enter the Initial Setting Time for the mix in the Maturity Start Time field e Calmetrix 1000 User Manual 2014 Calmetrix Inc All rights reserved Logger 10085422 MixID 1 concrete w bag Tmix 10 0 degC 02 00 11 55 23 40 47 34 71 26 95 13 Step 4 Press Save model i and give it a descriptive name This model is now stored your computer and you can use it to predict strength for the same mix Step 5 Go to the Data Analysis tab Click on Add Analysis logs I to upload the calorimetry data for the mix you just saved in Step 1 above Select the Model you saved in Step 5 Now you can enter curing ages in the corresponding fields j and get a strength estimate prediction in the k fields below 31 os 03 20 __ 16 00 120 00 672 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 15 06 j 118 41 673 05 CS 97 3 exp 635942 EA 0 41 R2 0 9997 IMPORTANT NOTE P Cal Strength automatically extrapolates the curves for curing ages that are greater than the total duration of the calorimetry log period This allows you to make compressive strength predictions However a minimum log time of 96 hours is recom
14. ater be selected when predicting setting times for the same mix design using the Data Analysis tab 177 d 95 0552 00 00 d 00 00 Na 00 00 00 00 NN m A 00 00 00 00 m A 0 00 00 i 00 00 00 00 I ll A M 00 00 00 00 Calmetrix 1000 User Manual 2014 Calmetrix Inc All rights reserved 6 You can later open and edit an existing model by clicking on Load Model g Now you re ready to use the setting time prediction tool Once a model has been established for a given mixture it can be used to predict early and late times of set of the same mixture by using only a calorimetry curve To use the prediction tool proceed as follows 1 Click on the Data Analysis tab 2 Click on Add Analysis Logs A to select the calorimetry data for which you would like to make a setting time prediction and click OK You can add up to 8 calorimetry curves Mix 18 Redaimed Water with 194 CC loz ewt Delvo Mix 2 20 Reclsimed Water with 2 CC toz cwt Mx 17 Control Recamed Water with 2 Chore 0 0 degC Mx 150 _ f Mx 18 Reciamed Water wth 1 CC lozjowt Devo Tmx 0 0 Mx Tme25 149 I Caibraton Fie M 20 Reciamed Water with 2 CC loz cwt Tmax 0 0 degC Mx Tme 25Mar 149 Gelectcaibratentie po e O ecane A TT ________________________ _ ___
15. esults from different test runs C HANDLING INSTRUCTIONS Sample containers must be clean on the outside before inserting them into the P Cal Cement or admixtures on the outside will cause the sample container to get stuck in the calorimeter with expensive and time consuming repair as a result All containers must be closed tightly to avoid accidental spillage AVOID transporting the P Cal with samples inside as equipment might be damaged If you have to move or transport the unit during a test with a sample loaded inside the unit make sure to keep the P Cal upright in the same position during the entire duration of the test for consistency in thermal insulation Calmetrix 1000 User Manual 2014 Calmetrix Inc All rights reserved D OPERATING THE EQUIPMENT CALCOMMANDER SOFTWARE Connect the P Cal to your computer using the USB cable provided Open the CalCommander software You will see the Welcome Screen FPLalogger FP CaRepots Fipcaset Pcalstencth t callogger icarepors icaae resiset rcxsseem setinos calmetrix CALCOMMANDER CALORIMETRY SOFTWARE Always connect instruments to the computer BEFORE you start Calcommander and unplug them AFTER you exit Calcommander Version 1 46 Build Date 19 May 2012 2012 Calmetrx Inc Parts 2012 National Instruments Corp Exit Calcommander Click on the F P Cal Logger tab on top Familiarize yourself with the main window of the P Cal Logger
16. ger to update the graph at any time during testing Create or Print PDF reports or Save the current graph as an image file Calmetrix 1000 User Manual 2014 Calmetrix Inc All rights reserved E OPERATING YOUR UNIT FOR THE FIRST TIME Calibration and Connections Your P Cal uses thermistors for temperature measurements and does not require any re calibration over time To verify that internal connections have not been compromised during shipping a test run should be conducted by loading the unit with a sample of hot water Verify that the sensor is logging the correct water temperature After thirty minutes the temperature reading should stabilize and should be within 2 F of the actual water temp rature measured before inserting the sample into the calorimeter Check for error signals LOW BATTERY or BAD SENSOR CONTACT on the upper right hand side of the P Cal Logger screen Refer to Section J Troubleshooting to address any issues 13 SAMPLE PREPARATION AND RECOMMENDED TEST PROCEDURE IMPORTANT NOTE Testing should be conducted with a standard container of your choice You can use standard 4 x8 100 mm x 200 mm cylinders or any type of container of any size that fits in the instrument made of glass metal or plastic and that can be tightly closed However it is important to use the same container for all your testing if you want to be able to compare data from tests conducted at different times
17. mended for predictions Examples of applications Example 1 you ran a calorimetry curve for 15 days You would like to get estimates of compressive strength at 1 3 7 and 28 days For 1 3 and 7 day strength the software will use actual values from the calorimetry curve For 28 days the software will use an extrapolated value Example 2 you ran a calorimetry curve for 72 hours 3 days and you would like to know an estimate of compressive strength at 18 hours 36 hours and 5 days You can easily estimate compressive strength 32 Calmetrix 1000 User Manual 2014 Calmetrix Inc All rights reserved for 18 hours and 36 hours but the extrapolated value for 5 days would not be reliable A minimum log time of 96 hours 4 days is required for strength prediction The following are useful papers we recommend as a resource for data interpretation and examples of uses for your P Cal field calorimeter and Calorimetry in general 1 Sandberg P and Liberman S Monitoring and Evaluation of Cement Hydration by Semi Adiabatic Field Calorimetry Concrete Heat Development Monitoring Prediction and Management ACI SP 241 2 Atlanta GA April 2007 pp 13 24 2 Riding K A Poole J L Juenger M C G Schindler A K and Folliard K J Calorimetry Performed On Site Methods and Uses ACI SP 241 3 Poole J L Hydration Study of Cementitious Materials Using Semi Adiabatic Calorimetry PhD Dissertatio
18. n The University of Texas at Austin Austin TX 2007 4 Wang K Ge Z Grove J Ruiz M Rasmussen R and Ferragut T Developing a Simple and Rapid Test for Monitoring the Heat Evolution of Concrete Mixtures for Both Laboratory and Field Applications National Concrete Pavement Technology Center Ames IA January 2007 58 pp 5 Cost V T and Gardiner A Practical Concrete Mixture Evaluation via Semi Adiabatic Calorimetry 2009 Concrete Technology Forum Focus on Performance Prediction Cincinnati OH May 2009 21 pp 6 Cost T Thermal Measurements of Hydration Concrete Mixtures A Usefule Quality Control Tool for Concrete Producers NRMCA Publication 2PE004 33 J TROUBLESHOOTING P Cal is a very robust equipment and any issues can usually be solved easily without lengthy troubleshooting or repair times Most common problems can be identified on the F P Cal Logger module screen of your CalCommander software If the Low Battery light is lit You will have to change the battery as soon as possible The logger requires one 3 Volt CR 2032 lithium battery Press Stop logging when possible You will not be able to continue the current test run after you changed the battery Disconnect the equipment from your computer carefully unscrew the bottom metal plate of your P Cal and locate the logger a small rectangular box Unplug the gray sensor cables and the black USB cable from the logger Make sure yo
19. omes with a Base Model which can be used until a specific model has been determined for each mix Click on Update Readout Now f to initiate the live reading Check the Enable Auto Read box g for the software to retrieve data automatically Input the desired Auto Read time period in h Click on Reanalyze to recalculate data if you switch to a different model during the experiment 7 The software will automatically display the Initial and Final saw cutting times in the Initial TIS Time of Saw or Time of Set and Final TIS fields when reaching the temperature increments as defined in the model 27 l PREDICTING COMPRESSIVE STRENGTH P CAL STRENGTH P Cal 1000 comes with P Cal Strength CalCommander s software module for strength prediction P Cal Strength can be used to infer compressive strength for multiple curing ages or to predict future compressive strength gain from 4 5 physical test data points collected within the first 96 hours of testing Users will have to perform physical strength testing at 5 or more curing ages and can then use a calorimetry curve to estimate compressive strength at additional curing ages To be able to estimate compressive strength for a given mix you will first need to e Generate compressive strength data from physical tests and a calorimetry curve see Data Generation step by step below e Create the correlation model between compressive strength and degree of hydration see Modeling step b
20. s above for the final saw cut time parameters in section B When finished click on Save Model or Save Model As to store your model in the database for future use Now you can use your calorimeter for live determination of saw cutting times 25 26 Before the concrete mix is prepared connect your calorimeter to the computer wired USB connection Open CalCommander on the computer and proceed to the F P Cal Logger tab Click on Start Logging b When preparing a mix take a sample in the sample container you are usually using with your calorimeter Write down the mix time and introduce the capped cylinder in your F Cal Go to the F P Cal Set tab and proceed to the Live Analysis section W0tJan 01 00 ER po 01 00 oo 70 01 00 EB 67 01 00 E Calmetrix 1000 User Manual 2014 Calmetrix Inc All rights reserved 5 Introduce the Mix Time in c and the approximate time the equipment takes to stabilize before actual hydration measurements begin in d This is the time it takes to absorb any temperature shocks coming from the difference in temperature of the materials used in the mix and the temperature in the calorimeter before starting the experiment Unless conditions are extreme very hot materials or very cold outside temperature the default value of 1 hour should be a good estimate 6 Choose the model your would like to use in e Please note that the software c
21. software module FP Callogger FPLalset PLalStrength ICaloger tcxRepots CNAE rcaset rcasmengh Settings P Cal 1000 Calorimeter 10141425 zm eee een pees eee eee oes es ees eee eee ess IE Temperature 197 External gt Internal External 10 10282 77 03 120 10 Y 100 7 g ot area 4 Low Battery Bad Sensor Contact 50 Battery P 3 19 El Get Last Sensor Readings 207 D Y x 0 x 1 0 D 00 00 00 10 00 20 00 30 00 40 00 50 01 00 0130 01 20 01 30 0150 02 00 Time since Start hh mm Copy Configuration from Current Tab 4 Sample 1 Paste Configuraiton to Current Tab om man are Read Configuration from Logger mof Water Cement Rato 0 00 Auction Ple Cement Source ID Cement Content 0 00 bs c J Lbs icy ar SCM 1 Type None Percent 0 00 Source Stop Logging SCM2 None vl Percent 0 00 Source Read Data SCM3 Type None vl Percent 0 00 Source Dose 0 00 _ EM Dose 0 00 Mix Configuration area as slump Measure 0 00 Units inches Iv Unit Weight 0 Units bs cft ER 320 0 0 Sensor Enabled Save Graph 9 10 Mix Configuration area This part of the screen is used to capture all mix and sample related data a Enter the Mix data for each sample
22. sults and create reports F P Cal Set to infer thermal indicators of setting times e P Cal Strength to calibrate and predict strength for a given mix design Register your software by clicking on the Registration link in the Support Section at http www calmetrix com Fill out the form or email your Module ID numbers to registration calmetrix com We will email you your Registration numbers 5 You can find videos software updates and registration user manuals and other support items in the Support Section on our website at http www calmetrix com B GENERAL USE P Cal contains state of the art high efficiency insulation enabling quasi adiabatic conditions The equipment can be used with cement paste mortar of real concrete including coarse aggregates Please refer to Section F in this manual for recommendations of use with paste and mortar samples The temperature curve measured by P Cal provides a fingerprint of the chemical reactions in the sample that can be analyzed to infer specific reactions during early stages of hydration usually 24 to 36 hours after batching mixing Because of the very limited heat loss P Cal can be used for strength prediction based on the maturity concept P Cal can be used with a single sample container of 4 x8 100 mm x 200mm maximum size Smaller containers can be used but it is important to choose a standard container type and use the same for all tests so as to be able to compare r
23. u mark the gray sensor cables to remember to which of the two plugs they were connected Remove the logger then unscrew the logger case Lift the circuit board and carefully push the battery out with a small blunt instrument or pull it out with your fingernail Insert a new battery positive side facing up Carefully realign the logger case and re fasten the screws Reconnect the temperature sensors and USB cable making sure the connectors are pushed all the way in Place the logger back and reassemble the unit If Bad Sensor Contact is lit Stop logging withdraw the sample and proceed to unscrew the bottom part of the equipment as described in the Low Battery section above Locate the logger and ensure that the sensor connector is pushed in all the way If the problem persists call Calmetrix Customer Support at 1 617 203 2090 or email support calmetrix com If erratic or clearly erroneous values appear in the temperature readings Insert a sample cylinder filled with water measure the water temperature and start logging Start logging and wait for 30 minutes Check that the temperature readings are within a 2 F 1 C range of the actual water temperature If the sensor indicates a temperature out of that range proceed as with the Bad Sensor Contact section above If the problem persists contact Calmetrix customer support 34 Calmetrix 1000 User Manual 2014 Calmetrix Inc All rights reserved ENJOY YOUR P CA
24. uipment Calmetrix P Cal 1000 F Cal 4000 F Cal 8000 and Grace AdiaCal 16 Calmetrix 1000 User Manual 2014 Calmetrix Inc All rights reserved Once you selected and loaded all the curves you would like to visualize the following window will show 2289 a8 88 8 3 a N H N pi 88 83 888 858 Sampe Ems Logger 1123003 MixID Tmix 32 0 Logger 1123003 MixID Tmix 32 0 de Logger 1123003 MixID Tmix 32 0 de Logger 1123003 MixID Tmix 32 0 de b Actual log data is displayed in this area Temperatures are measured and logged every four minutes c Use Graphing tools to Zoom In Out click on the legend to change colors or style of graphs 17 18 d Click on the min max values on the x y axis to change the min and max to modify the range and scale e Save Selected Log as Text File saves the information including all temperature values from the currently selected Sample tab in a txt file Data from txt files can easily be copied and pasted in an application such as Microsoft Excel for spreadsheet calculations on plotted data f If the Mix Time or Mix Temperature fields are highlighted in yellow you should verify that the data captured in these fields is correct g If you edit data in any field a yellow light will appear until you permanently save the change by clicking on Save Updated Log Calmetrix 1000 User
25. y step below Data Generation Step 1 Place your P Cal unit s close to your mixing station and connect the first one to a computer with the CalCommander software installed then start CalCommander and select the Cal Logger tab Step 2 Type in all relevant mix information see section D Operating your Equipment in this manual for details Step 3 Mix concrete and prepare samples for compressive strength testing a minimum of two samples per age at a minimum of four to five ages Step 4 Make sure you entered the Mix Time in the corresponding fields in the F P Cal Logger tab and click on Start Logging then close CalCommander and disconnect USB cable from the computer 28 Calmetrix 1000 User Manual 2014 Calmetrix Inc All rights reserved Step 5 Unfold and uncoil the external temperature probes see fig A from underneath each of the P Cal s bottom and place them in the vicinity of the samples for compressive strength testing protected from direct sunshine and wind Bottom of P Cal 1000 External Probe Fig A location of external probe lf compressive strength samples are customarily moved to a curing tank fog room insert the external temperature probe into the curing tank or fog room The external probe should at all time measure the temperature of the curing location IMPORTANT NOTE Do not submerge the P Cal unit other than the external probe in water and do not place
Download Pdf Manuals
Related Search
Related Contents
高解像度 ネットワーク ドーム カメラ Trust Carve DIGITAL PRESSURE CONTROLLER MODEL 7250 MODEL 7250i OxiSelect™ Protein Carbonyl Fluorometric Assay Copyright © All rights reserved.
Failed to retrieve file