User manual CuroCell® 3
Contents
1. air mattresses operate according to the alternating pressure principle a technique that is well proven and been used for a long time Alternating pressure is when the pressure regularly alternates between the mattress cells to replicate body movement Variable pressure cycles are set up where there is little or no pressure This is optimal for skin recovery or to prevent skin damage Always read the user manual prior to use 1 1 General information This system is a medical device and is CE marked in accordance with MDD 93 42 EEC According to statutory regulations made by the authorities regarding medical devices the manufacturer is required to report all accidents or incidents involving there products We would be very grateful for all information involving accidents or incidents relating to our products by reporting them immediately to us here at Care of Sweden The system has been tested and approved according to the following standards IEC 600601 1 IEC 60601 1 2 SS EN ISO 12182 SS EN ISO 14971 SS EN ISO 10993 EN 597 1 EN 597 2 EN 14126 1 2 Intended use CuroCell 3 is an alternating pressure overlay mattress and can be used in a wide range of health care environments to prevent pressure ulcers as well as treatment up to and including category 3 pressure ulcers It can also be equipped with an integrated foam core to be used directly on the bed frame The mattress is meant for use for persons of a recommended minimum length
2. care of sweden Supporting Life User manual 2015 03 09 CuroCell 3 Alternating pressure system Item number user manual 95 001229 31345026787 CE Contents Important safety precautions Introduction Product description Assembly Installation Operation Cleaning and reconditioning Storage Maintenance Troubleshooting Technical description Spare parts Options and accessories Warranty Other information Oo won A W 11 12 13 14 15 17 18 19 19 IMPORTANT SAFETY PRECAUTIONS Read all instructions prior to use Warning To minimize the risk of fire electric shocks or bodily injury rp The product must only be used for its intended purpose as described in this manual The product may not be combined assembled or repaired with other parts eg pump and mattress accessories or spare parts other than those described in this manual or other documentation from Care of Sweden 2 Do not use the product in wet rooms or where there is a risk of the pump coming into contact with water or another liquid Never handle a pump that has come into contact with water liquid Pull the plug out of the electrical socket immediately and send the product to an authorized service technician for servicing 3 DO NOT USE in the vicinity of or in contact with fire sources hot surfaces such as fire burning cigarettes or hot lamps Although the product has undergone fire testing it could become damaged if it comes into con
3. set cycle time one cycle 10 minutes Pump is noisy Check how the pump is hung on the bed Resonance can occur vibrations are felt in parts of the bed Remove the pump and listen to see whether this makes a difference Can be resolved by putting the pump on a flat steady surface or by placing a towel between the pump and bed If your questions are not answered by the above information please contact Care of Sweden or your local dealer 14 9 Technical description 9 1 Technical specification PUMP SPECIFICATION Model CuroCell 3 Voltage 240 V 50 60 Hz Power consumption 0 17A Type of fuse T1AL 250 V Cycle periods 10 min Dimension LxWxH 30x 10 5 x 22 5 cm Weight 3 4 kg Surroundings Temperature Operation 10 C 35 C Storage 15 C 50 C Shipping 15 C 70 C Humidity Operation 20 80 non condensing Storage 10 90 non condensing Atmospheric pressure 700hPa 1013 25hPa Classification Class Il Type BF IP21 Model CuroCell 3 Dimension LxWxH 80 85 90x200x13 cm Weight 5 4 kg 80x200 cm NOTE Care of Sweden retains the right to modify the product specification at any time WARNING Follow operating and maintenance instructions carefully to prevent injury as well as damage to the product or other equipment 9 2 Labelling The mattress is labeled with product name article number production date size bar code as well as ca
4. this purpose 6 Put the mattress together If cells for any reason have loosened from the tubes these must be replaced according to drawing in chapter 2 1 FOAM CORE Clean the affected area with a mild cleaning solution for this purpose such as alcohol with or without tensides lsopropanol 70 oxidating solutions or chlorine max 1 Press out the water carefully Let the foam core dry in a warm ventilated area without direct exposure of sunlight The foam core needs to be completely dry before it is used again NOTE Check the hygiene cover each time the product is cleaned If damaged it must be replaced or repaired Also check the pump tube connectors and power cord during cleaning Damaged parts must be replaced or repaired 6 Storage The mattress and pump are best stored in the product bag optional extra Handle the packed product with care Do not put anything heavy on top of it 12 7 Maintenance 7 1 General 1 Check the power cord and plug for damage and scratches 2 Check that the mattress cover is intact and that the cover and air cells are correctly assembled 3 Check that the air flow alternates between each outlet on the side of the pump 4 Check for damaged or pinched tubes and connectors Contact the manufacturer or your local dealer for spare parts 7 2 Changing fuses 1 Pull the plug out of the electrical socket before replacing a fuse 2 Remove the cover for the fuses with a screwdri
5. aning solution for this purpose such as alcohol with or without tensides lsopropanol 70 oxidating solutions or chlorine max 1 Excrements and blood spots must be removed immediately with cold water Carefully follow local instructions and usage instructions for the cleaner Q i Bet K Multi layer covers should be divided before washing Wash with like colors Chlorine and phenol based cleaners could have a negative impact on the PU surface z and should be avoided If chlorine is used we recommend a mixture of max 1 11 INNER COVER AND DISTANCE FABRIC accessories 6 O u at amp Clean the affected area with a mild cleaning solution for this purpose such as alcohol with or without tensides Isopropanol 70 oxidating solutions or chlorine max 1 5 2 Reconditioning PUMP See section 5 1 for instructions MATTRESS 1 Disconnect the tubes and let the air out also open the CPR valve 2 Remove the cover See instructions in chapter 5 1 for cleaning covers If the product is equipped with a foam core this should be removed and cleaned separately 3 Move the mattress to a clean area which is suitable for cleaning 4 Wipe off the cells all tubes and the CPR valve with a mild cleaning solution for this purpose such as alcohol with or without tensides Ilsopropanol 70 oxidating solutions or chlorine max 1 5 Wipe off and disinfect all the work surfaces with an alcohol based cleaning agent intended for
6. el and Pan Pacific Pressure Injury Alliance Prevention and Treatment of Pressure Ulcers Quick Reference Guide Emily Haesler Ed Cambridge Media Osborne Park Western Australia 2014 1 3 Other information Action plan Pressure ulcers To achieve the best possible result for this product it is recommended that a well planned and well structured working environment be used In our guideline Action plan you will find examples which can be applied and put into practice when using our products Mattress guide This practical guide gives you as a care giver help in choosing the correct mattress from the Care of Swedens range CuroCell 3 is a mattress in function group C This information can be found on our website www careofsweden com or you can order a copy from our Customer services by sending your details to info careofsweden se 2 Product description 2 1 Pump and mattress 1 CPR rapid deflation a Head b Foot 2 Pump 3 Mattress 4 Pump connector 1a Tubing system with CPR in head 3 static head cells 9 ventilated cells optional 1b Tubing system with CPR in foot 3 static head cells 9 ventilated cells optional CuroCell 3 with an integrated foam core and tubing system with CPR in foot 3 static head cells If the product is equipped with ventilated cells these are placed here 9 pieces Foam core 1 5 6 7 8 1 Quick connector unit air 3 Off On 0 1 5 Power cord 7 Bed hooks 2 Ala
7. hand side pump and mattress NOTE This connector is held in place with the elastic strap 4 6 Transport function If the patient must be moved in bed the tubes can be connected together detached from the pump and the mattress put in transport mode We recommend using this function only for short periods of time 10 4 8 Heel function with check valves option This feature allows pressure relief of the heels Disconnect the air cell which is directly under the heel This air cell will then be emptied for air Press the bottom 1 and disconnect the plug Let the cell remain in its place in the mattress The air cell is activated by assembling the connection again 5 Cleaning and reconditioning It is important to follow the instructions below before using the product for another user Consult the hygiene responsible or the supplier for help and instructions when unsure 5 1 Cleaning PUMP WARNING Do not open the pump housing risk of electric shock Service and maintenance must be performed by authorised technicians Do not get the pump wet or submerge it in any liquid eee Wipe of the pump with a damp cloth and a mild cleaning solution for this purpose such as alcohol with or without tensides lsopropanol 70 oxidating solutions or chlorine max 1 If using a different agent choose one that does not affect the pump housing MATTRESS COVER Covers can be wiped off with a mild cle
8. he plug to an approved electrical socket 220 V 4 Operation Note Always read the user manual prior to use 4 1 Start up 1 Start the pump by pressing 1 On 2 Air starts to fill up the mattress It takes about 30 minutes to inflate the mattress 3 If necessary the patient can lie on the mattress even when not inflated in this case the underlay will carry the load When using a mattress with an air underlay this must be inflated Note We recommend that the comfort dial be set to 8 max when inflating a mattress 4 2 Comfort setting The air pressure in the mattress can be adjusted as required ALWAYS consult the prescriber or other authorised personnel for the correct setting 1 Use the comfort dial to set a suitable air pressure based on the kg nsi e settings table on the pump 0 40 1 2 2 When the green indicator lamp for 50 60 3 4 70 80 5 6 normal pressure lights the mattress aaa r e has reached the set pressure 110 120 7 3 _ 130 160 8 8 l J NOTE e If the head of the bed is raised pressure must be increased at least one 1 level e When only parts of the mattress are under load e g if the patient is an amputee the pressure should be increased 1 or 2 levels e For recommended user weights refer to 1 2 4 3 Hand check checking the pressure setting Hand check is used to ensure that the mattress is properly inflated This should be done regularly once
9. lated Air cell ventilated Air cell non ventilated Air cell ventilated Size 80x200 cm 85x200 cm 90x200 cm 80x13 cm 80x13 cm 85x13 cm 85x13 cm 90x13 cm 90x13 cm Please make a note on your order when ordering ventilating cells CC 8217 CC 8201 Fuse set of 2 Filter CC3 CC4 washable If other spare parts are needed please contact Care of Sweden or your local dealer for further assistance 17 11 Options and accessories 11 1 Options Comfort fabric cover For additional mattress comfort Heel section Options available to relieve sensitive heels Emptying the air cell with release valve NOTE These options are only available with new orders Cable holder A pocket along one long side of the mattress that holds the power cord in place to minimise the risk of tripping over the cord or the cord becoming trapped under the wheels of the bed NOTE This option is only available with new orders Sturdy transport bag Transport and storage bag Ventilated cells Low Air Loss Micro ventilation in 9 of the 17 cells Cover With welded seams 11 2 Accessories Art no Description Size CC 907210 Comfort fabric cover 80x200 cm CC 907212 Comfort fabric cover 85x200 cm CC 907214 Comfort fabric cover 90x200 cm CCc 9093 Transport bag CC 9203 Elevating pad 80 cm Cc 9201 Elevating pad 85 cm CC 9202 Elevating pad 90 cm CC 900220 Heel function with valve 4 cells CC 900222 Heel function with valve 3 cell
10. of 120 cm where the recommended user weight is up to 160 kg We recommend that position changes are made regularly The time interval must be evaluated by the responsible personnel depending on the patient s status diagnosis and general condition The mattress should only be used lying in the lengthwise direction on the mattress with the feet at the end marked with the feet symbol The mattress must always be cleaned between patients NOTE e For certain patients such as amputees are the recommended minimum weight not valid These users may however have other requirements as the entire mattress is not being loaded Please refer to sections 4 2 and 4 3 for operating control e When used together with positioning pillows read the user manual carefully to ensure correct positioning of the patient e If you are using this product with an evacuation sheet it is the user s responsibility to ensure that it is safe to use for evacuation e Be careful when using bed rails or other protection on the bed so that the mattress does not become trapped or damaged e The mattress must not be lifted with a person on it or used for any other type of transport other than a possible evacuation using an evacuation sheet e This mattress is unsuitable for use during x ray examinations because of the risk of blurred images or artefacts that may lead to diagnostic errors 1 National Pressure Ulcer Advisory Panel European Pressure Ulcer Advisory Pan
11. per shift after changing the position in the bed or changing the comfort settings ko a t Slide a flat hand directly underneath the user s pelvic region Check that the user is not bottoming out and that there is sufficient air space to the underlying mattress A distance of one to two fingers is recommended If the distance is not sufficient increase the comfort setting to the next upper level Wait 10 minutes before carrying out a new hand check 4 4 Alarms When the pressure in the mattress drops to an abnormally low level the yellow low pressure lamp lights and an acoustic alarm sounds The alarm is deactivated by pressing the alarm off button If the problem is not resolved within 3 minutes the alarm will sound again The yellow lamp flashes until the pressure is corrected and reset See section 8 for troubleshooting WARNING If the air pressure remains low check the mattress and tubes for damage Repair if necessary or contact the manufacturer or your local dealer for help p 4 5 CPR In the event of an emergency requiring CPR cardiopulmonary resuscitation Quickly pull the CPR strap remove the quick connectors at the side of the pump to deflate the mattress as fast as possible On this product there are two possible positions for the CPR valve at the head end 1a or the foot end 1b Position 1a At the head end of the mattress Position 1b On tubes between the on the right side user s right
12. re and wash instructions The product also displays what side is up down and head foot 9 3 Symbol Definitions Read the user manual gt Not in domestic refuse follow recycling instructions Place on top of existing mattress Machine wash at 70 C w Do not tumble dry Wipe off Temperature range Manufacturer GROUp AS G Inma gt Yuna Group of function X XX A Recommended user weight x Counteract shear force Machine wash at 95 C Class Il equipment Double Insulated l XXXX XX XX Year Month Day CE CE marked acc to MDD 93 42 EEC Category Patient information category pa Heel function A lt 1 Chlorine gt Do not iron Type BF equipment The mattress should only be used lying in the lengthwise direction on the mattress 16 i Foot placement 9 Recycling ry Directly on bed base ex Do not wash Q Tumble dry RA Do not dry clean Paa IP code 10 Spare parts Size 80x200 cm 85x200 cm 90x200 cm Art no Cc 904020 CC 9046 CC 9047 CC 901001 CC 901020 CC 901002 CC 901021 CC 901003 CC 901022 Top part Olivia Bottom part CuroCell CC 904010 CC 903010 CC 9040 CC 9030 CC 9041 CC 9031 Description Top part CuroCell LF low friction Top part CuroCell LF low friction Top part CuroCell LF low friction Air cell non ventilated Air cell ventilated Air cell non venti
13. rm off 4 Static button 6 Fuses 2 8 Air filter 2 3 Pump functions 1 Power ON OFF 1 0 Einoras E arw aaia Button to turn the power on or off STATIC 4 5 2 Lamp for normal pressure NORMAL PRESSURE 3 6 At start up Lights when the mattress has reached the set N 7 A pressure and is ready for use f LJ X oe 8a 1 During operation Steady green light while pressure is normal a 3 Comfort setting Suitable air pressure is set using the comfort dial 1 8 Guide values are found in the table on the pump NOTE If the head of the bed is raised pressure must be increased at least one 1 level 4 Static mode When the STATIC button is pressed the alternating process is terminated and the pressure between the cells evens out The function can be used in association with patient care repositioning transferring in and out of bed Turn off this function by pressing the button It will return automatically to alternating mode after approx 30 minutes 5 Low pressure The pump has an acoustic and light alarm that is triggered when pressure in the mattress is too low The alarm continues to sound until the ALARM OFF button is pressed on the side of the pump In the event of an alarm check the following Disconnected pinched tubes open CPR valve leakage at air cell connector 6 Alarm off This turns off the acoustic alarm It will be activated again after about 3 minutes if
14. s Top part CuroCell LF low friction Top part with increased sliding function can be used for example heavy immobile patients 11 3 Other accessories To ensure the best level of comfort we recommend use of accessories from Care of Sweden since our products are designed to work well together Please contact Care of Sweden or your local dealer for accessories or more information on our products 18 12 Warranty 12 1 Scope This CuroCell system is covered by a 2 year warranty for manufacturing defects The warranty does not apply to normal wear and tear or product impairment or damage stemming from negligence or improper handling care 12 2 Service and maintenance This CuroCell mattress is a medical device in accordance with MDD 93 42 EEC that is used as an aid in preventing and treating pressure ulcers For this reason we recommend that the pump be regularly serviced and inspected to maintain functionality and performance Like other technical devices a pump that is properly cared for works better and lasts longer Service and maintenance must always be performed by Care of Sweden or one of its authorised technicians Only use spare parts that have been approved by Care of Sweden Use of non approved spare parts invalidate the warranty A warranty claim must be submitted to Care of Sweden before the product is returned Please contact Care of Sweden or your local dealer if you have questions regarding warranty and ser
15. tact with fire sources 4 Position and use the product in a manner that prevents it from becoming trapped or damaged NOTE Be particularly aware of trapping damage when using bed rails 5 Never use the product if the pump power cord or plug is defective the pump casing is damaged or it is not functioning properly Contact a service technician for examination and repair 6 Regularly check product functionality through a hand check see section 4 3 Also check that the comfort setting is correct as indicated in the table on the pump 7 Continual monitoring is required when the product is used for individuals needing special supervision such as children 8 The mattress are protected with a hygiene cover Please try to avoid the simultaneous use of multiple hygiene cover with this mattress Because of the risk of vapour permeability can be destroyed 9 The hygiene cover does not allow liquid or air to penetrate but is vapour permeable Make sure that the patient is positioned correctly to avoid the risk of suffocation 10 Prevent damage to the hygiene cover from sharp objects 11 If the cover are supplied with handles on the long sides theses handles are only to be used for lifting and moving the mattress itself All other use will be at your own responsibility and is not covered by the product warranty eee eee ee eee reer rer errr terre rer rer rer rere rere er rere terete erie terre rer rer rery 1 Introduction CuroCell
16. the pressure does not return to normal In case of problems with the alarm see chapter 8 for troubleshooting 3 Assembly Installation When unpacking the mattress check that none of the parts are damaged In the event of damage contact the manufacturer or your local dealer before using the product 1 Place the mattress on top of an existing mattress and fasten it by passing the elastic straps around the corners of the underlying mattress If the mattress has an integrated foam core it can be placed directly on the bed frame Fix the mattress to the bed frame using the bands on the cover Note e Make the bed with sheets for added comfort e Check that the CPR valve is closed e The mattress must be attached to the bed s moving parts if it is height adjustable e Check that all the snaps in each cell and cover are fastened correctly Hang the pump at the foot end of the bed or place it on a flat stable surface Route the power cord so that there is no risk of it causing someone to trip being run over by the bed wheels or getting pinched when the bed is raised or lowered If a cable holder is used optional open the snaps position the power cord in the pocket and close the snaps Connect the air tubes from the mattress to the side of the pump There is an audible click sound when the connectors are properly attached Check that the power switch on the side is set to O position Off Connect t
17. ver Pull out the fuse holder 3 Fit a new fuse of the same type T1AL 250 V and screw the cover back on There are two fuses 1 Remove the small round cover plate at the back of the pump 2 Lean the pump backwards and the filter slides out of its holder 3 Place the new cleaned filter in the holder and refit to the cover plate If the pump is used in a dirty environment the filter should be checked regularly 13 8 Troubleshooting Problem Solution The pump won t start Check that the power cord is connected to the power socket Check the fuses Alarm is on Check that the quick connectors to the air tubes mattress is deflating on the side of the pump are properly connected Check that the CPR valve valves are closed Open the mattress and check for leaks from air cells or tubes Check that all tube connectors are properly fitted and connected No air in head cells When the mattress is inflated the 3 static head cells are filled last Patient is bottoming The comfort setting may be set incorrectly Increase the pressure and wait a few minutes Do a hand check Refer to section 4 3 Check that the CPR valve is properly closed Mattress moves around Check that the air mattress is secured to the underlying mattress with the elastic straps Not if the mattresses is equipped with an integrated foam core Some cells have less air This is normal since the air flow alternates between every other cell for a
18. vice 12 3 Life expectance This product s expected life is estimated to be 5 years 13 Other information 13 1 Disassembly and recycling Except for certain parts of the pumps energy recovery is possible for almost all material in CuroCell products through incineration in waste incineration facilities Pump A used CuroCell pump must not be disassembled It must instead be taken to a recycling centre and sorted as electronic waste Mattress A used CuroCell mattress must be taken to a recycling centre and sorted as combustible waste For more information please contact Care of Sweden or your local dealer 13 2 Returns and complaints Complaints are to be reported to the supplier Return of a product shall follow the supplier s instructions and can only be done after the complaint has been reported Damages due to the transportation should be reported to the transportation agent 19 wd Manufactured by care of sweden Supporting Life Contact Phone 46 771 106600 Fax 46 325 12840 Email info careofsweden se Internet www careofsweden com Address Care of Sweden AB P O Box 146 SE 514 23 Tranemo SWEDEN Visit address Fabriksgatan 5A SE 514 33 Tranemo SWEDEN Cargo address Byns v g 4A SE 514 33 Tranemo SWEDEN Supplied by Care of Sweden and CuroCell are trademarks of Care of Sweden AB Care of Sweden AB 2015 20
Download Pdf Manuals
Related Search
Related Contents
SpikeDB Version 1.6 User Manual Bedienungsanleitung Silver Schmidt 取扱説明書(698.6 KB) TAFCO WINDOWS NU2-067S-I Instructions / Assembly 601-336C - CDN 3083-B100&CDN 3083 Instruction Manual Bedienungsanleitung Manuel d`utilisation caldera mural a gas estanca “touch control” 本体取扱説明書 Copyright © All rights reserved.