1. AA or NiMH batteries last for about 160 on hours or up to 6000 tests The following pages will explain in detail the operation of your FC Series portable breath tester Unlock the Power of Alcohol Testing Front View with Mouthpiece Large Graphic LCD Screen displays functions and results Mouthpiece E d AUTO TEST B Y Battery Icon indicates 00011 Mouthpiece Removal Tab battery level eecocccccecccccecccecee Buttons set eere A calibration standard fee Cee 0 o ee Seis Execute Button enables you to perform Function Button e cc op fichas scrolls through menus and saves changes Power Button ccccesccccccccscces turns unit on and off also returns you to main testing mode Wrist Strap Attachment Loop Features Large Graphic LCD Display Capable of showing numbers letters icons and plain English text messages e Automated Calibration Software controlled adjustments no tools necessary Auto Test Mode The easiest way to take a test Tester automatically takes deep lung sample when subject is at end of breath e Manual Test Mode Enables operator to control exact point of breath sample Passive Test Mode Checks for the presence of alcohol in the breath or in an open container without using a mouthpiece User Selectable Test Order Allows choice of either Auto Test or Passive Tes
2. E Table of Contents Introduction A 4 Front VW ita 5 Feat S oraire E E T ee ere eee 6 Preparation Installing Dat MES ninio id ina ida 7 T rning the FC On and OR ita 7 Observmg the SUBI A aaa 8 Attaching a PV CO nn ai 8 Breath Testing Modes Testing Modes EXPO raid di 9 End of Breath Precise Volume Explained essen 9 COTO UC HS an Automatic TES ann aiei nan aidai aaa A ia 10 Manual Override During an Automatic Test oncnnnicinnnnnnnicinnicararoncan nro raro nanonannoronracca nano 10 Conducting a Man al TES nait A sentent aide 11 Conducting a Passive TS dio 11 Operation II rae eaa aiie ees 12 ICONS UE RS ii a ut on nn nn 12 TESE RSR e a ea be eo aime 12 Viewing Previous Test Results jisisdacic sos sunsaieisatiaccasieeasatacuneiestansaadsdctiaanansuansiaaraedadepnasaaalaadinecenandl 12 Removing the Moliets da 12 User Settings Setting the Default Test OS ne ed vu ad 13 Setting the Tigger MOSS nano di 13 Calibration Settings Calibration Cal Check EX Ines it 14 Wet Bath and Dry Gas Explained ninia aid 14 15 Selecting the Calibration Standard usina ii 15 Selectina the Standard PE di 16 Table of Contents Wet Bath Calibration Wet Bath Simulator Set Up ai ni iia 16 Wet Calibration is anni 17 ide A A 18 Dry Gas Calibration Ibi Gas Tank Set UD NN 19 Dry Gas e AAA a an au da dt ie 19 20 Dr Gas Calibration Checa as 20 21 Maintenance A A 21 C
3. Press the Execute button to select between END OF BREATH and PRECISE VOLUME Press the Function button to save the setting Momentarily press Power to return to the testing mode User settings flow chart is shown on p 26 Calibration Cal Check Explained Calibration of an FC sets your unit to a known alcohol concentration to enable accurate BAC Breath Alcohol Concentration results You can use dry gas or wet bath solution to calibrate your FC1O Most commonly used solutions are 100 080 or 040 BAC The FC1O must be between 68 and 95 F 20 35 C to calibrate Lifeloc recommends you calibrate your FC Once every 12 months regardless of how many tests you have performed Or at intervals specified by your Internal Policies Quality Assurance Plan or State Regulations Or after two failed Calibration Checks A Calibration Check simply verifies the FC was calibrated correctly and is within the acceptable accuracy range Calibration check is also referred to as External Calibration Check Accuracy Check Verification and Cal Check Lifeloc recommends you perform a calibration check on your FC Once every 30 days Or at intervals specified by your Internal Policies Quality Assurance Plan or State Regulations Wet Bath and Dry Gas Explained You can calibrate and check your FC10 using either the wet bath or dry gas method You must first set your FC1O to recognize which me
4. 160 hours or up to 6000 tests We recommend storing the FCIO in temperatures between 32 130 F 0 55 C Accessories amp Supplies Lifeloc offers a complete line of accessories and supplies for your FC10 including For the most current listing of supplies for your FC10 go to www lifeloc com Click on Supplies gt FC Series e www lifeloc com store supplies fcseries html You can place your order online via fax to 303 431 1423 or call Customer Service at 303 431 9500 or 800 722 4817 a Mouthpieces Rubber Grip Carrying Cases Certified Simulator Dry Gas Standards Wet Bath Cleaning Wipes Solution and Calibration Kits Simulators User Settings Chart Calibration Display Chart Perform a Function Cal Check Execute Execute XXX Check Manual Test Function Perform a Calibration Execute XXX Calibrate Passive Test Function il Function Function Trigger Mode Cal Cal Execute Result Function Function Display LCD Function Settings Standard Function Function Calibration Function Function Order Function Function Lifel c 12441 West 49th Avenue Suite 4 Wheat Ridge CO 80033 303 431 9500 800 722 4872 fax 303 431 1423 Hours of Operation 7 30 am 4 30 pm MST If you are calling outside of these hours please leave us a voice message We will contact you the following business day From our single location in
5. SAR PP TN 21 Batterie A A en 21 On Screen Messages Messages Expand ou 22 Service Lifeloc Factory Wa ANNE nb ani ia ain ul 23 Extended Service Pi is 25 SOVICE ND ETE 23 General Usage FC10 Supplies Accessories and Supplies ainia an ann 25 Menu Charts User Settings Chal ani 26 Calibration Display Ena io 26 Attention FC10 Operator Congratulations on your purchase of a Lifeloc FC10 For over 25 years Lifeloc Technologies has been providing advanced alcohol testing equipment training to Law Enforcement and Corrections Professionals We are the leader in product innovation precision instruments ease of use Five Star Customer Care The FCIO breath alcohol tester is manufactured in Wheat Ridge Colorado by Lifeloc Tech nologies Inc Lifeloc offers premium quality products combined with exceptional service and technical support The FCIO is a state of the art breath alcohol tester that is software based and incorporates unique cutting edge technologies Because of the advanced FC10 design Results on a positive test register within 10 seconds You can take another test within 30 seconds after a positive Repeatability is not compromised e Your FC10 will automatically take the test when it senses a deep lung sample is delivered e Your FC10 will provide an accurate test or else explain to you why it cannot and even provide suggestions on how to proceed to complete an accurate test on your subject
6. Wheat Ridge Colorado we manufacture and service our products with you in mind Every product is designed with quality and ease of use as our priorities And our 5 Star Service is the fastest in the industry Lifeloc breath testers are used across the US and in over 35 countries We carry all of the accessories and supplies for your breath test equipment Please call us to reorder supplies and accessories or for information on purchasing additional testers
7. a 15 minute observation period in which they are not permitted to put anything in their mouth should be observed before testing This will ensure all residual alcohol from any source has completely dissipated and test results will be valid Attaching a Mouthpiece Remove the mouthpiece from its wrapper making sure not to touch the end which the subject will be blowing into Attach the mouthpiece to the port on the back of the FC1O Line up the mouthpiece port over the holes in the back of the mouthpiece Press in place Ensure it is securely attached Breath Testing Modes Explained The FCIO is capable of conducting Automatic Manual and Passive tests Automatic Test is the easiest way to achieve a deep lung sample The FC1O monitors the subject s breath and automatically takes the sample near the end of the breath flow Manual Test is normally used only when the subject is unable to provide a sufficient breath sample for the automatic test Passive Test is a quick screen to detect alcohol but is not designed to quantify the results Passive results are reported as POS if alcohol is detected NEG if alcohol is not detected In this mode no mouthpiece is required End of Breath and Precise Volume Explained In Auto Test mode the unit can be set to take a sample either When the subject nears the end of the exhalation End of Breath or e When it detects 1 5L of breath Precise Vol
8. instructions on page 16 Calibration Pump Disconnect long tube from the output port Attach a mouthpiece adapter securely to the output port on the simulator lid Refer to your simulator manual for location Turn the FC1O on Attach a mouthpiece to the back of the unit Press the Function button until display reads CALIBRATION then press the Execute button Display reads WET CHECK Press the Function button until display reads WET CALIBRATE Verify the CAL STANDARD is set to the concentration of certified alcohol solution you will be using when you calibrate To select new Calibration Standard see page 16 Slide the FC10 mouthpiece over the mouthpiece adaptor on the simulator Be prepared to blow into the tube for up to 10 seconds Start blowing Blow through the input tube or use a calibration pump to create and maintain 1 2 of bubbles on the surface of the solution Press Execute to proceed and start a 3 second countdown Press Execute to take a sample Continue blowing for another 3 seconds Stop blowing If successful the display will read CAL COMPLETE If no alcohol is detected display reads INVALID CALIBRATION Please repeat calibration setup and test starting with wet bath simulator setup Disconnect the FC10 and wait at least two minutes before conducting a Cal Check to verify the accuracy of your calibration Note Following calibration a Cal Check is r
9. over the phone Repairs are completed within 5 days for up to 5 units or it s free Notice The FCIO is a professional device designed to be used by trained operators in conjunction with a specified periodic maintenance and calibration calibration check regimen Use by untrained operators or without periodic calibration or calibration checks may result in invalid results or incorrect interpretation of results FC10 is not to be used by children under 12 years of age DO NOT DRINK AND DRIVE Lifeloc strongly recommends that no vehicle or machinery be operated after alcohol consumption Even small quantities of alcohol can result in driving impairment The FCIO is not waterproof and should not be immersed in or exposed to excessive water The FC1O is not suitable for use in a potentially explosive environment The FC10 cannot be used inside an oxygen tent If instrument will not be used for more than 6 months remove battery to avoid damage to the instrument caused by leaking battery acid Disposal of Instrument At the end of the instrument s service life e Do Not dispose of the FC10 as unsorted municipal waste e Dispose of the FCIO in accordance with national waste disposal regulations Specifications uri NUR O aa 2 6 x 5 x 125 66 x 127 x 32 mm Weight w batteries 9 oz 255 grams Measurement Range 000 to 600 BAC PRE CUA ii ie 005 BAC up to 100 BAC 5 for 100 400 BAC Battery LIIS isa nai Approximately
10. second countdown Press the Execute button again to take a sample Continue holding the Regulator button for another 5 seconds Stop pressing gas flow will stop If successful the display will read CAL COMPLETE If no alcohol is detected display reads INVALID CALIBRATION Please check gas level on your tank regulator and replace tank if empty Repeat dry gas calibration instructions on page 19 Disconnect the equipment and wait at least two minutes before conducting a Cal Check Note Following calibration a Cal Check is required before the unit allows you to conduct any tests Calibration display flow chart is shown on p 26 Performing a Dry Gas Calibration Check Prepare the dry gas tank according to its instructions on page 19 Turn FC1O on Attach a mouthpiece to the back of the unit Press the Function button until display reads CALIBRATION Press the Execute button Display reads DRY CHECK Verify the Calibration Standard is set to the corrected BAC Using the Altitude Correction Factor ACF chart on the dry gas tank calculate the corrected BAC Corrected BAC tank BAC x ACF To select a new Calibration Standard see page 16 Attach the FCIO mouthpiece to the regulator by sliding it over the mouthpiece adapter ensuring a snug fit Be prepared to press the Gas Tank Regulator button for up to 1O seconds Start pressing gas flow will start Press the Execute button to proceed and sta
11. 0 and 210 BAC If no alcohol was detected the display will read INVALID CHECK Repeat wet bath calibration check instructions Calibration display flow chart is shown on p 26 Dry Gas Tank Set Up Attach the regulator to the dry gas tank Attach the small tube to the output port on the regulator Securely fit the mouthpiece adaptor to the small tube on the regulator Note Do not store the dry gas tank with the regulator attached The regulator is not designed as the primary sealing mechanism Leaving the regulator attached may result in a leak Performing a Dry Gas Calibration Prepare the dry gas tank according to its instructions above Turn FC1O on Attach a mouthpiece to the back of the unit Press the Function button until the display reads CALIBRATION Press the Execute button Display reads DRY CHECK Press the Function button until the display reads DRY CALIBRATE Verify the Calibration Standard is set to the corrected BAC Using the Altitude Correction Factor ACF chart on the dry gas tank calculate the corrected BAC Corrected Standard tank BAC x ACF To select a new Calibration Standard see page 16 Attach the FC10 mouthpiece to the regulator by sliding it over the mouthpiece adaptor ensuring a snug fit Be prepared to press the Gas Tank Regulator button for up to 1O seconds Start pressing gas flow will start Press the Execute button on FC to proceed and start a 5
12. alcohol level is graphed and will be displayed before the result Viewing Previous Test Results LAST TEST RESULT wr 020 BAC E Press to remove mouthpiece M Mente Press the Function button until LAST TEST RESULT is displayed The results of the last test will be displayed Momentarily press the Power button to return to the testing mode Removing the Mouthpiece Remove the mouthpiece by pushing straight down on the tab at the right of the display screen DO NOT BEND the tab Setting the Default Test Order Test Order 1 Default Test Order 2 Test Order 3 Auto Test Passive Test Auto Test Manual Test Auto Test Manual Test Passive Test Manual Test Press the Function button until the display reads SETTINGS Press the Execute button Display reads TIME Press the Function button until the display reads DISPLAY SETTINGS Press the Execute button Display reads LCD CONTRAST Press the Function button Display reads TEST ORDER Press the Execute button to toggle between Test Order 1 2 and 3 above Press the Function button to save the changes Momentarily press Power to return to the testing mode Please note By selecting Test Order 3 you will disable passive testing User settings flow chart is shown on p 26 Setting Trigger Mode From the SETTINGS display press the Execute button Press the Function button until the display reads TRIGGER MODE
13. equired before the unit allows you to conduct any tests Calibration display flow chart is shown on p 26 Performing a Wet Calibration Check Prepare the wet bath simulator according to its instructions on page 16 Disconnect long tube from the output port Attach a mouthpiece adapter securely to the output port on the simulator lid Refer to your simulator manual for location Turn the FC1O on Attach a mouthpiece to the back of the unit Press the Function button until display reads CALIBRATION then press the Execute button Display reads WET CHECK Verify the CAL STANDARD is set to the concentration of certified alcohol solution you will be using when you check the calibration To select a new Calibration Standard see page 16 Slide the FC10 mouthpiece over the mouthpiece adaptor on the simulator Be prepared to blow into the tube for up to 10 seconds Start blowing Blow through the input tube or use a calibration pump to create and maintain 1 2 of bubbles on the surface of the solution Press Execute to proceed and start a 5 second countdown Press Execute to take a sample Continue blowing for another 5 seconds Stop blowing Read the result It should be within 005 BAC of the standard used If your solution is greater than 100 BAC accurate results will be within 5 Example A 100 BAC solution should read between 095 and 105 BAC A 200 BAC solution should read between 19
14. ers alcohol free cleaning wipes specific for your unit See page 25 for more details Batteries Your FC10 default setting is for four AA batteries However you do have the ability to use NiMH rechargeable batteries as well To use NiMH insert the batteries as directed and power on the unit Press the Function button until the display reads SETTINGS Press the Execute button Display shows TIME Press the Function button until the display shows BATTERY TYPE Press the Execute button to change your battery settings Press the Function button to save your settings Momentarily press the Power button to return to the testing mode The four batteries in your FC10 should last for about 160 hours of on time or up to 6000 tests It is recommended you use high quality alkaline batteries with your unit Messages Explained Message lt 1 3L Retest or Try Manual Test gt 06 ess External Interference Flow Error Retry amp Blow Steadily e gt Invalid Calibration Cal Check Log Empty Low Battery sessresesseg Pump Reset Needed J e Temperature 4B gt Timeout ess gt Explanation Breath flow ended before the subject blew 1 3 liters of breath Instruct subject to try again or use manual test mode BAC is unusually high and above 0 6 BAC Subject may require medical attention External interference has been detected Move to a different location and try again Exhalati
15. of certified solution or as your altitude corrected standard if using a dry gas tank See p 14 Press the Function button to save the settings Momentarily press Power to return to the testing mode Once you set the standard you do not have to set it again unless you change solutions or elevation dry gas only Selecting the Standard Type Press the Function button until display reads CALIBRATION Press the Execute button The display reads either WET CHECK or DRY CHECK Press the Function button until display reads CAL SETTINGS Press the Execute button The display reads CAL STANDARD Press the Function button until display reads STANDARD TYPE Press the Execute button to choose between DRY GAS and WET BATH Press the Function button to save settings Momentarily press Power to return to the testing mode Wet Bath Simulator Set Up ror calibration amp Calibration Check Pour a bottle of certified alcohol solution into the simulator jar and hand tighten lid Connect long tube from input port to output port so no alcohol escapes while simulator heats up Plug in the simulator and turn it on The simulator automatically heats the solution to 34 C 93 2 F in about 5 to 10 minutes Proper operating temperature is important for accuracy so be certain to check temperature before proceeding Performing a Wet Bath Calibration Prepare the wet bath simulator according to its
16. on not complete or interrupted Instruct subject to blow steadily as long as they can No alcohol was detected Repeat setup then retest There are no results in the memory Battery voltage is too low to take a test Replace batteries The pump needs to be reset Follow onscreen instructions When calibrating or taking a test unit is outside of temperature limits Calibration range 68 95 F 20 35 C Testing range 32 130 F O 55 C User blew for gt 15 seconds Please try again blow harder Lifeloc Factory Warranty The FC1O comes with a one year limited parts and labor warranty effective on the date of purchase by the end user The Warranty covers Parts and labor on covered repairs Software updates as applicable Air freight back to the customer after the unit is repaired U S only The Warranty does not cover Freight to the Lifeloc factory Misuse abuse negligence or accidents Extended Service Plans Extended Service Plans are available for your FC10 These provide complete coverage for an additional year at a reasonable cost and include free factory diagnostic checks Call Lifeloc for details or purchase online at www lifeloc com technical service html Service If your FC1O should require repairs or maintenance Lifeloc is here for you Just an email or phone call will put you in contact with our technical support personnel Many minor adjustments can be made
17. rt a 5 second countdown Press the Execute button again to take a sample Continue holding the Regulator button for another 3 seconds Stop pressing gas flow will stop Read the result It should be within 005 BAC of the corrected standard used If your solution is greater than 100 BAC accurate results will be within 5 Example A 082 BAC corrected standard should read between 077 and 087 BAC If no alcohol is detected display reads INVALID CHECK Please check gas level on your tank regulator and replace tank if empty Repeat dry gas calibration check instructions on page 20 Calibration display flow chart is shown on p 26 Fuel Cells Fuel cells are highly durable sensors that are capable of providing accurate breath alcohol results for years There are however a few precautions you should take to make certain that these devices perform for the longest period of time possible Use the device Fuel cells like moisture so it is a good idea to take tests periodically to provide needed moisture to the fuel cell especially in dry climates You do not need alcohol just breath Avoid cigarette smoke Make certain no one is permitted to blow cigarette smoke into the unit This can damage or destroy the fuel cell Cleaning Use of a mild disinfectant cleaner and a soft cloth on the outside of the case is recommended periodically to keep your unit clean Do not use alcohol to clean the unit Lifeloc off
18. t default mode Auto Shut Off Preserves battery life Fast Simple Operation While the FC10 contains a host of features it is still very easy to use Automatic Backlight Easy viewing of test results either day or night Exceptional Battery Life Up to 160 hours or 6000 test operation using four AA or NiMH rechargeable batteries Installing Batteries Press in and down on the battery door located on the back of the FC1O Install the four AA Alkaline batteries in the direction of the symbols in the battery case Close the case by pushing up on the battery door until it locks shut See page 21 for directions on using rechargeable batteries Turning the FC On and Off Press and hold the Power button on the bottom of the front of the unit until it beeps The FCIO performs an automatic internal diagnostics check To turn the unit off press and hold the Power button until it beeps twice The unit will shut down Note Momentarily pressing the Power button when the unit is on will return you to the main menu Observing the Subject The FCIO provides a highly accurate reading of breath alcohol acquired by sampling deep lung air Readings will also detect residual mouth alcohol To prevent mouth alcohol from affecting a test make certain that the subject is not allowed to put anything in their mouth for 15 minutes prior to taking a test If the subject just took a drink
19. thod or which Standard Type you will be using Once you choose the Standard Type the FC10 will store that information in memory and you do not have to set it again unless you change to a different method of performing a calibration calibration check Dry Gas Calibration requires that prior to calibration you enter the corrected Standard Value based on your altitude or elevation Using the chart on the outside of the canister multiply the number next to your elevation by the standard Example Denver Colorado s elevation is 5200 ft above sea level and dry gas standard 100 BAC Correction factor from tank is 820 e Corrected Standard 100 x 820 082 If you move to a significantly different altitude before calibrating you will have to change the standard in the FC1O Wet Bath Calibration does not require altitude correction Note The FC10 comes from the factory set for Wet Bath type and 100 BAC Calibration Standard Selecting the Calibration Standard Press the Function button until display reads CALIBRATION Press the Execute button The display reads either WET CHECK or DRY CHECK Press the Function button until display reads CAL SETTINGS Press the Execute button The display reads CAL STANDARD Cal Standard 0 100 c N 7 Sae Use the or button to change the number to the BAC level of standard you will be calibrating to It should be the same as on the bottle
20. to the mouthpiece firmly and steadily for as long as they can When they are near the end of their breath press the Execute button Read the result Please note manual test mode is pressure activated If the subject does not blow air into the mouthpiece the test can not be conducted Conducting a Passive Test No Mouthpiece Turn the FC1O on Press the Function button if necessary until the display reads PASSIVE TEST Hold the FC1O sample port orange colored opening labeled Port on the T back of the FC10 about 4 inches from the f subject s mouth J Have the subject blow toward the port Press the Execute button while the subject is blowing Read the result The unit will only display POS or NEG It will not display the actual numerical result Note A passive test can also be done over an open container to detect the presence of alcohol Breath Flow As the subject blows into the mouthpiece the FC1O will show a graph of the breath flow on the display as well as showing the amount of liters Liters are shown in the upper left corner Test Results After the alcohol is graphed the test result is displayed The result will remain on the screen until the Function or Power button is pressed The last test result is retained in memory until the next test is administered Alcohol Curve If the FC1O detects alcohol the
21. ume Both will give accurate results Precise Volume may work better with uncooperative subjects To select END OF BREATH or PRECISE VOLUME mode see Trigger Mode instructions on p 13 Conducting an Automatic Test Turn the FC10 on Verify the display reads AUTO TEST Attach the mouthpiece to the back of the unit Instruct subject to blow into the mouthpiece firmly and steadily for as long as they can But not necessarily as hard as they can The unit calculates volume and will give an error message if the subject cannot reach 1 3 liters Read the result After taking a test the FC10 will display the results in large numbers on the display Press the Function button to return to the test mode Manual Override during an Automatic Test Note This feature allows the completion of a test in the occasional instance when the subject may have diminished lung capacity and cannot activate the Auto Test Turn the FCIO on Attach a mouthpiece to the back of the unit and verify the display reads AUTO TEST Instruct the subject to blow into the mouthpiece firmly and steadily for as long as they can When they are near the end of their breath press the Execute button Read the result Conducting a Manual Test Turn the FC1O on Attach the mouthpiece to the back of the unit Press the Function button until the display Er reads MANUAL TEST Instruct the subject to blow in
Download Pdf Manuals
Related Search
Related Contents
AgScan,_Inc - Drone Laws Blog by Antonelli Law Manual - Ecotech [U4.81.03] Opérateur CALC_CHAM_ELEM User Manual MANUAL DE INSTRUÇÕES KA 214 pvd 600 / pvm user manual in original 取扱説明書(PDF) Mode d`emploi Exélia Art Contours d`oreille 表紙~24ページ一括 ※約2MBのファイルです。 Canada - Buyandsell.gc.ca Copyright © All rights reserved.