User manual for FA-665B (power supply)

image

Contents

1. For example knob 6 is at the position of counterclockwise to end or current of slave output excesses current limiting protection point at this time the voltage of slave output will not track the voltage of master So knob 6 should be rotated clockwise to maximum then the two outputs are in series Bv series connection if there is power output proper leads corresponding to output power should be used to short connect the negative terminal of master output with positive terminal of slave output reliabiv Since it is shorted bv a switchinside the unit current will pass on the shorted switch when there is power output This will affect the reliabilitv of the unit 2 2 3 Parallel using of the two adjustable outputs Press in switchs 15 at this time the two output are in parallel adjust voltage adjustment 20 of master output the voltage of two wavs keep same and slave output CC indicator 3 lights on When the two outputs are in parallel the CC adjustment 6 of slave output does not work When used as CC supply simply adjust the CC adjustment 21 of master output at this time output current of both master and slave are controlled by it and are same output current is up to double of independent s maximum current While the two outputs in parallel proper leads corresponding to output power should be used to short reliably the two positive terminal and the Two negative terminals of master slave output separately so as to
2. 0 MI2066 FA 665B ADJ USTABLE DC REGULATED POWER SUPPLY FA 665B USER S MANUAL SAFETY NOTES Read the user s manual before using the equipment mainly SAFETY RULES paragraph The symbol A on the equipment means SEE USER S MANUAL In this manual may also appear as a Caution or Warning symbol WARNING AND CAUTION statements may appear in this manual to avoid injury hazard or damage to this product or other property USER S MANUAL VERSION Version Date September 2015 FA 665B USER S MANUAL SAFETY RULES A These instruments fulfill the regulations of CE LVD EN 61010 2001 and CE EMC EN 55022 1998 A1 2000 EN 55024 1998 EN61000 3 2 2000 EN61000 3 3 1995 To ensure safe operation of the equipment and eliminate the danger of serious injury due to short circuit arcing the following safety precautions must be observed Damages resulting from failure to observe these safety precautions are exempt from any legal claims whatever Prior to connection of the equipment to the mains outlet check that the available mains voltage corresponds to the voltage setting of the equipment Connect the mains plug of the equipment only to a mains outlet with earth connection Do not place the equipment on damp or wet surfaces Do not subject the equipment to direct sunlight or extreme temperatures Do not subject the equipment to extreme humidity or dampness Replace a defective fuse onl
3. current designed in high stability and performance circuit In constant voltage state the output voltage can be arbitrarily adjusted from O V on in the nominal range and in the state of constant current the output current can be adjustable from 0 A on in the nominal range The two outputs can be connected in parallel or in series while the master used for voltage or current adjustment The maximum output voltage in series is double of independent s and the maximum output current in parallel is double too There are volt and Amp meters LED for indicating each of the two outputs with high accuracy The one fixed way outputs 5 V voltage Due to the single chip integrated regulator this output has good stability and ripple factor and has reliable overload protection to protect the unit against being damaged whenever overload or short circuit The unit features in small size good performance novel appearance and etc it is the ideal power supply unit for science investigation college factory electronic appliance maintenance and etc September 2015 1 USER S MANUAL FA GG5B E mer B stans M m PARAL m j s O j l l SLAVE po i f y j MASTER 0 30W0 5A es D 30VID SA A OO OOO Be DOW EE OOM AAA O 068 Figure 1 Front panel description Volt display indicating slave output voltage Amp display indicating slave output current Slave constant current indicator or tw
4. make load connected reliably with the two parallel outputs If the load is only connected to one of the output terminals unbalance mav be caused to current of the two outputs this mav also damage the series parallel switchs 15 The LED displav is in three digits To get more accurate measuring value vou should calibrate bv external circuit with precision measuring instrument September 2015 5 USER S MANUAL FA 665B8 CAUTION This unit has excellent protection function 5V output has reliable protection for current limit and short The two adjustable outputs have current limit protection As there is controlling circuit for regulating transistor s power loss in the circuit when short circuit occurs the power loss on large power transistors is not very high it can t cause any damage to the unit But there is still power loss when Short circuit in order to reduce aging and energy consumption so this situation should be found as soon as possible and turn off power then exclude the faults When operating is finished put it in a dry place of good ventilation and keep it clean If it is not in use for a long period pull off the power supply plug for storage For maintenance input voltage must be cut off 6 September 2015 FA G65B USER S MANUAL Specifications MM Input voltage 110 VAC 127 VAC x 10 60 Hz 220 VAC 240 VAC x 10 50 Hz switchable Output voltage 2x0to30V Output current Oto5A Line regula
5. the indicator will light if the fixed output is over current Master constant voltage adjustment adjusting master output voltage Master constant current adjustment adjusting master output current value adjusting the current limit protection point Master constant voltage indicator the LED illuminates when the master output is in voltage regulated state Master constant current indicator the LED illuminates when the master output is in current regulated state Volt display indicating master output voltage Amp display indicating master output current USER S MANUAL FA GG5B 2 2 i Operating method 2 2 1 E I ndependence use of two adjustable output Set switchs 15 to spring out position When the adjustable output is used as CV output first should rotate clockwise the CC adjustment 6 and 21 to maximum then turn on power switch 18 adjust CV adjustment 5 and 20 till output voltage reach required voltage value At this time the CC state indicator 3 and 23 go out and the CV state indicator 4 and 22 light on Used as CC output after turning on power switch 18 first rotate clockwise the CV adjustment 5 and 20 to maximum while rotate counterclockwise the CC adjustment 6 and 21 to minimum then connect the required load again adjust clockwise adjustment 6 and 21 till output current reach the required current value At this time the CV state indicator 4 and 22 go out and the CC stat
6. X PROMAX ELECTRONICA S L Francesc Moragas 71 75 08907 L HOSPITALET DE LLOBREGAT Barcelona SPAIN Tel 93 184 77 00 Tel Intl 34 93 184 77 02 Fax 93 338 11 26 Fax Intl 34 93 338 11 26 http www promaxelectronics com e mail promax O promaxelectronics com
7. e indicator 3 and 23 light on Used as the CV output in general the CC adjustment 6 and 21 should be set to maximum but for this unit the current limiting protection point can also be set arbitrarily Setting procedure turn on power rotate counterclockwise the CC adjustment 6 and 21 to minimum then make the positive and negative output terminal in short connection and rotate clockwise the CC adjustment 6 and 21 till output current equal to the required current limiting protection point so the current limiting protection point is well set 2 2 2 l Series using of the two adjustable outputs Set Switchs 15 in series mode according to the front panel At this time turn the master voltage adjustment 20 and the slave out voltage tracks strictiv the master output voltage and the output voltage can be up to double of independent s maximum voltage voltage between terminal 14 and 9 Before the series connecting it must be examined if the negative terminal of both master and slave output are connected to case grounded terminal if they are must be disconnect otherwise short circuit will be caused in the slave output when the two outputs are connected is series 4 September 2015 USER S MANUAL When the two outputs are in series the voltage is controlled bv master output but current adjustment of two outputs is still independent Therefore attention should be paid to the position of the CC adjustment 6
8. ectric shock DIRECT AND ALTERNATING GROUND TERMINAL PROTECTIVE CONDUCTOR CAUTION REFER TO MANUAL FRAME TERMINAL FUSE EQUIPMENT OR COMPONENT EQUIPOTENTIALITV Ke TO BE RECVCLED Descriptive Examples of Over Voltage Categories Catl Low voltage installations isolated from the mains Catll Portable domestic installations Cat ill Fixed domestic installations Cat IV Industrial installations September 2015 FA 665B USER S MANUAL TABLE OF CONTENTS 1 INTRODUCTION sg a ren torera ts rie 1 1 1 Description LEE 1 2 OPERATION cccccce cece cece ee eeeeeseeeeeeeeeeeeteesguaeeeeeesetssgggaeereeesetsesagggeenernees 2 2 1 Control and description of front pannel e cece ee eee cece tees eeeeeeeeeeeeeeees 2 2 2 Operating method eee ee eect eee cnn rr 4 2 2 1 Independence use of two adjustable output ee eee e tent eee 4 2 2 2 Series using of the two adjustable outputs ne eee nn 4 2 2 3 Parallel using of the two adjustable outputs ee seen eee e eee 5 3 SPECIFICATIONS i en nn A ee ee en eon ete 7 4 MALNFENANGE ki stent ne itre itini o 8 4 1 Cleaning Recommendations 6 2 8 September 2015 FA 665B USER S MANUAL ADJUSTABLE DC REGULATED POWER SUPPLY FA 665B 1 INTRODUCTION The model FA 665B high precision DC regulated power supply has three ways output Of which two ways are adjustable and one way is fixed The two adjustable outputs can also be selected for constant voltage or constant
9. o ways parallel state indication the LED illuminates when the slave output is in current regulated state or the two adjustable outputs is in parallel Slave constant voltage indicator the LED illuminates when the slave output is in voltage regulated state Slave constant voltage adjustment adjusting slave output voltage Slave constant current adjustment adjusting slave output current value adjusting the current limit protection point Output indicator when the output buttom is pressed the indicator will light to show that the power supply has output Output on off buttom control the output state of the power supply Slave output terminal connecting the negative termina of load 2 September 2015 USER S MANUAL September 2015 Case ground connecting the case to ground Slave output terminal connecting the positive terminal of load Master output terminal connecting the negative terminal of load Case ground connecting the case to ground Master output terminal connecting the positive terminal of load Control switch for selecting the two adjustable outputs independent series parallel Fixed 5V output terminal connecting the negative terminal of load Fixed 5V output terminal connecting the positive terminal of load Power switch the unit is ON when this button switch is depressed while CV LED 4 22 or CC LED 3 23 illuminating Fixed output over current indicator
10. tion CC 2 x 10 3 3 mA Fixedoutput LO mV O Two adjustable outputs CV 1x10 4 2 mV CV 1x10 4 5 mV CC 2 x 10 3 3 mA CC 2 x 10 3 5 mA 10 mV Ripple and noise Two adjustable output CV 0 5 mV rms CV 1 0 mV rms CC lt 3mA rms Fixed output 10 mV rms Protection Current limit Display accuracy Volt indication 0 2 Rdg 2 digits 2 5 Full Scale Amp indication 1 0 Rdg 2 digits 2 5 Full Scale Mechanical features Dimensions W x H x D 60 x 160 x 340 O kg Accesories 1 x Quick Reference Guide 2 x Fuses Note Equipment specifications are set in these environmental operating conditions Operation outside these specifications are also possible Please check with us if you have specific requirements RECOMMENDATIONS ABOUT THE PACKING It is recommended to keep all the packing material in order to return the equipment if necessary to the Technical Service September 2015 7 USER S MANUAL FA 665B8 4 MAINTENANCE A 4 1 Cleaning Recommendations To clean the cover take care the instrument is disconnected Do not use scented hydrocarbons or chlorized solvents The cover should be cleaned by means of a light solution of detergent and water applied with a soft cloth Dry thoroughly before using the system again Do not use for the cleaning alcohol or its derivatives these products can attack the mechanical properties of the materials and diminish their useful time of life 8 September 2015 AxPROMA
11. y with a fuse of the original rating Never short circuit fuse or fuse housing Do not exceed the maximum permissible input rating Conduct measuring works only in dry clothing and in rubber shoes i e on isolating mats Comply with the warning labels and other info on the equipment Do not insert metal objects into the equipment by way of the ventilation slots Do not place water filled containers on the equipment danger of short circuit in case of knock over of the container Do not operate the equipment near strong magnetic fields motors transformer etc Do not subject the equipment to shocks or strong vibrations Keep hot soldering iron or guns away from the equipment Allow the equipment to stabilize at room temperature before taking up measurement important for exact measurement Do not modify the equipment in any way Do not place the equipment face down on any table or work bench to prevent damaging the controls at the front September 2015 USER S MANUAL FA GG5B Opening the equipment and any service and repair work must be performed by qualified service personal Repair work should be performed in the presence of a second person trained to ad minister first aid if needed Power supplies do not belong to children hands Symbols related with safety DIRECT CURRENT ON Supply ALTERNATING CURRENT OFF Supply DOUBLE INSULATION Class II protection CAUTION Risk of el