Home
542K, PDF
Contents
1. and z2 be the two stage positions Si and S2 the measured FWHMs Then the correct focus z 9 can be found as zo z1 S1 S1 S2 x z2 z 1 The IRAF script samifocus cl does this calculation assuming that the stage positions are 12 and 20 Just type cl gt samifocus S1 S2 and read the answer If the best focus zo differs from the nominal value by more than 2mm something is wrong You need to examine and possibly adjust the LGS range To change the WFS stage position by 1 mm decrease the range by 30m with LGS near 7 km Once the focusing is done take the in focus sky image in closed loop and measure the FWHM of the star This informs you how well SAM is performing tonight 3 6 Target observation Once all 4 loops are closed and work stably and the focus is verified the astronomer can take science data with SAMI It is almost equivalent to classical imaging with a few caveats The SAM operator monitors the loops and informs the astronomer of relevant events LCH interrupts switch off the laser typically for less than a minute Time remaining until the next interrupt in seconds is displayed in the middle right part of ICS GUI The number turns yellow 60s before the interrupt The science exposure should be stopped or paused the DM and LLT loops should be opened before the laser switches off Otherwise the DM voltages usually go into saturation to fix this just open the AO loop When the laser comes back it is usually s
2. 4 Acquisition of guide stars details Acquisition of guide stars in SAM is one of the most difficult tasks owing to the small capture field of each probe only 3 Accurate determination of telescope pointing through field offset see Sect 3 2 is the key to success 4 1 Field geometry and GMAP software Tel PA EC Features _v FO Figure 6 Sky map displayed by the GMAP software and the geometry of the SAM guide probes Evoke the GMAP software by the button GMAP Catalogs in the GS TAB of ICS A small grey window appears In this window select the Skymap from the File menu to display the stars 4 This display is shown in Fig 6 Click on the TEL and PA buttons to refresh the pointing Select USNO A from the catalogs menu in the upper left You can adjust the scale by the small up down triangles in the upper right corner When the SAM position angle on the sky is 0 North points up and East to the left in GMAP To match the orientation in the SAMI image invert the X axis in the DS9 display The technical patrol field of the SAM guide probes has a size of 100x100mm The X Y coor dinates of the probes in mm are counted from the field center X points to the left to the East at PA 0 Y points up The lower right corner of the patrol field blue dashed line in Fig 6 has coordinates 50 50 the upper left corner has coordinates 50 50 The GP1 patrols the first half of the field negative X GP2 patrols pos
3. Fig 5 Center it on the green cursor with LLT M1 X Y motions arrows in the LLT TAB First use the medium speed then the small speed The arrows are inverted if the LGS is on the right side press the right arrow to center it After each correction command the system becomes non responsive for a few seconds while the LLT M1 motors move Remove the reference mirror The WFS should now see the laser light do not forget to arm the range gate and to power the Pockels cell Close the LLT M3 loop to center the spots Then close the main AO loop At this moment the LLT motors are automatically switched off Wait for at least 10s to verify stable AO loop operation and then authorize the astronomer to take the data 3 5 Focusing in closed loop Focusing in closed loop can be skipped if you are sure of the focus Focus should to be adjusted or checked to compensate for filter thickness in SAMI or when the air temperature changed substantially To check the focus take two sky images with SAMI exposure 10s to 30s in closed loop with the WES stage positions of 12 mm and 20 mm thus bracketing the nominal setting of 16mm You do not need to open the AO loop while moving the WES focus in the WFS TAB The idea is that two intentially defocused images should have equal resolution if the 16 mm stage setting in the middle is correct Display each image in IRAF command mscd and measure FWHM of the same isolated star command mscexam Let z
4. circular motion So if you found the star for GP1 and start acquiring the second star for GP2 do not yet close the tt loop If you guide on two stars the system tries to keep the average error signal near zero When the centering of the probes is not accurate we see that the two fake stars are displaced from the center in opposite directions To correct move one of the probes slightly no need to open the tt loop But never move the GPs during science exposure this will smear the image 4 4 What can go wrong with GS acquisition A failure to acquire guide stars happened in SAM more than once for various reasons Most likely the field offset was wrongly determined because either i the star was not identified correctly between the GMAP and SAMI image or ii the star had wrong coordinates e g it moved away since the catalog was made The correspondence between the coordinate origin X Y 0 0 in the GP plane and pixels in the SAMI is given by the parameters SAMIFCX SAMIFCY in ICS they are written in the icsoft ini file These can change because of temperature flexure or DM tilt It is a good idea to verify the field center before the run by taking the image of a fiducial source in GP1 Sect 5 5 Set the rotator angle at 0 and issue the ROTFLAT command in the OP TAB DM menu Place GP1 at the center by the Center GP command in the OP TAB GP1 menu Turn on the dimmer set intensity to 4 Take SAMI image with 0 2s exposure in the r or V
5. filter Display the image in IRAF and measure the coordinates of the star IRAF reports them in un binned pixels even when binning 12 is effective Enter the new coordinates in the command line of the Maintenance tab in ICS SET SAMIFCX 1891 SET SAMIFCY 2027 The calculated GP positions use the SAM position angle as reported by the TCS The actual angle might differ by as much as 1 which leads to an error of 1mm at the edge of the technical field Note that the angle between the SAMI Y axis and the GP Y axis is only 0 14 The guide star selected from the GMAP USNO catalog may not exist on the sky Try a different star Or better try a different catalog e g load the 2MASS catalog from on line server in the catalog selection menu of GMAP If the scripts FindGP report failure it usually means that there is no star in the vicinity of the GP 5 Calibrations and tests For training and troubleshooting a number of useful procedures can be performed as covered briefly in this Section 5 1 WES bias Open remotely the RTSoft WFS GUI by using the buttons on top of the ICS GUI Select Tools Bias calibration and run the calibration while the WFS receives no light e g when the laser is shuttered or the reference mirror blocks the beam This image is subtracted from every WFS frame so it is important to get it right As the bias is not very stable we recommend to re calibrate it every night 5 2 TurSim SAM has a buil
6. on the SAMdocs page The electronics of AOM is located in boxes attached to it and in the SAM rack SAM uses 3 PC computers and several software modules running on those PCs The ultraviolet laser guide star LGS is created by the 10 W UV laser attached to the SOAR truss inside the laser box Its beam goes towards the Laser Launch Telescope LLT behind the SOAR secondary mirror and is launched in the sky The laser power supply and chiller are located in a rack attached to the SOAR fork The light path inside AOM Fig 2 shows how turbulence is compensated The first active element is the SOAR tertiary mirror M3 that corrects tip tilt like in other SOAR instruments However SAM does not use the standard ISB guider and has its own two guide probes inside the AOM right at the focal plane The light of science target is then re imaged on the CCD by means of two off axis parabolic mirrors OAPs and a deformable mirror DM which corrects defocus and other wavefront deformations After OAP2 the corrected image can be directed either to SAMI or to the visitor instrument by a movable output selector mirror The LGS is created at 7 km so its image in SOAR is formed behind the normal focus in fact close to OAP1 The ultraviolet LGS light is reflected by OAP1 and DM and then is separated from the science beam by a dichroic Special optics collects the LGS light in the wave front sensor WFS to measure how flat is the wavefront and to adjust the DM s
7. preparing for the next LGS acquisition 3 8 At the end of the LGS night Close laser shutter Switch off the laser diodes then the 220V power to the laser in LAS TAB Run the SLEEP script available in the OP TAB SHUTTER and OP TAB AUX menus This script turns off the Pockels cell gating and switches off its high voltage It switches off the high voltage to the GPs and moves them to the park positions The DM voltage is set to zero The ACAM is powered off the live display image disappears The SAM environmental shutter is closed Finally the script energizes the LLT closes its shutter and turns the LLT power off Watch the messages to make sure that the sleep sequence finished successfully otherwise do it manually The SAM operator has to switch off the power to the DM driver the Pockels cell driver and the APD power supply in the soarenv 1 power control GUI Call the LCH to announce the time when the laser operation stopped Do not forget to submit the end of night report 3 9 At the end of SAM run The last night of the SAM run is over Exit ALL software of SAM otherwise the modules stay connected to the TCS increase the communication load and may interfere with operation of other instruments However do not switch off SAM computers Go to soarenv 1 and switch off all SAM hardware except the PXI chassis the chassis runs always Go to the IR platform turn the key in the laser rack off remove it and leave hanging
8. the user of the current system status and performance SAM has four servo loops in total Tip and tilt tt are compensated by the SOAR M3 and off loaded to the telescope mount The LGS spots are maintained centered by the LLT loop And finally turbulence is compensated by the AO loop Opening closing all loops is done by the loop buttons in the middle left part of the ICS GUI 3 Observing Procedure SAM is a complex system It should be prepared for each run switched on and checked by a trained person We assume here that this is done and that SAM is up and running ready to be used at night 3 1 In the evening Targets During the day load the Predictive Avoidance Maintenance PAM file in the LCH soft ware These files are received by the soarpam e mail alias from the Laser Clearing House LCH in response to our requests The target coordinates for the night cannot be changed Yes you can point the telescope anywhere but the laser propagation is only allowed in the direction of PAM targets during authorized PAM time windows The target coordinate list for this night or for the whole run must be also loaded in the SOAR TCS send the list to soarops ctio noao edu Software connections power up Connect by VNC to the ICS verify that the ICS GUI is running and not showing error messages One error message about WFS not reading frames will certainly be displayed Reset the WFS OP TAB WES to fix it Switch on power to the sen
9. will increase and an image of a star will appear in the GP1 window in the upper central part of the ICS GUI This image is fake it only shows conveniently the e and e signals by positioning the fake star in the black square The signals and the flux are displayed To center the star select the GP1 control and use the arrows in the center of GP TAB The medium and low speeds two and one points are most useful Move the probe towards the fake star Each correction takes a few seconds during this time the arrows are greyed out to prevent further commands The tt loop can use only one probe GP1 or GP2 but normally it guides on the average signal of both probes Select one or both probes by toggling Lock GP1 and Lock GP2 buttons among the loop buttons near the SOAR M3 button When you close the tt loop the sliders showing the M3 position start to move What if we are unlucky and the GP1 sees no signal Then press the FindGP1 button This script moves the SOAR M3 on a circular pattern with increasing radius of 1 3 and 5 If the star is close it sweeps over the probe and the signal modulation is detected The probe is moved automatically towards the star and the message of success is displayed If it fails the star is not close to the probe and it makes little sense to look for it by moving the probe around randomly and slowly Note that when we use FindGP1 or FindGP2 the tt loop must not yet be closed to allow M3
10. B DM menu this compensates for the natural DM flexure at current Nasmyth rotator position Use the LGS in open loop but with the closed LLT loop to measure the defocus a4 coefficient is displayed under the WFS image in the upper right part of the ICS GUI If the telescope is well focused the coefficient fluctuates around zero estimated average less than 0 3 microns If the average is not zero adjust the SOAR focus apply offset in ADC TAB If a4 gt 0 apply positive focus offset roughly 50 100 times the a4 value The tt loop must be closed ADC is not normally used with SAM but is available if needed The caveat is that ADC distorts the measurement of the field offset so it should be removed during pointing Once the guide stars are acquired you can move the ADC IN and use AUTO SET command in the OP TAB ADC menu Do not forget to move the ADC OUT for the next target Clouds or strong high altitude turbulence hence poor resolution can make observations with SAM worthless In such cases switch to the backup science program 3 7 Ready for the next target Once the observations of the target are finished the SAM operator prepares for the next target Open all loops the button OPEN ALL LOOPS does this Close the laser shutter in the LGS TAB Switch off the high voltage of GPs and move them out of the field To do this you can use the item DONE TARGET in the OP TAB DM menu This script also moves the reference mirror in the WFS IN
11. SAM User Guide Prepared by A Tokovinin Version 1 Date August 12 2013 File soar commissioning doc user guide guide tex Contents 1 How to use this guide 2 SAM at a glance 3 Observing Procedure Sel RN A A de Sa Aa 3 2 Pomting the targeth jk gece ra a e te a ts 3 3 Guide star acquisition ee 3 4 UGS acquisition s e ene geht Bae e eal Kone alec Bad ee ae ae 3 5 Focusing in closed loop a 3 6 Target observation so sceana a A E ee eee ee 3 7 Ready for the next target 2 a 3 8 At the end of the LGS might 0 02002 002 eee eee 3 9 AttheendofSAMrun 0 0000 000 0 ee 4 Acquisition of guide stars details 4 1 Field geometry and GMAP software ee 4 2 Guide probes and their control 2 ee 4 3 Select and acquire a guide star a 4 4 What can go wrong with GS acquisition 2 2 0 ee 5 Calibrations and tests Bil WS DIAS hus a ae oe Bote See aoe ae ee ad ee Eee Be atlas Cb ee A UPS ik ack Rel SB Bee ot gee hae de ae Ae Gwe Be los ele Be A es De ae Be 633 FlatiDM es aca Belts ek OR Pee eke BE eae Sak Be eee a 5 4 Interaction matrix si 62s Aa eek ehh nodes Se Siete Sa ee wok A 5 5 Imaging GPl with SAMI 0 020 000 00022 eee 6 Troubleshooting 7 Other topics N CW UO MWONNADAH OH w 10 ER E 12 12 13 13 13 13 14 14 14 15 1 How to use this guide This document is written for SAM operator the person who operat
12. e ICS manual to learn more RTSoft ae TT Loop WFS Window Help R 6 a gt 2 8 Guide probes PE A ontro e 4 oe ER ra soa a Fox WS o 3 3 flux 0 00 0 00 6 00 stage Loop 0 00 0 00 Mode Coef um Static um Totalfum butions EEE ES aae ae oas ZemiKe iran 0 116 0 000 0 116 i E romana Messages fri toners eran E LAS 43 6 6 Loop PARSER ap GS miai Bi A diagn Tabs astin cade probes WES TURSIM LLTALASER abces Teemetry Maintenance LCH status H I E 04 0 8 0 24 0 Output gt en TEAN baza TCS selector j a info 0 00 0 00 Scripts R IR 4 o AUX menu Field offset Figure 4 The user interface to SAM ICS GUI The lower left part of the GUI has various tabs with different information and controls You will use 4 tabs OP TAB shown in the figure with a pictogram of SAM GP TAB guide star acquisition WFS TAB WFS and TurSim LAS TAB LLT and laser control and ADC TAB ADC and TCS commands The remaining two tabs are informative or technical We refer to these TABS below in this short hand notation In the OP TAB various menus are available by pointing the mouse on a component and right click The additional AUX menu is available by clicking in the empty right corner of the operational diagram These menus are denoted here as e g OP TAB WFS OP TAB DM etc The remaining part of the ICS GUI is always visible independently of the tabs to inform
13. e means that GP1 is controlled To switch on the HV first press the green button APD High Voltage The upper indicators in the upper central part of the main GUI Fig 4 become green Then toggle the Reset APD switch and the APDs start counting the numbers change With the HV off some constant negative count is displayed When the GPs point to blank sky or even in the dark the count rate is not zero For correct calculation of er ey the background signal must be measured This is done by pointing the sky switching the HV on on both probes and toggling the TAKE BIAS button in the GS TAB Fig 7 Bias calibration takes few seconds after this the displayed counts show fluctuations near zero The bias can change depending on the sky brightness especially when the Moon is up its calibration must then be repeated 11 4 3 Select and acquire a guide star Select a suitable guide star in the GMAP click on it Useful brightness range is between R 187 faintest and R 11 too bright can trigger the over light protection Remember that the magnitudes in GMAP can be wrong by 1 2 Suppose that the star is in the right half of the field Click on the P1 button in GMAP to send the star coordinated to GP1 Go to GS TAB and press the Acquire button of GP1 The probe will be positioned on the star see the messages and note the X Y coordinates of the probe Turn the HV of GP1 on If the probe sees the star the flux
14. ermine the correspondence between SAM pixels and GP coordinates Set the rotator angle at 0 and issue the ROTFLAT command in the OP TAB DM menu Place GP1 at the center by the Center GP command in the OP TAB GP1 menu Turn on the dimmer set intensity to 4 Take SAMI image with 0 2s exposure in the r or V filter If needed adjust the focus of the GP1 z coordinate to get the sharp image If the FWHM of this star is less than 2 5 un binned pixels the SAM optics is tuned well 6 Troubleshooting Tips are given in this Section for solving common problems occurring during SAM operation SAMI gets empty sky image Check the output selector mirror on OP TAB verify that SOAR and ISB are configured for SAM LGS is not seen Check that the laser is powered both 220V and diodes in LAS TAB that both shutters are opened and that the LCH status is green Check that the LLT cover is opened To verify that the laser is actually emitting go to the SLGM GUI and see the signal of the LLT M2 photo diode should be around 2 4 If you do see the LGS with ACAM and can center it but see no spots in the WFS check the Pockels cell settings HV and range gate To test the Pockels cell use the reference beam in the WFS OP TAB WFS menu or TurSim Laser power control does not respond to the commands The control of the laser 220V power is capricious insist by repeating the command Sometimes the status of 220V laser power is reported wrongl
15. es the SOAR Adaptive Module during observations It does not explain how SAM works except for the brief introduction in the next Section 2 because the operator knows this already The operator is also familiar with the Instrument Control Software ICS and knows something about SAM Imager SAMI Technical support of SAM its calibration and preparation for the run are done by someone else Go directly to the observing procedure in Section 3 and execute it step by step For more informa tion follow the links to more detailed explanations here and in other documents Once you are very familiar with SAM just consult the cheat sheet with most frequently used commands and numbers Information about SAM SAM web page denoted as SAM throughout this guide http www ctio noao edu new Telescopes SOAR Instruments SAM SAM document navigator SAMdocs SAM samdocs samdocs html SAM manuals SAM ao_fl_archive html Instrument Control Soft User Manual SAM archive ICsoft UserManual pdf Adaptive Optics tutorial at CTIO http www ctio noao edu atokovin tutorial index html 2 SAM at a glance Instrument control Laser box Laser power Chiller Figure 1 Main components of SAM The main part of SAM Adaptive Optics Module AOM is attached to the optical ISB of SOAR It delivers turbulence corrected images to the built in imager SAMI or to the visitor instrument HRCam see Fig 1 clickable version of this Figure is
16. hape accordingly Figure 3 illustrates major software modules of SAM The RTC and HRCam computers are located in the SOAR computer room the AOM computer is housed in the rack on the SOAR platform Tip tilt SOAR M3 Visitor CCD instrument Imager Figure 2 Schematic diagram of the SAM AOM Software modules communicate between themselves and with the outside world e g with the SOAR TCS via common Internet In addition each computer has its own connections to the SAM hardware by fibers serial links and dedicated Ethernet lines these connections are not shown in the Figure Four different software applications run on the AOM but the user normally interacts only with the top level module Instrument Control Software ICS The low level applications control the AOM and LGS components respectively The LCH application imposes restrictions on the laser propagation All applications are accessed via VNC in separate work spaces and all must be running to operate SAM RTSoft SAMI Soft HRCAM soft RTC HRCam Ethernet TCS Figure 3 Computers and main software modules of SAM The astronomer takes data with SAMI using SAMI soft SIMAGER in coordination with the SAM operator SAMI is also used by the SAM operator to determine pointing corrections for the guide star acquisition Figure 4 shows the main features of the SAM user interface read th
17. itive X The patrol fields of both probes overlap near X 0 The red dashed square shows the science field of SAMI The probes can go inside the science 4GMAP description at SAM archive GS3 UserManual 211110 pdf 10 field in which case they partially block the image however they must not be closer than 30 to the center otherwise they start blocking the LGS Current positions of the probes are marked by yellow lines in the display 4 2 Guide probes and their control Figure 7 The GP tab of the ICS GUL Each guide probe GP consists of a 2x2 array of lenslets coupled by fibers to the avalanche photo diodes APDs which detect the light of a guide star When the star is centered at the lenslet crossing the four signals are equal Normalized differences of the fluxes in two directions e and ey are the error signals that drive the SOAR M3 during closed loop guiding The APDs detect light i e count photons only when they are powered HV switched on Excessive light can permanently damage the APDs There is a hardware protection against over light but you should not rely on it and operate the probes carefully avoiding over light Never switch on the HV during daytime and during telescope slewing The control of GPs is done via the GP TAB Fig 7 Its left and right halves refer to the GP1 and GP2 the controls in the middle apply to one or both probes by selecting the option under the arrows the left pointing yellow triangl
18. ngle Identify a star in the image with a star in the GMAP display preferably close to the center click on the star in GMAP to get its coordinates Then go to the SAMI GUI point the cursor in the GUI and press F3 Point this same star in the DS9 display press c Then go back to GMAP and press FO The newly determined values of field offset pointing offset in the image plane in mm are displayed in the lower right part of ICS GUI Typical values of FO are under 5 mm 3 3 Guide star acquisition Acquire two guide stars and close the tip tilt loop See detailed instructions in Sect 4 The slider display in the middle of the ICS GUI shows that the M3 position is changing in response to the tt control signals Once the tt loop is closed also close the mount loop to off load M3 tilts to the SOAR mount Telescope operator should verify that the TCS is receiving correction pulses from SAM 3 4 LGS acquisition File Windows ifs ean Controls Figure 5 LGS acquisition and centering with ACAM The camera control window on the left opens the image display The un gated LGS image is a fuzzy spot with some rays around it Energize the LLT motors in LGS TAB it takes a few seconds Move the reference beam mirror in the WFS beam in the OP TAB WFS menu or with buttons in the WFS TAB Propagate the laser open shutter in the LGS TAB You will see the fuzzy LGS spot in the ACAM display
19. pen the viewer of the LGS acquisition camera ACAM Click the buttons Acquisition camera and ON in the WFS TAB wait a few seconds The small GUI of the viewer appears connects to the camera and shows its identification Then click on the eye icon to open the live camera display see Fig 5 below Set range to 7km and and gate to 0 1km in the LAS TAB Switch on the Pockels cell HV and arm the gate pulses Also issue the command SpotOffset in the OP TAB AUX menu this introduces a correction in the spot offsets needed for better performance with the LGS 3 2 Pointing the target Point telescope to the science target The LCH status on the right in the ICS GUI normally becomes green laser propagation is authorized and shows the time remaining until the next closure in seconds It is necessary to determine the exact telescope pointing for the guide star GS acquisition see Sect 4 for details Using SAMI take a short exposure 10s to 30s in open loop and without guiding See SAM archive SAMLGS UploadingPA Mfiles v1 docx pdf Cpmputer IPs listed at SAM samdocs computers html 3See instruction for LCH communication in some broad band filter Display this image in IRAF Note that the SAMI X axis points to the East at SAM position angle 0 so to image orientation in DS9 with the GMAP invert the X axis in the DS9 display Menu Zoom Invert X The orientation of both SAMI and GMAP images depends on the SAM position a
20. sitive SAM electronics DM driver Pockels cell driver APD power VNC to soarenv 1 use Op Nasmyth Left tab in the power control GUI It is a good idea to refresh the WFS bias see Sect 5 1 to flatten the DM Sect 5 3 and to take the corrected image of GP1 with SAMI Sect 5 5 Laser If this is the first night of the run insert the key in the laser power supply rack access from the IR platform and turn it ON However this should have beeen done already while the SAM laser was tested for the run About 30 minutes before the start energize the laser 220V ON in the LAS TAB Wait about 1 min before the laser status begins to display then switch the laser diodes on for warming up The lighted sign in the SOAR control room now shows that the laser is working The laser shutters must be closed at this time Call the LCH and announce the time when the laser will be propagated expected start and end propagation time UT Wake up SAM When the SOAR dome is opened and the telescope active optics is being tuned open the LLT cover energize the LLT then open the cover in LAS TAB then turn the LLT power off also open the SAM environmental cover OP TAB SHUTTER You can use the Wake script accessible in the OP TAB SHUTTER menu SOAR must be configured for work with SAM M3 points to the optical Nasmyth M4 in the optical ISB is in the IN position To prepare for the night open the GMAP GUI button GMAP Catalogs in the GS TAB Also o
21. t in simulator of the laser source TurSim It is used for checking the health of the AO system and for internal flattening of the DM Deploy the TurSim arm by the command TurSim In in the OP TAB TurSim menu Set the LGS to 7 km with a wide gate of 0 5km to get more photons in the WFS TAB If the laser is off use the Auto setting in the same TAB to simulate laser sync pulses The star should be visible in the WES If it is not there deploy the reference mirror WFS TAB button Ref BEEAM IN center the source with ACAM and move the mirror out again Center the source roughly by the jog arrows in the WFS TAB trying to get the first two Zernike coefficients as displayed under the WFS spot image under 0 5 in absolute value Note that the jog directions change Close the AO loop and watch how the a4 defocus becomes zero while az and az are not affected You can also simulate turbulence by moving in the Disk 1 and making it rotate with 40 speed by commands in the OP TAB TurSim menu Recording the loop data and WFS spots command REC in OP TAB AUX menu can be done to document the loop performance Do not forget to move the disk out and to stop its rotation 5 3 Flat DM While the AO loop is closed on TurSim without any disk in the beam you can record the flat DM shape Use the command Record as flat in the OP TAB DM menu Always record flat with the 13 rotator angle at 0 otherwise the correction of the DM flexure
22. ufficient to simply close again the LLT loop spots appear in the WFS then the AO loop Sometimes you will need to re acquire the LGS using ACAM Sect 3 4 Recording loop data during normal science operation is always a good idea to document SAM performance and gather statistics To do this go to the OP TAB AUX menu and use the command REC Record loop data on each target at the beginning then every 0 5 hour or so Unexpected malfunction may happen for several reasons Insecure connection to the TCS causes the loss of tt guiding the WFS flux may drop due to cirrus clouds the AO loop may go wild diverge because of this etc All such events require actions on the part of SAM operator to bring the system back to normal operation Section 6 Dither of science images may be required To dither go to the ADC TAB enter the dither value in arcseconds It has be small e g 2 so that the displacements of guide probes do not bring them outside allowed range This risk is higher if one of the GPs is already close to the edge To dither simply press the arrows in the ADC TAB and watch how the tt loop is opened probes and telescope are moved and the loop is closed again automatically The dither takes about 15s during this time no science data can be taken Open loop observations are possible with some precautions You need to tune the DM and focus the telescope for optimum OL performance After pointing the target do ROTFLAT in the OP TA
23. will not be right Also try to record the flat after the DM had been powered and kept flat or working in closed loop for about 30 min or more This is important because the DM has a hysteresis and creep The DM focus depends strongly on temperature so a flat recorded on one day can show defocus on another day Exact calibration of the flat DM is important only for observations in the open loop in closed loop the DM shape is adjusted automatically 5 4 Interaction matrix When the TurSim source is in the beam without disks and centered and the DM is reasonably flat la4 lt 0 2 you can measure the reaction of the WFS to the changing voltage on each DM electrode This is called interaction matriz and used to control the AO loop Open remotely the RTSoft AO Loop control GUI topmost buttons in the ICS select the Inter action matrix calibration from the menu and execute it with 5 V amplitude Watch how the spots move when the electrodes are poked When this operation is finished you need to save the inter action matrix to compute the new reconstructor use threshold of 0 01 to apply and to save the reconstructor All these operations are done from the same AO loop RTSoft GUI Close it when done 5 5 Imaging GP1 with SAMI The guide probes have fiducial light sources that can be projected inside SAM and imaged with SAMI This helps to check the quality of the optics including the flatness of the DM and to det
24. y A motor refuses to move when commanded the message Command rejected on state active appears in the ICS GUI Open the AOM GUI select the motor and issue the Stop command This usually fixes the problem If not try exiting AOM and re starting it again 14 Crazy counts in the APDs are displayed in the ICS GUI even when APDs are not powered This happens when the PXI chassis is not connected to the RTC Power cycle the chassis reboot the RTC start again RTSoft AO loop diverges voltages on the DM go to saturation spots disappear Is the LGS still on or shut down by the LCH Check that the DM driver is powered look in the SAM rack if necessary Check that the LLT loop is closed before the AO loop so that the spots are well centered If the loop diverges still check that it works with TurSim Sec 5 2 Tip tilt loop diverges when the information on the SAM position angle received from the TCS is wrong This happens rarely Try to close the loop again repeat the last science exposure that was spoiled by divergent tt stars with tails Mount offoad does not work the SOAR operator does not see the corrections in the TCS GUI the SOAR M3 sliders slowly drift to saturation The reason is in SOAR TCS its fixes are not documented so far 7 Other topics To be covered here or in separate documents SAM checklist SAM cheat sheet LCH communication instructions Observing Tool 15
Download Pdf Manuals
Related Search
Related Contents
Online Order User Manual Samsung SGH-I560V Εγχειρίδιο χρήσης HOW TO PROCEED WITH TROUBLESHOOTING News Release BPU 3050A Plancha vibratoria reversible 30kN, 50cm/20in A25E 4x4 Product Brochure French 聴撤競明藩 大搬裁断機 - WordPress.com lab7 - Cristinel Ababei Computer Systems DELL Latitude E5430 Copyright © All rights reserved.
Failed to retrieve file