1. PERIMETER Operation Manual Dear Users Thank you very much for choosing EPR 1800 automatic computer perimeter manufactured by us For your security and benefit please read the lt Operation Instruction gt as well as all the datum of the instrument carefully before using it If you do not operate the instrument according to the Operation Instruction we shall not take any responsibility About lt Operation Instruction gt of this Instrument The copyright of the operation instruction belongs to us The content of the operation instruction is written according to the physical goods If you can not understand some of the content or clause or if you meet technical problems when using it please do not hesitate to contact us We have the right of interpreting and revising this operation instruction 10 Content DTP OGUC Oie aa E ahe naan dees ENS 3 INO UC rare E E EA 4 Technical service of Parameters cc eee cee eee ees 5 PSA ALOU a e E OE 7 Function of the Software 0 0 cece ccc e ee cee eee eee ees 19 OPEO SUCDS are E 23 EE ANIC S E EE E E T ET 29 TFET PRC AIO yere a EE REEERE 31 Parameter Introduction eiseres e eee ence eeees 35 DCC lar ANIONS senna ae ah 42 1 Introduction 1 1 Brief Introduction The system of EPR 1800 fully automatic perimeter is a new generation based on the old generation It has two examination resource colors kinds test program and strategy It has the ch
2. False NEG Errors False Negative errors occur when the patient does not respond to a suprathreshold stimulus in an area where the threshold has already been measured The interpretation of false negative errors is not as clear as that of false positive ones because they can be produced by a variety of sources Visual field test results of patients whose false negative errors exceed 33 are not considered reliable 2 False Positive Errors False POS Errors False positive errors occur if patients respond when no stimulus is presented For the purposes of this study we define a false positive response as randomly occurring independent of stimulus presentation and hence independent of any monitored response window The system will count it and if it s over 20 the test report will be treated as unreliable Visual field test results of patients whose false positive errors exceed 33 are not considered reliable 3 Fixation Loss Fixation losses occur when the patient s eye wanders from the fixation target Visual field test results of patients whose fixation losses exceed 20 are not considered reliable 4 Percentage Expressed as Letter P Less than 5 people within 100 people have this kind of visual situation Less than 2 people within 100 people have this kind of visual situation Less than 1 people within 100 people have this kind of visual situation Less than 0 5 people within 100 people have this kind of visual situatio
3. After testing you should save the test result by clicking at and then you can diagnosis Threshold Query o E kd D agnes agnas 2014032700 Let swinya seim 20140327003 Let sakura sokui 20140327004 Let yolendalll 20140327010 Lek agnes agne 20140327012 Let Tate Dewiaton Potem Dewaton T T 5 j4 5 f 2 gt lt 4 T6 5 1 sis J 3 5 6 3 16 o l 14 0 1 D 1 7 4 1 5 p b 0 1 2i 4 1 0 1 2 1 AME o a Si pe mA a NRA 6 onl bi 5 2 Gi t 1 0 2 D 3 ap 22 3 6 4 61 pe 881 to 8 5 2 O D 404 93 6i 74 6 53176 Tf GHT Bordertre MD 156 PSO3 95 Waiting Room Hongo Neame Patent numba yolanda yole 20140327008 Let yolanda 201403270 Let Agnes Jao 20140327013 Let agnes agne 20140327014 Let nes e 20140327015 Let agressa RID kib GPA Analyses Sex Faltale Age 30 Program 242 Syotegy Auto Thrashok Cursor Sge lhis 6 3 3 Click at the choose the test report you need and preview There are Single 3in1 and Overview test report optional 6 3 4 If your patient have come to you and test for several times you should use the Overview report to see the progressing of patient For doing more than one test with a same patient you should click 3 to proceed For GPA analysis at least 3 reports of 24 2 or 30 2 of same client shall be generated before running this function pa 6 3 5 Patient List a Through this section you can delete a patient profile
4. D for patient data security 5 perimeter InstallShield Wizard X Setup Type Choose the setup type that best suits your needs Please select a setup type Complete All program features will be installed Requires the most disk space Custom Choose which program features you want installed and where they will be installed Recommended for advanced users InstallShield o 17 iy perimeter InstallShield Wizard Ready to Install the Program The wizard is ready to begin installation Click Install to begin the installation If you want to review or change any of your installation settings dick Back Click Cancel to exit the wizard InstallShield perimeter InstallShield Wizard TX InstallShield Wizard Completed The InstallShield Wizard has successfully installed perimeter Click Finish to exit the wizard Ji 18 5 Function of the Software 5 1 Main Function of the Software 5 1 1 Visual Checking The main functions of this module are for visual checking statistic analysis of the checking result storing the data and printing 5 2 2 File Management The main functions are file searching report comparison and printing document deleting 5 2 Introduction of the Operation System Click the 2 twice continuously on desktop to start the software it will Launch perimete enter the software directly There are five interface of EPR 1800 software Doy
5. PMID 03202 PIHO LOIS 93570 1315021 3 10 2 Main applications Macular disease retinal disease optic nerve disease advanced glaucoma Test Range 0 10 Test points 68 dots 4 60 4 Main applications Retinal disease glaucoma Test Range 30 60 Test points 60 dots 36 Plas a n 1108 B15 09 84507 BINAE D1s7 bs Sesoes 01564 A F1MG9 PINDI BIDS BLAS 1 BIOS 018H7 B1105 019889 IDe PAOA ns04 aasi 16 831507 Ae F305 9 B1740 1172043017009 D185 56 1188 438 190030 198 658 1 205 0478 210035 216 OS B 288 58 1252058 I BH SE 5 Macula program Main applications Macular disease Test Range 0 5 Test points 16 dots 6 Nasal step program Main applications Glaucoma Test Range 30 50 Test points 14 dots PSTD s pasia ie F320 M3m 37 2 Screening test program 1 C 40 Main applications Common test Test Range 0 30 Test points 40 dots 2 C 76 Main applications Common test glaucoma optic nerve disease Test Range 0 30 Test points 76 dots 38 3 FF 81 Main applications Full field screening test retinal disease glaucoma optic nerve disease Test Range 0 60 Test points 81 dots 4 FF 120 Main applications Full field screening test retinal disease glaucoma optic nerve disease Test Range 0 60 Test points 120 dots 39 5 FF 135 Main applications Full field screening
6. 0degree 3 2 Relative humidity lt 85 3 3 Atmospheric pressure 700hPa 1060hPa 3 4 Power AV 100 240V Frequency 50 60Hz 3 5 Input power lt 400VA 3 4 Characteristics 3 4 1 The instrument belongs to type I BF Model common instrument 3 4 2 The instrument is supplied by single phase net power 3 4 3 The form of the instrument is intermittent working form 4 Installation 4 1 Hardware 4 1 1 Picture of the Whole Instrument 4 1 2 Introduction of Hardware Perimeter stimulus unit Laptop computer with large capability memory and hard disk CD ROM driver USB ports and one combination port Laser jet printer A The Front Picture of the Stimulator Power On Off USB Port A Main Spare Parts Printer Responder Serial Port Wire Power wire Structure Flowchart 4 2 Installation Environment 4 2 1 The instrument must be installed in the flat ground with no slope 4 2 2 The instrument must be installed in the clean quiet and dry room 4 2 3 The instrument must be installed in the dark room where nothing can be seen within one meter 4 2 4 The instrument must be installed with special ground wire 4 3 Hardware Setup 4 3 1 Connect Perimeter and Computer with the new USB Serial Port Wires we provide Power on off Power supply USB connecting port 4 3 2 Connect the Power of the Perimeter Fix the Responder into the connector On Sess con zie e ie Lee G 4 3 3 Take ou
7. 327008 Patient name yolanda yolanda01 Sex FeMale Age fzo Eye Left Device Connected Testing interface 20 Start test 5s Common f n Range l O atop Stop the test Dot Number Strategy Full Threshold Center point detection Off p The initial strategy Age Related ShortWave Detect off B Stimulation Color Stimulation Colord 7 Fixation Mode a Save the test result Stimulation Size B Fixation Loss Monitor Off p Blackground color White M Eye Move Alarm Mode No Action M Setting the parameter gt DET T A Select Program EYI Other E 60 Test another eye 2 aano Diagnosis inputting TA Nasal step gt System HEaintain gt Language Select ani _ Conci Hosptali Name abc Double name Double Name For EU USA and AUS etc market which need the first name and family name inputting Gaze Tracking EyeMove Tick on for activating the gaze tracking function which will generate gaze tracking curve and Auto pupil measurement function pall Strategy Raneri Option ACE J Hsne e dees eter Moe aoe e Recaro Te User defined test interface 5 3 Several important programs and parameters of visual testing First choose a Testing Program threshold test screen test special custom Once the program is selected go to the Parameter to choose a Strategy full threshold auto threshold fast auto threshold accordingly 22 6 Ope
8. Thrashok Terestod Gevone 7 Bids D a gt i 5 BID 5 ae i eS OFA a Bs naman mm ISo nnan Nane Faon unbelEw 7 He SB BNSEN MU saret agre i wW enero eit 2114037700 Led or 75 N sakar cah 2401704 Let n WHIM HS BR PUREE RE Cel Let r nin 5 corms ogra 2TA Lad sanac agrei 37101000 Let Toe Devation r Dewi 4 4 4 l 5 4 j 4 l 1 Zid i 4 1 9 Bs OGT Sovel MOD 56 PED SES We trg Roce Wong Nare PaesisimbelEye sonet agre 3NI Let 3D Image preview eo A Ciear oir Soc Fevsio Age jE Pragan I5 Syypy Ao Theirod Cumar Sine Tle wie raaa Home interface 19 Interface Bar Back to Home interface Back to Examination interface User define test interface Printing System setting and About us B Ol bree Sasa aaa Patient info area Recheck the same eye of same patient Edit Delete Selected patient data All patient list Last record HOR oS a C A A5 a _ Waiting area Register new patient Start check Edit Delete Monitoring Window 5 Save Report QA Parameter al Program Select fon 7 Test another eye n Diagnosis Chin Rest Control Sang s Auto pupil monitoring The system will adjust the chinrest automatically according pa to pupil monitoring Doctors do not need to monitor the whole testing after starting the test Sece ID o o as Gh iy Ve ee a e 5 94 Camere tnumber 201 40
9. angerous Voltage Notes Look through the file Earth wire D gt gt amp 3 Technical Service of the Parameter 3 1 Transport and Storage Prevent the instrument from damp being inverted and being shaken violently Keep it in the room in which the relative humidity is less than or equal to 85 percent the environment temperature is between 5 C 40 C the scope of air pressure 760 hPa 1060 hPa and well ventilated without corrosion gases If the instrument need moving or transporting in a short distance you should take apart all the connection wires and be transported by single If the instrument must be transported in a long distance re pack it into its packaging box and then transport it 3 2 Performance of Perimeter Scanner 3 2 1 Radius of stimul 3 2 2 ator 30cm 3 2 3 Stimulating source of white LED fixing spot is green 3 2 4 Stimulating intensity 0 db 40 db 3 2 5 Intensity of background illumination 31 5asb 3 2 6 Rear projection and the diameter of light spot is Goldman III 3 2 7 The number of stimulating and the stimulating time A 390 spots B Stimulating retention time 0 2s Adaptive C Stimulating spacing interval 0 2s Adaptive Window of eye position tracking White black CCD directly tracking the testing eye 3 2 8 0 90 scope can be tested in one time 3 2 9 The trip of chin rest up down gt 60mm right left gt 30mm 3 3 Working Environment of the Perimeter 3 1 Environment temperature Sdegree 4
10. aracteristics of full function high precision and speed Besides the above characteristics the whole system also has the characteristic of high dependability and steady performance 1 2 Registering Information i Name EPR 1800 perimeter 1 3 Purpose of Instrument This instrument is used for examining the change of visual field which may be hurt by glaucoma visual disease disease of brain surgery and disease of retina 2 Notes 2 1 Caution The local voltage must measure up to the standard voltage required by us If the voltage is not steady please install a steady voltage instrument We shall not take any responsibility for the damage due to the unstable voltage 2 2 Notes To avoid being damaged by the environment Damp Dusty Liquid under the sun and so on the instrument should be putted at the dry place Do not let the liquid or any other small objects run into the instrument otherwise these objects may make the inner parts of the instrument short circuit and even make the users get an electric shock or even cause a fire hazard 2 3 Caution Without the permission of us do not open the box of the instrument or we will not take the consequences 2 4 Notes The instrument can only be installed in the dark room And it can only be operated by those who have been trained by engineers of us or our authorized distributor fai It belongs to BF common instrument according to the degree of protecting voltage D
11. by clicking the icon of 2 and find recall a patient profile by clicking the icon of after you inputting the keywords Query li e agnes agnes 20140327001 Left swinya swiny 20140327003 Left sakura saku 20140327004 Lert yolanda011 20140327010 Left agnes agnes 20140327012 Left 28 7 Maintenance 7 1 Common Problems Treating methods Failed to setup USB device Re set USB device Voltage isn t enough or too Replace power high Can t start system or Memory lost or damaged Insert again or replace the system doesn t work computer memory card Infect virus Use anti virus software to scan virus Crashes while working or Open too many show that the program is wrong application programs re start the computer Hard disk 1s damaged Replace the hard disk Nothing in the screen LCD monitor doesn t display any Properly connect the wire when opening it g switch on power Connection wire is damaged Replace the Connection The responder does wire not work Pressing button is damaged Replace the pressing key The controlling card in the Replace or maintain the stimulator does not work controlling card 7 2 Maintenance 7 2 1 You should firstly turn on the power switch of the monitor and then turn on the power switch of the main frame when opening the machine When closing it you should firstly log out and then turn off the power supply of the monitor and the main frame 7 2 2 Sca
12. e Progam w aina Theesheld Sroegy Theeshokd tnumber an enero Patent name yoands yorida Sex Febtaie Age zH Eye Lett Device 24 6 3 Starting Checking 6 3 1 Before you start the test you should Click the Register Gt icon on the main interface to setup the patient profile and input the info of patient a patient profile will be saved automatically after you clicking Save Patient Infomation Waiting Room Serial number 20140328001 x First Name Doris a8 x Family name Stefan x l Birth 1974 1 10 yolanda yole 20140327009 Left Eye yolanda011 20140327011 Left a T Doctor v Agnes Zhao 20140327013 Left Right Left agnes agne 201 4032701 4 Left i IOP 1 0 CD m2 1 0 ABS 3 1415926 agnes agne 20140327015 Left CDR Ais Patient with myopia hyperopia and astigmatism will need to correct their diopter before starting the testing our perimeter set trial lens holder for this function Click on the icon of and putting patient s real refractive diopter then click on the Auto Calc to get the corrected lens you need to put on perimeter trial lens holder Auto Calc sphere Cylinder Axis Right Left r s For example if we inputting 5 0D under Sphere of Right eye click on Auto Calc the system will generate the corrected lens need to adapt for patient is 1 75D Doctor should test the patient with a 1 75D corrected lens 6 3 2 After save patient data you can find the pati
13. e in partition D otherwise the software will go wrong 4 4 3 After you install the computer system software Insert install disk we provided into CD ROM and copy all the files we provide to your computer s D partition After the installation please Do Keep the disc well 4 4 4 Software installation procedure 4 4 4 1 Install USB capture card driver for computer A If you run Win XP system please install the capture card driver as followed steps 1 Right click at My computer select Manage and enter in Computer Management Comp Open Explore Search Map Network Drive Disconnect Network Drive Create Shortcut Delete Rename Properties 2 Click at Computer Management And select Imaging devices in the right window Computer Management m File Action View Window Help e gt Om em m Computer Management Local p System Tools Event Viewer Shared Folders Local Users and Groups Performance Logs and Alert Device Manager Removable Storage Disk Defragmenter Disk Management vices and Applications DER l8 x SK 46102BSE2CF2 Computer Se Disk drives 3 Display adapters yj Human Interface Devices IDE ATA ATAPI controllers Imaging devices gP USB Device D gt Keyboards I T Mice and other pointing devices l Monitors 1 8 Network adapters 7 Ports COM amp LPT Mi Processors Sound video and game controllers Se Storage volumes z Syst
14. either immediately or in the future Microsoft strongly recommends that you stop this installation now and contact the hardware vendor for software that has passed Windows Logo testing 7 Click Finish to end the installation 13 Hardware Update Wizard Completing the Hardware Update Wizard The wizard has finished installing the software for a UCAP2820 Video Click Finish to close the wizard 8 Image shows like below means the driver installed successfully Computer Management Selec m File Action View Window Help e Am 76 m ALLRA m Computer Management Local 4 SK A6102B5E2CF2 iis System Tools Computer Event Viewer S Disk drives Shared Folders Display adapters Local Users and Groups gj Human Interface Devices Performance Logs and Alert IDE ATA ATAPI controllers a Device Manager 33 Imaging devices S Storage eB UCAP2820 Video Removable Storage 5 gt Keyboards Disk Management B Services and Applications i Monitors B Network adapters Y Ports COM amp LPT SB Processors Sound video and game controllers ge Storage volumes System devices Universal Serial Bus controllers Disk Defragmenter H Mice and other pointing devices F B If you run Win7 system please install the capture card driver as followed steps 1 Open the profile Driver 14 Gad a Windows T 2 Double click at profile DRV to open it and d
15. em devices ae od Hitje eh Universal Serial Bus controllers 3 Right click at USB Driver and Click at Update Driver as followed pic 11 g Computer Management m File Action Yiew Window Help e AMTS 2TA RA m Computer Management Local SK A6102B5E2CF2 ih System Tools Computer E fj Event Viewer lt Disk drives Shared Folders Display adapters Local Users and Groups 3 Human Interface Devices Performance Logs and Alert IDE ATA ATAPI controllers Device Manager S Imaging devices yy Storage D EIA g Removable Storage te Keyboarc Disk Defragmenter its Mice and Disable Disk Management E Monitors Uninstall E Services and Applications B Network Jy Ports CC SB Processot Properties Sound vid S Storage volumes System devices Universal Serial Bus controllers Scan for hardware changes C Computer Management 3 File Action View Window Help e fS EMA RA m Computer Management Local ie SK A6102BSE2CF2 z i Spm Hardware Update Wizard Event Viewer E3 Shared Folders wz res Users and Groups 7 Welcome to the Hardware Update Egy Performance Logs and Alert nS Wizard SS ay Storage Windows will search for current and updated software by looking on your computer on the hardware installation CD or on Removable Storage the Windows Update Web site with your permission Disk Defragmenter Read our privacy p
16. ent record at Waiting Room 25 Waiting Room moO yolanda yole 20140327009 Left yolandaQ11 20140327011 Agnes Zhao an 40327013 Left agnes agne 20140327014 Left agnes agnes 20140327015 Left Click on the patient colume you wish to enter and click on enter the testing interfaceThe default testing program is 30 2 and strategy is auto threshold if you want to change to other program click at the or pica cat select the program you want to use as following image A Select Program Threshold Screen Speciel Custom and choose as below Param Setting Laz Common Range Dot Number Strategy Full Threshold Center point detection off The initial strategy Short Wave Detect Off gt Stimulation Color Fast Auto Threshold Fixation Mode Center Point x Stimulation Size Fixation Loss Monitor Off x Blackground color white l Eye Move Alarm Mode No Action x OK cnoi Notes The program and strategy will automatically resume to default value once currently testing finished System will not memory your last program and strategy setting Operator need to proceed this process if default value needed to be 26 changed 6 3 3 Then click to start testing During the testing doctor can use the function of Auto Puil monitoring so that do not need to monitor the whole process personally The system will adjust the chinrest the headrest automatically 6 3 4
17. hin the normative database e General reduction of sensitivity Visual Fields VF are described to have generalized reduction of sensitivity when both conditions for outside normal limits are not met and the best region of the VF is depressed to a level at the 99 5th percentile within individuals of the normative database e Abnormally high sensitivity The GHT is described as having abnormally high sensitivity when the overall sensitivity in the affected region of the VF is better than 99 5 of individuals within the normative population e Within normal limits VFs are described as being within normal limits when none of the above conditions are met 9 Bebie Curves 33 Bebie curves Bebie or cumulative defect curves are useful in detecting diffuse depression of the visual field The curve is a graphic ranking of the defect the difference between the measured threshold and the age corrected normal threshold for each point in the visual field The x axis represents the rank of the defect from smallest left side to largest right side The y axis represents the magnitude of the defect 10 Mean Defects MD The average of the examination value of all spots minus normal value it shows the condition of the patient s vision sensitivity comparing with those of the same age 11 Short Floating SF It shows the light sensitivity deviation appearing in one perimeter examination process it shows the reaction consiste
18. matically This will guarantee the test accuracy Do not treat it as an system error and do not enter into testing procedure while self checking in processing the system will not run correctly Enter the testing only after self checking finished 6 2 2 Input the patient s information 6 2 3 Operator should inform patient the examination process before the examination 1 The patient should be familiar with the responder demonstrate him the right way of responding the way of clicking the responder and make him relax as soon as possible 2 Demonstration him the fixing sight 3 Cover the patient s eye which will not be tested with an eye shroud 4 Put the patient s jaw on the right side of the chin rest when examining the left eye put the patient s jaw on the left of the chin rest when examining the right eye If the patient testing Esterman binocular program patient can either put their chin on left or right chinrest but doctor need to adjust the cross in monitoring window aimed at center of patient s nose bridge 5 Put his forehead on the ribbon of the forehead bracket slightly 6 The eyeball shall stare at the center fixation light and cannot move only glance to perceive the stimulus light and respond to it by pressing the responder 7 eto oes Pi 7 _ _ CompletadCount 75 20 t CS False Pos Count fon FalseNeg Cout a0 Fwaton Loss Rae 0 o ooo Check Time jo Pupi Damate B2 Cursor Size iw
19. n 5 Total deviation The difference between a patient s threshold sensitivity and the age corrected normal sensitivity from the perimeter s internal normative database at each tested location of the visual field 6 Pattern deviation PD 32 The localized loss at each tested point after the removal of the effects of any generalized loss pattern deviation decibel dB values are the total deviation values minus the general value 7 Pattern standard deviation PSD It is a measurement of the degree which the shape of the patient s measured field or hill of vision departs from the NORMAL age corrected reference field model The value is expressed in decibels and any value of 2dB or greater will have a P value next to it indicating the significance of the deviation 8 Glaucoma Hemifield Test GHT It is for automated evaluation of single static threshold visual field test results in glaucoma It 1s also constructed to detect field loss that is symmetric around the horizontal meridian e Outside normal limits The GHT is described as outside normal limits when differences between a matched pair of corresponding zones exceeds the difference found in 99 of the normal population or when both members of a pair of zones are more abnormal than 99 5 of the individuals with the normative population e Borderline The GHT is described as borderline when matched pairs of zones are abnormal at the 97th percentile wit
20. n disk and arrange pieces in a certain period 7 2 3 Keep air clean dry use air conditioner if possible 7 2 4 If the instrument has not been used for a long time you should supply power for the main frame at intervals Usually three times a week four hours one time 7 2 5 If there 1s something wrong with the instrument please contact us immediately 29 or ask the special maintainers to maintain Notes You should clean the chin rest before and after each time you use it 7 3 Fusing Parts Model of fuse 5 XX 20mm T2A L250V Replace the fuse Screw off the cover of fuse replace it with a new good fuse then cover again see the picture of the perimeter 30 S Interpretation of the report Single zhongxian Name dd 3 o DOB 1989 01 01 1D 20140721011 Eye Left Program Select 24 2 Be E ain Fixation Monitor On Stimulus Ill White Pupil Diameter 5 7 Date 2014 07 21 Fixation Target Center Background 31 5 ASB Visual Acuity Time 16 26 50 Fixation Losses 0 7 Strategy Auto Threshold RX DS DCX Age 25 False Pos Errors 0 5 False Neg Errors 0 4 Test_Duration 03 17 Fovea Off 30 GHT Within Normal Limits 2 411 9 2 Zits 2 8 0 1 1 27 3 0 1 2 MD 1 08 2 1 0 0 0 4122 PSD 1 59 1 4 3 2 2 1 2 1 3 0 310 0 1 2 2 4 214 2 4 1 21 1 Total Deviation Pattern Deviation lt 5 lt 2 BB lt 1 E lt 0 5 Diagnosis Doctor Version V2 0 31 1 False Negative Errors
21. ncy in the course of examining The bigger the value is the worse the cooperation the patient shows The short wave will become height in the scope of abnormal visual field When more scope in the visual field become abnormal or the degree of abnormal increases the whole wave will become higher 12 Decibel dB In perimetry the intensity of a stimulus expressed as 0 1 log unit of attenuation of the maximal available stimulus the higher the dB the dimmer the stimulus intensity 13 Fixation monitoring Assessing the ability of the patient to maintain gaze by the experimenter s observation 34 9 Introduction of test parameter 1 Threshold test program 1 30 2 recommend Main applications Common test glaucoma optic nerve disease retinal disease x Test Range 0 30 Dee TOS Test points 76 dots oR i Ao 817221 1s bB 0188 621 2 24 2 Main applications Common test glaucoma optic nerve disease Test Range 0 24 poe Test points 54 dots aa BIRON PIEDI P1620 18021 191 015 saog 35 Aa 110015 PRLS oly 1299 015 82 21 ihe 28 015 gO 550 wis 94500 agon IAR Iaoa vied BILO1S BRP BED BETES RO 1g 831502 SASO raa egaa aada Sag Igos imz Iag M ELI M18 10a B9015 Aog TT 1 ee POR O 8 72hF 845013 beam psrag B45 021 pes sen SOLS PBD PEDA PRLS D4 PA209 025901 828 DSN Dg t B262 02 P 278 0210 23423
22. olicy Disk Management gp Services and Applications Can Windows connect to Windows Update to search for software Yes this time only Yes now and every Mame nec a device No not this time Click Next to continue 5 A window Hardware Update Wizard will pop up select Install the software automatically Recommended then click Next 12 C Computer Management File Action View Window Help e f6 828 amp 3 28 4 m Computer Management Local ge SK A6102BSE2CF2 rs System Tools R Event Viewer Hardware Update Wizard Shared Folders Local Users and Groups Performance Logs and Alert gt Device Manager SS This wizard helps you install software for Removable Storage Disk Defragmenter USB Device Disk Management Ta Services and Applications a If your hardware came with an installation CD lt 4 or floppy disk insert it now What do you want the wizard to do Install the software automatically Recommended D Install from a list or specific location Advanced Click Next to continue 6 A window pop up like bellow select Continue Anyway Hardware Installation Ad The software you are installing for this hardware UCAP2820 Video has not passed Windows Logo testing to verify its compatibility with Windows XP Tell me why this testing is important Continuing your installation of this software may impair or destabilize the correct operation of your system
23. ouble click at t to start installation USB Video Audio Device Driver InstallShield Wizard V W E i a em Sm Welcome to the InstallShield Wizard for USB Video Audio Device Driver The InstallShield Wizard will install USB Video Audio Device Driver on your computer To continue click Next 3 Click Next to continue USB Video Audio Device Driver InstallShield Wizard InstallShield Wizard Complete Setup has finished installing USB Video Audio Device Driver on your computer Cancel 15 4 Click Finish icon restart the computer A 4 4 4 2 Open the file and double click on icon 5 to start the installation process InstallShield Wizard Preparing to Install perimeter Setup is preparing the InstallShield Wizard which will guide you through the program setup process Please wait Extracting perimeter msi eee perimeter InstallShield Wizard Ea Welcome to the InstallShield Wizard for perimeter The InstallShield R Wizard will install perimeter on your computer To continue click Next WARNING This program is protected by copyright law and international treaties cack Newt 16 ie perimeter InstallShield Wizard X Destination Folder Click Next to install to this folder or dick Change to install to a A Install perimeter to cient anoe InstallShield lt Back Next gt Canel Relocate the file path recommend Disc
24. ration Steps Operation Steps The patient s preparation The doctor s Preparation Patient s examination Treating after examination Print report 6 1 Patient s Preparation 6 1 1 Perimeter examination should be done in the dark room or the result is not precise 6 1 2 Before examination the patient should be relaxed and know the whole examination process 6 1 3 Automatic perimeter will adjust itself and control the testing stimulus light automatically according to the operator s requirement Through the parameters setup Note As most of the ophthalmic patients are old men whose comprehension may be not very good we should tell them the whole examining process carefully before the examination so that they will not be nervous and in a state of a totally relaxation during the process of examination Since the patient should stare at the fixing spot for a long time and another hand will also became nervous they will be likely to feel fatigue If this happened the patient would not see things clearly which may influence the examination result So during the examination process let the patient relax and tell them the right reaction This is very important for a successful examination 6 2 Doctor s Preparation 23 6 2 1 Power on the perimeter and Enter into the perimeter system as the way in 5 2 Notes After powering on the perimeter it will take 3 4 minutes for a self checking by running the stimulus lights auto
25. t the fixed foa inside the Perimeter r ia 4 4 Software Setup 4 4 1 Windows System Requirements English Simplified Chinese Version Window XP 32 bit Professional English Simplified Chinese Version Windows 7 32bit 64bit Professional English Simplified Chinese Version Windows 7 32bit 64bit Ultimate or English Simplified Chinese Version Windows 8 32bit 64bit Professional PS Our perimeter software only support English Simplified Chinese Version Windows system if user running others language Windows system there will be unreadable code showed during install procedure and on software or other unpredictable error will be occur Please use relevant system we recommended 4 4 2 Hardware Requirements Before you install PERIMETER V 2 0 make sure your computer meets the following minimum requirements CPU Mainboard Intel chipset Processor gt 1 7Ghz multicore Dual core 4 threaded Memory minimum gt 2GB Hard disk Rotational Speed gt 7200 RPM Solid state drives without this parameter Caching gt 2MB Space gt 500GB Display Supporting 1440 900 resolution or greater 1440 900 recommended Perimeter software show normally on resolution of 1440 900 1600 900 1920 1080 USB 2 0 Port A Make sure there are at least 2 partition C and D existing in the hard disk Otherwise the software will run error Software will go wrong A Make sure there is at least 5GB free space for data storag
26. test Test Range 0 55 Temporal side 87 Test points 135 dots 3 Specialty test program 1 Esterman monocular Main applications Driver test for single eye Test Range 0 60 Temporal side 75 Test points 100 dots 40 2 Esterman binocular Main applications Driver test for double eyes Test Range Double temporal side 150 Test points 120 dots 3 Superior 36 Main applications Screening upper visual field Main applications upper visual field 60 Test points 36 dots 41 10 Declarations We can provide you with the information of those parts need maintained 1 We will provide maintenance and enquiry free for one life 2 We will maintain the machine for free for one year since the date of purchasing if the machine is operated according to the operation instruction 3 During the maintenance we will charge fee for the maintenance under the following conditions e Do not use maintain store the instruments according to operation instruction Take apart or amend the instruments without the permission of us which cause damage Damages is caused by accidents use wrongly or caused by other major nature factors e Please forgive us for not informing you 1f the design or the assigned type changes 42 ms PERIMETER Ef m N 9990 NW 14 Street Suite 105 Doral FL 33172 Telephone 888 334 4640