DIGORA® Optime DXR-50 001 User`s Manual
Contents
1. 0 ccc cec cee eeceeeeeeseeeeeeeeeaeeseeseeeaeeaes 4 2 1 Main parts and controls cece ccc cec cece eeceece eee eceeeseeseeceeseeceeeeeeeeegeeteeseeneees 4 ZZ IRCCS SS Ol le Sai aca cnens yaaceaie a son eatca cane arane aire sneeeie eae oan cece ie aot gaaleaatecesece 5 2 2 Display symbols and what they mean ccccceecsecceeceeeeeeseeeeeeseeeaeeseeeaeees T 2 9 USNO ME SYSTEM aene E a T O ands 10 Preparing MNeSySlE NM eorr r a ATEN ARAE AN 10 Preparing an IP fOr ExpOsSUTe aeaaeai o aa a EA 11 Magno DIALS Nolder S aoaaa A IEE 13 TAKING AN eXPOSU C aaa Goadvedan etiam ats 14 REJANG AMMANI Pilale O 16 Single USEF configurato esnea eaaa a ara a 16 M lticonnectconiguratioN snina E EA 19 Removing IPs from the plate collector o nonnononnennnnonnnnnnnsnennsnrerrnrenrernn 20 StoNdDyMOdG sioe a scatebanwetaiue 20 FREUEVGS 1ASE IMACS inanan E A 21 Snu COW MENT asn a 22 IP erasing mode Initial erasing of the IPS cc ecceccceeeceeeeeeeeeeeseeeaes 22 EROT eee er Ce eee ae OR Re ree UTE AOD RRR ee TEE TER DT eee re ane enn eee RoE 23 User s manual 204253 II Imaging plate unit I SCUD ODHON S aaa 24 S21 LOOUMMUIMD CMG an a a atetG ciel lateraacaaaes 24 Selecting the tooth numbering feature c ccc ceecceecceeeeeeeeeeeeeeseeseeees 24 Using the tooth numbering feature ccccccceccceeeceeeeeeeae esse eeaeeeeeesaeeaes 24 9 Z TRO OOIMMON orais tcacacntseg eaten a dtd O 25 39 Mag
2. Direct connection method uses the unit s n 1 After positioning the unit connect it to the PC s in the local area network using the Ethernet cable s not supplied 2 Switch the unit on The dental imaging software animation will appear This indicated that the unit is not communicating with the PC s in the network 3 PC Install the dental imaging software to be used in the PC s 4 PC Open the dental imaging software and select the scanner setup window CANnel Settings Driver Info Status Scanner Oleconmected Y ETZION Serial No B Installation and setup Imaging plate unit 5 PC From the scanner setup window select the Settings tab to open the Scanner Connection page Scanner settings Driver Info Scanner Connection Direct Connection settings Scanner senal number Computer network connection IntelF PRO1000 MT Network Connection C Direct Conmectior 6 PC Select Direct Connection Scanner Settings Driver Info Scanner Connection Direct Connection settings Scanner seral number g og00007 Computer network connection Intel R PAO TOO0 MT Network Connection f Direct Connectior Key the serial number of the unit into the Scanner serial number field The serial number of the unit will appear on the unit display when the unit is switched on It can also be found on the type label on the back of the unit Make sure that the Com
3. _ Food grade polyethylene Latex free A 2 User s manual 204253 A Technical Specifications Imaging plate unit A 3 Main dimensions User s manual 204253 A 3 Imaging plate unit A Technical Specifications Guidance and manufacturer s declaration electromagnetic emissions The DXR 50 is intended for use in the electromagnetic environment specified below The customer or the user of the DXR 50 should assure that it is used in such an environment Emissions test RF emissions CISPR 11 Compliance Electromagnetic environment guidance The DXR 50 uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Harmonic emissions IEC 61000 3 2 Class A Voltage fluctuations flicker emissions IEC 61000 3 3 Complies The DXR 50 is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes User s manual 204253 A Technical Specifications Imaging plate unit Guidance and manufacturer s declaration electromagnetic immunity The DXR 50 is intended for use in the electromagnetic environment specified below The customer or the user of the DXR 50 should assure that it is used in such an environment Immunit
4. lf there is an untransferred image in the unit s memory the unit cannot be shut down The image must be transferred first Refer to section Retrieve last image IP erasing mode Initial erasing of the IPs If an IP is being used for the very first time or has not been used for 24 hours and has not been exposed to ambient light the erasing procedure must be carried out before the IP is used The erasing procedure removes any fogging due to background radiation The dedicated erasing mode erases the IP but does not send the image blank to the dental imaging software Press and hold down the Start key for several seconds until the erase mode animation appears on the unit display The unit is now in the erase mode Hold the IP by its edges or use a protective cover and position it so that the side with the metal disk the back is on the right Insert the IP into the unit door remove the protective cover if used Amagnet will hold the IP in position and then automatically slide it into the unit User s manual 204253 2 Intraoral imaging plate unit Errors A Error User s manual 204253 Imaging plate unit NOTE It may take longer to erase IPs using the erasing mode than the normal read and erase mode This is to ensure that IPs that have not been used recently are erased properly After the IP has been erased it will be ejected from the unit The IP can now be used to take an exposure To ex
5. 2 Intraoral imaging plate unit Imaging plate unit If the exposure time is too short images will be noisy Such images may still usable for somediagnostic purposes If the exposure time is too long images will be too dark or will show patient movement These images will not be good enough for accurate diagnostic examination The exposure values table below should only be used as a guide Recommended exposure values s for DC x ray units 60 kV 7 mA 70 kV 7 mA Adult Child Adult Child Adult Child Adult Child Bitewing Maxillary incisor Maxillary cuspid Maxillary molar Occlusal Mandibular incisor 0 200 Mandibular cuspid 0 250 Mandibular molar 0 250 For AC x ray units increase the exposure times by about 30 3 Protect yourself from radiation and take the exposure 4 After the exposure remove the sealed hygiene bag from the patient s mouth CAUTION If there is a risk of cross contamination wash disinfect and dry the hygiene bag before opening it User s manual 204253 15 Imaging plate unit 2 Intraoral imaging plate unit Reading an imaging plate Single user configuration NOTE If the unit is operated in a multiconnect configuration please refer to Multiconnect configuration first to reserve the scanner 1 i S r qi ae 16 Pull the flap to open the hygiene bag Keep the IP in the protective cover so that you do not touch the IP or allow it to be exposed to
6. ambient light and then slide the protective cover and IP out of the hygiene bag Hold the protective cover and IP so that the white side of the protective cover is on the right The metal disk on the IP will also be on the right Insert the protective cover and IP into the unit door A magnet will hold the IP in the correct position Slide the protective cover off of the IP and leave the IP in position half way in the unit door User s manual 204253 2 Intraoral imaging plate unit N Progress animation User s manual 204253 Imaging plate unit As soon as the protective cover is removed the unit detects that the IP is in the unit door and will switch from the standby state to the ready state ready to use and automatically slide the IP into the unit CAUTION If the IP does not slide into the unit after the protective cover has been removed the IP is misaligned or has been placed in the unit the wrong way round Reposition the IP and insert it into the unit again If after repositing the IP it still does not slide into the unit press the Start key to manually switch the unit from the standby to the ready state and the IP will then slide into the unit when the protective cover is removed The busy animation will appear on the display which indicates that the imaging plate is being read After few seconds a preview image will appear on the unit display PC A read out progress window will appear o
7. ethanol 70 96 EtOH 30 4 DI WATER or ethanol anhydride Wipe the IP as described above and then wipe dry NEVER USE abrasive chemicals or materials to clean IPs Unsuitable cleaning solutions methods may damage or destroy the IPs or leave residues on the sensitive surface that may appear on the images 2 Imaging plate unit 4 4 Storage 4 5 Replacement 4 5 Disposal 28 4 Handling and care of imaging plates Unpacked exposed to ambient light in the dedicated storage box Below 33 C 80 RH and shielded from X rays and ultraviolet radiation If an IP is stored for over 24 hours in a hygiene bag or in a location that is shielded from ambient light the IP must be erased to remove any potential fogging before being used to take an exposure Replace an IP if the image shows marks spots or dots which still appear on images even after the IP has been properly cleaned the IP is mechanically damaged scratched or badly bent IPs must be disposed of in accordance with all local national and international regulations regarding the disposal of non environmentally friendly or hazardous materials Phosphor substance under the top coat on the sensitive side of IPs must not be swallowed User s manual 204253 5 Unit care and maintenance Imaging plate unit 5 Unit care and maintenance WARNING Switch the unit off and disconnect it from the main power supply before cleaning or disinfecting t
8. The complete system comprises the following the DIGORA Optime digital imaging plate unit the unit SOREDEX dental Imaging Plates IPs protective covers hygiene bags and other related imaging plate accessories APC not supplied in which suitable dental imaging software Alocal area network LAN cable not supplied will be required if the system is to be used ina network The unit is a laser scanning device designed to automatically read reusable SOREDEX dental intraoral imaging plate sizes 0 1 2 and 3 After reading the images can be viewed on the PC using the dental imaging sofware The unit can either be connection directly to the PC or to network via the LAN cable The unit can be set up to work with a single PC the single user configuration or with several PCs the multiconnect multiple user configuration With the multiconnect configuration up to eight PCs can be used one ata time with the unit This manual describes how to use both configurations Imaging plate unit Introduction NOTES Only personnel trained and authorized by the manufacturer of the unit are allowed to install and configure the unit Only use the imaging plates protective covers and hygiene bags supplied by the manufacturer of the unit Please read the section 7 Warnings and precations before using the unit 1 2 System installation Positioning the unit Do not position the unit in direct sunlight or near b
9. have now reserved the unit and it can now be used to read an image NOTE If the multiconnect icon is YELLOW it indicates that the unit has been reserved by someone else Wait until the unit is free If you wish to know who has reserved the unit place the cursor on the multiconnect icon while yellow and the message Scanner reserved by XXXX where XXXX is the multiconnect ID of the PC that has reserved the unit will appear NOTE If the multiconnect icon is GRAY it indicates that the unit has not been switched on or is just starting up and is not yet ready for use Switch the unit on and or wait until the unit is ready for use 19 Imaging plate unit 2 Intraoral imaging plate unit 3 PC To release the unit click the GREEN multiconnect icon and then click the Free the scanner unit button Free the scanner Removing IPs from the plate collector When removing IPs from the plate collector hold them by their edges Alternatively pull the plate collector out of the unit and tip the IPs out onto a flat clean surface If they are not to be reused immediately or within a short period of time store them in their storage box Standby mode 20 If no IP is inserted into the unit within a certain period of time the unit will beep several times and the status light will start to flash When it turns yellow the unitis in the standby mode The unit will automatically exit the standby mode as soon as a ne
10. not active The Ethernet connection consists of correctly installed cabling and any switches and or hubs used basically the complete Ethernet link lf the link is active it does not necessarily mean that the unit is physically connected to PC It only means that the unit is physically connected to something that is Ethernet compatible hub switch another PC etc CAUSE SOLUTION First remove and then reconnect the Ethernet cable between the unit and PC to see if this solves the problem i Physical connection is okay link active but the the Dental imaging software animation flashing yellow appears on the unit display The unit is not configured correctly Reconfigure Imaging plate unit oe B Installation and setup Cabling not correct link not active and the Unit connection animation appears on the unit display The unit is not physically connected to the PC Connect the unit to the PC The PCs Ethernet connector s is faulty There is usually a green link LED near the PCs Ethernet connector s on the NIC If the LED is not on replace the NIC PROBLEM The unit display is not active but the power on off status light is on green or yellow CAUSE SOLUTION Green LED defective display or its cabling Replace Yellow LED unit is in standby energy save mode
11. survey a should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol R NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicated theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the DXR 50 is used exceeds the applicable RF compliance level above the DXR 50 should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as reorienting of relocating the DXR 50 gt Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3 V m A 6 User s manual 204253 A Technical Specifications Imaging plate unit Recommended separation distances between portable and mobile RF communications equipment and the D XR 50 The DXR 50 is intended for use in an electromagnetic environment in which radiated RF disturbanc
12. DIGORA Optime DXR 50 001 Digital intraoral imaging plate system User s Manual Document no 204253 rev 2 0908 DIGORA Optime DXR 50 001 Digital intraoral imaging plate system User s Manual 0537 This unit is marked according to the Medical Device Directive 93 42 EEC if the unit contains the CE mark Document number 204253 rev 2 0908 Original approved English language version Manufactured by SOREDEX P O BOX 148 FIN 04301 TUUSULA FINLAND Tel 358 0 10 270 2000 Fax 3589701 5261 Imaging plate unit Soredex endeavours to produce product documentation that is accurate and up to date However our policy of continual product development may result in changes to products that are not reflected in the product documentation There fore this document should not be regarded as an infallible guide to current product specifications Soredex maintains the right to make changes and alterations without prior notice II User s manual 204253 Imaging plate unit Contents Tz THE DIGORACG ODUM 6 veccosscascussmeusnaisamiddnenstaaearencsadtannsyaranndiaceknoanss 1 TeV OCU CUO srren a A E OE EE 1 k2 SY SUSI MVS UA AUN OU a tate ah ral dakar autadiaddneca tained 2 POSIIONIMNG TIS UNIE tesntanaceeeentveceseoatasasasstarcnssdeoedeoiiaut AOTAN 2 OSIMOMMAG ING ir Onna oi fad aaa eee aes sadeletese a iad ae uae aaa 3 JNEC S ee eee i enn Pen Cen PRE oR enone ene Reena eer Pees 3 2 Intraoral imaging plate unit
13. Protective cover J Remove disconnect Insert connect Busy Unit in operation Progress animation Unit in erasing mode Check Something wrong or take alternative action A 8 User s manual 204253 2 Intraoral imaging plate unit Imaging plate unit Dental imaging software software not open not ready or waiting for user action Unit connection Not connected or connection not working Rotate Error state and error number Error A Check documentation supplied with the unit Unit in service mode Service technicians only NY 2 User s manual 204253 9 Imaging plate unit 2 Intraoral imaging plate unit 2 3 Using the system For optimum performance only use IPs protective covers and hygiene bags designed for this unit and supplied by authorized distributors The manufacturer of this unit will not be held responsible for any problems caused by using accessories from other manufacturers Proper handling cleaning and storage of the IPs ensures the best image quality and maximum service life of the IPs Refer to section 4 Handling and care of imaging plates Preparing the system 1 PC Switch on the PC connected to the unit 2 PC Open the dental imaging software and a new or existing patient card where you wish to store the intraoral images NOTE If you are using the system for the first time you may wish to check and or change the setup options See section 3 S
14. Start key I1 1 r Unit door and plate carrier Imaging plate collector Rear of unit Power supply connector for the power supply unit PSU A N CAUTION Only use the PSU supplied with the unit or an approved spare PSU supplied by an authorized distributor S Refer to technical specifications RJ 45 connector for Ethernet cable direct connection to PC network LAN connection Ethernet cable not included Refer to technical specifications PC not included with digital imaging software that conforms to the MDD 4 User s manual 204253 2 Intraoral imaging plate unit Imaging plate unit 2 2 Accessories For additional information about the accessories listed below contact your authorized dealer Not all accessories are available for all units Imaging plates IPs Equivalent to film sizes 0 1 2 and 3 IDOT version An identification code IDOT is printed i onthe IP and will appear on the intraoral image The IDOT identification mark allows the IP used for an exposure to be easily identified and removed if it damaged Standard STD version The standard version has no identification mark Protective covers For IPs 0 1 2 and 3 gt l Hygiene bags For IPs 0 1 2 and 3 User s manual 204253 5 Imaging plate unit 2 Intraoral imaging plate unit CAUTION For optimum performance only use IPs protective covers and hygiene bags supplie
15. Startup animation will appear on the unit display PC When the Startup animation clears click OK to connect the PC to the unit Now connect the other PCs in the network to the unit Just enter the IP address into the IP field and then click OK to connect the PC to the unit it is not necessary to hold down the Start key and click the Send to Scanner button with the other PCs once the unit has already got ans IP address B Installation and setup Imaging plate unit 3 Troubleshooting PROBLEM The unit does not come on The units power on off status light and display are off CAUSE SOLUTION The main power supply to the unit is off or the unit is not switched on i Ifthe LED onthe PSU is not on it is not receiving power from the mains Switch the mains power on li Ifthe LED on the PSU is on switch the unit on iii If status light and display still do not come on check the unit s membrane control panel and cabling Replace if faulty PROBLEM Animation displaying either Unit connection or Dental imaging software flashing yellow appear on the unit display and the unit will not work CAUSE SOLUTION e Defective RJ45 cable Replace Imaging plate unit B Installation and setup PROBLEM The IP connection between the unit and the PC does not work CAUSE SOLUTION Check that the Direct connection method was configured correctly If all the setting are correct but the connection still does not work
16. d by the manufacturer of the unit or the manufacturer s authorized distributors The manufacturer will not be held responsible for problems caused by using accessories from other manufacturers IP storage box For storing IPs safely and conveniently Microfiber cloth For cleaning IPs IP holders For bitewing periapical and endodontic exposures optional not included See section Imaging plate holders for more information User s manual 204253 2 Intraoral imaging plate unit Imaging plate unit 2 2 Display symbols and what they mean During use symbols and animations will appear on the unit display These indicate the status of the unit help you to operate the unit correctly show user mistakes and corrective actions display error codes display a preview image The main symbols are Startup During startup the unit serial number IP address and other information will appear on the unit display OBABAO SJ09 CORE ver xx Multiconnect wait Multiconnect configuration The unit is not reserved by any PC in the system Multiconnect reservation Multiconnect configuration The unit has been reserved by a PC e g PC number 2 Unit door Protective cover and imaging plate Yellow remove protective cover User s manual 204253 7 Imaging plate unit 2 Intraoral imaging plate unit i an s Imaging plate a Yellow wrong way round rotate ee ac l l l l l l
17. e installation by starting image capture using the imaging software If the Use Multi Connect was selected the Workstation identifier of the PC 1 8 being used will appear on the unit display IP method using the unit IP address If your system does not allow the direct connection method to be used to connect the PC s connection can be done using an IP address 1 Follow steps 1 to 5 from the previous section Direct connection method uses the unit s n 2 PC From the the Settings tab select IP based and then select the Enable changing IP address box Scanner Connection Direct Conmectior M Direct Connection settings Scanner senal number w network connectors o ee Scanner F address iW Enable changing IP address 1 92 168 2 11 Send to Scanner Scanner settings Driver Infa Imaging plate unit Scanner P address W Enable changing IP addi f 92 168 2 11 OBABAO S J09 CORE ver xx B Installation and setup Obtain an IP address for the unit from your network administrator and key it into the IP field in the Scanner IP address area NOTE The PC and the unit must be in the same subnet when setting the IP address of the unit PC Unit Press and hold down the Start key on the unit and then click the Send to Scanner button on the settings window Send to Scanner You will hear a beep which indicates that the PC is now sending the IP address the unit The
18. es are controlled The customer or the user of the DXR 50 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the DXR 50 as recommended below according to the maximum output power of the communications equipment Rated maximum output power of 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz transmitter W d 1 2 gt P d 1 2 gt P d 23 P 0 01 0 12 0 12 0 23 1 1 2 1 2 2 3 10 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people User s manual 204253 A 7 Imaging plate unit A Technical Specifications A 8 User s manual 204253 B Installation and setup Imaging plate unit Appendix B Installation and setup Installation and setup must only be carried out by service personnel trained and approved by the manufacturer of the unit 1 System installation Positioning the
19. etup options O 3 Press the on off key to switch the unit on The startup animation will appear on the display and the unit will carry out self test during which the IP carrier will slide out of the unit door OBABAO SJ09 CORE ver xx 10 User s manual 204253 2 Intraoral imaging plate unit Imaging plate unit When the status light turns green and ready animation indicating IP insertion appears on the unit display the unit is ready to use in the ready state NOTE lf the ready animation does not appear check the system setup described in the installation instructions Preparing an IP for exposure eee ee eee User s manual 204253 IMPORTANT NOTE If the IP is being used for the very first time or if it has not been used within the the last 24 hours it must be erased before use to remove any fogging caused by background radiation See section Erasing an imaging plate CAUTION When handling IPs protective covers and hygiene bags take all appropriate measures and precautions to prevent cross contamination 1 Place the IP you wish to use onto a protective cover The light blue side sensitive of the IP must face and be placed on the half of the protective cover that is the same shape as the IP Fold the half of the protective cover with the semicircular cutout over onto the IP The metal disk on the back black side of the IP must appear in the semicircular cut out 11 Imaging plate
20. fication Class 1 or 2 equipment depending on the classification lIEC60601 1 of the PSU No applied part Continuous operation IPXO enclosed equipment without protection against ingress of liquids Laser Safety Classification CLASS 1 LASER PRODUCT EN 60825 1 2007 ae 4 3 Ip mm Interface cable For Ethernet connections use an unshielded CAT6 LAN cable so that multiple chassis must not be connected The PC Ethernet switch to which unit ts connected to should be approved appropriately e g EN 60950 IEC 60950 UL 60950 Operating environment 10 C 40 C 30 90 RH 700 1060 mbar Storage transportation 10 C 50 C 0 90 RH 500 1080 mbar environment Operating current Less than 1 5 A User s manual 204253 A 1 Imaging plate unit A Technical Specifications A 2 Imaging Plates and hygiene bags Imaging plates IP Siz Size0 _ f Sizet Size Size 3 Image size KB 64m 325 458 606 IH Storage environment Imaging plates must be stored in their protective box below 33 C The box must be kept closed to remain dust free Photo stimulable phosphor uniformly coated on a support en plastic material Shielded with a top coat layer on active surface and encapsulated with lacquer around edges Imaging plates are industrial waste and must be disposed of in accordance with local and national regulations concerning the disposal of such material Hygiene bags Material
21. he unit Do not allow liquids to enter the unit 5 1 Cleaning the unit Use a non abrasive cloth moistened with either cool or lukewarm water soapy water mild detergent butylalcohol or ethanol ethyl alcohol 70 96 to clean the unit After cleaning wipe the unit with a non abrasive cloth moistened with water Never use solvents or abrasive cleaners to clean the unit Never use unfamiliar or untested cleaning agents If you are not sure what the cleaning agent contains DO NOT use it If you use a Spray cleaning agent DO NOT spray it directly into the unit door 5 2 Disinfecting unit CAUTION Wear gloves and other protective clothing when disin fecting the unit Wipe the unit with a cloth dampened with a suitable disinfectant solution such as ethanol 96 Never use abrasive corrosive or solvent disinfectants All sur faces must be dried before the unit is used WARNING Do not use any disinfecting sprays as the vapor could ignite and cause injury Disinfecting techniques for both the unit and the room where the unit is used must comply with all local and national regulations and laws concerning such equip ment and its location User s manual 204253 29 Imaging plate unit 30 5 Unit care and maintenance 5 3 Maintenance 5 4 Repair 5 5 Disposal The unit does not require any maintenance If the unit is damaged or malfunctions in any way it must only be repaired by service personne
22. imized image quality can be maintained irrespecective of who takes the image time is saved and profitability increased Problems caused by manually positioning the IP include incorrect vertical alignment distortion cone cut off poor projection standardization inferior image quality For more information on imaging plate holders and systems contact you dealer User s manual 204253 13 Imaging plate unit 2 Intraoral imaging plate unit Taking an exposure 14 i Place the IP in its sealed hygiene bag into the appropriate imaging plate holder and then insert it into the patient s mouth in the position for the image you wish to take Note that the back of the sealed hygiene bag the black side must face the X ray source L N T oe Noe 2 Select exposure values appropriate for the exposure you are taking The system will produce excellent images even if the exposure values differ considerably from the optimum values In most cases the same exposure values can be used for virtually all imaging purposes For normal everyday use select the Adult Bitewing exposure time from the following table If required the exposure time can be increased for very large patients and reduced for children The optimium exposure values also depends on the performace of the X ray unit being used and may vary by 1 step from the values in the following table User s manual 204253
23. it the erasing mode either wait 15 seconds for the unit to automatically exit the mode or press and hold down the Start key until the erase animation disapears lf there is a problem with the unit the error state symbol and an error number will appear on the unit display Restart the unit The error should clear If not contact your authorized distributor for assistance PC If the message Degraded image quality appears on the PC display during or after an IP is read it indicates that the image may not be within the factory set quality limits If the image looks okay no action is necessary If the message appears frequently contact your authorized distributor for assistance 23 Imaging plate unit 3 Setup options 3 Setup options There are several setup options in the dental imaging software that allow you to set the image quality to your requirements To select the setup options 1 From the dental imaging software you are using select Setup gt Scanner page 3 1 Tooth numbering Numbers can be assigned to the teeth that appear on the image Selecting the tooth numbering feature 1 Inthe Image Scanning section click the Show Image Preview and Dental Chart check box 2 Click OK to close the window and activate tooth numbering Using the tooth numbering feature After an IP has been read a window will open that shows the image and a tooth location map 1 Click the tooth teeth on the map that corresp
24. l author ized by the manufacturer of the unit At the end of the useful working life of the unit and or its accessories make sure that you follow national and local regulations regarding the disposal of the unit its accessories parts and materials The unit includes some or all of the following parts that are made of or include materials that are non environmentally friendly or hazardous electronic circuit boards electronic components User manual 204253 6 Symbols that appear on or in the unit Imaging plate unit 6 Symbols that appear on or in the unit 0537 TERTE W User s manual 204253 DANGEROUS VOLTAGE LASER RADIATION IMPORTANT INFORMATION Refer to user s manual Direct current Eternet connector RJ45 straight cable CE 0537 Symbol MDD 93 42 EEC This unit is marked according to the Medical Device Directive 93 42 EEC if the unit contains the CE mark ETL symbol This symbol indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately Please contact an authorized representative of the manufacturer for information concerning the decommissioning of your equipment 31 Imaging plate unit 7 Warnings and precations 7 Warnings and precations THE UNIT IS ACLASS 1 LASER PRODUCT Note When covers are removed the unit is a class 3B laser product avoid exposure to the laser beam CAUTION Use
25. n stored on an IP after an exposure Is sensitive to light Handle IPs in the same way as you would handle CDs or DVDs Hold IPs by their edges using your finger tips Protect the sensitive side of the IP from scratches wear and mechanical damage stains dirt dust and fingerprints NOTE Stains fingerprints on the sensitive side can degrade the diagnostic value of the image Dust wear on the sensitive side can appear as white grey spots on the image User s manual 204253 4 Handling and care of imaging plates Imaging plate unit 4 3 Cleaning User s manual 204253 Unprotected IPs must not come in contact with the patient the patient s saliva or any other bodily fluids Always use hygiene bags and protective covers with IPs when taking exposures to eliminate cross contamination maintain image quality maximize IP service life If IPs are handled and stored correctly cleaning should not be necessary or can be kept to an absolute minimum Clean the IP if there you see any visible marks or stains which are not obvious scratches on the image or if there is a reason to believe that the IP is contaminated Use the microfibre cloth supplied First very gently wipe the IP in a backwards forwards movement in both the widthwise and lengthwise directions and then finish with a circular wiping movement For stubborn marks and stains that cannot be be removed using the microfibre cloth use 70 96
26. n the PC display After a few seconds the image will appear in the dental imaging software The image can now be saved Refer to the documentation supplied with the dental imaging software you are using 17 Imaging plate unit 18 9 2 Intraoral imaging plate unit CAUTION If the metal disk on the rear of the IP can be seen on the digital image it indicates that the IP was exposed from the wrong side CAUTION RETRIEVING IMAGES If the image is not transferred to the PC because of a network PC or software failure the image can be retrieved from the units memory as long as the unit is NOT switched off For information on how to do this see section Retrieve last image After the IP has been read it will be automatically erased and then ejected from the unit into the plate collector After the IP has been ejected the ready animation will reappear on the display The unit is now ready to read the next IP User s manual 204253 2 Intraoral imaging plate unit Imaging plate unit Multiconnect configuration 1 User s manual 204253 PC To reserve the unit click the GREEN multiconnect icon which is in the bottom right hand corner of the PC display The Connect window will appear PC In the Connect window click the Reserve the Scanner button to reserve scanner Reserve the scanner The window will disappear and the GREEN multi connect icon will start to flash This indicates that you
27. nationally harmonized IEC standards 32 User s manual 204253 7 Warnings and precations Imaging plate unit Unit not suitable for use in the presence of flammable anaesthetic mixture with air or with oxygen or nitrous oxide For ethernet connections use an unshielded CAT6 LAN cable so that multiple chassis must not be connected The PC Ethernet switch to which unit is connected to should be approved appropriately e g EN 60950 IEC 60950 UL 60950 After installation check that the IEC 60601 1 leakage current levels are not exceeded In order to maintain safe and correct functioning of the unit only the power supply unit PSU delivered with the unit or distributed by authorized dealers Please refer to the unit technical specifications for a list of the PSUs If this device will be used with 3rd party imaging application software not supplied by the manufacturer the 3rd party imaging application software must comply with all local laws on patient information software This includes for example the Medical Device Directive 93 42 EEC and or FDA if applicable Medical electrical equipment needs special precautions regarding EMC and needs to be installed according to EMC information User s manual 204253 33 Imaging plate unit 7 Warnings and precations 34 User s manual 204253 A Technical Specifications Imaging plate unit Appendix A Technical Specifications A 1 Unit Intraoral imaging plate unit DXR 50 Classi
28. ns interruptions it is recommended that the DXR 50 be powered from an uninterruptible power supply or a battery Power frequency magnetic field should be at levels characteristic of atypical location ina typical commercial or hospital environment NOTE U is the a c mains voltage prior to application of the test level User s manual 204253 Imaging plate unit A Technical Specifications Guidance and manufacturer s declaration electromagnetic immunity The DXR 50 is intended for use in the electromagnetic environment specified below The customer or the user of the DXR 50 should assure that it is used in such an environment Immunity IEC 60601 test Compliance Electromagnetic environment guidance test level level Portable and mobile RF communications equipment should be used no closer to any part of the DXR 50 including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance Conducted 3 Vms 3 V d 1 2 P RF 150 kHz to es 61000 80 MHz d 1 2 JP 80MHzto 800 MHz d 2 3 P 800 MHz to 2 5 GHz paved on an where Pis the maximum output power rating RF 80 MHz to IEC 61000 25 GHz of the transmitter in watts W accord ing to the 4 3 transmitter manufacturer and dis the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site
29. o ROCES SING eea es oee a a ELTE 25 4 Handling and care of imaging plates 0 00n0nnannannannnnennannnnennnnnn 26 A We 92 eee eee et oe ee ee 26 AZ NANING ee E ee ee een eee eee ee 26 4 S Cleaner ne 27 A A SONA EE A E E A A E E E aoennte qatuns ete 28 AD REPII CS MEN enna E A 28 AS MIS DOS A a a eo pe 28 5 Unit Care ANd MAINTENANCE ccc ccc ccc ecc eee eeceeceeceeeeeceeceeceeseeeeesees 29 5 1 Cleaning NE UNM ies s5 ccncciccd caddie ndsssdsedcsaiesedcsabeges Sand nocd apd ades a5bgedendekieledasddes sis 29 SA IS Iinilic CUM UN Seah nad aces eetcenseetaconeacaasedeercceesenece anes aes i 29 29 WIAINLE MANIC asien tas adasaheaedeanieasacarieeiccaieaviaseh foul ates eai 30 HES CGI ea a E cane nsecaate noanisap aaniaagiasenesasoteeeaGaes 30 DD DISD OS all sasaiatucanasteacanvartiossian dea caoaaceenacicas A E EROE O 30 6 Symbols that appear on or in the UNIt ee eeceee seca eeeeeeeeeaees 31 7 WarningS and precatlions canssa ia a a 32 Appendix A Technical Specifications ccccccceceeeeeeeeeeeeeeeeees A 1 Appendix B Installation and setup cccccececeeseeeeeeeseeseeeeeeereees B 1 IV User s manual 204253 Introduction Imaging plate unit 1 The DIGORA Optime 1 1 Introduction User s manual 204253 This manual describes how to use the DIGORA Optime digital imaging plate unit the unit which is part of the digital intraoral imaging plate system the system
30. of controls or adjustments or performance of procedures other than those specified herein may result in hazardous laser radiation exposure e When handling imaging plates protective covers and hygiene bags always take the appropriate hygiene measures and precautions to prevent cross contamination e The imaging plates are harmful if swallowed e Donot move or knock the unit when it is reading an imaging plate e This unit must only be used to read image plates supplied by the manufacturer and must not be used for any other purpose NEVER use imaging plates protec tive or hygiene bags from other manufacturers e This unit or its accessories must not be modified altered or remanufactured in any way e Only the manufacturer s authorized service personnel are authorized to carry out maintenance and repair of the unit There are no user serviceable parts inside the unit e Infection control procedures must be observed when using accessories such as film holders x ray tube guides and imaging plates When using accessories always follow the manufacturer s instructions on how to use the accessory and prevent cross contamination from one patient to another e This unit can interfere with other devices due to its EMC characteristics e Other devices can interfere with this unit due to their EMC characteristics e This unit complies with IEC 60601 1 standard Accessory equipment connected to this device must be in compliance with the related
31. ond to the tooth teeth in the image Tooth numbers will be assigned to the selected teeth The tools at the top of the window allow the image to be manipulated 2 Click OK to save the tooth numbers with the image 24 User s manual 204253 3 Setup options 3 2 Resolution Imaging plate unit In the Image Scanning section select either Super or High default Super gives a pixel size of 35u This results in images with better resolution but more memory is required as the image files are larger High a pixel size of 64u This results in images with less noise especially if short exposure times are used Click OK to close the window and accept the selected resolution 3 3 Image Processing User s manual 204253 1 In the Image processing click the Noise Filter ing check box Noise filtering is normally always selected It makes images smoother when they are taken at lower radiation doses Click OK to close the window and select the noise filtering 25 Imaging plate unit 4 Handling and care of imaging plates 4 Handling and care of imaging plates 4 1 General 4 2 Handling 26 The correct use handling cleaning and storage of imaging plates guarantees the best image quality and maximum service life of the imaging plates X ray exposures DO NOT cause IPs to age The light coloured side of the IP is the side which stores the x ray image It is the SENSITIVE side Image informatio
32. puter network connection that provides the LAN network connection is selected Imaging plate unit Mult Connect i Use Multi Connect Workstation identiter Scanner Sutorelease timeout seconds B Installation and setup PC If the unit is to be used with several PCs select the Use Multi Connect check box and select a unique Workstation identifier number between 1 and 8 for the PC being configured from the drop down list Addition workstation information for example user name location etc can be entered into the field next to the work station identifier number IMPORTANT NOTE If only one PC is connected to the unit do not select the Use Multi Connect check box Frw2251 616 a r d 4 F b G a Cancel The Scanner Autorelease timeout is the length of time that the unit will remain reserved and unused by a PC before the PC automatically released the unit so that it can be used by another PC in the system the scanner can be reserved in advance from another PC The default setting is 40 seconds This can be changed by keying in a new value Click OK to connect the PC to the unit NOTE An automatic technique will automatically locate the unit within the local area network and connect the PC B Installation and setup Imaging plate unit 9 Repeat the above process for all the other PCs in the network Make sure that you give each PC a different Workstation identifier 10 Check th
33. right light Sunlight or bright light must not be allowed to shine directly on the unit door into which the IPs are inserted Position the unit on a stable flat surface so that vibrations will not degrade the image quality The unit can also be attached to a wall under or on a shelf with the optional mounting kit The unit must not be positioned so that it it touching other equipment It must not be placed on top of or under other equipment The unit can be positioned within the environment in which the patient is examined and treated patient environment User s manual 204253 Introduction Imaging plate unit Positioning the PC The PC connected to the unit should not be used in the patient environment The minimum horizontal distance between the patient and the PC is 1 5 m 4 5 ft The minimum vertical distance between the patient and the PC is 2 5 m 6 5 ft Other devices User s manual 204253 DO NOT connect any other devices to the unit or the PC connected to the unit that are not part of the supplied system notsupplied by the manufacturer of the unit notrecommended by the manufacturer of the unit Imaging plate unit 2 Intraoral imaging plate unit 2 Intraoral imaging plate unit 2 1 Main parts and controls Power on off key and status light O green on flashing entering the standby mode yellow in the standby mode ial Proximity sensor Display jz
34. unit Position the unit on a stable flat surface so that vibrations will not degrade the image quality The unit can also be attached to a wall under orona shelf using the optional scanner mounting Kit pt no 204138 Do not position the unit in direct sunlight or near bright light Sunlight or bright light must not be allowed to shine directly on the unit door into which the IPs are inserted The unit must not be positioned so that it touching other equipment The unit must not be placed on top of or under other equipment The unit can be positioned within the environment in which the patient is examined and treated patient environment Positioning the PC s not supplied The PC s connected to the unit should not be used in the patient environment The minimum horizontal distance between the patient and the PC s is 1 5 m 4 5 ft The minimum vertical distance between the patient and the PC s is 2 5 m 6 5 ft Imaging plate unit B Installation and setup Other devices DO NOT connect any other devices to the unit or the PC s connected to the unit that are not part of the supplied system notsupplied by the manufacturer of the unit notrecommended by the manufacturer of the unit 2 Connecting the unit to a PC LAN The procedure for connecting the unit to a single PC or several PCs in a local area network LAN is exactly the same except that every PC in the LAN needs to be given a unique ID number
35. unit 2 Intraoral imaging plate unit 3 Turn the protective cover and IP over so that the black side of the protective cover is uppermost This makes it easier to slide the protective cover and IP into the hygiene bag 4 Slide the protective cover and IP into the hygiene bag as far as they will go Make sure that the black side of the protective cover is on the same side as the black side of the hygiene bag 5 Peel off the cover paper from the sealing tape and then fold the flap along the pre formed line over and onto the sealing tape 6 Press and slide your thumb along the tape to ensure that the flap is properly sealed 7 Turn the sealed hygiene bag over and check that the IP and protective cover are in the correct position You must be able to see the light side of the protective cover and the metal disk on the IP 12 User s manual 204253 2 Intraoral imaging plate unit Imaging plate unit Imaging plate holders It is recommended that imaging plate holders be used to ensure accurate IP positing and consistently good images quality Using imaging plate holders improves image quality because the IP is positioned correctly in relation to the tooth there is no positioning guesswork the IP is not bent and thus distortion is eliminated the IP cannot move in relation to the X ray unit images are standardized and reproduceable there is no overlapping nor cone cut off P wear and tear is min
36. use the IP connection method The unit has been configured to work with one PC only and does not recognize any other PCs Reconfigure the system for multiple users PROBLEM The unit is connected to the PC but cannot scan images CAUSE SOLUTION Reserve the unit and check that it is configured for multiple users If not reconfigure the system Ifthe Dental imaging software animation flashing yellow appears it indicates that a patient has not been selected for the scanned image s Select a patient If the Insert animation appears on the unit display while inserting the IP but the unit does not insert the IP into the unit nor start scanning the reflective sensor in the unit door is faulty refer to the service manual B Installation and setup Imaging plate unit PROBLEM Local area network subnet configuration problem CAUSE SOLUTION i Ifthe Dental imaging software animation flashing yellow appears and cable s router s are known to be okay ask your network administrator for assistance ii Conflict mismatch IP addresses and or subnet mask of the unit PC Use command lt ipconfig gt from the command prompt Start gt gt Run gt gt cmd in Windows to get information about the active network settings Change either the IP address of the PC or the unit so that they are both in the same subnet If the problem persists ask your network administrator for assistance PROBLEM Ethernet link
37. w imaging plate is inserted into the unit or when the Start key is pressed After four hours the unit will exit the standby mode and automatically shut down User s manual 204253 2 Intraoral imaging plate unit Imaging plate unit Retrieve last image If the last image read is not transferred to the PC because of a network communication PC or software failure the last image read can be retrieved IMPORTANT NOTE The LAST read image can only be retrieved if the unit is left on If the unit is switched off the image will be lost To retrieve the last scanned image 1 User s manual 204253 Correct the problem that caused the communication failure When the connection between the unit and the PC is re established the last read image will automatically be transferred to the PC PC If the image is not automatically transferred to the PC select the Setup gt Scanner page from the dental imaging software your are using PC In the Retrieve Last Image field click the check box to retrieve the last read image NOTE If required you can select different parameters e g resolution show image preview etc for the image to be retrieved PC Click OK to close the Setup window The last read image will be transferred to the PC 21 Imaging plate unit 22 2 Intraoral imaging plate unit Shutting down the unit O ie Press and hold power on off key until the indicator light goes off NOTE
38. y test Electrostatic discharge ESD IEC 61000 4 2 IEC 60601 test level 6 kV contact 8 kV air Compliance level 6 kV contact 8 kV air Electromagnetic environment guidance Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast transients bursts IEC 61000 4 4 2 kV for power supply lines 1 kV for input output lines 2 kV for power Supply lines 1 kV for input output lines Mains power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 Voltage dips short interruptions and voltage variations On power supply lines IEC 61000 4 11 Power frequency 50 60 Hz magnetic field IEC 61000 4 8 1 kV differential mode lt 5 U gt 95 dip in Ur for 0 5 cycle 40 Ur 60 dip in Ur for 5 cycles 70 Ur 30 dip in Ur for 25 cycles lt 5 Ut gt 95 dip in Ur for 5 sec 1 kV differential mode lt 5 Ur gt 95 dip in Ur for 0 5 cycle 40 Ur 60 dip in Ur for 5 cycles 70 Ur 30 dip in Ur for 25 cycles lt 5 Ur gt 95 dip in Ur for 5 sec Mains power quality should be that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or hospital environment If user of the DXR 50 requires continued operation during power mai