Home

Valor 1000 Series Instruction Manual (Spanish) (French) (German

image

Contents

1. Compliance to the following standards is indicated by the corresponding mark on the product Sandor FCC Note This product conforms to the EMC directive 2004 108 EC and the Low Voltage Directive 2006 95 EC The complete declaration of Conformity is available from Ohaus Corporation AS NZS4251 1 Emission AS NZS4252 1 Immunity Disposal In conformance with the European Directive 2002 96 EC on Waste Electrical and Electronic Equipment WEEE this device may not be disposed of in domestic wasie This also applies to countries outside the EU per their specific requirements Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment If you have any questions please contact the responsible authority or the distributor from which you purchased this device should this device be passed on to other parties for private or professional use the content of this regulation must also be related Thank you for your contribution to environmental protection This equipment has been tested and found to comply with the limits for a Class A digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance wi
2. OHAUS Valor 1000 Series Instruction Manual Spanish French German Italian OHAUS Valor 1000 Series EN 1 1 INTRODUCTION This manual contains installation operation and maintenance instructions for the Ohaus Valor 1000 Series Scales Please read the manual completely before using the scale 1 1 Safety Precautions Please follow these safety precautions e Verify AC Adapter input voltage matches the local AC power supply Do not immerse the scale in water or other liquids Do not operate the scale in hostile environments Do not drop loads on the platform Do not place the scale upside down on the pan service should only be performed by authorized personnel Disconnect the scale from the power supply when cleaning 1 2 Controls Figure 1 1 Controls 6 7 segment 6 digit back LOD w units of measure enuniaiors fe Powerinptjock SSS o Lnnraenig OO EN 2 OHAUS Valor 1000 Series TABLE 1 2 CONTROL FUNCTIONS Button Action Function ZERO short Press sets display to zero Enter Menu Press Confirms settings TARE short Press Tares weight of item on pan H Menu Press Decrement over and under settings M short press stores displayed weight in accumulation memory t Display accumulation data when display is at zero Units Menu Press Increment over and under settings Long Press Change unit of measure while long pressing SELECT button SELECT short press Shift to next digit when setting over and
3. Series EN 7 5 MAINTENANCE Caution before cleaning turn off the scale remove the AC adapter 5 1 Cleaning The housing may be cleaned with a cloth dampened with a mild detergent if necessary Do not use solvents chemicals alcohol ammonia or abrasives to clean the housing or control panels 5 2 Troubleshooting The following table lists common problems and possible causes and remedies If the problem persists contact Ohaus or your authorized dealer TABLE 5 1 TROUBLESHOOTING Scale will not turn on No power to scale Verify connections and voltage Battery power used up Connect power and charge the battery Poor accuracy Improper calibration e Unstable Perform calibration e Move scale to environment Suitable location Unable to calibrate Unstable environment Move the scale to suitable location Incorrect calibration mass Use correct calibration mass scale flashes Bat Battery discharged Connect power and charge the battery Battery fails to charge Battery is defective Have battery replaced by authorized service dealer Pan has load during power on Remove weight from pan and re zero Weight on pan exceeds capacity Remove weight from the pan Invalid under and over limit settings Re enter under and over limits 5 3 Service Information If the troubleshooting section does not resolve or describe your problem contact your authorized Ohaus service agent For service assistance or technical support in the United States call toll f
4. desired unit icon is displayed Press ZERO to set the display to zero Put an empty container on the pan and then press TARE to tare the container weight Add material to the container The display shows the weight of the material in selected unit of measure 3 5 Accumulate Mode Use this mode to store the weight and HI LO SUM statistics from a series of samples Press ZERO to set the display to zero Place the sample on the pan Press M to store the weight of the sample Storage is confirmed by the display n X where x is the sample number The enunciator above SUM text will illuminate fo let user know there are readings in memory Remove the sample wait for stable zero reading then place the next sample on the pan Press M to store the weight of the OHAUS Valor 1000 Series EN 5 next sample Repeat this process until all samples have been weighed Review the accumulation data by pressing M when the display is at zero The display shows the number of samples n x Press M fo view the maximum sample weight Hxxxxx Press M fo view the minimum sample weight Lxxxxx Press M to view the sum of weights Xxxxx Press M to return to weighing Clear the accumulated data by pressing ZERO when the number of samples is displayed 3 6 Check Weigh Mode Use this mode to compare samples to a target weight range Press ZERO to set the display to zero Place a sample on the pan and read the weight and under accept over sta
5. in a dry environment dealer Risk of explosion can occur if replaced with the wrong type or connected improperly Dispose of the lead acid battery according to local laws and regulations CAUTION Battery is to be replaced only by an authorized Ohaus service EN 4 OHAUS Valor 1000 Series 2 4 Initial Calibration When the scale is operated for the first time a Span calibration is recommended to ensure accurate weighing results Before performing the calibration be sure to have the appropriate calibration weight Refer to section 4 3 for Span calibration procedures 3 OPERATION 3 1 Turning Scale On Off To turn the scale on flip toggle switch below right front keyboard The scale performs a display test momentarily displays software version and then enters the active weighing mode To turn the scale off toggle same switch fo off position 3 2 Zero Operation Zero is set under the following conditions e Automatically at Power On Cinitial Zero e Manually by pressing the ZERO button Press the ZERO button to zero the weight display The scale must be stable and weight 2 of capacity to accept zero operation 3 3 Manual Tare Place a container on the weighing pan then press TARE The display will show a net value of OU To clear the Tare value press TARE with the pan empty 3 4 Weighing Mode Use this mode to measure the weight of a sample in the selected unit of measure Press and hold SELECT and press M Unit until the
6. the warranty period shall begin on the date of shipment to the authorized dealer No other express or implied warranty is given by Ohaus Corporation Ohaus Corporation shall not be liable for any consequential damages As warranty legislation differs from state to state and country fo country please contact Ohaus or your local Ohaus dealer for further details OHAUS Ohaus Corporation 19A Chapin Road P O Box 2033 Pine Brook NJ 07058 2033 USA Tel 973 377 9000 Fax 973 944 7177 With offices worldwide Con oficinas alrededor del mundo Avec des bureaux dans le monde entier Weltweite Gesh ftsstellen Con uffici in tutto il mondo www ohaus com 8 0 2 5 1 2 2 0 P N 80251220 Ohaus Corporation 2007 all rights reserved todos los derechos reservados tous droits r serv s Alle Rechte vorbehalten tutti i diritti riservati Printed in China Impreso en la China Imprim en Chine Gedruckt in China Stampato in Cina
7. under values i i Long Press Change unit of measure with short press of M buiton j 2 INSTALLATION 2 1 Package Contents e Valor 1000 Scale e Stainless steel Pan e AC Power Adapter e Instruction Manual e Warranty Card 2 2 Location Use the scale on a firm steady surface Avoid locations with excessive air current vibrations heat sources or rapid temperature changes Adjust the leveling feet so the bubble is centered in the circle Level Bubble Level Feet Figure 2 1 Leveling the Scale OHAUS Valor 1000 Series EN 3 2 3 Power Connect the AC Adaptor to the AC mains supply Connect the plug to the DC jack on the bottom of the scale The scale may be operated on the AC mains supply or the built in rechargeable battery The scale must be turned on to charge the battery DC Mains Figure 2 2 Power Connection 2 3 1 Battery Power The scale can be operated on the internal rechargeable battery when AC power is not available The scale will automatically switch fo battery operation if there is a power failure or the power cord is removed for up to 12 hours During charging the battery Indicator displays the charge level see table 1 1 and the scale can be operated during charging The battery is protected against over charging and the scale can remain connected to the AC power line Before using the scale for the first time the battery should be fully charged NOTE e Charging the scale must only be performed
8. m the setting and return to weighing 4 3 CALIBRATION span Calibration Turn on the scale When display flashes all segments 888888 press and hold ZERO and SELECT at the same time SET will be displayed at the end of power on sequence Press the SELECT key until SET F1 is shown on the display Press ZERO key to display A x x divisions in grams 1 2 5 10 20 50 Press ZERO key to accept the default division setting which is shown on the display Press ZERO key fo display b X x decimal point 0 1 2 3 Press ZERO key to accept the default decimal point setting which is shown on the display Press ZERO key to display Cxxxxx xxxxx sSpan weight or capacity of scale Press ZERO key to accept the default span calibration weight setting which is shown on the display or press M or TARE to increment or decrement the span calibration weight setting Press ZERO to display dx x Zero tracking setting 0 9 from 0 125d to 2d Press ZERO key to accept the default setting which is shown on the display or press M or TARE to increment or decrement setting Default setting is 3 which 0 5d Press ZERO and display will show Load O Make sure pan is empty then press ZERO key to accept Display will show CAL O briefly then Load F will be displayed Press ZERO to display Fxxxxx xxxxx Span weight Put displayed weight on the scale and press ZERO to accept After completion scale will be in weighing mode OHAUS Valor 1000
9. ree 1 800 526 0659 between 8 00 AM and 5 00 PM EST An Ohaus product service specialist will be available to provide assistance Outside the EN 8 OHAUS Valor 1000 Series USA please visit our web site www ohaus com to locate the Ohaus office nearest you 6 TECHNICAL DATA The technical data is valid under the following ambient conditions Ambient temperature O C to 40 C Maximum Relative humidity 80 for temperatures up to 31 1C decreasing linearity to 50 relative humidity at 40 C Height above sea level Up to 2000m Operability is assured at ambient temperatures between 0 C and 40 C Power AC adapter 9 VDC 500 mA output Protection Protected against dust and water Pollution degree 2 Installation category Class III EMC See Declaration of Conformity 6 1 Specifications TABLE 6 1 Specifications cacy ts Reay os PP Weigh Accumulate Checkweigh Center of zero Stability Net Sum Over Accept Under Check Weighing Display Indicators unit and Battery charge status Display Size 20 mm 0 8 in digits Tare Range To capacity by subtraction Stabilization Time lt 3 seconds Auto Zero Tracking User selectable from 0 1d to 2d 32 F to 104 F 0 C to 40 C at 10 to 95 relative humidity Operating Temp Humidity Non condensing up to 2 5 miles 4000m above sea level ES Power Internal rechargeable sealed lead acid battery pan Size mm 4 0 8 8 OHAUS Valor 1000 Series EN 9 6 2 Compliance
10. th the instruction manual may cause harmful interference to radio communications EN 10 OHAUS Valor 1000 Series Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense This Class A digital apparatus complies with Canadian ICES 003 Cet appareil num rique de la classe A est conforme la norme NMB 003 du Canada ISO 9001 Registration In 1994 Ohaus Corporation USA was awarded a certificate of registration to ISO 9001 by Bureau Veritus Quality International BVQI confirming that the Ohaus quality management system is compliant with the ISO 9001 standard s requirements On May 15 2003 Ohaus Corporation USA was re registered to the ISO 9001 2000 standard LIMITED WARRANTY Ohaus products are warranted against defects in materials and workmanship from the date of delivery through the duration of the warranty period During the warranty period Ohaus will repair or at its option replace any component s that proves to be defective at no charge provided that the product is returned freight prepaid to Ohaus This warranty does not apply if the product has been damaged by accident or misuse exposed to radioactive or corrosive materials has foreign material penetrating to the inside of the product or as a result of service or modification by other than Ohaus In lieu of a properly returned warranty registration card
11. tus on display Refer fo section 4 7 fo Set limits 3 7 Changing Units of Measure Press and hold the M and SELECT buttons to display the next available measuring unit 4 SETTINGS The scale allows the user to customize several settings including check weigh limits backlight operation and calibration 4 1 Over and Under Limits Press and hold ZERO and press SELECT until SET is displayed Press SELECT to display SET FO Press ZERO to view the upper limit 000000 Press e Y io edit the upper limit Press ENTER to confirm the setting and display the lower limit 000000 Press e 3 to edit the lower limit Press ENTER to confirm the lower setting and display will show Run X where x is the alarm setting In this case x O Turns off check weigh mode 1 under accept over indicators are displayed alarm beeps when weight is outside accept range 2 under accept over indicators are displayed alarm beeps when weight is within accept range 3 under accept over indicators are displayed alarm is off EN 6 OHAUS Valor 1000 Series Press SELECT to change the alarm setting Press ENTER fo confirm setting and display End Press and hold ZERO key then press SELECT to return to weighing 4 2 Backlight Press and hold ZERO button until b9 x is displayed where x O backlight is turned off 1 backlight is on during weighing 2 backlight is always on Press TARE to change the setting Press ZERO to confir