Manual brochure MAG2000 (ENG) - Sanitaria Ortopedia Campo
Contents
1. Carnage and storage Carriage precautions MAG2000 is a portable device so it does not need any particular carriage precautions However we recommend to put away MAG2000 and its accessories in their own bag after every treatment We recommend to not roll up power supply and applicators cables Storage precautions MAG 2000 is protected till following environmental conditions Outside of the packaging Temperature from 5 to 40 C Rel humidity from 30 to 80 Pressure from 500 to 1060 hPa Inside of the packaging Temperature from 10 to 55 C Rel humidity from 10 to 90 Pressure from 500 to 1060 hPa Disposal The equipment is subjected to WEEE regulations see the symbol oe on the label concerning separate waste collection when disposing this product please use the designed areas for disposing electronic waste or contact the manufacturer Troubleshooting If it is used in accordance with the instructions of the user manual MAG2000 does not need a particular regular maintenance If you find any malfunctioning using MAG2000 please follow these instructions I A C E R Srl 24 MNPG52 04 a n kO MEDICAL DIVISION e check the main integrity by connecting a running device to the same main e check the connection with power supply and connection cables integrity e check the correct connection between MAG2000 and the applicator or the applicators e check all the operations have been done properly e every t
2. by connecting the applicator cable to one of two plugs CH1 CH2 placed on device upper side Connect the power supply cable to the main then connect the power supply plug to the circular connector placed on device upper side near to ON OFF switch Connect the power supply plug to the main 110 230 VAC 50 60 Hz Move the ON OFF switch placed on device upper side to the ON position I TECH logo and programs menu will be displayed on screen Select the therapy program running through programs menu using A an Y buttons and put yourself on desired program Osteoporosis Arthrosis Arthritis Cerv Arthrosis Articular pain Cervicalgias Sprains Press OK Screen will display base setting with therapy time 2 hours and magnetic field intensity 50 int These mean values are recommend by IACER in order to start immediately the treatment I A C E R Srl 13 MNPG52 04 I TECH 8 Press OK key The device will start the treatment and on screen will be displayed the magnet icon with magnetic field flux Display green light indicates that therapy is running RST SECT TCE 7 PTET eas T F ai pet pera sitter 9 At the end of therapy the screen will display automatically the menu programs Attention it s possible to stop temporary the therapy at any time pressing OK key at least for 2 seconds Press again OK key to continue the treatment During pause time green led turns off till the treatment
3. 04 MEDICAL DIVISION e InP1 P20 programs a longed treatment with an intensity higher than 60 can heat the 3 solenoids belt and this aspect makes therapy less comfortable we recommend to space out treatments and to not go over 2 consecutive hours of therapy e In P21 P35 programs we recommend to use the professional solenoids couple available as optional accessories if you want to adjust an intensity higher than 100 and a treatment time longer than 2 hours e Do not adjust intensity higher than 50 if you use magnetotherapy carpet optional accessory for prolonged treatments Adjustable programs instructions With MAG2000 you can adjust time therapy and magnetic field intensity parameters as indicated in the following steps 1 Connect the applicator or the applicators to the device by connecting applicator cable to one of two plugs CH1 CH2 placed on device upper side Connect the power supply cable to the main then connect the power supply plug to the circular connector placed on device upper side near to ON OFF switch Connect the power supply plug to the main 110 230VAC 50 60 Hz Move the ON OFF switch placed on device upper side to the ON position I TECH logo and programs menu will be displayed on screen Select therapy program running through programs menu using A and buttons and put yourself on desired program I A C E R Srl 17 MNPG52 04 I TEGR Osteoporosis Arthrosis Arthritis
4. Cerv Arthrosis Articular pain Cervicalgias Sprains 6 Press OK Screen will display base setting with therapy time 2 hours and magnetic field intensity 50 int These parameters can be modified as follows a Press OK button screen displays a moving key icon on the left side b Press A and Via to adjust the desired therapy hours from 0 to 24 and confirm by pressing OK key Screen will highlight the therapy minutes I A C E R Srl 18 MNPG52 04 TECH c Press A and Via to adjust the desired therapy minutes from 0 to 59 and confirm by pressing OK key Screen will highlight treatment intensity d Press A any ici to adjust the treatment intensity from 5 to 100 Gauss on P1 P20 programs from 5 to 150 Gauss on P21 P35 programs and confirm by pressing OK key 7 Display will show the step 6 screen highlighting the key icon on the left side press 8 Press OK the device will start the treatment displaying on screen the magnet icon with magnetic field flux Green light indicates that therapy is running I A C E R Srl 19 MNPG52 04 I TEGR 9 At the end of therapy the screen will display automatically the menu programs Attention it s possible to stop temporary the therapy at any time pressing OK key at least for 2 seconds Press again OK key to continue the treatment During pause time green led turns off till the treatment restarts Attention it s possible to get out from the treatment
5. E R Srl 21 MNPG52 04 m MEDICAL DIVISION 26 treat 20Hz 20 fre tree 24 hours _ 7 treat 30Hz 30 fre tree 24 hours _ 28 Treat 40 Hz 40 fre tree 24 hours _ 29 treat so Hz 50 fre tree 24hours 30 treat 6oHz 60 free free 24hours 31 treat 7oH2 70 fre tree 24hours 3 Treat 80 Hz 80 fre tree 24hours 3 Treat go Hz 90 fre tree 24 hours _ 34 treat 100Hz 100 fre tree 24hours 35 Autoscanm 2 6 24hours Autoscan program allows to adjust the desired therapy time then it will start automatically a frequency cycle from 10 Hz to 100 Hz with a time therapy of 5 minutes for each frequency It s an ideal program for the regeneration of both hard tissues bones and soft tissues tendons ligaments in the same treatment Therapy duration values are recommended by 1 A C E R S r l however the user can adjust the time as he prefers MAG2000 uses therapy time values working frequency values and field intensity values coming from scientific and medical literature as result of well known sperimentations and clinical evaluations Barker Lunt 1983 Bassett Pawluk Pilla 1974 Bassett Valdes Hernandez 1982 Setup language selection Move the ON OFF switch placed on device upper side to the ON _ position IMmediately after keeping pressed the button until the language list appears on the display Release the button select the
6. January 2010 Notified body Cermet Via di Cadriano 23 40057 Cadriano di Granarolo BO Italy Certification Path Annex II MAG2000 is a Class lla equipment with reference to Directive 93 42 EEC MDD annex IX rule 9 and following modifications Martellago 03 06 10 Legal Representative Mario Caprara FI I A C E R Srl 5 MNPG52 04 po MEDICAL DIVISION Specifications MAG 2000 has the following specifications e Class Ila equipment Directive 93 42 CEE Annexed IX rule 9 and following modifications e Class II applied part type BF Classif CEI EN 60601 1 e P21 protection equipment against solids dust and liquids penetration e Equipment and accessories not subjected to sterilization e Use of the equipment is prohibited close to flammable substances when mixed with air with nitrous oxide or when mixed with any flammable agents and in environments with high concentrations of oxygen e Continuous operating mode equipment e Equipment not suited to be used in external Purpose Clinical purpose Therapeutic Use Clinic Hospital and domestic use MAG2000 is indicated for the treatment rehabilitation and functional recovery of the following pathologies e wrist hand shoulder foot ankle and knee articulation e skeletal motor apparatus e arthrosis e atrophies and muscular dystrophy e bursitis e bruises I A C E R Srl 6 MNPG52 04 NL MEDICAL DIVISION e degeneration
7. of locomotor apparatus e sprains e periarthritis e benign lesions and muscular tears e tendinitis MAG2000 is particularly suitable for the treatment and the care of the osteoporosis and all the pathologies on bony tissues Cellulite treatment is not inserted in the CE0476 marking of the device device Cellulite treatment can be used only for beauty purposes Thanks to its high magnetic field intensity MAG2000 is particularly suitable for the treatment of bone fractures also with rigid bandages or braces MAG2000 device is indicated both for professional physiotherapists medics etc and for domestic user In case of home therapy we recommend using the device exclusively on medical therapist suggestion Expected lifetime 5 years Technical features Power supply Power UE24WCP 150120SPA 15VDC 1 2A Max absorbed current 0 6 A Insulation class CEI EN 60601 1 II Applied part CEI EN 60601 1 BF Dimensions mm 180x110x50 Field intensity Adjustable on increasing level up to 100 Gauss each channel in P1 P20 programs Adjustable on increasing level up to 150 Gauss each channel in P21 P35 programs Squared wave frequency 1 100 Hz Therapy time Adjustable by user In frequency programs from 21 to 35 maximum magnetic field intensity is 150 Gauss for each channel with solenoids couple applicator I A C E R Srl 7 MNPG52 04 I TECH MEDICAL DIVISION Intensity frequency and time values are given with 20 of accur
8. A C E R Srl 29 MNPG52 04 h 9 MEDICAL DIVISION For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in Watts W according to the transmitter manufacturer Note 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Warranty Make reference to the national laws for any warranty conditions by contacting the national distributor or directly the manufacturer IACER a DE t B MAG2000 All rights reserved MAG2000 and uenient sinision logos are owned by I A C E R Srl and are registered I A C E R Srl 30 MNPG52 04 I A C E R Srl 31 MNPG52 04 ILA C E R S r l Sede operativa 30030 Martellago VE Via S Pertini 24 A Tel 39 041 5401356 Fax 39 041 5402684 Sede legale S Marco 2757 30124 Venezia Cod Fisc P IVA IT 00185480274 R E A VEN 120250 M VE001767 Cap Soc 110 000 00 i v ww w iacer it iacer iacer it I A C E R Srl 32 MNPG52 04
9. USER MANUAL MNPG52 04 Edition 27 06 2013 Magnetotherapy model MAG2000 LA C E R Srl www iacer it www itechmedicaldivision com I A C E R Srl Via S Pertini 24 A 30030 Martellago VE ITALY Tel 39 041 5401356 Fax 39 041 5402684 e mail iacer iacer it http www itechmedicaldivision com Summary Introduction Magnetotherapy Technical Specifications Declaration of conformity Specifications Purpose Technical features Labelling Label details Symbols description Kit contents How to use Warnings Electromagnetic interference Contraindications and side effects Summary DCODOANDWDAGA AR Q 12 Quick use of the device with presetted parameters Adjustable programs instructions Stored programs list Set up language selection Maintenance Functioning control Cleaning Carriage and storage Disposal Troubleshooting Assistance Spare parts EMC tables Warranty I A C E R Srl MNPG52 04 iE MEDICAL DIVISION Introduction Magnetotherapy It s a long time that low frequency and high intensity pulsed electromagnetic fields have met maximum scientific consent in chronic and degenerative diseases treatment Magnetotherapy uses low frequency and high intensity pulsed electromagnetic fields induced by electric current on a bobbin due to its characteristics the electromagnetotherapy is universally recognized as the most suitable technique for the treatment of the bony pathol
10. acy Environmental conditions of operation Temperature from 5 to 40 C Relative humidity from 10 to 93 Pressure from 500 to 1060 hPa Labelling Label 1 back I A C E R Srl 8 MNPG52 04 SEZ F O E SE a He 7 Lt are y D ii di I me wee MEDIGAL DIVISION Label details Label 1 e MAGNETOTHERAPY MODEL MAG2000 SN 000001 EXT POWER SUPPLY UE24WCP 150120SPA output 15VDC 1 2A O E ee 21304 P2 1 I T EGR LA C E R Sri via S Pertini 24 4 30031 Martellago VE ITALY Label 2 Label 3 Label 4 DC 15V 1 2 A ON OFF CH1 CH2 UE24WCP 150120SPA Symbols description Attention consult operating instructions Product subject to WEEE regulations concerning separate waste collection of electronic equipment O Class Il equipment Applied part type BF T 476 Compliance with Directive 93 42 EEC and following modifications I A C E R Srl 9 MNPG52 04 ale MEDICAL DIVISION Admission temperatures FA Relative humidity _ al Fabricant information iS Protection level against solids dusts and liquids entrance Kit contents MAG2000 contains e N MAG2000 device e N power supply e N user and maintenance manual e N elastic therapeutic belt with 3 solenoids J N 1 carriage bag Moreover solenoids couple and magnetotherapy carpet are available as accessories Visit website www iacer it to obtain more information How to use Wamings Please read careful
11. at any time by NA pressing once key screen will display the base setting step 6 By I pressing again key screen will display programs menu step 5 Attention MAG2000 recognizes applicators correct connection During the treatment screen displays connection state under the magnet icon The presence of symbol V near to the channel number 1 or 2 indicates correct connection and applicator recognition The symbol X near to the channel number 1 or 2 indicates a not correct connection of applicator or its absence or even its incorrect functioning see Functioning control paragraph I A C E R Srl 20 MNPG52 04 Stored programs list Pre adjusted values Recommended values Pathology Time Treatment Treatment Hours cycles interval Osteoporosis 50 2 6 30 24hours 2 larthrosis 15 2 6 20 24hours S arthritis 30 2 6 20 24hours amp Cervicalarthrosis 5 2 6 15 Articular pain 25 Cervicalgias T Sprains 50 amp Fractures 50 2 6 9 Epicondyitis 45 2 6 2 1 Epitrocieitis 40 2 6 2 Intercostal contusions Lumbalgy Lumbar pain 14 Shoulder arthrosis 30_ Knee arthrosis Periarthritis 17 coxarthrosis 50 2 6 20 24hours 18 Muscular atrophy 35 2 6 20 24hours Ci Sl contracture 20 osteonecrosis 50 2 6 20 24hours Treat 1 Hz 4 Treat 3 Hz Treat 5 Hz 24 Treat 10Hz 10 15 Treat 15 Hz I A C
12. chosen language by using the A buttons Press OK key to confirm your selections I A C E R Srl 22 MNPG52 04 a n kO MEDICAL DIVISION Maintenance Functioning control Mag2000 equipment offers a magnet in order to control the device functioning Control procedure 1 switch on the device in according to user manual safety prescriptions 2 starta treatment in according to user manual instructions 3 get the magnet and put it near to applicator 4 check magnet vibration it will be proportional to selected treatment frequency Please contact the manufacturer in case of magnet vibration absence Cleaning Clean the equipment from the dust using a dry soft cloth Resistant stains can be removed using a sponge soaked in solution of water and alcohol 20 We recommend to disconnect the applicator from the device before cleaning the elastic therapeutic belt with 3 solenoids or the circular cases of professional solenoids couple e Extract the cable with 3 solenoids by removing the 2 silver studs through a screwdriver or open the circular cases through lateral ZIp e Clean the tissue using water and mild soap and wait for the complete drying before reconnecting the applicators ATTENTION always respect the applicators polarity paying attention to insert the bobbins with the side indicated by symbol turned to the green side of elastic belt therapeutic side I A C E R Srl 23 MNPG52 04 Piro MEDICAL DIVISION
13. end the treatment and consult a doctor Do not use on irritated skin or on open wound e Do not connect the device and its accessories to other devices not mentioned in the present user manual e Keep it out of children and animal s reach Avoid the device and accessories exposition to excessive direct light and dust Make reference to Maintenance paragraph e In case of prolonged treatments even up to 8 hours we suggest to use intensity lower than 50 In this case the efficacy is given by the prolonged treatment time rather than maximum adjustable intensity e High magnetic field intensity higher than 80 100 Gauss are recommended for short treatments up to 2 hours or in presence of rigid bandages braces WARNING Disconnect the power supply from the main after each treatment The manufacturer is considered responsible of the performances reliability and safety of the instrument only if I A C E R Srl 11 MNPG52 04 a a gE ir We Ag my ME ia 4 a oe is m Han B i an I 5 Il ig e Possible additions changes and or reparations are effected from authorized personnel e The electric plant of the environment in which MAG2000 is inserted it is conforming to the national laws e The instrument is employed in conformity to the instructions contained in this manual Electromagnetic interference The device doesn t produce and doesnt receive interferences from other equipments However it s opportune to use the instr
14. ly the user manual before using MAG2000 We recommend to visit magnetotherapy section on website www itechmedicaldivision com in order to obtain other information Take care of what follows I A C E R Srl 10 MNPG52 04 m mC MEDICAL DIVISION e Take care of position and meaning of the labels on MAG2000 Do not damage the therapeutic belt acting on connection cable and avoid to roll up the cable around the belt or the device e Check the integrity of the power supply before use Avoid the use in case of damage to the case or to the wire e Avoid the use of MAG2000 to people not educated through the reading of the manual e Avoid the use of MAG2000 contemporary with ointments containing free ions of magnetic metals e Do not use the device in damp environments and or in presence of inflammable agents e Do not wear metallic objects during therapy e Take care to place the green side of the therapeutic belt on the skin e Use only cables and applicators supplied by the Manufacturer Inadequate cables and applicators could damage the device and or could be hazardous for the patient e The user must periodically verify cables and applicators insulation and control their integrity eventually by contacting the manufacturer e Pay attention to use belt connexion cables and power supply strangulation danger e The materials used for the device production had passed the toxicity regulations In case of allergic reactions please susp
15. ns Cispr 11 Harmonic emissions IEC 61000 3 2 Voltage fluctuations flicker emissions IEC 61000 3 3 MAG 2000 uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment MAG 2000 is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes The MAG 2000 is suitable for use in all establishments other than domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Electromagnetic immunity MAG 2000 is intended for use in the electromagnetic environment specified below The customer or the user of MAG 2000 should assure that is used in such environment Immunity test Test level EN Compliance level 60601 1 2 6kV contact 8kV air Electrostatic discharge ESD EN 61000 4 2 6kV contact 8kV air Electrical fast transient burst IEC 61000 4 4 2kV for power supply lines I A C E R Srl 27 2kV per power supply lines Electromagnetic environment guidance Floors should be wood concrete or ceramic tile If floor are covered with synthetic material the relative humidity should be at least 30 Mains power quality should be at that of a typical commercial o
16. ogies in particular for the osteoporosys Pulsed electromagnetic fields induce biological modifications on biological membrane in order to re establish correct cellular functions According to different authors experiences in osteoporosys a considerable disease regression is evident from the sixth treatment and moreover it s evident an important increase of BMD Bone Mass Density The magnetic field high value Gauss generated by the device allows treatments in presence of braces or plaster bandage MAG 2000 Is a device able to satisfy home user needs thanks to its high magnetic intensity and its use facility I A C E R Srl 4 MNPG52 04 7 fa O MEDICAL DIVISION Technical Specifications Manufacturer I A C E R S r l Via S Pertini 24 a 30030 Martellago VE Tel 39 0415401356 Fax 39 0415402684 www itechmedicaldivision com lacer iacer it IACER S r l is an Italian medical devices manufacturer CE medical certificate n MED24021 issued by Cermet notified body n 0476 Declaration of conformity IACER S r l headquartered in Italy via S Pertini 24 A 30030 Martellago VE declares on its own responsibility that MAG2000 is manufactured in conformity with Directive 93 42 EEC MDD dated 14 June 1993 D Lgs 46 97 dated 24 February 1997 Attuazione della Direttiva 93 42 CEE concernente i dispositivi medici Annex Il as modified by Directive 2007 47 CE dated 5 September 2007 D Lgs 37 2010 dated 25
17. r hospital environment MNPG52 04 Impulses 1kV differential 1kV differential Mains power quality EN 61000 4 5 mode mode should be at that of a typical commercial or hospital environment Voltage dips short lt 5 Ur lt 5 Ur Mains power quality interruptions and gt 95 dips of Ur gt 95 dips of Ur should be at that of voltage variations on per 0 5 cycles per 0 5 cycles a typical commercial power supply input Or hospital lines 40 Ur 40 Ur environment IEC 61000 4 11 60 dips of Ur 60 dips of Ur If the user of the per 5 cycles per 5 cycles MAG2000 requires continued operation 70 Ur 70 Ur during power mains 30 dips of Ur 30 dips of Ur interruptions it is per 25 cycles per 25 cycles recommended that MAG2000 be lt 5 Ur lt 5 Ur powered from an gt 95 dips of Ur gt 95 dips of Ur uninterruptible per 5 seconds per 5 seconds power supply or a battery Mains power 3 A m electromagnetic field EN 61000 4 8 Mains power quality should be at that of a typical commercial Or hospital environment r f immunity MAG 2000 is intended for use in the electromagnetic environment specified below The customer or the user of MAG 2000 should assure that is used in such environment 60601 1 2 level environment guidance EN 61000 4 6 150kHz to 80MHz 150kHz to 80MHz I A C E R Srl 28 MNPG52 04 RF Radiata 3 Veff from 3 Veff from communications equipment EN 61000 4 3 80MHZ to 2 5GHz 80MH
18. restarts Attention it s possible to get out from the treatment at any time pressing once key screen will display the base setting step 6 By I pressing again key screen will display the programs menu step 5 Attention MAG2000 recognizes applicators correct connection During the treatment screen displays connection state under the magnet icon I A C E R Srl 14 MNPG52 04 I TEGN The presence of symbol Vaca to the channel number 1 or 2 is an indicator of correct connection and applicator recognition The symbol X near to the channel number 1 or 2 indicates a not correct connection of applicator or its absence or even its incorrect functioning see Functioning control paragraph Therapeutic belt and solenoids couple positioning Here below a list of main positions for the therapeutic belt and for the solenoids couple Wrap the belt around the area to be treated or position the belt on the area for example in vertebral column treatment During this phase take care to place the green side of the therapeutic belt on the skin The professional solenoids couple have to be placed in opposite positions on the area to be treated Also in this case take care to place the green side of the therapeutic belt on the skin Back lumbar I A C E R Srl 15 MNPG52 04 I TEGR Shoulder Cervical Tibia fibula Femur head Vertebral column Lumbar Suggestions for a correct use I A C E R Srl 16 MNPG52
19. ument keeping it at least at 3 meters from televisions monitors cellular telephones or any other electronic equipment The device has to be installed and work in compliance with the information about the electromagnetic compatibility mentioned in the user manual Please read the EMC Table paragraph The use of accessories transducer and cables different from the ones specified except for the transducers and cables supplied by the fabricant as spare parts for internal components can cause an emissions increase or a device immunity decrease The device should not be used next to or overlapped to other devices If it is necessary to use the device next to or overlapped to other devices it should be checked the normal operation Contraindications and side effects Patient in pregnancy tuberculosis juvenile diabetes viral in acute phase illnesses mycosis cardiopathic subjects tumours serious arrhythmias or pacemaker carriers children metallic prosthesis carriers acute infections epileptics different medical prescriptions excepted No significant side effects are known of nor are reported particular contraindications for excessive time length using the device Quick use of the device with presetied parameters We recommend to read the following instructions for an easy and quick use of MAG2000 I A C E R Srl 12 MNPG52 04 I T EGH MEDICAL DIVISION 6 Connect the applicator or the applicators to the device
20. wo years we suggest a complete check of device contact the manufacturer or locator dealer If you find any problems contact immediately the National Distributor or the manufacturer at the following address A C E R S r l Via S Pertini 24 a 30030 Martellago VE ITALY Tel 39 0415401356 Fax 39 0415402684 www itechmedicaldivision com iacer iacer it Assistance Every intervention on device must be performed by manufacturer For any assistance intervention contact the National Distributor or the manufacturer at the following address A C E R S r l Via S Pertini 24 a 30030 Martellago VE ITALY Tel 39 0415401356 Fax 39 0415402684 www itechmedicaldivision com iacer iacer it You can get any technical documentation on spare parts but only prior business authorization Spare parts For original spare parts contact the National Distributor or the manufacturer at following address I A C E R Srl 25 MNPG52 04 m MEDICAL DIVISION ILA C E R S r l Via S Pertini 24 a 30030 Martellago VE ITALY Tel 39 0415401356 Fax 39 0415402684 www itechmedicaldivision com iacer iacer it To preserve product warranty functionality and product safety we recommend to use only original spare parts I A C E R Srl 26 MNPG52 04 EMC tables Electromagnetic emission Compliance Electromagnetic environment guidance Class A Complies RF emissions Cispr 11 RF emissio
21. z to 2 5GHz should be used no closer to any part of MAG 2000 including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance d 1 2 VP 150kHz to 80MHz d 1 2 P 80 MHz to 800 MHz d 2 3 P 800 MHz to 2 5 GHz where P is the maximum output power rating of the transmitter in Watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m Field strangths from fixed RF transmitters are determined by an electromagnetic site survey should be less than the complicance level in each frequency rage Interference may occur in the vicinity of equipment marked with the Co following symbol A Recommended separation distances between portable and mobile communications equipment and the MAG 2000 MAG 2000 is intended for the use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of IMAG 2000 can help prevent electromagnetic interferences by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and MAG 2000 as recommended below according to the maximum output power of the communication equipment Rated maximum Separation distance according to the frequency of the power of the transmitter m transmitter W 150kHz to 80MHz 80MHz to 800MHz 800MHz to d 1 2 P d 1 2 P I