1. 10 Le gt 1 HCG801 E fm Page 10 Monday May 9 2005 6 49 PM 2 Inserting and Replacing Batteries Slide the battery cover in the direction of the arrow and lift the end of the cover upwards Note Do not pull too hard on the cover Insert two LRO3 AAA batteries as indicated Always use LRO3 AAA batteries Note Make sure that the polarity of the batteries is correct Close the battery cover Lower the battery cover 1 then while gently pressing the cover down slide it in the direction of the arrow 2 mal P WZ AIS 2 UN 2 HCG801 E fm Page 11 Monday May 9 2005 6 49 PM 2 Inserting and Replacing Batteries Battery life and replacement When the Please insert new batteries message appears replace both batteries with new ones e Always turn the unit off before replacing the batteries Dispose of the used batteries according to applicable local regulations Battery life Two new LRO3 AAA batteries will last for approximately 400 measurements If measurements are taken once a day at room temperature 22 C The batteries enclosed in the package are used for demonstration purpose It is possible these batteries will therefore not last for 400 measurements If the batteries are removed for more than one hour the settings and any ECG data sto
2. HCG801 E fm Page 1 Monday May 9 2005 6 49 PM OMRON Hcs s01 Portable cordless single channel ECG Monitor Model HCG 801 e Instruction Manual EN Thank you for purchasing the OMRON Portable ECG Monitor Before using this unit for the first time please be sure to read this instruction manual carefully and use the unit properly Please keep this instruction manual on hand for future reference OMRON assumes no responsibility for personal injury or damage A es sustained by or through use of this product er D HCG801 E fm Page 2 Monday May 9 2005 6 49 PM Contents Before Using the Unit Introduetion 255 ai 3 Notes ON Safety cece cgessneetceretastececcevevcaecectsanetenecestneyccetesseieetecestdueccsesseteeesteveseere 4 T KinnowsYour Unit era niet anni cele ties 6 Operating Instructions 2 Inserting and Replacing Batteries 10 3 Setting the Date and Time sms 13 4 Using SD Memory Card ss sssssnennnnennennneeennnnes 15 4 1 Inserting and Ejecting SD Memory Card 15 4 2 Formatting an SD Memory Card 2 2222 ae aan 19 5 Important Notes on Taking a Measurement 21 6 Taking a Measurement 7 Displaying ECG Data enter 8 Adjusting the Settings 8 17 Basic Op ration Ativan aaa 8 2 Date and Time 8 4 Brightness en 8 5 Contrast ns 8 6 Language 8 7 Date Format 8
3. oe 07 oe on Back to front Upside down Back to front and upside down Insert the card until it clicks into position Write Protection Locking the SD Memory Card e Make sure that the write protect switch is not locked before inserting the SD memory card If the write protect switch is locked down you will not be able to store new ECG data on the SD memory card Caution When inserting the SD memory card do not try to force the card into the slot if it is upside down back to front etc This could damage the card or the unit itself P WZ AIS J a A HCG801 E fm Page 17 Monday May 9 2005 6 49 PM 4 Using SD Memory Card 3 Close the SD memory card slot cover OMRON ucs s01 17 Z D V UN ZU 1 nil D 1 HCG801 E fm Page 18 Monday May 9 2005 6 49 PM 4 Using SD Memory Card To eject the SD memory card Check that the device is turned off Caution Only eject an SD memory card when the unit is turned off Open the SD memory card slot cover and press the memory card The lock is released and the card is ejected slightly D OMRON ucs s01 Remove the memory card me DO Close the SD memory card slot cover P WZ AIS gt HCG801 E fm Page 19 Monday May 9 2005 6 49 PM 4 Using SD Memory Card 4 2
4. HCG801 E fm Page 44 Monday May 9 2005 6 49 PM 9 Managing ECG Data 9 3 Copying data from the SD memory card to a PC E Copying data from the SD memory card to a PC 1 Please follow these instructions to copy data from the SD memory card to a PC Make sure the PC is equipped with an SD memory card port or connect a commercially available SD memory card reader writer to your PC Create a new folder on your PC where you want to save a copy of the ECG data Insert the SD memory card and browse to the SD memory card data as explained in the instruction manual of your PC and or software The ECG data is located in the hcg801 data folder where stands for the character of a removable disk Select all files and copy them in the folder you have created earlier 2 When copying is finished remove the SD memory card The SD memory card can now be formatted as described in 4 2 Formatting an SD Memory Card on p 19 if you wish to delete the data ig I 2 Als 5 HCG801 E fm Page 45 Monday May 9 2005 6 49 PM 9 Managing ECG Data Copying data from a PC to the SD memory card Please follow these instructions to copy data from a PC to the SD memory card 1 Make sure the PC is equipped with an SD memory card port or connect a commercially available SD memory card reader writer to your PC 2 Insert the SD memory card in the SD port of yo
5. taken while the SD memory card was not inserted Note The unit s internal memory can only store a maximum of five measurement results 1 Press the Y D button and select Data copy to SD E Date format EMessase 0verwrite mErase all data Data copy to 5 2 Press the OK button The conformation message Copy data into SD is displayed E EI l Data copy to SD Copy data into SD 3 Press the W C button to select YES Notes If the SD memory card is full the Overwrite data message may appear Refer to p 38 for details By default NO is highlighted ECG data stored on an SD memory card can be damaged by static electricity and other causes Copy data to PC or other device for additional safety ae 1 a ce I HCG801 E fm Page 43 Monday May 9 2005 6 49 PM Press the OK button The ECG data is copied to the SD memory card Copying Caution SD memory Never remove the SD memory card while data is being DO NOT REMOVE SD copied to it If the SD memory card is removed while data is being copied the data may be lost or damaged Notes The copied data is displayed and the screen from step 1 appears If the same ECG data already exists on the SD memory card the data is not copied 43 i a a ig 44 I 2 as
6. Press START to resume the measurement The indicated number of ECG data has been copied to an SD memory card The indicated number of ECG data has been erased No action required Wait until the message disappears 1 Number of readings from 0 to 5 The number displayed represents the amount of copied ECG data 2 Number of readings from 1 to 300 The number displayed represents the amount of erased ECG data 1 D de ils 5 HOG801 E fm Page 47 Monday May 9 2005 6 49 PM I Warning Messages 10 Care and Maintenance Formatting SD memory DO NOT REMOVE SD The SD memory card is being formatted Wait until the message disappears Do not remove the SD memory card until formatting has been completed Storing on SD memory DO NOT REMOVE SD ECG data is being written saved to the SD memory card Wait until the message disappears Do not remove the SD memory card until data storage has been completed Copying SD memory DO NOT REMOVE SD ECG data is being copied to the SD memory card Wait until the message disappears Do not remove the SD memory card until data copying has been completed Data in device memory is being erased ECG data stored on the unit is being erased Wait until the message disappears Data in SD is being erased DO NOT REMOVE SD until erasing finishes ECG data stored on the SD memory card is
7. formaldehyde high density ozone etc This device does not require calibration during the expected life cycle of five years aa 2 UND han HCG801 E fm Page 53 Monday May 9 2005 6 49 PM 12 Technical Data Model Induction system Heart rate range Accuracy Enclosure ingress protection Display Memory Electrical Safety Power Supply Battery life Operating Temperature and Humidity Storage Temperature and Humidity Weight Dimensions Attachments Note Specifications may be changed without prior notice e Disposal of this product and used batteries should be carried out in accordance with the national regulations for the disposal of electronic products l CE0197 This device supports SDA standard SD memory cards OMRON Healthcare Inc is a member of the SD Card Association HCG 801 Bipolar single channel 2 to 200 beats min Bandwidth 0 05 to 40Hz Sampling Rate 125Hz IP20 Graphic LCD 5 measurements using the unit s internal memory 300 measurements using the supplied SD memory card Internally powered equipment BF type device 2 LRO3 AAA batteries Approximately 400 measurements with alkaline batteries When measurements are taken once a day at room temperature 22 C 10 to 40 C 30 to 85 RH 20 to 60 C 10 to 95 RH Approximately 130g including batteries 121 mm W x 67 mm H x 24 mm D Batteries 2 x LRO3 AAA SD
8. tie D HCG801 E fm Page 26 Monday May 9 2005 6 49 PM 6 Taking a Measurement Common Causes of Inaccurate Measurement Results The following situations can result in inaccurate measurement results or errors e The index finger is not placed e The unit is held incorrectly closely on the electrode e The unit is being held in the left hand Always use your right hand to hold the unit when taking a measurement The rear of the unit is placed The chest electrode is against the against the chest finger and the finger electrode is against the chest WZ AD 2 as HCG801 E fm Page 27 Monday May 9 2005 6 49 PM 6 Taking a Measurement Measurement Result There are 13 messages regarding the measurement result A letter of the alphabet indicates each of the 13 possible messages This letter is displayed together with the time and date of the measurement when the GRAPH menu is displayed Important Do not interpret measurement results yourself consult your doctor Stable waveform Fast heart rate Fast and irregular heart rate Fast heart rate and deviating waveform Fast and irregular heart rate deviating waveform Slow heart rate Slow and irregular heart rate Slow heart rate and deviating waveform Slow and irregular heart rate deviating waveform Irregular heart rate Irregular heart rate and deviating waveform Deviating wav
9. 2005 15 00 ie 5 1 05 15 00 4 1 05 15 00 34 1 05 15 00 27 1 05 15 00 1 1 05 15 00 Each stored ECG data is displayed by date time and a character representing the measurement result refer to p 27 for the overview of character and measurement result tie 2 2 Als gt HCG801 E fm Page 29 Monday May 9 2005 6 49 PM 7 Displaying ECG Data ia Press the Y C gt button to move down the list and the lt J amp button to move up The most recent measurement result is displayed at the top of the list and is automatically highlighted with any previous measurement results listed below it in chronological order When the SD memory card is in the SD memory card slot the list of ECG data stored on the card is displayed If there is no SD memory card in the card slot the ECG data stored in unit memory is displayed 3 Ifyou press the Y C gt button you can scroll through Bf 1 2005 15 00 Es fr the ECG waveform Press the lt J amp button to return to the previous waveform display Two waveforms are displayed the lower smaller waveform being a view of the whole 30 second measurement At the bottom of the screen is a bar indicating the progress of the measurement The current heart rate is shown in the lower right corner of the screen The shaded area in the progress bar indicates the section of the ECG waveform being displayed 4 Press the OK
10. 8 Messages 8 9 MOVENNTITE are ri amer era ns dd Hr da re n on de pe ER es den een Eee 9 Managing ECG Data nuunsssnreennnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnsnnnnnnnnnannnnnn 9 1 nErasingsE CG Date a Ri 9 2 Copying Data to an SD Memory Card 9 3 Copying data from the SD memory card to a PC Care and Maintenance 10 Care and Maintenance davnwttessisssotasisssanneaucaviteaseGeosaevaeiuenseanans 46 10 1 Lists and Explanations of Messages 46 10 2 Troubleshooting f 2 4 stan EEE a we ai Aletsch eee er nn En et 50 11 Maintenance and Storage ss snnnns 52 12 Technical Data iaei saisit aaa aaraa aian aseina Tenian avail 53 13 General Information About the Heart and ECG Measurements 54 ha E a gt D HCG801 E fm Page 3 Monday May 9 2005 6 49 PM Introduction Your OMRON HCG 801 is an easy to use electrocardiograph unit that can record and store electrocardiogram ECG measurements of your heart rhythm Each ECG reading records about 30 second measurement and these ECG readings can help your doctor monitor your condition It also includes guidance screens to help you obtain good results when taking measurements The monitor is supplied with an SD memory card that can store up to 300 measurements which include the ECG data along with the date and time of measurement Storing ECG data on the SD m
11. Formatting an SD Memory Card You will need to format the SD memory card before you use it for the first time and in the following circumstances e If you use an SD memory card that has been used by other devices e If a memory card error occurs Notes When you format an SD memory card any data stored on the card is erased Always copy any data you wish to keep to a computer before formatting an SD memory card When you format an SD memory card format it with the unit Do not format an SD memory card with a computer 1 Press the MENU button press the Y button to select SET menu then press the OK button GRAPH men 2 Press the W C gt button to scroll through the settings and select Formatting SD Formatting SD ce gt HCG801 E fm Page 20 Monday May 9 2005 6 49 PM 4 Using SD Memory Card 20 I 3 Press the OK button The message All data in SD will be deleted Format SD is displayed Press the Y C gt button to select YES then press the OK button If you press the OK button while NO is selected the SD memory card will not be formatted E E E Formatting SD All data in SD will be deleted Format SD By default NO is high lighted Press the OK button A message warning you not to remove the SD memory card while it is being formatted is displayed and the SD memory card
12. SD 31 12 05 13 30 Adjust the settings then press the OK button to confirm the changes The SET menu can be used to adjust the following settings or manage ECG data stored on the SD memory card or unit Date and Time Volume Brightness Contrast Language Date Format Message Overwrite Erase all data Data copy to SD Formatting SD Notes Refer to 9 Managing ECG Data on p 39 for details on Erase all data and Data copy to SD Refer to 4 Using SD Memory Card on p 15 for details on Formatting SD To return to the menu screen from the SET menu screens press the MENU button To return to the previous screen highlight Back then press the OK button or MENU button EN e If you press the MENU button without first pressing the OK button after making the changes to a setting the changes to the setting will not be implemented 31 re a ce HCG801 E fm Page 32 Monday May 9 2005 6 49 PM 8 Adjusting the Settings 8 2 Date and Time Select this item if you need to reset the date and time for some reason For details on setting the date and time refer to p 13 Note You will be prompted to set the date and time when you first insert the batteries Volume ay The current volume level is displayed The mark indicates the at the bottom of the screen volume level The default setting is the maximum volume level Press the l
13. is formatted Formatting can take up to eight seconds to complete Caution Never remove an SD memory card while it is being formatted Formatting SD memory DO NOT REMOVE SD a tie D I HCG801 E fm Page 21 Monday May 9 2005 6 49 PM 5 Important Notes on Taking a Measurement In order to obtain a good ECG reading it is important that measurements are taken correctly Before taking a measurement for the first time read these instructions carefully and make sure you observe the following points each time you take a measurement e Make sure that the electrodes are directly touching your skin If your hands or skin are dry wipe them with a damp towel so that they are slightly moist e If the electrodes are dirty wipe any dirt off with a soft cloth moistened with disinfectant alcohol or a cotton swab Take care not to use too much disinfectant alcohol when cleaning the electrodes Do not wipe or get disinfectant alcohol on parts other than the electrodes Do not move during measurement Movement including talking coughing or sneezing during measurement can affect the measurement results If the position of your finger or arm is inappropriate during measurement the results may be unreliable so be sure to check their position before taking a measurement Refer to p 22 and p 23 After taking a measurement disinfect the electrodes by wip
14. memory card Storage pouch Instruction manual Quick reference card Registration card Guarantee card Lx e Uses bitmap font from Ricoh Co Ltd e The SD logo is a registered trademark e This unit confirms to EMC Standard IEC 60601 1 2 2001 zn zu 2 p y 2 13 General Information About the Heart and ECG Measurements The heart is a muscular pump controlled by electrical impulses generated by the body It is divided into two by the septa and each side has two chambers an atrium and a ventricle linked by a one way valve The left atrium and ventricle control oxygenated blood and the right atrium and ventricle de oxygenated used blood The electrical impulse that causes the heart to beat spreads across the atria causing the left and right atrium to contract and pump blood into the left and right ventricles respectively The two ventricles then contract and pump blood out of the heart The heart muscle then relaxes or re polarises allowing blood to fill up the heart again F atrium Right atrium Left ventricle Right ventricle Septa 7 An ECG monitor is able to measure the electrical impulse as it passes across and through the heart causing the heart to beat An ECG does not measure the movement of your heartbeat but rather the electrical activity that causes the heart to beat The measurements recorded by an ECG monitor when combined with a medical examination c
15. your chest when taking a measurement Displays the wave pattern after measurement and displays menus and setting options Displays or closes the menu screen Moves the desired menu or wave pattern to the left or up and decreases the number of the selected field VD button Moves the desired menu or wave pattern to the right or down and increases the number of the selected field OK button Selects the highlighted menu option or ECG data Clears error messages Battery cover Open this cover to insert batteries Finger electrodes Place your right index finger over these when taking a measurement START button Starts measurement SD memory card slot cover Open this cover to insert the SD memory card mal P WZ AIS amp HCG801 E fm Page 7 Monday May 9 2005 6 49 PM amp 1 Know Your Unit a Replacements Parts L SD memory card M Storage pouch WZ AD gt ou 7 a HCG801 E fm Page 8 Monday May 9 2005 6 49 PM 1 Know Your Unit Display The screen is used to display the guidance screens and results of ECG readings adjust settings and manage ECG data stored on the SD memory card or in unit memory Menus gt ESGRAPH meni SET_menu Menu screen The main menu screen is displayed when you press the MENU button GRAPH menu This allows you to view ECG readings s
16. 5 Press the D button for two seconds to turn off the power Press for two seconds The unit beeps then the power is turned off Note If you forget to turn off the power the unit will turn itself off automatically after five minutes P WZ AD ig 2 as HCG801 E fm Page 15 Monday May 9 2005 6 49 PM 4 Using SD Memory Card The unit includes an SD memory card for storing ECG data which can then be viewed by your doctor later This section describes how to insert and eject an SD memory card and how to format a card so it can be used to store ECG data Notes lf you need to purchase a new SD memory card consult your Omron Customer Service Department Tel 0870 7502771 Do not store data other than ECG data from the unit on the card 4 1 Inserting and Ejecting SD Memory Card The SD E3 mark is displayed in the top right corner of the screen while the SD memory card is inserted and the unit is turned on Inserting the SD memory card 1 Check that the unit is turned off Caution Never insert an SD memory card while the unit is turned on o HCG801 E fm Page 16 Monday May 9 2005 6 49 PM 4 Using SD Memory Card 2 Open the SD memory card slot cover and insert the SD memory card into the slot Notes OMRON Hcc 801 e Check that the card is correctly aligned and insert it into the slot as shown E A a
17. an help your doctor monitor your heart condition The ECG measurements recorded by the unit are NOT designed or intended for medical diagnosis Conditions such as arrhythmia and Ischemia can only be diagnosed by a doctor through a special examination HCG801 E fm Page 54 Monday May 9 2005 6 49 PM aa 2 UND U HCG801 E fm Page 55 Monday May 9 2005 6 49 PM 13 General Information About the Heart and ECG Measurements a About the ECG Waveform The ECG waveform shows the rhythm of your heartbeat during the 30 second measurement and displays the electrical activity causing the heart to beat The waveform for each heartbeat shows the progress of the electrical impulse across and through the heart Ventrcular contraction N QRS complex Atrial contraction P wave NAN Re polarization T wave The first peak indicates the spread of the impulse over the atria and the beginning of their contraction This is known as the P wave The second peak indicates the spread of the impulse over the ventricles and the beginning of their contraction This is known as the QRS complex The third peak indicates the activity as the heart relaxes re polarization and is known as the T wave What is Arrhythmia Arrhythmia is a condition where the heartbeat rhythm is abnormal due to flaws in the bio electrical system that drives the heartbeat Typical symptoms are skipped h
18. ata is erased and the screen from step 1 appears Note If you press the OK button after selecting NO the screen from step 1 appears Erasing Selected ECG Data Use the GRAPH menu if you only want to delete a selected set of ECG data For details on displaying the GRAPH menu follow step 1 in 7 Displaying ECG Data on p 28 1 Press the Y D button to highlight the data you want to erase 2 Press the MENU button The menu screen is displayed 3 Press the W C gt button and select Erase data 40 Ie T 2 UN 2 HCG801 E fm Page 41 Monday May 9 2005 6 49 PM 4 Press the OK button E S 1 2005 15 00 A 9 Managing ECG Data The confirmation message Erase selected data is displayed rase selected data Press the Y C gt button and select YES then press the OK button The ECG data is erased and the screen from step 1 i highlighted appears Note If you press the OK button after selecting NO the screen from step 1 appears Important Never remove an SD memory card while data is being erased 41 ce HCG801 E fm Page 42 Monday May 9 2005 6 49 PM 9 Managing ECG Data 9 2 Copying Data to an SD Memory Card You can copy ECG data from the unit s internal memory to an SD memory card This allows you to copy data to the SD memory card if measurements have been
19. ay The current setting is displayed at the bottom of the screen The default setting is Display 1 Press the lt J amp button or Y C gt button to highlight the Display or Hide then press the OK button to confirm your selection Note Even if Hide is selected results messages are saved by the unit 37 HCG801 E fm Page 37 Monday May 9 2005 6 49 PM iz a ce 8 Adjusting the Settings 8 9 Overwrite The unit will overwrite the oldest stored reading when the SD memory card or unit memory is full A message will warn that this will happen you can choose to hide or display this message The default setting is to hide the message E ES Overwrite E Date format amp Hessage Display The current setting is displayed at the bottom of the screen 1 Press the lt J amp button or Y C gt button to highlight Display or Hide then press the OK button to confirm your selection Caution While Hide is selected the unit will overwrite erase and replace the oldest measurement data with the results of the new measurement without displaying a message when the SD memory card or unit memory are full About the Overwrite message If you select Display the unit will display a message before you take a measurement asking if you want to overwrite existing data with the new measurement data Select Yes to overwrite the oldest ECG data with the results o
20. being erased Wait until the message disappears Do not remove the SD memory card until data erasing has been completed Reading data DO NOT REMOVE SD The unit is loading ECG data stored on the SD memory card Wait until the message disappears Do not remove the SD memory card until the data has been loaded by the unit No SD is inserted There is no SD memory card in the card slot Turn off the power and insert the SD memory card Never insert the SD memory card while the power is turned on 47 Ti gt UND a HCG801 E fm Page 48 Monday May 9 2005 6 49 PM 10 Care and Maintenance 48 I The unit is unable to access data SD is write protected because the SD memory card inserted is write protected Refer to the instruction supplied with the memory card and insert the SD memory card with write protection turned off The battery power has run low No saved data and the ECG data stored on the unit has disappeared Insert new batteries and reset the Date and Time and other settings Any ECG data that was stored on the unit cannot be recovered The battery power has run low Set clock and the Date and Time setting on the unit has disappeared Insert new batteries and reset the Date and Time from the SET menu The battery power has run low No saved data and the ECG data stored on the Set clock unit has disappeared or the c
21. button again to view the measurement f 5 1 2008 15 00 E WA comments and heart rate Stable waveform Note You can specify whether a result message is displayed or not Refer to p 37 for details Heart rate 60 nin 5 Press the OK button once more to return to the GRAPH menu list re ce gt HCG801 E fm Page 30 Monday May 9 2005 6 49 PM 8 Adjusting the Settings 8 1 30 I You can use the SET menu to adjust various settings You can press the START button to begin measurement even while the SET menu is displayed Basic Operation The following procedure describes how to select and adjust a setting Refer to the following pages for details of the options available for each setting Press the MENU button while the unit is on to display the menu screen eS GRAPH men Initially GRA menu is highlighted Press the Y C gt button to select SET menu Press the OK button Initially the Date and Time setting is highlighted E a 2 2 Als 5 HCG801 E fm Page 31 Monday May 9 2005 6 49 PM 8 Adjusting the Settings ia Press the lt J 4 or WD gt button to select the desired setting then press the OK button to see the options for that setting Date and Time Volume Brightness Cont rast Language 31 12 05 13 30 lt aOverwrite mErase all data Data copy to
22. eartbeats premature contraction an abnormally rapid tachycardia or slow bradycardia pulse This can be caused by heart disease ageing physical EN predisposition stress lack of sleep fatigue etc Arrhythmia can only be diagnosed by a doctor through a special examination What is Ischemia Ischemia is a condition in which insufficient oxygen is supplied to parts of the heart or other parts of the body This is usually due to a blockage or partial blockage of an artery Ischemia can only be diagnosed by a doctor through a special examination 55 zn zu 2 Sp y 7 gt HCG801 E fm Page 56 Monday May 9 2005 6 49 PM Customer services OMRON HEALTHCARE UKLTD Opal Drive Fox Milne Milton Keynes MK 15 0DG United Kingdom Omron Customer Service Department Tel 0870 7502771 OMRON Medizintechnik Handelsgesellschaft m b H Windeckstra e 81a D 68163 Mannheim Germany OMRON SANTE FRANCE 14 rue de Lisbonne F 93561 Rosny sous Bois Cedex France Manufacturer OMRON HEALTHCARE CO LTD 24 Yamanouchi Yamanoshita cho Ukyo ku Kyoto 615 0084 Japan EU representative OMRON HEALTHCARE EUROPE BV Kruisweg 577 2132 NA Hoofddorp The Netherlands www omron healthcare com ha ee IM HCG 801 1 04 05 1647415 5A a UND
23. eform Analysis impossible Please measure again 3 x sa a o a ao o v Recommendation on measurement result In addition to the above messages the device will display the heart rate in beats per minute Finally the device will display one of the following messages ER Store data Yes No This message will be displayed when the unit can find none or only sleight deviations from normal This data can be saved by selecting Yes and pressing the OK button Data is stored This message will be displayed when the unit can find moderate deviation from normal The unit will automatically store the data for future analysis by your doctor Data is stored Please show your Dr This message will be displayed when the unit can find significant deviation from normal The unit will automatically store the data for future analysis by your doctor 27 al gt HCG801 E fm Page 28 Monday May 9 2005 6 49 PM 7 Displaying ECG Data The GRAPH menu is used to display the list of ECG data stored on the SD memory card or in unit memory 1 Press the MENU button while the unit is turned on the GRAPH menu is highlighted If you then press the OK button the list of ECG data is displayed 2 Selectthe ECG data you wish to view from the GRAPH menu list then press the OK button to display the detailed view of the waveform rs El 5 1
24. emory card or in unit memory allows you or your doctor to view them later Important Please note that while the measurements from an ECG monitor are a useful guide for your doctor they cannot detect all changes in heart conditions Always consult your doctor if you notice any change in your condition regardless of the measurement results Consult a doctor even if Stable waveform is displayed as the measurement result if you have been diagnosed with a heart condition The illustrations used in this manual may differ slightly from the appearance of the actual product gt HCG801 E fm Page 4 Monday May 9 2005 6 49 PM Notes on Safety This unit is designed for monitoring of the ECG waveform It is not designed or intended for medical diagnosis Never use the unit in critical situations Self diagnosis and treatment is dangerous The measurement results are for monitoring purposes only and should never be used as a basis for starting or modifying treatment without independent confirmation by medical examination The values displayed by the unit are the values at the time of measurement Medical conditions can change suddenly If you notice any change in your condition consult your doctor regardless of the measurement results Warning Do not use with a cardiac pacemaker Do not use with a defibrillator Do not use in the presence of flammable anest
25. ens when you press the D button The batteries are worn out The batteries are inserted incorrectly Replace both batteries with new ones Insert the batteries with their polarities correctly aligned Measurement starts but the message Analysis impossible Please measure again appears The measurement data could not be analysed Check the instructions and repeat measurement Measurement finishes but no result is shown The electrodes are not making good contact with your finger or chest Check the instructions and repeat measurement Your body is too tense Relax and repeat the measurement Electrode contact was broken before 30 seconds of data has been recorded Make sure that your finger and chest maintain contact with the electrodes for the full 30 seconds of the measurement The chest electrode has been applied over clothing Apply the chest electrode to bare skin Your skin is dry Moisten your skin with a damp towel The ECG signal is too weak Make sure that the electrodes are applied correctly However for some individuals measurement may not be possible mal S Z y ECG data has disappeared HCG801 E fm Page 51 Monday May 9 2005 6 49 PM The batteries are worn out or have been removed for over one hour 10 Care and Maintenance The unit will not be able to store data if left with worn batteri
26. es for an extended period Replace the batteries with new ones as soon as possible after the warning message appears ECG data has disappeared from an SD memory card The card was exposed to a strong electromagnetic field electric noise or static damaging the data stored on the card Or the card is defective Erased data cannot be recovered Format the SD memory card again Other problems Turn off the unit and try again If the data and time or other settings have been reset to their default values set them again If the problem still occurs consult your Omron Customer Service Department Tel 0870 7502771 51 1 gt U HCG801 E fm Page 52 Monday May 9 2005 6 49 PM 11 Maintenance and Storage e Clean the unit with a cloth lightly dampened with water disinfectant alcohol or detergent then wipe it dry with a dry cloth Do not wipe the SD memory card slot cover Wipe the electrodes using a cloth dampened with disinfectant alcohol Do not wipe the unit with benzene gasoline paint thinner concentrated alcohol or other volatile detergents Do not place objects on top of the unit This could damage the unit Do not disassemble the unit There are no user serviceable parts Repairs should only be carried out by authorized personnel Do not sterilise this unit in an autoclave ultraviolet steriliser or gas steriliser EOG
27. es with their polarities reversed The chest electrode will be impeded by excessive body hair regularly remove hair at the electrode contact point Omron does not recommend the use of contact gel General e Avoid extremes in temperature and humidity Do not use this unit in locations subject to high or low temperatures or humidity Use at a temperature within 10 to 40 C and between 30 and 85 RH Do not sterilise this unit in an autoclave or gas steriliser EOG formaldehyde high density ozone etc EN Do not wash this unit with water Do not store the unit in the following ambient conditions Locations exposed to direct sunlight Locations subject to high temperatures and high humidity Wet or damp locations where water may get on the unit Dusty locations Near fires or open flames Locations exposed to strong vibration Locations exposed to strong electromagnetic fields gt amp iii as HCG801 E fm Page 6 Monday May 9 2005 6 49 PM a gt DQ Power Cancel button Chest electrode Display MENU button lt J button Know Your Unit Main Unit L al L al gt Turns the power on off and cancels measurement Place this against
28. f the new measurement If you select No measurement is cancelled Ril HCG801 E fm Page 38 Monday May 9 2005 6 49 PM a a ce HCG801 E fm Page 39 Monday May 9 2005 6 49 PM 9 Managing ECG Data 9 1 Erasing ECG Data Erasing All ECG Data You can use the SET menu to erase all the ECG data stored on the SD memory card or in unit memory Note If an SD memory card is inserted in the unit all data on erased the card will be 1 Press the W C gt button to scroll through the settings and select Erase all data Date format amp Hessage lt aOverwrite Data copy to SD 2 Press the OK button The message Erase all data in SD memory is displayed Note If the SD memory card is not inserted and measurements are being stored in unit memory the message Erase all data in the device is B E mErase all data Erase all data in SD memory displayed By default NO is highlighted 39 1 a a 2 lt HCG801 E fm Page 40 Monday May 9 2005 6 49 PM 9 Managing ECG Data E 3 Press the Y C gt button and select YES then press the OK E ET button mErase all data Erase 5 data The number of sets of data to be deleted is displayed 4 Press the Y C gt button and select YES then press the OK button Allthe ECG d
29. g a Measurement 4 Place the chest electrode on bare skin about 5 cm below your left nipple Notes Make sure that you are relaxed and that your chest is free of tension Press the chest electrode close against your bare chest Do not place the electrode on clothing as this will result in an inaccurate measurement Chest electrode About 5 cm Press closely to your chest Female Locate the ECG monitor at the lower end of the breastbone and move horizontally to the left to the middle of the left part of the chest If necessary gently lift the breast and place the Chest Electrode under the left breast Unless the chest electrode comes in contact with the brassiere it is not necessary to remove it if you are unsure please consult your doctor Q WZ AD Le gt 6 Taking a Measurement 5 While maintaining the same posture press the START button Measurement starts During measurement Measurement ends The unit beeps every second TT 2 s About 30 seconds Am 6j 1 2005 15 00 A The measurement ends automatically indicated by 4 rapid beeps The measurement takes about 30 seconds to complete The unit beeps every second during measurement During measurement the screen displays the ECG waveform and the progress bar at the bottom of the screen displays measurement progress and a guide to the heart rate Depending on
30. hetics drugs or pressurised oxygen such as in a hyperbaric chamber ultraviolet steriliser or oxygen tent Do not attempt self diagnosis of the measurement results and analysis always consult your doctor Self diagnosis may lead to a deterioration in your condition e Do not use on people with sensitive skin or allergies This may result in exanthema skin irritation or eczema Keep out of reach of small children and people who can not express their consent Do not use for any purpose other than obtaining an electrocardiograph Do not expose the unit to strong shocks or vibrations or drop or step on the unit Do not use batteries of a type other than that specified in this manual Do not apply over clothing or touch the electrodes with your left hand f D I HCG801 E fm Page 5 Monday May 9 2005 6 49 PM Notes on Safety E Do not take measurements when there are drops of water on your skin such as from sweat or after bathing Do not take measurements where the unit will be exposed to strong electromagnetic forces Caution Do not expose the unit to static electricity Always disperse any static electricity from your body before handling the unit Do not take measurements in a moving vehicle Do not use a cellular phone near the unit e Do not disassemble repair or modify the unit Do not insert batteri
31. if necessary The date and time are important indicators of when a measurement was taken Note The date and time format can be changed to different territorial standards Refer to 8 7 Date Format on p 36 Press the D Power Cancel button to turn the unit on When the unit is used for the first time or if the Date and Time setting has been reset the Date and Time setting screen is automatically displayed and the day digits are highlighted Press the Y C gt or lt J 4 button to set the day Press the WD gt button to increase the number Press the lt J 4 button to decrease the number D HCG801 E fm Page 14 Monday May 9 2005 6 49 PM 3 Setting the Date and Time 3 Press the OK button to confirm the day setting The day setting is set and the month setting is highlighted The display cycles through the settings day gt month gt year gt hour gt minute with each press ofthe OK button Note If the setting for the date format has been changed refer to 8 7 Date Format on p 36 the order in which items are highlighted may differ from the example above 4 Inthe same way press the W C gt or lt J amp button to set the month year hour and minute After you complete the date and time setting the screen from step 1 appears A 31 4 b 2005 Press the MENU button to return to the Menu screen at any time 0 00
32. ing them with a soft cloth or cotton swab moistened with disinfectant alcohol If you do not disinfect the electrodes there is a risk of skin disease Note If you have problems applying the unit so that the chest electrode is flat against your chest please ask your doctor for advice 21 ki gt HCG801 E fm Page 22 Monday May 9 2005 6 49 PM 6 Taking a Measurement 1 Press the D button to turn the unit on The unit displays a reminder that self measurement is not the same as medical diagnosis and then the guidance screens are displayed The guidance screens are displayed twice Note You can press the START button to begin a measurement even while the guidance screens are being displayed 2 Make sure that you are in a relaxed position The ideal posture is one where you are seated comfortably with your back straight Note You can also take a measurement while you are standing or while lying down 3 Place your finger so that it fits closely over the two Fit finger closely on the finger electrodes electrodes Notes Even if you are left handed use your right hand to take a measurement e Take care not to press the START button until you are ready to take the measurement x Do not press START button until ready to take measurement P UND ail HCG801 E fm Page 23 Monday May 9 2005 6 49 PM 6 Takin
33. lock has been reset Error Messages Error Message Please insert new The batteries are worn out batteries Insert new batteries and reset the Date and Time and other settings Replace both batteries with new ones An error occurred while trying to read ECG data stored in unit memory Device memory error Cannot read data An error occurred while trying to erase ECG data stored in unit memory Device memory error Cannot erase data An error occurred while trying to save ECG data to unit memory Device memory error Cannot write data Try again If the error still occurs consult your OMRON service representative ejfe HCG801 E fm Page 49 Monday May 9 2005 6 49 PM Error Message 10 Care and Maintenance SD memory error Cannot read data An error occurred while trying to read ECG data stored on the SD memory card SD memory error Cannot erase data An error occurred while trying to erase ECG data stored on the SD memory card SD memory error Cannot write data An error occurred while trying to save ECG data on the SD memory card Format the SD memory card Refer to p 19 Copy any important data to a PC before formatting the SD memory card as all data on the card will be erased Use a new different SD memory card Always turn off the unit before removing or inserting SD memory cards Error f
34. ng the Settings 8 6 Language D This setting specifies the language to be used for the screen display Date and Time i Volume Brightness gt Cont rast Language The currently selected language is The default language is English displayed at the bottom of the screen Press the lt J amp button or Y C gt button to select the desired language then press the OK button to confirm your selection 35 tie gt D 8 7 36 IP gt HCG801 E fm Page 36 Monday May 9 2005 6 49 PM 8 Adjusting the Settings Date Format You can select one of four formats DD MM YY HH MM 24 hour clock DD MM YY HH MM AM PM MM DD YY HH MM AM PM or YY MM DD HH MM AM PM E gt 12 31 05 1 30PH 05 12 31 1 30PM Date format amp Hessage aOverwrite mErase all data Data copy to SD 31 12 05 13 30 The date is displayed at the bottom ofthe screen using the currently selected format DOMY HEMM 24 hour clock Press the lt J amp button or WD gt button to select the desired format then press the OK button to confirm your selection E a ce 8 Adjusting the Settings 8 8 Messages This setting specifies whether to display or hide the message that appears after a measurement has been completed E ElDate format S0verwrite E gt mErase all data Data copy to SD Displ
35. ormatting SD memory An error occurred when formatting the SD memory card Use a new different SD memory card Always turn off the unit before removing or inserting SD memory cards SD memory error SD memory unknown The SD memory card may be damaged Format the SD memory card Refer to p 19 Copy any important data to a PC before formatting the SD memory card as all data on the card will be erased Use a new different SD memory card Always turn off the unit before removing or inserting SD memory cards SD memory full Please delete unnecessary data The SD memory card is full and there is not enough space to save new ECG data The number of data in SD is over the limit Please delete unnecessary data The SD memory card is full and there is not enough space to save new ECG data Either delete unnecessary data or use a new SD memory card Always turn off the unit before removing or inserting SD memory cards Device error The unit may have been damaged or is otherwise unable to operate normally Turn the unit off then on again If this error still occurs consult your Omron Customer Service Department Tel 0870 7502771 gt 10 Care and Maintenance 10 2 Troubleshooting 50 Le HCG801 E fm Page 50 Monday May 9 2005 6 49 PM Refer to this table for problems that do not result in a message on the display Nothing happ
36. red in the unit may be erased If the date and time setting is erased be sure to reset it before using the unit again Any ECG data erased from the unit cannot be recovered The battery life can be confirmed in the upper right of the display Je Remaining power is low Full power Half remaining Replace the batteries with new ones Warning If battery fluid should get in your eyes immediately rinse with plenty of clean water Consult a doctor immediately EN Caution Do not use batteries not specified for this unit Do not insert the batteries with the polarities in the wrong direction Do not dispose of batteries in fire If battery fluid should get on your skin or clothing immediately rinse with plenty of clean water 11 o p D HCG801 E fm Page 12 Monday May 9 2005 6 49 PM 2 Inserting and Replacing Batteries Remove the batteries from this unit when you are not going to use it for a long period of time approximately three months or more Do not use batteries of a different type together Do not use new and used batteries together a E gt 7 gt Le HCG801 E fm Page 13 Monday May 9 2005 6 49 PM 3 Setting the Date and Time Always set the date and time before using the unit for the first time Check the date and time are correct before using the unit reset them
37. t J 4 button or Y C gt button to adjust the volume The volume increases as the mark moves to the right The sound is muted off when the mark is to the far left a tie ce HCG801 E fm Page 33 Monday May 9 2005 6 49 PM 8 4 Brightness E Date and Time Volume The current brightness level is displayed The mark indicates the at the bottom of the screen brightness level The default setting is the maximum brightness level 8 Adjusting the Settings E Press the lt J 4 button or Y C gt button to adjust the brightness The brightness can be set to one of nine levels The brightness increases as the amp marker moves to the right ha 33 a tie gt 34 I HCG801 E fm Page 34 Monday May 9 2005 6 49 PM 8 Adjusting the Settings 8 5 Contrast E Date and Time Volume Brightness E gt The current contrast level is displayed The u mark indicates the at the bottom of the screen contrast level The default setting is medium contrast level Press the lt J amp button or Y C gt button to adjust the contrast The contrast can be set to one of nine levels The contrast increases as the 4 amp marker moves to the right 1 ki a a oe HCG801 E fm Page 35 Monday May 9 2005 6 49 PM 8 Adjusti
38. the waveform situations can occur in which the device cannot assess and display the heart rate during the measurement Notes e During measurement the device will only respond to the D button to discontinue the measurement If the contact of the electrodes becomes loose or the conditions change during measurement the measurement may not be correct Important Keep still and do not move until the measurement is complete 24 gt Ril HCG801 E fm Page 24 Monday May 9 2005 6 49 PM aa 2 UND HCG801 E fm Page 25 Monday May 9 2005 6 49 PM 6 Taking a Measuremen measurement result message is displayed when m gf 1 2005 15 00 E the measurement has been completed Refer to p Stable waveform 27 for details If you press the OK button while the result message is displayed the list of ECG data stored on the SD Heart rate 60 min memory card or unit is displayed Store data Yes No Note If the Please insert new batteries error message is displayed immediately after measurement the ECG data may not be Example result message stored Ifthis happens please repeat the measurement after replacing the batteries 6 Pressthe D button for two seconds to turn the unit off The unit beeps then the power is turned off Note If you forget to turn the unit off it will automatically shut itself off after five minutes Zr
39. tored on the SD memory card or in unit memory p 28 SET menu This allows you to set the date and time and adjust other settings p 30 GUIDE This displays the guidance screens that show you how to take a measurement Date and Time Volume Brightness Cont rast Language 17 1 05 0 00 Settings SET menu screen The setting menu screen is displayed when you select the SET menu and press the OK button Q Back This option takes you back to the previous screen from the SET menu R Close This option closes the menu screen P gt gt HCG801 E fm Page 9 Monday May 9 2005 6 49 PM ECG Graph 1 Know Your Unit This screen is displayed when you use the GRAPH menu to select ECG data stored on the unit S Calibration guide Serves as a guide when your doctor analyses the stored ECG data T Full 30 second waveform U Measurement time indicator V SD memory card indicator Displayed when the SD memory card is in the card slot W Battery level indicator X ECG waveform Y Heart rate bpm beats per minute Note During measurement the screen displays the ECG waveform and the progress bar at the bottom of the screen displays measurement progress and a guide to the heart rate Depending on the waveform situations can occur in which the device cannot assess and display the heart rate during the measurement 2 UND
40. ur PC and confirm that there is an hcg801 folder in the SD memory card If there is not format the SD memory card in the HCG 801 as described in section 4 2 of the HCG 801 instruction manual 3 Browse to the folder where you saved the ECG data in the PC Select all files and copy them to the SD memory card in the ncg801 data folder where stands for the character of a removable disk 4 When copying is finished insert the SD memory card in the HCG 801 as described in 4 1 Inserting and Ejecting SD Memory Card on p 15 5 Do not store more than 300 measurement files in this folder on the SD memory card When more than 300 files are copied in this folder the HCG 801 will display an error message when activated 45 E d HCG801 E fm Page 46 Monday May 9 2005 6 49 PM 10 Care and Maintenance 10 1 Lists and Explanations of Messages 46 I This section includes a list of all messages that may appear on the display Refer to these if you are not sure what a displayed message means Notice Messages Analysing The ECG data is being analysed No action required Wait until analysis ends Can not measure Please follow the instructions for use A ECG wave pattern could not be detected Check the instructions and repeat measurement Measurement was stopped Push START to begin measurement again The measurement was interrupted or cancelled