SERVICE MANUAL Rocket CRAFT™ Oocyte

image

Contents

1. test gauge 7 Activate the Standard a ONY Vacuum footswitch observe the green light is illuminated I E 4 8 Rotate the control knob B clockwise to increase to a ana the desired value on the c gauge See table below L OI 9 To decrease the value rotate anticlockwise 10 When the desired value is reached release the footswitch and tubing set occlusion 11 Activate the Standard V m f witch Test Indicated WARNING acuu ootswitc Value EE Ensure the vacuum 19 R dth ith Mecor e vacuum levels are within the E 50mmHg specified accuracy registered tolerance before Ee 13 Repeat from 6 for each indicated in the table right Specified accuracy 10 of test value Test Gauge 2 Copyright 2012 13 Rocket Medical plc All rights reserved ZDOCK226 Rev 4 30 09 13 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 9 FUSE REPLACEMENT WARNING ELECTRIC SHOCK HAZARD The equipment is to be Two 0 5A fuses al S used only with electrical located in the main power systems complying with socket mounted on the all IEC CEC and NEC rear panel requirements Using a small screwdriver carefully crar ease out the fuse carrier ROCKET MEDICAL WATFORD WOP ZEN N to reveal the fuses FAN To ensure continued protection REPLACE WITH FUSES IMPORTANT OF IDENTICAL RATING 22 Copyright 2012 13 Rocket Medical plc All rights reserved ZDOCK226 Rev 4
2. Off Motor 220VAC 50Hz 110VAC 60Hz model dependant Vacuum outlet port Pump head including valve housing Internal silencer with filter chamber Main power input Copyright 2012 13 Rocket Medical plc All rights reserved ZDOCK226 Rev 4 30 09 13 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 3 GENERAL DESCRIPTION The Rocket CRAFT Oocyte Aspiration Pumps have been developed to provide smooth low volume vacuum at a pre determined negative pressure Vacuum is activated by a foot operated toggle air switch controlled by the surgeon performing the oocyte collection The range of vacuum is infinitely variable from 10 200mmHg in medium vacuum mode and at a pre set 400mmHg in high vacuum mode Rocket CRAFT Oocyte Aspiration Pumps require either a water trap set Supplied or R57685 Disposable Rocket CRAFT Water Trap Sets supplied separately sterile and for single patient use A suitable collection test tube for use with oocyte needle sets is B D Falcon test tube No 2001F 17 x 100mm 3 1 INDICATIONS For the generation of high vacuum low volume suction between 10 400 mmHg to permit the aspiration of follicular fluid oocytes and ovarian fluid as part of the treatment of infertility relating to IVF and other gynaecological procedures 3 2 CONTRAINDICATIONS Not intended for use where ovarian aspiration or the aspiration of ovarian fluid is contraindicated For short term operation on
3. of the DEVICE Rocket CRAFT Oocyte Aspiration Pumps must be placed on a secure level surface away from sources of heat water splashes mists or cooling vents Do not expose to direct sunlight Do not expose to direct sunlight Do not expose to flammable gases Operating temperature Range 5 C and 35 C Copyright 2012 13 Rocket Medical plc All rights reserved ZDOCK226 Rev 4 30 09 13 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 4 10 SYMBOLS USED ON CRAFT SUCTION amp DUO VAC PUMPS These tables describe their respective meanings Mains Power ON OFF This device is Type B CE Mark 4 11 SYMBOLS USED ON R57685 PATIENT CONNECTION SET A TERILE A Read the Instruction for Use before connection and use Device is for Single Use Only Device is sterilised by Ethylene Oxide The device is Latex Free CE Mark Copyright 2012 13 Rocket Medical plc All rights reserved ZDOCK226 Rev 4 30 09 13 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 5 GENERAL INFORMATION 5 1 General Description The Rocket CRAFT Oocyte Aspiration Pumps have been developed to provide smooth low volume high vacuum at a pre determined negative pressure Vacuum is activated by a foot operated toggle air switch controlled by the surgeon performing the oocyte collection The range of vacuum is infinitely variable from 10 200mmHg in medium vacuum mode and at a pre set 440mmHg in hi
4. 13 Rocket Medical plc All rights reserved ZDOCK226 Rev 4 30 09 13 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 2 2 Rocket CRAFT DUO VAC Suction Pump Internal View E Vacuum gauge 0 700mmHg normal range 50 440mmHg Illuminated O I Mains Power On Off Motor 220VAC 50Hz 110VAC 60Hz model dependant Vacuum outlet port Pump head including valve housing Internal silencer with filter chamber EO PE een ee Se Main power input Copyright 2012 13 Rocket Medical plc All rights reserved ZDOCK226 Rev 4 30 09 13 Rocket CRAFT Oocyte Aspiration Pumps Service Manual Rocket Craft Suction Pump with earlier metal footswitch Illuminated O I Mains Power On Off Vacuum Display mmHg High Vacuum 440mmHg 1 control button ee dn Vacuum Control clockwise to increase anticlockwise to decrease the set value 5 Water trap connection port recommended for use with R57685 Rocket Craft Pump Water Trap Sets 6 Water Trap Set comprising glass bottle bung angled connectors silicone tube set 7 Footswitch connection port Medium Vacuum Standard 50 250mmHg footswitch Silicone tube set with metal luer connector Copyright 2012 13 Rocket Medical plc All rights reserved ZDOCK226 Rev 4 30 09 13 EE AE oe IN e Rocket CRAFT Oocyte Aspiration Pumps Service Manual Vacuum gauge 0 700mmHg normal range 50 440mmHg Illuminated O I Mains Power On
5. 30 09 13 Rocket CRAFT Oocyte Aspiration Pumps Service Manual A WARNING ELECTRIC SHOCK HAZARD Do not immerse the device o WARNING Do NOT attempt to sterilise the device WARNING Protect the device from ingress of liquid Should any liquid enter the device discontinue use immediately and refer to an authorised service agent WARNING No user serviceable parts inside A HAZARD Do NOT include used consumables as these pose a significant contamination risk e IMPORTANT A decontamination certificate MUST be included with every returned pump Repair or servicing cannot be commenced unless the service agent is in possession of this certificate 10 CLEANING THE PUMP CASING At the end of each clinical session turn off the device at the front panel and disconnect the device from mains power supply Using an aqueous solution of 70 alcohol IMS or isopropyl BP moisten a cloth and wipe all external surfaces of the device If the surface has become contaminated with proteinaceous material remove with a light detergent solution before surfacing cleaning with alcohol Do NOT use a 100 alcohol or any other solvent to clean the device as this may cause damage to the casing surface and display Prevent any fluid from entering the device ll YEAR OF MANUFACTURE The year in which the device was manufactured is indicated by the first 2 numbers of the serial number For examp
6. AFT Oocyte Aspiration Pumps Service Manual 4 Undo and remove the centre Pozi drive screw to release the valve plate 5 Carefully remove valve plate from head moulding with a twisting motion to separate the two halves 6 Valve gasket and valve plate Note the orientation of the cut outs on the valve gasket in relation IMPORTANT to the valve plate 7 Lift off pump head gasket and retain safely Copyright 2012 13 Rocket Medical plc All rights reserved ZDOCK226 Rev 4 30 09 13 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 8 Undo and remove the centre Pozi drive screw from the pump diaphragm retaining plate 9 Retain retaining plate and screw discard old diaphragm 10 Replace diaphragm with P N DO1 053 11 Re assemble and tighten sufficient to secure the oe retaining plate DO NOT OVERTIGHTEN IMPORTANT or use excessive force Copyright 2012 13 Rocket Medical plc All rights reserved ZDOCK226 Rev 4 30 09 13 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 6 3 Reassembling the valve set e 1 Ensure the valve plates n Se are orientated as in the LINO image right e VERY IF THIS ORIENTATION IS IMPORTANT REVERSED THE VALVE AND PUMP WILL FAIL TO OPERATE 2 Replace the valve gasket with P N VO1 002 Note the orientation of the cut outs on the valve gasket IMPORTANT in relation to the valve plate 3 Reassemble the valve
7. G Rocketmedical SERVICE MANUAL Rocket CRAFT Oocyte Aspiration Pumps d Ei O G Q Rocket Medical plc Sedling Road Washington Tyne amp Wear NE38 9BZ England T 44 0 191 419 6988 F 44 0 191 419 6989 E customerservices rocketmedical com www rocketmedical com Registered Office Rocket Medical plc Imperial Way Watford WD24 ANN England Company No 03276608 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 1 CONTENTS Ree vn C 1 GENERALASSEMBEY E 3 2 1 Rocket CRAP I DUO VAC Suction PUMP iersacdtesndsesacece eenegen Eege geeletegee eech 3 2 2 Rocket CRAFT TM DUO VAC Suction Pump Internal VieW sse eee eee ee eee 4 ee rr REI el H ONID E 5 2 4 Rocket CRAFT TM Suction Pump Internal View 6 3 GENERAL DESCRIPTION E 7 Sek WIE IER 7 92 ONTARIO LE 7 EEN 7 4 GENERAL INFORMATION WEE 8 ACOPERI Ceram eer eee A A E A E ee E E ee ee ee ee eee 8 AZ WOE AE EE 8 4o MANUAL eai Eet 8 4A MANUFAC TURE E 8 k ERCE AGENT EE 8 BG SUPPLY VOLTAGE e Ke EE 9 4 7 ELECTROMAGNETIC COMPATIBILITY E 9 LEPA ACGIN EE 9 4 9 POSITIONING and PLACEMENT of the DEVICE sss 9 4 10 SYMBOLS USED ON CRAFT SUCTION A DUO VAC PUMPS xxe 10 4 11 SYMBOLS USED ON R57685 PATIENT CONNECTION SET esse 10 E een ue ME 11 EGT SSI EE 11 RA leae en EE 11 e E Ol RANI ee EE 11 eK Ee 11 Ee 12 6 1 Replacing the filter element 12 6 2 Replacing the pump valve amp diaphragm set 14 6 3 Reassembling the E 17 7 REPAIRS S REPLACE
8. MENTS EE 19 7 1 Replacing the DUO VAC Footswitch Connechors sse eee eee 19 7 2 Replacing the gauge in DUO VAC amp Suction Pump modela esse sees ee eee 20 RINAL TE PIN EEN 21 9 FUSE REPLACEMENT soiien eO EEEE E EEEE 22 10 CLEANING THE PUMP CASING EE 23 11 YEAR OF MANUFACTURE E 23 12 RETURNING THE PUMP FOR FACTORY REPAIR OR SERVICE sse ee eee 23 De La ETT 24 14 TRANSPOR eebe 24 1 WARRANT E 25 16 TECHNICAL e e le LE 26 Ie e CAS IFEA HON EE 26 16 2 SPECIFICATION EE 26 DRA WIN O E 27 ke GENERALAS SEMBL CTT 27 17 2 GENERAL Re EE 28 17 9 GENERA CIN S TALLA TION KEE 29 Copyright 2012 13 Rocket Medical plc All rights reserved ZDOCK226 Rev 4 30 09 13 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 2 GENERAL ASSEMBLY 2 1 Rocket CRAFT DUO VAC Suction Pump Soe oe hy E N Rocket Craft DUOVAC Pump with footswitch supplied after 2013 Illuminated O I Mains Power On Off Vacuum Control Dial clockwise to increase anticlockwise to decrease the set value Vacuum Display mmHg Footswitch connection ports Water trap connection port recommended for use with R57685 Rocket Craft Pump Water Trap Sets Water Trap Set comprising glass bottle bung angled connectors silicone tube set Recommended for use with R57685 Rocket Craft Pump Water Trap Sets Silicone tube set with metal luer connector Footswitch BLACK Standard 50 250mmHg and WHITE High Vacuum 440mmHg foot switch Copyright 2012
9. Rev 4 30 09 13 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 7 2 Replacing the gauge in DUO VAC amp Suction Pump models Remove the upper casing See Section 6 Lift the front panel out of the retaining grooves on the lower case Loosen and remove the securing nuts or screws ringed right Draw the gauge through the front panel whilst supporting the mounting plate Detach the gauge from the centre tube by carefully pulling the tubing from the rear connection Replace the gauge reconnect the centre tubing and insert through the front panel Replace gauge mounting plate and tighten securing nuts or screws E A CRAFT DUO VAC SUCTION UNIT eo Ky Rocket Craft DUO VAC gauge assembly We A gt O U a Z T L m SCH gege H lf mr Rocket Craft Suction Pump gauge assembly Copyright 2012 13 Rocket Medical plc All rights reserved ZDOCK226 Rev 4 30 09 13 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 8 FINAL TESTING 1 Final testing should be completed with a validated and independently calibrated electronic vacuum gauge or similar device 2 Required operating scale 10 500mmHg in 1mmHg increments 3 Attach the vacuum test gauge as indicated in the image below to the external vacuum port 4 Attach the footswitch as for normal operation 5 Power device on 6 Occlude the tube to the
10. gh vacuum mode Rocket CRAFT Oocyte Aspiration Pumps require either a water trap set Supplied or R57685 Disposable Rocket CRAFT Water Trap Sets supplied separately sterile and for single patient use A suitable collection test tube for use with oocyte needle sets is B D Falcon test tube No 2001F 17 x 100mm 5 2 Indications For the generation of low flow high vacuum between 10 440 mmHg to permit the aspiration of follicular fluid oocytes and ovarian fluid as part of the treatment of infertility relating to IVF and other gynaecological procedures 5 3 Contraindications Not intended for use where ovarian aspiration or the aspiration of ovarian fluid is contraindicated For short term operation only This device is not indicated for continuous drainage 5 4 References Craft McLeod F Edmonds K Human embryo transfer technique Lancet 1961 ii 1104 5 Craft Diahanbakch O McLeod F et al Human pregnancy following oocyte and sperm transfer to the uterus Lancet 1992 1 1031 3 Craft I 1984 Clinical Methodology British Journal of Hospital Medicine 90 102 Reeves G Scott R T et al 1989 Journal of Assisted Reproduction and Genetics Volume 6 Number 6 December 1989 Copyright 2012 13 Rocket Medical plc All rights reserved ZDOCK226 Rev 4 30 09 13 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 6 ANNUAL SERVICE 6 1 Replacing the filter element 1 Lay the pump on a c
11. he Products have been repaired other than by Rocket Medical plc or other than in accordance with written instructions provided by Rocket Medical plc or altered by anyone other than Rocket Medical plc or if the Products have been subject to misuse negligence or accident Rocket Medical plc s sole and exclusive obligation and Buyer s sole and exclusive remedy under the above warranties is limited to repairing or replacing free of charge at Rocket Medical plc s option Products which are reported to Rocket Medical plc by mail telephone or email and which if so advised by Rocket Medical plc is thereafter returned with a statement of the observed deficiency not later than seven 7 days after the expiration date of the warranty to Rocket Medical plc during normal business address transport charges prepaid and which upon Rocket Medical plc s examination is not found to conform with the above warranties Rocket Medical plc shall not be otherwise liable for any damages including but not limited to incidental damages consequential damages or special damages There are no express or implied warranties which extend beyond the warranties herein above set forth Rocket Medical plc makes no warranty of merchantability or fitness for a particular purpose with respect to the Products or parts thereof 25 Copyright 2012 13 Rocket Medical plc All rights reserved ZDOCK226 Rev 4 30 09 13 26 Rocket CRAFT Oocyte Aspiration Pum
12. head and tighten the centre screw Copyright 2012 13 Rocket Medical plc All rights reserved ZDOCK226 Rev 4 30 09 13 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 4 Reseat the pump head gasket P N GO1 001 5 Replace the valve head assembly onto the gasket ensure the holes are correctly aligned and the gasket with seated correctly 6 Tighten the valve head screws in a diagonal pattern to ensure the pump head is correctly seated without distortion 7 Re fit the pump silencer to the head module IMPORTANT Copyright 2012 13 Rocket Medical plc All rights reserved ZDOCK226 Rev 4 30 09 13 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 7 REPAIRS amp REPLACEMENTS 7 1 Replacing the DUO VAC Footswitch Connectors The front panel connectors of the DUO VAC can be subject to damage through misuse or repeated over stressing of the retaining clips on connection of the footswitch Replacement is a straight forward operation 1 Remove the upper casing See Section 6 2 Lift the front panel out of the retaining grooves on the lower case 3 Remove tubing by carefully pulling from connector 4 Loosen and remove securing nut 5 Replace connector locate and tighten securing nut 6 Re fit tubing by pushing over connector inlet Copyright 2012 13 Rocket Medical plc R a ON VACUUM FOOT CONTROL Li All rights reserved g ZDOCK226
13. le a serial number starting 11180776 indicates the device was manufactured in 2011 2 RETURNING THE PUMP FOR FACTORY REPAIR OR SERVICE All devices to be returned must be prepared as described below for the protection of the servicing team and for safety during transport 1 Surface clean the pump as described in the Section above 2 Seal ina plastic bag and seal within a second plastic bag 3 Place in the original packaging 4 Enclose the following information e Contact name e Centre address e Decontamination Certificate e Description of the fault or service required e Accompanying Order to authorise servicing contact your local Customer Services Team for details 23 Copyright 2012 13 Rocket Medical plc All rights reserved ZDOCK226 Rev 4 30 09 13 Rocket CRAFT QOocyte Aspiration Pumps Service Manual 13 STORAGE The device must be stored and operated in temperatures 5 C to 35 C The device must be stored in a clean dry condition ideally in its original packaging which should be retained to return the unit for servicing Protect the device from ingress of liquid Should any liquid enter the device discontinue use immediately and refer to an authorised service agent The device must be transported in temperatures 5 C to 385 C The device is FRAGILE and must be transported in its original packaging to ensure protection If the original packaging is not available please contact you
14. lean surface protecting the casing from scratches 2 Remove 4 screws from WARNING the casing underside and ELECTRIC SHOCK retain HAZARD 3 Carefully separate the case halves ensuring the front and rear fascia plates remain located in the lower shell 4 Locate filter housing G 5 Using fingers lift the top away from filter chamber CAUTION Avoid the use of sharp tools as these may damage the O ring Copyright 2012 13 Rocket Medical plc All rights reserved ZDOCK226 Rev 4 30 09 13 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 6 Life out the filter assembly and remove the filter 7 Fit new replacement filter P N E05 000 8 Replace filter ensuring the top seats securely into the chamber 9 Avoid trapping or pinching the O ring 10 Check the top is correctly seated Copyright 2012 13 Rocket Medical plc All rights reserved ZDOCK226 Rev 4 30 09 13 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 6 2 Replacing the pump valve amp diaphragm set 1 Locate the pump silencer housing and carefully remove with a gentle twisting movement 2 Using a diagonal method remove the 4 screws securing the pump valve head 3 Lift the valve head away from the pump body to reveal the pump head gasket and diaphragm assembly Copyright 2012 13 Rocket Medical plc All rights reserved ZDOCK226 Rev 4 30 09 13 Rocket CR
15. ly The device is NOT intended for continuous drainage 3 3 REFERENCES Craft McLeod F Edmonds K Human embryo transfer technique Lancet 1961 ii 1104 5 Craft Diahanbakch O McLeod F et al Human pregnancy following oocyte and sperm transfer to the uterus Lancet 1992 i 1031 3 Craft I 1984 Clinical Methodology British Journal of Hospital Medicine 90 102 Reeves G Scott R T et al 1989 Journal of Assisted Reproduction and Genetics Volume 6 Number 6 December 1989 Copyright 2012 13 Rocket Medical plc All rights reserved ZDOCK226 Rev 4 30 09 13 Rocket CRAFT Oocyte Aspiration Pumps Service Manual WARNING READ THIS MANUAL CAREFULLY Please familiarise yourself with the contents of this manual before attempting to use the device Failure to observe these instructions may result in damage to the pump or cause injury to the patient or user This device should only be used by suitably qualified personnel L WARNING ELECTRIC SHOCK HAZARD The equipment is to be used only with electrical systems complying with all IEC CEC and NEC requirements CAUTION Any adjustment modification or repairs to the equipment should be carried out by authorised service agents Disposal of this device must be undertaken with regard to the WEEE directive 2002 96 EC 4 GENERAL INFORMATION 4 1 COPYRIGHT This manual contains information that is subject to cop
16. ps Service Manual 16 TECHNICAL SPECIFICATIONS 16 1 CLASSIFICATION IEC 60601 1 Degree of protection against electric shock Type B 16 2 SPECIFICATIONS Power Input to Pump 220 240 VAC 50Hz 0 12A Maximum current 2 5A Environmental conditions e Temperature 5 C to 35 C e Atmospheric Pressure Range 700 1060hPA Dimensions e W 264mm e H 124mm e D 164mm Weight e Unit 1 9Kg e Foot Switch 0 51Kg Vacuum Ranges e Medium vacuum 20mmHg to 200mmHg in 20mmHg increments e High vacuum 400mmHg Vacuum Range Accuracy 10 of value Copyright 2012 13 Rocket Medical plc All rights reserved ZDOCK226 Rev 4 30 09 13 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 17 DRAWINGS CAS SE OD PLEASE TER 70 HAD DEPT F OE TALS REFER TO CABLE SETPAATELIST AND DRAWING FOR LETTER REFERENCES MOTELS ATO T RT TER TO ET REL e BAT SA E PCHD Pes LAT L MS IEMACASE Le E P Lee 3 CPT zum 2 P PEM TO ORPES R E A TEE 71 SEA LETO ARRE EIR a A SE CATE C OE TME 29 mahn LAT EWE TER B GUE CTC SOB TEN OG Ae EU Le EWE TER F E LOA GE FOE EES EE EAR 8 XUI WELL FE IEN 86 17 1 GENERAL ASSEMBLY 2 30 09 13 Rev 4 ZDOCK226 All rights reserved Copyright 2012 13 Rocket Medical plc Rocket CRAFT Oocyte Aspiration Pumps Service Manual 17 2 GENERAL ELECTRICAL TE TET MTE 18 RY d PSL SSS Le 1 0 Rw 9d SE 5 m A m CD v D Q lt Di O 2 S 2
17. r local Customer Services Agent who will provide replacement packaging Weight e Unit 1 9Kg Dimensions Foot Switch e W 264mm 0 57Kg DUOVAC plastic e H 124mm 0 34kg Craft Suction plastic e D 164mm 0 51Kg DUOVAC metal 0 38kg Craft Suction metal Protect the device from ingress of liquid Should any liquid enter the device discontinue use immediately and refer to an authorised service agent 24 Copyright 2012 13 Rocket Medical plc All rights reserved ZDOCK226 Rev 4 30 09 13 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 15 WARRANTY Rocket CRAFT Oocyte Aspiration Pumps are sold by Rocket Medical plc under the warranties set forth in the following paragraphs Such warranties are extended only with respect to the purchase of the Products directly from Rocket Medical plc as new merchandise and are extended to the first Buyer thereof other than for resale For a period of twelve 12 months from the date of shipment the Products are warranted to be free from functional defects in materials and workmanship and to conform to the description of the Products contained in the operating manual and accompanying labels provided the same is properly operated under conditions of normal use that annual maintenance and service is performed at an authorised Rocket Medical plc service facility Removal of any QC seal voids the warranty The foregoing warranties shall not apply if t
18. spiration Pumps Service Manual WARNING ELECTRIC SHOCK HAZARD The equipment is to be used only with electrical systems complying with all IEC CEC and NEC requirements L WARNING ELECTRIC SHOCK HAZARD Do not immerse the device ie Be EXPLOSIVE WARNING Device can cause explosion in the presence of flammable gases 4 6 SUPPLY VOLTAGE SELECTION The device operates at a voltage 110V 60Hz or 220 240 VAC 50Hz 0 12A depending on model Ensure that the correct power cord is connected 4 7 ELECTROMAGNETIC COMPATIBILITY Rocket Craft Suction Pumps comply with the electromagnetic compatibility EMC limits for medical devices as specified by IEC 60601 1 2 2001 These limits are designed to provide a reasonable degree of protection against harmful interference found in typical medical installations Medical electrical equipment requires special precautions regarding EMC and the device must be installed positioned and operated according to the instructions contained in this manual to ensure continued electromagnetic compatibility The device must be operated according to the instructions contained in this manual to ensure continued electromagnetic compatibility 4 8 PACKAGING The packaging has been designed to allow secure transportation of the pump and its accessories After unpacking re assemble and retain the packaging for transport for servicing when required 4 9 POSITIONING and PLACEMENT
19. yright All rights reserved This manual should not be photocopied duplicated or distributed completely or in part without the written approval of Rocket Medical plc 4 2 MODEL NUMBERS Rocket Craft Suction Pump Complete 110v R29654 Rocket Craft Suction Pump Complete 240v R29655 Rocket Craft Duo Vac Suction Pump 240v R29660 Rocket Craft Duo Vac Suction Pump 110v R29661 4 3 MANUAL REVISION Revision 1 First Publication 15 02 12 Revision 2 Addition of drawings 29 06 12 Revision 3 Update of tolerance specifications 23 09 13 Revision 4 Update images with new footswitch 30 09 13 4 4 MANUFACTURER Rocket Medical plc Sedling Road WASHINGTON Tyne amp Wear NE38 9BZ UK 4 5 SERVICE AGENTS Rocket Craft Pumps typically require only annual maintenance however it is recommended they are serviced and calibrated annually at a Rocket Medical plc service facility Failure to service the pump at the indicated intervals may invalidate the Warranty Customer Services Rocket Medical plc Sedling Road WASHINGTON NE38 9BZ ENGLAND Tel 44 0 191 419 6988 Fax 44 0 191 419 6989 Email homesales rocketmedical com US Office Rocket Medical Unit 3 150 Recreation Park Drive HINGHAM MA 02043 USA Tel 1 781 749 6223 Fax 1 781 749 6235 Email usa rocketmedical com Copyright 2012 13 Rocket Medical plc All rights reserved ZDOCK226 Rev 4 30 09 13 Rocket CRAFT Oocyte A
20. zp 3 W Hed L ogr 28 Copyright 2012 13 Rocket Medical plc All rights reserved ZDOCK226 Rev 4 30 09 13 Rocket CRAFT Oocyte Aspiration Pumps Service Manual 17 3 GENERAL INSTALLATION OUTLINE DRAWING 1 9 Kg ALL DIMENSIONS IN MILLIMETRE Surface texture values in um Aa Tolerance unless otherwise stated Whole numbers D B One decimal placa d 0 3 Two decimal places f 0 1 Angular h Zp IF IN DOUBT ASK NG OUTLINE DRAWI 3 S METERA METAAL STL N A DO NOT SCALE ENEAK SHARP EWES lt p 29 30 09 13 Rev 4 ZDOCK226 All rights reserved Copyright 2012 13 Rocket Medical plc