(Jerome) 860 Hydrogen Sulfide Monitor User`s Manual246255

image

Contents

1. 5 SPECIFICA TIONS GG niei ees ee tiene edi tex eee eos void Eo beso e ideo EEEa 6 SENSOR CROSS SENSITIVITY INTERFERENCES ccscsccccsssscccccecccees 7 INITEIAE TURN ON petiti sae ca qp u ne oben Co Ue eoe Ve ceca a eo Ue eoe Eb veo eee ve PERSE DOE 8 USER INTERPACB aieiaeo E Dites 9 Status ISI REI soancaacscceuesceceseseseccucsesene 9 Men I St RTT T T TER 9 Parameter Editino ositos m 9 Monu buco a TN 10 CALIBRA TION enirn iaee a aa aO a 13 Calibration Procedure seesseosseesseessecssecssecssesosesssesssesssesssesssesseesseosseosseosseossesssessse 14 Calibration Check Procedure seeoseesseosseossoossoossosssesssecssesssesssosssesseesseosseesssossessoo 15 CLEANING ciones aee a H M 15 DATA LOGGING SETUP i seeedieseevibee even ceo eu eS Oed opo E EAEE Naa 16 IOCLEAR THE LOGGED DA TA iniiis eieren el eee Feweos evi aoaeiaa aa iesu 16 LOADING SOETWA RE seanina niece eo iecosed ea esp Uie cose toa qemose vios 17 I oadins JC S Ee ien ici io E vi REFERS ER aE E ROI MFEPESEU VUES EE XPEEEEU EE ER PANI DE TEE 18 DATA TRANSFER oreca a eae ie Poe Eos 22 Jerome 500 tO PC nieto v a Ev rb d niii metit aM 22 JCS LITEOPERA TION i neceieeetsuc eeboeoueiune eel uv epe Hu ducato eo ve eov ue tua espe Uv vae oed n Pa e Ua Soo v 23 Toolbar Explanation c 23 Plot Manipulation Pr 24 BATTERY REPLACEMENT s icsccssccsestsezs cesticesesieed aacctdcavenisedssustice
2. Do not touch the sensor at any time Use a damp disposable towel to clean the case be careful not to scratch the LCD area Soap and water can be used to rinse the unit but ensure that the sensor remains dry by installing the calibration plug with red caps on the barbs Do not use a soap with a lotion and or abrasives Do not use solvents To disinfect the case cover the sensor with the calibration plug with caps on the hose barbs and use isopropyl alcohol Do not use products such as Lysol Page 15 of 30 9 DATA LOGGING SETUP From the STATUS display press SELECT to access the menu When LOG is displayed press SELECT again to access the LOG menu Press DOWN to get to the INTERVAL menu Press SELECT and a time interval will be displayed 2 10 30 sec 1 2 5 10 30 min or 1 hour The interval determines how often the sample value is logged to the data file Press UP or DOWN until the desired interval is displayed and press SELECT gt The instrument displays SAVED and returns to INTERVAL nal ee ae e Oo Press UP to BEGIN and press SELECT Press SELECT again to CONFIRM LOGGING is displayed on the LCD The instrument continues logging until ON or OFF is held When LOGGINGi is stopped the instrument will perform the INITIAL TURN ON display as detailed on page 8 10 TO CLEAR THE LOGGED DATA The logged data should be cleared from the 860 on a regular basis to avoid any confusion du
3. Page 5 of 30 3 SPECIFICATIONS The Jerome 860 Hydrogen Sulfide Monitor is designed to continually monitor and record hydrogen sulfide concentrations in remote locations The unit provides automatic unattended long term monitoring of hydrogen sulfide gases for contained or hard to reach areas The compact battery powered module is portable water resistant and intrinsically safe The heart of the 860 is a family of durable electro chemical sensors Each sensor has a standard detection range and is also capable of measuring sporadic spikes in H2S concentrations Refer to the table below for the specific range and spike reading capabilities of the different models The 860 is easily programmed through a user friendly touch panel to control sample interval date and time and calibration modes A PC is not required to program the 860 The module measures and stores over 100 000 H2S data points For additional graphing and reporting functionality the data may be downloaded through an infrared port to a PC To download data to a PC JCS Lite must be installed and running on the PC Sensor Electrochemical Hydrogen Sulfide Resolution X860 0001 0 3 ppm X860 0002 0 1 ppm HS Measurement Range X860 0001 Standard 0 0 to 200 0 ppm Occasional spikes up to 1000 0 ppm X860 0002 Standard 0 0 to 50 0 ppm Occasional spikes up to 500 0 ppm Accuracy X860 0001 amp X860 0002 1 ppm or 496 at STP 760mm Hg 20 C 2 ppm or 8 max Response
4. USER MANUAL JEROME 860 HYDROGEN SULFIDE ANALYZER OPERATION MANUAL January 2011 ARIZONA INSTRUMENT LLC 3375 N Delaware Street Chandler AZ 85225 USA 800 528 7411 602 470 1414 f 602 281 1745 WWW azic com Email General azi azic com International inth azic com Customer Service support azic com ARIZONA INSTRUMENT LLC 3375 N Delaware St Chandler AZ 85225 800 528 7411 sales azic com www azic com MADE IN THE USA AZI P N 700 0051 D Last update January 2011 JEROME 860 HYDROGEN SULFIDE MONITOR Operation Manual PROPRIETARY RIGHTS NOTICE This manual contains valuable information and material developed by Arizona Instrument LLC for use with the Jerome 860 Hydrogen Sulfide Monitor No part of this manual may be reproduced or transmitted in any form or by any means electronic mechanical or otherwise This includes photocopying and recording or in connection with any information storage or retrieval system without the express written permission of Arizona Instrument LLC ALL RIGHTS RESERVED Copyright 2003 2010 Arizona Instrument LLC An up to date electronic copy of this manual can be found on the AZI web site at http www azic com downloads aspx Page 2 of 30 dE db E D Eu 8 9 10 11 12 13 14 15 Table of Contents INTRODUCTION TO THE JEROME 860 ccccscsssssssssssssssssssssssesescsssssssessssssesees 4 WARNINGS READ THIS BEFORE USING INSTRUMENT
5. boxes and packing materials are available from AZI e Pack the Jerome instrument only in a Jerome shipping container AZI WILL NOT BE RESPONSIBLE FOR SHIPPING DAMAGE IF YOU RETURN THE INSTRUMENT IMPROPERLY PACKAGED OR SHIPPED YOU SHOULD INSURE IT FOR FULL VALUE e Check for any damage and confirm receipt of all parts on your packing list Contact AZI Customer Service at 800 528 7411 602 470 1414 or through e mail at Support azic com if you have any questions Page 4 of 30 2 WARNINGS READ THIS BEFORE USING INSTRUMENT e OREDUCE THE RISK OF IGNITION OF A FLAMMABLE OR EXPLOSIVE ATMOSPHERE PERFORM ALL MAINTENANCE ONLY IN A KNOWN NON HAZARDOUS LOCATION e THE 860 IS NOT DESIGNED TO OPERATE AS A SAFETY DEVICE IT CONTAINS NO ALARM FEATURES OR ALERTS IT ONLY PROVIDES DATALOGGING FUNCTIONS e PLEASE CLEAN THE UNIT BEFORE RETURNING IT TO THE FACTORY FOR ANY REASON SEE BELOW FOR CLEANING GUIDELINES e Useadamp disposable towel to clean the case be careful not to scratch the LCD area Soap and water can be used to rinse the unit but ensure that the sensor remains dry by installing the calibration plug with the red caps on the barbs Do not use a soap with a lotion and or abrasives Do not use solvents e To disinfect the case cover the sensor with the calibration plug with caps on the barbs and use isopropyl alcohol Do not use products such as Lysol e Do not touch the sensor e The sensor requires oxygen to ope
6. Time 9096 of sampled concentration within 35 seconds Sample Interval User selectable at 2 10 30 seconds 1 2 5 10 30 minutes or 1 Environmental Range NS 4 50 C 20 to 8096 humidity normal operation non condensing 80 to 10096 humidity for short periods SEE WARNINGS SECTION FOR DETAILS Interface User programmable through touch panel Data transfer through infrared IrDA to compatible PC Data Capacity 100 000 data points Programmable Features Date Time Sample Interval and Calibration data Display 8 digit alphanumeric LCD oize Approximately 7 5 L x 4 25 W x 2 5 H or 19cm L x 10 8 cm W x 6 4 cm H Weight About 1 4 pounds 0 6 kg Page 6 of 30 Casework Molded plastic Battery Alkaline battery pack of four 1 5V AA cells Lasts 1 year with 1 minute sample interval PC Requirements for JCS JCS Lite is compatible with Windows XP and Windows 2000 Lite JCS Lite is not compatible with Windows Vista or Windows 7 The PC must have an IR port or a USB IR adapter Water Resistance The case is watertight and will float However do not submerge the instrument The sensor surface must be dry to read gas Certification Standards UL 913 Sixth Edition C22 2 No 157 92 UL Classified for Hazardous Locations in the USA and Canada File Number E235395 Warranty 1 year factory parts and labor Factory Calibration Annual sensor check for response and accuracy and replacement of battery Field Calibration Certified standard gas Reco
7. at the time of delivery and for a period of one 1 year thereafter to be free from defects in material or workmanship and shall conform to seller s specifications or such other specifications as seller has agreed to in writing Seller s obligations with respect to claims under this warranty shall be limited at seller s option either to the replacement of defective or non conforming product or to an appropriate credit for the purchase price thereof subject to the provisions of seller s Warranty Policy as amended from time to time said Policy being incorporated herein by reference Returned products under warranty claims will be shipped to seller s plant by buyer at buyer s expense and shall be accompanied by a statement of the reason for the return and an approved Return Material Authorization Number issued by seller Buyer remains responsible for payment for products not accepted for warranty adjustment handling costs and freight costs associated therewith Notwithstanding the foregoing no warranty shall be enforceable in the event that product has been subjected to environmental or stress testing by buyer or any third party without written approval of seller prior to such testing Further no warranty shall be enforceable if the alleged defect is found to have occurred because of misuse neglect improper installation repair alteration accident or improper return handling procedure by buyer Discontinued product is warranted only for a cred
8. capture data Place the Jerome 860 and the IrDA receiver with their IrDA ports facing each other about 6 apart Refer to DATA TRANSFER on page 22 for details Erases all previously logged data Enter the calibration gas concentration from the gas bottle This value is used for the SPAN calibration For calibration Ensure the 860 s temperature is stable Wait four 4 minutes from turn on in a known clean environment 0 ppm and then SELECT ZERO NOTE ZERO can be used to re zero the 860 without calibration gas but do not press zero unless the 860 s temperature is stable Hefer to CALIBRATION on page 13 for details Calibration After zeroing connect the 860 to the calibration gas wait four 4 minutes after turning gas flow on and SELECT SPAN The 860 will be calibrated to the previously entered concentration value and the field calibration date is updated Hefer to CALIBRATION on page 13 for details Page 11 of 30 MENU SUBMENU SPAN CHK INFO BAT VOLT FLD CAL FACT CAL BAT CHG SYSTEM DATE TM time TEMP C F ID RESET FACTORY If there is no user interaction and the 860 sits idle for 8 minutes it will automatically turn OPTIONS Calibration Check Connect the 860 to the calibration gas wait four 4 minutes after turning the gas flow on and SELECT SPAN CHK The reading is recorded along with the date and this reading can be viewed in the downloaded data file If the 860 is going to be left
9. in the field a SPAN calibration can also be performed immediately after the SPAN CHK Refer to CALIBRATION on pages 13 15 for details Displays the battery pack s cell voltage e New battery voltage can be up to 1 6V e Low battery voltage is 1 05V to 0 95V e Critically low battery voltage is below 0 95V View date of last field calibration View date of last factory calibration View date of last battery change Edit or enter the current month day year hour minute and second Select temperature displayed in either F or C Factory default is F Settings used to identify the 860 e INSTRMNT contains up to 40 characters that are used to identify the 860 instrument or location e The SN contains the serial number of the 860 Do not edit e The GAS contains the gas type being monitored H2S Do not edit this field Hesets the 860 and clears any errors Factory use only DO NOT EDIT OR CHANGE DATA or SETTINGS Itself off unless it is logging If the 860 is logging as indicated on the display the instrument will NOT shut off automatically Page 12 of 30 7 CALIBRATION Calibrate the Jerome 860 before each use to increase the accuracy of captured data Long term exposure to high gas concentrations can cause the sensor to drift downwards To verify sensor calibration perform an optional calibration check at the end of each use The following items are used during 860 calibration and supplied in
10. F accept the agreement do not accepi the agreement 4 f you agree to the terms of the license Select Destination Location S el e ct 8 acce pt th e ag reemen t an d Cl ick Where should Arizona Instrument JCS Lite be installed N ext Setup will install Arizona Instrument JCS Lite into the following folder gt The destination directory dialog will To continue click Next IF you would like to select a different folder click Browse display It is recommended that you icc ries accept the default settings At least 10 8 MB of free disk space is required Page 19 of 30 5 Click Next to continue gt The start menu dialog will display This will determine where the JCS Lite Start shortcut will be placed in your start menu It is recommended you accept the default settings 6 Click Next to continue gt The select additional tasks dialog will display Check the Create a desktop icon box if you want to create a shortcut to JCS Lite on your desktop Check the Create a Quick Launch icon box if you want to create a shortcut on your Taskbar 7 Click Next to continue gt A screen with your selected items is displayed Verify the information is correct 8 Click Install to install JCS Lite using these settings gt The installation of JCS Lite will start Page 20 of 30 Select Start Menu Folder Where should Setup place the program s shortcuts Setup wi
11. OTE The CD also contains PocketPC software and a link to install Microsoft ActiveSync It is not necessary to install the PocketPC software or ActiveSync PocketPC s are no longer available from AZI and they are not necessary to use the 860 The software is included on the CD only to support previous 860 customers who may still be using a PocketPC 1 Remove the CD from its protective cover and place it into the computer s CD drive 2 On most Windows operating systems the following display should appear AI Arizona Instrument Jerome 860 Software Installation Disk Select a link below to Install m l Arizona bns trument acd Microsoft ActiveSync rec i If the above link does not work search for ActiveSync at Install PocketPC Software watts microsoft com Install IC S Lite This CD also contains documentation in PDF format Adobe Acrobat Reader 4 0 or better is required 3 Ifthe setup program does not start within a few minutes gt Browse to the CD ROM drive using My Computer Windows Explorer or Start Run Double click on Install htm to display the above intro screen Page 17 of 30 1 From the intro screen gt Click on the Install JCS Lite icon some computer operating systems will Iu EUER DS a show a security warning If the CD Bey MN came from AZI or one of its authorized distributors in the same packaging as the Jerome 860 or was received directly through the mail cl
12. arameter e Press ESCAPE to move left or not save the parameter e Press UP or DOWN to change the alphanumeric digit Page 9 of 30 After the 860 completes the initial turn on display use UP or DOWN to scroll through the STATUS items Press SELECT and then use UP or DOWN to access the four main menu items STATUS LOG CAL INFO and SYSTEM Press SELECT again to choose a menu item Each of these main menu items have sub menu items as listed in the following diagram and explained in the table on the following pages 860 MENU STRUCTURE LOG CAL INFO SYSTEM BEGIN CONC BAT VOLT DATE TM INTERVAL ZERO FLD CAL TEMP C F OUTPUT SPAN FACT CAL CLEAR SPAN CHK BAT CHG RESET FACTORY Page 10 of 30 MENU STATUS Available at turn on and Escape LOG CAL SUBMENU GAS PPM MONTH DAY TIME TEMP BEGIN INTERVAL OUTPUT CLEAR CONC ZERO SPAN OPTIONS Measured Gas Concentration Current Date Current Time in a 24 hour clock format Current Temperature in F or C Begin logging LOGGING is displayed on the LCD The instrument continues logging until ON or OFF is held down When logging is stopped the 860 will perform the INITIAL TURN ON display Refer to DATA LOGGING SETUP on page 16 for details Select a logging interval of 2 10 30 sec 1 2 5 10 30 min 1 hour Factory default is 1 minute Output the logged data to the PC Place the 860 on any flat surface Set JCS Lite to
13. ates from 11 1un 2003 11 10 40 To 11 Jun 2003 11 25 22 Ceen Das Sun J s Display Times from nog 33 59 9 1 C Hon Display Days from gt Temperature scale and m Upper Limits gt C ursor type V Display Upper Limits 7 Mark Out of Limit Points with Wide Line __SetUpperLimit Click To Change Colors r Graph Line Color Sequence Axes Background Clipboard Graph Background Restore to 00 C0 00 ER Title mmm Default m 2 gt mmm Color 1 80 80 00 ED 00 00 FF Delete Color 1 FF FF 00 Insert Before Color 1 DO FF FF Change Color 1 FF O0 FF Copy Color 1 to ClipBoard AE 1 Jun 2003 Insert After Color 1 Program Options Page 25 of 30 AZI Jerome 860 JCS Lite File 860AZI PREFERRED CUSTOME 9 he Program Option allows the erom user to TE cin m Maximize the graph on the zl He alal 51s e zl 18 Screen gt Indicate by message that ee erie data files are being saved v Scale Graph to Fit Paper Size when Printed gt Fit the graph to the selected orinter paper size and Data File Storage determine which directory r Use System Data Fie Access within the computer will be Oe or vate eee eee ea used for file storage sete Use Speciied Decoy Directory C Documents and Settings djohnson My Documents Browse Page 26 of 30 The battery is considered field replaceable however extreme care must be take
14. bitmaps Press OK or Exit when done a Annotate Graph Op a l x Page 24 of 30 5 Click on the tabs at the bottom of the window to change the display between Graph Data Graph Options and Program Options 1 k Graph J Data Graph Options Fragram Options 6 The Graph tab presents the graph as shown above ATA AA A CNN TENE UN in spreadsheet format slala lull Aa ela ts el zi ial AZI PREFERRED CUSTOMER Serial Number 860000300 Gas H25 Date JUN 11 2003 11 33 20 Result DATE TIME GAS ppm Temp E MarkerInfo 11 Jun 2003 11 18 08 11 Jun 2003 11 18 12 11 Jun 2003 11 18 16 i Find Point Selected On Graph 11 Jun 2003 11 18 20 l Print Data 11 Jun 2003 11 18 24 11 Jun 2003 11 18 28 11 Jun 2003 11 18 32 11Jun 2003 11 18 36 11 Jun 2003 11 18 40 11 Jun 2003 11 18 44 8 The Graph Options allow the graph Maei to be customized by leli eas HA ale telele alel gt C O O r Mark Data Points on Graph J Scale Y Axis when zooming p Selected Point Cursor J m Display Temperature in 34 C Diamond C Circle Square C Filled Box Celsius C Farenheit F C Vettical Line Horizontal Line Horizontal amp Vertical Lines gt S e e cte d d ata by d ate a n d m Display Period Options Graph Type 1 Display All Data Continuous Graph p Stat Week Gn tl I e l nte rval S C Display D
15. cesckeddasccsaessesese 27 WARRANEI oaan E I EL 29 Page 3 of 30 1 INTRODUCTION TO THE JEROME 860 The Jerome 860 is a portable gas monitor that detects hydrogen sulfide gas A menu driven program allows you to change displayed information and operating parameters With the Jerome Communication Software JCS Lite installed on a computer data may be collected directly from the Jerome 860 for graphing and reporting There are two models in the 860 family with different detection ranges Model otandard Detection Range opike Reading Capability X860 0001 Up to 200 0 ppm Up to 1000 0 ppm X860 0002 Up to 50 0 ppm Up to 500 0 ppm Refer to SPECIFICATIONS on page 6 for more details on the two models This manual contains details that will optimize the results and the life of your Jerome 860 Read and refer to the manual for complete details on operation maintenance and troubleshooting The Jerome 860 is easy to operate and ready for use upon receipt from the factory e Remove the instrument from the packing material Retain all packaging materials for any future shipment of the instrument If the instrument is returned to AZI for any reason it must be placed in the original packaging materials that have been tested and proven to be effective protection during shipment e Call AZI Customer Service at 800 528 7411 or 602 470 1414 for Return Material Authorization RMA information prior to returning a unit e For all shipments
16. e to downloading the same data more than once Refer to DATA TRANSFER on page 22 for details on downloading data pt CAUTION The CLEAR action is non reversible Ensure the desired data has downloaded without error and is available on the PC before clearing the logged data in the instrument The data cannot be returned to the Jerome 860 from the PC From the STATUS display press SELECT to access the menu When LOG is displayed press SELECT again to access the LOG menu Press DOWN repeatedly until CLEAR is displayed Press SELECT and then press SELECT again to CONFIRM The 860 will display CLEARING and then CLEAR Press ESCAPE to return to the LOG menu DS S Data can also be cleared from the instrument by the PC if JCS Lite is configured for the automatic clearing action when data is downloaded successfully To enable this feature in JCS Lite the Delete Data icon on the Program Options tab must be checked Page 16 of 30 11 LOADING SOFTWARE The CD included with the 860 contains this manual and the JCS Lite software The JCS Lite software must be loaded on the computer in order for the sample data to be transferred from the Jerome 860 to the PC for graphing and reporting The computer either desktop or laptop must have e Microsoft Windows XP or Windows 2000 e An IR port either built in or through a USB IR adapter e JCS Lite installed the installation is detailed in this section PLEASE N
17. emove Tolerance Marks Out of Range Data Refresh Plot Lok d ES BU m V ge fo thems i y p Page 23 of 30 2 x Plot Manipulation B AZI Jerome 860 JCSLite File 860AZzI PREFERRED CUSTO Bl x File Graph Help al Sis Ee gg amp lal telel l Sl 6 Vv LJ ri a d Rn bs 5 5 5 11 Jun 2003 11 Jun 2003 11 Jun 2003 11 Jun 2003 11 10 40 11 15 11 20 11 29 Data Period From 11 Jun 2003 11 10 40 to 11 Jun 2003 11 25 22 Gas Temperature Gas Limits Graph Options Program Options 1 Clicking a spot on the graph of sampled H2S or temperature readings will draw a box or other shape as selected in Graph Options around the closest actual data point Information about that data point will be displayed in detail near the bottom of the window 2 Use the arrow buttons gt to move the graph or click on the centering button 4 to place the selected data point in the center of the display 3 Use the Zoom In Zoom Out buttons rto change the amount of data visible on the Print Plot screen gt To expand the Y Axis the Scale Y axis Aree Estreno Diele li Options tab must be checked DE 4 Right click on a data point to annotate the Remove a Marker graph Clicking the right mouse button will bring up the Annotate Graph Options menu shown Use this menu to add comments or bitmaps to a data point or to the graph or to selectively remove comments or
18. he property of their respective owners Lysol is a registered trademark of Reckitt Benckiser Inc Tygon is a registered trademark of Saint Gobain Performance Plastics Corporation Windows and ActiveSync are registered trademarks of Microsoft Corporation in the United States and other countries Arizona Instrument LLC Jerome 860 Part Number 700 0051 D January 2011 If you have any questions regarding the operation of this instrument please call our toll free number from the USA and Canada 800 528 7411 or e mail at support azic com Internationally call 602 470 1414 or fax 480 804 0656 Page 30 of 30
19. hout calibration gas Page 14 of 30 Calibration Check Procedure After each period of sampling a calibration check should be performed to determine the accuracy of the sampling operation The difference between the calibration check and the calibration concentration as listed on the calibration gas cylinder can then be used to determine the percentage of error if any developed during the sample period To perform the calibration check 1 2 3 Do not install the calibration plug at this point Turn the Jerome 860 ON in a known clean environment 0 ppm Wait four 4 minutes from power up gt Verify that the gas reading is near 0 0 PPM 4 Install the calibration plug into the sensor socket 5 6 Wait four 4 minutes Turn on the calibration gas Hecord the displayed gas reading and compare it to the calibration gas concentration Optional To log the reading and date for the Calibration Check SELECT the CAL SPAN CHK menu item at this point in the procedure Turn off the calibration gas and disconnect the calibration plug from the instrument To calculate the percentage error in the previous samples use this equation Calibratio nGasConcentration CalibrationCheck Reading x100 PercentError CalibrationGas Concentration For example a calibration check reading of 24ppm with a calibration gas concentration of 25ppm would indicate a final error of 4 25 24 00 4 8 CLEANING
20. ick on the YES button If you are still not able to install the software Yes e Check that your browser s security settings allow you to download software If your security settings are set too high you will not be able to download any executable programs by way of your browser If you set them at a medium level you will be able to download your software e Please see the Help menu in your browser for more information on security If you are still unable to install the software e Use My Computer to select the CD drive containing the AZI CD e Double click the JCSLite directory e Double click on setup exe Call AZI Customer service at 800 528 7411 602 470 1414 or e mail at support 2azic com for assistance The publisher cannot be determined due to the problems Authenticode signature not found Page 18 of 30 2 he welcome message will display Z Setup Arizona Instrument JCS Lite Arizona Instrument JCS Lite Welcome to the Arizona Instrument JCS Lite Setup Wizard 3 Click Next to continue the JCS Lite C M installation gt The license agreement will display Please read the license agreement on carefully License Agreement Arizona Instrument LLC Software License Aqreement for JCS Lite PLEASE READ THE TERMS OF THIS SOFTWARE LICENSE AGREEMENT LICENSE WHICH IS PRESENTED IELECTRONICALL Y WHEN INSTALLING THE SOFTWARE BY ICI ICKING THF I Accent tha Anrsemant AUTTON YOU AR
21. instrument in the back of a pickup truck to allow the instrument to temperature stabilize prior to field calibration Page 13 of 30 Calibration Procedure ELE NOTE CAL ZERO can be used to re zero the instrument Do not install the calibration plug at this point Enter the calibration gas concentration in the CAL CONC menu item This value is used for the SPAN calibration See USER INTERFACE on page 9 for additional detail on navigating the menu system Turn on the instrument in a known clean environment 0 ppm and wait at least four 4 minutes SELECT the CAL ZERO menu item and press SELECT again to CONFIRM CAL ZERO is momentarily displayed and the 860 displays the PPM status item If you attempt to CAL ZERO when gas is applied CAL FAIL is momentarily displayed and the action is aborted gt The instrument should display 0 0 PPM Install the calibration plug into the sensor socket Turn on the calibration gas Wait an additional four 4 minutes and then SELECT the CAL SPAN menu item press SELECT again to CONFIRM gt CAL SPAN is momentarily displayed and the 860 displays the PPM status item If you attempt to CAL SPAN when gas is not applied CAL FAIL is momentarily displayed and the action is aborted gt The instrument is now calibrated with the concentration value entered above Turn off the calibration gas and remove the calibration plug from the instrument wit
22. it or replacement at seller s option THE EXPRESS WARRANTIES GRANTED ABOVE SHALL EXTEND DIRECTLY TO BUYER AND NOT TO BUYER S CUSTOMERS AGENTS OR REPRESENTATIVES AND EXCEPT FOR WARRANTY OF TITLE IS IN LIEU OF ALL OTHER WARRANTIES WHETHER EXPRESSED OR IMPLIED INCLUDING ANY IMPLIED WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE AND MERCHANTABILITY SUCH OTHER WARRANTIES BEING SPECIFICALLY DISCLAIMED BY SELLER IN NO EVENT SHALL EITHER PARTY S LIABILITY FOR ANY BREACH OR ALLEGED BREACH OF THIS AGREEMENT EXCEED THE TOTAL EXTENDED PRICE OR PRICES SHOWN ON UNFILLED ORDERS NOR SHALL EITHER PARTY BE LIABLE FOR ANY SPECIAL INCIDENTAL OR CONSEQUENTIAL DAMAGES RESULTING FROM BREACH OR ALLEGED BREACH Notwithstanding the foregoing if any product covered by order s placed hereunder is designated as developmental prototype or experimental no warranty whatsoever except a warranty of title to component materials will be applicable thereto and buyer shall indemnify seller for any claims for liability asserted seller in connection therewith The foregoing state the entire liability of seller in connection with products supplied hereunder Page 29 of 30 TRADEMARK AND COPYRIGHT PROTECTION Copyright 2003 2010 Arizona Instrument LLC All Rights Reserved All specifications subject to change without notice Jerome Arizona Instrument AZI and the stylized AZI are all registered trademarks of Arizona Instrument LLC All other trademarks are t
23. ll create the program s shortcuts in the following Start Menu folder To continue click Next If you would like to select a different folder click Browse Arizona Instrument Lite Select Additional Tasks Which additional tasks should be performed Select the additional tasks you would like Setup to perform while installing Arizona Instrument JCS Lite then click Next Additional icons Create a desktop icon Create a Quick Launch icon Ready to Install Setup is now ready to begin installing Arizona Instrument JCS Lite on your Ly computer Click Install to continue with the installation or click Back if you want to review or change any settings Destination directory C Program Filesszl Start Menu folder Arizona Instrument JCS Lite Additional tasks Create a desktop icon Create a Quick Launch icon Installing Please wait while Setup installs amp rizona Instrument JCS Lite on your computer Registering files Cancel 9 When the installation is complete the Setup Complete d ialog will appear Instrument JCS Lite Setup Wizard 1 0 Check the au nch Arizona nstru ment JCS Setup has finished installing Arizona Instrument JCS Lite on your computer The application may be launched by selecting the installed icons Lite box to start JCS Lite immediately or uncheck the box if you are not ready to run JCS Lite 11 Click Finish Click Finish to exit Setup NOTE On so
24. made A running total of downloaded bytes will be displayed EU Received 5455 Bytes 7 As soon as all data is transferred the graph of that data will display 04 Jan 2010 05 Jan 2010 06 Jan 2010 Graph Options Program Options Page 22 of 30 1 Open JCS Lite on the PC using the Start menu Start Programs AZI JCS Lite or the shortcuts on the desktop or Taskbar AZI Jerome 860 JCS Lite File Graph Help 2 Click on File and then Open Data File on PC or click the eer Open A Data File button in the center of the JCS Lite Read Data From Jerome 860 window ee Merge Multiple Files Into Single File Configure Program Print Plat Printer Setup Save Data File As Gave Plot As Exit 3 The list of data files stored on the computer will display similar to the illustration at right Lc e ck Ee 9860000300_6 6 2003 9 22 43 c s 860000300_6_6_2003_12_33_ m 4 Double click on the desired file to view its contents in JCS Lite us Hes sal ajal t lel gt 2l ml Icon Meaning Open an existing data file on the PC or connected network Get data from Jerome 860 Instrument Read Data File From PocketPC not typically used save Graph to Disk as a BMP or PNG file oave Data to Disk Print Graph Center Highlighted Data Point in Graph Set Active Y Axis Zoom In Zoom Out Move Viewing Area Slowly nudge Up Down Left Right Restore Original Graph Scale Add R
25. me Windows operating systems a box may be displayed asking if you want to reboot or restart the computer at this time Select YES to restart and finalize the installation 12 The first time JCS Lite is run the following screen will typically appear As explained above in most situations the PocketPC and ActiveSync are not being used so click OK Do Not Ever Warn Me Again to avoid seeing this message in the future m Active Sync Not Installed Activesync Is Not Properly Installed And The PC Will Not Be Able To Communicate With The Pocket PC Device UF Do Not Ever Wan Me Again Page 21 of 30 12 DATA TRANSFER Jerome 860 to PC To transfer data from the 860 to the PC 2 1 Open JCS Lite on the PC using the Start menu Open A Data File Start Programs AZI JCS Lite or the shortcuts on the desktop or Taskbar 2 The three open data options will be displayed in the center of the window Click on the Read Data From Jerome X 860 button NOTE Ensure that no other IR devices such AT Em LE as V remotes etc are operating in the area or the data will be interrupted 3 The Waiting for Jerome 860 to connect over IRDA eee prompt will display 4 On the 860 navigate to the LOG OUTPUT menu item and press SELECT then press SELECT again to CONFIRM e 5 Place the 860 about 6 inches from the computer s IR port as shown 6 The data will transfer when the IR connection is
26. mmended before each use Field Replaceable Parts Battery Pack However extreme caution is required to protect the intrinsically safe fuse any time the case is open The slightest short will destroy the fuse which will require factory replacement of the safety barrier Accessories Field Calibration Kit All specifications subject to change without notice 4 SENSOR CROSS SENSITIVITY INTERFERENCES The sensor response to various potentially interfering gases is detailed below and on the next page X860 0001 X860 0002 pian 200 PPM SENSOR 50 PPM SENSOR Carbon 300 ppm lt 1 5 ppm Monoxide Sulfur Dioxide 35 ppm Dioxide 0 1ppm toO ppm 0 2ppm Hydrogen 10 000 ppm 15 ppm Hydrogen 10 ppm 1 4 ppm to 0 5 ppm 0 9 ppm to 0 3 ppm Cyanide Hydrogen 5 ppm 0 ppm 0 ppm Chloride Ethylene 100 ppm For example in a 300 ppm carbon monoxide environment H2S readings displayed by the X860 0001 may be elevated by up to 6 ppm from the actual H2S level present Page 7 of 30 To power on the 860 press and hold the SELECT button for 1 second e At power UP the display flashes in order gt AZI gt J860 H S gt VX XX where X XX is the version number of the 860 s operating program and gt X X PPM where X X is the detected concentration of H2S If only a line of dashes is displayed the SELECT button is not being held long enough Hold the SELECT button until AZI a
27. n whenever the instrument is open The slightest electrical short on the circuit board in the instrument will blow the intrinsically safe fuse If this occurs the instrument will not function until the safety barrier is replaced at the factory The date and time instrument serial number and desired temperature display units F or C will need to be reentered after the battery is replaced To change the battery e Flip the instrument over and remove all 10 screws securing the two halves of the case together e Flip the instrument back over and carefully lift off the front case Note the two halves of the case are connected by a cable Page 27 of 30 e Remove the four screws securing the battery cover and remove it e Depress the connector to unplug the battery and then remove it e Plug the new battery into the connector e Make sure the two desiccant pouches remain in the positions shown and then reassemble the instrument by reversing the above steps After assembly is complete turn on the instrument to verify that the intrinsically safe fuse is intact If the instrument powers on the fuse is intact The date and time instrument serial number and desired temperature display units F or C will need to be reentered Refer to USER INTERFACE on page 9 for guidance Page 28 of 30 15 WARRANTY Arizona Instrument LLC seller warrants to buyer that Jerome products delivered pursuant to this agreement
28. ppears Page 8 of 30 Status Items Normally the instrument displays the gas reading after power up or exiting the menu system Press UP and or DOWN to view the other items date time and temperature If the battery is low 1 049V to 0 95V cell voltage a LOW BAT warning message is displayed Other items may still be viewed with the UP DOWN buttons and the instrument continues to operate When the low battery warning first occurs there is approximately 1096 battery capacity left If the battery voltage is below 0 95V a system ERROR message is displayed When a system ERROR occurs it is displayed on the status line and the instrument stops operating The most likely system error is a critically low battery voltage of less than 0 95V Contact AZI Customer Service at 800 528 7411 602 470 1414 or e mail at support 2azic com for details and assistance Menu Items The following key presses are used to edit parameters e Press SELECT to enter the menus e Press UP or DOWN to scroll through the menu items e Press SELECT to pick a submenu parameter for editing or an action item Action items like LOG or OUTPUT require a second press of SELECT to CONFIRM the action Press ESCAPE to cancel the action or selection e Press ESCAPE to exit a menu Parameter Editing The following key presses are used to edit the parameters e Press SELECT to move right or save the p
29. rate Note the sensor can operate for a short time without oxygen for example during a calibration with H2S in nitrogen e n extreme humidity 80 to 100 periodically remove the instrument from operation every 2 weeks to allow the sensor to stabilize in an H2S free environment for 48 hours Water buildup in the sensor occurs at high humidity and the sensor seals will burst if not allowed to expel the water in a dry environment e In high temperature high humidity environments periodically remove the instrument from operation every 2 days to allow the sensor to stabilize e f measurement accuracy is important calibrate the unit before each use and check the calibration at the end of the logging period e The battery is considered field replaceable however extreme care must be taken When the instrument is open the slightest electrical short will destroy the intrinsically safe fuse If this occurs the safety barrier must be replaced See Section 14 BATTERY REPLACEMENT on page 27 for battery replacement instructions e During the annual factory calibration the battery is replaced Because the battery will last for a little over one year at one 1 minute logging intervals the battery should not need field replacement between factory calibrations e f an enclosure seal or surface is damaged return the instrument to AZI for repair H2S gas and water are corrosive and will corrode the circuitry if allowed to intrude through a damaged seal
30. the 860 Accessory Kit AZI P N Y990 0188 and or 860 Calibration Kit AZI P N Y990 0184 e 25 ppm Hes Calibration gas AZI P N 340 0010 contains 25ppm H S in nitrogen The bottle contains 57 liters and should last for at least 38 calibrations e 100 ppm H2S Calibration Gas is also available AZI P N 340 001 1 e The Regulator AZI P N 345 0159 is screwed onto the gas bottle and is reusable The regulator provides 0 3 L minute of gas e The Calibration plug AZI P N 600 0233 is attached to the regulator using the included 1 8 ID Tygon tubing AZI P N 2500 3001 e The Cal Gas Storage case AZI P N 990 0182 holds the Calibration Kit WARNING The calibration gas contains H5S The personal exposure limit is 10 ppm DO NOT inhale the output of the calibration bottle This low concentration can cause headache dizziness or nausea Higher concentrations 200 300 ppm can result in respiratory arrest leading to coma or unconsciousness Exposures for more than 30 minutes at concentrations of greater than 700 ppm have been fatal Continuous inhalation of low concentrations of H2S may cause olfactory fatigue or paralysis rendering the detection of its presence by odor ineffective The calibration should be performed under a fume hood or in a well ventilated area such as outdoors If the calibration is performed outside allow the instrument to equilibrate to the ambient temperature before calibration If traveling to a site store the

Download Pdf Manuals

image

Related Search

Related Contents

Manual de Instalação do Projuris 7  Capitolato speciale d`appalto - ALLEGATO C  User Manual - Larson Systems Inc.  Big Brain Academy - Manual  Samsung HT-910 User Manual  Liebert Mini-Mate2™ - PTS Data Center Solutions  ATW-T107/P 取扱説明書  AVAYA 3904 TELEPHONE THE 3904 TELEPHONE  JW-101ZT/102ZT/103ZT/104ZT/105ZT/106ZT/107ZT 取扱説明書  Installation, Betriebs- und Wartungsanweisungen  