Home
RKI Eagle 2 Intrinsically Safe Portable Gas Detector User Manual
Contents
1. This probe is designed for use where it is necessary to put the probe tip in areas not accessible with the 32 inch telescoping probe with dust filter and applications where the probe tube must be collapsible for storage See General Parts List on page 84 for probe ordering information 20 Optional Accessories EAGLE 2 Operator s Manual External Dilution Fittings Two external dilution fittings are available for the EAGLE 2 a 1 1 dilution fitting and a 3 1 dilution fitting They are designed to mate with the inlet fitting and accept a sample hose or probe The fittings are made with brass and nickel plated brass and are appropriate for use with the four standard gases The 1 1 fitting is normally used when it is necessary to introduce air into a sample that has no oxygen or a very low level of oxygen such as a nitrogen purged sample Both the 1 1 and 3 1 fittings can also be used when one of the target gas levels in the sample area will likely be present in a concentration above the detection range for that gas Since the fittings partially consist of unplated brass they are not appropriate for detection of elevated levels of H S or of gases that are easily absorbed such as Cl or SO TIL 1 1 Dilution Fitting 3 1 Dilution Fitting Figure 14 1 1 and 3 1 Dilution Fittings EAGL
2. _ m pee CoA a m AN VA Oxygen Oxyg en Sensor A Sensor O oy Bracket CO2 Scrubber L LEL Sensor and Sensor Bracket Figure 38 Typical Infrared CO Sensor Location The infrared CO sensor is installed in a single sensor flow chamber which is located in the area next to the standard 4 sensor flow chamber This area can accommodate up to three single sensor flow chambers Figure 38 above illustrates a typical infrared CO sensor location in front of the pump The infrared CO flow chamber may also be installed in one of the other two sensor chamber locations depending on the particular version of the EAGLE 2 Some infrared CO instrument configurations do not include the 4 sensor flow chamber Infrared CO Sensor The infrared CO sensor is a cylindrical sensor with a diffusion opening on the front and a 12 pin circular connector on the back A 12 wire cable plugs into the back of the infrared CO sensor with a circular connector that includes a locking lever The other end of the cable plugs into an infrared sub PCB see description below that is installed on the main PCB The sensor is held in the infrared flow chamber by a bracket on standoffs Infrared Sub PCB The infrared sub PCB is a circuit board that is installed on the main PCB in 224 Description EAGLE 2 Operator s Manual one of the 3 sub PCB positions when an infrar
3. 1 From the main menu place the cursor in front of CHANGE PASSWORD 126 Using Setup Mode EAGLE 2 Operator s Manual 2 Press and release POWER ENTER RESET The Password Protection Screen appears PASSWORD PROTECTION UP DOWN THEN ENTER OFF 3 Use AIR A YES or RANGE YW SHIFT to display the desired setting 4 If you selected OFF press and release POWER ENTER RESET to save the setting and return to the main menu If you selected ON continue with Step 5 5 Press and release POWER ENTER RESET The Set Password Screen appears The factory set pass password of 0000 is at the bottom of the screen with the first O flashing SET PASSWORD 0000 Use AIR A YES or RANGE YW SHIFT to display a number from 0 to 9 Press and release POWER ENTER RESET to enter the selection and advance to the next number 8 Repeat Step 6 and Step 7 to select the remaining numbers When you press and release POWER ENTER RESET to enter the last number the password is saved and you return to the main menu Restoring the Default Settings Each of the EAGLE 2 setup parameters such as the auto calibration values zero and span settings or parameters in Setup Mode has a default setting For the items in Setup Mode the default settings are the same as the standard factory settings If you want to return the EAGLE 2 to its default configuration it is possible to do so by using the Default Settings menu item in Setup Mo
4. Specifications Table 1 Standard Sensor Specifications Combustible Gas Oxygen Os Hydrogen Carbon Methane CH Sulfide H2S Monoxide CO Calibration Standard Detection 0 100 LEL 0 40 volume 0 100 0 ppm 0 500 ppm Range Reading 1 LEL 0 1 volume 0 5 ppm 1 ppm Increment Alarm 1 10 LEL 19 5 volume 10 0 ppm 25 ppm Factory Setting Alarm 2 50 LEL 23 5 volume 30 0 ppm 50 ppm Factory Setting When calibrating the EAGLE 2 with the Auto Calibration or the Single Calibration method the calibration gas value must be equal to or higher than the alarm 1 setting See Updating the Alarm Point Settings on page 114 for instructions to change the alarm points if necessary for the desired calibration gas value Table 2 EAGLE 2 Specifications Sampling Sample Draw Method Response T90 Within 30 Seconds Time Display Graphics LCD Display Operating 20 C to 50 C Below 85 RH Without Condensation Temperature amp Humidity Indication Combustible Gas LEL Catalytic Type Sensor Accuracy e 10 C to 40 C 5 of full scale e 20 C to 50 C 6 of full scale Combustible Gas ppm Catalytic Type Sensor e 25 ppm or 5 of reading whichever is greater Oxygen e 0 5 Oo Hydrogen Sulfide e 5 of reading or 2 ppm H3S whichever is greater Carbon Monoxide e 5 of reading or 5 ppm CO whichever is greater EAGLE 2 Operator s Manual Specifications 3 Safety Regulatory T 186718
5. 226 Start Up and Normal Operation EAGLE 2 Operator s Manual 5 Put the caps back on the scrubber Infrared CO Calibration An infrared CO channel can be calibrated using the auto calibration method or the single calibration method To calibrate an infrared CO channel using the single calibration method see Calibrating Using the Single Calibration Method on page 64 and follow the instructions for calibrating a single channel If your instrument is a multi channel instrument that includes one or more infrared CO channels RKI Instruments Inc recommends using the auto calibration method for convenience The calibration instructions below show a 5 channel instrument which has the four standard channels LEL oxygen H S CO and an infrared CO channel for 0 5 vol To use the auto calibration method to calibrate a multi channel instrument that includes an infrared CO channel do the following NOTE If your instrument has more than one infrared CO sensor you will need a calibration cylinder for each sensor 1 See Calibration Supplies and Equipment on page 56 for a description of the necessary calibration supplies In addition to an appropriate multi gas cylinder that is used to calibrate any active standard channels you will also need a cylinder to calibrate the infrared CO channel See Table 25 on page 237 for available cylinders Make sure your calibration cylinder is appropriate for the infrared CO detection ra
6. CSA classified as Intrinsically Safe Exia Class Groups A B C amp D Temperature Code T3C Power e Four C size alkaline batteries standard Supply e Four C size Ni MH batteries optional Continuous Alkaline Batteries 16 Hours Non Alarm Operation Fully Charged patie e Ni MH Batteries 18 Hours Non Alarm Operation Fully Charged ours 25 C High impact Plastic RF Shielded Dust and Weather Proof Standard 5 foot hose Accessories Hydrophobic probe e Shoulder Strap Optional Rechargeable NiMH Batteries Accessories 115 VAC Charger 12 VDC Charger Hoses of Various Lengths See General Parts List on page 84 Dilution Fitting 1 1 and 3 1 Various Probes See General Parts List on page 84 Data Logger Management Program Windows 2000 XP and Vista Maintenance Data Loader Program Windows 2000 XP and Vista IrDA USB Cable for connecting to a computer when using the Data Logger Management Program and Maintenance Data Loader Program not needed if computer has an infrared port Dimensions Approximately 171 H x 65 W x 39 D mm 5 6 H x 2 5 W x 1 5 D and Weight Approximately 310 g 11 oz About this Manual The EAGLE 2 Operator s Manual uses the following conventions for notes cautions and warnings NOTE Describes additional or critical information CAUTION Describes potential damage to equipment WARNING Describes potential danger that can result in injury or death
7. Chapter 4 Calibration Mode on page 55 for instructions to calibrate the EAGLE 2 When you are done with the calibration and exit Calibration Mode the unit will begin the startup sequence If the calibration was successful the screen above will not appear again until the unit is due for calibration If the calibration was not successful the screen above will again appear in the startup sequence e Ifthe unit is due for calibration and CAL PAST DUE ACT is set to NOTIFICATION ONLY then the following alert screen displays and the buzzer sounds in a double pulsing pattern CALIBRATION DATE IS PAST DUE Press and release the POWER ENTER RESET button to acknowledge the alert and continue with the startup sequence 282 Start Up Leak Check Mode EAGLE 2 Operator s Manual 12 The Date Time Screen appears for a few seconds 9 12 2008 15 00 00 13 If USER STATION ID is turned on see Turning the User Station ID Function On or Off on page 117 the ID Screen appears for a few seconds USER ID MIKE STATION ID PUMP 1 SERIAL NUMBER E2A515 If USER STATION ID is turned off only the serial number is shown 14 If the EAGLE 2 experiences a catalytic combustible sensor failure during start up a screen indicating which sensor failed appears and the buzzer sounds a pulsing tone twice per second In the example below the H S sensor has failed FAILED SENSOR S lt CH4 gt lt g
8. Dead Battery Alarm e Gas readings replaced by Double pulsing tone RECHARGE INSTRUMENT once per second BATTERIES Alarm LED arrays flash in a double pulsing pattern once per second Sensor Failure e FAILED SENSOR S appears Double pulsing tone at the top of the display and once per second the failed sensor s are indicated e Alarm LED arrays flash in a double pulsing pattern once per second Resetting and Silencing Alarms You can set the EAGLE 2 s gas alarms as latching or self resetting alarms see Updating the Alarm Latching Setting on page 116 e Self resetting alarms ALARM LATCHING set to SELF RESET Self resetting alarms automatically shut off and reset when the gas reading falls below or rises above for an oxygen low alarm the alarm setting You cannot reset self resetting alarms with the POWER ENTER RESET button You can set self resetting alarms with or without the alarm silence feature see Updating the Alarm Silence Setting on page 117 e Latching alarms ALARM LATCHING set to LATCHING Latching alarms will remain in effect until the gas reading falls below or rises above for an oxygen low alarm the alarm setting and they are reset with the POWER ENTER RESET button You can set latching alarms with or without the alarm silence feature see Updating the Alarm Silence Setting on page 117 ALARM SILENCE On and Alarms Set to LATCHING ALARM SILENCE set to ON and ALARM LATCHING set to LATC
9. CALIBRATION DATE IS PAST DUE PERFORM CALIBRATION To perform a calibration press and release the AIR A YES button The EAGLE 2 will enter Calibration Mode and the LCD will show the Calibration Mode main menu See Chapter 4 Calibration Mode on page 55 for instructions to calibrate the EAGLE 2 When you are done with the calibration and exit Calibration Mode the unit will begin the startup sequence If the calibration was successful the screen above will not appear again until the unit is due for calibration If the calibration was not successful the screen above will again appear in the startup sequence To continue without performing a calibration press and release the DISPLAY ADJUST NO button e Ifthe unit is due for calibration and CAL PAST DUE ACT is set to MUST CALIBRATE then the following screen displays and the buzzer sounds in a double pulsing pattern CALIBRATION DATE IS PAST DUE ENTER TO PERFORM CALIBRATION The EAGLE 2 cannot be used until a successful calibration has been performed Press and release the ENTER button to enter Calibration Mode See Chapter 4 Calibration Mode on page 55 for instructions to calibrate the EAGLE 2 When you are done with the calibration and exit Calibration Mode the unit will begin the startup sequence If the calibration was successful the screen above will not appear again until the unit is due for calibration If the calibration was not success
10. INSTRUMENTS EAGLE 2 Operator s Manual Part Number 71 0154RK Revision Q Released 12 4 13 www rkiinstruments com WARNING Read and understand this instruction manual before operating instrument Improper use of the gas monitor could result in bodily harm or death Periodic calibration and maintenance of the gas monitor is essential for proper operation and correct readings Please calibrate and maintain this instrument regularly Frequency of calibration depends upon the type of use you have and the sensor types Typical calibration frequencies for most applications are between 1 and 3 months but can be required more often or less often based on your usage EAGLE 2 Operator s Manual Table of Contents Chapter 1 Introduction 3 sieew eis ies Sarees Oa soe oe Gree ols 1 OVEVIEW oles eiked eas ou a aa entero ei eae eae ee Gea ee eet s 1 ADOUL Re EAGLE 2 aca cseier eae enue eee ea a e co3 tad Pee l Specification Sss saa eea Meena EE LD ETE REK ADE O EREE ee ween 3 About This Manya i ernennt e a E A wen EAEE E 4 Chapter 2 Description essssesssocscsssoosessooseeeoo 6 OVORVIEW s ernannte nan Vor etn hut edhe Oe ek a E O AN OEE nets 6 Instr ment Descriptions s aiea E a baa E ERA UREA 6 CASE naimen iea keea oo ea e deck dete ete Mis dale Bde ate 8 AN EE E dno ih EE AE E Ath dat T EEES 8 LCD REM rere a ee eee a T ee ere A ee N 9 Control BULIONS lt 0 Rad Fate Rade 24 Pik be bk tah BG Bhs 10 Printed C
11. LEL or Y volume range 3 Set the catalytic combustible channel s units to be ppm by using the RANGE YW SHIFT button 4 Perform a demand zero as described in Performing a Demand Zero on page 27 5 Proceed to monitor for gas as described in Monitoring an Area on page 30 Measuring Mode Alarms This section covers alarm indications in Measuring Mode It also describes how to reset the EAGLE 2 after an alarm has occurred and how to respond to an alarm condition NOTE False alarms may be caused by radio frequency RF or electromagnetic EMI interference Keep the EAGLE 2 away from RF and EMI sources such as radio transmitters or large motors EAGLE 2 Operator s Manual Measuring Mode Alarms 35 Alarm Indications The EAGLE 2 will sound an alarm and flash the LED arrays when one of the target gas concentrations rises above the Low Alarm level or in the case of oxygen falls below the Low Alarm level for that gas The EAGLE 2 also sounds an alarm and flashes the LED arrays when one of the target gas concentrations rises above the High Alarm level and when the STEL and TWA alarm levels are reached for CO and HS NOTE If an alarm condition occurs while you are in Display Mode the EAGLE 2 will automatically bring up the alarm screen instead When a failure condition occurs such as a sensor failure low flow or dead battery condition the unit will also sound an alarm and flash the LED arrays The table
12. A screen appears asking you to confirm that you want to restore the default configuration ARE YOU SURE YOU WANT TO SET DEFAULT CONFIGURATION 5 Press and release AIR A YES The screen will indicate that the default configuration has been restored and return to the main menu 128 Using Setup Mode EAGLE 2 Operator s Manual Updating the Lunch Break Setting With LUNCH BREAK set to OFF factory setting the EAGLE 2 automatically starts new TWA and PEAK reading collection and resets the time in operation at startup With LUNCH BREAK set to ON the Resume Measurements Screen displays during startup From this screen you can choose to continue accumulating TWA and PEAK readings and the time in operation from the last time the EAGLE 2 was used or start collecting new readings and reset the time in operation 1 From the main menu place the cursor next to LUNCH BREAK 2 Press and release POWER ENTER RESET The Lunch Break Screen appears LUNCH BREAK UP DOWN THEN ENTER OFF 3 Use AIR A YES or RANGE VY SHIFT to display the desired setting 4 Press and release POWER ENTER RESET to save the setting and return to the main menu Updating the Span Factor Setting With SPAN FACTOR set to ON factory setting the EAGLE 2 will display the span adjustment range for a sensor in the calibration results screen while in Single Calibration The span adjustment shows how low and how high the reading could hav
13. If the O ring came out place it back in making sure it is seated in its groove 16 Place the electrolyte reservoir assembly back in the sensor body making sure that the ridges on the electrolyte reservoir assembly line up with the grooves inside the sensor body CAUTION Verify that the electrolyte reservoir assembly is properly aligned before inserting it into the sensor body Forcing an electrolyte reservoir assembly into its sensor body may damage the electrolyte reservoir assembly or the sensor body 17 Screw the collar of the ESM 01 sensor back on 18 Plug the ESM O1 sensor back into the cable 19 Insert the sensor into the ESM 01 flow chamber and push it in until it bottoms out 20 Line up the holes in the ESM 01 sensor bracket with the two standoffs on EAGLE 2 Operator s Manual Maintenance 201 the ESM 01 chamber 21 Install the two sensor bracket screws 22 Confirm that the main PCB is seated in its slots and that its bottom edge is resting on the bottom of the bottom case If the main PCB is not seated properly it may be damaged when the top case is re installed 23 Make sure that the top case gasket is fully seated in its groove and carefully put the top case back on the bottom case If you have any difficulty mating the top and bottom cases inspect the placement of the main PCB and the placement of the top case gasket 24 Turn the EAGLE 2 upside down and tighten the three case screws to secure the top ca
14. Replacing the Charcoal Filter he ke pena eae serene we 79 Checking the Combustible Gas Sensor s Condition 80 Replacinesa Sensor s eure iati arua dete e aa k E coo e es 80 General Parts List ioia eee Nake kee a E EEEE E 84 Appendix A Calibrating with a Sample Bag 86 OVCIVICW amp Abad Oawid sie EE Sele dis soe Bae oot ota UE ER wee 86 Calibration Supplies and Equipment 2 20 0 e eee eee 86 Entering Calibration Mode 00 0 e cece eee 87 Calibrating Using the Auto Calibration Method 88 Setting the Fresh Air Reading 0 6 ccc eee 88 Performing a Span Adjustment in Auto Calibration 89 Calibrating Using the Single Calibration Method 94 Setting the Fresh Air Reading 0 6 ccc eens 94 Performing a Span Adjustment in Single Calibration 96 Parts EIS peace hee Ou ha EE ae ate ee meeps E gaieale eae aes 100 Appendix B Setup Mode eessssssssssosesssosesesooeee 101 OVCIVIEW saren eiin bak a a eS aches a eRe a AE Oe a AANA 101 Tips tor Using Set p Modes so r res perae alee tee ae ete a 102 Using Setup Mode inc cne 5h a Ghd Gala andra 8S Stine i bE ere BE 103 Setting the Date and Time 0 6 ene 104 Setting the Date Format 36 an 684 Bae ae ORR oe A 104 Setting the Battery Type a s t52 54 bee SRA BSR lod Gene ad bake 105 Configuring the Channels 0 65 scat ita hn does Os Pace vows 105 Config
15. Rock it back and forth gently if necessary to pull it out Take care not to pull the cable connector from the PID sub PCB The PID sensor protrudes from one end of the sensor adapter Grasp the old PID sensor firmly and pull it out of the sensor adapter Carefully line up the new PID sensor s pins with the sockets in the bottom of the sensor adapter and gently lower the sensor into the adapter until you feel it contact the bottom Do not attempt to push the sensor in farther once it makes contact with the bottom of the adapter until you are sure that the sensor s pins are engaged with the sockets If you feel that the pins did not engage the sockets slightly rotate the sensor back and forth without putting pressure on it until you feel the pins engage the sockets Push the sensor into the sockets until it bottoms out EAGLE 2 Operator s Manual Maintenance 171 12 Insert the sensor adapter into the PID flow chamber and push it in until it bottoms out 13 Plug the PID cable connector into the PID sub PCB on the main PCB 14 Confirm that the main PCB is seated in its slots and that its bottom edge is resting on the bottom of the bottom case If the main PCB is not seated properly it may be damaged when the top case is re installed 15 Make sure that the top case gasket is fully seated in its groove and carefully put the top case back on the bottom case If you have any difficulty mating the top and bottom cases inspect the p
16. The EAGLE 2 Operator s Manual is organized as follows e Chapter 1 is an introduction to the EAGLE 2 e Chapter 2 describes the components of the EAGLE 2 4 About this Manual EAGLE 2 Operator s Manual e Chapter 3 describes the operation of the EAGLE 2 e Chapter 4 describes Calibration Mode which allows you to calibrate the EAGLE 2 s active channels e Chapter 5 describes the EAGLE 2 s maintenance requirements and procedures e Appendix A describes calibration of the EAGLE 2 using a sample bag e Appendix B describes Setup Mode which allows you to configure different parameters of the EAGLE 2 e Appendix C describes the sub PCBs that are installed for optional sensors and how they affect gas and channel configuration e Appendix D describes the PID sensors and the operation of the EAGLE 2 with a PID sensor installed e Appendix E describes the ESM 01 sensors and the operation of the EAGLE 2 with an ESM 01 sensor installed e Appendix F describes the TC sensors and the operation of the EAGLE 2 with a TC sensor installed e Appendix G describes the infrared CO sensors and the operation of the EAGLE 2 with an infrared CO sensor installed e Appendix H describes the infrared methane sensors and the operation of the EAGLE 2 with an infrared methane sensor installed e Appendix I describes the infrared hydrocarbon sensor and the operation of the EAGLE 2 with an infrared hydrocarbon sensor installed e Appendix J de
17. only the serial number is shown 14 If the EAGLE 2 experiences a sensor failure during start up a screen indicating which sensor failed appears and the buzzer sounds a pulsing tone twice per second In the example below the TC volume CH4 sensor has failed FAILED SENSOR S lt gt lt gt lt gt lt CH4 gt ENTER TO CONTINUE 274 Start Up Bar Hole Mode EAGLE 2 Operator s Manual If one of the combustible sensors LEL or TC fails it is not possible to enter Bar Hole Mode Press and release the POWER ENTER RESET button to acknowledge the failure and return to the Mode Select Screen Replace the failed sensor as soon as possible If the oxygen sensor fails press and release the POWER ENTER RESET button to acknowledge the failure and continue to Bar Hole Mode The gas reading for the oxygen sensor will be replaced by XXX Replace the failed sensor as soon as possible If any other sensor that is installed fails press and release the POWER ENTER RESET button to acknowledge the failure and continue to Bar Hole Mode Replace the failed sensor s as soon as possible for use in Normal Mode 15 The EAGLE 2 is now operating in Bar Hole Mode The pump is off and the following screen appears BAR HOLE MODE CH4 0 LEL OXY 20 9 vol ENTER MEASURE ADJUST PURGE Only the methane and oxygen channels are displayed NOTE The units for the methane channel can be changed using the
18. oxygen CO and HS sensors at the same time automatically with no need for a zero oxygen source you can use the auto calibration feature with a 4 gas cylinder If the HS channel is not active then a 3 gas cylinder may be used for auto calibration This chapter includes instructions for auto calibration with a demand flow regulator and a 4 gas 56 Calibration Supplies and Equipment EAGLE 2 Operator s Manual cylinder This chapter also includes instructions for calibrating one channel at a time using single calibration Entering Calibration Mode To enter Calibration Mode do the following 1 Find a fresh air environment This is an environment free of toxic or combustible gases and of normal oxygen content 20 9 2 While in Measuring Mode press and hold the RANGE W SHIFT button then press the DISPLAY ADJUST NO button and release both buttons 3 Ifthe unit prompts you for the password enter it by using the AIR A YES and RANGE YW SHIFT buttons to select each password number and then pressing and releasing POWER ENTER RESET to enter the number and move on to the next one 4 The Calibration Mode Screen displays with the cursor next to AUTO CALIBRATION CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION NOTE The following screens illustrate a four gas EAGLE 2 for detection of CH LEL using catalytic sensor oxygen HS and CO Your EAGLE 2 may displa
19. sensor failed to calibrate FAILED SENSOR ENTER TO CONTINUE lt CO2 gt The buzzer and alarm LEDs activate in a double pulsing pattern Press and release the POWER ENTER RESET button to reset the alarm and continue to the Calibration Mode Screen Attempt to calibrate the infrared CO channel again If the failure continues investigate the cause See Troubleshooting on page 70 41 Disconnect the tubing from the probe 42 Unscrew the demand flow regulator from the calibration cylinder 43 Use the RANGE W SHIFT button to place the cursor next to the NORMAL OPERATION menu option then press and release the POWER ENTER RESET button to return to Measuring Mode EAGLE 2 Operator s Manual Infrared CO Calibration 235 Maintenance This section includes a procedure for replacing an infrared CO sensor Replacing the IR CO Sensor or Changing Sensor Type L 2 3 10 11 12 13 14 15 Verify that the EAGLE 2 is off Place the EAGLE 2 upside down on a flat surface or hold it upside down Unscrew the three case screws until they disengage from the top case They are captive screws so they will not fall off of the bottom case Turn the EAGLE 2 right side up and carefully lift the top case away from the bottom case Be careful not to lift it so far that it pulls on the main PCB with the cable that connects the top case to the main PCB Lay the top case down next to the bottom case to allow
20. 2 completes its startup sequence it is in Measuring Mode In Measuring Mode the EAGLE 2 continuously monitors the sampled atmosphere and displays the gas concentrations present for its target gases In a low light environment press and release any button to turn on the display backlight See Updating the Backlight Delay Setting on page 119 to program backlight duration If the Confirmation Alert feature is turned on in the Setup Mode menu see Updating the Confirmation Alert Setting on page 126 the EAGLE 2 beeps periodically to confirm that it s operating EAGLE 2 Operator s Manual Measuring Mode Normal Operation 29 Monitoring an Area 1 Start up the EAGLE 2 as described above in Start Up on page 22 It is now in Measuring Mode CH4 O LEL OXY 20 9vol H2S 0 0ppm CO Oppm 2 Take the EAGLE 2 to the monitoring area Put the probe tip in the area to be monitored NOTE If you notice that water has entered the flow system through the probe replace the probe s hydrophobic filter and inspect the O ring for filter particles See Replacing the Hydrophobic Probe s Particle Filter and Hydrophobic Filter Disk on page 75 for instructions to replace the hydrophobic filter 3 Wait 10 15 seconds and observe the display for gas readings If a reading is observed allow the reading to stabilize to determine the gas concentrations present NOTE Response time increases with the length of the
21. 4 Continue to hold the AIR A YES button until the display prompts you to release it The EAGLE 2 will set the fresh air reading for the CH channel Start up is complete and the unit is now ready for monitoring Leak Testing In Leak Check Mode the EAGLE 2 only displays combustible gas readings The readings are displayed in both numerical and bar graph form The bar graph displays readings up to 5 000 ppm while the numerical indicator displays readings up to 50 000 ppm in the following increments e 5 ppm increments from 0 ppm to 200 ppm e 10 ppm increments from 200 ppm to 1 000 ppm e 50 ppm increments from 1 000 ppm to 10 000 ppm 284 Leak Testing EAGLE 2 Operator s Manual e 250 ppm increments from 10 000 ppm to 50 000 ppm As the gas concentration increases from 0 ppm the alarm LEDs begin to blink in unison with the buzzer s pulsing The blinking pulsing rate increases as the gas reading increases If desired the buzzer can be turned off in Leak Check Mode In a low light environment press and release any of the buttons to turn on the display backlight See Updating the Backlight Delay Setting on page 119 to program backlight duration If CONFIRMATION BEEP is turned on in the Setup Mode menu the EAGLE 2 beeps once every 15 minutes to confirm that it s operating Locating a Leak 1 Start up the EAGLE 2 as described above in Start Up Leak Check Mode on page 279 2 Move the probe tip back and forth al
22. At the end of the sample period the pump will shut off and the peak methane and minimum oxygen levels monitored during the sample period will be displayed Bar Hole Mode also allows you to initiate an air purge cycle to purge gas from the EAGLE 2 after a sample is taken See Setting the Bar Hole Measurement Time on page 125 to change the bar hole sample time In a low light environment press and release the RANGE W SHIFT button to turn on the display backlight Although the backlight will turn on when any button is pressed other buttons may initiate an undesired operation sequence See Updating the Backlight Delay Setting on page 119 to program backlight duration If CONFIRMATION ALERT is turned on in the Setup Mode menu the EAGLE 2 will alert you based on the setting you choose once every 15 minutes to confirm that it s operating Performing a Bar Hole Test 1 Start up the EAGLE 2 as described in Start Up Bar Hole Mode on page 270 2 Take the EAGLE 2 to the bar hole that will be tested 276 Bar Hole Testing EAGLE 2 Operator s Manual 3 Insert the probe into the bar hole and press and release the ENTER button The pump will turn on the display will indicate MEASURING below the oxygen reading and the sample period will begin with the sample period counting down in seconds in the lower right corner of the display The CH channel will be displayed in vol BAR HOLE MODE CH4 0 vol OXY 20 9 vo
23. Battery Charger Conditions Amber LED Green LED Status BLINKING FAIL OFF ON CONTINUOUS OPERATION Replacing the Hydrophobic Probe s Particle Filter and Hydrophobic Filter Disk Inspect the probe s internal components if you notice that the EAGLE 2 s pump sounds bogged down or if an unexplained low flow alarm occurs Replace the particle filter if it appears to be dirty Replace the hydrophobic filter disk if it appears dirty or saturated with liquid Replace the O rings in the probe if either of them appears damaged 1 Grasp each end of the clear probe body firmly and unscrew the two halves from each other One half includes a plastic tube fitting and the probe tube The other half includes a metal fitting that mates with the EAGLE 2 inlet fitting EAGLE 2 Operator s Manual Replacing the Hydrophobic Probe s Particle Filter and Hydro O ring Inside wy Hydrophobic Filter Disk Particle Filter Not Included in PID Probe O ring Probe Body Figure 19 Replacing the Particle Filter and Hydrophobic Filter Disk 2 Remove the white hydrophobic filter disk from the top of the particle filter or from the probe body 3 Remove the particle filter from the probe body 4 Clean the inside of the probe body if necessary 5 Hold the probe half that has the plastic tube fitting and the probe tube with the fitting and tube facing down 6 Place the new cone sh
24. CAT 0 50000ppm CH4 p 50000 ppm CH4 gt 50000 ppm GAS A 11000 ppm HEX 10000 ppm CAT 40000 ppm H2 q END gt 18000 ppm IBU rm A YES All of the gases except for the one with the asterisk next to it are pre defined The gas with the asterisk next to it is user defined The user defined gas can be used if the desired gas is not one of the pre defined gases Below is a brief description of each choice EAGLE 2 Operator s Manual 50000 ppm CH4 This selection is for methane CH and is the standard factory setting The LEL for methane is 5 volume or 50 000 ppm If the gas is configured for this choice the methane elimination feature is inactive and the Methane Elimination Mode Screen will not appear in Display Mode See Methane Elimination Mode Screen on page 43 for a description of the Methane Elimination Mode Screen See Appendix J Methane Elimination Mode for a description of Methane Elimination Mode 11000 ppm HEX This selection is for hexane The LEL for hexane is 1 1 or 11 000 ppm If the gas is configured for this choice the methane elimination feature is active and the Methane Elimination Mode Screen is accessible in Display Mode See Methane Elimination Mode Screen on page 43 for a description of the Methane Elimination Mode Screen See Appendix J Methane Elimination Mode for a description of Methane Elimination Mode 40000 ppm H2 This selection is for hydrogen H2 The LEL for
25. CO 6 Move the cursor next to the sensor you want to calibrate with the AIR A YES and RANGE YW SHIFT buttons In the example below the CH4 sensor is selected for span adjustment SELECT SENSOR TO CALIBRATE ESCAPE gt CH4 OXY H2S CO If you do not want to proceed with the span adjustment press and release the DISPLAY ADJUST NO button or place the cursor next to ESCAPE and press and release POWER ENTER RESET to return to the Calibration Mode Screen To proceed with the calibration continue with the next step 66 Calibrating Using the Single Calibration Method EAGLE 2 Operator s Manual 7 Press and release the POWER ENTER RESET button to proceed to the Single Calibration Gas Value Screen for the selected channel The calibration gas value is flashing SINGLE CALIBRATION CH4 50 LEL UP DOWN TO ADJUST CALIBRATION VALUE ENTER WHEN DONE 8 If necessary adjust the calibration gas value to match the cylinder concentration with the AIR A YES and RANGE VY SHIFT buttons NOTE The calibration gas value cannot be set lower than the low alarm setting If the calibration gas value listed on the calibration cylinder is lower than the current low alarm setting enter Setup Mode and change the low alarm setting See Updating the Alarm Point Settings on page 114 for instructions If you need to change the alarm point setting only to perform a calibration make sure that you change th
26. Charging Jack a gt L Battery Case Thumbscrew in 7 l Infrared Port Figure 3 Component Location Front amp Back EAGLE 2 Operator s Manual Instrument Description 7 Case The EAGLE 2 s sturdy high impact plastic case is radio frequency RF resistant and is suitable for use in many environmental conditions indoors and out The case is dust proof and water resistant It s two main components the top case and bottom case are held together with three screws located on the bottom case The interface between the top case and bottom case is gasketed A sturdy well balanced handle on the top case allows you to hold the instrument comfortably A clear plastic window is located on the top case to the front of the handle for viewing the LCD A removable battery case is located at the rear of the bottom case A thumbscrew secures the battery case to the bottom case The interface between the battery case and the bottom case is gasketed Six raised feet on the bottom of the case four on the bottom case and two on the battery compartment raise the EAGLE 2 slightly from the surface on which it rests Sensors The EAGLE 2 uses up to six sensors to monitor combustible gas oxygen O2 carbon monoxide CO hydrogen sulfide H2S and various other toxic gases simultaneously The sensors are located inside the EAGLE 2 bottom case and are installed in the flow chamber The sensors described below are the four standard sensors Se
27. Instruments Inc recommends using the auto calibration method for convenience The calibration instructions below show a 5 channel instrument which has the four standard channels LEL oxygen H S CO and an infrared hydrocarbon channel The standard factory calibration for the autoranging infrared hydrocarbon sensor is to 50 LEL isobutane If you need maximum accuracy in the vol range the sensor may be calibrated to 50 vol isobutane To use the auto calibration method to calibrate a multi channel instrument that includes an infrared hydrocarbon channel do the following 1 See Calibration Supplies and Equipment on page 56 for a description of the necessary calibration supplies In addition to an appropriate multi gas cylinder that is used to calibrate any active standard channels you will also need a cylinder to calibrate the infrared hydrocarbon channel See Table 32 on page 265 for available cylinders Make sure your calibration cylinder is appropriate for the infrared hydrocarbon detection range 2 Find a fresh air environment This is an environment free of toxic or combustible gases and of normal oxygen content 20 9 3 While in Measuring Mode press and hold the RANGE W SHIFT button then press the DISPLAY ADJUST NO button and release both buttons 4 If the unit prompts you for the password enter it by using the AIR A YES and RANGE VY SHIFT buttons to select each password number and then pressing and releasing POW
28. K Switch Connector Pump Connector m C L og 4 a fess o O00 To Float Probe Connector IR Communication Sensor Transmitter ee 2a a 5 Figure 5 Main PCB The main PCB is located on the right side of the bottom case It slides into guiding grooves at the bottom front and rear of the bottom case It is held in place by the top case Connectors for the sensors differential pressure switch pump display PCB optional float probe and I S barrier PCB are located on the main PCB In addition an IR transmitter receiver is located at the front of the PCB behind the IR port on the front of the bottom case Alarm LEDs Two sets of red alarm LEDs light emitting diodes are visible through two raised frosted plastic lenses in the top case Each set has two LEDs They are above the LCD one above the left corner and one above the right corner The alarm LEDs alert you to gas low battery and failure alarms Infrared Communications Port An infrared IR communications port is located on the left front of the bottom case when the instrument is viewed from the front The data transmitted through the port is in standard IrDA protocol A computer s 12 Instrument Description EAGLE 2 Operator s Manual infrared port or an IrDA USB cable connected to a USB port can be used to download data saved by the EAGLE 2 to a computer using the Eagle 2 Data Logger Manag
29. LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm y 16 Place the cursor next to the channel whose gas value you want to change using the AIR A YES and RANGE YW SHIFT buttons 17 Press and release the POWER ENTER RESET button to select the EAGLE 2 Operator s Manual ESM 01 Calibration 185 channel The calibration gas value begins to flash 18 Use the AIR A YES and RANGE YW SHIFT buttons to adjust the calibration gas setting to the desired value NOTE The calibration gas value cannot be set lower than the low alarm setting If the calibration gas value listed on the calibration cylinder is lower than the current low alarm setting enter Setup Mode and change the low alarm setting See Updating the Alarm Point Settings on page 114 for instructions If you need to change the alarm point setting only to perform a calibration make sure that you change the alarm point setting back to its original value once the calibration has been performed 19 Press and release the POWER ENTER RESET button to save the change The calibration gas value stops flashing 20 Repeat Step 16 through Step 19 for any other channels that need to be changed 21 When you are done adjusting the calibration gas values move the cursor down past the bottom of the screen next to END ADJUST AUTO CALIBRATION VALUES gt END A 22 Press and release the POWER ENTER RESET button The following screen appears DO YOU WANT TO ST
30. MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION If the ESM 01 channel cannot be adjusted to the proper value a screen displays that indicates a calibration failure and indicates that the ESM 01 sensor failed to calibrate FAILED SENSOR ENTER TO CONTINUE lt SO2 gt The buzzer and alarm LEDs activate in a double pulsing pattern Press and release the POWER ENTER RESET button to reset the alarm and continue to the Calibration Mode Screen Attempt to calibrate the ESM 01 channel again If the failure continues investigate the cause See Troubleshooting on page 70 34 Disconnect the tubing from the probe 35 Unscrew the demand flow regulator from the calibration cylinder 36 Use the RANGE YW SHIFT button to place the cursor next to the NORMAL OPERATION menu option then press and release the POWER ENTER RESET button to return to Measuring Mode 198 ESM 01 Calibration EAGLE 2 Operator s Manual Maintenance The SO and HCN ESM 01 sensors are the only ESM 01 sensors that include user serviceable parts This section includes a procedure for replacing an ESM 01 sensor and for replacing the H2S scrubber in the SO and HCN sensors When replacing a sensor you may either replace it with another of the same sensor or you may install a different ESM 01 sensor If a different one is installed the EAGLE 2 will load the sensor parameters and configure the ESM 01 channe
31. Mode so methane elimination cannot be turned off In addition the relative response feature is inactive even if it is set to ON in the Setup Mode Relative Responses menu item CAUTION The 50000 ppm GAS gas configuration is normally set at the factory for very specific applications Consult RKI Instruments Inc before configuring the gas for 50000 ppm GAS e 10000 ppm CAT This selection is a user defined selection The factory setting is 10000 ppm CAT with the asterisk indicating that it is user defined If this selection has been updated in the field it will appear differently but the asterisk will always remain next to the menu item to indicate it is user defined If you choose the user defined selection the unit will prompt you to enter four parameters a three character gas name the LEL value in terms of ppm the response factor relative to methane and the detector voltage When you configure the gas as the user defined choice the methane elimination feature is active and the Methane Elimination Mode Screen is accessible in Display Mode See Methane Elimination Mode Screen on page 43 for a description of the Methane Elimination Mode Screen See Appendix J Methane Elimination Mode for a description of Methane Elimination Mode 110 Using Setup Mode EAGLE 2 Operator s Manual 5 Use AIR A YES and RANGE YW SHIFT to move the cursor next to the desired gas for the catalytic channel 6 If you placed the curso
32. Mode Screen If you do want to continue with the calibration proceed with the next step 7 Press and release the POWER ENTER RESET button to proceed to the Single Calibration Gas Value Screen for the selected channel The calibration gas value is flashing SINGLE CALIBRATION CH4 50 LEL UP DOWN TO ADJUST CALIBRATION VALUE ENTER WHEN DONE 8 If necessary adjust the calibration gas value to match the cylinder concentration using the AIR A YES and RANGE YW SHIFT buttons For this example the calibration gas value is entered as 50 LEL NOTE The calibration gas value cannot be set lower than the low alarm setting If the calibration gas value listed on the calibration cylinder is lower than the current low alarm setting enter Setup Mode and change the low alarm setting See Updating the Alarm Point Settings on page 114 for instructions If you need to change the alarm point setting only to perform a calibration make sure that you change the alarm point setting back to its original value once the calibration has been performed 9 Press and release the POWER ENTER RESET button to proceed to the Single Calibration Apply Gas Screen CAL IN PROCESS is flashing SINGLE CALIBRATION APPLY GAS CH4 0 LEL CAL IN PROCESS ENTER WHEN DONE 10 Fill the gas collection bag by screwing the fixed flow regulator or dispensing valve onto the calibration cylinder and turning the knob counterclockwise 11 Al
33. NORMAL OPERATION If the PID channel cannot be adjusted to the proper value a screen displays that indicates a calibration failure and indicates that the PID sensor failed to calibrate FAILED SENSOR ENTER TO CONTINUE lt IBL gt The buzzer and alarm LEDs activate in a double pulsing pattern Press and release the POWER ENTER RESET button to reset the alarm and continue to the Calibration Mode Screen Attempt to calibrate the PID channel again If the failure continues investigate the cause See Troubleshooting on page 70 37 Disconnect the tubing from the probe 38 Unscrew the demand flow regulator from the calibration cylinder 39 Use the RANGE YW SHIFT button to place the cursor next to the NORMAL OPERATION menu option then press and release the POWER ENTER RESET button to return to Measuring Mode EAGLE 2 Operator s Manual PID Calibration 151 Calibrating with a 5 Gas Cylinder L See Calibration Supplies and Equipment on page 56 for a description of the necessary calibration supplies See Table 17 on page 178 for available cylinders Make sure your calibration cylinder is appropriate for the PID detection range Find a fresh air environment This is an environment free of toxic or combustible gases and of normal oxygen content 20 9 While in Measuring Mode press and hold the RANGE W SHIFT button then press the DISPLAY ADJUST NO button and release both buttons If the unit
34. RANGE VY SHIFT button Press and release the RANGE VW SHIFT button until the desired units ppm LEL or vol are displayed NOTE If measuring the combustible gas in ppm units for maximum sensor stability allow 3 5 minutes for the sensor to warm up The small increment size in the lower range of a ppm measurement can cause instability if the unit is not properly warmed up Performing a Demand Zero Bar Hole Mode Before using the EAGLE 2 it is a recommended to set the fresh air readings for the target gases by performing a demand zero This will set the CH4 channel to zero and the OXY channel to 20 9 1 Find a fresh air environment This is an environment free of toxic or combustible gases and of normal oxygen content 20 9 2 Turn on the unit as described above in Turning On the EAGLE 2 Bar Hole Mode on page 270 EAGLE 2 Operator s Manual Start Up Bar Hole Mode 275 3 Press and hold the AIR A YES button The display will indicate that a demand zero is taking place and prompt you to hold the AIR A YES button 4 Continue to hold the AIR A YES button until the display prompts you to release it The EAGLE 2 will set the fresh air reading for the CH4 and oxygen channels Start up is complete and the unit is now ready for bar hole testing Bar Hole Testing In Bar Hole Mode the you can initiate sampling for a fixed time period to monitor for methane and oxygen in a bar hole The factory set time is 30 seconds
35. Remove the tubing from the rigid tube on the probe 30 Unscrew the 4 gas cylinder from the demand flow regulator EAGLE 2 Operator s Manual Infrared Hydrocarbon Calibration 261 31 If you want to change the infrared hydrocarbon channel s calibration gas value follow Step 15 Step 24 above beginning with the infrared hydrocarbon Calibration Gas Value Screen below instead of the standard channel Calibration Gas Value Screen CAL GAS VALUE HC 50 LEL ENTER TO BEGIN CAL 32 With the infrared hydrocarbon Calibration Gas Value Screen displayed press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen for the infrared hydrocarbon channel with CAL IN PROCESS flashing CAL IN PROCESS HC 0O LEL ENTER WHEN DONE If you do not want to proceed with the calibration press and release the DISPLAY ADJUST NO button to return to the infrared hydrocarbon Cal Gas Values Screen If you do want to continue with the calibration proceed to the next step 33 Screw the infrared hydrocarbon calibration cylinder onto the demand flow regulator 34 Connect the tubing from the regulator to the rigid tube on the probe Allow the EAGLE 2 to draw gas for one minute 35 Press and release the POWER ENTER RESET button to set the span adjustment for the infrared hydrocarbon channel to the programmed value 262 Infrared Hydrocarbon Calibration EAGLE 2 Operator s Manual 36 I
36. TC sensor There are 3 pre defined options and 1 user defined option The user defined option will always have an asterisk next to it TC 0 100 0vol CH4 gt 100 0 vol CH4 100 0 vol H2 10 0 vol H2 10 0 vol TC ay 11 Use AIR A YES and RANGE VY SHIFT to move the cursor next to the desired gas 12 If you placed the cursor next to one of the pre defined gases press and release POWER ENTER RESET to select the gas and proceed to Step 17 If you placed the cursor next to the user defined gas with the asterisk press POWER ENTER RESET and proceed with Step 13 13 The user defined gas setup screen appears with the first character of the gas name flashing The current gas name and range are shown on the top line of the screen TC 0 100 0vol CH4 CHANGE TO NAME TC FULL_SCALE 10 0 EAGLE 2 Operator s Manual Configuring the TC Gas in Setup Mode 219 WARNING Do not use the user defined gas configuration without consulting RKI Instruments Inc 14 Enter the gas name There are 3 characters available for the gas name The factory setting of TC uses only 2 characters Use AIR A YES and RANGE VY SHIFT to display the desired character then press POWER ENTER RESET to enter the displayed character and move to the next character Repeat until all three characters are entered When the last character is entered the full scale value will be flashing 15 Use AIR A YES and RANGE YW SHIFT to display the
37. When the last period disappears the unit is finished clearing the data and will display the following screen for a few seconds then return to Measuring Mode CLEARED Data Logging NOTE The EAGLE 2 only logs data while in Normal Mode If the EAGLE 2 is used in Leak Check Mode or Bar Hole Mode no downloadable data will be logged while it is in either of these two modes The EAGLE 2 features the ability to log data to its internal memory and download it to a computer via the infrared communications port on the front of the unit It logs gas readings during normal operation alarm data and calibration data To utilize the EAGLE 2 s downloading capability you will need the Eagle 2 EAGLE 2 Operator s Manual Data Logging 53 Data Logger Management Program and a computer with an infrared port or a USB port that runs one of the following operating systems Windows 2000 Windows XP or Windows Vista If your computer has an infrared port then no additional accessories are needed to download data from the EAGLE 2 If your computer does not have an infrared port but does have a USB port a USB IrDA adapter cable can be used to download data from the EAGLE 2 using the USB port The Eagle 2 Data Logger Management Program is available from RKI Instruments Inc The adapter cable is also available from RKI Instruments Inc or may be purchased from an electronic supply website The data logging capacity depends on how often t
38. You will note that if methane elimination is enabled the Catalytic Sensor Relative Response screen no longer appears even if the relative response feature in Setup Mode is turned on While methane elimination is enabled the Catalytic Sensor Relative Response screen will not appear in Display Mode even if the relative response feature in Setup Mode is enabled 7 Continue to press the DISPLAY button until the Measuring Mode screen appears HEX O LEL ME OXY 20 9vol H2S 0 0ppm CO Oppm When the instrument is reading in methane elimination an ME appears in line with the catalytic combustible channel reading Allow 2 minutes for the combustible sensor to stabilize 9 Perform a demand zero See Performing a Demand Zero on page 27 for instructions 10 Monitor for the target gas 11 If you wish to monitor for any other gas while in methane elimination mode a conversion factor must be taken into consideration With the EAGLE 2 calibrated to hexane use Table 33 below to determine the concentrations of other target gases Multiply the display reading by the factor in the appropriate column to determine the actual reading for that gas For example if you are using the EAGLE 2 in methane elimination mode to detect isobutane and the display reads 10 LEL the actual isobutane reading is 10 x 0 59 5 9 LEL isobutane Table 33 Methane Elimination Mode Conversion Factors Hexane Calibration Target Gas LEL Factor P
39. access to the flow system Locate the infrared CO sensor It has a twelve wire cable with a connector that mates to an infrared sub PCB that is installed on the main PCB and is normally located next to the pump Figure 38 on page 224 shows an infrared CO sensor in a typical location Unscrew and remove the two screws that hold down the infrared CO sensor bracket Grasp the sensor firmly and pull it out of the infrared CO flow chamber Rock it back and forth gently if necessary to pull it out Take care not to pull the cable from the sub PCB Rotate the locking lever counterclockwise on the cable connector that mates to the infrared CO sensor to unlock it Unplug the old infrared CO sensor from the cable Connect the new infrared CO sensor to the sensor cable and rotate the locking lever clockwise to lock the connector Insert the sensor into the infrared CO flow chamber and push it in until it bottoms out Line up the holes in the infrared CO sensor bracket with the two standoffs on the infrared CO chamber Install the two sensor bracket screws Confirm that the main PCB is seated in its slots and that its bottom edge is resting on the bottom of the bottom case If the main PCB is not seated 236 Maintenance EAGLE 2 Operator s Manual properly it may be damaged when the top case is re installed 16 Make sure that the top case gasket is fully seated in its groove and carefully put the top case back on the bottom cas
40. avoid leaks that can affect the calibration See Figure 19 on page 76 for an illustration of the internal parts of the probe 184 ESM 01 Calibration EAGLE 2 Operator s Manual 13 Move the cursor next to the AUTO CALIBRATION menu item by using the AIR A YES button CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 14 Press and release the POWER ENTER RESET button to display the Calibration Gas Values Screen CAL GAS VALUES CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm ENTER TO BEGIN CAL The gas concentrations displayed in the Calibration Gas Values Screen must match the gas concentrations listed on the 4 gas calibration cylinder If all concentrations match go to Step 25 If one or more concentrations do not match continue with Step 15 If you do not want to continue with the calibration press and release the DISPLAY ADJUST NO button to return to the Calibration Mode Screen NOTE The RKI 4 gas cylinder typically contains 12 O by volume When using the auto calibration method be sure to set the OXY auto calibration value to agree with the concentration listed on the cylinder s label not zero 15 To adjust the values on the screen press and hold the RANGE W SHIFT button then press the DISPLAY ADJUST NO button and release both The following screen appears with the cursor next to CH4 ADJUST AUTO CALIBRATION VALUES gt CH4 50
41. below summarizes the types of alarms produced by the EAGLE 2 and their indications Table 6 Alarm Types and Indications Alarm Type Visual Indications Low Alarm e ALRM1 appears next to gas reading Audible Indication Pulsing tone once per f second Concentration of gas rises above the Low Alarm setting or falls below the Low Alarm setting for Oo High Alarm Concentration of gas rises above the High Alarm setting TWA or STEL Concentration of CO or HS rises above the TWA or STEL alarm setting Over Range 36 Measuring Mode Alarms Alarm LED arrays flash once per second Backlight turns on ALRM2 appears next to gas reading Alarm LED arrays flash twice per second Backlight turns on Alarm LED arrays flash once per second Backlight turns on TWA or STEL appears next to gas reading OVER appears next to gas reading Gas reading indicates full scale Alarm LED arrays flash twice per second Backlight turns on Pulsing tone twice per second Pulsing tone once per second Pulsing tone twice per second EAGLE 2 Operator s Manual Table 6 Alarm Types and Indications Alarm Type Visual Indications Audible Indication Low Flow e The display indicates FAIL Double pulsing tone LOW FLOW LEVEL once per second e Alarm LED arrays flash ina double pulsing pattern once per second e Backlight turns on Low Battery Warning e BATT appears vertically along None the left side of LCD
42. by using AIR A YES and RANGE YW SHIFT to select each password number and then pressing and releasing the POWER ENTER RESET button to enter it and move on to the next number until all of the numbers are entered The main menu displays It displays six menu items at a time gt SET DATE amp TIME SET DATE FORMAT SET BATTERY TYPE CONFIGURE CHANNELS CONFIGURE GASES CATALYTIC UNITS y 6 Use the RANGE VY SHIFT button to move the cursor down the menu to CONFIGURE GASES 7 Press and release POWER ENTER RESET The Configure Gases Screen appears with the cursor flashing next to CAT the catalytic sensor If an EAGLE 2 has a TC sensor installed one of the three optional sensor types OP1 OP2 or OP3 will indicate it is a TC In the example below OP is shown as a TC sensor CONFIGURE GASES gt CAT CH4 CAT OPI CH4 TC O go Oo I oT i xw lt 218 Configuring the TC Gas in Setup Mode EAGLE 2 Operator s Manual 8 Use RANGE VW SHIFT to move the cursor down the menu to the TC sensor CONFIGURE GASES CAT CH4 CAT gt OP1 CH4 TC OP2 OP3 Y 9 To change the TC sensor gas configuration press and release POWER ENTER RESET 10 A screen appears that indicates the detection range for the currently configured gas In the example below the TC sensor is currently configured for 0 100 volume CH4 The screen also shows the gas configuration choices for the
43. channel is roughly calibrated to an alternate gas For example if the catalytic channel is setup for and calibrated to methane you can select hexane from a gas list accessible from the Catalytic Sensor Relative Response Screen in Display Mode so that the catalytic channel responds to gas as if it were calibrated to hexane See Catalytic Sensor Relative Response Screen on page 43 for instructions to use the relative response feature The factory setting for CATALYTIC SENSOR RELATIVE RESPONSE is OFF 1 From the main menu place the cursor next to RELATIVE RESPONSE Press and release POWER ENTER RESET The Catalytic Sensor Relative Response screen appears with the current setting flashing CATALYTIC SENSOR RELATIVE RESPONSE UP DOWN THEN ENTER OFF 2 Use AIR A YES or RANGE YW SHIFT to display the desired setting ON or OFF 3 Press and release POWER ENTER RESET to save the setting and return to the main menu Updating the Alarm Point Settings This menu item allows you to update one or more alarm points the reading at which the EAGLE 2 recognizes the alarm condition 1 From the main menu place the cursor next to ALARM POINTS 2 Press and release POWER ENTER RESET The Change Alarm Point Settings Screen appears and all detection channels are displayed CHANGE ALARM POINT SETTINGS gt 1 CH4 2 OXY 3 H2S 4 CO END 114 Using Setup Mode EAGLE 2 Operator s Manual 3 Move the cursor next
44. contact RKI Instruments Inc for further instructions Display Mode Two other operating modes are accessible when the EAGLE 2 is in Measuring Mode They are Display Mode and Calibration Mode This section describes using the EAGLE 2 in Display Mode In Display Mode you can display peak readings display the minimum operating and current battery voltage select how the active channels are displayed on the LCD turn the catalytic LEL sensor on or off if there is a TC or infrared combustible channel along with a catalytic combustible channel enable or disable methane elimination mode if the catalytic combustible gas channel is configured appropriately in the CONFIGURE GASES item in Setup Mode temporarily configure the catalytic combustible channel for a target gas other than the one used for calibration if the RELATIVE RESPONSE item in Setup Mode is set to ON display STEL readings HS and CO only display TWA readings H S and CO only display alarm settings select the user ID if the USER STATION ID item in Setup Mode is set to ON select the station ID if the USER STATION ID item in Setup Mode is set to ON display time in operation display date and time display remaining data logging time and clear data logger memory if the DATA LOG MEMORY item in Setup Mode is set to ON Tips for Using Display Mode To enter Display Mode and scroll from one screen to the next or skip an item when a question is asked press a
45. contaminate the window with finger oil Verify that the EAGLE 2 is off 2 Place the EAGLE 2 upside down on a flat surface or hold it upside down 3 Unscrew the three case screws until they disengage from the top case They are captive screws so they will not fall off of the bottom case 4 Turn the EAGLE 2 right side up and carefully lift the top case away from the bottom case Be careful not to lift it so far that it pulls on the main PCB with the cable that connects the top case to the main PCB 5 Lay the top case down next to the bottom case to allow access to the flow system 6 Locate the PID sensor adapter It has a five wire cable and is normally located next to the pump The cable has a connector that mates to a PID sub PCB that is installed on the main PCB Figure 25 on page 137 shows a PID sensor in a typical location 7 Grasp the sensor adapter firmly and pull it out of the PID flow chamber Rock it back and forth gently if necessary to pull it out Take care not to pull the cable connector from the PID sub PCB 8 The PID sensor protrudes from one end of the sensor adapter Grasp the PID sensor firmly and pull it out of the sensor adapter 9 Place the PID sensor face down on a flat clean working surface 10 Hold the PID sensor steady on the working surface with one hand and EAGLE 2 Operator s Manual Maintenance 165 using the other hand locate the tabs on the electrode stack removal tool and insert them into th
46. continue with the calibration proceed to the next step 26 Connect the tubing from the demand flow regulator to the rigid tube on the probe Allow the EAGLE 2 to draw gas for one minute 27 Press and release the POWER ENTER RESET button to set the span adjustment for each channel to the programmed values 148 PID Calibration EAGLE 2 Operator s Manual 28 If all channels passed calibration the following screen sequence occurs AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END Y CAL GAS VALUE IBL 100 ppm ENTER TO BEGIN CAL If any of the sensors cannot be adjusted to the proper value a screen displays that indicates a calibration failure and lists the sensor s that failed to calibrate In the example below the oxygen and H S channels failed calibration The other sensors calibrated normally FAIL SENSOR lt gt lt OXY gt lt H2S gt lt gt ENTER TO CONTINUE The buzzer and alarm LEDs activate in a double pulsing pattern Press and release the POWER ENTER RESET button to reset the alarm and continue to the Calibration Value Screen for the PID channel After calibrating the PID channel by following the instructions below attempt to calibrate the standard channels again If the failure continues investigate the cause See Troubleshooting on page 70 29 Remove the tubing from the rigid tube on the probe 30 Unscrew the 4 gas cylinder f
47. display SELECT SENSOR CH4 CALIBRATION TO CALIBRATE DONE ESCAPE 35 98 LEL gt CH4 OXY REMOVE GAS H2S CO Calibrating Using the Single Calibration Method In the example above the EAGLE 2 could have adjusted the reading as low as 35 LEL and as high as 98 LEL If the span adjustment is not successful a screen displays that indicates a calibration failure FAILED SENSOR 11 43 LEL lt CH4 gt ENTER TO CONTINUE EAGLE 2 Operator s Manual In the example above the EAGLE 2 could have adjusted the reading as low as 11 LEL and as high as 43 LEL Since the calibration gas value entered was 50 LEL the unit failed the span adjustment The buzzer and alarm LED arrays activate in a double pulsing pattern Press and release the POWER ENTER RESET button to reset the alarm and return to the Select Sensor Screen Attempt to calibrate again If the failure continues investigate the cause See Troubleshooting on page 70 NOTE The adjustment range will only appear in the calibration results screen if the Span Factor item in Setup Mode is set to ON If the Span Factor is set to OFF the adjustment range will not appear 19 Disconnect the tubing from the EAGLE 2 s probe 20 Repeat Step 6 through Step 19 for any other channels you want to calibrate Make sure you use an appropriate calibration cylinder for each channel CAUTION When calibrating the oxygen channel verify the concentration
48. displays with the cursor next to AUTO CALIBRATION CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 6 Move the cursor to the PERFORM AIR ADJUST menu item by using the RANGE W SHIFT button CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION gt PERFORM AIR ADJUST NORMAL OPERATION 208 TC Calibration EAGLE 2 Operator s Manual 7 Press and release the POWER ENTER RESET button The following screen appears PERFORM AIR ADJUST 8 Press and release the AIR A YES button to continue If you do not want to continue press the DISPLAY ADJUST NO button and the unit will return to the Calibration Mode Screen 9 The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds then indicate that the operation is complete before returning to the Calibration Mode Screen ADJUSTING ZERO RELEASE AIR BUTTON ADJUSTING ZERO COMPLETE CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION gt PERFORM AIR ADJUST NORMAL OPERATION 10 Install the demand flow regulator onto the multi gas calibration cylinder 11 Connect the sample tubing to the demand flow regulator 12 Install the probe on the EAGLE 2 inlet fitting Make sure the probe is EAGLE 2 Operator s Manual TC Calibration 209 complete with internal O ring and membrane and that the two halves of the probe are tighten
49. eg Shas cert ge e anh dren ad Se ONa dhl oA a 52 Data Logging Screen aio d oak gdm Bees PALE ares ES eee es 52 Wala LOSSIS cist au Uk era K E Mee eb E OEE ER eS 53 Chapter 4 Calibration Mode ccc cece ccc cccccccnces 55 OVCIMIW hoe se taka aa aes de Beane eee eee tea tae oe ee 55 Calibration Supplies and Equipment 0000 e eee eee 56 Entering Calibration Wigde s cee seas yak eae LAE Sele A URE eS 57 Calibrating Using the Auto Calibration Method 58 Setting the Fresh Air Reading 00 cee nes 58 Performing a Span Adjustment in Auto Calibration 59 Calibrating Using the Single Calibration Method 64 Setting the Fresh Air Reading nnnn ewok Wala Bs Row es 64 Performing a Span Adjustment in Single Calibration 65 Table of Contents EAGLE 2 Operator s Manual Chapter 5 Maintenance cc cece w cece cece eee neee 70 OVCEVIEW cone we nwa eehowing we sida cela Gaye rete meee ee 70 Troubleshooting 2 522 ede TAB b eee oes Cee TRE Ste aes 70 Replacing or Recharging the Batteries 0 0 00 00048 72 Replacing the Batteries 2 saan ine ape ee ad ealee eos hed 72 Recharging the Ni MH Batteries 60 cc eee 74 Replacing the Hydrophobic Probe s Particle Filter and Hydrophobic FUNC DISK 4 ccs eek e oa ed Pace hh eM ae Male Oks REGO ER E ache 75 Replacing the Hydrophobic Filter 0 0 20 e eee eee 77
50. for example the RKI 4 gas calibration cylinder The EAGLE 2 includes default auto calibration settings for most target gases For gases without a default auto calibration value the setting is 0 NOTE You can also update auto calibration settings in Calibration Mode Updating the auto calibration gas values in Calibration Mode is normally done when performing a calibration Updating these settings in Setup Mode allows you to update the settings without performing a calibration 1 From the main menu place the cursor next tt ADJ AUTOCAL VALUES 2 Press and release POWER ENTER RESET The Adjust Auto Calibration Values Screen appears The auto calibration value for each channel is shown ADJUST AUTO CALIBRATION VALUES gt CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm y 3 Use AIR A YES or RANGE VW SHIFT to place the cursor next to the auto calibration value you want to change 4 Press and release POWER ENTER RESET The auto calibration value 118 Using Setup Mode EAGLE 2 Operator s Manual begins to flash indicating it can be adjusted 5 Use AIR A YES and RANGE YW SHIFT to adjust the auto calibration value to the desired value 6 Press and release the POWER ENTER RESET button to accept the value 7 Repeat Step 3 Step 6 for each auto calibration value you want to change If you want to return to the main menu at any time without saving any changes press and release the DISPLAY ADJUST NO button un
51. high range sensor and 2 00 1 00 and 0 00 for a low range sensor The PID sensor fails to calibrate Cleaning the PID Sensor s Lamp The PID sensor has not come out of its warm up state The auto calibration values may not match the cylinder gas concentration auto calibration only The sample gas is not reaching the sensor because of a bad connection The calibration cylinder may be out of gas or is outdated The lamp may need to be cleaned The electrode stack may need to be replaced Clean the lamp Replace the lamp Replace the electrode stack If the difficulties continue contact RKI Instruments Inc for further instruction Make sure the EAGLE 2 has been properly set up for calibration Check all calibration tubing for leaks or for any bad connections Verify that the calibration cylinder contains an adequate supply of sample Clean the lamp Replace the electrode stack If the difficulties continue contact RKI Instruments Inc for further instruction Clean the lamp if you notice a significant drop in sensitivity from one scheduled calibration to another or if you are not able to calibrate the PID channel See the PID EAGLE 2 Spare Parts on page 178 for lamp cleaning kit ordering information The lamp cleaning kit includes the following items e anelectrode stack removal tool 160 Maintenance EAGLE 2 Operator s Manual e a small vi
52. hydrogen is 4 or 40 000 ppm If the gas is configured for this choice the methane elimination feature is inactive the Methane Elimination Mode Screen will not appear in Display Mode and the catalytic sensor voltage is set to 1 1 volts The standard catalytic sensor voltage is 2 4 volts Because the detector voltage is set to 1 1 volts the catalytic sensor will not respond significantly to methane and many other combustible gases but will respond to hydrogen Using Setup Mode 109 WARNING Do not configure the catalytic sensor gas to hydrogen if you are monitoring for general hydrocarbons Only use this selection if you are monitoring exclusively for hydrogen or if you do not want to see a significant response to other combustible gases e 18000 ppm IBU This selection is for isobutane The LEL for isobutane is 1 8 or 18 000 ppm If the gas is configured for this choice the methane elimination feature is inactive and the Methane Elimination Mode Screen will not appear in Display Mode See Methane Elimination Mode Screen on page 43 for a description of the Methane Elimination Mode Screen See Appendix J Methane Elimination Mode for a description of Methane Elimination Mode e 50000 ppm GAS This selection is for a generic combustible gas with the LEL set to 50000 ppm If the gas is configured for this choice the methane elimination feature will be active but the Methane Elimination Mode Screen will not appear in Display
53. integral cable that connects a switch in the float end to the plug at the other end This plug connects to the float probe socket that is adjacent to the EAGLE 2 s IR port A quick connect fitting at the same end of the assembly connects to the EAGLE 2 s inlet fitting The float probe switch at the opposite end of the 12 foot hose shuts off the pump if the probe begins to be submerged into a liquid Float Probe Quick Connect Fitting N 12 foot Hose Ze Fad C 4 Figure 43 Float Probe Assembly To use the float probe assembly CAUTION Drawing water gasoline or other liquids into the EAGLE 2 will cause damage 1 Attach the quick connect fitting to the EAGLE 2 s inlet fitting Insert the plug into the socket that is adjacent to the IR port Lower the probe into the tank or vessel Lower the probe very slowly keeping the float switch vertical to allow the float switch to activate if necessary Dilution Fitting 1 1 CAUTION When measuring oxygen readings remove the dilution fitting or use your finger to seal the small dilution hole on the side of the dilution fitting The catalytic combustible gas sensor requires oxygen to operate In environments where there is not enough oxygen to operate the combustible gas sensor for example a tank purged
54. it makes contact with the bottom of the adapter until you are sure that the sensor s pins are engaged with the sockets If you feel that the pins did not engage the sockets slightly rotate the sensor back and forth without putting pressure on it until you feel the pins engage the sockets 26 Push the sensor into the sockets until it bottoms out 27 Insert the sensor adapter into the PID flow chamber and push it in until it bottoms out 164 Maintenance EAGLE 2 Operator s Manual 28 Confirm that the main PCB is seated in its slots and that its bottom edge is resting on the bottom of the bottom case If the main PCB is not seated properly it may be damaged when the top case is re installed 29 Make sure that the top case gasket is fully seated in its groove and carefully put the top case back on the bottom case If you have any difficulty mating the top and bottom cases inspect the placement of the main PCB and the placement of the top case gasket 30 Turn the EAGLE 2 upside down and tighten the three case screws to secure the top case to the bottom case 31 Calibrate the PID channel as described in PID Calibration on page 143 Replacing the PID Sensor s Lamp If cleaning the PID lamp does not resolve any calibration problems you may be having the lamp may need to be replaced Perform the following procedure to replace the PID lamp NOTE Do not touch the new lamp window the flat end with your fingers as this may
55. item by using the AIR A YES button CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 18 Press and release the POWER ENTER RESET button to display the Calibration Gas Values Screen CAL GAS VALUES CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm ENTER TO BEGIN CAL The gas concentrations displayed in the Calibration Gas Values Screen must match the gas concentrations listed on the 4 gas calibration cylinder If all concentrations match go to Step 27 If one or more concentrations do not match continue with Step 17 If you do not want to continue with the calibration press and release the DISPLAY ADJUST NO button to return to the Calibration Mode Screen NOTE The RKI 4 gas cylinder typically contains 12 O by volume When using the auto calibration method be sure to set the OXY auto calibration value to agree with the concentration listed on the cylinder s label not zero 19 To adjust the values on the screen press and hold the RANGE W SHIFT button then press the DISPLAY ADJUST NO button and release both The following screen appears with the cursor next to CH4 ADJUST AUTO CALIBRATION VALUES gt CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm y 20 Place the cursor next to the channel whose gas value you want to change using the AIR A YES and RANGE YW SHIFT buttons 21 Press and release the POWER ENTER RESET button
56. not want to proceed with the calibration press and release the DISPLAY ADJUST NO button to return to the Cal Gas Values Screen If you do want to continue with the calibration proceed to the next step 26 Connect the tubing from the demand flow regulator to the rigid tube on the probe Allow the EAGLE 2 to draw gas for one minute 27 Press and release the POWER ENTER RESET button to set the span adjustment for each channel to the programmed values EAGLE 2 Operator s Manual ESM 01 Calibration 195 28 If all channels passed calibration the following screen sequence occurs AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END Y CAL GAS VALUE SO2 5 ppm ENTER TO BEGIN CAL If any of the sensors cannot be adjusted to the proper value a screen displays that indicates a calibration failure and lists the sensor s that failed to calibrate In the example below the oxygen and H S channels failed calibration The other sensors calibrated normally FAIL SENSOR lt gt lt OXY gt lt H2S gt lt gt ENTER TO CONTINUE The buzzer and alarm LEDs activate in a double pulsing pattern Press and release the POWER ENTER RESET button to reset the alarm and continue to the Calibration Value Screen for the ESM 01 channel After calibrating the ESM 01 channel by following the instructions below attempt to calibrate the standard channels again If the failure c
57. of oxygen listed on the cylinder s label For oxygen free samples 100 nitrogen for example set the oxygen zero setting to 0 0 21 After the last channel is calibrated disconnect the calibration tubing from the probe 22 With the Select Sensor Screen displayed place the cursor next to ESCAPE using the AIR A YES button SELECT SENSOR TO CALIBRATE gt ESCAPE CH4 OXY H2S CO 23 Press and release the POWER ENTER RESET button to return to the Calibration Mode Screen 24 Use the RANGE YW SHIFT button to place the cursor next to the NORMAL OPERATION menu item then press and release the POWER ENTER RESET button to return to Measuring Mode EAGLE 2 Operator s Manual Calibrating Using the Single Calibration Method 99 Parts List Table 12 Sample Bag Calibration Kit Spare Parts 06 1248RK 03 Calibration kit tubing 3 foot length 81 0090RK 01 Calibration cylinder 3 gas mix LEL Oxygen CO 34 liter steel 81 0154RK 04 Calibration cylinder 4 gas mix LEL Oxygen CO H2S 34 liter aluminum 81 1001RK Dispensing valve for 17 34 liter steel cylinders 81 1051RK 60 Regulator with gauge and knob 34 liter aluminum 58 liter 103 liter cylinders 6 LPM 81 1126RK Gas bag with clamp and hose barb 9 x 9 2 liter 81 5302RK LV Calibration kit for LEL Oxy CO unit w gas bag 34 Liter 81 5401RK LV Calibration kit for LEL Oxy HsS CO unit w gas bag 34 Liter 100 Parts List EAGLE 2 Operator s Manu
58. of the flow chamber easily enough using this method grasp it with a pair of pliers and rock it back and forth slightly while pulling on it CAUTION If using pliers to remove a sensor be careful not to damage the sensor in case you find that the sensor is still functional and does not need to be replaced e Hold the main PCB to support it where the sensor cable connects to it e Grasp the connector on the end of the sensor cable and pull the connector away from the main PCB to disconnect it from the main PCB 10 Install the new sensor 11 To install the LEL sensor do the following e Plug the replacement sensor into the sensor connector on the LEL sensor cable 82 Replacing a Sensor EAGLE 2 Operator s Manual e Insert the LEL sensor into the LEL sensor chamber in the flow chamber e Line up the holes in the LEL sensor bracket with the two standoffs on either side of the LEL sensor chamber e Install the two sensor bracket screws tightening them a little at a time alternately to push the sensor into its chamber evenly 12 To install the oxygen sensor do the following e Confirm that the sealing O ring is still in the bottom of the oxygen sensor chamber in the flow chamber and insert the oxygen sensor face down into the chamber e Route the sensor cable the same way the old sensor cable was routed and connect it to the main PCB Make sure to support the main PCB when making the connection e Reinstall the oxygen se
59. or obstruction The internal hydrophobic filter is dirty The pump is malfunctioning The auto calibration values may not match the cylinder gas concentrations auto calibration only The charcoal filter is saturated causing an elevated CO reading The sample gas is not reaching the sensors because of a bad connection The calibration cylinder may be out of gas or is outdated The sensor for the affected channel s may need replacement A memory error has occurred EAGLE 2 Operator s Manual Inspect the probe tube for any obstructions Inspect the hydrophobic filter disk in the probe and replace if necessary Inspect the sample hose for kinks or obstructions and replace if necessary Inspect the internal hydrophobic filter and replace if necessary If difficulties continue contact RKI Instruments Inc for further instruction Check all calibration tubing for leaks or for any bad connections Make sure the EAGLE 2 has been properly set up for calibration Change the charcoal filter Verify that the calibration cylinder contains an adequate supply of fresh test sample If the fail condition continues replace the sensor s If the difficulties continue contact RKI Instruments Inc for further instruction Press and hold the RANGE V SHIFT button then press the DISPLAY ADJUST NO button and release both The Enter Password Screen will appear Enter the passw
60. release POWER ENTER RESET to return to the main menu Start Up 1 Press and briefly hold down the POWER ENTER RESET button Release the button when you hear a beep 2 The LCD will show the following screen for about ten seconds GAS MONITOR INSTRUMENTS EAGLE 2 3 The Battery Voltage Screen appears for a few seconds BATTERY MIN 4 3 VOLTS BATTERY NOW 5 2 VOLTS 4 The Inert Mode Selection Screen will then be displayed MODE SELECT UP DOWN THEN ENTER NORMAL MODE Use the AIR A YES or RANGE v SHIFT button to display the mode you wish to enter If you have not made a selection after 20 seconds the instrument will begin to beep and the LEDs will begin to flash They will continue until you make a selection 296 Start Up EAGLE 2 Operator s Manual WARNING The EAGLE 2 is not a gas monitoring device until you select a mode and enter it 5 If you select Normal Mode the warm up sequence will continue as described in Start Up on page 22 unless Leak Check or Bar Hole Mode is active If Leak Check or Bar Hole Mode is active the Leak Check Bar Hole Mode Select Screen will appear immediately after the Normal Mode selection gt NORMAL MODE BAR HOLE MODE LEAK CHECK MODE Use the AIR A YES or RANGE vw SHIFT button to display the mode you wish to enter 6 If you select Inert Mode the warm up sequence will continue as described in Start Up on page 22 The oxygen alarms in the alarm point scre
61. s Manual Maintenance 169 16 17 18 19 20 21 22 23 Figure 34 Reinstalling Electrode Lamp Twisting the lamp slightly during insertion will help to ensure the lamp window is snug against the stack s front electrode The lamp should be supported by the O ring Continuing to hold the electrode stack between your forefinger and thumb carefully insert the lamp into the lamp cavity in the sensor ensuring that the lamp remains in position Press in the electrode stack firmly to ensure that the stack wing clips are engaged and the faces of the stack and sensor body are flush Carefully line up the PID sensor s pins with the sockets in the bottom of the sensor adapter and gently lower the sensor into the adapter until you feel it contact the bottom Do not attempt to push the sensor in farther once it makes contact with the bottom of the adapter until you are sure that the sensor s pins are engaged with the sockets If you feel that the pins did not engage the sockets slightly rotate the sensor back and forth without putting pressure on it until you feel the pins engage the sockets Push the sensor into the sockets until it bottoms out Insert the sensor adapter into the PID flow chamber and push it in until it bottoms out Confirm that the main PCB is seated in its slots and that its bottom edge 170 Maintenance EAGLE 2 Operator s Manual 24 25 26 is resting on the bottom of the bottom
62. that the finger cots be used on the fingers handling the lamp Finger cots are included with the lamp cleaning kit EAGLE 2 Operator s Manual Maintenance 161 10 Hold the PID sensor steady on the working surface with one hand and using the other hand locate the tabs on the electrode stack removal tool and insert them into the slots on the side of the PID sensor near the face Figure 27 Using Removal Tool 11 Squeeze the removal tool to push the tabs into the sensor slots until the electrode stack and lamp are released 12 Carefully lift the PID sensor body away from the electrode stack and lamp Take care not to touch the lamp window the flat end of the lamp with your fingers If the lamp remains lodged in the sensor body carefully remove it with tweezers N J in Lamp Window Figure 28 Lamp Window Location 13 If the spring in the lamp cavity comes out place it back into the lamp cavity 14 Hold the lamp in one hand being careful not to touch the lamp window with your fingers 15 With the other hand collect a small amount of aluminum oxide powder on a cotton swab 162 Maintenance EAGLE 2 Operator s Manual 16 Use this cotton swab to polish the PID lamp window Use a circular motion applying light pressure to clean the lamp window Do not touch the lamp window with your fingers Figure 29 Polishing the Electrode Lamp Window 17 Continue polishing until you can hear a sque
63. the EAGLE 2 s infrared hydrocarbon sensor and includes instructions to use an EAGLE 2 that has an infrared hydrocarbon sensor installed It also includes instructions to replace an infrared hydrocarbon sensor Table 30 Infrared Hydrocarbon Sensor Specifications LEL Configuration 0 100 LEL HC 1 LEL 10 LEL 50 LEL Table 31 Infrared Hydrocarbon Sensor Specifications Autoranging Configuration 0 100 LEL HC 1 LEL 10 LEL 50 LEL NOTE The gas name on the infrared hydrocarbon channel will normally read HC However depending on the instrument s shipment date and any special requirements your instrument may read IBU or some other gas name Target Gases The infrared HC sensor is a general hydrocarbon sensor It is setup for and factory calibrated to isobutane Description Table 30 and Table 31 above list the available infrared hydrocarbon sensor configurations The infrared hydrocarbon sensor is a smart sensor that stores the target gas The infrared hydrocarbon sensor is installed in a single sensor flow chamber which is located in the area next to the standard 4 sensor flow chamber This area can accommodate up to three single sensor flow chambers Figure 41 above illustrates a typical infrared hydrocarbon sensor location in front of the pump The infrared hydrocarbon flow chamber may also be installed in one of the other two sensor chamber locations depending on the particular version of the EAGLE 2 Som
64. the change Catalytic Sensor Relative Response Screen This screen displays only if RELATIVE RESPONSE in Setup Mode is set to ON see Updating the Catalytic Sensor Relative Response Setting on page 114 Use this screen to temporarily change the gas configuration of the EAGLE 2 Operator s Manual Display Mode 43 catalytic combustible channel SELECT RELATIVE RESPONSE TO CALIBRATED GAS FOR CATALYTIC SENSOR You can select from a list of gases whose response relative to the configured gas normally methane is programmed into the EAGLE 2 s memory This includes several pre defined gases and 5 gases that can be entered into the EAGLE 2 in the field using the Eagle 2 Maintenance Data Loader Program In order to program a field defined gas into the EAGLE 2 gas testing must be performed to determine the gas response factor relative to methane See the Eagle 2 Maintenance Data Loader Program Operator s manual for details regarding the gas testing and programming user defined gases into the EAGLE 2 s relative response list The last five items in the gas list are reserved for field defined gases The relative response feature enables you to temporarily monitor for the selected gas without having to recalibrate the EAGLE 2 The EAGLE 2 will clear the gas configuration change when it is turned off and will return to the programmed configuration when it is turned on again Because of normal variation between sensors
65. the faces of the stack and sensor body are flush Carefully line up the PID sensor s pins with the sockets in the bottom of the sensor adapter and gently lower the sensor into the adapter until you feel it contact the bottom Do not attempt to push the sensor in farther once it makes contact with the bottom of the adapter until you are sure that the sensor s pins are engaged with the sockets If you feel that the pins did not engage the sockets slightly rotate the sensor back and forth without putting pressure on it until you feel the pins engage the sockets Push the sensor into the sockets until it bottoms out Insert the sensor adapter into the PID flow chamber and push it in until it bottoms out Confirm that the main PCB is seated in its slots and that its bottom edge is resting on the bottom of the bottom case If the main PCB is not seated properly it may be damaged when the top case is re installed EAGLE 2 Operator s Manual Maintenance 167 24 Make sure that the top case gasket is fully seated in its groove and carefully put the top case back on the bottom case If you have any difficulty mating the top and bottom cases inspect the placement of the main PCB and the placement of the top case gasket 25 Turn the EAGLE 2 upside down and tighten the three case screws to secure the top case to the bottom case 26 Calibrate the PID channel as described in PID Calibration on page 143 Replacing the Electrode Sta
66. these relative response factors are typical factors If you use this feature the response to the selected gas will not be as accurate as it would be if you configured and calibrated the catalytic combustible channel to the target gas NOTE For maximum accuracy configure and calibrate the EAGLE 2 s catalytic combustible channel to the desired target gas 1 With the Relative Response Screen displayed press and release AIR A YES A list of gases will appear on the screen with EXIT at the top of the list There are multiple screens of gases gt EXIT ACETONE BENZENE BUTYL ACRYLATE BUTYL ACETATE 2 BUTYL ALCOHOL 44 Display Mode EAGLE 2 Operator s Manual The following is the complete list of factory defined gases Table 7 Relative Response Gas List Acetone Cumene Isobutane Methyl Ibutyl Mixed Xylenes Ketone Benzene Ethylene Isopropanol Nonane Field Defined Gas Dichloride Butyl Acrylate Ethyl Alcohol Field Defined Gas oranan maoan memoron ia 2 Use the AIR A YES or RANGE v SHIFT buttons to move the cursor next to the desired gas 3 Press and release POWER ENTER RESET The catalytic combustible channel will be configured to the selected gas and the EAGLE 2 will proceed to the STEL Screen This configuration will be in force until either a different gas is selected in Display Mode or the unit is turned off NOTE If a PID sensor is installed in the EAGLE 2 and CATALYTIC SENSOR RELATIVE RESPONSE
67. to ALARM LATCHING 2 Press and release POWER ENTER RESET The Alarm Latching Screen appears ALARM LATCHING UP DOWN THEN ENTER LATCHING Use AIR A YES or RANGE YW SHIFT to display the desired setting 4 Press and release POWER ENTER RESET to save the setting and return to the main menu 116 Using Setup Mode EAGLE 2 Operator s Manual Updating the Alarm Silence Setting With ALARM SILENCE set to ON factory setting pressing and releasing any button silences the buzzer when the EAGLE 2 is in alarm The LEDs continue to flash and the display continues to show the alarm When the gas concentration falls below the alarm level pressing and releasing POWER ENTER RESET clears all alarm indications for that alarm If an alarm condition occurs you may still enter Display Mode by pressing the DISPLAY button The buzzer will be silenced but the LEDs will continue to flash If you return to Measuring Mode and there is still an alarm condition the LEDs will continue to flash and the buzzer will remain off Once the condition clears press POWER ENTER RESET to clear the alarm indications With ALARM SILENCE set to OFF you cannot silence the buzzer If an alarm condition occurs and you enter Display Mode the buzzer will not be silenced and the LEDs will continue to flash Upon return to Measuring Mode if there is still an alarm condition you must wait until it clears before you can press POWER ENTER RESET to clear the al
68. to AUTO CALIBRATION CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 182 ESM 01 Calibration EAGLE 2 Operator s Manual 6 Move the cursor to the PERFORM AIR ADJUST menu item by using the RANGE YW SHIFT button CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION gt PERFORM AIR ADJUST NORMAL OPERATION 7 Press and release the POWER ENTER RESET button The following screen appears PERFORM AIR ADJUST 8 Press and release the AIR A YES button to continue If you do not want to continue press the DISPLAY ADJUST NO button and the unit will return to the Calibration Mode Screen EAGLE 2 Operator s Manual ESM 01 Calibration 183 9 The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds then indicate that the operation is complete before returning to the Calibration Mode Screen ADJUSTING ZERO RELEASE AIR BUTTON ADJUSTING ZERO COMPLETE CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION gt PERFORM AIR ADJUST NORMAL OPERATION 10 Install the demand flow regulator onto the multi gas calibration cylinder 11 Connect the sample tubing to the demand flow regulator 12 Install the probe on the EAGLE 2 inlet fitting Make sure the probe is complete with internal O ring and membrane and that the two halves of the probe are tightened firmly together to
69. to select the 230 Infrared CO Calibration EAGLE 2 Operator s Manual channel The calibration gas value begins to flash 22 Use the AIR A YES and RANGE YW SHIFT buttons to adjust the calibration gas setting to the desired value NOTE The calibration gas value cannot be set lower than the low alarm setting If the calibration gas value listed on the calibration cylinder is lower than the current low alarm setting enter Setup Mode and change the low alarm setting See Updating the Alarm Point Settings on page 114 for instructions If you need to change the alarm point setting only to perform a calibration make sure that you change the alarm point setting back to its original value once the calibration has been performed 23 Press and release the POWER ENTER RESET button to save the change The calibration gas value stops flashing 24 Repeat Step 17 through Step 21 for any other channels that need to be changed 25 When you are done adjusting the calibration gas values move the cursor down past the bottom of the screen next to END ADJUST AUTO CALIBRATION VALUES gt END A 26 Press and release the POWER ENTER RESET button The following screen appears DO YOU WANT TO STORE NEW VALUE S IN MEMORY FOR FUTURE CALIBRATIONS PRESS YES OR NO 27 If you select YES by pressing and releasing the AIR A YES button the changes that you made will be saved in the EAGLE 2 s memory as the new a
70. to the channel of the alarm point you want to update Press and release POWER ENTER RESET The channel s alarm points are displayed in this example for the catalytic combustible sensor channel CH4 0 100 LEL gt LO ALRM 10 LEL HI ALRM 50 LEL LO ALRM 5000 ppm HI ALRM 25000 ppm 4 Move the cursor next to the alarm point or alarm operation oxygen only that you want to update If you selected the oxygen channel you can set the alarm rising falling operation in addition to the alarm points NOTE If Inert Mode is active you can change the oxygen alarm points for both Normal Mode and Inert Mode For more information about changing Inert Mode alarm settings see Appendix N Using the EAGLE 2 in Inert Mode on page 293 OXY 0 40 0 vol gt FALLING AND RISING LO ALRM 19 5 vol HI ALRM 23 5 vol END 5 Press and release POWER ENTER RESET The alarm point or alarm operation oxygen only will begin to flash 6 Use AIR A YES and RANGE VW SHIFT to adjust the alarm point or alarm operation oxygen only to the desired setting Keep the following in mind e The low alarm cannot be set higher than the high alarm and the high alarm cannot be set lower than the low alarm e Any alarm setting can be turned off by adjusting it to its lowest setting The setting will be displayed as OFF e In addition to setting the oxygen alarm points you can also select one of the following operation
71. to the sub PCB with a 12 position connector The sub PCB plugs into the main PCB and is held in place with a screw flat washer lock washer There are no user serviceable parts on the infrared sub PCB Start Up and Normal Operation For instructions to startup and use an EAGLE 2 that includes an infrared methane sensor reference Start Up on page 22 Measuring Mode Normal Operation on page 29 and Measuring Mode Alarms on page 35 Detection Ranges The infrared methane sensor can be factory set to detect gas in a 0 100 LEL configuration an autoranging configuration or a 0 100 0 volume configuration The autoranging configuration detects gas on a 0 100 LEL and a 5 0 100 0 vol scale The gas reading will be displayed in LEL until the gas level reaches 100 LEL or 5 0 vol for methane Once the gas reading is above 100 LEL it is displayed in vol Alarm points exist for the LEL range but do not exist for the vol range Catalytic LEL Sensor Screen When either a TC sensor or an infrared combustible sensor is installed in an EAGLE 2 along with a catalytic combustible LEL sensor the user has the option of turning off the catalytic combustible LEL sensor in Display Mode If the unit is going to be used for sampling known high levels of combustible gas or in areas with known catalytic sensor poisons such as silicone vapors the catalytic combustible sensor should be turned off Even though this 240 Start U
72. tubing to the demand flow regulator 12 Install the probe on the EAGLE 2 inlet fitting Make sure the probe is complete with internal O ring and membrane and that the two halves of EAGLE 2 Operator s Manual Infrared Methane Calibration 243 the probe are tightened firmly together to avoid leaks that can affect the calibration See Figure 19 Replacing the Particle Filter and Hydrophobic Filter Disk on page 76 for an illustration of the internal parts of the probe 13 Move the cursor next to the AUTO CALIBRATION menu item by using the AIR A YES button CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 14 Press and release the POWER ENTER RESET button to display the Calibration Gas Values Screen CAL GAS VALUES CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm ENTER TO BEGIN CAL The gas concentrations displayed in the Calibration Gas Values Screen must match the gas concentrations listed on the 4 gas calibration cylinder If all concentrations match go to Step 25 If one or more concentrations do not match continue with Step 15 If you do not want to continue with the calibration press and release the DISPLAY ADJUST NO button to return to the Calibration Mode Screen NOTE The RKI 4 gas cylinder typically contains 12 O by volume When using the auto calibration method be sure to set the OXY auto calibration value to agree with the
73. variety of toxic gases This appendix describes the EAGLE 2 s ESM 01 sensors and includes instructions to use an EAGLE 2 that has one or more ESM 01 sensors installed It also includes instructions to replace an ESM 01 sensor Table 18 ESM 01 Sensor Specifications Target Gas Detection Reading Alarm 1 Alarm 2 STEL TWA Range ppm Increment Factory Factory ppm ppm Setting Setting Ammonia 0 75 0 0 5 ppm 12 0 ppm 25 0 ppm 25 NH3 Arsine 0 1 50 0 01 ppm 0 20 ppm 0 50 ppm O OFF AsH3 Chlorine 0 3 00 0 02 ppm 0 50 ppm 1 50 ppm 0 50 Clo Hydrogen 0 15 0 0 1 ppm 3 00 ppm 5 00 ppm OFF Cyanide HCN Phosphine 0 1 00 0 01 ppm 0 10 ppm 0 30 ppm 0 30 PH3 50 35 FF 1 00 4 7 1 00 Sulphur 0 6 00 0 05 ppm 1 00 ppm 2 00 ppm 1 00 0 Dioxide SOo Description Table 18 above lists the available ESM 01 sensors The ESM O01 is a smart sensor that stores sensor parameters including the target gas detection range alarm points and calibration settings in its memory So a sensor can be calibrated at the factory and shipped as a replacement sensor without the need to calibrate the sensor when it is installed as long as it is installed during the sensor s valid calibration period which is typically 3 months In addition you can change an existing ESM 01 channel from one type of ESM 01 sensor to another and the EAGLE 2 will automatically load all the sensor parameters and configure the ESM 01 channel for the new sensor withou
74. when GOODBYE appears on the display When GOODBYE disappears and the backlight turns off the unit is off 286 Leak Testing EAGLE 2 Operator s Manual Appendix M Tank Tester Model The EAGLE 2 Tank Tester model is intended for checking tanks or vessels that may contain residual hydrocarbon vapors or water or may have been purged of oxygen It is supplied as an LEL catalytic only unit or an LEL catalytic oxygen unit You can also use this model as a standard EAGLE 2 gas monitor by connecting the standard hose and probe and selection Normal Mode in the Inert Mode Selection Screen Table 34 Alarm Points for Tank Tester Configuration 10 LEL 50 LEL 1100 ppm 5500 ppm Normal Mode Oxygen Oxy 19 5 falling Oxy 23 5 rising Inert Mode Oxygen Oxy 5 0 rising Oxy 10 0 rising Description The tank tester model has an additional socket on the front on the housing to accommodate connection of the float probe assembly ae ee A 7 N _ Quick Connect Float mee Inlet Fitting Probe Socket IR Pot 1 Figure 42 EAGLE 2 Tank Tester Version This model includes the following non standard components EAGLE 2 Operator s Manual Description 287 Float Probe Assembly The float probe assembly helps prevent liquid from being drawn into the EAGLE 2 The float probe assembly is 12 feet long The hose between both ends of the float probe has an
75. you to select between alkaline and Ni MH batteries Since the discharge characteristics of alkaline and Ni MH batteries are different the EAGLE 2 uses this setting to ensure that the low battery warning is in effect long enough before a dead battery alarm to allow the user to change the batteries without a dead battery alarm occurring This setting has no effect on battery charging 1 From the main menu place the cursor next to SET BATTERY TYPE 2 Press and release POWER ENTER RESET The Battery Type screen appears with the current setting flashing BATTERY TYPE UP DOWN THEN ENTER ALKALINE 3 Use AIR A YES or RANGE VY SHIFT to display the desired setting 4 Press and release POWER ENTER RESET to save the setting and return to the main menu Configuring the Channels This menu item allows you to set the channel type for each of the six channels or to turn one or more channels off Although the standard EAGLE 2 is factory configured for four channels combustible gas catalytic sensor oxygen HS and CO with channels 5 and 6 turned off the EAGLE 2 can be factory and field configured for a variety of active channels and detector types It is not normally necessary to change the factory channel EAGLE 2 Operator s Manual Using Setup Mode 105 configuration CAUTION Before changing the channel configuration confirm that the correct sensors and electronic hardware are installed in the EAGLE 2 and that its cons
76. 0 VAC if an appropriate plug adapter is provided The AC charging station is shown below in Figure 7 AC Adapter PANASS Charging Module Figure 7 EAGLE 2 AC Charger EAGLE 2 Operator s Manual Optional Accessories 17 DC Charger An optional DC powered charger is available with a vehicle plug 12 VDC adapter It uses the same charging module as the standard AC charger Vehicle Plug DC Adapter Charging Module Figure 8 EAGLE 2 DC Charger AC DC Charger A charger is also available that includes both the AC adapter and the 12 VDC vehicle plug adapter The charging module is the same as the one used for the AC charger and the DC charger Optional Probes Various optional probes designed for specific applications are available for the EAGLE 2 They include the following e 30 inch aluminum probe Ce Figure 9 30 Inch Aluminum Probe This probe is designed for applications where it is necessary to put the probe tip in areas that are out of reach with the standard probe A small breather hole near the end of the probe tube prevents interruption of sampling and a low flow alarm if the probe tip is blocked 18 Optional Accessories EAGLE 2 Operator s Manual e 30 inch stainless steel probe Figure 10 30 Inch Stainless Steel Probe This probe is physically the same as the 30 inch aluminum probe and is intended for applications where a high level of corrosion resistance is require
77. 000 ees 152 Maintenances sesetan esd svenwentonebaneee ay ond 8 ete OSE os 160 Troubleshooting cae silo he et apitecigs Ons rer wate wane a asks PE E Si 160 Cleaning the PID Sensor s Lamp 0 000 cee 160 Replacing PID Sensor LAMD s ice Sc ee ole te Bw eae Aan eA 165 Replacing Electrode Slack ei2 suis a ese OSS RSLS dy SE 168 Replacing the PID Sensor 0 cnn 171 Configuring the PID Gas in Setup Mode 000 4 172 Partsi EDS teed he MSNA heed eG E Rie eh A ele hae le a AS E A 178 Appendix E ESM 01 Toxic Sensors 0 0 cc ee eeeeeee 179 OVI VIC Ws ecstasy a eis ee en oe ep a ech es es 179 Des ription s hits hte Sale ewes a Eh alee let eke Cina Dian erect Mead 179 ESM OrSEnSOr marawani de Wes Ge od Cuil EER EE Gaal EENS 180 ESM OL SUD PC Be hn ticsndetins gawd tides wate oe HO cae ei Reeser 181 Start Up and Normal Operation 00 e eee ee eens 181 ESM OL Calibrations seo v5 9 enet eke ees te E eat eds ve 181 Calibrating with a 4 Gas Cylinder and an ESM O1 Cylinder 182 Calibrating with a 5 Gas Cylinder 0 0 6 0c cee 191 Maintenance ae ent a E e E aa ds See E a 199 Replacing the ESM 01 Sensor lt 02 se bd Ns ai es ee aE 199 Replacing the HS Scrubber in the SO and HCN Sensors 200 Parts ESE iba soa sds ree a Gane ake Rae e Hee Ae ete ae see 202 Appendix F TC Sensors ccc cece ccc c cece cece sseees 204 OVCIVIEW cc e a E aus k olsen a EEEE ewe A Bay erate ee oe 204
78. 1 Install the demand flow regulator onto the calibration cylinder 2 Connect the sample tubing to the demand flow regulator 3 Install the probe on the EAGLE 2 inlet fitting Make sure the probe is complete with internal O ring and membrane and that the two halves of the probe are tightened firmly together to avoid leaks that can affect the calibration See Figure 19 Replacing the Particle Filter and Hydrophobic Filter Disk on page 76 for an illustration of the internal parts of the probe EAGLE 2 Operator s Manual Calibrating Using the Auto Calibration Method 59 4 Move the cursor next to the AUTO CALIBRATION menu item by using the AIR A YES button CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 5 Press and release the POWER ENTER RESET button to display the Calibration Gas Values Screen CAL GAS VALUES CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm ENTER TO BEGIN CAL The gas concentrations displayed in the Calibration Gas Values Screen must match the gas concentrations listed on the 4 gas calibration cylinder If all concentrations match go to Step 16 If one or more concentrations do not match continue with Step 6 If you do not want to continue with the calibration press and release the DISPLAY ADJUST NO button to return to the Calibration Mode Screen NOTE The RKI 4 gas cylinder typically contains 12 O by volume When using
79. 1 Calibration Gas Value Screen below instead of the standard channel Calibration Gas Value Screen CAL GAS VALUE NH3 25 ppm ENTER TO BEGIN CAL 32 With the ESM 01 Calibration Gas Value Screen displayed press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen for the ESM 01 channel with CAL IN PROCESS flashing CAL IN PROCESS NH3 Oppm ENTER WHEN DONE If you do not want to proceed with the calibration press and release the DISPLAY ADJUST NO button to return to the ESM 01 Cal Gas Values Screen If you do want to continue with the calibration proceed to the next step 33 Screw the ESM 01 calibration cylinder onto the demand flow regulator 34 Connect the tubing from the regulator to the rigid tube on the probe Allow the EAGLE 2 to draw gas for two minutes If you notice that the gas reading stabilizes sooner you can reduce the gas application time to the time it takes the gas reading to stabilize 35 Press and release the POWER ENTER RESET button to set the span adjustment for the ESM 01 channel to the programmed value EAGLE 2 Operator s Manual ESM 01 Calibration 189 36 If the ESM 01 channel passed calibration the following screen sequence occurs AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END Y CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION I
80. 1 flow chamber and push it in until it bottoms out Line up the holes in the ESM 01 sensor bracket with the two standoffs on EAGLE 2 Operator s Manual Maintenance 199 14 15 16 17 18 the ESM 01 chamber Install the two sensor bracket screws Confirm that the main PCB is seated in its slots and that its bottom edge is resting on the bottom of the bottom case If the main PCB is not seated properly it may be damaged when the top case is re installed Make sure that the top case gasket is fully seated in its groove and carefully put the top case back on the bottom case If you have any difficulty mating the top and bottom cases inspect the placement of the main PCB and the placement of the top case gasket Turn the EAGLE 2 upside down and tighten the three case screws to secure the top case to the bottom case Calibrate the ESM O1 channel as described in ESM 01 Calibration on page 181 Replacing the H2S Scrubber in the SO and HCN Sensors 1 2 3 Verify that the EAGLE 2 is off Place the EAGLE 2 upside down on a flat surface or hold it upside down Unscrew the three case screws until they disengage from the top case They are captive screws so they will not fall off of the bottom case Turn the EAGLE 2 right side up and carefully lift the top case away from the bottom case Be careful not to lift it so far that it pulls on the main PCB with the cable that connects the top case to the main PCB La
81. 2 TC Calibration EAGLE 2 Operator s Manual 27 Press and release the POWER ENTER RESET button to set the span adjustment for each channel to the programmed values 28 If all channels passed calibration the following screen sequence occurs AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END Y CAL GAS VALUE CH4 50 0 vol ENTER TO BEGIN CAL If any of the sensors cannot be adjusted to the proper value a screen displays that indicates a calibration failure and lists the sensor s that failed to calibrate In the example below the oxygen and HS channels failed calibration The other sensors calibrated normally FAIL SENSOR lt gt lt OXY gt lt H2S gt lt gt ENTER TO CONTINUE The buzzer and alarm LEDs activate in a double pulsing pattern Press and release the POWER ENTER RESET button to reset the alarm and continue to the Calibration Value Screen for the TC channel After calibrating the TC channel by following the instructions below attempt to calibrate the standard channels again If the failure continues investigate the cause See Troubleshooting on page 70 29 Remove the tubing from the rigid tube on the probe 30 Unscrew the 4 gas cylinder from the demand flow regulator EAGLE 2 Operator s Manual TC Calibration 213 31 If you want to change the TC channel s calibration gas value follow Step 15 Step 24 above beginning with the T
82. 20 9vol ALRM2 E R T 290 Start Up and Operation EAGLE 2 Operator s Manual If the instrument is sampling a fresh air environment the oxygen channel will be in alarm For more information about Inert Mode see Appendix N Using the EAGLE 2 in Inert Mode on page 293 8 When monitoring for combustible gas in an inerted tank be sure the dilution fitting is installed and multiply the combustible gas reading by 2 When monitoring for oxygen in an inerted tank remove the dilution fitting or seal the dilution hole with your finger Alarms The EAGLE 2 Tank Tester model has two sets of oxygen alarm settings one for Normal Mode and one for Inert Mode The Inert Mode oxygen alarms are factory set at 5 0 by volume rising and 10 0 by volume rising Normal Mode oxygen alarm setpoints also exist and take effect when you select Normal Mode at start up The rising Inert Mode alarms are used to monitor a purged vessel to alert you to a rising oxygen condition The oxygen level in a fresh air environment is above both Inert Mode oxygen alarm points so the EAGLE 2 Tank Tester version will go into alarm when turned on in Inert Mode in a fresh air environment To silence the alarm press the RESET button The audible alarm silences but the alarm lights continue to flash and the display screen continues to indicate an oxygen alarm If one of the alarm levels is newly exceeded the audible alarm sounds again For inform
83. 3 Methyl Acrylate 1 19 Cyclohexane 2 72 0 71 Methyl Ethyl 2 66 0 74 Ketone Cumene 4 46 Methyl Isobutyl 3 33 Ketone Ethylene 5 21 Mixed Xylenes 3 47 Dichloride Ethyl Alcohol Nonane 0 68 Ethyl Chloride Pentane 0 41 Ethyl Acrylate Propane 0 51 Hexane Styrene 0 80 Hydrogen Toluene 0 70 Monomer e The EAGLE 2 provides the catalytic combustible sensor with some protection against exposure to high levels of combustible gas which can damage the sensor It does this by turning off the combustible sensor power temporarily when it determines that an over scale more than 100 LEL concentration of combustible gas is present that may damage the sensor Nevertheless concentrations of combustible gas of more than 100 LEL can still affect the zero level or calibration of the combustible sensor if the concentration is high enough CAUTION Do not expose the catalytic combustible sensor to high concentrations of combustible gas such as that from a butane lighter Exposure to high concentrations of combustible gas may adversely affect the performance of the sensor 32 Measuring Mode Normal Operation EAGLE 2 Operator s Manual CAUTION Any rapid increase in the combustible gas reading on the catalytic combustible channel followed by a declining or erratic reading may indicate a gas concentration above the LEL which may be hazardous e Some gases such as silicone vapors chlorinated hydrocarbons and sulphur compounds can contamina
84. 43RK 02 81 0143RK 04 81 1054RK 81 1055RK 82 0003RK 82 0300RK 178 Parts List Table 17 PID EAGLE 2 Spare Parts Electrode stack 0 50 ppm 2 pack Electrode stack 0 2 000 ppm 2 pack Replacement lamp 0 50 ppm Replacement lamp 0 2 000 ppm Adapter assembly for PID sensor PID sensor 0 50 ppm VOC PID sensor 0 2 000 ppm VOC PID probe particle filter removed Teflon lined sample hose 5 feet Teflon lined sample hose 10 feet Teflon lined sample hose 15 feet Teflon lined sample hose 20 feet Calibration cylinder isobutylene 100 ppm in air 34 liter steel Calibration cylinder isobutylene 100 ppm in air 103 liter steel Calibration cylinder isobutylene 100 ppm in air 34 liter aluminum Calibration cylinder isobutylene 10 ppm in air 34 liter steel Calibration cylinder isobutylene 10 ppm in air 103 liter steel Calibration cylinder isobutylene 10 ppm in air 34 liter aluminum Calibration cylinder 5 gas IBL CHy Oo H2S CO 58 liter aluminum Calibration cylinder 5 gas IBL CHy Oo HoS CO 34 liter aluminum Regulator demand flow type for 58 and 103 liter aluminum or steel and 34 liter aluminum cylinder Regulator demand flow type for 17 and 34 liter steel cylinder Electrode stack removal tool Lamp cleaning kit with electrode stack removal tool EAGLE 2 Operator s Manual Appendix E ESM 01 Toxic Sensors Overview The ESM 01 sensors are used to monitor levels of a
85. A YES or RANGE V SHIFT button to display Inert Mode and press and release POWER ENTER RESET MODE SELECT UP DOWN THEN ENTER NORMAL MODE NOTE To change the Normal Mode oxygen other alarm points select Normal Mode and see Updating the Alarm Point Settings on page 114 OXY 0 40 0 vol INERT ALARM gt LO 5 0 vol HI 10 0 vol END 10 Press and release POWER ENTER RESET The alarm point or alarm operation oxygen only will begin to flash 11 Use AIR A YES and RANGE YW SHIFT to adjust the alarm point to the desired setting Keep the following in mind e The low alarm cannot be set higher than the high alarm and the high alarm cannot be set lower than the low alarm e Any alarm setting can be turned off by adjusting it to its lowest setting The setting will be displayed as OFF 12 If you want to continue with the change press and release POWER ENTER RESET to accept the setting If you want to exit this screen without saving any change to the alarms press and release DISPLAY ADJUST NO until you return to the Change Alarm Point Settings Screen EAGLE 2 Operator s Manual Alarms 295 13 When you are done making changes use RANGE YW SHIFT to move the cursor next to END 14 Press and release POWER ENTER RESET to save the new settings and return to the Change Alarm Point Settings Screen 15 Use RANGE VW SHIFT to move the cursor next to END 16 Press and
86. AGLE 2 automatically resumes accumulating the peak and TWA readings The EAGLE 2 will also continue to keep track of operating time including the operating time EAGLE 2 Operator s Manual Start Up 23 from the last time the EAGLE 2 was used See Time in Operation Screen on page 51 for more information about how the EAGLE 2 tracks the operating time The short term exposure limit STEL reading is reset each time the EAGLE 2 is turned on e To reset the accumulation of these measurements press and release the DISPLAY ADJUST NO button before the countdown reaches 0 8 The gas alarm setpoints are displayed by three screens in sequence the Low Alarm Screen High Alarm Screen and STEL TWA Alarm Screen Each screen remains on the LCD for three seconds A CH4 10 LEL LL OXY 19 5 vol OA H2S 10 0 ppm WR CO 25 ppm M CH4 10 vol S A CH4 50 LEL HL OXY 23 5 vol IA H2S 30 0 ppm GR CO 50 ppm HM CH4 50 vol S ALARMS STEL amp TWA H2S ppm 15 0 10 0 CO ppm 200 25 9 After the alarm screens if CAL REMINDER is turned on the screen that appears next depends on how CAL PAST DUE ACT is set in the Setup Mode Menu see Updating the Calibration Past Due Action Setting on page 123 24 Start Up EAGLE 2 Operator s Manual e Ifthe unit is due for calibration and CAL PAST DUE ACT is set to CONFIRM TO CAL then the following screen displays and the buzzer sounds in a double pulsing pattern
87. Appendix F TC Sensors Overview The TC sensors are used to monitor high levels of combustible gas This appendix describes the EAGLE 2 s TC sensor and includes instructions to use an EAGLE 2 that has a TC sensor installed It also includes instructions to calibrate and replace a TC sensor Table 20 TC Sensor Specifications Target Gas Detection Reading Alarm 1 Alarm 2 Range Increment Factory Factory Setting Setting Methane 0 100 0 vol VvOl OFF OFF CH4 0 0 5 Hydrogen 0 100 0 vol 0 5 vol 10 0 vol 50 0 vol H Hydrogen 0 10 0 vol 0 1 vol 1 0 vol 5 0 vol H Description The TC sensor is installed in a single sensor flow chamber which is located in the area next to the standard 4 sensor flow chamber This area can accommodate up to three single sensor flow chambers Figure 37 below illustrates a typical TC sensor location in front of the pump The TC flow chamber may also be installed in one of the other two sensor chamber locations depending on the particular version of the EAGLE 2 Some TC sensor instrument configurations do not include the 4 sensor flow chamber 204 Overview EAGLE 2 Operator s Manual Top Case Not Shown x Charcoal Filter TC Sensor Pump _ Standard Flow Chamber Y O E f O SS all 2 CO Sensor ya A Oxygen Sensor A N
88. B connects to that sub PCB with a cable There are four types of sub PCBs each supporting one type of optional sensor e The PID sub PCB supports a low range or a high range PID photo ionization detector sensor e The ESM 01 sub PCB supports any of the ESM 01 family toxic sensors e The TC sub PCB supports the EAGLE 2 TC thermal conductivity sensor e The Infrared IR sub PCB supports any of the EAGLE 2 IR sensors for combustible gas or CO3 There are no user serviceable parts on the sub PCBs Channel Setup amp Sub PCBs When a sub PCB is installed in a particular location it is reflected in the CONFIGURE CHANNELS and CONFIGURE GASES menu items in Setup Mode Sub PCBs and CONFIGURE CHANNELS A channel configured as an optional sensor type supported by a sub PCB is indicated by OP1 OP2 or OP3 specifying which sub PCB position is mapped to that channel OP1 indicates a channel supported by the SUB1 PCB OP2 indicates a channel supported by the SUB2 PCB and OP3 indicates a channel supported by the SUB3 PCB The example LCD screen below illustrates the first screen in the CONFIGURE CHANNELS menu item in Setup Mode and shows a channel configuration for LEL catalytic oxygen HS and a PID sensor with the sub PCB for the PID sensor installed in location SUB1 CONFIGURE CHANNELS gt CH1 CAT CH4 CH2 OXY OXY CH3 H2S H2S CH4 OP1 PID y If you configure a channel as an optional type that is map
89. C Calibration Gas Value Screen below instead of the standard channel Calibration Gas Value Screen CAL GAS VALUE CH4 50 0 vol ENTER TO BEGIN CAL 32 With the TC Calibration Gas Value Screen displayed press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen for the TC channel with CAL IN PROCESS flashing CAL IN PROCESS CH4 0 0 vol ENTER WHEN DONE If you do not want to proceed with the calibration press and release the DISPLAY ADJUST NO button to return to the TC Cal Gas Values Screen If you do want to continue with the calibration proceed to the next step 33 Screw the new TC calibration cylinder onto the demand flow regulator 34 Connect the tubing from the regulator to the rigid tube on the probe Allow the EAGLE 2 to draw gas for one minute 35 Press and release the POWER ENTER RESET button to set the span adjustment for the TC channel to the programmed value 214 TC Calibration EAGLE 2 Operator s Manual 36 If the TC channel passed calibration the following screen sequence occurs AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END Y CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION If the TC channel cannot be adjusted to the proper value a screen displays that indicates a calibration failure and indicates that the TC sensor failed to cali
90. CH4 10 vol S A CH4 50 LEL HL OXY 23 5 vol IA H2S 30 0 ppm GR CO 50 ppm HM CH4 50 vol ALARMS STEL amp TWA H2S ppm 15 0 10 0 CO ppm 200 25 CH4 vol OFF OFF 11 After the alarm screens if CAL REMINDER is turned on the screen that appears next depends on how CAL PAST DUE ACT is set in the Setup Mode Menu see Updating the Calibration Past Due Action Setting on page 123 272 Start Up Bar Hole Mode EAGLE 2 Operator s Manual e Ifthe unit is due for calibration and CAL PAST DUE ACT is set to CONFIRM TO CAL then the following screen displays and the buzzer sounds in a double pulsing pattern CALIBRATION DATE IS PAST DUE PERFORM CALIBRATION To perform a calibration press and release the AIR A YES button The EAGLE 2 will enter Calibration Mode and the LCD will show the Calibration Mode main menu See Chapter 4 Calibration Mode on page 55 for instructions to calibrate the EAGLE 2 When you are done with the calibration and exit Calibration Mode the unit will begin the startup sequence If the calibration was successful the screen above will not appear again until the unit is due for calibration If the calibration was not successful the screen above will again appear in the startup sequence To continue without performing a calibration press and release the DISPLAY ADJUST NO button e Ifthe unit is due for calibration and CAL PAST DUE ACT is set to MUST CA
91. Configuring the Combustible Gas on page 108 for instructions to setup the catalytic combustible channel gas CAUTION The EAGLE 2 catalytic combustible sensor experiences a significant gas response drop when it is changed from full response mode to methane elimination mode The EAGLE 2 must be calibrated in both full response mode and in methane elimination mode to account for this To properly calibrate the instrument do the following 1 With the EAGLE 2 catalytic combustible channel set up for hexane perform a demand zero and a calibration while in full response See Chapter 4 Calibration Mode on page 55 for instructions 2 Enable methane elimination in Display Mode 3 Allow the unit to stabilize for 2 minutes 4 Perform a demand zero and a calibration while in methane elimination mode See Chapter 4 Calibration Mode on page 55 for instructions EAGLE 2 Operator s Manual Calibration 269 Appendix K Using the EAGLE 2 in Bar Hole Mode Overview This chapter explains how to operate the EAGLE 2 in Bar Hole Mode Bar Hole Mode is used to perform consistent checks of bar holes when tracking down underground gas leaks When the EAGLE 2 is in Bar Hole Mode only the combustible and oxygen sensors are displayed If an EAGLE 2 is intended for bar hole testing it is shipped with Bar Hole Mode or both Bar Hole Mode and Leak Check Mode enabled so that the operator must choose which operational mode to use when the un
92. DESCHpPlIOhs ss os sauces esata e ead eee ete aes dos ade S 204 TE CUSOT oo x ates ahs Deed Seek Ace ae Ge EE ENA eee Brotha ee 205 EO SUD PCB naan e te Reais ba ah ye aoe fs Totti by eS ES 205 Start Up and Normal Operation 00 0 e eee eee eee 206 Catalytic LEL Sensor Screen 44 0504 bape stat oe Weaee eae tees 206 EAGLE 2 Operator s Manual Table of Contents TC Ca libratoi s vecnst scary ot bento eid Ware Sal vee a lade Ae OR ee ae 207 Ma inte anc ft aioe ride fice hein AE GS Ae De COREE OLS eta E ate ee 216 Replacing the TC Sensor 25 vnb xn bea ea ae eee bale ok Baek 216 Configuring the TC Gas in Setup Mode 0000 eae 217 Pants PAS aah Gace e e in od Ae ata Are OS teens Osa GE ele EE 222 Appendix G Infrared Carbon Dioxide Sensors 223 OVEIVI EW a eie eann E E EE E E EE E AE see E TER 223 Des ription esen erreen oaa en a e a teh E k 223 Infrared Carbon Dioxide Sensor 0 0 00 eee 224 Infrared SUD P CBs Be vty aids wba eal EAE Bae RE ES RS 224 COs SCUDDER as hata ec y died ee Reni aa EE pode a aE 225 Start Up and Normal Operation 0 0000 0 e sede ese eee eeeee 226 Performing a Demand Zero for Carbon Dioxide Sensors 226 Infrared Carbon Dioxide Calibration 0 000 e eee eee 227 Maintenant eri desir ke dnt eedd ennt EE EASE ea E EEEE ENEE EA 236 Replacing the IR CO Sensor or Changing Sensor Type 236 Parteist 24 4 tae h bee whee take oe OE a oE oe we
93. E 2 Operator s Manual Optional Accessories 21 Chapter 3 Operation Overview This chapter explains how to use the EAGLE 2 to perform confined space entry monitoring or general area monitoring in Normal Mode There are three operational modes in Normal Mode Measuring Mode Display Mode and Calibration Mode While in Normal Mode the unit is normally operating in Measuring Mode Display Mode and Calibration Mode are accessible from Measuring Mode Display Mode is described in this chapter Calibration Mode is described in Chapter 4 Calibration Mode on page 55 Special versions of the EAGLE 2 can also operate in Leak Check Mode and Bar Hole Mode See Appendix K Using the EAGLE 2 in Bar Hole Mode and Appendix L Using the EAGLE 2 in Leak Check Mode for operating instructions for Bar Hole and Leak Check Mode respectively Start Up This section explains how to start up the EAGLE 2 get it ready for operation and turn it off NOTE The screens illustrated in this section are for a standard 4 gas unit The screens displayed by your EAGLE 2 may be slightly different Turning On the EAGLE 2 To illustrate certain functions the following description of the EAGLE 2 start up sequence assumes that the following menu items in Setup Mode are turned on LUNCH BREAK CAL REMINDER and USER STATION ID If any of these items are turned off then the corresponding screens will not appear The EAGLE 2 may be used with
94. E 2 inlet fitting Make sure the probe is complete with internal O ring and membrane and that the two halves of the probe are tightened firmly together to avoid leaks that can affect the calibration See Figure 19 Replacing the Particle Filter and Hydrophobic Filter Disk on page 76 for an illustration of the internal parts of the probe EAGLE 2 Operator s Manual Infrared Hydrocarbon Calibration 257 13 Move the cursor next to the AUTO CALIBRATION menu item by using the AIR A YES button CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 14 Press and release the POWER ENTER RESET button to display the Calibration Gas Values Screen CAL GAS VALUES CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm ENTER TO BEGIN CAL The gas concentrations displayed in the Calibration Gas Values Screen must match the gas concentrations listed on the 4 gas calibration cylinder If all concentrations match go to Step 25 If one or more concentrations do not match continue with Step 15 If you do not want to continue with the calibration press and release the DISPLAY ADJUST NO button to return to the Calibration Mode Screen NOTE The RKI 4 gas cylinder typically contains 12 O by volume When using the auto calibration method be sure to set the OXY auto calibration value to agree with the concentration listed on the cylinder s label not zero 15 To ad
95. E 2 upside down on a flat surface or hold it upside down Unscrew the three case screws until they disengage from the top case They are captive screws so they will not fall off of the bottom case Turn the EAGLE 2 right side up and carefully lift the top case away from the bottom case Be careful not to lift it so far that it pulls on the main PCB with the cable that connects the top case to the main PCB Lay the top case down next to the bottom case to allow access to the flow system Locate the infrared methane sensor It has a twelve wire cable with a connector that mates to an infrared sub PCB that is installed on the main PCB and is normally located next to the pump Figure 40 on page 239 shows an infrared methane sensor in a typical location Unscrew and remove the two screws that hold down the infrared methane sensor bracket Grasp the sensor firmly and pull it out of the infrared methane flow chamber Rock it back and forth gently if necessary to pull it out Take care not to pull the cable from the sub PCB Rotate the locking lever counterclockwise on the cable connector that mates to the infrared methane sensor to unlock it Unplug the old infrared methane sensor from the cable Connect the new infrared methane sensor to the sensor cable and rotate the locking lever clockwise to lock the connector Insert the sensor into the infrared methane flow chamber and push it in until it bottoms out Line up the holes in the infr
96. EAGLE 2 in Bar Hole Mode and Appendix L Using the EAGLE 2 in Leak Check Mode for discussions of Bar Hole Mode and Leak Check Mode respectively 1 From the main menu place the cursor next to LC BH MODE SELECT 124 Using Setup Mode EAGLE 2 Operator s Manual 2 Press and release POWER ENTER RESET The Leak Check Bar Hole Mode Screen appears LEAK CHECK BAR HOLE MODE UP DOWN THEN ENTER OFF 3 Use AIR A YES or RANGE VY SHIFT to display the desired setting 4 Press and release POWER ENTER RESET to save the setting and return to the main menu Setting the Bar Hole Measurement Time This setting indicates the length of time the unit will sample when a bar hole measurement is initiated in Bar Hole Mode It can be set to 30 factory setting 45 or 60 seconds 1 From the main menu place the cursor next tt BH MEASURING TIME 2 Press and release POWER ENTER RESET The Bar Hole Measuring Time Screen appears BAR HOLE MEASURING TIME UP DOWN THEN ENTER 30 SECONDS 3 Use AIR A YES or RANGE YW SHIFT to display the desired setting 4 Press and release POWER ENTER RESET to save the setting and return to the main menu Zero Follower Settings The ZERO FOLLOWER setting is not intended for field adjustment The default setting for most target gases is ON The default setting for carbon dioxide channels and some configurations of non standard toxic gas channels is OFF The oxygen channel does n
97. ENTER TO CONTINUE The buzzer and alarm LED arrays activate in a double pulsing pattern Press and release the POWER ENTER RESET button to reset the alarm and return to the Calibration Mode Screen Attempt to calibrate again If the failure continues investigate the cause See Troubleshooting on page 70 20 Disconnect the tubing from the probe 21 Unscrew the demand flow regulator from the calibration cylinder 22 Use the RANGE VY SHIFT button to place the cursor next to the NORMAL OPERATION menu option then press and release the POWER ENTER RESET button to return to Measuring Mode EAGLE 2 Operator s Manual Calibrating Using the Auto Calibration Method 63 Calibrating Using the Single Calibration Method Single Calibration allows you to calibrate one channel at a time This is useful if you only want to calibrate one or two channels Setting the Fresh Air Reading 1 While in the Calibration Mode Screen move the cursor to the PERFORM AIR ADJUST menu item by using the RANGE W SHIFT button CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION gt PERFORM AIR ADJUST NORMAL OPERATION 2 Press and release the POWER ENTER RESET button The following screen appears PERFORM AIR ADJUST 3 Press and release the AIR A YES button to continue If you do not want to continue press the DISPLAY ADJUST NO button and the unit will return to the Calibration Mode Screen 64 Calibrati
98. ER HIGH RANGE 11 Use AIR A YES and RANGE YW SHIFT to set the correct range low or high for the sensor that is installed Press and release POWER ENTER RESET A screen appears with gas configuration choices for the PID sensor The screen below illustrates the first four choices for a high range PID PID O 2000ppm IBL gt 1000 ppm ACT 1000 ppm BNZ 1500 ppm DSL 15000 ppm ETA Y There are 17 choices that can be viewed on five screens Table 16 below lists all of the gas configuration choices 174 Configuring the PID Gas in Setup Mode EAGLE 2 Operator s Manual Table 16 PID Gas Configurations Gas Name LCD Abbreviation iE Raak poe Acetone ACT 30 00 ppm 1000 ppm Benzene BNZ 25 00 ppm 1000 ppm Diesel Fuel No 1 DSL 40 00 ppm 1500 ppm Ethanol ETA 400 0 ppm 15000 ppm Gasoline GSL 50 00 ppm 2000 ppm Isobutylene IBL 50 00 ppm 2000 ppm Standard Factory Gas Configuration Isopropanol IPA 200 0 ppm 5000 ppm JP 5 Fuel JP5 30 00 ppm 1000 ppm Methyl Ethyl Ketone MEK 40 00 ppm 1500 ppm Toluene TOL 25 00 ppm 1000 ppm N Hexane HEX 200 0 ppm 5000 ppm Propylene PRL 50 00 ppm 2500 ppm Styrene STY 20 00 ppm 500 ppm Tetrachloroethylene PCE 30 00 ppm 1000 ppm Trichloroethylene TCE 30 00 ppm 1000 ppm Vinyl Chloride VCM 100 0 ppm 4000 ppm PID PID 50 0 ppm 2000 ppm User Defined Gas Loaded with Maintenance Program 12 Use AIR A YES and RANGE YW SHIFT to place the cur
99. ER ENTER RESET to enter the number and move on to the next one EAGLE 2 Operator s Manual Infrared Hydrocarbon Calibration 255 5 The Calibration Mode Screen displays with the cursor next to AUTO CALIBRATION CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 6 Move the cursor to the PERFORM AIR ADJUST menu item by using the RANGE W SHIFT button CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION gt PERFORM AIR ADJUST NORMAL OPERATION 7 Press and release the POWER ENTER RESET button The following screen appears PERFORM AIR ADJUST 8 Press and release the AIR A YES button to continue If you do not want to continue press the DISPLAY ADJUST NO button and the unit will return to the Calibration Mode Screen 256 Infrared Hydrocarbon Calibration EAGLE 2 Operator s Manual 9 The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds then indicate that the operation is complete before returning to the Calibration Mode Screen ADJUSTING ZERO RELEASE AIR BUTTON ADJUSTING ZERO COMPLETE CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION gt PERFORM AIR ADJUST NORMAL OPERATION 10 Install the demand flow regulator onto the multi gas calibration cylinder 11 Connect the sample tubing to the demand flow regulator 12 Install the probe on the EAGL
100. ESET button to return to Measuring Mode Calibrating Using the Single Calibration Method Single Calibration allows you to calibrate one channel at a time This is useful if you only want to calibrate one or two channels Setting the Fresh Air Reading 1 While in the Calibration Mode Screen move the cursor to the PERFORM AIR ADJUST menu item by using the RANGE W SHIFT button CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION gt PERFORM AIR ADJUST NORMAL OPERATION 94 Calibrating Using the Single Calibration Method EAGLE 2 Operator s Manual 2 Press and release the POWER ENTER RESET button The following screen appears PERFORM AIR ADJUST 3 Press and release the AIR A YES button to continue If you do not want to continue press the DISPLAY ADJUST NO button and the unit will return to the Calibration Mode Screen 4 The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds then indicate that the operation is complete before returning to the Calibration Mode Screen ADJUSTING ZERO RELEASE AIR BUTTON ADJUSTING ZERO COMPLETE CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION gt PERFORM AIR ADJUST NORMAL OPERATION EAGLE 2 Operator s Manual Calibrating Using the Single Calibration Method 95 Performing a Span Adjustment in Single Calibration L Slide the tubing clamp onto the tubing and connect the tu
101. Gas Values Screen If you do want to continue with the calibration proceed to the next step 26 Connect the tubing from the demand flow regulator to the rigid tube on the probe Allow the EAGLE 2 to draw gas for one minute 27 Press and release the POWER ENTER RESET button to set the span adjustment for each channel to the programmed values 156 PID Calibration EAGLE 2 Operator s Manual 28 If all channels passed calibration the following screen sequence occurs AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END Y CAL GAS VALUE IBL 100 ppm ENTER TO BEGIN CAL If any of the sensors cannot be adjusted to the proper value a screen displays that indicates a calibration failure and lists the sensor s that failed to calibrate In the example below the oxygen and H S channels failed calibration The other sensors calibrated normally FAIL SENSOR lt gt lt OXY gt lt H2S gt lt gt ENTER TO CONTINUE The buzzer and alarm LEDs activate in a double pulsing pattern Press and release the POWER ENTER RESET button to reset the alarm and continue to the Calibration Value Screen for the PID channel After calibrating the PID channel by following the instructions below attempt to calibrate the standard channels again If the failure continues investigate the cause See Troubleshooting on page 70 EAGLE 2 Operator s Manual PID Calibrati
102. HING EAGLE 2 Operator s Manual Measuring Mode Alarms 37 are the factory settings When the EAGLE 2 goes into gas alarm press and release any button to silence the buzzer If the gas concentration was still above the alarm level when the button was pressed the LED arrays continue to flash and the EAGLE 2 continues to display the current alarm level The gas reading must fall below or rise above for an oxygen low alarm an alarm setting before you can reset the alarm When the alarm condition passes press and release the POWER ENTER RESET button to reset the alarm The LED arrays turn off and the EAGLE 2 alarm indications on the display turn off ALARM SILENCE Off and Alarms Set to LATCHING The gas reading must fall below or rise above for an oxygen low alarm an alarm setting before you can reset the alarm When the alarm condition passes press and release the POWER ENTER RESET button to reset the alarm The LED arrays and buzzer turn off and the EAGLE 2 alarm indications on the display turn off ALARM SILENCE On and Alarms Set to SELF RESETTING When the EAGLE 2 goes into gas alarm press and release any button to silence the buzzer The POWER ENTER RESET button will not reset the alarm When the gas reading falls below or rises above for an oxygen low alarm an alarm setpoint the alarm will automatically reset The LED arrays turn off and the EAGLE 2 alarm indications on the display turn off With ALARM SILENCE Off and Ala
103. LE 2 Operator s Manual Display Mode 41 Battery Voltage Screen The Battery Voltage Screen displays the minimum operating voltage and the current battery voltage Fully charged alkaline batteries typically indicate 6 0 volts fully charged Ni MH batteries typically indicate 5 2 volts This screen also displays during the startup sequence BATTERY MIN 4 3 VOLTS BATTERY NOW 5 2 VOLTS Gas Display Screen The Gas Display Screen gives you the option to select how the active channels are displayed SELECT GASES DISPLAYED You can display all of them on the screen at the same time one at a time with automatic scrolling or one at time with manual scrolling The factory setting displays all of the active channels at the same time To select how to display the active channels do the following 1 With the Gas Display Screen displayed press and release the AIR A YES button The following screen will appear with the cursor blinking SELECT gt DISPLAY ALL SCROLL AUTO SCROLL MANUAL 2 Use the AIR A YES and RANGE vw SHIFT buttons to place the cursor next to the desired choice 3 Press and release the POWER ENTER RESET button The unit continues to the STEL Screen and the display configuration will reflect your choice when you return to Measuring Mode 42 Display Mode EAGLE 2 Operator s Manual If you do not want to change the setting press and release the DISPLAY ADJUST NO butto
104. LEL OXY 20 9 vol H2S 0 0 ppm CO 0 m ENTER WHEN DONE If you do not want to proceed with the calibration press and release the DISPLAY ADJUST NO button to return to the Cal Gas Values Screen If you do want to continue with the calibration proceed to the next step 17 Connect the tubing from the demand flow regulator to the rigid tube on the probe Allow the EAGLE 2 to draw gas for one minute NOTE If you are calibrating the catalytic combustible channel with a gas concentration of 1000 ppm or lower you must use a 24 inch humidifier tube to connect the demand flow regulator to the rigid tube on the probe 18 Press and release the POWER ENTER RESET button to set the span adjustment for each channel to the programmed values 62 Calibrating Using the Auto Calibration Method EAGLE 2 Operator s Manual 19 If all channels passed calibration the following screen sequence occurs AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END Y CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION If any of the sensors cannot be adjusted to the proper value a screen displays that indicates a calibration failure and lists the sensor s that failed to calibrate In the example below the oxygen and H2S channels failed calibration The other sensors calibrated normally FAIL SENSOR lt gt lt OXY gt lt H2S gt lt gt
105. LIBRATE then the following screen displays and the buzzer sounds in a double pulsing pattern CALIBRATION DATE IS PAST DUE ENTER TO PERFORM CALIBRATION The EAGLE 2 cannot be used until a successful calibration has been performed Press and release the ENTER button to enter Calibration Mode See Chapter 4 Calibration Mode on page 55 for instructions to calibrate the EAGLE 2 When you are done with the calibration and exit Calibration Mode the unit will begin the startup sequence If the calibration was successful the screen above will not appear again until the unit is due for calibration If the calibration was not successful the screen above will again appear in the startup sequence e If the unit is due for calibration and CAL PAST DUE ACT is set to EAGLE 2 Operator s Manual Start Up Bar Hole Mode 273 NOTIFICATION ONLY then the following alert screen displays and the buzzer sounds in a double pulsing pattern CALIBRATION DATE IS PAST DUE Press and release the POWER ENTER RESET button to acknowledge the alert and continue with the startup sequence 12 The Date Time Screen appears for a few seconds 9 12 2008 15 00 00 13 If USER STATION ID is turned on see Turning the User Station ID Function On or Off on page 117 the ID Screen appears for a few seconds USER ID MIKE STATION ID PUMP 1 SERIAL NUMBER E2A515 If USER STATION ID is turned off
106. Make sure that the float probe assembly is hanging vertically and that the float device is located at the bottom of the probe EAGLE 2 Operator s Manual Start Up and Operation 289 4 The Battery Voltage Screen appears for a few seconds BATTERY MIN 4 3 VOLTS BATTERY NOW 5 2 VOLTS 5 The Inert Mode Selection Screen will then be displayed MODE SELECT UP DOWN THEN ENTER NORMAL MODE Use the AIR A YES or RANGE vw SHIFT button to display the mode you wish to enter To use the EAGLE 2 for tank tester applications ensure that the float probe is connected and select Inert Mode To use the EAGLE 2 in Normal Mode select Normal Mode and refer to Measuring Mode Normal Operation on page 29 If you have not made a selection after 20 seconds the instrument will begin to beep and the LEDs will begin to flash They will continue until you make a selection WARNING The EAGLE 2 is not a gas monitoring device until you select a mode and enter it 6 The warm up sequence will continue as described in Start Up on page 22 The oxygen alarms in the alarm point screens will be the alarm points for Inert Mode and not for Normal Mode 7 When the warm up is complete and the instrument is operating in Inert Mode the screen will appear as shown below if the instrument is sampling a fresh air environment The screen is shown with LEL and oxygen channels Your unit may only have LEL I CH4 O LEL N OXY
107. NG Do not plug the charger cable into a battery pack that contains alkaline batteries Do not attempt to charge alkaline batteries The charging module can either be used with an AC adapter or a vehicle plug DC adapter Both adapters plug into the module which then plugs into the EAGLE 2 L Plug the power adapter into either an AC outlet or into a vehicle outlet depending on which charger is being used Make sure the switch on the module is set to BATT CHARGER Make sure the EAGLE 2 is off Make sure the adapter and module are connected Insert the module s round plug into the EAGLE 2 s charging jack as shown in Figure 18 below NOTE The battery pack does not need to be attached to the EAGLE 2 case in order to charge It may be charged separately This allows a spare battery pack to be charged while the EAGLE 2 is in use Figure 18 Connecting the EAGLE 2 to the Charger 74 Replacing or Recharging the Batteries EAGLE 2 Operator s Manual 6 While the batteries are charging the green indicator LED will be off and the amber one will be on 7 The charging module has an internal timeout feature set at 9 5 hours A full charge should be reached in less than 9 5 hours When a full charge has been reached both the green and yellow LEDs will be on 8 Ifcharging should fail the green indicator LED will be off and the amber one will be blinking Table 10 summarizes the battery charger conditions Table 10
108. NTER RESET button When you hear a beep release the buttons EAGLE 2 Operator s Manual Description 293 3 The LCD will show the following screen for a few seconds with the S in the lower right corner indicating the unit is entering Setup Mode GAS MONITOR INSTRUMENTS EAGLE2 S 4 The S will then disappear and the following screen will appear for a few seconds GAS MONITOR INSTRUMENTS EAGLE 2 5 Ifthe unit prompts you for the password enter it by using the AIR A YES or RANGE YW SHIFT buttons to select each password number and then pressing and releasing the POWER ENTER RESET button to enter it and move on to the next number until all of the numbers are entered The main menu displays It displays six menu items at a time gt SET DATE amp TIME SET DATE FORMAT SET BATTERY TYPE CONFIGURE CHANNELS CONFIGURE GASES CATALYTIC UNITS y 6 Use the AIR A YES or RANGE VW SHIFT button to move the cursor to the ALARM POINTS menu item 7 Press and release POWER ENTER RESET The Change Alarm Point Settings Screen appears and all detection channels are displayed CHANGE ALARM POINT SETTINGS gt 1 CH4 2 OXY 3 H2S 4 CO END 294 Alarms EAGLE 2 Operator s Manual 8 Move the cursor next to the oxygen channel Press and release POWER ENTER RESET 9 A mode select screen will appear and prompt you to choose between Normal Mode and Inert Mode Use the AIR
109. NTER RESET to return to the Configure Gases screen or continue from Step 4 on page 108 to select a new gas 16 Use RANGE YW SHIFT to place the cursor next to END and press POWER ENTER RESET to return to the main menu 17 Be sure to perform a calibration on the catalytic combustible channel using an appropriate calibration cylinder for the target gas Setting the Catalytic Detection Units This menu item allows you to display the combustible gas units on the catalytic sensor channel as ppm LEL vol or selectable between the three units the CHANGE OK option The factory setting is CHANGE OK 1 From the main menu place the cursor next to CATALYTIC UNITS 2 Press and release POWER ENTER RESET The Catalytic Units screen appears with the current setting flashing at the bottom of the screen CATALYTIC UNITS UP DOWN THEN ENTER CHANGE OK 3 Use AIR A YES or RANGE VY SHIFT to scroll through the choices CHANGE OK vol ONLY LEL ONLY and ppm ONLY 4 When the desired setting is on the screen press and release POWER ENTER RESET to save the setting and return to the main menu EAGLE 2 Operator s Manual Using Setup Mode 113 Updating the Catalytic Sensor Relative Response Setting This menu item allows you to turn the catalytic sensor relative response feature on and off The catalytic sensor relative response feature enables you to change the catalytic sensor s response to gas on the fly so that the catalytic
110. O button while in a screen where you are entering or updating parameters to exit the screen without saving any changes Using Setup Mode WARNING The EAGLE 2 is not in operation as a gas detector while in Setup Mode 1 Take the EAGLE 2 to a non hazardous location and turn it off if it is on 2 Press and hold the AIR A YES and RANGE VY SHIFT buttons then press and hold the POWER ENTER RESET button When you hear a beep release the buttons 3 The LCD will show the following screen for a few seconds with the S in the lower right corner indicating the unit is entering Setup Mode GAS MONITOR INSTRUMENTS EAGLE2 S 4 The S will then disappear and the following screen will appear for a few seconds GAS MONITOR RK INSTRUMENTS EAGLE 2 EAGLE 2 Operator s Manual Using Setup Mode 103 5 Ifthe unit prompts you for the password enter it by using the AIR A YES and RANGE VY SHIFT buttons to select each password number and then pressing and releasing the POWER ENTER RESET button to enter it and move on to the next number until all of the numbers are entered The main menu displays It displays six menu items at a time gt SET DATE amp TIME SET DATE FORMAT SET BATTERY TYPE CONFIGURE CHANNELS CONFIGURE GASES CATALYTIC UNITS Y 6 Use the AIR A YES or RANGE YW SHIFT button to move the cursor up and down the menu items and to view additional menu items A down arrow in the lower ri
111. ON be p Oxygen Sensor E LEL Sensor and Bracket H2S Sensor Sensor Bracket Figure 37 Typical TC Sensor Location TC Sensor The TC sensor s appearance is exactly the same as that of the LEL sensor Its housing includes a sintered metal flame arrestor on one end that allows gas to diffuse into the sensor On the other end five pins extend from the sensor The TC sensor can be distinguished from the LEL sensor by the part number imprinted on it The TC sensor part number is TE 7568 while the LEL part number is NC 6260B The sensor cable connects to pins on one end and terminates in a four position connector on the other end which plugs into the TC sub PCB see description below The sensor bracket is installed over the TC sensor to keep it seated in place NOTE The LEL and TC sensors and sensor cables are identical in appearance Take care not to plug the LEL sensor cable into the port on the TC sub PCB and not to plug the TC sensor cable into the LEL port on the main PCB TC Sub PCB The TC sub PCB is a circuit board that is installed on the main PCB in one of the 3 sub PCB positions when a TC sensor is used with the EAGLE 2 The EAGLE 2 Operator s Manual Description 205 TC sensor cable connects to the sub PCB The sub PCB plugs into the main PCB and is held in place with a screw flat washer lock washer There are no user serviceable parts on the TC sub PCB Start Up and Normal Operation For instructions to start
112. ORE NEW VALUE S IN MEMORY FOR FUTURE CALIBRATIONS PRESS YES OR NO 23 If you select YES by pressing and releasing the AIR A YES button the changes that you made will be saved in the EAGLE 2 s memory as the new auto calibration gas values If you select NO by pressing and releasing the DISPLAY ADJUST NO button the changes you made will be used for any calibrations performed during the current operating session only The EAGLE 2 will delete the 186 ESM 01 Calibration EAGLE 2 Operator s Manual changes when the unit is turned off and will load the previous set of auto calibration values when it is turned on again 24 When you make your selection and press the desired button the unit returns to the Calibration Gas Values Screen CAL GAS VALUES CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm ENTER TO BEGIN CAL 25 Press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen with CAL IN PROCESS flashing CAL IN PROCESS CH4 0 LEL OXY 20 9 vol H2S 0 0 ppm CO 0 m ENTER WHEN DONE If you do not want to proceed with the calibration press and release the DISPLAY ADJUST NO button to return to the Cal Gas Values Screen If you do want to continue with the calibration proceed to the next step 26 Connect the tubing from the demand flow regulator to the rigid tube on the probe Allow the EAGLE 2 to draw gas for one minute 27 Press and release th
113. PID flow chamber Rock it back and forth gently if necessary to pull it out Take care not to pull the cable connector from the PID sub PCB 8 The PID sensor protrudes from one end of the sensor adapter Grasp the PID sensor firmly and pull it out of the sensor adapter 9 Place the PID sensor face down on a flat clean working surface 168 Maintenance EAGLE 2 Operator s Manual NOTE At this point it is recommended that the finger cots be used on the fingers handling the lamp Finger cots are included with the lamp cleaning kit 10 Hold the PID sensor steady on the working surface with one hand and using the other hand locate the tabs on the electrode stack removal tool and insert them into the slots on the side of the PID sensor near the face Figure 33 Using Removal Tool 11 Squeeze the removal tool to push the tabs into the sensor slots until the electrode stack and lamp are released 12 Carefully lift the PID sensor body away from the electrode stack and lamp If the lamp remains lodged in the sensor body carefully remove it with tweezers 13 If the spring in the lamp cavity comes out place it back into the lamp cavity 14 Discard the old electrode stack 15 Hold the new electrode stack between the thumb and forefinger of one hand and place the window end of the lamp inside the O ring seal in the new electrode stack with the other hand as shown below Take care not to touch the lamp window EAGLE 2 Operator
114. PM Factor Acetone 1 32 Benzene 1 10 Target Gas LEL Factor PPM Factor Isobutane 0 97 Isopropanol 1 55 Methane No No response response Butyl Acrylate 1 52 Butyl Acetate 1 58 2 Butyl Alcohol 1 30 EAGLE 2 Operator s Manual Monitoring in Methane Elimination Mode 267 Target Gas 1 Butyl Alcohol Cyclohexane Cumene Ethylene Dichloride Ethyl Alcohol Ethyl Chloride Ethyl Acrylate Hexane Hydrogen PPM Factor LEL Factor 1 74 2 18 1 25 Methyl Ethyl Keytone Keytone 2 04 11 50 1 71 2 04 1 68 Nonane Pentane Propane 1 00 1 75 Styrene Toluene Vinyl Acetate Monomer Target Gas Methyl Acrylate Methyl Isobutyl Mixed Xylenes PPM Factor LEL Factor 1 30 2 11 1 32 1 42 1 36 Low Low response response 0 92 2 17 1 36 Vapor pressure too low for significant LEL reading WARNING The EAGLE 2 s alarms are initiated by the display reading not the factored reading If you are monitoring for isobutane as in the above example and the low alarm is set for 10 LEL the EAGLE 2 will initiate a low alarm at 5 9 LEL isobutane display reading of 10 LEL 12 To return to full response mode return to the Methane Elimination Mode screen in Display Mode and press and release the AIR A YES or RANGE v SHIFT button to toggle the setting to DISABLED 13 Press and release the POWER ENTER RESET button The unit will save the setting and proceed to the
115. RM 5000 ppm HI ALRM 25000 ppm EAGLE 2 Operator s Manual RANGE vy SHIFT CHANNEL 1 CH4 0 100 LEL LO ALRM OFF vol lt p HI ALRM OFF vol Display Mode 47 4 Press and release the AIR A YES button to scroll through screens that display the rest of the active channels alarm settings CHANNEL 1 CHANNEL 2 CH4 0 100 LEL a OXY 0 40 0 vol LO ALRM 10 LEL LO ALRM 19 5 vol HI ALRM 50 LEL HI ALRM 23 5 vol LO ALRM 5000 ppm HI ALRM 25000 ppm Y AIR AIR E E CHANNEL 4 CHANNEL 3 CO 0 500 ppm H2S 0 100 0 ppm LO ALRM 25 ppm LO ALRM 10 0 ppm HI ALRM 50 ppm lt HIALRM 30 0 ppm STEL ALRM 200 ppm E STEL ALRM 15 0 ppm TWA ALRM 25 ppm YES TWA ALRM 10 0 ppm 48 Display Mode Select User ID Screen This screen displays only if USER STATION ID in the Setup Mode menu is set to ON see Turning the User Station ID Function On or Off on page 117 Use this screen to select a user ID from the user ID list in the EAGLE 2 s memory The current user ID is displayed A user ID can be up to 16 characters long The EAGLE 2 can store up to 32 user IDs SELECT USER ID CURRENT USER ID JOHN The user ID provides a way to identify the EAGLE 2 user during a data logging session If the user ID is changed during an operating session a new data session is initiated with the new user ID attached to it This allows you to ch
116. SELECT SENSOR CH4 CALIBRATION TO CALIBRATE DONE ESCAPE 35 98 LEL gt CH4 OXY REMOVE GAS H2S CO In the example above the EAGLE 2 could have adjusted the reading as low as 35 LEL and as high as 98 LEL If the span adjustment is not successful a screen displays that indicates a calibration failure FAILED SENSOR 11 43 LEL lt CH4 gt ENTER TO CONTINUE In the example above the EAGLE 2 could have adjusted the reading as low as 11 LEL and as high as 43 LEL Since the calibration gas value entered was 50 LEL the unit failed the span adjustment The buzzer and alarm LED arrays activate in a double pulsing pattern Press and release the POWER ENTER RESET button to reset the alarm and return to the Select Sensor Screen Attempt to calibrate again If the failure continues investigate the cause See Troubleshooting on page 70 NOTE The adjustment range will only appear in the calibration results screen if the Span Factor item in Setup Mode is set to ON If the Span Factor is set to OFF the adjustment range will not appear 13 Disconnect the tubing from the EAGLE 2 s probe 14 Repeat Step 6 through Step 13 for any other channels you want to calibrate Make sure you use an appropriate calibration cylinder for each sensor 68 Calibrating Using the Single Calibration Method EAGLE 2 Operator s Manual CAUTION When calibrating the oxygen sensor verify the concentration of oxyge
117. T is set in the Setup Mode Menu see Updating the Calibration Past Due Action Setting on page 123 e If the unit is due for calibration and CAL PAST DUE ACT is set to CONFIRM TO CAL then the following screen displays and the buzzer sounds in a double pulsing pattern CALIBRATION DATE IS PAST DUE PERFORM CALIBRATION To perform a calibration press and release the AIR A YES button The EAGLE 2 will enter Calibration Mode and the LCD will show the Calibration Mode main menu See Chapter 4 Calibration Mode on page 55 for instructions to calibrate the EAGLE 2 When you are EAGLE 2 Operator s Manual Start Up Leak Check Mode 281 done with the calibration and exit Calibration Mode the unit will begin the startup sequence If the calibration was successful the screen above will not appear again until the unit is due for calibration If the calibration was not successful the screen above will again appear in the startup sequence To continue without performing a calibration press and release the DISPLAY ADJUST NO button e Ifthe unit is due for calibration and CAL PAST DUE ACT is set to MUST CALIBRATE then the following screen displays and the buzzer sounds in a double pulsing pattern CALIBRATION DATE IS PAST DUE ENTER TO PERFORM CALIBRATION The EAGLE 2 cannot be used until a successful calibration has been performed Press and release the ENTER button to enter Calibration Mode See
118. THEN ENTER 2 40 The sensor voltage setting defines whether the catalytic sensor voltage is set for full response 2 40 volts or methane elimination 1 30 volts If the sensor voltage is set to 2 40 volts the unit will default to Full Response Mode when turned on but the methane elimination feature can be turned on in the Methane Elimination Mode Screen in Display Mode If the sensor voltage is set to 1 3 volts the unit will default to Methane Elimination Mode when turned on but the methane elimination feature can be turned off in the Methane Elimination Mode Screen in Display Mode 13 Use AIR A YES or RANGE YW SHIFT to display the desired catalytic sensor voltage 1 30 volts or 2 40 volts 14 Press and release POWER ENTER RESET The confirmation screen appears In the example below the user defined gas has been selected and defined as propane with the gas name set to PRO CHANGE TO PRO PRESS YES OR NO 112 Using Setup Mode EAGLE 2 Operator s Manual 15 If you want to accept the gas configuration change press and release AIR A YES The unit will return to the Configure Gases screen CONFIGURE GASES gt CAT PRO CAT OPI OP2 OP3 Y If you do not want to accept the gas configuration change press and release DISPLAY ADJUST NO to return to the screen with the gas choices shown in Step 4 on page 108 You can either scroll down to END and press POWER E
119. WER ENTER RESET button The following screen appears DO YOU WANT TO STORE NEW VALUE IN MEMORY FOR FUTURE CALIBRATIONS PRESS YES OR NO 23 If you select YES by pressing and releasing the AIR A YES button the changes that you made will be saved in the EAGLE 2 s memory as the new auto calibration gas values If you select NO by pressing and releasing the DISPLAY ADJUST NO button the changes you made will be used for any calibrations performed during the current operating session only The EAGLE 2 will delete the changes when the unit is turned off and will load the previous set of auto calibration values when it is turned on again 24 When you make your selection and press the desired button the unit returns to the Calibration Gas Values Screen CAL GAS VALUES CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm ENTER TO BEGIN CAL 25 Press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen with CAL IN PROCESS flashing CAL IN PROCESS CH4 0 LEL OXY 20 9 vol H2S 0 0 ppm CO 0 ppm ENTER WHEN DONE If you do not want to proceed with the calibration press and release the DISPLAY ADJUST NO button to return to the Cal Gas Values Screen If you do want to continue with the calibration proceed to the next step 26 Connect the tubing from the demand flow regulator to the rigid tube on the probe Allow the EAGLE 2 to draw gas for one minute 21
120. a sample hose or with the probe installed directly to the inlet fitting Determine which configuration works best for your application 1 Connect the sample hose or probe to the EAGLE 2 s quick connect inlet fitting 2 If using a sample hose connect the probe to the sample hose s quick connect fitting 3 Press and briefly hold down the POWER ENTER RESET button Release the button when you hear a beep 22 Overview EAGLE 2 Operator s Manual 4 The LCD will show the following screen for about ten seconds GAS MONITOR INSTRUMENTS EAGLE 2 5 The Battery Voltage Screen appears for a few seconds BATTERY MIN 4 3 VOLTS BATTERY NOW 5 2 VOLTS 6 The Active Gases Screen appears for a few seconds indicating which channels are active and their target gas ACTIVE GASES CH4 OXY H2S CO 7 If LUNCH BREAK is turned on see Updating the Lunch Break Setting on page 129 the Resume Measurement Screen appears The unit counts down from 5 seconds in the lower right corner of the LCD to the right of MEASUREMENTS LUNCH BREAK MODE ON RESUME PEAK AND TWA MEASUREMENTS 2 e To continue accumulating peak and time weighted average TWA readings from the last time the EAGLE 2 was used press and release the AIR A YES button before the countdown reaches 0 or allow the countdown to reach 0 If you do not press the AIR A YES button within the 5 second countdown the E
121. age 279 1 Connect the sample hose to the EAGLE 2 s quick connect inlet fitting 2 Connect the bar hole probe to the sample hose s quick connect fitting 3 Press and briefly hold down the POWER ENTER RESET button Release the button when you hear a beep 4 The LCD will show the following screen for about ten seconds GAS MONITOR INSTRUMENTS EAGLE 2 5 The Battery Voltage Screen appears for a few seconds BATTERY MIN 4 3 VOLTS BATTERY NOW 5 2 VOLTS 6 The Mode Select Screen displays gt NORMAL MODE BAR HOLE MODE LEAK CHECK MODE 7 The cursor will flash in front of NORMAL MODE Use the RANGE VY SHIFT button to move the cursor next to BAR HOLE MODE 8 With Bar Hole Mode selected press and release the POWER ENTER RESET button to begin the Bar Hole Mode startup sequence EAGLE 2 Operator s Manual Start Up Bar Hole Mode 271 NOTE If no button is pressed for 20 seconds the unit will proceed into whichever mode has the cursor in front of it 9 The Active Gases Screen appears for a few seconds indicating which channels are active and their target gas ACTIVE GASES CH4 OXY H2S CO CH4 10 The gas alarm setpoints are displayed by three screens in sequence the Low Alarm Screen High Alarm Screen and STEL TWA Alarm Screen Each screen remains on the LCD for three seconds S A CH4 10 LEL LL OXY 19 5 vol OA H2S 10 0 ppm WR CO 25 ppm M
122. air the display units should always switch to the previously selected units If you selected the methane channel to be displayed in vol then after 15 seconds the reading will remain in vol e When the purge is complete the screen will return to the initial Bar Hole Mode screen BAR HOLE MODE CH4 0 vol OXY 20 9 vol ENTER MEASURE ADJUST PURGE 7 If other bar holes will be tested proceed to the next bar hole and repeat Step 3 Step 6 8 To cancel a bar hole measurement or fresh air purge that is in progress press and release the DISPLAY ADJUST NO button 9 To exit Bar Hole Mode and return to the Mode Select Screen at any time press and hold the RANGE W SHIFT button for 5 seconds Turning Off the EAGLE 2 Bar Hole Mode 1 Press and hold the POWER ENTER button 2 The unit will initiate a bar hole measurement Keep holding the POWER ENTER button The buzzer will sound and the LCD back light will flash for about five seconds 3 Release the button when GOODBYE appears on the display When GOODBYE disappears and the backlight turns off the unit is off 278 Bar Hole Testing EAGLE 2 Operator s Manual Appendix L Using the EAGLE 2 in Leak Check Mode Overview This chapter explains how to operate the EAGLE 2 in Leak Check Mode Leak Check Mode is used to pinpoint small leaks of combustible gas from valves flanges connections and other potential leak points When the EAGLE 2 is in Leak Check Mode
123. aking sound made by the cotton swab moving over the window surface This usually occurs after about 15 seconds of polishing 18 Remove the residual powder from the lamp window with a clean cotton swab Take care not to touch the tip of the cotton swab that is used to clean the lamp as this may contaminate it with finger oil 19 Ensure the lamp is completely dry and any visible signs of contamination are removed before reinstalling EAGLE 2 Operator s Manual Maintenance 163 20 Hold the electrode stack between the thumb and forefinger of one hand and place the window end of the lamp inside the O ring seal in the electrode stack with the other hand as shown below Figure 30 Reinstalling the Electrode Lamp 21 Twisting the lamp slightly during insertion will help to ensure the lamp window is snug against the stack s front electrode The lamp should be supported by the O ring 22 Continuing to hold the electrode stack between your forefinger and thumb carefully insert the lamp into the lamp cavity in the sensor ensuring that the lamp remains in position 23 Press in the electrode stack firmly to ensure that the stack wing clips are engaged and the faces of the stack and sensor body are flush 24 Carefully line up the PID sensor s pins with the sockets in the bottom of the sensor adapter and gently lower the sensor into the adapter until you feel it contact the bottom 25 Do not attempt to push the sensor in farther once
124. al Appendix B Setup Mode Overview This appendix describes the EAGLE 2 in Setup Mode In Setup Mode you can set the date and time set the date format set the battery type configure the channels configure the gas for a catalytic or PID channel set the detection units for the catalytic channel turn the catalytic sensor relative response feature on or off set the alarm points change the alarm latching setting turn the alarm silence feature on or off turn the user station ID function on or off set the auto calibration values set the backlight delay time turn the automatic fresh air adjust feature on or off set the data logging interval time turn the data logger overwrite feature on or off turn the data log memory clear feature on or off adjust the display contrast turn the calibration reminder feature on or off set the calibration past due action set the calibration interval select the leak check bar hole mode operation setting set the bar hole measurement time turn the zero follower on or off for each channel set the zero suppression level for each channel except oxygen EAGLE 2 Operator s Manual Overview 101 turn the confirmation alert feature on or off turn the password feature on or off and set the password reset the instrument parameters to their default settings turn the lunch break function on or off turn the span factor on or off select the language return to normal operation The EAGLE 2
125. al of aluminum oxide powder e 40 cotton swabs e 10 finger cots Perform the following procedure to clean the PID lamp L 2 3 9 Verify that the EAGLE 2 is off Place the EAGLE 2 upside down on a flat surface or hold it upside down Unscrew the three case screws until they disengage from the top case They are captive screws so they will not fall off of the bottom case Turn the EAGLE 2 right side up and carefully lift the top case away from the bottom case Be careful not to lift it so far that it pulls on the main PCB with the cable that connects the top case to the main PCB Lay the top case down next to the bottom case to allow access to the flow system Locate the PID sensor adapter It has a five wire cable and is normally located next to the pump The cable has a connector that mates to a PID sub PCB that is installed on the main PCB Figure 25 on page 137 shows a PID sensor in a typical location Grasp the sensor adapter firmly and pull it out of the PID flow chamber Rock it back and forth gently if necessary to pull it out Take care not to pull the cable connector from the PID sub PCB The PID sensor protrudes from one end of the sensor adapter Grasp the PID sensor firmly and pull it out of the sensor adapter Place the PID sensor face down on a flat clean working surface NOTE Do not touch the lamp window with your fingers as this may contaminate the window with finger oil At this point it is recommended
126. alues on the screen press and hold the RANGE W SHIFT button then press the DISPLAY ADJUST NO button and release both The following screen appears with the cursor next to CH4 ADJUST AUTO CALIBRATION VALUES gt CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm y 16 Place the cursor next to the channel whose gas value you want to change using the AIR A YES and RANGE YW SHIFT buttons 17 Press and release the POWER ENTER RESET button to select the 146 PID Calibration EAGLE 2 Operator s Manual channel The calibration gas value begins to flash 18 Use the AIR A YES and RANGE YW SHIFT buttons to adjust the calibration gas setting to the desired value NOTE The calibration gas value cannot be set lower than the low alarm setting If the calibration gas value listed on the calibration cylinder is lower than the current low alarm setting enter Setup Mode and change the low alarm setting See Updating the Alarm Point Settings on page 114 for instructions If you need to change the alarm point setting only to perform a calibration make sure that you change the alarm point setting back to its original value once the calibration has been performed 19 Press and release the POWER ENTER RESET button to save the change The calibration gas value stops flashing 20 Repeat Step 15 through Step 19 for any other channels that need to be changed 21 When you are done adjusting the calibration gas values move the c
127. ample hose is not needed for monitoring a particular area the probe may be installed directly to the inlet fitting Sample hose lengths are available from 5 feet standard length to 125 feet see General Parts List on page 84 A teflon lined hose is provided with all units that contain a PID sensor This hose must be used when operating a PID EAGLE 2 see Appendix D PID Sensors on page 136 EAGLE 2 Operator s Manual Standard Accessories 15 CAUTION Sample hose lengths of more than 125 feet are not recommended for the EAGLE 2 because of flow rate reduction and increased response time Consult RKI Instruments Inc for hose lengths longer than 125 feet The probe includes a replaceable particle filter and hydrophobic filter disk that prevent particulates and water from entering the EAGLE 2 s flow system See Replacing the Hydrophobic Probe s Particle Filter and Hydrophobic Filter Disk on page 75 for instructions to replace the particle filter and hydrophobic filter disk NOTE When using the probe with a PID EAGLE 2 be sure that the particle filter is not installed fo Sample Hose c B j _7 Probe es aS SE Figure 6 Sample Hose and Hydrophobic Probe Optional A
128. ample to avoid false CO readings If false or elevated CO readings are noticed especially in the presence of H S change the charcoal filter The charcoal inside the filter cannot be replaced the entire filter must be replaced Exhaust Fitting The exhaust fitting is located below the inlet fitting It routes the gas sample out of the EAGLE 2 It includes a female 10 32 thread that can be used for the installation of a hose barb or other type of fitting that has a male 10 32 thread so that the exhaust can be routed to a particular location with flexible tubing if desired Standard Accessories Standard accessories include the shoulder strap the sample hose and the hydrophobic probe Shoulder Strap A comfortable elastic shoulder strap clips to the EAGLE 2 at the front and rear of the top case It clips to stainless steel rings that are installed in features on the top case It can be removed from the EAGLE 2 by opening the clip at each end of the strap and removing it from the strap ring at the front and rear of the top case Hose and Probe A 5 foot polyurethane sample hose and a 10 inch hydrophobic probe are included as standard The hose has a male quick connect fitting on one end and a female quick connect fitting on the other end The probe has a male quick connect fitting Normally the male end of sample hose is installed in the EAGLE 2 inlet fitting and the probe is installed in the female end of the hose However if the s
129. an 50 000 ppm the ppm unit reading will indicate 50 000 ppm and also indicate an overscale condition 34 Measuring Mode Normal Operation EAGLE 2 Operator s Manual The LEL and volume unit readings will still increase up to 100 LEL and 15 volume respectively which are equivalent to 150 000 ppm The catalytic combustible sensor is slightly affected by humidity This is not apparent when the EAGLE 2 is used for LEL or volume detection but because of the high sensitivity in the ppm range significant humidity changes can affect the ppm reading especially in the lower part of the range Take care to allow the unit to acclimate to a new environment for about a minute and perform a demand zero in a fresh air location when you move between areas of different humidity CAUTION If the catalytic combustible channel is being calibrated with a gas concentration of 1000 ppm or lower the a 24 inch humidifier must be used for accurate calibration See Chapter 4 Calibration Mode on page 55 for more explanation and instructions for calibration To monitor for combustible gas in the ppm range 1 Start up the EAGLE 2 as described in Start Up on page 22 2 Allow the EAGLE 2 to run for 3 5 minutes after the startup sequence is complete and it is in Measuring Mode This allows the catalytic combustible sensor to stabilize sufficiently for ppm monitoring NOTE This extra stabilization period is not necessary if monitoring in the
130. an still stress the catalytic LEL sensor The catalytic LEL sensor can be enabled or disabled in the Catalytic LEL Sensor screen in Display Mode The default setting is enabled To change the setting do the following NOTE The Catalytic LEL Sensor setting is reset when the EAGLE 2 is turned off When the EAGLE 2 is turned on this setting is always ENABLED 1 Use the DISPLAY ADJUST NO button to enter Display Mode and scroll to the Catalytic LEL Sensor screen The current setting will be flashing CAT LEL SENSOR ENABLED 2 Use the AIR A YES or RANGE VY SHIFT button to toggle to the desired setting 254 Start Up and Normal Operation EAGLE 2 Operator s Manual If set to DISABLED the gas reading for the catalytic LEL channel will be replaced by dashes 3 Press and release POWER ENTER RESET to save the setting and return to the main menu 4 Use the DISPLAY ADJUST NO button to scroll through the rest of Display Mode and enter Normal Operation Infrared Hydrocarbon Calibration An infrared hydrocarbon channel can be calibrated using the auto calibration method or the single calibration method To calibrate an infrared hydrocarbon channel using the single calibration method see Calibrating Using the Single Calibration Method on page 64 and follow the instructions for calibrating a single channel If your instrument is a multi channel instrument that includes an infrared hydrocarbon channel RKI
131. and 50 of the full scale value For example if you are calibrating the catalytic combustible gas channel your calibration cylinder should have a combustible gas concentration between 10 LEL and 50 LEL An oxygen free source such as 100 nitrogen is recommended for setting the oxygen zero NOTE The catalytic combustible channel can be set up for and calibrated to a number of different combustible gases See Configuring the Combustible Gas on page 108 for instructions Be sure that you are using an appropriate calibration cylinder for the target gas of the catalytic combustible channel CAUTION When using auto calibration with the standard 4 gas EAGLE 2 although the EAGLE 2 can be calibrated with an oxygen concentration of up to 19 5 RKI Instruments Inc recommends that the multi gas cylinder have an oxygen concentration in the range of 10 16 oxygen e A demand flow regulator to provide adequate sample gas flow e Non absorbent tubing e A 24 inch humidifier tube if you are calibrating the catalytic combustible channel with a gas concentration of 1000 ppm or lower WARNING If you are using a calibration kit that includes a gas bag and a fixed flow regulator or dispensing valve do not apply gas directly to the EAGLE 2 with the regulator or dispensing valve or damage to the pump will result See Appendix A Calibrating with a Sample Bag on page 86 for instructions to properly use a gas bag kit To calibrate the LEL
132. ange the user ID during operation and have each user ID that was used during an operating session saved for the corresponding data See the Eagle 2 Data Logger Management Program Operator s Manual for a detailed description of data logging and the user ID The user ID list cannot be edited using the EAGLE 2 user interface The Eagle 2 Maintenance Data Loader Program is required to define or change user IDs in the user ID list For a detailed description of editing the list of EAGLE 2 Operator s Manual user IDs stored in the EAGLE 2 see the Eagle 2 Maintenance Data Loader Program Operator s Manual To select a different user ID 1 With the Select User ID Screen displayed press and release the AIR A YES button A screen appears that includes the current user ID which is indicated by the cursor next to it Bruce gt John Mike The user IDs are displayed in groups of six The previous group of six is displayed when the cursor is moved up past the top of the LCD The next group of six is displayed when the cursor is moved down past the bottom of the LCD The list will not wrap around to the previous screen if the cursor is moved up from the first user ID or to the next screen if the cursor is moved down from the last user ID Any of the user IDs in the list that have not been changed from the factory setting will be shown as dashes 2 Use the AIR A YES and RANGE YW SHIFT buttons to move the cursor up and dow
133. annel configured for 0 100 vol CH4 To use the auto calibration method to calibrate a multi channel instrument that includes a TC channel do the following 1 See Calibration Supplies and Equipment on page 56 for a description of the necessary calibration supplies In addition to an appropriate multi gas cylinder that is used to calibrate any active standard channels you will also need a cylinder to calibrate the TC channel See Table 21 on page 222 for available cylinders Make sure your calibration cylinder is EAGLE 2 Operator s Manual TC Calibration 207 appropriate for the TC detection range If the EAGLE 2 is intended for use in a landfill RKI Instruments Inc recommends using the carbon dioxide methane mix calibration cylinder because it is representative of gases present in a landfill NOTE If your instrument has more than one TC sensor you will need a calibration cylinder for each sensor 2 Find a fresh air environment This is an environment free of toxic or combustible gases and of normal oxygen content 20 9 3 While in Measuring Mode press and hold the RANGE W SHIFT button then press the DISPLAY ADJUST NO button and release both buttons 4 Ifthe unit prompts you for the password enter it by using the AIR A YES and RANGE YW SHIFT buttons to select each password number and then pressing and releasing POWER ENTER RESET to enter the number and move on to the next one 5 The Calibration Mode Screen
134. anual for a detailed description of data logging and the station ID The station ID list cannot be edited using the EAGLE 2 user interface The Eagle 2 Maintenance Data Loader Program is required to define or change station IDs in the station ID list For a detailed description of editing the list of station IDs stored in the EAGLE 2 see the Eagle 2 Maintenance Data Loader Program Operator s Manual To select a different station ID 1 With the Select Station ID Screen displayed press and release the AIR A YES button A screen appears that includes the current station ID which is indicated by the cursor next to it Pump 2 gt Pump 3 Valve 5 The station IDs are displayed in groups of six The previous group of six is displayed when the cursor is moved up past the top of the LCD The next group of six is displayed when the cursor is moved down past the bottom of the LCD The list will not wrap around to the previous 50 Display Mode EAGLE 2 Operator s Manual screen if the cursor is moved up from the first station ID or to the next screen if the cursor is moved down from the last station ID Any of the station IDs in the list that have not been changed from the factory setting will be shown as dashes 2 Use the AIR A YES and RANGE YW SHIFT buttons to move the cursor up and down the screen and scroll through the available station IDs to find the desired station ID 3 When the desired station ID is display
135. aped particle filter into the probe body so that the wide part of the filter is facing up NOTE Do not install the particle filter into a probe intended to be used with a PID EAGLE 2 76 Replacing the Hydrophobic Probe s Particle Filter and Hydrophobic Filter 7 Place the new filter disk flat on top of the particle filter Make sure it is centered over the particle filter 8 Carefully screw the other half of the probe body onto the half with the particle filter and filter disk while keeping the probe oriented vertically to keep the disk centered 9 When you feel the O ring being compressed grasp both ends of the probe and tighten them together very firmly to ensure a seal 10 To test the seal do the following e install the probe on the EAGLE 2 e startup the EAGLE 2 e confirm that a low flow alarm occurs when you cover the end of the probe tube with your finger e if alow flow alarm does not occur hand tighten the probe further e ifalow flow alarm still does not occur when you cover the probe tube with your finger disassemble the probe inspect the placement of the O rings and filter disk reassemble the probe and re test it Replacing the Hydrophobic Filter Replace the hydrophobic filter inside the bottom case when it becomes dirty or clogged An unexplained low flow alarm may indicate that the hydrophobic filter is dirty and needs to be replaced 1 Verify that the EAGLE 2 is off 2 Place the EAGLE 2 upside d
136. ared methane sensor bracket with the two standoffs on the infrared methane chamber Install the two sensor bracket screws Confirm that the main PCB is seated in its slots and that its bottom edge is resting on the bottom of the bottom case If the main PCB is not seated properly it may be damaged when the top case is re installed 250 Maintenance EAGLE 2 Operator s Manual 16 Make sure that the top case gasket is fully seated in its groove and carefully put the top case back on the bottom case If you have any difficulty mating the top and bottom cases inspect the placement of the main PCB and the placement of the top case gasket 17 Turn the EAGLE 2 upside down and tighten the three case screws to secure the top case to the bottom case 18 Calibrate the infrared methane channel as described in Infrared Methane Calibration on page 241 Parts List Table 29 Infrared CH EAGLE 2 Parts List 47 5051RK Infrared CH sensor cable DEM 3313 3 Infrared CH4 sensor 0 100 LEL or 0 100 LEL 5 0 100 0 vol autoranging 81 0012RK 01 Calibration cylinder 50 LEL CH in air 34 liter steel 81 0012RK 03 Calibration cylinder 50 LEL CH in air 103 liter steel 81 0013RK 01 Calibration cylinder 50 vol CH in No 34 liter steel 81 0013RK 05 Calibration cylinder 50 vol CH in No 58 liter steel EAGLE 2 Operator s Manual Parts List 251 Appendix I Infrared Hydrocarbon Sensor Overview This appendix describes
137. arm indications 1 From the main menu place the cursor next to ALARM SILENCE 2 Press and release POWER ENTER RESET The Alarm Silence Option Screen appears ALARM SILENCE OPTION UP DOWN THEN ENTER ON 3 Use AIR A YES or RANGE YW SHIFT to display the desired setting 4 Press and release POWER ENTER RESET to save the setting and return to the main menu Turning the User Station ID Function On or Off With USER STATION ID set to ON the ID Screen displays during start up and the Select User ID Screen and Select Station ID Screen appear in Display Mode The ID s can be selected in Display Mode With USER STATION ID set to OFF factory setting the ID Screen does not display during start up and the Select User ID Screen and Select Station ID Screen do not appear in Display Mode 1 From the main menu place the cursor next to USER STATION ID EAGLE 2 Operator s Manual Using Setup Mode 117 2 Press and release POWER ENTER RESET The User and Station ID s Screen appears USER AND STATION ID S UP DOWN THEN ENTER OFF 3 Use AIR A YES or RANGE VY SHIFT to display the desired setting 4 Press and release POWER ENTER RESET to save the setting and return to the main menu Updating the Autocal Values The EAGLE 2 stores calibration gas value settings This allows you to calibrate all EAGLE 2 channels simultaneously with a calibration cylinder that contains all required target gases
138. ata to the data logger when the data logger memory is full 1 From the main menu place the cursor next to DATA LOG OVERWRITE 2 Press and release POWER ENTER RESET The Overwrite Log Data Screen appears OVERWRITE LOG DATA WHEN MEMORY IS FULL UP DOWN THEN ENTER ON 3 Use AIR A YES or RANGE YW SHIFT to display the desired setting 4 Press and release POWER ENTER RESET to save the setting and return to the main menu Updating the Data Log Memory Setting With DATA LOG MEMORY set to ON factory setting the Data Logging Screen in Display Mode asks whether you want to clear the logged data in addition to showing the log time remaining see Data Logging Screen on page 52 With DATA LOG MEMORY set to OFF the Data Logging Screen only shows the remaining log time and does not give you the opportunity to clear the logged data 1 From the main menu place the cursor next to DATA LOG MEMORY 2 Press and release POWER ENTER RESET The Prompt to Clear Data Log Memory Screen appears PROMPT TO CLEAR DATA LOG MEMORY UP DOWN THEN ENTER ON 3 Use AIR A YES or RANGE YW SHIFT to display the desired setting EAGLE 2 Operator s Manual Using Setup Mode 121 4 Press and release POWER ENTER RESET to save the setting and return to the main menu Updating the LCD Contrast Setting The contrast setting controls the LCD contrast Select the setting so the characters on the display are easy
139. ated by CAT or one of the sub PCBs indicated by OP1 option 1 OP2 option 2 and OP3 option 3 OP1 indicates that the SUB1 PCB supports the sensor OP2 indicates that the SUB2 PCB supports the sensor and OP3 indicates that the SUB3 PCB supports the sensor The OP1 OP2 and OP3 identifiers also indicate that the sensor is an optional sensor The catalytic sensor is a standard sensor Each of these is always listed 134 Channel Setup amp Sub PCBs EAGLE 2 Operator s Manual e The currently configured gas The catalytic sensor is always defined and configured even if it is not setup as active in CONFIGURE CHANNELS If any of the Sub PCBs are not installed then the corresponding sensor is undefined and the next two fields are filled with dashes In the example above the catalytic sensor is configured as methane CH3 the sensor supported by the SUB1 PCB is configured as isobutylene IBL the sensor supported by the SUB2 PCB is configured as ammonia NH3 and there is no sub PCB installed in position SUB3 e The sensor type The sensor type can be one of the following catalytic CAT photo ionization detector PID ESM 01 toxic ESM infrared IR or thermal conductivity TC Although any installed sensor type will be listed in CONFIGURE GASES the only sensor types whose gas can be configured are catalytic PID and TC sensor types If you select an ESM O1 toxic or IR type of sensor to configure in this menu item a
140. ation about viewing and changing the standard alarms used in Normal Mode see Updating the Alarm Point Settings on page 114 For information about viewing and changing Inert Mode oxygen alarms see Appendix N Using the EAGLE 2 in Inert Mode on page 293 Calibration Use a hexane calibrating source to calibrate the combustible gas LEL range Use a 100 nitrogen calibrating source to set the zero reading for the oxygen channel RKI Instruments Inc recommends using the Single Calibration method to calibrate the EAGLE 2 Tank Tester model See Chapter 4 Calibration Mode on page 55 The instructions in that section call for calibrating the EAGLE 2 with methane When following these instructions be sure to calibrate with hexane rather than methane NOTE Do not calibrate the EAGLE 2 Tank Tester model with the dilution fitting attached to the inlet fitting EAGLE 2 Operator s Manual Alarms 291 Parts List Table 35 lists part numbers for replacement parts and accessories of the EAGLE 2 s Tank Tester model Table 35 Parts List EAGLE 2 Tank Tester Model 80 0405RK Dilution fitting 1 1 80 0802RK Float probe 12 foot 292 Parts List EAGLE 2 Operator s Manual Appendix N Using the EAGLE 2 in Inert Mode Inert Mode is used to measure the combustible gas and or oxygen level in a purged environment The oxygen alarms in this mode are both increasing and are generally set at 5 0 and 10 0 Descripti
141. ause of the way that the main PCB circuitry is arranged if the CO sensor is plugged into the H2S socket and the H2S sensor is plugged into the CO socket the sensors will still operate properly and the CO and H2S readings will still appear on the channels that are programmed for those gases LCD A digital LCD liquid crystal display is visible through a clear plastic window in the top case The LCD simultaneously shows the gas reading for all installed sensors The LCD also shows information for each of the EAGLE 2 s operating modes EAGLE 2 Operator s Manual Instrument Description 9 Control Buttons Four control buttons are located below the LCD They are from left to right POWER ENTER RESET DISPLAY ADJUST NO AIR 4 YES and RANGE v SHIFT Table 3 EAGLE 2 Control Button Functions POWER ENTER e turns the EAGLE 2 on and off RESET silences and resets audible alarm if Alarm Latching is set to Latching and Alarm Silence is set to ON enters instructions values and settings into the EAGLE 2 s microprocessor DISPLAY ADJUST activates Display Mode NO silences and resets audible alarm if Alarm Latching is set to Latching and Alarm Silence is set to ON enters instructions into the EAGLE 2 s microprocessor AIR A YES activates the demand zero function adjusts the EAGLE 2 s fresh air reading silences and resets audible alarm if Alarm Latching is set to Latching and Alarm Silence is set to ON enters instructions i
142. ay aed eat 284 Locating OTIC oe oer Peels tad fe oO aa ee in se atthe heath 285 Turning the Buzzer On and Off In Leak Check Mode 285 Turning Off the EAGLE 2 Leak Check Mode 4 286 Appendix M EAGLE 2 Tank Tester Model 287 PESTIN is ae keds ans wos Ful OWES AAS ROARS RE Re T ales 287 Float Probe Assembly ox ote eS 2 AGRE ERE RAR OA EOL ER PERS 288 Dilution Fitting I cdg Stare hey EAS aac anil Dy bao oh ANs ead doen taba 288 EAGLE 2 Operator s Manual Table of Contents Sat UP oneee aati OR Vege rete TE ee ence en Meee tae sea 289 PLAINS sohbet hehe ena TS eR OEE ERE ER SOURS WER Os E 291 Calibration wii vedo Ms ies Aan ed baie eee GA eee oe Oe eee 291 Parts MAGC pete bits AT E E Witenes Gh were e a ens 292 Appendix N Using the EAGLE 2 in Inert Mode 293 Deseriptiom as cons sd beets es a set bts a nels E 2 ae 293 PMMA INS af iar eater wht E e a tend ates a wha We et a aS tet 293 SaN I sheet Bike beats re Bik ae Eat BS ein Cadre ND uke GG rd nae a en DE EY 296 Operationens ty OO rae geet re Batt hu sees te ere Rae 298 WARNING Understand manual before operating Substitution of components may impair intrinsic safety To prevent ignition of a hazardous atmosphere batteries must only be changed or charged in an area known to be nonhazardous Not tested in oxygen enriched atmospheres above 21 NOTE RKI Instruments Inc recommends that you refer to ISA RP12 13 Part II 1987 o
143. bing to the sample bag s inlet Leave the clamp unclamped for now Connect the other end of the tubing to the regulator s or dispensing valve s hose barb fitting Install the probe on the EAGLE 2 inlet fitting Make sure the probe is complete with internal O ring and membrane and that the two halves of the probe are tightened firmly together to avoid leaks that can affect the calibration See Figure 19 Replacing the Particle Filter and Hydrophobic Filter Disk on page 76 for an illustration of the internal parts of the probe Move the cursor next to the SINGLE CALIBRATION menu item by using the AIR A YES button CALIBRATION MODE AUTO CALIBRATION gt SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION Press and release the POWER ENTER RESET button The Select Sensor Screen appears with the cursor flashing SELECT SENSOR TO CALIBRATE gt ESCAPE CH4 OXY H2S CO Move the cursor next to the sensor you want to calibrate with the AIR A YES and RANGE VY SHIFT buttons In the example below the CH sensor is selected for span adjustment SELECT SENSOR TO CALIBRATE ESCAPE gt CH4 OXY H2S CO If you do not want to proceed with the span adjustment press and release the DISPLAY ADJUST NO button or place the cursor next to ESCAPE 96 Calibrating Using the Single Calibration Method EAGLE 2 Operator s Manual and press and release POWER ENTER RESET to return to the Calibration
144. brate FAILED SENSOR ENTER TO CONTINUE lt CH4 gt The buzzer and alarm LEDs activate in a double pulsing pattern Press and release the POWER ENTER RESET button to reset the alarm and continue to the Calibration Mode Screen Attempt to calibrate the TC channel again If the failure continues investigate the cause See Troubleshooting on page 70 37 Disconnect the tubing from the probe 38 Unscrew the demand flow regulator from the calibration cylinder 39 Use the RANGE YW SHIFT button to place the cursor next to the NORMAL OPERATION menu option then press and release the POWER ENTER RESET button to return to Measuring Mode EAGLE 2 Operator s Manual TC Calibration 215 Maintenance The TC sensor does not include any user serviceable parts This section includes a procedure for replacing the TC sensor Replacing the TC Sensor L 2 3 10 11 12 13 14 15 Verify that the EAGLE 2 is off Place the EAGLE 2 upside down on a flat surface or hold it upside down Unscrew the three case screws until they disengage from the top case They are captive screws so they will not fall off of the bottom case Turn the EAGLE 2 right side up and carefully lift the top case away from the bottom case Be careful not to lift it so far that it pulls on the main PCB with the cable that connects the top case to the main PCB Lay the top case down next to the bottom case to allow access to the flo
145. can be recharged by using the EAGLE 2 charger see Replacing or Recharging the Batteries on page 72 NOTE Use of batteries or battery chargers not specified by RKI Instruments Inc will void the CSA classification and may void the warranty WARNING To prevent ignition of a hazardous atmosphere batteries must only be changed or charged in an area known to be nonhazardous EAGLE 2 Operator s Manual Instrument Description 13 Flow System The EAGLE 2 flow system consists of the inlet fitting hydrophobic filter pump internal tubing differential pressure DP switch sensor chamber charcoal filter and exhaust fitting Inlet Fitting The inlet fitting is on the right front when viewed from the front of the bottom case It is a nickel plated brass quick connect fitting It mates with either the sample hose or with the hydrophobic probe Hydrophobic Filter The hydrophobic filter is located in the bottom case above the sensors Normally the hydrophobic probe accessory see Hose and Probe on page 15 will prevent water and particulate contamination from entering the flow system but if the probe is not used the hydrophobic filter will stop water and particulates from penetrating further into the flow system If it becomes dirty or water logged replace it see Replacing the Hydrophobic Filter on page 77 Pump A diaphragm pump inside the rear of the bottom case draws the sample to the sensors It can dra
146. case If the main PCB is not seated properly it may be damaged when the top case is re installed Make sure that the top case gasket is fully seated in its groove and carefully put the top case back on the bottom case If you have any difficulty mating the top and bottom cases inspect the placement of the main PCB and the placement of the top case gasket Turn the EAGLE 2 upside down and tighten the three case screws to secure the top case to the bottom case Calibrate the PID channel as described in PID Calibration on page 143 Replacing the PID Sensor L 2 3 10 11 Verify that the EAGLE 2 is off Place the EAGLE 2 upside down on a flat surface or hold it upside down Unscrew the three case screws until they disengage from the top case They are captive screws so they will not fall off of the bottom case Turn the EAGLE 2 right side up and carefully lift the top case away from the bottom case Be careful not to lift it so far that it pulls on the main PCB with the cable that connects the top case to the main PCB Lay the top case down next to the bottom case to allow access to the flow system Locate the PID sensor adapter It has a five wire cable and is normally located next to the pump The cable has a connector that mates to a PID sub PCB that is installed on the main PCB Figure 25 on page 137 shows a PID sensor in a typical location Grasp the sensor adapter firmly and pull it out of the PID flow chamber
147. ccessories Several optional accessories are available for the EAGLE 2 They include rechargeable Ni MH batteries battery chargers various special probes and dilution fittings The most commonly used optional accessories are described below Detailed instructions regarding the use of these and other available accessories are included in other parts of this manual Data logging accessories are briefly described in Data Logging on page 53 16 Optional Accessories EAGLE 2 Operator s Manual Rechargeable Ni MH Batteries Rechargeable Ni MH batteries are available for the EAGLE 2 A fully charged set of Ni MH batteries will power the EAGLE 2 for 18 hours The batteries will last for a minimum of 500 charge cycles See General Parts List on page 84 for ordering information Battery Chargers Three battery chargers are available for the EAGLE 2 to charge the optional Ni MH batteries the standard AC charger a DC charger with a vehicle plug adapter and an AC DC charger with a vehicle plug adapter AC Charger The standard AC charger consists of the charging module which includes all of the charging circuitry and an AC adapter The charging module includes a five foot cable with a connector on the end that mates with the EAGLE 2 charging socket The AC adapter plugs into a 115 VAC wall outlet and connects to the charging module with a jack on the end of a five foot DC output cable The AC adapter will also work for 100 VAC or 22
148. ce the EAGLE 2 upside down on a flat surface or hold it upside down Unscrew the three case screws until they disengage from the top case They are captive screws so they will not fall off of the bottom case Turn the EAGLE 2 right side up and carefully lift the top case away from the bottom case Be careful not to lift it so far that it pulls on the main PCB with the cable that connects the top case to the main PCB Lay the top case down next to the bottom case to allow access to the flow system Locate the infrared hydrocarbon sensor It has a twelve wire cable with a connector that mates to an infrared sub PCB that is installed on the main PCB and is normally located next to the pump Figure 41 on page 253 shows an infrared hydrocarbon sensor in a typical location Unscrew and remove the two screws that hold down the infrared hydrocarbon sensor bracket Grasp the sensor firmly and pull it out of the infrared hydrocarbon flow chamber Rock it back and forth gently if necessary to pull it out Take care not to pull the cable from the sub PCB Rotate the locking lever counterclockwise on the cable connector that mates to the infrared hydrocarbon sensor to unlock it Unplug the old infrared hydrocarbon sensor from the cable Connect the new infrared hydrocarbon sensor to the sensor cable and rotate the locking lever clockwise to lock the connector Insert the sensor into the infrared hydrocarbon flow chamber and push it in until it bo
149. ck The electrode stack can last for the life of the PID sensor if the EAGLE 2 is used in a very clean controlled environment When used in a heavily contaminated or dirty environment the electrode stack may only last a month A contaminated electrode stack will cause a drop in sensitivity which can cause problems calibrating the PID channel The electrode stack should be replaced if the PID sensor shows signs of contamination even after cleaning or replacing the lamp NOTE Do not touch the new lamp window the flat end with your fingers as this may contaminate the window with finger oil Verify that the EAGLE 2 is off 2 Place the EAGLE 2 upside down on a flat surface or hold it upside down 3 Unscrew the three case screws until they disengage from the top case They are captive screws so they will not fall off of the bottom case 4 Turn the EAGLE 2 right side up and carefully lift the top case away from the bottom case Be careful not to lift it so far that it pulls on the main PCB with the cable that connects the top case to the main PCB 5 Lay the top case down next to the bottom case to allow access to the flow system 6 Locate the PID sensor adapter It has a five wire cable and is normally located next to the pump The cable has a connector that mates to a PID sub PCB that is installed on the main PCB Figure 25 on page 137 shows a PID sensor in a typical location 7 Grasp the sensor adapter firmly and pull it out of the
150. combustible channel with a gas concentration of 1000 ppm or lower you must use a 24 inch humidifier tube to connect the gas bag to the rigid tube on the probe 23 Press and release the POWER ENTER RESET button to set the span adjustment to the programmed values 24 If all channels passed calibration the following screen sequence occurs AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END Y CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION EAGLE 2 Operator s Manual Calibrating Using the Auto Calibration Method 93 If any of the sensors cannot be adjusted to the proper value a screen displays that indicates a calibration failure and lists the sensor s that failed to calibrate In the example below the oxygen and HS channels failed calibration The other sensors calibrated normally FAIL SENSOR lt gt lt OXY gt lt H2S gt lt gt ENTER TO CONTINUE The buzzer and LED arrays activate in a double pulsing pattern Press and release the POWER ENTER RESET button to reset the alarm and return to the Calibration Mode Screen Attempt to calibrate again If the failure continues investigate the cause See Troubleshooting on page 70 25 Disconnect the tubing from the probe 26 Use the RANGE VW SHIFT button to place the cursor next to the NORMAL OPERATION menu option then press and release the POWER ENTER R
151. concentration listed on the cylinder s label not zero 244 Infrared Methane Calibration EAGLE 2 Operator s Manual 15 To adjust the values on the screen press and hold the RANGE W SHIFT button then press the DISPLAY ADJUST NO button and release both The following screen appears with the cursor next to CH4 ADJUST AUTO CALIBRATION VALUES gt CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm y 16 Place the cursor next to the channel whose gas value you want to change using the AIR A YES and RANGE YW SHIFT buttons 17 Press and release the POWER ENTER RESET button to select the channel The calibration gas value begins to flash 18 Use the AIR A YES and RANGE YW SHIFT buttons to adjust the calibration gas setting to the desired value NOTE The calibration gas value cannot be set lower than the low alarm setting If the calibration gas value listed on the calibration cylinder is lower than the current low alarm setting enter Setup Mode and change the low alarm setting See Updating the Alarm Point Settings on page 114 for instructions If you need to change the alarm point setting only to perform a calibration make sure that you change the alarm point setting back to its original value once the calibration has been performed 19 Press and release the POWER ENTER RESET button to save the change The calibration gas value stops flashing 20 Repeat Step 16 through Step 19 for any other channels that ne
152. configure the gas for a catalytic TC thermal conductivity or PID photo ionization detector sensor Only a catalytic sensor is used in a standard EAGLE 2 so only a catalytic sensor can be configured in a standard EAGLE 2 PID or TC sensors can only be configured if additional hardware not in a standard EAGLE 2 is factory installed 1 From the main menu place the cursor next to CONFIGURE GASES 2 Press and release POWER ENTER RESET The Configure Gases Screen appears with the cursor flashing next to CAT CONFIGURE GASES gt CAT CH4 CAT OPI OP2 OP3 y The OP1 OP2 and OP3 options are not defined in a standard 4 gas EAGLE 2 These options are only defined when hardware specific to optional sensors is factory installed in the EAGLE 2 See Appendix C Sub PCBs on page 132 for a description of this optional hardware and how it affects CONFIGURE GASES 3 To change the catalytic sensor gas configuration press and release POWER ENTER RESET 4 A screen appears with gas configuration choices for the catalytic channel CAT 0 50000ppm CH4 gt 50000 ppm CH4 11000 ppm HEX 40000 ppm H2 18000 ppm IBU Y For each gas the LEL lower explosive limit and gas name is displayed The 108 Using Setup Mode EAGLE 2 Operator s Manual LEL is shown in terms of ppm The available choices are on two screens RANGE SHIFT CAT 0 50000ppm CH4
153. d in the long probe tube e 4 foot stainless steel probe Figure 11 4 Foot Stainless Steel Probe This probe is designed for areas where it is necessary to put the probe tip in areas that are out of reach for even the 30 inch probes A stainless steel probe tube is used because the length of the probe tube requires a high degree of rigidity Stainless steel is more rigid than other normally used materials A small breather hole near the end of the probe tube prevents interruption of sampling and a low flow alarm if the probe tip is blocked e Barhole probe Figure 12 Barhole Probe This probe is designed specifically for barhole testing See Appendix K Using the EAGLE 2 in Bar Hole Mode for an in depth discussion of using the EAGLE 2 in Bar Hole Mode EAGLE 2 Operator s Manual Optional Accessories 19 e 10 inch probe with dust filter Figure 13 10 Inch Probe With Dust Filter This probe is designed for use where drawing water or moisture into the EAGLE 2 is not a concern Instead of a hydrophobic filter a cotton dust filter is used e 32 inch telescoping probe with dust filter This probe is designed for use where it is necessary to put the probe tip in areas not accessible with the 10 inch probe with dust filter and applications where the probe tube must be collapsible for storage e 7 foot telescoping probe with dust filter a
154. d the maximum setting is 365 days The factory setting is 90 days 1 From the main menu place the cursor next to CAL INTERVAL EAGLE 2 Operator s Manual Using Setup Mode 123 2 Press and release POWER ENTER RESET The Set Calibration Interval Screen appears SET CALIBRATION INTERVAL UP DOWN THEN ENTER 90 DAYS 3 Use AIR A YES or RANGE YW SHIFT to display the desired setting Tip Press and hold AIR A YES or RANGE Y SHIFT to rapidly scroll through the settings 4 Press and release POWER ENTER RESET to save the setting and return to the main menu Updating LC BH Mode Setting With LC BH MODE SELECT set to LC amp BH the Mode Select Screen appears when the unit is turned on You are able to select from Normal Mode Leak Check Mode and Bar Hole Mode With LC BH MODE SELECT set to BAR HOLE ONLY the Mode Select Screen appears when the unit is turned on You are able to select from Normal Mode and Bar Hole Mode With LC BH MODE SELECT set to LEAK CHECK ONLY the Mode Select Screen appears when the unit is turned on You are able to select from Normal Mode and Leak Check Mode With LC BH MODE SELECT set to OFF the Mode Select Screen does not appear when the unit is turned on and the unit goes into Normal Mode after the start up sequence This setting is factory set to OFF when a unit is shipped unless the instrument is ordered for bar hole measurement or leak checking use See Appendix K Using the
155. de Returning the EAGLE 2 to its default configuration can be useful if various setup parameters have been changed in the field and you want to return the EAGLE 2 to its original configuration as shipped from the factory EAGLE 2 Operator s Manual Using Setup Mode 127 The standard default gas configuration is LEL oxygen H S CO If you have turned any channels off or have added channels to your EAGLE 2 you will have to re setup your EAGLE 2 to the desired gas combination if you restore the EAGLE 2 to its default configuration There are some special EAGLE 2 configurations that may have a different default configuration than the standard Consult RKI Instruments Inc for information regarding non standard default configurations WARNING When the EAGLE 2 is restored to its default configuration the zero and span values for each channel are reset You must recalibrate all active channels if you restore the EAGLE 2 to its default configuration 1 From the main menu place the cursor in front of DEFAULT SETTINGS 2 Press and release POWER ENTER RESET The Restore Default Configuration Screen appears asking if you want to restore the default configuration RESTORE DEFAULT CONFIGURATION 3 If you do not want to restore the default configuration press and release DISPLAY ADJUST NO to return to the main menu If you do want to restore the default configuration continue with Step 4 4 Press and release AIR A YES
156. demand flow regulator 12 Install the probe on the EAGLE 2 inlet fitting Make sure the probe is complete with internal O ring and membrane and that the two halves of the probe are tightened firmly together to avoid leaks that can affect the calibration See Figure 19 Replacing the Particle Filter and Hydrophobic Filter Disk on page 76 for an illustration of the internal parts of the probe EAGLE 2 Operator s Manual PID Calibration 153 13 Move the cursor next to the AUTO CALIBRATION menu item by using the AIR A YES button CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 14 Press and release the POWER ENTER RESET button to display the Calibration Gas Values Screen CAL GAS VALUES CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm ENTER TO BEGIN CAL The gas concentrations displayed in the Calibration Gas Values Screen must match the gas concentrations listed on the 4 gas calibration cylinder If all concentrations match go to Step 25 If one or more concentrations do not match continue with Step 15 If you do not want to continue with the calibration press and release the DISPLAY ADJUST NO button to return to the Calibration Mode Screen NOTE The RKI 4 gas cylinder typically contains 12 O by volume When using the auto calibration method be sure to set the OXY auto calibration value to agree with the concentration listed on the c
157. desired full scale value It may be any value from 10 to 100 in increments of 10 16 Press and release POWER ENTER RESET to enter the full scale value The confirmation screen appears In the example below the user defined gas has been selected and defined as propane with the gas name set to PRO CHANGE TO PRO PRESS YES OR NO 17 If you want to accept the gas configuration change press and release AIR A YES The unit will return to the Configure Gases screen CONFIGURE GASES gt CAT CH4 CAT OP1 PRO TC OP2 OP3 lt If you do not want to accept the gas configuration change press and release DISPLAY ADJUST NO to return to the screen with the gas choices shown in Step 10 on page 219 You can either scroll down to END and press POWER ENTER RESET to return to the Configure Gases screen or continue from Step 10 on page 219 to select a new gas 18 Use RANGE YW SHIFT to place the cursor next to END 19 Press and release POWER ENTER RESET to return to the main menu 20 Use RANGE YW SHIFT to place the cursor in front of NORMAL OPERATION at the bottom of the main menu 220 Configuring the TC Gas in Setup Mode EAGLE 2 Operator s Manual 21 Press and release POWER ENTER RESET 22 A screen appears that asks if you want to save the changes you have made SAVE ALL CHANGES IN MEMORY NOTE If you entered Setup Mode and did not make any changes the abo
158. dicate higher than 50 000 ppm the equivalent of 33 LEL and 5 volume for this ratio but will continue to indicate LEL readings up to 100 LEL and volume readings up to 15 volume the equivalent of 150 000 ppm if the display units are changed to LEL or volume In addition all adjustable ppm parameters cannot be set higher than 50 000 ppm 10 Press and release POWER ENTER RESET to enter the ppm ratio The FACTOR for the gas begins to flash The FACTOR for the gas is the EAGLE 2 Operator s Manual Using Setup Mode 111 response factor for the user defined gas relative to methane The response factor must be obtained by testing the user defined gas and comparing its response to methane This parameter is used by the relative response feature See Catalytic Sensor Relative Response Screen on page 43 for a description of the relative response feature and how to use it See Combustible Gas Detection on page 31 for a list of response factors for several common hydrocarbon gases that have already been tested 11 Use AIR A YES and RANGE YW SHIFT to increase or decrease the response factor to the desired number 12 Press and release POWER ENTER RESET to enter the response factor The sensor voltage setting screen appears with the sensor voltage flashing The current gas and range is shown at the top of the screen and the current sensor voltage is shown at the bottom of the screen CAT 0 50000ppm CH4 SETTING EV UP DOWN
159. during the startup sequence when calibration is due See Updating the Calibration Reminder Setting on page 122 CAUTION BEFORE EACH DAY S USAGE SENSITIVITY IN THE LEL RANGE MUST BE TESTED ON A KNOWN CONCENTRATION OF THE COMBUSTIBLE TARGET GAS METHANE EQUIVALENT TO 25 50 OF FULL SCALE CONCENTRATION the LEL full scale is 100 LEL ACCURACY MUST BE WITHIN 0 to 20 OF ACTUAL ACCURACY MAY BE CORRECTED BY FOLLOWING THE CALIBRATION INSTRUCTIONS FOR THE COMBUSTIBLE CHANNEL BELOW If the combustible channel passes the above response test and does not require calibration the unit should still be calibrated periodically The optimum frequency of calibration depends heavily on how the EAGLE 2 is used For example instruments used daily may need to be calibrated weekly or monthly while instruments that are used only a few times a year may need to be calibrated before each use Typical calibration frequencies range from monthly to quarterly Make sure to perform the combustible channel response test as described above and make sure to develop a calibration schedule tailored to your application that takes this test and required calibration resulting from this test into account EAGLE 2 Operator s Manual Overview 55 Calibration Supplies and Equipment To calibrate the EAGLE 2 you will need e Known calibrating samples of the gases being detected The combustible and toxic gas samples should have concentrations between 10
160. e Appendix D PID Sensors for a description of the PID sensors Appendix E ESM 01 Toxic Sensors for a description of the ESM 01 toxic sensors Appendix F TC Sensors for a description of the TC sensors Appendix G Infrared Carbon Dioxide Sensors for a description of the IR CO sensors Appendix H Infrared Methane Sensor for a description of the IR methane sensors and Appendix I Infrared Hydrocarbon Sensor for a description of the IR hydrocarbon sensor The standard sensors use different detection principles as described below Catalytic Combustible Gas Sensor LEL Sensor The catalytic combustible gas LEL sensor detects combustible gas in the LEL range It uses a catalytic element for detection The reaction of gas with oxygen on the catalyst causes a change in the resistance of the element which changes the current flowing through it The current is amplified by the EAGLE 2 s circuitry converted to a measurement of combustible gas concentration and displayed on the LCD The LEL sensor housing includes a sintered metal flame arrestor on one end that allows gas to diffuse into the sensor On the other end five pins extend from the sensor The sensor cable connects to these pins on one end and terminates in a four position connector on the other end which plugs into the HC socket on the main PCB see Main PCB on page 12 8 Instrument Description EAGLE 2 Operator s Manual Oxygen Sensor T
161. e The calibration gas value stops flashing 11 Repeat Step 7 through Step 10 for any other channels that need to be changed 12 When you are done adjusting the calibration gas values move the cursor down past the bottom of the screen next to END ADJUST AUTO CALIBRATION VALUES gt END A 13 Press and release the POWER ENTER RESET button The following screen appears DO YOU WANT TO STORE NEW VALUE S IN MEMORY FOR FUTURE CALIBRATIONS PRESS YES OR NO 14 If you select YES by pressing and releasing the AIR A YES button the changes that you made will be saved in the EAGLE 2 s memory as the new auto calibration gas values If you select NO by pressing and releasing the DISPLAY ADJUST NO button the changes you made will be used for any calibrations performed during the current operating session only The EAGLE 2 will delete the EAGLE 2 Operator s Manual Calibrating Using the Auto Calibration Method 61 changes when the unit is turned off and will load the previous set of auto calibration values when it is turned on again 15 When you make your selection and press the desired button the unit returns to the Calibration Gas Values Screen CAL GAS VALUES CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm ENTER TO BEGIN CAL 16 Press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen with CAL IN PROCESS flashing CAL IN PROCESS CH4 0
162. e If you have any difficulty mating the top and bottom cases inspect the placement of the main PCB and the placement of the top case gasket 17 Turn the EAGLE 2 upside down and tighten the three case screws to secure the top case to the bottom case 18 Calibrate the infrared CO channel as described in Infrared CO Calibration on page 227 Parts List Table 25 Infrared CO EAGLE 2 Parts List 47 5051RK Infrared CO sensor cable DEM 3313 1 Infrared CO sensor 0 5 vol DEM 3313 4 Infrared COs sensor 0 10 000 ppm DEM 3313 5 Infrared COs sensor 0 60 vol 81 0071RK 01 Calibration cylinder 5000 ppm CO in nitrogen 34 liter 81 0071RK 03 Calibration cylinder 5000 ppm CO in nitrogen 103 liter 81 0072RK 01 Calibration cylinder 2 5 vol COs in nitrogen 34 liter 81 0072RK 03 Calibration cylinder 2 5 vol CO in nitrogen 103 liter 81 0073RK 01 Calibration cylinder 15 vol CO in nitrogen 34 liter 81 0073RK 03 Calibration cylinder 15 vol CO in nitrogen 103 liter EAGLE 2 Operator s Manual Parts List 237 Appendix H Infrared Methane Sensor Overview This appendix describes the EAGLE 2 s infrared methane sensors and includes instructions to use an EAGLE 2 that has an infrared methane sensor installed It also includes instructions to replace an infrared methane sensor Table 26 Infrared Methane Sensor Specifications LEL Configuration 0 100 LEL CH4 1 LEL 10 LEL 50 LEL Table 27 Inf
163. e POWER ENTER RESET button to set the span adjustment for each channel to the programmed values EAGLE 2 Operator s Manual ESM 01 Calibration 187 28 If all channels passed calibration the following screen sequence occurs AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END Y CAL GAS VALUE NH3 25 ppm ENTER TO BEGIN CAL If any of the sensors cannot be adjusted to the proper value a screen displays that indicates a calibration failure and lists the sensor s that failed to calibrate In the example below the oxygen and H S channels failed calibration The other sensors calibrated normally FAIL SENSOR lt gt lt OXY gt lt H2S gt lt gt ENTER TO CONTINUE The buzzer and alarm LEDs activate in a double pulsing pattern Press and release the POWER ENTER RESET button to reset the alarm and continue to the Calibration Value Screen for the ESM 01 channel After calibrating the ESM 01 channel by following the instructions below attempt to calibrate the standard channels again If the failure continues investigate the cause See Troubleshooting on page 70 29 Remove the tubing from the rigid tube on the probe 30 Unscrew the 4 gas cylinder from the demand flow regulator 188 ESM 01 Calibration EAGLE 2 Operator s Manual 31 If you want to change the ESM 01 channel s calibration gas value follow Step 15 Step 24 above beginning with the ESM 0
164. e alarm point setting back to its original value once the calibration has been performed 9 Press and release the POWER ENTER RESET button to proceed to the Single Calibration Apply Gas Screen CAL IN PROCESS is flashing SINGLE CALIBRATION APPLY GAS CH4 0O LEL CAL IN PROCESS ENTER WHEN DONE 10 Connect the tubing from the demand flow regulator to the rigid tube on the probe Allow the EAGLE 2 to draw gas for one minute NOTE If you are calibrating the catalytic combustible channel with a gas concentration of 1000 ppm or lower you must use a 24 inch humidifier tube to connect the demand flow regulator to the rigid tube on the probe 11 Press and release the POWER ENTER RESET button to perform the calibration 12 When the span adjustment is made the EAGLE 2 calculates the range of adjustment minimum and maximum it could have made based on its EAGLE 2 Operator s Manual Calibrating Using the Single Calibration Method 67 response level to the applied gas This calculated range is independent of the calibration gas value that was entered in Step 8 and Step 9 above The adjustment range is included on the result screen to indicate the condition of the sensor If the calibration gas value is in the adjustment range the span adjustment will pass If the calibration gas value is out of the adjustment range the span adjustment will fail If the span adjustment is successful the following screens display
165. e been adjusted With SPAN FACTOR set to OFF this span adjustment does not appear in the calibration results screen 1 From the main menu place the cursor next to SPAN FACTOR 2 Press and release POWER ENTER RESET The Span Factor Screen appears SPAN FACTOR DISPLAY UP DOWN THEN ENTER ON EAGLE 2 Operator s Manual Using Setup Mode 129 3 Use AIR A YES or RANGE YW SHIFT to display the desired setting 4 Press and release POWER ENTER RESET to save the setting and return to the main menu Updating the Language Setting This setting allows you to select the language for the EAGLE 2 s user interface The available choices are English factory setting Spanish French Italian and German 1 From the main menu place the cursor next to SELECT LANGUAGE 2 Press and release POWER ENTER RESET The Select Language Screen appears with the cursor in front of the current language SELECT LANGUAGE gt ENGLISH ESPANOL FRANCAIS ITALIANO DEUTSCH y 3 Use AIR A YES or RANGE V SHIFT to move the cursor in front of the desired language If you do not wish to select a new language either press and release DISPLAY ADJUST NO or move the cursor all the way to the bottom of the list in front of END and press and release POWER ENTER RESET The unit will return to the main menu 4 Press and release POWER ENTER RESET to save the new language setting and return to the main menu The EAGLE 2 s user i
166. e infrared hydrocarbon instrument configurations do not include the 4 sensor flow chamber 252 Overview EAGLE 2 Operator s Manual Top Case Not Shown Charcoal Filter Infrared HC Sensor Pump __ lt Standard Flow Chamber OE OS 7 aaa Dii hs S Fan J a a O D y peli O ap HEH CO Sensor a Oxygen Sensor oN Oxygen Sensor LEL Sensor and Bracket H2S Sensor Sensor Bracket Figure 41 Typical Infrared Hydrocarbon Sensor Location Infrared Hydrocarbon Sensor The infrared hydrocarbon sensor is a cylindrical sensor with a diffusion opening on the front and a 12 pin circular connector on the back A 12 wire cable plugs into the back of the infrared hydrocarbon sensor with a circular connector that includes a locking lever The other end of the cable plugs into an infrared sub PCB see description below that is installed on the main PCB The sensor is held in the infrared flow chamber by a bracket on standoffs Infrared Sub PCB The infrared sub PCB is a circuit board that is installed on the main PCB in one of the 3 sub PCB positions when an infrared hydrocarbon sensor is used with the EAGLE 2 The infrared hydrocarbon sensor connects to the sub PCB with a 12 position connector The sub PCB plugs into the main PCB and is held in place with a scre
167. e operation is complete before returning to the Calibration Mode Screen ADJUSTING ZERO RELEASE AIR BUTTON ADJUSTING ZERO COMPLETE CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION gt PERFORM AIR ADJUST NORMAL OPERATION Performing a Span Adjustment in Auto Calibration 1 Slide the tubing clamp onto the tubing and connect the tubing to the sample bag s inlet fitting Leave the clamp unclamped for now EAGLE 2 Operator s Manual Calibrating Using the Auto Calibration Method 89 2 Install the probe on the EAGLE 2 inlet fitting Make sure the probe is complete with internal O ring and membrane and that the two halves of the probe are tightened firmly together to avoid leaks that can affect the calibration See Figure 19 Replacing the Particle Filter and Hydrophobic Filter Disk on page 76 for an illustration of the internal parts of the probe 3 Move the cursor next to the AUTO CALIBRATION menu item by using the AIR A YES button CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 4 Press and release the POWER ENTER RESET button to display the Calibration Gas Values Screen CAL GAS VALUES CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm ENTER TO BEGIN CAL The gas concentrations displayed in the Calibration Gas Values Screen must match the gas concentrations listed on the 4 gas calibration cy
168. e selected channel In the example below the cursor has been moved next to the NOT USED selection CHANNEL 3 H25 OPI OP OP3 gt NOT USED The OP1 OP2 and OP3 options are not defined for a standard 4 gas EAGLE 2 These options are only defined when hardware specific to optional sensors is factory installed in the EAGLE 2 The in the gas name field indicates that hardware necessary to support an optional sensor is not installed If your EAGLE 2 supports one or more of these optional sensors the target gas will appear instead of See Appendix C Sub PCBs on page 132 for a description of this optional hardware and how it affects CHANNEL CONFIGURATION Press and release POWER ENTER RESET to select the channel configuration In the example below channel 3 has been turned off by selecting NOT USED gt CONFIGURE CHANNELS CH1 CAT CH4 CH2 OXY OXY CH3 NOT USED CH4 CO CO y EAGLE 2 Operator s Manual Repeat Step 5 and Step 6 for any other channels you want to configure Using Setup Mode 107 8 Use RANGE VW SHIFT to move the cursor next to the END menu item NOTE If you want to exit to the main menu without saving any channel configuration changes press and release DISPLAY ADJUST NO 9 Press and release POWER ENTER RESET to save the changes and return to the main menu Configuring the Combustible Gas This menu item allows you to
169. e slots on the side of the PID sensor near the face Figure 31 Using Removal Tool 11 Squeeze the removal tool to push the tabs into the sensor slots until the electrode stack and lamp are released 12 Carefully lift the PID sensor body away from the electrode stack and lamp If the lamp remains lodged in the sensor body carefully remove it with tweezers 13 If the spring in the lamp cavity comes out place it back into the lamp cavity 14 Discard the old PID lamp NOTE At this point it is recommended that the finger cots be used on the fingers handling the lamp Finger cots are included with the lamp cleaning kit 15 Hold the electrode stack between the thumb and forefinger of one hand and place the window end of the new lamp inside the O ring seal in the electrode stack with the other hand as shown below Take care not to touch the lamp window 166 Maintenance EAGLE 2 Operator s Manual 16 17 18 19 20 21 22 23 Figure 32 Reinstalling the Electrode Lamp Twisting the lamp slightly during insertion will help to ensure the lamp window is snug against the stack s front electrode The lamp should be supported by the O ring Continuing to hold the electrode stack between your forefinger and thumb carefully insert the lamp into the lamp cavity in the sensor ensuring that the lamp remains in position Press in the electrode stack firmly to ensure that the stack wing clips are engaged and
170. eS Beale ads eee De nee BE 254 Infrared Hydrocarbon Calibration 0 0 0 0 eee eee eens 255 Maintenances is cae toe a nE EE EAN RE E ANAT awd abl Pe 264 Replacing the IR Hydrocarbon Sensor 000 ees 264 Parts Eist Sick its ents tons be ted Gre peta tucks waar accent maces wie gnar a ee 265 Appendix J Methane Elimination Mode 06 266 Overview 53 chee St Hence a bus AOA ER OM EL Eee OEE NESE t Se eee aS 266 Monitoring in Methane Elimination Mode 4 4 266 Calibrations e ee den ea e iis de Sealed aes don ei eS 269 Appendix K Using the EAGLE 2 in Bar Hole Mode 270 OVERVIEW cas EOE us EM EE ee tena oe eres teats Ba ee ages 270 Stare Up Bar Hole Moder 2 casi s aces teu g eae eireed aaa ees 270 Turning on EAGLE 2 Bar Hole Mode 00000055 270 Performing a Demand Zero Bar Hole Mode 4 275 Bat Hole Testing ce ek edw tee Cees Coke hee eet Bhat eae he es 276 Performing a Bar Hole Test i odio Natta acetylene ea eves dade 276 Turning off the EAGLE 2 Bar Hole Mode 4 4 278 Appendix L Using the EAGLE 2 in Leak Check Mode 279 QVCEVICW Au pei eae ane Bo Mee eee ede E ns mn ed Ried Re miare Beek 279 Start Up Leak Check Mode ununnnannnnan aae 279 Turning On the EAGLE 2 Leak Check Mode 279 Performing a Demand Zero Leak Check Mode 254 Peale SUING etcetera heen acts a Sd tating RSet echt Stone ee It aOR
171. ed place the cursor next to it press and release the POWER ENTER button NOTE To exit the selection screen without saving a change press and release the DISPLAY ADJUST NO button You will return to the Select Station ID screen without saving the station ID change 4 The unit will save the selected station ID as the current one and proceed to the Time in Operation Screen Time in Operation Screen The Time In Operation Screen displays the length of time since the EAGLE 2 was turned on if the lunch break feature is turned off With the lunch break feature turned off the time in operation is reset when the EAGLE 2 is turned off See Updating the Lunch Break Setting on page 129 for a description of the lunch break feature TIME IN OPERATION 45 MINUTES If the lunch break feature is turned on the time in operation will only be reset if you do not choose to resume the peak and TWA measurements when the EAGLE 2 is turned on in the Resume Measurement Screen described in Step 7 on page 23 in Turning On the EAGLE 2 If you choose to resume the peak and TWA measurements during startup the EAGLE 2 will include the time in operation when the unit was last turned off in the current time in operation EAGLE 2 Operator s Manual Display Mode 51 Date Time Screen The Date Time Screen displays the current date and time 9 26 2009 09 08 35 Data Logging Screen CAUTION Once you clear the data logger
172. ed CO sensor is used with the EAGLE 2 The infrared CO sensor connects to the sub PCB with a 12 position connector The sub PCB plugs into the main PCB and is held in place with a screw flat washer lock washer There are no user serviceable parts on the infrared sub PCB CO Scrubber A carbon dioxide scrubber is mounted to the exterior side of EAGLE 2s that are factory shipped with carbon dioxide sensors for the ranges of 0 5 vol and 0 10 000 ppm NOTE EAGLE 2s with a range of 0 60 vol CO do not include a scrubber since the normal background of CO in air is negligible when compared to the full scale of these units This scrubber is for use when setting the carbon dioxide sensor s zero reading only Two black vinyl caps cover either end of the carbon dioxide scrubber To prolong the life of the scrubber be sure the caps are installed while the scrubber is not in use or while it is being stored Replace the scrubber when it turns from white to a violet color End Caps Figure 39 Carbon Dioxide Scrubber EAGLE 2 Operator s Manual Description 225 Start Up and Normal Operation For instructions to startup and use an EAGLE 2 that includes an infrared CO sensor reference Start Up on page 22 Measuring Mode Normal Operation on page 29 and Measuring Mode Alarms on page 35 Follow these instructions keeping the following special considerations in mind e A background level of CO exists in fres
173. ed firmly together to avoid leaks that can affect the calibration See Figure 19 Replacing the Particle Filter and Hydrophobic Filter Disk on page 76 for an illustration of the internal parts of the probe 13 Move the cursor next to the AUTO CALIBRATION menu item by using the AIR A YES button CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 14 Press and release the POWER ENTER RESET button to display the Calibration Gas Values Screen CAL GAS VALUES CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm ENTER TO BEGIN CAL The gas concentrations displayed in the Calibration Gas Values Screen must match the gas concentrations listed on the 4 gas calibration cylinder If all concentrations match go to Step 25 If one or more concentrations do not match continue with Step 15 If you do not want to continue with the calibration press and release the DISPLAY ADJUST NO button to return to the Calibration Mode Screen NOTE The RKI 4 gas cylinder typically contains 12 O by volume When using the auto calibration method be sure to set the OXY auto calibration value to agree with the concentration listed on the cylinder s label not zero 210 TC Calibration EAGLE 2 Operator s Manual 15 To adjust the values on the screen press and hold the RANGE W SHIFT button then press the DISPLAY ADJUST NO button and release both The following screen a
174. ed to be changed 21 When you are done adjusting the calibration gas values move the cursor down past the bottom of the screen next to END ADJUST AUTO CALIBRATION VALUES gt END EAGLE 2 Operator s Manual Infrared Methane Calibration 245 22 Press and release the POWER ENTER RESET button The following screen appears DO YOU WANT TO STORE NEW VALUE IN MEMORY FOR FUTURE CALIBRATIONS PRESS YES OR NO 23 If you select YES by pressing and releasing the AIR A YES button the changes that you made will be saved in the EAGLE 2 s memory as the new auto calibration gas values If you select NO by pressing and releasing the DISPLAY ADJUST NO button the changes you made will be used for any calibrations performed during the current operating session only The EAGLE 2 will delete the changes when the unit is turned off and will load the previous set of auto calibration values when it is turned on again 24 When you make your selection and press the desired button the unit returns to the Calibration Gas Values Screen CAL GAS VALUES CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm ENTER TO BEGIN CAL 25 Press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen with CAL IN PROCESS flashing CAL IN PROCESS CH4 0 LEL OXY 20 9 vol H2S 0 0 ppm CO 0 ppm ENTER WHEN DONE If you do not want to proceed with the calibration p
175. eed to be installed in the EAGLE 2 but if there are batteries installed they must be NiMH batteries and they must have a charge WARNING Use the EAGLE 2 charger s continuous operation mode to 3 4 5 6 power the EAGLE 2 only if NiMH batteries are installed in the EAGLE 2 or if no batteries are installed in the EAGLE 2 Do not use the charger for continuous operation if alkaline batteries are installed Place the EAGLE 2 in the area to be monitored Plug the power adapter into either an AC outlet or into a vehicle outlet depending on which charger is being used Set the switch on the module to CONT OPERAT Make sure the EAGLE 2 is off Make sure the adapter and module are connected Make sure that the NiMH batteries are either charged or removed NOTE If the batteries are not charged the EAGLE 2 will not turn on and will instead give a Charge Batteries indication when it is powered up after Step 7 below 28 Using the Battery Charger for Continuous Operation EAGLE 2 Operator s Man 7 Insert the module s round plug into the EAGLE 2 s charging jack as shown in Figure 15 below Figure 15 Connecting the EAGLE 2 to the Charger 8 See Chapter 3 Operation on page 22 for instructions for start up and operation of the EAGLE 2 9 While the charging module is powering the EAGLE 2 its amber LED will be off and its green LED will be on Measuring Mode Normal Operation When the EAGLE
176. ement Program See the Data Logger Management Program operator s manual for data logging and downloading instructions Buzzer A solid state electronic buzzer is located on the front of the top case It is a panel mounting type of buzzer and is water resistant and sealed to the inside of the top case with an O ring The buzzer sounds for gas alarms malfunctions low battery voltage and as an indicator during use of the EAGLE 2 s many display and adjustment options Battery Case amp Batteries Four C size alkaline batteries standard or optional rechargeable C size Ni MH batteries power the EAGLE 2 They are installed in the battery case which is located at the rear of the bottom case The battery case is secured to the bottom case with a thumbscrew Instrument run time is dependent upon battery type At 25 C alkaline batteries power the EAGLE 2 for 16 hours of non alarm operation Ni MH batteries will power the EAGLE 2 for 18 hours of non alarm operation The current battery voltage is viewable in Display Mode see Display Mode on page 40 When the EAGLE 2 detects low battery voltage a low battery warning is activated When battery voltage is too low for operation the EAGLE 2 sounds a dead battery alarm The alkaline or Ni MH batteries can be accessed for replacement by unscrewing the thumbscrew that secures the battery case to the bottom case and pulling the battery case away from the bottom case The Ni MH batteries
177. en seconds GAS MONITOR INSTRUMENTS EAGLE 2 5 The Battery Voltage Screen appears for a few seconds BATTERY MIN 4 3 VOLTS BATTERY NOW 5 2 VOLTS 6 The Mode Select Screen displays gt NORMAL MODE BAR HOLE MODE LEAK CHECK MODE 7 The cursor will flash in front of NORMAL MODE Use the RANGE VY SHIFT button to move the cursor next to LEAK CHECK MODE 8 With Leak Check Mode selected press and release the POWER ENTER RESET button to begin the Leak Check Mode startup sequence NOTE If no button is pressed for 20 seconds the unit will proceed into whichever mode has the cursor in front of it 9 The Active Gases Screen appears for a few seconds indicating which channels are active and their target gas ACTIVE GASES CH4 OXY H2S CO CH4 280 Start Up Leak Check Mode EAGLE 2 Operator s Manual 10 The gas alarm setpoints are displayed by three screens in sequence the Low Alarm Screen High Alarm Screen and STEL TWA Alarm Screen Each screen remains on the LCD for three seconds A CH4 50 LEL HL OXY 23 5 vol IA H2S 30 0 ppm GR CO 50ppm HM CH4 50 vol S A CH4 10 LEL LL OXY 19 5 vol OA H2S 10 0 ppm WR CO 25 ppm M CH4 10 vol S ALARMS STEL amp TWA H2S ppm 15 0 10 0 CO ppm 200 25 CH4 vol OFF OFF 11 After the alarm screens if CAL REMINDER is turned on the screen that appears next depends on how CAL PAST DUE AC
178. ens will be the alarm points for Inert Mode and not for Normal Mode 7 When the warm up is complete and the instrument is operating in Inert Mode the screen will appear as shown below if the instrument is sampling a fresh air environment I CH4 O LEL N OXY 20 9vol ALRM2 EH2S 0 0ppm RCO Oppm T If the instrument is sampling a fresh air environment the oxygen channel will be in alarm EAGLE 2 Operator s Manual Start Up 297 Operation See Measuring Mode Normal Operation on page 29 for operating instructions keeping in mind that the Inert Mode oxygen alarm settings are different You can access the Inert Mode Selection Screen while in Inert Mode or Normal Mode by pressing and holding the RANGE w SHIFT button MODE SELECT UP DOWN THEN ENTER NORMAL MODE Use the AIR A YES or RANGE v SHIFT button to display the mode you wish to enter and press and release the POWER ENTER RESET button 298 Operation EAGLE 2 Operator s Manual Warranty RKI Instruments Inc warrants the EAGLE 2 sold by us to be free from defects in materials workmanship and performance for a period of two years from the date of shipment from RKI Instruments Inc This includes the instrument and the original sensors Replacement parts are warranted for 1 year from the date of their shipment from RKI Instruments Inc Any parts found defective within their warranty period will be repaired or replaced at our o
179. erator s Manual 19 A screen appears that asks if you want to save the changes you have made SAVE ALL CHANGES IN MEMORY NOTE If you entered Setup Mode and did not make any changes the above screen will still appear In this case press and release DISPLAY ADJUST NO to proceed to exit Setup Mode and begin the EAGLE 2 s startup sequence 20 If you do not want to save the changes press and release DISPLAY ADJUST NO The unit will begin its startup sequence without saving the changes If you do want to save the changes press and release AIR A YES and continue with the next step 21 A confirmation screen appears asking if you are sure you want to save the changes ARE YOU SURE YOU WANT TO SAVE ALL CHANGES IN MEMORY 22 If you want to save the changes press and release AIR A YES to save the changes A screen will appear for a few seconds indicating that the changes have been saved and the unit will begin its startup sequence If you do not want to save the changes press and release DISPLAY ADJUST NO to proceed to the unit s startup sequence without saving changes EAGLE 2 Operator s Manual Configuring the PID Gas in Setup Mode 177 Parts List Part Number 33 0560RK 33 0561 RK 51 1500RK 51 1501RK 57 2005RK 61 0300RK 01 61 0300RK 02 80 0131 RK 20 80 0605RK 80 0610RK 80 0615RK 80 0620RK 81 0103RK 01 81 0103RK 03 81 0103RK 04 81 0104RK 01 81 0104RK 03 81 0104RK 04 81 01
180. et faut des 237 Appendix H Infrared Methane Sensors e000 238 OVER VICW foes ht acti On E eraa aa e batts Dea te kell ach ete EET aceite Ae 238 Target Gases ch Sie Geis teh tht lsat he steal Deku d dee shee RS Soak ba hh bees 238 DeSCrpPUOMin ves sede eri more ook WON ai wie pede eee ee ee wie 239 Infrared Methane Sensor os 4 as bate eb aha i GAO 240 Infrared SUD PC Bix ond sae eee Re Rene 4 enue GA eee he eRe 240 Start Up and Normal Operation e ode ced sg Pe Wes eles 240 Detection Ranges wishes 6h Six dos Dib vile ed D tnd cep thle hn a We oP ash Athen eons 240 Catalytic LEL Sensor Screen 502 65s eyed ect vede Clee ava ves 240 Infrared Methane Calibration x5 ouo seo ee do Ck eee Sie eae bees 241 Maintenances 2 34 0 4 82 5 odie bin eet a 8 bee ee Ha he ROE ARERE CS 250 Replacing the IR Methane Sensor 0 000 eee 250 Parts Lists Pee Ee hig eae sade eet ene ae eek es eae 251 Appendix I Infrared Hydrocarbon Sensor sosssessses 252 OVEVICW Sc nd e EE a A A E GENEE RINE 252 Tarcet Gases i Bile aoe Gs e a a Wee alata ae 252 DOSER PUOM 4 2 3a ha she ere ne HES EWE ORS E RE RE eek 252 Table of Contents EAGLE 2 Operator s Manual Infrared Hydrocarbon Sensor 3 i ie elie ey BS PA eee 253 Infrared Sub PCB l a nonan 253 Start Up and Normal Operation escv0d dues eke eae does Ven eeales 254 0 100 LEL 2 0 30 0 Yovol Autoranging 0 254 Catalytic LEL Sensor Screen lt 2 ia2 nce bene
181. f the ESM 01 channel cannot be adjusted to the proper value a screen displays that indicates a calibration failure and indicates that the ESM 01 sensor failed to calibrate FAILED SENSOR ENTER TO CONTINUE lt NH3 gt The buzzer and alarm LEDs activate in a double pulsing pattern Press and release the POWER ENTER RESET button to reset the alarm and continue to the Calibration Mode Screen Attempt to calibrate the ESM 01 channel again If the failure continues investigate the cause See Troubleshooting on page 70 37 Disconnect the tubing from the probe 38 Unscrew the demand flow regulator from the calibration cylinder 39 Use the RANGE YW SHIFT button to place the cursor next to the NORMAL OPERATION menu option then press and release the POWER ENTER RESET button to return to Measuring Mode 190 ESM 01 Calibration EAGLE 2 Operator s Manual Calibrating with a 5 Gas Cylinder The calibration instructions below show a5 channel instrument which has the four standard channels LEL oxygen H S CO and an ESM 01 channel for sulfur dioxide SO gt L See Calibration Supplies and Equipment on page 56 for a description of the necessary calibration supplies See Table 19 on page 202 for available cylinders Make sure your calibration cylinder is appropriate for the ESM 01 detection range Find a fresh air environment This is an environment free of toxic or combustible gases and of normal oxygen conte
182. f the infrared hydrocarbon channel passed calibration the following screen sequence occurs AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END Y CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION If the infrared hydrocarbon channel cannot be adjusted to the proper value a screen displays that indicates a calibration failure and indicates that the infrared hydrocarbon sensor failed to calibrate FAILED SENSOR ENTER TO CONTINUE lt HC gt The buzzer and alarm LEDs activate in a double pulsing pattern Press and release the POWER ENTER RESET button to reset the alarm and continue to the Calibration Mode Screen Attempt to calibrate the infrared hydrocarbon channel again If the failure continues investigate the cause See Troubleshooting on page 70 37 Disconnect the tubing from the probe 38 Unscrew the demand flow regulator from the calibration cylinder 39 Use the RANGE YW SHIFT button to place the cursor next to the NORMAL OPERATION menu option then press and release the POWER ENTER RESET button to return to Measuring Mode EAGLE 2 Operator s Manual Infrared Hydrocarbon Calibration 263 Maintenance This section includes a procedure to replace an infrared hydrocarbon sensor Replacing the IR Hydrocarbon Sensor L 2 3 10 11 12 13 14 15 Verify that the EAGLE 2 is off Pla
183. ful the screen above will again appear in the startup sequence e If the unit is due for calibration and CAL PAST DUE ACT is set to EAGLE 2 Operator s Manual Start Up 25 NOTIFICATION ONLY then the following alert screen displays and the buzzer sounds in a double pulsing pattern CALIBRATION DATE IS PAST DUE Press and release the POWER ENTER RESET button to acknowledge the alert and continue with the startup sequence 10 The Date Time Screen appears for a few seconds 9 12 2008 15 00 00 11 If USER STATION ID is turned on see Turning the User Station ID Function On or Off on page 117 the ID Screen appears for a few seconds USER ID MIKE STATION ID PUMP 1 SERIAL NUMBER E2A515 If USER STATION ID is turned off only the serial number is shown 12 If the EAGLE 2 experiences a sensor failure during start up a screen indicating which sensor failed appears and the buzzer sounds a pulsing tone twice per second In the example below the HS sensor has failed FAILED SENSOR S ENTER TO CONTINUE If you wish to continue press and release the POWER ENTER RESET 26 Start Up EAGLE 2 Operator s Manual button to acknowledge the failure The gas reading for the failed sensor will be replaced by XXX Replace the failed sensor as soon as possible 13 The EAGLE 2 is now monitoring for gas in Measuring Mode The Normal Operation Screen appears displaying
184. g 20 Repeat Step 16 through Step 19 for any other channels that need to be changed 21 When you are done adjusting the calibration gas values move the cursor down past the bottom of the screen next to END ADJUST AUTO CALIBRATION VALUES gt END A 22 Press and release the POWER ENTER RESET button The following screen appears DO YOU WANT TO STORE NEW VALUES IN MEMORY FOR FUTURE CALIBRATIONS PRESS YES OR NO 23 If you select YES by pressing and releasing the AIR A YES button the changes that you made will be saved in the EAGLE 2 s memory as the new auto calibration gas values 194 ESM 01 Calibration EAGLE 2 Operator s Manual If you select NO by pressing and releasing the DISPLAY ADJUST NO button the changes you made will be used for any calibrations performed during the current operating session only The EAGLE 2 will delete the changes when the unit is turned off and will load the previous set of auto calibration values when it is turned on again 24 When you make your selection and press the desired button the unit returns to the Calibration Gas Values Screen CAL GAS VALUES CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm ENTER TO BEGIN CAL 25 Press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen with CAL IN PROCESS flashing CAL IN PROCESS CH4 0 LEL OXY 20 9 vol H2S 0 0 ppm CO 0 m ENTER WHEN DONE If you do
185. g See Updating the Alarm Point Settings on page 114 for instructions If you need to change the alarm point setting only to perform a calibration make sure that you change the alarm point setting back to its original value once the calibration has been performed 9 Press and release the POWER ENTER RESET button to save the change The calibration gas value stops flashing 10 Repeat Step 6 through Step 9 for any other channels that need to be changed 11 When you are done adjusting the calibration gas values move the cursor down past the bottom of the screen next to END ADJUST AUTO CALIBRATION VALUES gt END A EAGLE 2 Operator s Manual Calibrating Using the Auto Calibration Method 91 12 Press and release the POWER ENTER RESET button The following screen appears DO YOU WANT TO STORE NEW VALUE IN MEMORY FOR FUTURE CALIBRATIONS PRESS YES OR NO 13 If you select YES by pressing and releasing the AIR A YES button the changes that you made will be saved in the EAGLE 2 s memory as the new auto calibration gas values If you select NO by pressing and releasing the DISPLAY ADJUST NO button the changes you made will be used for any calibrations performed during the current operating session only The EAGLE 2 will delete the changes when the unit is turned off and will load the previous set of auto calibration values when it is turned on again 14 When you make your selection and pres
186. ght corner of the LCD or an up arrow in the upper right corner of the LCD indicates that there are additional menu items accessible by moving the cursor down past the last menu item on the LCD or up past the first menu item on the LCD Setting the Date and Time 1 From the main menu place the cursor next to SET DATE amp TIME 2 Press and release POWER ENTER RESET The date and time will be displayed with the last two digits of the year flashing SET DATE amp TIME 12 21 2009 MM DD YYYY 11 02 12 3 Use AIR A YES or RANGE YW SHIFT to display the desired year 4 Press and release POWER ENTER RESET to save the setting The month setting flashes 5 Repeat Step 3 and Step 4 to enter the month day hours minutes and seconds settings The main menu displays after you enter the seconds setting Setting the Date Format The date can be displayed in two ways month day year factory setting or day month year 1 From the main menu place the cursor next tt SET DATE FORMAT 104 Using Setup Mode EAGLE 2 Operator s Manual 2 Press and release POWER ENTER RESET The Set Date Format screen appears with the current setting flashing SET DATE FORMAT UP DOWN THEN ENTER MM DD YYYY 3 Use AIR A YES or RANGE YW SHIFT to display the desired setting 4 Press and release POWER ENTER RESET to save the setting and return to the main menu Setting the Battery Type This menu item allows
187. h air The low range sensors will display a reading in fresh air Table 23 below indicates typical gas readings in fresh air Table 23 Carbon Dioxide Fresh Air Readings Sensor Approximate Fresh Air Range Reading 0 5 vol 0 04 vol 0 10 000 ppm 400 ppm 0 60 vol 0 0 vol e When you perform a demand zero during start up operation or calibration of a 0 10 000 ppm or 0 5 vol CO instrument you must use the CO scrubber provided with the instrument to remove background CO from the air being sampled See Performing a Demand Zero for Carbon Dioxide Sensors below e Since there is a background of CO in air do not use the AUTO FRESH AIR ADJ feature that can be turned on and off in Setup Mode The factory setting for the feature is off Performing a Demand Zero for Carbon Dioxide Sensors When setting the zero reading the carbon dioxide scrubber mounted to the side of the EAGLE 2 allows you to eliminate carbon dioxide normally found in fresh air To perform a demand zero do the following 1 Remove the black caps from the ends of the carbon dioxide scrubber Be sure to grab the scrubber by the EAGLE 2 fitting so that the fitting and tubing do not come off 2 Connect the carbon dioxide scrubber directly to the EAGLE 2 s inlet fitting 3 Wait one minute for the fresh air sample to flow through the carbon dioxide scrubber then press AIR A YES to set the zero reading 4 Remove the scrubber from the inlet fitting
188. hae ae g bs 2B eG Pee AS Ce bata eS 30 Using Optional Sample Hoses 1 0 0 00 cc eens 30 Using Exhaust Tubing rois ac atoradees eye Vole cadres Mew ws 31 Combustible Gas Detection 0 0 ccc eee ee 31 Monitoring Combustible Gas in the PPM or Volume Ranges 33 Measuring Mode Alarms cic stirs pate be de CAE RE OREER OES ASO ES 35 Alarm Indications 2 6 356555 B88 5g Rhee ha ae oN eed Boke aOR ae 36 Resetting and Silencing Alarms ss css eee eon Pee Me ene 37 Responding to Alarms csc hg See eta NG eb has Pe kee Gs 38 Display Mode Srann eers cenean eb bates oie k E E 8 nt eek ek wes 40 Tips jor Using Display Mode x weaiaiiees pia 28 ee Las eoiess tke 40 Peak Sereen tod amp sa 5 ined tees ai rien b Wea De ok tise sew ek wie ee 41 Battery Voltage Screen tos ns celine a a et ne deta heli e ee a aoa ces ae 42 Gas Display Screen i assar tenet Bsr kb tate dhs ot ata nl ah BS At a Seeds 42 Catalytic LEL Sensor Screen 1a 3 Chee MA 8G ARS Dee BA ee SS 43 Methane Elimination Mode Screen 00000 ees 43 Relative Response Screen 2s aca oo ee koa d DEORE eS BEN 43 STEL SCTEEN a o Gee s ex hee hi GAS un eae BR ey ae a Gece CERO A 46 TWA Sereen ears ot Sects gushes Gus yes ad wae ee eS 46 View Alarm Settings Screen 0 6 0c eens 46 Select User ID Screen 2 04 GEG ae BM tee eth Naan na 48 Select Station ID Screen oi ss eo oa we Oe SHAS EE ee ERS 50 Time in Operation Screen o 33 5 b 4 bee e 8A BERK R Aa e REDE 5 D te lime STEEN aii De
189. he Cal Gas Values Screen If you do want to continue with the calibration proceed to the next step 26 Connect the tubing from the demand flow regulator to the rigid tube on the probe Allow the EAGLE 2 to draw gas for one minute 27 Press and release the POWER ENTER RESET button to set the span adjustment for each channel to the programmed values 260 Infrared Hydrocarbon Calibration EAGLE 2 Operator s Manual 28 If all channels passed calibration the following screen sequence occurs AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END Y CAL GAS VALUE HC 50 LEL ENTER TO BEGIN CAL If any of the sensors cannot be adjusted to the proper value a screen displays that indicates a calibration failure and lists the sensor s that failed to calibrate In the example below the oxygen and H S channels failed calibration The other sensors calibrated normally FAIL SENSOR lt gt lt OXY gt lt H2S gt lt gt ENTER TO CONTINUE The buzzer and alarm LEDs activate in a double pulsing pattern Press and release the POWER ENTER RESET button to reset the alarm and continue to the Calibration Value Screen for the infrared hydrocarbon channel After calibrating the infrared hydrocarbon channel by following the instructions below attempt to calibrate the standard channels again If the failure continues investigate the cause See Troubleshooting on page 70 29
190. he EAGLE 2 stores data how many channels are active and how often the EAGLE 2 is turned on and off The table below illustrates how much data logging time is available for the various interval times It assumes that the unit is setup with four sensors is only turned on once and there are no alarm occurrences See Updating the Data Log Interval Setting on page 120 for instructions on setting the data logging interval time Table 8 Data Logging Capacity 4 gas EAGLE 2 Interval Time Data Logging Time 5 seconds 239 hours 10 days 10 seconds 479 hours 20 days 20 seconds 959 hours 40 days 30 seconds 1439 hours 60 days 1 minute 2879 hours 120 days 3 minutes 8639 hours 360 days 5 minutes 14 399 hours 600 days 10 minutes 28 798 hours 2 000 days For a complete description of the Data Logger Management Program and procedures for downloading data to a computer see the Eagle 2 Data Logger Management Program Operator s Manual 54 Data Logging EAGLE 2 Operator s Manual Chapter 4 Calibration Mode Overview This section describes the EAGLE 2 in Calibration Mode In Calibration Mode you can move through a menu of screens to do the following e Perform a span adjustment on all channels simultaneously using auto calibration e Perform a span adjustment on one channel at a time using single calibration e Perform a fresh air zero adjustment NOTE You can set up the EAGLE 2 to alert you
191. he O sensor is a galvanic type of sensor A membrane behind the openings on the sensor face allows gas to diffuse into the sensor at a rate proportional to the partial pressure of oxygen The oxygen reacts in the sensor and produces a voltage proportional to the concentration of oxygen The voltage is measured by the EAGLE 2 s circuitry converted to a measurement of gas concentration and displayed on the LCD The sensor includes a short cable that terminates in a round 7 position connector It mates with the OXY pins on the main PCB see Main PCB on page 12 CO and HS Sensors The CO and H3S sensors are electrochemical sensors that consist of three precious metal electrodes in a dilute acid electrolyte A gas permeable membrane covers the sensor face and allows gas to diffuse into the electrolyte The gas reacts in the sensor and produces a current proportional to the concentration of the target gas The current is amplified by the EAGLE 2 s circuitry converted to a measurement of gas concentration and displayed on the LCD The CO and H S sensors are physically very similar Except for their markings and wire colors they look almost identical A three position connector at the end of a 2 wire cable from each sensor plugs into a socket on the main PCB The sockets on the main PCB for the CO and H3S sensors are labeled CO and H2S Normally the CO connector plugs into the CO socket and the H2S plugs into the H2S socket However bec
192. he View Alarm Settings Screen gives you the option to view the gas alarm settings for all active channels VIEW ALARM SETTINGS To view the gas alarm settings do the following 1 With the View Alarm Settings Screen displayed press and release the 46 Display Mode EAGLE 2 Operator s Manual AIR A YES button The following screen appears showing Channel 1 alarm points CHANNEL 1 CH4 0 100 LEL LO ALRM 10 LEL HI ALRM 10 LEL LO ALRM 5000 ppm HI ALRM 25000 ppm 2 Ifthe channel description is too long to fit across the screen press and release the POWER ENTER RESET button to view the remainder of the displayed channel s description Press and release the POWER ENTER RESET button again to return to the previous screen CHANNEL 1 CH40 100 LEL LO ALRM 10 LEL HI ALRM 10 LEL LO ALRM 5000 ppm HI ALRM 25000 ppm POWER ENTER RESET CHANNEL 1 METHANE LO ALRM 10 LEL HI ALRM 10 LEL LO ALRM 5000 ppm HI ALRM 25000 ppm 7 3 Ifthe number of alarm settings is too many to display on one screen a down arrow will appear in the lower right corner of the display indicating that there are additional alarm points Press and release the RANGE V SHIFT button to scroll down and display the remainder of the gas alarm settings for the displayed channel and again to return to the previous screen CHANNEL 1 CH4 0 100 LEL LO ALRM 10 LEL HIALRM 10 LEL LO AL
193. he bottom case If you have any difficulty mating the top and bottom cases inspect the placement of the main PCB and the placement of the top case gasket 17 Turn the EAGLE 2 upside down and tighten the three case screws to secure the top case to the bottom case 18 Calibrate the TC channel as described in TC Calibration on page 207 Configuring the TC Gas in Setup Mode The TC channel can be configured for 3 different pre defined gases and 1 user defined gas in the CONFIGURE GASES menu item in Setup Mode To change the gas configuration of the TC channel in Setup Mode do the following WARNING Do not use the user defined gas configuration without consulting RKI Instruments Inc WARNING The EAGLE 2 is not in operation as a gas detector while in Setup Mode Take the EAGLE 2 to a non hazardous location and turn it off if it is on 2 Press and hold AIR A YES and RANGE YW SHIFT then press and hold POWER ENTER RESET When you hear a beep release the buttons 3 The LCD will show the following screen for a few seconds with the S in the lower right corner indicating the unit is entering Setup Mode GAS MONITOR INSTRUMENTS EAGLE2 S EAGLE 2 Operator s Manual Configuring the TC Gas in Setup Mode 217 4 The S will then disappear and the following screen will appear for a few seconds GAS MONITOR INSTRUMENTS EAGLE 2 5 Ifthe unit prompts you for the password enter it
194. hich has the four standard channels LEL oxygen H S CO and a high range PID channel calibrated to and setup for isobutylene IBL There are two ways to calibrate an EAGLE 2 with a PID channel installed The first way described below in Calibrating with a 4 Gas Cylinder and a PID Cylinder on page 143 uses a 4 gas cylinder and an appropriate PID cylinder The second way described below in Calibrating with a 5 Gas Cylinder on page 152 uses a multi gas cylinder that contains the target gas for all installed sensors Follow the set of instructions appropriate for your cylinder configuration Calibrating with a 4 Gas Cylinder and a PID Cylinder 1 See Calibration Supplies and Equipment on page 56 for a description of the necessary calibration supplies In addition to an appropriate multi gas cylinder that is used to calibrate any active standard channels you will also need a cylinder to calibrate the PID channel See Table 17 on page 178 for available cylinders Make sure your calibration cylinder is appropriate for the PID detection range 2 Find a fresh air environment This is an environment free of toxic or combustible gases and of normal oxygen content 20 9 3 While in Measuring Mode press and hold the RANGE W SHIFT button then press the DISPLAY ADJUST NO button and release both buttons 4 If the unit prompts you for the password enter it by using the AIR A YES and RANGE YW SHIFT buttons to select each password
195. ide 0 4000 4000 100 Oe Se a The last choice in each list PID can be defined by the user and loaded in the EAGLE 2 using the Eagle 2 Maintenance Data Loader Program See the Eagle 2 Maintenance Data Loader Program Operator s Manual 3 Use the AIR A YES or RANGE v SHIFT button to move the cursor next to the desired gas 4 Press and release POWER ENTER RESET The PID channel will be configured to the selected gas and the EAGLE 2 will proceed to the STEL Screen This configuration will be in force until either a different gas is selected in Display Mode or the unit is turned off 142 PID Relative Response Feature EAGLE 2 Operator s Manual PID Calibration A PID channel can be calibrated using the auto calibration method or the single calibration method To calibrate a PID channel using the single calibration method see Calibrating Using the Single Calibration Method on page 64 and follow the instructions for calibrating a single channel NOTE The factory setting for the Span Factor menu item in Setup Mode is OFF for an EAGLE 2 with a PID channel So the range of adjustment is not displayed when a single calibration is performed for any channel if a PID sensor is installed If your instrument is a multi channel instrument that includes a high range PID channel RKI Instruments Inc recommends using the auto calibration method for convenience The calibration instructions below show a 5 channel instrument w
196. ing on the front and a 12 pin circular connector on the back A 12 wire cable plugs into the back of the ESM 01 sensor with a circular connector that includes a locking lever The other end of the cable plugs into an ESM 01 sub PCB see description below that is installed on the main PCB The sensor is held in the ESM 01 flow chamber by a bracket on standoffs 180 Description EAGLE 2 Operator s Manual ESM 01 Sub PCB The ESM 01 sub PCB is a circuit board that is installed on the main PCB in one of the 3 sub PCB positions when an ESM 01 sensor is used with the EAGLE 2 The ESM 01 sensor connects to the sub PCB with a 12 position connector The sub PCB plugs into the main PCB and is held in place with a screw flat washer lock washer There are no user serviceable parts on the ESM O01 sub PCB Start Up and Normal Operation For instructions to startup and use an EAGLE 2 that includes an ESM 01 sensor reference Start Up on page 22 Measuring Mode Normal Operation on page 29 and Measuring Mode Alarms on page 35 Follow these instructions keeping the following special considerations in mind e Several of the gases that can be monitored with an ESM O1 are absorbed in the longer EAGLE 2 sample hoses Do not use sample hoses that are longer than the standard 5 foot hose without consulting RKI Instruments Inc e If your EAGLE 2 has more than one ESM 01 sensor installed it is possible that both sensors will respond to so
197. intenance Data Loader Program The Eagle 2 Maintenance Data Loader Program Operator s Manual contains a chart of response factors relative to isobutylene for many gases If the desired gas is not in this chart gas testing must be performed to determine the gas response factor relative to isobutylene See the Eagle 2 Maintenance Data Loader Program Operator s manual for details regarding the gas testing and programming user defined gases into the EAGLE 2 s relative response list Because of normal variation between sensors these relative response factors are typical factors If you use this feature the response to the selected gas will not be as accurate as it would be if you configured and calibrated the PID channel to the target gas For maximum accuracy configure and calibrate the EAGLE 2 s PID channel to the desired target gas PID Sensor Relative Response Screen in Display Mode To use the relative response feature for the PID sensor enter display mode and select the desired gas as described below 1 With the EAGLE 2 in Measuring Mode press and release the DISPLAY ADJUST NO button repeatedly until you arrive at the PID Sensor Relative Response Screen SELECT RELATIVE RESPONSE TO CALIBRATED GAS FOR PID SENSOR 140 PID Relative Response Feature EAGLE 2 Operator s Manual 2 With the PID Sensor Relative Response Screen displayed press and release AIR A YES A list of gases will appear on the screen wi
198. ircuit Boards PCBs iials she 608 GBS ER GAL AEE on Il Alarm LEDS e ORS NE E ee eR Me re ane a PES ey O ae 12 Infrared Communications Port xis 28 eo R ee A ee ERA ORE 12 BULLE curare een Ie Pad elas efecto Gn dA uaa aie arse A a tate rae 13 Battery Case amp Batteries 26 64 inlaw BA apse ash ws scans Dae Phas weds 13 Flow System iniy derio in e a ka edoa e a Mt E hatha 14 Standard ACCESSO IES ae a a E REEE E EEEE A E AE A AUE 15 Shoulder Str p aatan ee Sod dk a bid id Doh de ea ee add Siew Sob au 15 Hoseand PHONE Satie dooe sashes crate ren eaea e e shane een oan ges 15 Optional Accessories susunan u uneneen 16 Rechargeable Ni MH Batteries 0 0 00 000s 17 Battery Chargers en eases aioe Oe E note Sy CE as E S 17 Optional Probes ub 208d wea kWh 2d Ws LEROY 8 eg KOREA 18 External Dilution Fittings ck ocak ete th tare 9 toe tela an cca we gl wiblat de dee 21 Chapter 3 Operation i 9406 160 4 0044 44064594 0 4400s eens 22 ONECTVICW ac cnt tenten d ESTEE Saena EEEE E E ENT a EERE 22 Star UDe atest macs Rete a is A cs Pauls EE estat ga che E te ares A es airy techs eae 22 Turning On the EAGLE 2 ipoe udaa Se ed oa a Gla a A 22 Performing a Demand Zero 1 1 cc cece 27 Turning Off the FAGLE 2 eae ne ae eat an ee ee ee rete ae 27 Using the Battery Charger for Continuous Operation 28 EAGLE 2 Operator s Manual Table of Contents Measuring Mode Normal Operation 0 0 0 e eee eee 29 Monitoring an Areda
199. is factory set to suit most applications Update settings in Setup Mode only if required for your specific application The description of each item below indicates the factory setting for each item Tips for Using Setup Mode When in the main menu the cursor gt flashes in front of a menu item indicating that the item is selected Use the RANGE WY SHIFT button to move the cursor down through the main menu and submenu items and to lower values or change the setting in a specific option Use the AIR A YES button to move the cursor up through the main menu and submenu items and to raise values or change the setting in a specific option A down arrow in the lower right corner or an up arrow in the upper right corner of the LCD indicates that additional menu items can be viewed by pressing and releasing the RANGE YW SHIFT button in the case of the down arrow or the AIR A YES button in the case of the up arrow The example below illustrates a down arrow in the lower right corner gt SET DATE amp TIME SET DATE FORMAT SET BATTERY TYPE CONFIGURE CHANNELS CONFIGURE GASES CATALYTIC UNITS Y Use the POWER ENTER RESET button to enter a selected menu item with the cursor next to it and to enter and save settings during programming An adjustable parameter that is flashing can be adjusted with the AIR A YES and RANGE W SHIFT buttons 102 Tips for Using Setup Mode EAGLE 2 Operator s Manual e Press the DISPLAY ADJUST N
200. is set to ON in Setup Mode a PID Sensor Relative Response Screen appears after the Catalytic Sensor Relative Response Screen If CATALYTIC SENSOR RELATIVE RESPONSE is set to OFF the PID Sensor Relative Response Screen will still appear after the Methane Elimination Mode Screen See PID Relative Response Feature on page 139 for a description of the relative response feature for the PID sensor NOTE If Methane Elimination Mode is enabled the Catalytic Sensor Relative Response screen does not appear EAGLE 2 Operator s Manual Display Mode 45 STEL Screen The STEL Screen displays the short term exposure limit STEL readings for H3S and CO only The STEL reading is the average reading over the last 15 minutes S H2S 0 0ppm TCO Oppm E L TWA Screen The TWA Screen displays the time weighted average TWA readings for H3S and CO only T H2S 0 0ppm WCO Oppm A The TWA reading is the average reading over the last 8 hours If 8 hours have not elapsed since the last time the TWA reading was cleared the average is still calculated over 8 hours The missing time is assigned a 0 value for readings If the lunch break feature is turned off the TWA is cleared when the EAGLE 2 is turned off The lunch break feature enables the EAGLE 2 to remember TWA readings when it is turned off so it can continue them when it is turned on again See Turning On the EAGLE 2 on page 22 View Alarm Settings Screen T
201. it is turned on see Updating LC BH Mode Setting on page 124 Start Up Bar Hole Mode This section explains how to start up the EAGLE 2 in Bar Hole Mode and get it ready for operation Turning On the EAGLE 2 Bar Hole Mode WARNING If one or more sensors other than a catalytic LEL TC or oxygen sensor is installed these sensors will not be active while the EAGLE 2 is in Bar Hole Mode WARNING Gas alarms are not active when the EAGLE 2 is in Bar Hole Mode The following description of the EAGLE 2 start up sequence assumes that the following menu items in Setup Mode are turned on CAL REMINDER and USER STATION ID If either of these items is turned off then the corresponding screen will not appear NOTE In order for BAR HOLE MODE to appear as a selection in the Mode Select Screen in Step 3 below the EAGLE 2 must have both a catalytic LEL sensor and a TC sensor installed In addition both the catalytic LEL and TC channels must be configured for methane CHy in the CONFIGURE GASES menu item in Setup Mode If an EAGLE 2 has only the catalytic LEL sensor installed or if the TC channel is configured for a gas other than CHy then BAR HOLE MODE will not appear as a choice as shown in Step 3 below and 270 Overview EAGLE 2 Operator s Manual NORMAL MODE and LEAK CHECK MODE will be the only choices displayed In this case see Turning On the EAGLE 2 on page 22 or Turning On the EAGLE 2 Leak Check Mode on p
202. just the values on the screen press and hold the RANGE W SHIFT button then press the DISPLAY ADJUST NO button and release both The following screen appears with the cursor next to CH4 ADJUST AUTO CALIBRATION VALUES gt CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm y 16 Place the cursor next to the channel whose gas value you want to change using the AIR A YES and RANGE YW SHIFT buttons 17 Press and release the POWER ENTER RESET button to select the 258 Infrared Hydrocarbon Calibration EAGLE 2 Operator s Manual channel The calibration gas value begins to flash 18 Use the AIR A YES and RANGE YW SHIFT buttons to adjust the calibration gas setting to the desired value NOTE The calibration gas value cannot be set lower than the low alarm setting If the calibration gas value listed on the calibration cylinder is lower than the current low alarm setting enter Setup Mode and change the low alarm setting See Updating the Alarm Point Settings on page 114 for instructions If you need to change the alarm point setting only to perform a calibration make sure that you change the alarm point setting back to its original value once the calibration has been performed 19 Press and release the POWER ENTER RESET button to save the change The calibration gas value stops flashing 20 Repeat Step 16 through Step 19 for any other channels that need to be changed 21 When you are done adjusting the calibra
203. l MEASURING 30 SEC 4 After 15 seconds of sampling if the CH reading is less than 5 vol the CH channel will automatically begin displaying in the units you selected earlier If you selected the methane channel to be displayed in vol then after 15 seconds the reading will remain in vol 5 At the end of the sample period the pump will shut off and the audible alarm will sound then the peak methane reading and the minimum oxygen reading for the sample period will be displayed BAR HOLE MODE CH4 0 vol OXY 20 9 vol ENTER MEASURE ADJUST PURGE A gt tm 6 Ifa high concentration of methane is encountered a fresh air purge can be performed to purge the hose probe and EAGLE 2 of gas before the next bar hole test To perform a purge do the following e Remove the probe from the barhole so the instrument will draw fresh air EAGLE 2 Operator s Manual Bar Hole Testing 277 e Press and release the DISPLAY ADJUST NO button The display will now indicate FRESH AIR PURGE below the oxygen reading and the purge time will begin counting down from 30 seconds in the lower right corner of the display BAR HOLE MODE CH4 0 vol OXY 20 9 vol FRESH AIR PURGE 30 SEC e After 15 seconds of sampling if the CH4 reading is less than 5 vol the CH channel will automatically begin displaying in the units you selected earlier Since performing a fresh air purge draws fresh
204. l applications 222 Parts List EAGLE 2 Operator s Manual Appendix G Infrared Carbon Dioxide Sensors Overview The infrared CO sensors are used to monitor levels of carbon dioxide This appendix describes the EAGLE 2 s infrared CO sensors and includes instructions to use an EAGLE 2 that has one or more infrared CO sensors installed It also includes instructions to replace an infrared CO sensor Table 22 Infrared CO Sensor Specifications Range Increment Alarm 1 Alarm 2 STEL TWA Factory Setting Factory Setting ppm ppm 0 5 00 vol 0 02 vol 0 5 vol 3 0 vol 0 10 000 ppm 25 ppm 5 000 ppm 5 000 Description Table 22 above lists the available infrared CO sensors The infrared CO sensor is a smart sensor that stores sensor parameters including the target gas and detection range So you can change an existing infrared CO channel from one range to another and the EAGLE 2 will automatically load all the sensor parameters and configure the infrared CO channel for the new sensor without the need to enter CONFIGURE CHANNELS or CONFIGURE GASES in Setup Mode See Replacing the IR CO Sensor or Changing Sensor Type on page 236 for instructions to replace or change an infrared CO sensor EAGLE 2 Operator s Manual Overview 223 Top Case Not Shown v CO Sensor Infrared CO2 Sensor Pump Charcoal Filter H2S Sensor 4 _ lt Standard Flow Chamber Gf Ea E a a
205. l for the new sensor Replacing the ESM 01 Sensor or Changing Sensor Type L 2 3 10 11 12 13 Verify that the EAGLE 2 is off Place the EAGLE 2 upside down on a flat surface or hold it upside down Unscrew the three case screws until they disengage from the top case They are captive screws so they will not fall off of the bottom case Turn the EAGLE 2 right side up and carefully lift the top case away from the bottom case Be careful not to lift it so far that it pulls on the main PCB with the cable that connects the top case to the main PCB Lay the top case down next to the bottom case to allow access to the flow system Locate the ESM O1 sensor It has a twelve wire cable with a connector that mates to an ESM O1 sub PCB that is installed on the main PCB and is normally located next to the pump Figure 35 on page 180 shows an ESM 01 sensor in a typical location Unscrew and remove the two screws that hold down the ESM 01 sensor bracket Grasp the sensor firmly and pull it out of the ESM 01 flow chamber Rock it back and forth gently if necessary to pull it out Take care not to pull the cable from the sub PCB Rotate the locking lever counterclockwise on the cable connector that mates to the ESM O1 sensor to unlock it Unplug the old ESM 01 sensor from the cable Connect the new ESM O1 sensor to the sensor cable and rotate the locking lever clockwise to lock the connector Insert the sensor into the ESM 0
206. l give an indication at start up if calibration is past due and require the user to decide whether to perform a calibration or continue and use the EAGLE 2 without calibrating Press and release DISPLAY ADJUST NO to continue without calibrating or AIR A YES to perform a calibration With CAL PAST DUE ACT set to MUST CALIBRATE if the unit is due for calibration the EAGLE 2 will give an indication at start up that calibration is past due and prompt you to press and release POWER ENTER RESET to enter Calibration Mode and perform a calibration Using any other button will have no effect A successful calibration must be performed in order to use the instrument With CAL PAST DUE ACT set to NOTIFICATION ONLY the EAGLE 2 will give an indication at startup that calibration is past due You must press and release POWER ENTER RESET to acknowledge the indication and proceed with the startup sequence 1 From the main menu place the cursor next to CAL PAST DUE ACT 2 Press and release POWER ENTER RESET The Calibration Past Due Action Screen appears CALIBRATION PAST DUE ACTION UP DOWN THEN ENTER CONFIRM TO CAL Use AIR A YES or RANGE YW SHIFT to display the desired setting Press and release POWER ENTER RESET to save the setting and return to the main menu Updating the Calibration Interval This setting defines the amount of time between calibrations The time can be set in 1 day increments The minimum setting is 1 day an
207. lacement of the main PCB and the placement of the top case gasket 16 Turn the EAGLE 2 upside down and tighten the three case screws to secure the top case to the bottom case 17 Calibrate the PID channel as described in PID Calibration on page 143 Configuring the PID Gas in Setup Mode The standard PID channel is configured for and calibrated to isobutylene If calibration to a different gas is required for an application the PID channel can also be configured for other gases in the CONFIGURE GASES menu item in Setup Mode To change the gas configuration of the PID channel in Setup Mode do the following WARNING The EAGLE 2 is not in operation as a gas detector while in Setup Mode Take the EAGLE 2 to a non hazardous location and turn it off if it is on 2 Press and hold AIR A YES and RANGE VW SHIFT then press and hold POWER ENTER RESET When you hear a beep release the buttons 3 The LCD will show the following screen for a few seconds with the S in the lower right corner indicating the unit is entering Setup Mode GAS MONITOR INSTRUMENTS EAGLE2 S 172 Configuring the PID Gas in Setup Mode EAGLE 2 Operator s Manual 4 The S will then disappear and the following screen will appear for a few seconds GAS MONITOR INSTRUMENTS EAGLE 2 If the unit prompts you for the password enter it by using AIR A YES and RANGE YW SHIFT to select each password number and then p
208. lease the POWER ENTER RESET button The following screen appears PERFORM AIR ADJUST 10 Press and release the AIR A YES button to continue If you do not want to continue press the DISPLAY ADJUST NO button and the unit will return to the Calibration Mode Screen 228 Infrared CO Calibration EAGLE 2 Operator s Manual 11 The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds then indicate that the operation is complete before returning to the Calibration Mode Screen ADJUSTING ZERO RELEASE AIR BUTTON ADJUSTING ZERO COMPLETE CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION gt PERFORM AIR ADJUST NORMAL OPERATION 12 Remove the CO scrubber from the EAGLE 2 inlet fitting 13 Put the caps back on the scrubber 14 Install the demand flow regulator onto the multi gas calibration cylinder 15 Connect the sample tubing to the demand flow regulator 16 Install the probe on the EAGLE 2 inlet fitting Make sure the probe is complete with internal O ring and membrane and that the two halves of the probe are tightened firmly together to avoid leaks that can affect the calibration See Figure 19 Replacing the Particle Filter and Hydrophobic Filter Disk on page 76 for an illustration of the internal parts of the probe EAGLE 2 Operator s Manual Infrared CO Calibration 229 17 Move the cursor next to tte AUTO CALIBRATION menu
209. linder If all concentrations match go to Step 15 If one or more concentrations do not match continue with Step 5 If you do not want to continue with the calibration press and release the DISPLAY ADJUST NO button to return to the Calibration Mode Screen NOTE The RKI 4 gas cylinder typically contains 12 O by volume When using the auto calibration method be sure to set the OXY auto calibration gas value to agree with the concentration listed on the cylinder s label not zero 90 Calibrating Using the Auto Calibration Method EAGLE 2 Operator s Manual 5 To adjust the values on the screen hold down the RANGE VW SHIFT button then press the DISPLAY ADJUST NO button and release both The following screen appears with the cursor next to CH4 ADJUST AUTO CALIBRATION VALUES gt CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm y 6 Place the cursor next to the channel whose gas value you want to change using the AIR A YES and RANGE VW SHIFT buttons 7 Press and release the POWER ENTER RESET button to select the channel The calibration gas value begins to flash 8 Use the AIR A YES and RANGE YW SHIFT buttons to adjust the calibration gas setting to the desired value NOTE The calibration gas value cannot be set lower than the low alarm setting If the calibration gas value listed on the calibration cylinder is lower than the current low alarm setting enter Setup Mode and change the low alarm settin
210. low the gas to dispense until the gas collection bag is a little over half full EAGLE 2 Operator s Manual Calibrating Using the Single Calibration Method 97 98 12 13 14 15 Turn the knob clockwise to stop the gas flow clamp down the hose clamp and remove the regulator or dispensing valve from the cylinder Disconnect the tubing from the regulator or dispensing valve Open the hose clamp on the gas bag tubing Connect the tubing from the gas bag to the rigid tube on the probe Allow the gas to flow for one minute NOTE If you are calibrating the catalytic combustible channel with a gas 16 17 concentration of 1000 ppm or lower you must use a 24 inch humidifier tube to connect the gas bag to the rigid tube on the probe Press and release the POWER ENTER RESET button to make the span adjustment When the span adjustment is made the EAGLE 2 calculates the range of adjustment minimum and maximum it could have made based on its response level to the applied gas This calculated range is independent of the calibration gas value that was entered in Step 8 and Step 9 above The adjustment range is included on the result screen to indicate the condition of the sensor If the calibration gas value is in the adjustment range the span adjustment will pass If the calibration gas value is out of the adjustment range the span adjustment will fail 18 If the span adjustment is successful the following screens
211. lter endcap and pull it off of the charcoal filter 8 Grasp the top of the charcoal filter firmly and pull it out of the flow chamber Insert the replacement charcoal filter into the filter position in EAGLE 2 Operator s Manual Replacing the Charcoal Filter 79 10 11 2 the flow chamber and push it in until it bottoms out Insert the charcoal filter endcap into the end of the charcoal filter and push it in until it bottoms out Confirm that the main PCB is seated in its slots and that its bottom edge is resting on the bottom of the bottom case If the main PCB is not seated properly then it may be damaged when the top case is re installed Make sure that the top case gasket is fully seated in its groove and carefully put the top case back on the bottom case If you have any difficulty mating the top and bottom cases inspect the placement of the main PCB and the placement of the top case gasket Turn the EAGLE 2 upside down and tighten the three case screws to secure the top case to the bottom case Checking the Combustible Gas Sensor s Condition If you suspect that the combustible sensor has been contaminated or may be reaching the end of its operational life do the following to confirm it is still operating properly L Perform a calibration using single calibration as described in Calibrating Using the Single Calibration Method on page 64 When you perform the span adjustment note the adjustment
212. lytic combustible channel s units by pressing and releasing RANGE W SHIFT CH4 O LEL RANGE CH4 0 00vol OXY 20 9vol hA OXY 20 9vol H2S 0 0ppm H2S 0 0ppm CO Oppm gt CO Oppm RANGE SHIFT CH4 Oppm OXY 20 9vol H2S 0 0ppm CO Oppm Monitoring Combustible Gas in the PPM Range There are special considerations that must be taken into account when monitoring combustible gas in the ppm range with the catalytic combustible channel Because of the high sensitivity in the lower part of the ppm range the catalytic combustible channel needs more time than the warm up period to stabilize after the EAGLE 2 is turned on if it is going to be used for ppm level detection The reading increments in the ppm range are smallest in the lower part of the range and increase as the reading increases as follows e 5 ppm increments from 0 ppm to 200 ppm e 10 ppm increments from 200 ppm to 1 000 ppm e 50 ppm increments from 1 000 ppm to 10 000 ppm e 250 ppm increments from 10 000 ppm to 50 000 ppm If the catalytic combustible channel is configured for the user defined gas in Setup Mode see Configuring the Combustible Gas on page 108 the ppm ratio defined for the gas must be considered For example if the ppm ratio is set higher than 50 000 ppm then when the display units are set as ppm the reading will not go above 50 000 ppm which is equivalent to 33 LEL and 5 Yvolume So if the gas reading is higher th
213. me of the same gases at varying levels Make sure you understand any issues like this that may exist in your particular instrument e The SO and HCN ESM 01 sensors include a replaceable H S scrubber disk inside the sensor face The SO and HCN sensors respond to HS so the H2S scrubber disk removes H S from the sample to avoid false SO and HCN readings See Replacing the H S Scrubber in the SO and HCN Sensors on page 200 for instructions to replace the H S scrubber ESM 01 Calibration An ESM 01 channel can be calibrated using the auto calibration method or the single calibration method To calibrate an ESM O1 channel using the single calibration method see Calibrating Using the Single Calibration Method on page 64 and follow the instructions for calibrating a single channel If your instrument is a multi channel instrument that includes one or more ESM 01 channels RKI Instruments Inc recommends using the auto calibration method for convenience There are two ways to calibrate an EAGLE 2 with an ESM 01 channel installed The first way described below in Calibrating with a 4 Gas Cylinder and an ESM 01 Cylinder on page 182 uses a 4 gas cylinder and an IBL cylinder The second way EAGLE 2 Operator s Manual Start Up and Normal Operation 181 described below in Calibrating with a 5 Gas Cylinder on page 191 uses a multi gas cylinder that contains the target gas for all installed sensors Follow the set of inst
214. message will appear indicating that the gas configuration for this sensor is automatically detected by the EAGLE 2 and that manual configuration is not possible Both of these sensor types have the gas configuration stored in memory onboard the sensor ESM amp IR SENSOR ARE AUTOMATICALLY DETECTED MANUAL SELECTION IS NOT POSSIBLE See the appropriate optional sensor appendix for a complete description of configuring the gas for that type of sensor EAGLE 2 Operator s Manual Channel Setup amp Sub PCBs 135 Appendix D PID Sensors Overview The PID photo ionization detector sensors are used for applications where high sensitivity is needed to monitor ppm levels of VOCs volatile organic compounds This appendix describes the EAGLE 2 s PID sensors and includes instructions to use an EAGLE 2 that has a PID sensor installed It also includes instructions to maintain and replace a PID sensor Table 13 EAGLE 2 PID Sensor Specifications Target Gas Sensor Detection Reading Alarm 1 Alarm 2 Type Range Increment Factory Setting Factory Setting VOCs 0 50 00 ppm 0 02 ppm 5 00 ppm 10 00 ppm Isobutylene Calibration High Range 0 2000 ppm 1 ppm 400 ppm 1000 ppm Standard Description Two types of PID sensors can be used with the EAGLE 2 a low range high sensitivity sensor and a high range low sensitivity sensor see Table 13 for specifications The PID sensor is installed in a single sensor flo
215. modes low alarm decreasing and high alarm increasing FALLING AND RISING low and high alarm EAGLE 2 Operator s Manual Using Setup Mode 115 decreasing BOTH FALLING low and high alarm increasing BOTH RISING The factory setting is FALLING AND RISING e In order to appropriately set the calibration gas value during a calibration the low alarm setting must be lower than the desired calibration value 7 If you want to continue with the change press and release POWER ENTER RESET to accept the setting If you want to exit this screen without saving any change to the alarms press and release DISPLAY ADJUST NO until you return to the Change Alarm Point Settings Screen 8 Repeat Step 4 Step 7 for any additional changes you want to make 9 When you are done making changes use RANGE VW SHIFT to move the cursor next to END 10 Press and release POWER ENTER RESET to save the new settings and return to the Change Alarm Point Settings Screen 11 Use RANGE VW SHIFT to move the cursor next to END 12 Press and release POWER ENTER RESET to return to the main menu Updating the Alarm Latching Setting With ALARM LATCHING set to LATCHING factory setting the EAGLE 2 remains in alarm condition until the alarm condition passes and the POWER ENTER RESET button is pressed With ALARM LATCHING set to SELF RESET the EAGLE 2 automatically resets an alarm when the alarm condition passes 1 From the main menu place the cursor next
216. monitor and our warranty is limited to replacement of parts or our complete goods EAGLE 2 Operator s Manual Warranty 299
217. n and the unit will return to the Gas Display Screen Catalytic LEL Sensor Screen This screen appears only when either a TC sensor or an infrared combustible sensor is installed in an EAGLE 2 along with a catalytic combustible LEL sensor See Appendix F TC Sensors Appendix H Infrared Methane Sensor or Appendix I Infrared Hydrocarbon Sensor for a description of this screen and instructions to use it Methane Elimination Mode Screen This screen displays only if the EAGLE 2 catalytic combustible channel is setup for one of the gases in the CONFIGURE GASES menu item in Setup Mode that supports methane elimination See Configuring the Combustible Gas on page 108 METHANE ELIMINATION MODE DISABLED The standard setup for methane CH does not support methane elimination When applicable use this screen to enable and disable the methane elimination feature See Appendix J Methane Elimination Mode for more discussion of the methane elimination feature 1 With the Methane Elimination Mode Screen displayed press and release the AIR A YES or RANGE v SHIFT button to toggle to the desired setting ENABLED or DISABLED 2 Press and release the POWER ENTER RESET button The unit will save the setting and proceed to the next menu item If you changed the setting and do not want to save the change press and release the DISPLAY ADJUST NO button to continue to the next menu item without saving
218. n listed on the cylinder s label For oxygen free samples 100 nitrogen for example set the oxygen calibration value to 0 0 15 After the last channel is calibrated disconnect the calibration tubing from the probe then unscrew the demand flow regulator from the calibration cylinder 16 With the Select Sensor Screen displayed place the cursor next to ESCAPE using the AIR A YES button SELECT SENSOR TO CALIBRATE gt ESCAPE CH4 OXY H2S CO 17 Press and release the POWER ENTER RESET button to return to the Calibration Mode Screen 18 Use the RANGE YW SHIFT button to place the cursor next to the NORMAL OPERATION menu item then press and release the POWER ENTER RESET button to return to Measuring Mode EAGLE 2 Operator s Manual Calibrating Using the Single Calibration Method 69 Chapter 5 Maintenance Overview This chapter describes troubleshooting procedures for the EAGLE 2 It also includes procedures for replacing and recharging the batteries and replacing various consumable parts WARNING RKI Instruments Inc recommends that service calibration and repair of RKI instruments be performed by personnel properly trained for this work Replacing sensors and other parts with original equipment does not affect the intrinsic safety of the instrument Troubleshooting The troubleshooting table describes error messages symptoms probable causes and recommended action for problems you may encoun
219. n method be sure to set the OXY auto calibration value to agree with the concentration listed on the cylinder s label not zero 15 To adjust the values on the screen press and hold the RANGE W SHIFT button then press the DISPLAY ADJUST NO button and release both The following screen appears with the cursor next to CH4 ADJUST AUTO CALIBRATION VALUES gt CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm y EAGLE 2 Operator s Manual ESM 01 Calibration 193 16 Place the cursor next to the channel whose gas value you want to change using the AIR A YES and RANGE VW SHIFT buttons 17 Press and release the POWER ENTER RESET button to select the channel The calibration gas value begins to flash 18 Use the AIR A YES and RANGE YW SHIFT buttons to adjust the calibration gas setting to the desired value NOTE The calibration gas value cannot be set lower than the low alarm setting If the calibration gas value listed on the calibration cylinder is lower than the current low alarm setting enter Setup Mode and change the low alarm setting See Updating the Alarm Point Settings on page 114 for instructions If you need to change the alarm point setting only to perform a calibration make sure that you change the alarm point setting back to its original value once the calibration has been performed 19 Press and release the POWER ENTER RESET button to save the change The calibration gas value stops flashin
220. n the screen and scroll through the available user IDs to find the desired user ID 3 When the desired user ID is displayed place the cursor next to it press and release the POWER ENTER RESET button NOTE To exit the selection screen without saving a change press and release the DISPLAY ADJUST NO button You will return to the Select User ID screen without saving the user ID change 4 The unit will save the selected user ID as the current one and proceed to the Select Station ID Screen EAGLE 2 Operator s Manual Display Mode 49 Select Station ID Screen This screen displays only if USER STATION ID in the Setup Mode menu is set to ON see Turning the User Station ID Function On or Off on page 117 Use this screen to select a station ID from the station ID list in the EAGLE 2 s memory The current station ID is displayed A station ID can be up to 16 characters long The EAGLE 2 can store up to 128 station IDs SELECT STATION ID CURRENT STATION Pump 2 The station ID provides a way to identify a location where monitoring was done during a data logging session If the station ID is changed during an operating session a new data session is initiated with the new station ID attached to it This allows you to change the station ID during operation and have each station ID that was used during an operating session saved for the corresponding data See the Eagle 2 Data Logger Management Program Operator s M
221. nd release the DISPLAY ADJUST 40 Display Mode EAGLE 2 Operator s Manual NO button e To enter an item when a question is asked press and release the AIR A YES button e To change a flashing parameter use either the AIR A YES button or RANGE WY SHIFT button NOTE Each screen displays for 20 seconds If you do not press a button within 20 seconds the EAGLE 2 automatically returns to Measuring Mode Peak Screen The peak screen displays the highest lowest for oxygen concentrations detected since the EAGLE 2 was turned on Peak readings are stored in the EAGLE 2 s memory until a higher level is detected lower for oxygen the peak reading is cleared or the EAGLE 2 is turned off The lunch break feature enables the EAGLE 2 to save peak readings when it is turned off so it can continue them when it is turned on again See Turning On the EAGLE 2 on page 22 CH4 O LEL P OXY 20 9vol E H2S 0 0ppm ACO Oppm K To clear the peak readings do the following 1 With the Peak Screen displayed press and release the POWER ENTER RESET button The following screen will appear CLEAR PEAK READINGS 2 Press and release the AIR A YES button The peak readings will be reset and the unit will return to the Peak Screen If you do not want to clear the peak readings press and release the DISPLAY ADJUST NO button and the unit will return to the Peak Screen without clearing the peak readings EAG
222. ne adjusting the calibration gas values move the cursor down past the bottom of the screen next to END ADJUST AUTO CALIBRATION VALUES gt END A 22 Press and release the POWER ENTER RESET button The following screen appears DO YOU WANT TO STORE NEW VALUE S IN MEMORY FOR FUTURE CALIBRATIONS PRESS YES OR NO 23 If you select YES by pressing and releasing the AIR A YES button the changes that you made will be saved in the EAGLE 2 s memory as the new auto calibration gas values If you select NO by pressing and releasing the DISPLAY ADJUST NO button the changes you made will be used for any calibrations performed during the current operating session only The EAGLE 2 will delete the EAGLE 2 Operator s Manual PID Calibration 155 changes when the unit is turned off and will load the previous set of auto calibration values when it is turned on again 24 When you make your selection and press the desired button the unit returns to the Calibration Gas Values Screen CAL GAS VALUES CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm ENTER TO BEGIN CAL 25 Press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen with CAL IN PROCESS flashing CAL IN PROCESS CH4 0 LEL OXY 20 9 vol H2S 0 0 ppm CO 0 m ENTER WHEN DONE If you do not want to proceed with the calibration press and release the DISPLAY ADJUST NO button to return to the Cal
223. next menu item 14 Continue to press the DISPLAY button until the Measuring Mode screen appears HEX O LEL OXY 20 9vol H2S 0 0ppm CO Oppm The ME is no longer next to the catalytic combustible channel reading 268 Monitoring in Methane Elimination Mode EAGLE 2 Operator s Manual indicating that the unit is measuring in full response mode 15 Allow 2 minutes for the combustible sensor to stabilize before monitoring the target gas 16 Perform a demand zero See Performing a Demand Zero on page 27 for instructions 17 Monitor for the target gas NOTE The EAGLE 2 retains the methane elimination mode setting when it is turned off The methane elimination mode setting ENABLED or DISABLED will remain in effect until it is changed in Display Mode So if methane elimination mode is enabled when you turn off the EAGLE 2 it will remain enabled when you turn the EAGLE 2 on again Calibration The EAGLE 2 stores calibration data for the instrument both in methane elimination mode and in full response mode When using the instrument for applications where methane elimination mode is used it is most common to calibrate to hexane If you are planning to use the EAGLE 2 in methane elimination mode RKI Instruments Inc recommends that you setup and calibrate the instrument to hexane in both full response and methane elimination mode unless your application requires a different setup and calibration See
224. ng Using the Single Calibration Method EAGLE 2 Operator s Manual 4 The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds then indicate that the operation is complete before returning to the Calibration Mode Screen ADJUSTING ZERO RELEASE AIR BUTTON Y ADJUSTING ZERO COMPLETE Y CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION gt PERFORM AIR ADJUST NORMAL OPERATION Performing a Span Adjustment in Single Calibration 1 Install the demand flow regulator onto the calibration cylinder 2 Connect the sample tubing to the demand flow regulator 3 Install the probe on the EAGLE 2 inlet fitting Make sure the probe is complete with internal O ring and membrane and that the two halves of the probe are tightened firmly together to avoid leaks that can affect the calibration See Figure 19 Replacing the Particle Filter and Hydrophobic Filter Disk on page 76 for an illustration of the internal parts of the probe EAGLE 2 Operator s Manual Calibrating Using the Single Calibration Method 65 4 Move the cursor next to the SINGLE CALIBRATION menu item by using the AIR A YES button CALIBRATION MODE AUTO CALIBRATION gt SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 5 Press and release the POWER ENTER RESET button The Select Sensor Screen appears with the cursor flashing SELECT SENSOR TO CALIBRATE gt ESCAPE CH4 OXY H2S
225. nge 2 Find a fresh air environment This is an environment free of toxic or combustible gases and of normal oxygen content 20 9 3 While in Measuring Mode press and hold the RANGE W SHIFT button then press the DISPLAY ADJUST NO button and release both buttons 4 Ifthe unit prompts you for the password enter it by using the AIR A YES and RANGE YW SHIFT buttons to select each password number and then pressing and releasing POWER ENTER RESET to enter the number and move on to the next one EAGLE 2 Operator s Manual Infrared CO Calibration 227 5 The Calibration Mode Screen displays with the cursor next to AUTO CALIBRATION CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 6 Move the cursor to the PERFORM AIR ADJUST menu item by using the RANGE W SHIFT button CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION gt PERFORM AIR ADJUST NORMAL OPERATION 7 Remove the black caps from the ends of the carbon dioxide scrubber Be sure to grab the scrubber by the EAGLE 2 fitting so that the fitting and tubing do not come off 8 Attach the CO scrubber to the inlet fitting of the EAGLE 2 and allow the instrument to draw gas for 1 minute before performing an air adjust NOTE Attaching the CO scrubber to the inlet fitting eliminates the background CO found in fresh air and allows the EAGLE 2 to obtain an accurate zero reading 9 Press and re
226. nsor bracket and tighten both bracket screws firmly 13 To install the HS and CO sensors do the following e Insert the sensor face down into the sensor chamber in the flow chamber e Push the sensor in until it bottoms out e Route the sensor cable the same way the old sensor cable was routed and connect it to the main PCB Make sure to support the main PCB when making the connection 14 Confirm that the main PCB is seated in its slots and that its bottom edge is resting on the bottom of the bottom case If the main PCB is not seated properly then it may be damaged when the top case is re installed 15 Make sure that the top case gasket is fully seated in its groove and carefully put the top case back on the bottom case If you have any difficulty mating the top and bottom cases inspect the placement of the main PCB and the placement of the top case gasket 16 Turn the EAGLE 2 upside down and tighten the three case screws to secure the top case to the bottom case 17 Calibrate the new sensors as described in Chapter 4 Calibration Mode on page 55 EAGLE 2 Operator s Manual Replacing a Sensor 83 General Parts List Table 11 lists part numbers for the EAGLE 2 s replacement parts and accessories Table 11 General Parts List 33 0156RK 01 Filter element hydrophobic disk for standard 80 0131RK 10 probe pack of 5 33 0173RK Internal hydrophobic filter 33 1200RK Particle filter for standard 80 0131RK 10 p
227. nt 20 9 While in Measuring Mode press and hold the RANGE W SHIFT button then press the DISPLAY ADJUST NO button and release both buttons If the unit prompts you for the password enter it by using the AIR A YES and RANGE VY SHIFT buttons to select each password number and then pressing and releasing POWER ENTER RESET to enter the number and move on to the next one The Calibration Mode Screen displays with the cursor next to AUTO CALIBRATION CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION Move the cursor to the PERFORM AIR ADJUST menu item by using the RANGE W SHIFT button CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION gt PERFORM AIR ADJUST NORMAL OPERATION EAGLE 2 Operator s Manual ESM 01 Calibration 191 7 Press and release the POWER ENTER RESET button The following screen appears PERFORM AIR ADJUST 8 Press and release the AIR A YES button to continue If you do not want to continue press the DISPLAY ADJUST NO button and the unit will return to the Calibration Mode Screen 9 The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds then indicate that the operation is complete before returning to the Calibration Mode Screen ADJUSTING ZERO RELEASE AIR BUTTON ADJUSTING ZERO COMPLETE CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION g
228. nterface will now be in the newly selected language NOTE If you select a language other than English a prompt will appear during startup that allows you to change the language back to English if desired Exiting Setup Mode 1 From the main menu place the cursor in front of NORMAL OPERATION at the bottom of the menu 2 Press and release POWER ENTER RESET 130 Using Setup Mode EAGLE 2 Operator s Manual 3 A screen appears that asks if you want to save the changes you have made SAVE ALL CHANGES IN MEMORY NOTE If you entered Setup Mode and did not make any changes the above screen will still appear In this case press and release DISPLAY ADJUST NO to proceed to exit Setup Mode and begin the EAGLE 2 s startup sequence 4 If you do not want to save the changes press and release DISPLAY ADJUST NO The unit will begin its startup sequence without saving the changes If you do want to save the changes press and release AIR A YES and continue with the next step 5 A confirmation screen appears asking if you are sure you want to save the changes ARE YOU SURE YOU WANT TO SAVE ALL CHANGES IN MEMORY 6 Ifyou want to save the changes press and release AIR A YES to save the changes A screen will appear for a few seconds indicating that the changes have been saved and the unit will begin its start up sequence If you do not want to save the changes press and release DISPLAY ADJUST NO
229. nto the EAGLE 2 s microprocessor moves the cursor on the LCD up the screen increases the value of a parameter available for adjustment scrolls through parameter options RANGE SHIFT changes the detection units of the combustible gas channel when Catalytic Units is set to CHANGE OK in Setup Mode silences and resets audible alarm if Alarm Latching is set to Latching and Alarm Silence is set to ON enters instructions into the EAGLE 2 s microprocessor moves the cursor on the LCD down the screen decreases the value of a parameter available for adjustment scrolls through parameter options 10 Instrument Description EAGLE 2 Operator s Manual Printed Circuit Boards PCBs Display PCB Main PCB Figure 4 EAGLE 2 PCBs The EAGLE 2 s PCBs analyze record control store and display the information collected The main PCB and I S barrier PCB are located in the bottom case The display PCB is located in the top case The display PCB and I S barrier PCB are not user serviceable and are not involved in any user performed maintenance The main PCB is not user serviceable but it is involved in the replacement of sensors so it is described below EAGLE 2 Operator s Manual Instrument Description 11 Display connector H2S Connector H2S ya CO Connector CO HH i LEL Connector HC a a a y L S Barrier Connector ne pe Se seer Oxygen Connector Pins OXY
230. nual General Parts List 85 Appendix A Calibrating with a Sample Bag Overview The EAGLE 2 can be calibrated with a gas bag calibration kit instead of a demand flow regulator kit This appendix describes how to use a sample bag calibration kit to calibrate the EAGLE 2 A parts list at the end of this appendix lists spare parts for the calibration kit Calibration Supplies and Equipment Calibration Tubing Regulator A Gas Ba iz Calibration Cylinder 9 y j O Hose Clamp Figure 23 Gas Bag Calibration Kit 86 Overview EAGLE 2 Operator s Manual To calibrate the EAGLE 2 you will need e Known calibrating samples of the gases being detected The combustible and toxic gas samples should have concentrations between 10 and 50 of the full scale value For example if you are calibrating the catalytic combustible gas channel your calibration cylinder should have a combustible gas concentration between 10 LEL and 50 LEL An oxygen free source such as 100 nitrogen is recommended for setting the oxygen zero NOTE The catalytic combustible channel can be set up for and calibrated to a number of different combustible gases See Configuring the Combustible Gas on page 108 for instructions Be sure that you are using an appropriate calibration cylinder for the target gas of the catalytic combustible channel CAUTION When using auto calibration with the standard 4 gas EAGLE 2 although
231. number and then pressing and releasing POWER ENTER RESET to enter the number and move on to the next one EAGLE 2 Operator s Manual PID Calibration 143 5 The Calibration Mode Screen displays with the cursor next to AUTO CALIBRATION CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 6 Move the cursor to the PERFORM AIR ADJUST menu item by using the RANGE W SHIFT button CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION gt PERFORM AIR ADJUST NORMAL OPERATION 7 Press and release the POWER ENTER RESET button The following screen appears PERFORM AIR ADJUST 8 Press and release the AIR A YES button to continue If you do not want to continue press the DISPLAY ADJUST NO button and the unit will return to the Calibration Mode Screen 144 PID Calibration EAGLE 2 Operator s Manual 9 The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds then indicate that the operation is complete before returning to the Calibration Mode Screen ADJUSTING ZERO RELEASE AIR BUTTON ADJUSTING ZERO COMPLETE CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION gt PERFORM AIR ADJUST NORMAL OPERATION 10 Install the demand flow regulator onto the multi gas calibration cylinder 11 Connect the sample tubing to the demand flow regulator 12 Install the probe on
232. o the proper value a screen displays that indicates a calibration failure and indicates that the PID sensor failed to calibrate FAILED SENSOR ENTER TO CONTINUE lt IBL gt The buzzer and alarm LEDs activate in a double pulsing pattern Press and release the POWER ENTER RESET button to reset the alarm and continue to the Calibration Mode Screen Attempt to calibrate the PID channel again If the failure continues investigate the cause See Troubleshooting on page 70 34 Disconnect the tubing from the probe 35 Unscrew the demand flow regulator from the calibration cylinder 36 Use the RANGE V SHIFT button to place the cursor next to the NORMAL OPERATION menu option then press and release the POWER ENTER RESET button to return to Measuring Mode EAGLE 2 Operator s Manual PID Calibration 159 Maintenance The PID sensor includes user serviceable parts They are the lamp and the electrode stack The following sections include troubleshooting instructions as well as procedures for cleaning the lamp replacing the lamp replacing the electrode stack and replacing the PID sensor Troubleshooting The troubleshooting table describes symptoms probable causes and recommended action for problems you may encounter with the PID sensor Symptoms Probable Causes Recommended Action PID channel doesn t exit its warm up state and the reading continues to scroll through 200 100 and 0 for a
233. ompts you for the password enter it by using the AIR A YES and RANGE VY SHIFT buttons to select each password number and then pressing and releasing POWER ENTER RESET to enter the number and move on to the next one The Calibration Mode Screen displays with the cursor next to AUTO CALIBRATION CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION Move the cursor to the PERFORM AIR ADJUST menu item by using the RANGE W SHIFT button CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION gt PERFORM AIR ADJUST NORMAL OPERATION 242 Infrared Methane Calibration EAGLE 2 Operator s Manual 7 Press and release the POWER ENTER RESET button The following screen appears PERFORM AIR ADJUST 8 Press and release the AIR A YES button to continue If you do not want to continue press the DISPLAY ADJUST NO button and the unit will return to the Calibration Mode Screen 9 The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds then indicate that the operation is complete before returning to the Calibration Mode Screen ADJUSTING ZERO RELEASE AIR BUTTON ADJUSTING ZERO COMPLETE CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION gt PERFORM AIR ADJUST NORMAL OPERATION 10 Install the demand flow regulator onto the multi gas calibration cylinder 11 Connect the sample
234. on Inert Mode is factory activated for instruments that require it The instrument can still be used in Normal Mode for other applications It is recommended that either an IR CHy or an IR HC sensor be installed in a unit that is used to monitor combustible gas in Inert Mode since it does not require oxygen to work properly The catalytic LEL sensor does not operate at oxygen concentrations below 10 volume If it is necessary to use the catalytic LEL sensor in Inert Mode a dilution fitting must be installed Installing a dilution fitting will affect the oxygen reading since you re introducing oxygen into the sample Alarms The oxygen channel alarm points in Inert Mode are different from those in Normal Mode All other alarm point settings remain unchanged Since the application for Inert Mode is to detect a rising oxygen level in purged environments both oxygen alarms are set to rising The factory set alarm point is 5 0 volume for the low alarm and 10 0 volume for the high alarm These alarm points are user adjustable in Setup Mode See Updating the Alarm Point Settings on page 114 for instructions to set the alarm points for channels other than oxygen or to set the Normal Mode oxygen alarm points Below are instructions to set the Inert Mode oxygen alarms 1 Take the EAGLE 2 to a non hazardous location and turn it off if it is on 2 Press and hold the AIR A YES or RANGE VY SHIFT buttons then press and hold the POWER E
235. on 157 29 If you want to change the PID channel s calibration gas value follow Step 5 Step 24 above beginning with the PID Calibration Gas Value Screen below instead of the standard channel Calibration Gas Value Screen CAL GAS VALUE IBL 100 ppm ENTER TO BEGIN CAL 30 With the PID Calibration Gas Value Screen displayed press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen for the PID channel with CAL IN PROCESS flashing CAL IN PROCESS IBL 94 ppm ENTER WHEN DONE If you do not want to proceed with the calibration press and release the DISPLAY ADJUST NO button to return to the PID Cal Gas Values Screen If you do want to continue with the calibration proceed to the next step 31 Since calibration gas has already been flowing to the PID sensor while the other channels were being calibrated the PID sensor reading should be stable and ready for adjustment 32 Press and release the POWER ENTER RESET button to set the span adjustment for the PID channel to the programmed value 158 PID Calibration EAGLE 2 Operator s Manual 33 If the PID channel passed calibration the following screen sequence occurs AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END Y CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION If the PID channel cannot be adjusted t
236. ong the area where a leak is suspected 3 Observe the display reading If the gas level increases the numerical reading will increase the bar graph level will increase to the right and the beeping and buzzer pulsing frequency will increase 4 Use the increasing and decreasing of the reading to locate the leak point 5 To exit Leak Check Mode and return to the Mode Select Screen at any time press and hold the RANGE SHIFT button for 5 seconds Turning the Buzzer On and Off in Leak Check Mode The alarm buzzer can be turned off and on when the EAGLE 2 is in Leak Check Mode This setting only applies to Leak Check Mode and does not affect buzzer operation in Normal or Bar Hole Mode The buzzer setting is displayed in the lower left corner as BUZZER ON when the buzzer is on or BUZZER OFF when the buzzer is off When the EAGLE 2 is turned off or you exit Leak Check Mode it remembers the buzzer setting So when the EAGLE 2 is turned on again or you return to Leak Check Mode after operating in Normal Mode or Bar Hole Mode the buzzer has the same setting it did the last time it was in Leak Check Mode To turn the buzzer off or on while in Leak Check Mode press and release the DISPLAY ADJUST NO button EAGLE 2 Operator s Manual Leak Testing 285 Turning Off the EAGLE 2 Leak Check Mode 1 Press and hold the POWER button 2 The buzzer will sound and the LCD back light will flash for about five seconds 3 Release the button
237. only the catalytic combustible sensor is active If an EAGLE 2 is intended for tracking down leaks it is shipped with Leak Check Mode or both Leak Check Mode and Bar Hole Mode enabled so that the operator must choose which operational mode to use when the unit is turned on see Updating LC BH Mode Setting on page 124 Start Up Leak Check Mode This section explains how to start up the EAGLE 2 in Leak Check Mode and get it ready for operation Turning On the EAGLE 2 Leak Check Mode CAUTION If one or more sensors other than a catalytic combustible sensor is installed these sensors will not be active while the EAGLE 2 is in Leak Check Mode The following description of the EAGLE 2 start up sequence assumes that the following menu items in Setup Mode are turned on CAL REMINDER and USER STATION ID If either of these items is turned off then the corresponding screen will not appear 1 Connect the standard probe to the EAGLE 2 s quick connect inlet fitting NOTE Use the standard probe not the bar hole probe when using the EAGLE 2 in Leak Check Mode 2 Ifasample hose is used connect the sample hose to the EAGLE 2 s quick connect inlet fitting and connect the probe to the sample hose s quick connect fitting 3 Press and briefly hold down the POWER ENTER RESET button Release the button when you hear a beep EAGLE 2 Operator s Manual Overview 279 4 The LCD will show the following screen for about t
238. ontinues investigate the cause See Troubleshooting on page 70 196 ESM 01 Calibration EAGLE 2 Operator s Manual 29 If you want to change the ESM 01 channel s calibration gas value follow Step 15 Step 24 above beginning with the ESM 01 Calibration Gas Value Screen below instead of the standard channel Calibration Gas Value Screen CAL GAS VALUE SO2 5 ppm ENTER TO BEGIN CAL 30 With the ESM 01 Calibration Gas Value Screen displayed press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen for the ESM 01 channel with CAL IN PROCESS flashing CAL IN PROCESS SO2 3 ppm ENTER WHEN DONE If you do not want to proceed with the calibration press and release the DISPLAY ADJUST NO button to return to the ESM 01 Cal Gas Values Screen If you do want to continue with the calibration proceed to the next step 31 Since calibration gas has already been flowing to the ESM O1 sensor while the other channels were being calibrated the ESM 01 sensor reading should be stable and ready for adjustment 32 Press and release the POWER ENTER RESET button to set the span adjustment for the ESM 01 channel to the programmed value EAGLE 2 Operator s Manual ESM 01 Calibration 197 33 If the ESM 01 channel passed calibration the following screen sequence occurs AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END Y CALIBRATION
239. ord 1994 to proceed to the Set Default Screen Press and release the AIR A YES button twice to restore the defaults See Restoring the Default Settings on page 127 for a description of issues to consider when restoring the defaults If difficulties continue contact RKI Instruments Inc for further instruction Troubleshooting 71 Replacing or Recharging the Batteries WARNING To prevent ignition of a hazardous atmosphere batteries must only be changed or charged in an area known to be nonhazardous Replace or charge the batteries when the EAGLE 2 indicates that it is in low battery warning When in low battery warning BATT appears vertically along the left side the LCD CH4 O LEL B OXY 20 9vol A H2S 0 0ppm T CO Oppm T Replacing the Batteries NOTE Use Duracell Procell PC 1400 alkaline batteries Duracell MN 1400 alkaline batteries Energizer E93 or Energizer EN93 or RKI Instruments Inc 49 1330RK Ni MH batteries to maintain the CSA classification of the EAGLE 2 Use of other batteries or mixing alkaline and rechargeable batteries will void the CSA classification and may void the warranty 1 Turn off the EAGLE 2 WARNING Do not remove the batteries while the EAGLE 2 is on 2 Loosen the battery case thumbscrew by turning it counterclockwise until it disengages from the bottom case If necessary use a coin or large flat blade screwdriver to loosen it 72 Replacing o
240. ot support this feature Zero Suppression Settings The ZERO SUPRESSION setting is not intended for field adjustment The typical setting is 2 of full scale The oxygen channel does not support this feature EAGLE 2 Operator s Manual Using Setup Mode 125 Updating the Confirmation Alert Setting With CONFIRMATION ALERT set to BEEP AND LIGHT the EAGLE 2 beeps and flashes the LED arrays once every 15 minutes to verify that it is operating With CONFIRMATION ALERT set to LIGHT ONLY the EAGLE 2 flashes the LED arrays once every 15 minutes to verify that it is operating With CONFIRMATION ALERT set to BEEP ONLY the EAGLE 2 beeps once every 15 minutes to verify that it is operating With CONFIRMATION ALERT set to OFF factory setting the EAGLE 2 does not sound a confirmation alert 1 From the main menu place the cursor next to CONFIRMATION ALERT 2 Press and release POWER ENTER RESET The Confirmation Alert Screen appears CONFIRMATION ALERT UP DOWN THEN ENTER OFF 3 Use AIR A YES or RANGE VY SHIFT to display the desired setting 4 Press and release POWER ENTER RESET to save the setting and return to the main menu Turning the Password Function On or Off With CHANGE PASSWORD set to ON the EAGLE 2 prompts you for a password when you enter Calibration Mode or Setup Mode With CHANGE PASSWORD set to OFF factory setting no password is required to enter Calibration Mode or Setup Mode
241. ots and that its bottom edge is resting on the bottom of the bottom case If the main PCB is not seated properly then it may be damaged when the top case is re installed 10 Make sure that the top case gasket is fully seated in its groove and carefully put the top case back on the bottom case If you have any difficulty mating the top and bottom cases inspect the placement of the main PCB and the placement of the top case gasket 11 Turn the EAGLE 2 upside down and tighten the three case screws to secure the top case to the bottom case 78 Replacing the Hydrophobic Filter EAGLE 2 Operator s Manual Replacing the Charcoal Filter Verify that the EAGLE 2 is off 2 Place the EAGLE 2 upside down on a flat surface or hold it upside down 3 Unscrew the three case screws until they disengage from the top case They are captive screws so they will not fall off of the bottom case 4 Turn the EAGLE 2 right side up and carefully lift the top case away from the bottom case Be careful not to lift it so far that it pulls on the main PCB with the cable that connects the top case to the main PCB 5 Lay the top case down next to the bottom case to allow access to the flow system 6 Locate the charcoal filter It is next to the CO sensor at the front of the flow chamber TK ao Endcap hie Charcoal Filter T LLIKA Z OS Figure 21 Replacing the Charcoal Filter 7 Grasp the black charcoal fi
242. own on a flat surface or hold it upside down 3 Unscrew the three case screws until they disengage from the top case They are captive screws so they will not fall off of the bottom case 4 Turn the EAGLE 2 right side up and carefully lift the top case away from the bottom case Be careful not to lift it so far that it pulls on the main PCB with the cable that connects the top case to the main PCB 5 Lay the top case down next to the bottom case to allow access to the flow system EAGLE 2 Operator s Manual Replacing the Hydrophobic Filter 77 6 Locate the hydrophobic filter It is over the oxygen sensor Note which side of the hydrophobic filter has the RKI logo and part number This is the inlet side and should be facing toward the front of the EAGLE 2 RKI Logo amp Part Number on This Side Grey Connecting Hydrophobic Filter Tube 2X 6 i l 1 oy ole E d Q d Bee a a O Bottom Case Shown Without Tubing For Clarity Figure 20 Replacing the Hydrophobic Filter Pull the grey connecting tubes off of each end of the filter and remove it Install the new filter with the red RKI logo and part number on the inlet side of the flow chamber facing the front of the EAGLE 2 Make sure to push the grey connection tubes all the way onto the filter s hose barbs 9 Confirm that the main PCB is seated in its sl
243. p and Normal Operation EAGLE 2 Operator s Manual sensor has its own protective shut off exposure to high levels of combustible gas can still stress the catalytic LEL sensor The catalytic LEL sensor can be enabled or disabled in the Catalytic LEL Sensor screen in Display Mode The default setting is enabled To change the setting do the following NOTE The Catalytic LEL Sensor setting is reset when the EAGLE 2 is turned off When the EAGLE 2 is turned on this setting is always ENABLED 1 Use the DISPLAY ADJUST NO button to enter Display Mode and scroll to the Catalytic LEL Sensor screen The current setting will be flashing CAT LEL SENSOR ENABLED 2 Use the AIR A YES or RANGE VY SHIFT button to toggle to the desired setting If set to DISABLED the gas readings for the catalytic LEL channel will be replaced by dashes 3 Press and release POWER ENTER RESET to save the setting and return to the main menu 4 Use the DISPLAY ADJUST NO button to scroll through the rest of Display Mode and enter Normal Operation Infrared Methane Calibration An infrared methane channel can be calibrated using the auto calibration method or the single calibration method To calibrate an infrared methane channel using the single calibration method see Calibrating Using the Single Calibration Method on page 64 and follow the instructions for calibrating a single channel If your instrument is a multi channel inst
244. ped to a sub PCB position that has no sub PCB installed the channel will be undefined and that channel will indicate a sensor failure during the startup sequence The EAGLE 2 Operator s Manual Channel Setup amp Sub PCBs 133 CONFIGURE CHANNELS LCD screen below illustrates this situation CONFIGURE CHANNELS gt CH1 CAT CH4 CH2 OXY OXY CH3 H2S H2S CH4 OPI y The to the right of OP1 indicates that no sub PCB is installed in position SUB1 See Configuring the Channels on page 105 for a complete description of the CONFIGURE CHANNELS Setup Mode menu item Sub PCBs and CONFIGURE GASES The CONFIGURE GASES menu item in Setup Mode allows you to configure the gas for a catalytic TC or PID sensor that is installed There are four possible sensors that can be configured depending on the particular version of the EAGLE 2 The LCD screen below illustrates the first screen in the CONFIGURE GASES menu item in Setup Mode Currently Configured Gas Supporting Circuitry Std or Optional CONFIGURE GASES l gt CAT CH4 CAT OP1 IBL PID OP2 NH3 ESM OP3 Sensor Type The four possible sensors are a catalytic sensor and three optional sensors They are listed with the following information from left to right optional The circuitry that supports the sensor and whether it is standard or This can be the catalytic circuit on the main PCB indic
245. ppears with the cursor next to CH4 ADJUST AUTO CALIBRATION VALUES gt CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm y 16 Place the cursor next to the channel whose gas value you want to change using the AIR A YES and RANGE YW SHIFT buttons 17 Press and release the POWER ENTER RESET button to select the channel The calibration gas value begins to flash 18 Use the AIR A YES and RANGE YW SHIFT buttons to adjust the calibration gas setting to the desired value NOTE The calibration gas setting value be set lower than the low alarm setting If the calibration gas value listed on the calibration cylinder is lower than the current low alarm setting enter Setup Mode and change the low alarm setting See Updating the Alarm Point Settings on page 114 for instructions If you need to change the alarm point setting only to perform a calibration make sure that you change the alarm point setting back to its original value once the calibration has been performed 19 Press and release the POWER ENTER RESET button to save the change The calibration gas value stops flashing 20 Repeat Step 16 through Step 19 for any other channels that need to be changed 21 When you are done adjusting the calibration gas values move the cursor down past the bottom of the screen next to END ADJUST AUTO CALIBRATION VALUES gt END EAGLE 2 Operator s Manual TC Calibration 211 22 Press and release the PO
246. prompts you for the password enter it by using the AIR A YES and RANGE VY SHIFT buttons to select each password number and then pressing and releasing POWER ENTER RESET to enter the number and move on to the next one The Calibration Mode Screen displays with the cursor next to AUTO CALIBRATION CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION Move the cursor to the PERFORM AIR ADJUST menu item by using the RANGE W SHIFT button CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION gt PERFORM AIR ADJUST NORMAL OPERATION Press and release the POWER ENTER RESET button The following screen appears PERFORM AIR ADJUST Press and release the AIR A YES button to continue 152 PID Calibration EAGLE 2 Operator s Manual If you do not want to continue press the DISPLAY ADJUST NO button and the unit will return to the Calibration Mode Screen 9 The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds then indicate that the operation is complete before returning to the Calibration Mode Screen ADJUSTING ZERO RELEASE AIR BUTTON ADJUSTING ZERO COMPLETE CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION gt PERFORM AIR ADJUST NORMAL OPERATION 10 Install the demand flow regulator onto the multi gas calibration cylinder 11 Connect the sample tubing to the
247. ption free of charge This warranty does not apply to those items which by their nature are subject to deterioration or consumption in normal service and which must be cleaned repaired or replaced on a routine basis Examples of such items are Absorbent cartridges Filter elements disks or sheets Pump diaphragms and valves Warranty is voided by abuse including mechanical damage alteration rough handling or repair procedures not in accordance with the instruction manual This warranty indicates the full extent of our liability and we are not responsible for removal or replacement costs local repair costs transportation costs or contingent expenses incurred without our prior approval THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS EXPRESSED OR IMPLIED AND ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF RKI INSTRUMENTS INC INCLUDING BUT NOT LIMITED TO THE WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE IN NO EVENT SHALL RKI INSTRUMENTS INC BE LIABLE FOR INDIRECT INCIDENTAL OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY This warranty covers instruments and parts sold to users only by authorized distributors dealers and representatives as appointed by RKI Instruments Inc We do not assume indemnification for any accident or damage caused by the operation of this gas
248. r Oxygen Sensor One lt Oxygen Sensor LEL Sensor and Bracket H2S Sensor Sensor Bracket Figure 40 Typical Infrared Methane Sensor Location The infrared methane sensor is installed in a single sensor flow chamber which is located in the area next to the standard 4 sensor flow chamber This area can accommodate up to three single sensor flow chambers Figure 40 above illustrates a typical infrared methane sensor location in front of the pump The infrared methane flow chamber may also be installed in one of the EAGLE 2 Operator s Manual Description 239 other two sensor chamber locations depending on the particular version of the EAGLE 2 Some infrared methane instrument configurations do not include the 4 sensor flow chamber Infrared Methane Sensor The infrared methane sensor is a cylindrical sensor with a diffusion opening on the front and a 12 pin circular connector on the back A 12 wire cable plugs into the back of the infrared methane sensor with a circular connector that includes a locking lever The other end of the cable plugs into an infrared sub PCB see description below that is installed on the main PCB The sensor is held in the infrared flow chamber by a bracket on standoffs Infrared Sub PCB The infrared sub PCB is a circuit board that is installed on the main PCB in one of the 3 sub PCB positions when an infrared methane sensor is used with the EAGLE 2 The infrared methane sensor connects
249. r Recharging the Batteries EAGLE 2 Operator s Manual l Figure 16 Removing the Battery Case co 3 Pull the battery case away from the bottom case The thumbscrew is captive and will not fall out 4 Carefully remove the old batteries Verify that the battery compartment and electrical contacts are clean 5 Carefully install the new C size batteries Follow the battery diagram inside the battery case Make sure the batteries are pushed in all the way AMO Te I a O or O Cc SIZE C SIZE C SIZE SIZE oF qr s T _ Z Figure 17 Installing the Batteries 6 Reinstall the battery case onto the bottom case tightening the thumbscrew firmly by hand so that there is no visible gap between the battery case and the bottom case EAGLE 2 Operator s Manual Replacing or Recharging the Batteries 73 Recharging the Ni MH Batteries CAUTION Use with Ni MH battery p n 49 1330RK Charge only with RKI charger model 49 2175RK 49 2176RK or 49 2177RK Use of other rechargeable batteries or chargers or charging of other rechargeable batteries in the EAGLE 2 will void the warranty WARNI
250. r an equivalent international recommended practice for guidance in the use of combustible gas detection instruments Table of Contents EAGLE 2 Operator s Manual Chapter 1 Introduction Overview This chapter briefly describes the EAGLE 2 gas monitor This chapter also describes the EAGLE 2 Operator s Manual this document Table 1 at the end of this chapter lists the specifications for the EAGLE 2 About the EAGLE 2 Using an advanced detection system consisting of up to six gas sensors the EAGLE 2 sample draw gas monitor is capable of detecting the presence of combustible gas oxygen O2 carbon monoxide CO hydrogen sulfide H S and various other toxic gases simultaneously The EAGLE 2 s rugged reliable and easy to use design makes it ideally suited for a wide range of applications including sewage treatment plants utility manholes tunnels hazardous waste sites power stations petrochemical refineries mines paper mills drilling rigs and fire fighting stations The EAGLE 2 offers a full range of features including e Simultaneous monitoring of one to six gases The standard configuration includes four sensors for combustible gas 7LEL ppm and volume oxygen carbon monoxide CO and hydrogen sulfide H2S e Choice of three operating modes e Normal Mode for typical confined space or area monitoring Normal Mode is the standard factory setting e Bar Hole Mode for checking of bar holes when sea
251. r next to one of the pre defined gases press and release POWER ENTER RESET to select the gas and proceed to Step 15 If you placed the cursor next to the user defined gas with the asterisk press POWER ENTER RESET and proceed with Step 7 7 The user defined gas setup screen appears with the first character of the gas name flashing The current gas name and range are shown on the top line of the screen CAT 0 10000ppm CH4 CHANGE TO NAME CAT 10000 150000 10000 ppm RATIO 1 00 FACTOR 8 Enter the gas name Use AIR A YES and RANGE YW SHIFT to display the desired character then press POWER ENTER RESET to enter the displayed character and move to the next character Repeat until all three characters are entered When the last character is entered the ppm ratio value will be flashing 9 Use AIR A YES and RANGE YW SHIFT to display the desired ppm value This value is called the ppm ratio and must be the ppm equivalent of the LEL for the gas being defined For example if you are defining propane the LEL for propane is 21 000 ppm so you must enter 21000 ppm 2 1 volume NOTE If you define a gas whose LEL is above 50 000 ppm the LEL reading in Measuring Mode will reflect the defined ppm ratio but the ppm reading in Measuring Mode will not indicate above 50 000 ppm For example if you set the ratio to be 150 000 ppm and set the catalytic combustible channel to display the reading in ppm the gas reading will not in
252. range on the result screen as described in Step 12 and Step on page 68 A new sensor can typically be adjusted to more than twice the calibration gas concentration If the result screen indicates that the EAGLE 2 could not adjust the combustible gas reading to be at least 10 higher than the calibration gas concentration then the sensor should be replaced as soon as possible Replacing a Sensor Verify that the EAGLE 2 is off Place the EAGLE 2 upside down on a flat surface or hold it upside down Unscrew the three case screws until they disengage from the top case They are captive screws so they will not fall off of the bottom case Turn the EAGLE 2 right side up and carefully lift the top case away from the bottom case Be careful not to lift it so far that it pulls on the main PCB with the cable that connects the top case to the main PCB 80 Checking the Combustible Gas Sensor s ConditionEAGLE 2 Operator s Manual 5 Lay the top case down next to the bottom case to allow access to the flow system 6 Locate the sensor you want to replace and remove it from the flow chamber H2S Sensor LEL Sensor Cable LEL Sensor Bracket cope le Oxygen Sensor Bracket LEL Sensor Oxygen Sensor Main PCB H2S Connector Flow Block Figure 22 Replacing a Sensor 7 To remove the LEL sensor do the following e Unscrew and remove the two screws that hold down the LEL sensor bracket e Gra
253. rared Methane Sensor Specifications Autoranging Configuration 0 100 LEL CH 1 LEL 10 LEL 50 LEL Table 28 Infrared Methane Sensor Specification Volume Configuration Target Gases The infrared methane sensor is setup for and factory calibrated to methane There are gases that the sensor will still detect and respond to There are also gases that the methane sensor will not detect or respond to Lists of the gases falling in each of these respective categories can be found below Consult RKI Instruments Inc for combustible gases not listed below The infrared methane sensor is known to respond to the following combustible gases e ethane e hexane e IPA e isobutane e MEK 238 Overview EAGLE 2 Operator s Manual e propane e toluene The infrared methane sensor is known to not or to poorly respond to the following combustible gases e acetylene e hydrogen e styrene Description Table 26 and Table 27 above list the available infrared methane sensor configurations The infrared methane sensor is a smart sensor that stores the target gas Top Case Not Shown a Charcoal Filter Infrared CH4 al gre A Standard Flow Chambers YO pare I TE e RE oal i N T Tox N A U z mal OP IE 88 CHI OK ES 7 al E A E V W A A A x Co Z i CO Senso
254. rching for underground gas leaks e Leak Check Mode for locating leaks in valves and piping e Sample drawing pump with up to 125 foot range e Liquid crystal display LCD for complete and understandable information at a glance e Ultrabright alarm LEDs e Distinctive audible alarm for dangerous gas conditions or unit malfunction e Microprocessor control for reliability ease of use and advanced capabilities EAGLE 2 Operator s Manual Overview 1 e Data logging functions when used in Normal Mode e Alarm trend data when used in Normal Mode e STEL and TWA when used in Normal Mode and over range alarms e Peak readings when used in Normal Mode e Built in time function e Lunch break feature e RF shielded high impact plastic case e CSA classified for Class I Division I Groups A B C and D hazardous atmospheres WARNING The Model EAGLE 2 detects oxygen deficiency elevated levels of oxygen combustible gases carbon monoxide and hydrogen sulfide all of which can be dangerous or life threatening When using the EAGLE 2 you must follow the instructions and warnings in this manual to assure proper and safe operation of the unit and to minimize the risk of personal injury Be sure to maintain and periodically calibrate the EAGLE 2 as described in this manual NOTE ONLY THE COMBUSTIBLE GAS DETECTION PORTION OF THIS INSTRUMENT HAS BEEN ASSESSED FOR PERFORMANCE 2 About the EAGLE 2 EAGLE 2 Operator s Manual
255. ress and release the DISPLAY ADJUST NO button to return to the Cal Gas Values Screen If you do want to continue with the calibration proceed to the next step 26 Connect the tubing from the demand flow regulator to the rigid tube on the probe Allow the EAGLE 2 to draw gas for one minute 246 Infrared Methane Calibration EAGLE 2 Operator s Manual 27 Press and release the POWER ENTER RESET button to set the span adjustment for each channel to the programmed values 28 If all channels passed calibration the following screen sequence occurs AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END Y CAL GAS VALUE CH4 50 LEL ENTER TO BEGIN CAL If any of the sensors cannot be adjusted to the proper value a screen displays that indicates a calibration failure and lists the sensor s that failed to calibrate In the example below the oxygen and HS channels failed calibration The other sensors calibrated normally FAIL SENSOR lt gt lt OXY gt lt H2S gt lt gt ENTER TO CONTINUE The buzzer and alarm LEDs activate in a double pulsing pattern Press and release the POWER ENTER RESET button to reset the alarm and continue to the Calibration Value Screen for the infrared methane channel After calibrating the infrared methane channel by following the instructions below attempt to calibrate the standard channels again If the failure continues investigate the ca
256. ressing and releasing the POWER ENTER RESET button to enter it and move on to the next number until all of the numbers are entered The main menu displays It displays six menu items at a time gt SET DATE amp TIME SET DATE FORMAT SET BATTERY TYPE CONFIGURE CHANNELS CONFIGURE GASES CATALYTIC UNITS y Use the RANGE V SHIFT button to move the cursor down the menu to CONFIGURE GASES Press and release POWER ENTER RESET The Configure Gases Screen appears with the cursor flashing next to CAT the catalytic sensor If an EAGLE 2 has a PID sensor installed one of the three optional sensor types OP1 OP2 or OP3 will indicate it is a PID In the example below OP1 is shown as a PID sensor CONFIGURE GASES gt CAT CH4 CAT OP1 IBL PID E e z Ww N i T i wm lt EAGLE 2 Operator s Manual Configuring the PID Gas in Setup Mode 173 8 Use RANGE VW SHIFT to move the cursor down the menu to the PID sensor CONFIGURE GASES CAT CH4 CAT gt OP1 IBL PID OP2 OP3 y 9 To change the PID sensor gas configuration press and release POWER ENTER RESET 10 A screen appears that indicates the PID type low range or high range flashing and the detection range for the currently configured gas In the example below the PID sensor is currently configured as a high range sensor for 0 2 000 ppm isobutylene PID 0 2000ppm IBL UP DOWN THEN ENT
257. rms Set to SELF RESETTING When the EAGLE 2 goes into gas alarm the POWER ENTER RESET button will not silence or reset the alarm When the gas reading falls below or rises above for an oxygen low alarm an alarm setpoint the alarm will automatically reset The LED arrays and buzzer turn off and the EAGLE 2 alarm indications on the display turn off Responding to Alarms This section describes response to gas over range battery and sensor failure alarms Responding to Gas Alarms 1 Determine which gas alarm has been activated 2 Follow your established procedure for an increasing gas condition or a decreasing oxygen condition 3 If necessary reset the alarm using the POWER ENTER RESET button once the alarm condition has passed 38 Measuring Mode Alarms EAGLE 2 Operator s Manual Responding to Over Range Alarms WARNING An over range condition may indicate an extreme combustible gas toxic gas or oxygen concentration Confirm anormal condition with a different EAGLE 2 or with another gas detecting device 1 Determine which channel is in alarm 2 Follow your established procedure for an extreme gas condition 3 Reset the alarm using the POWER ENTER RESET button once the alarm condition has cleared 4 Calibrate the EAGLE 2 as described in Chapter 4 Calibration Mode on page 55 5 Ifthe over range condition continues or if you are not able to successfully calibrate the unit you may need to replace the sen
258. robe 33 2002RK 01 Humidifier 24 inch for calibration of catalytic combustible channel with gas concentrations of 1000 ppm or lower rtratyeopbobe mer Parc eror serasa ome 49 2175RK 115 220 VAC charger 49 2176RK 12 VDC charger 49 2177RK 115 220 VAC and 12 VDC charger 65 0601RK Oxygen sensor 84 General Parts List EAGLE 2 Operator s Manual Table 11 General Parts List cont 65 2005RK Carbon monoxide CO sensor Setup Program and all operator s manuals eT 80 0406RK Dilution fitting 3 80 05XXRK Sample hose Replace XX with length in feet 5 foot hose is standard Available lengths for the EAGLE 2 are 3 4 5 6 10 15 20 25 30 35 40 50 75 100 and 125 feet 81 0090RK 01 Calibration cylinder steel 34 liter three gas CH4 O CO 81 0090RK 03 Calibration cylinder steel 103 liter three gas CH4 O gt CO 81 0154RK 02 Calibration cylinder aluminum 58 liter four gas CH4 O gt HaS CO 81 0154RK 04 Calibration cylinder aluminum 34 liter four gas CH4 O2 HaS CO 81 1054RK Regulator demand flow type for 58 and 103 liter aluminum or steel and 34 liter aluminum calibration cylinders 81 5302RK Calibration kit for LEL Oxy CO unit w demand flow regulator 103 liter cylinder 81 5401RK Calibration kit for LEL Oxy H5S CO unit w demand flow regulator 58 liter cylinder ES 87RW H2S Hydrogen Sulfide H2S sensor NC 6260B LEL combustible sensor catalytic EAGLE 2 Operator s Ma
259. rom the regulator EAGLE 2 Operator s Manual PID Calibration 149 31 If you want to change the PID channel s calibration gas value follow Step 5 Step 24 above beginning with the PID Calibration Gas Value Screen below instead of the standard channel Calibration Gas Value Screen CAL GAS VALUE IBL 100 ppm ENTER TO BEGIN CAL 32 With the PID Calibration Gas Value Screen displayed press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen for the PID channel with CAL IN PROCESS flashing CAL IN PROCESS IBL Oppm ENTER WHEN DONE If you do not want to proceed with the calibration press and release the DISPLAY ADJUST NO button to return to the PID Cal Gas Values Screen If you do want to continue with the calibration proceed to the next step 33 Screw the PID calibration cylinder onto the demand flow regulator 34 Connect the tubing from the regulator to the rigid tube on the probe Allow the EAGLE 2 to draw gas for one minute 35 Press and release the POWER ENTER RESET button to set the span adjustment for the PID channel to the programmed value 150 PID Calibration EAGLE 2 Operator s Manual 36 If the PID channel passed calibration the following screen sequence occurs AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END Y CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST
260. rough 200 100 and 0 for a high range sensor and 2 00 1 00 and 0 00 for a low range sensor Allow the sensor to continue warming up for 5 minutes If it exits its warm up 138 Start Up and Normal Operation EAGLE 2 Operator s Manual state the alternating numbers will be replaced by a gas reading If the sensor doesn t exit its warm up state after 5 minutes see Troubleshooting on page 160 e If using a probe be sure to use the PID probe that does not have a particle filter installed If it is necessary to use a sample hose for any reason you must use the teflon lined hose that is supplied with the PID instrument A 5 foot hose is supplied as standard 10 15 and 20 foot hoses are also available See the Parts List at the end of this appendix for ordering information e If your EAGLE 2 is a multigas unit that is used for monitoring of combustible gases in the LEL range the PID channel will indicate an upscale reading if these gases are present If LEL concentrations of combustible gas are present the PID channel may indicate an overscale reading e The PID sensor will also respond to H3S so if H S is present the PID channel may indicate an upscale reading depending on the concentration present The standard calibration for a PID channel is to isobutylene A PID channel can be factory setup for and calibrated to other gases Consult RKI Instruments Inc for other available PID configurations and to specify the de
261. ructions appropriate for your cylinder configuration NOTE If your instrument has more than one ESM 01 sensor you will need a calibration cylinder for each sensor and follow the instructions in Calibrating with a 4 Gas Cylinder and an ESM 01 Cylinder on page 182 Calibrating with a 4 Gas Cylinder and an ESM 01 Cylinder The calibration instructions below show a5 channel instrument which has the four standard channels LEL oxygen H S CO and an ESM 01 channel for ammonia NH3 L See Calibration Supplies and Equipment on page 56 for a description of the necessary calibration supplies In addition to an appropriate multi gas cylinder that is used to calibrate any active standard channels you will also need a cylinder to calibrate the ESM 01 channel See Table 19 on page 202 for available cylinders Make sure your calibration cylinder is appropriate for the ESM 01 detection range Find a fresh air environment This is an environment free of toxic or combustible gases and of normal oxygen content 20 9 While in Measuring Mode press and hold the RANGE W SHIFT button then press the DISPLAY ADJUST NO button and release both buttons If the unit prompts you for the password enter it by using the AIR A YES and RANGE VY SHIFT buttons to select each password number and then pressing and releasing POWER ENTER RESET to enter the number and move on to the next one The Calibration Mode Screen displays with the cursor next
262. rument that includes an infrared methane channel RKI Instruments Inc recommends using the auto calibration method for convenience The calibration instructions below show a 5 channel instrument which has the four standard channels LEL oxygen H S CO and an infrared methane channel configured for autoranging The standard factory calibration for the autoranging infrared methane sensor is to 50 LEL If your instrument is configured for autoranging and you need EAGLE 2 Operator s Manual Infrared Methane Calibration 241 maximum accuracy in the vol range the sensor may be calibrated to 50 vol To use the auto calibration method to calibrate a multi channel instrument that includes an infrared methane channel do the following L See Calibration Supplies and Equipment on page 56 for a description of the necessary calibration supplies In addition to an appropriate multi gas cylinder that is used to calibrate any active standard channels you will also need a cylinder to calibrate the infrared methane channel See Table 29 on page 251 for available cylinders Make sure your calibration cylinder is appropriate for the infrared methane detection range Find a fresh air environment This is an environment free of toxic or combustible gases and of normal oxygen content 20 9 While in Measuring Mode press and hold the RANGE W SHIFT button then press the DISPLAY ADJUST NO button and release both buttons If the unit pr
263. s Manual 36 If the infrared methane channel passed calibration the following screen sequence occurs AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END Y CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION If the infrared methane channel cannot be adjusted to the proper value a screen displays that indicates a calibration failure and indicates that the infrared methane sensor failed to calibrate FAILED SENSOR ENTER TO CONTINUE lt CH4 gt The buzzer and alarm LEDs activate in a double pulsing pattern Press and release the POWER ENTER RESET button to reset the alarm and continue to the Calibration Mode Screen Attempt to calibrate the infrared methane channel again If the failure continues investigate the cause See Troubleshooting on page 70 37 Disconnect the tubing from the probe 38 Unscrew the demand flow regulator from the calibration cylinder 39 Use the RANGE YW SHIFT button to place the cursor next to the NORMAL OPERATION menu option then press and release the POWER ENTER RESET button to return to Measuring Mode EAGLE 2 Operator s Manual Infrared Methane Calibration 249 Maintenance This section includes a procedure to replace an infrared methane sensor Replacing the IR Methane Sensor L 2 3 10 11 12 13 14 15 Verify that the EAGLE 2 is off Place the EAGL
264. s the desired button the unit returns to the Calibration Gas Values Screen CAL GAS VALUES CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm ENTER TO BEGIN CAL 15 Press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen CAL IN PROCESS is flashing CAL IN PROCESS CH4 0 LEL OXY 20 9 vol H2S 0 0 ppm CO 0 ppm ENTER WHEN DONE If you do not want to proceed with the calibration press and release the DISPLAY ADJUST NO button to return to the Calibration Gas Values Screen If you do want to continue with the calibration proceed to the next step 16 Connect the gas bag tubing to the regulator s or dispensing valve s hose barb fitting 92 Calibrating Using the Auto Calibration Method EAGLE 2 Operator s Manual 17 Fill the gas collection bag by screwing the fixed flow regulator or dispensing valve onto the calibration cylinder and turning the knob counterclockwise 18 Allow the gas to dispense until the gas collection bag is a little over half full 19 Turn the knob clockwise to stop the gas flow clamp down the hose clamp and remove the regulator or dispensing valve from the cylinder 20 Disconnect the tubing from the regulator or dispensing valve 21 Open the hose clamp on the gas bag tubing 22 Connect the tubing from the gas bag to the rigid tube on the probe Allow the gas to flow for one minute NOTE If you are calibrating the catalytic
265. sample hose Long sample hoses will require more time to show a response at the EAGLE 2 The maximum sample hose length recommended for the EAGLE 2 is 125 feet Consult RKI Instruments Inc for longer sample hose lengths 4 Ifa gas alarm occurs take appropriate action See Responding to Alarms on page 38 Using Optional Sample Hoses The standard sample hose for the EAGLE 2 is 5 feet long Optional hoses are available up to 125 feet long If you are considering using a longer hose keep in mind that a longer hose will increase the EAGLE 2 s response time and the flowrate may decrease close to the low flow alarm point CAUTION Sample hose lengths of more than 125 feet are not recommended for the EAGLE 2 because of flow rate reduction and increased response time Consult RKI Instruments Inc for hose lengths longer than 125 feet 30 Measuring Mode Normal Operation EAGLE 2 Operator s Manual The chart below illustrates how response time is affected by the sample hose length Table 4 EAGLE 2 Response Time vs Sample Hose Length Typical Time to 90 Probe amp 100 Foot Hose Probe amp 125 Foot Hose 75 seconds Using Exhaust Tubing The EAGLE 2 s exhaust fitting has a female 10 32 thread to allow for the installation of a hose barb fitting with a 10 32 thread to which a flexible exhaust tube can be connected If you utilize this feature the tubing used must have a minimum internal diameter of 1 8 inch RKI Ins
266. scribes the operation of the EAGLE 2 in Methane Elimination Mode e Appendix K describes the operation of the EAGLE 2 in Bar Hole Mode e Appendix L describes the operation of the EAGLE 2 in Leak Check Mode e Appendix M describes the Tank Tester Model of the EAGLE 2 e Appendix N describes the operation of the EAGLE 2 in Inert Mode EAGLE 2 Operator s Manual About this Manual 5 6 Overview Chapter 2 Description Overview This chapter describes the EAGLE 2 instrument and accessories Instrument Description The EAGLE 2 includes the case sensors LCD control buttons printed circuit boards alarm LEDs infrared communication port buzzer battery case and batteries and flow system E C i ie o S e OO Alarm LEDs Inlet Fitting a E C LCD Display en ee _ Control Buttons at Pa A Ge Ne A N J y P N ENTER ADJUST A Vv RESET NO YES SHIFT j x F J Figure 1 Component Location Top View EAGLE 2 Operator s Manual Case Screw 3X Hidden From view A Case Foot 6X Bottom Case Figure 2 Component Location Side View Buzzer E 7 T RA Inlet Fitting i Exhaust Fitting a
267. se to the bottom case 25 Calibrate the ESM 01 channel as described in ESM 01 Calibration on page 181 Parts List Table 19 ESM 01 EAGLE 2 Parts List 33 7120RK H2S scrubber disk for ESM 01DH D HCN sensor 33 7121RK H2S scrubber disk for ESM 01DH F SO2 sensor 47 5015RK ESM 01 sensor cable ESM 01DH ASH3 Arsine sensor 0 1 50 ppm ESM 01R NH3 Ammonia sensor 0 75 0 ppm ESM 01DH D HCN Hydrogen cyanide sensor 0 15 0 ppm ESM 01DH PH3 Phosphine sensor 0 1 00 ppm ESM 01DH F SO2 Sulfur dioxide sensor 0 6 00 ppm ESM K01 CL2 Chlorine sensor 0 3 00 ppm 81 0142RK 02 Calibration cylinder 5 gas SOs CH4 O2 H2S CO 58 liter aluminum 81 0142RK 04 Calibration cylinder 5 gas SO CH4 Oo H2S CO 34 liter aluminum 81 0170RK 02 Calibration cylinder 5 ppm SOz in nitrogen 58 liter 81 0170RK 04 Calibration cylinder 5 ppm SO in nitrogen 34 liter 81 0175RK 02 Calibration cylinder 10 ppm NH3 in nitrogen 58 liter 81 0175RK 04 Calibration cylinder 10 ppm NH3 in nitrogen 34 liter 202 Parts List EAGLE 2 Operator s Manual 81 0185RK 02 Calibration cylinder 0 5 ppm PHg in nitrogen 58 liter 81 0192RK 02 Calibration cylinder 2 ppm Cl in nitrogen 58 liter 81 0192RK 04 Calibration cylinder 2 ppm Cl in nitrogen 34 liter 81 0196RK 02 Calibration cylinder 10 ppm HCN in nitrogen 58 liter 81 0196RK 04 Calibration cylinder 10 ppm HCN in nitrogen 34 liter EAGLE 2 Operator s Manual Parts List 203
268. sing methane elimination mode CAUTION The EAGLE 2 catalytic combustible sensor experiences a significant gas response drop when it is changed from full response mode to methane elimination mode The EAGLE 2 must be calibrated in both full response mode and in methane elimination mode to account for this See Calibration on page 269 for further instructions Monitoring in Methane Elimination Mode To monitor an area using methane elimination mode do the following 1 Turn on the EAGLE 2 as described in Start Up on page 22 2 Make sure that the EAGLE 2 catalytic combustible channel is set to monitor a gas in which methane elimination is an option For a list of these gases and a procedure to select one see Configuring the Combustible Gas on page 108 A common target gas is hexane Once a gas that allows for methane elimination is selected for the catalytic combustible channel the methane elimination mode screen becomes part of Display Mode 3 Use the DISPLAY button to enter Display Mode and to scroll through the screens until you reach the Methane Elimination Mode screen The current setting is displayed METHANE ELIMINATION MODE DISABLED 4 Press and release the AIR A YES or RANGE VW SHIFT button to toggle the setting to ENABLED 5 Press and release the POWER ENTER RESET button The unit will save the setting and proceed to the next menu item 266 Overview EAGLE 2 Operator s Manual 6
269. sired PID configuration when a unit is ordered The display screen below illustrates an EAGLE 2 with a low range PID sensor installed IBL 0 00ppm It is possible to temporarily configure a PID channel for a target gas other than the factory setting in Display Mode See the next section PID Relative Response Feature for instructions to use this feature PID Relative Response Feature The relative response feature enables you to change the PID sensor s response to gas on the fly so that the PID channel is roughly calibrated to an alternate gas This is done by temporarily changing the gas configuration of the PID channel You can select from a list of gases whose response relative to the configured gas normally isobutylene is programmed into the EAGLE 2 s memory For example if the PID channel is setup for and calibrated to EAGLE 2 Operator s Manual PID Relative Response Feature 139 isobutylene IBL you can select isopropyl alcohol IPA from a gas list accessible from the PID Relative Response Screen in Display Mode so that the PID channel responds to gas as if it were calibrated to isopropyl alcohol The EAGLE 2 will clear the gas configuration change when it is turned off and will return to the programmed configuration when it is turned on again The gas list for the relative response feature includes several pre defined gases and 1 gas that can be entered into the EAGLE 2 in the field using the Eagle 2 Ma
270. sor next to the desired gas In the example below the cursor is placed next to isopropanol IPA PID O 2000ppm IBL 2000 ppm GSL 2000 ppm IBL gt 5000 ppm IPA 1000 ppm JP5 Y EAGLE 2 Operator s Manual Configuring the PID Gas in Setup Mode 175 13 Press and release the POWER ENTER RESET button The LCD will ask for confirmation that you want to change the configured gas CHANGE TO IPA PRESS YES OR NO 14 To change the gas configuration press and release AIR A YES The unit will return to the Configure Gases Screen reflecting the new gas configuration CONFIGURE GASES CAT CH4 CAT gt OPI IPA PID OP2 OP3 lt If you do not want to change the gas configuration press and release DISPLAY ADJUST NO The unit will return to the first gas configuration choice screen with the gas configuration unchanged PID O 2000ppm IBL gt 1000 ppm ACT 1000 ppm BNZ 1500 ppm DSL 15000 ppm ETA Y Select a different gas and go to Step 13 or press and release DISPLAY ADJUST NO to return to the Gas Configuration screen 15 Use RANGE YW SHIFT to place the cursor next to END 16 Press and release POWER ENTER RESET to return to the main menu 17 Use RANGE YW SHIFT to place the cursor in front of NORMAL OPERATION at the bottom of the main menu 18 Press and release POWER ENTER RESET 176 Configuring the PID Gas in Setup Mode EAGLE 2 Op
271. sor that has triggered the over range alarm 6 If the over range condition continues after you have replaced the sensor contact RKI Instruments Inc for further instructions Responding to Battery Alarms WARNING The EAGLE 2 is not operational as a gas monitoring device during a dead battery alarm Take the Model EAGLE 2 to a non hazardous area and replace or recharge the batteries as described in Replacing or Recharging the Batteries on page 72 The EAGLE 2 is fully functional during a low battery warning However only a limited amount of operating time remains approximately 1 2 hours The amount of time depends on how often the LCD backlight is used and how often the unit is responding to alarm conditions Recharge the Ni MH batteries or replace the alkaline batteries as soon as possible as described in Replacing or Recharging the Batteries on page 72 NOTE Alarms and the LCD back light consume battery power and reduce the amount of operating time remaining Responding to Sensor Failure Alarms 1 Determine which sensor has triggered the sensor failure alarm 2 Try calibrating the sensor first as described in Chapter 4 Calibration Mode on page 55 before replacing it EAGLE 2 Operator s Manual Measuring Mode Alarms 39 If the sensor failure continues replace the sensor as described in Replacing a Sensor on page 80 If the sensor failure condition continues after you have replaced the sensor
272. sp the LEL sensor connector and gently pull it up until it either disengages from the LEL sensor or the LEL sensor comes out of the flow chamber with the connector e Ifthe sensor came out with the connector remove the sensor from the connector EAGLE 2 Operator s Manual Replacing a Sensor 81 e Ifthe sensor stayed in the flow chamber grasp the sensor and pull it out of the flow chamber 8 To remove the oxygen sensor do the following e Unscrew the two screws that hold the oxygen sensor bracket a few turns so that you can rotate and remove the oxygen sensor bracket Make sure to note the routing of the oxygen sensor cable to the main PCB so that you can route the replacement sensor cable the same way Also make sure that the O ring in the bottom of the flow chamber does not come out with the sensor e Move the hydrophobic filter towards the bottom case side wall and pull the oxygen sensor out of the flow chamber e Hold the main PCB to support it where the oxygen sensor cable connects to it e Grasp the connector on the end of the sensor cable and pull the connector away from the main PCB to disconnect it from the main PCB 9 To remove the H S and CO sensors do the following e Grasp the sensor firmly and rock it back and forth slightly while pulling on it Make sure to note the routing of the sensor cable to the main PCB so that you can route the replacement sensor cable the same way e If the sensor does not come out
273. t lt gt lt gt ENTER TO CONTINUE If the catalytic combustible sensor fails it is not possible to enter Leak Check Mode Press and release the POWER ENTER RESET button to acknowledge the failure and return to the Mode Select Screen Replace the failed sensor If any other sensor that is installed fails press and release the POWER ENTER RESET button to acknowledge the failure and continue to Leak Check Mode Change the failed sensor s as soon as possible for use in Normal Mode EAGLE 2 Operator s Manual Start Up Leak Check Mode 283 15 The EAGLE 2 is now operating in Leak Check Mode The pump is on and the following screen appears LEAK CHECK MODE CH4 0 ppm 0 5000 BUZZER ON NOTE For maximum sensor stability allow 3 5 minutes for the sensor to warm up The small increment size in the lower range of a ppm measurement can cause instability if the unit is not properly warmed up Performing a Demand Zero Leak Check Mode Before using the EAGLE 2 it is recommended to set the fresh air reading for the target gas by performing a demand zero This will set the CH channel to zero 1 Find a fresh air environment This is an environment free of toxic or combustible gases and of normal oxygen content 20 9 2 Turn on the unit as described above in Turning On the EAGLE 2 Leak Check Mode 3 Press and hold the AIR A YES button The display prompts you to hold the AIR A YES button
274. t PERFORM AIR ADJUST NORMAL OPERATION 10 Install the demand flow regulator onto the multi gas calibration cylinder 11 Connect the sample tubing to the demand flow regulator 12 Install the probe on the EAGLE 2 inlet fitting Make sure the probe is complete with internal O ring and membrane and that the two halves of 192 ESM 01 Calibration EAGLE 2 Operator s Manual the probe are tightened firmly together to avoid leaks that can affect the calibration See Figure 19 on page 76 for an illustration of the internal parts of the probe 13 Move the cursor next to the AUTO CALIBRATION menu item by using the AIR A YES button CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 14 Press and release the POWER ENTER RESET button to display the Calibration Gas Values Screen CAL GAS VALUES CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm ENTER TO BEGIN CAL The gas concentrations displayed in the Calibration Gas Values Screen must match the gas concentrations listed on the 4 gas calibration cylinder If all concentrations match go to Step 25 If one or more concentrations do not match continue with Step 15 If you do not want to continue with the calibration press and release the DISPLAY ADJUST NO button to return to the Calibration Mode Screen NOTE The RKI 4 gas cylinder typically contains 12 O by volume When using the auto calibratio
275. t Settings 0 0 0 0 ccc cee 127 Updating the Lunch Break Setting 0 0 06 c ne 129 Updating the Span Factor Setting 0 0c ns 129 Updating the Language Setting 0 0 0 0 0c cen 130 Exiting Setup Mode ahs hc hua ca btn oka and bod Go Aa ae ew aan 130 Appendix C Sab PCBS 3 assianane seer eee eers kenen pow aes 132 ONCIVICW siente aa na a ened tenth Kees hye ey oe eee es 132 WESCHIPU OMe iets atack rat NETO gw ud eda weterenstet Bane Rar quate EY 132 Channel Setup and Sub PCBs scuece dane gees dees dea pales 133 Sub PCBs and CONFIGURE CHANNELS o an 133 Sub PCBs and CONFIGURE GASES 0 0000 c cece 134 Appendix D PID Sensotsy is crits ie her aa E iw ews 136 OVI VIOW ES ciety Setting ted akc deena we tree GE on ie ee ete ee es 136 WESCMIPUOM bs eat Rowell sabe ie Be oe ate Ee eh ei Rises eters Med 136 PID Sensor and Sensor Adapter 0 000 cc eens 137 PVPS SUD COD 6s an E ES E TE I EEEE 137 Table of Contents EAGLE 2 Operator s Manual TOUTED PVG DC yt cae his amp Seat ate ly r Reaotc ta i ta tty SB eg OG OE OEA 137 Start Up and Normal Operation 00 c eee eee ee eee 138 PID Relative Response Feature 30540 cue ye ee keG ede e eee ws 139 PID Sensor Relative Response Screen in Display Mode 140 PID Calibration s 6 hans ues e820 treo 2G old edhe yA deheycaiped 00 aTa a ach whee 143 Calibrating with a 4 Gas Cylinder and a PID Cylinder 143 Calibrating with a 5 Gas Cylinder 0 0
276. t board that is installed on the main PCB in one of the 3 sub PCB positions when a PID sensor is used with the EAGLE 2 The PID sensor adapter connects to the sub PCB with a 5 position connector The sub PCB plugs into the main PCB and is held in place with a screw flat washer lock washer There are no user serviceable parts on the PID sub PCB PID Probe Several of the gases that can be monitored with a PID are easily absorbed in the EAGLE 2 s standard sample hose and standard probe One example of this is styrene Because of this RKI Instruments Inc recommends that you EAGLE 2 Operator s Manual Description 137 use the probe intended for PID use that does not have a particle filter installed and that you install the probe directly to the inlet fitting when monitoring for gas with the PID sensor The figure below shows an exploded drawing of the PID probe Sa Figure 26 PID Probe Start Up and Normal Operation For instructions to startup and use an EAGLE 2 that includes a PID sensor reference Start Up on page 22 Measuring Mode Normal Operation on page 29 and Measuring Mode Alarms on page 35 Follow these instructions keeping the following special considerations in mind e After the instrument s warm up period is over and the gas readings are displayed on the screen if the PID sensor is still in its warm up state the gas reading for the PID channel will scroll th
277. t the need to enter CONFIGURE CHANNELS or CONFIGURE GASES in Setup Mode EAGLE 2 Operator s Manual Overview 179 See Replacing the ESM 01 Sensor or Changing Sensor Type on page 199 for instructions to replace or change an ESM O1 sensor Top Case Not Shown Pump lt Standard Flow Chamber JOYE ene Charcoal Filter ESM 01 lel Tall Oxygen Sensor VORE Er e 7 Oxygen Sensor LEL Sensor and Bracket H2S Sensor Sensor Bracket Figure 35 Typical ESM 01 Sensor Location The ESM 01 sensor is installed in a single sensor flow chamber which is located in the area next to the standard 4 sensor flow chamber This area can accommodate up to three single sensor flow chambers Although the EAGLE 2 can support up to three ESM 01 sensors many combinations are impractical for various reasons including sensor cross sensitivity to other gases Consult RKI Instruments Inc for practical combinations Figure 35 above illustrates a typical ESM 01 sensor location in front of the pump The ESM O1 flow chamber may also be installed in one of the other two sensor chamber locations depending on the particular version of the EAGLE 2 Some ESM 01 instrument configurations do not include the 4 sensor flow chamber ESM 01 Sensor The ESM 01 sensor is a cylindrical sensor with a diffusion open
278. te the detection elements inside the combustible sensor damaging the sensor and result in reduced response to combustible gas Make every effort to avoid these gases The catalytic combustible sensor has an integral HS scrubber for protection from H S exposure resulting from normal use but you should avoid exposure to high levels of H S and other sulphur compounds Monitoring Combustible Gas in the PPM or Volume Range The standard factory configuration for the EAGLE 2 allows the user to use the RANGE YW SHIFT button to change the displayed detection units of the catalytic combustible channel between LEL ppm and volume It is possible to disable this capability and set the EAGLE 2 to display only one of the detection units by using the Catalytic Units menu item in Setup Mode See Setting the Catalytic Detection Units on page 113 for instructions to set this Setup Mode Menu item The detection range of the combustible catalytic channel when set for ppm or Yovolume will correspond to 0 100 LEL for the configured gas For example the LEL for methane CH3 is 5 volume or 50 000 ppm So if the catalytic combustible channel is displayed in terms of volume the full scale is 5 00 and if it is displayed in terms of ppm the full scale is 50 000 EAGLE 2 Operator s Manual Measuring Mode Normal Operation 33 If the Catalytic Units menu item in Setup Mode is set to CHANGE OK the standard factory setting then you can change the cata
279. ter with the EAGLE 2 Symptoms Probable Causes Table 9 Troubleshooting the EAGLE 2 Recommended Action e The LCD is blank The LCD shows abnormally high or low readings but other gas detection instruments do not 70 Overview The unit may have been turned off The alkaline batteries may need to be replaced or the Ni MH batteries recharged The EAGLE 2 may need to be recalibrated The sensor for the affected channel s may need replacement To turn on the unit press and briefly hold the POWER ENTER RESET button If the unit does not turn on replace the alkaline batteries or recharge the Ni MH batteries If the difficulties continue contact RKI Instruments Inc for further instruction Recalibrate the unit If the difficulties continue replace the sensor for the affected channel s and calibrate the affected channel s If the difficulties continue contact RKI Instruments Inc for further instruction EAGLE 2 Operator s Manual Symptoms Probable Causes Table 9 Troubleshooting the EAGLE 2 Recommended Action The unit indicates flow failure and does not recover when POWER ENTER RESET is pressed and released Auto calibration or single calibration fails e Display indicates SYSTEM FAIL 12 during startup The probe tube is clogged The hydrophobic filter disk in the probe is dirty The sample hose has a kink
280. th EXIT at the top of the list gt EXIT ACETONE BENZENE DIESEL FUEL NO 1 ETHANOL GASOLINE There are multiple screens of gases The following is the complete list of factory defined gases along with their detection ranges low alarm high alarm STEL and TWA settings Table 14 is a list of the low range values and Table 15 is a list of the high range values Table 14 Low Range Relative Response Gas List Target Gas Detection Alarm 1 Factory Alarm 2 Factory STEL Range ppm Setting ppm Setting ppm ppm te Diesel Fuel 0 40 00 2 00 3 00 NO 1 o To KAE o o o fo o fo o To Methyl Ethyl 0 40 00 2 00 3 00 ZE eee Tetrachloro 0 30 00 0 24 1 00 ethylene Trichloro 0 30 00 ethylene Vinyl Chloride 0 100 00 EAGLE 2 Operator s Manual PID Relative Response Feature 141 Table 14 Low Range Relative Response Gas List O PD 0 50 00 OFF OFF OFF OFF Table 15 High Range Relative Response Gas List Target Gas Detection Alarm 1 Factory Alarm 2 Factory STEL Range ppm a ppm a ppm Ea Acetone 01000 1000 1000 ao Fuel 2 1500 NO 1 Ethanol 0 15000 15000 1000 1500 1000 a a Cree oew o foo o e E R AR E R JP 5 Fuel 5 Fuel 01000 1000 OOo uo 21 a bl Mil ll Mil Ketone ee feje E S So e E F Styrene 050 500 Tetrachloro 1000 T oe ethylene Trichloro 0 1000 ethylene Vinyl Chloride Vinyl Chlor
281. the DISPLAY ADJUST NO button to enter Display Mode and scroll to the Catalytic LEL Sensor screen The current setting will be flashing The screen below indicates the warning that appears when a TC sensor and a catalytic LEL sensor are both installed Since the TC sensor only reads in volume if the catalytic LEL sensor is disabled there will be no alarms for the LEL range CAT LEL SENSOR xk WARNING NO LEL ALARMS IF CAT LEL IS DISABLED ENABLED Use the AIR A YES or RANGE Y SHIFT button to toggle to the desired setting If set to DISABLED the gas reading for the catalytic LEL channel will be replaced by dashes Press and release POWER ENTER RESET to save the setting and return to the main menu Use the DISPLAY ADJUST NO button to scroll through the rest of Display Mode and enter Normal Operation TC Calibration A TC channel can be calibrated using the auto calibration method or the single calibration method To calibrate a TC channel using the single calibration method see Calibrating Using the Single Calibration Method on page 64 and follow the instructions for calibrating a single channel If your instrument is a multi channel instrument that includes one or more TC channels RKI Instruments Inc recommends using the auto calibration method for convenience The calibration instructions below show a 5 channel instrument which has the four standard channels LEL oxygen H S CO and a TC ch
282. the EAGLE 2 is started up in the presence of a target gas the readings and alarms will not be accurate or reliable 1 From the main menu place the cursor next to AUTO FRESH AIR ADJ 2 Press and release POWER ENTER RESET The Fresh Air Adjust Screen appears FRESH AIR ADJUST ON POWERUP UP DOWN THEN ENTER OFF 3 Use AIR A YES or RANGE YW SHIFT to display the desired setting 4 Press and release POWER ENTER RESET to save the setting and return to the main menu Updating the Data Log Interval Setting This setting indicates how often the EAGLE 2 saves readings to the data logger The following interval times can be selected 10 minutes 5 minutes 3 minutes 1 minute 30 seconds 20 seconds 10 seconds or 5 seconds The factory setting is 30 seconds 1 From the main menu place the cursor next to DATA LOG INTERVAL 2 Press and release POWER ENTER RESET The Data Log Interval Screen appears DATA LOG INTERVAL UP DOWN THEN ENTER 30 SECONDS 3 Use AIR A YES or RANGE YW SHIFT to display the desired setting 4 Press and release POWER ENTER RESET to save the setting and return to the main menu 120 Using Setup Mode EAGLE 2 Operator s Manual Updating the Data Log Overwrite Setting With DATA LOG OVERWRITE set to ON factory setting the EAGLE 2 writes over the oldest data with new data when the data logger memory is full With DATA LOG OVERWRITE set to OFF the EAGLE 2 stops saving d
283. the EAGLE 2 can be calibrated with an oxygen concentration of up to 19 5 RKI Instruments Inc recommends that the multi gas cylinder have an oxygen concentration in the range of 10 16 oxygen e A gas collection bag with hose clamp e A6LPM fixed flow regulator or a dispensing valve e Calibration tubing To calibrate the combustible gas oxygen CO and HS sensors at the same time automatically with no need for a zero oxygen source you can use the auto calibration feature with a 4 gas cylinder If the HS channel is not active then a 3 gas cylinder may be used for auto calibration This document includes instructions for auto calibration with a fixed flow regulator or dispensing valve a sample bag and a 4 gas cylinder This document also includes instructions for calibrating one channel at a time using single calibration Entering Calibration Mode To enter Calibration Mode do the following 1 Find a fresh air environment This is an environment free of toxic or combustible gases and of normal oxygen content 20 9 2 While in Measuring Mode press and hold the RANGE W SHIFT button then press the DISPLAY ADJUST NO button and release both buttons EAGLE 2 Operator s Manual Entering Calibration Mode 87 3 Ifthe unit prompts you for the password enter it by using the AIR A YES and RANGE YW SHIFT buttons to select each password number and then pressing and releasing POWER ENTER RESET to enter the number and move on to
284. the EAGLE 2 inlet fitting Make sure the probe is complete with internal O ring and membrane and that the two halves of the probe are tightened firmly together to avoid leaks that can affect the calibration See Figure 19 Replacing the Particle Filter and Hydrophobic Filter Disk on page 76 for an illustration of the internal parts of the probe EAGLE 2 Operator s Manual PID Calibration 145 13 Move the cursor next to the AUTO CALIBRATION menu item by using the AIR A YES button CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION 14 Press and release the POWER ENTER RESET button to display the Calibration Gas Values Screen CAL GAS VALUES CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm ENTER TO BEGIN CAL The gas concentrations displayed in the Calibration Gas Values Screen must match the gas concentrations listed on the 4 gas calibration cylinder If all concentrations match go to Step 25 If one or more concentrations do not match continue with Step 15 If you do not want to continue with the calibration press and release the DISPLAY ADJUST NO button to return to the Calibration Mode Screen NOTE The RKI 4 gas cylinder typically contains 12 O by volume When using the auto calibration method be sure to set the OXY auto calibration value to agree with the concentration listed on the cylinder s label not zero 15 To adjust the v
285. the auto calibration method be sure to set the OXY auto calibration value to agree with the concentration listed on the cylinder s label not zero 6 To adjust the values on the screen press and hold the RANGE W SHIFT button then press the DISPLAY ADJUST NO button and release both The following screen appears with the cursor next to CH4 ADJUST AUTO CALIBRATION VALUES gt CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm y 7 Place the cursor next to the channel whose gas value you want to change using the AIR A YES and RANGE YW SHIFT buttons 8 Press and release the POWER ENTER RESET button to select the 60 Calibrating Using the Auto Calibration Method EAGLE 2 Operator s Manual channel The calibration gas value begins to flash 9 Use the AIR A YES and RANGE YW SHIFT buttons to adjust the calibration gas setting to the desired value NOTE The calibration gas value cannot be set lower than the low alarm setting If the calibration gas value listed on the calibration cylinder is lower than the current low alarm setting enter Setup Mode and change the low alarm setting See Updating the Alarm Point Settings on page 114 for instructions If you need to change the alarm point setting only to perform a calibration make sure that you change the alarm point setting back to its original value once the calibration has been performed 10 Press and release the POWER ENTER RESET button to save the chang
286. the current gas reading for each target gas CH4 O LEL OXY 20 9v0l H2S 0 0ppm CO Oppm Performing a Demand Zero Before using the EAGLE 2 it is recommended to set the fresh air readings for the target gases by performing a demand zero This will set the CHy H3S and CO channels to zero and the OXY channel to 20 9 1 Find a fresh air environment This is an environment free of toxic or combustible gases and of normal oxygen content 20 9 2 Turn on the unit as described above in Turning On the EAGLE 2 3 Press and hold the AIR A YES button The LCD prompts you to continue holding the AIR A YES button and the buzzer will pulse while you hold the button 4 Continue to hold the AIR A YES button until the LCD prompts you to release it The EAGLE 2 will set the fresh air reading for all channels Start up is complete and the unit is now ready for monitoring Turning Off the EAGLE 2 1 Press and hold the POWER ENTER RESET button 2 The buzzer will pulse for about five seconds 3 Release the button when GOODBYE and the RKI logo appear on the display When GOODBYE and the RKI logo disappear and the backlight turns off the unit is off EAGLE 2 Operator s Manual Start Up 27 Using the Battery Charger for Continuous Operation The battery charger can be used with an AC adapter or a vehicle plug DC adapter to run the EAGLE 2 in continuous operation instead of charging the batteries Batteries do not n
287. the infrared CO Calibration Gas Value Screen displayed press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen for the infrared CO channel with CAL IN PROCESS flashing CAL IN PROCESS CO2 0 00 vol ENTER WHEN DONE If you do not want to proceed with the calibration press and release the DISPLAY ADJUST NO button to return to the infrared CO Cal Gas Values Screen If you do want to continue with the calibration proceed to the next step 37 Screw the infrared CO calibration cylinder onto the demand flow regulator 38 Connect the tubing from the regulator to the rigid tube on the probe Allow the EAGLE 2 to draw gas for the appropriate time shown in the table below Table 24 IR CO Sensor Calibration Times Range Calibration Time 39 Press and release the POWER ENTER RESET button to set the span 234 Infrared CO Calibration EAGLE 2 Operator s Manual adjustment for the infrared CO channel to the programmed value 40 If the infrared CO channel passed calibration the following screen sequence occurs AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END Y CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION If the infrared CO channel cannot be adjusted to the proper value a screen displays that indicates a calibration failure and indicates that the infrared CO
288. the next one 4 The Calibration Mode Screen displays with the cursor next to AUTO CALIBRATION CALIBRATION MODE gt AUTO CALIBRATION SINGLE CALIBRATION PERFORM AIR ADJUST NORMAL OPERATION NOTE The following screens illustrate a 4 gas EAGLE 2 for detection of CH LEL using catalytic sensor oxygen H2S and CO Your EAGLE 2 may display slightly different screens Calibrating Using the Auto Calibration Method This method allows you to calibrate the CH LEL catalytic combustible sensor oxygen HS and CO sensors simultaneously It is designed for use with the RKI 4 gas calibration cylinder and is the quickest and most convenient method to calibrate the EAGLE 2 Setting the Fresh Air Reading 1 While in the Calibration Mode Screen move the cursor to the PERFORM AIR ADJUST menu item by using the RANGE W SHIFT button CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION gt PERFORM AIR ADJUST NORMAL OPERATION 88 Calibrating Using the Auto Calibration Method EAGLE 2 Operator s Manual 2 Press and release the POWER ENTER RESET button The following screen appears PERFORM AIR ADJUST 3 Press and release the AIR A YES button to continue If you do not want to continue press the DISPLAY ADJUST NO button and the unit will return to the Calibration Mode Screen 4 The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds then indicate that th
289. til you return to the main menu 8 Use RANGE V SHIFT to move the cursor next to the END menu item Press and release the POWER ENTER RESET button to save the changes and return to the main menu Updating the Backlight Delay Setting This setting indicates the length of time the LCD illuminates when you press any button The minimum setting is 0 seconds the maximum setting is 255 seconds The factory setting is 30 seconds 1 From the main menu place the cursor next to BACKLIGHT DELAY 2 Press and release POWER ENTER RESET The Backlight Delay Screen appears BACKLIGHT DELAY UP DOWN THEN ENTER 30 SECONDS 3 Use AIR A YES and RANGE VW SHIFT to adjust the time to the desired setting 4 Press and release POWER ENTER RESET to save the setting and return to the main menu Updating the Auto Fresh Air Setting This setting allows you to configure the EAGLE 2 so that a fresh air adjustment takes place automatically as part of the instrument startup sequence If AUTO FRESH AIR ADJ is set to ON the EAGLE 2 performs a fresh air adjustment at the end of the startup sequence before entering Normal Operation The factory setting is OFF EAGLE 2 Operator s Manual Using Setup Mode 119 WARNING If the automatic fresh air feature is turned on you must startup the EAGLE 2 in a known fresh air environment an environment free of toxic or combustible gases and of normal oxygen content 20 9 If this feature is on and
290. tion gas values move the cursor down past the bottom of the screen next to END ADJUST AUTO CALIBRATION VALUES gt END A 22 Press and release the POWER ENTER RESET button The following screen appears DO YOU WANT TO STORE NEW VALUE S IN MEMORY FOR FUTURE CALIBRATIONS PRESS YES OR NO 23 If you select YES by pressing and releasing the AIR A YES button the changes that you made will be saved in the EAGLE 2 s memory as the new auto calibration gas values If you select NO by pressing and releasing the DISPLAY ADJUST NO button the changes you made will be used for any calibrations performed during the current operating session only The EAGLE 2 will delete the EAGLE 2 Operator s Manual Infrared Hydrocarbon Calibration 259 changes when the unit is turned off and will load the previous set of auto calibration values when it is turned on again 24 When you make your selection and press the desired button the unit returns to the Calibration Gas Values Screen CAL GAS VALUES CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm ENTER TO BEGIN CAL 25 Press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen with CAL IN PROCESS flashing CAL IN PROCESS CH4 0 LEL OXY 20 9 vol H2S 0 0 ppm CO 0 ppm ENTER WHEN DONE If you do not want to proceed with the calibration press and release the DISPLAY ADJUST NO button to return to t
291. to proceed to the unit s start up sequence without saving changes EAGLE 2 Operator s Manual Using Setup Mode 131 Appendix C Sub PCBs Overview An EAGLE 2 that has one or more of the standard four sensors catalytic LEL oxygen H S and CO and no optional sensors does not have any sub PCBs installed The sub PCBs are used to add circuitry to the EAGLE 2 that supports various optional sensors An EAGLE 2 has provisions to install up to three sub PCBs Although an EAGLE 2 with three sub PCBs installed can theoretically support seven sensors if you include the standard four sensors the EAGLE 2 is only capable of having six active channels Description The main PCB can accept up to three sub PCBs The sub PCBs are installed below a row of connectors near the top of the main PCB The three positions are labeled on the main PCB silkscreen from right to left as SUB1 SUB2 and SUB3 These labels are not visible when the sub PCBs are installed ef Ao Figure 24 Sub PCB Positions The sub PCBs plug into the main PCB with a multiposition connector and are 132 Overview EAGLE 2 Operator s Manual retained in place with a screw flat washer lock washer Any sub PCB can be installed in any of the sub PCB positions A sensor that is supported by a sub PC
292. to see It can be set from 1 to 15 The factory setting is 8 The higher the setting the darker the characters and LCD background 1 From the main menu place the cursor next tt ADJUST CONTRAST 2 Press and release POWER ENTER RESET The Adjust Contrast Screen appears ADJUST CONTRAST UP DOWN THEN ENTER 8 3 Use AIR A YES or RANGE YW SHIFT to adjust the setting so that the characters on the LCD are easy to see 4 Press and release POWER ENTER RESET to save the setting and return to the main menu Updating the Calibration Reminder Setting With CAL REMINDER set to ON factory setting the EAGLE 2 will give an indication at start up if it is due for calibration The type of indication will depend on the CAL PAST DUE ACT setting see the next menu item below 1 From the main menu place the cursor next to CAL REMINDER 2 Press and release POWER ENTER RESET The Calibration Reminder Screen appears CALIBRATION REMINDER UP DOWN THEN ENTER ON 3 Use AIR A YES or RANGE YW SHIFT to display the desired setting 4 Press and release POWER ENTER RESET to save the setting and return to the main menu 122 Using Setup Mode EAGLE 2 Operator s Manual Updating the Calibration Past Due Action Setting This item defines what indication is given during start up when calibration is due and CAL REMINDER is set to ON With CAL PAST DUE ACT set to CONFIRM TO CAL factory setting the EAGLE 2 wil
293. truction and flow system are appropriate for the installed sensors Operation of the EAGLE 2 with a flow system or construction not compatible with the installed sensors will result in inaccurate readings Consult RKI Instruments Inc if you cannot confirm either of these items 1 From the main menu place the cursor next to CONFIGURE CHANNELS 2 Press and release POWER ENTER RESET The Configure Channels screen appears with the cursor flashing next to CH1 CONFIGURE CHANNELS gt CH1 CAT CH4 CH2 OXY OXY CH3 CH4 H2S H2S CO CO y The standard 4 gas configuration is shown below CONFIGURE CHANNELS CH1 CAT CH4 CH2 OXY OXY CH3 H2S H2S gt CH4 CO CO y RANGE SHIFT p AIR A YES CONFIGURE CHANNELS gt CH5 NOT USED CH6 NOT USED END 3 Use AIR A YES or RANGE YW SHIFT to move the cursor next to the channel you want to configure 106 Using Setup Mode EAGLE 2 Operator s Manual 4 Press and release POWER ENTER RESET The available configuration options are shown In the example below channel 3 has been selected for configuration CHANNEL 3 H2S gt CAT CH4 OXY OXY H2S H2S CO CO RANGE Vv SHIFT p A AIR A YES CHANNEL 3 H25 gt OP1 A OP OP3 NOT USED 5 Use AIR A YES and RANGE VW SHIFT to move the cursor next to the desired configuration for th
294. truments Inc recommends using flexible polyurethane tubing with a maximum exhaust tube length of 20 feet Consult RKI Instruments Inc for exhaust tubing lengths longer than 20 feet Combustible Gas Detection There are three issues to keep in mind when monitoring for combustible gas e The catalytic combustible sensor will respond to any combustible gas The standard calibration gas for the EAGLE 2 catalytic combustible channel is methane CH If the instrument is calibrated to a different combustible gas such as hexane or propane the gas name for the catalytic combustible channel will reflect the target gas The table below lists the conversion factors for several hydrocarbon gases if the EAGLE 2 is calibrated to methane To use this table multiply the display reading on the combustible gas channel by the factor in the appropriate row to obtain the actual gas concentration For example if you are detecting pentane and the display reads 10 LEL for the catalytic combustible channel you actually have 10 LEL x 1 38 13 8 LEL pentane present EAGLE 2 Operator s Manual Measuring Mode Normal Operation 31 Table 5 Full Response Mode Conversion Factors Methane Calibration Target Gas LEL Factor PPM Factor Target Gas LEL Factor PPM Factor Acetone 0 74 Isobutane 0 54 Benzene 0 62 Isopropanol 0 87 Butyl Acrylate 0 85 Methane 1 00 Butyl Acetate 0 89 Methanol 1 79 2 Butyl Alcohol 0 73 Methyl Acetate 0 87 1 Butyl Alcohol 1 2
295. ttoms out Line up the holes in the infrared hydrocarbon sensor bracket with the two standoffs on the infrared hydrocarbon chamber Install the two sensor bracket screws Confirm that the main PCB is seated in its slots and that its bottom edge is resting on the bottom of the bottom case If the main PCB is not seated properly it may be damaged when the top case is re installed 264 Maintenance EAGLE 2 Operator s Manual 16 Make sure that the top case gasket is fully seated in its groove and carefully put the top case back on the bottom case If you have any difficulty mating the top and bottom cases inspect the placement of the main PCB and the placement of the top case gasket 17 Turn the EAGLE 2 upside down and tighten the three case screws to secure the top case to the bottom case 18 Calibrate the infrared hydrocarbon channel as described in Infrared Hydrocarbon Calibration on page 255 Parts List Table 32 Infrared HC EAGLE 2 Parts List 47 5051RK Infrared HC sensor cable DEM 3313 2 Infrared HC sensor 0 100 LEL 2 0 30 0 vol autoranging 81 0018RK 01 Calibration cylinder 50 LEL isobutane in air 34 liter steel 81 0018RK 03 Calibration cylinder 50 LEL isobutane in air 103 liter steel EAGLE 2 Operator s Manual Parts List 265 Appendix J Methane Elimination Mode Overview For applications where methane is an interfering gas you can set the EAGLE 2 to eliminate most response to methane u
296. uence occurs AUTO CALIBRATION AUTO CALIBRATION DONE REMOVE GAS END Y CAL GAS VALUE CO2 2 50 vol ENTER TO BEGIN CAL If any of the sensors cannot be adjusted to the proper value a screen displays that indicates a calibration failure and lists the sensor s that failed to calibrate In the example below the oxygen and H S channels failed calibration The other sensors calibrated normally FAIL SENSOR lt gt lt OXY gt lt H2S gt lt gt ENTER TO CONTINUE The buzzer and alarm LEDs activate in a double pulsing pattern Press and release the POWER ENTER RESET button to reset the alarm and continue to the Calibration Value Screen for the infrared CO channel After calibrating the infrared CO channel by following the instructions below attempt to calibrate the standard channels again If the failure continues investigate the cause See Troubleshooting on page 70 33 Remove the tubing from the rigid tube on the probe 34 Unscrew the 4 gas cylinder from the demand flow regulator EAGLE 2 Operator s Manual Infrared CO Calibration 233 35 If you want to change the infrared CO channel s calibration gas value follow Step 17 Step 26 above beginning with the infrared CO Calibration Gas Value Screen below instead of the standard channel Calibration Gas Value Screen CAL GAS VALUE CO2 2 50 vol ENTER TO BEGIN CAL 36 With
297. up and use an EAGLE 2 that includes a TC sensor reference Start Up on page 22 Measuring Mode Normal Operation on page 29 and Measuring Mode Alarms on page 35 Follow these instructions keeping the following in mind e The TC sensor is used to monitor combustible gases above their LEL lower explosive limit When monitoring the target gas in volume be aware that you may be monitoring gas levels that are potentially explosive Catalytic LEL Sensor Screen When either a TC sensor or an infrared combustible sensor is installed in an EAGLE 2 along with a catalytic combustible LEL sensor the user has the option of turning off the catalytic combustible LEL sensor in Display Mode If the unit is going to be used for sampling known high levels of combustible gas or in areas with known catalytic sensor poisons such as silicone vapors the catalytic combustible sensor should be turned off Even though this sensor has its own protective shut off exposure to high levels of combustible gas can still stress the catalytic LEL sensor The catalytic LEL sensor can be enabled or disabled in the Catalytic LEL Sensor screen in Display Mode The default setting is enabled To change the setting do the following NOTE The Catalytic LEL Sensor setting is reset when the EAGLE 2 is turned off When the EAGLE 2 is turned on this setting is always ENABLED 206 Start Up and Normal Operation EAGLE 2 Operator s Manual L Use
298. uring the Combustible Gas nes 108 Setting the Catalytic Detection Units 0 0 000 cee 113 Updating the Relative Response Setting nnannnnnnana 114 EAGLE 2 Operator s Manual Table of Contents Updating the Alarm Point Settings 0 006 ccc nee 114 Updating the Alarm Latching Setting 0 000 000 ee 116 Updating the Alarm Silence Setting 0 000 c eee 117 Turning the User Station ID Function On or Off 0 0 0 0 0 44 117 Updating the Autocal Values 0 0 cc tenes 118 Updating the Backlight Delay Setting 0 0 60 e eee 119 Updating the Auto Fresh Air Setting 0000 ccc eee 119 Updating the Data Log Interval Setting 0 000 cee 120 Updating the Data Log Overwrite Setting 4000 0s 121 Updating the Data Log Memory Setting 0 0 0 00 ee 121 Updating the LCD Contrast Setting 000 ccc eee 122 Updating the Calibration Reminder Setting 0 0044 122 Updating the Calibration Past Due Action Setting 123 Updating the Calibration Interval 0 0 ce 123 Updating LC BH Mode Setting 000s 124 Setting the Bar Hole Measurement Time 000 0005 125 Zero Follower Settings 5 55 cohen cactini 1 hos eo tee F thls Gohan erat hia ent 125 Zero Suppression Settings in pods Matas oh ha Daa eee 125 Updating the Confirmation Alert Setting 0 0 000 cee 126 Turning the Password Function On or Off 0 0600 ees 126 Restoring the Defaul
299. ursor down past the bottom of the screen next to END ADJUST AUTO CALIBRATION VALUES gt END A 22 Press and release the POWER ENTER RESET button The following screen appears DO YOU WANT TO STORE NEW VALUE S IN MEMORY FOR FUTURE CALIBRATIONS PRESS YES OR NO 23 If you select YES by pressing and releasing the AIR A YES button the changes that you made will be saved in the EAGLE 2 s memory as the new auto calibration gas values If you select NO by pressing and releasing the DISPLAY ADJUST NO button the changes you made will be used for any calibrations performed during the current operating session only The EAGLE 2 will delete the EAGLE 2 Operator s Manual PID Calibration 147 changes when the unit is turned off and will load the previous set of auto calibration values when it is turned on again 24 When you make your selection and press the desired button the unit returns to the Calibration Gas Values Screen CAL GAS VALUES CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm ENTER TO BEGIN CAL 25 Press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen with CAL IN PROCESS flashing CAL IN PROCESS CH4 0 LEL OXY 20 9 vol H2S 0 0 ppm CO 0 m ENTER WHEN DONE If you do not want to proceed with the calibration press and release the DISPLAY ADJUST NO button to return to the Cal Gas Values Screen If you do want to
300. use See Troubleshooting on page 70 29 Remove the tubing from the rigid tube on the probe 30 Unscrew the 4 gas cylinder from the demand flow regulator EAGLE 2 Operator s Manual Infrared Methane Calibration 247 31 If you want to change the infrared methane channel s calibration gas value follow Step 15 Step 24 above beginning with the infrared methane Calibration Gas Value Screen below instead of the standard channel Calibration Gas Value Screen CAL GAS VALUE CH4 50 LEL ENTER TO BEGIN CAL 32 With the infrared methane Calibration Gas Value Screen displayed press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen for the infrared methane channel with CAL IN PROCESS flashing CAL IN PROCESS CH4 0 LEL ENTER WHEN DONE If you do not want to proceed with the calibration press and release the DISPLAY ADJUST NO button to return to the infrared methane Cal Gas Values Screen If you do want to continue with the calibration proceed to the next step 33 Screw the infrared methane calibration cylinder onto the demand flow regulator 34 Connect the tubing from the regulator to the rigid tube on the probe Allow the EAGLE 2 to draw gas for 90 seconds 35 Press and release the POWER ENTER RESET button to set the span adjustment for the infrared methane channel to the programmed value 248 Infrared Methane Calibration EAGLE 2 Operator
301. uto calibration gas values If you select NO by pressing and releasing the DISPLAY ADJUST NO button the changes you made will be used for any calibrations performed during the current operating session only The EAGLE 2 will delete the EAGLE 2 Operator s Manual Infrared CO Calibration 231 changes when the unit is turned off and will load the previous set of auto calibration values when it is turned on again 28 When you make your selection and press the desired button the unit returns to the Calibration Gas Values Screen CAL GAS VALUES CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm ENTER TO BEGIN CAL 29 Press and release the POWER ENTER RESET button to proceed to the Calibration In Process Screen with CAL IN PROCESS flashing CAL IN PROCESS CH4 0 LEL OXY 20 9 vol H2S 0 0 ppm CO 0 ppm ENTER WHEN DONE If you do not want to proceed with the calibration press and release the DISPLAY ADJUST NO button to return to the Cal Gas Values Screen If you do want to continue with the calibration proceed to the next step 30 Connect the tubing from the demand flow regulator to the rigid tube on the probe Allow the EAGLE 2 to draw gas for one minute 31 Press and release the POWER ENTER RESET button to set the span adjustment for each channel to the programmed values 232 Infrared CO Calibration EAGLE 2 Operator s Manual 32 If all channels passed calibration the following screen seq
302. ve screen will still appear In this case press and release DISPLAY ADJUST NO to proceed to exit Setup Mode and begin the EAGLE 2 s startup sequence 23 If you do not want to save the changes press and release DISPLAY ADJUST NO The unit will begin its startup sequence without saving the changes If you do want to save the changes press and release AIR A YES and continue with the next step 24 A confirmation screen appears asking if you are sure you want to save the changes ARE YOU SURE YOU WANT TO SAVE ALL CHANGES IN MEMORY 25 If you want to save the changes press and release AIR A YES to save the changes A screen will appear for a few seconds indicating that the changes have been saved and the unit will begin its startup sequence If you do not want to save the changes press and release DISPLAY ADJUST NO to proceed to the unit s startup sequence without saving changes EAGLE 2 Operator s Manual Configuring the TC Gas in Setup Mode 221 Parts List Table 21 TC EAGLE 2 Parts List 47 5010RK TC LEL sensor cable 81 0013RK 01 Calibration cylinder 50 vol CH in No 34 liter steel 81 0013RK 05 Calibration cylinder 50 vol CH in No 58 liter steel 81 0023RK 01 Calibration cylinder H2 8 volume in nitrogen 34 liter steel 81 0024RK 01 Calibration cylinder Hs 100 volume 34 liter steel 81 0025RK 01 Calibration cylinder 35 CO2 50 CHy4 balance No 34 liter steel intended for landfil
303. w system Locate the TC sensor It has a five wire cable with a connector that mates to a TC sub PCB that is installed on the main PCB and is normally located in a single sensor flow chamber next to the pump Figure 37 on page 205 shows a TC sensor in a typical location Unscrew and remove the two screws that hold down the TC sensor bracket Grasp the TC sensor connector and gently pull it up until it either disengages from the TC sensor or the TC sensor comes out of the flow chamber with the connector If the sensor came out with the connector remove the sensor from the connector If the sensor stayed in the flow chamber grasp the sensor and pull it out of the flow chamber Plug the replacement sensor into the sensor connector on the TC sensor cable Insert the TC sensor into the TC sensor chamber in the flow chamber Line up the holes in the TC sensor retaining bracket with the two standoffs on either side of the TC sensor chamber Install the two sensor retaining screws tightening them a little at a time alternately to push the sensor into its chamber evenly Confirm that the main PCB is seated in its slots and that its bottom edge is resting on the bottom of the bottom case If the main PCB is not seated 216 Maintenance EAGLE 2 Operator s Manual properly it may be damaged when the top case is reinstalled 16 Make sure that the top case gasket is fully seated in its groove and carefully put the top case back on t
304. w chamber which is located in the area next to the standard 4 sensor flow chamber This area can accommodate up to three single sensor flow chambers Figure 25 below illustrates a typical PID sensor location in front of the pump The PID flow chamber may also be installed in one of the other two sensor chamber locations depending on the particular version of the EAGLE 2 Some PID instrument configurations do not include the 4 sensor flow chamber 136 Overview EAGLE 2 Operator s Manual Top Case Not Shown x Charcoal Filter PID Sensor Fs T Pump i Standard Flow Chamber A OKC i O q W JO CO Sensor Oxygen Sensor Ss Oxygen Sensor g N LEL Sensor and Bracket H2S Sensor gt Sensor Bracket Figure 25 Typical PID Sensor Location PID Sensor amp Sensor Adapter The PID sensor is a cylindrical sensor with a diffusion opening on the front and 3 pins on the back It is plugged into a sensor adapter with a 5 wire cable that terminates in a 5 position connector The connector plugs into a PID sub PCB see description below that is installed on the main PCB The sensor adapter allows installation of the PID sensor into the PID flow chamber The sensor adapter is held in the PID flow chamber with two O rings which also seal around the sensor adapter PID Sub PCB The PID sub PCB is a circui
305. w flat washer lock washer There are no user serviceable parts on the infrared sub PCB EAGLE 2 Operator s Manual Description 253 Start Up and Normal Operation For instructions to startup and use an EAGLE 2 that includes an infrared hydrocarbon sensor reference Start Up on page 22 Measuring Mode Normal Operation on page 29 and Measuring Mode Alarms on page 35 0 100 LEL 2 0 30 0 vol Autoranging The infrared hydrocarbon sensor can be factory set to detect gas in a 0 100 LEL configuration or an autoranging configuration The autoranging configuration detects gas on a 0 100 LEL and a 2 0 30 0 vol scale The gas reading will be displayed in LEL until the gas level reaches 100 LEL or 2 0 vol for isobutane Once the gas reading is above 100 LEL it is displayed in vol Alarm points exist for the LEL range but do not exist for the vol range Catalytic LEL Sensor Screen When either a TC sensor or an infrared combustible sensor is installed in an EAGLE 2 along with a catalytic combustible LEL sensor the user has the option of turning off the catalytic combustible LEL sensor in Display Mode If the unit is going to be used for sampling known high levels of combustible gas or in areas with known catalytic sensor poisons such as silicone vapors the catalytic combustible sensor should be turned off Even though this sensor has its own protective shut off exposure to high levels of combustible gas c
306. w sample from as far as 125 feet from the EAGLE 2 CAUTION Sample hose lengths of more than 125 feet are not recommended for the EAGLE 2 because of flow rate reduction and increased response time Consult RKI Instruments Inc for sample hose lengths longer than 125 feet Internal Tubing The flow system includes polyurethane tubing to route the sample between the various components of the flow system The internal sample tubing is not user serviceable Differential Pressure DP Switch The DP switch is inside the front of the bottom case It senses the EAGLE 2 s flowrate by monitoring the pressure drop between points in the flow system When the flowrate becomes too low for safe operation of the EAGLE 2 a set of contacts inside it open and the EAGLE 2 indicates a low flow alarm Sensor Chamber A PVC block in the bottom case is configured to accept the four gas sensors It routes the sample to each sensor The LEL sensor and the oxygen sensor are retained in the sensor chamber by brackets The CO and H3S sensors are each pushed past two sealing O rings into the chamber and are retained by 14 Instrument Description EAGLE 2 Operator s Manual the O ring compression force Charcoal Filter The charcoal filter is located in the front of the flow chamber next to the CO sensor It contains activated charcoal The CO sensor will respond if exposed to H S and certain hydrocarbon gases The charcoal filter scrubs these gases out of the s
307. with an inerting gas the 1 1 dilution fitting adds sufficient oxygen by blending ambient air with the incoming sample The standard dilution fitting dilutes at a ratio of 1 1 one part air to one part sample The dilution fitting is not an integral part of the float probe 288 Description EAGLE 2 Operator s Manual assembly and must be installed on the EAGLE 2 inlet fitting before installing the float probe assembly when measuring gas in an inert atmosphere To attach the dilution fitting 1 Attach the dilution fitting s male quick connect fitting to the EAGLE 2 s inlet fitting 2 Attach the float probe assembly to the opposite end of the dilution fitting NOTE When using the dilution fitting multiply the combustible gas reading LEL or PPM by 2 to determine the actual combustible gas concentration Always remove the dilution fitting or seal the dilution hole with your finger to measure for oxygen Start Up and Operation 1 Press and briefly hold down the POWER ENTER RESET button Release the button when you hear a beep 2 The LCD will show the following screen for about ten seconds GAS MONITOR RKI INSTRUMENTS EAGLE 2 3 The Connect Float Probe Screen appears for a few seconds prompting you to confirm that the float probe s quick connect fitting is connected to the EAGLE 2 s inlet fitting and that the float probe s plug is connected to the EAGLE 2 CONNECT FLOAT PROBE
308. y slightly different screens EAGLE 2 Operator s Manual Entering Calibration Mode 57 Calibrating Using the Auto Calibration Method This method allows you to calibrate the CH LEL sensor oxygen HS and CO sensors simultaneously It is designed for use with the RKI 4 gas calibration cylinder and is the quickest and easiest method to calibrate the EAGLE 2 Setting the Fresh Air Reading 1 While in the Calibration Mode Screen move the cursor to the PERFORM AIR ADJUST menu item by using the RANGE W SHIFT button CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION gt PERFORM AIR ADJUST NORMAL OPERATION 2 Press and release the POWER ENTER RESET button The following screen appears PERFORM AIR ADJUST 3 Press and release the AIR A YES button to continue If you do not want to continue press the DISPLAY ADJUST NO button and the unit will return to the Calibration Mode Screen 58 Calibrating Using the Auto Calibration Method EAGLE 2 Operator s Manual 4 The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds then indicate that the operation is complete before returning to the Calibration Mode Screen ADJUSTING ZERO RELEASE AIR BUTTON ADJUSTING ZERO COMPLETE CALIBRATION MODE AUTO CALIBRATION SINGLE CALIBRATION gt PERFORM AIR ADJUST NORMAL OPERATION Performing a Span Adjustment in Auto Calibration
309. y the top case down next to the bottom case to allow access to the flow system Locate the ESM 01 sensor It has a twelve wire cable with a connector that mates to an ESM O1 sub PCB that is installed on the main PCB and is normally located next to the pump Figure 35 on page 180 shows an ESM 01 sensor in a typical location Unscrew and remove the two screws that hold down the ESM O1 sensor bracket Grasp the sensor firmly and pull it out of the ESM 01 flow chamber Rock it back and forth gently if necessary to pull it out Take care not to pull the cable from the sub PCB Rotate the locking lever counterclockwise on the cable connector that mates to the ESM 01 sensor to unlock it 10 Unplug the ESM 01 sensor from the cable 11 The ESM 01 sensor consists of an electrolyte reservoir assembly retained 200 Maintenance EAGLE 2 Operator s Manual in the sensor body by a threaded collar on the connector end of the sensor Unscrew the collar from the sensor body a Sensor Body Groove H2S Scrubber Electrolyte Reservoir Assembly Collar Figure 36 ESM 01 Sensor Component Location 12 Grasp the connector and lift the electrolyte reservoir assembly out of the sensor body 13 Turn the body upside down to remove the HS scrubber If it does not come out push on it from the top of the upside down body 14 Discard the old H5S scrubber 15 Carefully place the new H2S scrubber in the end of the sensor body
310. ylinder s label not zero 15 To adjust the values on the screen press and hold the RANGE W SHIFT button then press the DISPLAY ADJUST NO button and release both The following screen appears with the cursor next to CH4 ADJUST AUTO CALIBRATION VALUES gt CH4 50 LEL OXY 12 0 vol H2S 25 0 ppm CO 50 ppm y 16 Place the cursor next to the channel whose gas value you want to change using the AIR A YES and RANGE YW SHIFT buttons 17 Press and release the POWER ENTER RESET button to select the 154 PID Calibration EAGLE 2 Operator s Manual channel The calibration gas value begins to flash 18 Use the AIR A YES and RANGE YW SHIFT buttons to adjust the calibration gas setting to the desired value NOTE The calibration gas value cannot be set lower than the low alarm setting If the calibration gas value listed on the calibration cylinder is lower than the current low alarm setting enter Setup Mode and change the low alarm setting See Updating the Alarm Point Settings on page 114 for instructions If you need to change the alarm point setting only to perform a calibration make sure that you change the alarm point setting back to its original value once the calibration has been performed 19 Press and release the POWER ENTER RESET button to save the change The calibration gas value stops flashing 20 Repeat Step 15 through Step 19 for any other channels that need to be changed 21 When you are do
311. you cannot retrieve any data previously stored in the data logger The Data Logging screen displays the time remaining until the data logger memory is full and asks if you want to clear the data logger memory DATA LOGGING 200 HRS REMAINING ERASE LOG MEMORY To return to Measuring Mode while at the Data Logging Screen press and release the DISPLAY ADJUST NO button To clear the data logger memory do the following 1 With the Data Logging Screen displayed press and release the AIR A YES button The following screen appears asking you to confirm that you want to clear the data logger memory LOG MEMORY CLEAR DATA LOG 2 If you do not want to clear the data logger memory at this point or at Step 3 below press and release the DISPLAY ADJUST NO button The unit will return to Measuring Mode 52 Display Mode EAGLE 2 Operator s Manual 3 If you want to clear the data logger memory press and release the AIR A YES button The following screen appears asking you to reconfirm that you want to clear the data logger memory LOG MEMORY ARE YOU SURE YOU WANT TO PERMANENTLY ERASE DATA FROM MEMORY 4 If you want to clear the data logger memory press and release the AIR A YES button The unit will display the following screen as it clears the data PLEASE WAIT 5 The periods below PLEASE WAIT will disappear one at a time from right to left as the data is cleared
Download Pdf Manuals
Related Search
Related Contents
Mode d`emploi Manuel d`utilisation du SEFRAM 9710 Ps5005 GB-NL-FR-D Philips AX5003/10 User's Manual 高精度ハンディ圧力表示器 モデル CPH6400 製品仕様書 Peavey Unity Passive User's Manual Philips InStyle Suspension light 69053/48/16 Copyright © All rights reserved.
Failed to retrieve file