USER MANUAL

image

Contents

1. POWER VAC ASPIRATOR Model No PM61 shown PM61E SAVE THESE INSTRUCTIONS ACAUTION Federal USA law restricts this device to sale by or on the order of a physician PRECISION MEDICAL 300 Held Drive Tel 001 610 262 6090 Northampton PA 18067 USA Fax 001 610 262 6080 www precisionmedical com CONTENTS RECEIVING 2 INTENDED USE noniine a tiaa 2 READ ALL INSTRUCTIONS BEFORE USING 2 SAFETY INFORMATION WARNINGS AND 2 Saa AON E E E E E 4 COMPONENT IDENTIFICATION ooeec 5 GROUNDING 6 OPERATING 6 CLEANING MAINTENANCE 7 RETURNS A aE 8 DISPOSAL INSTRUCTIONS 8 TROU BEESHOO TING 9 REPLACEMENT PARTS 9 LIMITED WARRANTY c c0ccccceecceceeeeseececeueeseeeeeseeeeeeeeeaneeeees 10 POWER VAC RECEIVING INSPECTION Remove the Precision Medical Inc Power Vac Aspirator from the packaging and inspect for damage If there is any damage DO NOT USE and contact your Provider INTENDED USE To supply a continuous vacuum source within the state
2. Suction Tubing 1 Defective vacuum gauge 1 Clogged Hydrophobic Filter assembly 2 Canister overfilled a Contact your Provider 1 Replace Hydrophobic Filter assembly 2 Empty Canister Unable to adjust s Defective Vacuum Contact your vacuum control Provider REPLACEMENT PARTS LIST Description Part Patient Suction Tubing 1955 Inline Hydrophobic Filter amp Tube Assembly 502690 Inline Hydrophobic Filter 1690 Suction Canister 2000 cc with lid 502687 Small Filter 502391 Large Filter 502471 Fuse 503240 POWER VAC BAVA W Ny LIMITED WARRANTY AND LIMITATION OF LIABILITY Precision Medical Inc warrants that Power Vac Aspirator the Product will be free of defects in workmanship and or material for the following period Five 5 years from shipment Should any failure to conform to this warranty appear within the applicable period Precision Medical Inc shall upon written notification thereof and substantiation that the goods have been stored installed maintained and operated in accordance with Precision Medical Inc s instructions and standard industry practice and that no modifications substitutions or alterations have been made to the goods correct such defect by suitable repair or replacement at its own expense ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES The representatives of Precision Medical Inc or any retailers ar
3. Switch Maximum Volume 2 000 cc with respect to protection from electrical shock degree of protection against electrical shock degree of protection against ingress of liquids continuous This Power Aspirator meets the EMC requirements of EN55001 1 1998 Specifications are subject to change without prior notice POWER VAC COMPONENT IDENTIFICATION On Off Switch Vacuum Gauge Suction Canister Control Knob Inline Hydrophobic Filter Air Filters AC Inlet Fuse Holder Power Cord Power Cord Hour Retainer Meter ACAUTION If for any reason any label becomes illegible or lost contact Precision Medical Inc POWER VAC GROUNDING INSTRUCTIONS The Power Vac Aspirator should be grounded In the event of an electrical short circuit grounding reduces the risk of electrical shock by providing an escape wire for the electrical current This Aspirator is equipped with a cord having a grounding wire with a grounding plug The plug must be plugged into an outlet that is properly installed and grounded ADANGER Improper use of the grounding plug can result in a risk of electrical shock AWARNING Ground reliability can only be achieved when the power cord is connected to an equivalent receptacle marked Hospital Grade OPERATING INSTRUCTIONS CAUTION Inspect the Power Vac Aspirator for visual damage DO NOT USE if damaged 1 Position the Aspirator near a
4. properly grounded electrical receptacle Allow a minimum of 4 in 10 cm clearance on all sides to ensure adequate air flow for cooling of the Aspirator 2 Make sure the On Off Switch is in the OFF position 3 Connect the Inline Hydrophobic Filter as follows e Connect the short tubing to the outlet on the pump Install Inline Hydrophobic Filter make sure filter is oriented per directions on filter e Connect one of the long tubing to the filter and other end to the elbow fitting e Connect the elbow fitting to the center of lid on Suction Canister NOTE The Inline Hydrophobic Filter prevents liquid from entering the vacuum pump ACAUTION DO NOT run the Power Vac Aspirator without the Inline Hydrophobic Filter This will void the warranty 4 Plug the female end of the Power Cord into the A C inlet on the back of the Aspirator 5 Plug the male end into a grounded receptacle 6 Push the On Off Switch to the ON position 7 Block the patient connection on the Suction Canister and adjust the Vacuum Control Knob to the desired vacuum To INCREASE vacuum Turn knob CLOCKWISE To DECREASE vacuum Turn knob COUNTERCLOCKWISE Vacuum should be set as prescribed by your Physician or Healthcare Professional ASPIRATOR 8 Connect one end of the Patient Suction Tubing to the Patient Tubing Connection on the Suction Canister 9 Connect the other end of the Patient Suc
5. as been dropped or damaged DO NOT use near any type of flame or flammable explosive substances or in the presence of a flammable anesthetic e DO NOT allow the Aspirator to exceed Operating or Storage Temperature specifications e Keep the cord away from heated surfaces AWARNING e Connect the Aspirator to a properly grounded outlet See GROUNDING INSTRUCTIONS e Close supervision is necessary when this Aspirator is used by on or near children or invalids Use this Aspirator only for its Intended Use as described in this manual DO leave the Aspirator unattended while operating DO cover air vents Restriction of air flow will cause unit to overheat causing the thermal reset to shut down the Aspirator POWER VAC SPECIFICATIONS Operating Vacuum Range Dimensions Length Width Height Weight Shipping Weight Electrical 0 26 inHg 0 660 mmHg 15 9 in 40 3 cm 7 4 in 18 7 cm 8 1 in 25 7 cm 21 Ibs 9 6 kg 25 108 11 3 kg PM61 108 VAC to 132 VAC 60 Hz 5 Amps PM61E 207 VAC to 253 VAC 50 Hz 2 4 Amps Operating Temperature Range 50 F to 104 F 10 C to 40 C Storage Temperature Range 4 F to 104 F 20 C to 40 C Humidity Filter Safety Systems Suction Canister Equipment Classification Class Type B IPXO Mode of Operation Maximum 95 Non condensing Inline Hydrophobic Automatic Resetting Thermal Limit
6. by your Healthcare Professional or Healthcare Facility protocol AWARNING If the Aspirator is used for treatment of infectious disease biohazard precaution should be followed Contact your Provider Please Recycle LY TROUBLESHOOTING If the Power Vac Aspirator fails to function consult the Troubleshooting Guide below There are no serviceable parts inside DO NOT open case If problem cannot be solved consult your Provider Problem Probable Cause Remedy The Aspirator will not operate 1 On Off switch is in the OFF position 2 Power Cord is not connected properly 3 Thermal limit exceeded 4 Blown Fuse 1 Push On Off switch to the ON position 2 Check plug at wall receptacle and AC Inlet on the back of Aspirator 3 Shut the Aspirator OFF Wait 15 minutes check that air filters are clean and clear Replace Fuse With Compressor running no vacuum at Patient Connection Tubing a Control Knob completely turned counterclockwise 2 Hydrophobic Filter assembly not connected properly 3 Suction Canister not connected or assembled properly 4 Defective Suction Canister a Increase vacuum with Control Knob 2 Confirm and tighten connections 3 Confirm and tighten connection assembly 4 Replace Suction Canister No vacuum on gauge vacuum at Patient Connection Tubing Vacuum on gauge no vacuum at Patient
7. d operating vacuum range to aspirate fluids in medical suctioning procedures by trained healthcare professionals READ ALL INSTRUCTIONS BEFORE USING This manual instructs a Professional to install and operate the Power Vac Aspirator This is provided for your safety and to prevent damage to the Aspirator If you do not understand this manual DO NOT USE the Aspirator and contact your Provider SAFETY INFORMATION WARNINGS AND CAUTIONS ADANGER Indicates an imminently hazardous situation which ifnot avoided will result in death or serious injury AWARNING ndicatesa potentially hazardous situation which if not avoided could result in death or serious injury ACAUTION ndicatesa potentially hazardous situation which if not avoided may result in minor or moderate injury CAUTION Used without the safety alert symbol indicates a potentially hazardous situation which if not avoided may result in property damage CONSULT ACCOMPANYING DOCUMENTS PE Ground Fuse Type B equipment No Alternating Current ASPIRATOR ADANGER TO REDUCE RISK OF BURNS ELECTROCUTION FIRE OR INJURY TO PERSONS e Always unplug the Power Vac Aspirator after using e DO use the Aspirator if exposed to water Unplug immediately if immersed in water DO NOT attempt to repair the Aspirator Any attempt could result in electrical shock DO use if the Aspirator or the power cord h
8. dical com ISO 13485 Certified 502603 07 23 09 PRINTED IN USA
9. e not authorized to make oral warranties about the merchandise described in this contract and any such statements shall not be relied upon and are not part of the contract for sale Thus this writing is a final complete and exclusive statement of the terms of that contract THIS WARRANTY IS EXCLUSIVE AND ISIN LIEU OF ANY WARRANTY OF MERCHANTABILITY FITNESS FOR A PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY WHETHER EXPRESS OR IMPLIED lt gt gt 5 lt gt lt 2 gt lt lt 5595 23 Precision Medical Inc shall not under any circumstances be liable for special incidental or consequential damages including but not limited to lost profits lost sales or injury to person or property Correction of non conformities as provided above shall constitute fulfillment of all liabilities of Precision medical Inc whether based on contract negligence strict tort or otherwise Precision Medical Inc reserves the right to discontinue manufacture of any product or change product materials designs or specifications with out notice Precision Medical Inc reserves the right to correct clerical or typographical errors without penalty POWER VAC ASPIRATOR PRECISION MEDICAL Precision Medical Inc 300 Held Drive Northampton PA 18067 USA Toll Free Tel 800 272 7285 Toll Free Fax 800 352 1240 Tel 001 610 262 6090 Fax 001 610 262 6080 www precisionme
10. emove Air Filters from their recessed areas and wash in warm soapy water and rinse 3 Absorb excess water from filters with a dry towel 4 Return filters to their original locations AT 8500 HOURS Return the Power Vac Aspirator to your Provider for compressor maintenance including pump Call Precision Medical Inc for Preventative Maintenance Services POWER VAC CAUTION e Excessive dust build up on external filters will reduce the performance of the Aspirator If this occurs clean or replace with new filters DO use where saline mist could be drawn into the Aspirator The Air Filters do not filter out saline Saline causes rapid deterioration of the Aspirator and will void the warranty EXTERNAL CLEANING 1 Disconnect all connections before cleaning 2 Clean exterior surfaces of the Aspirator with a cloth dampened with mild detergent and water 3 Wipe dry with a clean cloth 4 Store the Aspirator in a clean area free from grease oil and other sources of contamination RETURNS Returned products require a Returned Goods Authorization RGA Number contact Precision Medical Inc All returns must be packagedin sealed containers to prevent damage Precision Medical Inc will not be responsible for goods damaged in transit Refer to Precision Medical Inc Return Policy available on the Internet www precisionmedical com DISPOSAL INSTRUCTIONS Dispose of Bio waste products as prescribed
11. tion Tubing to the suction catheter and begin procedure as instructed by Healthcare Provider NOTE DO NOT overfill the Canister 10 When procedure is completed press the On Off Switch to the OFF position 11 Empty the Suction Canister as follows e Disconnect patient tubing and elbow fitting from the top of the Canister e Remove the Canister from its bracket and empty as instructed by Physician or Healthcare Provider 12 Reassemble Suction Canister and Tubing in reverse order CAUTION Canister is single patient use only CLEANING MAINTENANCE AWARNING e Turn the Power Vac Aspirator OFF and unplug from the electrical receptacle before cleaning DO NOT immerse the Aspirator into liquid Inline Hydrophobic Filter This filter cannot be cleaned If it becomes contaminated clogged or wet it must be replaced It is recommended the tubing from the canister to the filter be replaced when the filter is replaced To replace Inline Hydrophobic Filter 1 Disconnect filter from short and long tubing 2 Dispose of filter as directed by your Healthcare Professional or your Healthcare Facilities protocol 3 Install new filter Part 1690 Make sure filter is oriented per directions on filter Then reconnect tubing and elbow Tubing is single patient use only DO NOT clean WEEKLY 1 Check both Air Filters on top rear of cabinet for dust build up 2 If dust build up occurs unplug the Aspirator r