1. When reach its standard working life the treatment effects will be not so obvious so we suggest a change of new treatment handles Besides the treatment handles that reached their standard working life will not be included in our guarantee 19 6 1 Skin test Before receiving treatment the client has to do a skin test The content is as the following 1 If do whole face treatment pls choose ear root part as the test area If it is chest neck or other body areas pls choose the parts that will not affect appearance as the test area 2 When the testing area is chosen pls clean it and cover a thin layer of gel and then make the treatment handle fully touch the testing area shooting 2 3 light spots 1 3 light spots overlap is acceptably After2 3 minutes pls observe the testing area If a slight redness could be seen it indicates that the used energy is suitable and the operator could use this energy and parameter to do treatments for the client If nothing happens on the testing area it indicates a lower energy and the operator should increase energy 1 2J every time If an excessive redness occurs immediately it indicates a super high energy and the operator should reduce the energy 3 If doing hair removal a cut of hair to 1 2mm is a first thing and then clean the target area Shooting the treatment area with trial light spots if a slight redness and a burning of the hair on the whole area occur it indicates the energy
2. interface shown as pic 3 Note When water circulation has problems the system will alarm Error 1 will appear which shows the water circulation is not smooth or little and the equipment can not work normally The pic in the operation interface show the actual flow size if it is o It means that the water 13 circulation is not smooth at all If the working temperature is higher than the set temperature alarm set alarm temperature is 60 C the system will also alarm Error 2 will appear indicating the temperature is too high not suitable for treatment Parameter Function HR J cm i Counter pP information Pic 3 The picture 3 is operation interface show the runing state and parameters including working model water flow energy pulse quatity and pulse width also can set the system running parameters change system running state The client can set the parameters as following Function select 4 kinds of functions HR SH VR PR click the corresponding function the system will go into the related function and do treatment The operator can also adjust corresponding parameters according to actual situation and save Reaching to better result Aly Energy adjust click the adjust botton he up arrow representative f Jem increase the down arrow representative decrease Som this botton shows Energy adjust the adjusting rage is 10 60J cm2 14 Width parameters asju
3. removal etc The treatment theory is the selective absorption effect The system is controlled by MCU It is composed of mainframe touch screen handles Operation and controlling part is in the mainframe The controller in it can monitor the machine comtinuously Controlling system TFT touch screen key switch and emergency stop etc Power source Supply the voltage the core part of the machine Waterway Help the treatment handle cooling Power warehouse store and release voltage Screen Display and adjust working mode and parameter Treatment handle Xenon crystal switch and transfer tools 1 2 operation Introduction Surplus for limbs all over the body such as the torso armpit hair removal Changes of aging skin The treatment of pigment pathological changes Chapter 2 Safety 2 1 Electrical and mechanical safety Considerable thought was given during the design and producing of the system to minimize the hazard for both operator and clientele The system requires 220 V single phase power source the strongest output Current is not beyond 10A On the maintenance panel Automatic electric leakage was installed when electric is leaking it will protect the system Open the machine again you need raise up electric leakage device Dangerous voltage presents inside the console during course of the treatment After turning off the system dangerous voltage still left Removing the covers creates a safety hazard 2 2 Movement and Tran
4. Main Structure HR handle Touch screen Key switch Chapter 3 System Description p Ka SR handle VR handle Emergency stop Handle pull W l ETE 7 a a Vents Water level observation window The structure is shown above touch screen emergency stop key switch treatment handles etc 3 2 Component Description Touch screen show and adjust the info of operation system Emergency stop Close the power source when urgent thing happened Key switch open or close the machine Handle pull carry the machine Main engine cooling system power supply controlling system integrated circuit etc HR handle changeable painless remove all kinds of hair SR handle changeable skin rejuvenation skin tightening VR handle changeable vascular removal Filler add pure water into the machine Water level observation window monitor the water yield Vents adding drawing off the water open it then you can add water Outlet let out the water Power line connection connecting the power line change fuse Protection switch protect the safety of the machine HR Connection install HR handle SR Connection install SR handle VR Connection install VR handle Chapter 4 Installation and Commissioning 4 1 installation The system can be installed in hospitals medical clinics and beauty salons little preparation is needed before installation When you purchase our company s instruments you
5. Three handles IPL machine NBW I323 User Manul A NUBWAY Beauty Equipment Table of Content Table of Content LDV SUSI OV ELV OW oraaa EE ANE AE EO ONERE 1 PL y stems IMTOdU Ct Onara ante a aaa OL 12 Operation IntroduciiOiiesinnnenn annn a a 1 P SALCLY A AE A A T A 2 1 Electrical and Mechanical Safety ccccccccsssssssssscccccsssssssssssssesesesssscececceees 1 2 2 Movement and Transportation Safety cccccccccccccccceeceeeeeeeeeeeeeeeesesssssseeees 2 BO OPC safety cas cvatucias seedenstiase anssdendsiudac ead A eneesesaeane 2 DA NV ATIVE S adecceswsisrsdehen E E a 72 Oy SVStem Descriptio Meannan a a tetanic eae SC Do Main Struct UFE sicacsen sie eaa coo ntenev anette cau tewsanie nade seis eae ee eras 3 3 2 COMPONENTS Description eseesressserseescessssssssssesssssseeseeseeceseescesesssssess A 6 4 Installation amp Commissioning seeseeeeseeeeeeeesssosscrreeessserreressssssereeeesss A Tasta On ranea a a Osea soe T eats 6 A CON MIS SONIN in T E E E N A 1 2 pe Technical ParametenS iein a N a a 18 Bot POC MICHAL Paranie e Srann E a T T 18 5 2 Relevant requirement Seniai E TA 19 6 Operation PreceGUres siscctiscsesatecseestedsciedassieiniasee tiie O 19 OLOKUE CS sahcaietatelae vasa ba ctiacegeat iesane T eo aieee 20 6 2 Treatment COU Ceannain rset tastes esas esas sees 21 gt TS Ta Sass stoerce eer EO nase ae edo ican EOE 21 OGA POW CE Ollora he ele seta ese se eee eee 7 Treatm
6. can enjoy our free onsite installation and commissioning services The system should be installed and commissioned by professional and technical personnel authorized by our company Installation work includes 4 1 1 check equipment list Equipment lists Mainframe 1 Hr treatment handle 1 SR treatment handle 1 VR treatment handle 1 Switch key 2 Protective glasses 1 Power cord 1 Rubber ring 4 Spare fuse 2 Package box 1 Water funnel 1 4 1 2 Be familiar with the installation requirements 4 1 2 Be familiar with the installation requirements Before unpacking the Instrument you should check whether the accessories are complete the installation location meets the requirements of this section The system should be installed away from heat vents and keep an appropriate distance from other objects the temperature should keep about 20 C The instrument is preset the rated voltage of the installation place before delivery according to the user s requirements The corresponding electrical equipment requirements are as follows Single phase 220V 10 10A 50Hz 110v 10 The equipment lands the ground through the ground wire of the power sheath passing the wall line Good ground connection is essential for safe operation The input voltage can not have transients voltage and current spikes collapse or surge To ensure that the system is in the best condition the room temperature shou
7. djustments of Ip Hair removal can protect epidermis and remove thin and thick hairs 6 Square large spot can remove hair quickly and increase treatment speed Chapter 8 Indications and contra indications 8 1 Indications Pigment removal Speckles age spot sun induced freckles and all kind pigment sourced from cosmetics most effective Blood vessel removal Red blood streak bottlenose erythematic spot and so on Anti ageing Wrinkles acne scar low skin elasticity Trichopore and so on Hair removal on limbs body face and lips 8 2 Contra indication E Pregnant woman Using photosensitive medicine Recently tanned skin within the last month History of keloid scarring Diabetes unless under control Cancer particular skin cancer Heart disease epilepsy Chapter 9 Post treatment care 9 1 No exposure to sunlight Clients should use highest factor sunscreen and protect the treated area from exposure to sunlight sunbed or other artificial tanning for one month following treatment Tanning after treatment sessions may enhance melanin regeneration which may result in hyper pigmentation 9 2 Free from always touching and pressing Clients must do not press and scratch the treated area against infection 9 3 Makeup Makeup immediately after the treatment is not allowed for better curative effect advising no use of makeup for four to five days following treatment The skin is sensitive during this period Remove the makeup when s
8. doing treatment pls touch the treatment area tightly but avoid pressure on the area 6 When doing treatment the light spots should be continuous 1 3 overlap is acceptable 7 Itis allowed to change the direction of treatment handle to achieve full treatment 8 For darker skin D and D2 should be lengthened appropriately 9 For whiter skin D1 and D2 should be shortened appropriately 6 4 Power off Put the treatment handle back to its place Push the button Ready to Standby turning off the handle light Turn the key switch to OFF Clean the light guide crystal with cotton or soft cloth Chapter 7 Treatment principle and features 7 1 Treatment principle The Intense Pulse light delivered by 1323 can penetrate into deep skin and take effects on the targeted tissues After an absorbtion of energy the pigment cells will be broken and absorbed by the body gradually When the blood vessels are exposed to intense light the hemoglobin will selectively absorb the energy The photo thermal effect makes the blood vessels closed and then absorbed by the body gradually The normal skin seldom absorbs the delivered energy so the normal skin will not be hurt The Intense Pulse light takes effects on pigments and blood vessels Meanwhile it can stimulate the proliferation of collagen and improve the rearrangement of elastic fiber finally achieving the purpose of skin rejuvenation wrinkles removal and pores shrinki
9. ent Principle and Features ccc cee cee cee cee cee eee sesse 22 Fol A POALMMOTIC F PINCID Ceiri i uan e a i 22 PER E E E E E a CS E E NE E E E AA E E ade ELI I A E OO 8 Indications and Contraindications cccccccccccccescescesscescscescesseeseess 24 Sande ON E A AS 24 8 3 Contra IN GICAL ONS aerar E sas oensauatedodascuaes 24 O P st treatment Cale carna aaa a aa a a 10 Maintenance and troubleshooting e cee csccccsecseecessesceessessseees 25 LOA Clea MINS the AIT Msveiicsetersast cocvnccaweitts dace E E Ea 25 10 2 Cleanine the cista eoan a naa eaaese alco sean ade 25 10 9 Troubleshooting solutio Nresnneurieeeneiri e saa ees 25 IOA Trouble SOO EO asrorini e e e a an 25 10 5 Error TMLOPMA ION eessen e 26 10 6 The cooling system Maintenancce sssssseseeeececcecccccecceeeeeeeeeeeeeaaeas 27 10 7 ouch Screen Calibrations octet tenia a a A IARC ORIOL Usa cncdrananctalenutsataucnceroana tant asuni cae aan aimae Ch comer NE eee TIE Chapter1 System Overview Thanks for your purchasing NBW I323 Intense Pulse hair removal machine Believe that the machine can give you a believable reliable function and satisfied repay This is the introduction of NBW I323 machine in this chapter Note The person operating or keeping it must read this manual carefully before using 1 1 System Introduction Name Intense pulse hair removal machine skin rejuvenation vascular removal spot skin Whitening hair
10. gments group of the diseased tissue is much more than the normal tissues Its temperature rise after a absorption of lights is higher than normal skin Using the temperature difference between them to close diseased blood vessels and break down pigments without hurting normal tissues 1323 Treatment scopes Vascular disease telangiectasias reddening of skin and red nose caused by rosacea Pigmentary changes freckles skin spots poikiloderma sun damage and skin aging pores rough skin lack of elasticity and acne Improve facial skin aging restore skin s luster and elasticity and regain youth This machine could be used to treat black spots and black eye shorten the recovery time of laser resurfacing This machine could do hair removal This machine could treat face neck chest hands and 4 limbs 7 2 Features 1 IPL Hair removal is the less pain compared with all the other hair removal methods Clinical cases indicates most patients just feel like they are been beating by a rubber band 2 The advantage of IPL Hair removal is its permanent hair removal Using the theory that Intense pulse light could penetrate into dermis and subcutaneous fatty tissue take effects on the deep hair follicles of different areas to remove the hair on any area 3 IPL Hair removal has the advantage of no injury to skin 4 IPL Hair removal has the advantage that its pigmentation after hair removal is closing to our skin 23 5 The long pulse a
11. ine is not finished ERROR 4 Water temperature is overheat not suitable for further treatment 16 The system status display from left to right they are water flow button water temperature alarm button precombustion button Water flow button use to show the size of the flow of water in real time Water temperature alarm button hen the temperature of water is too high the icon will be red Bie the system will display ERROR 1 Precombustion button when you click precombustion button Ell will appear when the Xenon lamp precombusting successfully Charging complete button after the Xenon lamp precombuste successfully the machine is charging while finish charging you will see this the button M Start working After you set the right treatment parameters press the button to be Standby When the machine finishes precombusting and charging the button you will see is L then the system enters the treatment status At this time you can not change the parameters press the trigger button on the treatment handle the corresponding handle will flashes according to the set parameter and the buzzer will beep The total count and single count will be increased on the screen When you release the trigger button the flashes will stop Press the button to be wen the system will enter standby status when you can modify the treatment parameters if you need 17 Note If the machine is not flashing prohibit s
12. is suitable if no redness and a burning of the hair on the part area occur it indicates the energy is low and the operator should enhance energy Generally the energy of trial light spots starts from 15J cm2 Reference paramters Treatment Interval white skin Yellow skin times time 25 35J c 20 30J em 15 25J cm 4 6 times 30 days m2 2 2 20 22 28J c 20 25J em 18 25J cm 3 5 times 21 days m2 2 2 VR 25 30J c 20 25J em 18 25J cm 2 4 times 30 days m2 2 2 Pulse reference set Notes It is not suggested to do treatments on the hair whose color is lighter than the surrounding skin 6 6 1 2 3 2 Treatment course 1 It is suggested to do 4 6 courses every 28 days a course according to hair growth cycle and different treatment areas 2 After a month the patients can provide their feedback to the clinics If needed a second treatment could be done to achieve better results Notes After the hair is removed it will have some burning sensation within 3 hours But pls do not worry It does not need any special care generally it will disappear after 3 hours 3 Attentions Completely protect the client s eyes when doing treatment When doing treatment pls starts from the thicker skin and then extends to the surrounding A second scan on the lesion area is acceptable 21 4 It s better to avoid eyelashes and do not repeat the light spots when treat forehead 5 When
13. ld be maintained at about 20 C relative humidity should not exceed 80 The thermal power of the system is about 200W during work so it s the best if the treatment rooms are air conditioned Do not allow the instrument being irradiated by prolonged direct sunlight nor being placed near a heater 4 1 3 installation steps 1 Carefully open the package of the machine place the machine in a preselected position 2 Take out the handle bracket and install it in the corresponding position check whether the connection is good or not shown as below picture Note intall the handles first and then adding water 3 Plug in the power cord 4 Insert the switch key and the Chassis key 5 Fulfill water A First prepare 8 liters distilled or purified water Running water or mineral water is not allowed B Unscrew the waterinlet of the machine C Unscrew the overflow vents at he back of the machine You can fill in water only when the overflow vents is open D Fill in water with the water funnel delivered along with the machine E Fill in water until the overflow vents overflows shows the water is enough You can also observe by the water level viewing window 10 F Do not tighten the overflow mouth at first please wait until the machine finish debugging then tighten the overflow mouth The inside of the machine is a closed water loop if tighten overflow mouth at first there will be air inside the wa
14. ng Intense pulse light hair removal is different from laser hair removal The delivered light is broad spectrum 560 1200nm intense pulse light Its adjustable scope is wide so it can do permanent hair removal to different colors and different depth of skin hair Its principle is that the pigments in hair follicle and hair shaft has an selective absorption of light and the photo thermal effect will increase the temperature of 22 hair follicles make the hair cells dead and achieve the effects of permanent hair removal I323 has a breakthrough aiming at the common problem of burning skin During treatment each light spot is made of soft pulsed lights the maximum pulse width is 20ms The long pulse width ensures the treated area has a longer time to absorb the light and heat Then a soft treatment can be achieved to improve the treatment effects and greatly reduce the skin burning which can avoid the chromatosis and improve the treatment safety 1323 has 2 working principles Biological stimulation when the intense pulsed light applies to the skin the produced photochemical effect can make a molecular structure chemical change inside the dermis collagen fibers and elastic fibers to restore its original elasticity In addition its photo thermal effect can strengthen the function of blood vessels and improve its circulation so as to achieve the therapeutic effects of removing wrinkles and shrinking pores Photo thermal effect The pi
15. nge the purified or distilled water once a month by outfall you may find it at the down back of the unit You may find 2 parts one is blue another is black push the blue part with your left hand and pull the black part with your right hand the black part will be taken off the unit water out then 10 4Troubleshooting The chart below shows some possible trouble and solution accordingly Check the power line loose or not Machine cannot open Check fuse If add enough water Check if the followed the order of adding water Water warning Add water mouth release air inside Handle working but no Check water cycling 25 some dS Check rubber ring broken or not 1 check if you press Work Button on LCD Handle no flash 2 check if the counter work well if counter not working the handle button broken Check whether the power supply voltage is too low clean the crystal check if the energy you chose is too low Output energy weak check if the handle need to be changed Check if the setting protecting Temperature is too low normal 35C Warning in shot time The environment temperature is too high For other problem please contact us please don t open the machine yourself 10 5 Error information If the system malfunctions during start up or normal operation turn the system off immediately and contact with our service department Only the technician authorized by our company can service the system Otherwise we bear no resp
16. onsibility 10 6 Replace the fuse wire Turn off all power switch unplug the power cord open the fuse holder with a small flathead screwdriver and then take the fuse wire in the fuse holder Take out the spare 15A 250V fuse wire in the accessories sent to you along with the machine j LA kis i 4 3 ki lt a j D 26 10 6 The cooling system maintenance Please check the fan regularly fan damage will cause machines overheating causing the machine doesn t work or even completely damaged Regularly check the cooling water if necessary please add 10 7 Touch Screen Calibration Turn on the machine without any operation continue press the button on the handle 11 times the LCD will appear blue On the up left corner of the LCD you can find a write press the it will run to right press again it will run to the bottom press 3times then reopen the machine to finish it Chapter 11 Contact Us Beijing Nubway S amp T Co Ltd Web www nubway com Email nubway nubway com Address Room 805 No 12 Yabao Rd Chaoyang District Beijing 100020 China Tel 010 10 5120 6522 Fax 010 10 5120 6522 27
17. sportation safety The unit contains delicate components and no allowance of long transportation after first installation and adjustment During the movement the shock absorbing package originated from the manufacturer must be used Otherwise it might cause adverse effect to the system 2 3 Optical safety Intense pulse light with wavelength of 420 1200nm sourced from the system can create safety hazard All personnel including clientele in the treatment room must wear protective glasses whenever the unit is in use Intense pulse light emission presents an eye hazard Take all necessary precautions in areas where the system is being used Never permit reflective objects such as jewelry mirrors and watches to reflect the intense pulse light Never look directly at the intense pulse light emitted from the treatment handle even when wearing the protective glasses Never direct the intense pulse light at any skin except the test patch and the target area 2 4 Warnings gt Allow access to the treatment room only to those essential to the procedure and well trained in the required safety procedures gt Make sure that all treatment room staffs are familiar with the system s control and know how to shut down the system instantly gt Clear identify that no igniting combustible materials present in and around the treatment area gt Make sure the caution signs are posted in a prominent place of the treatment room 3 1
18. st Parameter this botton is parameter display areas The functions as belowing P_AMT FAO output quatity 1 15 P1 PANN first width 1 30ms D1 PAN first delay 5 50ms M tput second width 1 30ms EA second delay 5 50ms Please choose corresponding function buttons above adjust the output parameter through button P Set parameters Press the set button in the operation interface then goes into the parameter set interface Shown as Pic 4 Pic 4 This interface is adjustment of the same pulse width The button is water flow regulation The set range is 5 10 There will be ERROR 1 appearing if the actual water flow is less than the set amount Note Please do not adjust the Water flow alarm value press AE interface button to save the changed parameters and get back to the previous Press ay button not to save the changed parameters get back to the previous interface directly Clear Total Counter Counting display refer counting display button the Total Couner is for the total count Clear is for single count Press Clear can clear the single count System information display referes system informationbutton used to show all warning information Including ERROR 1 Water circulation has problems ERROR 2 Precombustion fails you need cancel the Precombustion and Precombustion again ERROR 3 Charging of the mach
19. taying in the working state for a long time Shut down when the treatment is completed click button the system will return to the initial interface then turn off the key switch is OK Note Do not force the machine shut down during working Chapter 5 Technical Parameters 5 1 Technical specifications This system has multi pulse and multi wavelength its technical specifications is 10 60J cem2 lt 2 8 34mm SR VR 16 57 HR lt 0 2mm Crystal 0 30 C temperature Semi conductor water Air Power supply AC220V 10 15A 50 60Hz 110V 10 25A50 60Hz 5 2 Relevant Requirements Running environment of this system Environmental temperature 10 C 30 C Relative humidity no more than 80 Storage environment of this system Environmental temperature O C 40 C Relative humidity no more than 80 Chapter 6 Operating Procedures Before receiving treatment a low energy skin test should be done first After selecting the appropriate energy covering the treatment area with a thin layer of gel and then making the treatment handle work at the target area pushing the button switch to do treatment repeatedly The output frequency is generally 1p 3s After treatment it is better to build a treatment file for your clients to improve the effects during the following treatments Pls note the working life of treatment handles The standard working life of each treatment handle is about 2 years or 50000 spots
20. ter tank which will make water do not circulate cause machine failure After the above water filling steps please start power and commission Note Please install the treatment handles as shown below Press the Spring as the arrow pointed in order to avoid the rubber ring leaking water During the use dragging the handle is strictly prohibited 11 4 2 Commissioning 1 Make sure the emergency shut off knob is Upward bounce Turn right the key clockwise to start the system then the system starts to work you can hear the sound of water circulating if normal The touch play screen shows as following picture 1 i Pic 1 Pic 1 shows the initialization process of the machine please check whether the the 12 connecting place between the machine and treatment handles is leaking or not If there is water leakage stop the commissioning of the machine immediately Take off the treatment handles check whether the rubber ring inside the handle joint is good or not If broken please replace a new one Note 1 Make sure the handle in place before commissioning Plug out handles is strictly prohibited during treatment Note 2 Make sure the red emergency stop button is up or the machine can not supply power 2 The system will enter language select interface after initialization shown as Picture 2 Select Language Pic 2 3 Please press the pic on this interface then the system will go into the operation
21. welling and reddening appearing at the treated site against infection and pigmentation In the mean time advise client take some medicine such as VitC VitE 9 4 Food Advising not eat the hot and spicy food because the skin is sensitive during this period 24 9 5 Other Advise that improve other causes of disease such as adjusting the endocrine situation by taking Chinese herb of medicine taking few medicines that include VitC and Vit E Caution According to the normal physiological process clients should take the next treatment 21 days later The whole treatment needs 3 5 sessions Due to physical difference between individuals a few of clients may have temporary discomfort such as blister deposit of pigment swelling and so on Advise you to use ice bag to cover the treated area for 30 minutes right after treatment No make up in 5 days after treatment Chapter 1o Maintenance and troubleshooting 10 1 Cleaning the unit Clean the system exterior at least once a week Wipe all surfaces with soft damp non abrasive cloth Mild cleansing solution may be used Not spill liquids in the unit 10 2 Cleaning the crystal on hand piece The crystal on hand piece should be kept clean at all times and should be cleaned after each treatment To clean the crystal on hand piece first dry it with a lint free cloth Then clean it with a cloth moistened with water free alcohol and let it dry thoroughly 10 3 Changing of the cooling water Cha