MAN02 MAN15 MAN30 Manometer
Contents
1. 2 15 Manometer Users Manual Mode d emploi Bedienungshandbuch Manual d Uso Manual de uso 2 MAN15 Manometer Users Manual MANO2 15 30 Rev001 2009 Amprobe Test Tools All rights reserved Limited Warranty and Limitation of Liability Your Amprobe product will be free from defects in material and workmanship for 1 year from the date of purchase This warranty does not cover fuses disposable batteries or damage from accident neglect misuse alteration contamination or abnormal conditions of operation or handling Resellers are not authorized to extend any other warranty Amprobe s behalf To obtain service during the warranty period return the product with proof of purchase to an authorized Amprobe Test Tools Service Center or to an Amprobe dealer or distributor See Repair Section for details THIS WARRANTY IS YOUR ONLY REMEDY ALL OTHER WARRANTIES WHETHER EXPRESS IMPLIED OR STAUTORY INCLUDING IMPLIED WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE OR MERCHANTABILITY ARE HEREBY DISCLAIMED MANUFACTURER SHALL NOT BE LIABLE FOR ANY SPECIAL INDIRECT INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES ARISING FROM ANY CAUSE OR THEORY Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages this limitation of liability may not apply to you
2. Differential Pressure a measurement of the difference between two pressures i e use differential pressure sensor to measure gauge pressure by leaving one process connection open to atmosphere and connecting the second sensor port to your system 5 n ou Press momentarily and the units will cycle through 5 InH20 psi mbar bar mmH20 which are indicated on the bottom of the display For 02 Press momentarily and the units will cycle through bar mmHg ozin2 Kgcm2 psi inH2O kpa ftH20 inHg 20 mbar which are indicated on the bottom of the display See fig 7 amp 8 For MAN15 and MAN30 Press momentarily and the backlight illuminates for approx 30 seconds then turns off automatically Press momentarily to decrease the figure when calibration is being performed CALIBRATION MODE Calibration mode is only applicable for a standard manometer calibrator or any qualified meter calibration facility for annual calibration 1 First manually set the display to zero no pressure applied to the connector refer to the manual zero procedure 2 Turn the meter off 3 Press amp simultaneously CA appears on the display See Fig 9 the meter enters to the calibration mode make sure the units setting on PSI to start positive pressure calibration The meter has defaulted as 1 6 psi for 2 12 00 psi for MAN15 24 00 psi for MAN30 calibration point the adj
3. Repair All test tools returned for warranty or non warranty repair or for calibration should be accompanied by the following your name company s name address telephone number and proof of purchase Additionally please include a brief description of the problem or the service requested and include the test leads with the meter Non warranty repair or replacement charges should be remitted in the form of a check a money order credit card with expiration date or a purchase order made payable to Amprobe Test Tools In Warranty Repairs and Replacement All Countries Please read the warranty statement and check your battery before requesting repair During the warranty period any defective test tool can be returned to your Amprobe Test Tools distributor for an exchange for the same or like product Please check the Where to Buy section on www amprobe com for a list of distributors near you Additionally in the United States and Canada In Warranty repair and replacement units can also be sent to a Amprobe Test Tools Service Center see address below Non Warranty Repairs and Replacement US and Canada Non warranty repairs in the United States and Canada should be sent to a Amprobe Test Tools Service Center Call Amprobe Test Tools or inquire at your point of purchase for current repair and replacement rates In USA In Canada Amprobe Test Tools Amprobe Test Tools Everett WA 98203 Mississauga ON L4Z 1X9 Tel 877 AMPROBE
4. 267 7623 Tel 905 890 7600 Non Warranty Repairs and Replacement Europe European non warranty units be replaced by your Amprobe Test Tools distributor for a nominal charge Please check the Where to Buy section on www amprobe com for a list of distributors near you European Correspondence Address Amprobe Test Tools Europe In den Engematten 14 79286 Glottertal Germany Tel 49 0 7684 8009 0 Correspondence only no repair or replacement available from this address European customers please contact your distributor AAAMPROBE manometer aus 82 88 58 Indicator Pushbutton 6 Record pushbutton e Data Hold pushbutton Differential pushbutton e Unit Selector pushbutton Q Backlight pushbutton Primary data screen displays pressure value Minus pressure display MAX MIN pressure recorded REC starts recording mode and displays max min pressure recorded DC power input jack USB output port USB cable not included H M S 88 88 88 displays time in Hour Minute Second eoo 0000 W Pressure unit indication Low battery indicator DIF Differential pressure mode Positive pressure connection input Negative pressure connection input Data Hold freezes pressure reading MANO2 MAN15 MAN30 Manometer CONTENTS INTRODUCTION eee eo eI epee 2 FEATURES a ean ing alee Aa 2 S
5. measurement of difference of two pressures 5selectable units of measure in H20 psi mbar bar mmH20 For 02 e 11 selectable units of measure bar mmHg ozin2 Kgcm2 psi inH2O ftH2O inHg cmH20 mbar For MAN15 and MAN30 Standard Accessories Please check that the tubing is not leaking or damaged before use STANDARD ACCESSORIES The package contains Manometer 9V Battery Users manual 2connection hoses 4mm ID x 6mm OD x 500mm Carrying Case QUICK START Unscrew battery compartment on the rear of the instrument and insert the battery Replace cover and secure with screw 1 Press 1 to switch the instrument 2 Press QNID to select unit of pressure measurement required For zeroing press for three seconds 3 Press for differential pressure measurement 4 Press to freeze the reading on the display Press again to cancel feature 5 Press to start a recording press again to display MAX value of the recording session press again to display MIN value of the recording session and press again to return to real time recording mode The Instrument records only the Max and the Min values for each session To record all values for a recording session use the Download Suite Software Press and hold for 3 seconds to turn clock feature off Note Clock feature available with gauge pressure only not differential The instrument will automatically switch off after 20
6. TANDARD ACCESSORIES iunea rone a a a Ei entente entente eerte enn 2 QUICK START EE 3 AUTO POWER OER aiina E E 3 OPERATION MODE oaoa A ie sen 3 CALIBRATION MODE 5 E aie AA 4 Calibration point refererice 2 aaaea eee a ae 5 MANUAL ZERO SETTING TROUBLESHOOTING jo 5 Replacing The Battery edel 6 USB PC INTERFACE CAPABILITIES Download Suite Software installation 6 Operation E EN 6 MAINTENANCE eerte de tere eese eS 7 TL WEE H Ki eut e EE H gest ierten EEN a H Congratulations on your purchase of the Amprobe 2 MAN15 MAN30 manometers These instruments are portable battery operated pressure measuring device They are ideal for technicians measuring pressure level Medical equipment Computer peripherals pneumatic controls INTRODUCTION The meter will display all LCD segments when it is first turned on for approx 3 seconds e The is divided into two distinct sections One large primary top screen and one smaller right bottom screen relative Clock The two display areas will constantly update with the pressure measurements FEATURES The meter measures Gauge pressure a measurement of pressure that is referred to ambient pressure Differential pressure a
7. e display to zero no pressure applied to the connector press the GOLD key for 2 seconds The meter should display 0 000 from right to left See Fig 13 Then the LCD display shows normal mode TROUBLE SHOOTING Cannot Power on Check the battery connections Replace the battery or attach Optional AC adaptor Low BAT indication Replace the battery when LCD displays BAT No Display Make sure the battery is functional checkif sleep mode is active Refer to disable sleep mode on page 6 to turn it off check if the tubing is connected to the meter tightly Err 1 For the pressure value exceeding the maximum range Err 1 appears on the display See Fig 14 Do not exceed rated over pressure range of manometer Sensor may be damaged Err 2 For the measurement pressure less than minimum range Err 2 will appear See Fig 15 Err 3 For a differential pressure value larger than maximum display Err 3 appears on the display See Fig 16 Err 4 When you set zero make sure you have disconnected the tubing If you see an 4 appear on the display it means the manometer is damaged See Fig 17 Note Err 4 Will be also appear If the tubing is connected during zero set E10L or E2UL When you see these errors while operating computer software it means pressure source is less or over than the range of the instrument Replace your 9 volt battery when The BAT icon appears on the right o
8. ed temperature range 0 50 32 122 F Operating temperature range 0 50 C 32 122 F Storage temperature range 20 60 C 4 140 F Operating Humidity Max 8096 RH Power Supply 9 volt battery Exceeding Maximum pressure will cause permanent sensor damage Range 0 2 psi MANO2 0 15 psi MAN15 0 30 psi MAN30 Power Supply Conversion amp Resolution esu ers e penes me eos Accuracy 0 3 of full scale at 25 C 77 F Repeatability 0 2 0 5 of full scale Linearity Hysteresis 0 29 to 1 of full scale Combined accuracy 1 0 of full scale Dimension 72 x 182 x 30mm Unit Weight Approx 220 0 485lb with battery Response Time 0 5 seconds Format Baud rate 2400 bit sec Data bit 8 Stop bit 1 Parity none 00 10 35 v V V V inH20 psi mbar bar inHzO psi mbar bar mmH20 Fig 1 Fig 2 Fig 3 D VV V VM VV V VM V V VM inH20 psi mbar bar mmH20 inHzO psi mbar bar mmH20 inH20 psi mbar bar mmH O Fig 4 Fig 5 Fig 6 v VV V VV VV V VM 9 V V inHz2O psi mbar bar inHz2O psi mbar bar 20 psi mbar bar mmH20 Fig 7 Fig 8 Fig 9 VV V VV VV V VV VV V V inH20 psi mbar bar inH20 psi mbar bar inHzO psi mbar bar mmH20 Fig 10 Fig 11 Fig 12 VV V VM inH20 psi mbar bar mmH O Fig 13 F
9. f the screen The meter will not power Use of the back light causes the BAT icon to appear Replacing The Battery Even if the battery was recently replaced check its voltage level if you get no response from your instrument To replace the battery 1 Remove the tubing from the instrument 2 Lay the instrument face down on a clean flat surface 3 Remove battery cover Remove battery from instruments that you do not plan to use for a month or more Do not leave battery in instrument USB PC INTERFACE CAPABILITIES The USB cable and the Download Suite software are required to transfer data to a PC The USB port is located on the right side of the instrument The USB Cable is not included It can be purchased separately as an optional accessory Download Suite Software installation Insert the Download Suite CD into the CD ROM drive To install the software follow the on screen instructions Operation To open the program double click the Download Suite icon To transfer data to a PC follow the on screen instructions MAINTENANCE The meter is calibrated in house before shipping When properly maintained the meter will maintain its accuracy as specified Cleanin Use das cloth and mild soap to clean the case of the MAN15 Manometer do not use harsh detergents or abrasives as these may mar the finish or damage the unit s case with an adverse chemical reaction SPECIFICATIONS Operating Conditions e Compensat
10. ig 14 Fig 16 AMPROBE VE 888 IANOMETER Fig 15 Fig 17 Visit www Amprobe com e Catalog e Application notes e Product specifications e User manuals Please Recycle
11. minutes unless sleep mode is disabled 6 Press 25 to turn ON the back light It automatically switches off after 30 seconds AUTO POWER OFF Sleep mode function This instrument will turn off automatically in approx 20 minutes after power on if there are no key activations For recording or operating over longer periods of time you can disable the sleep mode by pressing D and simultaneously before powering on An n will appear in the middle of the screen at which time you can release the button See Fig 1 The disable sleep mode will be invalid after power off OPERATION MODE See Fig 18 1 Turns instrument on and off 2 Press momentarily and relative clock starts in the lower right screen REC is displayed in the middle left of screen Fig 2 The other functions are locked out except Power and Backlight Press momentarily again and the unit cycles through MAX Fig 3 and MIN Fig 4 and back to current pressure reading the record mode is displayed on the LCD Press and hold for 3 seconds to turn off the record function and return to normal mode 3 Press momentarily to freeze the pressure reading Fig 5 4 Press momentarily DIF appears on top of the LCD and the display indicates the relative zero relative zero causes the value of the display to show as 0 0 only the amount of pressure change will be indicated Press momentarily again and the unit returns to the normal mode of pressure differential see Fig 6
12. ustable pressure range is from 1 5 to 1 7 for 2 11 70 to 12 30 for MAN15 23 40 to 24 60 for If calibration pressure source is not 12 00 psi increase by pressing key or decrease pressing key to set calibration point as required Save the calibration point by pressing key SA and small CA appears on the display in 2 seconds See Fig 10 The meter auto skips to the negative pressure point for next calibration mode Follow the same procedure as step 4 for the negative pressure calibration point The LCD now displays 12 00 and small CA See Fig 11 Perform the necessary calibration figure Refer to your pressure standard if needed Again save the calibration point by pressing key SA and CA appears in 2 seconds and then press End and CA appear in another 2 seconds The meter turns back to the normal mode See Fig 12 If you can t save by pressing QNID key i e no SA appeared please check the followings The calibration pressure source could be referred to the following table if you entered the right positive pressure or negative pressure If you want to skip positive calibration when in the calibration mode press ON key to skip to negative calibration point Calibration point reference Model PsiRange Calibration point Recommended MAN15 0 15 11 70 12 30 manzo 0 30 23 40 24 60 MANUAL ZERO SETTING When you set th
Download Pdf Manuals
Related Search
Related Contents
Oracle Service Cloud - Analytics Performance Best Practices 取扱説明書 - Psn About This Manual T'nB ADVI034347 User Manual - Elation Professional Samsung 400TS-3 Flat Panel Television User Manual EXSOUND Pearl II Digital RF Wireless User's Guide Apple A1303 Cell Phone User Manual Radica Games TETRIS 76069 User's Manual Copyright © All rights reserved.