Operation and Service Manual

image

Contents

1. a ee ee ies nee Syed 8 Kit GComtentsss ni AA ta Aa oe tek a Bool ae 8 Initial Inspector AA A ie 8 Repacking for Shipment css cio dae ide 9 Agilent Technologies Packing Materials oooooooococorcorco cect eee eens 9 Non Agilent Packing Materials secsi se idad eee da ewe es 9 Characters cia iia Aas Sets ei eg baled dag A e 10 Electrical Characteristic sec ss SA oe AG eA SIA SO ERAS GS Gh Soa se A 10 Pin Depth cece isos Wes Wis Mae ee Stine Gite oy Sy ee A eee 10 Operating Conditions i226 ck te oe oS wales e The bo rn 10 Storage Conditjons it eee hh ede bates a aot tered ged detente iia 10 Carrying Case Dimensions s ipes igh e ee nba eR eG Sb epee es 11 2 Operation Operating Precautions mthe nett eect A E OE ani 14 Handling and Storage ocre e e Be ands athe bmw ee Sonne 14 Connection Techniques 032 pants Bike o dal eae en eee 14 Checking Before Connection ii A BORE Ee Sb Ea Be ep bees 14 Visual Inspections 43 5 c te gachage ah da 14 Cl aning Connectors sss ced sc rr hae oe A ae ORS ROE IES RENE CER eS oN 15 Making Connection viril is 16 Connecting terminations directory o oooooooooooorrrr cect ete 16 Connecting airline and terminations 0 1 2 0 0 cee errereen rererere 17 Calibration Data Diskette Format 0 0 00 cee eect teen eee ae 18 3 Service Calibration A A op dice a rae a od cag nat ned els ices Soe antes Re ae 20 Calibration Period a 6 ke ek eS ere hele Ras mele Wate se ote B
2. other defects notify your nearest Agilent Technologies office If the shipping container is damaged or the cushioning material shows signs of stress notify the carrier as well as the Agilent Technologies office Keep the shipping materials for the carrier s inspection 8 Chapter 1 General Information Repacking for Shipment Repacking for Shipment This section provides information on how to repackage the performance test kit for shipment Agilent Technologies Packing Materials Containers and materials identical to those used in factory packaging are available from Agilent Technologies If the instrument is being returned to Agilent Technologies for servicing attach a tag indicating the type of service required return address model number and full serial number Also mark the container FRAGILE to ensure careful handling In any correspondence refer to the unit by model number and full serial number Non Agilent Packing Materials The following general instructions should be used for re packing with commercially available packing materials 1 Ifshipping to a Agilent Technologies office or service center attach a tag indicating the type of service required return address model number and full serial number 2 Use a strong shipping container A double walled carton made of 350 pound test material is adequate 3 se enough shock absorbing material 3 to 4 inch layer around all sides of the case to provide a firm cushio
3. 6 N cm A 1 2 inch open end wrench may be necessary to hold the airline stationary c Gently inserts the airline center conductor into termination s center conductor Mate the outer conductors Then torque the connection to 136 N cm Chapter 2 17 Operation Calibration Data Diskette Format Calibration Data Diskette Format The performance test kit calibration data is stored in the diskette for automated performance test This section provides the diskette format information that is necessary to use the diskette Table 2 1 outlines the calibration data diskette format Table 2 1 Diskette Format Outline Media 3 5 inch Double side Initialize Format DOS ASCII File Bssss DAT BINARY File 16190Bssss 1 ssss is least 4 digit of the kit s serial number 2 For backward compatibility of old calibration file 18 Chapter 2 Service In this chapter maintenance of 16190B is described 19 Service Calibration Calibration The performance test kit is calibrated when shipped from the factory The performance test kit must be calibrated at a facility that satisfies the calibration uncertainties given in Table 3 1 and Table 3 2 Agilent Technologies calibration laboratories satisfy all of these calibration uncertainties For complete information on calibration contact your nearest Agilent Technologies office Table 3 1 Impedance
4. Agilent 16190B Performance Test Kit Operation and Service Manual Second Edition ie Agilent Technologies Agilent Part No 16190 90020 July 2001 Printed in Japan NOTE Notices The information contained in this document is subject to change without notice This document contains proprietary information that is protected by copyright All rights are reserved No part of this document may be photocopied reproduced or translated to another language without the prior written consent of the Agilent Technologies Agilent Technologies Japan Ltd 1 3 2 Murotani Nishi Ku Kobe shi Hyogo 651 2241 Japan Copyright Agilent Technologies Japan Ltd 2000 2001 Manual Printing History The manual s printing date and part number indicate its current edition The printing date changes when a new edition is printed Minor corrections and updates that are incorporated at reprint do not cause the date to change The manual part number changes when extensive technical changes are incorporated February 2000 First Edition July 2001 Second Edition Safety Summary The following general safety precautions must be observed during all phases of operation service and repair of this instrument Failure to comply with these precautions or with specific WARNINGS elsewhere in this manual may impair the protection provided by the equipment In addition it violates safety standards of design manufacture and intended use of the in
5. Lifting out 7mm connector sleeve al Sleeve we yf ae ART anny JCI sl 428 7a0j039 2 While pressing terminations from the top as shown inFigure 2 2 firmly connect it to the 7 mm terminal clockwise viewed from the top Termination connection MN E D_ A E gt oie A Oli 4287a0j043 16 Chapter 2 Operation Making Connection 3 Hold the torque wrench with your thumb and index finger behind the groove in the handle 4 Tighten the connection until the torque wrench handle begins to break It is not necessary to fully break the handle to reach the specified torque Connecting airline and terminations 1 Connect the airline and terminations to the 7mm connector as following procedure a Fully retract the threads on the 7mm connector Then insert the marked side tip of the airline center conductor Figure 2 3 Airline connection to 7mm connector WAR Encino WAR PacKaan b Gently cover the airline center conductor with the airline outer conductor with the logo side down To prevent damage don t let the conductor scrape the edge of the outer conductor Mate the outer conductors then torque the connection to 13
6. Magnitude IZI Admittance Magnitude IYI Calibration Uncertainty Frequency OPEN SHORT 50 Q Airline with Airline with MHz OPEN SHORT 1 0 75 2 5 mQ 0 15 0 18 1 5 10 0 80 3 5 mQ 0 15 0 18 0 40 100 0 80 14 mQ 0 15 0 18 0 24 200 0 80 25 mQ 0 20 0 20 0 29 300 0 80 30 mQ 0 25 0 25 0 29 500 0 80 50 mQ 0 35 0 35 0 34 600 0 85 50 mQ 0 45 0 85 0 48 800 0 85 50 mQ 0 45 0 85 No Calibration 1000 0 85 50 mQ 0 45 0 50 0 50 1300 0 85 100 mQ 0 45 No Calibration 0 80 1600 0 90 100 mQ 0 45 0 80 1 0 1800 0 90 100 mQ 0 45 0 80 0 80 2000 1 50 200 mQ 1 0 1 2 1 0 2200 1 50 200 mQ 1 0 1 2 No Calibration 2400 1 50 200 mQ 1 0 1 2 1 3 2600 1 50 200 mQ 1 0 1 3 1 3 2800 1 50 200mQ 1 0 No Calibration 1 3 3000 1 50 200 mQ 1 0 1 3 No Calibration 1 Calibration uncertainty when shipped from the factory 20 Chapter 3 Service Calibration Table 3 2 Impedance Phase 0 Admittance Phase 0 Calibration Uncertainty Frequency 50 Q Airline with Airline with MHz OPEN SHORT 1 1 5 mrad 1 8 mrad 15 mrad 10 1 5 mrad 1 8 mrad 4 0 mrad 100 1 5 mrad 1 8 mrad 2 4 mrad 200 2 0 mrad 2 0 mrad 2 9 mrad 300 2 5 mrad 2 5 mrad 2 9 mrad 500 3 5 mrad 3 5 m
7. apter 1 General Information Characteristics Carrying Case Dimensions The carrying case dimensions are 350 mm W x 100 mm H x 270 mm D Chapter 1 11 General Information Characteristics 12 Chapter 1 Operation This chapter provides operating information for the Agilent 16190B Performance Test Kit 13 NOTE Operation Operating Precautions Operating Precautions There are several precautions that must be observed to protect the devices in this kit and the instruments being tested Handling and Storage Handle and store the devices in this kit with great care Their continued performance and accuracy depend on maintaining very precise mechanical tolerances When not in use place the airline center conductor in its tube replace the airline outer conductor and the termination s protection caps and store the devices in the carrying case Connection Techniques Extreme care should be taken when making connections or disconnections with the performance test kit devices because the mechanical tolerances and electrical performance of their connectors are better than those of most other 7 mm connectors Checking Before Connection Visual Inspection Visually inspect and if necessary clean all connectors each time a connection is made Metal particles from the connector threads may find their way onto the mating plane surfaces when a connection is disconnected If the inspection shows that the
8. center collet needs to be replaced use only precision 6 slot collets PN 85050 20001 If the inspection shows that the airline center conductor tip needs to be replaced replace the tip PN 1250 2006 Do not use damaged connectors Document Microwave Connector Care PN 08510 90064 explains in detail how to care for microwave connectors 14 Chapter 2 WARNING Operation Cleaning Connectors Cleaning Connectors Clean connectors are essential for ensuring the integrity of RF coaxial connections Use the following procedure to clean the connectors 1 Use compressed air or nitrogen Use compressed air or nitrogen to loosen particles on the connector mating plane surfaces Clean air can t damage a connector or leave particles or residues behind Always use protective eywear when using compressed air or nitrogen You can use any source of clean dry low pressure compressed air or nitrogen that has an effective oil vapor filter liquid condensation trap placed just before the outlet hose Ground the hose nozzle to prevent electrostatic discharge and set the air pressure to less than 414 kPa to control the velocity of the air stream High Velocity streams of compressed air can cause electrostatic effects when directed into a connector 2 Clean the connector threads Use a lint free swab or cleaning cloth moistened with isopropyl alcohol to remove any dirt or stubborn contaminants on a connector that can t be removed w
9. ethe nee ES 21 REP ALR cre sectarian A ets 22 Contents General Information This chapter provides the Agilent Technologies 16190B Performance Test Kit description characteristics and related general information General Information Description Description Agilent Technologies 16190B is a performance test kit designed to verify the impedance measurement accuracy of impedance analyzers or impedance meters that have a 7 mm measurement terminal Kit Contents The 16190B contents are listed in Table 1 1 Table 1 1 Contents of the 16190B Description Agilent Part No Qty Airline 50 Q 7mm Not Assigned 1 Cap Protection 1401 0123 2 50 Q Termination Not Assigned 1 Open Termination Not Assigned 1 Short Termination Not Assigned 1 Cap Termination Protection 16190 25011 3 Wrench 1 2 and 8 15 Open End 8710 1770 1 Carrying Case 16190 60250 1 Floppy Diskette for Calibration Data Not Assigned 1 Calibration Report Not Assigned 1 Operation Note This manual 16190 90010 1 Initial Inspection Inspect the shipping container for damage If the shipping container or cushioning material is damaged it should be kept until the shipping contents have been checked for completeness and the performance test kit has been checked mechanically and electrically The shipping contents should consist of the items shown in Table 1 1 If the shipment is incomplete or if there is mechanical damage or
10. ith compressed air or nitrogen a Apply a small amount of isopropyl alcohol to the lint free cleaning swab b Clean the connector threads c Let the alcohol evaporate then blow the threads dry with a gentle stream of clean low pressure compressed air or nitrogen 3 Clean the mating plane surfaces Apply a small amount of isopropyl alcohol to a new swab and clean the center and outer conductor mating plane surfaces 4 Dry the connector After cleaning blow the connector dry with a gentle stream of clean compressed air or nitrogen Always completely dry a connector before you reassemble or use it 5 Reinspect Inspect the connector again to be sure that no particles or residue Chapter 2 15 CAUTION NOTE Figure 2 1 Figure 2 2 Operation Making Connection Making Connection When making connections with the verification devices turn the nut on the device never turn the device itself Doing so can cause damage to the device or to the attaching connector When making connections a 3 4 inch 136 N cm torque wrench PN 8710 1766 is necessary A 1 2 inch open end wrench may also be necessary to hold one device stationary while torquing the nut on the other device Connecting terminations directory 1 The connector nut outside the 7 mm connector on the test head mounted on the top of the test fixture stand should be turned clockwise viewed from the top then lift connector sleeve completely out
11. n and to prevent movement inside the container 4 Seal the shipping container securely Mark the shipping container FRAGILE to ensure careful handling 6 In any correspondence refer to the unit by model number and full serial number Chapter 1 9 General Information Characteristics Characteristics This section provides the characteristics of the Agilent 16190B performance test kit Electrical Characteristic Table 1 2 list the performance test kit electrical characteristics Table 1 2 Electrical Characteristics OPEN Termination 0 63 pF 5 at 1 MHz 50 Q Termination 50 Q 0 5 at 1 MHz Airline with OPEN Termination 7 3 pF 5 at 1 MHz Airline with SHORT Termination 18 0 nH 5 at 1 MHz Pin Depth Table 1 3 shows the kit pin depth of the kit Table 1 3 Pin Depth OPEN Termination 0 00 mm to 0 05 mm 50 Q Termination 0 00 mm to 0 05 mm Airline 50 Q 7mm 0 010 mm to 0 0025 mm Operating Conditions The kit must be operated under the ambient environmental conditions in Table 1 4 Table 1 4 Operating Conditions Temperature 23 C 5 C Relative Humidity lt 70 at 28 C Altitude 0 m to 4500 m Storage Conditions The kit must be stored under the ambient environmental conditions in Table 1 5 Table 1 5 Storage Conditions Temperature 40 C to 75 C Relative Humidity lt 95 at 40 C Altitude O m to 15000 m 10 Ch
12. part or all of the product Note denotes important information It calls attention to a procedure practice condition or the like which is essential to highlight Certification Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology to the extent allowed by the Institution s calibration facility or to the calibration facilities of other International Standards Organization members Warranty This Agilent Technologies instrument product is warranted against defects in material and workmanship for a period corresponding to the individual warranty periods of its component products Instruments are warranted for a period of one year Fixtures and adapters are warranted for a period of 90 days During the warranty period Agilent Technologies will at its option either repair or replace products that prove to be defective For warranty service or repair this product must be returned to a service facility designated IMPORTANT by Agilent Technologies Buyer shall prepay shipping charges to Agilent Technologies and Agilent Technologies shall pay shipping charges to return the product to Buyer However Buyer shall pay all shipping charges duties and taxes for products returned to Agilent Technologies from ano
13. rad 3 4 mrad 600 4 5 mrad 8 5 mrad 4 8 mrad 800 4 5 mrad 8 5 mrad No Calibration 1000 4 5 mrad 5 0 mrad 5 0 mrad 1300 4 5 mrad No Calibration 8 0 mrad 1600 4 5 mrad 8 0 mrad 10 mrad 1800 4 5 mrad 8 0 mrad 8 0 mrad 2000 10 mrad 12 mrad 10 mrad 2200 10 mrad 12 mrad No Calibration 2400 10 mrad 12 mrad 13 mrad 2600 10 mrad 13 mrad 13 mrad 2800 10 mrad No Calibration 13 mrad 3000 10 mrad 13 mrad No Calibration 1 Calibration uncertainty when shipped from the factory Calibration Period The suggested initial interval for calibration is 12 months or sooner The actual need for calibration depends on the use of the kit After reviewing the results of the initial calibration you may establish a different calibration interval that reflect the usage and wear of the kit Chapter 3 21 Service Repair Repair The performance test kit can be repaired only at Agilent Technologies facilities For complete information on repair contact your nearest Agilent Technologies office 22 Chapter 3
14. strument Agilent Technologies Company assumes no liability for the customer s failure to comply with these requirements 16190B complies with INSTALLATION CATEGORY I and POLLUTION DEGREE 2 in IEC61010 1 16190B is an INDOOR USE product e DO NOT Substitute Parts Or Modify Instrument Because of the danger of introducing additional hazards do not install substitute parts or perform unauthorized modifications to the instrument Return the instrument to a Agilent Technologies Sales and Service Office for service and repair to ensure that safety features are maintained e Dangerous Procedure Warnings WARNING CAUTION NOTE Warnings such as the example below precede potentially dangerous procedures throughout this manual Instructions contained in the warnings must be followed Safety Symbol General definitions of safety symbols used on the instrument or in manuals are listed below Instruction Manual symbol the product is marked with this symbol when it is necessary for the user to refer to the instrument manual This warning sign denotes a hazard It calls attention to a procedure practice condition or the like which if not correctly performed or adhered to could result in injury or death to personnel This Caution sign denotes a hazard It calls attention to a procedure practice condition or the like which if not correctly performed or adhered to could result in damage to or destruction of
15. ther country Agilent Technologies warrants that its software and firmware designated by Agilent Technologies for use with an instrument will execute its programming instruction when property installed on that instrument Agilent Technologies does not warrant that the operation of the instrument or software or firmware will be uninterrupted or error free Limitation of Warranty The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer Buyer supplied software or interfacing unauthorized modification or misuse operation outside the environmental specifications for the product or improper site preparation or maintenance No other warranty is expressed or implied Agilent Technologies specifically disclaims the implied warranties of merchantability and fitness for a particular purpose Exclusive Remedies The remedies provided herein are buyer s sole and exclusive remedies Agilent Technologies shall not be liable for any direct indirect special incidental or consequential damages whether based on contract tort or any other legal theory Assistance Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products For any assistance contact your nearest Agilent Technologies Sales and Service Office Addresses are provided at the back of this manual Contents 1 General Information Description cai A A A